1 2
Note: A Record Notes screen will appear prior to starting a measurement if Audio
Notes have been activated. Refer to User Manual for more details.
Run Screens
The Play symbol is displayed at the top left-hand corner of the screen, and the
Pause symbol is displayed at the bottom left-hand corner of the screen. The lefthand soft key ( ) is the Pause/Run control and toggles between these two modes when operated.
The screen displays PAUSED!
when measurements are paused. The incremental run number and its duration are displayed at the top right-hand corner of the screen. If the Run Duration Timer is set, an icon that shows a stopwatch is displayed next to the countdown timer. To enable or disabled the Run Duration Timer, use the Settings screens (refer to Figure 6). If the Run Duration Timer is not set, the digital timer clock counts up until you stop the measurement run manually (by pressing the Run/Stop
3 4
Figure 9. Memory screens
Taking Measurements
(Refer to Figure 10)
The measurement options and screen details will vary, depending on the instrument type you are using.
The instrument must be in the Measurement Stop mode (with red bars at the top and bottom of the screen) before you can start a measurement.
Press the Run/Stop key ( ) to start making measurements. The screen shows green bars at the top and bottom.
Figure 11. Graph and Main screens
For instruments that include this option, the broadband values can be viewed as a bargraph or as a graphical time history on the Graph screen. Press the View key ( ) to toggle between the Graph and the Main screens.
Certain instrument types provide two additional Octave (or 1/3 Octave) screens (refer to Figure 12). Use the View key ( ) to toggle the screens. Use the left/right navigation
Each bar in the graph represents an octave band measured in Hz or kHz. To view a table of the octave values, press the View key ( ).
In 1/3 Octave mode, use the up/down navigation keys to change the displayed frequencies on the table.
Some instruments in the CEL-63x range can display all the screens shown in
Figure 11 and Figure 12.
The Marker screen allows you to tag noise data and record audio, for example noise
Figure 10. Example Stop and Run screens
Figure 12. 1/3 Octave graph and list views from a road vehicle, with a specific identity using one of four markers.
When the measurement run has finished, press the Run/Stop key ( ). A screen is displayed asking you to confirm the action - press Yes to end the run.
Storage
Switch the CEL-63x instrument off when you are not using it to record measurements, and remove the batteries if the instrument will not be used for extended periods of time.
Always protect the instrument from physical damage and from water.
To protect the microphone, you should always keep the windshield fitted.
More information
For more information about any of the topics summarised in this Quick Start guide and additional functions, refer to the User Manual.
Instrument types
Table 1 below lists the instrument types and shows the capabilities of each instrument in the range. Please remember that the features that are available to you and the screens that you can see depend on the type of instrument you are using.
Application
Model number
Cumulative Results
Period Results
Profile Results
Statistical Parameters (Ln%)
Audio Notes
Marker Events
Level Events
External Events
SLM Mode
ONLINE Mode
NNR Mode
Timers
Table 1. CEL-63x instrument types, applications and capabilities
CASELLA MEASUREMENT
Regent House, Wolseley Road, Kempston, Bedford, MK42 7JY, U.K.
Phone: +44 (0) 1234 844 100
Fax: +44 (0) 1234 841 490
E-mail for technical support: [email protected]
For online Technical Support: http://supportdesk.casellacel.com
Document number FG40-04 © 2011 Casella Measurement
www.casellameasurement.com
1
2
Field Guide
CEL-63x Series Sound Level Meters
5
6
3
4
A
B
7
1. Windshield (covering the removable microphone)
2. Removable amplifier 5. Soft keys and B
3. On/Off key
4. Display
6. Navigation keys
7. Run/Stop key
Figure 1. The CEL-63x series
Introduction
This Quick Start guide helps you to familiarise yourself with the CEL-63x instrument quickly, so that you can begin to use it without delay.
The guide assumes this is the first time you have used the instrument. It therefore includes instructions to install the batteries, switch ON and set up the instrument to be ready for use, calibrate it, take measurements, and store it safely.
We want you to get the best possible performance from the CEL-63x instrument, and we recommend you refer to the Operator’s Manual for detailed information.
Think Environment. Think Casella.
Fit the batteries
The CEL-63x instrument requires three AA batteries.
Remove the battery compartment cover and fit the batteries using the correct polarities. Refit the battery compartment cover.
Note: The batteries can be zinc carbon, alkaline or rechargeable. DO NOT mix battery types.
Switch the instrument ON
Press the On/Off key. Confirm the battery condition symbol shows there is a good level of charge in the batteries.
The initialisation screen is displayed for approximately 10 seconds, and then the measurement screen is displayed in the STOP mode (with red bars at the top and the bottom of the screen). Press the Menu key to access the settings, memory and instrument status. The screens displayed depend on the model of instrument in use.
Settings screen
MENU SETTINGS LANGUAGE
4
1 2 3
A
5
B A
C
B A
C
C C
1. Settings 2. Memory results 3. Instrument status
A. Exit / Back B. Select / Edit
C. Navigation keys
Figure 2. Settings screen
C
B
Set language
(Refer to Figure 2)
Select Settings B
Language B language. Press Select ( ). Press Exit A
C
C
Set frequency modes
(Refer to Figure 2)
Some instruments in the CEL-63x range allow you to select either Octave or
1/3 Octave with which to take measurements. Select 1 Settings B
C Octave 5 B
C Select B Exit A
Set the time and date
(Refer to Figure 3)
Select Settings B C Set
Clock B C
C Save B Exit
1
2
1
A B A
Figure 4. Calibration
Select the Measurement Stop screen. Fit the acoustic calibrator fully over the removable microphone. Press the calibrator’s On/Off key . The CEL-63x instrument detects the signal and activates the calibration screen (with yellow bars at displays the word PASSED when finished. Press Exit ( ) to return to the Stop screen. Press and hold the calibrator’s On/Off key to switch the calibrator off.
Remove the acoustic calibrator.
Setup
(Refer to Figure 5)
Select Settings B C Setup measurement view for your local legislation. Refer to the User Manual for detailed information about the measurement views.
Note : This instrument simultaneously records all noise functions for every measurement it completes. The Measurement View screens simply define which functions are displayed on the measurement screens.
Measurement control
(Refer to Figure 6)
Select Settings and use the navigation keys to select the Measurement control
B Measure -
Start/Stop symbol . Press Edit B Mode (Key
Press / Fixed Duration / Timers). Press Save ( ). Use the navigation keys to select additional Fixed duration or Timers options. Press Edit ( ), and use the navigation keys to set the run duration. Press Save ( ) to continue (if required).
1
2
Figure 6. Run duration
Measurement data sets
(Refer to Figure 7)
Cumulative measurements made by the instrument apply for the entire measurement duration. They do not show how the sound level changed during the period of measurement. The instrument can also make Periodic measurements, which store measurements acquired during a repeating interval of time. When one interval ends, the next one begins until the periodic measurements stop.
The instrument’s Profile recording function provides an additional time-history recording channel, which may operate in parallel with either Cumulative or Periodic data recording. The Profile time history channel can provide faster recording resolution but for a reduced set of noise parameters.
Select Settings and use the navigation keys to select the Measurement control
B Data Sets symbol . Set the Periodic and Profile measurement intervals and press Exit .
Figure 3. Setting the time and date
Calibration
(refer to Figure 4)
Calibration is auto-set to the reference level as set within the Settings menu. Refer to the manual to set the calibration reference level.
1
2
1
2
1
Figure 8. Memory results
Memory results
(Refer to Figure 8)
Press the Menu key to access the settings, memory and instrument status. Use the navigation keys to select the Memory Results symbol and press Select B continue. Use the navigation keys to select a result in the list, and press Select ( ) to select the measurement.
Use the navigation keys to select View Results (see Figure 9) and press Select B to display the selected memory results. For additional data press View ( ). The screen displays detailed, read-only data for the selected measurement. Press View
Memory Results. Use the navigation keys to move the vertical line across the graph.
Each graph bar represents an octave level measured in Hz or kHz. Press View ( ) again to view tabular octave results.
Use the navigation keys to toggle between functions such as L and L . Press View
Press Exit ( ) to leave the Memory Results , or press View B Memory
Results screens again. Note that the display screens will vary according to the model of CEL-63x instrument you are using.
Delete Memory Results
Refer to the User Manual for details and instructions to delete the results stored in the instrument’s memory.
Figure 5. Settings options Figure 7. Measurement data sets