Operator Manual NC001689

March 2001

Issue 3

CHATILLON

®

DFM Series

CADET

®

CD1 Series

- NOTICE -

This operator’s manual may be used for either the CHATILLON ® DFM Series or AMETEK CADET ® CD1 Series digital force gauges. The setup, operation, calibration and maintenance instructions for these different brands of products are identical.

MODEL NUMBERS & CAPACITIES

CHATILLON ® DFM Series

Model

DFM2

DFM10

Capacity x Resolution

2 lb x 0.002 lb

1 kg x 0.001 kg

10 N x 0.01 N

10 lb x 0.01 lb

5 kg x 0.005 kg

50 N x 0.05 N

DFM50 50 lb x 0.05 lb

25 kg x 0.02 kg

250 N x 0.2 N

DFM100 100 lb x 0.1 lb

50 kg x 0.05 kg

500 N x 0.5 N

Note: Force gauges that use 220V source power for charging will have the suffix “-220V” after the base model number, e.g. DFM2-220V.

CADET ® CD1 Series

Model

CD1-002

CD1-010

CD1-050

CD1-100

Capacity x Resolution

2 lb x 0.002 lb

1 kg x 0.001 kg

10 N x 0.01 N

10 lb x 0.01 lb

5 kg x 0.005 kg

50 N x 0.05 N

50 lb x 0.05 lb

25 kg x 0.02 kg

250 N x 0.2 N

100 lb x 0.1 lb

50 kg x 0.05 kg

500 N x 0.5 N

- NOTICE -

Force reading accuracy may be affected by a low battery condition. BEFORE using the force gauge for the first time, fully charge the rechargeable NiCd battery. Plug the charger into the gauge’s charge jack and into an

AC outlet.

Always charge the battery whenever the BAT indicator is highlited. A minimum charge of 8 hours is recommended.

INTRODUCTION

The CHATILLON ® DFM Series (or CADET ® CD1 Series) has been designed to be easy to setup, operate and maintain. The instrument is ideal for tensile or compression measurement. The instrument may be supplied for either 110V or 220V operation. Four base models are available for both the CHATILLON and CADET brands, each model is distringuished by a different capacity. Capacities may be displayed in either lbs, kg or N measurements.

The instrument is supplied with a carrying case, battery charger, hex key, extension rod, hook and flat head attachments.

ELECTRICAL SAFETY

These instruments have been assessed against the essentail heath and safety requirements of the Low Voltage and the EMC Directives listed below. Based on conformity with the listed directives, the DFM and CD1 are compliant with the following:

Directive

EMC Directive (89/336/EEC)

Test Standard

EN 50081-1

EN 50082-1

EMC Generic Emissions Standard

EMC Generic Immunity Standard

Low Voltage Directive (73/23/EEC) EN 61010-1 Safety Requirement for Electrical Equip

SPECIFICATIONS

Accuracy: +0.5% of Full Scale lbs, kg, N Selectable Units:

Overload Capacity: 110% of rated Capacity

Overload Protection: 175% of rated Capacity

Tare Capacity:

Display:

Sampling Rate:

10% of full scale in lb mode

LCD, updated 8 times/ second

1000 Hz

Deflection:

Temperature Range:

Variable, <0.014-inches

(max) at full load

+40°F to +110°F

(+5°C to +45°C)

Temperature Stability: 0.06% per °F

Power:

Battery Life (*):

115V, 60 Hz, <100 mA

220V, 50-60 Hz, <100 mA

8-10 hours between charges

Weight:

Warranty:

1.3 lbs (0.6 kg)

1 year

HANDHELD SETUP

The DFM/CD1 features a reversible split housing. The DFM/CD1 is factory assembled with the shaft pointed down for stand mounting applications. If the gauge is to be used for handheld testing, it may be desireable to reorient the shaft so that it points upward.

-WARNING -

Extreme care should be taken when removing the housing. Make sure that you do not pull, twist or pinch any of the wires inside the housing.

To change the shaft orientation perform the following:

1.

Remove the two hex screws on the backside of the bottom housing using the Hex Key provided with your gauge.

2.

Carefully, rotate the top housing 180-degrees making sure not to bind, twist or pull any of the wires.

3.

Reassemble the housing, making sure that no wires are pinched between the top and bottom housings.

1 /

2

"

#10-32 Thd.

Mtg. Hole

(2 places)

TEST STAND MOUNTING

Two #10-32 tapped holes and a hole for a 3/16-inch dowel pin are provided on the back of the force gauge.

These are used for mounting the gauge to a test stand.

