DW701M manual - produktinfo.conrad

SIZE:72*97MM 正面
SAFETY PRECAUTIONS
Babies, young children or individuals who cannot express
their consent are not suitable to take blood pressure
measurement.
Individuals with serious arteriosclerosis are not suitable to
take blood pressure measurement.
Deluxe Automatic Wrist
Blood Pressure Monitor
Blood pressure readings may change in case of pregnancy.
Pregnant women can consult their doctor before taking
measurement.
Instruction Manual
Self-measurement is not medical treatment. If there are
unusual values, please consult your doctor.
Ref : INOTP02
Handling batteries properly:
- As soon as old batteries run out, replace with new
batteries.
- Do not use old and new batteries together.
- Align the polarities of batteries correctly.
- When the unit will not be used for more than 3 months,
remove the batteries. Otherwise, batteries may leak and
cause damage to the unit.
QUICK START GUIDE
Avoid smoking, eating or exercising for 30 minutes before
taking measurement.
Date/ Time
User 1~4
Memory Average
Low Battery
Indicator
Pulse Symbol
2. Roll up the sleeve to bare your
wrist if necessary. Take caution
not to constrict the blood flow.
3. Pull on the end of the cuff until
it wraps securely around you wrist.
4. Place your elbow on a table and palm facing upward, so
that the cuff will be at the same level as your heart. You
can use the storage case to support your arm during the
measurement.
1. Press and hold the” /AVG“button for a few seconds until
the Year display flashes.
Start Pulse
Memory
One Touch Operation
Irregular Heartbeat Detector
Blood Pressure Mode and Clock Mode
4 User / 396 Total Memory
4 Alarm Clock Settings
5 Mode Averaging Function – Hour, Day, Week, Month,
and Overall
• WHO Blood Pressure Indicator
• Auto-off Power Saving Feature
3
TAKING MEASUREMENT
Note:
- You will not be able to set alarm clock if Clock Function is
set to “OFF”.
- You can set a maximum of 4 alarms.
Note:
- Avoid smoking, eating or exercising for 30 minutes before
taking measurement.
- Do not move or talk during measurement.
- Measurement can be taken on either arm.
“button to select the current year.
3.1 Sitting Correctly
3. Press the” /AVG“button to confirm Year Setting.
2. Insert 2 AAA batteries with the
polarities in correct direction.
Set
Power
Features
2.3 Alarm Setting
Note:
- Set the correct date and time before taking a measurement.
- Date and time should be reset after replacing new batteries.
2. Press the” M “or”
Diastolic
Memory No.
(1~99 readings)
2
2.2 Date & Time Setting
1. Squeeze slightly and pull to
remove the battery cover.
Memory Recall
(Up)
Page Select
(Down)
Systolic
•
•
•
•
•
•
5. Press the “ON/OFF” button to
turn on the unit. Then press the
“START” button. The cuff will
automatically inflate and the
measurement will start.
1
2.1 Battery Installation
Monitor
1. Sit upright in a chair with your
both feet on the floor.
6. When the measurement is complete, the cuff will
automatically deflate and your systolic and diastolic
pressure values and pulse rate will display.
PREPARATION
OVERVIEW
4. Follow the same steps to set the Month, Date, Hour and
Minute.
5. You can choose Clock Function “ON” or “OFF” while the
monitor is not in use. Press“M “or”
“button to select.
Press the” /AVG “button to confirm.
3. Replace the battery cover. Press
it lightly until the lock securely
clicks into place.
6. To disable Alarm, Press the”M “or”
“button to select
“OFF” and press the“ /AVG“button to confirm.
7. To set 1st Alarm, Press the”M “or”
“button to select
“ON”. Press the” /AVG“button to confirm.
8. When Hour digit flashes, press the” M “or”
“button to
select Hour. Press the” /AVG“button to confirm.
1. Relax
2. Sit upright in a chair with your both feet on the floor.
3. Roll up the sleeve to bare your wrist if necessary. Take
caution not to constrict the blood flow.
4. Pull on the end of the cuff until it wraps securely around
you wrist. Do not over tighten the cuff. Allow 1-2 cm or ½
inch between the cuff and your hand.
9. When Minute digit flashes, press the” M “or”
“button
to select Minute. Press the” /AVG“button to confirm.
Note: Replace batteries when the battery indicator “
shows low voltage.
”
Noted: If you select Clock Function “ON”, you will continue at
the Alarm Setting section. Otherwise, setting will end.
10. Follow the same steps to set the 2nd, 3rd, and 4th Alarm.
Note:
- Press the “ON/OFF” button anytime during clock setting
alarm setting to start over if an error occurs.
- Each alarm clock lasts 45 seconds, or you can turn it off
by pressing the “ON/OFF” button.
4
5
5. Place your elbow on a table and palm
facing upward, so that the cuff will be
at the same level as your heart. You
can use the storage case to support
your arm during the measurement.
6
7
SIZE:72*97MM 反面
Irregular Heartbeat
3.2 User Setting
Regular but Above Normal Heartbeat
This unit allows a maximum of 4 users to store measurement
records in different User Pages.
This unit has a unique feature that alerts user of irregular
heartbeat detection during measurement.
1. Press the “ON/OFF” button to turn on the unit.
2. Press”
“button to switch between User Page 1, 2, 3
and 4.
Note:
An irregular heartbeat is defined as
a heartbeat rhythm that has a
variation of more than 25% from
the average rhythm detected
during measurement.
1. Press the “ON/OFF” button to turn on the unit.
2. Press the “START” button. The cuff will automatically inflate
and the measurement will start.
3. When the measurement is complete, the cuff will
automatically deflate and your systolic and diastolic
pressure values and pulse rate will display.
4. Remove the wrist cuff and take note of your measurement
results.
Note: You can press the “ON/OFF ” button to stop
measurement at any time.
3.4 Heartbeat Readings
8
If such irregular rhythm occurs twice
or more during the same
measurement, the display screen will
show the pulse value and “IH” symbol
alternately.Example: (90 / IH)
Note: Please consult your doctor if irregular heartbeat
occurs often.
Grade 3 hypertension(severe) (Red)
95
90
85
Grade 1 hypertension(mild)
(Purple)
High- normal(Green)
Normal
(Sky Blue)
Optimal(Blue)
120 130
140 150
160
Systolic Pressure
170 180
mmHg
10
TROUBLESHOOTING
STORAGE & MAINTENANCE
Problem
No display when
you press the
Power/Start button
Cause
Solution
Have the batteries
run out?
Replace new
batteries.
Are the battery
polarities wrong?
Re-align batteries
in correct polarities.
“Err P” displayed
Fail to inflate
Check if the cuff is
properly connected.
One Hour
1 d
One Day
7 d
One Week
“Err 1” displayed
Deflate too rapidly
Return for servicing.
30 d
One Month
“Err 2” displayed
Movement during
measurement
Do not move during
measurement
ALL
All Readings
Signal interference
Remove interfering
source e.g. mobile
phones, magnets.
4.3 Delete Past Readings
1. Press the “
” button to select the User Page you want
to delete.
2. Press and hold the” M ”&”
“button simultaneously for
5 seconds. Symbol of “ALL EE” will be displayed and all past
readings stored in that User Page will be deleted.
100
80
1 h
User can delete all the past readings stored in each User
Page.
105
“Err 3” displayed
Incorrect results
Measure again.
Note: If your problem cannot be solved by the above,
consult your store of purchase. Do not disassemble
the unit.
Keep away the unit from direct sunlight, extreme
temperatures, humidity or moisture.
Use a dry, soft cloth to clean the unit, or if desired, use a
cloth lightly dampened with water.
Do not use alcohol, benzene, thinner or other volatile
liquids to clean the unit.
1. Press the“ M “button to enter memory function.
2. If the User Page is not what you want to view, press the
”
“ button to select.
3. Press the” M “button to display the next sets of values.
Past readings are displayed from the most recent to the
oldest.
4. This can stored up to 99 sets of past readings for each user
(4 x 99).
11
Storage Environment
-20°C to 60°C, 10% to 95% RH
Operation Environment 10°C to 40°C, 30% to 85% RH
3.15” (L) x 2.36” (W) x 3.15” (H) or
External Dimensions
80 x 60 x 80 mm
125g (batteries excluded)
Net Weight
Cuff Dimensions
Wrist circumference of
135~220mm
Package Content
Instruction Manual, Battery Set,
Warranty Card
Do not wash or expose the arm cuff to liquid.
Remove batteries from the unit when it will not be used
for more than 3 months.
SPECIFICATIONS
Model No.
Measurement Method
Measurement Range
Accuracy
Inflation
Pressure Detection
Power Supply
Auto Power Off
13
After measurement, the last set of measured values will be
saved automatically after measurement when the unit is
turned off by pressing “ON/OFF” button or by automatic
power-off after 3 minutes.
4.1 View Past Readings
Grade 2 hypertension(moderate)
(Yellow)
9
Average Values of
12
mmHg
Regular but Below Normal Heartbeat
4.2 View Average Readings
Symbol on Display
This unit features a 6-color LED light indicator to display
blood pressure level in accordance with the WHO (World
Health Organization) classification.
110
Pulse
1. Press the “ /AVG” button to enter Average function.
2. If the User Page is not what you want to view, press the
“
” button to select.
3. Press the “ /AVG” button to view different Average Values.
5 different Average Values are available:
MEMORY FUNCTIONS
3.5 WHO Light Indicator
If the monitor detects a pulse rate
below 60 beats per minute during
measurement, the monitor will
display the pulse value and “LO”
symbol alternately.
Example: (58 / LO)
Besides Systolic and Diastolic
values, Pulse will also be
displayed after a measurement.
WHO Light Indicator
Example: If your measurement reading is 122/75, the LED light
will display the color in blue.
Diastolic Pressure
3.3 Taking Measurement
WHO Classification
Color
Optimal --------------------Blue
Normal ---------------------Sky Blue
High-normal --------------Green
Mild hypertension -------Purple
Moderate hypertension ---Yellow
Severe hypertension ----Red
If the monitor detects a pulse rate
above 100 beats per minute during
measurement, the monitor will
display the pulse value and “HI”
symbol alternately.
Example: (102 / HI)
DW701M
Oscillometric
Pressure: 20 – 280 mmHg
Pulse Rate: 40 – 195 beats/min
Pressure: +/- 3 mmHg
Pulse Rate: +/- 5%
Pump driven
Semi-conductor
2x AAA batteries
3 minutes after last operation
14
Find our products on
www.inocare.fr
Manufactured in accordance with International Standards
0123
Importer:
Manufacturer:
INOVAXION
10 rue Jean Cottin 75018 - Paris - France
Truly Instrument Ltd - Hong Kong
EC-Representative:Hong Kong Truly Semiconductor Ltd.
1-2 Alpha Buildings, The Buntings Kestrel Drive,
Stowmarket Suffolk, IP14 5GZ, UK
Download PDF