PROLITE-24
User’s Guide
nm
Fiber Test UTP
Mini Optical Power Meter, Visual Fault Locator
& RJ45 connection tester dBm
PROLITE-24
Description of the controls and elements.
>2s REF
UNIT
LED
LIGHT
VFL
P R O L I T E -
24
Fiber Test UTP www.promaxelectronics.com
1
2
3
Introduction
The PROLITE-24 is a multifunctional optical power meter. It integrates the functions of a Visual Fault Locator (VFL) and RJ45 Ethernet
Cable Tester. It also has a flashlight. Its small size (pocket size) makes it a very lightweight, strong and durable tool. It is an instrument that is specially suitable for applications such as FTTx broadband installation and maintenance tests and comprehensive tests of fiber optic network cable.
Description of the controls and elements.
nm dBm
>2s REF
UNIT
LED
LIGHT
VFL
P R O L I T E 24
Fiber Test UTP
4
5
6
1
2
Short press to turn on / long press to turn off (Short press to turn off the
Auto-off function when starting up); In user calibration mode, short press to save the current calibration value.
Backlight switch used to turn on/off the backlight.
3
4
5
Switches VFL on/off, short press to switch CW/1Hz / 2Hz; Short press in user calibration mode to decrease 0.05dB.
Short press to switch measurement unit among dBm, dB and mW; Long press to set current measurement value as reference value; Short press in user calibration mode to increase 0.05dB.
Short press to switch the measurement unit between dBm, dB and mW;
Long press to select current measurement value as reference value; Short press in calibration mode to increase 0.05 dB.
6 Select measurement wavelength: 850/1300/1310/1490/1550/1625/1650 nm
www.promaxelectronics.com
PROLITE-24
0 DG0414_ 11-22
1
5
2
4
3
1 VFL Interface.
2 LED flashlight.
3 Optical Power Meter Interface.
4 Type-C USB (Internal control).
5
RJ45 interface and RJ45 far-end module, used for network line sequence pairing.
Specifications
OPTICAL POWER METER
Detector type
Calibrated Wavelength
Measurement range
Frequency identification
Resolution
Accuracy
Measurement units
Measurement modes
Optical connector
VISUAL FAULT LOCATOR
Output wavelength
Output power
Output mode
Optical port
TESTER ETHERNET
Connection
Operating mode
OTHERS
Display
Flashlight
Calibration
POWER SUPPLY
MECHANICAL FEATURES
Dimensions
Weight
InGaAs
850 / 1300 / 1310 / 1490 / 1550 / 1625 / 1650 nm
From -70 dBm to +6 dBm
270 Hz / 330 Hz / 1 kHz / 2 kHz
0.01 dBm
± 5 % dBm / dB / mW
Normal or based on a user-defined reference value
2.5 mm universal interface (support FC, SC and ST connectors)
650 nm ±3 nm
2 mW
CW / 1 Hz / 2 Hz
2.5 mm universal interface (support FC, SC and ST connectors)
Connection of the ends of the ethernet cable to the remote module, and to the equipment
Sequential test of the 8 wires of the RJ-45 cable
Backlit LCD
LED flashlight
Selectable calibration mode
3 x AAA batteries
Auto power-off 10 minutes
67 (W.) x 104 (H.) x 28 (D.) mm
140 gr
www.promaxelectronics.com