Bulletin No. 1567-544

LSM-902/6900 Laser Scan Micrometer

LSM-902 (Measurement unit) & LSM-6900 (Display unit) are the most accurate

1" (25mm) measurement range Laser Scan Micrometer in the world.

LSM-902

Adjustable Workstages (optional)

LSM-6900

Features:

Measuring range – .004 to 1" (0.1 to 25mm) covers most precision parts.

Repeatability – ±.000002" (±0.05µm) highest repeatability for a 1" (25mm) LSM in the world.

Linearity – ±.00002" (±0.5µm) over the entire range, and higher linearity in the small range.

High Resolution – Improves measurement accuracy

■ Stable edge detection by using new algorithm.

Multi-function type display.

Model and Order No.

Model

LSM-902/6900

Order No.

544-496A

Display type mm/inch

Application of laser products

[FDA]/Class II

LSM-902/6900 Specifications and Accessories

LSM-902

Measuring range

Reading (Selectable)

Repeatability / [*1]

Linearity / [*1]

Positional error / [*1] [*5]

Measuring region

Number of scans

Laser wavelength

Laser scanning speed

LSM-6900

Display

Whole range

Small range

.004~1.0" (0.1~25mm)

.000001~.0005" (0.01~10µm)

±.000002" (±0.05µm) / [*2]

±.000020" (±0.5µm) / [*3]

±(0.3 + 0.1 ▲ D)µm [ ▲ D:mm]

±(.000012 + .0001 ▲ D)" [ ▲ D: inch]

±.000020" (±0.5µm)

±0.6 x 1.0" (±1.5 x 25mm) / [*6]

800/sec

650nm [Visible LD], 1.5mW (peak)

2240"/sec (56m/sec)

/ [*3] [*4]

Fluorescent display 16-digit + 11-digit, Guidance LEDs

Segment designation

Edge designation

1 to 7 (1 to 3 for Transparent material)

1 to 255

Averaging method

Simple averaging

Moving averaging

1 to 2048

32 to 2048

Tolerance judgement (GO,±NG); Multi-limit judgement (7 classes); Offset/Zero-set; Abnormal data elimination; Automatic workpiece detection; Reference value setting; Data output condition; Laser power deterioration; Sample measurement;

Measuring function [*7]

Scanning signal monitor connector

Remote interlock connector

Power switch

Built-in interface

Statistical processing; Dual-program measurement; Automatic measurement using edge mode; Workpiece position display; Transparent object measuring; Key-lock function; mm/inch changeover; Dual-gauge calibration; Selection of resolution; Judgement in ready state; Display of a comma to mark the thousandths position; Mastering; None-display unnecessary digits; Group judgement

Provided as standard (with the plug)

Provided as standard (with the plug)

Key switch used

RS-232C; Foot switch connector; I/O analog interface

Optional interface

DCU slot

Expansion slot (1-slot)

Digimatic code output unit (2-ch)

2nd I/O analog I/F; BCD I/F; GP-IB I/F;

Power supply AC100V~240V±10%, 50/60Hz, 40VA

[*1]: Accuracy inspection environment/Temperature 20°C±1°C, Humidity: 50%±10% [*2]: The repeatability is determined by the value for ±2

σ at the measurement interval of 1.28 sec.

[*3]: Specified at the center of the measurement region. [*4]: ▲ D: Diameter difference to master gage. [*5]: An error due to workpiece shift either in the optical axis direction or in the scanning direction.

[*6]: The area given by [measuring range on the optical axis] x [measuring range in the scanning direction.] [*7]: The combination of functions is limited, details are described in the user's manual.

STANDARD ACCESSORIES

Order No.

Item

02AGC320A

930966

Signal cable (5m/16ft)

Power cable

934626

956042

214938

02AGC401

GHD lead wire

Short plug to Remote-interlock

Plug for Remote-interlock/MP121M by Marushin

Plug for Scanning signal connector/MP-105LC

02AGC604

02AGC605

99MBC071A

99MBC075A by Marushin

Power ON/OFF key

Fuse 1A (Time-lag)

User's manual for LSM-6900 (English)

User's manual for LSM-902 (English / FDA)

Quantity

1

1

1

1

1

1

1

1

1

1

OPTIONAL ACCESSORIES

Order No.

02AGD180

Item

Calibration gauges set for LSM-902

02AGD270

02AGD280

02AGC840

02AGC880

Workstage for LSM-501/503/902

Adjustable Workstage for LSM-902

Digimatic code output unit (2-ch)

2nd I/O analog interface unit

02AGC910

02AGC940

02AGD600B

223663

936937

937179T

BCD interface unit

GP-IB interface unit

Printer & cable set

Printer paper

Digimatic code output cable

Foot switch

Distributed By:

© 2003 Mitutoyo America Corporation, Aurora IL www.mitutoyo.com

Mitutoyo America Corporation

Michigan

Regional Offices

Illinois California

Phone: (734) 459-2810 Phone: (630) 978-5385 Phone: (626) 961-9661

Massachusetts Indiana North Carolina

Phone: (978) 692-8765 Phone: (317) 577-6070 Phone: (704) 875-8332

We reserve the right to change specifications and prices without notice.

11A-3 • Printed in USA • May 2003

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement