Kilotech KHS 200-300, KHS 200-60 User Manual

Add to My manuals
17 Pages

advertisement

Kilotech KHS 200-300, KHS 200-60 User Manual | Manualzz

Content

1. Introduction .............................................................. 1

Notice ................................................... 1

Safety Guide ........................................ 1

2. Specifications ........................................................... 2

Features ............................................... 2

Specifications ...................................... 3

Capacity and Resolution ................... 4

Schematic Diagram ............................ 4

Dimension ........................................... 5

3. Operation Guide ...................................................... 6

Power On ............................................. 6

Power Off............................................. 6

Tare ....................................................... 7

Hold ..................................................... 8

Unit Switch .......................................... 8

Zero ...................................................... 9

Setting .................................................. 9

4. Trouble-shooting ................................................... 12

5. Note .......................................................................... 13

1.

Introduction

Notice

Before you use the scale, please read this manual through carefully, and keep it properly for future use.

Safety Guide

For good performance and precise measurement, be careful with daily operation and maintenance. Note the following instructions:

 Do NOT overload the scale. This will damage the loadcell and void the warranty.

 Do NOT leave load hung on the scale for long. This will decrease the scale’s accuracy and shorten the loadcell’s life.

 Inspect shackle and hook before using.

 When the scale runs out of power, replace the battery with fresh ones.

 Do NOT use the scale under thunder or rain.

 Do NOT attempt to repair the scale yourself.

Contact Kilotech at 800-694-4445, 877-328-5988 or www.kilotech.com.

Page 1 of 13

2.

Specifications

Features

This scale is a combination of the sound and proven mechanical design, with today’s most advanced electronics to provide a superb feature set. It is versatile, reliable, accurate and easy to operate.

Quality

Construction

& design

Power

In accordance with OIML R76,

GB/T11883-2002. ISO9001-2000 certified quality system.

Stainless steel load receptor and aluminum-casting case for better safety.

0.75 inch - 20mm LCD, visual distance over 10m. Dismountable hook.

SMT technology, quality integrated circuit and dedicated weighing loadcell, ensures long stability time.

3*AA battery with low power consumption design.

Page 2 of 13

Specifications

Accuracy Class Chinese GB/T 11883-2002 Class III

Equivalent to OIML R76

Tare Range 100% F.S.

Auto Zero

Manual Zero

± 50% F.S.

± 2% F.S.

0.5e/s Zero-tracking

Reading Stable

Time

≤10 seconds

Auto-sleep

Auto-off

Overload

Max. Safety

Load

Ultimate Load

Battery Life

Sable and no action in 3 sec.

Sable and no action in 3 min.

100% F.S. + 9e

120% F.S.

300% F.S.

>150 hours

Battery 3*AA battery

Temp. (Op.) - 10 ℃ ~ + 40 ℃

Humidity (Op.) ≤80% at 20 ℃ (non-condensing)

Display 0.7 inch (17.78mm) numerical

Net Weight 620g

Page 3 of 13

Capacity and Resolution

Model

Max. Cap.

(kg/lb)

30/60 KHS 200-30

KHS 200-60 60/120

KHS 200-150 150/300

KHS 200-300 300/600

E

(kg/lb)

Division

(n)

0.01/0.02 3,000

0.02/0.05 3,000

0.05/0.1 3,000

0.1/0.2 3,000

Schematic Diagram

Power

Loadcell

AD C

PGA

CPU

LCD

Key

Page 4 of 13

Dimension

Page 5 of 13

3.

Operation Guide

Power On

Press for 1 second to power up, battery tests and initializes.

 Screen shows detection message while scale detects its load and auto-zero itself.

Power Off

Press for 1 second to power off scale.

 Display shows power-off message.

Page 6 of 13

Tare

Press to store weight as tare.

 “TARE” shows and weight turns zero.

 If weight is over

100%F.S., or less than 0, tare is not allowed.

 If weight is not stable or held, tare is not allowed.

 Taring will reduce scale’s apparent overloading range. E.g., if a 10kg container is tared and scale’s maximum capacity is

50kg, scale will overload at a new weight of

40.18kg (5000 – 1000 + additional 9 divisions).

Press to restore tare.

 “TARE” hides, and weight resumes in gross mode.

Page 7 of 13

Hold

Press to lock the display.

 “HOLD” shows and display is frozen.

Press again to resume display.

 “HOLD” hides and display resume refreshing.

Unit Switch

Press for 1 second to switch measuring unit between kg, lb and N.

 If scale is tared, or held, unit switching is not allowed.

Page 8 of 13

Zero

Press for 1 second to zero scale.

 “ZERO” shows and weight turns zero.

 If weight is tared, or not stable, or held, zero is not allowed.

 If weight is over Manual-Zero range, zero is not allowed.

Setting

Press and sim ultaneously, and hold the two buttons for 1 second, to enter Setting Mode.

Press to enter

Auto-Off option; and press to change

Auto-Off option.

Page 9 of 13

 Auto-Off can be set to 0 (disabled), 5 (5min),

15 (15min), 30 (30min), 60 (60min).

Press to enter

Auto-Zero option; and press to change

Auto-Zero option.

 Auto-Zero can be set to 0 (disabled), 2

(2%F.S.), 4 (4%F.S.), 10 (10%F.S.), 20

(20%F.S.), 50 (50%F.S.).

Press to enter

Manual-Zero option; and press to change

Manual-Zero option.

 Manual-Zero can also be set to 0 (disabled),

2 (2%F.S.), 4 (4%F.S.), 10 (10%F.S.), 20

(20%F.S.), 50 (50%F.S.).

Press to enter

Backlight option; and press to change

Backlight option.

 Backlight can be set to off (disabled), 5

(5sec), 15 (15sec), 30 (30sec), 60 (60sec), on

Page 10 of 13

(never).

 When Backlight is set to 5, 15, 30, or 60, scale turns off backlight in corresponding seconds after it is stable.

 When Backlight is set to off, backlight is disabled, and never turns on.

 When Backlight is set to on, scale never turns off backlight, no matter it is stable or not.

Press to exit the

Settings Mode.

Page 11 of 13

4.

Trouble-shooting

Symptom blank display when

On/Off is pressed no action taken after TARE or

HOLD is pressed unstable readings reading is not zero without load large error in weight reading

Possible Causes discharged battery defective battery defective ON/OFF key

Solution replace battery press ON/OFF key for long defective TARE or

HOLD key clean TARE or

HOLD key scale in motion scale is damped dry the scale dust on PCB board clean PCB board unstable system power load-cell stressed too much or too long stabilize the load and scale longer warm-up time hang the scale in storage scale is not zeroed before loading re-calibration needed improper unit keep the scale unload and reboot re-calibrate the scale switch to proper unit

Page 12 of 13

5.

Note

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

__________________________________________________

Page 13 of 13