Pre-Installation Guide

(Service Installation)

Part Number: 15-2353-074

Rev: 1, May - 2015

Copyright © 2015 Teledyne Technologies Incorporated

All rights reserved. Reproduction, adaptation, or translation without permission is prohibited, except as allowed under copyright laws. Printed in the U.S.A.

Updated Information

The information contained in this document is subject to change without notice.

Trademarks

The companies indicated own the following trademarks:

QuickTrace ..........................................Teledyne Leeman Labs

All other brands or product names are trademarks or registered trademarks of their respective owners.

Teledyne Leeman Labs

110 Lowell Road

Hudson, NH 03051 U.S.A

www.teledyneleemanlabs.com

Sales Support: 800-634-9942

Technical Support: 800-533-6267

Main: 603-886-8400

Fax: 603-886-4322

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation

Guide (Service Installation)

1.1 Introduction

Thank you for purchasing a Teledyne Leeman Lab’s QuickTrace

®

M-8000 CVAF mercury analyzer. The purpose of this guide is to assist new users in the preparation of their laboratory prior to the installation of the Teledyne Leeman Lab’s QuickTrace

®

M-8000

CVAF mercury analyzer by the Teledyne Leeman Lab’s Service Engineer.

In preparation for the installation of your analyzer, please review this pre-installation guide to ensure that your laboratory is prepared. After the installation, please retain a copy of this guide for your records.

If there is any doubt about the information provided, please contact your local Service

Representative or Teledyne Leeman Labs using information in Section 1.15 "Contact

Information".

1.2 Receiving the Instrument

Your analyzer will be installed by a Teledyne Leeman Lab’s Service Engineer. Please do not unpack any boxes without consulting the Teledyne Leeman Lab’s Customer

Support Department (see Section 1.15 "Contact Information" if necessary).

The Service Engineer will be responsible for reviewing the shipment against the packing list. The Service Engineer cannot be responsible for this task, nor can Leeman Labs be responsible for any missing items, if boxes have been opened or removed before the arrival of the installation engineer.

The Service Engineer is a skilled professional who will install your equipment, verify that it is operating to specifications, and train your personnel in its basic operation. Your preparation enables you to use his/her visit to the best advantage.

NOTE

The connection of the QuickTrace® M-8000 CVAF mercury analyzer to your company’s internal communication network/server is not part of the normal installation process and is best left to a trained Information Technology (IT) professional.

Introduction

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

1

1.3 Pre-Installation Requirements

This section summarizes the site requirements necessary for proper installation of the

QuickTrace

®

M-8000 mercury analyzer. Specific details of these requirements are included in the sections which follow.

Table 1-1 QuickTrace® M-8000 Site Requirements

Dimensions

Width (Base)

Depth (Base)

Height

Weight

Electrical Requirements

Four Grounded Electrical

Outlets

Gas Supply

Type

Purity

48 cm (19 in

60 cm (24 in) a

20 cm (8 in) b

16.8 Kg (37 lbs)

100 - 240 VAC ±10%, 50/60 Hz

Ventilation (for ENC-500 Enclosure, if installed)

Recommended Rate 283 LPM (10 CFM)

Trunk Diameter 11 cm (4 in)

Inert Gas Regulator

Recommended Regulator

Pressure Range

Argon

Ultra-High Purity (99.999%)

Argon

Two-Stage

69 - 689 kPa (10 - 100 psi)

Connection Size 3.175 mm (1/8 in) OD a. The optional ENC-500 enclosure will require 81 cm (32 in) in base depth to allow for the exhaust chimney. Ideally, an unobstructed open island type of bench either 24 or 36 inches deep will suffice.

b. The optional ENC-500 enclosure w/elevating legs for the autosampler requires 86 cm (33.75 in) in height.

1.4 Customer Supplied PC Requirements

A customer supplied PC must meet the requirements in Table 1-2 "Minimum PC

Hardware/Software Requirements"..

Table 1-2 Minimum PC Hardware/Software Requirements

Operating System

Microsoft® Windows® 7 (32 and 64-bit) and Windows® 8.1 (64-bit only)

Random Access Memory (RAM) 2 GB Minimum

Video Processing

Processor

1024 x 768 with 24-bit color

Pentium Dual Core 2.3 GHz

Networking Communication

Ports

Additional Hardware

Two free USB communication ports, either serial and/or USB

Internet Browser

CD-ROM drive

Printer with Windows® compatible print driver

PS/2 or Bus Mouse

Microsoft® Internet Explorer® 4 or higher must be installed for the online Help to function

2

NOTE

Also refer to the Teledyne Leeman Labs Supplied Computer Specifications document for PC specifications best suited to the M-8000 mercury analyzer.

Pre-Installation Requirements

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

1.5 Choosing a Location

NOTE

Always position the equipment so that it is easy to disconnect the power cord.

1.5.1 Spacial Requirements According to Configuration

The QuickTrace

®

M-8000 Mercury Analyzer System includes the analyzer, PC (standard desk top mini tower) with monitor, reagent and rinse bottles, 10 L waste bottle

(supplied), and an optional autosampler and autosampler enclosure.

NOTE

30 cm x 30 cm (1 ft

2

) of floor space is required for the liquid waste receptacle.

The waste receptacle can be located directly below the analyzer or directly in front of the lab bench (in line with the analyzer’s peristaltic pump).

Approximate spacial requirements are shown in Table 1-3 "M-8000 Spacial

Requirements".

NOTE

Allow an additional 5 cm (2 in) behind the system for cable egress, ventilation, and access to power switches. If installing the optional ENC-500 enclosure, allow ~ 21 cm (8 in) behind the enclosure for ventilation (total minimum bench depth ≈ 32). If the bench does not provide enough depth for this additional clearance, a chase behind the bench can be used.

The ASX-520 autosampler is made to be placed on top of the M-8000 analyzer and the width and depth required is the same.

Choosing a Location

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

3

 If the M-8000 is being used with an ENC-500 enclosure with elevating legs the autosampler sits inside the enclosure (instead of on top of the M-8000 analyzer) and the enclosure sits over the M-8000 analyzer. In this configuration, enclosure height, width and depth requirements include the

M-8000 analyzer and ASX-520 autosampler.

Table 1-3 M-8000 Spacial Requirements

a

Width

Depth

Configuration M-8000 w/ENC-500 (with elevating legs) and

ASX-520 Rugged

Autosampler

56 cm (22 in)

60 cm (24 in) c

PC

92 cm (36 in)

44 cm (17 in)

Height (Unobstructed Space Above)

86 cm (33.75 in) 41 cm(16 in) a. 30 cm x 30 cm (1 ft

2

) of floor space is required for the liquid waste receptacle. The waste receptacle can be located directly below the analyzer or directly in front of the lab bench (in line with the analyzer’s peristaltic pump). b. PC width requirement can be reduced to 51 cm (20 in) by placing the mini tower of the pc behind the monitor.

c. The optional ENC-500 enclosure will require 81 cm (32 in) in base depth to allow for the exhaust chimney. Ideally, an unobstructed open island type of bench either 24 or 36 inches deep will suffice.

b

NOTE

Teledyne Leeman Labs highly recommends the ENC-500 for protection against sample contamination and removal of acid gases from the laboratory environment during analysis of sample batches. We also recommend the

ENC-500 for EPA 1631 and EPA 245.7 analysis in conjunction with a clean room, or at a minimum, a clean area dedicated for the preparation and analysis procedures.

4

Choosing a Location

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

Figure 1-1 Example Footprint of QuickTrace

®

M-8000 with ENC-500 enclosure

(with elevating legs)

56 cm (22 in)

Computer not shown

81 cm (32 in)

Figure 1-2 Example Footprint of QuickTrace

®

M-8000 without enclosure

48 cm (19 in)

Computer not shown

60 cm (24 in)

Choosing a Location

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

5

1.5.2 Work Surface Requirements

The analyzer must be placed on a sturdy countertop or table.

NOTE

Do not place the analyzer on a wheeled cart or folding table.

The work surface should be at least 61 cm (24 in) deep.

1.6 General Requirements

 A clean and dedicated hood for standard and sample preparations.

NOTE

Due to the likelihood of accelerated damage from corrosion and dust, locating the analyzer or autosampler in a fume hood with stagnant air will automatically void the warranty.

1.7 Electrical Requirements

NOTE

Do not apply power to the power supply until ready to operate the analyzer.

Place the QuickTrace

®

M-8000 within 1.2 meters (4 feet) of a standard power outlet. The standard configuration requires four power outlets, one each for the QuickTrace

®

M-8000 Mercury Analyzer, autosampler, computer, and monitor.

NOTE

An AC surge protected power strip with six outlets is strongly recommended.

The power cord set supplied with the QuickTrace

®

M-8000 and autosampler meets the requirements of the country where the instrument was purchased.

NOTE

If the instrument is to be used in a country other than the one specified at the time of ordering, obtain a new power cord set that meets the requirements of that country.

6

General Requirements

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

The QuickTrace

®

M-8000 mercury analyzer and autosampler use an auto-switching power supply that accepts AC input from 100-240 VAC ±10%, 50/60Hz. The specifications are shown in Table 1-4 "Voltage and Power Requirements".

Table 1-4 Voltage and Power Requirements

M-8000 Mercury Analyzer

AC Input

8A @ 120 VAC ±10%

4A @ 240 VAC ±10%

50/60 Hz

ASX-520 Rugged Autosampler Total

100 - 240 ±10% VAC

1.9A

50/60 Hz

9.9A @ 120 VAC ±10%

5.9A @ 240 VAC ±10%

50/60 Hz

WARNING

This equipment is designed for connection to a grounded (earthed) outlet.

The grounding type plug is an important safety feature. For continued protection against electrical shock or damage to the instrument, do not disable this feature.

The power requirements for the computer can be found on the label affixed to the bottom of the computer, or in the computer user’s manual.

1.8 Gas Supply Requirements

A source of Ultra-High Purity (99.999%) Argon is required. The gas regulator must be an inert 2-stage design and provide 69 - 689 kPa (10 - 100 psi), with a coupling for either a cylinder or Dewar capable of delivering 241 kPa (35 psi). The M-8000 is supplied with

1.82 meters (6 feet) of 1/8 inch (3.175 mm) OD gas supply tubing. If the regulator is located more than 1.82 meters (6 feet) from the analyzer, additional tubing will be required.

Gas Supply Requirements

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

7

1.9 Ventilation Requirements

1.9.1 Without ENC-500 Enclosure

During operation, the QuickTrace

®

M-8000 internally contains trace amounts of mercury vapor. To prevent inhalation of the vapor, the QuickTrace

®

M-8000 uses a solid

KMnO

4 absorbent trap located on the back of the instrument. This trap absorbs the mercury vapor prior to final exhaust, therefore no extra ventilation is required beyond that of a standard laboratory environment.

WARNING

1.9.2 With ENC-500 Enclosure

Gases exhausting from the QuickTrace® M-8000 cabinet, prior to the external Hg vapor trap (affixed to the rear cabinet panel) contain traces of mercury vapor and must be treated as such. Do not run the QuickTrace®

M-8000 unless exhausted gas is properly “scrubbed” or removed. Fill, maintain, and use the provided KMnO

4

absorbent trap or run a transfer line to a fume hood. It is the responsibility of the laboratory to supply the fill chemical and maintain the integrity of the mercury vapor trap.

NOTE

The ENC-500 enclosure is optional, but is highly recommended for use with the M-8000.

If an optional ENC-500 enclosure is being used, the lab must be equipped with the following:

 A ventilation trunk 11 cm (4 in) in diameter for connecting to the enclosure exhaust chimney.

 The ENC-500 is rated for an exhaust flow with a maximum draw of 453 LPM

(16 CFM) (recommended draw 283 LPM [10 CFM]).

NOTE

The ENC-500 enclosure includes a HEPA filter. The HEPA filter stops particulates from entering the enclosure as it draws room air to remove acid gas during the analysis of digested samples. Particulates can result in sample contamination and the biasing of results during ultra-trace analysis.

1.10 Environmental Conditions

The ambient temperature should be kept as stable as possible. Locating the QuickTrace

®

M-8000 directly in the path of an air conditioner or heater vent may cause baseline drift, and is not recommended.

8

Ventilation Requirements

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

1.11 Necessary Supplies

UHP Argon (Cylinder or Dewar)

99.999% purity Argon

Inert Gas Regulator

Inert 2-stage, 69 - 689 kPa (10 - 100 psi) secondary pressure gauge, with a coupling for either a cylinder or Dewar capable of delivering 241 kPa (35 psi).

NOTE

The M-8000 is supplied with 1.82 meters (6 feet) of 1/8 inch (3.175 mm) OD gas supply tubing. If the regulator is located more than 1.82 meters (6 feet) from the analyzer, additional tubing will be required.

AC Power Strip (surge protected) with six outlets (if sufficient wall outlets are not available).

Certified Mercury Standard Solution

100 or 1000 ppm

Hydrochloric acid (Minimum grade: Trace Metal specification @ 37%)

Recommended: Mallinckrodt/Macron, 6 each - 2.5 L, AR Select, Item # 5587-46

Nitric acid, trace metal grade (68-70%)

Recommended: Mallinckrodt/Macron, 2.5 L, AR (ACS), Item # 2704-46

Stannous chloride (crystals, Di-hydrate)

Recommended: Mallinckrodt/Macron, ACS 500 g, Item # 8176-04

Potassium permanganate; solid, crystalline for the hg exhaust vapor trap or suitable for mercury analysis i.e., EPA 245.1

Recommended: j.t baker 500 g bottle item number 3227-1

Precision adjustable air displacement pipettes, 10 to 100 µl (TD), 200 to 1000 µl

(TD), 400 to 2000 µl (TD), 1000 to 5000 µl (TD), 2000 to 10000 µl (TD)

Replacement tips for pipettes

Disposable plastic pipette droppers

Polypropylene Graduated Cylinders: 10, 25, 50, and 100 mL

Polypropylene or polyethylene small mouth bottle with cap (three each -500 mL)

Additional Chemical Compounds

NOTE

1.11.1

The sample preparation procedures of the intended analytical method may require additional chemical compounds. Refer to published method specifications.

Sources for Premixed 1631 and 245.7 Chemicals

Table 1-5 Premixed 1631 and 245.7 Chemical Sources

Supplier

Environmental Express

 o2si Smart Solutions

Website

http://www.envexp.com/ https://www.o2si.com/

Necessary Supplies

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

9

1.11.2 USEPA Method 1631 Users

Teledyne Leeman Labs recommends the following when using the QuickTrace

® mercury analyzer to perform USEPA Method 1631:

M-8000

 A purged water source.

Purged SnCl

2 for the reduction of Hg.

For optimal performance an in-line gold trap (PN SP8042) for trace Hg removal from the gas supply used to purge the reagents and reagent water should be used. The M-8000 is supplied with a gold trap for Hg removal from the carrier gas.

Hg pre-filter (PN 15-2145-003) for the gas supply when using gold trap modes of operation. This pre-filter should be used in conjunction with in-line gold trap SP8042.

NOTE

Teledyne Leeman Labs recommends using the Hg pre-filter to remove trace amounts of mercury from the instrument carrier gas and the gas used to purge reagents. The filter may also be used for the preparation of ultra-trace purged

DI water.

 A clean room, or at a minimum a clean area, dedicated for the preparation and analysis procedures.

1.12 Recommended Supplies

Volumetric Flasks 100 mL class A (TC) 6 each

Volumetric Flasks 1000 mL class A (TC) 1 each

Volumetric Flasks 2000 mL class A (TC) 1 each

Kimwipes

Weighing Balance, top loading (0.1 g readability or better, capacity > 200 g)

Stopwatch (for measuring liquid uptake rates)

Powder Funnel, wide-bore stem, small overall size

Wrenches, adjustable 12 in, 6 in and 4 in

Screw Drivers:

 1 Small Phillips-Head

1 Medium Phillips-Head

1 Long-Shank Medium Flat-Head

 1 Small Thin Flat-Head

Flow Meter 0-500 mL/min calibrated for argon gas with 1 mL readability

Deionized Water

Soda Lime 4-8 mesh

NOTE

Soda lime is used to preserve the integrity of the gold traps during USEPA

M ethod 1631 analysis. Because EPA 1631 is a per formance -based methodology, soda lime is not a necessary supply and may be eliminated.

10

Recommended Supplies

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

1.13 Installation Solution Preparations

NOTE

The reagent, rinse and standard solutions should be prepared during the unpacking and initial installation period. Teledyne Leeman Labs and customer owned polypropylene bottles and standard tubes should be used.

Table 1-6 Installation Standards, Reagents and Rinse Solutions

a

Solution

Stannous Chloride

Calibration Blank

Autosampler Rinse Solution

Reagent Capillary Rinse Solution

Reagent Capillary Rinse Solution

Preparation

1 L of 10% w/v stannous chloride in 7% v/v HCl

3% v/v HCl

2 L of 3% v/v HCl

500 mL of 10% v/v HNO

500 mL of 10% v/v HCl

Reagent Capillary Rinse Solution 500 mL of DI water

LC (ng/L) Secondary Mercury Standard 100 mL of 10,000 ng/L

3

LC (ng/L) Working Mercury Standard

Calibration Standards

100 mL of 200 ng/L

Shown in Table 1-7 "Standard Concentrations"

a. All bottles and vessels need to be cleaned prior to use.

Table 1-7 Standard Concentrations

EPA Method 1631

Standard Concentration

0.5 ng/L

1.0 ng/L

5.0 ng/L

25.0 ng/L

50.0 ng/L

100.0 ng/L

EPA Method 245.7

Standard Concentration

5.0 ng/L

10.0 ng/L

25.0 ng/L

50.0 ng/L

100.0 ng/L

200 ng/L

0.1 mL

0.2 mL

1.0 mL

5.0 mL

10.0 mL

20.0 mL

200 ng/L

1.0 mL

2.0 mL

5.0 mL

10.0 mL

20.0 mL

Blank 3% HCl

39.9 mL

39.8 mL

39.0 mL

35.0 mL

30.0 mL

20.0 mL

Blank 3% HCl

39.0 mL

38.0 mL

35.0 mL

30.0 mL

20.0 mL

1.14 Preparedness Statement

Failure to meet one or more of the pre-installation requirements may prevent your instrument from operating properly. During installation, if an installation engineer determines that one or more of the pre-installation requirements have not be met, the company reserves the right to delay installation until all requirements are satisfactorily met. Any time lost during installation, caused by failure to meet the pre-installation requirements, will be billed to your account. If you have any questions regarding these

requirements, please contact Teledyne Leeman Labs using Section 1.15 "Contact

Information".

Installation Solution Preparations

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

11

1.15 Contact Information

Teledyne Leeman Labs encourages you to contact us for any questions or concerns regarding your installation, or to order accessories and consumables. Please use the appropriate means of contact listed below for the most efficient service.

1.15.1 Teledyne Leeman Labs

110 Lowell Road

Hudson, NH 03051 U.S.A

Main: 603-886-8400

Fax: 603-886-4322 www.teledyneleemanlabs.com

1.15.2 Sales

US and International: +1 800-634-9942 or +1 603-886-8400

US and International Fax: +1 603-886-4322

E-mail: [email protected]

Visit our website at www.teledyneleemanlabs.com for a complete list of US and

International Sales Representatives.

1.15.3 Technical Support

US: +1 800-533-6267 (1-800-Leemans)

International: +1 603-886-8400

E-mail: [email protected]

1.15.4 Replacement Parts and Consumables

US and International: +1 800-533-6267

E-mail: [email protected]

12

Contact Information

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

1.16 Your Installation is Our Highest Priority

Streamlining the installation of your new QuickTrace

®

M-8000 CVAF mercury analyzer system is our highest priority! Teledyne Leeman Labs would like to thank you in advance for the preparation of your laboratory. Your attention to detail will assist in the installation of your instrument. If you have any questions or concerns we encourage you to contact us directly:

Jeff Forsberg

Teledyne Leeman Labs

Mercury Product Manager [email protected]

(402) 733-2829

Bob McCarthy

Teledyne Leeman Labs

Director Customer Support [email protected]

(603) 521-3267

Your Installation is Our Highest Priority

QuickTrace

®

M-8000 Mercury Analyzer Pre-Installation Guide (Service Installation)

13

14

Your Installation is Our Highest Priority