epump_operation_and_service_ma..

Add to my manuals
42 Pages

advertisement

epump_operation_and_service_ma.. | Manualzz

tyco

/ Healthcare

kENDALL

KANGAROO

e

Pump*

Enteral Feeding Pump

Operation and Service Manual

Enteral Feeding Pump

Operation and Service Manual

Enteral Feeding Pump

Operation and Service Manual

Enteral Feeding Pump

Operation and Service Manual

Enteral Feeding Pump

Operation and Service Manual

����

See Instructions

TABLE OF CONTENTS

Thank you for purchasing the KANGAROO

e

Pump* Enteral Feeding Pump. With proper

care, this device will provide you with years of precision service.

Table of Contents

Page

Section I - General Information ........................................................................1

Section II - Safety and Warnings ......................................................................3

Section III - Icon Identification .........................................................................4

Section IV - Initial Setup

Attaching the A/C Adapter Power Cord ........................................................................................................5

Battery Setup ...............................................................................................................................................5

Attaching Pole Clamp ..................................................................................................................................6

Section V - Instructions for Use

Quick Start ...................................................................................................................................................7

General Startup ...........................................................................................................................................8

Placement/Mounting ............................................................................................................................8

A/C Power Operation .............................................................................................................................8

Battery Power Operation ........................................................................................................................8

Power On/Off .......................................................................................................................................8

Keep or Clear Prior Pump Settings ...........................................................................................................8

Loading Pump Sets ......................................................................................................................................9

Prime Pump ...............................................................................................................................................10

Auto Priming ......................................................................................................................................11

Hold-To-Prime ...................................................................................................................................

11

Re-priming after Bag Empty .................................................................................................................11

Selecting Continuous or Intermittent Feed Mode .......................................................................................12

Continuous Mode (Adjust Feed) .................................................................................................................12

Continuous Mode Feeding ...................................................................................................................12

Continuous Mode Flushing ...................................................................................................................

13

Intermittent Mode (Adjust Bolus) ..............................................................................................................14

Intermittent Mode Feeding ...................................................................................................................

14

Intermittent Mode Feeding - Bolus Max .................................................................................................

15

Intermittent Mode Flushing ..................................................................................................................15

Run ............................................................................................................................................................16

Hold .................................................................................................................................................17

KANGAROO

e

Pump

*

Table of Contents

Page

Flush Now .........................................................................................................................................18

More Options .............................................................................................................................................19

Buzzer Alarm Volume .........................................................................................................................19

History ..............................................................................................................................................19

Language Selection .............................................................................................................................

19

Continuous/Intermittent Mode .............................................................................................................19

BioTech Menu .............................................................................................................................................20

Accessing the BioTech Screen ................................................................................................................20

BioTech Screen Data ............................................................................................................................20

Lock Settings Feature (Program Locked) .................................................................................................

20

Lock Screen Feature ......................................................................................................................................................21

Section VI - Re-Certification of Performance

System Performance Re-Certification ........................................................................................................21

Certifying the Accuracy of Pump Flow Rate ................................................................................................22

Section VII - Cleaning .................................................................................... 22

Section VIII - Battery Replacement ................................................................ 24

Section IX - Error/Warning/Informational Screens and Troubleshooting

System Error ...............................................................................................................................................25

Hold Error ...................................................................................................................................................26

Rotor Error .................................................................................................................................................27

Feed Error ..................................................................................................................................................27

Flush Error ..................................................................................................................................................27

Flow Error ..................................................................................................................................................28

Pump Set Dislodged Error ..........................................................................................................................28

Battery Low ...............................................................................................................................................29

Feeding Complete ......................................................................................................................................29

Pump Set Use >24 Hours Warning .............................................................................................................29

LED Array ...................................................................................................................................................29

Section X - Specifications and Symbols

Specifications .............................................................................................................................................30

Symbols on Pump and Accessories .............................................................................................................32

Section XI - Customer Service ......................................................................... 33

KANGAROO

e

Pump

*

Table of Contents

Section XII - Maintenance

Page

Side Door Enclosing Pump Set Loading Area ..............................................................................................33

Battery Pack ...............................................................................................................................................33

Battery Door ..............................................................................................................................................33

Power Cord .................................................................................................................................................33

Pole Clamp .................................................................................................................................................33

Section XIII - Service Part Numbers ................................................................ 34

Section XIV - Warranty ................................................................................... 34

Section XV - Electromagnetic Conformity Declaration ...................................... 35

KANGAROO

e

Pump

Section I - General Information

The KANGAROO ePump enteral feeding pump is a simple-to-use, precision enteral feeding pump. It is easily programmed to provide patients with either continuous or intermittent feeding, and can provide automatic flushing capability when used with KANGAROO ePump Feed & Flush Sets.

User Interface:

• Intuitive user interface

• Large, backlit LCD display

• Step-by-step prompts to guide programming and operation

• “Stoplight” LED array visually indicates pump status in a bright or darkened room

Ergonomics:

• Quiet operation

• Compact, aesthetic design

• Tabletop usage or IV pole mounting

• Simple loading of Pump Sets

• Transparent side door to protect Pump Set

• Integrated handle for easy transport

Features:

• MISTIC (Magnetic Intelligent Set Type Identification Connector) Feed-Only vs. Feed & Flush Set identification system to ensure match between the pump’s user interface and Pump Set type

• Automatic Anti-Free-Flow (AFF) System

• Audible alarm to indicate errors or Pump Set loading conditions

• Attitude-independent design eliminates need for drip chamber on pump set

• Sensor technology detects both upstream and downstream flow conditions

• Continuous feed, intermittent feed, and flush capability

• Auto-prime feature reduces the need for time-consuming manual priming

• Hold feature with programmable restart

• Stores previous 72 hours of feeding and flushing history

Service:

• Serial I/O port for data retrieval and software updates

• Uses replaceable, rechargeable NiMH Battery Pack for 15 hours of backup power

• Removable Power Cord

• Removable Pole Clamp

KANGAROO

e

Pump

*

1

2

Figure 1A. KANGAROO ePump features, front and rear views.

KANGAROO

e

Pump

*

Figure 1B. KANGAROO ePump Pump Set loading area.

Section II - Safety and Warnings

CAUTION: U.S. federal law restricts the sale of this device to physicians or to their direct representatives.

1. Read this booklet thoroughly before using the KANGAROO ePump enteral feeding pump.

2. Do not use this device near flammable anesthetics.

3. Disconnect power supply before cleaning or servicing.

4. Use only KANGAROO ePump enteral feeding pump Power Cord (with built-in A/C adapter) with pump. See

Section XIII - Service Part Numbers for replacement of Power Cord.

5. Use only KANGAROO ePump enteral feeding pump Pump Sets with this device. Pump is not compatible with other

Pump Sets.

6. For re-certification, see Section VI - Re-certification of Performance.

For other integrity checks, consult with a qualified Biomedical technician or contact the manufacturer

(Section XI - Customer Service).

7. For service or for technical information, please contact Customer Service (Section XI).

8. Do not open the main housing, as there are no user-serviceable parts inside. Opening may affect function of device

and voids the warranty. To replace battery, see Section VIII.

9. Dispose of old battery-powered electronic equipment in a manner consistent with institutional policy for expired equipment disposal.

KANGAROO

e

Pump

*

3

4

10. Cleaning frequency and practices must be consistent with institutional policy for cleaning of non-sterile devices. See

Section VII - Cleaning, for instructions on cleaning the KANGAROO ePump enteral feeding pump.

11. See icon descriptions in Section III-Icon Identification and Section X - Specifications and Symbols for additional safety information.

12. This device is designed to minimize the effects of uncontrolled electromagnetic interference and other types of interference from external sources. Avoid use of other equipment that may cause erratic operation or degradation in the performance of this device.

13. WARNING: The DB9 serial I/O port (see Figure 1A) is intended only for use by the manufacturer.

Any accessory equipment that can be connected to the DB9 Serial I/O port, and which does not comply with the equivalent safety requirements of this equipment, may lead to a reduced level of safety of the pump. While use of the pump’s DB9 port is not recommended for customers, the following should be considered in the event that the customer decides to connect to this port:

• The accessory is specifically certified as complying with the safety requirements for accessories used with the pump, whereas the safety certification of the accessory has been performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized national standard.

• Under no circumstances should the certified accessory be used while the pump is in use for enteral feeding of a patient.

Section III - Icon Identification

Run Mode

Vertically-moving droplet in

RUNNING screen indicates Run mode.

Battery Symbols

• Full charge

• 2/3 or less charge

• 1/3 or less charge

• Very low battery (15 minutes max.)

Auto Priming Indicator

When auto priming, the bar will progress to show auto prime feature is active.

Volume Setting

Indicates alarm volume setting.

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

Error Indicator

Pump operations are stopped until the error is resolved.

Information Indicator

This indicates information only, and does not require any immediate action.

Section IV - Initial Setup

Attaching the A/C Adapter Power Cord

Plug the Power Cord into the A/C Adapter Port on the back of the pump. See Figure 1A to locate port on pump. Make note of the location of the pins and tab on the plug to properly orient the plug.

Figure 2. Alignment of

A/C Adapter plug with pump’s adapter port.

The connector, which attaches to the back of the pump, has a spring-loaded locking sleeve that locks into the pump, thus helping to prevent the connector from being accidentally knocked out. To remove the connector, grasp the outer, sliding sleeve member and pull. Do not detach the connector by pulling on the wire only.

Battery Setup

The battery pack is fully charged before shipping, and is left unconnected to optimally preserve battery life.

WARNING: The battery should be connected before using the pump.

It is also recommended that the battery charge be “topped off” before battery-only operation is utilized. The battery pack will charge continuously whenever the pump is plugged into a wall outlet, including during normal use of the pump with

A/C power. Eight hours of charging is required to fully recharge the battery pack.

A new battery pack will provide approximately 15 hours of service before needing to be recharged.

KANGAROO

e

Pump

*

5

Attaching Pole Clamp

The KANGAROO ePump enteral feeding pump can be attached to a vertical pole using the Pole Clamp, included with the pump.

To attach the Pole Clamp to the KANGAROO ePump enteral feeding pump, align the hole on the Pole Clamp with the mounting hole on the back of the pump. Use a 3/8” round head bolt and rubber washer to fasten together, as shown in

Figure 3.

6

Figure 3. Attaching and using the Pole Clamp.

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

Section V - Instructions For Use

Quick Start

1. Press POWER on lower, right-hand corner of control panel.

2. Open blue door enclosing Pump Set loading area.

3. To load Pump Set (refer to Figure 4):

• Grasp finger tab on valve and insert into left pocket (1).

• Grasp black ring retainer and gently stretch tubing around rotor (2). Pull up on retainer and insert into right pocket (3).

• Press down on finger tab to ensure that valve is fully seated.

• Close blue door.

(Note: display status line should read SET LOADED).

4. To automatically prime the pump, pressPRIME PUMP andAUTO PRIME. For Feed & Flush sets, auto priming will prime both lines.

5. To directly control pump priming, pressPRIME PUMP and then press and hold the HOLD TO PRIME button(s). If a Feed & Flush Set is used, remember to first pressHOLD TO PRIME FLUSH until the flushing fluid has reached the valve, and then press HOLD TO PRIME FEED until the feeding fluid has reached past the valve and down to the distal end connector.

6. Use buttonson left side of LCD to set the feeding parameters. Set flushing parameters if a Feed & Flush Set is loaded. The continuous or intermittent feed mode selection is made under the MORE options menu.

7. PressRUN when ready to start. The screen will display RUNNING.

8. To stop, pressHOLD or press and hold the POWER button.

Finger

Tab

Black Ring

Retainer

(MISTIC)

Valve

1

3

Figure 4.

Installation of

Pump Set.

2

Rotor

KANGAROO

e

Pump

*

7

General Startup

Placement/Mounting

There are two recommended placement methods for the KANGAROO ePump enteral feeding pump.

• Attached to a vertical IV pole via the Pole Clamp peripheral device, included with the pump.

• Placed feet down on any horizontal surface.

A/C Power Operation

Plug the KANGAROO ePump enteral feeding pump into an A/C power outlet for normal operation. When the pump is

not plugged in, or if A/C power is interrupted, a built-in rechargeable battery will run the pump.

Battery Power Operation

If A/C power is not available to the pump, or if A/C power is interrupted, the pump will automatically operate on

backup battery power. The built-in battery recharges any time the pump is plugged into an outlet.

A fully-charged, new battery will supply 15 hours of backup power at 125 ml/hr before requiring recharging. Contact

Customer Service, Section XI, to replace the Battery Pack if battery life degrades.

Power On/Off

To power up the pump, press the POWER button on the lower right of the front panel. To power down the

pump, press and hold the POWER button until the pump display turns off.

Keep or Clear Prior Pump Settings

Immediately after powering up the pump, if the prior feed or flush settings were not cleared and the settings are not

locked in the BioTech Screen, the pump will give the following two options:

8

KEEP SETTINGS - Select this option to start with the same settings that were most recently programmed into the pump. Any of the previously-programmed settings can be modified, if needed. The feed totals,

mL FED” (Continuous Feed mode) or “BOLUSES FED” (Intermittent Feed Mode) are also maintained.

CLEAR SETTINGS - Select this option to reset all input feeding settings to zero. It will then be necessary to program all settings before running the pump. Note that only input settings are cleared. The feed totals,

mL FED” (Continuous Feed mode) or “BOLUSES FED” (Intermittent Feed Mode) will not be reset to zero.

To clear output feed totals, “mL FED” or “BOLUSES FED”:

• Program the input settings

• Start the pump RUNNING

• Immediately press the HOLD button.

• Select the top button, CLEAR VOL. FED

The status line of the pump will then show LOAD A SET (Figure 5), or, if the Pump Set is already loaded, it will show

SET LOADED (Figure 7).

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

Loading Pump Sets

The KANGAROO ePump enteral feeding pump will indicate LOAD A SET in the screen’s status line if a Pump Set is not installed. A blinking icon of a Pump Set will also appear in the upper left corner of the screen. Figure 5A shows the screen for continuous feeding mode and Figure 5B shows the screen for intermittent feeding mode.

Figure 5A. LOAD A SET opening menu screen for Continuous

Feeding mode.

Figure 5B. LOAD A SET opening menu screen for Intermittent

Feeding mode.

To load a Pump Set, follow the steps below:

1. Open the blue transparent door that encloses the rotor and Pump Set loading area.

2. Load the Pump Set by holding the finger tab on the valve, and inserting the valve down into the slot on the left.

(Figure 6A)

3. Grasp the MISTIC retainer end (black ring retainer) and wrap the tubing counterclockwise around the rotor.

(Figure 6B)

4. Carefully pull the MISTIC retainer end upwards to position it over the slot on the right, and then lower the retainer into the slot. (Figure 6B)

5. Press down on the valve’s finger tab to ensure proper seating.

6. Close blue transparent door.

KANGAROO

e

Pump

*

Figure 6A.

KANGAROO

e

Pump

*

Figure 6B.

9

The display should read SET LOADED and will show one of the following opening menus, depending on the type of

Pump Set detected, and whether continuous or intermittent feeding mode has been selected.

Figure 7A. Opening menu for Continuous Feed mode with a Feed-Only set.

Figure 7B. Opening menu for Continuous Feed mode with a Feed & Flush set.

Figure 7C. Opening menu for Intermittent Feed mode with a Feed-Only set.

Prime Pump

Figure 7D. Opening menu for Intermittent Feed mode with a Feed & Flush set.

The KANGAROO ePump enteral feeding pump may be primed automatically with a single button push, including the flushing line if a Feed & Flush Set is loaded. The pump can also be primed in a more interactive method using the holdto-prime feature.

After a Pump Set has been loaded and the status line shows SET LOADED, press PRIME PUMP in the opening menu

(Figure 7) to get one of the PRIME PUMP menus shown in Figure 8.

10

Figure 8A. PRIME PUMP menu for Feed-Only set.

Figure 8B. PRIME PUMP menu for Feed & Flush set.

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

Auto Priming

The AUTO PRIME option will be available if the pump senses that the currently installed Pump Set was not

previously primed, and that there is no fluid in the line near the rotor. If auto-priming is desired, press and

releaseAUTO PRIME to automatically prime the line(s). For Feed & Flush Pump Sets, the auto prime feature will

automatically prime both lines, starting with the flush line.

Press STOP to cancel AUTO PRIME.

When auto-priming is completed, the status line will show AUTO PRIME COMPLETE and will no longer show the

AUTO PRIME option. Be sure that the line(s) are fully primed. If not, use the hold-to-prime option(s) to complete

the priming, as described below.

Hold-to-prime

The hold-to-prime options allow for precise interactive control of the priming process.

Feed-Only Pump Sets

For Feed-Only Pump Sets, the screen will appear as in Figure 8A. Press and holdHOLD TO PRIME FEED until the feed line has been primed down to the stepped end connector at the end of the Pump Set.

Feed & Flush Pump Sets

For Feed & Flush Pump Sets, the display will appear as in Figure 8B. First press and holdHOLD TO PRIME

FLUSH to prime the flushing line, and then press and holdHOLD TO PRIME FEED to prime the feed line down to the stepped end connector at the end of the Pump Set.

If a Feed & Flush Pump Set is used, remember to first press HOLD TO PRIME FLUSH until the flushing fluid has reached the valve, and then pressHOLD TO PRIME FEED until the feeding fluid has reached past the valve and down to the stepped end connector. If the feeding line is primed first, subsequent priming of the flushing line will force the air between the flush solution bag and valve into the main line, which already has been primed with enteral formula.

Re-priming after Bag Empty

A Pump Set bag that has been emptied will trigger the FEED ERROR screen. In this condition the Pump Set

bag can be refilled to continue the feeding, but only after the Pump Set has been re-primed.

To re-prime the Pump Set, do the following:

• Disconnect the feeding line from the patient

• Refill the bag

• Press

CONTINUE to begin the pump RUNNING

• Press HOLD

• Press

ADJUST SETTINGS

• Press

HOLD TO PRIME to prime the line(s)

Pump Sets should not be reused after 24 hours of initial usage.

KANGAROO

e

Pump

*

11

Selecting Continuous or Intermittent Feed Mode

The KANGAROO ePump enteral feeding pump can be programmed to feed continuously or intermittently.

The continuous feeding mode will deliver the enteral nutrition at a steady rate, either until the programmed volume has been delivered or until the supply has been exhausted. The Intermittent feeding mode delivers boluses of enteral nutrition at programmed time intervals. The bolus volume and feed rate are also programmed.

To select continuous feed mode or intermittent feed mode, press MORE on the opening menu, shown in Figures

5 or 7, to access the MORE OPTIONS menu, Figure 9A.

The MORE option is available either before a Pump Set has been loaded (Figure 5) or after the Pump Set is loaded

(Figure 7). See subsection “More Options” for information on the other options in the MORE OPTIONS menu.

SelectCONTINUOUS/INTERMITTENT under the MORE OPTIONS menu, Figure 9A. Then press 

CONTINUOUS MODE orINTERMITTENT MODE from the SELECT MODE menu, Figure 9B. PressDONE to exit.

Figure 9A. MORE OPTIONS menu.

Figure 9B. SELECT MODE menu for continuous or intermittent mode selection.

Continuous Mode (Adjust Feed)

Continuous Mode Feeding

For CONTINUOUS feeding mode, the ADJUST FEED option is available in the opening menus (Figure 7A for

Feed-Only sets or Figure 7B for Feed & Flush sets). Select this option to program the feeding parameters from the

ADJUST FEED screen, Figure 10A.

PressFEED RATE in Figure 10A to define the rate of delivery. Press thebuttons in Figure 10B to program the

rate from 1 to 400 in increments of 1 ml/hr. PressENTER (top button) to exit the menu.

PressFEED VTBD in Figure 10A to define the Volume To Be Delivered. Press thebuttons in Figure 10C to

program the volume from 1 to 3000 ml in increments of 1 ml. PressENTER (top button) to exit the menu. If the

FEED VTBD is set to zero, the pump will run until the bag is emptied.

12

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

Figure 10A. ADJUST FEED menu. Figure 10B. Setting the Feed

Rate.

Continuous Mode Flushing

Figure 10C. Setting the Feed

Volume to be Delivered.

For CONTINUOUS feeding mode with a Feed & Flush Set loaded, the opening menu will appear as shown in

Figure 7B. Press ADJUST FLUSH in Figure 7B to program the flushing parameters from the ADJUST FLUSH

screen, shown in Figure 11A.

PressFLUSH VOLUME in Figure 11A to define the volume per flush occurrence. Press thebuttons in

Figure 11B to program the flush volume from 10 to 500 in increments of 1 ml. Press ENTER (top button) to exit

the menu.

Press FLUSH INTERVAL in Figure 11A to define the time interval between starts of flushing. Press thebuttons

in Figure 11C to program the time interval from 1 to 24 hours in increments of 1 hour. PressENTER (top button) to

exit the menu.

The pump will automatically limit flushing settings which exceed pump capabilities to deliver.

CAUTION: The KANGAROO ePump enteral feeding pump flushes at a rate of 1960 mL/hr (32.7 mL/minute). Use care when programming the flush volume so that it matches the patient’s need.

NOTE: The pump will automatically flush an additional 25 ml of water after completing the programmed delivery of formula, as long as at least one flush had occurred during formula delivery.

KANGAROO

e

Pump

*

Figure 11A. ADJUST FLUSH menu, continuous mode.

Figure 11B. Setting the Flush

Volume.

KANGAROO

e

Pump

*

Figure 11C. Setting the Flush

Time Interval.

13

Intermittent Mode (Adjust Bolus)

Intermittent Mode Feeding

For INTERMITTENT feeding mode, the ADJUST BOLUS option is available in the opening menus (Figure 7C for a

Feed-Only set or Figure 7D for a Feed & Flush set). Select this option to program the feeding parameters from the

ADJUST BOLUS screen, Figure 12A.

PressBOLUS RATE in Figure 12A to define the rate of delivery. Press thebuttons in Figure 12B to program the

delivery rate from 1 to 400 in increments of 1 ml/hr. Press ENTER (top button) to exit the menu.

PressNUMBER OF BOLUSES in Figure 12A to define the total number of boluses to be delivered. Press the

buttons in Figure 12C to program the number of boluses, up to a maximum of 99. PressENTER (top button) to

exit the menu. The number of boluses can be set as zero, in which case the pump will run until the bag is emptied.

PressVOLUME PER BOLUS in Figure 12A to define the Volume to be delivered in each bolus. Press

thebuttons in Figure 12D to program the volume per bolus from 1 to 1000 ml in increments of 1 ml.

PressENTER (top button) to exit the menu.

PressBOLUS INTERVAL in Figure 12A to define the time interval between starts of bolus deliveries. Press

thebuttons in Figure 12E to program the time interval between starts of bolus deliveries from 1 to 24 hours in

increments of 1 hour. PressENTER (top button) to exit the menu.

The pump will automatically limit bolus settings which exceed pump capabilities to deliver. For example, a bolus

delivery volume of 200 mL cannot be made once per hour if a rate of 150 mL/hr is already programmed. In this case,

the maximum volume allowed would be 150 mL.

14

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

Figure 12A. ADJUST

BOLUS menu.

Figure 12B. Setting the Bolus

Rate.

Figure 12C. Setting the Number of Boluses.

KANGAROO

e

Pump

*

Figure 12D. Setting the volume per Bolus.

Figure 12E. Setting the time interval between starts of bolus.

Intermittent Mode Feeding - Bolus Max

The BOLUS MAX option can be used to deliver a bolus at a high rate, similar to gravity feeding. PressBOLUS

MAX in the BOLUS RATE menu, Figure 12B, to program the BOLUS MAX mode. The feeding rate in Figure 7

will show “999” during RUNNING, indicating that the high feed rate is active.

Intermittent Mode Flushing

For INTERMITTENT feeding mode with a Feed & Flush Set, the opening menu, shown in Figure 7D, will show the

ADJUST FLUSH option. PressADJUST FLUSH in Figure 7D to program the flushing parameters from the

ADJUST FLUSH screen, Figure 13A.

KANGAROO

e

Pump

*

15

Figure 13A. ADJUST FLUSH menu, intermittent mode.

Figure 13B. Setting the Flush

Volume.

Figure 13C. Setting the Flush

Time Interval.

Note: The nature of two intermittent processes, periodic bolus feeding and periodic flushing, is such that occasional conflicts can arise where each process needs to be running at the same time. Because the Feed-Only type of Pump Set is available, and is dedicated only to feeding, the flushing process is assigned precedence over the feeding process for Feed & Flush pump sets. “Flushing precedence” means that a flush cycle will run to its end even if a bolus delivery is scheduled to begin. It also means that a flush cycle will interrupt a bolus feeding cycle that is already in progress. The feeding cycle will pick up where it was interrupted after the flushing cycle is complete. It is therefore possible that the net feeding volume is reduced from what may be intended, over a given time period.

PressFLUSH VOLUME in Figure 13A to define the volume per flush cycle. Press thebuttons in Figure 13B to program the flush volume from 1 to 500 in increments of 1 ml. Press ENTER (top button) to exit the menu.

PressFLUSH INTERVAL in Figure 13A to define the time interval between starts of flushing. Press thebuttons in

Figure 13C to program the time interval from 1 to 24 hours in increments of 1 hour. PressENTER to exit the menu.

The pump will automatically limit flushing settings which exceed pump capabilities to deliver.

CAUTION: The KANGAROO ePump enteral feeding pump flushes at a rate of 1960 mL/hr (32.7 mL/minute). Use care when programming the flush volume so that it matches the patient’s need.

NOTE: The pump will automatically flush an additional 25 ml of water after completing the programmed delivery of

formula, as long as at least one flush had occurred during formula delivery.

Run

The KANGAROO ePump enteral feeding pump will only run when a Pump Set is properly loaded and the necessary feeding parameters have been programmed. For a Feed & Flush Set, the flushing parameters do not have to be set for the pump to begin operating. If flushing parameters are set as zeros, there will be no flushing.

After the pump has been programmed, press

RUN to start operation. Figure 14 shows the RUNNING screen for continuous feeding with flushing. The status line will indicate RUNNING, and the droplet icon will move vertically on the screen. The green LED indicator will be illuminated to give a quick visual indication of positive status in a darkened room.

16

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

The RUN option can be selected from the following screens, if the proper parameters are programmed and the Pump Set is loaded:

SET LOADED screens (Figure 7)

ADJUST FEED screen, continuous mode (Figure 10A)

ADJUST FLUSH screens (Figure 11A, 13A)

HOLDING screens (Figure 15)

Figure 14. RUNNING mode screen for continuous feed mode.

Hold

Pump operation can be paused by pressingHOLD while it is RUNNING, such as in Figure 14. The status line will

indicate HOLDING and the yellow LED indicator will be lit up to give a quick visual indication of a warning status in a

darkened room.

Note: the pump’s alarm will sound if it has been left HOLDING for 10 minutes without further input. It can be

programmed to automatically restart after a specified time period by pressing theRESUME IN ___ MIN.

button.

There are five options available during HOLDING. See Figures 15A and 15B.

KANGAROO

e

Pump

*

Figure 15A. HOLDING mode options, continuous feeding mode.

Figure 15B. HOLDING mode options, intermittent feeding mode.

KANGAROO

e

Pump

*

17

Clear Volume Fed

PressCLEAR VOL FED to re-zero the mL FED output status on the right side of the screen of Figure 15A, or to re-zero the BOLUSES FED output status in Figure 15B.

Adjust Settings

Press ADJUST SETTINGS to return to one of the opening menus shown in Figure 7, thus allowing any setting to be modified. If the Pump Set has been removed after going into HOLD, the opening menu will show LOAD A SET, as seen in Figure 5.

Resume After Specified Time Interval

PressRESUME IN __ MIN to immediately input a time in minutes in which to resume operation. Each press of  RESUME IN MIN will increment the resume time by 5-minute increments, up to 30 minutes.

After “30” is reached, each additional button press will increment in larger time steps, up to a maximum of 240 minutes. No other input is needed to start the countdown, as the time immediately counts down after the incremental value is reached.

Run Mode

PressRUN to immediately restart pump operation. This option will not be available if the Pump Set is removed during HOLD.

View History

PressHISTORY to access the HISTORY screen. Use this screen to review feeding and flushing history.

See subsection “More Options” below for further information.

Flush Now

During the RUNNING mode, the pump can be diverted to immediately perform a flush (flush on demand) by

pressingFLUSH NOW in the RUNNING menu. See Figure 14. WhenFLUSH NOW is pressed, the screen

will prompt for the volume to be immediately flushed, and will show the previously programmed flush volume as

default. The screen will appear as the standard Flush input screen as shown in Figure 11B (or 13B). Setting

the FLUSH NOW flushing volume is temporary and will not change the main/periodic flushing volume previously

programmed. IfFLUSH NOW is pressed by accident, the screen will return to the RUNNING screen after 10

seconds, if there is no input for the flush volume.

18

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

More Options

PressMORE in one of the opening menus, Figures 5 or 7, to access the MORE OPTIONS screen, Figure 16.

Figure 16. MORE

OPTIONS

menu.

Buzzer Alarm Volume

PressBUZZER in the MORE OPTIONS menu to access the BUZZER screen and increase or decrease the alarm

buzzer volume. The alarm will be audible as the volume setting is changed. This sets the volume level for any

warning or error that may occur with the pump.

History

PressHISTORY in the MORE OPTIONS menu to access the HISTORY screen. Press thebuttons in Figure 17

to set the time history of interest. The totals of FEED mL and FLUSH mL will be shown for the requested history

time.

WARNING: A maximum of 72 hours of past history can be retrieved, excluding data from the current hour.

Figure 17. Feed and Flush totals for up to 72 hours of previous history, excluding data from the current hour.

Language Selection

PressLANGUAGES in the MORE OPTIONS menu to change the preferred language for the screen interface.

Continuous/Intermittent Mode

PressCONTINUOUS/INTERMITTENT In the MORE OPTIONS menu to access the SELECT MODE menu,

Figure 9B. Use this menu to select between continuous feeding mode and intermittent feeding mode. See subsection

“Selecting Continuous or Intermittent Feed Mode” for more information.

PressDONE in the MORE OPTIONS menu to exit the menu.

KANGAROO

e

Pump

*

19

BioTech Menu

The BioTech screen will show technical information about the pump, and has a lock feature to lock the Pump Settings against unauthorized changes.

Accessing the BioTech Screen

To get the BioTech screen, press the power button to start the pump, and press and hold the topbutton while the

startup sequence (kangaroo icon “hops” across the screen) executes. You must be pressing the topbutton at the

end of the sequence in order to access the BioTech menu. Figure 18 shows the BioTech screen.

Figure 18. BioTech screen, with

Pump Settings “locked” against unauthorized changes.

BioTech Screen Data

SN is the serial number of the pump.

SW shows the software version number in this pump.

FLASH shows the version number of the flash data installed in this pump.

CERT INTV indicates the number of times the pump has been powered on since the last re-certification. See

Section VI for details of re-certifying the pump.

Lock Settings Feature (Program Locked)

To “lock” the settings against unauthorized changes, pressLOCK SETTINGS to get the check mark, as shown in

Figure 18. The settings can only be changed again by pressingLOCK SETTINGS to remove the check mark.

If the settings are locked and a change is attempted, the screen will indicate PROGRAM LOCKED, and will not allow

access to the parameters, Figure 19.

The pump can be PRIMED or set to RUN, HOLD, etc. when settings are locked.

PressDONE to exit the BioTech menu. To re-enter this menu screen, power down the pump and then power it up

again and hold the topbutton, as described above.

20

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

Figure 19. Program Locked screen, which indicates the pump is in LOCK SETTINGS mode.

Lock Screen Feature

The KANGAROO ePump enteral feeding pump can be set to RUN and then locked out from unauthorized use. While the

LOCK SETTINGS feature prevents only changes to the input parameters, the LOCK SCREEN feature will lock out any button presses, including RUN or HOLD, except power off.

Use LOCK SCREEN when the pump is running and there is a need to prevent tampering such as putting the pump into an unwanted HOLD. The RUNNING screen will show a padlock to indicate that no changes can be made without unlocking the screen.

To lock the screen, use the following procedure while the pump is in RUNNING mode:

• Press and hold the secondbutton from the top for at least 2 seconds.

• Press and hold the thirdbutton from the top for at least 2 seconds.

• Press and hold the secondbutton again for at least 2 seconds.

To unlock the screen, use the same procedure above.

Figure 20. Padlock indicates pump is in

LOCK SCREEN mode.

Section VI - Re-Certification of KANGAROO ePump Performance

System Performance Re-Certification

A series of self-diagnostic checks can be run to re-certify pump performance. It is recommended that this

re-certification procedure be run at least once per year, or any time the pump is suspected of having improper

performance. See Section XIII-Service Part Numbers for ordering new Re-Certification Pump Sets.

Load the Re-Certification Pump Set. The pump will automatically recognize the Re-Certification Pump Set and give

KANGAROO

e

Pump

*

21

brief instructions to initiate the self-diagnostic procedure. Follow the instructions and review the summary

information at the end of the test.

Certifying the Accuracy of Pump Flow Rate

Use the following procedure to check the flow rate accuracy:

• Perform test at room temperature (72° F ± 3° F) with a new Pump Set.

• Fully fill a new Feed-Only Pump set bag with water.

• Load the Pump Set.

• Suspend the Pump Set bag so that the bottom of the bag is 18 inches above the pump.

• Make sure pump is plugged into A/C power (do not run test with battery power).

• Program a Continuous Mode feed rate (see below for suggested example rates).

• Run the pump for 15 minutes to achieve steady state operation.

• Then begin collecting the water in a calibrated measuring container, such as a large-capacity graduated cylinder, for

exactly 30 minutes.

NOTE: The amount of water collected in 30 minutes will be half of the value programmed for the hourly flow rate,

within ± 10%, or 0.5 ml, whichever is larger.

Rate Set Amount Expected in 30 Minutes

75 mL/hr 33.7 mL to 41.3 mL

150 mL/hr 67.5 mL to 82.5 mL

If the amount collected is outside the range, load a new Pump Set and re-run the test procedure to confirm results.

Section VII - Cleaning

Cleaning should be performed as needed. It may also be desireable to define cleaning intervals based on knowledge of the environment in which the pump is used. Only personnel trained in the cleaning of medical devices should perform cleaning.

CAUTION:

Do not immerse pump or power cord in water or other cleaning solution; clean using a damp (not wet) cloth or sponge. Failure to follow the cleaning procedures described herein could result in hazards to users. As with any AC powered electrical device, care must be taken to prevent liquid from entering the pump to avoid electrical shock hazard, fire hazard, or damage to electrical components.

If any of the following events occur, DO NOT USE the pump until it has been properly cleaned and serviced by personnel trained in servicing KANGAROO ePump enteral feeding pumps:

• wetting of the pump’s power cord or leakage into the pump interior during cleaning.

• spillage of large amounts of formula onto the pump exterior, or any spillage onto the power cord.

22

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

General Cleaning Directions

Cleaning of KANGAROO ePump enteral feeding pumps must be performed as follows:

CAUTION:

Disconnect pump from AC power source before cleaning. After cleaning, do not connect to AC power source until pump and power cord are thoroughly dry.

• A mild detergent should be used for general cleaning. If necessary, the pump may be cleaned with a 10:1 water

and hypochlorite mixture, however, repeated cleaning with this solution can damage the plastic pump case.

Isopropyl alcohol, applied with a damp (not wet) cotton swab, may be used for cleaning difficult-to-reach areas,

however, it should be used sparingly because repeated cleaning may damage the case.

• DO NOT USE strong cleaners such as

Spray-Nine,

Phisohex,

Hibiclens, or

Vesta-Syde because damage to the

pump case housing can result.

Directions for Cleaning Pump Housing

• Refer to General Cleaning Directions before starting.

• Clean outside surface with a damp (not wet) cloth or sponge and keep pump in upright position as much as

possible.

• Avoid excess moisture near pole clamp area.

• Do not allow cleaning solution to enter the vertical vents in the housing, underneath the handle.

Directions for Cleaning Pump Power Cord

• Refer to General Cleaning Directions before starting.

• Unless soiling is observed, the power cord should not be cleaned.

• If cleaning of the power cord is necessary, unplug from outlet and wipe the exterior surfaces of the wall plug

with a cloth dampened with isopropyl alcohol.

CAUTION:

Avoid exposing power cord to excess moisture, as this can lead to an electrical shock or fire hazard.

Directions for Cleaning Rotor Assembly

• Refer to General Cleaning Directions before starting.

• Open the blue door enclosing the Pump Set loading area.

• Loosen rotor set screw with 5/64” (2mm) allen wrench and gently pull rotor off shaft. After removing rotor, avoid

getting any moisture in the rotor shaft opening.

• Clean rollers thoroughly with warm soapy water, or isopropyl alcohol if necessary.

• Be sure all parts of rotor are completely dry before putting it back onto shaft.

• To replace rotor, align set screw on hub of rotor with the flattened section of the output shaft.

• Push rotor into place and tighten set screw (Do not over tighten).

KANGAROO

e

Pump

*

23

Preventative Maintenance

This pump must be periodically re-certified to assure proper functioning and safety. The recommended service interval is at least once per year. Re-certification may be done at the user’s Biomedical Engineering Department, an outside service, or by Tyco Healthcare Group LP Factory Service. To arrange for Tyco Healthcare Group LP Factory Service, call 1-800-448-0190.

In Canada, call 1-877-664-TYCO(8926)

If a pump malfunctions, please contact your THGLP Representative or call Customer Service for instruction.

Section VIII - Battery Replacement

See Figure 21 and do the following to replace the Battery Pack:

• Power down the unit.

• Loosen the Battery Door screw, on back of pump.

• Open Battery Door and disconnect wire harness. A small straight-bladed screw driver may be needed to disconnect

the connector tab.

• Slide Battery Pack out of compartment.

• Slide a new Battery Pack into compartment with wire harness extending out.

• Align connector with red wire at upper right.

• Push connector in until properly latched.

• Tuck wires into slot between connector and battery to insure wires are not pinched. Re-tighten Battery

Door screw.

Note: Replacement Battery Packs are available from Tyco Healthcare/Kendall Co. (See Section XIII - Service Part

Numbers).

Figure 21. Installation of Battery Pack.

WARNING: USE OF A BATTERY PACK NOT SUPPLIED BY TYCO HEALTHCARE/KENDALL CAN BE DANGEROUS AND

VOIDS ALL WARRANTIES AND PERFORMANCE SPECIFICATIONS.

The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is required to fully recharge the Battery Pack.

24

KANGAROO

e

Pump

*

Section IX - Error/Warning/Informational Screens and Troubleshooting

For operational errors, the KANGAROO ePump enteral feeding pump display will show the error indicator icon, with specific information regarding the error condition, and will sound the alarm.

The following conditions will trigger error conditions and will activate audible and visible alarms:

• Feed or Flush containers are empty

• Feed or Flush tube becomes occluded between bag and pump

• Tube becomes occluded between pump and patient

• Battery low (alarm sounds momentarily, about 15 minutes before shut-off)

• Unit is left in HOLD mode longer than 10 minutes without input

• Pump Set tubing is improperly loaded around the rotor

• System errors, see below

• Feeding complete

• The MISTIC connector is removed during AUTOPRIMING or RUNNING

The Error, Warning, and Information Screens are described below:

System Error

The system error screen is the most general form of error, Figure 22. The only way to exit from a System Error is to power down. No CONTINUE option is allowed because of the severity of the error. An error number is also displayed on the screen, for reference purposes. This number should be reported when calling the Customer Service Support Line.

Error # Related Sub-system

0 Internal timer stopped.

1 Flash Chip missing OR version incompatible.

2 Too many graphic screens in memory.

4 Never got a ADC ISR for motor current.

5 Never got a ADC ISR for battery level.

6 Never got a ADC ISR for battery temperature.

7 Never got a ADC ISR for Ultrasonic Upstream Occlusion or Bag-Empty checks.

8 Never got a ADC ISR for Ultrasonic during Downstream Occlusion.

9 RS232 Communications Transmit.

10 Flash Write Error.

11 Flash Read Error.

12 Stopcock Time Out.

13 Time Out for Buzzer Test

KANGAROO

e

Pump

*

25

Error # Related Sub-system

14 Time Out during downstream occlusion check, never got RotorOn3=0. May blow fuse.

20 Could not find Stop Cock Flush Position during prime routine.

21 Could not find Stop Cock Feed Position during prime routine.

22 Motor Error during Tubing AutoLoad.

23 Motor Failsafe Time Out Error. Possible damaged gearbox encoder or flex cable.

50 Graphic Display Error - Timeout during GDStatusCheck function.

51 Graphic Display Error - Timeout during GDAutoModeDataWriteCheck function.

52 Graphic Display Error - Timeout during GDWriteBytes function.

Figure 22. General System

Error screen. See Customer

Service section.

If the error cannot be resolved in a timely manner, press POWER to stop operation of the pump and put a different pump into service.

Hold Error

The HOLD ERROR screen will appear if the pump has been in HOLD mode, without input, for more than 10 minutes.

See subsection “Hold” for description of HOLDING mode.

PressCONTINUE to return to the HOLDING screen, where settings can be adjusted, the pump can be set to run immediately, or the pump can be set to run in a specified number of minutes. See Figure 15 to see the options on the

HOLDING menu screen.

If the error cannot be resolved, press POWER to stop operation of the pump and put a different pump into service.

Figure 23. Hold Error screen.

26

KANGAROO

e

Pump

*

Rotor Error

The ROTOR ERROR screen appears during RUNNING or PRIMING, when the pump has detected an unusual operating condition for the rotor. See Figure 24. The ROTOR ERROR generally results from a problem with the Pump Set tubing around the rotor.

Check that the Pump Set is not damaged, and re-load the tubing around the rotor as described in subsection “Loading

Pump Sets”. Press CONTINUE to restart.

If the error cannot be resolved, press POWER to stop operation of the pump and put a different pump into service.

Figure 24. Rotor Error screen.

Re-load the set tubing and press

CONTINUE.

Feed Error

The FEED ERROR screen appears when the enteral formula is no longer being delivered because the bag is empty or because of a clog between the pump and the bag. Check the feed line to find the occlusion causing the blockage. If the occlusion cannot be cleared, load a new Pump Set, prime it, and press

CONTINUE to restart.

If the error still cannot be resolved, press POWER to stop operation of the pump and put a different pump into service.

Figure 25. Feed Error screen.

Flush Error (with Feed & Flush Set only)

The FLUSH ERROR screen appears when the flushing solution is no longer being delivered because the flush bag is empty or because of a clog. Check the flush line to find the occlusion causing the blockage. If the occlusion cannot be cleared, load a new Pump Set, prime it, and press 

CONTINUE to restart.

KANGAROO

e

Pump

*

27

If the error still cannot be resolved, press POWER to stop operation of the pump and put a different pump into service.

Figure 26. Flush Error screen.

Flow Error

The FLOW ERROR screen appears when either the feeding or the flushing solution is no longer being delivered because of a clog. Check the line to find the occlusion causing the blockage. If the error cannot be resolved, check the valve pocket in the Pump Set loading area for dampness or dirt, which may be a cause for a false error. Clean and dry the valve pocket.

If the error still cannot be resolved, load a new Pump Set, prime it, and press

CONTINUE to restart.

If the error still cannot be resolved, press POWER to stop operation of the pump and put a different pump into service.

Figure 27. Flow Error screen.

Pump Set Dislodged Error

The PUMP SET DISLODGED screen will appear if the black ring retainer (MISTIC) is not properly loaded in the MISTIC pocket in the Pump Set loading area. Check and correct the positioning of the MISTIC retainer if possible. If the error cannot be resolved, load a new Pump Set, prime it, and press CONTINUE to restart.

If the error cannot be resolved, press POWER to stop operation of the pump and put a different pump into service.

Figure 28. Pump Set Dislodged Error screen.

28

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

KANGAROO

e

Pump

*

Battery Low

The BATTERY LOW screen appears when the battery needs to be recharged. There is approximately 15 minutes of battery life remaining when this screen appears.

Plug the power cord in to begin recharging. The pump will automatically return to the screen that was active prior to the error. The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is required to fully recharge the battery pack.

Figure 29. Battery Low error screen.

Indicates that the battery must be recharged immediately.

Feeding Complete

The FEEDING COMPLETE information screen appears after completion of the programmed feeding. PressPower

Down to turn off the pump. PressCONTINUE to return to the SET LOADED opening menu, Figure 7.

Figure 30. Feeding Complete notification screen.

Pump Set Use >24 Hours Warning

The Pump Set usage warning indicator will blink on the RUNNING screen if a Pump Set has been used for 24 or more hours. It is recommended to replace Pump Sets after this length of usage. This icon is only an informational message and does not require action.

Figure 31 shows an example of the RUNNING screen, where a Feed & Flush Set has been in use for more than 24 hours.

The informational warning indicator shows up in the upper left on the screen.

Figure 31. Running screen with the >

24 HRS of Pump Set Usage indicator in upper left of the screen.

KANGAROO

e

Pump

*

29

LED Array

The pump status LED array on the upper right of the pump gives a quick visual indication of the pump status, especially in darkened rooms.

A green light status indicates normal pump operation.

A yellow light status indicates a warning situation for either a low battery condition, the pump in HOLDING mode, or one of the following errors: HOLD ERROR, FEED ERROR, FLOW ERROR, and FLUSH ERROR.

A red light status indicates a critical pump error. The screen will show one of the following error messages: SYSTEM

ERROR, SET DISLODGED ERROR, or ROTOR ERROR.

Section X - Specifications and Symbols

Specifications

UL Listing

Medical Electrical Equipment

KANGAROO ePump* Enteral Feeding Pump

(1) Classified with respect to electrical shock, fire, and mechanical hazards in accordance with UL60601-1.

(2) Classified with respect to electrical shock, fire, mechanical and other specified hazards in accordance with CAN/CSA C22.2 No. 601.1

Type Infusion Device

Volumetric

Pumping Mechanism

Rotary Peristaltic

Pump Sets

KANGAROO ePump MISTIC Feed-Only Set or Feed & Flush Set

Feeding Formula Delivery Rate

1-400 ml/hr in 1 ml increments

Feeding Formula VTBD

1-3000 ml in 1 ml increments

Bolus Volume

1-1000 ml in 1 ml increments

Number of Boluses

1-99

Bolus Interval

4-24 hours in 1 hour increments

30

KANGAROO

e

Pump

*

Flushing Solution Dose Range

10-500 ml in 1 ml increments

Flushing Solution Interval Range

1-24 hr in 1 hr increments

Accuracy

+/-10% or 0.5 ml/hr, whichever is larger, with bag at 18” above pump, at room temperature 72°F +/- 3°F,

using a new pump set with less than the recommended 24 hours of maximum usage.

Occlusion Pressure

15 psi (103 kPa) Maximum

Dimensions

Height: 6.6” Width: 6.4” Depth: 4.6”

Weight

2.4 lbs (3.3 lbs with pole clamp)

Material

Housing: ABS/Polycarbonate blend

Pump Set Door: Polyester/ Polycarbonate blend

Power

Use AC adapter for wall outlet usage. Pump operates on 9V DC, 1.7 Amps.

Battery

New, fully charged NiMH Battery Pack delivers ≥ 15 hrs at 100 ml/hr feed rate. Approximately 15 minutes prior to

complete battery discharge, a low battery alarm will occur (see subsection “Battery Low” in Section IX). When

complete discharge occurs, the pump will automatically turn itself off.

The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is

required to fully recharge the battery pack.

Alarms

• System Error

• Flow Error

• Pump Set Dislodged Error

• Hold Error

• Rotor Error

• Feed Error

• Flush Error

• Battery Low

Alarm Volume

Minimum of 70 dBA at 1 meter in maximum volume orientation

KANGAROO

e

Pump

*

31

32

Operating Temperature

10° - 40° C (50° - 104° F) 75% R.H. non-condensing

Storage Temperature

0° - 50° C (32° - 122° F) 95% R.H. non-condensing

Type of Protection Against Electrical Shock

Class II, Internally-powered Equipment

Degree of Protection Against Electrical Shock

Type BF

Mode of Operation

Continuous or programmed intermittent operation

Degree of Protection Against Ingress of Fluids

Drip-proof-IPX1

Symbols on Pump and Accessories

The following symbols are found on the pump or on the accesories, such as the A/C adapter.

����������������

Refer to Accompanying

Documents

�������

���������

Type BF Protection (Degree of protection against electrical shock - there is no conductive connection to the patient)

1

0123

2

CE Mark - European

Community Certification

3

Set Loading Guide

Manufacturing Date Code

UL in USA and Canada

��

TUV Certification

EN60601-1

IEC60601-1

TUV Certification

SN

REF

~

Serial Number

Reorder number for the device located on the carton label

Alternating Current

Class II Equipment (degree of protection against electrical shock), Double Insulated

Direct Current (DC)

IPX1

Drip Proof (Degree of protection against ingress of fluids)

Power Button

Buttons for Selecting

Screen Prompts

Power Source

9V DC 1.7 Amp

DB9 Serial I/O

Communication Port

(Certification and

Programming)

Caution: For indoor use only

Functional Ground

KANGAROO

e

Pump

*

Section XI - Customer Service

The circuitry of the KANGAROO ePump enteral feeding pump is not customer serviceable. In particular, electronic assembly rework by non-authorized KANGAROO ePump enteral feeding pump technicians will likely affect accuracy.

Certain replacement items, as listed in Section XIII - Service Part Numbers, are available from the service centers listed below.

All service personnel must be properly trained and qualified with operation of the KANGAROO ePump enteral feeding pump. Improper service may impair operation of the pump.

Return for Repair

1. Call Customer Service for an Authorized Return Number and shipping instructions, using the appropriate phone number below:

2. Pack the instrument carefully and ship the insured parcel to:

United States

Tyco Healthcare Group LP

98.6 Faichney Drive

Watertown, NY 13601

1-800-448-0190 or (315) 788-5246

Canada

Kendall Division of Tyco/Healthcare Canada Inc.

151 Whitehall Drive

Markham, Ontario, Canada L3R9T1

1-800-268-9895 or (905) 479-5500

U.K. Service Center (Outside of U.S. and Canada)

Tyco Healthcare Group LP

20 Garryduff Road

Ballymoney, BT53 7AP

001-44-12656-63234

Section XII - Maintenance

For general maintenance issues not discussed below, contact Customer Service (Section XI).

WARNING: Do not open the main housing, as there are no user-serviceable parts inside. Opening of device may affect function of device and voids the warranty.

The following maintenance items/parts can be replaced

Top

Hinge

Pin

Top Door

Hinge by the customer on the KANGAROO ePump enteral feeding pump . See Section XIII for Service Part Numbers and

Section XI for contacting Customer Service.

Side Door enclosing Pump Set Loading area

To replace the Side Door, see Figure 32 and do the following:

Lower Hinge

Pin

Lower Door

Hinge

1. Place top Door Hinge onto the Top Hinge Pin.

Figure 32. Blue Door Replacement.

2. Carefully press down on the bottom of the door to spread the door hinges apart.

3. Place the bottom Door Hinge onto the Bottom Hinge Pin.

KANGAROO

e

Pump

*

33

Battery Pack

See Section VIII for full instructions on replacing Battery Pack.

Battery Door

To replace Battery Door, see Figure 33 and do the following:

1. Align the Battery Door Hinges with the long hinge pins.

2. Snap the hinges onto the hinge pins.

3. Screw down the door to hold in place.

Power Cord

See Section IV for initial setup, including Power Cord attachment.

Figure 33. Battery Door Replacement.

Pole Clamp

See Section IV for initial setup, including attachment of the Pole Clamp to the pump.

Section XIII - Service Part Numbers

To place an order for parts, or if technical assistance is required, call customer service.

The KANGAROO ePump enteral feeding pump contains a limited number of serviceable parts. User maintenance is to be performed only by appropriately qualified technical personnel.

Visit our web site at: www.TycoHealthcare.com

(1) KANGAROO ePump . . . . . . . . . . . . . . . . .382400

(2) Main Door (Blue transparent door). .F31877WT

(3) Battery Pack . . . . . . . . . . . . . . . . . . .F010484WT

(4) Battery Door (with Screw). . . . . . . . . F31878WT

Power Cord (with A/C Adapter). . . . . . .F010486WT

(5) Pole Clamp . . . . . . . . . . . . . . . . . . . . . A11164WT

Re-Certification Pump Set . . . . . . . . . . . . . . 776150

2

1

3

Figure 34. User Serviceable Parts.

5

4

Section XIV - Warranty

LIMITED WARRANTY: Kendall, a Tyco Healthcare Group LP company, warrants to the original purchaser (“Customer”) that this product will be free of defects in materials and workmanship, under normal use, for one (1) year from the date of original purchase from Kendall or its authorized distributor. If this product does not operate as warranted above during the applicable warranty period, Kendall may, at its option and expense, replace the defective part or product with a comparable part or product, repair the defective part or product, or, if neither replacement nor repair is reasonably available, refund to Customer the purchase price for the defective part or product. Dated proof of original purchase will be required. Kendall does not assume any liability for loss arising from unauthorized repair, misuse, neglect, or accident.

34

KANGAROO

e

Pump

*

Section XV - Electromagnetic Conformity Declaration

The KANGAROO ePump enteral feeding pump has been built and tested according to UL60601-1, E181931 CSA-C22.2

No.601.1-M90, and EN60601-1-2 Standards.

The KANGAROO ePump enteral feeding pump is intended for use in the electromagnetic environment specified below.

The user of the pump should assure that it is used in such an environment.

Emissions Test Compliance

RF emissions

(CISPR 11) Group 1

RF emissions

(CISPR 11) Class B

Harmonic emissions

Electromagnetic Environment - Guidance

The KANGAROO ePump uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The KANGAROO ePump is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

(IEC 61000-3-2) Class A

Voltage fluctuations/ flicker emissions

(IEC 61000-3-3) Complies

Radiated Disturbance Immunity

(EN60601-1-2 / IEC 61000-4-3:2002) Complies

Conducted Disturbance Immunity

(EN60601-1-2 / IEC 61000-4-6:2001) Complies

Power Frequency Magnetic Field Immunity

(EN60601-1-2 / IEC 61000-4-8:2001) Complies

Voltage dips and sags Immunity

(EN60601-1-2 / IEC 61000-4-11:2001) Complies

Electrical Fast Transient / Bursts Immunity

(EN60601-1-2 / IEC 61000-4-4:2001) Complies

Electrostatic Discharge Immunity

(EN60601-1-2 / IEC 61000-4-2:2001) Complies

Surge Immunity

(EN60601-1-2 / IEC 61000-4-5:2001) Complies

KANGAROO

e

Pump

*

35

Patents pending.

*KANGAROO ePump is a trademark of Sherwood Services AG

SPRAY-NINE is a trademark Spray-Nine Corporation

PHISOPHEX is a trademark of Smith Kline Beecham Corp.

HIBICLENS is a trademark of Tubifoam Ltd.

VESTA-SYDE is a trademark of Steris, Inc.

TYCO HEALTHCARE GROUP LP

MANSFIELD, MA 02048

MADE IN U.S.A.

© 2004 TYCO HEALTHCARE GROUP

LP. ALL RIGHTS RESERVED. 040415

IN U.S. 1-800-962-9888

www.TycoHealthcare.com

EC REP TYCO

HEALTHCARE U.K. LTD, GOSPORT, P013OAS, U.K.

SWISS DISTRIBUTOR: TYCO HEALTHCARE SWITZERLAND LTD,

ROOSTRASSE 53, POSTFACH CH-8832, WOLLERAU SWITZERLAND

182321

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals