TH2821A
Portable LCR Meter
OPERATION MANUAL
English
1 st
November 2005
Edition Rev 1.0.0
© Copyright 2005 Changzhou Tonghui Electronic Co., Ltd.
All rights reserved.
Contents
TH2821A OPERATION MANUAL
Contents............................................................2
How to Contact Us .......................................3
Incoming Inspection .....................................4
Notes on Use ...............................................5
Warranty:....................................................6
Chapter 1 Overview............................................7
1.1 Introduction...........................................7
1.2 Main Functions.......................................7
1.3 Specifications ........................................9
1.4 Environment Requirements ...................12
Chapter 2 Panel Illustration..............................13
2.1 LCD Display Illustration.........................13
2.2 Keyboard Illustration ............................15
Chapter 3 Operation.........................................17
3.1 Power on .............................................17
3.2 How to operate....................................18
3.2.1 First Key Functions .....................18
3.2.2 Second Key Functions: ................20
3.3 Battery recharge..................................30
3.4 Clearing Instruction..............................31
Appendix .........................................................32
2
TH2821A OPERATION MANUAL
How to Contact Us
Changzhou Tonghui Electronic Co., Ltd.
Address: No. 3, Tianshan Road, New District,
Changzhou, Jiangsu, China
Tel: 0086-519-5132222, 5113342, 5109592
Fax: 0086-519-5109972 http:// www.tonghui.com.cn
http:// www.tonghui.cn
Email: [email protected]
3
TH2821A OPERATION MANUAL
Incoming Inspection
Inspect the shipping container for damage. The contents of the shipment should be listed as follows. If the contents are incomplete, if there is damage or defect, please contact our company or your nearest Sales and Service Office.
Accessories
TH26028 DC Power Adapter
TH26027 4 terminal Kelvin test clip leads
1
1
LH-200H7C rechargeable battery (inside) 1
Manual 1 User
Options
TH26029 SMD component test fixture
4
Notes on Use
TH2821A OPERATION MANUAL z This meter is only for indoor use. z Turn off the TH2821A while switching the power supply between battery and DC adapter or replacing the battery. z Although internal circuit protection is provided, DC voltage or current may damage TH2821A. Before you measure a capacitor, be sure the capacitor is fully discharged. z Charging may disturb measurement result sometimes. z Nickel Metal Hydride rechargeable battery can be used for the power supply. TH2821A will not work normally when battery voltage is less than 6V. z The 12V AC to DC adaptor is recommended to be used for TH2821A power supply. z Perform Open and Short corrections for accurate measurement especially when test fixture is changed. z The functions locked with password are not accessible by users.
5
Warranty:
TH2821A OPERATION MANUAL
This instrument product is warranted against defects in material and workmanship for a period of two years from the date of shipment. During the warranty period, Our company will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility designated by our company.
Warranty limitation
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer,
Buyer-supplied software or interfacing, unauthorized modification or misuse, or improper site preparation or maintenance.
6
Chapter 1 Overview
Thank you for purchasing our product. To get the maximum performance from the instrument, please read this manual first, and keep this manual at hand.
1.1 Introduction
TH2821A is a microprocessor-controlled portable meter with low power consumption. It can measure six basic parameters, they are inductance L, capacitance C, resistance R, impedance |Z|, dissipation factor D and quality factor Q.
TH2821A can fulfill the measurement needs of various component manufacturers and maintenance technicians.
1.2 Main Functions
1. Test Parameter
L-Q, C-D, R-Q and Z-Q.
2. Correction
OPEN: multi-frequency correction of open circuit;
SHORT: multi-frequency correction of short circuit.
3. Display Mode
Direct —— direct measurement value;
∆
∆%
—— absolute deviation;
—— percent deviation.
4. Range Hold
When measuring a large number of components with the same nominal value, this function can effectively improve the measuring rate.
TH2821A OPERATION MANUAL
5. Comparator Function
Built-in 4 Bins comparator: NG, P1, P2 and P3.
6. Equivalent Circuit Mode
Both parallel and series equivalent circuit modes can be obtained.
7. Data Hold
This function can be used to freeze the current display value.
8. Alarm Mode
NG, P1, P2, P3 and OFF modes can be selected.
8
TH2821A OPERATION MANUAL
1.3 Specifications
Parameter
Frequency
Accuracy
Display
Measurement
Range
Test Level
(Range Auto and Open
Circuit)
L-Q, C-D, R-Q and Z-Q
100 Hz, 120 Hz , 1 kHz and 10 kHz
Basic Accuracy: 0.3%
5 digits display for both primary and secondary parameters
100 Hz,120 Hz 1 µH – 9999 H
L
1 kHz,10 kHz 0.1 µH – 999.9 H
C
100 Hz, 120Hz
1kHz,10 kHz
1 pF – 9999 µF
0.1 pF – 999.9 µF
R, |Z| 0.0001 Ω - 999.9 MΩ
D, Q
∆%
0.0001 – 9999
0.0001% - 9999%
10kHz 0.1 Vrms (1±15%)
120Hz
1kHz
0.3 Vrms (1±15%)
100Hz 0.42 Vrms (1±15%)
Ranging Mode Auto and Hold
Equivalent
Circuit
Parallel and Series
Display
Correction
Rate
Terminals
Direct, ∆ABS and ∆%
Open and Short Zeroing
Approx. 3 meas/sec
5 terminals
9
TH2821A OPERATION MANUAL
(Continued)
Comparator 4 Bins:NG, P1, P2 and P3
∆% -9999% - 99999%
Limit Setup
Range
L
C
R
Z
0.0001 µH – 99999 H
0.0001 pF – 99999 µF
0.0001 Ω - 99999 MΩ
0.0001 Ω - 99999 MΩ
Alarm Mode NG, P1, P2, P3 and OFF
Power Supply
9V rechargeable battery or DC12V(100 mA) adapter
Low Battery
Indication
Approx. 6V
Power
Consumption
Normal: Approx. 25 mA
Auto power-off: Approx. 500 nA
Auto Power
Off time
Approx. 5 minutes
Weight Approx. 400 g
Dimensions 200mm(L) × 95mm(W) × 40mm(D)
Table 1-1 Specifications
10
TH2821A OPERATION MANUAL
*
Note:
Primary parameter accuracy(A e
C: A e
= 0.3%(1+C x
L: A e
= 0.3% (1+ L
/C max x
/L
)
+C max min
/C
+ L min x
)
/L x
)
Z: A e
= 0.3% (1+ Z x
R: A e
/Z max
+ Z min
/Z x
)
= 0.3%(1+ R x
/R max
+ R min
/R x
)
Max and Min values are as follows:
Parameter Range
C max
Auto
80µF/f
C min
150pF/f max
L max
159H/f
L min
0.32mH/f
Z max
Z min
= R max
; Z min
= R min
1MΩ
1.59Ω
, Frequency unit: kHz. Here:Z
Secondary parameter accuracy
D
D e e
= A
= A e e
/3
(1+D x when D x
)/3 when D x
≤0.1
>0.1
Q e
= ±
1
Q x
µ
Q
× x
D
× e
D e
when Q x
·D e
<1
11
TH2821A OPERATION MANUAL
1.4 Environment Requirements
1 . Please do not operate TH2821A under the following environment conditions, as any of them will directly affect measuring precision or damage the meter: z
Please do not operate the instrument in places where is dusty, vibrant, under direct sunlight, or where there is corrosive air. z Although TH2821A has been specially designed for reducing the noise caused by AC power, the environment with low noise is still recommended.
If this can not be arranged, please make sure to use power filter for the AC-DC adaptor.
2. The TH2821A must be operated under the following environment conditions:
Temperature: 0°C ~ 40°C,
Humidity: ≤ 90% RH at 40°C.
3. Storage Temperature: -25°C ~ 50°C.
12
TH2821A OPERATION MANUAL
Chapter 2 Panel Illustration
2.1 LCD Display Illustration
Figure 2-1 LCD Display
13
TH2821A OPERATION MANUAL
No. Description No.
1
Battery Power
2
3
Indicator
Remote
Indicator
Beeper
9
Description
Ranging Mode Indicator
4
5
6
7
Indicator
Series/Parallel
Indicator
Comparator
Indicator
Secondary
Parameter
Indicator
Secondary
Parameter
Display
13 2 nd Function Indicator
Table 2-1 LCD Description
Others:
DH: Data hold indictor
CAL: Correction function indictor
MENU: Menu operation indictor
14
TH2821A OPERATION MANUAL
2.2 Keyboard Illustration
POWER
①
CLEAR
PARA
ESC
②
LIMIT
U ③
SHIFT
⑧
EQU
DH
⑦
▲
BUZZER
⑥
V
MENU
FREQ
④
ENTER
DISPLAY
AUTO
⑤
Figure 2-2 Keyboard
Â
Convention:
Key Function Conventions
1 st
2 nd
function: PARA
function: CLEAR
3 rd function:
}
15
No. Key
TH2821A OPERATION MANUAL
Function
SHIFT+
Key
○
○
○
○
PARA
U
FREQ
AUTO
On/Off
Parameter
Selection
Range Up
Frequency
Selection
Range Auto
Selection
CLEAR
LIMIT
MENU
Function
Correction key
Sorting Limit
Setup
Auxiliary
Menu
V
DH
Range Down BUZZER Alarm Mode
Data Hold EQU
Equivalent
Circuit
○ SHIFT 2 nd Function
The 3 rd
Table 2-2 Key Description
Functions:
}
, o , q , p , r and x
are valid during Menu setup and Data input operations.
16
TH2821A OPERATION MANUAL
Chapter 3 Operation
3.1 Power on
1. Press POWER key to turn on TH2821A.
2. The operation system version will be displayed.
3. At last the instrument enters the measurement state.
6. Parameter D
8. Unit
Figure 4-1 Measurement Display
Measurement display description:
1. Battery Supply
4. Series Circuit
3. Beeper ON
5. Sorting Result: P1
7. Secondary parameter
9. Range 0 (Auto)
10. Primary parameter 11. Frequency: 1 kHz
12. Parameter C
17
TH2821A OPERATION MANUAL
3.2 How to operate
3.2.1 First Key Functions
1. Parameter Setup:
Press PARA key to select the following measurement parameter combinations:L-Q, C-D, R-Q and Z-Q.
Units Description:
L µH mH H
C pF nF µF
R/|Z| Ω kΩ
Table 4-1 Units
MΩ
|Z| is the absolute value of impedance. Measurement value of L, C or R may be positive or negative. Negative capacitance value means that the device under test is actually an inductor; also negative inductance value means that the device under test is actually a capacitor. In theory
R should be positive constantly, under some condition, R may be negative due to over zero correction. Please carry out correct zero correction.
The maximum number of display digits is 5, but 5-digit is not always available and 4-digit is displayed sometimes.
The conversion is described in the following description:
From 4-digit to 5-digit:
When the first 2 digit of current value is less than 18.
From 5-digit to 4-digit:
When the first 2 digit of current value is more than 20.
2. Frequency Setup:
Use FREQ key to select the following test frequencies in turn: 100 Hz, 120 Hz ,1 kHz and 10 kHz.
18
TH2821A OPERATION MANUAL
3. Range Setup:
U, V and AUTO keys are used to set the measurement range. AUTO key toggles ranging mode between “Auto” and “Hold”. U and V keys are used to increase or decrease the measurement range, if the current ranging mode is “Auto”, then the ranging mode is changed to
“Hold” at the same time.
*
Note:
When ranging mode is set to HOLD, the measurement range is fixed at current range. Overload symbol “-----” will be displayed if the impedance under test exceeds the current effective measurement range or display range.
Range
No.
Range
Resistor
Range
Up
Range
Down
0 100kΩ
Ï
20kΩ
1
2
3
4
10kΩ
1kΩ
100Ω
20Ω
Ï
2kΩ
Ï
200Ω
Ï
20Ω
Ï
Ð
18kΩ
Ð
1.8kΩ
Ð
180Ω
Ð
18Ω
Ð
Table 4-2 Ranges
19
TH2821A OPERATION MANUAL
*
Note:
How to calculate the measurement range
Example: Assume capacitance C=210pF, dissipation D=0.0010 and test frequency f=1 kHz.
Solution:
Z
X
=
R
X
+ j
2 π
1 fC
X
No. 2.
Z
X
≈
2 π
1 fC
X
=
2 × 3 .
1416 ×
1
1000 × 210 × 10 − 9
≈ 757 .
9 Ω
From the Table 4-2, we can get the correct measurement range is
4. Data Hold
Press DH key to freeze the display, press DH key again to release.
3.2.2 Second Key Functions:
1. Correction Function z Press u key to select the second function, “SHIFT” will be lighted on the screen. z Press CLEAR key to enter the correction function, the following information will be displayed on the screen.
Figure 4-2 Correction Display
20
TH2821A OPERATION MANUAL z Clear (Clear) is displayed in the primary parameter display area, OpeN (OPEN), Short (Short) or Quit (Quit) will be displayed in the secondary parameter display area.
*
Note:
OpeN (OPEN), Short (Short) and Quit (Quit) are selected and displayed automatically by the meter according to the impedance value of the device under test. z Press
}
key to cancel the correction operation and return to the measurement state. Press start the correction measurement. x
key to z When correction measurement is finished, PASS or FAIL will be displayed on the screen, see Figure 4-3 z Press
Figure 4-3 Open correction Passed
}
key to abort the current correction data measurement and return to the measurement. The previous correction data will still be stored.
*
Note:
1. The correction function must be used for accurate measurement.
The correction function can eliminate the stray admittance
(capacitance, and inductance) and the residual impedance
(resistance and reactance) induced by test fixture, test leads and
21
TH2821A OPERATION MANUAL instrument itself. Perform the correction operation again if the measurement conditions are changed such as test fixture and environment temperature.
2. It is recommended to perform the open and short correction at the same time.
3. During short correction period, FAIL (FAIL) will be displayed in the secondary parameter display area when short correction is failed.
Make sure that the measurement contacts are shorted reliably and perform the short correction again.
4. TH2821A measures the correction data at all frequency points and all measurement ranges. The correction data is stored in the non-volatile memory. So you don’t have to perform the correction again, if the test conditions are not changed.
5. Open and short corrections are automatically selected by the instrument according to the impedance value under test. If there is a component in the fixture or if there is error with the instrument, Quit (Quit) will be displayed in the secondary parameter display area.
2. Equivalent Circuit z Press u key to select the second function, “SHIFT” will be lighted on the screen. z Press EQU key to select the Series or Parallel circuit mode. (Refer to Figure4-1)
*
Note:
1. The actual C, R and L are not the ideal pure C, R and L. Normally an actual component can be regarded as the combination of an ideal resistor and an ideal reactor in series or parallel circuit mode.
2. TH2821A can convert between the two different equivalent circuit modes using the following equations. The measurement values of the two different circuit modes maybe different under different quality factor Q (or dissipation factor D).
22
TH2821A OPERATION MANUAL
Capacitance Cp: from parallel to series
Cp
Circuit Mode:
Rp
Dissipation:
D
Series: C
S
=
=
1
2 π fC
P
( 1 +
D
R
P
2
=
)
C
P
1
Q
R
S
=
R
P
D
2 ( 1 +
D
2 )
Capacitance Cs:from series to parallel
Circuit Mode: Cs
Rs
Dissipation:
D
= 2 π fR
S
C
S
=
1
Q
Parallel:
R
C
P
P
=
= 1
R
S
( 1 +
( 1 +
D
2 )
C
S
D
2 ) D
2
Inductance Lp:from parallel to series
Lp
Circuit Mode:
Rp
Dissipation:
D
Series: L
S
R
S
2
R fL
P
P
= 1 ( 1 +
=
1
Q
D
2 )
L
P
=
23
R
P
D
2 ( 1 +
D
2 )
TH2821A OPERATION MANUAL
Inductance Ls:from series to parallel
Lp Rp
Dissipation:
D
Parallel:
L
P
=
=
2 π
R
S fL
S
( 1 +
=
1
Q
D
2 ) L
S
R
P
=
R
S
( 1 +
D
2 ) D
2
Here parameter with subscript s means the series mode, parameter with subscript p means the parallel mode.
3. From the above equations, we can conclude that the conversion between series and parallel is determined by D 2 or Q 2 (Q=1/D).
The value of D 2 or Q 2 directly determined the parameter values in different circuit mode.
Example:
Three capacitors have the same series capacitance: Cs=0.1µF, but their dissipations are different with each other: D1=0.0100,
D2=0.1000, D3=1.0000. According to the above equation, we can get their capacitance in parallel mode:
Cp1 = 0.09999 µF
Cp2 = 0.09901 µF
Cp3 = 0.05000 µF
We can find that Cs is almost the same with Cp when D is very small (D< 0.01), but when D is more than 0.01,Cp and Cs are different obviously. For example: When D = 0.1,the difference is
1%,but when D = 1,the difference is almost 50%.
3. Alarm Setup z Press u key to select the second function, “SHIFT” will be lighted on the screen.
24
TH2821A OPERATION MANUAL z Press BUZZER, the following information will be displayed on the LCD screen: z Use r and q
Figure 4-4 Alarm Setup
keys to select following alert modes in turn.
OFF Alert off
P1 P1 alert mode
P2 P2 alert mode
P3 P3 alert mode
Ng NG alert mode z Press x to save the alert setup and return to the measurement state. z press z Press
} u
to abort the alert setup without change.
4. Display Mode
key to select the second function, “SHIFT” will be lighted on the screen. z Press DISPLAY key to select the display mode: Direct,
Absolute deviation(∆), Percent deviation(∆%).
*
Note:
1. Absolute deviation (∆)display mode
25
TH2821A OPERATION MANUAL
Figure 4-5 Absolute deviation(∆)display mode
∆ =
X x
−
X std
Where X x
is the measurement value, X std
is the standard value.
2. Percent deviation(∆%)display mode
Fixture 4-6 Percent deviation(∆%)display mode.
∆ % =
X x
X
− std
X std
⋅ 100 %
26
TH2821A OPERATION MANUAL
5. Sorting Setup
Â
Key Convention:
Key Main Menu Data Input r Select the former item Select the left digit q Select the next item Select the right digit p o
Select parameter for
StD setup
Select parameter for
StD setup
Increase the digit, move the point left and set the unit
Decrease the digit, move the point right and set the unit x
Enter the sub menu Confirm the data input, and
} Return to the measurement state Return to the main menu
Table 4-4 Key convention in sorting setup z Press u key to select the second function, “SHIFT” will be lighted on the screen. z Press LIMIT to enter the sorting setup, P1~ will be flashing on the screen, and the value of P1~ is also displayed. z Use r, q to select the following sorting parameters:
P1_: Low limit of P1
P1~: High limit of P1
P2_: Low limit of P2
P2~: High limit of P2
P3_: Low limit of P3
P3~: High limit of P3
27
TH2821A OPERATION MANUAL
Q_: Low limit of Q
D~: High limit of D
StD: Standard value
}
key to abort the sorting setup and return to z Press the measurement state. z Press x key to select a parameter and enter the data input operation.
Figure 4-7 Sorting Limit Setup z Use r, q keys to select a digit or the point. z Use p , o keys to set the numeric value. z Press } to return to the main menu. z Press x to confirm and save the numeric data. z Select parameter STD to set the standard Value.
Figure 4-8 Standard Value Setup z Use r, q keys to select a digit, the point or the unit. z Use p , o keys to set the numeric data and unit. z Press
}
to return to the main menu.
28
z Press
TH2821A OPERATION MANUAL x
to confirm and save the numeric data.
Sorting
Setup
N D ≤ Dmax
Q ≥ Qmin
Y
P1min ≤ ∆% ≤ P1max
N
P2min ≤ ∆% ≤ P2max
N
P3min ≤ ∆% ≤ P3max
N
NG
Y
Y
Y
P1
P2
P3
END
Figure 4-8 Sorting Flow Chart
29
TH2821A OPERATION MANUAL
6. Other Functions: z Press u key to select the second function, “SHIFT” will be lighted on the screen. z Press MENU key, APO will be flashing on the screen.
Current delay time is also displayed. x
key to enter the auto power off function z Press setup, current delay time flashes on the screen. z Use p , o keys to select 5’, 10’, 20’,30’ or OFF. z Press x key to confirm the current delay time and return to the main menu.
*
Note:
When an external DC adapter is used, APO function is disabled automatically. z Use r, q keys to select “CAL” function. z Press x
key to enter the calibration function. This function is protected with password and is not available for users.
3.3 Battery recharge
*Note:
The instrument will not be started, when the battery is weakening. Recharge the battery immediately by using
DC Adaptor Power Supply. z Battery Power Indicator will flash when charging.
30
TH2821A OPERATION MANUAL z Charging may disturb measurement result sometimes. z The battery must be 9V series Nickel Metal Hydride rechargeable battery. For example GP20R8H and
LH-2007HC. 。 z 2~3 hours for charging and 5~6 hours for using.
3.4 Clearing Instruction
To clean the instrument, use a soft cloth slightly dipped in water. Do not spray cleanser directly onto the instrument, since it may leak into the cabinet and cause damage. Do not use chemicals containing benzine, alcohol or aromatic hydrocarbons.
31
TH2821A OPERATION MANUAL
Appendix
Message Code Table : clear Clear: Correction
OpeN Open: Open correction short Short: Short correction
P1~
P1_
Mg
Q_
D~
Std
P1 — : High limit of P1 Bin
P1_: Low limit of P1 Bin
NG: No-Good
Q_: Low limit of quality factor
D_: High limit of dissipation factor
Std: Standard value (Nominal value)
APO: Auto Power Off ApO
Cal
Psd
Alert
OM
Off pass faIl
CAL: Accuracy Calibration
PSD: Password
Alert
ON
OFF
Pass
Fail
Quit Quit
First Edition2005-11