- Frank`s Hospital Workshop

Add to my manuals
256 Pages

advertisement

- Frank`s Hospital Workshop | Manualzz
S/5 Aespire Anesthesia Machine
Technical Reference Manual
S/5 Aespire
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product
group code, the year of manufacture and a sequential unit number for identification.
AAA F 12345
This alpha character indicates the year of product manufacture
and when the serial number was assigned;
“D” = 2000, “E” = 2001, “F” = 2002, etc.
“I” and “O” are not used.
Aespire, ProTIVA, SmartVent, and Link-25 are registered trademarks of Datex-Ohmeda Inc.
Other brand names or product names used in this manual are trademarks or registered
trademarks of their respective holders.
07/04
1009-0356-000
Technical Reference Manual
Covers the following:
S/5 Aespire anesthesia machine
S/5 Aespire 100 anesthesia machine
S/5 ProTIVA anesthesia machine
This document is not to be reproduced in any manner, nor are the contents to be disclosed to
anyone, without the express authorization of the product service department, Datex-Ohmeda,
Ohmeda Drive, PO Box 7550, Madison, Wisconsin, 53707.
©
1009-0356-000 07/04
2004 Datex-Ohmeda Inc.
i
S/5 Aespire
Important
The information contained in this service manual pertains only to those models of products
which are marketed by Datex-Ohmeda as of the effective date of this manual or the latest
revision thereof. This service manual was prepared for exclusive use by Datex-Ohmeda service
personnel in light of their training and experience as well as the availability to them of parts,
proper tools and test equipment. Consequently, Datex-Ohmeda provides this service manual to
its customers purely as a business convenience and for the customer's general information only
without warranty of the results with respect to any application of such information. Furthermore,
because of the wide variety of circumstances under which maintenance and repair activities
may be performed and the unique nature of each individual's own experience, capacity, and
qualifications, the fact that customer has received such information from Datex-Ohmeda does
not imply in anyway that Datex-Ohmeda deems said individual to be qualified to perform any
such maintenance or repair service. Moreover, it should not be assumed that every acceptable
test and safety procedure or method, precaution, tool, equipment or device is referred to within,
or that abnormal or unusual circumstances, may not warrant or suggest different or additional
procedures or requirements.
This manual is subject to periodic review, update and revision. Customers are cautioned to
obtain and consult the latest revision before undertaking any service of the equipment.
Comments and suggestions on this manual are invited from our customers. Send your
comments and suggestions to the Manager of Technical Communications, Datex-Ohmeda,
Ohmeda Drive, PO Box 7550, Madison, Wisconsin 53707.
w CAUTION
Servicing of this product in accordance with this service manual should never be
undertaken in the absence of proper tools, test equipment and the most recent revision
to this service manual which is clearly and thoroughly understood.
Technical Competence
The procedures described in this service manual should be performed by trained and authorized
personnel only. Maintenance should only be undertaken by competent individuals who have a
general knowledge of and experience with devices of this nature. No repairs should ever be
undertaken or attempted by anyone not having such qualifications.
Datex-Ohmeda strongly recommends using only genuine replacement parts, manufactured or
sold by Datex-Ohmeda for all repair parts replacements.
Read completely through each step in every procedure before starting the procedure; any
exceptions may result in a failure to properly and safely complete the attempted procedure.
ii
07/04 1009-0356-000
Table of Contents
Important . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Technical Competence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
1 Introduction
1.1 What this manual includes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2 Standard service procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2.1 User’s Reference Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2.2 Technical Reference Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.3 What is an S/5 Aespire . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3.1 Aespire 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3.2 S/5 ProTIVA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.4 Configuration options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.4.1 Standard configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.4.2 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.5 Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.6 Symbols used in the manual or on the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
2 Theory of Operation
2.1 Theory overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2 Gas flow through the anesthesia machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.2 Physical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.2.3 Suction regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
2.2.4 System switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.2.5 Flow control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
2.3 Flow through the breathing system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2.3.1 Overview of flow paths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3.2 Manual ventilation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3.3 Mechanical ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3.4 Fresh gas and O2 flush flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1009-0356-000 07/04
2-11
2-12
2-15
2-18
iii
S/5 Aespire
3
Checkout Procedure
3.1 Inspect the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
3.2 Pipeline and cylinder tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
3.3 Flow control, pressure relief, O2 supply alarm, and flush flow tests . . . . . . . . . . . . . . . . . .3-4
3.3.1 With O2 monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3.3.2 Without O2 monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3.3.3 Pressure relief tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.3.4 O2 supply alarm test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.3.5 Flush Flow Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.4 Vaporizer back pressure test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3.5 Low-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.5.1 Negative low-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.5.2 ISO or BSI standard low-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3.6 Alarm tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.7 Breathing system tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.8 Auxiliary O2 flowmeter tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
3.9 Integrated Suction Regulator tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
3.10 Power failure test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
3.11 Electrical safety tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
4
Repair Procedures
4.1 Servicing the ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
4.2 How to bleed gas pressure from the machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.3 How to remove the rear panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.3.1 To remove the rear upper panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.3.2 To remove the lower access panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.4 How to remove the tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
4.5 Replace pipeline inlet filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
4.5.1 Replace pipeline inlet check valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
4.6 Change drive gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.7 Service the cylinder supply modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.7.1 Tightening procedure for high-pressure tube fittings . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.7.2 Replace primary regulator module (complete replacement) . . . . . . . . . . . . . . . . .4-8
4.7.3 Replace cylinder inlet filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
4.7.4 Replace cylinder check valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
4.7.5 Replace 3rd-gas cylinder supply module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.8 Replace system switch assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
iv
07/04 1009-0356-000
Table of Contents
4.9 Service the flowmeter module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.9.1 Remove front flowmeter panel shield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.9.2 Remove flowtubes for cleaning or replacement . . . . . . . . . . . . . . . . . . . . . . . . . .
4.9.3 Remove complete flowmeter head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.9.4 Replace flowmeter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.9.5 Replace flowmeter frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.9.6 Replace O2 supply switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-13
4-15
4-16
4-20
4.10.1 Repair manifold port valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.10.2 Checkout procedure for manifold port valve . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.10.3 Replace vaporizer manifold check valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.10.4 Replace vaporizer pressure relief valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.10.5 Replace vaporizer manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.11 Replace ACGO selector switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-26
4-27
4-28
4-30
4-31
4-32
4-21
4.9.7 Checkout procedure for O2 supply switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
4.9.8 Replace secondary regulator manifold or balance regulator manifold . . . . . . . 4-22
4.9.9 Replace O2 or N2O needle valves (on machines with N2O) . . . . . . . . . . . . . . . . . 4-23
4.9.10 Replace an Air needle valve on all machines or an O2 needle valve on machines
without N2O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
4.10 Service vaporizer manifold parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
4.12 Clean or replace ACGO port flapper valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
4.13 Reconfigure sample gas return line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-35
4.14 Replace the APL valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
4.15 Replace the bag support arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37
4.15.1 Servicing the bag support arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.15.2 Replace friction pad in lower bag arm assembly . . . . . . . . . . . . . . . . . . . . . . . .
4.15.3 Replace bag port housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.16 Replace auxiliary O2 flowmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-38
4-39
4-40
4-41
4.17 Replace the suction control module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-42
4.17.1 Front panel method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-42
4.17.2 Rear panel method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
4.18 Replace ABS breathing system components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
4.18.1 Replace Bag/Vent switch assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
4.18.2 Replace bellows base latch assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
4.19 Replace casters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-46
4.20 Replace task light and switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
4.20.1 To replace the task-light switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
4.20.2 To replace the task-light circuit board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
1009-0356-000 07/04
v
S/5 Aespire
4.21 Replace the display arm or display cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-48
4.21.1 Cable tie installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-48
4.21.2 Removing the display arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.21.3 Replacing a display cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.21.4 Installing the long arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.21.5 Installing the short arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.22 Replace display and cables in ProTIVA machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5
4-49
4-49
4-50
4-51
4-52
Maintenance
5.1 Aespire Planned Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.1.1 Every twelve (12) months . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.1.2 Every twenty-four (24) months . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
5.2 Auxiliary O2 flowmeter tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
5.3 Integrated Suction Regulator tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
6 Calibration
6.1 Primary Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
6.1.1 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
6.1.2 Testing Primary Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
6.1.3 Adjusting Primary Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
6.2 Secondary Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
6.2.1 Testing/Adjusting Secondary Regulators or Balance Regulators . . . . . . . . . . . . . .6-7
6.3 Flowmeter Needle Valve Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
6.3.1 O2 Needle Valve Calibration (Minimum Flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.3.2 N2O Needle Valve Calibration (Minimum Flow). . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
6.3.3 Air Needle Valve Calibration (Minimum Flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
6.3.4 Needle Valve Calibration (Maximum Flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
6.4 Link system calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
6.5 O2 Flush Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
6.6 Airway pressure gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.6.1 Zero the pressure gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.6.2 Checking the pressure gauge accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
7
Troubleshooting
7.1 General Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
7.2 Breathing System Leak Test Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
7.2.1 Breathing system leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
7.2.2 Breathing System Troubleshooting Flowcharts . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7
7.2.3 Leak Isolation Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
vi
07/04 1009-0356-000
Table of Contents
8 Illustrated Parts
8.1 Service tools — Anesthesia machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
8.1.1 Test Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
8.1.2 Test Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
8.1.3 Secondary regulator pilot pressure tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
8.2 External components - front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-6
8.3 External components - front view references . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
8.4 External Components - rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-8
8.5 Control module mounting for a ProTIVA machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-9
8.6 Aespire 100 - exclusive components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
8.6.1 AC Inlet (Aespire 100). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
8.6.2 Display mount (Aespire 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13
8.7 Front panel, gauges and system switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
8.8 Rear panel components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15
8.9 Tabletop components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16
8.10 Right-side Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-17
8.11 External components - lower assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18
8.12 Vent Engine Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
8.13 Display cables, serial board, AGSS flowtube, and sample return . . . . . . . . . . . . . . . . . 8-20
8.14 AC Power cords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-21
8.15 AC Inlet/Outlet Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-22
8.16 Pipeline inlet fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-24
8.17 Cylinder Gas Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-25
8.17.1 Cylinder inlet fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-26
8.18 Vaporizer manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-27
8.19 Flowmeter components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-28
8.19.1 Flowtube parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-30
8.19.2 Secondary regulator components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-32
8.20 ABS to machine Interface Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-34
8.20.1 Flush Regulator, Flush Valve, and ACGO Selector Switch . . . . . . . . . . . . . . . . . 8-35
8.21 Breathing system interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-36
1009-0356-000 07/04
vii
S/5 Aespire
8.22 Breathing System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-37
8.22.1 APL Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-37
8.22.2 Bag/Vent Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.3 Absorber canister . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.4 Flow Sensor Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.5 Breathing Circuit Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.6 Exhalation valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.7 Bellows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.8 Bellow base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.9 Bag Arms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.23 Drawer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8-38
8-39
8-40
8-41
8-42
8-43
8-44
8-45
8-46
8.24 Legris quick-release fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-47
8.25 Vent Drive and low-pressure tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-48
8.26 Tubing for use with Legris fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-50
8.27 Cables and harnesses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-52
8.28 Cables and harnesses (Aespire 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-54
8.29 Anesthetic Gas Scavenging System — AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-56
8.29.1 Passive AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.29.2 Adjustable AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.29.3 Active AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.30 Integrated Suction Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8-56
8-58
8-60
8-62
8.30.1 Major Components (Continuous and Venturi suction) . . . . . . . . . . . . . . . . . . . . 8-62
8.30.2 Suction Control Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-63
8.30.3 Venturi assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-64
8.31 Auxiliary O2 Flowmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-65
8.32 Display mounts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-66
8.33 Cable management arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-67
8.34 Display arm mounting kits for optional equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-68
9 Schematics and Diagrams
viii
07/04 1009-0356-000
1
In this section
Introduction
This section provides a general overview of the S/5 Aespire Anesthesia Machine.
1.1 What this manual includes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2 Standard service procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2.1 User’s Reference Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2.2 Technical Reference Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.3 What is an S/5 Aespire . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3.1 Aespire 100 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3.2 S/5 ProTIVA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.4 Configuration options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.4.1 Standard configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.4.2 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.5 Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.6 Symbols used in the manual or on the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
1009-0356-000
07/04
1-1
S/5 Aespire
1.1 What this manual includes
Anesthesia Machine
This manual covers the service information for the S/5 Aespire line of
anesthesia machines. It covers the following components:
• gas delivery components,
• breathing system components,
• frame component (except those strictly associated with a specific
ventilator),
• optional suction regulator and auxiliary O2 flowmeter.
Ventilator
The ventilator associated with the S/5 Aespire machine has its own Technical
Reference Manual:
• for the7100 Ventilator see manual 1006-0836-000.
S/5 ProTIVA
The ProTIVA machine is configured with standard Aespire machine
components, with the exception of the vaporizer manifold and the Ventilator/
Monitoring Display (7100 Control Module) mounting solution
(refer to Section 8.5).
B Braun equipment is not covered in this manual. Refer to B Braun service
documentation.
Other equipment
Other equipment may be attached to the system on the display mount, the
top shelf, or on the side dovetail rails. Consult separate documentation
relative to these items for details.
1.2 Standard service procedures
1.2.1 User’s Reference
Manuals
Some sections of this manual refer you to the User’s Reference Manual for the
S/5 Aespire. To expedite repairs, you must have, and be familiar with, the
User’s Reference Manuals for this product.
Refer to the S/5 Aespire User’s Reference Manual if you need further
information about the operation of the system.
1.2.2 Technical
Reference Manuals
You must first determine where a problem is located before you can determine
which Technical Reference Manual to use:
• Use this manual for machine and breathing system related issues.
• Use the 7100 Ventilator manual for ventilator related issues.
1-2
07/04 1009-0356-000
1 Introduction
1.3 What is an S/5 Aespire
The S/5 Aespire is a compact, integrated and intuitive anesthesia delivery
system. The ventilator portion provides mechanical ventilation to a patient
during surgery as well as monitoring and displaying various patient
parameters.
The system uses a microprocessor-controlled ventilator with internal
monitors, electronic PEEP, Volume Mode, and other optional features. A serial
interface permits communication to cardiovascular and respiratory gas
monitoring.
Note
Configurations available for this product depend on local market and
standards requirements. Illustrations in this manual may not represent all
configurations of the product.
The S/5 Aespire is not suitable for use in an MRI environment.
1.3.1 Aespire 100
The Aespire 100 machine is based on the standard S/5 Aespire machine with
the following exceptions to available features or options:
• does not include the RS232 Serial Interface
• does not include the Bi-level LED light strip
• the 2 Vap manifold is standard (does not support 1 Vap manifold)
• not available with AC power outlets (area used by AC Inlet)
• uses 4-inch casters instead of 5-inch casters
1.3.2 S/5 ProTIVA
1009-0356-000
07/04
The S/5 ProTIVA is a special adaptation of the S/5 Aespire machine for use
with B Braun intravenous drug delivery components.
1-3
S/5 Aespire
1.4 Configuration options
1.4.1 Standard
configuration
The standard configuration includes the following items.
Items marked with an asterisk (*) are not included in the Aespire 100
machine.
• 7100 Ventilator
• Advanced Breathing System (ABS)
• Auxiliary Common Gas Outlet (ACGO)
• Serial Interface - RS232*
• Bi-level LED light strip*
• Two large drawers
1.4.2 Options
Options include the following items.
Items marked with an asterisk (*) are not available in the Aespire 100
machine).
• selected software features
• vaporizer manifold (1 Vap* or 2 Vap)
• pipeline configurations (O2/N2O, O2/Air, or O2/N2O/Air)
• gas cylinder configurations (two inboard, one outboard)
• inboard configuration = O2/N2O, O2/Air, or O2/O2
• outboard configuration = N2O only
• manual bag (on support arm or on tube)
• gas scavenging (active, adjustable, passive, or venturi)
• a suction regulator (pipeline vacuum or venturi vacuum)
• an auxiliary O2 flowmeter
• localized electrical power outlets* (isolated or non-isolated)
• various display mounting solutions
1.5 Components
The following figures show the front and rear views of the machine.
There are some differences between models.
Figure 1-1 •S/5 Aespire (front view - left side)
Figure 1-2 •S/5 Aespire (front view - right side)
Figure 1-3 •S/5 Aespire (rear view)
Figure 1-4 •S/5 ProTIVA with a typical B Braun fluid manager (fm) system
1-4
07/04 1009-0356-000
1 Introduction
18
17
16
15
1
14
2
AB.74.054
13
12
11
10
9
8
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
7
6
5
4
3
Auxiliary common gas outlet (ACGO) switch
ACGO
Inspiratory check valve
Inspiratory flow sensor or flow port adapter
Canister (carbon dioxide absorbent)
Canister release
Expiratory flow sensor or flow port adapter
Leak test plug
Expiratory check valve
Breathing system release
Manual bag port
APL (adjustable pressure-limiting) valve
Bag/Vent switch
Bellows assembly
Pressure gauge (airway)
Sample gas return port
Scavenging flow indicator
RS-232 Serial port (not available in the Aespire 100 machine)
Figure 1-1 • S/5 Aespire (front view - left side)
1009-0356-000
07/04
1-5
S/5 Aespire
3
4
2
5
6
10
7
11
AB.74.053
1
9
8
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
ABS (Advanced Breathing System)
Flow controls
Ventilator Display/Control Module
Dovetail rails
Vaporizer
Pipeline pressure gauge(s) (upper row)
System switch
Cylinder pressure gauge(s) (lower row)
O2 Flush
Auxiliary O2 flowmeter
Suction regulator
Figure 1-2 • S/5 Aespire (front view - right side)
1-6
07/04 1009-0356-000
1 Introduction
AC Inlet for Aespire 100 machine
7
5
6
1
2
3
AB.74.004
4
5
6
7
1.
2.
3.
4.
5.
6.
7.
Circuit Breaker for Electrical Outlets*
Electrical Outlets*
Pipeline Connection(s)
Cylinder Supplies
System Circuit Breaker (AAC Inlet fuses for Aespire 100 machine)
Mains Inlet
Equipotential Stud
* Items marked with an asterisk (*) are not available in the Aespire 100 machine.
Figure 1-3 • S/5 Aespire (rear view)
1009-0356-000
07/04
1-7
AB.97.009
S/5 Aespire
5
3
4
2
AB.97.002
1
1.
2.
3.
4.
5.
Ventilator/Monitoring display
Syringe pumps (3)
fm controller
fm computer
fm segment
Figure 1-4 • S/5 ProTIVA with a typical B Braun fluid manager (fm) system
1-8
07/04 1009-0356-000
1 Introduction
1.6 Symbols used in the manual or on the equipment
w Warnings and w Cautions tell you about dangerous conditions that can
occur if you do not follow all instructions in this manual.
Warnings tell about a condition that can cause injury to the operator or the
patient.
Cautions tell about a condition that can cause damage to the equipment. Read
and follow all warnings and cautions.
Other symbols replace words on the equipment or in Datex-Ohmeda manuals.
No one device or manual uses all of the symbols. These symbols include:
m
L
l
n
M
N
†
p
x
y
P
Y
1009-0356-000
On (power)
A
Off (power)
Standby
Standby or preparatory state for part of
the equipment
“ON” only for part of the equipment
“OFF” only for part of the equipment
Direct current
Alternating current
Alarm silence button
Alarm silence touch key (Tec 6).
j
Type B equipment
J
Type BF equipment
D
w
wW
O
Type CF equipment
Caution, ISO 7000-0434
Attention, refer to product instructions,
IEC 601-1
Dangerous voltage
Protective earth ground
Electrical input
Earth ground
Electrical output
Frame or chassis ground
Pneumatic input
Equipotential
Pneumatic output
07/04
1-9
S/5 Aespire
+
t
Plus, positive polarity
Minus, negative polarity
Movement in one direction
Movement in two directions
Variability
Read top of float
Variability in steps
Vacuum inlet
g
This way up
Suction bottle outlet
o
Lamp, lighting, illumination
T
Cylinder
z
Lock
Z
Unlock
Isolation transformer
U
Close drain
u
Open drain (remove liquid)
134°C
Autoclavable
Í
Not autoclavable
q
t
REF
1-10
k
E
Linkage system
Risk of Explosion.
Inspiratory flow
O2 sensor connection
Stock Number
Low pressure leak test
r
R
Q
O2 +
SN
Mechanical ventilation
Bag position/ manual ventilation
Expiratory flow
O2 Flush button
Serial Number
07/04 1009-0356-000
1 Introduction
< 345 kPa
Alarm silence touch key
Volume alarms On/Off touch key
End case touch key
Menu touch key
Circle breathing circuit module
Bain/Mapleson D breathing circuit
module
The primary regulator is set to pressure
less than 345 kPa (50 psi)
Absorber on
Absorber off (CO2 Bypass active)
European Union Representative
1009-0356-000
07/04
< 414 kPa
The primary regulator is set to pressure
less than 414 kPa (60 psi)
CO2 Bypass Option
Systems with this mark agree with the
European Council Directive (93/42/
EEC) for Medical Devices when they
are used as specified in their
Operation and Maintenance Manuals.
The xxxx is the certification number of
the Notified Body used by DatexOhmeda’s Quality Systems.
1-11
Notes
1-12
07/04 1009-0356-000
2
In this section
Theory of Operation
2.1 Theory overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2 Gas flow through the anesthesia machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.2 Physical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.2.3 Venturi Suction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
2.2.4 System switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.2.5 Flow control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
2.3 Flow through the breathing system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2.3.1 Overview of flow paths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3.2 Manual ventilation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3.3 Mechanical ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3.4 Fresh gas and O2 flush flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1009-0356-000
07/04
2-11
2-12
2-15
2-18
2-1
S/5 Aespire
2.1 Theory overview
This section describes:
• The flow of gas through the anesthesia machine.
• The flow of gas through the breathing system.
• Electrical signals between the anesthesia machine, including the breathing
system, and the ventilator.
2.2 Gas flow through the anesthesia machine
2.2.1 Overview
Gas supplies
Refer to Figure 2-1.
Gas comes into the system through a pipeline (2) or cylinder (4) connection.
All connections have indexed fittings, filters, and check valves (one-way
valves). Gauges show the pipeline (1) and cylinder (3) pressures.
A primary regulator (5) decreases the cylinder pressures to approximately
pipeline levels. A pressure relief valve (6) helps protect the system from high
pressures.
To help prevent problems with the gas supplies:
• Install yoke plugs on all empty cylinder connections.
• When a pipeline supply is adequate, keep the cylinder valve closed.
O2 flow
Pipeline or regulated cylinder pressure supplies O2 directly to the ventilator
(7a for O2 drive gas) and the venturi suction (21a for O2 drive gas) supply
connection. An additional regulator (13) decreases the pressure for the flush
valve (14a) and the auxiliary flowmeter (25).
The flush valve supplies high flows of O2 to the fresh gas outlet (26 or 27)
when you push the flush button. The flush pressure switch (14b) monitors
activation of the flush valve.
When the system switch (8) is On, O2 flows to the rest of the system.
A secondary regulator (10) supplies a constant O2 pressure to the O2 flow
control valve (11). There is a minimum flow of 25 to 75 mL/min (for dual-tube
flowmeters) or 175 to 225 mL/min (for single-tube flowmeters) through the
O2 flowmeter (12).
The O2 pressure switch (9) monitors the O2 supply pressure. If the pressure is
too low, an alarm appears on the ventilator display.
2-2
07/04 1009-0356-000
2 Theory of Operation
Air and N2O flow
Pipeline or regulated cylinder pressure supplies Air directly to the ventilator
(7b for Air drive gas) and the venturi suction (21b for Air drive gas) supply
connection.
When the system switch (8) is On, air flows to the rest of the system.
A secondary regulator (18) supplies the Air flow control valve (19). Because
there is no balance regulator, air flow continues at the set rate during an O 2
supply failure.
A balance regulator (15) controls the N2O supply pressure to the N2O flow
control valve(16). The O2 secondary regulator pressure at a pilot port controls
the output of the balance regulator. The N2O output pressure drops with
decreasing O2 supply pressure and shuts off hypoxic gas flow before the O 2
supply pressure reaches zero.
A chain link system (Link-25) on the N2O and O2 flow controls (16, 11) helps
keep the O2 concentration higher than 21% (approximate value) at the
common gas outlet.
Mixed gas
The mixed gas goes from the flowmeter outlet, through the vaporizer manifold
and vaporizer (23) that is On, to the ACGO selector switch (E). A pressure relief
valve (24) sets the maximum outlet pressure.
The ACGO selector switch directs the mixed gas to the selected circuit — to the
breathing system (26) or to the ACGO (27).
1009-0356-000
07/04
2-3
S/5 Aespire
Key to Numbered Components
Key to Symbols
1.
2.
3.
4.
5.
6.
7.
Pipeline pressure gauge
Pipeline inlet
Cylinder pressure gauge
Cylinder inlet (maximum of 3 cylinders)
Primary regulator (cylinder pressure)
High-pressure relief valve (758 kPa / 110 psi)*
Supply connections for the ventilator
a. O2 drive gas
b. Air drive gas
8. System switch
9. Switch for low O2 supply pressure alarm (used with the ventilator)
10. O2 secondary regulator (207 kPa / 30 psi)*
11. O2 flow control valve
12. O2 flow tube(s)
13. O2 flush and auxiliary flowmeter regulator (241 kPa / 35 psi)*
14. O2 Flush
a. Flush valve
b. Pressure switch (used with the ventilator)
15. N2O balance regulator
16. N2O flow control valve
17. N2O flow tube(s)
18. Air secondary regulator (207 kPa / 30 psi)*
19. Air flow control valve
20. Air flow tube(s)
21. Supply connection for Venturi suction
a. O2 drive gas
b. Air drive gas
22. Vaporizer port valve
23. Vaporizer
24. Low-pressure relief valve (38 kPa / 5.5 psi)*
25. Auxiliary flowmeter (optional)
26. To ABS
27. To ACGO
28. Test port (primary regulator)
29. Test port (secondary/balance regulator)
* Approximate values
Pneumatic Connection
Filter
Direction of Flow
Check Valve
2-4
07/04 1009-0356-000
2 Theory of Operation
N2O
4
Air
4
O2
4
3
3
5
A
A
28
3
5
5
A
28
A
28
2
3
25
5
28
2
B
6
O2
4
B
6
1
2
B
6
1
1
21a
21b
7a
7b
13
8
14a
C
9
10
29
11
Link 25
12
15
29
16
14b
18
29
19
Link 25
17
26
20
27
22
22
24
D
22
AB.74.041
22
E
23
23
A. Cylinder Supplies
B. Pneumatic Manifold
C. Flowmeter Head
D. Vaporizer Manifold
E. ACGO Select Switch
Figure 2-1 • Pneumatic circuit
1009-0356-000
07/04
2-5
S/5 Aespire
Figure 2-2 shows the physical path that the gas takes.
AB.74.030
2.2.2 Physical
connections
O2
N2O
Air
N2O
O2
Air
Vent Drive
Figure 2-2 • Typical tubing connections - pictorial
2-6
07/04 1009-0356-000
2 Theory of Operation
2.2.3 Suction regulators
Pipeline vacuum
The suction regulator (shown in Figure 2-2) uses an external vacuum source.
Venturi Drive vacuum
The suction regulator (shown in Figure 2-3) uses an internal, venturi derived
vacuum source.
Drive gas (internally plumbed Air or O2) enters the Venturi Module (VM) at the
drive port (A). As the drive gas passes through the venturi module, a vacuum
is created at port B. The drive gas exits the venturi module at port C and is
exhausted outside the machine through the muffler (D).
The control port (E) on the venturi module responds to pneumatic signals from
the front panel switch on the Suction Control Module (SCM) to turn the venturi
vacuum drive gas on or off. The check valve (CV) helps prevent pressurization
of the suction circuitry if the exhaust is occluded or the venturi unit fails.
E
SCM
A
Venturi Drive Gas
Air or O2
B
CV
VM
Suction
AB.74.049
C
D
Figure 2-3 • Venturi suction
1009-0356-000
07/04
2-7
S/5 Aespire
2.2.4 System switch
The system switch has two positions: On and Standby.
In the Standby position
The switch:
• Turns off the ventilator (electrical).
• Stops O2 and Air to the flowhead (pneumatic).
• Without O2 pressure, the N2O balance regulator stops N2O.
In the On position
The switch:
• Turns on the ventilator (electrical).
• Supplies O2 and Air to the flowhead.
• With adequate O2 pressure, the N2O balance regulator supplies N2O.
System Switch
(electrical)
System Switch
(pneumatic)
O2 Out (Port 4)
Air Out (Port 4)
Wiring Harness
Air In (Port 3)
O2 In (Port 3)
(Rear View)
2-8
07/04 1009-0356-000
2 Theory of Operation
2.2.5 Flow control
Needle valves (one for each gas) adjust gas flows. Clockwise rotation
decreases flow. Counterclockwise increases flow. Mechanical stops set
minimum flows for all gases. The link system sets the maximum ratio of N 2O to
O2.
w WARNING
The Link 25 Proportioning System sets a minimum O2 concentration in
the fresh gas stream when only O2 and N2O are used. Use of an absorber
or another gas can still cause a hypoxic mixture to be delivered to the
patient, especially at low O2 flow rates.
Minimum flows
At minimum flow, two tabs prevent clockwise rotation of the valve stem. One
tab is on the stop collar; the other is on the valve body.
Valve Body
Stop Tabs
Valve
Stem
Stop
Collar
Link system
The chain link system helps assure an approximate minimum 1 to 3 ratio of
flow between O2 and N2O. When engaged (minimum O2 concentration), a tab
on the O2 knob is in contact with a tab on the O2 sprocket so that the O2 and
N2O knobs turn together:
• an increase in N2O flow causes an increase in O2 flow,
• a decrease in O2 flow causes a decrease in N2O flow.
Linkage Tabs
O2 Knob
1009-0356-000
07/04
2-9
S/5 Aespire
Higher concentrations of O2 are possible when the link system is not engaged:
either by reducing the N2O flow below the point of engagement or by
increasing O2 flow above the point of engagement.
When the N2O flow is below the point of engagement, increasing the N 2O flow
turns the O2 sprocket without changing the O2 flow. At the point of
engagement, the tab on the O2 sprocket makes contact with the tab on the O2
knob. Once the linkage is engaged, turning the N2O flow control
counterclockwise (increase in N2O flow) also turns the O2 knob
counterclockwise (increase in O2 flow) to maintain a nominal 25% minimum
O2 concentration.
Decreasing the N2O flow from the engagement point rotates the tab on the O2
sprocket away from the tab on the O2 knob. Increasing the O2 flow rotates the
knob tab away from the sprocket tab. Either action increases the O 2
concentration above 21%. Sufficiently decreasing O2 flow or increasing the
N2O flow brings the two tabs back into contact and engages the linkage.
The kick-in point is defined as the N2O flow at which the N2O valve becomes
engaged with the O2 valve flowing at 200 mL/min. This engagement point is
an arbitrary benchmark that assists in calibrating the proportioning system.
The position of the kick-in is set in the factory. During field calibration, you set
the O2 flow to 200 mL/min and the N2O flow to the kick-in flow (usually in the
range of 400 to 700 mL/min) and then install the sprockets with the O 2
knob/sprocket engaged.
Maximum flows
All gas flows in Canada require maximum flow stops. A maximum stop collar
on the body of the needle valve and a stop collar on the stem of the needle
valve set the maximum flow.
At maximum flow, a tab on the stop collar hits the tab on the maximum stop
collar and prevents you from turning the knob further counterclockwise. As
you decrease the flow, the valve stem moves toward the needle valve
assembly and clears the tab.
Maximum
Stop Collar
Valve Stem
Valve Body
Stop Collar
2-10
07/04 1009-0356-000
2 Theory of Operation
2.3 Flow through the breathing system
2.3.1 Overview of flow
paths
1009-0356-000
07/04
This section looks at three types of flow paths.
• Ventilation paths: How gas flows from the drive source (bag or bellows) to
and from the patient.
• Fresh gas paths: Fresh gas can flow from the machine interface directly to
the patient through the inspiratory check valve, or through the absorber into
the expiratory flow, or directly to an external circuit through the optional
auxiliary common gas outlet.
• Scavenged gas paths: APL or Pop-off.
2-11
S/5 Aespire
2.3.2 Manual ventilation
Manual inspiration
(Figure 2-4)
The Bag/Vent switch closes the ventilator path (B)..
Gas flows from the bag (1), through the absorber (2), into the breathing circuit
module, and through a unidirectional valve (inspiratory check valve) to the
patient (3).
During inspiration, fresh gas (FG) flows from the machine into the inspiratory
limb, upstream of the inspiratory check valve.
AP
FG
B
3
1
2
AB.82.026
3
1
B Bag/Vent switch to Bag
FG Fresh Gas
AP Airway Pressure
1 Flow to absorber
2 Flow from absorber
3 Inspiratory flow
Figure 2-4 • Gas flow during manual inspiration
2-12
07/04 1009-0356-000
2 Theory of Operation
Manual expiration
(Figure 2-5)
The Bag/Vent switch keeps the ventilator path closed (B).
Gas flows from the patient (4), through a unidirectional valve (expiratory
check valve), and into the bag (5).
During exhalation, fresh gas flows backwards through the absorber (FG) into
the expiratory limb, downstream of the expiratory check valve.
For machines that are plumbed to return sample gas to the breathing system,
the returned gas (SGR) enters the breathing system after the expiratory check
valve (Refer to section 4.13).
AP
FG
SGR
B
4
5
AB.82.027
4
FG
AP Airway Pressure
B Bag/Vent switch to Bag
FG Fresh Gas
SGR Sample Gas Return
4 Expiratory flow
5 Flow to bag
Figure 2-5 • Flow during manual expiration
1009-0356-000
07/04
2-13
S/5 Aespire
APL Valve
(Figure 2-6)
The APL valve sets a pressure limit for manual ventilation.
As you turn the APL knob, it puts more or less force on the APL disc and seat
(D/S). If the circuit pressure is too high (6), the disc and seat inside the
diaphragm opens and vents gas to the scavenging system (7).
D/S
6
7
AB.82.028
7
D/S APL disc and seat
6 APL flow
7 To scavenging
Figure 2-6 • Flow through the APL Valve
2-14
07/04 1009-0356-000
2 Theory of Operation
2.3.3 Mechanical ventilation
Mechanical inspiration
(Figure 2-7)
The Bag/Vent switch closes the manual path (V). Pilot pressure (P) closes the
exhalation valve.
Drive gas (D) pushes down on the bellows. Gas flows from the bellows (1),
through the absorber (2), and through a unidirectional valve (inspiratory
check valve) to the patient (3).
During inspiration, fresh gas flows into the inspiratory limb, upstream of the
inspiratory check valve.
AP
D
FG
1
3
3
V
2
P
AB.82.029
D
V Bag/Vent switch to Vent
P Pilot pressure
D Drive gas
FG Fresh Gas
AP Airway Pressure
1 Flow to absorber
2 Flow from absorber
3 Inspiratory flow
Figure 2-7 • Mechanical inspiration
1009-0356-000
07/04
2-15
S/5 Aespire
Mechanical expiration
(Figure 2-8)
Drive-gas flow stops and the exhalation valve opens. Exhaled gas flows from
the patient (4), through a unidirectional valve (expiratory check valve) and
into the bellows (5). Residual drive gas (D) flows out of the bellows to the
scavenging system (6).
If PEEP is selected, static pressure on the pilot port of the exhalation valve
sets the PEEP level.
During exhalation, fresh gas flows backwards through the absorber (FG) into
the expiratory limb, downstream of the expiratory check valve.
For machines that are plumbed to return sample gas to the breathing system,
the returned gas (SGR) enters the breathing system after the expiratory check
valve (Refer to section 4.13).
AP
FG
SGR
D
5
5
4
AB.82.030
6
FG
AP Airway Pressure
D Drive gas
FG Fresh Gas
SGR Sample Gas Return
4 Expiratory flow
5 Flow to bellows
6 To scavenging
Figure 2-8 • Flow through the APL Valve
2-16
07/04 1009-0356-000
2 Theory of Operation
Pop-off valve(Figure 2-9)
The pop-off valve limits the pressure inside the bellows to 2.5 cm H 2O above
the drive gas pressure. This normally occurs when the bellows reaches the top
of the housing at the end of exhalation.
Excess gas (7) vents to the scavenging system (6) through the pop-off valve
and the exhalation valve.
7
AB.82.031
6
6 Pop-off flow
7 To scavenging
Figure 2-9 • Flow through the pop-off valve
1009-0356-000
07/04
2-17
S/5 Aespire
2.3.4 Fresh gas and O2 flush flow
To ABS breathing system
(Figure 2-10)
Fresh gas (1) flows from the vaporizer manifold outlet to the ACGO Selector
Switch.
With the ACGO Selector Switch in the ABS position, fresh gas flow is
channeled to the breathing system.
AB.74.035
The output of the O2 Flush regulator (2) is channeled to the O2 Flush valve.
When activated, O2 flush flow joins the fresh gas flow in the ACGO Selector
Switch.
1
2
ABS
ACGO
Selector
Switch
Figure 2-10 • Fresh gas and O2 flush flow (to ABS)
2-18
07/04 1009-0356-000
2 Theory of Operation
Auxiliary Common Gas
Outlet(Figure 2-11)
Fresh gas (1) flows from the vaporizer manifold outlet to the ACGO Selector Switch.
With the ACGO Selector Switch in the ACGO position, fresh gas flow is channeled to the
ACGO outlet.
At the ACGO outlet, a small sample is diverted to the O 2 Sensor in the ABS for O2
monitoring.
The output of the O2 Flush regulator (2) is channeled to the O2 Flush valve.
When activated, O2 flush flow joins the fresh gas flow in the ACGO Selector Switch.
1
2
O2 Cell
O2 Sense
AB.74.034
ACGO
ACGO
Selector
Switch
Figure 2-11 • Fresh gas and O2 flush flow (to ACGO)
1009-0356-000
07/04
2-19
Notes
2-20
07/04 1009-0356-000
3
In this section
Checkout Procedure
3.1 Inspect the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
3.2 Pipeline and cylinder tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
3.3 Flow control, pressure relief, O2 supply alarm, and flush flow tests . . . . . . . . . . . . . . . . . .3-4
3.3.1 With O2 monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4
3.3.2 Without O2 monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3.3.3 Pressure relief tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.3.4 O2 supply alarm test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.3.5 Flush Flow Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
3.4 Vaporizer back pressure test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3.5 Low-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.5.1 Negative low-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
3.5.2 ISO or BSI standard low-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
3.6 Alarm tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
3.7 Breathing system tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
3.8 Auxiliary O2 flowmeter tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
3.9 Integrated Suction Regulator tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
3.10 Power failure test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
3.11 Electrical safety tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
w WARNINGS
After any repair or service of the Aespire system, complete all tests in this section.
Before you do the tests in this section:
• Complete all necessary calibrations and subassembly tests. Refer to the
individual procedures for a list of necessary calibrations.
• Completely reassemble the system.
If a test failure occurs, make appropriate repairs and test for correct operation.
1009-0356-000
07/04
3-1
S/5 Aespire
3.1 Inspect the system
Make sure that:
• The equipment is not damaged.
• All components are correctly attached.
• Pipeline gas supplies are connected and the pressures are correct.
• A supply of reserve O2 is provided and connected to the machine.
• Cylinder valves are closed on models with cylinder supplies.
w CAUTION
w WARNING
The upper shelf weight limit is 34 kg (75 lb).
Do not leave gas cylinder valves open if the pipeline supply is in use.
Cylinder supplies could be depleted, leaving an insufficient reserve
supply in case of pipeline failure.
AB.72.014
• Models with cylinder supplies have a cylinder wrench attached to the
system.
• Make sure the casters are not loose and the brake is set and prevents
movement.
3-2
07/04 1009-0356-000
3 Checkout Procedure
3.2 Pipeline and cylinder tests
w CAUTION
To prevent damage:
• Open the cylinder valves slowly.
• Do not force the flow controls.
If your system does not use cylinder supplies, do not do steps 2 and 3.
1. Disconnect the pipeline supplies and close all cylinder valves (if
equipped). If the pipeline and the cylinder pressure gauges are not at zero,
bleed all gasses from the system.
a.
b.
c.
d.
e.
f.
Connect an O2 supply.
Set the system switch to On.
Set the flow controls to mid range.
Make sure that all gauges but O2 are at zero.
Disconnect the O2 supply.
Make sure that the O2 gauge goes to zero. As pressure decreases,
alarms for low O2 supply pressure should occur.
2. Make sure that the cylinders are full:
a. Open each cylinder valve.
b. Make sure that each cylinder has sufficient pressure. If not, close the
applicable cylinder valve and install a full cylinder.
3. Test one cylinder at a time for high pressure leaks:
Set the system switch to Standby, which stops the O2 flow.
If equipped, turn the auxiliary O2 flow control fully clockwise (no flow).
If equipped, turn off venturi derived suction.
Open the cylinder.
Record the cylinder pressure.
Close the cylinder valve.
Record the cylinder pressure after one minute. If the pressure
decreases more than indicated below, there is a leak.
5000 kPa (725 psig) for ventilator drive gas.
690 kPa (100 psig) for non ventilator drive gas.
Install a new cylinder gasket and do this step again.
h. Repeat step 3 for all cylinders.
a.
b.
c.
d.
e.
f.
g.
w WARNING
Do not leave gas cylinder valves open if the pipeline supply is in use.
Cylinder supplies could be depleted, leaving an insufficient reserve
supply in case of pipeline failure.
4. Connect the pipeline supplies one at a time and ensure that the
corresponding gauge indicates pipeline pressure.
1009-0356-000
07/04
3-3
S/5 Aespire
3.3 Flow control, pressure relief, O2 supply alarm, and flush flow tests
If the system includes O2 monitoring, complete the flow control tests
in Section 3.3.1, “With O2 monitoring”.
If the system does not include O2 monitoring, complete the flow control
tests in Section 3.3.2, “Without O2 monitoring”.
3.3.1 With O2 monitoring
w WARNING
Nitrous oxide (N2O) flows through the system during this test. Use a safe
and approved procedure to collect and remove it.
1. Set up the gas scavenging system.
a. Connect the AGSS to a gas scavenging system.
b. Attach a patient circuit and plug the patient port.
c. Attach a bag to the bag port (or plug the bag port).
d. Set the Bag/Vent switch to Bag.
e. Adjust the APL valve to minimum.
2. Connect the pipeline supplies or slowly open the cylinder valves.
3. Turn all flow controls fully clockwise (minimum flow).
4. Set the ACGO selector switch to ABS.
5. Turn on the system.
6. Confirm that the O2 sensor measures 21% in room air and 100% in pure
O2. If not, calibrate the O2 sensor.
7. Make sure that:
• For a dual-tube O2 flowmeter,
the O2 flowtube shows 0.025 to 0.075 L/min.
• For a single-tube O2 flowmeter,
the O2 flowtube shows 0.175 to 0.225 L/min.
• The other flowtubes show no gas flow.
8. Set the flow controls to mid range of each flowtube and make sure that the
flowtube floats rotate and move smoothly.
Note
If the system does not include N2O, skip steps 9 and 10.
9. Check the Link proportioning system concentration (increasing N 2O flow).
Observe the following precautions:
a. Start with all valves at the minimum setting.
b. Adjust only the N2O flow control.
c. Increase the N2O flow as specified in the following table and make sure
the O2 concentration is in range.
3-4
07/04 1009-0356-000
3 Checkout Procedure
Note: Allow the O2 monitor to stabilize. At the lower flows, the O2
monitor may take up to 90 seconds to stabilize.
d. If you overshoot a setting, turn the O2 flow control clockwise until the
N2O flow decreases to the previous setting before continuing the test.
Set the N2O flow (L/min)
Measured O2
0.15 (dual flowtubes only)
21% minimum
0.5 (dual flowtubes only)
21% minimum
0.8
21% to 30%
1.0
21% to 30%
2.0
21% to 30%
6.0
21% to 30%
9.0
21% to 30%
10.Check the proportioning system concentration (decreasing O 2 flow).
Observe the following precautions:
a. Start with N2O valve at the maximum setting.
b. Adjust only the O2 flow control.
c. Decrease the O2 flow as specified in the table and make sure the O2
concentration is in the allowed range.
Note: Allow the O2 monitor to stabilize. At the lower flows, the O2
monitor may take up to 90 seconds to stabilize.
d. If you overshoot a setting, turn the N2O flow control counterclockwise
until the O2 flow increases to the previous setting before continuing the
test.
Set the O2 flow (L/min)
Measured O2
3.0
21% to 30%
2.0
21% to 30%
1.0
21% to 30%
0.3
21% to 30%
11. Check the linearity of the Air flow control.
• Turn the N2O flow control fully clockwise to minimum stop.
1009-0356-000
07/04
Set the O2 flow
(L/min)
Set the Air flow (L/min)
O2 monitor range
4.0
3.0
61% to 71%
3.5
6.0
45% to 55%
1.5
8.0
28% to 38%
3-5
S/5 Aespire
3.3.2 Without O2 monitoring
w WARNING
The following procedure will test for any significant malfunction of the
Link system but it will not confirm proper calibration. Periodic calibration
procedures using an accurate and properly calibrated O2 monitor must
be performed as recommended in the User’s Reference Manual, Part 2,
section 3 User Maintenance.
w
Nitrous oxide (N2O) flows through the system during this test. Use a safe
and approved procedure to collect and remove it.
1. Set up the gas scavenging system.
a. Connect the AGSS to a gas scavenging system.
b. Attach a patient circuit and plug the patient port.
c. Attach a bag to the bag port (or plug the bag port).
d. Set the Bag/Vent switch to Bag.
e. Adjust the APL valve to minimum.
2. Connect the pipeline supplies or slowly open the cylinder valves.
3. Turn all flow controls fully clockwise (minimum flow).
4. Set the ACGO selector switch to ABS.
5. Turn on the system.
6. Make sure that:
• For a dual-tube O2 flowmeter,
the O2 flowtube shows 0.025 to 0.075 L/min.
• For a single-tube O2 flowmeter,
the O2 flowtube shows 0.175 to 0.225 L/min.
• The other flowtubes show no gas flow.
7. Set the flow controls to mid range of each flowtube and make sure that the
flowtube floats rotate and move smoothly.
Note
If the system does not include N2O, skip steps 8 and 9.
8. Check the Link proportioning system (increasing N2O flow).
Observe the following precautions:
a. Start with all valves at the minimum setting.
b. Adjust only the N2O flow control.
c. Increase the N2O flow as specified in the following table and make sure
the O2 flow is as specified.
3-6
07/04 1009-0356-000
3 Checkout Procedure
d. If you overshoot a setting, turn the O2 flow control clockwise until the
N2O flow decreases to the previous setting before continuing the test.
Set the N2O flow control to
(L/min)
The O2 flow must be
greater than
(L/min):
0.8
0.2
2
0.5
4
1.0
10
2.5
9. Check the Link proportioning system (decreasing O2 flow).
Observe the following precautions:
a. Set the N2O flow to 9.0 L/min.
b. Set the O2 flow to 3 L/min or higher.
c. While reducing the O2 flow, set the N2O flow to the rates shown in the
table. The O2 flow must be greater than the minimum limits.
d. If you overshoot a setting, turn the N2O flow control counterclockwise
until the N2O flow increases to the previous setting before continuing
the test.
1009-0356-000
07/04
Set the N2O flow
(using the O2 flow control)
to (L/min)
The O2 flow must be
greater than
(L/min)
8.0
2.0
4.0
1.0
0.8
0.2
3-7
S/5 Aespire
3.3.3 Pressure relief
tests
To check the pressure relief valve (vaporizer manifold outlet).
1. Turn all flow controls fully clockwise (minimum flow).
2. Set the ACGO selector switch to ACGO.
3. Connect a gauge or a digital manometer to the
ACGO outlet using the positive pressure leak
test adapter.
4. Adjust the O2 flow to 0.5 L/min.
Test
Adapter
5. Verify that the test device reading stabilizes
within the following range:
31–60 kPa, 230–450 mm Hg, 4.5–8.5 psi.
6. Remove the test device and the adapter.
3.3.4 O2 supply alarm
test
1. Set all flow controls to 3 L/min.
2. Stop the O2 supply. (Disconnect the pipeline supply or close the cylinder
valve.)
3. Make sure that:
a. The low O2 supply alarm occurs.
b. The N2O (if equipped) and O2 flows stop. The O2 flow stops last.
c. Air (if equipped) flow continues.
d. Gas supply alarms occur on the ventilator if the ventilator uses O 2 as
the drive gas.
4. Turn all of the flow controls fully clockwise (minimum flow).
5. Reconnect the pipeline supplies.
3.3.5 Flush Flow Test
With Ventilator
1. Set the Bag/Vent switch to Vent.
2. Set the system switch to Standby.
3. Attach a patient circuit and plug the patient port.
4. Set the ACGO selector switch to ABS.
5. Ensure that the bellows is completely collapsed.
6. Measure the amount of time it takes to fill the bellows when the O 2 Flush
button is fully and continuously depressed.
7. Repeat the above measurement two more times (deflate bellows by
removing the plug from the patient port).
• The bellows should fill in 1.8 to 2.3 seconds.
3-8
07/04 1009-0356-000
3 Checkout Procedure
Without Ventilator:
1. Set the Bag/Vent switch to Bag.
2. Set the system switch to Standby.
3. Attach a patient circuit and plug the patient port.
4. Attach a 3-liter rebreathing bag on the bag arm or manual bag port
(if a 3-liter bag is not available, use the time specified for a1-liter bag times
the volume of the bag used).
5. Set the ACGO selector switch to ABS.
6. Ensure that the rebreathing bag is completely collapsed.
7. Adjust the APL valve to maximum.
8. Measure the amount of time it takes to fill the rebreathing bag when the
O2 Flush button is fully and continuously depressed.
Note: When the airway pressure gauge exceeds 10 cm H2O, the
rebreathing bag is full.
9. Repeat the above step two more times (deflate the rebreathing bag by
removing the plug from the patient port).
• A 3-liter bag should fill in 3.6 to 5.1 seconds.
• A 1-liter bag should fill in 1.2 to 1.7 seconds.
Possible Causes of Failure
• Large leak (if long filling time).
• Flush regulator setting (Section 6.5).
• Flush regulator cross-connection (if long filling time).
• ACGO selector valve inlet cross-connection (if short filling time).
From
O2 Supply
To
O2 Flush Valve
(and optional Auxiliary O2 Flowmeter
O2 Flush Output
ACGO
Selector
Switch
From
Vaporizer Manifold
1009-0356-000
07/04
3-9
S/5 Aespire
3.4 Vaporizer back pressure test
w WARNING
Anesthetic agent vapor comes out of the common gas outlet during this
test. Use a safe, approved procedure to remove and collect the agent.
1. Set up the gas scavenging system.
a. Connect the AGSS to a gas scavenging system.
b. Attach a patient circuit and plug the patient port.
c. Attach a bag to the bag port (or plug the bag port).
d. Set the Bag/Vent switch to Bag.
e. Adjust the APL valve to minimum.
2. Set the ACGO selector switch to ABS.
3. Set the system switch to On.
4. Set the O2 flow to 6 L/min.
5. Make sure that the O2 flow stays constant and the float moves freely.
6. Adjust the vaporizer concentration from 0 to 1% one click at a time. The O 2
flow must not decrease more than 1 L/min through the full range. If the O 2
flow decreases more than 1 L/min:
a. Install a different vaporizer and try this step again.
b. If the O2 flow decreases less than 1 L/min with a different vaporizer, the
malfunction is in the first vaporizer.
c. If the O2 flow also decreases more than 1 L/min with a different
vaporizer, the malfunction is in the Aespire system. Do not use the
system until it is serviced (repair vaporizer manifold port valve).
7. Complete steps 3 through 5 for each vaporizer and vaporizer position.
8. Set the system switch to Standby.
3-10
07/04 1009-0356-000
3 Checkout Procedure
3.5 Low-pressure leak test
Note
Perform either the “Negative low-pressure leak test” or the “ISO or BSI
standard low-pressure leak test. It is not necessary to perform both tests.
w WARNING
Do not use a system with a low-pressure leak. Anesthetic gas will go into
the atmosphere, not into the breathing circuit.
3.5.1 Negative lowpressure leak test
1. Test the leak test device:
a. Put your hand on the inlet of the leak test device. Push hard for a good
seal.
b. Squeeze the bulb to remove all air from the bulb.
c. If the bulb completely inflates in less than 60 seconds, replace the leak
test device.
max 280 cmH2O
2. Set the system switch to Standby.
3. Set the ACGO selector switch to ACGO.
ACGO Outlet
4. Turn off all vaporizers.
5. Test the anesthesia machine for low-pressure leaks:
a. Open the flow controls one and a half turns counterclockwise.
b. Connect the test device to the ACGO outlet.
c. Compress and release the bulb until it is empty.
d. The vacuum causes the floats to move. This is usual. If the bulb
completely inflates in 30 seconds or less, there is a leak in the lowpressure circuit.
6. Test each vaporizer for low-pressure leaks:
a.
b.
c.
d.
Set the vaporizer to 1%.
Repeat step 5.
Set the vaporizer to OFF.
Test the remaining vaporizers.
7. Disconnect the test device.
8. Turn all flow controls fully clockwise (minimum flow). Do not over tighten.
w WARNING
Agent mixtures from the low-pressure leak test stay in the system.
Always flush the system with O2 after the low-pressure leak test (1 L/min
for one minute).
Turn off all vaporizers at the end of the low-pressure leak test.
9. Flush the system with O2:
a.
b.
c.
d.
e.
1009-0356-000
07/04
Set the system switch to On.
Set the O2 flow to 1 L/min.
Continue the O2 flow for one minute.
Turn the O2 flow control fully clockwise (minimum flow).
Set the system switch to Standby.
3-11
S/5 Aespire
3.5.2 ISO or BSI standard
low-pressure leak test
w CAUTION
Do the positive pressure leak test at the ACGO outlet only.
1. Set the ACGO selector switch to ACGO.
2. Turn all flow controls fully clockwise (minimum flow).
3. Using the positive pressure leak test adapter, connect the ISO or BSI
specific leak test device to the ACGO outlet. Push the adapter into the
ACGO outlet throughout the test to get a good seal.
max 280 cmH2O
Leak Test Adapter
2
.5
2.5
.3
.2
0
L/min O2
1
1.5
.5
.4
3
.1
a
kP
3.5
Test Device
(Top View)
Test Device
(Front View)
4. Keep flowmeter of the test device vertical for accurate results.
5. Fully open the needle valve on the test device (counterclockwise).
w CAUTION
If the needle valve is not fully open, this test can damage the pressure
gauge on the test device.
6. Set the system switch to On.
3-12
07/04 1009-0356-000
3 Checkout Procedure
7. Open the O2 flow control and set a total flow of 0.4 L/min through the
flowmeter on the test device.
8. Make sure that the pressure gauge on the test device reads zero and that
all other flow controls are fully closed.
9. Close the needle valve on the test device until the test gauge reads:
ISO 5358
3 kPa
BSI 4272.3
20 kPa
10.If the flow through the test device is less than
0.35 L/min (ISO) or
0.3 L/min (BSI),
there is a low pressure leak in the anesthesia machine.
11. Fully open the needle valve on the test device to decrease the back
pressure.
12.Test each vaporizer for low-pressure leaks:
a. Set the vaporizer to 1%.
b. Repeat steps 7 through 10.
c. Turn the vaporizer OFF.
d. Test the remaining vaporizers.
13. Remove test tool and adapter.
w WARNING
Agent mixtures from the low-pressure leak test stay in the system.
Always flush the system with O2 after the low-pressure leak test (1 L/min
for one minute).
Turn all vaporizers OFF at the end of the low-pressure leak test.
14. Flush the system with O2:
a. Set the system switch to On.
b. Set the O2 flow to 1 L/min.
c. Continue the O2 flow for one minute.
d. Turn the O2 flow control fully clockwise (minimum flow).
e. Set the system switch to Standby.
1009-0356-000
07/04
3-13
S/5 Aespire
3.6 Alarm tests
1. Connect a test lung to the patient connection.
2. Set the Bag/Vent switch to Vent.
3. Set the system switch to On.
4. Set the controls:
• Ventilation Mode: Volume control (select from main menu)
• Ventilator:
- Tidal Vol: 400 ml
- Rate: 12
- I:E Ratio:1:2
- Plimit:40 cm H2O
- PEEP: OFF
• Anesthesia Machine:
- O2 flow: minimum flow
- All other gases: OFF
- ACGO selector switch to ABS
5. Push O2 Flush to fill the bellows.
6. Set the Bag/Vent switch to Bag and back to Vent.
7. Make sure that:
a. Mechanical ventilation starts.
b. A subatmospheric pressure alarm does not occur.
Note: With active gas scavenging, too much scavenging flow can cause
subatmospheric alarm.
c. The ventilator displays the correct data.
d. The bellows inflate and deflate during mechanical ventilation.
8. Set the O2 flow control to 5 L/min.
9. Make sure that:
a. The end expiratory pressure is approximately 0 cm H2O.
Note: Positive end expiratory pressure when PEEP is off may indicate
that the scavenging system is not removing enough gas.
b. The ventilator displays the correct data.
c. The bellows inflate and deflate during mechanical ventilation.
10. Test the low minute volume alarm:
a.
b.
c.
d.
e.
3-14
Go to the alarms menu.
Set the alarm limit for low minute volume to 6.0 L/min.
Make sure that a low minute volume alarm occurs.
Go to the alarms menu.
Set the low minute volume alarm to OFF.
07/04 1009-0356-000
3 Checkout Procedure
11. Test the high airway pressure alarm:
a. Set Plimit to less than the peak airway pressure.
b. Make sure that the high airway pressure alarm occurs.
c. Set Plimit to correct level.
12. Test the apnea and low airway pressure alarms:
a. Turn all flow controls fully clockwise.
b. Remove the test lung from the patient connection.
c. Other alarms such as low minute volume can occur.
d. Make sure that the low airway pressure and apnea alarms occur. The
apnea alarm occurs after 30 seconds.
13. Test the sustained airway pressure alarm:
a. Set the controls:
b.
c.
d.
e.
• APL valve — Closed (70)
• Bag/Vent switch — Bag
Mechanical ventilation stops when the Bag/Vent switch is set to Bag.
Occlude the bag port connector with a test plug.
Close the patient connection using the test plug located on the side of
the ABS and push the O2 Flush button.
Make sure that the sustained pressure alarm occurs after
approximately 15 seconds at the sustained pressure limit (6-30 cm
H2O varies with pressure limit).
14.Test the O2 monitor and alarms:
a. Remove the O2 sensor from the circuit module.
b. Make sure the sensor measures approximately 21% O2 in room air.
c. Set the low O2 alarm to 50%. Make sure a low O2 alarm occurs.
d. Set the low O2 alarm back to 21% and make sure that alarm cancels.
e. Put the O2 sensor back in the circuit.
f. Remove the test lung from the patient connection.
g. Set the High O2 alarm to 50%.
h. Push the flush button to fill the breathing system.
i. Set the O2 flow control to 2 L/min.
j. Make sure the high O2 alarm comes On.
k. Set the high O2 alarm back to 100% and make sure that alarm cancels.
l. After 2 minutes in pure O2, the O2 display reads approximately 100%.
m. Turn the O2 flow control fully clockwise (minimum flow).
15. Set the system switch to Standby.
1009-0356-000
07/04
3-15
S/5 Aespire
3.7 Breathing system tests
w WARNING
Objects in the breathing system can stop gas flow to the patient. This can
cause injury or death:
• Do not use a test plug that is small enough to fall into the breathing
system.
1. Verify that AGSS is operating. For systems that have a flow indicator on the
side, make sure that the flow indicator shows a flow in the green (normal)
region.
2. Zero the pressure gauge (Section 6.6.1).
Check Valves
3. Make sure that the check valves on the breathing circuit module work
correctly:
a. The Inspiratory check valve rises during inspiration and falls at the start
of expiration.
b. The Expiratory check valve rises during expiration and falls at the start
of inspiration.
Ventilator Bellows
4. Ventilator bellows test:
a.
b.
c.
d.
Set the system switch to Standby.
Set the Bag/Vent switch to Ventilator.
Set all flow controls to minimum.
Close the breathing circuit at the patient connection. Use the test plug
located on the side of the ABS.
e. Push the O2 flush button to fill the bellows.
f. The pressure must not increase to more than 15 cm H2O on the
pressure gauge.
g. If the bellows falls more than 100 mL/min (top of indicator), it has a
leak.
Service Mode Tests
5. Enter the Service Mode: Push and hold the adjustment knob on the
ventilator’s display and set the system switch to On.
a. Select and confirm “Service Modes.”
b. Follow the menu structure outline below to reach the adjustment for the
inspiratory flow valve. Select and confirm at each step.
• “Diagnostics Tests/Tools”
• “Breathing System Leak Test”
c. Follow the instructions on the screen.
d. The leak rate should be less than 200 mL/min.
For machines with a single-tube O2 flowmeter, the pressure reading
should reach 30 cm H2O at minimum flows greater than 200 mL/min.
Note: If test fails, see Section 7.2, “Breathing System Leak Test Guide”.
3-16
07/04 1009-0356-000
3 Checkout Procedure
Bag Circuit
6. Test the Bag circuit for leaks:
a. Set the system switch to On.
b. Set the Bag/Ventilator switch to Bag.
c. Plug the Bag port (use your hand or the approved test plug).
d. Close the APL valve (70 cm H2O).
e. Set the O2 flow to 0.25 L/min.
f. Close the patient connection (using a hand or test plug on the side of
the breathing system) and pressurize the bag circuit with the O 2 flush
button to approximately 30 cm H2O.
g. Release the flush button. The pressure must not decrease. A pressure
decrease large enough to see on the gauge indicates an unacceptable
leak.
Note: If test fails, see Section 7.2, “Breathing System Leak Test Guide”.
APL Valve
7. Test the APL valve:
a. Fully close the APL valve (70 cm H2O).
b. Set the total fresh gas flow to approximately 3 L/min and make sure
that the value on the inspiratory pressure gauge does not exceed 85 cm
H2O. Some pressure fluctuation is normal.
c. Fully open the APL valve (to the MIN position).
d. Set O2 flow to 3 L/min. Turn any other gases off.
e. Make sure that the value on the inspiratory pressure gauge is less than
approximately 5 cm H2O.
f. Push the O2 flush button. Make sure that the value on the inspiratory
pressure gauge stays less than 10 cm H2O.
g. Set the O2 flow to minimum and make sure that the value on the
inspiratory pressure gauge does not decrease below 0 cm H 2O.
8. Remove your hand or the test plug from the patient connection.
9. Set the System switch to Standby.
w WARNING
1009-0356-000
07/04
Make sure that there are no test plugs or other objects caught in
the breathing system.
3-17
S/5 Aespire
3.8 Auxiliary O2 flowmeter tests
1. Open the O2 cylinder valve or connect an O2 pipeline.
2. Rotate the flow control clockwise (decrease) to shut off the flow. The ball
should rest at the bottom of the flow tube and not move.
3. Rotate the flow control counterclockwise (increase). The ball should rise
immediately after rotation is begun. It should rise smoothly and steadily
with continued counterclockwise rotation. When a desired flow is set, the
ball should maintain in a steady position.
4. Occlude the auxiliary O2 outlet. The ball should rest at the bottom of the
flow tube and not move. A ball that does not rest at the bottom of the flow
tube indicates a leak and requires service.
5. Rotate the flow control clockwise to shut off the flow.
3.9 Integrated Suction Regulator tests
The gauge needle should come to rest within the zero range bracket when no
suction is being supplied. Gauges which do not comply may be out of
calibration.
1. Adjust the regulator setting to minimum.
2. Turn the mode selector to I (On).
3. Ensure the gauge remains less than 200 mmHg (26 kPa, 0.26 Bar).
4. Occlude the inlet.
5. Ensure the gauge remains less than 200 mmHg (26 kPa, 0.26 Bar).
6. Adjust the regulator in an increasing vacuum level.
7. The gauge should rise after rotation has begun. The gauge should rise with
continued rotation of the regulator adjustment.
8. Adjust the regulator setting to minimum.
9. Turn the Mode selector to O (Off).
3-18
07/04 1009-0356-000
3 Checkout Procedure
3.10 Power failure test
1. Connect the power cord to a mains outlet. The mains indicator on the
display comes on when AC Power is connected.
If the indicator is not on, the display assembly is not receiving AC power.
• Verify AC power to the machine (reset circuit breaker).
• Check fuses in display assembly’s inlet module.
• Check fuses in machine’s AC inlet assembly.
2. Set the system switch to On.
3. Unplug the power cord with the system turned on.
4. Make sure that the power failure alarm comes on.
5. Make sure the following message is on the ventilator display:
• “On Battery - Power OK?”
6. Connect the power cable again.
7. Make sure the alarm cancels.
3.11 Electrical safety tests
Make sure the system is completely assembled and all accessory devices are
connected to electrical outlets.
1. Connect an approved test device (e.g. UL, CSA, or AAMI) and verify that the
leakage current is less than:
Voltage
Max. Leakage Current
120/100 Vac
300 µAmps
220/240 Vac
500 µAmps
2. Make sure that the resistance to ground is less than 0.2Ω between an
exposed metal surface and the ground pin on the power cord.
1009-0356-000
07/04
3-19
Notes
3-20
07/04 1009-0356-000
4
In this section
Repair Procedures
This section covers the repair and replacement procedures for components of the
Aespire anesthesia machine.
4.1 Servicing the ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
4.2 How to bleed gas pressure from the machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.3 How to remove the rear panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.3.1 To remove the rear upper panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.3.2 To remove the lower access panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
4.4 How to remove the tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
4.5 Replace pipeline inlet filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
4.5.1 Replace pipeline inlet check valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
4.6 Change drive gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
4.7 Service the cylinder supply modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.7.1 Tightening procedure for high-pressure tube fittings . . . . . . . . . . . . . . . . . . . . . . . .4-8
4.7.2 Replace primary regulator module (complete replacement) . . . . . . . . . . . . . . . . .4-8
4.7.3 Replace cylinder inlet filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
4.7.4 Replace cylinder check valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
4.7.5 Replace 3rd-gas cylinder supply module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.8 Replace system switch assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
4.9 Service the flowmeter module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.9.1 Remove front flowmeter panel shield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.9.2 Remove flowtubes for cleaning or replacement . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
4.9.3 Remove complete flowmeter head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
4.9.4 Replace flowmeter modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
4.9.5 Replace flowmeter frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
4.9.6 Replace O2 supply switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
4.9.7 Checkout procedure for O2 supply switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
4.9.8 Replace secondary regulator manifold or balance regulator manifold . . . . . . . 4-22
4.9.9 Replace O2 or N2O needle valves (on machines with N2O) . . . . . . . . . . . . . . . . . 4-23
4.9.10 Replace an Air needle valve on all machines or an O2 needle valve on machines
without N2O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
1009-0356-000
07/04
4-1
S/5 Aespire
4.10 Service vaporizer manifold parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
4.10.1 Repair manifold port valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.10.2 Checkout procedure for manifold port valve . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.10.3 Replace vaporizer manifold check valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.10.4 Replace vaporizer pressure relief valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.10.5 Replace vaporizer manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.11 Replace ACGO selector switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-26
4-27
4-28
4-30
4-31
4-32
4.12 Clean or replace ACGO port flapper valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
4.13 Reconfigure sample gas return line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-35
4.14 Replace the APL valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
4.15 Replace the bag support arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37
4.15.1 Servicing the bag support arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.15.2 Replace friction pad in lower bag arm assembly . . . . . . . . . . . . . . . . . . . . . . . .
4.15.3 Replace bag port housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.16 Replace auxiliary O2 flowmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-38
4-39
4-40
4-41
4.17 Replace the suction control module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-42
4.17.1 Front panel method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-42
4.17.2 Rear panel method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
4.18 Replace ABS breathing system components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
4.18.1 Replace Bag/Vent switch assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
4.18.2 Replace bellows base latch assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
4.19 Replace casters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-46
4.20 Replace task light and switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
4.20.1 To replace the task-light switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
4.20.2 To replace the task-light circuit board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
4.21 Replace the display arm or display cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-48
4.21.1 Cable tie installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.21.2 Removing the display arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.21.3 Replacing a display cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.21.4 Installing the long arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.21.5 Installing the short arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.22 Replace display and cables in ProTIVA machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-2
4-48
4-49
4-49
4-50
4-51
4-52
07/04 1009-0356-000
4 Repair Procedures
w WARNING
To prevent fires:
• Use lubricants approved for anesthesia or O2 equipment, such as Krytox.
• Do not use lubricants that contain oil or grease; they burn or explode in high
O2 concentrations.
• All covers used on the system must be made from antistatic (conductive)
materials. Static electricity can cause fires.
w
Obey infection control and safety procedures. Used equipment may contain
blood and body fluids.
w
A movable part or a removable component may present a pinch or a crush
hazard. Use care when moving or replacing system parts and components.
w
Some internal parts have sharp edges and can cause cuts or abrasions. Use care
when servicing internal components.
w
After repairs are completed, always perform the checkout procedure. Refer to
Section 3 of this manual.
4.1 Servicing the ventilator
The Aespire anesthesia machine is configured with the 7100 Ventilator.
Service information for this ventilator is provided in a separate manual as detailed in
Section 1.2.2.
The 7100 Ventilator comprises the following components:
• The Display/Control module.
• The Serial Isolation and Connector Board, located in the upper structure.
• The Vent Engine, located in the rear portion of the breathing system.
• The Ventilator Monitoring Board, located under the tabletop.
Common machine components such as switches and sensors which provide input to
the ventilator are covered in this manual.
1009-0356-000
07/04
4-3
S/5 Aespire
4.2 How to bleed gas pressure from the machine
Before disconnecting pneumatic fittings, bleed all gas pressure from the machine.
1. Set the system switch to On.
2. Close all cylinder valves and disconnect all pipeline supplies from the source.
Note: If the machine includes N2O, do not disconnect the O2 pipeline.
If pipeline O2 is not available, open the O2 cylinder valve.
3. Turn the flow controls for all gases (except O2) at least one turn counterclockwise.
4. Ensure that all cylinder and pipeline gauges read zero before proceeding.
• For machines with N2O, disconnect the O2 pipeline supply from the source
(or close the O2 cylinder valve).
• Press the O2 flush button to bleed O2 from the system.
5. Set the system switch to Standby.
4.3 How to remove the rear panels
You must remove the rear upper panel to repair or replace many of the machine’s
components. To access the 3rd cylinder supply (if equipped) you must remove the
lower access panels.
4.3.1 To remove
the rear upper
panel
1. Bleed all gas pressure from the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. Disconnect all electrical cables.
4. To remove the rear panel, fully loosen the three captive screws that hold the panel in
place. Remove the panel.
• If the machine includes integrated suction, disconnect the two tube fittings from
the overflow safety trap manifold.
• If the machine includes electrical outlets, lower the panel and place it so that it
does not stress the power cable.
4.3.2 To remove
the lower access
panels
1. Bleed all gas pressure from the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. If present, remove the inboard cylinders.
4. Remove (two screws) the small upper access panel to access the 3rd cylinder
regulator (N2O) test port.
5. Remove (six screws) the large lower access panel to access the 3rd cylinder
regulator.
4-4
07/04 1009-0356-000
4 Repair Procedures
4.4 How to remove the tabletop
The tabletop is held in place with five captive screws along the periphery of the pan
assembly (accessed from below the rim of the tabletop).
• One screw (A) is in a deep recess at the right-rear corner of the tabletop.
• Two screws (B) are at the front of the tabletop: one screw is at the right corner of the
tabletop, one is near the O2 Flush button.
• To access the remaining two screws (C), you must remove the ABS: one screw is at the
left corner of the tabletop, one is near the APL Valve.
A
C
C
B
B
1009-0356-000
07/04
4-5
S/5 Aespire
4.5 Replace pipeline inlet filter
1. Remove the pipeline inlet fitting.
2. Pull the pipeline inlet filter out of the fitting. The o-ring should come out with the filter.
3. Install the new pipeline inlet filter in the pipeline inlet fitting. The new filter comes
with an o-ring.
4.5.1 Replace
pipeline inlet
check valve
1. Remove the rear panel (Section 4.3).
2. Remove the pipeline inlet fitting.
3. The Air and O2 pipeline manifolds include a drive gas connection at the back of the
manifold. Remove the drive gas tube or plug to access the check valve.
4. From the back of the pipeline manifold, use a thin tool to push out the check valve.
(For an N2O manifold, you will have to carefully apply pressure at the outlet of the
manifold — with a syringe for example — to gently force the check valve out of the
manifold).
5. Push the new check valve into the opening, using the same thin tool. The new check
valve includes an o-ring — orient it toward the pipeline inlet. Note: Make sure to push
the new check valve all the way back into the opening until it bottoms out on the
shoulder.
6. Install the pipeline inlet fitting.
7. Perform the checkout procedure (Section 3).
4-6
07/04 1009-0356-000
4 Repair Procedures
4.6 Change drive gas
w CAUTION
If you change the drive gas, you must also change the drive gas selection on the
ventilator service setup screen. Refer to Section 4 of the ventilator Technical
Reference manual.
• If the drive gas selection and the actual drive gas do not agree, volumes will
not be correct.
The ventilator will alarm with the message “Low Drive Gas Press” if the selected
drive gas pressure, either O2 or Air, is lost.
1. Remove the rear panel (Section 4.3).
Note:
The O2 and Air pipeline manifolds have a drive gas connection at the back. The
connection not in use is plugged.
2. Remove the plug from the new connection.
3. Disconnect the drive gas hose from the present connection.
4. Install the plug in this connection (pull on the plug to ensure that it is locked into the
fitting).
5. Reroute the drive gas hose so that it does not cause kinks in other tubing.
6. Connect the drive gas hose to the new connection (pull on the hose connector to
ensure that it is locked into the fitting).
7. Do a high-pressure leak test (Section 3.2).
8. Enter the service mode and select the correct drive gas.
9. Test the primary regulator. Verify that it functions within specifications now that it will
be supplying drive gas to the ventilator (Section 6.1).
10. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-7
S/5 Aespire
4.7 Service the cylinder supply modules
w WARNING
4.7.1 Tightening
procedure for
high-pressure
tube fittings
Be careful not to expose internal components to grease or oil (except Krytox or
equivalent).
The cylinder pressure gauge is connected to the cylinder supply through a copper tube
with fittings at both ends. Use the following tightening procedure whenever you are
replacing a cylinder supply or a cylinder pressure gauge.
1. Insert the tubing into the fitting until the ferrule seats in the fitting.
2. Tighten the nut by hand.
3. Continue tightening the nut with a wrench until it reaches the original position (about
1/4 turn). You will feel an increase in resistance at the original position.
4. After reaching the original position, tighten the nut just slightly.
Note
4.7.2 Replace
primary regulator
module (complete
replacement)
If you are installing a new tube that has not been tightened before, tighten the nut with
a wrench an additional 3/4 of a turn after the nut is finger tight.
1. Bleed all gas pressure from the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. Remove the rear panel (Section 4.3).
4. Disconnect the high-pressure cylinder gauge fitting.
5. Disconnect the output tube fitting.
6. Remove the three mounting screws and
lockwashers.
7. To reassemble, perform the previous
steps in reverse order.
• Tighten the high-pressure fitting as
detailed in Section 4.7.1
• Pull on the cylinder output fitting to
ensure it is locked in place.
8. Check the output of the regulator
BEFORE you install the rear panel.
Adjust if necessary (Section 6.1).
9. Perform the checkout procedure
(Section 3).
4-8
07/04 1009-0356-000
4 Repair Procedures
4.7.3 Replace
cylinder inlet filter
1. Open the cylinder yokes.
2. Remove the inlet adapter from the cylinder yoke, using a 4 mm hex wrench.
Note: A brass retaining ring keeps the filter inside the inlet adapter.
3. Thread a 6-mm screw (two turns only) into the brass retaining ring and pull it out.
w CAUTION
Be careful not to crush the filter. Do not thread in the screw more than two full
turns.
4. Remove the filter.
5. Install the new filter and brass retaining ring.
6. Install the inlet adapter in the cylinder yoke.
7. Perform the checkout procedure (Section 3).
4.7.4 Replace
cylinder check
valve
1009-0356-000
07/04
The cylinder check valve is not a replaceable item. If the check valve is defective, you
must replace the complete cylinder supply module.
4-9
S/5 Aespire
4.7.5 Replace
3rd-gas cylinder
supply module
1. Remove the lower rear cover (Section 4.3.2).
2. Disconnect the high-pressure cylinder gauge fitting.
3. Disconnect the output tube fitting.
4. Remove the three mounting screws and lockwashers.
Outlet tube fitting
Pressure gauge
fitting
5. To reassemble, perform the previous steps in reverse order.
• Tighten the high-pressure fitting as detailed in Section 4.7.1
• Pull on the cylinder output fitting to ensure it is locked in place.
6. Check the output of the regulator BEFORE you install the rear cover. Adjust if
necessary (Section 6.1).
7. Perform the checkout procedure (Section 3).
4-10
07/04 1009-0356-000
4 Repair Procedures
4.8 Replace system switch assembly
1. Bleed all gas pressure from the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. Remove the tabletop (Section 4.4).
4. Remove the gauge panel mounting screws and move the panel forward to access the
system switch.
System Switch
(electrical)
Gauge panel
5. Disconnect the wires from the electrical switch.
6. Back out the system switch mounting screws just enough to allow the knob collar to
be released.
7. While holding the switch assembly, push in the knob and turn it counterclockwise.
8. Pull the knob and collar out from the front and remove the switch assembly.
1009-0356-000
07/04
4-11
S/5 Aespire
9. Install the replacement switch assembly:
a. Loosen the two outside screws on the electrical module.
b. Insert the wires in the electrical module and tighten the screws.
c. Pull the wires on the electrical module to ensure that there is a good connection.
d. Turn back the system switch mounting screws until their tips recede.
e. Orient the switch assembly with the Air fittings toward the right and the O 2
fittings toward the left.
f. Install the switch assembly through the gauge panel.
g. Push the knob collar in with the indicator up and turn it clockwise until it locks.
h. Tighten the mounting screws. Make sure that the top edge of the switch
assembly is parallel to the top edge of the gauge panel.
i. Transfer the tubing from the old system switch to the new system switch on the
pneumatic module (pull on the tubing to ensure that it is locked into the
module).
O2 Out (Port 4)
Air Out (Port 4)
Wiring Harness
Air In (Port 3)
O2 In (Port 3)
(Rear View)
10. Test the replacement switch assembly:
a. Connect Air and O2 supplies.
b. Connect the power cable to an electrical outlet.
c. Set the system switch to On.
d. Increase the O2 and Air flow. Make sure that gas flows.
e. Make sure that you do not feel or hear any leaks.
f. Make sure that the display comes On.
g. Set the system switch to Standby.
h. Make sure all gas flow stops and the display turns Off.
11. Reinstall the gauge panel and the tabletop.
12. Perform the checkout procedure (Section 3).
4-12
07/04 1009-0356-000
4 Repair Procedures
4.9 Service the flowmeter module
4.9.1 Remove
front flowmeter
panel shield
1. Bleed all gas pressure from the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. The flowmeter panel is held in place with two latching tabs at the right side. To
remove the panel, release each latch by pushing it toward the center of the panel
with a thin rod (3-mm hex wrench) through the access hole in the shroud.
4. Remove the panel.
5. To reinstall the panel, engage the retaining tabs on the left side and press the right
side against the shroud to latch it in place.
4.9.2 Remove
flowtubes for
cleaning or
replacement
1009-0356-000
07/04
1. Remove the front flowmeter panel shield (Section 4.9.1).
2. To remove a flowtube assembly, push up on the tube just enough to clear the bottom
seal, pull out from the bottom until the tube clears the flowtube module, then pull
down slowly to release it from the module.
4-13
S/5 Aespire
w WARNING
Floats are calibrated to a specific tube. Keep each float with its tube. Replace
tube and floats together. Interchanging floats can cause incorrect readings.
Disassemble the flowtube assemblies only when service is required. Excessive
cleaning can remove the antistatic coating from inside the tube. Damage to the
float requires replacement of the entire flowtube.
3. Disassemble the flowtube assembly.
Spring
O-ring
Float Stop
Float stops are indexed to
flowtubes and to flowtube
modules by gas.
Tube
Tabs and slots at both
connections must be
aligned to install
flowtubes.
Float
Filter
O-ring
4. Completely clean, rinse, and dry the flowtube. Use hospital grade alcohol and a
flowtube brush.
5. Replace stops, filter and o-rings, as necessary. Lightly coat all o-rings with Krytox. Be
careful to not get any Krytox on the bottom of small flowtube float stops.
6. Reassemble the flowtube assembly.
7. Insert the flowtube, spring side first, into the top of the module with the scale
oriented forward.
8. Push up and slide the bottom of the flowtube into place on the bottom o-ring. It may
be necessary to rotate the tube to engage the index tabs.
Note: Be sure o-rings are inserted completely into the collar.
9. Push down on the tube to seat the bottom o-ring.
10. Reinstall the front flowmeter panel shield.
11. Perform the checkout procedure (Section 3).
4-14
07/04 1009-0356-000
4 Repair Procedures
4.9.3 Remove
complete
flowmeter head
1. Remove the rear panel (Section 4.3).
2. Disconnect the tubing at the rear of each gas module. The following example is a
back view of the flowmeter head.
Fresh gas out
N2O in
O2 in
Air in
O2 supply
switch
Fresh gas out
(for single-tube flowhead)
3. Disconnect the O2 supply switch harness. Note position of switch connections so
that you can reassemble correctly later.
Open
4. Remove the front flowmeter panel shield (Section 4.9.1).
5. Remove the mounting screw from each regulator manifold.
6. Pull the flowhead forward.
7. To reinstall, perform the previous steps in reverse order (pull on the tubing to ensure
it is locked into the fittings).
8. Check for proper alignment of the front flowmeter panel. If any of the needle valve
knobs rub against the flowmeter panel, reposition the flowhead to allow for proper
clearance.
9. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-15
S/5 Aespire
4.9.4 Replace
flowmeter
modules
1. Remove the complete flowmeter head (Section 4.9.3).
2. Refer to the following illustrations. Note that these illustrations show ANSI flowmeter
module positions. The order is reversed in ISO machines.
O2
O2/Air modules
N20
Air
O2/Air flowmeter modules are connected at the bottom with a long screw (A)
and nut (B) that is recessed. To remove, retain nut while loosening screw.
B
Note
A
The flowmeter modules are held together at the top with a u-clip. To separate the
modules, pivot the modules (front to back) 45 degree. The u-clip will disengage and
allow the modules to separate.
The flowmeter modules are interconnected at the top by a spacer tube. The o-ring on
the spacer tube makes a leak-tight seal.
U-Clip
O-ring
Top of flowmeter module
Spacer
Tube
4-16
07/04 1009-0356-000
4 Repair Procedures
Single-tube flowhead
The outlet fitting for a single-tube flowhead is not an integral component of the O 2
flowmeter. The outlet fitting is a separate component that includes an o-ring seal and is
held in place with two u-clips.
Small
U-Clip
O-ring
Large
U-Clip
3. To remove the Air flowmeter module:
a. Remove the screw located on the side of the Air flowmeter module.
b. Hold the flowmeter module with the flowtubes facing you.
c. Grasp the outer modules at the bottom of the regulator manifold and push the
left module away from you until the u-clip pops off and the module separates
from the other assemblies.
d. Pull the modules sideways to separate them at the top.
Save the u-clip, spacer tube, and the o-ring for reassembly.
4. To remove the O2 or N2O flowmeter module:
a. Set the O2 and N2O needle valves to their maximum position (counterclockwise).
b. Loosen the set screws on the N2O knob, then remove the knob.
c. Loosen the set screws on the N2O sprocket and the O2 knob.
d. To remove, grasp the O2 knob/sprocket, N2O sprocket, and chain as an
assembly. Remove as an assembly.
e. Remove the spacer from the N2O needle valve spindle.
f. Remove the pilot tube going to the balance regulator.
1009-0356-000
07/04
4-17
S/5 Aespire
Remove
N2O
O2
g. The O2 and N2O flowmeter modules are held together by a single screw. Remove
the screw located on the side of the O2 flowmeter module.
h. Hold the flowmeter modules with the flowtubes facing you.
i. Grasp the modules at the bottom of the regulator manifolds and push the left
module away from you until the N2O module separates from the O2 module.
j. Pull the modules sideways to separate them at the top.
Save the u-clip, spacer tube, and the o-ring for reassembly.
5. To reassemble the flowmeter modules, perform the previous steps in reverse order.
Note: The u-clips must be reinstalled with the barbed leg to the left as viewed from
the front.
U-Clip
O-ring
Top of flowmeter module
Spacer
Tube
6. Install the screw through the O2 flowmeter module to locks the O2 and N2O
flowmeter modules together.
7. Reattach the pilot tube that goes to the balance regulator.
8. Install the flowhead into the machine. Reconnect the tubing and the O 2 supply
switch harness.
9. Confirm needle valve calibration (Section 6.3).
10. Install the spacer on the N2O needle valve stem.
11. Install the chain on the O2 knob/sprocket assembly and the N2O sprocket.
12. Install the chain and sprockets on the needle valve stems as an assembly. Do not
tighten the set screws.
4-18
07/04 1009-0356-000
4 Repair Procedures
Sprocket/Chain
Assembly
13. Install the N2O knob. Snug one set screw to hold the knob in place.
14. Perform the link system calibration (Section 6.4).
15. Install the flowmeter panel shield.
16. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-19
S/5 Aespire
4.9.5 Replace
flowmeter frame
1. Remove the front flowmeter panel shield (Section 4.9.1).
2. Remove the complete flowmeter head (Section 4.9.3).
3. Separate the flowmeter modules as required (Section 4.9.4).
4. Remove the flowtubes (Section 4.9.2). Keep all the parts for reassembly.
5. Remove the gas identification panel by removing the two screws at the back of the
frame. Keep all the parts for reassembly.
6. Remove the flowmeter frame by loosening the four mounting screws at the back of
the regulator manifold.
Note: There is a retainer in each screw location that keeps each screw within the
manifold.
7. To reassemble, perform the previous steps in reverse order.
8. If replacing O2 or N2O frames, perform the link system calibration (Section 6.4).
9. Perform the checkout procedure (Section 3).
4-20
07/04 1009-0356-000
4 Repair Procedures
4.9.6 Replace O2
supply switch
The O2 supply switch is located on the O2 flowmeter’s regulator module.
1. Remove the upper rear panel (Section 4.3).
2. Remove the two mounting screws from the O2 supply switch.
Adjustment
screw
4-mm hex
3. Pull the O2 supply switch out of the regulator manifold.
4. Install the replacement O2 supply switch.
5. Tighten the screws.
6. Disconnect the leads from the old switch and reconnect them to the new switch.
7. Adjust the alarm threshold for the new O2 supply switch, as explained in the
checkout procedure below (Section 4.9.7).
8. Replace the rear panel.
9. Perform the checkout procedure (Section 3).
4.9.7 Checkout
procedure for O2
supply switch
1. Remove the upper rear panel (Section 4.3).
2. Attach a gauge to the O2 primary regulator test port. (On pipeline only machines,
attach the gauge to a 6-mm O2 port).
3. Adjust the O2 flow control to minimum flow (clockwise).
4. Install an O2 cylinder and open the cylinder valve (for pipeline only, connect O 2
pipeline source).
5. Turn the system on.
6. Close the cylinder valve (disconnect pipeline from source) and watch the test gauge
as the O2 pressure bleeds down slowly.
Note: The “No O2 pressure” alarm should occur between descending pressure of
221–193 kPa (32–28 psi).
7. If adjustment is required, set the adjustment screw so that the “No O 2 pressure”
alarm occurs at 207 ±7 kPa (30 ±1 psi).
8. Disconnect the gauge and plug the test port
9. To reassemble, perform the previous steps in reverse order.
10. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-21
S/5 Aespire
4.9.8 Replace
secondary
regulator manifold
or balance
regulator manifold
1. Remove the front flowmeter panel shield (Section 4.9.1).
2. Remove the complete flowmeter head (Section 4.9.3).
3. Separate the flowmeter modules (Section 4.9.4).
4. Remove the flowmeter frame from the regulator manifold by removing the four
screws at the rear of the regulator manifold (no need to remove flowtubes).
5. Remove the needle valve from the regulator manifold:
a. Unscrew the complete assembly together (stop collar, needle valve).
b. Replace the o-ring if necessary.
6. Screw the needle valve into the new regulator manifold.
7. Remove the plugs and balance regulator elbow fitting (and O 2 supply switch if an O2
module) from the old regulator manifold.
8. Install the plugs and balance regulator elbow fitting (and O 2 supply switch if an O2
module) into the new regulator manifold (pull on the plugs and fittings to ensure that
they are locked into the manifold).
9. Reinstall the flowmeter frame to the regulator manifold.
10. Reinstall all the flowmeter modules to the flowmeter head.
11. Reinstall the flowmeter head (Section 4.9.4).
12. Do the necessary calibrations (Section 6).
Necessary calibrations
Section
Secondary Regulator
6.2
O2 minimum flow
6.3.1
Maximum flow
6.3.4
Link system
6.4
13. Reinstall the front flowmeter panel shield.
14. Perform the checkout procedure (Section 3).
4-22
07/04 1009-0356-000
4 Repair Procedures
4.9.9 Replace O2
or N2O needle
valves (on
machines with
N2O)
For machines without N2O, refer to Section 4.9.10 for replacing the O2 needle valve.
1. Bleed all gas pressure from the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. Remove the front flowmeter panel shield (Section 4.9.1).
4. Set the O2 and N2O needle valves to their minimum position.
5. Loosen the set screws on the N2O knob, then remove the knob.
6. Loosen the set screws on the N2O sprocket and the O2 knob.
7. To remove, grasp the O2 knob/sprocket, N2O sprocket, and chain as an assembly.
Remove as an assembly.
8. Remove the spacer from the N2O needle valve spindle.
9. Loosen the set screws on the needle valve stop collar for the needle valve that is
being replaced.
10. Remove the stop collar.
11. To remove the needle valve from the flowmeter block, turn the needle valve
counterclockwise with a 16-mm wrench.
12. To install the new needle valve, turn it clockwise and tighten it with the wrench.
Note: Be sure the o-ring is properly located on the tip of the needle valve.
w WARNING
The O2 and N2O needle valves are not the same. Patient injury can result if the
wrong needle valve is installed in the flowmeter block. You can identify the N2O
needle valve by a groove located just below the top brass hex.
O2 needle valve has no groove
N2O needle valve has a groove
13. Install the stop collar on the new needle valve. Do not tighten the set screws.
14. Perform the needle valve calibration (Section 6.3).
1009-0356-000
07/04
4-23
S/5 Aespire
15. After calibrating the needle valve, put the spacer the N 2O needle valve spindle.
Spacer
16. Put the chain on the O2 knob/sprocket assembly and the N2O sprocket.
17. Install the chain and sprockets on the needle valve spindles as an assembly. Do not
tighten the set screws.
Sprocket/Chain
Assembly
18. Install the N2O knob. Do not tighten the set screws.
19. Perform the link system calibration (Section 6.4).
20. Install the flowmeter panel shield.
21. Perform the checkout procedure (Section 3).
4-24
07/04 1009-0356-000
4 Repair Procedures
4.9.10 Replace an
Air needle valve on
all machines or an
O2 needle valve on
machines without
N2O
For machines with N2O, refer to Section 4.9.9 for replacing the O2 needle valve.
1. Bleed all gas pressure from the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. Remove the front flowmeter panel shield (Section 4.9.1).
4. Loosen the set screws on the flow control knob and remove the knob.
5. Loosen the set screws on the stop collar and remove the collar.
6. If equipped, remove the maximum flow stop collar.
7. Using a 16-mm wrench, remove the needle valve by turning it counterclockwise.
8. Install the new needle valve and tighten.
Note: Be sure the o-ring is properly located on the tip of the needle valve.
9. If equipped, install the maximum flow stop collar (do not tighten).
10. Install the stop collar (do not tighten the screws).
11. Install the flow control knob on the shaft. Tighten one set screw to snug.
12. Reconnect the gas supplies.
13. Perform the flow control stop procedures explained in:
• Section 6.3.1 for O2.
• Section 6.3.3 for Air.
• Section 6.3.4 for maximum flow.
14. Install the flowmeter panel shield.
15. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-25
S/5 Aespire
4.10 Service vaporizer manifold parts
4.10.1 Repair
manifold port
valve
1. Set the system switch to Standby.
2. Remove the vaporizers from the vaporizer manifold.
3. Using a 14-mm wrench, carefully remove the valve nipple (threaded).
4. Disassemble as necessary to replace parts. The following illustration shows the
parts.
Note: The port valve
replacement kit includes the
valve cartridge assembly
and the seal. The kit does
not include o-rings.
O-ring
Nipple
Valve
Cartridge
Seal
Spring
O-ring
5. When installing a new valve cartridge assembly into the vaporizer manifold, put a
light coat of Krytox on the bottom portion of the cartridge. The bottom portion of the
cartridge is defined as the brass surface that is inserted in the lower spring. Note: Do
not apply Krytox to the valve seal.
6. Verify that the parts are free of dust and dirt.
7. To reassemble, perform the previous steps in reverse order.
8. Complete the port valve checkout procedure described below (Section 4.10.2).
4-26
07/04 1009-0356-000
4 Repair Procedures
4.10.2 Checkout
procedure for
manifold port
valve
Note
Use the Vaporizer Manifold Valve Test Tool to perform the checkout procedure for the
manifold port valve. This tool and test procedure are intended for use only when the
valve cartridge assembly is replaced.
This replacement and test procedure is a service action and is not part of the
maintenance program.
1. Set the system switch to Standby.
2. After replacing the valve cartridge assembly, remove the vaporizer port o-ring.
3. Attach the valve tester to the top of the valve by sliding the bottom of the tester onto
the o-ring groove.
4. Tighten the tester screw down onto the valve until the screw bottoms out on the top
of the valve. The tester o-ring should create a seal with the top of the valve.
5. Remove the upper rear panel (Section 4.3).
6. Remove the inlet tube from the vaporizer manifold.
7. Set the SCGO Selector switch to ACGO.
8. Test the negative low-pressure leak-test device:
a. Put you hand on the inlet of the leak-test device. Push hard for a good seal.
b. Remove all air from the bulb.
c. The bulb should not inflate in less than 60 seconds.
9. Attach the negative low-pressure leak-test device to the ACGO outlet.
10.Remove all air from the bulb. The bulb should not inflate in less than 45 seconds.
11. Remove the valve tester.
12. Reassemble the inlet tube, vaporizer port o-ring, and the upper rear panel.
13. Conduct a negative low-pressure leak test on the system (Section 3.5.1).
w WARNING
1009-0356-000
07/04
If the valve test tool is not removed before flowing gas through the system,
pneumatic head damage could result.
4-27
S/5 Aespire
4.10.3 Replace
vaporizer manifold
check valve
1. Set the system switch to Standby.
2. Remove the vaporizers from the vaporizer manifold.
3. Remove the upper rear panel.
4. Disconnect the tubing from the valve block.
5. Remove the valve block.
• To access the left-hand mounting screw (A), remove the right (viewed from front)
side panel (B).
• The right-side extrusion include an access hole (C) for removing the left-hand
mounting screw.
• Note: For early production machines that do not have an access hole in the
extrusion, you must remove the vaporizer manifold to remove the valve block.
C
A
B
4-28
07/04 1009-0356-000
4 Repair Procedures
Note
The valve body, o-ring, and flapper do not come out with the block. They stay intact at
the bottom of the vaporizer manifold.
6. Pull the flapper out of the valve body.
O-ring
Valve body
Flapper
Valve body
O-ring
Flapper
7. Using a hex wrench, put the wrench through one of the holes of the valve body and
pull down to remove the valve body and o-ring.
8. Verify that parts are free of dust and dirt.
9. Replace the flapper by inserting the flapper stem and gently pulling the stem until
the flapper secures to the valve body.
10. Lightly lubricate the o-ring with Krytox.
11. Place the lubricated o-ring on the valve body port at the bottom of the manifold.
12. Gently install the valve body in the manifold:
• Check that the o-ring makes a good seal between the manifold and the valve
body.
• Check that the flapper valve makes solid contact with the valve body.
13. Install the valve block.
14. Reconnect the tubing to the valve block. Pull on the tube to ensure that it is locked
in the fitting.
15. Install the vaporizer front panel.
16. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-29
S/5 Aespire
4.10.4 Replace
vaporizer pressure
relief valve
1. Set the system switch to Standby.
2. Remove the vaporizers from the vaporizer manifold.
3. Remove the upper rear panel (Section 4.3).
4. Using a 13mm open ended wrench, remove the vaporizer pressure relief valve by
turning counterclockwise.
5. Verify that the parts are free of dust and dirt.
6. Install a new vaporizer pressure relief valve.
7. To reassemble, perform the previous steps in reverse order.
8. Perform the checkout procedure (Section 3).
4-30
07/04 1009-0356-000
4 Repair Procedures
4.10.5 Replace
vaporizer manifold
1. Remove the upper rear panel (Section 4.3).
2. Remove the front flowmeter shield (Section 4.9.1).
3. Remove the right side panel (A).
4. From the front of the machine, remove the screw (B) at the right upright of the
flowhead bezel.
5. From the back of the machine, remove the two screws (C) that hold the vaporizer
manifold vertical support to the flowhead bracket.
6. From the back of the machine, remove the two screws (D) that hold the vertical
support to the vaporizer manifold.
7. Remove the vertical support from the machine.
8. While holding the vaporizer manifold, remove the two screws (E) at the right-hand
extrusion to release the manifold.
A
E
B
D
C
9. Install the new vaporizer manifold in reverse order. Do not fully tighten the screws
until they are all in place.
• Attach the new manifold to the right-hand extrusion (E).
• Attach the vertical support to the vaporizer manifold (D).
• Attach the bottom of the vertical support to the flowhead bracket (C).
• Attach the vertical support to the front bezel (B).
10. Tighten the mounting screws in the following order: E, D, C, B.
11. Reassemble the machine.
12. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-31
S/5 Aespire
4.11 Replace ACGO selector switch
Removal
1. Remove the tabletop (Section 4.4).
2. Clip the tie wraps (1) from the outlet barb fittings at the side of the switch.
5
6
3
2
4
1
3. Disconnect the fresh gas (2) and flush (3) tubes at the back of the switch.
4. Disconnect the wires from the ACGO mode microswitch ( 4) at the back of the selector
switch.
5. Disconnect the wires from the flush pressure switch (5) on top of the selector switch.
6. Set the ACGO selector switch to ABS.
7. Back out the selector switch mounting screws (6) until the tips are flush with the face
of the mounting casting.
8. While pushing the selector knob toward the machine and holding it steady, push the
valve body toward the knob and rotate it counterclockwise to separate the valve
body from the knob assembly.
9. Remove the knob assembly and protective shroud from the machine.
10. Remove the valve from the silicone output tubes.
Replacement
1. Remove the knob assembly from the valve body.
2. Back out the selector switch mounting screws until the tips are flush with the face of
the mounting casting.
3. Guide the outlet fittings of the valve body into their respective silicone tubes.
4. Hold the selector knob with the
indicator mark facing down. Turn the
chrome collar to its maximum
counterclockwise position (as
viewed from the front).
4-32
07/04 1009-0356-000
4 Repair Procedures
5. Place the shroud over the knob and
guide the assembly into the pan
opening.
6. Ensure that the indicators on the
shroud align with label on the pan
and the alignment tab mates with
the alignment hole in the pan.
7. While holding the knob assembly
steady against the pan, place the
valve assembly over the knob
actuator. Using moderate force
press the two assemblies together.
The knob should rotate to the ACGO
position.
Actuator
8. While continuing to force the
assemblies together, rotate the
knob assembly to the ABS position. The assemblies should snap into place.
9. Verify proper alignment of the knob with the setting indicators. Tighten the mounting
screws evenly to secure the switch assembly to the pan.
10. Secure the outlet tubing with tie wraps.
11. Connect the fresh gas and flush gas tubing. Pull on the tubing to ensure that it is
locked in the fitting.
12. Reconnect the wires to the ACGO mode microswitch at the back of the valve
(top two terminals).
13. Reconnect the wires to the flush pressure switch at the top of the valve
(upper and lower terminals).
14. Replace the tabletop.
Test procedure
1. Confirm that flush flow and 5 L/min fresh gas flow are diverted to the ACGO port and
the ABS in the respective knob positions.
2. Confirm that the ventilator display indicates ACGO mode when the valve is set in the
ACGO position.
3. Test the function of the flush pressure switch
(Service Mode - “Display Discrete I/O Signals”).
4. Perform the low-pressure leak test (Section 3.5).
5. Perform the checkout procedure (Section 3)
1009-0356-000
07/04
4-33
S/5 Aespire
4.12 Clean or replace ACGO port flapper valve
1. Remove the tabletop (Section 4.4).
2. Remove the ACGO cap mounting
screws.
3. Remove the cap.
Cap
4. Examine the flapper and disk for
obstructions or debris. Clean with
isopropyl alcohol if necessary; retest.
5. If leak persists, replace the flapper.
• Remove the flapper from the check
valve disk.
• Clean the new flapper with isopropyl
alcohol.
• Apply a drop of isopropyl alcohol to
the center hub of the new flapper.
• Before the alcohol evaporates, align
the center hub of the new flapper
with the center hole of the check
valve disc.
• While pressing the flapper against
the disc, use you fingernail to help
pull the hub through the disc from
the other side.
Hub
Flapper
O-ring
6. Lubricate the o-ring sparingly with Krytox (do not get Krytox on the flapper).
7. Insert the flapper assembly into the ACGO outlet with the flapper up.
8. Replace the cap.
4-34
07/04 1009-0356-000
4 Repair Procedures
4.13 Reconfigure sample gas return line
Sample gas return is directed to the scavenging system as a factory default. Perform
the following to reroute the sample gas back to the breathing system. Refer to “Tubing”
on page 9-8.
1. Remove the tabletop (Section 4.4).
2. Port 4 (A) of the ABS breathing system
is connected to the expiratory circuit,
downstream of the expiratory check
valve. As a factory default, Port 4 is
plumbed with a length of tubing that is
plugged (B) at the far end.
3. Remove the plug from the tube.
A
4. Find the sample return line at the leftrear corner of the pan assembly. The
sample return line includes an inline
connector (C) at the point where the
sample line goes down into the vent
engine housing.
B
C
5. Separate the scavenging tube,
removing the inline connector from the portion of the tube that extends into the vent
engine housing. Plug the open end of the scavenging tube with the plug removed
above.
6. Insert the inline connector from the sample return port into the open tube to Port 4.
Pull on the connector to ensure that it is securely connected.
7. Replace the tabletop.
8. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-35
S/5 Aespire
4.14 Replace the APL valve
1. Remove the ABS breathing system.
2. The APL valve is held in place with a
spring and a retainer (A) that snaps into
a recess in the lower body of the APL
valve. To release the retainer, place an
appropriately sized straight blade
screwdriver into the housing cutout (B).
Twist the screwdriver to release the
retainer.
B
3. Place the new APL valve into position
with the setting indicator facing
forward.
4. Place the spring into the retainer.
5. While holding the APL valve tight to the
housing, Snap the spring and retainer
onto the valve body from below.
A
6. Reinstall the ABS breathing system.
7. Perform the checkout procedure (Section 3).
4-36
07/04 1009-0356-000
4 Repair Procedures
4.15 Replace the bag support arm
1. Remove the ABS breathing system from the machine.
2. From the underside of the casting,
remove the two screws/
lockwashers (A) that hold the arm in
place.
• If either of the pins (see below)
remain in the casting, remove
them from the casting.
A
3. Install the new bag support arm
assembly.
• Position the bag arm over
mounting pattern of 4 small holes
in the support casting. The arm
should extend towards the front of
B
the machine. Align the two pins
(B) extending from the base of the
bag arm assembly, with two of the
small holes in the casting that are
in line with the APL valve.
• Lower the bag arm, pushing the two pins into the holes.
• From the underside of the casting, secure the bag arm with two M3x16 screws and
lockwashers.
4. Test the force required to swing the bag arm from side to side and adjust if necessary.
Note: The adjustment nut is
initially set so that 5-mm of
exposed thread extends from
the adjusting nut. With use,
the force required to move the
arm increases and may
require readjustment.
The force is adjusted by turning the
lock nut (8-mm socket) which is
accessible from underneath the
support casting. Turn clockwise to
increase the force and
counterclockwise to reduce the
force.
C
• Swing the bag arm sideways
through the 90 degree arc
permitted by its internal stop.
• Adjust to just enough friction to prevent the bag arm from swinging sideways as
the bag height is being changed. The bag arm height is changed by squeezing the
lock release lever (C) at the free end of the bag arm and rotating it to the desired
position.
5. Replace the ABS breathing system.
1009-0356-000
07/04
4-37
S/5 Aespire
4.15.1 Servicing
the bag support
arm
Service parts for the bag support arm include the upper and lower assemblies.
To replace either assembly:
1. Remove the bag support arm from the machine (Section 4.15).
2. To separate the upper assembly
from the lower assembly, use a
small (2.5-mm) pin punch from the
bottom to drive the dowel pin up and
out.
3. To assemble the bag arm, apply a
light coat of Krytox to the area of the
upper arm (A) that extends into the
lower arm (including the dowel pin
groove).
A
B
C
C
Groove
4. Insert the upper assembly into the lower assembly. Align the surface (B) of the upper
assembly with the surface (C) of the lower assembly.
5. Insert the dowel pin into the hole
(from the top side as shown). Drive
the dowel pin into the bag arm until
it is flush with the top surface.
4-38
07/04 1009-0356-000
4 Repair Procedures
4.15.2 Replace
friction pad in
lower bag arm
assembly
1. Remove the ABS breathing system from the machine.
2. Using an 8-mm socket, remove the
nut (A), shoulder washer, and spring
from the lower assembly.
3. Lift the bag support arm off of the
swivel post. Remove the old friction
pad.
A
4. Wipe any residue and friction
particles from the post.
5. Insert a new friction pad into the
base. Keep approximately
1 mm of space between the end of
the pad and the bottom of the base.
Note: Align the friction pad gap with
the seam (B) in the base. Position
the retaining screw so the pin (C) at
the base is perpendicular to the
seam.
6. With the bag support arm facing
forward, place the base of the arm
over the swivel post. Ensure that the
slot in the base (D) engages the tab
(E) on the swivel post.
7. Replace the spring, shoulder
washer and nut. Tighten the nut until
5 mm of thread extends beyond the
nut.
B
C
E
D
8. Follow the procedure in Section 4.15 to adjust the force required to swing the bag
arm from side to side.
1009-0356-000
07/04
4-39
S/5 Aespire
4.15.3 Replace
bag port housing
1. Remove the bag support arm
cover (A) — screw and
lockwasher from below.
A
2. Remove nut (B) to remove the
release lever (C).
E
3. Remove the retaining ring (D).
4. Slide the bag port housing (E)
off the end of the bag support
arm.
D
C
B
5. Before installing the new bag
port housing, clean and
lubricate sparingly with Krytox
the exposed metal end (F) and
the guide slot (G) of the bag
support arm.
F
G
6. Slide the new bag port housing
onto the bag arm. Secure it with
the retaining ring.
2-mm
7. Lubricate sparingly with Krytox
the pivot boss (H) before
replacing the release lever.
8. After replacing the release
lever, adjust the mounting nut
so that a 2-mm gap remains
between the lever and housing
when the release lever is fully
depressed.
H
9. Replace the bag arm cover.
4-40
07/04 1009-0356-000
4 Repair Procedures
4.16 Replace auxiliary O2 flowmeter
1. Bleed all gas pressure from the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. Remove the tabletop (Section 4.4).
4. Remove the adjustment knob from the flowmeter; pull forward.
5. Remove the gauge panel mounting screws and move the panel forward to access the
flowmeter.
B
C
A
6. Disconnect the inlet tube fitting (A).
7. Disconnect the tube (B) from the outlet fitting.
8. Remove the four screws (C) that hold the flowmeter mounting bracket to the front
panel.
9. Transfer the mounting bracket to the new flowmeter.
10. Reassemble in reverse order.
11. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-41
S/5 Aespire
4.17 Replace the suction control module
The suction control module can be replaced by removing the front panel, along with the
ABS and the tabletop, to gain access. Alternatively, if the situation warrants, the
suction control module can be accessed by removing the rear panel.
4.17.1 Front panel
method
1. Bleed all gas pressure from the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. Remove the tabletop (Section 4.4).
4. Remove the gauge panel mounting screws and move the panel forward to access the
suction control module.
C
(pilot)
A (suction - black)
B
(vacuum
white)
5. Disconnect the tubing from the suction control module.
• Tube (A) from overflow safety trap (suction).
• Tube (B) from vacuum source.
• If venturi drive, tube (C) from pilot valve adapter.
6. Remove the two mounting screws that hold the suction control module to the
mounting bracket.
7. Transfer the mounting bracket to the new suction control module.
8. Reassemble in reverse order.
9. Perform the checkout procedure (Section 3).
4-42
07/04 1009-0356-000
4 Repair Procedures
4.17.2 Rear panel
method
1. Lower the upper rear panel (Section 4.3).
2. Disconnect the white (A) vacuum and black (B) suction fittings from the rear panel.
Do not remove the tubing from the regulator.
3. If you are replacing a venturi drive suction control module, disconnect the tube (C)
from the pilot valve adapter.
F
(suction)
B
(suction)
D
A
(vacuum)
C
E
(vacuum)
4. Remove the two screws (D) that hold the suction control module to the mounting
bracket.
5. Remove the regulator assembly from the front panel.
6. Transfer the tubing to the new regulator:
• Attach the vacuum source tube (white fitting) to the lower connector (E).
• Attach the suction tube (black fitting) to the upper connector (F).
7. Guide the tubes into the front panel opening.
8. While holding the regulator assembly against the front panel, attach the retaining
bracket to the regulator. Tighten the screws to secure the regulator assembly.
9. If applicable, attach the control port tube to the pilot valve adapter (C).
10. Attach the vacuum and suction fitting to the rear panel manifold.
11. Replace the rear panel.
12. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-43
S/5 Aespire
4.18 Replace ABS breathing system components
4.18.1 Replace
Bag/Vent switch
assembly
1. Remove the ABS breathing system.
2. From the underside, remove the bellows base manifold (A) and fully loosen the two
captive screws (B) at the bag port side of the APL/BTV manifold.
B
B
A
3. From the topside, rotate the Bag/Vent switch cartridge counterclockwise until the
Bag/vent switch outlet port (C) clears the bellows housing.
C
4. Lift out the Bag/Vent switch cartridge from the housing.
5. Replace the Bag/Vent switch cartridge in reverse order.
6. Reinstall the ABS breathing system.
7. Perform the checkout procedure (Section 3).
4-44
07/04 1009-0356-000
4 Repair Procedures
4.18.2 Replace
bellows base latch
assembly
To replace the latch assembly, you must disassemble the bellows base assembly to the
point where you can remove the guide (A) and latch assembly (B) as a unit.
1. Remove the Bag/Vent switch cartridge (Section 4.18.1).
2. Remove the two remaining screws (C) that hold the APL/BTV manifold to the bellows
base assembly. Remove the APL/BTV manifol.
B
D
C
A
3. To remove the guide/latch assembly, remove two mounting screws (D) from the
underside. Remove two additional mounting screws from the topside. Remove the
guide/latch assembly from the bellows base assembly.
4. Separate the latch assembly from the guide assembly.
5. To install the new latch assembly, put the spring (E) into place in the guide assembly
(long leg down).
6. Place the latch assembly on the guide assembly so that the latch engages the short
leg of the spring. Secure the latch assembly (F) to the guide assembly.
E
F
7. Mount the guide/latch assembly into the bellows base assembly.
8. Reassemble the breathing system in reverse order.
9. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
4-45
S/5 Aespire
4.19 Replace casters
w WARNING
Replacing a caster requires at least two people to maneuver and tip the machine.
Personal injury and/or machine damage is possible if one person attempts this
procedure alone.
1. Disconnect all pipeline hoses from the wall and the machine, close all gas cylinders,
unplug the power cord, and set the system switch to standby.
w CAUTION
Remove the vaporizers before tipping the machine. If a vaporizer is inverted, it
must be set to 5% and purged for 30 minutes with a 5 L/min flow. The interlock
system prevents purging more than one vaporizer at a time.
2. Remove the absorber, the vaporizers, gas cylinders, drawers and all auxiliary
equipment.
w CAUTION
To prevent damage, do not tip the Aespire machine more than 10 degrees from
vertical.
3. Block the opposite wheels; then, block up the machine until there is enough room to
remove the defective caster.
To block up the machine, tip and slide blocks under the caster base. Raise both
sides evenly until the unit is high enough to remove the caster.
4. The casters are threaded into the base
and held with a Loctite compound.
Remove the caster with an
appropriately sized open-end wrench.
5. If required, clean the treads of the new
caster with denatured alcohol.
6. Apply Loctite 242 to the threads of the
new caster. Install the caster securely
into place.
7. Make sure the caster turns freely.
8. Carefully lower the machine to the floor.
9. Perform the checkout procedure (Section 3).
4-46
07/04 1009-0356-000
4 Repair Procedures
4.20 Replace task light and switch
Remove the four screws (A) that hold the task-light lens to the upper shelf.
A
A
A
D
4.20.1 To replace
the task-light
switch
A
B
C
1. Using a small needle-nose pliers, disconnect the switch harness from the task-light
circuit board connector (B).
2. Remove the two screws (C) that hold the switch retainer plate to the upper shelf.
3. Transfer the switch retainer plate to the new switch, counter-sunk side to the outside.
4. Mount the switch to the upper shelf.
5. Remount the task-light assembly. Ensure that the switch harness and the task-light
harness wires are positioned in their respective recesses and are not pinched under
the task-light lens.
4.20.2 To replace
the task-light
circuit board
1. Using a small needle-nose pliers,
• disconnect the switch harness from the task light circuit board connector (B).
• disconnect the task-light harness from the task light circuit board connector (D).
2. Slide the task-light circuit board out of the lens.
3. Slide the new task light into the lens, ensuring that the connectors are aligned with
the lens cutouts.
4. Plug the task-light harness and the switch harness into their respective connectors
on the task-light circuit board. Use a small screwdriver to push the connectors
securely into place.
5. Remount the task-light assembly. Ensure that the switch harness and the task-light
harness wires are positioned in their respective recesses and are not pinched under
the task-light lens.
1009-0356-000
07/04
4-47
S/5 Aespire
4.21 Replace the display arm or display cables
Cable replacement requires that you first remove the display arm from the dovetail
extrusion.
Before replacing the display arm, note the routing of the cables.
After replacing the display arm, ensure that the cables are dressed properly and do not
interfere with the motion of the display arm.
Follow the procedure in Section 4.21.1 for the recommended use of cable ties.
4.21.1 Cable tie
installation
1. Wrap the cable tie around the
cables in the indicated position.
Start the tail through the clamp as
shown.
2. Feed the tail between the cables.
3. Pull the tie tight and cut off the tail
below the outer surface of the
cables. Do not leave a sharp edge
exposed.
4-48
07/04 1009-0356-000
4 Repair Procedures
4.21.2 Removing
the display arm
If equipped, remove additional equipment from the arm before removing the arm.
1. Disconnect the cables from the display.
2. Remove the display from the display
arm.
3. Remove the cables from the cable
clamps.
4. Loosen the screws that secure the
display arm in the dovetail.
5. If required, use a rubber mallet to
tap the display arm out of the
dovetail.
4.21.3 Replacing
a display cable
1. Remove the three screws that hold
the dovetail extrusion to the upright.
Remove the extrusion to allow cable
replacement.
2. As require to access the particular
cable routing for replacement,
remove either (or all):
• the rear cover (Section 4.3),
• the tabletop (Section 4.4),
• or the AC Inlet module.
3. After replacing the cable, place the
cables in the notches in the order
shown.
• A - Serial Isolation
• B - Monitoring
• C - Vent Engine
• D - Power cable
A
B
C
D
4. Install the dovetail extrusion loosely
to hold the cables in place.
5. Adjust the cable length outside the
machine to approximately 66 cm.
6. Securely tighten the extrusion mounting screws.
1009-0356-000
07/04
4-49
S/5 Aespire
4.21.4 Installing
the long arm
1. Place the arm into the extrusion.
2. Use a rubber mallet to tap the arm
into place. Leave a 12-mm gap
between the lower edge of the arm
mounting plate and the end of the
dovetail.
12-mm
3. Tighten the mounting screws to
secure the display arm.
4. Remount the display.
5. Route the display cables neatly
through the cable clamps.
6. Attach the cables to the display.
7. If required, install cable tie in the
locations shown.
• Place one cable tie close to the
pivot of the arm.
• Place the second cable tie near
the display as shown.
8. Ensure that the cables are secured
so that they do not interfere with the
display arm through the entire range
of motion.
4-50
07/04 1009-0356-000
4 Repair Procedures
4.21.5 Installing
the short arm
1. Place the arm into the extrusion.
2. Use a rubber mallet to tap the arm
into place. Leave a 12-mm gap
between the lower edge of the arm
mounting plate and the end of the
dovetail.
12-mm
3. Tighten the mounting screws to
secure the display arm.
4. Remount the display.
5. Route the display cables neatly
through the cable clamps.
6. Attach the cables to the display.
7. If required, install cable tie in the
locations shown.
• Place one cable tie close to the
pivot of the arm.
• Place the second cable between
the two straps.
• Place the third cable tie near the
display as shown.
8. Ensure that the cables are secured
so that they do not interfere with
display through the entire range of
motion.
1009-0356-000
07/04
4-51
S/5 Aespire
4.22 Replace display and cables in ProTIVA machine
1. Remove two thumbscrews that
secure the display mounting bracket
to the no-vap manifold.
2. Lift the display slightly to disengage
the mounting pins.
3. Lower the display face down on the
worksurface.
4. Replace the display and
reassemble in reverse order.
5. To replace a cable, remove the
retaining block and feed the cable to
the back of the machine.
6. As require to access the particular
cable routing for replacement,
remove either (or all):
• the rear cover (Section 4.3),
• the tabletop (Section 4.4),
• or the AC Inlet module.
7. Attach the cable to the display.
8. When replacing the cable retainer,
ensure:
• that the cables fit properly in their feed-through slots (as detailed below),
• and that they extend out of the machine with minimal slack.
9. From the back of the machine, ensure that the cables are routed such that they do
not interfere with the replacement of the rear cover.
D C
B A
A B C D
A - Serial Isolation
B - Monitoring board
C - Vent Engine
D - Power cable
4-52
07/04 1009-0356-000
5
In this section
Maintenance
This section covers the regular maintenance procedures (minimum requirements)
needed to make sure that the Aespire Anesthesia Machine — including the ventilator —
operates to specifications.
5.1 Aespire Planned Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.1.1 Every twelve (12) months . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.1.2 Every twenty-four (24) months . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
5.2 Auxiliary O2 flowmeter tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
5.3 Integrated Suction Regulator tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
w WARNINGS
Do not perform testing or maintenance on the Aespire Anesthesia Machine while
it is being used on a patient. Possible injury can result.
Items can be contaminated due to infectious patients. Wear sterile rubber
gloves. Contamination can spread to you and others.
Obey infection control and safety procedures. Used equipment may contain
blood and body fluids.
1009-0356-000
07/04
5-1
S/5 Aespire
5.1 Aespire Planned Maintenance
Serial Number:
Date: (YY/MM/DD)
Hospital:
Performed by:
12 months
24 month
5.1.1 Every twelve (12)
months
Machine Parts Replacement
Machine Checks and Tests
/
/
______________________
Perform the following steps every 12 months.
Refer to the listed section in this manual. Perform the following step:
Replace the vaporizer port o-rings (Section 4.10.1)
(Kit Stock Number 1102-3016-000)
Refer to the Aespire User’s Reference Manual, Part 2.
Perform the following steps:
1. User maintenance listed below. Including disassembly, inspection,
cleaning and parts replacement as required (Section 3 and Section 2).
• AGSS Maintenance:
Empty any condensate from the reservoir (disposable item).
Inspect air brake for occlusion.
Inspect, clean or replace filter on active AGSS.
• Breathing Circuit Maintenance
• Bellows Assembly Maintenance
• Bellows Assembly Tests
• O2 Sensor Calibration
• Flow Sensor Calibration
Refer to listed sections in this manual.
Perform the following steps:
5-2
2. Inspect the system (Section 3.1)
3. Pipeline and cylinder tests (Section 3.2)
4. Flow control and pressure relief tests (Section 3.3)
5. Vaporizer back pressure test (Section 3.4)
6. Low-pressure leak test (Section 3.5)
7. Airway pressure gauge accuracy check (Section 6.6.2)
8. Alarm tests (Section 3.6)
9. Breathing systems tests (Section 3.7)
10. Auxiliary O2 flowmeter tests, if equipped with option (Section 5.2)
07/04 1009-0356-000
5 Maintenance
11. Integrated suction regulator tests, if equipped with option (Section 5.3)
12. Power failure test (Section 3.10)
13. Electrical safety tests (Section 3.11)
7100 Ventilator Checks, Tests
and Calibrations
Refer to the listed sections in the Aestiva 7100 Ventilator Service Manual.
Perform the following steps:
1. MOPV pressure relief valve test (Section 6.3)
2. From the Ventilator Service Mode menu, perform the following:
• Display Discrete I/O Signals. Verify proper operation of all switches.
(Section 4.10.2)
• Display Error Log. If any error codes have been logged follow the
appropriate troubleshooting procedures. Clear the error log. (Section
4.5)
• Adjust Drive Gas Regulator (Section 4.9.3)
• Airway Sensor Span (Section 4.9.4)
• PEEP Valve Calibration (Section 4.9.5)
• Inspiratory Valve Calibration (Section 4.9.6)
• Pressure Sensitivity (Section 4.9.7)
5.1.2 Every twenty-four
(24) months
7100 Ventilator Parts
Replacement
1009-0356-000
07/04
In addition to the 12-month requirements, replace the following parts every
24 months. All machine and ventilator parts should be replaced before
performing the checks, tests, and calibrations.
Refer to the listed sections in the Aestiva 7100 Ventilator Service Manual.
Perform the following steps:
1. Replace the internal backup battery (Section 7.2.3)
(Stock Number 1504-3505-000).
2. Replace the free breathing flapper valve (Section 6.2)
(Stock Number 0211-1454-100).
3. Replace the free breathing valve o-ring (Section 6.2)
(Stock Number 1503-3208-000).
5-3
S/5 Aespire
5.2 Auxiliary O2 flowmeter tests
1. Open the O2 cylinder valve or connect an O2 pipeline.
2. Rotate the flow control clockwise (decrease) to shut off the flow. The ball
should rest at the bottom of the flow tube and not move.
3. Rotate the flow control counterclockwise (increase). The ball should rise
immediately after rotation is begun. It should rise smoothly and steadily
with continued counterclockwise rotation. When a desired flow is set, the
ball should maintain in a steady position.
4. Rotate the flow control clockwise to shut off the flow.
Flow Accuracy Test
Note: To check flow accuracy, be sure that the flow test device is capable of
measuring 0 to 15 L/min with an accuracy of ±2% of reading.
1. Connect the flowmeter outlet to the flow test device.
2. Adjust the flowmeter so the center of the ball aligns with the selected test
point (observe that the ball maintains a steady position for 10 seconds).
3. The test device reading should be between the limits shown for each of the
selected settings in the table below.
Flow Tester Reading
Flowmeter Setting
L/min
Lower Limit
L/min
Upper Limit
L/min
1
0.52
1.48
3
2.56
3.44
5
4.60
5.40
10
9.70
10.30
maximum
(valve fully open)
12.00
-------
4. Rotate the flow control clockwise to shut off the flow.
5. Close the O2 cylinder valve or disconnect the O2 pipeline.
5-4
07/04 1009-0356-000
5 Maintenance
5.3 Integrated Suction Regulator tests
Note
There are two types of integrated suction systems for the Avance anesthesia
machine:
• Continuous Vacuum Regulator, Three-Mode, Pipeline Vacuum
• Continuous Vacuum Regulator, Three-Mode, Venturi Derived Vacuum
For Pipeline Vacuum systems,
a vacuum source of at least 500 mm Hg (67 kPa or 20 in Hg) is required for
testing. The supply open flow must be a minimum of 50 L/min.
For Venturi Derived Vacuum systems,
an O2 or Air source of at least 282 kPa (41 psi) is required for testing.
Gauge Accuracy
The gauge needle should come to rest within the zero range bracket when no
suction is being supplied. Gauges which do not comply may be out of
tolerance.
Note
To check gauge accuracy, be sure that the test gauge is capable of measuring
0 to 550 mm Hg with an accuracy of ±1% of reading.
1. Connect the suction patient port to the test gauge.
2. Turn the mode selector switch to I (ON).
3. Ensure that the vacuum test gauge is in agreement with the suction vacuum
gauge ± 38 mm Hg/5 kPa at the following test points.
Test points
Flow Test
Suction vacuum gauge
Test gauge tolerance
100 mm Hg (13.3 kPa)
62–138 mm Hg (8.3–18.4 kPa)
300 mm Hg (40 kPa)
262–338 mm Hg (35–45 kPa)
500 mm Hg (66.7 kPa)
462-538 mm Hg (61.6–71.7) kPa)
Note: To check flow accuracy, be sure that the flow test device is capable of
measuring 0–30 L/min.
1. Connect the patient port of the suction regulator to the flow test device.
2. Rotate the suction control knob fully clockwise (increase).
3. Turn the mode selector switch to I (ON) and verify that the flow rate is:
• at least 20 L/min.
4. Disconnect the test flowmeter.
(Tests continue on next page.)
1009-0356-000
07/04
5-5
S/5 Aespire
Regulation Test
1. Turn the mode selector switch to I (ON).
2. Occlude the patient port of the suction regulator.
3. Set the vacuum regulator gauge to 100 mm Hg/13 kPa.
4. Open and close the patient port several times.
5. With the patient port occluded, the gauge should return to 100 mm Hg/13
kPa within a tolerance of ± 10 mm Hg/1.3 kPa.
Vacuum Bleed Test
1. Occlude the patient port of the suction regulator.
2. Set the vacuum regulator gauge to 100 mm Hg/13 kPa.
3. Turn the mode selector switch to O (OFF) and observe the gauge needle. It
must return to the zero range bracket or stop pin within 10 seconds.
Vacuum Leak Test
1. Turn the mode selector switch to O (OFF).
2. Rotate the suction control knob a minimum of two full turns in the
clockwise direction (increase suction) to ensure its setting is not at the off
position.
3. Occlude the patient port of the suction regulator.
4. Observe the suction gauge, the needle should not move.
5. Rotate the suction control knob fully counterclockwise to ensure its setting
is at the fully off position.
6. Turn the mode selector switch to I (ON).
7. Observe the suction gauge, the needle should not move.
5-6
07/04 1009-0356-000
6 Calibration
w WARNING
After adjustments and calibration are completed, always perform the checkout
procedure. Refer to Section 3 of this manual.
In this section
This section covers calibration procedures for components of the Aespire anesthesia
machine.
6.1 Primary Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
6.1.1 Test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
6.1.2 Testing Primary Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
6.1.3 Adjusting Primary Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
6.2 Secondary Regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
6.2.1 Testing/Adjusting Secondary Regulators or Balance Regulators . . . . . . . . . . . . . .6-7
6.3 Flowmeter Needle Valve Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
6.3.1 O2 Needle Valve Calibration (Minimum Flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.3.2 N2O Needle Valve Calibration (Minimum Flow). . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
6.3.3 Air Needle Valve Calibration (Minimum Flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
6.3.4 Needle Valve Calibration (Maximum Flow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
6.4 Link system calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
6.5 O2 Flush Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
6.6 Airway pressure gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.6.1 Zero the pressure gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-24
6.6.2 Checking the pressure gauge accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
1009-0356-000
07/04
6-1
S/5 Aespire
6.1 Primary Regulators
Follow the procedure in Section 6.1.1 to gain access to the regulators. Then,
in Section 6.1.2, select the test that is appropriate for the regulator you are
testing.
w WARNING
When testing/adjusting N2O regulators, nitrous oxide flows through the
system. Use a safe and approved procedure to collect and remove it.
6.1.1 Test setup
w WARNING
w CAUTION
Wear safety glasses while test device is connected to the test port.
Be careful not to plug the output of the primary regulator without having
a pressure relief valve in the output circuit.
1. Set the system switch to Standby.
2. Disconnect all pipeline supplies.
3. Remove the upper rear panel (Section 4.3).
For 3rd gas cylinder regulators, also remove the lower rear panel.
4. If equipped, turn the auxiliary O2 flowmeter control fully clockwise (no flow).
5. Install a full cylinder in the cylinder supply to be tested. It is essential that
the cylinder be within 10% of its full pressure.
6. Remove the plug from the test port and connect a test device capable of
measuring 689 kPa (100 psi).
Remove Plug
6-2
07/04 1009-0356-000
6 Calibration
6.1.2 Testing Primary
Regulators
Test A
There are two variations of the test procedure for the primary regulators:
• Test A — For primary regulators that supply drive gas to the ventilator.
• Test B — For all gases not used to supply drive gas to the ventilator.
For primary regulators that supply drive gas to the ventilator (O 2 or Air):
Under low flow conditions, the output pressure of a properly adjusted/
functioning regulator should fall within specifications listed in step 4. Under
high flow conditions, the output pressure should not drop below the
specifications in step 12.
1. Remove the bellows assembly.
2. Slowly open the cylinder valve.
3. Set the system switch to On.
4. Low Flow Test: Set the fresh gas flow to 0.05 L/min (or minimum flow for
O2). When checking an Air regulator on systems that have a single
flowtube, open the needle valve 1/8 turn from the minimum stop to
achieve a flow close to 0.05 L/min.
• Close the cylinder valve and allow the pressure to decay to 2068 kPa
(300 psi) as indicated on the cylinder gauge (upper limit of the red band).
The flow may be temporarily increased to facilitate the decay.
• At the time that the cylinder pressure reaches 2068 kPa (300 psi), set
the system switch to Standby.
• Within one minute, the test device must stabilize between:
(60) DIN 372–400 kPa (54–58 psi)
(50) Pin Indexed 310–341 kPa (45.0–49.5 psi).
- If the test device pressure does not stabilize within one minute,
replace the cylinder supply.
- If the test device stabilizes within one minute, but the readings are
not within specifications, readjust the regulator (Section 6.1.3).
5. Slowly open the cylinder valve.
6. Enter the Service Mode:
(Push and hold the adjustment knob on the ventilator’s display and set the
system switch to On.)
7. Select and confirm “Service Modes.”
8. Follow the menu structure outline below to reach the adjustment for the
inspiratory flow valve. Select and confirm at each step.
• “Diagnostics Tests/Tools”
• “Valves - Test Tool”
• “Set Inspiratory Valve”
1009-0356-000
07/04
6-3
S/5 Aespire
9. High Flow Test: Rotate adjustment knob counterclockwise to obtain
65 (L/min):
• While watching the test device press confirm.
• After 2 seconds, select “Go to Diagnostic Tests/Tools Menu” and press
confirm to stop the gas flow.
• The minimum test device reading observed must be greater than:
(60) DIN 221 kPa (32 psi)
(50) Pin Indexed 207 kPa (30 psi)
Repeat this step three times.
If the test device reading under “high flow” conditions is less than
specified, readjust the regulator per the procedure in Section 6.1.3;
however, set the regulated pressure higher by the difference you noted in
this step plus 7 kPa (1 psi). This adjusts the “low flow” regulated output to
the high side of the specification so that the “high flow” regulated pressure
can fall within the specification.
If the regulator subsequently fails the “low flow” specification (step 4)
because the reading is too high, replace the cylinder supply.
10. Set the system switch to Standby.
11. Close the cylinder valve.
12. Bleed the system of all pressure.
13. Disconnect the test device and plug the test port (pull on the plug to
ensure it is locked in the fitting).
14. Replace the bellows assembly.
15. Replace the rear panel(s).
16. Perform the checkout procedure (Section 3).
Test B
For all gases not used to supply drive gas to the ventilator:
Under low flow conditions, the output pressure of a properly adjusted/
functioning regulator should fall within specifications listed in step 4. Under
high flow conditions, the output pressure should not drop below the
specifications in step 7.
1. If the cylinder supply being tested is N2O, connect a source of O2 and set
the O2 flow control to the minimum stop (pilot pressure for secondary
regulator).
2. Slowly open the cylinder valve for the regulator being tested.
3. Set the system switch to On.
6-4
07/04 1009-0356-000
6 Calibration
4. Low Flow Test: Set the flow of the gas being tested to 0.05 L/min (or
minimum flow for O2). When checking a regulator on systems that have a
single flowtube, open the needle valve 1/8 turn from the minimum stop to
achieve a flow close to 0.05 L/min.
• Close the cylinder valve and allow the pressure to decay to 2068 kPa
(300 psi) as indicated on the cylinder gauge (upper limit of the red band).
The flow may be temporarily increased to facilitate the decay.
• At the time that the cylinder pressure reaches 2068 kPa (300 psi), set
the system switch to Standby.
• Within one minute, the test device must stabilize between:
(60) DIN 372–400 kPa (54–58 psi)
(50) Pin Indexed 310–341 kPa (45.0–49.5 psi).
- If the test device pressure does not stabilize within one minute,
replace the cylinder supply.
- If the test device stabilizes within one minute, but the readings are
not within specifications, readjust the regulator (Section 6.1.3).
5. Slowly open the cylinder valve.
6. Set the system switch to On.
7. High Flow Test: Set the flow control valve to the maximum indicated flow
on the flow tube.
• The test device reading must be greater than:
(60) DIN 221 kPa (32 psi)
(50) Pin Indexed 221 kPa (32 psi)
- If the test device reading under “high flow” conditions is less than
specified, readjust the regulator per the procedure in Section 6.1.3;
however, set the regulated pressure higher by the difference you
noted in this step plus 7 kPa (1 psi). This adjusts the “low flow”
regulated output to the high side of the specification so that the
“high flow” regulated pressure can fall within the specification.
- If the regulator subsequently fails the “low flow” specification
(step 4) because the reading is too high, replace the cylinder supply.
8. Set the system switch to Standby.
9. Close the cylinder valve.
10. Bleed the system of all pressure.
11. Disconnect the test device and plug the test port (pull on the plug to
ensure it is locked in the fitting).
12. Replace the rear panel(s).
13. Perform the checkout procedure (Section 3).
1009-0356-000
07/04
6-5
S/5 Aespire
6.1.3 Adjusting Primary
Regulators
Important: Cylinder supplies in an Aespire machine must have all primary
regulators set to the same pressure range: (50) Pin Indexed or (60) DIN. If a
regulator is replaced, the replacement regulator must be set (as required) to
the same specification as the one removed.
Important: Install a full cylinder in the cylinder supply to be adjusted. It is
essential that the cylinder be within 10% of its full pressure.
If the cylinder supply being adjusted is N2O, connect a source of O2 and set
the O2 flow control to the minimum stop (pilot pressure for secondary
regulator).
To adjust the primary regulators, follow the procedure in Section 6.1.1 to gain
access to the regulators.
Do not attempt to adjust without flow.
1. Slowly open the cylinder valve.
2. Set the system switch to On.
Adjust
clockwise
to increase
setting
3. Set and maintain the fresh gas flow of the gas being tested to 0.05 L/min
(or minimum flow for O2). When adjusting a regulator on systems that have
a single flowtube, open the needle valve 1/8 turn from the minimum stop
to achieve a flow close to 0.05 L/min.
4. Close the cylinder valve and allow the pressure to decay to 2068 kPa (300
psi) as indicated on the cylinder gauge (upper limit of the red band). The
flow may be temporarily increased to facilitate the decay.
5. When the cylinder gauge reaches the upper limit of the red band, adjust
the regulator output pressure to:
(60) DIN 386–400 kPa (56–58 psi)
(50) Pin Indexed 327–341 kPa (47.5–49.5 psi).
Note: It may be necessary to open the cylinder valve and repeat steps 4
and 5 a number of times to achieve the above setting.
6. Test the regulator settings per the appropriate test in Section 6.1.2:
• Test A — For primary regulators that supply drive gas to the ventilator.
• Test B — For all gases not used to supply drive gas to the ventilator.
6-6
07/04 1009-0356-000
6 Calibration
6.2 Secondary Regulators
w WARNING
6.2.1 Testing/Adjusting
Secondary Regulators or
Balance Regulators
When testing N2O regulators, nitrous oxide flows through the system.
Use a safe and approved procedure to collect and remove it.
1. Set the system switch to Standby.
2. Remove the flowmeter panel shield (Section 4.9.1).
3. Remove the plug from the test port and connect a test device capable of
measuring 689 kPa (100 psi) using 1/8-inch nylon tubing.
Air
Adjust
O2
Adjust
Test Ports
4. Set the flow of the tested gas and of O2 as detailed in the chart.
5. Verify that the output of the tested regulator is within the range listed in the
chart.
Regulator
Output
Flow
Regulated gas
Flow
O2
O2
207 ±7 kPa (30 ±1 psi)
2 L/min
---------
Air
207 ±7 kPa (30 ±1 psi)
2 L/min
---------
N2 O
± 14 kPa (±2 psi) of O2 reading
10 L/min
4 L/min
6. If required, adjust the O2 and Air regulators to meet the above
specifications. The N2O regulator is not adjustable; replace if out of range.
Note: The adjustment screws for these regulators are self-locking.
7. Disconnect the test device and plug the test port (pull on the plug to ensure
it is locked in the fitting).
8. Perform the Flow Control Tests (Section 3.3).
1009-0356-000
07/04
6-7
S/5 Aespire
6.3 Flowmeter Needle Valve Calibration
You need to calibrate a needle valve:
• if you install a new one,
• if minimum and maximum flows are not within specifications.
6.3.1 O2 Needle Valve Calibration (Minimum Flow)
w CAUTION:
Do not force the needle valve against the seat. Overtightening the valve
can cause the minimum flow setting to drift out of specifications.
1. Set the system switch to Standby.
2. Remove the flowmeter panel shield (Section 4.9.1).
Note: If adjusting an existing needle valve,
• remove the N2O and O2 knob and sprocket assemblies
(on machines with O2 only flowhead, remove the O2 knob),
• and loosen the O2 stop collar setscrews.
3. Slide a stop collar onto the valve stem with the stop tab toward the valve.
Do not tighten setscrews.
Stop Collar
Stop
Tab
Valve Stop Tab
Valve Stem
4. Set the system switch to On.
5. Adjust the O2 needle valve to maintain a flow of
• 50 ±25 mL/min for dual tube flowmeters.
• 200 ±25 mL.min for single tube flowmeters.
6. Push the stop collar against the valve body.
6-8
07/04 1009-0356-000
6 Calibration
7. Turn the collar clockwise until the collar stop tab contacts the minimum
stop tab on the valve body. Do not turn the valve stem.
Collar Stop
Valve
Stop
Collar
Collar stop must be on
CCW side of valve stop.
8. Carefully pull the collar back so there is a slight gap between collar and the
valve body (but still engages the valve stop).
Collar
Setscrew
(example)
Gap
9. Tighten the collar setscrews. Start with the one opposite the tab if possible.
10. Turn the valve stem counterclockwise at least one revolution to make sure
the collar tab clears the valve stop.
Stop must clear
If the stop does not clear:
a. Turn the valve stem back to minimum position.
b. Loosen the collar setscrews.
c. Repeat steps 6 through 9.
11. Turn the valve stem clockwise to the minimum stop.
12. Verify that the flow is within the
• 50 ±25 mL/min range for dual tube flowmeters.
• 200 ±25 mL/min range for single tube flowmeters.
13. Set the maximum stop collar if necessary (Section 6.3.4).
Note: Maximum stop collars are required in Canada for all gas flow
controls.
14. Calibrate the Link proportioning system (Section 6.4)
(on machines with O2 only flowhead, replace and set the knob so that at
minimum flow the label text is horizontal; replace the flowmeter shield).
1009-0356-000
07/04
6-9
S/5 Aespire
6.3.2 N2O Needle Valve Calibration (Minimum Flow)
w WARNING
You must be in a well ventilated room or use a gas evacuation device at
this time. Anesthetic vapors exhausted into the room air can be harmful
to your health.
w CAUTION:
Do not force the needle valve against the seat. Overtightening the valve
can cause the minimum flow setting to drift out of specifications.
1. Disconnect all pipeline supplies and close all cylinder valves.
2. Remove the upper rear panel.
Note: If adjusting an existing needle valve,
• remove the N2O and O2 knob and sprocket assemblies,
• and loosen the O2 stop collar setscrews.
3. Remove the flowmeter panel shield (Section 4.9.1).
4. Disconnect the tube from the pilot port on the N2O regulator.
N2O pilot port
Disconnect
tube
N2O
O2
5. Disconnect the 4-mm outlet tube from the back of the N2O pipeline
manifold.
Connect to
N20 pilot port
Connect 4-mm outlet tube
6. Using a 4-mm tube/tee fixture (see Service Tools - Section 8.1.3), connect
a tube (tee end) to the N2O supply outlet at the back of the pipeline
manifold. Connect the outlet tube to the open connection on the tee
connector of the fixture. Connect the other end of the fixture to the N 2O pilot
port at the front of the flowhead. This setup supplies pilot pressure to the
N2O balancing regulator during the minimum stop calibration.
6-10
07/04 1009-0356-000
6 Calibration
7. Slide a stop collar onto the valve stem with the stop tab toward the valve.
Do not tighten setscrews.
Stop Collar
Stop
Tab
Valve Stop Tab
Valve Stem
8. Connect either an N2O pipeline or cylinder supply.
9. Slowly open the N2O cylinder valve.
Important: Do not connect the O2 pipeline or open the O2 cylinder valve.
10.Adjust the needle valve until the float is nearly touching the filter, but not
quite.
Flowtube
Float
N2O
Filter
11. Disconnect the tubing from the inlet of the vaporizer manifold (closest to
flowhead).
12. If the machine has an Air option, bleed down the air supply. Air can inflate
the bubble (next step) if it is not shut off.
13. Apply a small amount of leak detection fluid (Snoop) to the end of the tube
to form a bubble.
14. Turn the valve stem clockwise until the bubble no longer inflates. Do not
turn more than 10 degrees clockwise past this point.
10 degrees
N2O
1009-0356-000
07/04
6-11
S/5 Aespire
15. Push the stop collar against the valve body.
16. Turn the collar clockwise until the collar stop tab contacts the minimum
stop tab on the valve body. Do not turn the valve stem.
Collar Stop
Valve
Stop
Collar
Collar stop must be on
CCW side of valve stop.
17. Carefully pull the collar back so there is a slight gap between collar and the
valve body (but still engages the valve stop).
Collar
Setscrew
(example)
Gap
18. Tighten the collar setscrews. Start with the one opposite the tab if
possible.
19. Turn the valve stem counterclockwise at least one revolution to make sure
the collar tab clears the valve stop.
Stops must clear
If the stops do not clear:
a. Turn the valve stem back to minimum position.
b. Loosen the collar setscrews.
c. Repeat steps 14 through17.
6-12
07/04 1009-0356-000
6 Calibration
20. Turn the valve clockwise to the minimum stop.
21. Verify there is no flow at the end of the tube.
22. Thoroughly clean the end of the nylon tube and reconnect it to the
vaporizer manifold inlet.
23. Set the maximum stop collar if necessary (Section 6.3.4).
Note: Maximum stop collars are required in Canada for all gas needle
valves.
24. After calibrating minimum flow for N2O:
a. Close the cylinder valve and use the needle valve to bleed the
remaining gas.
b. Remove the test fixture connecting the N2O gas supply to pressure
balance regulator pilot port.
c. Reconnect the pilot tube to the N2O pilot port. Pull on the tubing to
ensure it is locked into the fitting.
Connect
N2O
O2
d. Reconnect the outlet tube to the N2O pipeline supply manifold.
25. Calibrate the Link proportioning system (Section 6.4).
1009-0356-000
07/04
6-13
S/5 Aespire
6.3.3 Air Needle Valve Calibration (Minimum Flow)
w CAUTION:
Do not force the needle valve against the seat. Overtightening the valve
can cause the minimum flow setting to drift out of specifications.
1. Set the system switch to Standby.
2. Disconnect all pipeline hoses and close all cylinder valves except for air.
3. Remove the flowmeter panel shield (Section 4.9.1).
4. Remove the upper rear panel.
Note: If adjusting an existing needle valve,
• remove the Air knob,
• and loosen the Air stop collar setscrews.
5. Slide a stop collar onto the valve stem with the stop tab toward the valve.
Do not tighten setscrews.
Stop Collar
Stop
Tab
Valve Stop Tab
Valve Stem
6. Set the system switch to On.
7. Adjust the needle valve until the float is nearly touching the filter, but not
quite.
Flowtube
Float
Air
Filter
8. Disconnect the tubing from the inlet to the vaporizer manifold (left end of
manifold).
9. Apply a small amount of leak detection fluid (Snoop) to the end of the tube
to form a bubble.
6-14
07/04 1009-0356-000
6 Calibration
10. Turn the needle valve clockwise until the bubble no longer inflates. Do not
turn more than 10 degrees clockwise past this point.
10 degrees
Air
11. Push the stop collar against the valve body.
12. Turn the collar clockwise until the collar stop tab contacts the minimum
stop tab on the valve body. Do not turn the valve stem.
Collar Stop
Valve
Stop
Collar
Collar stop must be on
CCW side of valve stop.
1009-0356-000
07/04
6-15
S/5 Aespire
13. Carefully pull the collar back so there is a slight gap between collar and the
valve body (but still engages the valve stop).
Collar
Setscrew
(example)
Gap
14. Tighten the collar setscrews. Start with the one opposite the tab if
possible.
15. Turn the valve stem counterclockwise at least one revolution to make sure
the collar tab clears the valve stop.
Stops must clear
If stops do not clear:
a. Turn the valve stem back to minimum position.
b. Loosen the collar setscrews.
c. Repeat steps 11 through 14.
16. Turn the valve stem clockwise to the minimum stop.
17. Verify there is no flow at the end of the tube.
18. Thoroughly clean the end of the nylon tube and reconnect it to the
vaporizer manifold inlet.
19. Set the maximum stop collar if necessary (Section 6.3.4).
Note: Maximum stop collars are required in Canada for all gas needle
valves.
20. Set the knob so that at minimum flow the label text is horizontal and the
knob is on an even plane (front to back) with the N2O and O2 knobs.
21. Replace the flowmeter panel shield and the rear panel.
6-16
07/04 1009-0356-000
6 Calibration
6.3.4 Needle Valve
Calibration (Maximum
Flow)
Note: Maximum stop collars are required in Canada for all gas needle valves.
1. Calibrate the needle valve for minimum flow:
• Section 6.3.1 for O2
• Section 6.3.2 for N2O
• Section 6.3.3 for Air
2. Turn the valve open 1/2 turn beyond the maximum indicated flow.
3. Position the maximum stop collar so that its hook contacts the stop collar
tab on the counterclockwise side. The hook and tab should have
overlapping contact of about 0.75 mm (about half the thickness of the stop
collar tab).
Maximum
Stop Collar
Stop Collar
Needle Valve
Assembly
Note: This illustration shows the maximum stop collar in a vertical position.
The actual position of the maximum stop collar may vary for each needle
valve.
4. Tighten the locking screw on the maximum stop collar.
5. Turn the valve one full turn clockwise to make sure the hook does not
contact the stop collar tab. If there is contact, move the maximum stop
collar slightly forward.
6. Verify that you can turn the valve open 1/2 turn beyond the maximum
indicated flow.
7. Turn the valve fully clockwise to the minimum stop.
1009-0356-000
07/04
6-17
S/5 Aespire
6.4 Link system calibration
Before you start, make sure that:
• All parts are correctly installed.
• Stops on needle valves are set correctly.
• The machine meets leak check requirements.
• Confirm that the O2 sensor measures 21% in room air and 100% in pure O2.
If not, calibrate the O2 sensor.
Note: All illustrations in this section show ANSI flowmeter module positions.
The order is reversed on ISO machines.
w WARNING
You must be in a well ventilated room or use a gas evacuation device at
this time. Anesthetic vapors exhausted into the room air can be harmful
to your health.
1. Set the system switch to Standby.
2. Remove the flowmeter panel shield (Section 4.9.1).
3. Put the plastic spacer on the N2O needle valve spindle.
4. Turn the O2 and the N2O needle valves clockwise to their minimum stop
position.
Spacer
5. Put the chain onto the O2 knob/sprocket assembly and the N2O sprocket.
Note: The N2O sprocket set screws should be away from the valve.
6-18
07/04 1009-0356-000
6 Calibration
6. Install the chain and sprockets onto the needle valve stems as an
assembly. Press the O2 knob/sprocket against the O2 minimum stop
collar.
Sprocket/Chain
Assembly
7. Tighten the setscrews in the O2 knob. Do not tighten the N2O sprocket
setscrews.
Note: If O2 label is on the knob, turn the knob so that the identification label
is horizontal before tightening the setscrews.
8. Turn on the O2 and the N2O gas supplies (pipeline or cylinder).
9. Set the system switch to On.
10. Adjust the needle valves:
• O2 needle valve: 200 ±10 mL/min.
• N2O needle valve: 600 ±25 mL/min.
1009-0356-000
07/04
6-19
S/5 Aespire
11. Turn the sprocket on the O2 knob sprocket assembly counterclockwise
until it stops against the tab on the O2 knob. Do not allow the N2O or O2
valve stems to rotate.
Turn
counterclockwise
12. Push the N2O sprocket against the plastic spacer.
13. Holding the O2 knob, rotate the N2O sprocket counterclockwise until all
slack is removed from the chain.
14. Lightly tighten both N2O sprocket setscrews.
15. Turn the N2O needle valve clockwise to the minimum stop position.
6-20
07/04 1009-0356-000
6 Calibration
16. Install the N2O knob. Turn the knob so that the identification label is
horizontal before tightening the setscrews.
17. Turn the N2O needle valve counterclockwise, and check that the oxygen
flow increases as N2O flow increases.
18. Turn the O2 needle valve clockwise, and check that the N2O flow
decreases as O2 decreases.
19. Check the proportioning system concentration (increasing N 2O flow).
Observe the following precautions:
• Start with both valves at the minimum setting.
• Adjust only the N2O needle valve.
• Increase the N2O flow as specified in the table below and make sure the
O2 concentration is in the allowed range.
Note: Allow the O2 monitor to stabilize. At the lower flows, the O2 monitor
may take up to 90 seconds to stabilize.
• If you overshoot a setting, turn the O2 needle valve clockwise until the
N2O flow decreases to the previous setting before continuing the test.
1009-0356-000
07/04
Set the N2O flow (L/min)
Measured O2
0.8
22% to 29%
1
22% to 29%
2
22% to 29%
6
22% to 29%
9
22% to 29%
6-21
S/5 Aespire
20. Check the proportioning system concentration (decreasing O 2 flow).
Observe the following precautions:
• Turn the N2O needle valve to the maximum setting.
• Adjust only the O2 needle valve.
• Decrease the O2 flow as specified in the table and make sure the O2
concentration is in the allowed range.
Note: Allow the O2 monitor to stabilize. At the lower flows, the O2 monitor
may take up to 90 seconds to stabilize.
• If you overshoot a setting, turn the N2O needle valve counterclockwise
until the O2 flow increases to the previous setting before continuing the
test.
Set the O2 flow (L/min)
Measured O2
3
22% to 29%
2
22% to 29%
1
22% to 29%
0.3
22% to 29%
If both tests meet the criteria, calibration is complete (go to the next step).
If either test fails to meet the criteria, return to step 10 and adjust N 2O to
a lower or higher setting.
If:
Then:
Concentration Low
Decrease N2O
Concentration High
Increase N2O
Note: Adjusting the regulator pressure is not recommended. It has little
effect on proportioning. If you have difficulty proportioning the system, you
may need to replace either or both needle valves.
21. Tighten N2O sprocket setscrews.
22. Set the system switch to Standby.
23. Turn off the O2 and the N2O gas supplies.
24. Check that all setscrews are tight.
25. Adjust all needle valves to minimum stop position.
26. Install flowmeter panel shield.
6-22
07/04 1009-0356-000
6 Calibration
6.5 O2 Flush Regulator
1. Bleed all gas pressure for the machine (Section 4.2).
2. Ensure that all cylinder and pipeline gauges read zero before proceeding.
3. Remove the upper rear panel (Section 4.3).
4. Remove the O2 Flush Regulator output tubing. Attach a 6-mm tee and a
test device to the open port.
Remove output tubing
Attach 6-mm tee and test device
Jam nut
Adjustment screw
5. Connect an O2 pipeline supply or slowly open the O2 cylinder valve.
6. Push the flush button just enough to achieve a slight flow or open the
auxiliary flowmeter if equipped with this option. Read the pressure shown
on the test device.
The pressure should be 241 ±7 kPa (35 ±1.0 psi).
7. If adjustment is required:
a. Loosen the adjustment screw’s jam nut.
b. Adjust the regulator (in small steps) to the above specification.
c. Tighten the jam nut.
d. Verify the reading.
8. Disconnect the pipeline supply or close the cylinder valve.
9. Bleed gas pressure by pushing the flush button; then, disconnect the tee
and test device.
10. Reattach the output tubing to the regulator.
11. Install the rear panel.
1009-0356-000
07/04
6-23
S/5 Aespire
6.6 Airway pressure gauge
6.6.1 Zero the pressure
gauge
1. Attach a patient circuit to the Breathing System. Leave the patient end
open.
2. Set the Bag/Vent switch to Bag.
3. Adjust the APL valve to maximum.
4. Remove the lens from the pressure gauge:
• Apply a slight pressure with your thumb and finger tips around the outer
edge of the lens.
• Turn the lens counterclockwise to remove it.
5. Adjust the pressure gauge to zero.
6. Plug the patient circuit.
7. Press and release the O2 flush button to sweep the needle across the
pressure gauge.
8. Remove the plug from the patient circuit to relieve the pressure in the circuit
and recheck the zero setting of the pressure gauge.
9. If required, repeat zero and span procedure.
10. Replace the lens cover.
Zero Adjust
6-24
07/04 1009-0356-000
6 Calibration
6.6.2 Checking the
pressure gauge accuracy
The accuracy of the airway pressure gauge can be checked by using the
following:
• a low-pressure test device (digital manometer or test gauge) with an
accuracy of ±2% of reading,
• a low-pressure supply source (typically a syringe),
• and an airway pressure gauge test adapter.
1. Ensure that the pressure gauge is zeroed (Section 6.6.1).
2. Remove the upper rear panel.
3. Remove the existing tube from the back of the pressure gauge and connect
the test adapter tube directly to the gauge.
4. Connect a low-pressure supply source (syringe) to one of the open tubes of
the test adapter.
5. Connect a low-pressure test device to the remaining open tube of the test
adapter.
6. Adjust the pressure source to the following pressures as read on the airway
pressure gauge. The test device gauge should read within the values
indicated.
Airway Pressure Gauge
Test Device
0 cm H2O
0 ±1 cm H2O
40 cm H2O
40 ±2 cm H2O
-5 cm H2O
-5 ±2 cm H2O
To
Test Device
To
Low-pressure Source
(Syringe)
To
Pressure Gauge Fitting
1009-0356-000
07/04
6-25
Notes
6-26
07/04 1009-0356-000
7
In this section
Troubleshooting
This section covers the troubleshooting procedures for the Aespire machine pneumatic
systems. For troubleshooting electrical systems, refer to the 7100 Ventilator Technical
Reference manual.
7.1 General Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
7.2 Breathing System Leak Test Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
7.2.1 Breathing system leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
7.2.2 Breathing System Troubleshooting Flowcharts . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7
7.2.3 Leak Isolation Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
1009-0356-000
07/04
7-1
S/5 Aespire
7.1 General Troubleshooting
w WARNING
Objects in the breathing system can stop gas flow to the patient. This can
cause injury or death:
• Do not use a test plug that is small enough to fall into the breathing
system.
• Make sure that there are no test plugs or other objects caught in the
breathing system.
Problem
Possible Cause
Action
High Pressure Leak
Pipeline leak
Use a leak detector or Snoop to check for source of leak.
Repair or replace defective parts.
O2 flush valve
Use a leak detector or Snoop to check for source of leak.
Make sure tubing connections are tight.
Replace valve if defective.
System switch
Use a leak detector or Snoop to check for source of leak.
Make sure tubing connections are tight.
Replace switch if defective.
Cylinder not installed properly
Make sure cylinder is correctly aligned.
Verify that tee handles are tight.
Cylinder gauges
Use a leak detector or Snoop to check for source of leak.
Replace gauge if defective.
Cylinder gaskets
Use a leak detector or Snoop to check for source of leak.
Replace gasket if defective.
Relief valves
Use a leak detector or Snoop to check for source of leak.
Replace valve if defective.
Vaporizer not installed properly
Reseat vaporizer if not installed properly.
Have vaporizer serviced at vaporizer center if vaporizer
leaks.
Missing or damaged o-ring on vaporizer
manifold
Check condition of o-ring.
Replace if missing or damaged.
Loose fill port
Check fill port. Tighten if loose.
Low Pressure Leak
(with vaporizer mounted)
7-2
07/04 1009-0356-000
7 Troubleshooting
Problem
Possible Cause
Low Pressure Leak
(with or without vaporizer)
Leaking port valve on vaporizer manifold Use the Vaporizer Manifold Valve Tester to check for leak.
See Section 4.10.2 for instructions.
If test fails, tighten, repair, or replace as needed.
Leak at flowmeter head
Action
If vaporizer manifold passed previous tests:
Remove tubing from input side of head and occlude
port.
Perform leak test. If test fails:
• Check for damaged o-rings between flowmeter
modules. Replace as needed.
• Check for damaged o-rings at top and bottom of flow
tubes. Replace as needed.
• Check for cracked flow tube. Replace as needed.
If secondary regulator leaks, replace the complete
module.
Note: An alternate method is to pressurize the system
and use a leak detector or Snoop to check for source of
leak.
Bellows leak
Breathing System Leak
Breathing System Leak
(Intermittent)
Leaking relief valve on vaporizer
manifold
Remove relief valve.
Occlude opening.
Perform leak test.
If test passes, replace valve.
Leaking flush valve
Attach pressure measuring device on CGO.
Replace valve if device shows increased pressure.
Leaking system switch
Attach pressure measuring device on CGO.
Replace switch if device shows increased pressure.
Pop-off valve diaphragm not sealing
properly
Disassemble pop-off valve; inspect and clean seats;
reseat; reassemble.
Bellows mounting rim loose
Remove rim and pop-off valve diaphragm; reseat
diaphragm; snap rim (2) into place.
Bellows improperly mounted or has a
hole or tear
Check that only the last bellows convolute is mounted to
the rim and that the ring roll is in the groove under the
rim. Inspect the bellows for damage; replace.
Absorber canister not installed properly
Install canister properly.
Soda lime dust on canister seals
Clean seals and mating surfaces.
ACGO O2 sense check valve
Replace.
N2O flow does not decrease Defective pilot regulator
with O2 flow
Unit displays low O2
pressure with pipeline but
not with cylinders
Low O2 supply switch
Unable to begin mechanical ABS not fully engaged
ventilation
No O2 supply
Defective Bag/Vent switch
1009-0356-000
07/04
Check pilot regulator. Replace if needed.
Check switch. Calibrate or replace as appropriate.
Remount ABS.
Check O2 supply.
Check Bag/Vent switch.
7-3
S/5 Aespire
7.2 Breathing System Leak Test Guide
Note
Always perform the low-pressure leak test (Section 3.5) on the machine
before proceeding with these breathing system leak tests.
The procedure in Section 7.2.1 helps you isolate the leak: to Bag Mode
components, to Vent Mode components, or to components that are common
to both modes.
• If you have a similar leak in both the bag mode and the ventilator mode, you
must consider the Flow Sensor Module, the Circuit Module, the Absorber
Canister area, and the bulkhead components (including CGO tubing).
Carefully inspect the circuit module for damaged seals or misassembly, and
the seating of the O2 sensor.
• If you have a larger leak in one area than the other (Vent or Bag), the leak is
most likely NOT in the Flow Sensor Module, the Circuit Module, the Absorber
Canister area, or the bulkhead ports.
Follow the troubleshooting flowcharts in Section 7.2.2 to determine the best
sequence of tests for locating a breathing system leak.
The procedures in Section 7.2.3 test specific components of the breathing
system for leaks.
w WARNING
Objects in the breathing system can stop gas flow to the patient. This can
cause injury or death:
• Do not use a test plug that is small enough to fall into the breathing
system.
• Make sure that there are no test plugs or other objects caught in the
breathing system.
7-4
07/04 1009-0356-000
7 Troubleshooting
7.2.1 Breathing system
leak test
w WARNING
This test checks for leaks in Vent Mode and Bag Mode components. It is part
of the overall checkout procedure, Section 3.7 “Breathing system tests.” It is
repeated here for testing convenience.
Objects in the breathing system can stop gas flow to the patient. This can
cause injury or death:
• Do not use a test plug that is small enough to fall into the breathing
system.
1. Verify that AGSS is operating. For systems that have a flow indicator on the
side, make sure that the flow indicator shows a flow in the green (normal)
region.
2. Zero the pressure gauge (Section 6.6.1).
Check Valves
3. Make sure that the check valves on the breathing circuit module work
correctly:
a. The Inspiratory check valve rises during inspiration and falls at the start
of expiration.
b. The Expiratory check valve rises during expiration and falls at the start
of inspiration.
Ventilator Bellows
4. Ventilator bellows test:
a.
b.
c.
d.
Set the system switch to Standby.
Set the Bag/Vent switch to Ventilator.
Set all flow controls to minimum.
Close the breathing circuit at the patient connection. Use the test plug
located on the side of the ABS.
e. Push the O2 flush button to fill the bellows.
f. The pressure must not increase to more than 15 cm H2O on the
pressure gauge.
g. If the bellows falls more than 100 mL/min (top of indicator), it has a
leak.
Service Mode Tests
5. Enter the Service Mode: Push and hold the adjustment knob on the
ventilator’s display and set the system switch to On.
a. Select and confirm “Service Modes.”
b. Follow the menu structure outline below to reach the adjustment for the
inspiratory flow valve. Select and confirm at each step.
• “Diagnostics Tests/Tools”
• “Breathing System Leak Test”
c. Follow the instructions on the screen.
d. The leak rate should be less than 200 mL/min.
For machines with a single-tube O2 flowmeter, the pressure reading
should reach 30 cm H2O at minimum flows greater than 200 mL/min.
Note: If test fails, see Section 7.2, “Breathing System Leak Test Guide”.
1009-0356-000
07/04
7-5
S/5 Aespire
Bag Circuit
6. Test the Bag circuit for leaks:
a. Set the system switch to On.
b. Set the Bag/Ventilator switch to Bag.
c. Plug the Bag port (use your hand or the approved test plug).
d. Close the APL valve (70 cm H2O).
e. Set the O2 flow to 0.25 L/min.
f. Close the patient connection (using a hand or test plug on the side of
the breathing system) and pressurize the bag circuit with the O 2 flush
button to approximately 30 cm H2O.
g. Release the flush button. The pressure must not decrease. A pressure
decrease large enough to see on the gauge indicates an unacceptable
leak.
Note: If test fails, see Section 7.2, “Breathing System Leak Test Guide”.
APL Valve
7. Test the APL valve:
a. Fully close the APL valve (70 cm H2O).
b. Set the total fresh gas flow to approximately 3 L/min and make sure
that the value on the inspiratory pressure gauge does not exceed 85 cm
H2O. Some pressure fluctuation is normal.
c. Fully open the APL valve (to the MIN position).
d. Set O2 flow to 3 L/min. Turn any other gases off.
e. Make sure that the value on the inspiratory pressure gauge is less than
approximately 5 cm H2O.
f. Push the O2 flush button. Make sure that the value on the inspiratory
pressure gauge stays less than 10 cm H2O.
g. Set the O2 flow to minimum and make sure that the value on the
inspiratory pressure gauge does not decrease below 0 cm H 2O.
8. Remove your hand or the test plug from the patient connection.
9. Set the System switch to Standby.
w WARNING
7-6
Make sure that there are no test plugs or other objects caught in
the breathing system.
07/04 1009-0356-000
7 Troubleshooting
7.2.2 Breathing System
Troubleshooting
Flowcharts
Start
Review Leak Test Guide (Section 7.2)
• Do machine low-pressure leak test
(Section 3.8)
• Isolate breathing system leak to
Bag Mode,Vent Mode, or Both
(Section 7.2.1)
Perform Leak Test 1
Verifying the integrity
of test tools
Perform Test 2
Low-pressure leak
testing the machine
Fail
Repair leak in
machine
Go to Flowchart 2
Bag
Mode
only
Leak in
Bag Mode,
Vent Mode,
or Both?
Vent
Mode
only
Go to Flowchart 3
AB.74.042
Pass
Both Bag and
Vent Mode
Perform Test 3
Testing the pressure gauge
and port U-cup seals
Fail
Inspect/Replace tubing to
gauge; Replace U-Cup seals
Pass
Go to Flowchart 4
Flowchart
1009-0356-000
1
07/04
7-7
S/5 Aespire
Leak in
Bag Mode only
Fail
Check or Replace
the Bag/Vent
Switch lower Seal
Yes
Pass
Re-Install all breathing circuit
components and repeat
Breathing System Leak Test
(Section 7.2.1)
AB.74.043
Perform Test 4
Testing the bag port,
APL Valve, and Bag/Vent
Switch, and Negative
Pressure Relief Valve
Does the
Bellows
Inflate?
No
Perform Test 5
Testing the bag
port cover seal
Pass
Fail
Inspect
APL/BTV Manifold O-Ring seal
and Negative Pressure Relief;
replace as necessary.
Replace APL components
and repeat Test 4.
If test fails the second time,
inspect APL component;
replace as necessary.
Flowchart
7-8
2
07/04 1009-0356-000
7 Troubleshooting
Leak in
Vent Mode only
Perform Test 6
Testing the bellows assembly,
and Bag/Vent Switch
Perform Test 8
Testing the bellows and
bellows Pop-off
Inspect Exhalation Valve
and Drive Circuit
AB.74.044
Pass
Fail
Pass
Re-Install all breathing circuit
components and repeat
Breathing System
Leak Test
(Section 7.2.1)
Perform Test 7
Testing the Bag/Vent Switch,
Negative Relief Valve,
Bellows Base Manifold,
bellows and Pop-off Valve
Pass
Fail
Pass
Check / Replace the
following Bag/Vent parts:
Upper Seal, Disk, and O-Ring.
If test fails the second time,
inspect Negative Relief and
APL/BTV Manifold O-Ring.
Perform Test 8
Testing the bellows and
bellows Pop-off
Fail
Check/Replace:
Pop-off diaphragm,
Bellows Housing,
U-Cup seal,
and Bellows integrity
Flowchart
1009-0356-000
3
07/04
7-9
S/5 Aespire
Leak in both Bag and Vent Mode
Common Areas: Flow Sensor
Module, Circuit Module, Soda Lime
Canister, Negative Pressure Relief
Perform Test 9
Testing the Flow Sensor
Module, Circuit Module, and
Soda Lime Canister
Pass
Perform Test 13
Testing the
Negative Pressure
Relief
Perform Test 10
Testing the Circuit
Module and Canister
Pass
Fail
Perform Test 11
Testing the Circuit
Module
Perform Test 14 and 15
Testing the Flow
Sensors, Bulkhead
connectors and tubing
Flowchart
7-10
Fail
Pass
Replace the Negative
Pressure Relief,
Bag/Vent Cartridge
O-Rings
Check/Replace the following
Bag/Vent parts:Upper Seal,
Disk, and O-Ring
Pass
Fail
Inspect Canister for
leaks
Perform Test 12
Testing the Inspiratory
side of the Circuit
Module and Canister
Pass
Fail
Inspect Expiratory
side of Circuit Module.
Repeat Test 11
Inspect Inspiratory
side of Circuit Module.
Repeat Test 11
AB.74.045
Fail
4
07/04 1009-0356-000
7 Troubleshooting
Leak in
Flow Sensor Module
or Circuit Module
Perform Tests 14 and 15
Testing the Flow Sensors
using the low-pressure
leak test device
Pass
Check/Replace
Flow Sensors or
bulkhead connector
O-Rings
Swap Module(s)
with another machine.
Repair or replace Module
Flowchart
1009-0356-000
Check/Replace the seals
on the Circuit Module and
the O-Ring on the O2 cell
Fail
Perform Breathing
System Leak Test
(Section 7.2.1)
Pass
AB.74.048
Fail
Done
5
07/04
7-11
S/5 Aespire
7.2.3 Leak Isolation Tests
The previous flowcharts refer you to the following tests.
These tests require the use of the Low Pressure Leak Test Device and the Leak
Test Tool Kit (refer to Section 8.1, "Service tools — Anesthesia machine".
The Leak Test Tool Kit includes:
• the Machine Test Tool
• the Circuit Test Tool
• and various Test Plugs
When performing these tests, ensure that the ACGO selector switch is set to
the ABS position.
Test 1
Verifying the integrity of the test tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
Test 2
Low-pressure leak testing the machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
Test 3
Testing the airway pressure gauge, and Port 1 and Port 3 u-cup seals . . . . . . . . . 7-15
Test 4
Testing the bag port cover, the APL valve, the Bag/Vent switch, and the negative
pressure relief valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
Test 5
Testing the APL diaphram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17
Test 6
Testing the bellows module and the Bag/Vent switch . . . . . . . . . . . . . . . . . . . . . . 7-18
Test 7
Testing the bellows, the bellows pop-off valve, the bellows base manifold, and the
Bag/Vent switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
Test 8
Testing the bellows assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-20
Test 9
Testing the flow sensor module, the circuit module, and the soda lime canister . 7-21
Test 10
Testing the circuit module and the canister . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-22
Test 11
Testing the circuit module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-22
Test 12
Testing the inspiratory side of the circuit module . . . . . . . . . . . . . . . . . . . . . . . . . . 7-23
Test 13
Testing the negative pressure relief valve. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
Test 14
Testing the flow sensors only. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-25
Test 15
Testing a flow sensor including the Ventilator Monitoring Assembly and interfacing
components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-26
w WARNING
Objects in the breathing system can stop gas flow to the patient. This can
cause injury or death:
• Do not use a test plug that is small enough to fall into the breathing
system.
• Make sure that there are no test plugs or other objects caught in the
breathing system.
w CAUTION
7-12
Do not use O2 Flush for leak isolation tests. Do not leave pressurized
systems unattended. High pressure and equipment damage may result.
07/04 1009-0356-000
7 Troubleshooting
Test 1
Verifying the integrity of the test tools
Machine Test Tool
Front View
Bulb Thru Port
(not connected to airway path)
Pressure Sense
Port
Airw
ay P
ath
Fresh Gas Port
Alignment Post
Back View
Bulb Thru Port
1. Verify integrity of low-pressure leak test device.
ath
ay P
Airw
Alignment Post
• Put your hand on the inlet of the leak test device. Push hard for a good
seal.
• Squeeze the bulb to remove all air from the bulb.
• If the bulb completely inflates in less than 60 seconds, replace the leak
test device.
2. Attach the low-pressure leak test device to the Machine Test Tool.
3. Plug the two pressure orifices.
4. Repeatedly squeeze and release the hand bulb until it remains collapsed.
5. If the bulb inflates in less than 30 seconds, locate and correct the leak.
1009-0356-000
07/04
7-13
S/5 Aespire
Test 2
Low-pressure leak testing the machine
1. Remove the breathing system from the machine.
2. Attach the Machine Test Tool (using only the Thru Port) and the lowpressure leak test device to Port 3 of the breathing system interface as
shown above.
Note: To prevent damage to the airway pressure gauge, ensure that the
gauge port (Port 1) is not connected to the Test Tool.
3. Set the ACGO selector switch to ABS.
4. Set the system switch to Standby.
5. Turn off all vaporizers.
6. Open the flow controls one and a half turns counterclockwise.
7. Compress and release the bulb until it is empty.
8. The vacuum causes the floats to move. This is usual. If the bulb completely
inflates in 30 seconds or less, there is a leak in the low-pressure circuit.
7-14
07/04 1009-0356-000
7 Troubleshooting
Test 3
Testing the airway pressure gauge, and Port 1 and Port 3 u-cup seals
Alignment
Post
1. Attach the Machine Test Tool to the breathing system interface ports (using
the alignment post) as shown above.
2. Turn all of the flow controls fully clockwise (minimum flow).
3. Set the system switch to On.
4. Occlude the tapered plug.
• At minimum flow, the airway pressure gauge reading should increase.
• If not, there is a leak in the tested circuit.
1009-0356-000
07/04
7-15
S/5 Aespire
Test 4
Testing the bag port cover, the APL valve, the Bag/Vent switch, and
the negative pressure relief valve
Plug
Bag Port
1. Separate the Bellows Module from the Circuit Module and re-install the
Bellows Module.
2. Occlude the Bag Port connector.
3. Connect the Machine Test Tool to the interface ports as shown above.
4. Set the Bag /Vent switch to Bag and close the APL Valve (70 cm H 2O).
5. Slowly increase the O2 flow to achieve 30 cm H2O.
• The leak rate is equal to the flow needed to maintain 30 cm H 2O.
• The leak rate should be less than 200 mL/min.
Note: If the bellows rises, it indicates a leak in the Bag /Vent Switch.
7-16
07/04 1009-0356-000
7 Troubleshooting
Test 5
Testing the APL diaphram
Plug
APL
Scavenging
Port
Plug
Bag Port
Note
If required, set up the Machine Test Tool and breathing system as shown in
Test 4.
1. Slide the Bellows Module away from the machine.
2. Remove the APL ramp and diaphragm.
3. Insert a Test Plug into the APL scavenging port, as shown above.
4. Slide the Bellows Module partially back onto the machine casting.
5. Ensure that the Bag Port is plugged and that the Bag/Vent switch is set to
Bag.
6. Slowly increase the O2 flow to achieve 30 cm H2O.
• Leak rate is equal to the flow needed to maintain 30 cm H 2O.
• The leak rate should be less than 200 mL/min.
1009-0356-000
07/04
7-17
S/5 Aespire
Test 6
Testing the bellows module and the Bag/Vent switch
1. Separate the Bellows Module from the Circuit Module and re-install the
Bellows Module.
2. Enter the Service Mode: Push and hold the adjustment knob on the
ventilator’s display and set the system switch to On.
a. Select and confirm “Service Modes.”
b. Follow the menu structure outline below to reach the adjustment for the
inspiratory flow valve. Select and confirm at each step.
• “Diagnostics Tests/Tools”
• “Breathing System Leak Test”
3. Follow the instructions on the screen.
• At step 5, connect the Machine Test Tool to the interface ports as shown
above, instead. Continue with steps 6 and 7 on the screen.
• The leak rate should be less than 200 mL/min.
7-18
07/04 1009-0356-000
7 Troubleshooting
Test 7
Testing the bellows, the bellows pop-off valve, the bellows base
manifold, and the Bag/Vent switch
1. Separate the Bellows Module from the Circuit Module.
2. Insert appropriate test plugs into the bellows base manifold as shown to
the left.
Note: Position the bellows assembly so that the bellows remain collapsed
as you plug the ports.
Plug
Plug
3. Set Bag/Vent switch to Vent.
4. Position the bellows upright with the bellows collapsed.
5. Connect the Machine Test Tool to the interface ports as shown above.
6. Slowly increase the O2 flow to achieve 30 cm H2O.
Note: The bellows will rise until the pressure equalizes.
• The leak rate is equal to the flow needed to maintain 30 cm H 2O.
• The leak rate should be less than 200 mL/min.
1009-0356-000
07/04
7-19
S/5 Aespire
Test 8
Note
Testing the bellows assembly
If required, set up the Machine Test Tool and breathing system as shown in
Test 7.
1. Remove the bellows base manifold from the Bellows Module.
2. Insert appropriate test plugs into the bellows base manifold as shown to
the left.
Plug
Note: Position the bellows assembly so that the bellows remain collapsed
as you plug the ports.
Plug
3. Connect the tapered plug of the Machine Test Tool to the bellows base inlet
as shown to the left.
Fresh Gas
4. Position the bellows upright with the bellows collapsed.
5. Slowly increase the O2 flow to achieve 30 cm H2O.
Note: The bellows will rise until the pressure equalizes.
• The leak rate is equal to the flow needed to maintain 30 cm H 2O.
• The leak rate should be less than 200 mL/min.
7-20
07/04 1009-0356-000
7 Troubleshooting
Test 9
Testing the flow sensor module, the circuit module, and the soda
lime canister
Plug
1. Separate the Bellows Module from the Circuit Module and re-install the
Circuit/Flow Sensor Module.
2. Connect short tubing between the inhalation and exhalation ports of the
breathing system.
3. Insert an appropriate test plug in the outlet port of the Circuit Module.
4. Enter the Service Mode: Push and hold the adjustment knob on the
ventilator’s display and set the system switch to On.
a. Select and confirm “Service Modes.”
b. Follow the menu structure outline below to reach the adjustment for the
inspiratory flow valve. Select and confirm at each step.
• “Diagnostics Tests/Tools”
• “Display A/D Channels”
c. Record the Inspiratory and Expiratory Flow actual values.
Note: The Inspiratory and Expiratory Flow actual values should be near
zero.
5. Slowly increase the O2 flow to achieve 30 cm H2O.
• The leak rate is equal to the flow needed to maintain 30 cm H 2O.
• The leak rate should be less than 200 mL/min.
6. Observe the Inspiratory and Expiratory Flow actual values. The values
should be near zero, as previously recorded in step 4.
Note: If one channel indicates flow, see “Inaccurate Volume Ventilation
Troubleshooting” in the 7100 Ventilator Service Manual.
7. Release Pressure.
1009-0356-000
07/04
7-21
S/5 Aespire
Test 10
Testing the circuit module and the canister
1. Remove the Flow Sensor module.
2. Connect the Circuit Test Tool to the Circuit Module as shown above.
3. Slowly increase the O2 flow to achieve 30cmH2O. The leak rate is equal to
the flow required to maintain 30 cmH2O.
• The leak rate should be less than 200 mL/min.
Test 11
Testing the circuit module
Plug
Plug
Plug
Note: If required, set up the machine as in Test 10.
1. Remove the Soda Lime Canister.
2. Using appropriate Test Plugs, plug the three canister ports in the Circuit
Module as shown above.
3. Slowly increase the O2 flow to achieve 30 cm H2O.
• The leak rate is equal to the flow needed to maintain 30 cm H 2O.
• The leak rate should be less than 200 mL/min.
7-22
07/04 1009-0356-000
7 Troubleshooting
Test 12
Testing the inspiratory side of the circuit module
Plug
Loop
Note: If required, set up the machine as in Test 10 and 11.
1. Connect the Circuit Test Tool to the Circuit Module as shown above.
2. Insert an appropriate test plug in the inspiratory outlet to the canister as
shown above.
3. Slowly increase the O2 flow to achieve 30 cm H2O.
• The leak rate is equal to the flow needed to maintain 30 cm H 2O.
• The leak rate should be less than 200 mL/min.
1009-0356-000
07/04
7-23
S/5 Aespire
Test 13
Testing the negative pressure relief valve
1. Separate the Bellows Module from the Circuit Module.
2. Remove the Bellows Interface Manifold.
3. Insert test plug (recessed end) into the rear Bag/Vent switch port as shown.
Plug
4. Install the Bellows Module.
5. Connect the Machine Test Tool to the interface ports and the Bellows
Module as shown above.
6. Set the Bag/Vent Switch to Vent.
7. Slowly increase the O2 flow to achieve 30 cm H2O.
• The leak rate is equal to the flow needed to maintain 30 cm H 2O.
• The leak rate should be less than 200 mL/min.
7-24
07/04 1009-0356-000
7 Troubleshooting
Test 14
Testing the flow sensors only
1. Remove the Flow Sensor Module.
2. Plug each Flow Sensor as shown above.
3. Connect the low-pressure leak test device to the open end of the Flow
Sensor.
4. Block the connector end of the Flow Sensor with your hand.
5. Compress and release the bulb until it is empty.
6. If the bulb inflates in 30 seconds or less, there is a leak in the flow sensor.
7. If there are no leaks in the flow sensors, go to Test 15.
1009-0356-000
07/04
7-25
S/5 Aespire
Test 15
Testing a flow sensor including the Ventilator Monitoring Assembly
and interfacing components
1. Remove Flow Sensors from the Flow Sensor Module.
2. Attach the Flow Sensor to the bulkhead connector.
3. Plug each Flow Sensor as shown.
4. Connect the low-pressure leak test device to the open end of the Flow
Sensor.
5. Compress and release the bulb until it is empty.
6. If the bulb inflates in 30 seconds or less, there is a leak. The leak may be
through the connector o-rings, in the internal tubbing, or in the Transducer
on the VMB.
7-26
07/04 1009-0356-000
8
In this section
Illustrated Parts
8.1 Service tools — Anesthesia machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
8.1.1 Test Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
8.1.2 Test Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-4
8.1.3 Secondary regulator pilot pressure tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
8.2 External components - front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-6
8.3 External components - front view references . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
8.4 External Components - rear view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-8
8.5 Control module mounting for a ProTIVA machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-9
8.6 Aespire 100 - exclusive components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
8.6.1 AC Inlet (Aespire 100). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-12
8.6.2 Display mount (Aespire 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-13
8.7 Front panel, gauges and system switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
8.8 Rear panel components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-15
8.9 Tabletop components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-16
8.10 Right-side Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-17
8.11 External components - lower assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18
8.12 Vent Engine Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-19
8.13 Display cables, serial board, AGSS flowtube, and sample return . . . . . . . . . . . . . . . . . 8-20
8.14 AC Power cords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-21
8.15 AC Inlet/Outlet Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-22
8.16 Pipeline inlet fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-24
8.17 Cylinder Gas Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-25
8.17.1 Cylinder inlet fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-26
8.18 Vaporizer manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-27
8.19 Flowmeter components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-28
8.19.1 Flowtube parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-30
8.19.2 Secondary regulator components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-32
8.20 ABS to machine Interface Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-34
8.20.1 Flush Regulator, Flush Valve, and ACGO Selector Switch . . . . . . . . . . . . . . . . . 8-35
1009-0356-000
07/04
8-1
S/5 Aespire
8.21 Breathing system interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-36
8.22 Breathing System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-37
8.22.1 APL Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.2 Bag/Vent Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.3 Absorber canister . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.4 Flow Sensor Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.5 Breathing Circuit Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.6 Exhalation valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.7 Bellows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.8 Bellow base . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.22.9 Bag Arms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.23 Drawer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8-37
8-38
8-39
8-40
8-41
8-42
8-43
8-44
8-45
8-46
8.24 Legris quick-release fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-47
8.25 Vent Drive and low-pressure tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-48
8.26 Tubing for use with Legris fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-50
8.27 Cables and harnesses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-52
8.28 Cables and harnesses (Aespire 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-54
8.29 Anesthetic Gas Scavenging System — AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-56
8.29.1 Passive AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.29.2 Adjustable AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.29.3 Active AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.30 Integrated Suction Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8-56
8-58
8-60
8-62
8.30.1 Major Components (Continuous and Venturi suction) . . . . . . . . . . . . . . . . . . . .
8.30.2 Suction Control Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.30.3 Venturi assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8.31 Auxiliary O2 Flowmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8-62
8-63
8-64
8-65
8.32 Display mounts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-66
8.33 Cable management arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-67
8.34 Display arm mounting kits for optional equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-68
8-2
07/04 1009-0356-000
8 Illustrated Parts
8.1 Service tools — Anesthesia machine
8.1.1 Test Devices
Item
Tool
Stock Number
1
Test flowmeter, 6–50 L/min (Suction Flow Test)
1006-8431-000
Not Shown
Low-pressure Leak Test Device
(negative pressure)
0309-1319-800
Low-pressure Leak Test Device
(positive pressure - ISO)
1001-8976-000
Low-pressure Leak Test Device
(positive pressure - BSI)
1001-8975-000
Flow test device capable of measuring 0–15 L/min
with an accuracy of ±2% of reading
Refer to section 3.8
Vacuum test gauge capable of measuring 0 to 550 mm Hg
with an accuracy of ±1% of reading
Refer to section 3.9
Test device capable of measuring 0–30 L/min (see Item 1 above)
Refer to section 3.9
Leakage current test device
Refer to section 3.11
Test device capable of measuring 689 kPa (100 psi)
Refer to section 6.1
Low-pressure test device (digital manometer or test gauge)
with an accuracy of ±2% of reading
Refer to section 6.6.2
Test Devices
1
1009-0356-000
07/04
8-3
S/5 Aespire
8.1.2 Test Tools
Item
Tool
1
Leak Test Tool Kit, ABS breathing system
1a Test Tool, bulkhead
1b Plug, tapered 27x12 mm
1c Plug, tapered 24x18 mm
1d Test Tool, circle module (2 each)
1e Plug, service B/S 11 mm (2 each)
1f Plug, service BTV 18 mm (2 each)
2
Plug, stopper
3
Adapter, positive low-pressure leak test
4
Vaporizer Manifold Valve Test Tool
Not Shown
Tool to help disconnect tubing from Legris fittings
Test Lung
Leak detection fluid, Snoop
Stock Number
1407-7013-000
1407-8500-000
1407-8505-000
1407-8506-000
1407-8502-000
1407-8504-000
1407-8503-000
2900-0001-000
1009-3119-000
1006-3967-000
2900-0000-000
0219-7210-300
obtain locally
1b
1d
1c
1a
1e
(2)
1f
(2)
3
4
2
8-4
07/04 1009-0356-000
8 Illustrated Parts
8.1.3 Secondary
regulator pilot pressure
tool
Assemble the secondary regulator pilot pressure tool using a 4-mm tee and
tubing as shown. This tool is used with N2O needle valve calibration.
1
3
2
Item
Description
Stock Number
1
Tee, 4 mm, tube/tube/tube
1202-3653-000
2
Tubing, 4 mm (approximately 450 mm - 18 inches)
1001-3060-000
3
Tubing, 4 mm (approximately 50 mm - 2 inches)
1001-3060-000
1009-0356-000
07/04
8-5
S/5 Aespire
8.2 External components - front view
4 (5, 6)
AB.74.016
3
16, 17
18, 19
11
7
AB.74.005
12 (13)
9
10 (15, 13)
8
2
1
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
8-6
14 (13)
Description
Caster, 125-mm with brake (front)
Caster, 125-mm no brake (rear)
Cover, cable channel
Upper shelf
Bolt, M6x40
Lockwasher, M6 internal
Gauge, airway pressure (includes mounting hardware)
Handle, side
Handle, Medirail
Screw, M6x12 Sems
Spacer
Screw. M6x70
Lockwasher M6 external
Screw, M6x20
Shim
Task Light PCB
Lens, Task Light
Screw
Switch Assembly, task light
Plate, switch mounting retainer
Screw
Stock Number
1006-3070-000
1006-3071-000
1009-3020-000
1009-3022-000
0144-2131-911
0144-1118-130
1009-3034-000
1009-3033-000
1009-3101-000
0144-2436-106
1009-3102-000
0144-2131-923
9213-0560-003
0144-2131-921
1009-3131-000
1009-5504-000
1011-3308-000
0142-4254-106
1009-5587-000
1009-3143-000
0140-6226-107
07/04 1009-0356-000
8 Illustrated Parts
8.3 External components - front view references
1
9
AB.74.016
8
14
2
10
15
11
AB.74.005
7
3
4
12
13
16
6
5
16
1009-0356-000
Item
Description
Section number
1
“Serial board”
Refer to section 8.13
2
“AGSS gauge, and sample return”
Refer to section 8.13
3
“Vent Engine Housing”
Refer to section 8.12
4
“Anesthetic Gas Scavenging System — AGSS”
Refer to section 8.29
5
“Breathing System”
Refer to section 8.22
6
“Drawer”
Refer to section 8.23
7
“Tabletop components”
Refer to section 8.9
8
“Display cables”
“Display mounts”
Refer to section 8.13
Refer to section 8.32
9
“Flowmeter components”
Refer to section 8.19
10
“Auxiliary O2 Flowmeter”
Refer to section 8.31
11
“Integrated Suction Regulator”
Refer to section 8.30
12
“Front panel, gauges and system switch”
Refer to section 8.7
13
“ABS to machine Interface Components”
Refer to section 8.20
14
“Vaporizer manifold”
Refer to section 8.18
15
“Right-side Components”
Refer to section 8.10
16
“External components - lower assembly”
Refer to section 8.11
07/04
8-7
S/5 Aespire
8.4 External Components - rear view
4
5
AB.74.004
6
2
1
8-8
3
7
Item
Description
Stock Number
1
AC Inlet
Refer to section 8.14
2
Pipeline Inlets
Label, pipeline inlet blank
Refer to section 8.16
1009-3197-000
3
Cylinder Gas Supplies
Refer to section 8.17
4
Electrical Power Outlet
Blank panel (no outlets)
Screw, M4x12
Refer to section 8.15
1011-3329-000
1009-3109-000
5
Suction items
Refer to section 8.30
6
Rear panel items
Refer to section 8.8
7
Thumbscrew
Ring, retainer
1406-3304-000
1406-3319-000
07/04 1009-0356-000
8 Illustrated Parts
8.5 Control module mounting for a ProTIVA machine
6
8
AB.97.002
7
4
5
3
1
1
2
Item
Description
Stock Number
1
Thumbscrew, M6x14
1505-3005-000
2
Bracket, mount 7100 display TIVA
1009-3290-000
3
Screw, M4x12 Pozidriv
Lockwasher, M4 external
0140-6226-111
9213-0540-003
4
Block, mount 7100 display TIVA
1009-3294-000
5*
Screw, M4x8 Pozidriv Flat HD
0140-6226-107
6
Clamp, cable TIVA
1009-3291-000
7
Screw, M6x14 SKT HD CAP
0144-2131-922
8
Manifold, display mount TIVA
1009-8232-000
* Apply Loctite 242.
1009-0356-000
07/04
8-9
S/5 Aespire
8.6 Aespire 100 - exclusive components
8-10
Item
Description
Stock Number
1
AC Inlet
Refer to section 8.6.1
2
Display mount components
Refer to section 8.6.2
3
Power supply, 40W universal
(the display/control module in the Aespire 100
uses a different power supply than in a standard
Aespire machine)
1609-3040-000
4
Flowhead components
(with the following exceptions, all components of
the O2 only flowhead are covered in the
“Flowmeter components” section.
Refer to section 8.19
4a
Knob, O2 (does not include label)
1006-3634-000
4b
Plate, flowtube blank
1006-1506-000
4c
Screw, tread forming 0.88 inch
(this screw is slightly longer to
compensate for the flowtube blank plate)
1009-3410-000
5
Blank Label Set
(when replacing the flowmeter bezel, use
appropriate label to cover unused flow control
positions)
1009-3409-000
6
Harnesses
Refer to section 8.28
7
Caster, non-locking 4-inch
1009-3038-000
8
Caster, locking 4-inch
1009-3039-000
9
Panel, lower rear cover
1009-3407-000
10
Screw, M4x8
1006-3178-000
11
Thumbscrew
1406-3304-000
12
Ring, retaining
1406-3319-000
07/04 1009-0356-000
8 Illustrated Parts
1
2
Power Supply
4
3
4c
4b
5
4a
6
10
7
9
11 (12)
8
1009-0356-000
07/04
8-11
S/5 Aespire
8.6.1 AC Inlet (Aespire 100)
11
12
13
10
9
2
4
7
8
3
1
6
5
8-12
Item
Description
Stock Number
1
AC Inlet, with line filter
1609-3019-000
2
Fuse drawer
1609-3020-000
3
Fuse, 2A
1009-5778-000
4
Screw, M3x8
9211-0530-083
5
Nut, M3 Keps
0144-3717-302
6
Transient Suppression board
1609-3101-000
7
Standoff, 52-mm Long
1609-3320-000
8
Screw, M4x8
0140-6226-113
9
Stud, Equal Potential, 6-mm
0208-0070-300
10
Screw, M4x12
1009-3109-000
11
Power Cord
Refer to section 8.14
12
Clamp, power cord retainer
1009-3405-000
13
Screw, M4x8
1006-3178-000
07/04 1009-0356-000
8 Illustrated Parts
8.6.2 Display mount (Aespire 100)
4
5
6
1
11 (12)
2
7
7
3
6
10
13
9
8
Item
Description
Stock Number
1
Dovetail insert
1009-3408-000
2*
Screw, M8x20 SKT HD
0144-2440-821
3*
Screw, M6x12 SKT HD
0144-2436-101
4
Extrusion, Vent bracket mount
1504-3514-000
5
Bolt, carriage
1006-1433-000
6
Casting, Vent Bracket, 20 degree (2)
1504-3526-000
7
Bearing, white plastic (2)
1006-3228-000
8
Handle, T-clamping
1301-3001-000
9
Washer, flat
9213-0180-006
10
Set screw, M6x10
0141-4227-111
11
Screw, M4x16
9211-0440-163
12
Lockwasher, M4
0144-1118-128
13
Clamp
1504-3527-000
* Apply Loctite 242
1009-0356-000
07/04
8-13
S/5 Aespire
8.7 Front panel, gauges and system switch
9
7
3
1
4
3
10
2
6
5
11
8
7
5
Item
Description
Stock Number
1
Panel, gauge front
1009-3018-000
2
Switch, D-O system
1006-8452-000
3
Gauge, low pressure (includes mounting hardware)
1009-3079-000
4
Connector, 1/8 inch Legris to 10-32
1006-3711-000
5
Gauge, high pressure (includes mounting
hardware)
1009-3080-000
6
Connector, 1/8 inch copper tube to 5/16-24
1006-3712-000
7
Label, gauge
Stock Number
(cylinder)
O2 ANSI
N2O ANSI
Air ANSI
1009-3081-000
1009-3082-000
1009-3083-000
1009-3199-000
1009-3201-000
1009-3200-000
O2 ISO
N2O ISO
Air ISO
1009-3202-000
1009-3082-000
1009-3203-000
1009-3204-000
1009-3201-000
1009-3205-000
O2 Neutral
N2O Neutral
Air Neutral
1009-3234-000
1009-3235-000
1009-3236-000
1009-3237-000
1009-3238-000
1009-3239-000
8
Tube Kit, copper tube and fittings (inboard cylinder)
Tube Kit, copper tube and fittings (3rd gas)
1006-8371-000
1006-8372-000
9
Plate, gauge blanking
1009-3045-000
10
Plate, gauge blank backing
1009-3147-000
11
Palnut
1009-3090-000
8-14
Stock Number
(pipeline)
07/04 1009-0356-000
8 Illustrated Parts
8.8 Rear panel components
1
2 (3)
4
8 (9)
5
14 (15, 16)
17
6 (7)
13 (15, 16)
10 (9)
12 (9)
11 (9)
Item
Description
Stock Number
1
Cover, rear upper
1009-3073-000
2
Cap, hose reel
1009-3075-000
3
Screw, M5.5x20
1009-3384-000
4
Strap, hook/loop
1009-3233-000
5
Screw, M6x1.0 captive
1009-3114-000
6
Door, access (not functional for Aespire)
1009-3074-000
7
Screw, M4x12
1009-3109-000
8
Spring, cantilever
1009-3124-000
9
Screw, M3x8
0142-4254-106
10
Cover, trap bottle (if no internal suction)
1009-3173-000
11
Cover, regulator yoke (if no regulator)
1009-3121-000
12
Plate, clip cover
1009-3185-000
13
Wrench, DIN cylinder (with cable)
1202-3651-000
14
Wrench, pin index cylinder (with cable)
0219-3415-800
15
Cable
1010-3049-000
16
Ferrule, cylinder wrench cable retainer
1001-3708-000
17*
Handle, P-grip
1009-3343-000
* Clean mounting surface with isopropyl alcohol.
1009-0356-000
07/04
8-15
S/5 Aespire
8.9 Tabletop components
6
7 (8, 9)
1
2 (3)
11
10
4 (5)
8-16
Item
Description
Stock Number
1
Tabletop, work surface
1009-3029-000
2
Screw, relieved
1504-3001-000
3
Washer, retainer
1009-3178-000
4
Window, check-valve
1009-3088-000
5
Palnut
1009-3090-000
6
Hook, breathing circuit
1009-3086-000
7
Bolt, shoulder
1009-3172-000
8
Washer, wave
1009-3035-000
9
Washer, Nylon
1009-3150-000
10
Clip, with tape (used with bag arm)
1009-8196-000
11
Clip, with tape (used with bag on hose)
1009-8197-000
07/04 1009-0356-000
8 Illustrated Parts
8.10 Right-side Components
5
4
6
7
1
1009-0356-000
2 (3)
07/04
Item
Description
Stock Number
1
Extrusion cover
1009-3021-000
2
Screw, M6x20
0144-2131-921
3
Lockwasher, M6 internal
0144-1118-130
4
Dovetail, RH upright
1009-3129-000
5
Screw, M4x10 self-tapping
1009-5534-000
6
Cover, pipeline inlet
1009-3091-000
7
Screw, M4x8
1006-3178-000
8-17
S/5 Aespire
8.11 External components - lower assembly
1 (3)
10 (11, 12)
13 (8, 9)
2 (3)
7 (8, 9)
6
4 (5)
8-18
Item
Description
Stock Number
1
Panel, access
1009-3059-000
2
Panel, service
1009-4141-000
3
Screw, M4x8
1006-3178-000
4
Thumbscrew
1406-3304-000
5
Ring, retaining
1406-3319-000
6
Cover, scavenger reservoir
1009-3027-000
7
Bracket, suction reservoir
1009-3107-000
8
Screw, M4x16
9211-0440-163
9
Lockwasher, M4 external
9213-0540-003
10
Clip, suction bag hose
1407-3327-000
11
Screw, M5x16 PAN HD
9211-8350-163
12
Lockwasher, M5 external;
0144-1118-220
13
Bumper, absorber
1009-3105-000
07/04 1009-0356-000
8 Illustrated Parts
8.12 Vent Engine Housing
1
7
6
8
5
4
2
10 (11)
3
9
12
13 (14, 15)
8
1009-0356-000
Item
Description
Stock Number
1
Vent Engine Cover Plate Assy
1407-7009-000
2
TAB GUIDE BELLOWS BASE
1407-3313-000
(2)
3
SCR M3X16 POSI DR PAN HD A4 SST
1504-3003-000
(2)
4
Cap, Plug
1406-3524-000
5
FITTING PNL MOUNT 3.18 HOSE BARB UNION
1504-3014-000
6
PLUG HOLE 15.9 DIA NYLON MICRO PLASTICS
1006-1473-000
7
PLATE CONN VENT
1407-3321-000
8
SCR M4X8 POZI-DR DIN84 PAN SERRATED
1006-3178-000
9
HARN VENT ENG BRD TO COIN SLOT
1009-5545-000
10
BLOCK LATCHING DSUB CONN
1504-3617-000
(2)
11
SCR 4-40 X 3/8 SKT BCG HD CAP
0144-2117-206
(2)
12
Bracket, scavenging guard in machines without scavenging
1407-3922-000
13
CLIP-SUCTION BAG HOSE
1407-3327-000
14
SCR M5 X 16 PAN PH HD SST
9211-8350-163
(2)
15
Lockwasher
0144-1118-220
(2)
07/04
Qty
(2)
(5)
8-19
S/5 Aespire
8.13 Display cables, serial board, AGSS flowtube, and sample return
8
10
9
1
2
3
1
2
3
4
5
6
7
4
13
12
11
14
15
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
8-20
Description
CABLE SER ISLN CONN BRD TO CTRL MOD 7100
CABLE MONITORING BOARD
CABLE PNEUMATIC ENGINE
Cable, power
“Harness, Serial ISO to O2 flush SW”
“Harness, Serial ISO to O2 supply SW”
“Harness, Serial ISO to on/ standby SW”
Serial Isolation Board
Standoff
Lockwasher
Screw, thread forming
Label, flow indicator AGSS
Label, flow indicator AGFS (for German variant)
Label, blank (for machines without flow indicator)
Flowtube, AGSS
Clip, AGSS flowtube
Coupling, Colder (Kit includes mounting nut)
Label set, blank (for machines without scavenging)
Label, AGFS (for German variant)
Stock Number
1009-5691-000
1504-5604-000
1504-5605-000
1009-5711-000
1009-5567-000
1009-5568-000
1009-5566-000
1009-5500-000
1202-3092-000
0144-1104-331
1009-3400-000
1406-3527-000
1009-3301-000
1009-3241-000
1406-3560-000
1009-3181-000
1009-8321-000
1009-3351-000
1009-3300-000
07/04 1009-0356-000
8 Illustrated Parts
8.14 AC Power cords
Aespire 100 machine
3
1
1
3
2
Item
Description
1
Power Cord
2
3
4
1009-0356-000
4
07/04
Stock Number
Australia and China, 220-240 VAC AS 3112 outlets
1006-3888-000
EURO and France, 220 VAC with CEE 7/7
1001-3380-000
India and South Africa, 220-240 VAC BS546
1006-3885-000
Japan and US, 100-120 VAC NEMA
1006-3907-000
Peru, 220-240 VAC NEMA
1006-3882-000
Swiss, 220-240 VAC SEV 1011
1006-3889-000
UK, 220-240 VAC BS1363
1006-3884-000
Clamp, power cord retainer
Screw, M4x8 PAN serrated
Clamp, power cord retainer
1009-3103-000
1006-3178-000
1009-3405-000
8-21
S/5 Aespire
8.15 AC Inlet/Outlet Components
Item
Description
Stock Number
1
Inlet, 100/120 AC, with line filter and 15 A circuit breaker
Inlet, 220/240 AC, with line filter and 8 A circuit breaker
1009-5698-000
1009-5757-000
2
Fuse, 2A - 5x20mm
Fuse holder
1009-5778-000
1009-5674-000
3
Circuit board, Inrush, 100-120V
1006-3245-000
Circuit board, Inrush, 220-240V
1006-3246-000
Circuit board, Transient Suppression, 100-120V
1006-3788-000
Circuit board, Transient Suppression, 220-240V
1006-3789-000
4
Filter, AC Line, 100-240V
1009-5675-000
5
Cable, line filter to Display Module
1009-5711-000
6
Harness, 100/120 V to Toroid (or to outlets harness)
Harness, 220/240 V to Toroid (or to outlets harness)
Harness, if no outlets
1009-5713-000
1009-5714-000
1009-5715-000
7
Harness, to 100/120 V outlets
Harness, to 220/240 V outlets
1009-5716-000
1009-5717-000
8
Outlet Receptacle, Australia, AS 3112
1001-3305-000
Outlet Receptacle, EURO, CEE 7/7
1202-3551-000
Outlet Receptacle, France, CEE 7/4
Support Frame, snap in
1006-4421-000
1006-4422-000
Outlet Receptacle, India and South Africa, BS 546
1006-3805-000
Outlet Receptacle, Japanese
1006-3578-000
Outlet Receptacle, NA, Nema 5-15
1006-3555-000
Outlet Receptacle, Swiss, SEV 1011
1006-3807-000
Outlet Receptacle, UK, BS1363
1001-3309-000
Circuit Breaker, 1A, Rocker
1009-5722-000
Circuit Breaker, 2A Rocker
1009-5721-000
Circuit Breaker, 3A Rocker
1009-5720-000
Circuit Breaker, 4A Rocker
1009-5719-000
10*
Toroid, 100-240V
Heatshrink tubing
1009-5692-000
1202-3268-000
11
Screw, M6x70
0144-2131-923
12
Lockwasher, M6
9213-0560-003
13
Washer
0402-1107-500
14
Cover, transformer
1009-3063-000
15
Screw, M4x8 DIN84 (for transformer cover)
1006-3178-000
16
Stud, Equal Potential, 6 mm
0208-0070-300
9
* Apply heatshrink tubing to terminals of black and white wires if not being used.
8-22
07/04 1009-0356-000
8 Illustrated Parts
5
9
8
4
7
2
1
10*
6
11 (12,13)
16
3
14
AS 3112
Australia/China
AA.96.228
15
Nema 5-15
Japanese
1009-0356-000
07/04
CEE 7/7
EURO
SEV 1011
Swiss,
CEE 7/4
France
BS1363
UK
BS 546
India and South Africa
Nema 5-15
NA
8-23
S/5 Aespire
8.16 Pipeline inlet fittings
7
1
6
2
3
4
5
Item
1
1
1
2
3
4
5
6
7
8-24
Description
Pipeline inlet - O2 fittings
Body, O2 DISS
Body, O2 NIST
Body, O2 DIN
Body, O2 G 3/8 BSPP
Pipeline inlet assembly O2 France
Pipeline inlet assembly O2 Canada
Pipeline inlet assembly O2 Australia
Pipeline inlet - N2O fittings
Body, N2O DISS
Body, N2O NIST
Body, N2O DIN
Body, N2O G 3/8 BSPP
Pipeline inlet assembly N2O France
Pipeline inlet assembly N2O Canada
Pipeline inlet assembly N2O Australia
Pipeline inlet Air fitting
Body, Air DISS
Body, Air NIST
Body, Air DIN
Body, Air G 3/8 BSPP
Pipeline inlet assembly Air France (service kit)
Pipeline inlet assembly Air Canada (service kit)
Pipeline inlet assembly Air Australia (service kit)
O-ring, bore seal
O2 and N2O
Air
Sintered metal filter with o-ring
Pipeline checkvalve with o-ring
Gas Inlet Manifold (replacement)
O2
N2 O
Air
Relief valve, 758 kPa (110 psi)
Screw, M4x20
Lockwasher, M4 external
Stock Number
-------------------1006-5149-000
1006-5158-000
1006-5161-000
1006-5170-000
1006-8363-000
1006-8360-000
1006-8396-000
-------------------1006-5150-000
1006-5159-000
1006-5162-000
1006-5171-000
1006-8362-000
1006-8359-000
1006-8397-000
-------------------1006-5151-000
1006-5160-000
1006-5163-000
1006-5172-000
1006-8361-000
1006-8358-000
1006-8398-000
--------------------0210-0479-300
0210-0539-300
1006-8351-000
1006-3160-000
1009-8066-000
1009-8067-000
1009-8068-000
1011-3049-000
0144-2124-218
9213-0540-003
07/04 1009-0356-000
8 Illustrated Parts
8.17 Cylinder Gas Supplies
10
9
1
3
5
8
7
2
4
6
Item
Description
Pin Index (Inboard)
DIN (Inboard)
1
Gas supply O2
1006-3201-000
1006-3207-000
DIN, Large Cylinder
(Inboard)
1006-3880-000
1
Gas supply N2O
1006-3202-000
1006-3208-000
1006-3881-000
1
Gas supply Air
1006-3203-000
Pin Index (Outboard)
1006-3209-000
-----
2
Gas supply N2O
1009-8210-000
Item
3
Description
Standoff (3 per supply)
Screw, M6x80 socket head cap (3 per supply)
Lockwasher, M6 external (for above screw)
Screw, M6x25 socket head cap (3 per supply)
Lockwasher, M6 external (for above screw)
Cylinder inlets (Pin Index or DIN for external cylinder)
Tee handle beige
Clamp, yoke
Spacer, gas block (2)
Screw, M8 x 25 long socket head cap (2)
Elbow fitting for cylinder pressure gauge (copper tube connection of gas supply)
Label Set, cylinder supply, O2
Stock Number
1009-3085-000
0144-2131-913
9213-0560-003
9211-0660-254
9213-0560-003
Refer to section 8.17.1
0219-3372-600
1001-4076-000
1001-4077-000
9211-0680-253
1006-3713-000
1006-3854-000
Label Set, cylinder supply, N2O
1006-3855-000
Label Set, cylinder supply, Air
1006-3856-000
4
5
6
7
8
9
10
1009-0356-000
07/04
8-25
S/5 Aespire
8.17.1 Cylinder inlet fittings
2 DIN (external cylinder)
1 Pin Index
2e
2a
1a
1b
1c
1d
2b
2d
2c
1e
2b
2h
2g
Item
Description
Stock Number
1
Cylinder inlets (Pin Index)
Gasket
O-ring
Adapter, inlet
Filter, sintered bronze
Retaining ring, filter
--------------------0210-5022-300
9221-3013-116
1001-4075-000
9914-6380-000
1001-5954-000
Cylinder inlets (DIN)
Screw, M8x16
Sealing ring (DIN)
DIN Adapter (O2)
--------------------0144-2140-242
1009-3356-000
1006-4000-000
DIN Adapter (N2O)
1006-4001-000
2d
2e
2f
DIN Adapter (Air)
O-ring, 0.687 ID, 0.812 OD
Filter, sintered bronze
Sealing ring, N2O DIN Conn 11
1006-4002-000
0210-0544-300
9914-6380-000
1202-3641-000
2g
Adapter, large cylinder N2O
1006-4028-000
2h
Adapter, large cylinder O2
1006-4027-000
1a
1b*
1c
1d
1e
2
2a
2b
2c
*
8-26
2f
Lubricate sparingly with Krytox
07/04 1009-0356-000
8 Illustrated Parts
8.18 Vaporizer manifold
6
5
3
4
7
2
8
AA.96.287
1
9
10
11
15
12
14
13
15
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
1009-0356-000
07/04
Description
Manifold assembly, complete, two position
Manifold assembly, complete, one position
O-ring, 0.687 inch ID 0.812 inch OD
Spring, compression
Valve kit, includes seal
Seal
O-ring, 14.3 mm ID
(Package of 6 o-rings)
Nipple, vaporizer port (New Style)
Screw, M2.5 - 0.45x6 PAN, Pozidriv, SST
Spring, Dzus
Seat, check valve
O-ring 27.1 OD 21.89 mm ID
Flapper
Housing
Screw, M4 x 30, cap head
Valve, relief, 5.5 psi, 7/16-20 THD
Flexible tubing, 1/4 inch, mixed gas
Stock Number
1006-8355-000
1009-8065-000
0210-0544-300
1006-3736-000
1006-8373-000
1006-3690-000
1102-3043-000
1102-3016-000
1006-4215-000
1006-3037-000
1102-3056-000
1006-1352-000
1006-3866-000
0211-1451-100
1006-1351-000
9211-0640-304
1006-4128-000
1001-3064-000
8-27
S/5 Aespire
8.19 Flowmeter components
6**
5*
Single-tube Flowhead Outlet
4
6**
11
12
3
9***
10 (5*)
13
2
8
14
7
1
Item
1
2
8-28
Description
Flowhead Module: includes regulator, flowtube module, flowtubes, needle
valve, intermodule tube and associated o-ring, and label plate; does not include
labels, link-25, or knobs (order separately).
O2 flowhead module with dual flowtubes
Stock Number
-------------------
1006-8380-000
O2 flowhead module with single flowtube
1009-8069-000
N2O flowhead module with dual flowtubes
1006-8381-000
N2O flowhead module with single flowtube
1009-8070-000
Air flowhead module with dual flowtubes
Air flowhead module with single flowtube
Secondary regulators/Balance Regulators
Regulator Kit, O2 (adjustable), without pressure switch
1006-8383-000
1006-8382-000
------------------1006-8341-000
Regulator Kit, N2O (pressure balancing)
1006-8344-000
Regulator Kit, Air (adjustable)
1006-8340-000
07/04 1009-0356-000
8 Illustrated Parts
Item
3
4
5*
6**
7
8
9***
10
11
12
13
14
Description
Flowtube Module: includes housing, o-rings, and plug ball; does not include,
flowtubes, label or label panel (order separately).
Flowtube module, O2 - dual
Stock Number
--------------------1006-8338-000
Flowtube module, O2 - single
1009-8234-000
Flowtube module, N2O - dual
1006-8337-000
Flowtube module, N2O - single
1009-8235-000
Flowtube module, Air - dual
Flowtube module, Air - single
Tube, intermodule connector
O-ring, intermodule connector
Clip, U-type self retaining
Screw, M5x30 (module mounting)
Screw, M5x55 (O2/N2O/3rd gas module connect)
1006-8333-000
1006-8334-000
1006-3628-000
1006-3613-000
1006-4350-000
1102-3049-000
1006-3607-000
Screw, M5x110 (O2 - Air module connect)
Washer, M5 (O2 - Air module connect)
Nut, M5 Keps (O2 - Air module connect)
Flowmeter labels
Label panel, flowmeter module
Side panel, (O2 - Air flowmeter module)
1006-3080-000
1006-1459-000
0144-3717-324
refer to chart below
1006-1290-000
1009-3186-000
Screw, label panel
Adapter, single-tube outlet
Clip, small
Elbow, 1/4 inch
Bezel, flowmeter 3-gas
Bezel, flowmeter 2-gas
1006-3608-000
1009-3056-000
1009-3309-000
1202-3804-000
1009-3104-000
1009-3108-000
* Lubricate sparingly with Krytox.
** Note orientation of item 6; with flowmeter facing forward, the barbs should face to the left.
*** Position the label on the panel so that the right edge of the label is flush with the right edge of the panel. The left
edge of the label will extend slightly beyond the left edge of the panel. When mounted in the flowmeter module,
the flowmeter label on the right should slightly overlap the flowmeter label directly to its left.
**** The O2 label set contains a label for both the dual-tube (A) and single-tube (B) modules. Make sure to install
the correct label.
Chart: Flowmeter Labels (include knob labels)
1009-0356-000
07/04
Stock Number
Stock Number
Stock Number
ANSI
Air (Yellow)
1006-0209-000
N2O (Blue)
1009-3209-000
O2 (Green)
1009-3210-000
ISO
O2 (Neutral)
1009-3211-000
N2O (Blue)
1009-3209-000
Air (Black/White)
1009-3242-000
Neutral
O2 (Neutral)
1009-3211-000
N2O (Neutral)
1009-3240-000
Air (Neutral)
1009-3240-000
8-29
S/5 Aespire
8.19.1 Flowtube parts
Item
Description
Stock Number
1
Large flowtube kits (includes float, filter, o-rings, tube)
---------------------
Flowtube kit, Air, large
1006-8325-000
Flowtube kit, N2O, large
1006-8329-000
Flowtube kit, O2, large
1006-8331-000
Small flowtube kits (includes float, filter, o-rings, tube)
---------------------
Flowtube kit, Air, small
1006-8326-000
Flowtube kit, N2O, small
1006-8330-000
Flowtube kit, O2, small
1006-8332-000
Single-tube flowtube kits (includes float, filter, o-rings, tube)
---------------------
2
3
Flowtube kit, single-tube Air flowmeters use the Large flowtube kits
Flowtube kit, N2O, single-tube, dual-taper with filter
1009-8199-000
Flowtube kit, O2, single-tube, dual-taper with filter
1009-8198-000
4
Spring, top of flowtubes
1006-3624-000
5
Float stop, O2 large
1006-1225-000
Float stop, N2O large
1006-1226-000
Float stop, Air large
1006-1227-000
6
O-ring, 17.6 OD, 12.37 ID, large flowtube, top
1006-3615-000
7
Filter, large flowtube
1006-3584-000
8
O-ring, 17.6 OD, 12.37 ID, large flowtube, bottom (red)
1006-3968-000
9
Float stop, O2 small
1006-1233-000
Float stop, N2O small
1006-1234-000
Float stop, Air small
1006-1235-000
10
O-ring, 11.26 OD, 6.02 ID, small flowtube, top
1006-3617-000
11
Filter, small flowtube
1006-3583-000
12
O-ring, 11.26 OD, 6.02 ID, small flowtube, bottom (red)
1006-3969-000
13
Ball, 6 mm (plug fresh gas end)
1006-1353-000
O-ring Kit, flowtubes (includes 4 each of top and bottom o-rings for large
flowtube and 3 each of top and bottom o-rings for small flowtube).
1006-8393-000
Silicon tube kit, long, including cable ties
1006-8378-000
Silicon tube kit, short, including cable ties
1006-8379-000
Not Shown:
8-30
07/04 1009-0356-000
8 Illustrated Parts
2
4
9
1
4
5
13
10
6
11
7
12
8
3
13
4
5
6
7
8
1009-0356-000
07/04
8-31
S/5 Aespire
8.19.2 Secondary
regulator components
Item
Description
Stock Number
1
Knob (N2O) (Air) without label
1006-3633-000
2
O2 Proportioning assembly
(includes knob, sprocket, set screws, without knob label)
1006-8339-000
3
Set screw
0141-4227-105
4
O2 Proportioner chain
1006-3610-000
5
Sprocket, N2O
1006-3625-000
6
Plug, 1/8 inch
1006-3611-000
7
Spacer, link system, N2O needle valve
1006-5140-000
8
Valve, needle (O2) (Air)
1006-8346-000
Valve, needle N2O (has notch around valve body)
1006-8345-000
9
O-ring, 10.1 ID 13.3 OD
9221-3010-116
10
O-ring, 0.250 inch ID 0.375 inch OD
0210-0687-300
11
Stop collar (all gases)
1006-3632-000
12
Maximum stop collar kit (includes item 13)
Maximum stop collars are required in Canada for all gas flow controls.
1006-8055-000
13
Mounting screw, M3x8, SKT HD CAP
1006-3865-000
14
Pressure switch, O2 supply alarm
1006-3623-000
15
O-ring, 0.250 inch ID 0.375 inch OD
0210-0687-300
16
Screws, M4x12 Pozidriv PAN
0140-6226-111
17
Plug, pressure switch cavity
1006-3665-000
18
Screws, M4x8 Pozidriv PAN
1006-3178-000
19
Fitting, O2 pilot, plug-in elbow
1006-3533-000
20
Fitting, O2 pilot, thread-in elbow
1006-3663-000
21
Tubing, 4-mm (RH head 144 mm - LH head 164 mm)
1001-3060-000
8-32
07/04 1009-0356-000
8 Illustrated Parts
4
5
3
6
3
1
2
7
13
8
3
11
9
12
8
10
16
17
15
18
14
20
AB.74.050
15
21
19
1009-0356-000
07/04
8-33
S/5 Aespire
8.20 ABS to machine Interface Components
20
19
3 (4, 5)
17 (18)
14
15
16
20
9
13
12
10 (11)
8
Item
1
2
6
7
Description
PCA VENT MONITORING BRD [TESTED VMB]
PCA B/S task light interface board
(used when no monitoring is ordered)
SCR M4X12 POZI-DR PAN SST TYPE 316
WASH/LOCK EXT M4 DIN 6797 SST TYPE 316
SPACER.171 IN ID 0.250IN OD 0.250 IN LG
ALUMINUM
Cables and harnesses
O2 Flush Valve
8
ACGO Selector Valve
3
4
5
9
Port, ACGO body
10
Screw, M4x30
11
Lockwasher, M4
12
Cap, ACGO check valve
13
Screw, M4x8
14
Disk, ACGO check valve
15
Flapper, ACGO check valve
16*
O-ring
17
Fitting, elbow barbed
18*
O-ring
19
Screw, M3x6
20
Tie wrap
* Lubricate sparingly with Krytox.
8-34
7
6
1 (or 2)
Stock Number
1009-8002-000
1009-5681-000
0140-6226-111
9213-0540-003
1202-3693-000
Refer to section 8.27
Refer to section
8.20.1
Refer to section
8.20.1
1009-3096-000
9211-0640-304
9213-0540-003
1009-3095-000
9211-1040-069
1009-3062-000
1009-3097-000
0210-0543-300
1009-3160-000
0210-0691-300
9211-1030-055
0203-5915-300
07/04 1009-0356-000
8 Illustrated Parts
8.20.1 Flush Regulator, Flush Valve, and ACGO Selector Switch
13
9a
(9b)
11
6
5
7 (8)
12
9c
9
2 (3, 4)
1
10
1009-0356-000
07/04
Item
Description
Stock Number
1
Flush valve, without button
1006-8357-000
2
Flush Button with rod
1011-3354-000
3
Spring
1006-3186-000
4
E-clip
0203-5225-300
5
Bracket
1011-3355-000
6
Screw, M4x8
1006-3178-000
7
Screw, M4x12
0140-6226-111
8
Lockwasher, M4
9213-0540-003
9
ACGO Selector Switch, complete
(without guard - item 10)
1009-3099-000
9a Flush pressure switch
1006-3972-000
9b O-ring
1006-3213-000
9c Screws
0144-2124-201
10
Guard
1009-3140-000
11
Tubing, silicone
1009-3164-000
12
Tie wrap
0203-5915-300
13
Regulator, O2 Flush
1011-3168-000
8-35
S/5 Aespire
8.21 Breathing system interface
17 16
19
18
(21, 22)
13 (11, 12)
9 (11*, 12)
13 (11*, 12)
20
2
6 (7, 8)
3 (4)
1
Item
Description
1
ASSEMBLY MAIN SUPPORT CASTING
2
Bolt, M6x16 flange
3
HANDLE GRIP
4
SCR M6X16 Sems
5
LATCH PUSH TO CLOSE
6
LATCH PUSH TO CLOSE W/MICROSWITCH
7
SCR SKT HD CAP M3-0.5X8 SST
8
WASHER LOCK EXTERNAL M3
9
Port, plug circuit
10
Port, fresh gas
11*
SEAL U-CUP 12.7 ID BCG 19.05 OD EPR
12
RING RET 15.88 SHAFT DIA TYPE E SST
13
Port, sample gas
14
Connector, BULKHEAD O2 CELL, with harness
15
RING RETAINING 9.53 SFT DIAMETER TYPE E SST
16
SW SUBMINITURE W/QDISC TERMINALS
17
SCR M2.5 X 10
18
BRACKET BTV SWITCH
19
LEVER BTV SWITCH
20
CAP BRACKET BTV
21
SCR SKT HD CAP M3-0.5X8 SST
22
WASHER LOCK EXTERNAL M3
* Lubricate sparingly with Krytox.
8-36
5
10 (11*, 12)
Stock Number
1407-7010-000
1009-3125-000
1407-3317-000
0144-2436-109
1407-3309-000
1407-3310-000
1006-3865-000
9213-0530-003
1407-3333-000
1407-3314-000
1407-3320-000
1406-3446-000
1407-3318-000
1009-5586-000
1406-3277-000
1406-3296-000
1009-3153-000
1407-3319-000
1407-3325-000
1407-3324-000
1006-3865-000
9213-0530-003
14 (15)
Qty
(5)
(2)
(2)
(2)
(4)
(4)
(2)
(2)
(2)
(2)
07/04 1009-0356-000
8 Illustrated Parts
8.22 Breathing System
8.22.1 APL Valve
1
9
20 (21)
8
22
13
20 (21)
2
16 (17)
14
3
15
AB.82.008
7
4, 5, 6
10
18
(19)
11 (12)
Item
1
2
3
4
5
6
7
8
9
10
10
11
12
13
14
15
16
17
18
19
20
21
22
1009-0356-000
07/04
Description
APL Valve Assy (includes items 2 through 6)
SPRING CPRSN 53.14 OD 36.8 L 1.48 N/MM
RETAINER SPRING APL
DIAPHRAGM APL
CAGE APL
POPPET APL VALVE
RAMP APL
COVER APL
MANIFOLD APL/BTV
Cover, Manifold APL/BTV (with 22-mm male bag port)
Cover, Manifold APL/BTV (with Australian bag port - 22 mm female)
COVER MANIFOLD APL/BTV
WEIGHT DEAD 14CM H2O BCG ABS NEG RELIEF
SEAL ABS NEG RELIEF VLV
O-RING 22 ID 30 OD 4 W SI 40 DURO
O-RING 88.49 ID 95.55 OD 3.53 W SILICONE 50 DURO
O-RING 1.049ID 1.255OD 0.103W EPDM NO 121
SCR M4X16 BT SKT HD SST TYPE 316
Lockwasher, M4 external
SCR THUMB M4 SHLDR 7.5 X 7
RING RETAINING 3.96 SFT DIA CRESCENT SST
SCR M4 X 40 FL HD SST PH
O-RING 2.9 ID 6.46 OD 1.78 W EP 70 DURO
SCR SEMS M4X8 BT SKT HD W/EXT L/W SST 316
Stock Number
1009-8200-000
1406-3328-000
1407-3404-000
1406-3331-000
1406-3333-000
1406-3332-000
1407-3400-000
1407-3405-000
1407-3401-000
1407-3402-000
1407-3412-000
1407-3402-000
1407-3406-000
1407-3407-000
1407-3104-000
1407-3403-000
1407-3408-000
0140-6226-115
9213-0540-003
1407-3410-000
1407-3411-000
0140-6226-122
1407-3409-000
0144-2436-108
QTY
(2)
(2)
(3)
(3)
(2)
(2)
(3)
8-37
S/5 Aespire
8.22.2 Bag/Vent Switch
AB.82.022
1
3
2
Item
8-38
4
Description
Stock Number
BTV Switch Cartridge
1407-7003-000
1
COVER BTV
1407-3500-000
2
SCR SEMS M4X8 BT SKT HD W/EXT L/W SST 316
0144-2436-108
3
O-RING 44.02 ID 51.1 OD 3.53 W SI 70 DURO
1407-3507-000
4
SEAL, BTV
1407-3506-000
QTY
(2)
07/04 1009-0356-000
8 Illustrated Parts
8.22.3 Absorber canister
5
4
3
6
3
AB.82.017
2
Item
1009-0356-000
07/04
1
Description
Stock Number
Absorber Canister, Reusable
1407-7004-000
1
CANISTER, CO2
1407-3200-000
2
O-RING 110.72 ID 117.78 OD 3.53 W EPR 50 DURO
1407-3204-000
3
FOAM, CO2 CANISTER (PKG 40)
1407-3201-000
4
COVER, CO2 CANISTER with LOCKING RING
(does not include items 5 and 6)
1407-3203-000
5
SCREEN, CO2 CANISTER COVER
1407-3205-000
6
SCREW, M3X8 FL PH HD SST
9211-0530-083
Qty
(2)
8-39
S/5 Aespire
AB.82.019
8.22.4 Flow Sensor Module
2
3
6
1
5
4
10
7
9
8
Item
8-40
Description
Stock Number
Qty
Flow Sensor Module (does not include Item 1)
1407-7001-000
1
Flow Sensor (plastic)
Flow Sensor (metal - autoclavable)
Flow Port Adapter
1503-3856-000
1503-3244-000
1503-3849-000
2
COVER FLOW SNSR
1407-3000-000
3
HOLDER FLOW SNSR UPPER
1407-3002-000
4
HOLDER FLOW SNSR LOWER
1407-3003-000
5
SCR THUMB M6X43 SST
1406-3304-000
6
SCR M4 X 10 SKT CAP BUTTON HEAD SST
0144-2117-718
(2)
7
CUFF FLOW SNSR
1407-3004-000
(2)
8
LATCH FLOW SNSR
1407-3001-000
9
SPR TORSION FLOW SNSR LATCH
1407-3005-000
10
RING TRUARC 0.188 SHAFT E-RING SST
0203-5225-300
07/04 1009-0356-000
8 Illustrated Parts
8.22.5 Breathing Circuit Module
1
3a
2
3b
3c
3
4, 5
3d
9
6 (11)
14
15
12
8
AB.82.021
13, 11
16
10
7
Item
11
Description
Stock Number
Breathing Circuit Module (*)
1407-7002-000
1
LENS CIRCUIT CHK VALVES
1407-3101-000
2
O-RING 44.02 ID 51.1 OD 3.53 W SI 70 DURO
1407-3507-000
3
Check Valve Assembly
1406-8219-000
3a RETAINER DISK 26.97D 12.7H 0.76T
1400-3017-000
3b DISC CHK V RVSBL 1.025D
0210-5297-100
3c SEAT UNIDIRECTIONAL V B/S
1406-3396-000
3d O-RING 20.35 ID 23.90 OD 1.78W
1406-3397-000
4
PLATE CIRCUIT FLANGE
1407-3110-000
5
SCR SEMS M4X8 BT SKT HD W/EXT L/W SST 316
0144-2436-108
6
HOOK LATCH
1407-3604-000
7** O-RING, 22 ID 30 OD 4 W SI 40 DURO
1407-3104-000
8** O-RING, 12.37 ID 17.6 OD
1006-3968-000
9
MANIFOLD CIRCUIT
1407-3100-000
10
PIN CANISTER PIVOT
1407-3109-000
11
RING TRUARC 0.188 SHAFT NO 5133-18H E-RING SST
0203-5225-300
12
LEVER CANISTER LATCH
1407-3102-000
13
PIN CANISTER LEVER
1407-3108-000
14* O2 Cell
6050-0004-110
O-ring, cell
1406-3466-000
15* Plug with o-ring (for units without circuit O2 sensing)
1503-3857-000
O-ring, plug
1407-3112-000
16* Cable, O2 Cell
1009-5570-000
* The O2 cell (or plug) and the cell cable are not included in the breathing circuit module.
** Lubricate sparingly with Krytox.
1009-0356-000
07/04
Qty
(2)
(2)
(2)
(2)
(2)
(2)
(6)
(2)
(5)
8-41
S/5 Aespire
8.22.6 Exhalation valve
9
10
11
12
13
AB.82.020
4
3
2
5 (6)
7
1
8
Item
Description
Stock Number
Qty
Exhalation Valve Assy
1407-7005-000
1
BASE EXHALATION VALVE
1407-3700-000
2
DIAPHRAGM ASSY EXH VALVE
1503-8121-000
3
SEAT EXHALATION VLV ABS
1407-3704-000
4
COVER EXHALATION VALVE
1407-3701-000
5
SCR M4X16 PH PAN HD SST TYPE 316
9211-0440-163
(3)
6
O-RING 2.9 ID 6.46 OD 1.78 W EP 70 DURO
1407-3409-000
(3)
7
SCR THUMB M6X43 10MM HEAD B/S
1406-3306-000
(2)
8
O-RING 4.47 ID X 8.03 OD 1.78 W EPR 70 DURO
1407-3703-000
(2)
9
RETAINER DISK 26.97D 12.7H 0.76T SST FLUTTER V
1400-3017-000
10
WEIGHT DEAD 10 CMH2O BCG PASSIVE AGSS
1406-3572-000
11
SEAT POSITIVE PRESS BCG VALVE PASSIVE AGSS
1406-3571-000
12*
O-RING OD19.16 BCG ID15.6 EPDM DURO 70 -016
1006-3616-000
13
RING RETAINING 19.05 SHAFT DIA SST
1406-3577-000
* Lubricate sparingly with Krytox.
8-42
07/04 1009-0356-000
8 Illustrated Parts
8.22.7 Bellows
AB.82.018
1
2
3
5
6
4
7
8
1009-0356-000
07/04
Item
Description
Stock Number
1
Bellows housing
1500-3117-000
2
Bellows
1500-3378-000
3
Rim
1500-3351-000
4
Pressure relief valve assy
1500-3377-000
5
Latch, base
1500-3352-000
6
Seal, base
1500-3359-000
7
Base, bellows
Refer to section 8.22.8
8
Manifold, bellows base
1407-3702-000
8-43
S/5 Aespire
8.22.8 Bellow base
1
3
2
1
Item
1
2
3
8-44
Description
Bellows Base Assy
1a Latch Assy
HOOK LATCH
E-Ring
Stock Number
1407-7006-000
1407-7007-000
1407-3604-000
0203-5225-300
07/04 1009-0356-000
8 Illustrated Parts
8.22.9 Bag Arms
1
2a
2
4
3a
2b
2d
2c
AB.82.034
3
3b
2e
3c
3d
3e
5
6
7
3f
8
Item
1
2
2a
Description
Bag Arm Assembly (complete)
Bag Arm Upper Assembly
Cover, bag port housing
Screw, M3x20
Lockwasher, M3 internal
2b Housing, bag port
2c Lever, lock release
2d Ring, retaining
2e Nut, M3 Nyloc
3
Bag Arm Lower Assembly
3a PAD FRICTION MATERIAL
3b SPRING CPRSN
3c WASHER, shoulder
3d WASHER LOCK M3 INT SST
3e SCREW M3X16 POSI DR PAN HD A4 SST
3f Nut, M5 Nyloc
4
Pin, dowel 3.18 DIA 31.8 L SST
Items if no Bag Arm
5
Clip, patient tubing
6
Washer, shoulder
7
Lockwasher, M4 external
8
Screw, M4x16
1009-0356-000
07/04
Stock Number
1009-8159-000
1407-7011-000
1407-3807-000
0140-6719-103
9213-0430-003
1407-3806-000
1407-3808-000
1406-3577-000
0144-3536-112
1407-7012-000
1407-3818-000
1406-3270-000
1407-3815-000
9213-0430-003
1504-3003-000
9212-0350-006
1407-3804-000
Qty
(2)
(2)
1407-3810-000
1407-3814-000
9213-0540-003
9211-0440-163
8-45
S/5 Aespire
8.23 Drawer
3
1
2
2
5 (6)
8-46
4 (6)
6
Item
Description
Stock Number
1
Slide, drawer
1009-3084-000
2
Screw, M4x8 Nyloc
1009-3183-000
3
Drawer, body
1009-3078-000
4
Drawer Front, lower (down arrow)
1009-3032-000
5
Drawer Front, upper (up arrow)
1009-3031-000
6
Screw, M4x12
1009-3109-000
07/04 1009-0356-000
8 Illustrated Parts
8.24 Legris quick-release fittings
Item
1
2
3
4
5
6
Description
Stock Number
Tees — (tube/tube/tube)
4 mm (N2O)
1202-3653-000
6 mm (O2)
1006-3544-000
8 mm (Air)
8 mm/6 mm/8 mm (SCGO pilot)
1006-3545-000
1009-3297-000
Tees — (tube/tube/standpipe)
6 mm (O2)
1006-3862-000
8 mm (Air - Drive gas)
1009-3370-000
Elbow — (tube/standpipe)
4 mm (N2O)
1006-3533-000
6 mm (O2)
1006-3534-000
8 mm (Air)
1/4 inch (mixed gas)
1/4 inch (45° - mixed gas)
1006-3535-000
1006-3737-000
1009-3368-000
Elbow — (tube/tube)
1/4 inch (mixed gas)
1202-3804-000
Y
6 mm (O2)
1009-3043-000
8 mm (Air)
8 mm Y with tailpiece
1009-3044-000
1009-3360-000
Plug
4 mm (N2O)
1006-3530-000
6 mm (O2)
1006-3531-000
8 mm (Air)
1006-3532-000
Note: Not every fitting is used in all machines.
1009-0356-000
07/04
8-47
S/5 Aespire
8.25 Vent Drive and low-pressure tubing
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
Description
Coupler, female - black
Coupler, male - black
Coupler, female - white
Coupler, male - white
Coupler, female - yellow
Coupler, male - yellow
Tee (male barb)
Cap, plug
Fitting, coupler barb ends
Plug, 4-mm
Tubing, low-pressure
Tubing (silicone)
Tubing (silicone)
151 mm - 1/4 inch
72 mm - 3/8 inch
42 mm - 3/8 inch
Tube Markings (factory build only)
Length — Size
125
VENT DRIVE
120
127
128
129
130
131
132
133
134
135
136
137
144
145
146
148
Aux O2 OUT
RGM return
unmarked
unmarked
AGSS flowtube
PAW
PAW
PAW
unmarked
unmarked
RGM to Scavenge
RGM to Circuit
Venturi Pilot
Venturi Drive
unmarked
unmarked
149
150
unmarked
unmarked
Length — Size
(black)
Stock Number
1503-3128-000
1503-3237-000
1503-3119-000
1503-3236-000
1503-3132-000
1503-3131-000
1009-3011-000
1406-3524-000
1009-3077-000
1006-3530-000
1605-1001-000
1009-3164-000
1009-3164-000
900 mm - 8 mm
1009-3296-000
(low-pressure)
(low-pressure)
(low-pressure)
(low-pressure)
(low-pressure)
(low-pressure)
(low-pressure)
(low-pressure)
(low-pressure)
(low-pressure)
(low-pressure)
(low-pressure)
(black)
(black)
Tygon
Tygon
250 mm - 1/4 inch
750 mm - 1/4 inch
300 mm - 1/4 inch
151 mm - 1/4 inch
750 mm - 1/4 inch
600 mm - 1/4 inch
260 mm - 1/4 inch
330 mm - 1/4 inch
25 mm - 1/4 inch
50 mm - 1/4 inch
200 mm - 1/4 inch
300 mm - 1/4 inch
330 mm - 4 mm
300 mm - 8 mm
260 mm - 1/2 inch
465 mm - 1/2 inch
1605-1001-000
1605-1001-000
1605-1001-000
1605-1001-000
1605-1001-000
1605-1001-000
1605-1001-000
1605-1001-000
1605-1001-000
1605-1001-000
1605-1001-000
1605-1001-000
1009-3363-000
1009-3296-000
6700-0005-300
6700-0005-300
Tygon
40 mm - 8 mm
180 mm - 1/2 inch
1001-3063-000
6700-0005-300
*Refer to section 8.26
** Sample gas return is directed to the scavenging system as a factory default. A qualified service representative
can reroute the sample gas back to the breathing system (refer to Section 4.13).
8-48
07/04 1009-0356-000
8 Illustrated Parts
120
From
O2 Supply
127
128
5
**
6 130
9
8
1
2
136
*
From
Flowhead
131
*
10
129
125
11
3
*
7
132
4
*
137
133
12
7
134
13
135
AB.74.033
*
125
144
145
149
146
AB.91.012
150
Venturi Drive
Suction
148
1009-0356-000
07/04
8-49
S/5 Aespire
8.26 Tubing for use with Legris fittings
Except for the Tygon tubing (Items 147 and 150), this tubing is a flexible,
Nylon-type tubing for use with quick-release fittings.
Item
Description
Length — Size
Stock Number
106*
107*
108**
109*
124**
126
N20 PLINE - N2O FLWMOD
N2O PLINE - N20 CYL
UNMARKED
N20 CYL - N2O FLWMOD
Unmarked
Unmarked
1200 mm - 4 mm
430 mm - 4 mm
40 mm - 4 mm
800 mm - 4 mm
250 mm - 4 mm
50 mm - 4 mm
1001-3060-000
1001-3060-000
1001-3060-000
1001-3060-000
1001-3060-000
1001-3060-000
113
114
115
116
117
118
123***
219
SW4-O2 FLWMOD
REGULATED O2
REGULATED O2
REGULATED O2
O2 PLINE - REG IN
O2 PLINE - O2 SW3
unmarked
O2 CYL - O2 PLINE
1350 mm - 6 mm
400 mm - 6 mm
600 mm - 6 mm
250 mm - 6 mm
330 mm - 6 mm
270 mm - 6 mm
175 mm - 6 mm
215 mm - 6 mm
1001-3062-000
1001-3062-000
1001-3062-000
1001-3062-000
1001-3062-000
1001-3062-000
1001-3062-000
1001-3062-000
111
110
112
AIR PLINE - AIR SW3
SW4-AIR FLWMOD
AIR CYL- AIR PLINE
460 mm - 8 mm
1350 mm - 8 mm
270 mm - 8 mm
1001-3063-000
1001-3063-000
1001-3063-000
100
101
102
AIR PLINE - AIR GAGE
N2O PLINE - N20 GAGE
O2 PLINE - O2 GAGE
470 mm - 1/8 inch
470 mm - 1/8 inch
470 mm - 1/8 inch
1006-3718-000
1006-3718-000
1006-3718-000
104
105
103
VAP OUT- ACGO
FLWMOD-VAP IN
FLUSH VLV-ACGO
840 mm - 1/4 inch
840 mm - 1/4 inch
280 mm - 1/4 inch
1001-3064-000
1001-3064-000
1001-3064-000
147
150
unmarked (Tygon)
unmarked (Tygon)
290 mm - 1/2 inch
180 mm - 1/2 inch
6700-0005-300
6700-0005-300
* With no N2O cylinder supply, Items 107 and 109 are replaced with Item 106.
** With an inboard N2O cylinder supply, Items 124 is replaced with Item 108.
*** With two inboard O2 cylinder supplies, Item 123 connects the second O2
cylinder supply to the first O2 cylinder supply.
8-50
07/04 1009-0356-000
8 Illustrated Parts
105
110
104
O2
N2O
103
Air
113
115
109
114
117
116
118
101
107
N2O
102
219
O2
124
100
147
111
150
112
126
1009-0356-000
07/04
AB.74.030
Air
8-51
S/5 Aespire
8.27 Cables and harnesses
8-52
Item
Description
Stock Number
1
Harness, Vent Monitoring board to ABS flow sensors
(includes tubing)
1009-8165-000
2
Cable, Vent Monitoring board
1504-5604-000
3
Cable, Serial Isolation
1009-5691-000
4
Cable, Pneumatic Vent Engine
1504-5605-000
5
Cable, power
1009-5711-000
6
Harness, Serial ISO to O2 flush SW
1009-5567-000
7
Harness, Serial ISO to O2 supply SW
1009-5568-000
8
Harness, Serial ISO to on/ standby SW
1009-5566-000
9
Harness, O2 CELL TO FTLR BRD HARN
1009-5586-000
10
Harness, BAG TO VENT SW TO HARN FTLR BRD
1009-5585-000
11
Harness, FLTR BRD TO B/S O2 SNSR AND SW
1009-5531-000
12
Harness, VENT ENG BRD TO CONN PLATE
1009-5545-000
13
Harness, Vent Mon Brd to Task Light
1009-5533-000
14
Harness, to 100/120 V outlets
Harness, to 220/240 V outlets
1009-5716-000
1009-5717-000
15
Harness, ACGO Switch
1009-5762-000
16
Harness, Task Light
1009-5584-000
17
Power Cord
Refer to section 8.14
07/04 1009-0356-000
8 Illustrated Parts
16
5
3
6
7
O2
N2O
Air
4
8
12
2
9
11
13
11
14
11
AB.74.047
white 10
black
white
black
15
1
17
1009-0356-000
07/04
8-53
S/5 Aespire
8.28 Cables and harnesses (Aespire 100)
8-54
Item
Description
Stock Number
1
Harness, Vent Monitoring board to ABS flow sensors
(includes tubing)
1009-8165-000
2
Cable, Vent Monitoring board
1504-5604-000
3
Cable, 7100 module to machine switches
1009-6060-000
4
Cable, Pneumatic Vent Engine
1504-5605-000
5
Cable, power to Control module
1009-5711-000
6
Power Cord, machine
Refer to section 8.14
7
Harness, ACGO Switch
1009-5762-000
8
Harness, O2 CELL TO FTLR BRD HARN
1009-5586-000
9
Harness, BAG TO VENT SW TO HARN FTLR BRD
1009-5585-000
10
Harness, FLTR BRD TO B/S O2 SNSR AND SW
1009-5531-000
11
Harness, VENT ENG BRD TO CONN PLATE
1009-5545-000
07/04 1009-0356-000
8 Illustrated Parts
5
3
6
4
O2
11
2
white 9
black
white
black
8
10
10
10
AB.74.069
7
1
1009-0356-000
07/04
8-55
S/5 Aespire
8.29 Anesthetic Gas Scavenging System — AGSS
8.29.1 Passive AGSS
Item
1
2
3
4
5
6
7
7a
7b
*7c
7d
8*
9*
10
11
12
13
Items 1 through 12 are included in all AGSS kits.
Description, Common Parts
Seal, Receiver Body
Reservoir
Seal and scavenging down-tube
Thumbscrew, M6x28.5
O-ring, 4.42 ID, 9.65 OD
Thumbscrew, M6x43
Valve, unidirectional (negative pressure relief)
Seat, Valve, Negative Pressure
Retainer, disc
O-ring, 20.35 ID, 23.90 OD
Disc, check-valve
O-ring, 22 ID, 30 OD silicone
O-ring, 21.95 ID, 25.51 OD
Screw, M4x8
Cap, 3.18 Barb, Silicone
Adapter, auxiliary inlet, 30-mm male to 30-mm male
Adapter, auxiliary inlet, 30-mm male to 19-mm male
Stock Number
1407-3901-000
1407-3903-000
1407-3904-000
1406-3305-000
1407-3923-000
1406-3304-000
1406-8219-000
1406-3396-000
1400-3017-000
1406-3397-000
0210-5297-100
1407-3104-000
1406-3558-000
9211-0640-083
1406-3524-000
M1003134
M1003947
Passive AGSS Specific Parts
14
Receiver, Passive/Adjustable
15
Plug Assembly, tethered
16
Screw, shoulder M3
17
Connector, 30-mm ISO, Male
18
Adapter, scavenging, 30-mm female to 19-mm male
1407-3908-000
1407-3909-000
1407-3915-000
1406-3555-000
1500-3376-000
Qty
(2)
(2)
(2)
(2)
(5 pack)
* Lubricate sparingly with Krytox
8-56
07/04 1009-0356-000
8 Illustrated Parts
1
2
3
6 (5)
8
7
11
4 (5)
17
15
(10, 9)
16
15
or
12
18
1009-0356-000
13
14
07/04
8-57
S/5 Aespire
8.29.2 Adjustable AGSS
Item
1
2
3
4
5
6
7
7a
7b
*7c
7d
8*
9*
10
11
12
13
Items 1 through 12 are included in all AGSS kits.
Description, Common Parts
Seal, Receiver Body
Reservoir
Seal and scavenging down-tube
Thumbscrew, M6x28.5
O-ring, 4.42 ID, 9.65 OD
Thumbscrew, M6x43
Valve, unidirectional (negative pressure relief)
Seat, Valve, Negative Pressure
Retainer, disc
O-ring, 20.35 ID, 23.90 OD
Disc, check-valve
O-ring, 22 ID, 30 OD silicone
O-ring, 21.95 ID, 25.51 OD
Screw, M4x8
Cap, 3.18 Barb, Silicone
Adapter, auxiliary inlet, 30-mm male to 30-mm male
Adapter, auxiliary inlet, 30-mm male to 19-mm male
Adjustable AGSS Specific Parts
14
Receiver, Passive/Adjustable
15
Plug Assembly, tethered
16
Screw, shoulder M3
17
Needle Valve Assembly (with DISS EVAC connector)
18
Bag with 30 mm male connector
Stock Number
1407-3901-000
1407-3903-000
1407-3904-000
1406-3305-000
1407-3923-000
1406-3304-000
1406-8219-000
1406-3396-000
1400-3017-000
1406-3397-000
0210-5297-100
1407-3104-000
1406-3558-000
9211-0640-083
1406-3524-000
M1003134
M1003947
Qty
(2)
(2)
(2)
1407-3908-000
1407-3909-000
1407-3915-000
1407-3918-000
8004460
* Lubricate sparingly with Krytox
8-58
07/04 1009-0356-000
8 Illustrated Parts
1
2
3
6 (5)
8
7
11
4 (5)
17
15
(10, 9)
16
or
12
13
18
14
1009-0356-000
07/04
8-59
S/5 Aespire
8.29.3 Active AGSS
Item
1
2
3
4
5
6
7
7a
7b
*7c
7d
8*
9*
10
11
12
13
Items 1 through 12 are included in all AGSS kits.
Description, Common Parts
Seal, Receiver Body
Reservoir
Seal and scavenging down-tube
Thumbscrew, M6x28.5
O-ring, 4.42 ID, 9.65 OD
Thumbscrew, M6x43
Valve, unidirectional (negative pressure relief)
Seat, Valve, Negative Pressure
Retainer, disc
O-ring, 20.35 ID, 23.90 OD
Disc, check-valve
O-ring, 22 ID, 30 OD silicone
O-ring, 21.95 ID, 25.51 OD
Screw, M4x8
Cap, 3.18 Barb, Silicone
Adapter, auxiliary inlet, 30-mm male to 30-mm male
Adapter, auxiliary inlet, 30-mm male to 19-mm male
Stock Number
1407-3901-000
1407-3903-000
1407-3904-000
1406-3305-000
1407-3923-000
1406-3304-000
1406-8219-000
1406-3396-000
1400-3017-000
1406-3397-000
0210-5297-100
1407-3104-000
1406-3558-000
9211-0640-083
1406-3524-000
M1003134
M1003947
Active AGSS Specific Parts
14
Receiver, with air brake
15
Seal, for filter and orifice
16
Filter
1407-3900-000
1407-3902-000
1406-3521-000
Active High Flow Specific Parts
17a
Connector, high flow M30 thread
18
Orifice, high flow
1406-3557-000
1407-3920-000
Active Low Flow with EVAC connector Specific Parts
17b
Connector, low flow EVAC
18
Orifice, low flow
1406-3597-000
1407-3919-000
Active Low Flow with 25 mm connector Specific Parts
17c
Connector, low flow 25 mm
18
Orifice, low flow
1406-3573-000
1407-3919-000
Qty
(2)
(2)
(2)
(2)
Active Low Flow with 12.7 mm hose barb connector Specific Parts
17d
Connector, low flow 12.7 mm (1/2 inch)
1406-3574-000
18
-noneActive Low Flow with 30 mm ISO male connector Specific Parts
17e
Connector, low flow 25 mm
1406-3555-000
18
Orifice, low flow
1407-3919-000
* Lubricate sparingly with Krytox
8-60
07/04 1009-0356-000
8 Illustrated Parts
1
2
3
6 (5)
8
7
11
4 (5)
17
(10, 9)
15
or
16
12
14
15, 18
17a
1009-0356-000
13
17b
07/04
17c
17d
17e
8-61
S/5 Aespire
8.30 Integrated Suction Regulator
8.30.1 Major Components (Continuous and Venturi suction)
8 (d)
8
(a,b,c)
2
9
17
3
4
5
suction
15
1
15
vacuum
15
14
suction
13
14
only with Venturi
15
11
vacuum
12
vacuum
10
suction
14
16
Item
1
2
3
4
5
6
7
8a
8b
8c
8d
9
10
11
12
13
14
15
16
17
8-62
6 (7)
Description
Suction Control Module
Venturi Assembly
Cover, blank (if no Suction)
Bracket, blank cover mounting
Screw, M4x10 self-tapping
Manifold
Screw, M5x20 BHSCS PT THD FORMING
Connector, NIST
Connector, Barb
Connector, Air Liquide
Muffler, for Venturi Drive
Coupling, Colder insert metal
Coupling, Colder body black
Coupling, Colder insert black
Coupling, Colder body white
Coupling, Colder insert white
Tubing, Tygon
Fitting, barb to 8-mm Legris
Cap, white
Overflow Safety Trap
Stock Number
Refer to section 8.30.2
Refer to section 8.30.3
1009-3271-000
1009-3270-000
1009-5534-000
1009-3123-000
1009-3384-000
1011-3524-000
0221-0702-300
1009-8292-000
1011-3511-000
1009-3135-000
1009-3373-000
1009-3374-000
1009-3371-000
1009-3372-000
Refer to section 8.26
1009-3137-000
1009-3385-000
6700-0647-000
07/04 1009-0356-000
8 Illustrated Parts
8.30.2 Suction Control Module
6
5
4
3b*
3a*
3
2
1a
*
7
8
1
9**
only with Venturi
10
9c
9b
9a
Item Description
Stock Number
1
1009-3227-000
1009-3228-000
6700-0133-500
Gauge, 760 mmHg
Gauge, 1 Bar
1a O-ring, Gauge (included with gauge assy, 2ea. required)
2
3
Control panel assembly, with suction regulator knob and mode control knob
Regulator Module (plugs into manifold assembly)
3a O-ring, Regulator Module, Large (included with regulator module)
3b O-ring, Regulator Module, Stem (included with regulator module)
1009-3213-000
6700-1225-800
6700-0136-500
0210-0527-300
4
Manifold Assembly, without Gauge and Regulator Module
1009-3277-000
5
Screw, #6 - 2 inch
1009-3340-000
6
Mounting bracket
1009-3255-000
7
Screw, #6 - 1 inch
1009-3339-000
8
Filter
0206-5159-300
9
Pilot valve adapter assembly (includes plunger, jam nut, and valve assembly)
1009-3278-000
10
Cap, white
1009-3192-000
* Lubricate the regulator module o-rings and the mating bore of the manifold sparingly with Dow 111 lubricant.
** Drop the plunger (9a), round end first, into the manifold. Thread the pilot valve into the manifold body. Set the
mode switch to raise the plunger. Adjust the pilot valve (9b) so that the plunger actuates the pilot valve
approximately half of its travel. Tighten the jam nut (9c).
1009-0356-000
07/04
8-63
S/5 Aespire
8.30.3 Venturi assembly
1
2
6, 7
13
12
3
5
8
4
9
11
16
15
10
14
17
18
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
8-64
Description
C-clip retainer, Truarc
Elbow fitting, 4-mm Legris
Cap
Spoppet
Seal, u-cup large
Orifice
Screen, 150 mesh monel
Seal, u-cup small
Body
Venturi
Elbow fitting, 8-mm Legris
O-ring, large
O-ring, small
Check valve
Bracket, Venturi mounting
Screw, M5x20 BHSCS PT THD FORMING
Cable tie
Tubing, Tygon
Stock Number
1500-3158-000
1006-3663-000
1011-5002-000
1011-5001-000
1503-3090-000
1011-3508-000
1001-3808-000
1503-3089-000
1011-5000-000
1011-3509-000
1011-3510-000
9221-3032-116
1503-3108-000
1011-8002-000
1009-3182-000
1009-3384-000
0203-5915-300
Refer to section 8.25
07/04 1009-0356-000
8 Illustrated Parts
8.31 Auxiliary O2 Flowmeter
5
1
6*
4
12
7
11
2 (3)
10
9
11
13
8
Item
Description
Stock Number
1
Flowmeter, 1-10 L/min, Complete with fittings installed
Flowmeter, 1-10 L/min, without fittings
1006-8424-000
1006-3841-000
2
Knob, gray
1011-3471-000
3
Set Screw
9211-0830-053
4
Nipple, Panel-Mount, Auxiliary O2 Outlet
1006-5177-000
5
Label, blank (if no Auxiliary O2)
1009-3243-000
6*
Nut, M12x1.75, SST
0144-3132-140
7**
Flowmeter Fitting, 1/8 NPTM straight adapter
Flowmeter Fitting, 1/8 NPTM elbow adapter
0204-8877-300
0204-8788-300
8**
Flowmeter Fitting Assembly, 6-mm Tubing Adapter
1006-8423-000
9
Plate, Flowmeter Mounting
1009-3126-000
10
Screw, 10-32 x 3/8
0140-6631-107
11
Screw, M4x10 self-tapping
1009-5534-000
12
Tubing (low-pressure) 250 mm - 1/4 inch
1605-1001-000
13
Plate, blank (if no Auxiliary O2)
1009-3128-000
* Apply Loctite 242.
** Apply Teflon tape.
1009-0356-000
07/04
8-65
S/5 Aespire
8.32 Display mounts
1
5
4
2a
3a
2b
3b
Item
Description
Stock Number
1
GCX Arm Kit, long
1009-3262-000
2
GCX Bracket Kit,
includes display mount (2a), cable guard (2b), and mounting hardware
1009-3263-000
3
GCX Arm Kit, short
includes display arm (3a) and fasteners (3b)
1009-3264-000
4
Extrusion, upper dovetail
1009-3113-000
5
Screw, M6x20
0144-2131-925
*
Cable ties
0203-5915-300
* Refer to Section 4.21 for mounting the display arms and proper cable dressing.
8-66
07/04 1009-0356-000
8 Illustrated Parts
8.33 Cable management arm
4 (5)
7 (2, 3, 8)
1 (2, 3)
6
Item
1009-0356-000
07/04
Description
Stock Number
Cable management arm, complete assembly
1009-8181-000
1
Bracket, cable management
1009-3261-000
2
Screw, M4x12
0140-6226-111
3
Lockwasher, M4 external
9213-0540-003
4
Extrusion, cable arm, front loading
1009-3247-000
5
Screw, M4x6 Nyloc
1009-3283-000
6
Retainer, multiple cable
1009-3252-000
7
Retainer, cable small
1009-3259-000
8
Washer, M4 flat
0144-1025-165
8-67
S/5 Aespire
8.34 Display arm mounting kits for optional equipment
2
AC.09.003
AC.09.002
1
3
AC.09.001
AC.09.004
4
8-68
Item
Description
Stock Number
1
Cardiocap 5 mount
1009-3265-000
2
S/5 Flat Panel mount
1009-3266-000
3
Spacelabs Flat Panel mount
1009-3267-000
4
Spacelabs PC Scout mount
1009-3268-000
07/04 1009-0356-000
9
In this section
Schematics and Diagrams
Schematics are subject to change without notice.
Circuit boards are available only as complete assemblies.
1009-0356-000
Figure 9-1
System connection block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-2
Figure 9-2
Gas scavenging circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-3
Figure 9-3
Electrical cabling block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-4
Figure 9-4
Pneumatic circuit diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Figure 9-5
Wiring harnesses. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-6
Figure 9-6
Wiring harnesses (Aespire 100). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-7
Figure 9-7
Tubing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-8
Figure 9-8
Schematic, AC Inlet module; 100–120 V (Non-isolated outlets or no outlets) . . . . . . . . . . . . .9-9
Figure 9-9
Schematic, AC Inlet module; 100–120 V (Isolated outlets. . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10
Figure 9-10
Schematic, AC Inlet module; 220–240 V (Non-isolated outlets or no outlets) . . . . . . . . . . . 9-11
Figure 9-11
Schematic, AC Inlet module; 220–240 V (Isolated outlets) . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
Figure 9-12
Schematic, AC Power (Aespire 100) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13
07/04
9-1
S/5 Aespire
Atmosphere
Free Breathing
Check Valve
APL valve
0-70 cm H2O
Mechanical Overpressure
Valve (110 cm H2O)
NO
Popoff
Valve
NC
Proportional PEEP Valve
Bag
Exhalation Valve
(2.0 cm H2O bias)
Supply Pressure Switch
Bag/Vent
10 cm H2O
PEEP Safety Valve
O2 Flush
Aux O2
0-10 LPM
(opt)
To Scavenging
(opt)
35 psi
Negative Pressure
relief valve
25 psi
@ 15 LPM
Gauge P
Expiratory
Flow Sensor
200 mL Reservoir
0-120 LPM Flow
O2 P-LINE
Gas Monitor
System Switch
Control Bleed to Ambient
1.0 LPM @ 3.0 cm H2O
if continuous (rate dependent)
110 psi
Relief
Absorber
Drain
P Gauge
0-10 LPM Drive Gas
0-10 LPM Patient and Fresh Gas
0-20 LPM Total Typical Flow
Inspiratory
Flow
Control
Valve
Ambient
O2 Cyl (opt)
NC
NO
P Gauge
Venturi
O2 Cyl (opt)
Patient
Venturi
5.4 psi
Vent
Drive
30 psi
Balance
Regulator
Air (opt)
30 psi
Gauge P
AIR P-LINE (opt)
P Gauge
110 psi
Relief
O2
Supply
Switch
Link 25
O2
Flush
Switch
NO
O2
Sensor
NC
Inspiratory
Flow Sensor
Link 25
ACGO
Selector
Valve
Air Cyl (opt)
Gauge P
ACGO
N2O P-LINE (opt, Std US)
Airway
Pressure P
Gauge
Selectatec Manifold
5.5 psi Pressure
Relief Valve
N2O Cyl (opt)
PT
Pneumatic
Engine
Board
PT
PT
AB.74.029
Flowmeter Module
(Dual Flowtubes Optional)
Insp. Flow Trans.
110 psi
Relief
Exp. Flow Trans.
P Gauge
Ventilator Monitoring Board
VAP
VAP
Figure 9-1 • System connection block diagram
9-2
07/04 1009-0356-000
9 Schematics and Diagrams
Scavenging
From System
Scavenging
From System
+10 cm H2O
Relief Valve
+10 cm H2O
Relief Valve
0.3 cm H2O
Entrainment
+10 cm H2O
Relief Valve
0.3 cm H2O
Entrainment
0.3 cm H2O
Entrainment
High or Low
Flow Connector to
Disposal System
Needle Valve
Assembly with
DISS EVAC
Connector
30 mm Male
Connector
Reservoir
Passive
Reservoir
Adjustable
Room
Air
High or Low
Filter Flow
Restrictor
Reservoir
Flow Indicator
AB.74.031
Scavenging
From System
Active
(See Note)
Key to Symbols
= Plugged port (1/8 inch) for sample gas return
= Plugged port (30 mm) for auxiliary breathing system scavenging
= Open port (30 mm) for auxiliary breathing system scavenging
Note: Active AGSS systems with a 12.7 mm connector do not include the
Flow Restrictor and the Flow Indicator.
Figure 9-2 • Gas scavenging circuits
1009-0356-000
07/04
9-3
S/5 Aespire
Inside Machine
AC Inlet
with
Circuit Breaker and
Line Filter
Surge
or
Inrush
Board
Fuses
Control Module
Universal
Power Supply
Line Filter
Fan
Outlet
Box
Serial Isolation
Connector Board
Isolation
Transformer
5mm x 20mm
T2L/250 V
+6V @ 5A Max
RS232
Serial Adapter Cable
6V Battery
System Switch
On/Standby
Software Upgrade
Module
O2 Supply Switch
Flash ROM
Speaker
O2 Flush Switch
7100 Control Board
Inspiratory Valve
Backlight
Vent
Engine
Board
PEEP Valve
Pneumatic Interface Cable
System Pressure Switch
LCD Display
320 X 240
PEEP Safety Valve
Vent Engine
Task Light
Monitoring Interface Cable
Bag/Vent
Switch
ABS On
Switch
Expiratory
Flow Sensor
Ventilator
Monitoring
Board
Rotary
Encoder
Membrane
Switches
AB.74.046
Power
Cord
Bulkhead
Connector
Inspiratory
Flow Sensor
O2 Sensor
Breathing System
ACGO
Switch
Tabletop
Figure 9-3 • Electrical cabling block diagram
9-4
07/04 1009-0356-000
9 Schematics and Diagrams
Key to Symbols
1.
2.
3.
4.
5.
6.
7.
Pipeline pressure gauge
Pipeline inlet
Cylinder pressure gauge
Cylinder inlet
Primary regulator (cylinder pressure)
High-pressure relief valve (758 kPa / 110 psi)*
Supply connections for the ventilator
a. O2 drive gas
b. Air drive gas
8. System switch
9. Switch for low O2 supply pressure alarm (used with the ventilator)
10. O2 secondary regulator (207 kPa / 30 psi)*
11. O2 flow control valve
12. O2 flow tube(s)
13. O2 flush and auxiliary flowmeter regulator (241 kPa / 35 psi)*
14. O2 Flush
a. Flush valve
b. Pressure switch (used with the ventilator)
15. N2O balance regulator
16. N2O flow control valve
17. N2O flow tube(s)
18. Air secondary regulator (207 kPa / 30 psi)*
19. Air flow control valve
20. Air flow tube(s)
21. Supply connection for Venturi suction
a. O2 drive gas
b. Air drive gas
22. Vaporizer port valve
23. Vaporizer
24. Low-pressure relief valve (38 kPa / 5.5 psi)*
25. Auxiliary flowmeter (optional)
26. To ABS
27. To ACGO
28. Test port (primary regulator)
29. Test port (secondary/balance regulator)
* Approximate values
Pneumatic Connection
Filter
Direction of Flow
N2O
4
Air
4
O2
4
3
3
5
A
A
28
3
5
5
A
28
A
28
2
3
25
5
28
2
B
6
O2
4
B
6
1
2
B
6
1
1
21a
21b
7a
7b
13
8
14a
C
9
10
29
11
Link 25
12
15
29
16
14b
18
29
19
Link 25
17
26
20
27
22
22
E
22
24
D
22
AB.74.041
Key to Numbered Components
23
23
A. Cylinder Supplies
B. Pneumatic Manifold
C. Flowmeter Head
D. Vaporizer Manifold
E. ACGO Select Switch
Check Valve
Figure 9-4 • Pneumatic circuit diagram
1009-0356-000
07/04
9-5
S/5 Aespire
Control/Display Module
Task Light
RS-232
O2
N2O
Air
VEB
ABS ON
white
black
white
black
VMB
AC
Inlet
AB.74.047
Bag/Vent
Key to Symbols
VEB = Vent Engine Board
VMB = Ventilator Monitoring Board
ACGO
ACGO = Auxiliary Common Gas Outlet
ACGO
Selector
Switch
Figure 9-5 • Wiring harnesses
9-6
07/04 1009-0356-000
9 Schematics and Diagrams
Control/Display Module
AC
Inlet
O2
VEB
Bag/Vent
ABS ON
white
black
white
black
AB.74.069
VMB
Key to Symbols
VEB = Vent Engine Board
VMB = Ventilator Monitoring Board
ACGO
ACGO = Auxiliary Common Gas Outlet
ACGO
Selector
Switch
Figure 9-6 • Wiring harnesses (Aespire 100)
1009-0356-000
07/04
9-7
S/5 Aespire
Auxiliary O2 Flowmeter
Sample
Return
AB.74.030
From
O2 Supply
Scavenging Downtube
Vaporizer Manifold
*
From
Flowmeter
Vent Engine
VEB
O2
N2O
Air
N2O
Ventilator
Monitoring
Board
AB.74.033
O2
ACGO
Air
ACGO
Selector
Switch
*
The Sample Gas Return is directed to the scavenging system as a
factory default. A qualified service representative can reroute the
sample gas back to the breathing system. Refer to Section 4.13.
Figure 9-7 • Tubing
9-8
07/04 1009-0356-000
9 Schematics and Diagrams
TRANSIENT SUPPRESSION PCA 100-120V
JMPR1
AC INLET MODULE
1
2
3
15A
SYSTEM BREAKER
(BRN)
LINE
FILTER
LINE OUT
LINE IN
RV2
150V
D6521ZOV151RA20
(ROCKER)
AC INLET
IEC 320
Differential
Mode
1
2
(BLU)
RV1
D6521ZOV141RA20
140V
NEUTRAL IN
Common
Mode
F1
1A
250VAC
FUSE2AG
LINE FILTER
L
CR1
1.5KE250C
250V
1500W
10%
R1
10
20%
2W
N
G
NEUTRAL OUT
T0
Display Module
L
3EHZ1
4
N
2A
FUSE
RV3
D6521ZOV141RA20
140V
1
2
GND (GRN/Y)
G1
U-5B
250V
15%
2A
FUSE
G2
U-5B
250V
15%
CHASSIS GROUND
EQUIPOTENTIAL STUD
CONNECTOR
OUTLET
BREAKERS
COUNTRY SPECIFIC
OUTLETS
AB.74.036
CHASSIS GROUND
Figure 9-8 • Schematic, AC Inlet module; 100–120 V (Non-isolated outlets or no outlets)
1009-0356-000
07/04
9-9
S/5 Aespire
INRUSH PCA 100-120V
AC INLET MODULE
1
2
3
R2
1.8
15A
SYSTEM BREAKER
(BRN)
LINE
FILTER
FUSIBLE
10%
5W
LINE IN
RV2
150V
D6521ZOV151RA20
(ROCKER)
AC INLET
IEC 320
Differential
Mode
1
2
(BLU)
G1
U-5B
250V
15%
RV1
D6521ZOV141RA20
140V
NEUTRAL IN
CR1
1.5KE250C
250V
1500W
10%
K1
4
6
LINE OUT
1
2A
FUSE
LINE FILTER
L
1
5
SOLID
STATE
DELAY
TIMER
JH1A-W-AC120V-Q
R1
10
20%
2W
IN
3
OUT
4
N
F1
1A
250VAC
FUSE2AG
3EHZ1
G
2
T1
KH111 LOAD
120V
NEUTRAL OUT
T0
Display Module
L
N
2A
FUSE
RV3
D6521ZOV141RA20
140V
Common
Mode
1
2
G2
U-5B
250V
15%
GND (GRN/YEL)
CHASSIS GROUND
OUTLET
BREAKERS
EQUIPOTENTIAL STUD
COUNTRY SPECIFIC
OUTLETS
CONNECTOR
CHASSIS GROUND
AB.74.038
ISOLATION TRANSFORMER
VIO
BLK
THERMAL FUSE
130C
BRN
RED
RED
BRN
THERMAL FUSE
130C
CONNECTOR
ORN
BLU
BLK
WHITE
NC
NC
Figure 9-9 • Schematic, AC Inlet module; 100–120 V (Isolated outlets)
9-10
07/04 1009-0356-000
9 Schematics and Diagrams
TRANSIENT SUPPRESSION PCA 220-240 V
AC INLET MODULE
1
2
3
JMPR1
8A
SYSTEM BREAKER
(BRN)
LINE
FILTER
LINE IN
LINE OUT
RV2
D65ZOV271RA140
275V
(ROCKER)
AC INLET
IEC 320
Differential
Mode
1
2
(BLU)
G1
U-9B
470V
15%
RV1
D65ZOV271RA140
275V
NEUTRAL IN
CR1
1.5KE440CA
440V
5%
1500W
R1
10
2W
20%
CERAMIC
F1
1A
250VAC
FUSE2AG
2A
FUSE
LINE FILTER
L
N
G
NEUTRAL OUT
T0
Display Module
L
3EHZ1
4
N
2A
FUSE
RV3
D65ZOV271RA140
275V
Common
Mode
1
2
GND (GRN/Y)
G2
U-9B
470V
15%
CHASSIS GROUND
EQUIPOTENTIAL STUD
CONNECTOR
OUTLET
BREAKERS
COUNTRY SPECIFIC
OUTLETS
AB.74.037
CHASSIS GROUND
Figure 9-10 • Schematic, AC Inlet module; 220–240 V (Non-isolated outlets or no outlets)
1009-0356-000
07/04
9-11
Notes
INRUSH PCA 220-240V
R2
6.8
AC INLET MODULE
1
2
3
8A
SYSTEM BREAKER
(BRN)
LINE
FILTER
10%
7W
FUSIBLE
LINE IN
RV2
D65ZOV271RA140
275V
(ROCKER)
AC INLET
IEC 320
Differential
Mode
1
2
(BLU)
G1
U-9B
470V
15%
RV1
D65ZOV271RA140
275V
NEUTRAL IN
CR1
1.5KE440CA
440V
5%
1500W
R1
10
2W
20%
CERAMIC
K1
4
LINE OUT
1
6
1
SOLID
STATE
DELAY
TIMER
LINE FILTER
L
T1
KH115
240V
5
JH1A-W-AC240V-Q
2A
FUSE
N
3
3EHZ1
G
4
T0
Display Module
L
N
2
F1
1A
250VAC
FUSE2AG
NEUTRAL OUT
2A
FUSE
RV3
D65ZOV271RA140
275V
Common
Mode
1
2
G2
U-9B
470V
15%
CHASSIS GROUND
GND (GRN/Y)
OUTLET
BREAKERS
EQUIPOTENTIAL STUD
COUNTRY SPECIFIC
OUTLETS
CONNECTOR
CHASSIS GROUND
AB.74.039
ISOLATION TRANSFORMER
VIO
BLK
THERMAL FUSE
130C
BRN
RED
RED
BRN
THERMAL FUSE
130C
CONNECTOR
ORN
BLU
BLK
WHITE
NC
NC
Figure 9-11 • Schematic, AC Inlet module; 220–240 V (Isolated outlets)
9-12
07/04 1009-0356-000
9 Schematics and Diagrams
TRANSIENT SUPPRESSION PCA 100-240V
LINE OUT
LINE IN
FUSE
2A
FUSE
LINE
FILTER
1
2
470V
15%
Differential
Mode
AC INLET
IEC 320
RV1
275V
NEUTRAL IN
F1
1A
250VAC
FUSE2AG
RV2
275V
CR1
1.5KE440CA
440V
1500W
5%
R1
10
20%
2W
T0
Display Module
C1
0.1UF
20%
275VAC
AB.74.062
2A
G1
U-9B
NEUTRAL OUT
EQUIPOTENTIAL STUD
CHASSIS GROUND
Figure 9-12 • Schematic, AC Power (Aespire 100)
1009-0356-000
07/04
9-13
Notes
9-14
07/04 1009-0356-000
S/5 Aespire Machine
Technical Reference Manual, English
1009 0356 000
07 04 B 01 01 02
Printed in USA
©Datex-Ohmeda, Inc. All rights reserved

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement