Casella 63x Series Sound Level Meter User manual


Add to my manuals
2 Pages

advertisement

Casella 63x Series Sound Level Meter User manual | Manualzz

1 2

Note: A Record Notes screen will appear prior to starting a measurement if Audio

Notes have been activated. Refer to User Manual for more details.

Run Screens

The Play symbol is displayed at the top left-hand corner of the screen, and the

Pause symbol is displayed at the bottom left-hand corner of the screen. The lefthand soft key ( ) is the Pause/Run control and toggles between these two modes when operated.

The screen displays PAUSED!

when measurements are paused. The incremental run number and its duration are displayed at the top right-hand corner of the screen. If the Run Duration Timer is set, an icon that shows a stopwatch is displayed next to the countdown timer. To enable or disabled the Run Duration Timer, use the Settings screens (refer to Figure 6). If the Run Duration Timer is not set, the digital timer clock counts up until you stop the measurement run manually (by pressing the Run/Stop

3 4

Figure 9. Memory screens

Taking Measurements

(Refer to Figure 10)

The measurement options and screen details will vary, depending on the instrument type you are using.

The instrument must be in the Measurement Stop mode (with red bars at the top and bottom of the screen) before you can start a measurement.

Press the Run/Stop key ( ) to start making measurements. The screen shows green bars at the top and bottom.

Figure 11. Graph and Main screens

For instruments that include this option, the broadband values can be viewed as a bargraph or as a graphical time history on the Graph screen. Press the View key ( ) to toggle between the Graph and the Main screens.

Certain instrument types provide two additional Octave (or 1/3 Octave) screens (refer to Figure 12). Use the View key ( ) to toggle the screens. Use the left/right navigation

Each bar in the graph represents an octave band measured in Hz or kHz. To view a table of the octave values, press the View key ( ).

In 1/3 Octave mode, use the up/down navigation keys to change the displayed frequencies on the table.

Some instruments in the CEL-63x range can display all the screens shown in

Figure 11 and Figure 12.

The Marker screen allows you to tag noise data and record audio, for example noise

Figure 10. Example Stop and Run screens

Figure 12. 1/3 Octave graph and list views from a road vehicle, with a specific identity using one of four markers.

When the measurement run has finished, press the Run/Stop key ( ). A screen is displayed asking you to confirm the action - press Yes to end the run.

Storage

Switch the CEL-63x instrument off when you are not using it to record measurements, and remove the batteries if the instrument will not be used for extended periods of time.

Always protect the instrument from physical damage and from water.

To protect the microphone, you should always keep the windshield fitted.

More information

For more information about any of the topics summarised in this Quick Start guide and additional functions, refer to the User Manual.

Instrument types

Table 1 below lists the instrument types and shows the capabilities of each instrument in the range. Please remember that the features that are available to you and the screens that you can see depend on the type of instrument you are using.

Application

Model number

Cumulative Results

Period Results

Profile Results

Statistical Parameters (Ln%)

Audio Notes

Marker Events

Level Events

External Events

SLM Mode

ONLINE Mode

NNR Mode

Timers

Table 1. CEL-63x instrument types, applications and capabilities

CASELLA MEASUREMENT

Regent House, Wolseley Road, Kempston, Bedford, MK42 7JY, U.K.

Phone: +44 (0) 1234 844 100

Fax: +44 (0) 1234 841 490

E-mail for technical support: [email protected]

For online Technical Support: http://supportdesk.casellacel.com

Document number FG40-04 © 2011 Casella Measurement

www.casellameasurement.com

1

2

Field Guide

CEL-63x Series Sound Level Meters

5

6

3

4

A

B

7

1. Windshield (covering the removable microphone)

2. Removable amplifier 5. Soft keys and B

3. On/Off key

4. Display

6. Navigation keys

7. Run/Stop key

Figure 1. The CEL-63x series

Introduction

This Quick Start guide helps you to familiarise yourself with the CEL-63x instrument quickly, so that you can begin to use it without delay.

The guide assumes this is the first time you have used the instrument. It therefore includes instructions to install the batteries, switch ON and set up the instrument to be ready for use, calibrate it, take measurements, and store it safely.

We want you to get the best possible performance from the CEL-63x instrument, and we recommend you refer to the Operator’s Manual for detailed information.

Think Environment. Think Casella.

Fit the batteries

The CEL-63x instrument requires three AA batteries.

Remove the battery compartment cover and fit the batteries using the correct polarities. Refit the battery compartment cover.

Note: The batteries can be zinc carbon, alkaline or rechargeable. DO NOT mix battery types.

Switch the instrument ON

Press the On/Off key. Confirm the battery condition symbol shows there is a good level of charge in the batteries.

The initialisation screen is displayed for approximately 10 seconds, and then the measurement screen is displayed in the STOP mode (with red bars at the top and the bottom of the screen). Press the Menu key to access the settings, memory and instrument status. The screens displayed depend on the model of instrument in use.

Settings screen

MENU SETTINGS LANGUAGE

4

1 2 3

A

5

B A

C

B A

C

C C

1. Settings 2. Memory results 3. Instrument status

A. Exit / Back B. Select / Edit

C. Navigation keys

Figure 2. Settings screen

C

B

Set language

(Refer to Figure 2)

Select Settings B

Language B language. Press Select ( ). Press Exit A

C

C

Set frequency modes

(Refer to Figure 2)

Some instruments in the CEL-63x range allow you to select either Octave or

1/3 Octave with which to take measurements. Select 1 Settings B

C Octave 5 B

C Select B Exit A

Set the time and date

(Refer to Figure 3)

Select Settings B C Set

Clock B C

C Save B Exit

1

2

1

A B A

Figure 4. Calibration

Select the Measurement Stop screen. Fit the acoustic calibrator fully over the removable microphone. Press the calibrator’s On/Off key . The CEL-63x instrument detects the signal and activates the calibration screen (with yellow bars at displays the word PASSED when finished. Press Exit ( ) to return to the Stop screen. Press and hold the calibrator’s On/Off key to switch the calibrator off.

Remove the acoustic calibrator.

Setup

(Refer to Figure 5)

Select Settings B C Setup measurement view for your local legislation. Refer to the User Manual for detailed information about the measurement views.

Note : This instrument simultaneously records all noise functions for every measurement it completes. The Measurement View screens simply define which functions are displayed on the measurement screens.

Measurement control

(Refer to Figure 6)

Select Settings and use the navigation keys to select the Measurement control

B Measure -

Start/Stop symbol . Press Edit B Mode (Key

Press / Fixed Duration / Timers). Press Save ( ). Use the navigation keys to select additional Fixed duration or Timers options. Press Edit ( ), and use the navigation keys to set the run duration. Press Save ( ) to continue (if required).

1

2

Figure 6. Run duration

Measurement data sets

(Refer to Figure 7)

Cumulative measurements made by the instrument apply for the entire measurement duration. They do not show how the sound level changed during the period of measurement. The instrument can also make Periodic measurements, which store measurements acquired during a repeating interval of time. When one interval ends, the next one begins until the periodic measurements stop.

The instrument’s Profile recording function provides an additional time-history recording channel, which may operate in parallel with either Cumulative or Periodic data recording. The Profile time history channel can provide faster recording resolution but for a reduced set of noise parameters.

Select Settings and use the navigation keys to select the Measurement control

B Data Sets symbol . Set the Periodic and Profile measurement intervals and press Exit .

Figure 3. Setting the time and date

Calibration

(refer to Figure 4)

Calibration is auto-set to the reference level as set within the Settings menu. Refer to the manual to set the calibration reference level.

1

2

1

2

1

Figure 8. Memory results

Memory results

(Refer to Figure 8)

Press the Menu key to access the settings, memory and instrument status. Use the navigation keys to select the Memory Results symbol and press Select B continue. Use the navigation keys to select a result in the list, and press Select ( ) to select the measurement.

Use the navigation keys to select View Results (see Figure 9) and press Select B to display the selected memory results. For additional data press View ( ). The screen displays detailed, read-only data for the selected measurement. Press View

Memory Results. Use the navigation keys to move the vertical line across the graph.

Each graph bar represents an octave level measured in Hz or kHz. Press View ( ) again to view tabular octave results.

Use the navigation keys to toggle between functions such as L and L . Press View

Press Exit ( ) to leave the Memory Results , or press View B Memory

Results screens again. Note that the display screens will vary according to the model of CEL-63x instrument you are using.

Delete Memory Results

Refer to the User Manual for details and instructions to delete the results stored in the instrument’s memory.

Figure 5. Settings options Figure 7. Measurement data sets

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Key Features

  • Measures sound levels in accordance with international standards
  • Easy-to-use interface with large backlit LCD display
  • Measures Leq, Lmax, Lmin, and other parameters
  • Stores and recalls data for later analysis
  • Generates reports in a variety of formats
  • Rugged and durable construction for use in harsh environments

Questions & Answers

2024-05-23

F F G

How do I pause and resume measurements?
Use the Pause/Run control feature to toggle between 'Pause' and 'Run' modes.
How do I set a time limit for measurements?
Enable the Run Duration Timer in the Settings screens to automatically stop measurements after a specified duration.
How do I switch between frequency bands?
Use the left/right navigation keys to navigate through frequency bands during measurements.
How do I create notes during measurements?
Use the Marker screen to record audio and create markers for noise data during measurements.
How do I start a measurement?
Press the Run/Stop key to initiate a measurement.
What models of the device are available?
The CEL-600 series includes the CEL-631 Environmental, CEL-630 Occupational, CEL-632 with logging, and CEL-633 with logging. Each model caters to specific applications.
How do I calibrate the device?
Fit the acoustic calibrator over the microphone, activate the calibration screen, and follow the instructions.

Related manuals

Frequently Answers and Questions

What are the applications of the CEL 63x Series Sound Level Meter?
The CEL 63x Series Sound Level Meter is ideal for a wide range of applications, including environmental monitoring, occupational health and safety, and product noise testing.
What are the key features of the CEL 63x Series Sound Level Meter?
The CEL 63x Series Sound Level Meter offers a variety of features that make it easy to use and effective for a variety of tasks. These features include: * A large, backlit LCD display that is easy to read in all lighting conditions. * A simple, intuitive user interface that makes it easy to set up and use the meter. * A wide range of measurement parameters, including Leq, Lmax, and Lmin. * The ability to store and recall data for later analysis. * The ability to generate reports in a variety of formats.
How do I use the CEL 63x Series Sound Level Meter?
The CEL 63x Series Sound Level Meter is easy to use. Simply follow these steps: 1. Turn on the meter by pressing the power button. 2. Select the desired measurement parameters by pressing the mode button. 3. Position the meter in the desired location. 4. Press the start button to begin measuring. 5. Press the stop button to stop measuring. 6. Press the recall button to view the measurement results. 7. Press the report button to generate a report.

advertisement