User`s Manual - Graham Field

JB0112-014
Medical Equipment
With respect to electric shock, fire
and mechanical hazards only in
accordance with UL-60601-1 and
CAN/CSA C22.2 No. 601.1
E303594
User’s Manual
Caution:
Federal law restricts this device to sale by or on order of a physician, or any other
practitioner licensed by the law of the State in which he practices to use or order the use of
this device.
NOTE:
Check ALL parts for shipping damage. If shipping damage is noted, DO NOT use.
Contact medical equipment provider for further instruction.
JB0112-014-INS-LABRevB07
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
Contents
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
AC Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Illustration of Important Parts . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
How to Operate Your JOHN BUNN Suction Unit . . . . . . . . . . .6
Cleaning Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
Specifications/Classifications/Symbols . . . . . . . . . . . . . . . . . .13
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
1. SAFETY INFORMATION
When using electrical products, especially when children are present, basic safety precautions
should always be followed. Read all instructions before using.
Important information is highlighted by these terms:
DANGER:
Urgent safety information for hazards that will cause serious injury or death.
WARNING:
Important safety information for hazards that might cause serious injury.
CAUTION:
Information for preventing damage to the product.
NOTE:
Information to which you should pay special attention.
READ ALL USER’S MANUAL BEFORE USING.
SAVE USER’S MANUAL
DANGER
To reduce the risk of electrocution:
1. Do not use while bathing.
2. Do not place or store product where it can fall or be pulled into a tub or sink.
3. Do not place in or drop into water or other liquid.
4. Do not reach for a product that has fallen into water. Unplug immediately.
WARNING
To reduce the risk of burns, electrocution, fire or injury to persons:
1. Close supervision is necessary when this product is used by, on, or near
children or invalids.
2. Use this product only for its intended use as described in this guide.
3. Never operate this product if:
a. It has a damaged power cord or plug.
b. It is not working properly.
c. It has been dropped or damaged.
d. It has been dropped into water.
4. Return the product to an authorized JOHN BUNN medical equipment service center
for examination and repair.
5. Keep the power cord away from heated surfaces.
6. Never use while drowsy or asleep.
SPECIFIC WARNINGS
TO REDUCE THE RISK OF ELECTRICAL SHOCK OR INJURY TO PERSONS, DO NOT
DISASSEMBLE THIS UNIT. Disassembly or attempted repairs if accomplished incorrectly
can create electrical shock hazard. Only qualified technicians should perform servicing of
this suction unit.
2
JB0112-014-INS-LABRevB07
DANGER
The JOHN BUNN Suction Unit is a vacuum suction device designed for the collection of
nonflammable fluid materials in medical applications only. Improper use during medical
applications can cause injury or death.
For all medical applications:
1. All suctioning should be done in strict accordance with appropriate procedures that have been
established by a licensed medical authority.
2. Some attachments or accessories may not fit the tubing supplied. All attachments or
accessories should be checked prior to use to assure proper fit.
2. AC POWER SOURCE
CAUTION
The Model JB0112-014 is used with a power supply allowing operation on AC voltage
(115 VAC, 60 Hz). However, the correct power source must be used to connect to
adaptable wall power.
3. ELECTRICAL
Compressor
If the overload protector shuts off the motor frequently, you may have a low voltage situation. Low
voltage can also be suspected when:
1. The motor does not get up to full power or speed.
2. Other motor operated appliances fail to operate properly. Too many motor operated appliances
on same circuit.
NOTE:
The motor of this compressor has a thermal overload protector. If the motor overheats, the
overload protector will shut off the motor. If this occurs, turn the ON/OFF switch to the OFF
position and allow the motor to cool (approximately 5 minutes). Restart compressor. If the
compressor fails to start, wait an additional 5 minutes before attempting to restart. If this fails,
inspect the fuses on your suction unit. If the unit still does not restart, call your medical
equipment provider immediately.
WARNING:
Proper operation of this product is necessary to avoid physical harm. Improper use can
cause personal injury and/or tissue damage. This product should not be used unless the
operator has been instructed by a physician or another qualified health care professional.
1. Familiarize yourself with this unit.
2. Place unit on table and push down hard so that suction feet hold unit firmly in place.
3. Before plugging power cord into electrical outlet, check if the ON/OFF switch is in the OFF position.
NOTE:
Before attempting to start a cold unit, let the unit reach room temperature in a heated area.
3
JB0112-014-INS-LABRevB07
4. INTRODUCTION
Your JOHN BUNN Suction Unit is a professional medical suctioning device that produces a
maximum vacuum of 560 mmHg. The unit is equipped with an 800 cc collection bottle and has
been designed for convenient and reliable usages.
Features:
• New anti-bacteria filter design helps reduce contamination.
• High quality vacuum regulator and anti-vibration vacuum gauge setting.
• Oil-less, permanent lubrication.
• Collection bottle with overflow protection.
• Superior performance and high reliability for continuous operation design.
• Easy carry and easy cleaning.
Following the recommended operating and maintenance procedures outlined in this user’s manual
will maximize the life of this product. Refer to the diagrams on page 6.
4
JB0112-014-INS-LABRevB07
5. ILLUSTRATION OF IMPORTANT PARTS
MODEL:
1. Vacuum Adjust Knob
2. Vacuum Gauge
3. Bacteria Filter
4. Connection Tubing
5. Cover
6. Patient Tubing
7. 800cc Collection Bottle
8. Holder
9. Fuse
10. Power Switch
11. Power Cord
12. Base
13. Rubber Feet
14. Cord Fixed Support
Figure 1
Figure 2
5
JB0112-014-INS-LABRevB07
6. SET-UP
1. Connect either end of the connecting tubing (item# 4) to the top of collection bottle connector
then connect the other end to the bacteria filter.
2. The patient tubing should be connected to collection bottle at the outlet labeled (patient). Please
assure that all connections are secure and without leaks before using.
7. HOW TO OPERATE YOUR JOHN BUNN SUCTION UNIT
1. Before use, check the specification label on the side of the suction unit to ensure that the
voltage and current indicated on the unit correspond with the voltage and current available.
Make sure the power switch located on the side of the unit is in the “Off” position before
connecting the unit to a power source.
Figure 3
“Off” position
2. Switch the power switch to “I” position (on), the unit will start running.
Figure 4
“On” position
6
JB0112-014-INS-LABRevB07
3. Adjust the vacuum level from 0 to 560 mm Hg by turning the vacuum adjust knob (clockwise to
increase vacuum and counter-clockwise to decrease vacuum). Referring to the vacuum gauge
while setting the desired level of vacuum. To accurately read the gauge, block the patient end of
the hose or cap off the collection bottle and allow the gauge to reach a stable vacuum reading.
Vacuum Gauge
Vacuum Adjust Knob
Figure 5
4. Place the patient tubing to the appropriate position of patient and start to suck sputum.
Patient End of the Hose
Figure 6
7
Cap
Collection Bottle
JB0112-014-INS-LABRevB07
5. When the sputum reaches the safe full level of collection bottle (800 cc), the suctioning will stop
sucking automatically (float shut-off).
6. Switch the power switch to “O” position (off), shut off the suction motor.
7. Remove the sputum from collection bottle after the motor stops turning completely.
(Please refer to the section “CLEAN INSTRUCTIONS” on page 9 for the directions of
how to clean the collection bottle).
8. Turn on the unit and start to run the suction unit again.
NOTE:
1. If the unit sustains a severe drop, the validity of the gauge must be checked. Inspect if the
float shut-off is still functioning properly when the liquid in the collection bottle reaches the
safe full level.
2. Always transport unit with vacuum adjust knob rotated fully clockwise in case the unit is dropped.
CAUTION:
1. When automatic float shut-off is activated, contents of the collection bottles should be
emptied. Further suctioning could cause damage to the vacuum pump.
2. Should fluid be aspirated back into the unit, send your suction to an authorized service
center for a detailed inspection as the vacuum pump may have been damaged.
8
JB0112-014-INS-LABRevB07
8. CLEANING INSTRUCTIONS
Clean Collection Bottle:
1. Turn off the suction unit and allow vacuum to drop, then disconnect AC power source.
2. With the collection bottle still in the holder, remove the collection bottle lid. The bottle can
now be taken out of the holder to be emptied.
Collection Bottle Lid
Holder
Collection Bottle
NOTE:
Collection bottle should be emptied after each use.
3. Collection bottle should be thoroughly cleaned after each use by one of the following three
methods:
a. Wash in a hot water/dishwashing detergent solution and rinse with clean, hot tap water. Then
wash in one part vinegar to three parts hot water solution. Rinse with hot tap water and air dry.
b. Wash with rubbing alcohol and air dry
c. Wash with a commercial (bacterial-germicidal) disinfectant; follow disinfectant manufacturer's
recommended instructions and dilution rates carefully.
Clean Suction Unit and Collection Bottle Lid:
1. Disconnect the suction unit from all external power sources.
2. Wipe the suction unit housing and collection bottle lid with a clean cloth and any commercial
(bacterial-germicidal) disinfectant.
CAUTION
Do not submerge unit in water as this will result in severe damage to the vacuum pump.
9
JB0112-014-INS-LABRevB07
Clean Tubing:
1. Disconnect the tubing from the unit.
2. Tubing should be rinsed thoroughly after every use by running hot tap water through it followed
by a solution of one part vinegar to three parts hot water.
3. Rinse with hot tap water and air dry.
4. Keep the outer surface of the tubing clean by wiping with a clean, damp cloth.
WARNING:
The waste in collection bottle may cause respiratory infection which must be handled
according to medical waste disposing regulation. Operator must wear gloves and a
medical class mask before cleaning the collection bottle, tubing and handle the wasting
unit and/or accessories.
9. MAINTENANCE
Inspect suction tubing and collection bottle for leaks, cracks, etc. before each use.
DANGER:
Electric shock hazard
Do not attempt to open or remove the suction cabinet. If service is required, return
your suction unit to an authorized service center. Opening or tampering with the unit
will void the warranty.
Bacteria Filter Change:
1. The bacteria filter should be replaced every two months. If overflow occurs, change
the filter immediately.
2. Remove the filter by disconnecting it from the tubing connected to the pressure adjusting
valve and the tubing connected to the collection bottle lid assembly.
3. Replace with a new JOHN BUNN bacteria filter and connect to above two tubing.
Additional filters can be purchased from an authorized dealer.
10
JB0112-014-INS-LABRevB07
Tubing
Vacuum
Adjust
Knob
Bacteria
Filter
Tubing
Collection
Bottle
NOTE
Use only JOHN BUNN filters. Do not substitute any other material for this bacteria filter. Improper
substitution may lead to contamination or poor performance.
Fuse Change:
1. Use only the same type of fuse to replace the broken fuse.
2. Fuse specification: Miniature Glass Fuse 5x20mm, Fast Blow (Quick Acting), F2A/250V.
WARNING
When this unit or accessories reach the end of their useful lives (depends on using
conditions), residues in collection bottle, tubing and filter may cause respiratory infection.
Dump the wasting collection bottle, tubing and filter into a specific waste bin labeled with
medical waste.
11
JB0112-014-INS-LABRevB07
10. TROUBLESHOOTING
Your JOHN BUNN Suction Unit contains no user-serviceable parts. If you believe your unit is not
working properly, BEFORE YOU RETURN IT TO THE MEDICAL EQUIPMENT PROVIDER WHERE YOU
PURCHASED IT OR TO JOHN BUNN, please take a few moments to check for these possible causes:
PROBLEM
ACTION
Motor runs, but no vacuum
1. Verify tubing connection security
2. Check for leaks or tubing kinks
3. Ensure that float shut-off is not activated
4. Check for bottle leaks and cracks
Unit does not turn on
Low Vacuum
1. Check power source and connection
2. Ensure that the wall outlet is live
3. Check power cord for damage
1. Use vacuum adjust knob to increase vacuum level
2. Check system for leaks
3. Adjust vacuum adjust knob and release
12
JB0112-014-INS-LABRevB07
11. SPECIFICATIONS / CLASSIFICATIONS
Dimensions
Weight
Maximum Vacuum
Flow Range
Sound Level
Electrical Requirements
Power Consumption
Bacteria Filter Pore Size
Collection Bottle Capacity
Operating Temperature
Operating Humidity
Operating Atmospheric Pressure
Storage Temperature
Storage Humidity
Storage Atmospheric Pressure
Fuse Specification
Equipment Classifications
(L) 375 x (W) 175 x (H) 250 mm
(14.8 x 6.8 x 9.8 in)
5.5 kg (12.1 lb)
560 ± 3% mm Hg
>44 LPM
<58 dBA
JB0112-014 AC 115V ~ 60Hz
JB0112-014 70 watts
0.45 microns
800 ml (cc)
32° ~ 104° F (0° ~ 40°C)
95% RH
10.2 ~ 15.4 psi ( 70 ~106 kpa )
14° ~ 158° F ( -10° ~ 70°C )
0 ~ 95%
7.3 ~ 15.4 psi ( 50 ~106 kpa )
Miniature Glass Fuse 5x20mm,
Fast Blow (Quick Acting), F2A/250 V
With respect to protection from electric shock Class I
13
JB0112-014-INS-LABRevB07
12. WARRANTY
Two years limited warranty
The compressor portion of the JOHN BUNN Suction Unit is warranted to be free from
defective workmanship and materials for a period of two years from date of purchase
and six months on accessories.
Any defective part(s) will be repaired or replaced at Graham-Field Health Products’ option if the unit
has not been tampered with or used improperly during that period. Make certain that any
malfunction is not due to inadequate cleaning or failure to follow the instructions. If repair is
necessary, please contact the medical equipment provider where you purchased it or JOHN BUNN.
NOTE
This warranty does not cover providing a loaner unit, compensating for costs incurred in rental
while said unit is under repair, or for labor costs incurred in repairing or replacing defective part(s).
THERE IS NO OTHER EXPRESS WARRANTY. IMPLIED WARRANTIES, INCLUDING THOSE OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO THE
DURATION OF THE EXPRESS LIMITED WARRANTY AND TO THE EXTENT PERMITTED BY LAW
ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED. THIS IS THE EXCLUSIVE REMEDY AND
LIABILITY FOR CONSEQUENTIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL
WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLUSION IS PERMITTED BY LAW. SOME
STATES (IN THE US) DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY
LASTS, OR THE LIMITATION OR EXCLUSION OF CONSEQUENTIAL OR INCIDENTAL DAMAGES,
SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU.
This warranty gives you specific legal rights, and you may also have other rights, which vary from
state to state (in the US).
Manufacturer's Note
Thank you for choosing a JOHN BUNN Suction Unit. We would like you to be a satisfied customer.
If you have any questions or comments, please send them to our address on the back cover.
For Service Call Your Authorized JOHN BUNN Medical Provider:
U.S.A., Corporate Headquarters:
Graham-Field Health Products
2935 Northeast Parkway
Atlanta, Georgia 30360
telephone: 800-347-5678, 770-447-1609
fax: 800-726-0601, 678-291-3232
http://www.grahamfield.com
JB0112-014-INS-LABRevB07