Technical Update

WATER,PR,10090 – 12/2009

PROCEDURE

Equipment Type:

WRO 300

Subject:

Updated Installation Procedure

From:

Richard Homol

A new installation procedure and checklist have been released for the WRO300.

Along with the procedure, the installation kit has also changed. The procedure and checklist are attached.

In the past, installation kits were shipped with a pressure regulator. The pressure regulator has been replaced with the Gambro flow regulator, WTW522001001

.

The flow regulator should be installed when the operating pressure is greater then 30 psi. It is not required at facilities where the operating pressure falls below

30 psi., but the device should be left with the customer for future use.

There have also been changes to the conductivity calibration, according to the new service manual, WTHCEN12073.

Other minor content changes have been made in anticipation of changes to the pretreatment components.

Please refer to Table 2, Tools List, to make sure you have all the necessary tools to perform the installation.

If you have any questions, please call Technical Support at 1-800-525-2623.

________________________________________________________________________

Rev. A

FORM #. TAA.FRM.0042

Effective.10.08.2008

WRO 300 Installation Procedure

WT300INST01 Rev. A Page 1 of 10

Introduction

This document provides instructions for completion of the WRO 300 installation. Steps are numbered to correspond with the installation checklist, WT300INSTLIST. This procedure must be used in conjunction with the WRO 300 Installation Guide

HCEN12115 (provided by the manufacturer and shipped with the WRO 300); the WRO

300 Service Manual, WTHCEN12073; the installation kit, WT300INSTKIT01, that is shipped with the WRO 300 in a separate box; and GXP with the latest version of the defaultUSxp file.

The WRO 300 Installation Guide provides information on WRO 300 setup and hose connections, as well as information on STARTUP after installation.

The WRO 300 Service Manual provides valuable information regarding internal menus, screen navigation, service pass codes and performance verifications.

WT300INSTKIT01 provides the parts necessary to complete the installation. Parts included in the kit are listed in Table 1.

It will be necessary to use GXP and a computer to configure the WRO 300.

Some of the parameters cannot be changed in the internal menu. See Tech Update Water

PR10081, or current Technical Update. The defaultUSxp file will allow the technician to configure all the machines properly and quickly. Also, certain values will need to be configured to clinic specific settings. You must use this program to properly complete the installation.

Table 1 Parts List, WT300INSTKIT01

Vendor Part #

WT300CAD

WT300HOSE

WTW522001001

K12755001

QTY Item Description

1 ea Collect Assist Device

1 ea Hose Assy for connecting WRO 300 to Dialysis Machine

1 ea Flow Regulator

1 ea Tube Removal Tool

9801-RTAG

9801-YTAG

WT300INSTLIST

WT300INST01

N/A

N/A

0500-092

0882-113

0883-113

0950-004

0766-315

1 ea Tag, Red Disinfect with Chain

1 ea Tag, Yellow Preservative with Chain

1 ea Installation Checklist

1 ea Installation Procedure

4 ea Bolt, Hex, SS, M6 x 25mm

4 ea Washer, Spring Lock, SS, M6

10 in Tubing, Clear Vinyl

2 ea HFC Hose Barb Insert, Polypropylene, 3/8” Hose Shutoff

2 ea HFC Hose Barb Body, Polypropylene, 3/8” Hose Straight Thru

8 ea Hose Clamp

1 Hose Adaptor, 3/8 barb x FGHT

WT300INST01 Rev. A Page 2 of 10

Tools Required

The tools listed in table 2 are needed to install the WRO 300.

The use of GXP is required to complete proper installation.

Table 2 Tools List

Torx screwdrivers T-10, T-20, & T-25

100psi test fixture, Pressure-Flow measurement tool

Installation Guide, shipped with WRO 300

25 bar pressure gauge

Quick Disconnect for RO Module

Not applicable

WT100TEST

HCEN12115

WT2510159

WTW507024001

Tube Removal Tool, supplied in Installation Kit

PC interface cable - Logging lead

PC cable adapter USB to serial adapter (If needed)

GXP Software with current defaultUS.xp files

Refer to Tech Update Water PR10056 and Water PR10081

Service Manual

Tech Update Vinegar Cleaning (if needed)

Tech Update Integrated Chemical Disinfection (if needed)

Tube cutting tool

K12755001

WTK18401001

Not Applicable

Can be downloaded from our website at tech.usa-gambro.com

WTHCEN12073

Water PR10075

Water PR10073R

N/A

Measuring Tape, 10 ft. N/A

Procedure

1. Assemble Gambro Cart and Prefilter Kits

1.1. Assemble the cart and install the pretreatment kits according to the instructions provided.

NOTE: If an alternative non-Gambro pretreatment system is to be used, it is the

responsibility of the customer to make sure all local, state, and federal regulations are followed. Use the quick-connect fittings, provided in the Installation Kit to attach the pretreatment assembly to the WRO 300. Use the quick-connect, valve stem, on the hose coming from the pretreatment components. See figure 1

NOTE: If the SS Accessory Box is to be installed, perform all of the Pretreatment and

WRO 300 installation steps and mount the WRO 300 to the cart before attaching the SS

Box. It is difficult to remove the cover of the WRO 300 after the box is installed.

NOTE: The use of carbon tanks will necessitate the installation of particle filtration

between the tanks and the inlet of the WRO 300. A 10” 5 micron filter is recommended.

WT300INST01 Rev. A Page 3 of 10

Figure 1

1.2. On the checklist, mark the type of pretreatment installed.

2. WRO 300 Assembly

Unpacking the WRO 300

NOTE: Complete instructions for unpacking and installing the WRO 300 are available

in the Installation Guide, HCEN12115, which is shipped with the WRO.

2.1. Carefully remove the WRO from the shipping container according to instructions found on the outside of the box and verify that all parts have shipped according to the chart found in section 1.1 of the Installation Guide.

2.2. Verify that all parts have shipped in the Installation Kit. See list above.

CAUTION! The WRO 300 contains a preservative which may cause irritation to eyes,

skin, etc. The internal sections of the machine may be under pressure due to altitude or atmospheric conditions. Use caution when removing the plugs to prevent contact with the preservative.

2.3. Remove the plugs from the back of the WRO using the tube removal tool

(K12215001) provided in the Installation Kit. See section 1.3 of the Installation

Guide.

Mounting the WRO 300

When mounting the WRO onto the Gambro cart make sure the WRO 300 is oriented in such a way that the back of the RO is on the same side of the cart as the cord handle. See

Figure 2.

WT300INST01 Rev. A Page 4 of 10

Figure 2

2.4. From underneath the cart base, use the four M6 bolts and lock washers to connect to the mounting holes in the bottom of the WRO 300.

2.5. Installation of Tubing and Product Loop

MAKE ALL TUBING CONNECTIONS USING THE HOSE CLAMPS

PROVIDED.

NOTE: Be certain to use the proper size clamps when attaching the tubing to the

WRO 300. The 6 ½ ft, ¼ in. ID tubing, 11mm hose clamps and smaller sized elbows are used for the product loop.

NOTE: Italized wording refer to specific steps found in the Installation Guide,

Section 1.3.2 Using the tubing provided, cut to appropriate length as described below.

2.5.1. Feed Water Inlet: Cut the feed line 3 ft long and attach a QC Body to one end of the hose, then connect it to the output connector of the pretreatment component.

WT300INST01 Rev. A Page 5 of 10

NOTE: Attach the QC Valve Stem to the outlet of the pretreatment line if

needed, then connect the hoses together.

2.5.2. Use an angled connector and clamp and attach the feed line to the back of the WRO 300 as shown in figure 1.5 of the Installation Guide

2.5.3. Product water outlet and return: Use the supplied 2 meter (6 1/2 ft) hose, cut it slightly off-center and attach the Y coupling.

2.5.4. Attach the 2 smaller angled connectors and attach to the WRO 300 as shown in figure 1.6 of the Installation Guide.

2.5.5. Drain Outlet: Cut the drain line in two parts. 1 part should be 1 ft long and attached to the WRO 300 using the angled connector.

Attach the QC Valve Stem to the opened end of the hose.

Cut the other part 8 ft. long.

Attach the QC Body to one end of the hose and connect the hoses together.

See figure 2 above.

2.5.6. Tank air vent: Use the 10 in. clear tubing supplied in the Installation Kit.

2.5.7. Incoming Water Line: Cut a section of hose 9 ft long and attach the hose adaptor. Connect the opened end of the hose to the input of the pretreatment system. Use the Quick Connect fittings if available. Connect the other end to a water source.

Note: The incoming water line may already be installed if the pretreatment was already assembled.

2.6. Install the power cord Be sure to run the cable underneath the handle.

2.7. Flush Carbon Pretreatment

2.7.1. Disconnect the QC connector from the drain line and attach it to the output of the first carbon component. Start with the Primary Carbon

Component, then do the Secondary Carbon Component if installed.

NOTE: If Gambro QC connectors have not been used on the pretreatment, it will

be necessary to flush the carbon components using what ever method that may be available.

NOTE: It is essential that the carbon be properly rinsed prior to starting the WRO

300. Carbon fines (carbon dust) present in new carbon filters and tanks can damage the RO membrane and pump if not properly rinsed out prior to use.

WT300INST01 Rev. A Page 6 of 10

2.7.2. Connect the water line to a water source and open the water valve partially to slowly purge the air from the pretreatment assembly.

2.7.3. Open the inlet water valve fully and allow the system to flush properly.

Black carbon fines will be observed at the drain. Allow carbon tanks to flush for 5 minutes after the drain line becomes clear. Carbon Filters should be flushed for an additional 2 minutes after the drain water is clear.

2.7.4. After flushing, turn off the water and proceed to the second carbon component or return the hoses to their original positions.

3. WRO 300 Rinse

3.1. Follow the instructions for rinsing the WRO 300 provided in section 1.5 of the

Installation Guide.

3.2. During Rinse observe the lowest pressure indicated on the pressure gauge located just before the RO. This is the Operating Pressure. Record this value on the checklist.

3.3. If the Operating Pressure is greater then 30 psi it is recommended to install the flow regulator. Operating pressures above 44 psi will require the installation of the flow regulator.

3.3.1. At the end of Rinse, close off the water supply and relieve the pressure in the pretreatment components.

3.3.2. Cut the tubing just below the back handle of the WRO and attach the flow regulator. See figure 2 above. (The arrow should point to the WRO 300.)

If not installed, this part should be left with the customer in case of future needs.

3.4. During Rinse verify there are no leaks in the pretreatment or in the WRO 300.

Repair any leaks as necessary. Keep checking this throughout the installation to ensure the system is left leak-free for the customer.

3.5. Rinse Completed. A full rinse must be completed as the machine has been packed in a preservation and anti-freeze solution.

3.6. Check for complete rinse-out: Refer to 1.5.1 in the Installation Guide.

3.7. Record the Product water conductivity value as shown on the display.

3.8. Configure the WRO 300 using GXP. Follow the procedure according to Tech

Update, Water PR10081 or the current Technical Update. Use the defaultUS.xp file to properly configure the WRO 300. Make any clinic specific changes as required.

WT300INST01 Rev. A Page 7 of 10

NOTE: DefaultUS.xp files are software version specific and will be update as new

software is released. Be sure to use the correct version when configuring the WRO 300.

At the time of this release current software version is 4.2

4. WRO 300 Performance Verification

NOTE: When attaching or removing devices using the quick connect fittings, ALWAYS make sure the system is in Standby. Pressure in the system can cause orings inside the connectors to become dislodged, causing leaks.

Install the 25 bar pressure gauge to the top of the membrane module and install the CAD to the product line with the valve closed.

4.1. Conductivity Offsets set to 0.

4.1.1. Enter the Internal Service mode and scroll to CALIBRATE COND.

Verify that the FW and PW offsets are set to 0.

4.1.2. Place the WRO 300 into Run. (this can be done while in the service mode) by holding down the RUN button for 3 seconds.

Figure 3

4.2. Verify Feed Water Conductivity

Adjust the CAD (see figure 3), to allow about 500 ml/min of product consumption.

Allow the system to run for 15 minutes.

Using a calibrated conductivity meter, take a water sample from the sample port located after the carbon filtration.

WT300INST01 Rev. A Page 8 of 10

Compare the reading to the temperature compensated value (T) on the display.

The values should be within 10% or 10 uS, whichever is greater of each other.

Adjust the COEFF as needed.

4.2.1. Record the COEFF value

4.2.2. Record the conductivity value.

4.3. Verify Product Water Conductivity.

Using a calibrated conductivity meter, take a water sample from the end of the product loop using the CAD.

Compare the reading to the display on the WRO 300.

The values should be within 10% or 1uS, whichever is greater of each other

Adjust the COEFF as needed

4.3.1. Record the COEFF value

4.3.2. Record the conductivity value.

4.4. Verify Product Pressure and Flow. Refer to the Service Manual, section 6.4 for information on adjusting the flow and pressure. Flow should be 1200 + 100 ml/min., with a loop pressure between 22 to 30 psi. Measure the flow for at least

30 seconds

4.4.1. Record the loop pressure

4.4.2. Record flow rate

4.5. Verify the Reject Pressure is less then 13 bar (190psi.). Record the value

4.6. Verify Proper Operation of the Chemical Pump.

Using water for testing, initiate a Minncare Disinfection procedure and verify 50 ml of volume uptake (+ 20 % ).

4.6.1. Put 100 ml of water into a graduated cylinder.

4.6.2. Insert the disinfect wand into the front of the WRO 300 and place the other end in the graduated cylinder

4.6.3. Use the arrow keys to select Minncare

4.6.4. Initiate the chemical disinfection buy pressing DIS

4.6.5. The WRO 300 will enter the chemical intake phase.

4.6.6. Verify the WRO 300 pulled in 50 ml (+ 10ml)

4.6.7. Record the volume.

WT300INST01 Rev. A Page 9 of 10

4.7. Clear the Rinse Flag using Internal Service

4.7.1. Enter Internal Service following the procedure listed in section 2.7 of the

4.7.2. Scroll to Clear Chem Flag and confirm.

4.7.3. Set CHEM FLAG to False using one of the arrow keys and confirm.

4.7.4. Exit the Internal Service mode by pressing the DIS button 3 times.

4.7.5. Cycle power using the power switch at the back of the machine.

Upon power up, the system should return to Standby.

4.8. Verify Leakage Current is less then 145 uA and record the value.

4.9. Verify Ground Resistance is less then 0.3 Ohms and record the value.

4.10. Install the SS Lock Box, if provided. Be sure the one way bolt heads are facing to the outside of the box (Mark N/A if the Lock Box is not installed.)

4.11. Advise Customer that AAMI analysis for Chemical and Bacterial Contaminants is required prior to use.

WT300INST01 Rev. A Page 10 of 10

WRO 300 INSTALLATION CHECKLIST

EQUIPMENT LOCATION_____________________ SERVICE DATE(S) _______________________________

SERIAL NUMBER___________________________ WORK ORDER__________________________________

MACHINE HOURS___________________________ TECHNICIAN: ___________________________________

A checkmark or a “X” indicates the step is completed or function is within the specifications listed on the checklist, procedure or

Service Manual. Record values where units are given.

1. ASSEMBLY OF GAMBRO CART AND PRETREATMENT KITS.

1.1. Gambro CART and pretreatment Kits assembled and installed per instructions.

(Mark N/A is a non-Gambro pretreatment system is used)

1.2. Type of pretreatment installed:

1.2.1. Carbon Filter

________

1.2.2. Dual Carbon Filters

1.2.3. Carbon Tank w Carbon Filter

1.2.4. Dual Carbon Tanks

1.2.5. Other

2. WRO 300 ASSEMBLY

2.1. Remove WRO from the shipping container and verify all parts have been supplied

2.2. Verify all parts have shipped in Installation Kit content

________

________

________

________

________

________

________

2.3. Remove plugs from back of WRO 300 ________

2.4. Mount the WRO 300 to the Gambro Cart. (Mark N/A if non-Gambro cart was used) ________

2.5. Hoses attached to the WRO 300 per Installation Guide and Installation Procedure

2.6. Power cord installed

2.7. Carbon Components Properly Flushed

________

________

________

3. WRO 300 RINSE

3.1. Initiate Rinse

3.2. Record Operating Pressure

3.3. Install Flow Regulator (Mark N/A if the flow regulator is not installed)

3.4. Check pretreatment and WRO for leaks

3.5. Rinse completed. Refer to section 1.5 of the installation guide.

3.6. Check for complete rinse out per Installation Guide 1.5.1

3.7. Record product water conductivity

3.8. Configure the WRO 300 using GXP and the current defaultUS.xp file

4. WRO 300 PERFORMANCE VERIFICATION

4.1. Conductivity Offsets set to zero

4.2. Verify feed water conductivity

4.2.1. Record the COEFF

4.2.2. Record conductivity value

4.3. Verify product water conductivity

4.3.1. Record the COEFF

4.3.2. Record conductivity value

4.4. Verify product pressure and flow

4.4.1. Record pressure setting

________

________ psi

________

________

________

________

________ uS

________

________

________

________

________ uS

________

________

________ uS

________

________ psi

4.4.2. Record flow rate

4.5. Verify reject pressure is less than 13 Bar (190 psi) Record the pressure

4.6. Verify proper operation of the chemical pump and record the uptake volume

4.7. Clear rinse flag using Internal Service

________ l/min

________ psi

________ ml

________

4.8. Verify leakage current does not exceed specification of 145 uS Record value

4.9. Verify ground resistance is below 0.3 ohms Record value

________ uA

________

4.10. Install SS lock box (if provided) (Mark N/A if the lock box is not installed) ________

4.11. Advise user that AAMI analysis for Chemical and Bacterial contaminates is required ________

Part Number: WT300INSTLIST REV: C