C-LOGIC

55

INSTRUCTION MANUAL

P

ocket-size digital multimeter

Safety information

This meter has been designed according to EN61010-1,

EN61010-2-030,EN61010-2-033, 300V CAT III and pollution2.

Follow all safety and operating instructions to ensure the meter is used safely and is kept in good condition With proper use and care, your digital multimeter will give you years of satisfactory service.

During use

• Never exceed the protection limit indicated in the specifications for each range of measurement.

• Never use the meter to measure voltages that might exceed 300V above earth ground in category III installations

• Always be careful when working with voltages above

60v dc or 30v ac rms. keep fingers behind the probe

barriers while measuring.

• Do not perform resistance measurements on live circuits.

• Inspect test leads and probes for cracks, breaks or crazes in the insulation before using the meter

• If the equipment is used in a manner not specified by manufacturer, the protection provided by equipment may be impaired.

Safety Symbols

Important safety information ,refer to the instruction manual.

Earth(ground) TERMINAL

01

Equipment protected throughout by double insulation.

Fuse must be replaced with ratings specified in the manual.

AC (Alternating Current)

CAT III

DC (Direct Current)

It is applicable to test and measuring circuits connected to the distribution part of the building's low-voltage MAINS installation.

European union directives

CONFORMS TO UL STD. 61010-1, 61010-2-030,

61010-2-033, 61010-031; CERTIFIED TO CSA STD.

C22.2 No.61010-1,61010-2-030, 61010-2-033,

61010-031.

FRONT PANEL

M300C

HIGH SAFETY

DMM

V V A

2

V kΩ

20

V kΩ

200

V kΩ mA

300V

2000kΩ

OFF

CAT III 300V 300V 200mA MAX

3

2

4

1

Maintenance

• Before opening case .always disconnect test leads from all energized circuits.

• For continuous protection against fire. Replace fuse only

with ratings: F 250mA/300V (Quick Acting)

• Never use the meter unless the back cover is in place and

fastened completely.

• Do not use abrasives or solvents on the meter .to clean it use only a damp cloth and mild detergent.

02

1. Range switch

This switch is used to select desired ranges as well as to turn on/off the meter .

2. Function switch switch for selecting functions.

3. LCD display

3 1/2digit.7segment.maximum 1999 counts.

4. Test leads

Red test lead for positive (+) and black test lead for negative (-).

03

General description

This compact digital multimeter is designed to measure AC and DC voltages .DC current .Resistance .

Diode and to perform audible continuity checks with accuracy and easy .

Small and lightweight .with a carrying case and test leads wound on its body. This instrument will provide you years of satisfactory service.

Specification

Accuracy is guaranteed for 1 year .

23°C ±5°C, Less than 75%RH.

DC Voltage

Range

2V

20V

200mV

300V

Resolution

1mV

10mV

100mV

1V

Accuracy

±0.5% of rdg ±1dgt

±0.8% of rdg ±1dgt

±0.8% of rdg ±1dgt

±0.8% of rdg ±1dgt

Overload protection:300V dc or rms ac for all ranges

AC Voltage

Range

200V

300V

Resolution

100mV

1V

Accuracy

±1.2% of rdg ±10dgts

±1.2% of rdg ±10dgts

Overload protection: 300V dc or rms ac for all ranges

Frequency range: 45Hz to 450Hz

Response: average responding .calibrated in rms of a sine wave

04

DC Current

Range

200mA

Resolution

0.1mA

Accuracy

±2.0% of rdg ±2dgts

Overload protection: 250mA/300V fuse

Resistance

Range Resolution

2k Ω

20k Ω

200k Ω

2000k Ω

1 Ω

10 Ω

100 Ω

1k Ω

Accuracy

±1.0% of rdg ±2dgts

±1.0% of rdg ±2dgts

±1.0% of rdg ±2dgts

±1.0% of rdg ±2dgts

Maximum open circuit voltage : 0.65V

overload protection: 250V rms ac for all ranges

Diode test

Range Description

Show the approx. forward voltage drop of the diode.

Overload protection: 250V rms ac

Diode test

Range Description

Built-in buzzer sounds when resistance is less than 50 Ω.

Overload protection: 250V rms ac

05

General characteristics

Maximum voltage between

Terminals and earth ground

Fuse protection

Power supply

Display

Measuring method

Over range indication

Polarity indication

Operating temperature

Storage temperature

Low battery indication

Size

Weight

CAT III 300V

F 250 mA/300V

12V battery .GP -23AX1

LCD .1999 counts .updates 2-3/sec dual-slope integration A/D converter only figure “1” on the display

“ - ” displayed for negative polarity

0°C to 40°C (32°F to 104 °F)

-10°C to 50°C (10°F to 122°F) appears on the display

120x70x18mm approx.110g

Operating instruction

DC Voltage measurement

1. Set the function switch at V position

2. Set the range switch at desired position. If the magnitude of voltage to be measured is unknown beforehand .

set the range switch at the highest position and then reduce until satisfactory reading is obtained.

3. Connect test leads across the source or load being measured. The polarity of red lead connection will be indicated at the same time as the voltage value.

4. When the range switch is set at 300V position. “HV”sign will appear on the display to remind user of high voltage measurement. Special attention should be paid.

06

AC Voltage measurement

1. Set the function switch at V~ position

2. Set the range switch at desired position .measurement

reading can be obtained at 2V and 20V positions, but the accuracy is not guaranteed.

3. Connect test leads across the source or load being measured and read the voltage value on the LCD display.

4. When the range switch is set at 300V position , a “HV” sign

will appear on the display to remind user of high voltage

measurement.

DC Current measurement

1. Set the function switch at A position.

2. Set the range switch at 200mA position .measurement

reading can be obtained at other positions. But the decimal point will be at the incorrect place.

3. Open the circuit in which the current is to be measured.

And connect test leads in series with the circuit.

4. Read current value on the LCD display along with the polarity of red lead connection.

Resistance measurement

1. Set the function switch at Ω position .(note: the polarity of red lead is positive “+”)

2. Set the range switch at desired position

3. Connect test leads across the resistor to be measured

and read LCD display.

4. If the resistor was being measured is connected to a circuit .turn off power and discharges all capacitors before applying test leads.

07

Diode test

1. Set the function switch at Ω position .(note: the polarity of red lead is positive “+”)

2. Set the range switch at position.

3. Connect the red test lead to the anode of the diode to be tested and the black lead to the cathode of the diode.

4. The approx. forward voltage drop of the diode will be displayed in mv.if the connection is reversed. Only figure

“1”will be shown

Audible continuity test

1. Set the function switch at Ω position.

2. Set the range at position.

3. Connect test leads to two points of the circuit to be tested.

if the resistance is less than 50 Ω. Buzzer will sound.

Battery & Fuse Replacement

If the sign appears on the LCD display .it indicates that the battery should be replaced. Remove the screw on the back cover and open the case .replace the exhausted battery with a new one of the same type.

Fuse rarely need replacement and blow almost always as a result of operator's error. open the case and replace the blown fuse with the ratings specified: F 250 mA/300V.

WARNING

disconnected from the measured circuit before removing the rear cover. Make sure the rear cover is tightly screwed before using the instrument.

Replacing test leads

Replace test leads if leads become damaged or worn.

WARNING

If the test leads need to be replaced, you must use a new one which should meet EN 61010-031 standard, rated CAT III 300V, 0.4A or better.

Warning

Before attempting to open the case .always be sure that test

leads have been disconnected form measurement circuits.

close case and tighten screws completely before using the

meter to avoid electrical shock hazard.

Accessories

Battery 1pcs (12V GP-23A)

Carrying Case 1pcs

Operating manual 1pcs

MGL EUMAN , S.L.

Parque Empresarial de Argame,

C/Picu Castiellu, Parcelas i-1 a i4

E-33163 Argame, Morcín

Asturias, España, (Spain)

08

00-05-4065

C-LOGIC

55

MANUAL DE INSTRUCCIONES

Multímetro digital compacto

Información de seguridad importante, referido al manual de instrucciones

Terminal de tierra (Ground)

01

Equip o protegido con doble aislamiento .

El fusible debe ser cambiado siguiendo los valores nominales especificados en el manual .

AC ( corriente alterna )

CAT III

DC ( corriente directa/continua )

Es aplicable a los circuitos de prueba y medición conectados a la parte de distribución de la instalación de red de baja tensión del edificio .

Directivas de la unión europea

De acuerdo a UL STD. 61010-1, 61010-2-030,

61010-2-033, 61010-031; CERTIFIED TO CSA STD.

C22.2 No.61010-1,61010-2-030, 61010-2-033,

61010-031.

PANEL FRONTAL

M300C

HIGH SAFETY

DMM

V V A

2

V kΩ

20

V kΩ

200

V kΩ mA

300V

2000kΩ

OFF

CAT III 300V 300V 200mA MAX

3

2

4

1

02 03

Rang o Resolu c i ó n

2V

20V

200mV

300V

1mV

10mV

100mV

1V

Precisión

±0.5%

±0.8%

±0.8%

±0.8% de de de de

rdg ±1dgt

rdg ±1dgt

rdg ±1dgt

rdg ±1dgt

Protección sobrecarga : 300V DC o rms AC para todos los rangos

Tensión AC

Rang

200V

300V o Resolu

100mV

1V c i ó n Precisión

±1.2% de rdg ±10dgts

±1.2% de rdg ±10dgts

Protección sobrecarga : 300V DC o rms AC para todos los rangos

Rango de frecuencia : 45Hz a 450Hz

Respuesta: respuesta promedio .calibrado en rms a sine wave

04

Corriente DC

Rang o

200mA

Resolu

0.1mA

c i

ó n

Precisión

±2.0% de

rdg ±2dgts

Protección sobrecarga

: fusible

250mA/300V

Resist e nc ia

Rang o

2k Ω

20k Ω

200k Ω

2000k Ω

Resolu c i

ó n

1 Ω

10

100

1k

Ω

Ω

Ω

Precisión

±1.0% de

rdg ±2dgts

±1.0% de

rdg ±2dgts

±1.0% de

rdg ±2dgts

±1.0% de

rdg ±2dgts

Tensión máxima con circuito abierto

: 0.65V

Protección sobrecarga

: 250V rms

AC para todos los rangos

Prueba diodo

Rang o

Descrip c i

ó n

Muestra aprox. la caída de tensión directa del diodo

.

Protección sobrecarga: 250V rms AC

Prueba diodo

Rango Descripción

Sonidos de timbre integrados cuando la resistencia es inferior a 50 Ω.

Protección sobrecarga: 250V rms AC

Características generales

Voltaje máximo entre terminales y tierra

Protección de fusibles

Fuente de alimentación

Pantalla

Método de medición

Indicación de sobre rango

Indicación de polaridad

Temperatura de funcionamiento

Temperatura de almacenamiento

Indicación de batería baja

Tamaño

Peso

CAT III 300V

F 250 mA/300V

Batería

12V.GP -23AX1

LCD .1999 cu e nt a s.

Refresh

2-3/sec convertidor A / D de integración de doble pendiente sólo figura "1" en la pantalla

Se muestra "-" para polaridad negativa

0°C a

40°C (32°F a

104 °F)

-10°C a

50°C (10°F a

122°F) aparece en la pantalla

120x70x18mm approx.110g

Instrucción de funcionamiento

Medición de tensión CC

Medición de tensión AC

Medición de corriente DC

Medición de la resistencia

05 06 07

Prueba diodo Reemplazo de cables de prueba

Reemplace los cables de prueba si los cables se dañan o desgastan.

ADVERTENCIA

Si es necesario reemplazar los cables de prueba, debe usar uno nuevo que cumpla con la norma EN 61010-031, con clasificación CAT III 300V, 0.4A o mejor.

Prueba audible de continuidad

1.

Fije el selector de función en la posición

Ω .

2.

Fije el rango en la posición

3.

Conecte los cables de prueba a dos puntos del circuito a probar.

Si la resistencia es menor de

50 Ω la sirena sonará

.

Reemplazo de batería y fusible

Si el símbolo aparece en la pantalla LCD. Indica que se debe reemplazar la batería. Quite el tornillo de la tapa trasera y abra la carcasa. Reemplace la batería agotada por una nueva del mismo tipo.

El fusible rara vez necesita ser reemplazado y casi siempre se funde como resultado de un error del operador. Abra la caja y reemplace el fusible fundido con los valores nominales especificados: F 250 mA / 300V.

ADVERTENCIA

sondas estén desconectadas del circuito medido antes de trasera esté bien atornillada antes de usar el instrumento.

Advertencia

Antes de intentar abrir la carcasa, asegúrese siempre de que los cables de prueba se hayan desconectado de los circuitos de medición. Cierre la caja y apriete los tornillos completamente antes de usar el medidor para evitar el riesgo de descarga eléctrica.

Acce sorios

Batería

Estuche de transporte

Manual de instrucciones l

1pcs (12V GP-23A)

1pcs

1pcs

MGL EUMAN , S.L.

Parque Empresarial de Argame,

C/Picu Castiellu, Parcelas i-1 a i4

E-33163 Argame, Morcín

Asturias, España, (Spain)

08

00-05-4065

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement