UNI-T UT203 Operating Manual
Add to my manuals47 Pages
Uni-Trend UT203 is a 3 3/4 digits digital clamp multimeter with steady operations and fashionable structure. It can measure AC/DC voltage, AC/DC current, frequency, duty cycle, resistance, diodes, and continuity. It features data hold, sleep mode, and relative mode for convenient and accurate measurements. With its auto-ranging capability and CAT II 600V, CAT III 300V safety rating, the UT203 is ideal for various electrical applications.
advertisement
P/N: 41416301
Registered Design No.: ZL02 3 357791.6
UT203/204
Digital Clamp Multimeters
Model UT203/204: OPERATING MANUAL
Table of Contents
Title
Overview
Unpacking Inspection
Safety Information
Rules For Safe Operation
International Electrical Symbols
The Meter Structure
Rotary Switch
Functional Buttons
The Effectiveness of Functional Buttons
Display Symbols
Measurement Operation
A. DC/AC Voltage Measurement
B. Measuring Resistance
C. Testing Diodes
D. Testing for Continuity
E. Frequency Measurement
F. Duty Cycle Measurement
G. DC/AC Current Measurement
1
Page
3
10
11
12
14
6
9
4
5
22
24
26
28
15
18
18
20
30
Title
Sleep Mode
Specifications
A. General Specifications
B. Environmental Requirements
Accuracy Specifications
A. DC Voltage
B. AC Voltage
C. Resistance
D. Diode Test
E. Continuity Test
F. Frequency
G. Duty Cycle
H. DC Current
I. AC Current
Maintenance
A. General Service
B. Replacing the Battery
2
Model UT203/204: OPERATING MANUAL
Page
40
40
42
44
37
37
38
39
44
45
32
33
33
34
35
35
36
Model UT203/204: OPERATING MANUAL
Overview
This Operating Manual covers information on safety and cautions. Please read the relevant information carefully and observe all the Warnings and Notes strictly.
Warning
To avoid electric shock or personal injury, read the “Safety Information” and “Rules for Safe Operation” carefully before using the Meter.
Digital Clamp Multimeter Model UT203/UT204 (hereafter referred to as “the Meter”) are 3 3/4 digits with steady operations, fashionable structure and highly reliable measuring instrument. The Meter uses large scale of integrated circuit with double integrated A/D converter as its core and has full range overload protection.
The Meter can not only measure AC/DC Voltage, AC/DC Current, Frequency, Duty
Cycle, Resistance, Diodes, Continuity but also it has Data Hold, Sleep Mode and
Relative Mode features.
UT204 has an extra True RMS feature.
3
Model UT203/204: OPERATING MANUAL
Unpacking Inspection
Open the package case and take out the Meter. Check the following items carefully to see any missing or damaged part:
Item
3
4
1
2
5
Test Clip
Carrying Bag
Description
English Operating Manual
Test Lead
9V Battery (NEDA1604, 6F22 or 006P)
Qty
1 piece
1 pair
1 pair
1 piece
1 piece
In the event you find any missing or damage, please contact your dealer immediately.
4
Model UT203/204: OPERATING MANUAL
Safety Information
This Meter complies with the standards IEC61010: in pollution degree 2, overvoltage category (CAT. II 600V, CAT. III 300V) and double insulation.
CAT. II: Local level, appliance, PORTABLE EQUIPMENT etc., with smaller transient overvoltages than CAT. III.
CAT. III: Distribution level, fixed installation, with smaller transient overvoltages than
CAT. IV
Use the Meter only as specified in this operating manual, otherwise the protection provided by the Meter may be impaired.
In this manual, a Warning identifies conditions and actions that pose hazards to the user, or may damage the Meter or the equipment under test.
A Note identifies the information that user should pay attention to.
International electrical symbols used on the Meter and in this Operating Manual are explained on page 9.
5
Model UT203/204: OPERATING MANUAL
Rules For Safe Operation
Warning
To avoid possible electric shock or personal injury, and to avoid possible damage to the Meter or to the equipment under test, adhere to the following rules:
l
Before using the Meter inspect the case. Do not use the Meter if it is damaged or the case (or part of the case) is removed. Look for cracks or missing plastic.
l
Pay attention to the insulation around the connectors.
Inspect the test leads for damaged insulation or exposed metal. Check the test
l
leads for continuity. Replace damaged test leads with identical model number or electrical specifications before using the Meter.
Do not apply more than the rated voltage, as marked on the Meter, between the terminals or between any terminal and grounding. If the value to be measured
l
is unknown, use the maximum measurement position and reduce the range step by step until a satisfactory reading is obtained.
When measurement has been completed, disconnect the connection between the test leads and the circuit under test, remove the testing leads away from the input terminals of the Meter and turn the Meter power off.
6
Model UT203/204: OPERATING MANUAL
l
The rotary switch should be placed in the right position and no any changeover
l
of range shall be made during measurement is conducted to prevent damage of the Meter.
Do not carry out the measurement when the Meter’s back case and battery compartment are not closed to avoid electric shock.
l
Do not input higher than 600V between the two Meter’s input terminal to avoid electric shock and damages to the Meter.
l
When the Meter working at an effective voltage over 60V in DC or 30V rms in AC, special care should be taken for there is danger of electric shock.
l l
Use the proper terminals, function, and range for your measurements.
Do not use or store the Meter in an environment of high temperature, humidity, explosive, inflammable and strong magnetic field. The performance of the Meter may deteriorate after dampened.
l l l
When using the test leads, keep your fingers behind the finger guards.
Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity and diode.
Replace the battery as soon as the battery indicator appears. With a low
l
battery, the Meter might produce false readings that can lead to electric shock and personal injury.
When servicing the Meter, use only the same model number or identical electrical specifications replacement parts.
7
Model UT203/204: OPERATING MANUAL
l
When servicing the Meter, use only the same model number or identical electrical
l
specifications replacement parts.
The internal circuit of the Meter shall not be altered at will to avoid damage of
l
the Meter and any accident.
Soft cloth and mild detergent should be used to clean the surface of the Meter
l l
when servicing. No abrasive and solvent should be used to prevent the surface of the Meter from corrosion, damage and accident.
The Meter is suitable for indoor use.
Turn the Meter off when it is not in use and take out the battery when not using
l
for a long time.
Constantly check the battery as it may leak when it has been using for some time, replace the battery as soon as leaking appears. A leaking battery will damage the Meter.
8
Model UT203/204: OPERATING MANUAL
International Electrical Symbols
AC (Alternating Current)
DC (Direct Current)
AC or DC
Grounding
Double Insulated
Warning. Refer to the Operating Manual
Deficiency of Built-In Battery
Continuity Test
Diode
Fuse
Conforms to Standards of European Union
9
Model UT203/204: OPERATING MANUAL
The Meter Structure (see figure 1)
1.
2.
3.
4.
5.
6.
7.
Input Terminals
LCD Display
Functional Buttons
Rotary Switch
Trigger: press the lever to open the transformer jaws. When the pressure on the lever is released, the jaws will close.
Hand Guards: to protect user’s hand from touching the dangerous area.
Transformer Jaw: designed to pick up the AC and
DC current flowing through the conductor. It could transfer current to voltage. The tested conductor must vertically go through the Jaw center.
(figure 1)
10
Model UT203/204: OPERATING MANUAL
Rotary Switch
Below table indicated for information about the rotary switch positions.
Rotary Switch Position
OFF
V
40A
/
Hz / Duty%
& 400A
Function
Power is turned off
AC or DC voltage measurement
Resistance measurement
: Diode test
: Continuity test
Frequency Measurement and Duty
Measurement
AC and DC current measurement range
11
Model UT203/204: OPERATING MANUAL
Functional Buttons
Below table indicated for information about the functional button operations.
Button
HOLD
Operation Performed
REL
is default to auto ranging measurement mode.
to step down the range.
and displays a result.
At Hz/Duty% range:
measurement mode.
12
Model UT203/204: OPERATING MANUAL
Button Operation Performed
SELECT
Press SELECT button to select the alternate functions marked in
turn the Meter on, it will disable the Sleep Mode feature.
13
Model UT203/204: OPERATING MANUAL
The Effectiveness of Functional Buttons
Not every functional buttons can be used on every rotary switch positions. Below table describe which functional buttons can be used on which rotary switch positions
Rotary Switch
Positions
V
/
Hz / Duty%
40A
400A
SELECT
l
N/A
l
N/A
l l
Functional Buttons
REL
l l
N/A
l l l
HOLD
l l l l l l
14
Model UT203/204: OPERATING MANUAL
Display Symbols (see figure 2)
15
3
2
4
14
1
7 8 9
16
(figure 2)
15
5
6
10
11
12
13
Model UT203/204: OPERATING MANUAL
7
8
9
5
6
Number
1
2
3
4
Symbol
AC
DC
%
Meaning
Indicator for AC voltage or current
Indicator for DC voltage
The battery is low.
Warning: To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears.
The Meter is in the auto range mode in which the Meter automatically selects the range with the best resolution.
Test of diode
The continuity buzzer is on
Indicator for Duty.
Data hold is active
Indicator for REL mode
16
Model UT203/204: OPERATING MANUAL
Number
10
Symbol
,k ,M
11
12
13
14
15
16
Hz
A mV, V
TRMS
OL
Meaning
: Ohm. The unit of resistance.
k : Kilohm. 1x103 or 1000 ohms
M : Megohm. 1x106 or 1,000,000 ohms
The unit of Frequency
Amperes (amps). The unit of current.
Volts. The unit of voltage.
mV: Millivolt. 1x10-3 or 0.001 volts
Indicates negative reading
Indicator for TRMS mode
The input value is too large for the selected range
17
Model UT203/204: OPERATING MANUAL
Measurement Operation
A. DC/AC Voltage Measurement (see figure 3)
Warning
To avoid harms to you or damages to the Meter from eletric shock, do not attempt to measure voltages higher than 600V AC/DC, although readings may be obtained.
The DC Voltage ranges are:
400mV, 4V, 40V, 400V and 600V.
The AC Voltage ranges are:
4V, 40V, 400V and 600V.
Red Black
(figure 3)
To measure DC voltage, connect the Meter as follows:
1.
Insert the red test lead into the Hz Duty% terminal and the black test lead into the COM terminal.
18
Model UT203/204: OPERATING MANUAL
2.
manual ranging measurement mode.
3.
Connect the test leads across with the object being measured.
The measured value shows on the display.
Note
l
When DC/AC voltage measurement has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.
19
Model UT203/204: OPERATING MANUAL
B. Measuring Resistance (see figure 4)
Warning
To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring resistance.
The resistance ranges are:
400 , 4k , 40k ,400k ,M and 40M .
Red Black
To measure resistance, connect the Meter as follows:
1.
Insert the red test lead into the Hz Duty%
terminal and the black test lead into the
(figure 4)
COM terminal.
2.
Set the rotary switch to . Resistance measurement is default to auto range mode,
3.
Connect the test leads across with the object being measured.
The measured value shows on the display.
20
Model UT203/204: OPERATING MANUAL
Note
l l
To obtain a more precise reading, you could remove the objects being tested from the circuit when measuring.
When resistance measurement has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.
21
Model UT203/204: OPERATING MANUAL
C. Testing Diodes (see figure 5)
Warning
To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before testing diodes.
Use the diode test to check diodes, transistors, and other semiconductor devices. The diode test sends a current through the semicondutor junction, then measure the voltage drop across the junction. A good silicon junction drops between 0.5V and 0.8V.
Red Black
(figure 5)
To test the diode out of a circuit, connect the Meter as follows:
1.
Insert the red test lead into the Hz Duty% terminal and the black test lead into the COM terminal.
22
Model UT203/204: OPERATING MANUAL
2.
Set the rotary switch to . Diode measurement mode is a default or presss
SELECT to select measurement mode.
3.
For forward voltage drop readings on any semiconductor component, place the red test lead on the component’s anode and place the black test lead on the component’s cathode.
Note
l l
To obtain a more precise reading, you could remove the objects being tested from the circuit when measuring.
When diode testing has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.
23
Model UT203/204: OPERATING MANUAL
D. Testing for Continuity (see figure 6)
Warning
To avoid damages to the Meter or to the devices under test, disconnect circuit power and discharge all the high-voltage capacitors before measuring continuity.
To test for continuity, connect the Meter as follows:
1.
Insert the red test lead into the Hz Duty%
terminal and the black test lead into the
2.
COM terminal.
Set the rotary switch to and press SELECT
Red Black
3.
button to select measurement mode.
The buzzer sounds if the resistance of a circuit
(figure 6)
4.
under test is less than 50
The buzzer may or may not sounds if the resistance of a circuit under test is between 50 to 100
5.
The buzzer does not sound if the resistance of a circuit under test is higher than
100
24
Model UT203/204: OPERATING MANUAL
Note
l
When continuity testing has been completed, disconnect the connection between the testing leads and the circuit under test and remove testing leads from the input terminals.
25
Model UT203/204: OPERATING MANUAL
E. Frequency Measurement (see figure 7)
Warning
To avoid harms to you or damages to the Meter from eletric shock, do not attempt to measure voltages higher than 600V AC/DC, although readings may be obtained.
The resistance ranges are:
10Hz, 100Hz, 1kHz, 10kHz, 100kHz, 1MHz and
10MHz.
Red Black
To measure frequency, connect the Meter as follows:
1.
Insert the red temperature probe into the Hz
Duty% terminal and the black temperature probe into the COM terminal.
(figure 7)
2.
Set the rotary switch to Hz.
3.
Connect the test leads across with the object being measured.
The measured value shows on the display.
26
Model UT203/204: OPERATING MANUAL
Note
l
When frequency measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
27
Model UT203/204: OPERATING MANUAL
F. Duty Cycle Measurement (see figure 8)
Warning
To avoid harms to you or damages to the Meter from eletric shock, do not attempt to measure voltages higher than 600V AC/DC, although readings may be obtained.
The duty cycle range is: 0.1%~99.9%.
Red Black
To measure duty cycle, connect the Meter as follows:
1.
Insert the red temperature probe into the Hz Duty%
terminal and the black temperature probe
into the COM terminal.
(figure 8)
2.
mode.
3.
Connect the test leads across with the object being measured.
The measured value shows on the display.
28
Model UT203/204: OPERATING MANUAL
Note
l
When duty cycle measurement has been completed, disconnect the connection between the testing leads and the circuit under test, and remove the testing leads away from the input terminals of the Meter.
29
Model UT203/204: OPERATING MANUAL
G. DC/AC Current Measurement (see figure 9)
The measuremnet ranges of current are: 40.00
A
To measure current, do the following:
40A
400A
1.
measuremnet mode is a default. Press SELECT to switch between DC and AC measurement mode.
2.
Hold the Meter tight, don’t release. The Hall components are very senstive not only to the magnet but also to heat and machines reaction force. Any shock will cause the changing in reading in the short time.
(figure 9)
3.
Press the lever to open the transformer jaw.
4.
Center the conductor within the transformer jaw, then release the Meter slowly until the trasnformer jaw is completely closed, Make sure the conductor to be tested is placed at the center of the transformer jaw, otherwise it will casue deviation.
The Meter can only measure one conductor at a time, to meausre more than one condutor at a time will cause deviation..
30
Model UT203/204: OPERATING MANUAL
Note
l l and displays a result.
When current measurement has been completed, disconnect the connection between the conductor under test and the jaw, and remove the conductor away from the transformer jaw of the Meter.
31
Model UT203/204: OPERATING MANUAL
Sleep Mode
To preserve battery life, the Meter automatically turns off if you do not turn the rotary switch or press any button for around 15 minutes.
The Meter can be activated by turning the rotary switch or pressing the button based on ‘The Effectiveness of Functional Buttons” on page 14. Press SELECT to activate the Meter will disable the Sleep Mode feature.
The Meter beeps 5 times in about 1 minute before entering Sleep Mode and it will have a 1 long beep just before entering Sleep Mode.
To disable the Sleep Mode function, press and hold SELECT button while turning on the Meter.
32
Model UT203/204: OPERATING MANUAL
Specifications
A. General Specifications:
l l l l l l l l l l l l
Maximum Voltage between any Terminals and grounding: Refer to different range input protection voltage.
Display: 3 3/4 digits LCD display, Maximum display 3999
Polarity: Automatically display.
Overloading: Display OL or –OL
Battery Deficiency: Display
Measurement Speed: Updates 3 times/second.
Measuremnet Deviation: The conductor being meaured is not placed in the center of the jawt during AC/DC current measurement, it will cause extra 1% deviation based on the stated accuracy.
Drop Test: 1 meter drop test passed.
Max. Jaw Size: 28mm diameter.
Projected Max. Current conductor size: 26mm diameter.
Electro-Magnetic: When carrying out measurement near the electro-magnetic, it may cause unstable or wrong reading.
Power: 1 x 9V battery (NEDA1604 or 6F22 or 006P)
33
Model UT203/204: OPERATING MANUAL
l l l l
Battery Life: typically 150hours (alkaline battery)
Sleep Mode (can be disabled)
Dimensions (H x W x L): 208mm x 76mm x 30mm.
Weight: Approximate 260g (battery included)
B. Environmental Requirements
l l l l
The Meter is suitable for indoor use.
Altitude: Operating: 2000m
Storage: 10000m
Safety/ Compliances: IEC 61010 CAT.II 600V, CAT.III 300V over voltage and double insulation standard.
Temperature and humidity:
Operating: 0ºC~30ºC ( 85%R.H);
30ºC~40ºC ( 75%R.H);
40ºC~50ºC ( 45%R.H);
Storage: -20ºC~+60ºC ( 85%R.H)
34
Model UT203/204: OPERATING MANUAL
Accurate Specifications
Accuracy: (a% reading + b digits), guarantee for 1 year.
Operating temperature: 23ºC 5ºC
Relative humidity: 85%R.H
Temperature coefficient: 0.1x(specified accuracy)/1ºC
A. DC Voltage
Range
400.0mV
4.000V
40.00V
400.0V
600V
Resolution
0.1mV
1mV
10mV
100mV
1V
Remarks: Input impedance: 10M
Accuracy
(0.8%+3)
(0.8%+1)
(1%+3)
Overload protection
600V DC/AC
35
Model UT203/204: OPERATING MANUAL
B. AC Voltage
Range
4.000V
40.00V
400.0V
600V
Resolution
1mV
10mV
100mV
1V
Accuracy
(1%+5)
Overload protection
600V DC/AC
(1.2%+5)
Remarks:
l
Input impedance: 10M // less than 100pF l l
Frequency response: 40Hz~400Hz.
Change to AC:
UT203:
Change to AC by using average response method. Input sine wave, then adjust the reading until it is same as the effective value.
UT203:
Combine AC and True RMS response method. Input sine wave to adjust. Non sine wave must follow the below data to adjust:
Peak factor: 1.4~2.0, add 1.0% on the stated accuracy
Peak factor: 2.0~2.5, add 2.5% on the stated accuracy
Peak factor: 2.5~3.0, add 4.0% on the stated accuracy.
36
Model UT203/204: OPERATING MANUAL
C. Resistance
Range
400.0
4.000k
40.00k
400.0k
4.000M
40.00M
Resolution
100m
1
10
100
1k
10k
Accuracy
(1.2%+2)
(1%+2)
(1.2%+2)
(1.5%+2)
Overload protection
600Vp
D. Diode Test
Range Resolution
1mV
Accuracy
Display forward voltage drop nearest value
Overload protection
600Vp
Remark: Open circuit voltage approximate 1.48V.
37
Model UT203/204: OPERATING MANUAL
E. Continuity Test
Range Resolution
100m
Accuracy
Around 50 , the buzzer beeps
Overload protection
600Vp
Remark:
l l
Open circuit voltage approximate 0.45V.
The buzzer may or may not beeps when the resistance of a circuit under test is l between 50 ~100
The buzzer will not beep when the resistance of a circuit under test is > 100 .
38
Model UT203/204: OPERATING MANUAL
F. Frequency
Range
10Hz
100Hz
1kHz
10kHz
100kHz
1MHz
10MHz
Resolution
0.001Hz
0.01Hz
0.1Hz
1Hz
10Hz
100Hz
1kHz
Remark:
Input Sensitivity as follows:
When 100kHz: 300mV rms;
When > 100kHz: 600mV rms
When > 1MHz: 800mV rms
Accuracy
(0.1%+3)
For reference only
Overload protection
600Vp
39
Model UT203/204: OPERATING MANUAL
G. Duty Cycle
Range
0.1%~99.9%
Resolution
0.1%
Accuracy
For reference only
Overload protection
600Vp
H. DC Current
Range
40.00A
400.0A
Resolution
0.01A
0.1A
Accuracy
(2%+5)
(2%+3)
Overload protection
400A DC/AC
Warning
The operating temperature must be 0ºC ~40ºC when measuring current.
Remark:
l
If the reading is positive, the current direction is from bottom to up. See figure 10, the front case face up while the bottom case face down. Hold the Meter tight, do now release. The Hall components are very sensitive not only to the magnet but
40
Model UT203/204: OPERATING MANUAL
l also to heat and machines reaction force. Any shock will cause the changing in reading in the short time. Follow the below procedure to measure current will be more precise:
Hold the Meter tight and press the lever to open the transformer jaw. Center the conductor within the transformer jaws, then release the Meter slowly until the transformer jaw is completely closed. Make sure the conductor to be tested is placed at the center of the transformer jaw, otherwise it will cause +1.0% deviation based on the stated accuracy.
Remove the transformer jaw.
l l l l
Repeat the above 1. procedure.
The obtained reading will be more precise.
41
Model UT203/204: OPERATING MANUAL
I. AC Current
Range Resolution Accuracy
40.00A
0.01A
(2.5%+8)
400.0A
0.1A
(2.5%+5)
Frequency Response
50Hz ~ 60Hz
Overload protection
400A DC/AC
Warning
The operating temperature must be 0ºC ~40ºC when measuring current.
Remark:
l
It may have 10 digits or less unstable or wrong digits, it will not affect measurement l result.
Hold the Meter tight, do now release. The Hall components are very sensitive not only to the magnet but also to heat and machines reaction force. Any shock will cause the changing in reading in the short time. Follow the below procedure to measure current will be more precise:
1.
Hold the Meter tight and press the lever to open the transformer jaw. Center the conductor within the transformer jaws, then release the Meter slowly until the transformer jaw is completely closed. Make sure the conductor to be tested is placed at the center of the transformer jaw, otherwise it will cause +1.0% deviation based on the stated accuracy.
42
Model UT203/204: OPERATING MANUAL
2.
3.
4.
5.
Remove the transformer jaw.
Repeat the above 1. procedure.
The obtained reading will be more precise.
l Change to AC:
UT203:
Change to AC by using average response method. Input sine wave, then adjust the reading until it is same as the effective value.
UT204:
Combine AC and True RMS response method. Input sine wave to adjust. Non sine wave must follow the below data to adjust:
Peak factor: 1.4~2.0, add 1.0% on the stated accuracy
Peak factor: 2.0~2.5, add 2.5% on the stated accuracy
Peak factor: 2.5~3.0, add 4.0% on the stated accuracy.
43
Model UT203/204: OPERATING MANUAL
MAINTENANCE
This section provides basic maintenance information including battery replacement instruction.
Warning
Do not attempt to repair or service your Meter unless you are qualified to do so and have the relevant calibration, performance test, and service information.
To avoid electrical shock or damage to the Meter, do not get water inside the case.
A. General Service
l l l l l
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents.
To clean the terminals with cotton bar with detergent, as dirt or moisture in the terminals can affect readings.
Turn the Meter power off when it is not in use.
Take out the battery when it is not using for a long time.
Do not use or store the Meter in a place of humidity, high temperature, explosive, inflammable and strong magnetic field.
44
Model UT203/204: OPERATING MANUAL
B. Replacing the Battery (see figure 10)
Warning
To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator “ ” appears.
Screw
Make sure the transformer jaw and the tets leads are disconected from the circuit being tested before opening the case bottom.
To replace the battery:
(figure 10)
1.
2.
3.
4.
5.
Turn the Meter off and remove all the connections from the input terminals
Turn the Meter’s front case down.
Remove the screw from the battery compartment, and separate the battery compartment from the case bottom.
Take out the old battery and replace with a new 9V battery (NEDA1604, 6F22 or
006P.
Rejoin the case bottom and the battery compartment, and reinstall the screw.
** END **
This operating manual is subject to change without notice.
45
Copyright 2005 Uni-Trend Group Limited.
All rights reserved.
Manufacturer:
Uni-Trend Technology (Dongguan) Limited
Dong Fang Da Dao
Bei Shan Dong Fang Industrial Development District
Hu Men Town, Dongguan City
Guang Dong Province
China
Postal Code: 523 925
Headquarters:
Uni-Trend Group Limited
Rm901, 9/F, Nanyang Plaza
57 Hung To Road
Kwun Tong
Kowloon, Hong Kong
Tel: (852) 2950 9168
Fax: (852) 2950 9303
Email: [email protected]
http://www.uni-trend.com
46
Model UT203/204: OPERATING MANUAL
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project