Operator Manual NC001689
March 2001
Issue 3
CHATILLON
®
DFM Series
CADET
®
CD1 Series
- NOTICE -
This operator’s manual may be used for either the CHATILLON ® DFM Series or AMETEK CADET ® CD1 Series digital force gauges. The setup, operation, calibration and maintenance instructions for these different brands of products are identical.
MODEL NUMBERS & CAPACITIES
CHATILLON ® DFM Series
Model
DFM2
DFM10
Capacity x Resolution
2 lb x 0.002 lb
1 kg x 0.001 kg
10 N x 0.01 N
10 lb x 0.01 lb
5 kg x 0.005 kg
50 N x 0.05 N
DFM50 50 lb x 0.05 lb
25 kg x 0.02 kg
250 N x 0.2 N
DFM100 100 lb x 0.1 lb
50 kg x 0.05 kg
500 N x 0.5 N
Note: Force gauges that use 220V source power for charging will have the suffix “-220V” after the base model number, e.g. DFM2-220V.
CADET ® CD1 Series
Model
CD1-002
CD1-010
CD1-050
CD1-100
Capacity x Resolution
2 lb x 0.002 lb
1 kg x 0.001 kg
10 N x 0.01 N
10 lb x 0.01 lb
5 kg x 0.005 kg
50 N x 0.05 N
50 lb x 0.05 lb
25 kg x 0.02 kg
250 N x 0.2 N
100 lb x 0.1 lb
50 kg x 0.05 kg
500 N x 0.5 N
- NOTICE -
Force reading accuracy may be affected by a low battery condition. BEFORE using the force gauge for the first time, fully charge the rechargeable NiCd battery. Plug the charger into the gauge’s charge jack and into an
AC outlet.
Always charge the battery whenever the BAT indicator is highlited. A minimum charge of 8 hours is recommended.
INTRODUCTION
The CHATILLON ® DFM Series (or CADET ® CD1 Series) has been designed to be easy to setup, operate and maintain. The instrument is ideal for tensile or compression measurement. The instrument may be supplied for either 110V or 220V operation. Four base models are available for both the CHATILLON and CADET brands, each model is distringuished by a different capacity. Capacities may be displayed in either lbs, kg or N measurements.
The instrument is supplied with a carrying case, battery charger, hex key, extension rod, hook and flat head attachments.
ELECTRICAL SAFETY
These instruments have been assessed against the essentail heath and safety requirements of the Low Voltage and the EMC Directives listed below. Based on conformity with the listed directives, the DFM and CD1 are compliant with the following:
Directive
EMC Directive (89/336/EEC)
Test Standard
EN 50081-1
EN 50082-1
EMC Generic Emissions Standard
EMC Generic Immunity Standard
Low Voltage Directive (73/23/EEC) EN 61010-1 Safety Requirement for Electrical Equip
SPECIFICATIONS
Accuracy: +0.5% of Full Scale lbs, kg, N Selectable Units:
Overload Capacity: 110% of rated Capacity
Overload Protection: 175% of rated Capacity
Tare Capacity:
Display:
Sampling Rate:
10% of full scale in lb mode
LCD, updated 8 times/ second
1000 Hz
Deflection:
Temperature Range:
Variable, <0.014-inches
(max) at full load
+40°F to +110°F
(+5°C to +45°C)
Temperature Stability: 0.06% per °F
Power:
Battery Life (*):
115V, 60 Hz, <100 mA
220V, 50-60 Hz, <100 mA
8-10 hours between charges
Weight:
Warranty:
1.3 lbs (0.6 kg)
1 year
HANDHELD SETUP
The DFM/CD1 features a reversible split housing. The DFM/CD1 is factory assembled with the shaft pointed down for stand mounting applications. If the gauge is to be used for handheld testing, it may be desireable to reorient the shaft so that it points upward.
-WARNING -
Extreme care should be taken when removing the housing. Make sure that you do not pull, twist or pinch any of the wires inside the housing.
To change the shaft orientation perform the following:
1.
Remove the two hex screws on the backside of the bottom housing using the Hex Key provided with your gauge.
2.
Carefully, rotate the top housing 180-degrees making sure not to bind, twist or pull any of the wires.
3.
Reassemble the housing, making sure that no wires are pinched between the top and bottom housings.
1 /
2
"
#10-32 Thd.
Mtg. Hole
(2 places)
TEST STAND MOUNTING
Two #10-32 tapped holes and a hole for a 3/16-inch dowel pin are provided on the back of the force gauge.
These are used for mounting the gauge to a test stand.
2 1 /
4
"
7"
TEST STAND ADAPTER KITS
Adapter kits are required when mounting your gauge to a CHATILLON or AMETEK test stand. Please see the
Parts List section for part numbers.
2 11 /
16
" 1 1 /
2
"
DISPLAY READOUTS v
T C PEAK BAT
0 0 0 0 LB
KG
N
Indicates TENSION in NORMAL Mode measured in lbs.
T C PEAK BAT v
0 0 0 0 LB
KG
N
Indicates COMPRESSION in NORMAL mode measured in kg.
v
T C PEAK BAT v
0 0 0 0 LB
KG
N
Indicates PEAK TENSION mode measured in N.
T C PEAK BAT v v
0 0 0 0 LB
KG
N
Indicates PEAK COMPRESSION mode measured in
N .
v
T C PEAK BAT v
0 0 0 0 LB
KG
N
Indicates LOW BATTERY condition. Battery MUST be fully charged whenever a Low Battery Condition occurs.
-WARNING -
Force reading accuracy may be affected by a low battery condition. BEFORE using the force gauge for the first time, fully charge the rechargeable NiCd battery. Plug the charger into the gauge’s charge jack and into an
AC outlet.
Always charge the battery whenever the BAT indicator is highlited. A minimum charge of 8 hours is recommended.
BASIC OPERATIONS
Turn Gauge On/Off
Select Units of Measure
Select Force Type (Display Mode)
Zero (Tare) the Displayed Force Value
Position side switch ON/OFF
Select UNITS until desired UNIT is highlited
< LB, KG, N >
Select PEAK until desired TYPE is highlited
< NORM, T-PK, C-PK>
Select ZERO to tare and reset the displayed value.
Zero All Force Values simultaneously.
Reset Current Peak Reading value.
-NOTICE -
Always apply load axially. Applying load on an angle or ecentric loading may cause an erroneous force reading.
OVERLOAD CONDITIONS
Extreme care should be taken to ensure that you do not overload the loadcell. Overload protection up to 175% of the gauge’s capacity is provided in compression mode only, however, any overload condition may damage your gauge. If you do overload the gauge, the display will flash with “0000” or “_ _ _ _”. If your gauge displays “-EP”
“-EPP” or “-EL”, contact your AMETEK or CHATILLON representative for service.
MAINTENANCE & CALIBRATION
You should routinely check and verify gauge accuracy by suspending a known weight from a hook adapter. If calibration is required, please contact AMETEK or your local AMETEK representative.
To maintain optimum performance, we recommend a factory calibration every 12 months.
- WARNING -
This force gauge is equipped with a rechargeable battery. The battery is recyclable. It may be unlawful to dispose of this battery into the municipal waste stream. Please check with yor local solid waste officials for details on how to properly dispose of the battery.
PARTS LIST
15
50 52
88 87
56
16
11
12
54
53
53 59
54
54
54
54 54
89
57
89
57
89
57
88 50
50
88 51
53
58
89 57
58
89 57
58
55
TEST STAND ADAPTER KITS
Adapter kits are required when mounting your gauge to a CHATILLON or AMETEK test stand. Please see the
Parts List section for part numbers.
Kit No.
NC000985
NC000985
LTCK-2
LTCK-2
HTCK-6
Chatillon Stand AMETEK Stand
TCM201 ACCUFORCE 201
TCD200
LTS
ACCUFORCE 200
MLTS
LTC
HTC
MLTS
MHTC
55
56
57
58
59
61*
87
Model DFM and CD1 Series Parts Listing
Ref.
No.
Part
No.
Description
11
12
15
16
50
51
52
53
54
88
89
90*
SPK-FMG-011A Adapter, Flat
SPK-FMG-012A
SPK-FMG-012B
Adapter, Hook
Adapter, Hook
SPK-FMG-015 Hex Key
SPK-FMG-016A Battery Charger, 115V
SPK-FMG-016B Battery Charger, 220V
SPK-DFM-050A Top Case Assembly
SPK-DFM-050B Top Case Assembly
SPK-DFM-050C Top Case Assembly
SPK-DFM-050D Top Case Assembly
SPK-DFM-051
SPK-DFM-052A
Keypad
Lens Assembly
SPK-DFM-052B
SPK-DFM-052C
SPK-DFM-052D
Lens Assembly
Lens Assembly
Lens Assembly
SPK-DFM-053 Printed Circuit Board
SPK-DFM-054A Loadcell Assembly
SPK-DFM-054B Loadcell Assembly
SPK-DFM-054C Loadcell Assembly
SPK-DFM-054D Loadcell Assembly
SPK-DFM-055 Power Jack Assembly
SPK-DFM-056 Extension Rod
SPK-DFM-057 Bottom Case Assembly
SPK-DFM-058 Battery Pack Assembly
SPK-DFM-059 EPROM, Programmed
SPK-DFM-061 Carrying Case
SPK-CD1-087A
SPK-CD1-087B
SPK-CD1-087C
SPK-CD1-087D
Lens Assembly
Lens Assembly
Lens Assembly
Lens Assembly
SPK-CD1-088A Top Case Assembly
SPK-CD1-088B Top Case Assembly
SPK-CD1-088C Top Case Assembly
SPK-CD1-088D Top Case Assembly
SPK-CD1-089 Bottom Case Assembly
SPK-CD1-090 Carrying Case
NOTE: Items with “*” after the Ref. No. are not shown.
DFM DFM DFM DFM
2 10 50 100
CD1
002
CD1
010
CD1
050
CD1
100
1
1
1
1
1
1
1
-
-
1
-
-
-
1
1
-
-
1
1
-
-
1
1
1
-
1
-
-
-
-
-
-
-
-
-
-
1
1
1
1
-
1
1
-
1
1
-
-
1
-
1
-
1
-
1
-
-
1
1
1
-
1
-
-
-
-
-
-
-
-
-
-
1
1
1
1
-
1
1
1
-
1
-
1
-
-
1
-
-
-
1
-
1
1
1
1
1
-
-
-
-
-
-
-
-
-
-
-
1
1
1
1
-
1
1
-
-
1
1
-
-
-
1
1
-
-
1
1
-
1
1
1
1
-
-
-
-
-
-
-
-
-
-
-
1
-
-
1
1
1
1
-
-
1
-
-
-
-
1
-
-
-
1
-
-
1
1
1
-
1
-
-
-
-
1
-
-
1
1
1
1
-
-
1
-
1
1
-
1
1
-
-
-
-
1
-
-
-
1
-
-
1
1
1
-
1
-
-
1
1
-
-
-
-
1
1
1
-
-
1
-
1
1
1
-
1
-
-
-
-
1
-
-
-
1
-
-
1
1
1
1
-
-
1
-
-
-
-
1
-
1
1
1
-
-
1
-
1
1
-
-
1
1
-
-
-
1
-
-
-
1
-
-
1
1
1
1
-
1
-
-
-
-
1
-
-
1
1
PRODUCT WARRANTY
This instrument is warranted against defects in workmanship, material and design for one (1) year from the date of delivery to the extent that
AMETEK will, at its sole option, repair or replace the instrument or any part thereof which is defective provided, however, that the warranty shall not apply to instruments subjected to tampering or abuse, or exposed to highly corrosive conditions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED AND AMETEK HEREBY DISCLAIMS ALL
OTHER WARRANTIES, INCLUDING, WITHOUT LIMITATION, ANY WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR
MERCHANTABILITY. AMETEK SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOT
LIMITED TO, ANY ANTICIPATED OR LOST PROFITS.
This warranty is voidable if the purchaser fails to follow any and all instructions, warnings or cautions in the instrument’s operation manual.
If a manufacturing defect is found, AMETEK will replace or repair the instrument or replace any defective part thereof without charge; however,
AMETEK’s obligation hereunder does not include the cost of transportation which must be borne by the customer. AMETEK assumes no responsibility for damage in transit, and any claims for such damage should be presented to the carrier by the purchaser.
TEST AND CALIBRATION INSTRUMENTS
AMETEK Test and Calibration
Instruments
8600 Somerset Drive
Largo, Florida 33773
USA
Tel
Tel
Fax
(727) 536-7831
(800) 527-9999
(727) 539-6882
AMETEK LLOYD
INSTRUMENTS
Forum House
12 Barnes Wallis Road
Segensworth East
Fareham
Hampshire PO 15 5TT
Tel
Fax
+44 (0) 1489 486399
+44 (0) 1489 8851118
AMETEK Precision
Instruments Europe GmbH
Rudolf-Diesel-Strasse 16
D-40670, Meerbusch
Germany
Tel +49 2159 9136 0
Fax +49 2159 9136 39
AMETEK Singapore Pvt. Ltd.
10 Ang Mo Kio Street 65
#05-12 TECHPOINT
Singapore
569059
Tel
Fax
+65 484 2388
+65 481 6588
Internet Addresses: www.ametek.com
www.chatillon.com
www.lloyd-instruments.co.uk
AMETEK and CHATILLON are registered trademarks of AMETEK, Inc.
Information within this document is subject to change without notice.
ISO 9001
Manufacturer
Pub No. NC002490
Issued 03/01
Copyright 2001, by AMETEK, Inc.
Printed in U.S.A.