Sierra Monitor Corporation 5100-26-IT Instruction Manual


Add to my manuals
63 Pages

advertisement

Sierra Monitor Corporation 5100-26-IT Instruction Manual | Manualzz

MODEL 5100-XX-IT

IT Series

TOXIC GAS SENSOR MODULE

Version 3.00A

APPLICABILITY & EFFECTIVITY

Effective for all Model 5100-XX-IT Modules manufactured after December 1, 2011.

5100-03-IT

5100-04-IT

5100-05-IT FM Approved

5100-06-IT

5100-08-IT

5100-10-IT

5100-12-IT

5100-25-IT

5100-26-IT

Instruction Manual Part Number T12020

Rev. C

Model 5100-XX-IT Toxic Gas Sensor Module

FM PERFORMANCE APPROVAL

ONLY THE FOLLOWING ITEMS, FUNCTIONS AND OPTIONS ARE FM * APPROVED

Model 5100-05-IT Hydrogen Sulfide Gas Sensor Module

Sensor Module

Model 5100-05-IT

Calibration Equipment

Model 1250-01

Model 1260-05

Model 1260-45

Model 5358-01

Notes:

Sensor Module - Hydrogen Sulfide

Gas Sensor Calibration Kit, Type A

Hydrogen Sulfide in N2, 25 ppm Gas Cylinder

Hydrogen Sulfide in N2, 10 ppm Gas Cylinder

Calibration Adapter

1) Apparatus must be installed in accordance with National Electrical Code.

* FM Approvals, a subsidiary of FM Global

 Project# 3034435

Model 5100-XX-IT Toxic Gas Sensor Module

TABLE OF CONTENTS

1.

PRODUCT DESCRIPTION ................................................................................................................................. 1

1.1 GENERAL ........................................................................................................................................................ 1

CONFIGURATION ........................................................................................................................ 1

1.3 THEORY OF OPERATION .............................................................................................................................. 2

1.4 MODES OF OPERATION ................................................................................................................................ 2

1.5 MECHANICAL ................................................................................................................................................. 3

1.6 INTERCONNECT WIRING .............................................................................................................................. 3

2.

CAUTIONS & WARNINGS ................................................................................................................................. 6

2.1 INTRODUCTION ............................................................................................................................................. 6

2.2 GAS SENSOR MODULES - GENERAL .......................................................................................................... 6

............................................................................................................................................................ 6

2.4 CALIBRATION FREQUENCY ......................................................................................................................... 6

3.

QUICK START .................................................................................................................................................... 7

3.1 OVERVIEW

3.2 WIRING

INSTALLATION ............................................................................................................................... 7

3.4 TRANSMITTER

3.5 START-UP

STABILIZATION ................................................................................................................................... 7

4.

INSTALLATION .................................................................................................................................................. 8

MODULE LOCATIONS ................................................................................................................... 8

............................................................................................................................................................ 9

4.3 ENCLOSURE

4.4 TRANSMITTER AND SENSOR INSTALLATION ......................................................................................... 10

4.5 MODULE ADDRESS SWITCH ...................................................................................................................... 12

5.

OPERATION ..................................................................................................................................................... 18

5.1 INTRODUCTION ........................................................................................................................................... 18

5.2 DATA

5.3 MAIN MENU .................................................................................................................................................. 19

5.4 SET-UP .......................................................................................................................................................... 20

5.5 MAINTENANCE SUB-MENU ........................................................................................................................ 23

6.

CALIBRATION .................................................................................................................................................. 24

6.1 CALIBRATION FREQUENCY ....................................................................................................................... 24

6.2 CALIBRATION PREPARATION .................................................................................................................... 24

6.3 CALIBRATION GAS DELIVERY METHODS ................................................................................................. 24

6.4 SENSOR EXPOSURE TO GAS .................................................................................................................... 24

6.5 CALIBRATION SUB-MENU........................................................................................................................... 25

7.

SERVICE ........................................................................................................................................................... 26

7.1 SENSOR MODULE CONFIGURATION ........................................................................................................ 26

7.2 ENCLOSURE REPLACEMENT .................................................................................................................... 26

7.3 TRANSMITTER REPLACEMENT ................................................................................................................. 27

7.4 SENSOR REPLACEMENT ............................................................................................................................ 27

7.5 INSTALLATION INSPECTION ...................................................................................................................... 27

7.6 INSPECTION AND TROUBLESHOOTING GUIDE ....................................................................................... 28

Contents

Model 5100-XX-IT Toxic Gas Sensor Module

8.

APPENDICES ................................................................................................................................................... 29

8.1 APPENDIX A: SPECIFICATIONS.................................................................................................................. 29

8.2 APPENDIX B: MODEL NUMBERS & PARTS LIST ....................................................................................... 32

8.3 APPENDIX C: LIMITED WARRANTY ............................................................................................................ 33

8.4 APPENDIX D: REMOTE SENSOR DRAWING ............................................................................................. 34

8.5 APPENDIX E: MODBUS MEMORY MAP ...................................................................................................... 35

8.6 APPENDIX F: HART ...................................................................................................................................... 36

8.7 APPENDIX G: MENU KEY FOR 5100-XX-IT GAS SENSOR MODULE ........................................................ 43

8.8 APPENDIX H: 5100-25-IT & 5100-26-IT GAS SENSOR MODULE ............................................................... 46

8.9 APPENDIX I: 5100-05-IT FM PERFORMANCE APPROVAL ........................................................................ 50

8.10 APPENDIX J: 5100-XX-IT SIL-2 CERTIFICATES ......................................................................................... 52

8.11 APPENDIX K: 5100-05-IT ATEX CERTIFICATE ........................................................................................... 55

LIST OF FIGURES

FIGURE 1 - 1: MODEL 5100-05-IT-AL TOXIC GAS SENSOR MODULE ALUMINUM – DIMENSIONS ................ 4

FIGURE 1 - 2: MODEL 5100-05-IT-SS TOXIC GAS SENSOR MODULE STAINLESS STEEL – DIMENSIONS .. 4

FIGURE 1 - 3: MODEL 5100-XX-IT TOXIC GAS SENSOR MODULE – MOUNTING OPTIONS ........................... 5

FIGURE 4 - 1: TYPICAL MOUNTING OPTIONS ..................................................................................................... 8

FIGURE 4 - 2: TRANSMITTER FACE PLATE ....................................................................................................... 12

FIGURE 4 - 3: INTERFACE BOARD CONNECTORS ........................................................................................... 13

FIGURE 4 - 4: RS-485: TERMINATION, BIAS JUMPER ...................................................................................... 13

FIGURE 4 - 5: MA CIRCUIT TYPES ...................................................................................................................... 14

FIGURE 4 - 6: 4-20 MA CIRCUIT TYPE CONNECTIONS FOR 5100-XX-IT ........................................................ 15

FIGURE 4 - 7: WRITING CONNECTIONS FOR MODBUS AND SENTRY INTERFACE ..................................... 16

FIGURE 4 - 8: WIRING CONNECTIONS FOR REMOTE ALARM RESET ........................................................... 17

FIGURE 5 - 1: FACE PLATE WITH OPERATOR KEY PAD ................................................................................. 18

FIGURE 6 - 1: MODEL 5358-01 CALIBRATION ADAPTER ................................................................................. 24

FIGURE 7 - 1: MODULE COMPONENTS ............................................................................................................. 26

FIGURE 8 - 1: REMOTE SENSOR DRAWING ...................................................................................................... 34

FIGURE 8 - 2: MODULE COMPONENTS ............................................................................................................. 46

FIGURE 8 - 3: SENSOR ASSEMBLY – EXPLODED VIEW .................................................................................. 48

FIGURE 8 - 4: MODEL 5100-25-IT-AL AND MODEL 5100-26-IT-AL .................................................................... 49

FIGURE 8 - 5: MODEL 5100-25-IT-SS AND MODEL 5100-26-IT-SS ................................................................... 49

FIGURE 8 - 6: ATEX............................................................................................................................................... 55

LIST OF TABLES

TABLE 4 - 1: SPECIFIC GRAITIES .......................................................................................................................... 8

TABLE 4 - 2: MINIMUM WIRE GAUGES .............................................................................................................. 10

TABLE 4 - 3: SENSOR MODULE ADDRESS SWITCH POSITIONS .................................................................... 12

TABLE 5 - 1: MASTER MENU ............................................................................................................................... 19

TABLE 5 - 2: OPERATION DISPLAY VALUES ..................................................................................................... 19

TABLE 5 - 3: SET-UP CONFIGURATION ............................................................................................................. 21

TABLE 5 - 4: MAINTENANCE MENU .................................................................................................................... 23

TABLE 6 - 1: CALIBRATION .................................................................................................................................. 25

TABLE 8 - 1: MODBUS MEMORY MAP ................................................................................................................ 35

Contents

Model 5100-XX-IT Toxic Gas Sensor Module

1. PRODUCT DESCRIPTION

1.1 GENERAL

The Model 5100-XX-IT Toxic Gas Sensor Module is a smart transmitter and member of the IT Series family and it offers a broad array of features including:

• Integral alphanumeric LED display

• Up to 180 day calibration frequency

• FM Approval for performance and safety

• SIL-2

• Optional integral alarm relays (3)

• 4-20 mA output

• Modbus ® RTU interface

• SMC Sentry interface

• Optional HART interface

• 316 Stainless steel enclosure option

• Remote sensor option

• Low maintenance and operation costs

• ATEX Approval (5100-05-IT)

The 5100-XX-IT is designed and approved for installation and operation in hazardous locations.

Members of the 5100-XX-IT Series Toxic Gas family include:

• 5100-03-IT Oxygen Gas Sensor Module

• 5100-04-IT Carbon Monoxide Gas Sensor Module

• 5100-05-IT Hydrogen Sulfide Gas Sensor Module

• 5100-06-IT Chlorine Gas Sensor Module

• 5100-08-IT Chlorine Dioxide Gas Sensor Module

• 5100-10-IT Sulfur Dioxide Gas Sensor Module

• 5100-12-IT Nitrogen Dioxide Gas Sensor Module

• 5100-25-IT Ammonia Gas Sensor Module

• 5100-26-IT Hydrogen Fluoride Gas Sensor Module

1.2 PRODUCT CONFIGURATION

Various module mounting configurations can be implemented without special fixtures. Where applicable, these options are factory configured prior to shipment. Mounting configuration can be selected by the installer or field technician and are fully described in this manual. Sensor must always be oriented downward.

®

Registered trademark of Schneider Electric

Page: 1

Model 5100-XX-IT Toxic Gas Sensor Module

1.3 THEORY OF OPERATION

Electrochemical sensors are fuel cell-like devices consisting of an anode, cathode, and electrolyte. The components of the cell are selected so a subject gas, allowed to diffuse into the cell, will cause a chemical reaction and generate a current. The cells are diffusion limited so the rate the gas enters the cell is solely dependent on the gas concentration. The current generated is proportional to the rate of consumption of the subject gas in the cell.

Sierra Monitor electrochemical sensors provide improved reliability by allowing the gas to diffuse into the sensor through a capillary port, rather than diffusing through membranes. The result is an extremely stable sensor with very low temperature and pressure coefficients and the capability to monitor gas as ppm.

1.4 MODES OF OPERATION

IT gas sensor modules can be installed on Sierra Monitor Sentry Model 5000 controllers Ver. 6.XX MFD after 9/1/95. When it is installed in a Sentry system the IT module must have a unique address which can be established by setting an address between 1 and 8 on the Module Address Switch accessible from the cover plate as illustrated in Figure 4-2. Figure 4-7 in this manual provides the wiring terminations for connections to the Sentry controller.

When the module is operated in conjunction with a Sentry controller, the alarm relay setup (See section

5.3) should be set to “Sentry”, allowing the Sentry controller to manage alarm relay action rather than the

5100-XX-IT Gas Sensor Module.

An RS-485 Modbus RTU serial interface allows direct connection to standard PLCs and DCSs. The

Module Address Switch (section 4.5) allows the user to select up to 15 different Modbus addresses. Also, an additional 254 Modbus addresses are available via menu selection. Figure 4-7 in this manual provides the wiring terminations for Modbus connections. The 5100-XX-IT provides the wiring terminations for

Modbus connections.

The analog 4-20 mA interface allows direct connection to standard controller and distributed system. The module is an active current source.

The standard configuration is set up for a 3-wire non-isolated connection. An optional 4-wire isolated connection is also available and can be enabled by changing JP1 and JP2. (Figure 4.6)

1.4.4 REMOTE SENSOR (APPENDIX D)

The Remote Sensor option can be used to remotely mount the sensor up to 15’ from the transmitter.

1.4.5 HART CONNECTION (APPENDIX F)

A HART interface option is available. Refer to Appendix F for information.

1.4.6 REMOTE ALARM RESET

An input is available for connection of remote alarm reset/acknowledge. Figure 4-8 provides the wiring termination for connecting the remote alarm reset. This only resets local alarms, not Sentry alarms.

1.4.7 OPTIONAL INTEGRAL RELAYS

The optional relays are integral to the gas sensor module and are rated as SPDT, 250VAC, 5 Amps* for the High Alarm and Low Alarm relays and SPDT, 250 VAC, 2 Amp for the Trouble relay.

* HART option relays are all SPDT, 250VAC, 2 Amp

If the gas sensor module is provided with the optional relays, it will include Terminal P4 on the interface board (Figure 4-3 and 4-4). Relay output connections are on P4.

Page: 2

Model 5100-XX-IT Toxic Gas Sensor Module

1.5 MECHANICAL

The sensor module is comprised of the following three primary components:

 ENCLOSURE

Standard on the 5100-XX-IT is an explosion-proof, rain-tight cast aluminum electrical housing (Figure 1-3) with three ¾” FNPT conduit hubs. The 5100-XX-IT-SS has a 316 Stainless Steel enclosure (Figure 1-2).

Both enclosure covers have a viewing window. The design of the enclosure allows 3-way mounting choices as shown in figure 1-3.

 TRANSMITTER ELECTRONICS

Electronic Assembly consisting of one printed circuit board assembly mounted under a cover plate, plugged into one field termination board. Connectors for wiring for power, signal interface and alarm relays are located on the bottom of the termination board .

 SENSOR ASSEMBLY

The sensor assembly includes an explosion proof housing containing the gas sensor and a wiring harness for connection to the transmitter. The sensor assembly threads into one hub of the enclosure. The exposed end of the sensor assembly is threaded to allow connection of a rain-shield or calibration gas.

1.6 INTERCONNECT WIRING

Not supplied with the sensor module, but necessary to the installation and operation is the two conductor wiring shielded which connects the module to its power source and controller. Before this wiring is installed it is important to read and understand the control system installation instructions to determine wiring requirements and alternatives.

1.7 POWER REQUIREMENTS

The modules operate on DC power between 10 VDC and 30 VDC. Regulated DC power must be supplied from a separate source or from an approved Sentry or IT controller.

Page: 3

Model 5100-XX-IT Toxic Gas Sensor Module

Figure 1 - 1: Model 5100-05-IT-AL Toxic Gas Sensor Module Aluminum – Dimensions

(Valid for all except 5100-25/26-IT – See Appendix G)

Figure 1 - 2: Model 5100-05-IT-SS Toxic Gas Sensor Module Stainless Steel – Dimensions

(Valid for all except 5100-25/26-IT – See Appendix G)

Page: 4

Model 5100-XX-IT Toxic Gas Sensor Module

Figure 1 - 3: Model 5100-XX-IT Toxic Gas Sensor Module – Mounting Options

NOTE

Module installation must be such that it is mounted vertical with sensor pointing downward.

Page: 5

Model 5100-XX-IT Toxic Gas Sensor Module

2. CAUTIONS & WARNINGS

2.1 INTRODUCTION

Although the IT Transmitter Modules are designed and constructed for installation and operation in industrial applications including “hostile” environments, caution should be taken to insure that the installation is made in compliance with this instruction manual and that certain procedures and conditions are avoided. This chapter discusses the necessary cautions. Read the entire chapter prior to installation of the equipment.

2.2 GAS SENSOR MODULES - GENERAL

Avoid installing sensor modules where they will be unnecessarily exposed to wind, dust, water (i.e. direct hose down), shock, or vibration. Observe temperature range limitations.

Sensors may be adversely affected by prolonged exposure to certain materials. Loss of sensitivity, or corrosion, may be gradual if such materials are present in low concentrations. These materials include:

Halides (compounds containing chlorine, fluorine, bromine, iodine), acid vapors, caustic liquids or mists.

Care has been taken by the manufacturer to ship modules in protective packaging to avoid contamination prior to installation. It is recommended that the modules remain protected during installation and that the covering be removed immediately prior to system start-up.

During normal use the O

2

, CO, H

2

S, and SO

2

sensors are protected from dirt and oil contamination by a sintered metal cover. If this cover becomes clogged, the response of the sensor will be reduced. Protect the sensor from contamination by careful placement, or by use of rain shields and dust shields.

Sensor modules must not be painted. Paint may contain compounds which will contaminate the sensor. Paint will cause clogging of the sintered metal cover and will cause difficulties during attachment of the calibration head or other maintenance activity. It is recommended that the module be tagged

"DO NOT PAINT"

.

2.3 WIRING

The manufacturer recommends that extra caution be taken where the installation is near any sources of electromagnetic or radio frequency interference. Precautions include:

• Avoid running sensor module cable close to high power cables, radio transmission lines, or cables subject to pulses of high current.

• Avoid running cables near large electric motors or generators.

• When the sensor module is to be operated in analog (4-20mA output) mode shielded cable is required.

• When shielding is used, it is recommended that shields be grounded at the controller and nowhere else.

• All splices must be via either a termination hardware system or soldered. Improperly spliced cable can result in corrosion, resistance changes and system errors. The use of wire nuts and crimp-on connectors is unacceptable.

NOTE

Installation and wiring must be in accordance with the National Electrical Code.

Voltage AC conductors are not to be run in the same conduit as voltage DC conductors.

2.4 CALIBRATION FREQUENCY

The 5100-XX-IT calibration frequency is six months (180 days). However, prudent gas detector maintenance practices normally suggest a simple recalibration following incidences of exposure to high levels of gas that would lead to alarm activation within any system utilizing the output of the gas sensor module.

Page: 6

Model 5100-XX-IT Toxic Gas Sensor Module

3. QUICK START

3.1 OVERVIEW

The gas sensor module has been supplied factory calibrated and ready for immediate installation and operation. An installer familiar with installation and operation of gas detection products can use this section to begin immediate use of the module.

3.2 WIRING

See section 4.2 to determine if 3-wire or 4-wire operation is necessary.

Provide two conductor twisted shielded wiring from the power supply/control device to the sensor module location. Use wire that is 18 AWG or larger.

3.3 MODULE INSTALLATION

The sensor can be mounted in a variety of configurations supported by the conduit. See figure 1-3 to determine which configuration is best for your specific application. The default configuration enables the modules to be put in line with other modules with the sensor element below the transmitter. To change the configuration simply remove the transmitter and rotate to the appropriate configuration and remount the standoffs and transmitter.

The module is designed to be installed on a ¾” conduit. Two important warnings:

The installation must meet any hazardous environment codes for electrical equipment.

The sensor module enclosure mounting must be spaced far enough from any vertical surface to allow removal and replacement of the sensor assembly which is threaded into one ¾” conduit entry.

Sensor housing must be oriented vertically pointing downward.

If module is installed outdoors it is recommended that it be sheltered from direct sunlight.

3.4 TRANSMITTER INSTALLATION

To install the transmitter printed circuit assembly into the housing, carefully turn the faceplate so that the printing is in the correct horizontal position for the mounting configuration and slide the assembly over the two stand-offs in the enclosure.

Hand tighten the two captive panel thumb screws into the stand-offs. Replace the enclosure cover prior to providing power to the transmitter

If the transmitter is installed in a classified hazardous area, replace the threaded cover prior to providing power.

3.5 START-UP & OPERATION

To begin operation of the sensor module activate the instrument loop with 14-30 VDC. Each time the sensor module is powered up it will perform a warm-up for approximately 1.0 minutes (60 seconds). During this time the display will read “Starting”. The loop output will be held at 4 mA.

NOTE:

If the sensor is uncalibrated, the startup display will state “UNCALIB” instead of “STARTING”

NOTE:

The 5100-06-IT, 5100-08-IT, 5100-25-IT and 5100-26-IT require 300 seconds warm-up period

After the warm-up period has expired, the display will indicate the gas concentration. Also, the instrument loop will be released to output current in the range of 4 to 20 mA. The actual current is linear with the gas concentration and depends on the selected range. For instance, when the range of 0 to 50 PPM has been selected, a current of 4 mA corresponds to 0 PPM and a current of 20 mA corresponds to 50 PPM.

3.6 ZERO STABILIZATION

All electrochemical sensors require at least 30 minutes on power prior to calibration. This allows the electrode potentials to equilibrate, resulting in a stable zero signal level.

Page: 7

Model 5100-XX-IT Toxic Gas Sensor Module

4. INSTALLATION

4.1 SENSOR MODULE LOCATIONS

NOTE

All IT modules are factory pre-configured and calibrated.

All modules are tagged to indicate the configuration including the sensor module number

Identify all components during unpacking and install using the factory configuration.

The gas sensor module utilizes a diffusion type sensor which should be located close to either the expected source or destination of the gas hazard. If the gas is heavier than air, the sensor module should be installed within 24 inches of the ground or floor. If it is lighter than air, move it above 6’.

N/A Air 1.00

5100-04-IT CO 0.97

After optimum locations are determined based on the above recommendations, consideration should be given to placing the sensor modules in locations which are accessible for calibration service. Slight adjustments to the location of the sensor module may have little impact on effectiveness but major effect on accessibility.

5100-05-IT H

5100-06-IT Cl

2

2

S 1.19

2.49

5100-08-IT ClO

5100-10-IT SO

2

2

3.09

2.26

• Modules should be placed close to the potential source of gas.

• Modules should be placed in areas accessible for calibration.

5100-12-IT NO

5100-25-IT NH

2

3

2.12

0.60

5100-26-IT HF 1.86

• Sensors should be pointed down and the conduit should include an inverse trap to reduce moisture

(condensation) from accumulating in the electronics enclosure.

Table 4 - 1: Specific Graities

• Remote calibration fitting (5360-00) should be used to facilitate calibration gas delivery. Run polyurethane tubing (1/4” O.D. x 1/8” I.D.) from fitting to an accessible location.

Figure 4 - 1: Typical Mounting Options

Page: 8

Model 5100-XX-IT Toxic Gas Sensor Module

4.2 WIRING

4.2.1 ANALOG 4-20 MA OPERATION (FIGURE 4-6)

The 4-20 mA output for the 5100-XX-IT can be either 3-wire (ISA Type 3 Circuit) or standard 4-wire (ISA

Type 4 Circuit) operation. If using 3-wire operation, use a minimum of 18 AWG, shielded, 3-conductor cable up to 2000’. For 4-wire operation, use a minimum of 2 each, 18 AWG, twisted, shielded, pair up to

2000’.

4.2.2 MODBUS OPERATION USING RS-485 CONNECTION

Use a minimum of 18 AWG, 2 conductor for DC power connection. No shield required. In addition use a minimum of 24 AWG, low capacitance, shielded data cable for RS485 half duplex communication. The installation may be planned in a manner which provides up to 32 sensor modules on a single home run.

TERMINATION RESISTOR JUMPERS:

Termination resistors are used in RS-485 wire runs to provide impedance matching. The IT series modules use a 120 Ohm resistor for this function. The cable being used for this RS-485 connection must have a minimum of 100 Ohm impedance with a maximum of 120 Ohms.

Installations where the cable length is under 100’, termination resistors may not be required. In installations where the cable length is greater than 100’, it is recommended to place the termination jumpers on the first device and last device on the RS-485 wire run. Termination jumpers must be removed from all other modules connected between the first and last device. The first device in the RS-485 multiplexed bus is usually a gas controller or PLC. Factory term resistor setting is “not enabled.”

BIAS JUMPERS: (BIAS A, BIAS B)

Bias resistors are used to force RS-485 receiver outputs to a known (fail-safe) state, when the bus is idle.

Bias jumpers are always installed in pairs as the bias must be placed on both the TX A and TX B lines.

Sierra Monitor’s IT series of toxic gas sensors automatically apply the bias jumpers, and are factory installed so that the bias is always enabled.

4.2.3 SENTRY OPERATION USING SMC SENTRY CONNECTION (FIGURE 4-7)

Use a minimum of 18 AWG, 3-conductor cable up to 2000’. The cable may or may not be shielded. We recommend shielded cable in circumstances that there could be RF or EM interference present. Shield to be terminated and grounded only at the Sentry controller. Shield must be cut and dressed at the module end so that no part of it comes in contact with the conduit or ground.

NOTE

Be sure to follow all local electric code and safety requirements when installing the 5100-XX-IT Gas Sensor Module

4.2.4 GENERAL

Install conduit as required by local code or construction specifications. Provide for splice boxes where multiple modules will be wired to a single run. Pull conductors of the correct gauge wire from the controller to each splice box and from the respective splice box to each planned module location. See for proper wire termination in the splice box. Twisted wire secured with wire nuts is not an acceptable splice.

NOTES

The drain wire of shielded cable must NOT be used as one of the conductors.

Installation and wiring must be in accordance with the National Electrical Code. Temperature rating of cable wire must be at least 75 o

C. If cable runs through higher temperature environments, it must be specified for that environment.

Page: 9

Model 5100-XX-IT Toxic Gas Sensor Module

4.3 ENCLOSURE INSTALLATION

To protect the transmitter and sensor assembly they should be removed from the enclosure and preserved until final installation and wiring termination.

Number of modules

1

2

3

4

Maximum length of wire run (feet)

500 1,000 2,000 3,000 5,000

18 18 16 16 14

18 18 14 12 xx

18 16 12 xx xx

16 14 12 xx xx

Table 4 - 2: Minimum Wire Gauges

Prior to installation and wiring:

1. Remove the transmitter from the module housing by:

• Unscrew the two captive panel screws on the faceplate.

• Lift the transmitter out of the enclosure.

• Unplug the sensor cable from transmitter connector P2.

• Remove the sensor assembly from the enclosure hub.

2. Install the module enclosure onto the end of the supply conduit and/or bolt into position as required.

NOTE

When enclosure earth grounding is required for the installation a grounding lug is located in the base of the enclosure. Install the earth ground under the green ground screw.

4.4 TRANSMITTER AND SENSOR INSTALLATION

When all pre-wire is complete:

1. Install sensor assembly in the open hub on the module enclosure. The sensor assembly thread must be fully seated into the hub and tightened to maintain explosion proof assembly.

2. Connect the wires which return to the controller to interface board connectors P1, P2, P3 and P4 according to Figure 4-3 and Table 4-4.

3. Connect the sensor assembly cable to top transmitter board connector P2.

4. Align the headers between the top transmitter board and the lower interface board and push together.

5. Turn rotary switch to correct sensor address if required.

6. Carefully return the transmitter to the enclosure installing it over the two stand-off’s. Tighten the retaining screws into the stand-offs.

7. Cycle power to accept module address change.

P1

1

2

3

PCB Label

Switch

4-20

IN +

IN -

IN +

4 IN -

5 GND

6

7

4-20

OUT +

OUT -

Function

Digital Input SW +

Digital Input SW -

4-20 mA Input +

4-20 mA Input -

Ground

4-20 mA Output +

4-20 mA Output -

Page: 10

Model 5100-XX-IT Toxic Gas Sensor Module

4

5

6

P2

1

2

3

PCB Label

RS 485

RS 485

+

-

S

+

-

S

Function

RS 485 (+) (A)

RS 485 (-) (B)

RS 485 shield (Isolated GND)

RS 485 (+) (A)

RS 485 (-) (B)

RS 485 shield (Isolated GND)

P3B PCB Label

1 P

2 S

3 G

P3A

4 P

5 S

6 G

Function

Sentry Signal or Communication

Sentry Signal or Communication

P4 Connections are only available when the optional Relays are included

P4

1

2

3

4

5

6

PCB Label

LO ALM

N/C

COM

N/O

N/C

Function

Low Alarm Relay NC

Low Alarm Relay COM

Low Alarm Relay NO

High Alarm Relay NC

High Alarm Relay COM

High Alarm Relay NO

HI ALM

COM

N/O

7

8

9

TRBL

N/C

COM

N/O

Trouble Alarm Relay NC *

Trouble Alarm Relay COM*

Trouble Alarm Relay NO*

* Trouble relay is fail safe so it is energized for normal operation, functions are labeled for normal operation.

8. Establish the module address according to section 4.5.

NOTES

The starting delay period normally takes approximately 3 minutes but under some circumstances can take longer.

Page: 11

Model 5100-XX-IT Toxic Gas Sensor Module

4.5 MODULE ADDRESS SWITCH

For digital interface applications the module address switch (or Modbus node) Figure 4-2 must be set per

Table 4-3:

P OSITION A DDRESS P OSITION A DDRESS

4

5

6

7

2

3

Sensor 2

Sensor 3

Sensor 4

Sensor 5

Sensor 6

Sensor 7

A

B

C

D

E

F

Sensor 10

Sensor 11

Sensor 12

Sensor 13

Sensor 14

Sensor 15

Software

Menu

Table 4 - 3: Sensor Module Address

Switch Positions

NOTES

For Sentry applications only sensor addresses 1-8 are allowed. If using Modbus output sensor addresses 1-15 are available. Position 0 allows the Modbus Address to be set by software

menu, in the range 16-254.

Figure 4 - 2: Transmitter Face Plate

Page: 12

Model 5100-XX-IT Toxic Gas Sensor Module

Figure 4 - 3: Interface Board Connectors

Figure 4 - 4: RS-485: Termination, BIAS Jumper

Page: 13

Model 5100-XX-IT Toxic Gas Sensor Module

TYPE 3 CIRCUIT (3 WIRE NON-ISOLATED)

TYPE 3 CIRCUIT (3 WIRE NON-ISOLATED)

Figure 4 - 5: mA Circuit Types

Page: 14

Model 5100-XX-IT Toxic Gas Sensor Module

Figure 4 - 6: 4-20 mA Circuit Type Connections for 5100-XX-IT

Page: 15

Model 5100-XX-IT Toxic Gas Sensor Module

5100-XX-IT TO SENTRY

(Dry contact only unpowered)

5100-XX-IT TO MODBUS DEVICE

Figure 4 - 7: Wiring Connections for Modbus and Sentry Interface

Page: 16

Model 5100-XX-IT Toxic Gas Sensor Module

5100-XX-IT Remote Alarm Reset (Unsupervised)

(Dry contact only unpowered)

P1

+

+

+

4.7K OHM (CUSTOMER SUPPLIED)

SWITCH IN

+

SWITCH IN

-

Figure 4 - 8: Wiring Connections for Remote Alarm Reset

Page: 17

Model 5100-XX-IT Toxic Gas Sensor Module

5. OPERATION

5.1 INTRODUCTION

The Gas Sensor Module utilizes a visual menu system operated by means of a magnet. A magnetic tool

(5358-50) is supplied for this purpose. The menu system is used to configure alarm set-points, calibrate the sensor module, and for maintenance procedures and alarms acknowledge.

5.2 DATA ENTRY KEYPAD

The module menu system is operated by means of directing the magnet stick toward each of four independent hall-effect magnetic switches. Each switch functions as if it is a manually activated panel key. The keys are located under the faceplate above and below the LED display and are labeled M , E , S and T as shown in Figure 5-1.

• Key : MODE

• Key ENTER

• Key : UP (+)

• Key : DOWN (-)

Mode

Enter

Up

Down

Figure 5 - 1: Face Plate with Operator Key Pad

Page: 18

Model 5100-XX-IT Toxic Gas Sensor Module

5.3 MAIN MENU

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

Key

M E

S T

M E

S T

M E

S T

M E

S T

Function Display

Mode

Enter

Up

Down

Switch [M]

Switch [E]

Switch [

Switch [

Description

] Previous Menu

] Next Menu

5100-XX First screen at power up: Model No.

M E

S T

VXX-XX-Second screen at power up: Version No.

STARTING Third screen at start up: Starting Delay

XXX PPM Normal condition - default display

Mode ALMRSET: Mode Function - Alarm Reset

Banner: "Press [E] to reset alarm"

M E

S T

RESET Alarm Reset

XXX PPM Default Display

Mode ALMRSET: Mode Function - Alarm Reset

Mode

Mode

Mode

Mode

Enter

CALIB:--

SETUP:--

MAINT:--

Mode Function - Calibrate

Mode Function - Set up alarms, relays, etc

Mode Function - Maintenance

EXIT-?-Exit Menu

XXX PPM Apply Selected Mode (Exit)

Reference

Table 6-1

Table 5-3

Table 5-4

XXX PPM Default Display

(Once a minute the sensor displays module address)

Table 5 - 1: Master Menu

Table 5-2 defines the key operational displays on the operator interface.

D ISPLAY

STARTING

XXX PPM

LXXX PPM

HXXX PPM

CXXX PPM

D ESCRIPTION

Delay from loss of power at start-up

Concentration

Low Alarm (Warning)

High Alarm (Alarm)

Measures gas, concentration exceeds

100% of Full Scale

Calibration Mode

Acknowledged Function

Table 5 - 2: Operation Display Values

If display shows “Start” instead of “Starting”, the module must be calibrated before using it.

Page: 19

Model 5100-XX-IT Toxic Gas Sensor Module

5.4 SET-UP

The sensor module set-points menu is used to initially set-up the alarm set points, relay actions, gas type and range, 4-20 mA action or RS-485/Sentry address and baud rates, Digital Input and Warm-up time. Use the [ ▲ ] or [ ▼ ] keys to select following menus and press [E] to select.

Alarms: Use the [ ▲ ] or [ ▼ ] keys to select Low Alarm (Warning) or High Alarm (Alarm) menu. Key

[ ▲ ] will adjust the setpoint upwards and Key [ ▼ ] will adjust the value downwards. Once it reaches the desired setpoint, Key [E] will accept it and ACK will appear.

Set-points can be configured using this menu up to the following values:

Factory Alarm Set Points

5100-03-IT 19.5% 16.5%

5100-04-IT 50 PPM 100 PPM

5100-05-IT

5100-06-IT

10 PPM

2.0 PPM

20 PPM

5.0 PPM

5100-08-IT

5100-10-IT

5100-12-IT

5100-25-IT

0.3 PPM

10 PPM

2 PPM

25 PPM

1.0 PPM

20 PPM

4 PPM

50 PPM

NOTE: To enable rising O2 Alarm set warning setpoint higher than 20.9%

Relays: Use the [ ▲ ] or [ ▼ ] keys to select High Alarm (Alarm) or Low Alarm (Warning) relay menu and press [E]. Use the [ ▲ ] or [ ▼ ] keys to select the correct alarm relay action for the application,

Latch, Sentry or Non-Latch. Selecting “Sentry” enables the Sentry controller to make all alarm action decisions. * indicates the current selection.

Range: Use the [ ▲ ] or [ ▼ ] keys to select Range menu and press [E]. When “Range” is selected menu provides any choices of ranges available for the gas type selected. Use the [ ▲ ] or [ ▼ ] keys to select the desired range. If the “User” range is selected, use the [ ▲ ] or [ ▼ ] keys to adjust the high end of the range desired.

Factory Range Set-up

5100-03-IT 5-25% 0-25%

5100-04-IT

5100-05-IT

5100-06-IT

Variable

Variable

None

5100-10-IT

5100-12-IT

Variable

None

4-20mA: Use the [ ▲ ] or [ ▼ ] keys to select Calib or CalibOut menu and press [E]. The “Calib” section of the menu allows the user to calibrate the 4 mA and 20 mA outputs. To calibrate the 4 mA and 20 mA outputs it is necessary to have an amp meter connected to the 5100-XX-IT and upon selecting the 4 mA output calibration then the [ ▲ ] or [ ▼ ] keys can be used to adjust the 4 mA reading on the amp meter until it reads 4 mA. Similar steps can then be performed for the 20 mA output. The CalibOut section allows the user to select the 4-20 mA output action desired during calibration. * indicates the current selected value. Available selections include:

Page: 20

Model 5100-XX-IT Toxic Gas Sensor Module

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

ƒ Track – the 4-20mA value tracks the calibration gas exposed to the gas sensor module

ƒ C2.50mA

– the 4-20mA value is held at 2.50mA during calibration

ƒ C4.00mA

– the 4-20mA value is held at 4.0mA during calibration.

RS-485 - Use the [ ▲ ] or [ ▼ ] keys to select Address or Baud rate menu and press [E]. Note that the

5100-XX-IT has a rotary switch on the faceplate and it is used to select addresses 1-15. When connected to Sentry the user can select 1-8 and when using Modbus RS-485 the user can select addresses 1-15. For Modbus addresses above 15, set the rotary switch to 0 and then use the

“Address” menu to select any address between 16 and 254. The Baud rate menu allows the user to select a baud rate of 38400, 19200, 9600, 4800 or 2400. * indicates current selection. (Sentry default is 9600)

DigInput – Use the [ ▲ ] or [ ▼ ] keys to select supervised or Non-supervised digital input and then press [E] to change.

Key Function Display

--0%LELDefault Display

Description

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

Mode

Mode

Mode

Enter

Down

Down

Down

Down

ALMRSET:

CALIB:--

SETUP:---

Alarms

Relays

Relays

4-20mA

Dig Input

Mode Function - Alarm Reset

Mode Function - Calibrate

Mode Function - Set Point Adjustments

S.P. Function - Alarm Adjust

S.P. Function - Relays Adjust

S.P. Function - Range Adjust

S.P. Function - 4-20 mA Adjust

S.P. Function - RS-485/Sentry Output Adjust

High Alarm Set Point Adjustment Example

M E

S T

M E

S T

Enter

Enter

H.Alarm

HASP:60-

S.P. Function - High Alarm Adjust

High Alarm Set Point: current = 60

Use S or T keys to adjust to new set point

M E

S T

M E

S T

Down (x5) HASP:55High Alarm Set Point: new = 55

Enter ACK Momentary Acknowledge of new Set Point

H.Alarm

S.P. Function - High Alarm Adjust

Reference

* A Below

* B Below

* C Below

* D Below

* E Below

*A

Enter

Down

Enter

Down

Down

Relays Set Point Adjustment Example

H. Relay S.P. Function - High Alarm Relay Adjust

L.Relay

Latch

Sentry

*Sentry

S.P. Function - Low Alarm Relay Adjust

Use S or T keys to adjust to new relay action

(Latch, Sentry, NonLatc) * indicates current

NOTE: Sentry indicates that Sentry controls relay action and not the IT Sensor Module

Low Alarm Relay set to Sentry

*B

Table 5 - 3: Set-Up Configuration

Page: 21

Model 5100-XX-IT Toxic Gas Sensor Module

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

Enter

Enter

Down

Down

Enter

Range

*100 PPM

10 PPM

USER

100 PPM

Gas Range Adjustment Example

S.P. Function - Range Adjust

Select [E] to select or S or T to select another and press [E]

Press [E] if selecting 0-10 PPM range

Press [E] if selecting User adjustable range

Select [E] to select or S or T to select another and press [E]

4-20 mA Adjustment Example

*C

Enter Calib S.P. Function - Calib Adjust

Enter Out: 4mA Use S or T keys to select 4 mA or 20mA

Enter 4mA Selects 4 mA

Enter

Mode

ACK

Calib

Momentary Acknowledge of new Set Point

S.P. Function - Calib Adjust

Down

Enter

CalibOut

Track

S.P. Function - Output during Calibration Adjust

Use S or T keys to select Track, C2.50mA, C4.00mA

*D

Track = Output during calibration tracks the calibration gas,

C2.50mA = Output during calibration is 2.50 mA, C4.00mA =

Output during calibration is 4.0 mA

Enter

Enter

Enter

Enter

Enter

Down

Enter

*Track

Address

* = Current selection

RS-485 Adjustment Example

S.P. Function - RS-485 Address Adjust

Addr:016 Use S or T keys to enter new address

ACK

Address

Baud

New address selected

S.P. Function - RS-485 Address Adjust

S.P. Function - RS-485 Baud Rate Adjust

*38400 Press [E] to select or [ S ] or [ T ] to select another

*E

Table 5-3: Set-Up Configuration (Continued)

Page: 22

Model 5100-XX-IT Toxic Gas Sensor Module

5.5 MAINTENANCE SUB-MENU

The maintenance menu enables the operator to view sensor and software versions. Sierra Monitor technical support has access to other values as needed. The maintenance menu operation is described in Table 5-4.

Reference Key Function Display

XXXPPM Default Display

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

Mode

Mode

Mode

Mode

Enter

Enter

ALMRSET:

CALIB:--

SETUP:-

MAINT:--

Ver1.00aA

CCC001

Description

Mode Function - Alarm Reset

Mode Function - Calibrate

Mode Function - Set Point Adjustments

Mode Function - Maintenance

Module Software Version Number

Module Custom Configuration Control number

Table 5 - 4: Maintenance Menu

Page: 23

Model 5100-XX-IT Toxic Gas Sensor Module

6. CALIBRATION

6.1 CALIBRATION FREQUENCY

The Gas Sensor Module has been calibrated in the factory prior to shipment. It is recommended that the user calibrate before placing into service. The sensor module must be calibrated every 180 days at a minimum.

Periodic functional tests are advisable for critical applications and hostile environments. Oxygen sensors will require more frequent adjustment.

The sensor module microprocessor software includes high-level self-checking algorithms which provide continuous sensor diagnostic and self-adjustment. Users may select to increase calibration frequencies based on low-drift experience during first two calibration periods.

6.2 CALIBRATION PREPARATION

Calibration of the sensor is accomplished by simple menu based steps and application of span gas.

N OTE

If an error is made during any stage of the calibration process, hold the magnet stick at the [M] for 10 seconds. A scrolling display will indicate “calibration aborted” and the sensor module will exit the calibration activity and return to normal operating mode. The calibration procedure can then be restarted.

Calibration must be performed only when the area is known to be clear of the gas of interest. When in doubt, use a portable instrument to confirm that there is no background.

For compliance with Factory Mutual (FM) Approvals, the Sierra Monitor Model 1250-01, 1260 -05 and 1260-45 are the FM Approved calibration gas delivery devices. Use the Model 5358-01 Calibration Adapter delivery fitting.

NOTE: Refer to Appendix H for information on use of permeation tube for 5100-

25-IT and 5100-26-IT calibration.

6.3 CALIBRATION GAS DELIVERY METHODS

Calibration gas can be delivered to the sensors via the Model 5358-01:

Calibration Adapter (Figure 6-1) - used with portable calibrators.

CALIBRATION PROCEDURE

The Calibration Menu is described on Table 6-1:

The procedure requires that the menu “keys” be activated using the magnet stick.

Each key press steps through the process of setting the zero value for clean air and then setting the span value.

At each of these steps, apply calibration gas of the value corresponding to the setting accepted on the sensor module display.

6.4 SENSOR EXPOSURE TO GAS

Figure 6 - 1: Model

5358-01 Calibration

Adapter

Calibration gas must be delivered to the sensor using the flow rate and duration listed in below:

Period

5100-03-IT Air 300 cc of Zero Air or exposure to Ambient Air (3 minutes)

5100-04-IT Carbon Monoxide 300 cc/min Until Stable (minimum 3 minutes)

5100-05-IT Hydrogen Sulfide 300 cc/min

5100-06-IT Chlorine

5100-08-IT* Chlorine

5100-10-IT Sulfur Dioxide

300 cc/min

300 cc/min

300 cc/min

Until Stable (minimum 3 minutes)

Until Stable (minimum 3 minutes)

Until Stable (minimum 3 minutes)

Until Stable (minimum 3 minutes)

5100-12-IT Nitrogen Dioxide 300 cc/min

5100-25-IT Ammonia

Until Stable (minimum 3 minutes)

See Appendix H

5100-26-IT Hydrogen Fluoride See Appendix H

Page: 24

Model 5100-XX-IT Toxic Gas Sensor Module

* Note: Use 10PPM Cl

2

to calibrate

6.5 CALIBRATION SUB-MENU

Key

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

M E

S T

Function Display

XXXPPM Default Display

Mode ALMRSET: Mode Function - Alarm Reset

Mode

Enter

Enter

Enter

Enter

CALIB:--

CAL-0PPM--

ACK

10PPM-SP

Mode Function - Calibrate

Banner: Apply zero gas, enter <E> when done (Not Applicable for 5100-03-IT)

Operation: Confirm area clear of gas, or apply zero air to sensor.

Zero gas setting acknowledged

Banner: Select span, enter <E> to calibrate sensor

Banner: (5100-03-IT) Specify Cal gas <E> when done

C 4 PPM Banner: Apply span gas, then enter <E> to calibrate gas sensor

C 4 PPM

Operation: Apply calbration gas.

CXXXPPM

Operation: As gas is applied the reading will increase - wait 3 minutes

CAL-OK Calibration Passed - now remove gas

W AIT-300

Operation: Five minute time out before sensor is returned to service.

Sub-Routines

Enter 25PPM-SP

Operation: Ready for user adjustment

Enter C 25PPM

Banner: Apply 25PPM gas, then enter <E> to calibrate gas sensor

C 25PPM

Operation: Apply calbration gas.

Enter

Mode

CAL-FAIL

W AIT-300

(Any)

Operation: No calibration gas applied, or sensor did not respond correctly.

Operation: Five minute time out before sensor is returned to service.

Operation: Hold magnet over Mode Switch for ten senconds to abort calibration

Banner: Calibration Aborted

XXXPPM

Description

10PPM-SP Banner: Select span, enter <E> when done

Down 25PPM-SP

Operation: Change Span Gas Value to 25PPM

Down Adj-SPAN

Operation: User adjustable value

Default Display

Table 6 - 1: Calibration

Reference

*

Sub A

Sub B

Sub A

Sub B

Sub C

* 5100-03-IT –

Oxygen Calibration: If area is known to be safe with adequate air flow, you may use the surrounding air as the calibration gas.

If Oxygen levels are unknown, it is recommended to use a Zero Air gas standard.

When calibrating an Oxygen module, there is no zero gas process in the calibration procedure. Only span adjustments can be made, 90 day calibration interval recommended. Normal ambient Air = 20.9% Oxygen.

Begin calibration procedure using the magnet on the Mode key until CALIB is displayed.

Enter to begin procedure. Select cal gas value (20.9), Enter to confirm.

Apply span gas for 3 minutes or use surrounding air if acceptable.

Select Enter when done, and procedure is complete.

NOTE

The Analog output values to be transmitted during the calibration procedure are selected in the maintenance configuration. Normal 20.9% value provides 17.3 mA.

Page: 25

Model 5100-XX-IT Toxic Gas Sensor Module

7. SERVICE

7.1 SENSOR MODULE CONFIGURATION

The gas sensor module is comprised of the following sub-assemblies (Figure 7-1):

5100-XX-IT Gas Sensor Module (See Appendix B)

SPL21810

SPL21823

Aluminum Enclosure

316SS Enclosure

XXXXXXX

5200-XX-IT

Sensor Assembly (See Appendix B)

Sensor (See Appendix B)

7.2 ENCLOSURE REPLACEMENT

The enclosure should be replaced if the cover threads or conduit threads have been damaged, or if the enclosure has been damaged sufficiently that it no longer meets the required NEMA classification.

To replace the enclosure follow the transmitter and sensor assembly removal instructions, remove the damaged enclosure from its conduit, install a new enclosure and continue the transmitter and sensor assembly replacement instructions.

Figure 7 - 1: Module Components

Page: 26

Model 5100-XX-IT Toxic Gas Sensor Module

7.3 TRANSMITTER REPLACEMENT

The transmitter assembly should be replaced when it is determined that it is unreliable, noisy or cannot be adjusted for calibration. This may occur due to age, corrosion or failed components.

To replace the transmitter assembly: a. Remove the cover of the main enclosure. b. Unscrew the two thumb screws in the top of the cover plate, lift the assembly and rotate 90 o wiring service loop.

to relieve the c. Remove all wires from bond, marking them carefully so they can be correctly replaced. d. Unplug the sensor connector from the transmitter. e. Reverse the preceding steps to install the new transmitter. f. Restore power and allow a minimum of 30 minutes for stabilization before re-calibration.

7.4 SENSOR REPLACEMENT

NOTE: Refer to Appendix H for information on sensor electrolyte replacement for 5100-25-IT and 5100-26-IT.

The gas sensor which is located inside the sensor assembly housing can be replaced without replacement of the housing. The gas sensor needs replacement when:

 The “CAL-FAIL” message appears after calibration.

 The sensor output signal is noisy, causing erroneous gas level readings.

 The “Sensor Failure” message displays.

To replace the sensor: a. Confirm that system power has been removed. b. Remove the gas sensor module enclosure cover. c. Unscrew the two thumb screws in the top of the faceplate, lift the transmitter assembly and rotate 90 o relieve the wiring service loop.

to d. Unplug the sensor connector from the transmitter. e. Unscrew the old sensor assembly from the enclosure conduit hub. Remove the sensor assembly with its harness. f. Unscrew sensor housing cover from the sensor. g. Carefully pull the old sensor straight up from the socket. h. Press the new sensor into the socket. i. Reverse the preceding steps to install the sensor assembly. j. Allow the new sensor to stabilize for a minimum of 30 minutes and then calibrate using the procedure in

Section 6.

7.5 INSTALLATION INSPECTION

Prior to system start-up or trouble shooting, the entire system should be visually inspected. The following are guidelines for that inspection:

• Controller installed in conformance to instruction manual recommendations.

• AC power is correctly grounded.

• Hot AC and relay connections have safety covers installed.

• All splices are soldered or via terminal block.

• Cabling is away from sources of electrical noise or RFI where possible.

7.5.3 SENSOR MODULE INSTALLATION

• Module installation in conformance with this manual.

• Modules accessible for calibration.

• Wiring terminations clean and correct.

Page: 27

Model 5100-XX-IT Toxic Gas Sensor Module

7.5.4 MOISTURE TRAPS AND RAINSHIELDS

• Conduit seals and drains installed to avoid moisture build up in electronics enclosure. Water accumulation in sensor module enclosures is a major cause of damage and system failures - take precautions to seal electrical conduits and provide moisture traps and drains to avoid water damage.

• Rain-shields installed where applicable.

• Regulated DC voltage to be applied to the sensor module must be between 10 VDC and 30 VDC.

7.6 INSPECTION AND TROUBLESHOOTING GUIDE

The inspection and troubleshooting guide can be used to step through the system start-up and to determine the corrective action if a fault occurs.

4.2.5 IF IR MODULE DOES NOT RESPOND TO GAS

1. Repeat calibration procedure.

2. Remove the gas and wait for the timer to completely count down.

3. Apply calibration gas and verify that the sensor sees calibration gas after calibration.

4. If the sensor still does not see gas, power cycle the unit and repeat calibration.

4.2.6 IF THE MODULE DISPLAYS “STARTING” FOR MORE THAN 1 HOUR

1. Make sure the sensor is placed in an ambient room temperature environment.

2. Power cycle the sensor.

3. Ensure that the sensor is not exposed to gas of interest during warm-up.

4.2.7 IF THE MODULE DOES NOT DISPLAY THE CORRECT %PPM

1. Power cycle the unit.

2. Recalibrate the sensor.

4.2.8 IF THE DISPLAY SHOWS ‘F’ – SENSOR MISSING OR BAD SENSOR

1. Power down the unit.

2. Open the enclosure and unplug the sensor assembly from the transmitter board.

3. Plug the sensor back into the transmitter board carefully and ensure a secure fit.

4. Check all other connections.

5. Power up the unit.

4.2.9 IF THE MODULE SHOWS “***CALIBRATION REQUIRED***”

1. Calibrate the module.

4.2.10 IF THE DISPLAY SHOWS ‘H’ (OR L) THEN THE LOCAL HIGH OR LOW ALARM IS ACTIVE

4.2.11 IF THE DISPLAY SHOWS “M” – NOT CALIBRATED

1. Calibrate the module.

4.2.12 IF THE DISPLAY SHOWS “C” – CALIBRATION MODE

1. Complete calibration and exit to operating mode.

4.2.13 IF THE DISPLAY SHOWS “S” – SENTRY CONNECTION

1. Check connections with Sentry Controller

4.2.14 DIAGNOSTIC LEDS

LED 1 = 1 Hz Heartbeat normal, 2 Hz Panic Error

2 = Computation process

3 = 1 Hz Heartbeat normal

4 = HART or Sentry comms activity

Page: 28

Model 5100-XX-IT Toxic Gas Sensor Module

8. APPENDICES

8.1 APPENDIX A: SPECIFICATIONS

Sensor:

Type: Electrochemical

Standard

Range(1)

Optional

Max Range

Zero Drift Repeatability Linearity Resolution

Respons e

Time(2)

Accuracy

Sensor

Life (3)

5100-03-IT O

2

5-25% Vol 0-25% Vol +/- 0.1% +/- 0.2% +/- 0.2% +/- 0.1%

5100-04-IT CO 0-500 PPM

5100-05-IT H

2

S (5)

0-100 PPM

5100-06-IT Cl

2

100 PPM

0-10.0 PPM 10.0 PPM

5100-08-IT ClO

2

0-3.0 PPM 3.0 PPM

5100-10-IT SO

2

0-100 PPM

5100-12-IT NO

2

0-20 PPM

5100-25-IT NH

3

(4)

0-100 PPM

5100-26-IT HF 0-10 PPM

1200 PPM +/- 1 PPM

100 PPM

20.0 PPM

100 PPM

10 PPM

+/- 0.5 PPM +/- 1 PPM

+/- 0.5 PPM

+/- 0.5 PPM

+/- 0.5 PPM

+/- 0.5 PPM

+/- 1.0 PPM

--

+/- 1 PPM

+/- 0.5 PPM

+/- 0.5 PPM

+/- 1 PPM

+/- 0.5 PPM

+/- 1.0 PPM

--

+/- 1 PPM

+/- 1 PPM

+/- 0.5 PPM

+/- 0.5 PPM

+/- 1 PPM

+/- 0.5 PPM

+/- 1.0 PPM

--

0.5 PPM

0.1 PPM

0.1 PPM

0.1 PPM

0.5 PPM

0.2 PPM

+/- 1 PPM

0.5 PPM

<10 sec. +/- 0.2%

<35 sec. +/- 1 PPM

<35 sec. +/- 1 PPM

2 years

2 years

2 years

<60 sec. +/- 0.5 PPM 2 years

<60 sec. +/- 0.5 PPM 2 years

<20 sec. +/- 0.5 PPM 2 years

<35 sec. +/- 0.5 PPM 2 years

<30 sec. +/- 1 PPM 1 year

<30 sec. +/- 1 PPM 1 year

(1) Optional ranges available

(2) Response time to 90% full signal value for applied concentration

(3) Sensor life typical for use at standard temperature and pressure with occasional exposure to gas of interest

(4) Diffusion via membrane. Requires electrolyte

(5) Max range, zero drift, linearity, resolution, and sensor life, have not been verified by FM Approvals.

Output:

Input:

Display:

Relays (Optional):

Analog Output:

Signal Output:

Remote Alarm Reset:

Fixed and Scrolling LED

5 Amp, High Alarm, Low Alarm, 2 Amp Trouble all SPDT

Analog 4-20 mA (Trouble 1.5mA, Calibration 2.5mA)

Loop Resistance 800 ohm

Sentry bus

Serial RS-485 Modbus RTU

Normally open digital input (supervised or non-supervised)

Electrical Data:

Operating Range:

Power consumption:

Connection type:

Input voltage:

5100-03-IT O

2

5100-04-IT CO

5100-05-IT H

2

S (5)

5100-06-IT Cl

2

5100-08-IT ClO

2

5100-10-IT SO

2

5100-12-IT NO

2

5100-25-IT NH

3

(4)

5100-26-IT HF (4)

2 watts

3 wire

24 VDC nominal: 10-30VDC

Ambient Temp ( o F) Ambient Temp ( o C)

5 to 122 o F

- 4 to 122 o

F

- 4 to 122 o

F

- 4 to 122 o

F

- 4 to 122 o

F

- 4 to 122 o

F

- 4 to 122 o F

+14 to 113 o

F

+14 to 113 o F

-15 to 50 o

-20 to 50 o

-20 to 50 o

-20 to 50 o

-10 to 45 o

-10 to 45 o

C

-20 to 50 o

C

-20 to 50 o

C

-20 to 50 o

C

C

C

C

C

C

Relative Humidity

15 – 90%

15 – 90%

15 – 90%

15 – 90%

15 – 90%

15 – 90%

15 – 90%

20 – 95%

20 – 95%

Page: 29

Model 5100-XX-IT Toxic Gas Sensor Module

Enclosure:

Material - Aluminum:

Material - Stainless Steel:

Dimensions:

Weight:

Housing:

Electrical Approvals:

Factory Mutual (FM):

UL:

ATEX (5100-XX-IT):

Polyester powder-coated, sand-cast, copper-free aluminum

316 Stainless Steel

8.3 x 4.8 x 4.9 inches Al (H x W x D)

(21.0 x 12.2 x 12.4 cm) Al

7.6 x 4.4 x 4.9 inches SS (H x W x D)

(19.3 x 11.2 x 12.4 cm) SS

2.7 lb. (1.3 Kg) (Aluminum enclosure)

5.4 lb. (2.6 Kg) 316 Stainless Steel

NEMA 4X, NEMA 7

Class 1, Div. I, Groups B, C, D

FM Standards 3600, 3615, 3810

UL Standard 508A

II 2 G Ex d IIB+H2 T6 Gb Ta = -20 o

C to +50 o

C

Performance Approvals:

FM Performance Approval: only)

5100-05-IT FM Approval Performance and Safety Approval

(Appendix I)

Remote Sensor Option (Appendix D):

Distance between sensor and transmitter:

5100-04-IT

5100-05-IT

5100-08-IT

5100-10-IT

5100-12-IT

5100-25-IT

Modbus:

Baud: 38400 (Adjustable 2400 to 38400 Baud)

Parity: None

1 bits:

Control:

8

None

Memory map: Appendix E

Warranty:

Limited Warranty:

Specifications subject to change without notice

2 years

Page: 30

Cross Sensitivity Data:

5100-04-IT Carbon Monoxide

Gas

HS

SO

NO

NO

H

2

C

2

2

2

H

4

HCN

HCl

Conc.

15 PPM

5 PPM

35 PPM

5 PPM

100 PPM

100 PPM

10 PPM

5 PPM

Response

~38 PPM

~3 PPM

~10 PPM

~-3 PPM

<60 PPM

<100 PPM

~5 PPM

0 PPM

5100-05-IT Hydrogen Sulfide

Gas

CO

Conc.

400 PPM

SO

NO

2

NO

2

Cl

H

2

2

C

2

NH

H

4

3

20 PPM

50 PPM

10 PPM

10 PPM

400 PPM

400 PPM

20 PPM

Response

<.1 PPM

<10 PPM

<3 PPM

<-30 PPM

<-25 PPM

1 PPM

<0.1 PPM

<0.1 PPM

5100-06-IT Chlorine

Gas

CO

H

2

SO

S

NO

2

NO

2

H

2

C

2

H

4

HCN

HCl

Conc.

300 PPM

15 PPM

5 PPM

35 PPM

5 PPM

100 PPM

100 PPM

10 PPM

5 PPM

5100-08-IT Chlorine Dioxide

Gas

Cl

2

Conc.

3.1 PPM

Response

0 PPM

0 to -3.38 PPM

-0.05 PPM

0 PPM

~5 PPM

0 PPM

0 PPM

0 PPM

0 PPM

Response

10 PPM

Model 5100-XX-IT Toxic Gas Sensor Module

5100-10-IT Sulfur Dioxide

Gas

H

2

S

NO

NO

2

Cl

2

NO

2

H

2

HCN

HCl

C

2

H

4

Conc.

15 PPM

35 PPM

5 PPM

5 PPM

5 PPM

100 PPM

10 PPM

5 PPM

100 PPM

Response

0 PPM

-7 to 0 PPM

-5 PPM

-1.5 to 0 PPM

~5 PPM

0 PPM

<5 PPM

0 PPM

0 PPM

5100-12-IT Nitrogen Dioxide

Gas Conc.

CO

H

2

SO

S

NO

2

Cl

2

H

2

HCN

300 PPM

15 PPM

5 PPM

35 PPM

1 PPM

100 PPM

10 PPM

HCl

C

2

H

4

5 PPM

100 PPM

Response

0 PPM

-1.5 to 0 PPM

-0.05 to 0 PPM

0 PPM

~1 PPM

0 PPM

0 PPM

0 PPM

0 PPM

5100-25-IT Ammonia

Gas

CO

SO

NO

2

Conc.

1000 PPM

15 PPM

100 PPM

Response

0 PPM

>40 PPM

0 PPM

NO

2

H

2

CH

4

60 PPM 0 PPM

PPM

4%

HCN 10 PPM 75 PPM

H

2

S 3 PPM 75 PPM

5100-26-IT Hydrogen Sulfide

Gas

Cl

2

HCl

NH

SiH

3

CO

4

SO

NO

NO

2

H

2

2

H

2

S

Conc.

5 PPM

5 PPM

75 PPM

15 PPM

1000 PPM

5 PPM

100 PPM

15 PPM

30 PPM

Response

8 PPM

9 PPM

0 PPM

0 PPM

0 PPM

9 PPM

0 PPM

2 PPM

0 PPM

Page: 31

Model 5100-XX-IT Toxic Gas Sensor Module

8.2 APPENDIX B: MODEL NUMBERS & PARTS LIST

Sensor Module

Relay/Connection Sensor Remote

Model Gas Enclosure

IT - - XX - XX X

5100 Series - 03 = O

2

- 04 = CO

- 05 = H

- 06 = Cl

2

2

S

- IT Series

- 08 = ClO

2

- 10 = SO

2

- A1 = AL ¾” NPT

- A2 = AL M20 x 1.5 - 02 = Relays Option

- S1 = SS ¾” NPT

- S2 = SS M20 x 1.5

- 01 = No Relays (std)

- 05 = HART Option

- 00 = Std.

- 12 = NO

- 25 = NH

- 26 = HF

2

3

- 0 = Standard - 0 = None

- 1 = Marine - 1 = Remote

- 2 = Conformal Sensor

Options

5311-00

5311-02 Rainshield with calibration port

Calibration Items 1250-01

1250-02

1250-03

1260-00

1260-04

Gas Sensor Calibration Kit, Type A (CO, H

2

Gas Sensor Calibration Kit, Type B (SO

2

Gas Sensor Calibration Kit, Type C (Cl

Gas Cylinder, Air, (Type A), 105 liters

2

)

S, Air)

, NO

2

)

Gas Cylinder, CO 100 PPM, (Type A) 57 liters

1260-05

1260-06

1260-10

1260-12

1260-13

5360-00

1256-01

1256-02

1265-03

Gas Cylinder, H

2

Gas Cylinder, Cl

2

S in N

, 5 PPM, (Type C) 104 liters

Gas Cylinder, SO

2

Gas Cylinder, NO

2

2

, 25 PPM, (Type A) 57 liters

, 50 PPM, (Type B) 74 liters

, 10 PPM, (Type B) 74 liters

Gas Cylinder, CO, 1000 PPM, (Type A), 105 liters

Calibrator Head Standard

Regulator Type A Calibrator

Regulator Type B Calibrator

Regulator Type C Calibrator

5358-01 Calibration Adapter - Direct, Standard

Calibration/Configuration Magnetic Tool, 4501-XX

9210-00

9211-09

SPX27057

Calibrator, Permeation Tube

Permeation Tube, Ammonia, NH

3

Kit, Sensor Recharge, for 5100-25-IT

SPX27061 Kit, Sensor Recharge, for 5100-26-IT

See Appendix D for 5100-25-IT and 5100-26-IT

Spare Parts

5200-03-IT

5200-04-IT

5200-05-IT

5200-06-IT

5200-10-IT

5200-12-IT

Sensor, for 5100-03-IT

Sensor, for 5100-04-IT

Sensor, for 5100-05-IT

Sensor, for 5100-06-IT / 5100-08-IT

Sensor, for 5100-10-IT

Sensor, for 5100-12-IT

SPL21834

SPL21832

SPL21830

SPL21833

SPL21844

SPL21845

SPM21878

Sensor Assembly, Aluminum, for 5100-03-IT - Oxygen

Sensor Assembly, Aluminum, for 5100-04-IT - CO

Sensor Assembly, Aluminum, for 5100-05-IT - H

2

S

Sensor Assembly, Aluminum, for 5100-06-IT / 5100-08-IT - Cl

2

& ClO

2

Sensor Assembly, Aluminum, for 5100-10-IT - SO

2

Sensor Assembly, Aluminum, for 5100-12-IT – NO

2

Sensor Assembly, Plastic, for 5100-26-IT

Sensor Assembly, 316SS, for 5100-03-IT - Oxygen

Sensor Assembly, 316SS, for 5100-04-IT - CO

SPM27082

SPM27083

SPM27074

SPM27084

SPM27087

SPM27088

Sensor Assembly, 316SS, for 5100-05-IT - H

2

S

Sensor Assembly, 316SS, for 5100-06-IT / 5100-08-IT - Cl

2

& ClO

2

Sensor Assembly, 316SS, for 5100-10-IT - SO

Sensor Assembly, 316SS, for 5100-12-IT - NO

SPL21824 Transmitter

2

2

SPL21810

SPL21823

SPL21825

SPL21829

Enclosure, Transmitter, Aluminum

Enclosure Transmitter, 316SS

Interface Board without Relay

Interface Board with Relay

Page: 32

Model 5100-XX-IT Toxic Gas Sensor Module

8.3 APPENDIX C: LIMITED WARRANTY

SIERRA MONITOR CORPORATION warrants its products to be free from defects in workmanship or material under normal use and service for two years after date of shipment. SMC will repair or replace without charge any equipment found to be defective during the warranty period. Final determination of the nature and responsibility for defective or damaged equipment will be made by SMC personnel.

All warranties hereunder are contingent upon proper use in the application for which the product was intended and do not cover products which have been modified or repaired without SMC approval or which have been subjected to accident, improper maintenance, installation or application, or on which original identification marks have been removed or altered. This Limited Warranty also will not apply to interconnecting cables or wires, consumables (i.e. calibration gases, batteries, sensors), nor to any damage resulting from battery leakage.

In all cases SMC’s responsibility and liability under this warranty shall be limited to the cost of the equipment.

The purchaser must obtain shipping instructions for the prepaid return of any item under this warranty provision and compliance with such instruction shall be a condition of this warranty.

Except for the express warranty stated above, SMC disclaims all warranties with regard to the products sold hereunder including all implied warranties of merchantability and fitness and the express warranties stated herein are in lieu of all obligations or liabilities on the part of SMC for damages including, but not limited to, consequential damages arising out of/or in connection with the use or performance of the product.

Page: 33

Model 5100-XX-IT Toxic Gas Sensor Module

8.4 APPENDIX D: REMOTE SENSOR DRAWING Figure 8 - 1: Remote Sensor Drawing

NOTE

Drawing using Stainless Steel enclosure available from Sierra Monitor.

Page: 34

Model 5100-XX-IT Toxic Gas Sensor Module

8.5 APPENDIX E: MODBUS MEMORY MAP

Table 8 - 1: Modbus Memory Map

40032

40033

40034

40035

40036

40037

40038

40039

40024

40025

40026

40027

40028

40029

40030

40031

40040

40041

40042

40043

40044

40045

Address

40001

40002

40011

40012

40013

40014

40015

40016

40017

40018

40003

40004

40005

40006

40007

40008

40009

40010

40019

40020

40021

40022

40023

Concentration

Temperature

Alarm Relay

Warning Relay

Warning Setpoint

Alarm Setpoint

CCC detail

Software Revision U

Software Revision L

Modbus Map Revision

Trouble Bits

Trouble

Alarm Immediate

Warning Immediate

Serial Number U

Serial Number L

Run time U

Run time L

Gas (MAX) value

Model

Range

Units

Register Name

Module Status

Gas Scale

Cal Concentration level

Calibration due, in days

Calibration Count

Display Gas Output

Gas (MIN) value

Alarm Time U

Alarm Time L

Warning Time U

Warning Time L

Trouble Time U

Trouble Time L

Max Gas Time U

Max Gas Time L

Min Gas Time U

Min Gas Time L

Rotary Switch

Restart Count

Alarm Relay Action Setup

Warning Relay Action Setup

4-20mA CalibOut Value

4-20mA TroubleOut Value

R

R

R

R

R

R

R

R

R

R

R

R

R

R/W

R

R

R

R

RW

RW

R

R

R

R

R

R

R

R

R

R

R

R

R

R

R/W

R/W

R/W

R/W

R

R

RW

RW

Read/Write

R

R

Description

Gas concentration multiplied by Gas Scale (e.g. 209 = 20.9%)

Temperature in degrees Celcius scaled by a factor of 10

Boolean indicating the Alarm relay status (0 = No Alarm, 1 - (High) Alarm).

Clearing bit will reset alarm.

Boolean indicating the Warning relay status (0 = No Alarm, 1 - Warning Alarm). Clearing bit will reset alarm.

Warning Alarm set point, used for activating Low Alarm multiplied by Gas Scale

Alarm set point, used for activating High Alarm multiplied by Gas Scale e.g. 0001, 0007, etc e.g. 201 denotes version 2.01

e.g. version aA is denoted by 0x6141 e.g. version 2 is denoted as v2 = 0x7602

0 = no trouble

1 for any trouble, 0 = no trouble

Like 40003, but never latched

Like 40004, but never latched

Serial Number - 24 bits

In seconds

In seconds

Gas concentration multiplied by Gas Scale

Always read as 51028

R

1 = %LEL, 2 = %VOL

A bit is defined for each of the following states:

(no bit set also means some sort of trouble)

Module State - In Calibration (0x50) = calibration failed

Module State - In Calibration (0x30) = calibration passed

Module State - In Calibration (0x10)

Module State - Trouble (0x08)

Module State - Warming (0x04)

Module State - RUN (ALARM) (0x02)

Module State - RUN (0x01)

Value like 1, 10 or 100

Gas concentration value used during calibration, multiplied by Gas Scale

Gets reset to 365 / 180 if cal successful, Less then 0 = calibration due

Counter increments if cal successful (0 = not calibrated)

Gas value seen on display (i.e. useful during calibration)

Minimum gas value multiplied by Gas Scale

In seconds

In seconds

In seconds

In seconds

In seconds

LS Byte = rotary switch; MS Byte = any soft / polling address assigned

Restart count

Alarm Relay Action Setup - Latch/Sentry/Non-latch

Warning Relay Action Setup - Latch/Sentry/Non-latch

U = denotes upper word

L = denotes lower word

Page: 35

8.6 APPENDIX F: HART

Page: 36

Model 5100-XX-IT Toxic Gas Sensor Module

Figure 8 - 2: HART

Model 5100-XX-IT Toxic Gas Sensor Module

Figure 8 - 3: 4-20 mA Circuits Types 5100-XX-IT – Connections - HART

Page: 37

Model 5100-XX-IT Toxic Gas Sensor Module

HART PROTOCOL MENU

HART (Highway Addressable Remote Transducer) Protocol is the global standard for sending and receiving digital information across analog wires between smart devices and control or monitoring system. HART is a bi-directional communication protocol that provides data access between intelligent field instruments and host systems. A host can be any software application from technician's hand-held device or laptop to a plant's process control, asset management, safety or other system using any control platform. The HART protocol used on this gas sensor module has been developed to comply fully with the HART specifications outlined by the HART Communication Foundation (HCF).

If the HART interface board has been provided with the gas sensor module, then HART will operate via the 4-

20 mA interface automatically. While no specific handheld device is necessary for HART operation, the gas sensor module does comply with the HART standard so any handheld or PC-based HCF certified tool will work.

The information below provides the operator instructions on connecting HART and the HART Protocol Menu.

Common   Practice   Commands  

The   following   Common   Practice   commands   are   implemented.

 

Command  

Number  

38  

48

48

48

48

48

48

 

 

 

 

 

 

Byte   Number

N/A

0

1

2

3

4

5

 

 

 

 

 

 

 

  Description  

Reset   Configuration   Changed   Flag  

Returns   Module   Status   

Module   State  ‐  In   Calibration          (0x10),   

                                                            (0x30   =   cal   passed)  

                                                            (0x50   =   cal   failed)  

Module   State  ‐  Trouble                  (0x08)  

Module   State  ‐  Warming                (0x04)  

Module   State  ‐  RUN   (ALARM)        (0x02)  

Module   State  ‐  RUN                         (0x01)  

Returns   Alarm   Relay   Status  

0x01   =   Alarm   Relay   ON  

0x02   =   Warning   Relay   ON  

0x03   =   Both   Relays   ON  

Returns   Trouble   Status   (non ‐ zero   value   indicates   Trouble)  

Returns   maximum   gas   value  

Returns   gas   value   used   during   calibration  

Returns   IT ‐ Series   Model   Number  

0x28   =   5100 ‐ IT ‐ 28  

 

Device   Specific   Commands  

The   Device   Specific   commands   are   used   for   accessing   the   unique   features   of   the   5100 ‐ IT   Series   of   gas   sensor  

  modules.

 

A   response   code   of   0   indicates   SUCCESS.

   This   applies   to   all   commands.

   As   per   HART   specification,   all   command   responses   will   have   a   response   code   and   status   byte,   plus   any   command   specific   data   bytes   as   indicated   in   the  

  following   tables.

 

Page: 38

Model 5100-XX-IT Toxic Gas Sensor Module

 

Device   Specific   Commands   Summary  

 

 

 

Command  

Number  

130  

131  

132  

133  

134  

135  

136  

137  

138  

139  

140  

141  

142  

143  

144  

Key

Set

Set

Set

Set

Set

 

 

 

 

 

  Press

Alarm

Warning

Alarm

Warning

Reset

Abort

Calibration

Apply

Apply

Set  

 

 

 

 

 

 

 

Level

Relay

Alarms

Calibration

ZERO

SPAN

RESERVED  

 

 

 

 

Level

Relay

 

 

Gas

Gas

Transducer  

 

Gas

 

 

Force   Gas   Value  

 

 

 

Action

 

 

Action

Level

Reset   Force   Gas   Value  

Read   MODBUS   RTU  

 

 

 

Register

Serial  

Description

 

Number  

145  

146  

147  

Set   Calibration   Mode   output   current  

Set   Trouble   Mode   output   current  

RESERVED  

148  

149  

RESERVED

Set   Gas  

 

Factor  

 

Command   130:   Key   Press   

This   command   mimics   the   front   plate   key   inputs.

 

Request   Data   Bytes:   

 

Byte  

0  

Format  

Unsigned ‐ 8   1=Enter  

2=Down  

4=Up  

8=Menu  

Description  

Command   Specific   Response   Data   Bytes:  

Byte  

0  

 

Command   131:   Set   Alarm   Level  

Request   Data   Bytes:  

Format   Description  

Unsigned ‐ 8   Returns   same   value   as   passed  

Byte  

0  

Format   Description  

Unsigned ‐ 8   Alarm   level,   range   0   to   60.

  Default   is   60.

 

Command   Specific   Response   Data   Bytes:  

Byte  

0  

Format   Description  

Unsigned ‐ 8   Returns   new   Alarm   Level.

 

Page: 39

Model 5100-XX-IT Toxic Gas Sensor Module

Command   132:   Set   Warning   Level  

Request   Data   Bytes:  

Byte   Format   Description  

0   Unsigned ‐ 8   Warning   level,   range   0   to   60.

  Default   is   20.

 

Command   Specific   Response   Data   Bytes:  

Byte   Format   Description  

0   Unsigned ‐ 8   Returns   new   Warning   Level.

 

 

Command   133:   Set   Alarm   Relay   Action  

Request   Data   Bytes:  

Byte   Format   Description  

0   Unsigned ‐ 8     0=Latching   (default)  

2=Non ‐ Latching  

Command   Specific   Response   Data   Bytes:  

Byte   Format   Description  

0   Unsigned ‐ 8   Returns   new   Alarm   Relay   Action   value  

 

Command   134:   Set   Warning   Relay   Action  

Request   Data   Bytes:  

Byte   Format   Description  

0   Unsigned ‐ 8   0=Latching   

2=Non ‐ Latching   (default)  

Command   Specific   Response   Data   Bytes:  

Byte   Format   Description  

0   Unsigned ‐ 8   Returns   new   Warning   Relay   Action   value  

 

Command   135:   Reset   Alarms  

This   command   will   reset   any   alarm   relays   that   have   been   latched,   providing   the   alarm   condition   is   no   longer   present   on   the   sensor.

 

Request   Data   Bytes:         None  

Command   Specific   Response   Data   Bytes:    None  

 

Command   136:   Abort   Calibration  

This   command   aborts   the   calibration   procedure.

 

Request   Data   Bytes:         None  

Command   Specific   Response   Data   Bytes:    None  

 

Command   137:   Set   Calibration   Gas   Level  

Request   Data   Bytes:  

Byte  

0  

Format   Description  

Unsigned ‐ 8   Calibration   Gas   Level,   range   0   to   100  

Command   Specific   Response   Data   Bytes:  

Byte

0  

 

 

 

 

Command   138:   Apply   ZERO   Gas  

Format   Description  

Unsigned ‐ 8   Returns   new   Calibration   Gas   Level  

Page: 40

Model 5100-XX-IT Toxic Gas Sensor Module

This   command   will   signal   the   module   to   accept   the   present   sensor   output   to   as   the   ZERO   gas   condition.

 

Request   Data   Bytes:         None  

Command   Specific   Response   Data   Bytes:    None  

 

Command   139:   Apply   SPAN   Gas  

This   command   will   signal   the   module   to   accept   the   present   sensor   output   to   as   the   SPAN   gas   condition.

 

Request   Data   Bytes:         None  

Command   Specific   Response   Data   Bytes:    None  

 

Command   141:   Force   Gas   Value  

Request   Data   Bytes:  

Byte  

0  

Format   Description  

Unsigned ‐ 8   Sets   the   Force   Gas   Value,   range   0   to   100  

Command   Specific   Response   Data   Bytes:  

Byte  

0  

Format   Description  

Unsigned ‐ 8   Returns   new   Force   Gas   Value  

 

Command   142:   Reset   Force   Gas   Value  

This   command   will   signal   the   module   to   set   the   Force   Gas   Value   to   0.

 

Request   Data   Bytes:         None  

Command   Specific   Response   Data   Bytes:    None  

 

Command   143:   Read   MODBUS   RTU   Register  

This   command   allows   the   reading   of   sensor   module   data   as   defined   in   the   MODBUS   RTU   register   map.

 

Request   Data   Bytes:  

Byte   Format   Description  

0 ‐ 1   Unsigned ‐ 16   Address   of   Requested   Modbus   Register  

40001   =   lowest   address   

65535   =   highest   address  

 

 

Command   Specific   Response   Data   Bytes:  

Byte   Format   Description  

0 ‐ 1   Unsigned ‐ 16   Returns   contents   of   selected   Modbus   Register  

 

Command   144:   Set   Transducer   Serial   Number  

This   command   allows   user   to   add   a   device   specific   serial   number,   if   required.

   This   is   different   from   the   serial   number   assigned   by   the   Manufacturer.

 

Request   Data   Bytes:  

Byte   Format   Description  

0 ‐ 2   Unsigned ‐ 24   Serial   Number  

Command   Specific   Response   Data   Bytes:  

Byte   Format   Description  

0 ‐ 2   Unsigned ‐ 16   Returns   new   Serial   Number  

 

 

 

Command   145:   Set   Calibration   Mode   output   current  

Page: 41

Model 5100-XX-IT Toxic Gas Sensor Module

 

 

 

 

 

 

 

 

 

 

 

Request   Data   Bytes:  

Byte   Format   Description  

0 ‐ 1   Unsigned ‐ 16   Current   output   required   during   calibration,   in   micro ‐ amperes  

Command   Specific   Response   Data   Bytes:  

Byte   Format   Description  

0 ‐ 1   Unsigned ‐ 16   Returns   new   Calibration   current   value  

 

Command   146:   Set   Trouble   Mode   output   current  

Request   Data   Bytes:  

Byte  

Byte  

Format  

0 ‐ 1   Unsigned ‐ 16   Current   output   required   to   indicate   Trouble,   in   micro ‐ amperes  

Command   Specific   Response   Data   Bytes:  

Format  

Description  

Description  

Unsigned ‐ 16   Returns   new   Trouble   current   value   0 ‐ 1  

 

Command   149:   Set   Gas   Factor  

Request   Data   Bytes:  

Byte   Format   Description  

0   Unsigned ‐ 8   Sets   the   Gas   Factor   value,   range   30 ‐ 200  

Command   Specific   Response   Data   Bytes:  

Byte  

0  

Format   Description  

Unsigned ‐ 8   Returns   new   Gas   Factor   value  

Version   1.00

  Created   13   July   2011

Page: 42

Model 5100-XX-IT Toxic Gas Sensor Module

8.7 APPENDIX G: MENU KEY FOR 5100-XX-IT GAS SENSOR MODULE

M : MODE

• Key : ENTER

• Key UP (+)

• Key DOWN (-)

MAIN MENU

Use the M key to access each of the 4 main sections of the menu.

M M M M M

| | | | |

|-

ALMRSET

: |- : |- : |-

EXIT?

ALARM RESET MENU

ALMRSET:

|

E – RESET (Resets current alarm)

CALIB MENU

CALIB

:

|

E – CAL-0PPM (Calibrate Zero)

|

E

– Ack (Acknowledges Zero)

|

*25 SPN (Last span gas used – initially set at 25 PPM)

| |

| ▲ 5 SPN (Select Span gas value from below)

| ▲ 10 SPN

| ▲ 25 SPN

| ▲ 50 SPN

| ▲ 75 SPN

| ▲ 100 SPN

| ▲ ADJ SPN (To select an alternate span gas value)

| |

| E 25-SPAN

| |

| ▲ 26-SPAN (Change span value to desired, once completed

E

starts calibration)

|

E

C XXXPPP (Reading the value of the calibration gas being applied)

|

|- CAL-OK (if successful) or CAL-FAIL (if unsuccessful)

Page: 43

Model 5100-XX-IT Toxic Gas Sensor Module

SETUP MENU

Once in the SETUP menu u se the ▲ or ▼ keys to access each of the 5 main sections of the SETUP menu.

▼ ▼ ▼ ▼ ▼

| | | | |

|- Alarms : |- : |- RS-485

SETUP:

|

E

– Alarms (Alarms Set-Point Adjustment)

| |

|

E

– H.Alarm (High Alarm Setup)

| |

| E --------------------------- OR--------------------------------------▼

| | |

| |- HASP: 20 (Current Hi Alarm set-point

| |

|- L. ALARM (Low Alarm Setup)

|

| ▲ or ▼ (To adjust to desired set-point)

| |

| E – Ack (Accepts set-point)

|

|

|

E

LASP:10 (Current Lo Alarm Set-point)

|

▲ or ▼ (To adjust to desired set-point)

|

E – Ack (Accepts set-point)

|

|----

▼ Relays (Relay Set Point Adjustment)

| |

|

E

----------------------------- OR----------------------------------------▼ :

| | |

| | - H. Relay (High Relay Setup)

| |

|- L. Relay (Low Relay Setup)

|

| E – Latch

| |

|

▲ or ▼ (To adjust to desired condition)

E

– Latch

|

or ▼ (To adjust to desired condition)

|

(Selections are Latch, NonLatc, or Senty)

| (* indicates current status, Sentry = Sentry controller controls alarm relay action)

|

|----

▼ Range (Gas Range Adjustment)

| |

| E *100 PPM (* = current selection)

| |

|

▲ or ▼ (To adjust to desired range, choices are 10 PPM or USER)

| If USER is selected use the ▲ or ▼ keys to change to desired value

|

Page: 44

Model 5100-XX-IT Toxic Gas Sensor Module

|

|----

▼ 4-20mA (Verifies 4mA or 20mA output and selects condition of output during calibration)

| ▼ -------or---------------

E

Mode:

| |

| |

| |

| |

| ▼ -------or---------------

E

Calib

| |

| |

| |

| |

| |

| |

| |

| |

| |

| |

| |

| |

| |

|

|-INPUT ( ▲ : or ▼ (To adjust select Input,

Output or Locked, * = selection)

|

|- Out: 4mA

|

|

--or—

E

(To calibrate analog output 4mA, |

then t o accept)

▼ --or-E (To calibrate analog output 20mA, | ▲ or ▼ to adjust level,

|

|- 4mA

then to accept)

▲ or ▼ to adjust level,

▼ --or—

E

(To calibrate analog input 4mA, | ▲ or ▼ to adjust level,

| then to accept)

|- 20mA

▼ --or-E (To calibrate analog input 20mA, | ▲ or ▼ to adjust level,

| | then to accept)

| |

| ▼ -------or---------------

E

CalibOut (Adjust 4-20mA output during Calibration)

| | |

| |

| |

| |

| |

| |

|- Track: ( ▲ or ▼ (To adjust action/mA value during calibration

Track = mA value tracks Gas value

Hold = holds mA value existing prior to calibration

C1.50mA = 1.50 mA during calibration

| |

| |

C4.00 mA = 4.00 mA during calibration

User mA = | ▲ or ▼ to adjust to desired mA value

| ▼ -------or---------------

E

TblOut (Adjust 4-20mA output during Trouble condition)

| |

| |

| |

|

|- T2.50mA: ( ▲ or ▼ (To select 2.50mA or User mA, * = current value)

User mA = | ▲ or ▼ to adjust to desired mA value

| |

| |

|----

▼ RS-485 (Select RS-485 address and baud rate)

| |

| | - Address

| |

| ▼ | -------or---------------

E

Addr:016 (Use ( ▲ or ▼ and

E

to select module address – 016 to 254)

| |

| |- Baud -------------- E *38400 (Use | ▲ or ▼ and E to select baud rate, * = current selection)

MAINT MENU

MAINT:

|

E – V0.90dF (Software version installed on module)

|

E

– CCC001 (Configuration number)

Page: 45

Model 5100-XX-IT Toxic Gas Sensor Module

8.8 APPENDIX H: 5100-25-IT & 5100-26-IT GAS SENSOR MODULE

1. Calibration of 5100-26-IT (Follow Instructions in Section 6)

2. Calibration of 5100-25-IT (Follow Instructions in Section 6)

3. Equipment Required

The following tools and equipment will be required for calibration:

Permeation Device Calibrator (Model 9210-00), with

Ammonia Permeation Tube (Model 9211-09).

Calibration Adapter (Model 5358-01)

4. Permeation Device Calibrator Operation

The calibrator is a portable device with a pump that operates from a

1.5 V alkaline size D battery. A fresh battery should allow 8 hours of continuous operation. The permeation device will probably be shipped separately, usually in a pipe with screw cap ends, one of which will be marked for opening. The permeation device should come fitted with a scrubber, but, as a precaution, open the pipe and remove the permeation device in a well ventilated area.

1. Remove the scrubber from the permeation device and attach the device to the stainless steel “Tee” assembly inside the calibrator box. Tighten the connection with a wrench, with the device resting inside one of the openings in the rack.

2. Close the door of the calibrator and turn on the pump switch.

3. Set the flow rate for the desired ppm of ammonia. The desired flow rate for calibration of the Model 5100-25-IT is 300-500 cc/min. and the concentration range should be set between the high alarm point and 100 ppm full scale.

The carrier gas flow rate can be calculated from the formula:

Figure 8 - 4: Module

Components

P = (CxF)/K

Where P = the permeation rate in nanograms/min., and C = ppm of span gas, F = carrier gas flow rate in ml./min. and K = 1.437 for ammonia. The permeation rate for the temperature read on the front of the calibrator can be obtained from permeation rate vs. temperature chart.

4. Run the gas calibrator for 30 min. before attaching the output hose to the ammonia sensor.

5. When the use of the calibrator is completed, do not turn off the pump unit until the permeation device has been removed from the “Tee” and the gas scrubber reattached. Failure to remove the permeation device when the pump has been turned off will contaminate the system with high amounts of the permeating gas and result in erroneous concentrations. The device, with scrubber, can be stored in the storage rack overnight, but should not stay in the gas calibrator for longer storage periods.

5. Sensor Assy. Rebuild Procedure

Follow procedures outlined in Section 6.0

Electrolyte/Membrane Replacement (See Fig. 8-5)

The electrolyte needs to be replaced at least every six months, or if there is evidence of low sensor output. The membrane and O-Ring should be replaced at the same time that the electrolyte is replaced.

Page: 46

Model 5100-XX-IT Toxic Gas Sensor Module

1. Equipment Required

- Electrolyte Recharge Kit (SPX27057 for Model 5100-25-IT and SPX27061 for Model 5100-26-

IT) consisting of the electrolyte, package of 5 membranes, 5 0-Rings and alcohol wipes to clean the electrode.

- Tweezers to aid in removing and replacing the membrane

2. Preparation

a. Transfer the necessary electrolyte into a plastic wash bottle.

b. Confirm that system power has been removed.

c. Remove the transmitter electronics board from the main housing and unplug the sensor harness from the transmitter electronics.

d. Unscrew the sensor assembly from the bottom of the enclosure.

e. Remove the assembly cover from the sensor assembly.

CAUTION: Only remove the sensor cover from the assembly, DO NOT attempt to unscrew the sensor from the mounting assembly or you could damage the wiring.

3. Electrolyte Replacement (See Fig. 8-4)

a. Invert the sensor assembly so that the membrane is upward.

b. Insert the sensor assembly in a clamp, if available, and unscrew the box nut counterclockwise, and take out the hold-down ring, membrane, O-Ring, and washer. Then pour out the old electrolyte into a beaker for disposal.

CAUTION: Avoid contact of the electrolyte with skin, eyes and clothing. In case of contact wash immediately - rubber gloves and safety glasses are recommended.

c. Clean the electrode with the alcohol wipes included in the electrolyte replacement kit. All oxidation residue must be removed before re-assembly. When dirt is found inside the sensor assembly, wash the dirt off with electrolyte.

d. Hold the sensor assembly with the sensing electrode up. Close the pressure compensation port with a tissue. Pour approximately 10cc of fresh electrolyte from the wash bottle into the sensor, and rinse the inside of the sensor. Empty the electrolyte into the beaker.

e. Place the old washer back over the sensor and place the O-ring on top of the washer. Then transfer one drop of electrolyte from the plastic wash bottle to the exposed electrode surface.

f. Place new membrane into the hold-down ring and carefully place the plastic hold-down ring on top of the sensor body with the ring ears in the slots of the sensor body. Then fasten the sensor cover over the assembly.

CAUTION: Never touch the membrane with your hands. Use rubber gloves or tweezers.

g. Invert the sensor so that the membrane is down. If this is the first time that electrolyte has been replace, there should be a round Mili Seal covering the pressure compensation screw. This Mili-Seal prevented loss of electrolyte during shipment. Remove and discard the round Mili Seal and remove the screw to expose the electrolyte supply port.

h. Transfer electrolyte from the plastic wash bottle into the electrolyte supply port (see Fig. 8-4) until the level is at the maximum level line on the side of the sensor body.

i. Replace the pressure compensation screw into the electrolyte supply port.

j. Replace the sensor housing cover.

k. Install the sensor assembly into the enclosure hub and tighten firmly.

Page: 47

Model 5100-XX-IT Toxic Gas Sensor Module

l. Reconnect the sensor harness to the transmitter, carefully install the transmitter into the housing and restore system power.

m. Recalibrate the transmitter following the instructions in Section 6.

Spare Parts

Figure 8 - 5: Sensor Assembly – Exploded View

SPX27057 Recharge Kit NH3

SPX27061 Recharge Kit HF

SPX57009 Membrane Kit (both)

Page: 48

Model 5100-XX-IT Toxic Gas Sensor Module

Figure 8 - 6: Model 5100-25-IT-AL and Model 5100-26-IT-AL

Figure 8 - 7: Model 5100-25-IT-SS and Model 5100-26-IT-SS

Page: 49

Model 5100-XX-IT Toxic Gas Sensor Module

8.9 APPENDIX I: 5100-05-IT FM PERFORMANCE APPROVAL

Page: 50

Page: 51

Model 5100-XX-IT Toxic Gas Sensor Module

8.10 APPENDIX J: 5100-XX-IT SIL-2 CERTIFICATES

Page: 52

Model 5100-XX-IT Toxic Gas Sensor Module

Page: 53

Model 5100-XX-IT Toxic Gas Sensor Module

Page: 54

Model 5100-XX-IT Toxic Gas Sensor Module

8.11 APPENDIX K: 5100-05-IT ATEX CERTIFICATE

Model 5100-XX-IT Toxic Gas Sensor Module

Figure 8 - 8: ATEX

Page: 55

Page: 56

Model 5100-XX-IT Toxic Gas Sensor Module

Page: 57

Model 5100-XX-IT Toxic Gas Sensor Module

Model 5100-XX-IT Toxic Gas Sensor Module

THIS PAGE INTENTIONALLY LEFT BLANK

Page: 58

Model 5100-XX-IT Toxic Gas Sensor Module

4-20 MA .................................................. 1, 14, 20, 22

5100-25-IT ........................................................ 24, 47

ALARM ........................................................... 1, 2, 20

AMMONIA ................................................................. 1

ANALOG ................................................................... 2

ANALOG 4-20 ........................................................... 9

ATEX .................................................................. 1, 56

AUXILIARY INTEGRAL RELAY ............................... 1

CALIBRATION .............................................. 8, 24, 33

CALIBRATION FREQUENCY ........................ 1, 6, 24

CALIBRATION GAS DELIVERY ............................ 24

CALIBRATION PREPARATION ............................. 24

CALIBRATION PROCEDURE ................................ 24

DATA ENTRY KEY PAD ........................................ 18

DESCRIPTION ......................................................... 1

DIGINPUT ............................................................... 21

ENCLOSURE .....................................1, 3, 10, 26, 31

FM APPROVAL ............................................ 1, 31, 51

FM PERFORMANCE ............................................. 31

GAS RANGE .......................................................... 22

HART ..................................................1, 2, 37, 39, 43

INSTALLATION ............................................ 7, 10, 27

MAINTENANCE ...................................................... 23

INDEX

MENU .................................................................... 44

MODBUS ............................... 1, 2, 16, 30, 31, 36, 37

MODULE ADDRESS SWITCH .............................. 12

OPERATION ................................................ 7, 18, 30

POWER ................................................................... 3

RANGE .................................................................. 20

RELAY ......................................................... 2, 20, 21

REMOTE ALARM .................................................. 17

REMOTE DISPLAY ................................................. 1

REMOTE SENSOR ........................................... 2, 31

RS-485 ................................................... 9, 21, 22, 30

SENSOR ...................................... 3, 6, 10, 26, 27, 30

SENTRY ............................................ 1, 9, 16, 30, 35

SERVICE ............................................................... 26

SIL-2 .................................................................. 1, 53

SPARE PARTS ................................................ 33, 49

SPECIFICATIONS ................................................. 30

START-UP ............................................................... 7

THEORY OF OPERATION ...................................... 2

TRANSMITTER ................................... 2, 3, 7, 10, 27

WARRANTY .................................................... 31, 34

WIRING ........................................................ 3, 6, 7, 9

Page: 59

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement