Eutech Thermo Scientific Temp 360 Instruction manual

Add to My manuals
44 Pages

advertisement

Eutech Thermo Scientific Temp 360 Instruction manual | Manualzz
Instruction Manual
Temp-360
RTD Data logging Thermometer
Part of Thermo Fisher Scientific
68X526601 Rev 2 06/10
TABLE OF CONTENTS
1.
2.
3.
4.
INTRODUCTION ............................... 1
SAFETY PRECAUTIONS.................. 2
SPECIFICATIONS ............................ 3
BATTERY INSTALLATION AND
REPLACEMENT................................ 6
5. INSERTING AND REMOVING
OPTIONAL RUBBER ARMOUR........ 7
6. ASSEMBLING OPTIONAL
HANDSFREE ACCESORIES............. 8
7. CONNECTING A RTD....................... 9
8. KEY FUNCTIONS ........................... 10
9. DISPLAY OVERVIEW ..................... 11
10. MEASUREMENT MODE ................. 13
11. HOLD FUNCTON ............................ 14
12. MIN, MAX, AND AVE FUNCTION ... 14
13. DATA LOGGING ............................. 14
14. SETUP MODE................................. 15
15. GENERAL SETUP SCREEN .......... 16
16. CALIBRATION SCREEN ................. 18
17. ALARMS SCREEN.......................... 20
18. DATA LOGGING SCREEN ............. 25
19. CALIBRATION REPORT SCREEN . 27
20. CLEAR / RESET SCREEN ............. 28
21. MAINTENANCE............................... 29
22. CLEANING ...................................... 29
23. BATTERIES..................................... 29
24. TROUBLESHOOTING..................... 30
25. ACCESSORIES ............................... 31
26. WARRANTY .................................... 32
27. PRODUCT RETURN ....................... 32
28. INNOCAL® CALIBRATION AND
REPAIR SERVICES ........................ 33
1. INTRODUCTION
This versatile hand-held
instrument provides highly
accurate temperature
measurements. The
instrument is designed for
easy operation and
includes the following
features:
•
•
•
•
•
•
•
•
•
•
•
•
•
Menu driven setup and operation
Data logging for up to 2000 points
USB output
Operator selection of Celsius or
Fahrenheit scale
Resolution of up to 0.01°C/ °F
Large backlit LCD and dot-matrix
graphic LCD display
Hold feature for temporarily retain a
reading
Displays minimum, maximum and
average readings
Field calibration capability
Disabling of Auto-Off function
Low battery warning
Three-pin Din connector input
Operates with a wide selection of
probes
-1-
2. SAFETY PRECAUTIONS
WARNING:
1. This instrument is designed to accept
low level signals supplied by standard
RTDs. Under NO circumstances
should the input voltage exceed the
specified 50V RMS.
2. To prevent ignition of a hazardous
atmosphere, batteries must only be
changed in an area known to be nonhazardous.
CAUTION:
1. Do not use or store this instrument in
microwave ovens or any abnormally
hot or cold areas.
2. Weak batteries should not be left in
the instrument. Dead batteries can
leak and cause damage to unit.
DANGER:
1. Voltages present at the RTD may
also be present at the battery
terminals. Always disconnect the RTD
when changing batteries.
-2-
3. SPECIFICATIONS
RTD Thermometers
Temperature Range:
-201 to 1210°C (-330 to 2210°F)
Out of range display:
---Resolution:
0.1 °C/°F: From -330.0 to -100 °C/°F
0.01 °C/°F: From -99.99 to 99.99 °C/°F
0.1 °C/°F: From 100.0 to 999.9 °C/°F
1 °C/°F: Above 1000 °C/°F
Accuracy:
±0.1 °C/ ±0.2 °F:
From -330.0 to -100 °C/°F
±0.03 °C/±0.06 °F:
From -99.99 to 99.99 °C/°F
±0.1 °C/ ±0.2 °F:
From 100.0 to 999.9 °C/°F
±1 °C/ ±2 °F: Above 1000 °C/°F
Display:
Backlit Dot-matrix 50mm X 37.2mm
Data Logging:
2000 points
Logging Interval:
2 sec to 60 min
Min/Max/Avg Function:
Yes
-3-
Auto Off (Adjustable Time):
Enable/Disable option available
Stability Criteria:
Yes, upon stability of 5 seconds
Display Update Rate:
0.5 sec per update
Input:
One three-pin DIN connector
Input Protection:
50V rms
Storage:
– 40°C to 65°C (– 40°F to 149°F)
Humidity:
10% to 90% (non-condensing)
Battery Life:
Size: Three AA, 1.5V; Alkaline
Life: 400 hours continuous, typical,
(without backlighting and buzzer on)
Dimensions:
Without Armor:
175mm (L) X 97mm (W) X 42mm (H)
With Armor:
180mm (L) X 102mm (W) X 52mm (H)
Weight with Batteries:
Without Armor: 267g
With Armor: 362g
-4-
Ingress Protection:
Meets IEC-529 IP-54 for dust
and water resistant enclosures
(with probe attached)
CE Compliance:
EN61326-1/A1: 1998
(EU EMC Directive)
-5-
4. BATTERY INSTALLATION AND
REPLACEMENT
The total battery life without
backlighting is about 400 hours.
Remaining battery power is indicated
by the battery life indicator.
Indicator
Voltage
Cell + 3 bars
More than 4.1 V
Cell + 2 bars
More than 3.6 V
Cell + 1 bars
More than 3.0 V
Cell + empty bars
More than 2.85 V
flashing
Adaptor power
Main power supply
supply
connected
USB power supply
USB power
Selected settings are stored in
memory and will remain in memory
even after power is turned off, or while
batteries are being replaced.
1.
2.
3.
4.
5.
Before changing battery, turn
instrument off and disconnect RTD.
Loosen screw and lift battery cover
off back of case.
Remove the three AA batteries.
Insert three new batteries observing
polarity.
Install cover and tighten screw.
-6-
5. INSERTING AND REMOVING
OPTIONAL RUBBER ARMOUR
1.
2.
To insert thermometer into the
optional rubber armor, slide in from
the top of meter before pushing the
bottom edges of meter down to set it
into position. Lift up the stand at the
back of meter for bench top
applications if necessary.
To remove thermometer from armor,
push out from the bottom edges of
meter until it is completely out of
boot.
-7-
6. ASSEMBLING OPTIONAL
HANDSFREE ACCESSORIES
You can use the optional magnets and
strap in the Handsfree Kit accessories
for hands free operations.
-8-
7. CONNECTING A RTD
Use the correct 100 Ω RTD (alpha =
0.003850) for your instrument. Using
an incorrect probe type will result in
erroneous readings. Insert the 3-pin
plug into the mating connector on the
top of the instrument.
If no probe is connected the display
will read “open”.
RTDs are sensitive at the tip or
sensing element. When taking
measurements, allow time for the
reading to stabilize. Multiplying the
time constant of the probe by 5 will
give you the approximate time
required.
-9-
8. KEY FUNCTIONS
F1
Step through Min, Max and
Avg readings
F2
Toggle between F and C
display
F3
Toggle between menu and
measure mode
hold
on/off
light
Freeze display
Turns meter on and off (press
and hold for 3 seconds to turn
off)
Press momentarily to turn on
backlight
recall▲
Recalls and steps through
stored readings
log▼
Stores current measured
value to memory
Note: Function keys change in setup
mode to provide advanced operation
flexibility. Text above key will indicate
function.
- 10 -
9. DISPLAY OVERVIEW
The dot matrix display features a large
primary display, smaller secondary
displays for channel info or
min/max/ave, and helpful annuciators
for added measurement data.
- 11 -
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
Power supply indicator
(Battery/Main adaptor/USB)
Date (format of mm – dd)
Time (hour:min)
Time format (am/pm/hrs)
Measurement mode
Hold function indicator
Stable indicator
Data logging indicator
PC data logging indicator
Main reading display
Current reading unit indicator
Min/Max/Avg elapsed time
Current Min/Max/Avg reading
indicator
Meter logging memory location
Measurement alarm active
indicator
Countdown time indicator with
countdown time
Min/Max/Avg display
Min/Max/Avg unit indicator
Function keys
- 12 -
10. MEASUREMENT MODE
On initial start-up the meter will display
the measured value in the primary
display.
Pressing the F2 key will toggle reading
between F and C display.
Pressing the F1 key initiates and
toggles through the Minimum,
Maximum, and Average Reading
modes.
Pressing the F3 key accesses the
Setup mode.
- 13 -
11. HOLD FUNCTION
Press the hold key to retain the reading
on the display. Press the hold key
again for normal operation.
12. MIN, MAX, AND AVE FUNCTION
Press the F1 key to toggle between the
minimum, maximum, and average
readings. The minimum and maximum
reading function is ideal for monitoring
unattended operations while continually
displaying every temperature change
that occurs. The minimum and
maximum values are sensed and
automatically stored.
To exit and clear this function, press the
F3 key to access the Menu functions.
See the CLEAR/RESET SCREEN
section for more details.
13. DATA LOGGING
Press the log ▼ key to store the current
reading to memory. The memory
indicator M = 1234 shows the memory
location for the next stored reading.
Press the recall ▲ key to review stored
readings.
See the DATA LOGGING section for
timed logging, and logging to a
computer. See the CLEAR/RESET
SCREEN section for information on
clearing stored readings.
- 14 -
14. SETUP MODE
To access the setup mode from
measurement mode press the “Menu”
key (F3).
Press the ▲▼ keys on the meter key
pad to scroll through options.
To enter a setup screen press the
“Select” key (F1).
To return to the measurement mode
press the “Meas” key (F3). The
following menu options are listed:
1. General setup
2. User field calibration
3. Alarm settings
4. Data logging settings
5. View user calibration report
6. Clear/Reset options
- 15 -
15. GENERAL SETUP SCREEN
The first page of the General Setup
screens let you set resolution, auto-off
time, and password.
Press the F1 key to indicate that you
want to change the setting of the
current parameter or press the recall
▲ key or log▼ key to move to the
next parameter.
Press the recall ▲ key or log ▼ key
to change the options.
Press the F2 key to choose the next
setting. When the option is set, press
the F1 key to accept the choice.
- 16 -
The screen below is used to
reset/change the password. If the user
forgets the password, 5586 can be
used to reset to a new value.
On the second screen you can set the
time and date.
- 17 -
16. CALIBRATION SCREEN
The thermometer is factory calibrated
and does not require calibration
before use. The Calibration function
allows a single point calibration of the
thermometer, at any temperature
point, to compensate for RTD offset
error. It is NOT necessary to perform
a field calibration to obtain the
specified meter accuracy. Use the
field calibration feature to improve
thermometer/probe accuracy or to
compensate for RTD drift.
Before going into the Calibration mode,
the user must enter the password.
Press the F2 key to change to the next
digit. (Default Password is 9900)
- 18 -
There are two calibration options:
Offset – adjusts at a single point. Offset
calibration can be performed at any
temperature in the offset range of
±10.00 °C or ±18.00 °F.
Slope – adjusts at two points. The two
calibration points must be at least
40.00 °C (72.00 °F) apart. The second
calibration point should be at a higher
value than the first calibration point.
Press the ▲ or ▼ key to adjust the
value to match a known temperature
standard. Press the F1 key to accept.
- 19 -
17. ALARMS SCREEN
There are two kinds of alarm settings
available in the alarm setting options:
MEASUREMENT ALARM
Disable or enable the alarm by pressing
the recall ▲ or log ▼ key and the F1
key to accept. Increase or decrease the
individual limit by pressing the recall ▲
or log ▼ key.
Meter in alarm mode
- 20 -
COUNTDOWN ALARM
You can enable/disable the countdown
alarm and set the countdown time from
5 seconds to 1 hour.
After setting (enabling) the countdown
alarm, the measurement screen will
look like this:
- 21 -
Press the F2 key to start the
countdown:
Press the F2 key in the middle of a
countdown to stop the process:
Restart the countdown by pressing the
“Start” key (F2).
Alternatively, press the “Menu” key
(F3) to go into the timer menu and
disable the alarm.
- 22 -
The buzzer will sound for 30 seconds
or until the “Reset” key (F2) is pressed
at the end of the countdown.
Flashing icon
indicates that
the countdown
has expired
To reset the timer, press the “Reset”
key (F2). The display will look like this:
To repeat functions, press the F2 key.
Note: The °C/ °F function is disabled
when the countdown function is
o
o
enabled. To activate the C/ F
function on the measurement screen,
see the COUNTDOWN ALARM
- 23 -
section and disable the countdown
alarm function.
When activated, the countdown timer
temporarily overrides the auto-off
function until the countdown is
completed or manually stopped. If the
meter is manually or automatically shut
off, the countdown alarm is
automatically set to “Disable” each
time the meter is shut off. The “Timer”
key will revert back to the “°C/°F” key.
- 24 -
18. DATA LOGGING SCREEN
Press the recall ▲ key or log ▼ key
to choose the logging methods as auto
or manual. If auto logging is selected,
use the recall ▲ key or log ▼ key to
set the time interval. Its range is from 2
minutes to 60 minutes.
DATA TRANSFER FROM METER TO
COMPUTER
- 25 -
Once the USB connection is
established with a PC, press the
“Select” key (F1) to download data
from Meter to PC using
HyperTerminal.
***NOTE: METER MUST BE TURNED
ON PRIOR TO CONNECTING USB
CABLE OR COMPUTER MAY NOT
RECOGNIZE THE INSTRUMENT***
- 26 -
19. CALIBRATION REPORT
SCREEN
The calibration report will show the
time and date along with results of the
last user calibration.
- 27 -
20. CLEAR / RESET SCREEN
Press the F1 key to select the data you
want to clear or reset. For calibration,
logged data and reset all, you will have
to enter the password to proceed.
(Default Password is 9900)
- 28 -
21. MAINTENANCE
Properly used, the thermometer
should maintain calibration indefinitely
and not require service other than
occasional cleaning of the housing and
changing of the batteries.
22. CLEANING
WARNING:
TO PREVENT IGNITION OF A
HAZARDOUS ATMOSPHERE BY
ELECTROSTATIC DISCHARGE,
CLEAN WITH DAMP CLOTH.
Do not clean with abrasives or
solvents. Use mild detergents, never
immerse nor use excessive fluid.
23. BATTERIES
If there is no display when the
thermometer is turned on, check the
condition of the three AA batteries.
Also check that the battery terminals
are clean and batteries are properly
installed. If replacement is necessary,
refer to the BATTERY
INSTALLATION AND
REPLACEMENT section for the
replacement procedure.
- 29 -
24. TROUBLESHOOTING
The following chart lists the most
probable faults. There are no internal
adjustments or user-replaceable parts.
FAULT
ACTION
NO
Display
Check condition of batteries.
Check that batteries are
inserted properly.
Display
shows
OVER /
UNDER
Out of range indication
Display
Shows
OPEN
No RTD connected in the
connector
Display
Shows
Err
If display shows this
message other than during
the field calibration mode,
please return the instrument
for servicing
Cannot
connect
to PC
Ensure that the meter is
powered on and measuring
prior to connecting to USB
port. Before and after the
driver is installed, your PC
may not recognize the Temp360 if the Temp-360 is not
powered on prior to
connecting the USB to the
computer.
- 30 -
25. ACCESSORIES
Replacement Meters and Meter
Accessories
Item
Part
Number
Temp 360 Thermometer
35426-60
Rubber Armour with Stand
35427-80
Handsfree Kit (Two Magnets and
a Strap)
35427-85
General Purpose Probe
(immersion into liquids)
08117-70
Penetration Probe
(meat, semi-soft materials)
08117-85
Surface Probe
(direct contact on hot surfaces)
08117-78
Air/gas Probe
08117-90
- 31 -
26. WARRANTY
The Manufacturer warrants this
product to be free from significant
deviations from published
specifications for a period of three
years. If repair or adjustment is
necessary within the warranty
period, the problem will be corrected
at no charge if it is not due to
misuse or abuse on your part as
determined by the Manufacturer.
Repair costs outside the warranty
period, or those resulting from
product misuse or abuse, may be
invoiced to you.
27. PRODUCT RETURN
To limit charges and delays, contact
the seller or Manufacturer for
authorization and shipping
instructions before returning the
product, either within or outside of
the warranty period. When returning
the product, please state the reason
for the return. For your protection,
pack the instrument carefully and
insure it against possible damage or
loss. The Manufacturer will not be
responsible for damage resulting
from careless or insufficient packing.
- 32 -
28. INNOCAL® CALIBRATION
AND REPAIR SERVICES
Optimum performance of your
temperature-measuring instrument is
not a timeless condition. To ensure
quality measurements, have your
instrument calibrated regularly. Trust
InnoCal® to satisfy your calibration
and equipment repair needs. With
over a decade of service, we've
helped thousands of customers meet
ISO, FDA, EPA, GLPs/cGMPs and
other quality standards.
Conformity*
ISO/IEC 17025:2005 accredited
NIST Handbook 150, 2000 Edition
ANSI/NCSL Z540-2-1997
NIST Technical Note 1297
ISO 9000:2000
Fast Service
Our substantial inventory of
replacement parts ensures a fast
turnaround and prevents costly
downtime. Most instruments serviced
in five business days!
Excellent Value
Get quality at a fair price. Our
InnoCal® NIST-traceable certificates
offer extensive test data on a broad
range of measurement parameters
without breaking the bank!
- 33 -
Reliable Support
Trust in our free diagnostic support
and troubleshooting advice. Our
factory-trained metrologists and
technicians have years of experience
and extensive technical data.
Convenient Reminders
It’s so easy to keep your instruments
functioning properly. Based on your
requirements, InnoCal will send you
a reminder when it’s time to re-certify
or service your instrument.
We provide you with the
documentation you need to meet
your most stringent quality
requirements for the control of
inspection, measuring, and test
equipment.
Certification includes certificate of
calibration with test data, including:
• description and identification of the
item certified
• condition of the item
• issue date
• identification of calibration
procedure
• calibration date
• as found/as left test data (where
applicable)
• signature of technician
• statement of estimated uncertainty
• list of equipment used to perform
calibration (including their
calibration dates)
- 34 -
With today's high quality standards
such as ISO 9000, certification is
becoming increasingly important.
Traceability is not a timeless
condition. It must be verified and
maintained over the life of the
calibration to ensure the highest
accuracy possible. When you have
your calibration done by InnoCal, we
will send you an automatic reminder
when it is time to recalibrate your
instrument.
Are your calibration certificates
good enough?
InnoCal surpasses the competition by
providing the most complete
certificates as required by NIST. All of
our certificates include measured
data and point-by-point measurement
uncertainty, and by request, we’ll
provide test accuracy and test
uncertainty ratios at no extra cost.
Call us today and see why InnoCal is
The Choice of Quality.
*See our Scope of Accreditation for
any limitations.
- 35 -
Calibration
test points
against
NISTtraceable
standards
Meter
only
Probe
only
System
(meter +
probe)*
Four test points
across range of
instrument.
0, 100, 165,
230 C
(-4, 32, 446,
770F)
17000-04
17001-04
17002-04
InnoCal—The Choice of Quality
1-866-InnoCal (1-866-466-6225)
InnoCalSolutions.com
- 36 -
- 37 -
- 38 -
- 39 -
TECHNICAL ASSISTANCE
If you have any questions about the
use of this product, contact the
Manufacturer or authorized seller.
For more information on OAKTON
Instruments Products, please contact
your nearest distributor or visit our web
site listed below:
Oakton Instruments
625 E Bunker Court
Vernon Hills, IL 60061
USA
Tel: (1) 888-462-5866
Fax: (1) 847-247-2984
[email protected]
www.4oakton.com
Distributed by:

advertisement

Related manuals

advertisement