UltraRAE
Specific Compound Monitor
UltraRAE is a quick, spot-check monitor, typically used by occupational health, safety
and environmental personnel. The combination filter tube and PID removes the disadvantages associated with colorimetric tubes
and gas chromatographs. Applications
include tank entry pre-screening during
refinery and plant maintenance, hazardous
material response, marine spill response and
refinery down-stream monitoring.
Interchangeable RAE Sep™
Filter Tubes
• Benzene
• Methylene Chloride (Halocarbon)
• Butadiene (Polymer manufacture)
Features
• Low cost of ownership
– Inexpensive analysis due to low cost
RAE-Sep™ tubes vs. gas detection
tubes
– Cheaper than portable GC and
no annual factory calibration
• Large keys operable with 3 layers
of gloves
• Large easy to read back light display
– Alarm activated
– Darkness or manually activated
• 3,000 point data logging storage
capacity for data download to PC
• Optional vibration alarm
• Snap-in rechargeable or alkaline battery
pack, 10 hours of operation with Smart
battery charging.
• The only compound specific monitor with
combination of interchangeable separation tubes and PID
Applications
• 3-10 times faster analysis vs. conventional
tube or portable gas chromatograph
methods
– Results obtained in seconds
(<75 seconds) rather than minutes
– Detects benzene down to 0.1 ppm
in 60 seconds
• Benzene
– Petrochemical
– Transportation industry – tankers,
railways, aircraft and barges
– Pipelines
– Military
– Rubber manufacturers
• 3,000 point datalogging storage capacity
for data download to PC compatible with
Windows™ 95, 98, 2000 and NT
• Methylene Chloride
– Paint strippers in industries from
airlines to furniture refinishers to
ship building
– Solvents for cleaning chemical tanks
– Solvents in the manufacture of
plastics, flexible foam, sealants,
and photographic film stock
– Adhesives in the assembly of
plastic parts
• Excellent range and linearity
– Good down to 0.1 ppm and up to over
200 ppm benzene (no more GC
column overload problems)
– More accurate than gas detection
tubes
• Internal pump
– Allows you to sample deep (up to 30ft
or 10m) into a confined space
– Prevents liquid flow into the monitor
– Warns if sample tubing is blocked
or crimped
1339 Moffett Park Drive • Sunnyvale, California 94089 USA
Tel: 408.752.0723 • Fax: 408.752.0724
Email: RaeSales@raesystems.com • www.raesystems.com
AL01 Rev. 8/99
Partial selection
• Butadiene
– Polymers industries
• Manufacturers of plastics, resin,
and rubber
Ordering Information
Model PGM 7200 Monitor
VOC
Benzene
Halocarbons
Butadiene
Range
0-2000 ppm
0-200 ppm
0-200 ppm
0-200 ppm
• Rechargeable NiMH battery pack
Resolution
0.1 ppm 0-99.9 ppm
1 ppm 100-2000 ppm
0.1 ppm
0.2 ppm
0.1 ppm
• Alkaline battery adapter (accepts
4 AA size alkaline batteries)
Response
Time
30 seconds
60 seconds
Compound
Dependent
75 seconds
Calibration
Gas
100 ppm
Isobutylene
5 ppm
Benzene
Compound
Dependent
5 ppm
Butadiene
UV Lamp
9.8, 10.6 or 11.7 eV
9.8 eV
11.7 eV
9.8 eV
300 ppm
Toluene*
300 ppm
Acetone*
150 ppm
Hexane*
• AC/DC charging adapter
• Inlet probe and water trap filter
• Operation and maintenance manual
• Soft carry case
Model PGM 7200 Kit
Interference None
Rejection*
*Refer to RAE Systems’ Technical Notes no: 127, 133 and 147 for more details.
Intrinsic Safety
UL, cUL Class 1, Division 1, Group A, B, C, D
EEx ia IIL C T4
Size
7.75” (19.7 cm) L x 2.75” (7 cm) W x
1.5” (3.8 cm) H
Weight
16 oz (455 gm) with battery pack
Detector
Photo-ionization sensor with vacuum UV lamp
Battery
Rechargeable, snap-in, field replaceable,
NiCd battery pack
Battery Charging
10 hours charge through built-in charger
Operating Hours
10 hours
Display
Large LCD, manual and alarm activated
back light
Repeatability
± 10% of reading or ± 0.1 ppm
RAE-Sep Tube
Field replaceable, bar coded, sealed at
both ends for storage
Keypads
1 operation key and 2 program keys
Alarm Setting
Two adjustable alarm limit levels
(Factory default 1 and 5 ppm)
Audible Alarm
90 dB buzzer
Visual Alarm
Flashing red LED
External Alarm
Optional plug in pen size vibration alarm
Datalogging
3000 points; includes sample number,
gas reading, time/date, specific vapor.
Header information includes monitor serial
number, user ID and site number.
Communication
Download data to PC and up load instrument
setup from PC through RS-232 link to
serial port on PC
Calibration
Two point field calibration of zero and span
Sampling Pump
Internal. Flow rate 400 cc/min. with VOC tube
Temperature
0° to 40°C (32° to 113°F)
Humidity
0% to 95% relative humidity (non-condensing)
™
Items included with Monitor plus:
• Calibration gas
• Calibration regulator with
flow controller
• Porous metal filter and O-ring kit
• Tool kit
• Hard transport case
Accessories
• Lamp cleaning kit
• Automotive charger/adapter
• Remote access probe
• External vibration alarm
UltraRAE
Specifications
4
ISO 9001
CERTIFIED
D I S T R I B U T E D B Y:
Download PDF
Similar pages