Associated Research, Inc started as a repair company servicing panel meters and
instruments back in 1936. Founder James Inman soon made Associated Research
a manufacturer when he designed the Vibrotest. Vibrotest, the first battery operated
Insulation Resistance Tester ever created, became standard issue on all U.S. Navy ships
during World War II. This was the beginning of what would become many industry
breakthroughs for Associated Research. As the number of electrical products increased
during the post-war economic boom, Mr. Inman realized the need for rugged quality
test instruments. AR was again on the leading edge this time with the introduction of
the Hypot® high potential tester for electrical safety testing. Hipot testing soon became
our forte. True to the beliefs that our founder, James F. Inman, instituted back in 1936,
we continue to offer the highest quality instruments on the market.
AR has continued to innovate and develop advanced products in the electrical safety
compliance testing industry from basic entry-level Hipot and Ground Bond testers to
automated multi-function electrical safety compliance analyzers. Our instruments are
designed to test to the requirements set forth by safety agencies such as UL, TUV, CSA,
IEC, VDE, BSI and European norms. We offer a variety of testers capable of performing
AC Hipot, DC Hipot, Insulation Resistance, Ground Bond, Ground Continuity, Line
Leakage and Functional Run Testing. We also provide Medical Safety Test systems,
modular scanning matrices, and our Autoware testing software to meet the testing
needs and requirements of our customers. We have developed the most complete line
of automation interfaces for our products, including Ethernet, GPIB, RS-232 and USB.
All of our testers also incorporate easy-to-use menus.
Our experience and leadership have helped to shape the industry and our dedication to product
development continues to improve the safety testing process for manufacturers around the globe. We are
committed to providing the highest quality, most advanced instrumentation on the market. Our instruments all
carry the CE Mark and all of our standard testers carry a Nationally Recognized Testing Laboratory (NRTL)
listing mark. This demonstrates our commitment to provide you with the safest and most innovative testers on
the market – a commitment our competitors simply cannot make. Furthermore, Associated Research, Inc. is
an ISO 9001 registered company which demonstrates our commitment to quality!
As the industry leader we also guarantee our customers'
satisfaction. If for any reason you are not completely satisfied
with your purchase within 45 days, you can receive a full refund,
exchange or credit towards another AR product, no questions
asked. Our instruments are backed by the industry’s best warranty.
We offer a standard one year warranty on all of our instruments
with the opportunity to extend that warranty for a full five years if
the instrument is returned to our factory service center each year
for its annual certification and safety inspection. We also offer a three year
warranty for those who are unable to take advantage of the factory certification
that is part of our extended warranty program.
AR is dedicated to making sure that your testing solution provides you with complete
satisfaction for years to come. This is why we offer the most comprehensive
service program in the industry. Our service program ensures that you receive
the value you deserve when using an AR tester.
We offer expedient calibration and repair work performed by expert
technicians who work exclusively with AR instruments. We have a host of
calibration options to match the broad needs of all of our customers. We
offer calibration services from our standard calibration to our accredited
calibration including ISO 17025 calibration, ANSI Z540.1 calibration,
CTL and Denan’s Law calibration. We have industry leading turn-around
time and guarantee 24-hour turn-around for standard calibrations. A
guaranteed 48-hour turn-around time is provided on all of our flat rate
repair work. If we don’t ship your instrument in time, we’ll pay the shipping
cost. In addition, all of our calibration and service work is backed by a 90
day service warranty.
We understand that finding the right electrical
safety testing solution can be a difficult
task. That is why we offer our customers a
comprehensive selection of educational
resources from regional seminars and online
webinars to on-site factory sponsored
training programs. We also offer an
extensive library of technical booklets, articles
and white papers. At AR, we want to ensure that
your investment is maximized and that your tester
performs safely and reliably.
COMMON SAFETY STANDARD REFERENCE CHART
Standard / Harmonized
Standard
Testing Type
IEC/UL 60601-1 3rd
Performance
Edition Medical Electrical
Equipment
Production
UL 60335-1 Household
Electrical Appliances
500 – 4000 VAC or
707 – 5656 VDC
No
Breakdown
1000 – 3000 VAC
Test
Time
60 s
1 or 60 s
GROUND BOND/CONTINUITY
Test Current V Limit
Max. R
Test Time
10-25 A
≤6V
≤ 0.1 Ω
5s
10-25 A
≤6V
≤ 0.1 Ω
5s
2.5 x Max
Over Current ≤ 12 V
Protection
≤ 0.1 Ω
120 s
50 uA
60 s
1000 VAC + 2 x rated V or
(1000 VDC + 2 x rated V) X 120%
50 uA
1 or 60 s
Performance
500 – 2400 VAC x rated V
+ 2400 VAC
No
Breakdown
60 s
≥ 10 A
≤ 12 V
0.1 – 0.2 Ω
≤ 120 s
Production
400 – 2500 VAC
5-30 mA
1s
≥ 10 A
≤ 12 V
0.1 – 0.2 Ω
No time
specified
Performance
500 V – 2400 VAC x rated V +
2400 VAC
No
Breakdown
60 s
40 A
≤ 6.5 V
≤ 0.5 Ω
120 s
Production
400 – 2500 VAC
5-30 mA
1s
40 A
≤ 12 V
0.1 – 0.2 Ω
No time
specified
Performance
500 – 4 x rated V + 2000 VAC
No
Breakdown
60 s
≥ 10 A
≤ 12 V
≤ 0.5 Ω
60 s
≤ 0.1 Ω
120 s
≤ 0.1 Ω
Continuity
IEC 60598-1 Luminaires
Production
1000 VAC - 1000 VAC
x 2 x rated V
Production
1200 VAC
IEC/UL 61010-1 &
Performance
CSA 22.2 No. 61010-1
Laboratory Control Test &
Measurement Equipment Production
Performance
Continuity
Not Specified - Responsibility of Manufacturer
Performance
UL 1598 Luminaires
EN 60204-1 Electrical
Equipment of Machines
Max I.
1000 VAC + 2 x rated V or
2000 VAC + 4 x rated V
Performance
IEC 61730-2 UL 1703
Photovoltaic Modules &
Panels
Production
IEC 60335-1 Household
Electrical Appliances
DIELECTRIC WITHSTAND
Test Voltage
820 – 1350 VAC or
1160 - 1900 VDC
2 x rated V or
1000 VAC
Production
No
Breakdown
60 s
1s
60 s
No
Breakdown 2 s ramp
2 s dwell
or 1 s
No
Breakdown
1s
30 A
≤4V
Continuity
25 or 30 A
≤ 10 V or ≤ 0.1 Ω or <4 V
≤ 12 V
0.133 Ω
60 or
120 s
Continuity
0.2 - 10 A
≤ 24 V
Refer to
No time
Section 18.2.2 specified
Not Specified - Responsibility of Manufacturer
500 VAC or 1000 VAC
60 s
UL 2202 Electric Vehicle Performance
+ 2 x rated V
No
Charging System
Breakdown
Equipment
Production 1000 – 1700 VAC + 3.4 x rated V
60 or 1 s
1200 VAC + rated V
or DC Equivalent
IEC 61851-1 Electric
Vehicle Conductive
Charging System
Performance
UL 45A Portable
Electrical Appliances
Performance
1000 VAC + 2 x rated V or
DC equivalent
Production
1000 - 3000 VAC
Production
No
Breakdown
≤ 60 A
≤ 12 V
Continuity
120 –
240 s
Continuity
60 s
Continuity
Not Specified - Responsibility of Manufacturer
No
Breakdown
EN 60950-1 EN 50116 Performance
Information Technology
Equipment
Production
1000 – 3000 VAC or
1414 – 4242 VDC
No
Breakdown
UL 60950-1 CSA 22.2
Performance
No. 60950-1 Information
Technology Equipment
Production
1000 – 3000 VAC or
1414 – 4242 VDC
No
Breakdown
60 s
Continuity
1s
Continuity
120 s
30 A
≤ 12 V
≤ 0.1 Ω
60 s
1-4s
25 A
≤ 12 V
≤ 0.1 Ω
1- 4 s
60 s
≤ 40 A
≤ 12 V
≤ 0.1 Ω
60 s
1–6s
Continuity
EARTH LEAKAGE
Test Voltage
Max I.
110% x rated
V
5-10 mA
INSULATION RESISTANCE
Test Time
N/A
Max rated V
10 uA – 1 mA
60s
V Limit
Min R
SUGGESTED MODEL #
ARI Instrument
Testing Type
N/A
8106 or MedTEST
Performance
N/A
8104 or 8204
Production
8106 or MedTEST
Performance
500 VDC or
Max rated V
40-400 MΩ
N/A
N/A
3770 or 7650
Production
1.06 x rated V 0.25 – 5.0 uA
N/A
7704 & 620L
Performance
N/A
N/A
7704
Production
1.06 x rated V 0.25 – 5.0 uA
N/A
7704 & 620L
Performance
N/A
N/A
7704
Production
8106
Performance
Hypot III or 7650
Production
8104 or 8204
Performance
N/A
Hypot III or 7650
Production
N/A
8106 or MedTEST
Performance
N/A
3765 or 7650
Production
8104 or 8204
Performance
Hypot III or 7650
Production
N/A
8106 or MedTEST
Performance
N/A
Hypot III or 7650
Production
8106 or MedTEST
Performance
Hypot III or 7650
Production
8106 or MedTEST
Performance
Hypot III or 7650
Production
8106 or MedTEST
Performance
8104 or 8204
Production
8106 or MedTEST
Performance
Hypot III or 7650
Production
Rated V
0.5 – 10 mA
60 s
500 VDC
1-4 MΩ
Standard / Harmonized
Standard
IEC/UL 60601-1 3rd
Edition Medical Electrical
Equipment
IEC 61730-2 UL 1703
Photovoltaic Modules &
Panels
IEC 60335-1 Household
Electrical Appliances
UL 60335-1 Household
Electrical Appliances
IEC 60598-1 Luminaires
Not Specified - Responsibility of Manufacturer
No time
specified
N/A
N/A
< 300 V
0.5 mA
N/A
No time
specified
N/A
500 VDC
≥ 2 MΩ
UL 1598 Luminaires
500 V
≥ 1 MΩ
Not Specified - Responsibility of Manufacturer
Rated V
0.5 - 0.75 mA
or 5 mA
N/A
Touch Current Only
60 s
500 V
≥ 1 MΩ or ≥
7 MΩ
Not Specified - Responsibility of Manufacturer
< 300 V
0.5 – 3.5 mA
60 s
N/A
< 300 V
0.25 – 3.5 mA
0.25 – 3.5 mA
N/A
≥ 50 KΩ
N/A
60 s
N/A
< 300 V
500 V
500 V
≥ 2 MΩ
N/A
60 s
500 V
N/A
≥ 2 MΩ
IEC/UL 61010-1 & CSA 22.2
No. 61010-1 Laboratory
Control Test & Measurement
Equipment
EN 60204-1 Electrical
Equipment of Machines
UL 2202 Electric Vehicle
Charging System Equipment
IEC 61851-1 Electric Vehicle
Conductive Charging
System
UL 45A Portable Electrical
Appliances
EN 60950-1 EN 50116
Information Technology
Equipment
UL 60950-1 CSA 22.2
No. 60950-1 Information
Technology Equipment
AC Hipot DC Hipot Insulation Resistance Ground Bond Functional Run Line Leakage
OMNIA II
®
8204
OMNIA®
8104
8105
8106
AC Hipot DC Hipot Insulation Resistance Ground Bond Functional Run Line Leakage
HypotMAX
®
7700
7704
7705
7710
7715
7720
AC Hipot DC Hipot Insulation Resistance Ground Bond Functional Run Line Leakage
HypotULTRA® III
7620
7650
AC Hipot DC Hipot Insulation Resistance Ground Bond Functional Run Line Leakage
Hypot® III
3705
3765
3770
3780
AC Hipot DC Hipot Insulation Resistance Ground Bond Functional Run Line Leakage
HYAMP® III
3130
3140
3160
LINECHEK® II
AC Hipot DC Hipot Insulation Resistance Ground Bond Functional Run Line Leakage
620L
MedTEST is our comprehensive Medical Electrical Safety Tester. MedTEST can be custom designed to meet all of your medical
safety testing needs. It complies with test requirements called out in common medical electrical safety specifications such as UL2601,
UL60601, IEC601-1, IEC60601-1 and EN60601-1.
USB/RS-232
GPIB
Ethernet/
Data Storage/
RS-485
Internal Interconnection
to External
Autoware
Scanner
Scanner
DualCHEK
Color TFT
Display
USB/RS-232
GPIB
Ethernet/
Data Storage/
RS-485
Internal Interconnection
to External
Autoware
Scanner
Scanner
DualCHEK
Color TFT
Display
USB/RS-232
GPIB
Ethernet/
Data Storage/
RS-485
Internal Interconnection
to External
Autoware
Scanner
Scanner
DualCHEK
Color TFT
Display
USB/RS-232
GPIB
Ethernet/
Data Storage/
RS-485
Internal Interconnection
to External
Autoware
Scanner
Scanner
DualCHEK
Color TFT
Display
GPIB
Ethernet/
Data Storage/
RS-485
Internal Interconnection
to External
Autoware
Scanner
Scanner
DualCHEK
Color TFT
Display
GPIB
Ethernet/
Data Storage/
RS-485
Internal Interconnection
to External
Autoware
Scanner
Scanner
DualCHEK
Color TFT
Display
OMNIA® II
8204
OMNIA®
8104
8105
8106
HypotMAX®
7700
7704
7705
7710
7715
7720
HypotULTRA® III
7620
7650
Hypot III
®
3705
3765
3770
3780
RS-232 Only
RS-232 Only
RS-232 Only
RS-232 Only
USB/RS-232
HYAMP®
3130
3140
3160
RS-232 Opt.
USB/RS-232
LINECHEK® II
620L
Visit our Product Selection Wizard, designed to help find the right testers for your application. Through a series of questions
we will be able to suggest the perfect tester for you. Go to www.asresearch.com and follow the link to the Product Selection Wizard.
An Electrical Safety Compliance Analyzer That Is As Unique As Your Application!
OMNIA II, our next generation of Electrical Safety Compliance Analyzers, is designed around the way you test. We
understand that every testing application is unique and finding the right tester can be difficult. OMNIA II model 8204
is a multi-function Electrical Safety Compliance Analyzer that provides you with customizable features and unmatched
functionality.
Model 8204 - 4 -in-1 Electrical Safety Compliance Analyzer that performs AC Hipot, DC Hipot,
40 Amp Ground Bond and Insulation Resistance testing in a single instrument.
Features and Benefits
Color TFT Display – the 8204 provides an 800 x 480
Color TFT display that makes setting up test files, viewing
results, and performing tests easier than ever. Colorcoded test steps clearly indicate pass/fail conditions.
Choose from 3 different color schemes to match your
preferences.
“My Menu” Interface – the 8204 is equipped with
our “My Menu” interface which allows operators to
personalize menu settings. Create shortcuts to your
favorite screens and preferences for unmatched ease
and speed in setup.
DualCHEK – the DualCHEK feature allows the user to
perform a simultaneous Hipot and Ground Bond test.
This can safely increase productivity and throughput
on the production line.
Multiple Automation Interfaces – the 8204 comes
standard with a USB/RS-232 interface. The 8204 is
also available with optional Ethernet, GPIB, Data
Storage, and printer output interfaces.
Multiple Language Settings – the 8204 is available in
two language settings. Users can select to view the
menu in English or Traditional Chinese.
Patented SmartGFI® function - The SmartGFI is a high
speed shutdown circuit that provides maximum operator
protection. If the circuit detects excessive leakage to
ground it shuts down the high voltage output in less than
1 millisecond. SmartGFI is automatically activated if the
DUT is not grounded with no need for operator input.
Expanded Test Memories – The 8204 allows users to
link a total of 10,000 test steps. This allows users to create
and save even the most complex test setups.
Patented Prompt and Hold function – The prompt and hold
function allows users to insert prompts or instructions as
part of the test cycle. This helps to eliminate operator error
when changing test leads is required as part of the test
cycle.
Patented CAL-ALERT® function - The patented CAL-ALERT
feature automatically alerts the user when the instrument
is due for calibration. This eliminates the need for manual
tracking of calibration dates.
Patented VERI-CHEK® function - The VERI-CHEK feature is
a menu-driven process by which the instrument’s failure
detectors are proven to be functioning properly, “verifying” the functionality of the electrical safety tester and
connected accessories.
CHARGE LO® - The Charge-LO Circuit sets a minimum
charging current which is based on the DC test voltage,
the rate of rise, and the capacitance of the DUT. This
circuit confirms that the DUT is connected when performing a test.
RAMP HI® - The Ramp-HI feature prevents the test from
failing during ramp up due to excessive charging current.
This allows you to charge the DUT as rapidly as possible
without causing false failures.
APT
AC Power Source
Compatible
www.aspowertechnologies.com
Safety agency listed.
Input Specifications
Voltage
115/230 VAC ± 15% Auto Range
Continuity Test Mode
Output Current
DC 0.1 A ± 0.00001 A
Frequency
50/60 Hz ± 5%
Resistance Display
Range: 1.00 - 10000 Ω
Fuse
115 VAC, 230 VAC - 5A/250 VAC Slow Blow
HI and LO-Limit
0.0 - 10000 Ω
Dwell Timer
Range: 0.0, 0.3 - 999.9 sec (0=continuous)
Milliohm Offset
Range: 0.00 – 10.00 Ω
Dielectric Withstand Test Mode
Output Rating
5 kV @ 50 mAAC
5 kV @ 20 mADC
Voltage Setting
Range: 0 - 5000 VAC
0 - 5000 VDC
Resolution: 1 V
Accuracy: ± (2% of Setting + 5 V)
HI and LO-Limit
AC Total
AC Real
DC
Range: 0.000-9.999 mA
Resolution: 0.001 mA
Accuracy: ± (2% of setting + 2 counts)
Range: 10.00 - 50.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of Setting + 2 Counts)
Range: 0.000-9.999 mA
Resolution: 0.001 mA
Accuracy: ± (3% of Setting + 50 µA)
Range: 10.00 - 50.00 mA
Resolution: 0.01 mA
Accuracy: ± (3% of Setting + 50 µA)
Range: 0.00 - 999.9 µA
Resolution: 0.1 µA
Accuracy: ± (2% of Setting + 2 Counts)
Range: 1000 - 20000 µA
Resolution: 1 µA
Accuracy: ± (2% of Setting + 2 Counts)
Insulation Resistance Test Mode
Output Voltage
Range: 30 - 1000 VDC
Charging Current
Maximum >20 mA peak
HI and LO-Limit
Range: 0.05-99.99 MΩ
Resolution: 0.01 MΩ
Range: 100.0 - 999.9 MΩ
Resolution: 0.1 MΩ
Range: 1000 - 50000 MΩ
Resolution: 1 MΩ
Charge-LO
0.000 - 3.500 µA or Auto Set
Ramp Timer
Ramp-Up: 0.1 - 999.9 secs
Ramp-Down: 0.0, 1.0 -999.9 secs
Dwell Timer
0, 0.5 - 999.9 (0=Continuous)
Delay Timer
0.5 - 999.9 secs
Ground Fault Interrupt GFI Trip Current: 5.0 mA max
HV Shut down Speed: < 1 ms
Ground Bond Test Mode
Output Voltage
Range: 3.00 - 8.00 VAC
Ramp HI
>20 mA peak maximum, ON/OFF User Selection
Output Frequency
50/60 Hz ± 0.1% , User Selection
Charge LO
Range: 0.0 - 350.0 µA DC or Auto Set
Output Current
Range: 1.00 - 40.00 A
Resolution: 0.01 A
Accuracy: ± (2 % of setting + 2 counts)
Output Regulation
(1% of output + 0.02A), Within maximum load limits,
±
and over input voltage range.
Maximum Loading
1.00 - 10.00 A,
10.01 - 30.00 A,
30.01 - 40.00 A,
HI and LO-Limit
Range: 0 - 150 for 30.01 - 40.00 A
Range: 0 - 200 for 10.01 - 30.00 A
Range: 0 - 600 for 6.00 - 10.00 A
Range: 0 - 600 for 5.99 - 1.00 A
Resolution: 1 mΩ
Accuracy: 6.00 - 40.00 A, ± (2% of setting + 2 Counts)
Accuracy: 1.00 - 5.99 A, ± (3% of setting + 3 Counts)
Milliohm Offset
Range: 0 - 200 mΩ
DC Output Ripple
< 4 % (5 kV/20 mA at Resistive Load)
Discharge Time
< 50 msec for no load,
< 100 msec for capcitor load
(all capacitance values in MAX load spec below)
Max Capacitive Load
DC Mode
1 µF < 1 kV
0.75 µF < 2 kV
0.5 µF < 3 kV
AC Output Waveform
Sine Wave. Crest Factor = 1.3 - 1.5
Output Frequency
50/60 Hz ± 0.1% , User Selection, 400/800 Hz Option
0.08 µF < 4 kV
0.04 µF < 5 kV
Output Regulation± (1% of output + 5 V), From no load to full load and Low
Line to High Line (combined regulation)
Dwell Timer
AC DC 0, 0.4 - 999.9 (0=continuous)
0, 0.3 - 999.9 (0=continuous)
Ramp Timer
Ramp-Up AC: 0.1 - 999.9
Ramp-Down AC: 0.0-999.9
Ramp-Up DC: 0.4 - 999.9
Ramp-Down DC: 0.0, 1.0-999.9
Ground Continuity
Current: DC 0.1 A ± 0.01A, fixed
Max. Ground Resistance: M
Ground Fault Interrupt GFI Trip Current: 5.0 mA max
HV Shut Down Speed: < 1 ms
Arc Detection
1 - 9 Ranges (9 is the most sensitive)
0 - 600 mΩ
0 - 200 mΩ
0 - 150 mΩ
General Specifications
Mechanical
Bench or rack mount with tilt up front feet
Dimensions3U (WxHxD) 16.93 x 5.24 x 19.69 in.
(430 x 133 x 500 mm)
InterfaceUSB/RS-232 Standard, GPIB, Ethernet, Data Storage
(RS-485), Printer Port Optional
Memory
Up-to 10,000 Steps can be linked
MultilingualChoose from English (Default) and Chinese
DisplayColor TFT LCD. 800 x 480 Resolution. Choose from 4
different color settings
For more information on testing to a specfic standard, refer back to the
Common Safety Standard Reference Chart.
Accredited calibration service available. Includes ISO 17025,
ANSI Z540.1-1994, CTL & Denan’s Law requirements.
Fully-Automated, Multi-Function Electrical Safety Compliance Analyzer
Multi-function safety compliance analyzers with an enhanced graphic LCD provide complete test setup and results with
an easy-to-use interface. OMNIA® provides 4-in-1, 5-in-1 and 6-in-1 testing solutions. Testers include AC Hipot, DC Hipot,
Ground Bond/Continuity, Insulation Resistance, Functional Run and Line Leakage tests. An optional internal scanner
is available for the 8104, 4-in-1 tester. An additional external modular scanner is available for use with all testers. All
testers come standard with USB and RS-232 interfaces. Ethernet, GPIB, and RS-485 interfaces are also available.
Model 8104 - OMNIA 4, 5 kV @ 40 mA AC, 5 kV @ 20 mA DC, IR Test, 40 Amp
Ground Bond & Optional HV & HC Scanner
Model 8105 - OMNIA 5, 5 kV @ 40 mA AC, 5 kV @ 20 mA DC, IR Test, 40 Amp
Ground Bond & Functional Run Test
Model 8106 - OMNIA 6, 5 kV @ 40 mA AC, 5 kV @ 20 mA DC, IR Test, 40 Amp
Ground Bond, Functional Run Test & Line Leakage Test
Features and Benefits
Patented SmartGFI® safety circuit protects
the operator from shock hazards
Real Current measurement allows
operators to monitor total and real
current on a single screen
Patented Prompt and Hold function
provides a unique method for performing
multiple steps during a test cycle
Patented CAL-ALERT® and
VERI-CHEK® features help
to ensure that your
instrument is calibrated
and stays within specs
Data storage card available for
storing and transferring test data
without a connection to a PC
Line Leakage tester with 7 different
measuring devices and RMS or PEAK
leakage measurements
50 Memories with 30 steps
per memory that can be
stored and recalled in any
alphanumeric combination
Can be easily connected to AR’s
SC6540, 620L or an APT Brand AC
Power Source to provide a customizable
test system
RAMP HI® and CHARGE LO®
testing for more effective
DC Hipot testing
USB/RS-232, GPIB, Ethernet, or
RS-485 automation interfaces
available
Autoware Testing Software available for
complete Automation Control
Perform Hipot/Line Leakage
without changing test leads
Patented Graphic LCD and
intuitive menu system to
simplify the entire testing
process from set-up to results
Cold Resistance Feature for Line to
Neutral Continuity Testing
APT
AC Power Source
Compatible
www.aspowertechnologies.com
Safety agency listed.
Input Specifications
Dielectric Withstand Test Mode (continued)
Voltage
115/230 V selectable, ± 10% variation
Frequency
50/60 Hz ± 5%
Max Capacitive Load
DC Mode
AC Output Waveform
1 µF < 1 kV
0.08 µF < 4 kV
0.75 µF < 2 kV
0.04 µF < 5 kV
0.5 µF < 3 kV
Sine Wave, Crest Factor = 1.3 - 1.5
Fuse
10 A Slow Blow 250 VAC
Output Frequency
60 or 50 Hz, User Selecatable
Output Regulation
± (1% of output + 5 V) from no load to full load and over
input voltage range
Dwell Timer
AC 0.4 - 999.9 sec (0 = Continuous)
DC 0.3 - 999.9 sec (0 = Continuous)
Ramp Timer
Ramp-Up: AC 0.1 - 999.9 sec, DC 0.4 - 999.9 sec
Ramp-Down: AC 0.0 - 999.9 sec, DC 0.0,1.0 - 999.9 sec
Ground Continuity
Current: DC 0.1 A ± 0.01 A, fixed
Max. Ground Resistance: 1 Ω ± 0.1 Ω, fixed
Dielectric Withstand Test Mode
Output Rating
5 kV @ 40 mAAC
5 kV @ 20 mADC
Voltage Setting
Range: 0-5000 VAC
0-5000 VDC
Resolution: 1 V
Accuracy: ± (2% of setting + 5 V)
Voltage Display
Range: 0.00 - 5.00 kV Full Scale
Resolution: 0.01 kV
Accuracy: ± (2% of reading + 10 V)
HI and LO-Limit
AC Total
AC Real
DC
Range: 0.000-9.999 mA
Resolution: 0.001 mA
Range: 10.00 – 40.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of setting + 2 counts)
Range: 0.000-9.999 mA
Resolution: 0.001 mA
Range:
10.00 – 40.00 mA
Resolution: 0.01 mA
Accuracy: ± (3% of setting + 50 µA)
Range: 0.0-999.9 µA
Resolution: 0.1 µA
Range: 1000 – 20000 µA
Accuracy: ± (2% of setting + 2 counts)
Current Display
AC Total
AC Real
DC
Range: 0.000 mA – 3.500 mA
Resolution: 0.001 mA
Range: 3.00 mA – 40.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of reading + 2 counts)
Range: 0.000 mA – 9.999 mA
Resolution: 0.001 mA
Range: 10.00 mA – 40.00 mA
Resolution: 0.01 mA
Accuracy: ± (3% of reading + 50 µA)
Range: 0.0 µA – 350.0 µA
Resolution: 0.1 µA
Range: 0.300 mA – 3.500 mA
Resolution: 0.001 mA
Range: 3.00 mA – 20.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of reading + 2 counts)
Ramp HI
>20 mA peak maximum, ON/OFF selectable
Charge LO
Range: 0.000 - 350 µA or Auto Set
DC Output Ripple
≤ 4% Ripple rms at 5 kV DC @ 20 mA, Resistive Load
Discharge Time
≤ 200 ms
Ground Fault Interrupt GFI Trip Current: 450 µA max (AC or DC)
HV Shut Down Speed: < 1 ms
Continuity Test Mode
Output Current
DC 0.1 A ± 0.00001 A
Resistance Display
Range: 0.00 – 10000.00 Ω
HI and LO-Limit
0.00 – 10000 Ω
Dwell Timer
Range: 0.0, 0.3 - 999.9 sec (0 = Continuous)
Milliohm Offset
Range: 0.00 – 10.00 Ω
Ground Bond Test Mode
Output Voltage
Range: 3.00 - 8.00 VAC
Output Frequency
50/60 Hz, user selectable
Output Current
Range: 1.00 - 40.00 A, Resolution: 0.01 A
Output Regulation
Accuracy: ± (1% of output + 0.02 A)
Within maximum load limits, and over input voltage range
Maximum Loading
1.00 - 10.00 A,
10.01 - 30.00 A,
30.01 - 40.00 A,
Current display
Range: 0.00 – 40.00 A
Resolution: 0.01 A
Accuracy: ± (3% of setting + 0.03 A)
0 - 600 mΩ
0 - 200 mΩ
0 - 150 mΩ
Ohmmeter Display
Range: 0 – 150 mΩ for 30.01 – 40.00 Amps
0 – 200 mΩ for 10.01 – 30.00 Amps
0 – 600 mΩ for 6.00 – 10.00 Amps
Resolution: 1 mΩ
Accuracy: ± (2% of reading + 2 mΩ)
Range: 0 – 600 mΩ for 1.00 – 5.99 Amps
Resolution: 1 mΩ
Accuracy: ± (3% of reading + 3 mΩ)
Ground Bond Test Mode (continued)
OMNIA 8105 and 8106 Functional Run Test Mode
HI and LO Limit
Range: 0 – 150 mΩ for 30.01 – 40.00 Amps
0 – 200 mΩ for 10.01 – 30.00 Amps
0 – 600 mΩ for 1.00 – 10.00 Amps
Resolution: 1 mΩ
Accuracy: Same as Ohmmeter Display
DUT Power
Voltage: 0 – 277 VAC Single Phase Unbalanced
(One Hot or Line conductor and One Neutral)
Current: 15 AAC max continuous
Short Circuit Protection: 23 AAC, Response Time < 3s
Delay Time Setting
Range: 0.2 – 999.9 seconds
Dwell Time Setting
Range: 0.1 – 999.9 seconds
(0 = Continuous)
Milliohm Offset
Range: 0 - 200 mΩ
Insulation Resistance Test Mode
Voltage Setting
Range: 50 - 1000 VDC
Charging Current
Maximum >20 mA peak
Resistance Display
Range: 0.05 MΩ – 50000 MΩ (4 Digit, Auto Ranging)
Resolution: 50 – 499 VDC
500 – 1000 VDC
MΩ
MΩ
MΩ
0.001
0.050 – 1.999
0.050 – 9.999
0.01
2.00 – 19.99
10.00 – 99.99
0.1
20.0 – 199.9
100.0 – 999.9
1 200 – 500001000 – 50000
Accuracy
50 – 499 V
0.05 M – 999.9 M
HI and LO Limit
Range: 0.05 M – 99.99 MΩ
Resolution: 0.01 M
Range: 100.0 M – 999.9 M
Resolution: 0.1 M
Range: 1000 M – 50000 M
Resolution: 1 M
(HI – Limit: 0 = OFF)
Accuracy: Same as Resistance Display Accuracy
Charge LO
Range: 0.000 - 3.500 µA or Auto Set
Ramp Timer
Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: 0.0, 1.0 - 999.9 sec
Delay Timer
1.0 - 999.9 sec (0 = Continuous)
± (7% of reading +2 counts)
500 – 1000 V
0.05 M – 999.9 M ± (2% of reading +2 counts)
1000 M – 9999 M ± (5% of reading +2 counts)
10000 M – 50000 M ± (15% of reading +2 counts)
Ground Fault Interrupt GFI Trip Current: 450 µA max (AC or DC)
HV Shut Down Speed: < 1 ms
General Specifications
Mechanical
Bench or rack mount with tilt up front feet
Dimensions
3U (WxHxD) 16.93 x 5.24 x 19.69 in.
(430 x 133 x 500 mm)
Weight 8104 - 26 kgs / 57.32 lbs.
8105/8106 - 29 kgs. / 63.93 lbs.
Interface
USB/RS-232 Standard, GPIB, Ethernet,
Data Storage (RS-485)
Memory
50 memories, 30 step/memory
Trip Point
Voltage: Volt-Hi
Settings Volt-LO Range: 0.0 – 277.0 VAC
Current:Amp-HI
Amp-LO Range: 0.0 – 15.00 AAC
Watts: Power-HI
Power-LO Range: 0 – 4200 W
Power Factor: PF-HI
PF-LO Range: 0.000 – 1.000
Leakage Current: Leak-HI
Leak-LO Range: 0.00 – 10.00 mA (0 = OFF)
OMNIA 8106 Line Leakage Test Mode
DUT Power
Voltage: 0 – 277 VAC
Current: 15 AAC max continuous
Short Circuit Protection: 23 AAC, Response Time <3s
Leakage Current Current Display rms or PEAK
Range 1: 0.0 µA – 999.9 µA
Resolution: 0.1 µA/step
Range 2: 1000 µA – 6000 µA
Resolution: 1 µA/step
Accuracy rms: DC to 100 kHz
± (1.5% of reading +3 counts)
>100k to 1 MHz
± 5% of reading, (10.0 µA – 6000 µA)
Measuring Device
A
UL544 Non Patient
B
UL544 Patient
C
UL2601-1, UL60601-1, IEC601-1, IEC60601-1 EN60601-1
DUL1563
E
UL1950, UL3101, UL61010, IEC950, IEC1010, IEC 60950, IEC61010, IEC60335-1, IEC60990 Fig4-U2
H
IEC60990 Fig5-U3
I
IEC60990 Fig3-U1
Accredited calibration service available. Includes ISO 17025,
ANSI Z540.1-1994, CTL & Denan’s Law requirements.
For more information on testing to a specfic standard, refer back to the
Common Safety Standard Reference Chart.
High Voltage or High Current Electrical Safety Compliance Analyzers
HypotMAX® is designed for automated applications requiring testers with either higher voltage or higher output current
capability. The HypotMAX® family includes two high current testers: the 7700 3-in-1 version with 500 VA AC output and
the 7704 4-in-1 version with 500 VA AC output. The high voltage testers are the 7705 10 kV AC Hipot, 7710 12 kV DC
Hipot, 7715 20 kV AC Hipot and the 7720 20 kV DC Hipot. All testers come standard with USB and RS-232 interfaces.
GPIB (IEEE-488) and other automation interfaces optional.
Model
Model
Model
Model
Model
Model
7700
7704
7705
7710
7715
7720
- 5 kV @ 100 mA AC (500 VA), 6 kV @ 10 mA DC & IR Test
- 5 kV @ 100 mA AC (500 VA), 6 kV @ 10 mA DC, IR, 30Amp GB
- 10 kV @ 20 mA AC
- 12 kV @ 10 mA DC
- 20 kV @ 10 mA AC
- 20 kV @ 5 mA DC
Features and Benefits
Patented SmartGFI® safety
circuit protects the operator
from shock hazards
50 memories that can be
stored and recalled. Multifunction testers include 8
steps per memory
Up to 20 kV AC or DC Hipot
testing for manufacturers
with higher voltage testing
requirements
USB/RS-232 or GPIB automation
interfaces available
RAMP HI® and CHARGE LO®
systems for more effective
DC Hipot testing
4 wire measurement and
milliohm offset for accurate
Ground Bond test results
(Model 7704)
500 VA testers available
for Higher Current Hipot
test applications
Autoware Testing Software
available for complete
Automation Control
Meets 200 mA short circuit
requirements
Input Specifications
Voltage 7700/7704 100/115/200/230 VAC ± 10%,
single phase, user selection
7705/7710
7715/7720 115/230 VAC ± 10%, single phase, user selection
Frequency Fuse 50/60 Hz ± 5%
7700/7704 15 Amp 250 V fast blow internal
7705/7710 6.3 Amp, 250 V Slow Blow
7715/7720
Dielectric Withstand Test Mode
Output Rating 7700/7704
7705
7710
7715
7720
5 kV @ 100 mAAC, 6 kV @ 10 mADC
10 kV @ 20 mAAC
12 kV @ 10 mADC
20 kV @ 10 mAAC
20 kV @ 5 mADC
Output
7700/7704
Adjustment
7705
7710
7715
7720
Range: 0 - 5 kV AC, 0 - 6 kVDC
Resolution: 1 V/step
Accuracy: ± (2% of setting + 5 V)
Range: 0 - 10 kVAC
Resolution: 10 V/step
Accuracy: ± (2% of setting + 10 V)
Range: 0 - 12 kVDC
Resolution: 10 V/step
Accuracy: ± (2% of setting + 10 V)
Range: 0 - 20 kVAC
Resolution: 10 V/step
Accuracy: ± (2% of setting + 10 V)
Range: 0 - 20 kVDC
Resolution: 10 V/step
Accuracy: ± (2% of setting + 10 V)
HI-Limit 7700/7704
and LO-Limit
Hi-Limit
LO-Limit
7705
7710
7715
7720
AC Range: 0.00 - 99.00 mA
Resolution: 0.01 mA/step
DC Range: 0 - 9999 µA
Resolution: 1 µA/step
Accuracy: ± (2% of setting + 2 counts)
AC Range: 0.000 - 9.999 mA
Resolution: 0.001 mA/step
DC Range: 0 - 999.9 µA
Resolution: 1 µA/step
Accuracy: ± (2% of setting + 2 counts)
Range 1: 0.0 - 9.999 mA
Resolution: 0.001 mA/step
Range 2: 10.00 - 20.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of setting + 2 counts)
Range 1: 0.000 - 999.9 µA
Resolution: 0.1 µA/step
Range 2: 1000 - 9999 µA
Resolution: 1 µA
Accuracy: ± (2% of setting + 2 counts)
Range: 0.00 - 9.999 mA
Resolution: 0.001 mA/step
Accuracy: ± (2% of setting + 2 counts)
Range 1: 0.0 – 999.9 µA
Resolution: 0.1 µA/step
Range 2: 1000 – 5000 µA
Resolution: 1 µA/step
Accuracy: ± (2% of setting + 2 counts)
Dielectric Withstand Test Mode (continued)
DC Ramp HI 7700/7704 12 mA peak maximum, (ON/OFF selectable all testers)
7710 13 mA peak maximum, 10 mADC, ON/OFF selectable
7720 6.75 mA peak maximum, 5 mADC, ON/OFF selectable
DC Charge LO 7700/7704 Range: 0.0 - 350 µADC or auto set
7710/7720
Arc Detection 7700, 7704, 7710, 7720: 1-9
7705: 1 - 9 at output voltage < 7.00 kV
1 - 8 at output voltage ≥ 7.00 kV
7715: 1 - 9 at output voltage < 15.00 kV
1 - 7 at output voltage ≥ 15.00 kV
Voltage 7700/7704 Range: 0.00 - 6.00 kV full scale
Display Resolution: 10 V/step
Accuracy: ± (2% of reading + 2 counts)
7705 Range: 0.00 - 10.00 kV Full scale
Resolution: 10 V
Accuracy: ± (2% of reading + 20 V)
7710 Range: 0.00 - 12.00 kV Full scale
Resolution: 10 V
Accuracy: ± (2% of reading + 2 counts)
7715 Range: 0.00 - 20.00 kV Full scale
Resolution: 10 V
Accuracy: ± (2% of reading + 20 V)
7720 Range: 0.00 - 20.00 kV Full scale
Resolution: 10 V
Accuracy: ± (2% of reading + 20 V)
Current 7700/7704 Auto Range
Display
AC Range 1: 0.000 mA - 3.500 mA
Resolution: 0.001 mA/step
Range 2: 3.00 - 99.00 mA
Resolution: 0.01 mA/step
DC Range 0.0 µA - 350.0 µA
Resolution: 0.1 µA/step
Range 2: 300 µA - 3500 µA
Resolution: 1 µA/step
Range 3: 3000 µA - 9990 µA
Resolution: 10 µA/step
7705 Auto Range
Range 1: 0.000 mA - 3.500 mA
Resolution: 0.001 mA
Range 2: 3.00 - 20.00 mA
Resolution: 0.01 mA
7710 Auto Range
Range 1: 0.0 - 350.0 µA
Resolution: 0.1 µA
Range 2: 300 - 3500 µA
Resolution: 1 µA
Range 3: 3000 mA - 9999 µA
Resolution: 10 µA
7715 Auto Range
Range 1: 0.000 mA - 3.500 mA
Resolution: 0.001 mA
Range 2: 3.00 - 10.00 mA
Resolution: 0.01 mA
7720 Auto Range
Range 1: 0.0 - 350.0 µA
Resolution: 0.1 µA
Range 2: 300 - 5000 µA
Resolution: 1 µA
Dielectric Withstand Test Mode (continued)
DC Output 7700/7704 4% Ripple rms at 6 kVDC @ 3.5 mA, Resistive load
Ripple 7710 < 5% Ripple at 12 kV @ 9999 µA, Resistive Load
7720 < 5% Ripple at 20 kV @ 4999 µA, Resistive Load
AC Output Waveform Sine Wave, Crest Factor = 1.3 - 1.5
AC Output 7705/7710 ± (1% of setting + 10 V) from no load to full load
Regulation 7715/7720
Output Frequency Range: 60 or 50 Hz, user selection
Accuracy: ± 1%
Output
7700/7704 ± (1% of output + 5 V) from no load to full load
Regulation 7705/7710 ± (1% of output + 10 V) from no load to full load
7715/7720 ± (1% of output + 10 V) from no load to full load
Discharge 7700/7704 < 200 m secs
Time 7710 No load < 400 ms
7720 No load < 500 ms
Dwell Timer 7700/7704 Range: 0, 0.3 - 999.9 sec (0 = Continuous)
Resolution: 0.1 sec increments
Accuracy: ± (0.1% + 0.05 sec)
7705/7710/7715/7720 AC Range: 0, 0.3 - 999.9 sec or min (0 = Continuous)
DC Range: 0, 0.4 - 999.9 sec or min (0 = Continuous)
Resolution: 0.1 second or minute increments
Accuracy: ± (0.1% + 1 count)
Ramp Timer 7700/7704 AC Range: 0.1 - 999.9 sec
DC Range: 0.4 - 999.9 sec
Resolution: 0.1 sec increments
Accuracy: ± (0.1% + 0.05 sec)
7705/7715 Range: 0.3 - 999.9 sec
7710/7720 Range: 0.4 - 999.9 sec
7705/7710/7715/7720 Resolution: 0.1 sec increments
Accuracy: ± (0.1% + 1 count)
Ground Continuity 7700 Current: DC 0.1 A ± 0.01 A, fixed
Max. Ground Resistance: 1 Ω ± 0.1 Ω, fixed
Ground Fault 7700/7704 GFI Trip Current: 450 µA max (AC or DC)
Interrupt HV Shut Down Speed: < 1 ms
7705/7710/7715/7720 GFI Trip Current: 1 mA max
HV Shut Down response: < 1 ms
Insulation Resistance Test Mode (Models 7700 & 7704 only)
Output Voltage Range: 100 - 1000 VDC
Resolution: 1 V/step
Accuracy: ± (2% of reading + 2 V)
Short Circuit Current Maximum: 12 mA peak
Voltage Display Range: 0 - 1000 V
Resolution: 1 V/step
Accuracy: ± (2% of reading + 2 counts)
Resistance Display Range: 1 - 9999 MΩ (4 digit, auto ranging)
Resolution: 500 VDC 1000 VDC
MΩ
MΩ
MΩ
0.001
1.000 - 5.388 1.000 - 9.999
0.01
1.40 - 53.88
2.80 - 99.99
0.1
14.0 - 538.8
28.0 - 999.9
1
140 - 9999
280 - 9999
Accuracy: ± (2% of reading + 2 counts) at test voltage
500 - 1000 V and 1 - 1000 MΩ
± (8% of reading + 2 counts) at test voltage
500 - 1000 V and 1000 - 9999 MΩ
± (8% of reading + 2 counts) at test voltage
100 - 500 V and 0 - 1000 MΩ
Insulation Resistance Test Mode (Models 7700 & 7704 cont.)
Charge-LO
HI-Limit
LO-Limit
Delay Timer
Ground Fault Interrupt
Range: 0.000 - 3.500 µA or auto set
Range: 0 - 9999 MΩ (0 = OFF)
Range: 1 - 9999 MΩ
Range: 0, 0.5 - 999.9 sec (0 = Continuous)
Resolution: 0.1 sec/step
Accuracy: ± (0.1% + 0.05 sec)
GFI Trip Current: 450 µA max (AC or DC)
HV Shut Down Speed: < 1 ms
Ground Bond Test Mode (Model 7704 only)
Output Voltage Range: 3.00 - 8.00 VAC
(Open Circuit Limit) Resolution: 0.01 V/step
Accuracy ± (2% of setting + 0.03 V) O.C. condition
Output Frequency Range: 50 or 60 Hz, user selection
Accuracy: ± 1%
Output Current Range: 3.00 - 30.00 AAC
Resolution: 0.01 A/step
Accuracy: ± (2% of setting + 0.02 A)
Current Display Range: 0.00 - 30.00 A
Resolution: 0.01 A/step
Accuracy: ± (2% of stetting + 0.03 A)
Resistance Display Range: 0 - 600 mΩ
Resolution: 1 mΩ/step
Accuracy: ± (2% of reading + 2 mΩ)
HI & LO Limit Range: 0 - 600 mΩ for 3 - 10 A
0 - 150 mΩ for 3 - 30 A
Resolution: 1 mΩ/step
Accuracy: ± (2% of setting + 2 mΩ)
Dwell Timer Range: 0, 0.5 - 999.9 sec (0 = Continuous)
Resolution: 0.1 sec/step
Accuracy: ± (0.1% + 0.05 sec)
Milliohm Offset Maximum Offset Capability: 200 mΩ
Resolution: 1 mΩ/step
Accuracy: ± (2% of setting + 2 mΩ)
General Specifications
Mechanical
Tilt up front feet
Dimensions 7700/7704
7705/7710/7715/7720
(WxHxD) 17 x 5.8 x 16.5 in. (432 x 147 x 419 mm)
(WxHxD) 16.93 x 5.24 x 15.75 in. (430 x 133 x 400 mm)
Weight
61.65 lbs (28 kgs)
68.75 lbs (31.25 kgs)
48.9 lbs (22 kgs)
48.9 lbs (22 kgs)
48.9 lbs (22 kgs)
48.9 lbs (22 kgs)
7700
7704
7705
7710
7715
7720
Interface
Standard USB/RS-232, Optional GPIB
Memory
7700/7704 50 memories w/8 Steps per memory 7705/7710 50 memories
7715/7720
Specifications subject to change without notice.
Accredited calibration service available. Includes ISO 17025,
ANSI Z540.1-1994, CTL & Denan’s Law requirements.
For more information on testing to a specfic standard, refer back to the
Common Safety Standard Reference Chart.
Fully-Automated Dielectric Withstand Analyzer
HypotULTRA®III is a multi-function dielectric analyzer with an enhanced graphic LCD. Choose from two models: the
7620 AC Hipot tester and the 7650 AC/DC/IR tester. Both testers include an optional 4-port or 8-port built-in scanner.
An additional external modular scanner is available for use with both testers. All testers come standard with USB and
RS-232 interfaces. Ethernet, GPIB, and RS-485 interfaces are also available.
Model 7620 -5 kV AC Hipot Tester. Internal 4 or 8 Port
Scanning Matrix available
Model 7650 -5 kV AC Hipot Tester, 5 kV DC Hipot Tester
& Insulation Resistance Tester. Internal 4 or 8 Port
Scanning Matrix available
Features and Benefits
Patented SmartGFI® safety circuit protects the operator from shock hazards
USB/RS-232, GPIB, Ethernet,
or RS-485 automation interfaces
available
Patented VERI-CHEK® feature
prompts the users through steps to validate the instrument’s operation
Data Storage card available
for storing and transferring test data without a connection to a PC
Patented Prompt and Hold
function provides a unique method for performing multiple steps during a test cycle
Patented CAL-ALERT®
alerts the operator when
the HypotULTRA III is due
for re-calibration
RAMP HI® and CHARGE LO®
for more effective DC
Hipot testing
Two Continuity Test modes
allow for simultaneous continuity tests during Hipot testing as well as point-to-point continuity testing
Graphic LCD and intuitive
menu system to simplify
the entire testing process
from set-up to results
50 memories with 30 steps
per memory that can be
stored and recalled in any alphanumeric combination
Real Current measurement
allows operators to monitor total and real current on
a single screen
Advanced functionality available
with an optional 4 or 8 port internal scanner
Autoware Testing Software
available for complete
Automation Control
Safety agency listed.
Input Specifications
Dielectric Withstand Test Mode
Voltage 115 / 230 VAC ± 10%, Automatically Selected
Dielectric Withstand Test Mode
Current Display
DC
Output Rating 5 kV @ 30 mAAC
5 kV @ 10 mADC for 7650 only
DC Output Ripple ≤ 4% Ripple rms at 5 kVDC @ 10 mA, Resistive Load
Frequency 50/60 Hz ± 5%
Fuse 4 Amp 250 V Slow Blow
Output Adjustment Range:
0 – 5000 VAC
0 – 5000 VDC for 7650 only
Resolution: 1 V
Accuracy: ± (2% of setting + 5 volts)
(Can be adjusted during operation.
Disabled when key lockout is active.)
Ramp-HI 12 mA peak maximum, ON/OFF selectable
Charge-LO Range:
0.0 - 350.0 µA DC or Auto set
Maximum & Minimum
Limits
AC Total Range 1: 0.000 – 9.999 mA
Resolution: 0.001 mA
Range 2: 10.00 – 30.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of setting + 2 counts)
AC Real Range 1: 0.000 – 9.999 mA
Resolution: 0.001 mA
Range 2: 10.00 – 30.00 mA
Resolution: 0.01 mA
Accuracy: (3% of setting + 0.05 mA) All Ranges
PF > 0.1
V > 250 VAC
DC Range 1: 0.0 – 999.9 μA for 7650 only
Resolution: 0.1 μA
Range 2: 1000 – 10000 μA for 7650 only
Resolution: 1 μA
Accuracy: ± (2% of setting + 2 counts)
Arc Detection Range:
1-9
Voltage Display Range:
0.00 - 5.00 kV Full Scale
Resolution: 10 V
Accuracy: ± (2% of setting + 2 counts)
Current Display Auto Range
AC Total Range 1: 0.000 mA – 3.500 mA
Resolution: 0.001 mA
Range 2: 3.00 – 30.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of reading + 2 counts)
AC Real Range:
0.000 mA – 30.00 mA
Resolution: 0.001 mA or 0.01 mA
Accuracy: ± (3% of reading + 0.05 mA) All Ranges
PF > 0.1
V > 250 VAC
Range 1:
Resolution:
Range 2:
Resolution:
Range 3:
Resolution:
Accuracy: (continued)
0.0 μA – 350.0 μA for 7650 only
0.1 μA
0.300 mA – 3.500 mA for 7650 only
0.001 mA
3.00 mA – 9.99 mA for 7650 only
0.01 mA
± (2% of reading + 2 counts)
Discharge Time ≤ 200 ms
Maximum Capacitive 1 μF----< 1 kV Load in DC Mode 0.75 μF----< 2 kV 0.5 μF----< 3 kV
0.08 μF----< 4 kV
0.04 μF----< 5 kV
AC Output Wave Form Sine Wave, Crest Factor = 1.3 - 1.5
Output Frequency Range:
Accuracy:
60 or 50 Hz, User Selection
± 0.1%
Output Regulation ± (1 % of output + 5 V)
From no load to full load and over input voltage range
Dwell Timer Range:
0.0, 0.4 - 999.9 sec (0 = Continuous)
Ramp Timer Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: AC 0.0 - 999.9 sec
DC: 0.0, 1.0 - 999.9 sec
0.0=OFF
Ground Continuity Current:
DC 0.1 A ± 0.01 A, fixed
Max. ground resistance:
1 Ω ± 0.1 Ω, fixed
Ground Fault Interrupt GFI Trip Current: 450 µA max (AC or DC)
HV Shut Down Speed:
< 1 ms
Insulation Resistance Test Mode
(Model 7650 Only)
Output Voltage Range:
50 - 1000 VDC
Resolution: 1 V
Accuracy: ± (2% of reading + 2 counts)
Short Circuit CurrentMaximum:
12 mA peak
Voltage Display Range:
0 – 1000 V
Resolution: 1 V
Accuracy: ± (2% of reading + 2 counts)
Insulation Resistance Test Mode
Model 7650 Only (continued)
Resistance Display Range: 0.05 MΩ - 50000 MΩ (5 Digit, Auto Ranging)
Resolution: 500 VDC
1000 VDC
MΩ
0.001
0.01
0.1
1
MΩ MΩ
0.050 - 9.999 0.100 - 9.999
1.00 - 99.99
1.00 - 99.99
10.0 - 999.9
10.0 - 999.9
100 - 50000
100 - 50000
Accuracy:
50 – 499 V
0.05 MΩ – 999.9 MΩ
± (7% of reading + 2 counts)
500 – 1000 V
0.10 MΩ – 999.9 MΩ
± (2% of reading + 2 counts)
1000 MΩ – 9999 MΩ
± (5% of reading + 2 counts)
10000 MΩ – 50000 MΩ
± (15% of reading + 2 counts)
Charge-LO Range:
0.000 - 3.500 µA or Auto Set
Maximum and Range: 0.0, 0.05 MΩ – 99.99 MΩ
Minimum Limits Resolution: 0.01 MΩ
Range: 100.0 MΩ – 999.9 MΩ
Resolution: 0.1 MΩ
Range: 1000 MΩ – 50000 MΩ
Resolution: 1 MΩ
(Max Limit: 0 = OFF)
Accuracy: Same as Resistance Display Accuracy
Ramp Timer Range:
Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: 0.0, 1.0 - 999.9 sec
Delay Timer Range:
0.0, 1.0 - 999.9 sec 0 = Continuous
Ground Fault GFI Trip Current: 450 µA max
Interrupt
HV Shut Down Speed: < 1 ms
Continuity Test Mode
Output Current DC 0.1 A ± 0.01 A Total Resistance*: 0.00-33.0 Ω
DC 0.01 A ± 0.001 A Total Resistance*: 31.0-330 Ω
DC 0.001 A ± 0.0001 A Total Resistance*: 310-2000 Ω
Resistance Display Range 1:0.00 – 19.99 Ω
Resolution: 0.01 Ω
Accuracy: ± (1 % of reading + 0.05 Ω)
Range 2: 20.0 – 199.9 Ω
Resolution: 0.1 Ω
Accuracy: ± (1 % of reading + 0.2 Ω)
Range 3: 200 – 2000 Ω
Resolution: 1 Ω
Accuracy: ± (1 % of reading + 2 Ω)
*Total Resistance of Test Leads, Fixture and DUT.
Continuity Test Mode (continued)
Maximum and Range 1: 0.00 – 99.99 Ω
Minimum Limits Resolution: 0.01 Ω
Accuracy: ± (1% of setting+0.05 Ω)
Range 2: 100.0 – 999.9 Ω
Resolution: 0.1 Ω
Accuracy: ± (1% of setting+0.2 Ω)
Range 3: 1000 – 2000 Ω
Resolution: 1 Ω
Accuracy: ± (1% of setting+2 Ω)
(Max Limit: 0 = OFF)
Dwell Timer Range:
0.0, 0.3 - 999.9 sec (0 = Continuous)
Milliohm Offset Range:
0.00 - 10.00 Ω
General Specifications
Mechanical Bench or rack mount (2U height) with tilt up front feet
Dimensions (w x h x d) 16.92 x 3.50 x 15.75 in (430 x 89 x 400 mm)
Weight 31.38 Lbs (14.23 kgs) variable with options
Interface Standard USB/RS-232
Optional Ethernet, GPIB, Data Storage (RS-485) or
Printer Port with Date and Time Stamp
Memory 50 memories, 30 steps/memory
Specifications subject to change without notice.
Accredited calibration service available. Includes ISO 17025,
ANSI Z540.1-1994, CTL & Denan’s Law requirements.
For more information on testing to a specfic standard, refer back to the
Common Safety Standard Reference Chart.
Production Line Dielectric Withstand Testers
Hypot® III is a bench top Dielectric Withstand tester with an enhanced graphic LCD. Four models are available: the
3705 AC Hipot tester, the 3765 AC/DC Hipot tester, the 3770 AC/DC Hipot tester with built-in Insulation Resistance,
and the 3780 500 VA AC Hipot tester. All testers feature an RS-232 interface for entry-level automation.
Model 3705 - 5 kV AC Hipot Tester
Model 3765 - 5 kV AC, 6 kV DC Hipot Tester
Model 3770 - 5 kV AC, 6 kV DC Hipot & Insulation Resistance Tester
Model 3780 - 500 VA (5 kVA @ 100 mA) AC Hipot Tester
Features and Benefits
RS-232 interface standard
for entry-level automation
Patented SmartGFI® safety
circuit protects the operator
from shock hazards
10 Memories with 3 Steps
per memory for storing and
recalling test parameters
PLC Remote Control for
simple remote operation
Patented VERI-CHEK® feature
prompts users through steps
to validate the instrument’s
operation
Interconnects with a HYAMP III
Associated Research Ground
Bond tester to form a
complete test system
Patented CAL-ALERT® feature
alerts the operator that the tester is due for re-calibration
Digitally controlled arc
detection circuit allows the
operator to program sensitivity
levels for detecting arcs
Built-in adjustable Continuity
test for checking basic
continuity
Graphic LCD provides intuitive
menu system to simplify the
entire testing process from
set-up to results
Remote Safety Interlock feature
prevents the high-voltage from
being activated without the
interlock enabled
Minimum and maximum trip
settings for safer and more
accurate testing
Comes complete with an
adapter box for products
terminated in a line cord
Meets 200 mA Short circuit
requirements (Model 3780)
Models 3705, 3765 & 3770.
Model 3780.
3705, 3765, 3770 Specifications
Input Specifications
Dielectric Withstand Test Mode (continued)
Voltage 115/230 VAC ± 10%, user selectable
Ground Continuity Current
Frequency 50/60 Hz ± 5%
Ground Continuity Range: 0.0 Ω - 1.50 Ω
Maximum Limit Resolution: 0.01 Ω
Minimum Limit Accuracy: ± (3% of setting + 0.02 Ω)
Fuse 3.15 A, fast acting 250 VAC
DC 0.1 A ± 0.01 A, fixed
Dielectric Withstand Test Mode
Ground Continuity Range: 0.0 Ω - 0.50 Ω
Auto Offset Resolution: 0.01 Ω
Accuracy: ± (3% of setting + 0.02 Ω)
Voltage Setting Range:
0 - 5.00 kVAC
0 - 6.00 kVDC
Resolution: 0.01 kV
Accuracy: ± (2% of setting + 5 V)
Insulation Resistance Test Mode
Output Rating 5000 V @ 20 mAAC
6000 V @ 7.5 mADC
Maximum Limit AC
DC
Range:
0.00 - 20.00 mA
Resolution: 0.01 mA
Range:
0 - 7500 µA
Resolution: 1 µA
Accuracy: AC and DC ± (2% of setting + 2 counts)
Minimum Limit AC
DC
Range:
0.000 - 9.999 mA
Resolution: 0.001 mA
Range: 0.0 - 999.9 µA
Resolution: 0.1 µA
Accuracy: AC and DC ± (2% of setting + 2 counts)
Arc Detection Range: 0 - 9, 0 disabled
Ground Fault GFI Trip Current: 450 µA max (AC or DC)
Interrupt HV Shut Down Speed: < 1ms
Current Display
AC
DC
Auto Range
Range 1: 0.000 - 3.500 mA
Resolution: 0.001 mA
Range 2: 3.00 - 20.00 mA
Resolution: 0.01 mA
Range 1: 0.0 µA - 350.0 µA
Resolution: 1: 0.1 µA
Range 2: 0.300 mA - 3.500 mA
Resolution: 0.001 mA
Range 3: 3.00 mA - 7.50 mA
Resolution: 0.01 mA
Accuracy: All Ranges ± (2% of reading + 2 counts)
DC Output Ripple ≤ 5% Ripple rms at 6 kVDC @ 7.5 mA, Resistive Load
Discharge Time
≤ 200 ms
The maximum capacitive load vs output voltage:
0.20 µF < 1 kV
0.050 µF < 4 kV
0.10 µF < 2 kV
0.040 µF < 5 kV
0.06 µF < 3 kV
0.015 µF < 6 kV
AC Voltage WaveformSine Wave, Crest Factor = 1.3 - 1.5
Voltage Setting Range: 30 - 1000 VDC
Resolution: 1 V
Accuracy: ± (2% of setting + 5 V)
Resistance Display Range: 1 - 9999 MΩ (4 Digit, Auto Ranging)
Resolution: 500 VDC - 1000 VDC
MΩ
MΩ
0.001
1.000 - 9.999
0.01
10.00 - 99.99
0.1
100.0 - 999.9
1
1000 - 9999
Accuracy: ± (2% of reading + 2 counts) at test voltage
500 - 1000 V and 1 - 999.9 MΩ
± (5% of reading + 2 counts) at test voltage
500 - 1000 V and 1000 - 9999 MΩ
± (8% of reading + 2 counts) at test voltage
30 - 500 V and 1 - 1000 MΩ
Maximum Limit Range: 0, 1 - 9999 MΩ (0=OFF)
Resolution: 1 MΩ
Accuracy: Same as Resistance Display
Minimum Limit Range: 1 - 9999 MΩ
Resolution: 1 MΩ
Accuracy: Same as Resistance Display
Ramp Timer Range:
Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: 1.0 - 999.9 sec (0=OFF)
Resolution: 0.1 sec
Accuracy: ± (0.1% of reading + 0.05 sec)
Delay Timer Range: 0, 0.5 - 999.9 sec (0 = Continuous)
Resolution: 0.1 sec
Accuracy: ± (0.1% of reading + 0.05 sec)
GFI Trip Current 450 µA max
HV Shut Down Speed < 1 ms
General Specifications
Output Frequency Range: 50 or 60 Hz, User Selectable
Mechanical
Bench or rack mount with tilt up feet
Output Voltage ± (1% of output + 5 V) from no load to full load and over
Regulation input voltage range.
Dimensions
(W x H x D) 8.46 x 3.5 x 14.57 in.
(215 x 89 x 370 mm)
Dwell Timer Range:
AC 0, 0.3 - 999.9 sec (0 = Continuous)
DC 0, 0.4 - 999.9 sec (0 = Continuous)
Accuracy: ± (0.1% of reading + 0.05 sec)
Weight 20.96 lbs (9.53 kgs)
Interface
RS-232 interface standard for entry-level automation
Memory
10 Memories, 3 steps per memory
Ramp Timer Range: Ramp-Up: 0.1 - 999.9 sec
Ramp-Down:AC 0.0 - 999.9 sec
DC 1.0 - 999.9 sec (0=OFF)
Accuracy: ± (0.1% of reading + 0.05 sec)
Specifications subject to change without notice.
3780 Specifications
Input Specifications
General Specifications
Mechanical
Bench or rack mount with tilt up feet
Frequency 50/60 Hz ± 5%
Dimensions
(W x H x D) 16.93 x 5.24 x 13.78 in.
(430 x 133 x 350 mm)
Weight 49 lbs (23 kgs)
Output Rating 5000 V @ 100 mAAC
Interface
RS-232 interface standard for entry-level automation
Voltage Setting Range:
0 – 5.00 kVAC
Resolution: 0.01 kV
Accuracy: ± (2% of setting + 0.01 kV)
(Adjustable during operation.
Disable when key lockout is active.)
Memory
10 Memories, 3 steps per memory
Maximum Limit AC Range:
0.00 – 99.99 mA
Resolution: 0.01 mA
Accuracy: ± (2% of setting + 6 counts)
Specifications subject to change without notice.
Voltage 115/230 VAC ± 15%, automatically selected
Fuse 15 Amp, Slow Blow 250 VAC
Dielectric Withstand Test Mode
Minimum Limit AC Range:
0.000 – 9.999 mA
Resolution: 0.001 mA
Accuracy: ± (2% of setting + 6 counts)
Arc Detection Range: 0 - 9, 0 disabled
Ground Fault GFI Trip Current: 450 µA max
Interrupt HV Shut Down Speed: < 1ms
Current Display
Auto Range AC
Range 1: 0.000 mA – 3.500 mA
Resolution: 0.001 mA
Accuracy: ± (2% of setting + 2 counts)
Range 2: 3.00 – 99.99 mA
Resolution: 0.01 mA
Accuracy: ± (2% of setting + 6 counts)
AC Voltage Wave Form Sine Wave, Crest Factor = 1.3 – 1.5
Output Frequency Range: 50 or 60 Hz, User Selectable
Accuracy: ± 0.1%
Output Voltage ± (1 % of output + 5 V)
Regulation from no load to full load and over input voltage range.
Dwell Timer Range:
0, 0.3 – 999.9 sec (0 = Constant)
Resolution: 0.1 sec
Accuracy: ± (0.1% of reading + 0.05 sec)
Ramp Timer Range: Ramp-Up: 0.1 - 999.9 sec
Ramp-Down:AC 0.0 - 999.9 sec
Resolution: 0.1 sec
Accuracy: ± (0.1% of reading + 0.05 sec)
Ground Continuity Current
DC 0.1 A ± 0.01 A, fixed
Ground Continuity Range: 0.0 Ω - 1.50 Ω
Maximum Limit Resolution: 0.01 Ω
Minimum Limit Accuracy: ± (3% of setting + 0.02 Ω)
Ground Continuity Range: 0.0 Ω - 0.50 Ω
Auto Offset Resolution: 0.01 Ω
Accuracy: ± (3% of setting + 0.02 Ω)
Output Short Circuit > 200 mA
Current
Accredited calibration service available. Includes ISO 17025,
ANSI Z540.1-1994, CTL & Denan’s Law requirements.
For more information on testing to a specfic standard, refer back to the
Common Safety Standard Reference Chart.
Production Line Ground Bond Testers
HYAMP® III is a microprocessor controlled Ground Bond tester with an enhanced graphic LCD. Three models are
available: the 3130 with 30 Amp output, the 3140 with 40 Amp output and the 3160 with 60 Amp output. Model
3130 also has an optional external RS-232 interface. HYAMP III can be interconnected with Hypot III to form a
complete test system.
Model 3130 - 30 Amp Ground Bond Tester
Model 3140 - 40 Amp Ground Bond Tester
Model 3160 - 60 Amp Ground Bond Tester
Features and Benefits
External RS-232 interface
available for entry-level
automation (Model 3130)
10 Memories with 3 Steps
per memory for storing and
recalling test parameters
Patented VERI-CHEK® feature
prompts users through steps
to validate the instrument’s
operation
Interconnects with an
Associated Research Hipot
tester to form a complete
test system
Graphic LCD provides intuitive
menu system to simplify the
entire testing process from
set-up to results
4 wire measurement and
milliohm offset for accurate
Ground Bond test results
Patented CAL-ALERT® feature
alerts the operator that the
tester is due for re-calibration
PLC Remote Control allows
operators to remotely control
the Ground Bond tester
Electronic Dwell timer for
more consistent and
reliable testing
Adjustable output current
and milliohm trip ranges to
meet all safety agency
specifications for Ground
Bond test requirements
Safety agency listed.
Input Specifications
Ground Bond Test Mode (Continued)
Voltage
115/230 VAC ± 10%, user selectable
Frequency
50/60 Hz ± 5%
Fuse - 3130
6.3 A, Slow Blow 250 VAC
Fuse - 3140
10 A, Slow Blow 250 VAC
Ohmmeter Display 3140 Range: 0 - 150 mΩ for 30.01 - 40.00 A
0 - 200 mΩ for 10.01 - 30.00 A
0 - 600 mΩ for 6.00 - 10.00 A
Resolution: 1 mΩ
Accuracy: ± (2% of reading + 2 mΩ)
Range: 0 - 600 mΩ for 1.00 - 5.99 A
Resolution: 1 mΩ
Accuracy: ± (3% of reading + 3 mΩ)
Fuse - 3160
15 A, Slow Blow 250 VAC
Ground Bond Test Mode
Output Rating
Current 3130: 1.00 - 30.00 AAC
Voltage 3130: 6 VAC, fixed
Current 3140: 1.00 - 40.00 AAC
Voltage 3140: 8 VAC, fixed
Current 3160: 1.00 - 60.00 AAC
Voltage 3160: 9 VAC, fixed
Resolution: 0.01 A
Regulation: ± (2% of setting + 0.02 A)
Output Frequency
Range: 50 / 60 Hz, User Selectable
Dwell Time Setting
Range: 0 and 0.5 - 999.9 secs
0 for continuous running
Resolution: 0.1 sec
Accuracy: ± (0.1% of setting + 0.05 secs)
Ohmmeter Display 3160 Range: 0 - 150 mΩ for 30.01 - 60.00 A
0 - 300 mΩ for 15.01 - 30.00 A
0 - 600 mΩ for 6.00 - 15.00 A
Resolution: 1 mΩ
Accuracy: ± (2% of reading + 2 mΩ)
Range: 0 - 600 mΩ for 1.00 - 5.99 A
Resolution: 1 mΩ
Accuracy: ± (3% of reading + 3 mΩ)
Timer Display Range: 0.0 - 999.9 secs
Resolution: 0.1 secs
Accuracy: ± (0.1% of reading + 0.05 secs)
General Specifications
Mechanical
Bench or rack mount with tilt up feet.
Dimensions 3130
(W x H x D) 8.5 x 4.0 x 15.5 in. (216 x 103 x 390 mm)
includes feet
Maximum & Minimum Range 3130: 0 - 120 mΩ for 1 - 30.00 A
Limits
0 - 510 mΩ for 1 - 10.00 A
Accuracy 3130: ± (2% of setting + 2 mΩ)
Range 3140: 0 - 150 mΩ for 30.01 - 40.00 A
0 - 200 mΩ for 10.01 - 30.00 A
0 - 600 mΩ for 1.00 - 10.00 A
Accuracy 3140: ± (3% of setting + 3 mΩ)
Range 3160: 0 - 150 mΩ for 30.01 - 60.00 A
0 - 300 mΩ for 15.01 - 30.00 A
0 - 600 mΩ for 1.00 - 15.00 A
Accuracy 3160: ± (3% of setting + 3 mΩ)
Dimensions 3140
(W x H x D) 8.5 x 4.0 x 16.9 in. (216 x 103 x 430 mm)
includes feet
Dimensions 3160
(W x H x D) 16.9 x 5.1 x 15.7 in. (430 x 130 x 400 mm)
includes feet
Weight 3130
19.15 lbs (8.7 kgs)
Weight 3140
30.9 lbs (14 kgs)
Weight 3160
49.40 lbs (22.40 kgs)
Offset Capability
Remote Control &
Signal Output
The following input and output signals are provided
through two 9 pin D type connectors:
1. Remote control: Test, Reset, Interlock, and
Withstand Processing
2. Remote recall of memory program #1, #2, and #3
3. Outputs: Pass, Fail, Test-in-process, Start Out
and Reset Out
Program Memory
10 Memories, 3 steps per memory
Interface
External RS-232 interface for Model 3130
Range: 0 - 100 mΩ
Resolution: 1 mΩ
Accuracy: ± (2% of setting + 2 mΩ)
Current Display 3130 Range: 0.00 - 30.00 A
Resolution: 0.01 A / step
Accuracy: ± (3% of reading + 0.03 A)
Current Display 3140 Range: 0.00 - 40.00 A
Resolution: 0.01 A
Accuracy: ± (3% of reading + 0.03 A)
Current Display 3160 Range: 0.00 - 60.00 A
Resolution: 0.01 A
Accuracy: ± (3% of reading + 0.03 A)
Ohmmeter Display 3130 Range: 0 - 510 mΩ
Resolution: 1 mΩ / step
Accuracy: ± (2% of reading + 2 mΩ)
Specifications subject to change without notice.
Accredited calibration service available. Includes ISO 17025,
ANSI Z540.1-1994, CTL & Denan’s Law requirements.
For more information on testing to a specfic standard, refer back to the
Common Safety Standard Reference Chart.
Fully-Automated Line Leakage Tester
The 620L is a stand alone Line Leakage tester with an enhanced graphic LCD which automates leakage testing in production
and laboratory environments. The 620L is configured for up to 40 Amps of current draw for DUT input power. It is designed
to test to most safety agency standards for Line Leakage testing. The 620L comes standard with USB and RS-232 interfaces.
Ethernet, GPIB, and RS-485 interfaces are also available.
Model 620L - Fully-Automated Line Leakage Tester
Features and Benefits
Test operators can configure
the 620L to perform all 8 required Line Leakage tests
Interconnection to SC6540
Modular Scanner provides
automated control of multiple
test points
Leakage current readings can
be monitored using both PEAK
and RMS measurements
Graphic LCD and intuitive
menu system to simplify the
entire testing process
Most common measuring
devices (MD’s) are already incorporated into the instrument’s intuitive
menu system
Patented CAL-ALERT® alerts the
operator that the 620L is due
for re-calibration
50 Memories with 30 steps
per memory can be stored and recalled in any alphanumeric combination
Compact 3U Rack
Mount Design
Optional Functional Run
Testing for additional measurements
Interconnection to APT Brand
AC Power Source
Handles up to 40 Amp
maximum continuous
DUT Current
Optional cold resistance
measurement capability
USB/RS-232, GPIB, Ethernet,
or RS-485 automation
interfaces available
Easily Interconnect to any
automated Associated
Research Hipot Tester
Autoware Testing Software
available for complete
Automation Control
APT
AC Power Source
Compatible
www.aspowertechnologies.com
Safety agency listed.
Input Specifications
Measuring Device Module
Voltage
115/230 VAC ± 10%, user selection
Frequency
50/60 Hz ± 5%
Fuse
2 A Slow Blow 250 VAC
MD1
UL544NP, UL484 , UL923, UL471, UL867, UL697
MD2UL544P
MD3
IEC 60601-1
MD4UL1563
MD5
IEC60990 Fig4 U2, IEC60950-1, IEC60335-1,
IEC60598-1,IEC60065, IEC61010
MD6
IEC60990 Fig5 U3, IEC60598-1
MD7
IEC60950, IEC61010-1 FigA.2 (2 kohm) for Run function
External MD
Basic measuring element 1 kohm
MD Voltage Limit 70 VDC
Line Conditions
Reverse Power
Switch
Neutral Switch
Ground Switch
Reverse polarity switch for normal condition
Neutral switch on/off selection for single fault
Ground switch on/off selection for class I single fault
DUT Power
Probe Settings
Surface to Surface (PH - PL)
Surface to Line
(PH - L)
Ground to Line (G - L)
Leakage Limit Settings
Touch Current
High/Low
Range:
0.0 µA - 999.9 µA / 1000 µA - 9999 µA /
Limit (RMS)
10.00 mA - 20.00 mA
Resolution: 0.1 µA/1 µA/0.01 mA
Touch Current
High/Low
Range:
0.0 µA - 999.9 µA/ 1000 µA - 9999 µA /
Limit (Peak)
10.00 mA - 30.00 mA
Resolution: 0.1 µA/1 µA/0.01 mA
0.0 - 277.0 V
AC Current
40 Arms max continuous
AC Voltage
High/Low Limit
Range:
0.0 - 277.0 V
Resolution: 0.1 V/step
AC Voltage Display Range:
0.0 - 277.0 V
Resolution: 0.1 V/step
Accuracy: ± (1.5% of reading + 2 counts), 30.0 - 277.0 V
Delay time setting Range:
0.5 – 999.9 sec
Resolution: 0.1 sec
Dwell time setting
Display
Touch Current
Display (RMS)
Range:
Resolution:
Accuracy:
Range:
Resolution:
Accuracy:
Range:
Resolution:
Accuracy:
Touch Current
Display (Peak)
Range:
Resolution:
Accuracy:
Range:
Resolution:
Accuracy:
Range:
Resolution:
Accuracy:
AC Voltage
0.0 µA - 550 µA, frequency DC, 15 Hz - 1 MHz
0.1 µA
DC: 15 Hz < f < 100 kHz: ±(2% of reading + 3 counts)
100 kHz < f < 1 MHz: ±5% of reading,
(10.0 µA - 999.9 µA)
400 µA - 8500 µA, frequency DC, 15 Hz - 1 MHz
1 µA
DC: 15 Hz < f < 100 kHz: ±(2% of reading + 3 counts)
100 kHz < f < 1 MHz: ±5% of reading,
(10 µA - 8500 µA)
8.00 mA - 20.00 mA, frequency DC, 15 Hz - 1 MHz
0.01 mA
DC: 15 Hz < f < 100 MHz: ±5% of reading,
(0.01 mA -20.00 mA)
0.0 µA - 550 µA, frequency DC - 1 MHz
0.1 µA
DC : ±(2% of reading +3 counts)
15 Hz < f < 1 MHZ : ±10% of reading +2 µA
400 µA - 8500 µA, frequency DC - 1 MHz
1 µA
DC: ±(2% of reading +3 counts)
15 Hz < f < 1 MHz : ±10% of reading +2 µA 8.00 mA - 30.00 mA, frequency DC - 100 kHz
0.01 mA
DC: ±(2% of reading +3 counts)
15 Hz < f < 100 kHz : ±10% of reading +2 counts
Range:
0, 0.5 – 999.9 sec (0=Continuous)
Resolution: 0.1 sec
Accuracy: ± (0.1% of reading + 0.05 seconds)
Failure Protection (Start-Up) - Neutral Voltage Check (Neutral-V)
Over current and ground current check (Line - OC)
General Specifications
Dimension
(W x H x D) 16.93 x 5.24 x 11.81
(430 x 133 x 300 mm)
Weight
26.45 lbs (12 kgs)
Display
320 X 240 graphic LCD
Mechanical
Bench or rack mount with tilt up feet
Memory
50 Memories, 30 steps per each memory
File locations can link 900 steps max
Interface
USB/RS232 Standard, Ethernet, GPIB, Data Storage
(RS-485) Optional
Specifications subject to change without notice.
Accredited calibration service available. Includes ISO 17025,
ANSI Z540.1-1994, CTL & Denan’s Law requirements.
For more information on testing to a specfic standard, refer back to the
Common Safety Standard Reference Chart.
High Voltage and High Current Modular Scanning Matrix
The SC6540 modular scanner is designed to automate multi-point and multi-product testing when using OMNIA,
HypotULTRA III, LINECHEK II or HypotMAX testers 7700 and 7704. There are 10 different configurations available that are
built off of two basic scanning configurations determined by the power module. A master scanner (M) is configured with its
own power module and controlled directly through automation software. It is available with either a USB/RS-232, GPIB, or
Ethernet Interface. A slave scanner (S) is configured without a power module and is controlled either by a master scanner or
the electrical safety tester. The SC6540 can be configured with 8 HV (high voltage), 16 HV, 8 HV and 8 GB (Ground Bond),
8 GB or 16 GB testing channels. The patented modular design provides a flexible testing solution that is configurable to a
manufacturer’s needs. Designed to interconnect with most of our automated electrical safety testers, the SC6540 allows for
automated multi-point or multi-product safety testing. This is an ideal solution for applications such as transformers, motors,
cables or any DUT that requires tests between various points.
Model SC6540 HNM - 8 Channel High Voltage Scanner
Model SC6540 HHM - 16 Channel High Voltage Scanner
Model SC6540 HGM - 8 Channel High Voltage Scanner
8 Channel High Current Scanner
Model SC6540 GNM - 8 Channel High Current Scanner
Model SC6540 GGM - 16 Channel High Current Scanner
Features and Benefits
Configurable scanning matrix
Multi-point or multi-product testing
capabilities
Automation interfaces for
Autoware® control
Point-to-point continuity tests
Compatible with Hipot, Ground
Bond, Line Leakage and Insulation
Resistance Tests
Up to 16 high voltage
switching channels on a single
scanner
High current outputs rated up
to 40 Amps
Up to 80 testing points from a
single power source
Compact 2U rack mount
design
Safety agency listed.
Modular Scanning Matrix Specifications
Input (Master only)
115 VAC (+/- 10%), 50/60 Hz, single phase
230 VAC (+/- 10%), 50/60 Hz, single phase
User selectable
Fuse (Master only)
250 V/2 A/fast-blow
PC Control (Master only)
Choice of Ethernet, GPIB or USB/RS-232
Scanner Control
Master: one scanner bus output controls
up to 4 additional slaves
Slave: one output and one input
Maximum HV Rating
5 kV AC and DC
Maximum HC Rating
40 Amps
Model SC6540 HNM*
8 Channel High Voltage Scanner
Model SC6540 HHM*
16 Channel High Voltage Scanner
No. of Possible Channels 8 or 16
HV Output Terminations
100 ft. reel HV cable rated for up to 30 kV
with 8 HV connectors
GND Output Terminations 20 terminals provided, to accept 10/12 AWG
hook-up wire (user supplied wire)
Model SC6540 HGM*
Temperature
32˚ - 104˚ F (0˚ - 40˚ C)
8 Channel High Voltage Scanner
8 Channel High Current Scanner
Humidity
0 - 80%
Altitude
6560 ft. (2000 m)
Dimensions
2U with tilt-up front feet
(WxHxD) 17 x 4.07 x 12.96 in.
(432 x 103 x 329 mm)
Weight Master: 20.05 lbs. max. (9.09 kgs)
(with 2 high voltage modules)
Slave: 15.45 lbs. max. (7.01 kgs)
(with 2 high voltage modules)
Configurations
The modular design allows for a variety of configurations. In addition to master
or slave configurations, the scanners can also be set-up with the following
configurations, 8 or 16 high voltage testing channels, 8 high voltage and ground
bond testing channels, and 8 or 16 ground bond testing channels.
The different
configurations
(shown right)
are indicated
by the following
alpha designators
M
H
HH
G
GG
N
S
= Master Scanner
= 8 High Voltage Channels
= 16 High Voltage Channels
= 8 Ground Bond Channels
= 16 Ground Bond Channels
= Empty Module
= Slave
Model SC6540 GNM*
8 Channel High Current Scanner
Model SC6540 GGM*
16 Channel High Current Scanner
*Also available in slave configuration.
Specifications subject to change without notice.
Accredited calibration service available. Includes ISO 17025,
ANSI Z540.1-1994, CTL & Denan’s Law requirements.
For more information on testing to a specfic standard, refer back to the
Common Safety Standard Reference Chart.”
Medical Safety Testing System
MedTEST™ provides advanced test functionality while increasing
efficiency for medical device manufacturers. It complies with
test requirements called out in common medical electrical safety
specifications such as UL2601, UL60601-1, IEC601-1, IEC60601-1,
EN60601-1 and more. It performs patient lead testing on medical
devices while continuously running your DUT. This feature can offer
significant time savings as the Device Under Test (DUT) does not need
to be powered down and up to perform the multiple leakage tests,
even in reverse polarity conditions.
MedTEST™ can accommodate several types of tests including Hipot,
Ground Bond and various Line Leakage tests. MedTEST™ provides
all this without requiring the test operator to disconnect and reconnect
test leads to perform the different tests.
Features and Benefits
MedTEST combines all of
the most common electrical
safety tests required by
safety agencies (AC Hipot,
DC Hipot, IR Test, Ground
Bond/Continuity, Line
Leakage and Functional Run
Test) into a single system
which can be enclosed in a
standard rack mount cabinet
Completely automated
scanner matrix setup
MedTEST can also perform
all B, BF and CF type
applied part tests
Integration with APT Brand
AC Power Source
Continuous Run capability
even in reverse polarity
Up to 40 Amp continuous
current capability
Advanced Software Control
with Autoware II
Expanded reporting capability
with barcode scanning, data
storage, report generation and
data printouts
Programmable test setups
with accompanying steps
7 built-in measuring devices
with external connection
capability
APT
AC Power Source
Compatible
www.aspowertechnologies.com
MedTEST TM provides advanced test functionality
while increasing efficiency for medical device manufacturers.
MedTEST is designed to run with our Autoware
software. This enables the user to have complete
computer control of the test system. MedTEST will
connect to a PC through a host of PC interfaces
including USB/RS-232, Ethernet or GPIB.
Autoware allows the MedTEST system to be remotely
programmed and set-up. It provides programmable
memories and steps which can be saved and
recalled resulting in more efficient testing. Each test
memory can store up to 30 test steps which can
be configured to perform any of the safety tests.
All these test steps can be linked together to form a
complete automated test sequence. Further, all test
memories can be linked together creating a virtually
unlimited number of test steps. Complete data
capture is also easily achieved with the MedTEST
system. Test results for every single test can be
viewed in a statistical format, exported for database
archiving, or even directly sent to a print report.
The MedTEST system is designed to provide you with a custom test solution. It can be configured to meet almost any application
and since the configuration incorporates existing Associated Research testers you don’t end up paying for a custom solution and
still maintain the ability to meet ever changing demands for your test equipment.
The MedTEST configuration begins with the
Associated Research 620L for Line Leakage
Testing functionality, The OMNIA 8204 for
Hipot and Ground Bond testing functionality
and an Associated Power Technologies AC
power source.
Depending on the requirements of your
application, MedTEST can be configured to
perform patient lead tests on a virtually endless
amount of points.
With its advanced functionality and the ability to improve test efficiency,
MedTEST is the most cost-effective solution on the market!
Line Conditions
Dielectric Withstand Test Mode
Power Switch
Reverse polarity switch for normal condition
Neutral Switch
Neutral switch on/off selection for single fault
Ground Switch
Ground switch on/off selection for class I single fault
Probe Settings
Surface to Surface (PH - PL)
Surface to Line
(PH - L)
Ground to Line (G - L)
Leakage Limit Settings
Touch Current
High/Low
Range:
0.0 µA - 999.9 µA / 1000 µA - 9999 µA /
Limit (RMS)
10.00 mA - 20.00 mA
Resolution: 0.1 µA / 1 µA / 0.01 mA
Touch Current
High/Low
Range:
0.0 µA - 999.9 µA / 1000 µA - 9999 µA /
Limit (Peak)
10.00 mA - 30.00 mA
Resolution: 0.1 µA / 1 µA / 0.01 mA
Measuring Device Module
MD1
UL544NP, UL484 , UL923, UL471, UL867, UL697
MD2UL544P
MD3
IEC 60601-1
MD4UL1563
MD5
IEC60990 Fig4 U2, IEC60950-1, IEC60335-1,
IEC60598-1, IEC60065, IEC61010
MD6
IEC60990 Fig5 U3, IEC60598-1
MD7
IEC60950, IEC61010-1 FigA.2 (2 kohm) for Run function
External MD
Basic measuring element 1 kohm
MD Voltage Limit 70 VDC
DUT Power
AC Voltage
0.0 - 277.0 V
AC Current
40 Arms max continuous
AC Voltage
High/Low Limit
Range:
0.0 - 277.0 V
Resolution: 0.1 V/step
AC Voltage Display Range:
0.0 - 277.0 V
Resolution: 0.1 V/step
Accuracy: ± (1.5% of reading + 2 counts), 30.0 - 277.0 V
Delay time setting Range:
0.5 – 999.9 sec
Resolution: 0.1 sec
Dwell time setting
Range:
0, 0.5 – 999.9 sec (0=Continuous)
Resolution: 0.1 sec
Accuracy: ± (0.1% of reading + 0.05 seconds)
Failure Protection (Start-Up) - Neutral Voltage Check (Neutral-V)
Over current and ground current check (Line - OC)
Output Rating
5 kV @ 50 mAAC
5 kV @ 20 mADC
Voltage Setting
Range: 0-5000 VAC
0-5000 VDC
Resolution: 1 V
Accuracy: ± (2% of setting + 5 V)
HI and LO-Limit
AC Total
AC Real
DC
Range: 0.000-9.999 mA
Resolution: 0.001 mA
Accuracy: ± (2% of setting + 2 counts)
Range: 10.00 - 50.00 mA
Resolution: 0.01 mA
Accuracy: ± (2% of Setting + 2 Counts)
Range: 0.000-9.999 mA
Resolution: 0.001 mA
Accuracy: ± (3% of Setting + 50 µA)
Range: 10.00 - 50.00 mA
Resolution: 0.01 mA
Accuracy: ± (3% of Setting + 50 µA)
Range: 0.00 - 999.9 µA
Resolution: 0.1 µA
Accuracy: ± (2% of Setting + 2 Counts)
Range: 1000 - 20000 µA
Resolution: 1 µA
Accuracy: ± (2% of Setting + 2 Counts)
Ramp HI
>20 mA peak maximum, ON/OFF selectable
Charge LO
Range: 0.000 - 350 µA or Auto Set
DC Output Ripple
≤ 4% Ripple rms at 5 kVDC @ 20 mA, Resistive Load
Discharge Time
< 50 msec for no load,
< 100 msec for capcitor load
(all capacitance values in MAX load spec below)
Maximum
Capacitive Load
1 µF < 1 kV
0.75 µF < 2 kV
0.50 µF < 3 kV
Output Frequency
AC Output Waveform
50/60 Hz ± 0.1% , User Selection, 400/800 Hz Option
Sine Wave, Crest Factor = 1.3 - 1.5
Output Regulation
± (1% of output + 5 V) from no load to full load and over
input voltage range
Dwell Timer
AC 0.4 - 999.9 sec (0 = Continuous)
DC 0.3 - 999.9 sec (0 = Continuous)
Ramp Timer
Ramp-Up AC: 0.1 - 999.9
Ramp-Down AC: 0.0-999.9
Ramp-Up DC: 0.4 - 999.9
Ramp-Down DC: 0.0, 1.0-999.9
Ground Continuity
Current: DC 0.1 A ± 0.01 A, fixed
Max. Ground Resistance: 1 Ω ± 0.1 Ω, fixed
0.08 µF < 4 kV
0.04 µF < 5 kV
Ground Fault Interrupt GFI Trip Current: 5.0 mA max
HV Shut Down Speed: < 1 ms
Continuity Test Mode
Insulation Resistance Test Mode
Output Current
DC 0.1 A ± 0.00001 A
Output Voltage
Range: 30 - 1000 VDC
Resistance Display
Range: 0.00 – 10000.00 Ω
Charging Current
Maximum >20 mA peak
HI and LO-Limit
0.00 – 10000 Ω
Dwell Timer
Range: 0.0, 0.3 - 999.9 sec (0 = Continuous)
Milliohm Offset
Range: 0.00 – 10.00Ω
HI and LO-Limit
Range: 0.05-99.99 MΩ
Resolution: 0.01 MΩ
Range: 100.0 - 999.9 MΩ
Resolution: 0.1 MΩ
Range: 1000 - 50000 MΩ
Resolution: 1 MΩ
Charge-LO
0.000 - 3.500 µA or Auto Set
Ground Bond Test Mode
Ramp Timer
Ramp-Up: 0.1 - 999.9 secs
Ramp-Down: 0.0, 1.0 -999.9 secs
Output Voltage
Range: 3.00 - 8.00 VAC
Dwell Timer
0, 0.5 - 999.9 (0=Continuous)
Output Frequency
50/60 Hz ± 0.1% , User Selection
Delay Timer
0.5 - 999.9 secs
Output Current
Range: 1.00 - 40.00 A
Resolution: 0.01 A
Accuracy: ± (2 % of setting + 2 counts)
Ground Fault Interrupt GFI Trip Current: 5.0 mA max
HV Shut down Speed: < 1 ms
Output Regulation
(1% of output + 0.02A), Within maximum load limits, and
±
over input voltage range.
Maximum Loading
1.00 - 10.00 A,
10.01 - 30.00 A,
30.01 - 40.00 A,
HI and LO-Limit
Range: 0 - 150 for 30.01 - 40.00 A
Range: 0 - 200 for 10.01 - 30.00 A
Range: 0 - 600 for 6.00 - 10.00 A
Range: 0 - 600 for 5.99 - 1.00 A
Resolution: 1 mΩ
Accuracy: 6.00 - 40.00 A, ± (2% of setting + 2 Counts)
Accuracy: 1.00 - 5.99 A, ± (3% of setting + 3 Counts)
Milliohm Offset
Range: 0 - 200 mΩ
0 - 600 mΩ
0 - 200 mΩ
0 - 150 mΩ
AC Power Source
AC Power Source
Up-to 4 kVA compatible power sources available.
Configuration
AC Power Source configuration depends on application.
MedTEST hardware is configured for testing products with one side of the
supply mains at earth potential (Fig 10 UL60601-1).
MedTEST hardware is configured for unbalanced 220 V DUT input power.
Specifications subject to change without notice.
Custom Configurations available. Contact us for details.
General Specifications
Interface
USB/RS-232 Standard, Ethernet, GPIB Optional
Safety
Built-in SmartGFI® circuit
Memory
50 memories, 30 step/memory
For more information on testing to a specfic standard, refer back to the
Common Safety Standard Reference Chart.
System 3400
The system 3400 is a manual electrical safety test system. The
base system includes: HYAMP® III 3140, 40 Amp Ground Bond
tester, interconnect cables and a receptacle box with a choice of
four Hypot® III Dielectric Withstand testers.
Model
Model
Model
Model
34-05
34-65
34-70
34-80
- HYAMP® III 3140 with Hypot® III 3705
- HYAMP® III 3140 with Hypot® III 3765
- HYAMP® III 3140 with Hypot® III 3770
- HYAMP® III 3140 with Hypot® III 3780
System 3100
The system 3100 is a manual electrical safety test system. The
base system includes: HYAMP® III 3130, interconnect cables
and a receptacle box with a choice of four Hypot® III Dielectric
Withstand testers.
Model
Model
Model
Model
31-05
31-65
31-70
31-80
- HYAMP® III 3130 with Hypot® III 3705
- HYAMP® III 3130 with Hypot® III 3765
- HYAMP® III 3130 with Hypot® III 3770
- HYAMP® III 3130 with Hypot® III 3780
System 6000
The system 6000 is a manual electrical safety test system. The
base system includes: HYAMP® III 3160, 60 Amp Ground Bond
tester and interconnect cables with a choice of four Hypot® III
Dielectric Withstand testers.
Model
Model
Model
Model
60-05
60-65
60-70
60-80
- HYAMP® III 3160 with Hypot® III 3705
- HYAMP® III 3160 with Hypot® III 3765
- HYAMP® III 3160 with Hypot® III 3770
- HYAMP® III 3160 with Hypot® III 3780
Safety agency listed.
Electrical Safety Compliance Software
Software that works in tandem with Associated Research, Inc.
Testers to automate a testing workstation. Autoware allows you to capture,
store and analyze test data and results!
Autoware II is a user friendly software solution designed to work with Associated Research’s
OMNIA II, OMNIA, LINECHECK II, and SC6540 instruments. Autoware II software automates
testing procedures and increases production line efficiency by providing the ability to manage
and remotely program instruments. Autoware II allows for the automatic loading of test files
and the ability to drag and drop test files to and from the instrument and PC. It also allows for
the operation and monitoring of the instrument in real time. Autoware II provides the backup
assurance manufacturers need to prove their equipment meets electrical safety requirements
established by safety agencies by allowing for statistical analysis, advanced print out, and
archival of test data.
Autoware II is also compatible with Associated Power Technologies, Inc.’s 6000, 7000 &
300XAC series of AC power sources for RUN and LLT testing.
Features and Benefits
Choice of Ethernet, USB/RS-232 or
GPIB communication interfaces
Pre-configured Instrument
Verification Files
Graphical Drag and Drop User
Interface
Customizable Menu Display
Batch Processing
Automated Instrument Identification
and setup
Control Panel Screen
Results Screen
Two Versions Available (Full or
Data Logging)
Intelligent Error Handling
System Settings
Autoware S9870 is used in tandem with Associated Research Inc.’s QUADCHEK
II, HypotMAX, HypotULTRA III, LINECHEK and RUNCHEK series of instruments.
38814 - H
igh Voltage Pistol
Probe with Switch
This high voltage test gun has a dual action
test trigger that controls the retractable probe
tip and activates the high voltage output of
the instrument. The unique ergonomic shape
makes this probe easy and comfortable to use.
35822 - Foot Switch
Ideal for use in applications where an
operator needs to perform a test while
using test probes.
39077 - Test Cabinet
This test cabinet is used with any
Associated Research rack mountable
tester.
38081 - High Voltage Probe
38082 - Return Probe
38539 - 2 Wire 40 Amp Ground Bond Probe
38538 - 4 Wire 40 Amp Ground Bond Probe
Retractable high voltage and return probes
for safe high voltage testing up to 5 kV. The
probe tips allow for safe and easy testing
by extending and retracting with the
press of a button.
These fixed tip probes allow for easy
contact with the grounding points of the
DUT. Separate test buttons to use in various
positions. Three different tip styles included.
36541, 38482, 36544, 38777,
38578 - Adaptor Box
CBLSR-05M - M
agnetic Hipot Return
Adapter boxes allow for safe and easy
testing of line cord terminated products.
All necessary connections are made
within the adapter box helping to avoid
connection errors, with configurations
that satisfy a variety of needs.
39067 - Test Enclosure
This DUT Enclosure is designed with a safety
interlock that will eliminate the potential for
electrical shock when used correctly. The
interlock will disable the high voltage if the
enclosure door is open. The DUT Enclosure
is designed to be used with all interlocked
Associated Research Hipot testers.
Cable
CBLHR-05M
- Magnetic Ground Bond
Return Cable
We have available test leads that are
terminated in a magnetic tip that allow for
easy application of the test lead to the DUT.
These test leads are available in lengths
from 5 to15 feet.
39001 - Small Shipping Case
39085 - Large Shipping Case
We offer a large shipping case and a
small shipping case each with custom
inserts to protect your Associated
Research tester. They are used for the
shipping and transportation of an
Associated Research tester.
A Powerful Online Resource!
-
Convenient Online Ordering
Request a Video Demonstration
Schedule a Calibration
See New Product Introductions
View Product Specials and Promotions
Join our Informative Seminars & Webinars!
Sign-up for our Feature-packed Newsletter
Take a Virtual Tour of our Instruments
View Instructional Videos
We have local sales offices throughout
the world to serve you more efficiently.
To find your nearest representative visit the
“Local Sales Offices” section of our web site at
www.asresearch.com or call us toll-free at 1-800-858-8378
DISTRIBUIDOR EN ESPAÑA Y PORTUGAL:
Pallars 65-71
08018 Barcelona, Spain
T. (+34) 93 320 80 55
F. (+34) 93 320 80 56
www.wavecontrol.com
13860 West Laurel Drive,
Lake Forest, IL U.S.A. 60045
Tel: +1-847-367-4077 Fax: +1-847-367-4080
E-mail: info@asresearch.com
For more information visit us at www.asresearch.com
or call us toll-free at 1-800-858-TEST(8378)
VER 0112