Vertical Gel Electrophoresis Systems


Add to my manuals
23 Pages

advertisement

Vertical Gel Electrophoresis Systems | Manualzz

user

manual

Vertical Gel  

Electrophoresis 

Systems 

Twin Plate Vertical Gel Units

Mini Size 10 × 10 cm Twin Plate Vertical Gel Unit, 

Cat. No. E4100

Mini Size 10 × 10 cm Twin Plate Vertical Gel Unit with Casting Base, 

Cat. No. E4100-Y

Cooled Mini Size 10 × 10 cm Twin Plate Vertical Gel Unit, 

Cat. No. E4101

Cooled Mini Size 10 × 10 cm Twin Plate Vertical Gel Unit with Casting Base, 

Cat. No. E4101-Y

Standard Size 20 × 20 cm Twin Plate Vertical Gel Unit, 

Cat. No. E4200

Standard Size 20 × 20 cm Twin Plate Vertical Gel Unit with Casting Base, 

Cat. No. E4200-Y

Cooled Standard Size 20 × 20 cm Twin Plate Vertical Gel Unit, 

Cat. No. E4201

Cooled Standard Size 20 × 20 cm Twin Plate Vertical Gel Unit with Casting Base, 

Cat. No. E4201-Y

um

DNTV100-IM/Rev. A0/08-11

Contents

Safety Warnings and Precautions 

General Information 

Packing Lists 

System Details 

Using the Denville Vertical Gel Electrophoresis Units 

A.  Setting Up the Vertical Unit 

B.  Care and Maintenance 

C.  Gel Plate Preparation 

D.  Gel Plate Assembly 

E.  Casting Using the Gel-running Module 

F.  Casting 

G.  Gel Pouring 

H.  Sample Loading 

I.  Gel and Buffer Volumes/Run Conditions 

J.  At the End of the Run 

Ordering Information 

Warranty 

Disposal 

13

14

15

16

10

11

8

9

18

19

7

8

6

6

3

5

ii

1

 pi

 pii

Safety Warnings and  

Precautions

Important user information

Please read this entire manual to fully understand the safe  and effective use of this product.

The exclamation mark within an equilateral triangle is  intended to alert the user to the presence of important  operating and maintenance instructions in the literature  accompanying the instrument.

Should you have any comments on this manual, we will be  pleased to receive them at:

Denville Scientific, Inc. 

P.O. Box 4588 

Metuchen, NJ 08840-4588 

Denville Scientific, Inc. reserves the right to make changes in  the specifications without prior notice.

When used correctly, these units pose no health risk.

However, these units can deliver dangerous levels of electricity  and are to be operated only by qualified personnel following  the guidelines laid out in this instruction manual.

The unit must never be used without the safety lid correctly  in position.

The unit should not be used if there is any sign of damage to  the external tank or lid.

These units comply with the statutory CE safety directives: 

73/23/EEC: low voltage directive: IEC 1010-1:1990 plus  amendment 1:1992 

EN 61010-1:1993/BS EN 61010-1:1993

General Information

Mini Size 10 × 10 cm Twin Plate 

Vertical Gel Units 

The Denville Mini Size gel units are a dual-gel unit, with all  the important features required by research scientists. They  are flexible and come either cooled or un-cooled and with or  without a casting base. They can run most 8 × 10 cm or  

10 × 10 cm commercially available pre-cast polyacrylamide  gels, in a wide range of cassette thickness, due to the novel  clamping arrangement. Two screws make gel clamping  simple without using separate clamps. Self-sealing snap-on  connectors prevent the spills, which commonly occur when  disconnecting the cooling liquid.

Operational

Buffer Volume 

Upper   Lower  

  Max  

Approx.   Voltage 

Model  Chamber  Chamber  Gel Vol.  (Volts) 

Max

Current 

(mAmps)

E4100 

E4100-Y   

100 ml 

E4101 

E4101-Y

1000 ml  7 ml 

   

80 –150  25 – 45  1 gel 

50 – 85  2 gels 

 p1

 p2

Standard Size 20 × 20 cm Twin Plate 

Vertical Gel Units 

The Denville Standard Size gel units are dual-gel units, with  all the important features required by research scientists. They  are flexible and come either cooled or un-cooled and with  or without a casting base. The gels are in a large format for  increased sample throughput and longer separation. High  resolution can be maintained with the excellent cooling  features. Self-sealing snap-on connectors prevent the spills,  which commonly occur when disconnecting the cooling liquid.

Operational

Model 

Buffer Volume 

Upper   Lower  

  Max  

Approx.   Voltage 

Chamber  Chamber  Gel Vol.  (Volts) 

E4200 

E4200-Y   

650 ml  3500 ml  35 ml 

   

E4201 

E4201-Y   

Max

Current 

(mAmps)

90  –120 

Stacking gel 

120  –180 

Resolving gel

20 – 30 per gel 

Packing Lists

The packing lists should be referred to as soon as the unit is  received to ensure that all components have been included. The  unit should be checked for damage when received. Be sure to  keep all packaging material for damage claims for to use should  it become necessary to return the unit. Please contact your  supplier if there are any problems or missing items.

E4100 Series Mini Size 10 × 10 cm Twin Plate 

Vertical Gel Units

Mini Size 10 × 10 cm Twin Plate Vertical Gel Unit (E4100)

Package includes: Tank, lid, gel-running module, a black cable,  a red cable, 2× plain glass plates, 2× notched glass plates,  

2× 1 mm thick 12-sample combs, 4× 1 mm spacers,  

2× spacer aligners, and a dummy plate.

Mini Size 10 × 10 cm Twin Plate Vertical Gel Unit with 

Casting Base (E4100-Y)

Package includes: Tank, lid, gel-running module, caster base  with silicone seals, a black cable, a red cable, 2× plain glass  plates, 2× notched glass plates, 2× 1 mm thick 12-sample  combs, 4× 1 mm spacers, 2× spacer aligners, and a   dummy plate.

Cooled Mini Size 10 × 10 cm Twin Plate Vertical  

Gel Unit (E4101)

Package includes: Cooled tank, lid, gel-running module, a black  cable, a red cable, 2× plain glass plates, 2× notched glass  plates, 2× 1 mm thick 12-sample combs, 4× 1 mm spacers,  

2× spacer aligners, and a dummy plate.

Cooled Mini Size 10 × 10 cm Twin Plate Vertical  

Gel Unit with Casting Base (E4101-Y)

Package includes: Cooled tank, lid, gel-running module, caster  base with silicone seals, a black cable, a red cable,  

2× plain glass plates, 2× notched glass plates, 2× 1 mm thick 

12-sample combs, 4× 1 mm spacers, 2× spacer aligners, and a  dummy plate.

 p3

 p4

E4200 Series Standard Size 20 × 20 cm Twin Plate 

Vertical Gel Units

Standard Size 20 × 20 cm Twin Plate Vertical Gel Unit 

(E4200)

Package includes: Tank, lid, gel-running module, a black cable,  a red cable, 2× plain glass plates, 2× notched glass plates,  

2× 1 mm thick 24-sample combs, 4× 1 mm spacers,  

2× spacer aligners, and a dummy plate.

Standard Size 20 × 20 cm Twin Plate Vertical Gel Unit  with Casting Base (E4200-Y)

Package includes: Tank, lid, gel-running module, caster base  with silicone seals, a black cable, a red cable, 2× plain glass  plates, 2× notched glass plates, 2× 1 mm thick 24-sample  combs, 4× 1 mm spacers, 2× spacer aligners, and a   dummy plate.

Cooled Standard Size 20 × 20 cm Twin Plate Vertical  

Gel Unit (E4201)

Package includes: Cooled tank, lid, gel-running module, a  black cable, a red cable, 2× plain glass plates, 2× notched  glass plates, 2x 1 mm thick 24-sample combs, 4× 1 mm  spacers, 2× spacer aligners, and a dummy plate.

Cooled Standard Size 20 × 20 cm Twin Plate Vertical  

Gel Unit with Casting Base (E4201-Y)

Package includes: Cooled tank, lid, gel-running module, caster  base with silicone seals, a black cable, a red cable,  

2× plain glass plates, 2× notched glass plates, 2× 1 mm thick 

24-sample combs, 4× 1 mm spacers, 2× spacer aligners, and  a dummy plate.

System Details

Usage guidance and restrictions

• Maximum altitude 2000 m.

• Temperature range between 4 °C and 60 °C.

•  Maximum relative humidity 80% for temperatures up to 

31 °C decreasing linearly to 50% relative humidity at 40 °C.

• Not for outdoor use.

This apparatus is rated POLLUTION DEGREE 2 in accordance  with IEC 664. POLLUTION DEGREE 2, states that: “Normally  only non-conductive pollution occurs. Occasionally, however,   a temporary conductivity caused by condensation must   be expected”.

 p5

 p6

Using the Denville Vertical Gel  

Electrophoresis Units

A. Setting Up the Vertical Unit 

Instructions for attaching electrode leads to the lid:

1   Note the position of the lid on the unit. This shows the  correct polarity and the correct orientation of the cables,  black is negative and red is positive. 

2   Remove the lid from the unit. Note if the lid is not  removed, fitting the cables may loosen the gold plug and  cause damage to the electrode.

3   Screw the cables into the tapped holes as fully as possible  so that there is no gap between the lid and the leading  edge of the cable fitting.

4  Refit the lid.

B. Care and Maintenance

Units are best cleaned using warm water and a mild  detergent. Water at temperatures above 60 °C can cause 

damage to the unit and components. The units should  not be left in detergents for more than 30 minutes. The tank  should be thoroughly rinsed with warm water and distilled  water to prevent build up of salts but care should be taken not  to damage the electrode. Vigorous cleaning is not necessary or  advised. Air drying is recommended before use.

Caution! The units should never come into contact with the  following cleaning agents, these will cause irreversible and  cumulative damage:

Acetone, Phenol, Chloroform, Carbon tetrachloride, Methanol, 

Ethanol, Isopropyl alcohol, Alkalis.

Caution! DEPC is a suspected carcinogen. Always wear gloves  and safety glasses.

 p7

 p8

C. Gel Plate Preparation

1   Clean the plates, spacers and combs in mild laboratory  detergent. DO NOT use abrasive creams or scourers. If a  particularly clean finish is required (e.g. for silver-stained  gels) glass plates can be soaked in chromic acid overnight,  rinse with water then wipe successively with ethanol,  acetone and ethanol again. NEVER allow organic solvents  or chromic acid come into contact with the plastic  components.

2   The notched glass plate can be siliconized in a fume hood  with dimethyldichlorosilane if required to assist in plate  separation after the run.

3   Handle clean plates with gloved hands (remove any  fingerprints with acetone).

D. Gel Plate Assembly

1   On a clean level bench, position the two side spacers flush  with the edges of the rectangular glass plate and then  overlay the notched plate.

2   The gel plates can be sealed either with tape, or by  clamping greased spacers between the plates with bulldog  clips, or by using a casting base (see Casting Using the 

Gel-running Module, page 9).

3   For tape sealing, hold or clamp the plates firmly and  seal the edges of the gel cassette with gel sealing tape. 

The tape should be applied smoothly with no wrinkles. 

Reinforce the corners by overlapping extra pieces of tape  onto the glass. Grease or fingerprints will prevent a good  seal being formed.

4   If greasing is the preferred method, smear a little silicone  grease or Vaseline over the spacers before assembly and  use the long spacer to seal the bottom of the gel and  clamp with bulldog clips. Note that the side spacers will  seem too long if this sealing method is chosen. The side  spacers should be cut to size — make sure the cut is a  clean right angle.

E. Casting Using the Gel-running Module

1   Loosen the clamping plates on the gel-running module  by untightening the screws. Place the gel-running module  on its side on a level bench surface, and slide the gel  cassettes, comprising the glass plates and spacers, into the  gel-running module until they meet notched overhangs. 

The notched glass plates should face the centre of the  gel-running module (GRM). Ensure the glass plates are 

level with the bottom of the GRM. It is important  that the plates are square in the GRM and on a flat  surface before installing in the caster base (step 4).

2   Put the gel-running module in an upright position on the  bench surface. If using plain glass plates without bonded  spacers, use the spacer aligner to push the spacers so that  they are perfectly aligned with the bottom of the glass  plates (there is no need to do this with plain glass plates  with bonded spacers). Tighten the screws, using your other  hand to hold the plates and spacers in position on the  bench surface.

3   Once the gel cassettes are tightened into the gel-running  module, invert the gel-running module to check that the  plates and spacers are aligned correctly in the gel-running  module. Readjust the plates and spacers if necessary.

4   Place the gel-running module onto the casting base with  the silicone seals sitting in the grooves and the cam-pin  levers pointing into the bench surface.

5   Slot the cam pins into position, turning them in a clockwise  direction so that the gel-running module and gel cassettes  are drawn onto the silicone seals. The cam pins should then  point upwards at 180° to the bench surface.

6   Check for leaks by adding 2 ml distilled water into each  gel cassette. The glass plates should be firmly embedded  within the silicone seals, ensuring that the gel cassettes  remain leak-free. Discard the water and pour the  acrylamide gels.

 p9

F. Casting

1. Place each plain glass plate on  a level bench surface, followed by  the spacers and then the notched  glass plate.

2. Place the gel-running module (GRM)  on its side on the bench and insert the  glass plate cassette. Lightly tighten the  screws before turning over the GRM to  repeat on the other side.

3. If not using plain glass plates with  bonded spacers, get a spacer aligner to  align the spacers flush with the vertical  edges of the plain glass plates before  tightening the screws within the GRM.

4. Run your forefinger along the bottom  edge of the glass plates to ensure that  they are flush with bottom edge of each  spacer — if not, repeat step 3.

5. Lower the assembled GRM onto  the casting base with the cam pins  pointing downwards into the bench  surface.

6. Turn the cam pins through 180° to  secure the GRM onto the casting base 

— as you turn the cam pins you will  experience increased resistance.

7. Pour the acrylamide gel between  the glass plates.

8. Carefully insert the appropriate  comb between the glass plates and  allow the gel to set for at least  

30 minutes.

9. After carefully removing the  combs, release the GRM by turning  the cam pins in opposite direction  on the casting base.

10. Place the GRM into the gel tank,  ensuring that there is sufficient  buffer to cover the bottom of the  glass plates.

11. Fill the inner buffer chamber  within the GRM so that it just covers  the top of the gel — the gel is now  ready to be loaded.

12. Replace the lid and insert the  power cables into the power supply,  making sure that it is first switched  off — the unit is now ready for   electrophoresis.

 p10

G. Gel Pouring

1   For reproducibility and uniform polyacrylamide  crosslinking, we recommend using Electran grade 

materials and degassing gel solutions before use. 

Acrylamide solutions should be stored in a cool, dark  environment, such as a refrigerator, and allowed to reach  room temperature before pouring. Avoid exposure to heat  and sunlight.

2   Polymerization conditions should be adjusted to effect  polymerization within about 5  –10 minutes. Test a small  volume in a vial before pouring the gel. As a rough guide 

100 ml of degassed 6% acrylamide gel will set in about  

5 minutes at room temperature when gently mixed  with 450 µl of freshly prepared 10% (w/v) ammonium  persulphate plus 200 µl TEMED. The setting time increases  to about 10 minutes if the TEMED volume is reduced to 

100 µl, and to approximately 15 minutes with 75 µl. The  amount of catalyst may need to be reduced under warm  conditions. Do not pour under direct sunlight.

3   Gel pouring can be carried out directly in a gel-casting  unit (included with some models and available as optional  accessory) or by clamping a taped gel into the tank unit.

4   Run the acrylamide separating gel mix slowly down the  inside edge of the gel cassette. Avoid aeration. If a  stacking gel is to be used, carefully overlay the separating  gel to a depth of 3 – 5 mm with 1× separating gel buffer  or water-saturated butanol (If no stacking gel is to be  used, insert the required comb into the solution at the top  of the gel cassette before polymerization). 

5   Following polymerization of the separating gel,  pour off the overlay layer (rinse off the butanol with  electrophoresis gel buffer) and pour a stacking gel. 

Insert the comb ensuring bubbles are not trapped. Once  the stacking gel has polymerized the gel may be used  immediately.

 p11

6   If tape has been used, remove the gel tape from the  the seal between the glass and the silicone gasket. Clean  both the silicone gasket, located on the upper buffer 

 into place. Secure the gel plates into position in the 

• Turn all power supply controls off and disconnect the

 damage to the unit!

8   Add the appropriate volume of running buffer to  

Maximum fill lines or above the top of  the rectangular gel plate if using glass  plates smaller than 10 × 10 cm.

unregulated.

• Do not operate with buffer temperature above 45 °C.

All plastic parts are rated for 45 °C continuous duty.

Circulate coolant through the heat exchanger during electrophoresis to minimize heating. Overheating will cause irreparable damage to the unit!

• Only accessories and parts approved or supplied by

Hoefer, Inc, may be used for operating, maintaining, and servicing this product.

 p12

p3

H. Sample Loading

1   Samples should be prepared using the relevant sample  buffer and procedures.

2   If the sample may contain insoluble material, centrifuge at 

12000 × g for 5 minutes. If this stage is omitted samples  may streak during electrophoresis.

3   Carefully remove the sample comb and immediately flush  the wells with electrophoresis buffer from a syringe.

Table 1 provides a guide to the amounts of protein that can be  successfully applied when using a Tris Glycine SDS system to a 

5-mm wide, 1 mm gel slot.

Table 1

Comb 

Single  

Band 

1 mm × 5 mm wide  1– 6 µg 

1.5 mm × 5 mm wide  1– 10 µg 

Multiple 

Bands 

30 – 60 µg 

50 –100 µg 

Sample 

Vol.

< 40 µl

< 60 µl

The volumes of sample wells can be estimated by simply multiplying  the width, thickness and depth of the well.

4   Load the samples using a gel loading pipette tip. If possible  avoid taking liquid from the pellet area at the bottom of  the tube. During sample loading the pipette tip should be 

1–2 mm above the bottom of the well to minimise dilution  of the sample and to keep the sample as a tight layer.

5   Fill unused wells with the equivalent volume of sample  buffer to maintain uniform electrical resistance across   the gel.

6   Replace the safety lid firmly making sure that the electrical  connectors form a good contact.

7   Connect the electrophoresis apparatus to the power pack  and connect the power pack to the mains supply. Turn  all settings to zero before turning on the mains supply. 

Adjust the controls to the desired settings. Follow the  manufacturer’s instructions.

 p13

 p14

I. Gel and Buffer Volumes/Run Conditions

Guidelines for operating conditions are given in General 

Information on page 1, however, conditions vary according  to the number of gels, their composition, length and cross  sectional area. The current required will increase proportionally  to the number of gels or gel thickness providing that the  voltage is not limiting, e.g. 2 gels require twice the current of 1,  but the same voltage. Longer gels require proportionally higher  voltages. By increasing the gel concentration the electrical  resistance is increased and the rate of migration decreases. 

Higher voltages may be applied, but take care not to overheat  the gel. The conductivity of gels using non-dissociating buffer  systems often vary and therefore conditions need to be  determined empirically.

1   The run conditions are to be taken as a guideline only and  apply to SDS Tris-glycine gels. If the plates become hot  reduce the power settings.

2   If a native gel is being used, it may be necessary to   pre-electrophorese the gel for 15– 40 minutes prior to  sample loading.

3   For SDS gels do not pre-electrophorese the gel.

J. At the End of the Run

1   Stop the run, turn the power supply settings to zero, turn  off the power source and disconnect the power leads.

2   Remove the safety lid by gripping the edges and pressing  on the locating lugs with your thumbs.

3   Remove the internal gel running unit by lifting vertically.

4   After unclamping the gel and removing the tape (if used),  separate the plates with a strong, broad blade (not metal,  as this will damage the plates). If you are using notched or  eared plates DO NOT pry them apart at the ears. 

5   After removing the gel for staining, clean the plates  thoroughly and rinse in distilled water. A clean sheet of  foam rubber placed in the bottom of the sink serves as a  useful support and minimizes the risk of glass plate damage.

or carefully decant the buffer away from the electrical 

ORGANIC SOLVENTS. Avoid direct contact of objects with 

• Circulate only water or 50/50 water/ethylene glycol used for cleaning. DO NOT use abrasive creams or scourers. 

• Do not connect the heat exchanger to a water tap the connectors are clean and dry before use or storage.

unregulated.

• Do not operate with buffer temperature above 45 °C.

All plastic parts are rated for 45 °C continuous duty.

Circulate coolant through the heat exchanger during electrophoresis to minimize heating. Overheating will cause irreparable damage to the unit!

• Only accessories and parts approved or supplied by

Hoefer, Inc, may be used for operating, maintaining, and servicing this product.

 p15

p3

Ordering Information

E4100 Series Mini Size 10 × 10 cm Twin Plate Vertical Gel Unit  

Replacement Parts and Accessories

Combs

8 Sample wells, 0.75 mm 

10 Sample wells, 0.75 mm 

12 Sample wells, 0.75 mm 

16 Sample wells, 0.75 mm 

8 Sample wells, 1.0 mm 

10 Sample wells, 1.0 mm 

12 Sample wells, 1.0 mm 

16 Sample wells, 1.0 mm 

8 Sample wells, 1.5 mm 

10 Sample wells, 1.5 mm 

12 Sample wells, 1.5 mm 

16 Sample wells, 1.5 mm 

8 Sample wells, 2.0 mm 

10 Sample wells, 2.0 mm 

12 Sample wells, 2.0 mm 

16 Sample wells, 2.0 mm 

E4122-8MC

E4122-10

E4122-12

E4122-16MC

E4123-8MC

E4123-10

E4123-12

E4123-16MC

E4124-8MC

E4124-10

E4124-12

E4124-16MC

E4125-8MC

E4125-10

E4125-12

E4125-16MC

Spacers

0.75 mm Spacer set, 10 × 100 mm 

1.0 mm Spacer set, 10 × 100 mm 

1.5 mm Spacer set, 10 × 100 mm 

2.0 mm Spacer set, 10 × 100 mm 

Plates

Plain 10 × 10 cm glass with bonded 0.75 mm spacers, pkg/2 

Plain 10 × 10 cm glass with bonded 1 mm spacers, pkg/2 

Plain 10 × 10 cm glass with bonded 1.5 mm spacers, pkg/2 

Plain 10 × 10 cm glass with bonded 2 mm spacers, pkg/2 

Plain glass 10 × 10 cm, pkg/2 

Notched glass 10 × 10cm, pkg/2 

E4161

E4162

E4163

E4164

E4151

E4152

E4153

E4154

E4155-P

E4155-N

 p16

E4200 Series Standard Size 20 × 20 cm Twin Plate Vertical Gel Units 

Replacement Parts and Accessories

Combs

18 Sample wells, 0.75 mm 

24 Sample wells, 0.75 mm 

36 Sample wells, 0.75 mm 

48 Sample wells, 0.75 mm 

18 Sample wells, 1.0 mm 

24 Sample wells, 1.0 mm 

36 Sample wells, 1.0 mm 

48 Sample wells, 1.0 mm 

18 Sample wells, 1.5 mm 

24 Sample wells, 1.5 mm 

36 Sample wells, 1.5 mm 

48 Sample wells, 1.5 mm 

E4222-18MC

E4222-24

E4222-36MC

E4222-48

E4223-18MC

E4223-24

E4223-36MC

E4223-48

E4224-18MC

E4224-24

E4224-36MC

E4224-48

Spacers

0.75 mm Spacer set, 20 × 200 mm 

1.0 mm Spacer set, 20 × 200 mm 

1.5 mm Spacer set, 20 × 200 mm 

Plates

Plain 20 × 20 cm glass with bonded 0.75 mm spacers, pkg/2 

Plain 20 × 20 cm glass with bonded 1 mm spacers, pkg/2 

Plain 20 × 20 cm glass with bonded 1.5 mm spacers, pkg/2 

Plain glass 20 × 20 cm, pkg/2 

Notched glass 20 × 20 cm, pkg/2 

E4261

E4262

E4263

E4251

E4252

E4253

E4255-P

E4255-N

 p17

 p18

Warranty

Denville warrants that the unit you have received has been  thoroughly tested and meets its published specification.

This unit (excluding all accessories) is warranted against  defective material and workmanship for a period of twelve 

(12) months from the date of shipment ex factory.

Denville will repair all defective equipment returned during  the warranty period without charge, provided the equipment  has been used under normal laboratory conditions and in  accordance with the operating limitations and maintenance  procedures outlined in this instruction manual and when not  having been subject to accident, alteration, misuse or abuse. 

No liability is accepted for loss or damage arising from the  incorrect use of this unit. Denville’s liability is limited to the  repair or replacement of the unit or refund of the purchase  price, at Denville’s option. Denville is not liable for any  consequential damages.

Denville reserves the right to alter the specification of its  products without prior notice. This will enable us to implement  developments as soon as they arise.

Denville’s products are for research use only.

A return authorization must be obtained from Denville before  returning any product for warranty repair on a freight-  prepaid basis.

Disposal

 This equipment is marked with the crossed out wheeled  bin symbol to indicate that it must not be disposed of with  unsorted waste.

 Instead it is your responsibility to dispose of your equipment  correctly at lifecycle-end by handing it over to an authorized  facility for separate collection and recycling. 

 It is also your responsibility to decontaminate the  equipment in case of biological, chemical and/or radiological  contamination, so that the persons involved in the disposal  and recycling of the equipment are protected from any  potential hazard.

For further information about where you can drop off  your waste equipment, please contact your local Denville  representative.

By doing so, you will help to conserve natural and  environmental resources and you will ensure that your  equipment is recycled in a manner that protects human health. 

Thank you 

 p19

Denville Scientific, Inc.

P.O. Box 4588

Metuchen, NJ 08840-4588

Toll Free: 1-800-453-0385

Phone: 1-908-757-7577

Fax: 1-908-757-7551

Email: [email protected]

Web: www.densci.com

© 2011 Denville Scientific, Inc.  

All rights reserved.

Printed in the USA.

 p20