2 1 /

4

"

7"

TEST STAND ADAPTER KITS

Adapter kits are required when mounting your gauge to a CHATILLON or AMETEK test stand. Please see the

Parts List section for part numbers.

2 11 /

16

" 1 1 /

2

"

DISPLAY READOUTS v

T C PEAK BAT

0 0 0 0 LB

KG

N

Indicates TENSION in NORMAL Mode measured in lbs.

T C PEAK BAT v

0 0 0 0 LB

KG

N

Indicates COMPRESSION in NORMAL mode measured in kg.

v

T C PEAK BAT v

0 0 0 0 LB

KG

N

Indicates PEAK TENSION mode measured in N.

T C PEAK BAT v v

0 0 0 0 LB

KG

N

Indicates PEAK COMPRESSION mode measured in

N .

v

T C PEAK BAT v

0 0 0 0 LB

KG

N

Indicates LOW BATTERY condition. Battery MUST be fully charged whenever a Low Battery Condition occurs.

-WARNING -

Force reading accuracy may be affected by a low battery condition. BEFORE using the force gauge for the first time, fully charge the rechargeable NiCd battery. Plug the charger into the gauge’s charge jack and into an

AC outlet.

Always charge the battery whenever the BAT indicator is highlited. A minimum charge of 8 hours is recommended.

BASIC OPERATIONS

Turn Gauge On/Off

Select Units of Measure

Select Force Type (Display Mode)

Zero (Tare) the Displayed Force Value

Position side switch ON/OFF

Select UNITS until desired UNIT is highlited

< LB, KG, N >

Select PEAK until desired TYPE is highlited

< NORM, T-PK, C-PK>

Select ZERO to tare and reset the displayed value.

Zero All Force Values simultaneously.

Reset Current Peak Reading value.

-NOTICE -

Always apply load axially. Applying load on an angle or ecentric loading may cause an erroneous force reading.

OVERLOAD CONDITIONS

Extreme care should be taken to ensure that you do not overload the loadcell. Overload protection up to 175% of the gauge’s capacity is provided in compression mode only, however, any overload condition may damage your gauge. If you do overload the gauge, the display will flash with “0000” or “_ _ _ _”. If your gauge displays “-EP”

“-EPP” or “-EL”, contact your AMETEK or CHATILLON representative for service.

MAINTENANCE & CALIBRATION

You should routinely check and verify gauge accuracy by suspending a known weight from a hook adapter. If calibration is required, please contact AMETEK or your local AMETEK representative.

To maintain optimum performance, we recommend a factory calibration every 12 months.

- WARNING -

This force gauge is equipped with a rechargeable battery. The battery is recyclable. It may be unlawful to dispose of this battery into the municipal waste stream. Please check with yor local solid waste officials for details on how to properly dispose of the battery.

PARTS LIST

15

50 52

88 87

56

16

11

12

54

53

53 59

54

54

54

54 54

89

57

89

57

89

57

88 50

50

88 51

53

58

89 57

58

89 57

58

55

TEST STAND ADAPTER KITS

Adapter kits are required when mounting your gauge to a CHATILLON or AMETEK test stand. Please see the

Parts List section for part numbers.

Kit No.

NC000985

NC000985

LTCK-2

LTCK-2

HTCK-6

Chatillon Stand AMETEK Stand

TCM201 ACCUFORCE 201

TCD200

LTS

ACCUFORCE 200

MLTS

LTC

HTC

MLTS

MHTC

55

56

57

58

59

61*

87

Model DFM and CD1 Series Parts Listing

Ref.

No.

Part

No.

Description

11

12

15

16

50

51

52

53

54

88

89

90*

SPK-FMG-011A Adapter, Flat

SPK-FMG-012A

SPK-FMG-012B

Adapter, Hook

Adapter, Hook

SPK-FMG-015 Hex Key

SPK-FMG-016A Battery Charger, 115V

SPK-FMG-016B Battery Charger, 220V

SPK-DFM-050A Top Case Assembly

SPK-DFM-050B Top Case Assembly

SPK-DFM-050C Top Case Assembly

SPK-DFM-050D Top Case Assembly

SPK-DFM-051

SPK-DFM-052A

Keypad

Lens Assembly

SPK-DFM-052B

SPK-DFM-052C

SPK-DFM-052D

Lens Assembly

Lens Assembly

Lens Assembly

SPK-DFM-053 Printed Circuit Board

SPK-DFM-054A Loadcell Assembly

SPK-DFM-054B Loadcell Assembly

SPK-DFM-054C Loadcell Assembly

SPK-DFM-054D Loadcell Assembly

SPK-DFM-055 Power Jack Assembly

SPK-DFM-056 Extension Rod

SPK-DFM-057 Bottom Case Assembly

SPK-DFM-058 Battery Pack Assembly

SPK-DFM-059 EPROM, Programmed

SPK-DFM-061 Carrying Case

SPK-CD1-087A

SPK-CD1-087B

SPK-CD1-087C

SPK-CD1-087D

Lens Assembly

Lens Assembly

Lens Assembly

Lens Assembly

SPK-CD1-088A Top Case Assembly

SPK-CD1-088B Top Case Assembly

SPK-CD1-088C Top Case Assembly

SPK-CD1-088D Top Case Assembly

SPK-CD1-089 Bottom Case Assembly

SPK-CD1-090 Carrying Case

NOTE: Items with “*” after the Ref. No. are not shown.

DFM DFM DFM DFM

2 10 50 100

CD1

002

CD1

010

CD1

050

CD1

100

1

1

1

1

1

1

1

-

-

1

-

-

-

1

1

-

-

1

1

-

-

1

1

1

-

1

-

-

-

-

-

-

-

-

-

-

1

1

1

1

-

1

1

-

1

1

-

-

1

-

1

-

1

-

1

-

-

1

1

1

-

1

-

-

-

-

-

-

-

-

-

-

1

1

1

1

-

1

1

1

-

1

-

1

-

-

1

-

-

-

1

-

1

1

1

1

1

-

-

-

-

-

-

-

-

-

-

-

1

1

1

1

-

1

1

-

-

1

1

-

-

-

1

1

-

-

1

1

-

1

1

1

1

-

-

-

-

-

-

-

-

-

-

-

1

-

-

1

1

1

1

-

-

1

-

-

-

-

1

-

-

-

1

-

-

1

1

1

-

1

-

-

-

-

1

-

-

1

1

1

1

-

-

1

-

1

1

-

1

1

-

-

-

-

1

-

-

-

1

-

-

1

1

1

-

1

-

-

1

1

-

-

-

-

1

1

1

-

-

1

-

1

1

1

-

1

-

-

-

-

1

-

-

-

1

-

-

1

1

1

1

-

-

1

-

-

-

-

1

-

1

1

1

-

-

1

-

1

1

-

-

1

1

-

-

-

1

-

-

-

1

-

-

1

1

1

1

-

1

-

-

-

-

1

-

-

1

1

PRODUCT WARRANTY

This instrument is warranted against defects in workmanship, material and design for one (1) year from the date of delivery to the extent that

AMETEK will, at its sole option, repair or replace the instrument or any part thereof which is defective provided, however, that the warranty shall not apply to instruments subjected to tampering or abuse, or exposed to highly corrosive conditions.

THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED AND AMETEK HEREBY DISCLAIMS ALL

OTHER WARRANTIES, INCLUDING, WITHOUT LIMITATION, ANY WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR

MERCHANTABILITY. AMETEK SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT

LIMITED TO, ANY ANTICIPATED OR LOST PROFITS.

This warranty is voidable if the purchaser fails to follow any and all instructions, warnings or cautions in the instrument’s operation manual.

If a manufacturing defect is found, AMETEK will replace or repair the instrument or replace any defective part thereof without charge; however,

AMETEK’s obligation hereunder does not include the cost of transportation which must be borne by the customer. AMETEK assumes no responsibility for damage in transit, and any claims for such damage should be presented to the carrier by the purchaser.

TEST AND CALIBRATION INSTRUMENTS

AMETEK Test and Calibration

Instruments

8600 Somerset Drive

Largo, Florida 33773

USA

Tel

Tel

Fax

(727) 536-7831

(800) 527-9999

(727) 539-6882

AMETEK LLOYD

INSTRUMENTS

Forum House

12 Barnes Wallis Road

Segensworth East

Fareham

Hampshire PO 15 5TT

Tel

Fax

+44 (0) 1489 486399

+44 (0) 1489 8851118

AMETEK Precision

Instruments Europe GmbH

Rudolf-Diesel-Strasse 16

D-40670, Meerbusch

Germany

Tel +49 2159 9136 0

Fax +49 2159 9136 39

AMETEK Singapore Pvt. Ltd.

10 Ang Mo Kio Street 65

#05-12 TECHPOINT

Singapore

569059

Tel

Fax

+65 484 2388

+65 481 6588

Internet Addresses: www.ametek.com

www.chatillon.com

www.lloyd-instruments.co.uk

AMETEK and CHATILLON are registered trademarks of AMETEK, Inc.

Information within this document is subject to change without notice.

ISO 9001

Manufacturer

Pub No. NC002490

Issued 03/01

Copyright 2001, by AMETEK, Inc.

Printed in U.S.A.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement