Krautkramer USM Vision+ User`s Manual 3 MB

Add to my manuals
164 Pages

advertisement

Krautkramer USM Vision+ User`s Manual 3 MB | Manualzz
GE
Measurement & Control
Ultrasonic Inspection
Krautkramer USM Vision+
User’s Manual
110N1532 Rev. 3
July 2015
Krautkramer USM Vision+
Portable Phased Array Ultrasonic Testing
User’s Manual
110N1532 Rev. 3
(Software Version 9.4.1)
July 2015
www.gemeasurement.com
©2015 General Electric Company. All rights reserved.
Technical content subject to change without notice.
[no content intended for this page]
ii
Contents
Chapter 1. USM Vision Overview
1.1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Chapter 2. Initial Startup and Operation
2.1
Instrument Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2.2
Setting up the USM Vision+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.3
Connecting a Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.4
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.5
2.6
2.4.1
Operation Using the External Power Supply . . . . . . . . . . . . . . . . . . . . .6
2.4.2
Operation Using Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2.4.3
Checking the Battery Charge Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2.4.4
Exchanging Batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.4.5
Charging Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2.5.1
USB Sockets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2.5.2
Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
The Trackballs and Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2.6.1
Trackballs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
2.6.2
Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.7
Software Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.8
Preparing the USM Vision+ for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.9
Starting the USM Vision+. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.10 Establishing Base Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.11 Checking System Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.12 Exiting the Software and Shutting Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Chapter 3. Phased Array (PA) Mode
3.1
The Phased Array Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2
The Gates Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.3
Weld Geometry in the Display Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
3.3.1
3.4
The Readings Sub-Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
The File Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3.4.1
The Load/Store Pass Sub-Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
USM Vision+ User’s Manual
iii
Contents
3.5
Color Palette Editor (Amplitude and True Depth). . . . . . . . . . . . . . . . . . . . . . . 40
3.5.1
3.6
Color Rulers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
3.6.1
3.7
User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Amplitude Color Bars . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
3.6.2
Amplitude Color Palette for RF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
3.6.3
Depth Color Bars . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Step Control Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.7.1
Valid Value Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.7.2
Number of Beams in Pattern. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.7.3
Grey Out Logic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.7.4
User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
3.8
Sound Velocity Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.9
Range Trigger: IP Acquisition and IF Gate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
3.10 New Viewer Display Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Chapter 4. Calibration
4.1
Calibrating a 0° Linear Scan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
4.2
Calibrating a 20° Linear Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
4.3
Calibrating a -20° to 20° Sector Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
4.4
Calibrating a 40° to 70° Sector Scan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4.4.1
4.5
4.6
Sensitivity Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
4.5.1
1-point TCG (0° Linear Scan) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
4.5.2
1-point TCG = ACG (40° to 70° Sector Scan) . . . . . . . . . . . . . . . . . . . . 91
Sector Scan TCG Recording . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
4.6.1
4.7
Creating a Dual Sector Scan for Weld Inspection. . . . . . . . . . . . . . . 83
Verifying TCG Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
4.6.2
TCG Amplitude Evaluation Levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
4.6.3
Echo Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Encoder Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Chapter 5. Specifications
5.1
General Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
5.2
I/O Connector (LEMO ECG.2B.314.CLV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125
iv
USM Vision+ User’s Manual
Contents
Appendix A. Creating User Accounts
A.1
Setting up User Accounts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
A.2
Setting up a User’s Access Rights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Appendix B. Calibrating the Touchscreen
B.1
Recalibrating the Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 137
Appendix C. Environmental Compliance
C.1
Waste Electrical and Electronic Equipment Directive . . . . . . . . . . . . . . . . . 141
C.2
Battery Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
C.2.1
What do the Markings Mean? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142
C.2.2
The Risks and Your Role in Reducing Them . . . . . . . . . . . . . . . . . . . 142
Appendix D. Glossary
USM Vision+ User’s Manual
v
Contents
[no content intended for this page]
vi
USM Vision+ User’s Manual
Preface
Important Notice
The information in this section must be read and understood by all users of
GE Measurement & Control ultrasonic test equipment. Failure to follow these
instructions can lead to errors in measurements or other test results. Decisions based
on erroneous results can, in turn, lead to property damage, personal injury or death.
General Warnings
Proper use of ultrasonic test equipment requires three essential elements:
• Selection of the correct test equipment
• Knowledge of the specific “test application requirements”
• Operator training about welds
This operating manual provides instruction in the basic setup and operation of the
GE equipment. There are, however, additional factors which affect the use of
ultrasonic test equipment. Specific information regarding these additional factors is
beyond the scope of this manual. The operator should refer to textbooks on the
subject of ultrasonic testing for more detailed information.
Operator Training
Operators must receive adequate training before using ultrasonic test equipment.
Operators must be trained in general ultrasonic testing procedures and in the setup
and performance required by a particular test. Operators must understand:
• Sound wave propagation theory
• Effects of the velocity of sound in the test material
• Behavior of the sound wave where two different materials are in contact
• Areas covered by the sound beam
More specific information about operator training, qualification, certification, and
test specifications is available from various technical societies, industry groups, and
government agencies.
USM Vision+ User’s Manual
vii
Preface
Testing Limitations
In ultrasonic testing, information is obtained only from within the limits of the
sound beam. Operators must exercise great caution in making inferences about the
test material outside the limits of the sound beam. For example, when testing large
materials, it may be impossible or impractical to inspect the entire test piece.
When a less-than-complete inspection is to be performed, the operator must be
shown the specific areas to inspect. Inferences about the condition of areas not
inspected, based on data from the evaluated areas, should only be attempted by
personnel fully trained in applicable statistical and probability techniques. In
particular, materials subject to erosion or corrosion, in which conditions can vary
significantly in any given area, should only be evaluated by fully trained and
experienced operators.Sound beams reflect from the first interior surface
encountered. Because of part geometry and overlapped flaws or overlapped
surfaces, thickness gauges may measure the distance to an internal flaw rather than
to the back wall of the material. Operators must take steps to ensure that the entire
thickness of the test material is being examined.
Operators must be familiar with the use of ultrasonic couplants. Testing skills must
be developed so that couplant is used and applied in a consistent manner to
minimize variations in couplant layer thickness and errors in test results. Calibration
and actual testing should be performed under similar coupling conditions, using a
minimum amount of couplant and applying consistent pressure on the transducer.
viii
USM Vision+ User’s Manual
Preface
Safety Information
ATTENTION! This instrument is designed only for materials
testing. Any use for medical applications or other purposes is
not allowed.
This instrument may only be used in industrial environments.
This instrument can be operated with batteries or while
plugged into an electrical outlet using the AC charger. The
power supply unit has the electrical safety class II.
Only authorized personnel should open the unit.
This product is not rated for use in an explosive atmosphere /
environment.
Use caution when using the harness while climbing — there is a
risk of strangulation.
The neck strap is not intended to be used when climbing with
the instrument.
If a support stand closes on a user’s fingers, it can cause injury.
Software
According to the current state of the art, software is never completely free from
errors. Before using any software-controlled test equipment, please make sure that
the required functions operate perfectly in the intended application.
USM Vision+ User’s Manual
ix
Preface
Defects, Errors and Exceptional Stresses
If you have reason to believe that safe operation of your instrument is no longer
possible, you have to disconnect the instrument and secure it against unintentional
re-connection. Remove the batteries if necessary.
A safe operation is no longer possible if:
• The instrument shows visible damage
• The instrument no longer operates perfectly
• The instrument has been subjected to prolonged storage under adverse
conditions like exceptional temperatures, especially high air humidity or
corrosive environmental conditions
• The instrument has been subjected to heavy stresses during transportation
x
USM Vision+ User’s Manual
Preface
Battery Safety Information
ATTENTION! The power for this ultrasonic instrument can be
supplied by lithium-ion batteries. Read these safety
instructions and the product operating manual carefully.
Do not open or dismantle batteries.
Do not expose batteries to heat above 80°C or fire. Avoid
storage in direct sunlight.
Do not short-circuit a battery.
Do not store batteries haphazardly in a box or drawer where
they may short-circuit each other or be short-circuited by other
metal objects.
Do not remove a battery from its original packaging until
required for use.
Do not subject batteries to mechanical shock.
In case of a battery leaking, do not allow the liquid to come in
contact with the skin or eyes. If contact has been made, wash
the affected area with copious amounts of water and seek
medical advice.
Charge only with the charger provided with the equipment.
Follow the instructions in the operating manual for inserting
the batteries into the instrument, and the note indicating the
charging of the batteries in the instrument.
Observe the plus (+) and minus (-) marks on battery and
equipment and ensure correct use.
Do not mix batteries of different manufacture, capacity, size or
type within this instrument.
Keep batteries out of the reach of children.
Keep batteries clean and dry.
USM Vision+ User’s Manual
xi
Preface
Battery Safety Information (cont.)
ATTENTION! The power for this ultrasonic instrument can be
supplied by lithium-ion batteries. Read these safety
instructions and the product operating manual carefully.
Use the battery only in the application for which it was
intended.
When possible, remove the battery from the equipment when
not in use.
Do not store batteries longer than 1 month in discharged state.
Do not store batteries longer than 6 month without recharge.
The battery must be recycled or disposed of properly, according
to the national and local regulations. If you have any questions,
contact your local dealer.
The batteries must be disposed of only in the discharged state
to the collection point. In case of not fully discharged batteries,
there is a short-circuit risk. Short-circuits can be prevented by
isolation of contacts with adhesive tape.
FCC Compliance Statement
This device complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions:
1.
This device may not cause harmful interference.
2.
This device must accept any interference received, including interference that
may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at his
own expense.
xii
USM Vision+ User’s Manual
Preface
Service
The ultrasonic flaw detector USM Vision+ is manufactured by:
GE Sensing & Inspection Technologies GmbH
Robert-Bosch-Straße 3
50354 Hürth
Germany
T +49 (0) 22 33 601 111
F +49 (0) 22 33 601 402
The USM Vision+ is manufactured according to the state-of-the-art methods using
high-quality components. Thorough in-process inspections or intermediate tests and
a quality management system certified to DIN EN ISO 9001 ensure an optimum
quality of conformance and workmanship of the instrument.
Should you nevertheless detect an error on your instrument, power the instrument
off and remove the batteries. Inform your local GE customer service and support,
indicating the error and describing it.
Keep the shipping container for any repairs possibly required which cannot be made
on the spot.
If there is anything special that you would like to know about the use, handling,
operation, and specifications of the instruments, please contact your nearest GE
representative or contact one of the service centers listed on the rear cover of this
manual.
USM Vision+ User’s Manual
xiii
Preface
Typographical Conventions
Note: These paragraphs provide additional information about the topic which is
helpful but is not essential to proper completion of the task.
Important: These paragraphs provide emphasis to instructions that are essential
to proper setup of the equipment. Failure to follow these instructions
carefully may cause unreliable performance.
WARNING! These paragraphs indicate a potentially hazardous
situation which can result in serious personal injury or death, if it is
not avoided.
CAUTION! These paragraphs indicate a potentially hazardous
situation which can result in minor or moderate injury to
personnel or damage to the equipment, if it is not avoided.
xiv
USM Vision+ User’s Manual
Chapter 1. USM Vision Overview
Chapter 1. USM Vision Overview
1.1 Introduction
The USM Vision+ phased array inspection system (see Figure 1 below) combines
phased array and conventional ultrasonic testing (UT) for imaging testing results
and encoded data recording plus on-screen evaluation. The system works in 16/128
configuration; that is, it offers 16 physical channels, where, in multiplex mode,
probes with a maximum of 128 elements can be driven. An intuitive menu structure
and layout simplify operation, even when multiple groups were chosen (up to a
maximum of 12).
Users enter data via either a touchscreen, or through six soft buttons (F1-F6) located
below the screen, and two trackballs.
Figure 1: The USM Vision+
Note: This manual is subject to revision to reflect product updates and additions.
Please contact your local sales agent or visit our web page,
www.ge-mcs.com, to find the most recent revision.
USM Vision+ User’s Manual
1
Chapter 1. USM Vision Overview
[no content intended for this page]
2
USM Vision+ User’s Manual
Chapter 2. Initial Startup and Operation
Chapter 2. Initial Startup and Operation
2.1 Instrument Overview
Figure 2 below outlines the controls available on the USM Vision+ front panel.
2
1
3
7
6
8
4
5
Figure 2: USM Vision Front Panel
Table 1: Key to USM Vision+ Front Panel (see Figure 2 above)
Number
Front Panel Control
1
Trackball with two keys to control the focus in the user interface
2
Touch-sensitive screen (touchscreen), for direct operation of the
graphic interface
3
Trackball with two keys, for direct access to values in controls having
focus
4
Function keys, assignment programmable via software
5
Power key (for switching the meter ON and OFF)
6
Gain increase and decrease keys
7
Image freeze
8
Data storage, report generation
USM Vision+ User’s Manual
3
Chapter 2. Initial Startup and Operation
2.1 Instrument Overview (cont.)
1
2
6
5
4
3
Figure 3: USM Vision Back Panel
Table 2: Key to USM Vision+ Back Panel (see Figure 3 above)
Number Front Panel Control
4
1
LEMO 00 COAX sockets, for conventional UT use in single and T/R
mode
2
PC interfaces: Ethernet, USB, Lemo 0S for connecting VGA and
Power Supply
3
Prop-up stand and handle, for transportation and inclined installation
4
Battery compartment at the bottom,
for holding one or two lithium-ion batteries
5
Phased array probe connector
6
I/O connector, encoder input
USM Vision+ User’s Manual
Chapter 2. Initial Startup and Operation
2.2 Setting up the USM Vision+
The USM Vision+ has a prop-up stand and handle at the rear that locks into place in
different positions.When completely folded out, it becomes a transport handle for
the instrument. You can set up the USM Vision+ at different angles to have the best
possible view of the screen.
Important: Place the USM Vision+ on a stable flat surface. When operating on a
stable flat surface, make sure that the stand is secure, and do not
touch the instrument on its back. Due to the applied force, the prop-up
stand might move and could either snap to the back or the instrument
could drop on the surface. When opening the handle, be sure to grasp
it firmly to avoid unwanted movements.
The device is intended for indoor or outdoor use. Select a suitable location for
installation that guarantees meeting the environmental conditions. The ambient
temperature must be between 0 and +45°C. The relative humidity must not exceed
95%.
Because the USM Vision+ generates heat during operation, ensure that there is
adequate ventilation and enough clearance between the unit and heat-sensitive
objects or equipment.
Avoid direct heat, heat accumulation and overheating by direct sunlight or other
heat sources. Ensure that there is adequate and unhindered circulation of air.
Ensure that no dirt or only dry, nonconducting dirt accumulates on the instrument,
in particular at the connectors.
The following condition must be met for safe operation:
• No iron or steel dust must penetrate into the instrument, in particular at the
connectors. Apply protective caps on any connectors that are not in use.
USM Vision+ User’s Manual
5
Chapter 2. Initial Startup and Operation
2.3 Connecting a Probe
A wide range of probes manufactured by GE may be used in combination with the
USM Vision+, provided the suitable connecting cable is available.The connector
sockets for one or more probes are located on the top side of the USM Vision+.
When connecting probes having only one transducer element, you can use either
connector socket. The connectors for transmitter and receiver are marked with a red
ring (= receiver) and a black ring (= transmitter).
2.4 Power Supply
the USM Vision+ can be powered using an external power supply or by up to two
lithium-ion batteries. You can connect the USM Vision+ to the mains supply system
when the battery is in the instrument. A discharged battery will be charged in this
situation, while the instrument remains in operation.
The power supply unit provided is intended only for indoor operation.
2.4.1
Operation Using the External Power Supply
The external power supply automatically adjusts to any AC input voltage between
90 V and 240 V nominal voltage.
Connect the USM Vision+ to the mains socket-outlet using the external power
supply with the appropriate power cable. The socket connector is under the cover on
the top side of the USM Vision+. Proceed as follows:
• Loosen the knurled screw in the cover on the top side of the instrument and
open the cover completely.
• Plug the Lemo connector of the external power supply unit into the socket
connector +15V until it snaps into place with a clearly-audible click.
• When pulling off the Lemo plug, withdraw the metal bushing on the plug
first in order to release the lock.
• When transporting the USM Vision+, always shut the cover and fasten the
knurled screw.
Note: When powering from mains operation, it is recommended that you use a
battery in the USM Vision+ in order to prevent a system failure and data
loss in case of a power failure.
6
USM Vision+ User’s Manual
Chapter 2. Initial Startup and Operation
2.4.2
Operation Using Batteries
Use either one or preferably two lithium-ion batteries for the optional battery
operation mode. These batteries each have a high capacity. Consequently, two
lithium-ion batteries ensure a long operating time for the instrument.
2.4.2a
Inserting Batteries
The battery compartment is located at the bottom of the instrument and the cover is
fixed with one knurled screw. To install a battery, proceed as follows:
• Loosen the knurled screw in the cover on the bottom side of the instrument.
• Fold the cover down, and you will see two battery compartments.
• Insert a battery into either the left-hand or the right-hand battery
compartment. Be sure to position the battery so that the contacts point back
and downwards.
• Close the cover and fasten the knurled screw.
2.4.3
Checking the Battery Charge Level
The lithium-ion battery is equipped with a battery charge indicator, which is located
on the front of the battery. Five LCD segments indicate the battery charge level,
with the number of filled LCD segments showing the charging level as follows:
• 5 segments – battery charge level 100... 81%
• 4 segments – battery charge level 80... 61%
• 3 segments – battery charge level 60... 41%
• 2 segments – battery charge level 40... 21%
• 1 segments – battery charge level 20... 1%
USM Vision+ User’s Manual
7
Chapter 2. Initial Startup and Operation
2.4.4
Exchanging Batteries
Important: If you remove both batteries during operation and the instrument is
not connected to the mains supply, all unsaved data will be lost.
You can exchange one battery during operation as follows:
• First, insert a fully charged battery into an empty compartment.
• Then, remove the other battery.
2.4.5
Charging Batteries
You can charge the lithium-ion battery either directly in the instrument or by means
of an external battery charger. Several batteries are charged in succession.
If there is a battery in the instrument, the charging process starts automatically as
soon as you connect the external power supply. You can carry out ultrasonic tests
and charge a battery at the same time.
The charging time is approximately 6 hours per battery. Charging time is
independent of operation. The charging time applies to ambient temperatures from
25 °C to 30°C. Please keep in mind that the batteries are not charged to their full
capacity at high temperatures.
Charging lithium-ion batteries is also possible with the external battery charger
recommended and provided by GE Measurement & Control.
Note: You will find information on how to handle the external battery charger in
the documentation supplied with the charger.
8
USM Vision+ User’s Manual
Chapter 2. Initial Startup and Operation
2.5 Interfaces
Three USB sockets, a VGA socket and one network interface are located behind the
cover on the top side of the USM Vision+.
2.5.1
USB Sockets
The three USB Type A sockets offer multifunctional connection possibilities for
items such as a mouse, keyboard, printer or external data carriers.
Important: Do not use the USB Type B connector, as it may damage the USM
Vision+ or your computer.
Note: You will find information on the relevant software in the documentation
provided by the device manufacturer.
2.5.2
Network
For data transfer, you can connect the USM Vision+ with to Ethernet network. The
basic network functions are already set up, but have to be adapted and configured to
specific local network requirements.
2.6 The Trackballs and Touchscreen
2.6.1
Trackballs
The USM Vision+ features two trackballs to operate the graphical software
interface. The two trackballs work independently of each other.
Each trackball has a large button and a small button. Buttons on both sides of the
trackball have the same function. The big button confirms any selection or performs
the action of the button having the focus, while the small button cancels an action.
You can operate both trackballs simultaneously, such as moving the focus with the
right trackball while triggering functions with the buttons of the left trackball.
USM Vision+ User’s Manual
9
Chapter 2. Initial Startup and Operation
2.6.2
Touchscreen
The USM Vision+ is equipped with a touchscreen, enabling direct operation of the
menus appearing on the screen. This operation by direct touch replaces operation by
means of a mouse (selecting and clicking). A mouse pointer is therefore not
necessary. To select or to mark an element of the user interface, just touch the
corresponding point of the screen briefly.
CAUTION! Do not touch the touchscreen with any hard or
sharp-edged objects (e.g., ball-point pen or screwdriver). They
might severely damage the touch-sensitive surface. Do not apply
high pressure to the screen, as the touchscreen needs only slight
pressure to react.
2.7 Software Installation
The USM Vision+ is delivered completely configured and ready for operation.
2.8 Preparing the USM Vision+ for Use
When you unpack the USM Vision+, carefully remove the instrument, the probe,
the power supply and the cables from the shipping containers. Before discarding
any of the packing materials, account for all components and documentation listed
on the packing slip. If anything is missing or damaged, contact GE Measurement &
Control immediately for assistance
After you first receive the USM Vision+, you must insert the batteries into the
battery compartment on the bottom of the unit. You must then charge the batteries,
using the power supply included with the unit. When you start the USM Vision+,
you will need a user name and password from your system administrator to log onto
the system. Before beginning an inspection, you must also import an inspection
plan into the unit from a linked PC or USB memory stick and attach the probe to the
port at the top of the USM Vision+.
10
USM Vision+ User’s Manual
Chapter 2. Initial Startup and Operation
2.9 Starting the USM Vision+
To start the USM Vision+, press the Power button in the lower right corner of the
front panel (see Figure 2 on page 3). The instrument is switched ON and the USM
Vision+ software boots up.The initialization screen appears similar to Figure 4
below.
Figure 4: USM Vision Initialization Screen
Once loaded, the screen displays two security levels: Inspector and Administrator
(complete access), as shown in Figure 5 on page 12. For additional entry levels
please refer to Appendix A of this manual.
USM Vision+ User’s Manual
11
Chapter 2. Initial Startup and Operation
2.9 Starting the USM Vision+ (cont.)
Figure 5: Login Screen
When you press the Admin option, you will need to enter a password, as shown in
Figure 6 below. If you press the Inspector option, you will be asked to enter your
user name via the keyboard.
Figure 6: Entering the Password
12
USM Vision+ User’s Manual
Chapter 2. Initial Startup and Operation
2.9 Starting the USM Vision+ (cont.)
To enter the password, press the keyboard icon in the upper left corner of the screen.
The onboard keyboard opens (Figure 7 below).
Figure 7: The On-screen Keyboard
The default password is “1234”. (To change the user password, refer to Appendix
A.) Once the system has confirmed your password, the Windows screen opens
(Figure 8 below) providing access to all functions and applications.
Figure 8: The Windows Desktop Screen
USM Vision+ User’s Manual
13
Chapter 2. Initial Startup and Operation
2.9 Starting the USM Vision+ (cont.)
To enter the USM Vision+ application, press on the icon “EchoLoader”. The next
screen is the USM Vision+ boot screen.
Figure 9: USM Vision+ Boot Screen
Once confirmed, you will view the boot screen shown in Figure 9, then go to the
option selection screen in Figure 10 below. The available options are PA (phased
array) and conventional. This manual will describe all functions for the PA
application (see Chapter 3); subsequent revisions will cover future applications.
Figure 10: Option Selection Screen
14
USM Vision+ User’s Manual
Chapter 2. Initial Startup and Operation
2.10 Establishing Base Settings
To establish base settings (language, date and time, measurement units) for the
USM Vision+, press the Main Menu button and then the Settings option (Figure 11
below).
Figure 11: The Main Menu Button
Then press General. The General Settings option (shown in Figure 12 below)
opens.
Figure 12: The General Settings Window
USM Vision+ User’s Manual
15
Chapter 2. Initial Startup and Operation
2.10 Establishing Base Settings (cont.)
• To change the language, press the Language drop-down menu and then press the
desired language: English, Japanese, Chinese, Portuguese, Deutsch (German),
Italian, Francais (French), Espanol (Spanish), Russian and Arabic.
Note: The first release will have only the English and German GUIs; GUIs for
other languages will follow shortly.
• To change the date, press the Month drop-down menu and select the current
month. For the Day and Year, press the associated drop-down menu and use the
keypad to enter the day.
• To change the time, press the Hour or Minute drop-down menu. Then use the
associated keypad to enter the hour (up to 23) and the minute (up to 59).
• To select the unit of measurement, press the Unit drop-down menu. You can
choose either millimeters or inches.
• To select the decimal point style, press the Decimal option button for a period (.)
or a comma (,).
• To alter the color of the screen, press the Color Scheme drop-down menu. You
can select from standard, light and dark options.
• The Startup Option determines the startup of the conventional channel. You can
select either the default setting or the startup with the last used setting.
When you have completed entering settings, press Close. the USM Vision+ returns
to the Main Menu.
16
USM Vision+ User’s Manual
Chapter 2. Initial Startup and Operation
2.11 Checking System Information
The System Information screen contains basic data about your USM Vision+, such
as serial number, hardware and software version, and the most recent calibration
date. To access this data from the Main Menu, press Settings and then System
Information. The screen appears similar to Figure 13 below. Press Close to return
to the Main Menu.
Figure 13: System Information Window
Note: If you are contacting GE for service, please have all system information
data ready to simplify remote assistance.
USM Vision+ User’s Manual
17
Chapter 2. Initial Startup and Operation
2.12 Exiting the Software and Shutting Down
Important: Always exit the software first before shutting down the instrument. Not
following this sequence may result in a loss of data.
From the Main Menu, you have two options to close the USM Vision+: Logout or
Shutdown.
• Selecting Shutdown will shut down the entire system.
• Selecting Logout will log off the current user.
Note:
To switch off the USM Vision+ without shutting down the software or the
operating system, keep the Power button on the front panel depressed for at
least 4 seconds.
Switching off without shutting down the software may be required as an extreme
exception to this procedure (e.g., if the instrument no longer responds). In this case,
all unsaved data is lost.
18
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
Chapter 3. Phased Array (PA) Mode
3.1
The Phased Array Menu
Figure 14 below illustrates the phased array menu.
Figure 14: The Phased Array Menu
The available menus are presented sequentially in this chapter, but you may proceed
directly to any specific menu. To do so, see the Menu Map in Figure 15 on page 20
and follow the instructions in the appropriate section:
USM Vision+ User’s Manual
19
20
Electronic Scan
Aperture
Delay Law
Colors
Scan Setup
Import/Delete
Views
Encoder Cal.
Load/Store
Pass
Display
Scan
File
Close
Delay - Record
Delay - Setup
Calibration
Gates
Pulser
Curvature
Range
Geometry
UT
Wedge
Probe
Run
Cursors
TCG - Setup
Groups
Receiver
Weld Overlay
TCG - Record
Sub menus
Weld
Geometry
TCG - Verify
Readings
TCG – Levels
3.1
Part
Main
Chapter 3. Phased Array (PA) Mode
The Phased Array Menu (cont.)
Figure 15: PA Mode Menu Map
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
3.1
The Phased Array Menu (cont.)
Table 3: Menu Layout and Readings
Main
Menu
Sub Menu
Function
Values
Dim
Comment
PART
Material Vel. 100 - 15000 m/s,
m/s
in/ms
Currently used sound
velocity of the test object
Thickness
Thickness of the test object
mm, in
Shape
flat, convex,
concave
Shape of the test object
Beam into
curv.
no, yes
Sound beam directed to the
direction of curvature
Outer Diam.
Calc Delay
Law
mm, in
execute
Outer diameter of the test
object
Start the delay law
calculation
PROBE
Probe Name
alpha
num.
Name of the probe in use
Probe SN
alpha
num.
Serial number of the probe
in use
Probe Freq.
0.5 - 20
MHz
Frequency of the probe in
use
No. of
Elements
1 - 128
integer
Element count of the probe
in use
Pitch
mm, in
Element pith of probe in use
Elevation
mm, in
Element length of the probe
in use
USM Vision+ User’s Manual
21
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
Values
Dim
Comment
alphanum.
Name of the wedge / delay
in use
Wedge Angle 0 - 45
°
Wedge angle
Wedge
Velocity
m/s,
in/ms
longitudinal sound velocity
in the wedge / delay
Wedge Front
mm, in
Distance from center of the
array to the wedge (probe)
front
Z-Offset
mm, in
Distance from center of the
array to the coupling surface
WEDGE
GEOMETRY Wedge Name
500 - 6000
m/s
1st Elem.
Pos.
Position of the 1st element
CURVATURE Wedge Name
Shape
alphanum.
flat, convex,
concave
Radius
22
Name of the wedge / delay
in use
Shape of the wedge / delay
mm, in
Radius of the wedge / delay
in primary axis
Beam Dir.
To left, to
right
Direction of the beam
Calc Delay
Law
execute
Start the delay law
calculation
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
UT
RANGE
Range Start
Start of the range (A-scan
and frame)
Range Mode Auto,
Manual
automatic range setting
related to no. of legs
Values
Range End
PULSER
mm, in
Comment
End of the range (A-scan
and frame)
Legs
1 - 10
Number of legs to be
displayed in the frame
Range
Trigger
IP, IF
Trigger of range and gates
form IP of IE
Pulser
Voltage
30 - 150
IP Width
Mode
Auto,
Manual
Pulser Width 30 - 1260
RECEIVER
Dim
PRF Mode
Auto,
Manual
PRF
15 - 10000
Range
Trigger
IP, IF
volts
Pulse width automatically
calculated form the probe
frequency
ns
Pulse width
Pulse repetition frequency
automatically calculated
Hz
Pulse repetition frequency
Trigger of range and gates
form IP of IE
Rectification full, pos,
neg, RF
Filter
Voltage of the initial pulse
Signal rectification
MHz
Frequency filter
Video Filter
off, on
Signal smoothing
Auto 80%
execute
adjust gain to get echo in
gate A to 80% FSH
USM Vision+ User’s Manual
23
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
DELAY
LAW
Sub Menu
Function
Values
APERTURE
Start
Element
1 - 128
Defines the 1st element to
be used in current setup
Aperture
1 - 16
Number of elements used in
the virtual probe
Focal Depth
Dim
mm, in
Comment
Depth of the focal point
Pin Offset
0 - 127
Physical pin number for first
element in current group
Calc Delay
Law
execute
Start the delay law
calculation
Sector
linear
TR
TOFD
Defines type of electronic
scan
Dual mode
Through transmission mode
for TOFD
ELECTRONIC Type
SCAN
Sector: Angle Start
linear: Angle
TR:
TOFD: Angle
-86
'-86 to 86
Sector: Angle Stop
linear:
TR: Receiv.
TOFD Offset
Receiv.
Offset
86
°
Beam angle (start)
°
Beam angle stop
°
Angle increment from shot
to shot
with TR-probes the order of
elements
'-86 to 86
1 - 127
1 - 127
Sector: Angle Step
0.1 - 5
linear:
TR: Receiv. Order parallel,
turned
TOFD:
24
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
Values
Sector:
linear:
TR:
TOFD PCS
Calc Delay
Law
GROUPS
Dim
Comment
mm/in
Probe Center Separation
execute
Group Name
Start the delay law
calculation
alpha
num.
Name of the current group
Group
Display
current, all
Type of display for more
than one group
Copy
execute
Copy current group to
create a new one with
similar parameters
Rename
alpha
num.
Rename duplicated group
Delete
execute
delete current group
Gate Select
A, B, I
Select gate for parameter
change
GATES
Gate Start
mm, in
Start of selected gate
Gate Width
mm, in
Width of selected gate
Gate Thresh. 0 - 95
%
Threshold of selected gate
TOF Mode
Peak, Flank,
J-Flank
TOF measurement mode of
selected gate
Gate Logic
off, positive,
negative
Logic of selected gate
USM Vision+ User’s Manual
25
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
Values
Dim
Comment
Reference
Refl.
Depth,
Radius, SDH
Type of reference reflector
to be used for calibration
SDH Dia.
1 - 25
Diameter of the SDH, if
selected
CALIBRATION
DELAY
SETUP
Ref. Distance
mm, in
Sound path or depth of
selected reference reflector
Tolerance
mm, in
wanted calibration
tolerance
Clear Cal.
DELAY
RECORD
26
execute
delete an existing
calibration
Gate A Start
mm, in
Start of selected gate
Gate A Width
mm, in
Width of selected gate
Tolerance
mm, in
wanted calibration
tolerance
Start
(Record)
execute
Start the calibration
procedure, or record the
reference signals
Store
execute
Store the calibration values
Clear Cal.
execute
delete an existing
calibration
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
Values
TCG SETUP
Reference
Refl.
Depth, SDH
SDH Dia.
Reference
Ampl.
10 - 100
Tolerance
TCG
RECORD
Dim
Comment
Type of reference reflector
to be used for calibration
mm, in
Diameter of the SDH, if
selected
%
Wanted screen height of the
reference reflector
mm, in
wanted calibration
tolerance
Clear TCG
execute
Delete an existing TCG
calibration
TCG
execute
Toggle TCG ON/OFF
Reference
No.
1 - 16
integer
current number of the TCG
reference (point)
Target Depth
mm, in
depth or sound path of the
current reference reflector
Gate A Start
mm, in
Start of selected gate
Beam Section
°
Angular range for TCG
recording
Start
(Record)
execute
Start the TCG recording
procedure, or record the
reference amplitudes
End (Store)
execute
finalize the TCG recording
and store the TCG
parameters
USM Vision+ User’s Manual
27
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
Values
TCG VERIFY Check TCG
Finish
execute
Beam Angle
Beam Select
TCG Curve
28
Dim
Comment
Start TCG verification
procedure (envelope of TCG
references) or finish the
verification
°
integer
off, on
Shot number or angle for
TCG verification
TCG level display
TCG LEVELS TCG Level 1
dB
additional TCG line
TCG Level 2
dB
additional TCG line
TCG Level 3
dB
additional TCG line
TCG Level 4
dB
additional TCG line
TCG Curve
off, on
display TCG lines or not
Transfer
Corr.
dB
apply transfer correction
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
Values
Dim
Comment
VIEWS
Beam Angle, relates to
Beam Select Angle Start
and Stop
Shot number or beam angle
Views
A+E/S, E/S,
A+E/S+C,
A+E/S+B,
A+E/S+B+C
selects screen views, e.g.,
A-Scan, A-Scan + E- or
S-Scan, etc.
View
Correction
angle corr./
volume corr.
type of display for inclined
scanning in the electronic
scan (frame)
DISPLAY
Display Gates no, yes
Gate display (still active,
even when switched off)
Data Source
Gate A, B,
Acquisition
Data to be used for scanning
Data Type
AMP, TOF
Type of data to be used with
scanning
COLORS
Amp Palette
Color palette for echo
amplitudes
TOF Palette
Color palette for TOF values
Cursor
Display
USM Vision+ User’s Manual
show, hide
Display of the measurement
cursors
29
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
Values
CURSORS
View Select
Top, Side,
Frame
Dim
Select view for image
evaluation
Beam Angle relates to
Beam Select Angle Start
and Stop
Shot number or beam angle
Top+Side Cursor X1
Side+Frame Cursor Y1
mm, in
X = mechanical scanning
coordinate
Y = transverse coordinate
(parallel to primary axis)
Frame+Side Cursor Z1
Top+Frame Cursor Y1
mm, in
Z = depth coordinate
Top+Side Cursor X2
Side+Frame Cursor Y2
mm, in
Frame+Side Cursor Z2
Top+Frame Cursor Y2
mm, in
WELD
OVERLAY
Show
Overlay
yes, no
Display overlay in the
electronic scan (Frame)
Origin Offset
X
mm, in
Start position of the
mechanical scan
Origin Offset
Y
mm, in
Distance between Y-origin
(e.g. weld center) and front
of probe
Flip Weld
Side
execute
Change probe position to
the other side of the weld
Beam Dir.
to Left,
to Right
Direction of the beam
Weld Type
30
Comment
V, X, J
Type of the weld to be
overlaid
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Dim
Comment
WELD
Dimension A
GEOMETRY
mm, in
specific weld parameter
Dimension B
mm, in
specific weld parameter
Dimension C
mm, in
specific weld parameter
Dimension D
mm, in
specific weld parameter
Dimension E
mm, in
specific weld parameter
Dimension F
mm, in
specific weld parameter
READINGS
Function
Reading 1
Values
see reading
list
Reading no. 1 to be
displayed on screen
Reading 2
Reading no. 2 to be
displayed on screen
Reading 3
Reading no. 3 to be
displayed on screen
Reading 4
Reading no. 4 to be
displayed on screen
Reading 5
Reading no. 5 to be
displayed on screen
Reading 6
Reading no. 6 to be
displayed on screen
USM Vision+ User’s Manual
31
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
Values
ENCODER
CAL.
Scan Mode
timed,
positional
Dim
Comment
SCAN
Type of mechanical
scanning
Encoder Dir. clockwise,
counter
clockwise
Count direction of wheel
encoder
Encoder
Counts
mm/
Tick,
in/Tick
Distance between two
encoder ticks
Scan
Increment
mm, in
Distance between two
consecutive data
acquisitions with encoded
scanning
Cal. Distance
mm, in
Distance for encoder
calibration
Start
Calibrate
Stop
Calibrate
SCAN SETUP Scan Mode
Scan vs.
Array
timed,
positional
Type of mechanical
scanning
Perpendic-ul
ar, parallel
mechanical scan
perpendicular or parallel to
primary (array) axis
Origin Offset
X
mm, in
Start position of the
mechanical scan
Scan Length
mm, in
Length of current
mechanical scan
Views
32
Start or stop encoder
calibration
A+E/S, E/S,
A+E/S+C,
A+E/S+B,
A+E/S+B+C
selects screen views, e.g.,
A-Scan, A-Scan + E- or
S-Scan, etc.
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
Sub Menu
Function
Values
RUN
Scan Start
Scan Stop
execute
Start or stop the mechanical
scan
Auto Name
off, on
use given name with
sequential no. with
consecutive storage of scans
or setups
Store
execute
Start storage
Store as
execute
Start store a file with name
editing
Flip Weld
Side
execute
Change probe position to
the other side of the weld
Clear Data
execute
clear current scan data
USM Vision+ User’s Manual
Dim
Comment
33
Chapter 3. Phased Array (PA) Mode
Table 3: Menu Layout and Readings (cont.)
Main
Menu
FILE
Sub Menu
Function
Values
LOAD/
Name
STORE PASS
IMPORT/
DELETE
Comment
alphanum.
Group (file) name
Auto Name
off, on
use given name with
sequential no. with
consecutive storage of scans
or setups
Store
execute
Start storage
Store as
execute
Start store a file with name
editing
Load
execute
Load selected file
Delete
execute
Delete selected file
Name
Group (file) name
Export One
execute
export selected file
Export All
execute
export all files containing
mechanical scan data
Target Drive
34
Dim
specifies target drive
Eject
execute
ejects specified drive
Import
execute
imports all files containing
mechanical data form target
drive to the system memory
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
3.2
The Gates Menu
A typical screen for the Gates Menu is shown in Figure 16 below.
Figure 16: Gates Menu
The available options for this menu are listed in Table 4 below.
Function
Gate Select
Gate Start
Gate Width
Gate Thresh.
TOF Mode
Gate Logic
USM Vision+ User’s Manual
Table 4: Gates Menu Options
Description
Select desired gate
Sets gate start value
Sets gate width value
Sets gate threshold
Selects point for TOF measurement
Selects the way of gate threshold violation
35
Chapter 3. Phased Array (PA) Mode
3.3
Weld Geometry in the Display Menu
The weld geometry is set according to the type of weld, as listed in Table 5 below.
Figure 17 below shows examples of the three weld types, with the locations of the
dimensions.
Table 5: Weld Geometry
Dimension Dimension Dimension Dimension Dimension Dimension
A
B
C
D
E
F
Single V
Root
Height
—
Half Root
Width
Half Cap
Width
—
—
—
Top Height
Bottom
Height
Half Top
Cap Width
Half Root
Width
Half
Bottom
Cap Width
Prep. Angle
Root
Height
Half Root
Width
Shoulder
Width
Radius
—
Double V
J Weld
Single V
Double V
J Weld
Figure 17: Weld Types
36
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
3.3.1
The Readings Sub-Menu
A typical screen for the Readings sub-menu is shown in Figure 18 below.
Figure 18: Display-Readings Sub-Menu
You can select up to six readings (each from 20 parameters) for display on the
screen. The available options for this menu are listed in Table 6 below.
Reading
Table 6: Options for Readings Sub-Menu
Description
None
Reading empty
Selected Beam
Beam number
Amp A (%)
Amplitude of echo in Gate A in % FSH
Amp A to TCG (%)
Amplitude of echo in Gate A in % compared to TCG
Amp A to TCG (dB)
Amplitude of echo in Gate A in dB compared to TCG
Sound Path A
Sound Path of echo in Gate A
Surf. Dist. A
Surface (projected) distance of echo in Gate A, relative to Origin
Offset Y
Corr. Depth A
Corrected depth relating to echo in gate A (part thickness
considered)
USM Vision+ User’s Manual
37
Chapter 3. Phased Array (PA) Mode
Reading
Table 6: Options for Readings Sub-Menu (cont.)
Description
Uncorr. Depth A
Uncorrected depth relating to echo in gate A
Amp B(%)
Amplitude of echo in Gate B in % FSH
Sound Path B
Sound Path of echo in Gate B
Surf. Dist. B
Surface (projected) distance of echo in Gate B, relative to Origin
Offset Y
Corr. Depth B
Corrected depth relating to echo in gate B (part thickness
considered)
Uncorr. Depth B
Uncorrected depth relating to echo in gate B
S-Diff B-A
Sound path difference between echoes in gate A and B
Amp A/B (dB)
Ratio of amplitudes in Gate A and B in dB
Amp I (%)
Amplitude of echo in gate I (Interface gate)
Scan Pos.
Scan position
Scan Pos. (°)
Scan position related to circumference of a round object (rod,
pipe)
Encoder counts
actual encoder value
Missing Lines (%)
Missing scan lines in a scan recording
38
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
3.4
The File Menu
3.4.1
The Load/Store Pass Sub-Menu
A typical screen for the Load/Store Pass sub-menu is shown in Figure 19 below.
Figure 19: File—Load/Store Pass Sub-Menu
The available options for this sub-menu are listed in Table 7 below.
Function
Table 7: Load/Store Pass Sub-Menu Options
Description
Name
Name of currently loaded file; opens data base to search for the
wanted file
Auto Name
Selects manual file name editing or automatic file name editing with
“Store”
Store
Stores the current data with the current name, or adds incremental
number
Store as
Stores the current data with a new name (on-board keyboard)
Load
Loads file entered in Name
Delete
Deletes current file
USM Vision+ User’s Manual
39
Chapter 3. Phased Array (PA) Mode
3.5
Color Palette Editor (Amplitude and True Depth)
3.5.1
3.5.1a
User Interface
Predefined Color Palettes
Color Palettes that come with the instrument are predefined and cannot be
changed by the operator. The operator can only choose from the given list
for Amplitude and Depth.
3.5.1b
Custom Color Palettes
Unlike predefined palettes, the “Custom” Color palettes can be changed by
the operators. They can specify the number of colors and specify, for each
color, at which value it will end and the color value. There will be a selector
for predefined colors, but they can also access the RGB values for each
color.
3.5.1c
Color Interpolations
Some of the predefined Color Palettes support a linear interpolation between
colors. This option is not available for custom colors, which will be limited
to colors that switch on a threshold.
3.5.1d
Function Bar: Color Selection
In the Color Selection menu, users can switch between the Amplitude or
Depth Mode Palettes, and select which palette should be used for the
selected mode. For a custom palette, they can set the number of colors.
Figure 20: Color Selection Menu
40
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
3.5.1e
Function Bar: Color Ranges
Figure 21: Color Range Menu
In the Color Range menu, the operator can specify the upper value for the selected
color range, select a predefined color for it, or modify the RGB value of the color in
the selected range.
Range End [% LSH] will be replaced by Range End [%Depth displayed] for
predefined depth palettes and changes Range End [mm] for the “Custom” depth
palette. Range End [mm] is valid from 0.0 to 6000.0 (as is RangeEnd in
UT->Range).
Color Index value is limited from 1 to “Number of Colors”. With color control, the
user selects from a list of seven predefined colors: black, red, green, blue, yellow,
white and gray. Changing any of the red, green or blue values will set Color to
Custom. Values for red, green and blue controls range from 0 to 255.
3.5.1f
Grey Out Logic
Predefined Color Palettes cannot be modified. Parameters allowing the operator to
inspect the current palette settings are operable (like Color Index). Custom Color
Palettes can only be modified while the USM Vision + is not scanning and not
analyzing.
3.5.1g
Storage
The selected Color Palette Names, along with their characteristics, will be stored in
the settings file. Thus, after loading an instrument setting or a scan, the color
definitions remain as they were when the file was stored.
USM Vision+ User’s Manual
41
Chapter 3. Phased Array (PA) Mode
3.6
Color Rulers
Figure 22: Color Ruler
3.6.1
Amplitude Color Bars
Figure 23: Predefined Amplitude Color Bars
The predefined color bars in the instrument are displayed in Figure 23 above. The
“Custom” color palette is displayed as it is initially set before the operator has
started to change it.
42
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
3.6.2
Amplitude Color Palette for RF
If Rectification is set to “RF,” the A-Scan displays the range between -100% LSH
and +100% LSH. The amplitudes in the Frame-View are always displayed rectified.
Therefore the Amplitude Color palette in RF repeats the positive color values for
the negative amplitude range.
Figure 24: RF Amplitude Color Palette
3.6.3
Depth Color Bars
Figure 25: Depth Color Bars
The Color Palette for Depth values (Figure 25) appears at the right side of the
A-Scan. On any signal peak (or flank event) in the A-Scan, the color that will be
displayed within the Top-View will show up at the same horizontal position in the
color bar. The Depth Color Bar is only displayed when “Depth Mode” is selected as
Data Type for the Top-View. Depth Color Palette is not displayed at all when
calibrating Delay or TCG.
USM Vision+ User’s Manual
43
Chapter 3. Phased Array (PA) Mode
3.7
Step Control Functionality
The linear and T/R pattern describe an aperture sequence of sending and receiving a
beam with the same angle and focal depth from different elements of a phased array
probe.
In USM Vision+ ver. 9.4.0, users could specify the first probe element where the
first beam aperture should start. The pattern contained all apertures starting at the
first element and the following probe elements, until the last element of the aperture
reached the last element of the probe. Therefore the first element of the aperture
increased by one from beam to beam.
With version 9.4.1, the operator can set the increment to a value other than one.The
pattern can thus contain a lower number of beams, because the USM Vision +
reaches the last probe element more quickly.
3.7.1
Valid Value Range
The minimum value for this control is 1 the maximum is (No of channels - 1) which
calculates to 127 for the current USM Vision+ hardware.
3.7.2
Number of Beams in Pattern
The number of beams in a linear or T/R pattern is:
Number of Beams =
(Number of Probe Elements - (Start Element -1) - Aperture) / Linear Step + 1
3.7.3
Grey Out Logic
Control is greyed out during any calibration and when the USM Vision + is in scan
or analyze mode.
3.7.4
User Interface
An additional control has been added for Linear as well as for T/R pattern selected.
Figure 26: User Interface Control
44
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
3.8
Sound Velocity Measurement
Figure 27: Sound Velocity Measurement
Sound velocity measurement is the topmost calibration menu. The interface appears
similar to Figure 27 above.
The operator enters backwall, radius, side drilled hole or flat bottom hole as Ref.
Type, and the distances /depths of the two reference reflectors as Ref. Dist D1 and
Ref. Dist D2. In addition, if the scan is other than sector, the operator needs to select
the beam for the measurement in the Display Views menu.
The instrument will set the natural beam angle when the operator presses the
START-button. Beam selection will be grayed out while sound velocity
measurement is active. The operator can choose the threshold of gate A as well as
the TOF measurement mode for gate A. Peak-mode is the typical setting for sound
velocity measurement.
In the example on the following pages, the linear probe 115-000-766 has been used
on a 10 mm aluminum plate, started with 6400 m/s as material velocity in menu
Part. When pressing the START-button, the instrument switches gate A on, and sets
it around the first reference distance (Start = D1 - 2 mm, if possible, and
width = 4 mm).
USM Vision+ User’s Manual
45
Chapter 3. Phased Array (PA) Mode
3.8
Sound Velocity Measurement (cont.)
The instrument activates the A-Scan + Frame - layout, and displays the envelope
curve in the A-Scan. An inverted V-icon is displayed while sound velocity
measurement is active. At any time the gain may be changed, and the envelope
curve will be generated again.
Figure 28: Envelope Curve in A-Scan
At any time the sound velocity measurement can be canceled by pressing the
CANCEL button. In this case, the former layout, start and width of gate A are
restored.
The Set UT range button enables users to quickly set a UT range around the two
reference distances / depths, when the current display does not show them:
Range start = (D1 - 2mm) - 0.25 * ((D2 + 2mm) - (D1 - 2mm))
Range end = (D2 + 2mm) + 0.25 * ((D2 + 2mm) - (D1 - 2mm)).
The operator now maximizes the first reference echo, and presses the STORE D1
button: The instrument will display the message “Measured TOF is invalid”, when
no echo has been detected in Gate A (This should only occur when the echo does
not intersect the gate bar in case of gate TOF measurement mode = flank, or in case
of “loss of echo in interface gate”, when range trigger = IF in menu range, which are
not typical settings for sound velocity measurement).
46
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
3.8
Sound Velocity Measurement (cont.)
Figure 29: Recording TOF
The instrument records the TOF of the highest echo of the envelope curve, within
the TOF range given by gate A start to gate A end, when the measured TOF value is
valid and positions gate A around the value of 'Ref Dist. D2' (Start = D2 - 2 mm, if
possible, and Width = 4 mm).
The operator now maximizes the second reference echo, and presses the STORE D2
button: The instrument will display the message “Measured TOF is invalid”, when
no echo has been detected in Gate A (for the same reasons as mentioned in D1
above).
The instrument records the TOF of the highest echo of the envelope curve, within
the TOF range given by gate A start to gate A end, when the measured TOF value is
valid, and calculates the sound velocity according to formula (2) of M. Berke
related document, and displays it as grayed out as Sound Velo (see Figure 30 on the
next page).
USM Vision+ User’s Manual
47
Chapter 3. Phased Array (PA) Mode
3.8
Sound Velocity Measurement (cont.)
Figure 30: Sound Velocity Calculation
The operator can now accept the measured sound velocity or cancel the value to
repeat the steps as desired.
With ACCEPT the measured sound velocity is confirmed and assigned to the actual
part’s material velocity. The delay law calculation is then automatically applied, and
the sound velocity measurement is complete. The former layout, start and width of
gate A are restored, and a V-icon, which is not inverted, will be shown.
48
USM Vision+ User’s Manual
Chapter 3. Phased Array (PA) Mode
3.8
Sound Velocity Measurement (cont.)
Figure 31: Return to Former Layout
The V-icon will be shown as long as the delay law calculation will not be called
again. As a result, the part material velocity is stored, but the state “sound velocity
was measured” is not stored, so that the V-icon will not be shown after re-load.
USM Vision+ User’s Manual
49
Chapter 3. Phased Array (PA) Mode
3.9
Range Trigger: IP Acquisition and IF Gate
The parameter “Range Trigger” in the UT Range menu has an additional mode, “IP
Acquisition and IF Gate.” In this mode, the A-Scan starts with the initial pulse (IP),
while gates A and B, and the TCG are triggered by the echo in the interface gate.
Thereby, the area between initial pulse and echo in the interface gate is displayed in
A-Scan and frame view, when the Range Start is set up accordingly.
To enable this mode, operators must select a linear scan with phasing angle = 0
degree, and the interface gate must be switched on.
Gate A and B are dynamically displayed in the A-Scan, which means that their gate
bars move as the echo in the interface gate moves. Gate A and B are not displayed
in frame view or side view. Frame view, side view and A-Scan are related to initial
pulse and are in sync.
When the top view will be generated from gated data (Gate A or B), it will be
relative to the echo in the interface gate. The readouts of gate A and B are also
relative to the echo in the interface gate.
As the start value of gate A or B must be a positive value, and gate A and B are
triggered by the echo in the interface gate, it is not possible to position gate A or B
before the echo in the interface gate.
3.10 New Viewer Display Options
Some additional viewer functions are available in version 9.4.1. These options
appear highlighted in light green in Figure 32 on the next page, while the options
available in version 9.4.0 are in white.
50
USM Vision+ User’s Manual
ignored
Amp
A+E/S+B
A+E/S+B+C (a)
A+E/S+B+C (d) Depth
A+E/S+B+C (d) Depth
A+E/S+B+C (d) Depth
Depth
Depth
Depth
A+E/S+C (d)
A+E/S+C (d)
A+E/S+C (d)
Acquisition
Gate A
Gate B
ignored
Acquisition
Acquisition
Gate A
Gate B
Data Source Data Tpe
ignored
ignored
Amp
Acquisition
Settings Scheme 9.4.0
Frame + Side + 2 * Top
Frame + 2* Top
Frame + Side + Top
Frame + Side
Frame + Top
Layout
Frame
Top View Config
--Amp Acquisition
Amp Gate A
Amp Gate B
Depth Acquisition
Depth Gate A
Depth Gate B
Depth Gate B-A
--Amp Acquisition
Amp Gate A
Amp Gate B
Depth Acquisition
Depth Gate A
Depth Gate B
Depth Gate B-A
Amp & Depth Acquisition
Amp & Depth Gate A
Amp & Depth Gate B
Amp & Depth Acquisition
Amp & Depth Gate A
Amp & Depth Gate B
Settings Scheme 9.4.1
A-Scan
Viewer 2
Viewer 3
Viewer 4
Viewer 3
Viewer 2
Viewer 2
Screen Appearance
A-Scan
A-Scan
Viewer 1
Viewer 1
USM Vision+ User’s Manual
Viewer 1
Layout
A+E/S
A+E/S+C (a)
Viewer 1
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Frame Amplitude
Chapter 3. Phased Array (PA) Mode
3.10 New Viewer Display Options (cont.)
Figure 32: Viewer Display Options
51
Chapter 3. Phased Array (PA) Mode
[no content intended for this page]
52
USM Vision+ User’s Manual
Chapter 4. Calibration
Chapter 4. Calibration
Successful calibration is vital to accurate weld inspection. The USM Vision+ can
perform both linear and sector scan calibration, as well as sensitivity, ACG, TCG
and encoder calibration.
4.1
Calibrating a 0° Linear Scan
This procedure calibrates a linear scan of 0° with probe 115-500-016 in direct
coupling.
1. The first step is to set the Part, Probe, Wedge, UT and Delay Law menus.
Table 8 below lists the appropriate settings for the menus.
Table 8: Menu Settings for Linear Scan 0°
Menu
Parameter
Value
Part
Mat. Velocity
5920 m/s
Part
Thickness
100 mm
Part
Shape
Flat
Part
Beam into curve.
No
Probe
Probe Name
115-500-016
Wedge Name
Custom
Wedge
Wedge Angle
0°
Wedge
Wedge Velocity
2730 m/s
Wedge
Wedge Front
0 mm
Wedge
Z-Offset
1 mm
Wedge
1st Elem. Pos.
Low
Shape
Flat
Beam Dir.
To Right
Range Start
0 mm
UT
Range Mode
Manual
UT
Range End
100 mm
Wedge
Wedge
Sub-Menu
Geometry
Curvature
Wedge
UT
USM Vision+ User’s Manual
Range
53
Chapter 4. Calibration
Table 8: Menu Settings for Linear Scan 0° (cont.)
Menu
Parameter
Value
UT
No. of Legs
1
UT
Range Trigger
IP
Pulser Voltage
90 V
UT
IP Width Mode
Auto
UT
PRF Mode
Manual
UT
PRF
2000 Hz
Rectification
Full Wave
UT
Filter
0.5 –11.5 MHz
UT
Video Filter
Off
Start Element
1
Delay Law
Aperture
16
Delay Law
Focal Depth
100 mm
Delay Law
Pin Offset
0
Electronic Scan
Linear
Angle
0°
UT
UT
Delay Law
Delay Law
Delay Law
Sub-Menu
Pulser
Receiver
Aperture
Electronic Scan
2. The symbol
indicates that element delays need to be calculated. Press
the Calc. Delay Law icon. The system now calculates the individual
element delays for the current setup and stores the values.
IMPORTANT:
As long as the Delay Law Calculation is pending, many functions of
the instrument are blocked!
Since all entered values are correct related to the probe and part, the system should
show almost correct signals, when coupled to the 25 mm K1. Gate parameters have
been set to measure the two echoes from 25mm and 50 mm using gate A and B.
54
USM Vision+ User’s Manual
Chapter 4. Calibration
4.1 Calibrating a 0° Linear Scan (cont.)
The current readings for the setup shown in Figure 33 below are:
• Beam = 10 (shot number)
• SA^ = sound path for max. echo in gate A
• SB^ = sound path for max. echo in gate B
• SBA = Sound path difference
• A% = echo height in gate A
• B% = echo height in gate B
Figure 33: Menu Setup for Calibration
Gate parameters have been set to measure the two echoes from 25 mm and 50 mm
using gate A and B.
3. Now you must set up the calibration itself, according to the parameters in
Table 9 on the next page.
USM Vision+ User’s Manual
55
Chapter 4. Calibration
4.1 Calibrating a 0° Linear Scan (cont.)
Table 9: Calibration Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
Delay - Setup
Reference Refl.
Depth
Calibration
Ref. Distance
25 mm
Calibration
Tolerance
1mm
Figure 34: Calibration Setup
Note: GE recommends setting the gain so that the optimized 25mm echo is at
~80% FSH.
4. Press Start to start recording the calibration echo.
Table 10: Calibration Recording
56
Menu
Sub-Menu
Calibration
Delay - Record
USM Vision+ User’s Manual
Chapter 4. Calibration
4.1 Calibrating a 0° Linear Scan (cont.)
Figure 35: Start Recording
5. As calibration begins, the Start icon changes to Record. A new window
shows the measured distances for all shots versus the beam number. The red
curve shows the measured sound paths for the current probe position.
Move the probe slowly to reach a constant coupling over the probe coupling
area. The red curve may change slightly. The gate measurement point should
be set to Peak.
USM Vision+ User’s Manual
57
Chapter 4. Calibration
4.1 Calibrating a 0° Linear Scan (cont.)
Figure 36: Window with Measured Distances
6. When you have found all sound paths, press Record to let the system
calculate the necessary delay line correction for each shot (beam).
Figure 37: Start Recording
58
USM Vision+ User’s Manual
Chapter 4. Calibration
4.1 Calibrating a 0° Linear Scan (cont.)
7. Press Store if all measured sound paths are within the tolerance band.
Figure 38: Storing the Calibration
8. The system is now calibrated, confirmed by the calibration symbol
You can perform a calibration check using K1/100 mm.
.
Figure 39: Calibration Check
USM Vision+ User’s Manual
59
Chapter 4. Calibration
4.2
Calibrating a 20° Linear Scan
This procedure calibrates a linear scan of 20° with probe 115-500-016 in direct
coupling.
Calibration block: DAC vertical
Thickness:
105 mm
Reference:
3 mm SDHs
Sound velocity: 5920 m/s
1. The first step is to set the Part, Probe, Wedge, UT and Delay Law menus.
Table 11 below lists the appropriate settings for the menus.
Table 11: Menu Settings for Linear Scan 20°
Menu
Parameter
Value
Part
Mat. Velocity
5920 m/s
Part
Thickness
105 mm
Part
Shape
Flat
Part
Beam into curve.
No
Probe
Probe Name
115-500-016
Wedge Name
Custom
Wedge
Wedge Angle
0°
Wedge
Wedge Velocity
2730 m/s
Wedge
Wedge Front
0 mm
Wedge
Z-Offset
1 mm
Wedge
1st Elem. Pos.
Low
Shape
Flat
Beam Dir.
To Right
Range Start
0 mm
UT
Range Mode
Manual
UT
Range End
120 mm
UT
No. of Legs
1
Wedge
Wedge
Sub-Menu
Geometry
Curvature
Wedge
UT
60
Range
USM Vision+ User’s Manual
Chapter 4. Calibration
Table 11: Menu Settings for Linear Scan 20° (cont.)
Menu
Sub-Menu
Parameter
Value
Range Trigger
IP
Pulser Voltage
90 V
UT
IP Width Mode
Auto
UT
PRF Mode
Manual
UT
PRF
2000 Hz
Rectification
Full Wave
UT
Filter
0.5 - 11.5 MHz
UT
Video Filter
Off
Start Element
1
Delay Law
Aperture
16
Delay Law
Focal Depth
50 mm
Delay Law
Pin Offset
0
Electronic Scan
Linear
Angle
20°
UT
UT
Pulser
UT
Receiver
Delay Law
Delay Law
Aperture
Electronic Scan
Delay Law
2. The symbol
indicates that element delays need to be calculated. Press
the Calc. Delay Law icon. The system now calculates the individual
element delays for the current setup and stores the values.
IMPORTANT:
As long as the Delay Law Calculation is pending, many functions of
the instrument are blocked!
Since all entered values are correct related to the probe and part, the system should
show almost correct signals, when coupled to the DAC block.
USM Vision+ User’s Manual
61
Chapter 4. Calibration
4.2 Calibrating a 20° Linear Scan (cont.)
The current readings for the setup shown in Figure 40 below are:
•
•
•
•
•
•
Beam = 10 (shot number)
SA^ = sound path for max. echo in gate A
SB^ = sound path for max. echo in gate B
SBA = Sound path difference
A% = echo height in gate A
B% = echo height in gate B
Figure 40: Menu Setup for Calibration
3. Set up the calibration according to the parameters in Table 12 below.
Table 12: Calibration Parameters
62
Menu
Sub-Menu
Parameter
Value
Calibration
Delay - Setup
Reference Refl.
SDH
Calibration
SDH Diam.
3 mm
Calibration
Ref. Distance
55 mm
Calibration
Tolerance
1 mm
USM Vision+ User’s Manual
Chapter 4. Calibration
4.2 Calibrating a 20° Linear Scan (cont.)
Figure 41 below shows the echo from the SDH in 55 mm.
Figure 41: SDH Echo
4. Press Start to begin recording.
Table 13: Calibration Recording
USM Vision+ User’s Manual
Menu
Sub-Menu
Calibration
Delay - Record
63
Chapter 4. Calibration
4.2 Calibrating a 20° Linear Scan (cont.)
Figure 42: Start Recording
Note: After you press Start, Gate A will automatically be set to cover the reference
reflector (here the SDH in 55 mm).
5. As calibration begins, the Start icon changes to Record. A new window
shows the measured distances for all shots versus the beam number. The red
curve shows the measured depths for the current probe position.
Move the probe slowly to hit the SDH with every shot (beam). The red curve
will change to correspond.
64
USM Vision+ User’s Manual
Chapter 4. Calibration
4.2 Calibrating a 20° Linear Scan (cont.)
Figure 43: Window with Measured Distances
6. When you have found all depths, press Record to let the system calculate
the necessary delay line correction for each shot (beam).
Figure 44: Start Recording
USM Vision+ User’s Manual
65
Chapter 4. Calibration
4.2 Calibrating a 20° Linear Scan (cont.)
7. Press Store if all measured depths are within the tolerance band.
Figure 45: Storing the Calibration
8. The system is now calibrated, confirmed by the calibration symbol
You can perform a calibration check using a 105 mm corner.
.
Figure 46: Calibration Check
66
USM Vision+ User’s Manual
Chapter 4. Calibration
4.2 Calibrating a 20° Linear Scan (cont.)
9. Sensitivity linearization uses the same procedure as with 0°. Instead of the
back wall echo at constant depth, you use the 3 mm SDH at 55 mm in the
DAC block.
Table 14: Amplitude Linearization (1-Point TCG)
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Setup
Reference Refl.
SDH
Calibration
SDH Diam.
3 mm
Calibration
Reference Ampl.
80%
Calibration
Tolerance
1 mm
USM Vision+ User’s Manual
67
Chapter 4. Calibration
4.3
Calibrating a -20° to 20° Sector Scan
This procedure enables calibration of a -20° to 20° sector scan with probe B2SPA.
Calibration block: Half cylinder
Radius:
50 mm
Sound velocity: 5840 m/s
1. The first step is to set the Part, Probe, Wedge, UT and Delay Law menus.
Table 15 below lists the appropriate settings for the menus.
Table 15: Menu Settings for -20° to 20° Sector Scan
Menu
Parameter
Value
Part
Mat. Velocity
5840 m/s
Part
Thickness
100 mm
Part
Shape
Flat
Part
Beam into curve.
No
Probe
Probe Name
Custom *
Probe
Probe Frequency
2 MHz
Probe
No. of Elements
16
Probe
Pitch
1.5 mm
Probe
Elevation
24 mm
Wedge Name
Custom
Wedge
Wedge Angle
0°
Wedge
Wedge Velocity
2730 m/s
Wedge
Wedge Front
0 mm
Wedge
Z-Offset
2.3 mm
Wedge
1st Elem. Pos.
Low
Shape
Flat
Beam Dir.
To Right
Wedge
Wedge
Wedge
68
Sub-Menu
Geometry
Curvature
USM Vision+ User’s Manual
Chapter 4. Calibration
Table 15: Menu Settings for -20° to 20° Sector Scan (cont.)
Menu
Sub-Menu
Parameter
Value
UT
Range
Range Start
0 mm
UT
Range Mode
Manual
UT
Range End
120 mm
UT
No. of Legs
1
UT
Range Trigger
IP
Pulser Voltage
90 V
UT
IP Width Mode
Auto
UT
PRF Mode
Manual
UT
PRF
2000 Hz
Rectification
Full Wave
UT
Filter
0.5 –11.5 MHz
UT
Video Filter
Off
Start Element
1
Delay Law
Aperture
16
Delay Law
Focal Depth
100 mm
Delay Law
Pin Offset
0
Electronic Scan
Sector
Delay Law
Angle
0°
Delay Law
Angle Start
-20°
Delay Law
Angle Stop
20’
Delay Law
Angle Step
1°
UT
Pulser
UT
Receiver
Delay Law
Delay Law
Aperture
Electronic Scan
2. The symbol
indicates that element delays need to be calculated. Press
the Calc. Delay Law icon. The system now calculates the individual
element delays for the current setup and stores the values.
USM Vision+ User’s Manual
69
Chapter 4. Calibration
4.3 Calibrating a -20° to 20° Sector Scan (cont.)
IMPORTANT:
As long as the Delay Law Calculation is pending, many functions of
the instrument are blocked!
Since all entered values are correct related to the probe and part, the system should
show almost correct signals, when coupled to the 50 mm half-circular block, and
optimizing the echo from radius 50 mm. Gate values have been set to measure the
two echoes from 50 mm and 100 mm using gate A and B.
The current readings for the setup shown in Figure 47 below are:
• Beam = 21 (beam number, here 0°)
• SA^ = sound path for max. echo in gate A
• SB^ = sound path for max. echo in gate B
• SBA = Sound path difference
• A% = echo height in gate A
• B% = echo height in gate B
Figure 47: Menu Setup for Calibration
70
USM Vision+ User’s Manual
Chapter 4. Calibration
4.3 Calibrating a -20° to 20° Sector Scan (cont.)
3. Set up the calibration according to the parameters in Table 16 below.
Table 16: Calibration Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
Delay - Setup
Reference Refl.
Radius
Calibration
Ref. Distance
50 mm
Calibration
Tolerance
1mm
Note: GE recommends selecting an angle of 0° (at the center of the angular range)
and setting the gain so that the optimized 50 mm echo is at ~80% FSH.
Figure 48: Calibration Setup
4. Press Start to begin recording.
Table 17: Calibration Recording
USM Vision+ User’s Manual
Menu
Sub-Menu
Calibration
Delay - Record
71
Chapter 4. Calibration
4.3 Calibrating a -20° to 20° Sector Scan (cont.)
Figure 49: Recording Setup
Note: The probe’s primary (active) axis is in the radial direction (cable in axial
direction).
Figure 50: Probe Direction
5. As calibration begins, the Start icon changes to Record. A new window
shows the measured distances for all shots versus the beam number. The red
curve shows the measured sound paths for the current probe position.
Move the probe slowly to collect the maximum amplitude for every angle.
The red curve will change to correspond.
72
USM Vision+ User’s Manual
Chapter 4. Calibration
4.3 Calibrating a -20° to 20° Sector Scan (cont.)
Figure 51: Window with Measured Sound Paths
6. When you have found all maximum amplitudes, press Record to let the
system calculate the necessary delay line correction for each angle.
Figure 52: Start Recording
USM Vision+ User’s Manual
73
Chapter 4. Calibration
4.3 Calibrating a -20° to 20° Sector Scan (cont.)
7. Press Store if all measured sound paths are within the tolerance band.
Figure 53: Storing the Calibration
8. The system is now calibrated, confirmed by the calibration symbol
.
You can perform a calibration check using a SDH in 65 mm or an SDH full
screen (Figure 54).
Figure 54: Full Screen Calibration Check
74
USM Vision+ User’s Manual
Chapter 4. Calibration
4.4
Calibrating a 40° to 70° Sector Scan
This procedure enables calibration of a 40° to 70° sector scan with probe MWB4PA
(built-in wedge).
1. The first step is to set the Part, Probe, Wedge, UT and Delay Law menus.
Table 18 below lists the appropriate settings for the menus.
Table 18: Menu Settings for 40° to 70° Sector Scan
Menu
Parameter
Value
Part
Mat. Velocity
3250 m/s
Part
Thickness
100 mm
Part
Shape
Flat
Part
Beam into curve.
No
Probe
Probe Name
69142
Wedge Name
Custom
Wedge
Wedge Angle
43.2°
Wedge
Wedge Velocity
2730 m/s
Wedge
Wedge Front
16.1 mm
Wedge
Z-Offset
5.3 mm
Wedge
1st Elem. Pos.
Low
Shape
Flat
Beam Dir.
To Right
Range Start
0 mm
UT
Range Mode
Manual
UT
Range End
100 mm
UT
No. of Legs
1
UT
Range Trigger
IP
Pulser Voltage
90 V
IP Width Mode
Auto
Wedge
Wedge
Sub-Menu
Geometry
Curvature
Wedge
UT
UT
UT
USM Vision+ User’s Manual
Range
Pulser
75
Chapter 4. Calibration
Table 18: Menu Settings for 40° to 70° Sector Scan (cont.)
Menu
Parameter
Value
UT
PRF Mode
Manual
UT
PRF
2000 Hz
Rectification
Full Wave
UT
Filter
0.5 –11.5 MHz
UT
Video Filter
Off
Start Element
1
Delay Law
Aperture
16
Delay Law
Focal Depth
100 mm
Delay Law
Pin Offset
0
Electronic Scan
Sector
Delay Law
Angle Start
40°
Delay Law
Angle Stop
70’
Delay Law
Angle Step
1°
UT
Delay Law
Delay Law
Sub-Menu
Receiver
Aperture
Electronic Scan
2. The symbol
indicates that element delays need to be calculated. Press
the Calc. Delay Law icon. The system now calculates the individual
element delays for the current setup and stores the values.
IMPORTANT:
As long as the Delay Law Calculation is pending, many functions of
the instrument are blocked!
Since all entered values are correct related to the probe and part, the system should
show almost correct signals, when coupled to K2, and optimizing the echo from
radius 25 mm. Gate values have been set to measure the two echoes from 25 mm
and 100 mm using gate A and B. The part thickness must be equal or larger than the
sound path related to the calibration distance! After calibration, the thickness can be
set to the real value of the test object.
76
USM Vision+ User’s Manual
Chapter 4. Calibration
4.4 Calibrating a 40° to 70° Sector Scan (cont.)
The current readings for the setup shown in Figure 55 below are:
• Beam = 21 (beam number, here 60°)
• SA^ = sound path for max. echo in gate A
• SB^ = sound path for max. echo in gate B
• SBA = Sound path difference
• A% = echo height in gate A
• B% = echo height in gate B
Figure 55: Menu Setup for Calibration
USM Vision+ User’s Manual
77
Chapter 4. Calibration
4.4 Calibrating a 40° to 70° Sector Scan (cont.)
3. Set up the calibration according to the parameters in Table 19 below.
Table 19: Calibration Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
Delay - Setup
Reference Refl.
Radius
Calibration
Ref. Distance
25 mm
Calibration
Tolerance
1mm
Note: GE recommends selecting an angle of 55° (at the center of the angular
range) and setting the gain so that the optimized 25 mm echo is at ~80%
FSH.
Figure 56: Calibration Setup
4. Press Start to begin recording.
Table 20: Calibration Recording
78
Menu
Sub-Menu
Calibration
Delay - Record
USM Vision+ User’s Manual
Chapter 4. Calibration
4.4 Calibrating a 40° to 70° Sector Scan (cont.)
Figure 57: Recording Setup
5. As calibration begins, the Start icon changes to Record. A new window
shows the measured distances for all shots versus the beam angle. The red
curve shows the measured sound paths for the current probe position.
Move the probe slowly to collect the maximum amplitude for every angle.
The red curve will change to correspond.
Figure 58: Window with Measured Sound Paths
USM Vision+ User’s Manual
79
Chapter 4. Calibration
4.4 Calibrating a 40° to 70° Sector Scan (cont.)
6. When you have found all maximum amplitudes, press Record to let the
system calculate the necessary delay line correction for each angle.
Figure 59: Start Recording
7. Press Store if all measured sound paths are within the tolerance band.
Figure 60: Storing the Calibration
80
USM Vision+ User’s Manual
Chapter 4. Calibration
4.4 Calibrating a 40° to 70° Sector Scan (cont.)
8. The system is now calibrated, confirmed by the calibration symbol
.
You can perform a calibration check using K1 at 40° (Figure 37) 55° (Figure
38) or 70°.
Figure 61: Calibration Check at 40° (99.9 mm)
Figure 62: Calibration Check at 55° (99.9 mm)
USM Vision+ User’s Manual
81
Chapter 4. Calibration
4.4 Calibrating a 40° to 70° Sector Scan (cont.)
Figure 63: Calibration Check at 70° (99.5 mm)
82
USM Vision+ User’s Manual
Chapter 4. Calibration
4.4.1
Creating a Dual Sector Scan for Weld Inspection
Once you have correctly set up and calibrated the first group, you can then “clone”
a second group from the first, e.g., for a double side simultaneous scan (dual scan).
You must use a so-called “Y-probe” (two probes sharing one connector). This
example uses the “Y-version” of the probe used for the single sector scan
calibration, as described in section 4.4, order number P/N 115-500-013.
Here, the two 16 element probes are connected via the Y-cable with one socket. On
the socket, probe #1 (left) uses pins 1 – 16, probe #2 (right) uses pins 33 – 48. Pin
Offset for probe #2 is 32.
Set up the 115-500-013 probe on 36° wedge (69438) with:
• Calibration on K2
• TCG recording with 4 points (10, 25, 40, 55mm)
• Sector from 40° - 70°, stored to file “2Sector_left”
Figure 64 below shows the setup for probe #1 on the DAC block to verify the
correct calibration and probe position.
Figure 64: Setup for Probe #1
USM Vision+ User’s Manual
83
Chapter 4. Calibration
4.4.1
Creating a Dual Sector Scan for Weld Inspection (cont.)
1. Set the parameters as shown in Table 21 below.
Menu
Delay Law:
Groups
Table 21: Menus and Actions for Dual Sector Scan
Action / Function
Action
Comment
Copy
The system creates a second
group having exactly the same
parameters as group #1.
Rename
Enter:
The new group is now called
sector right “sector right”.
Ok
Delay Law:
Aperture
Pin Offset
= 32
Probe #2 (right) starts now at pin
33 on the combined socket.
Display: Weld
Overlay
Beam dir.
= To Left
The beam direction of probe #2
now looks to the left.
Origin Offset Y
= +40 mm
The probe offset to the weld
center is now equal to probe #1,
but opposite.
Group Name
Select
previous
group
Delay Law:
Groups
Rename
Enter:
The first (original) group has now been
sector left renamed to “sector left”.
Ok
File: Load/Store Store As
Enter:
2Sector
The setup including 2 groups is now stored
with the name “2Sector”.
2. Set up the dual sector scan in the following sequence, shown in Figure 65 on
the next page:
• Weld geometry: Double V
• Dimensions: A = E = 0 mm, B = C = 12 mm, D = F = 9 mm, T = 24 mm
• UT Range: 20 mm - 52 mm, 3 Legs, Probe offset to weld center = 40 mm
84
USM Vision+ User’s Manual
Chapter 4. Calibration
4.4.1
Creating a Dual Sector Scan for Weld Inspection (cont.)
Figure 65: Weld Overlay with Flip Button
The scan result appears as shown in Figure 66 below. The scan data is stored in the
file: “2Sector_weld scan” The current position (blue cursor) displays a root defect.
Figure 66: Scan Result with Root Defect
USM Vision+ User’s Manual
85
Chapter 4. Calibration
4.5
Sensitivity Calibration
Sensitivity calibration is used to compensate any differences in sensitivity for every
shot (beam) of the current group, based on a known reference reflector in a given
depth.
Sensitivity differences are related to:
• Production differences for the individual elements in the Phased Array probe
• Beam steering
• Changes of delay line length due to shifting the virtual probe along the array
(electronic linear scan)
Sensitivity calibration is achieved by using the TCG calibration for just one single
given reference reflector (→ 1-Point TCG).
Before the TCG calibration can be started, the Delay Law Calculation (DLC) and
the Delay calibration must have been completed successfully. The examples in this
section are for a linear scan 0° with probe 115-500-016 in direct coupling and for a
sector scan 40° to 70° with MWB4PA.
4.5.1
1-point TCG (0° Linear Scan)
TCG recording for just one reflector will linearize the sensitivity for all shots
(beams) of the current group.
1. Set the calibration parameters as described in Table 22 below.
Table 22: Calibration Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG- Setup
Reference Refl.
Depth
Tolerance
5%
Calibration
86
USM Vision+ User’s Manual
Chapter 4. Calibration
4.5.1
1-point TCG (0° Linear Scan) (cont.)
Figure 67: TCG-Setup Sub-Menu
2. Press Start. Gate A will automatically be set to cover the reference reflector:
here, the BE in 25 mm.
Table 23: Recording Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG- Record
Reference
Distance
25 mm
USM Vision+ User’s Manual
87
Chapter 4. Calibration
4.5.1
1-point TCG (0° Linear Scan) (cont.)
Figure 68: Start Recording
3. The Start icon changes to Record. A new window shows the measured
amplitudes for all shots versus the beam number. The red curve shows the
measured sound amplitudes for the current probe position. Move the probe
slightly to reach a constant coupling over the probe coupling area. The red
curve will change slightly.
Figure 69: Window with Measured Amplitudes
88
USM Vision+ User’s Manual
Chapter 4. Calibration
4.5.1
1-point TCG (0° Linear Scan) (cont.)
4. When you have found all maximum amplitudes, press Record to let the
system calculate the necessary amplitude correction for every shot (beam).
Figure 70: Amplitude Recording
5. Press Store if all measured amplitudes are within the tolerance band. The
electronic scan now performs an equal sensitivity for all 49 shots (beams).
Figure 71: Store Recording
USM Vision+ User’s Manual
89
Chapter 4. Calibration
4.5.1
1-point TCG (0° Linear Scan) (cont.)
6. Press End to finalize TCG calibration and store the amplitude correction
values. The End button turns grey, and the system is ready for use.
Figure 72: End Calibration
90
USM Vision+ User’s Manual
Chapter 4. Calibration
4.5.2
1-point TCG = ACG (40° to 70° Sector Scan)
TCG recording for a single reflector— here, the 1.5 mm SDH in the K1 calibration
block—will linearize the sensitivity for all angles (beams) of the current group →
ACG = Angle Corrected Gain.
1. Set the calibration parameters as described in Table 24 below.
Table 24: Calibration Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Setup
Reference Refl.
SDH
Calibration
SDH Diam.
1.5 mm
Calibration
Tolerance
5%
Figure 73: TCG Setup Sub-Menu
2. Press Start. Set the gain to receive an echo amplitude of ~80%.
Note: Gate A will automatically be set to cover the reference reflector (here the
SDH in 15 mm) after you press Start.
USM Vision+ User’s Manual
91
Chapter 4. Calibration
4.5.2
1-point TCG = ACG (40° to 70° Sector Scan) (cont.)
Table 25: Calibration Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Record
Ref. Distance
15 mm
Figure 74: Recording Setup
3. Start changes to Record. A new window shows the measured amplitudes
for all shots versus the beam angle. The red curve shows the measured echo
amplitudes for the current probe position. Move the probe slowly to record
the maximum reference amplitude for every angle. The red curve will
change correspondingly. Reduce the gain if the amplitude exceeds 100%
FSH.
92
USM Vision+ User’s Manual
Chapter 4. Calibration
4.5.2
1-point TCG = ACG (40° to 70° Sector Scan) (cont.)
Figure 75: Window with Measured Amplitudes
4. When you have found all maximum amplitudes, press Record to let the
system calculate the necessary amplitude correction for every angle (beam).
Re-check the amplitudes by recording the amplitudes again, and press
Record repeatedly.
Figure 76: Recording Underway
USM Vision+ User’s Manual
93
Chapter 4. Calibration
4.5.2
1-point TCG = ACG (40° to 70° Sector Scan) (cont.)
5. Press Store if all measured amplitudes (red curve) are finally within the
tolerance band.
Figure 77: Store Calibration
6. Press End to finalize TCG calibration and store the amplitude correction
values. The electronic scan now performs an equal sensitivity for all 31
angles (beams)→ ACG = Angle Corrected Gain. The End button turns grey.
94
USM Vision+ User’s Manual
Chapter 4. Calibration
4.5.2
1-point TCG = ACG (40° to 70° Sector Scan) (cont.)
Figure 78: End Calibration
The following screens illustrate amplitude verification at 40°, 55° and 70°.
Figure 79: Amplitude Verification at 40°
USM Vision+ User’s Manual
95
Chapter 4. Calibration
4.5.2
1-point TCG = ACG (40° to 70° Sector Scan) (cont.)
Figure 80: Amplitude Verification at 55°
Figure 81: Amplitude Verification at 70°
96
USM Vision+ User’s Manual
Chapter 4. Calibration
4.6
Sector Scan TCG Recording
TCG recording is used to compensate any differences in sensitivity for every shot
(beam) of the current group, based on several known reference reflectors of the
same size, but at different known depths. Apart from ACG (Angle Corrected Gain),
the losses due to beam divergence and sound attenuation will be compensated
within the range defined by the distances of the reference reflectors.
Before the TCG calibration can be started, the Delay Law Calculation (DLC) and
the Delay calibration must have been completed successfully.
In case one TCG point has already been recorded (e.g., for ACG), recording may be
continued by adding new reference reflectors in other depths.
IMPORTANT:
You must continue an already existing TCG with the same reflector
type and size! If you cannot, clear the existing TCG and start a
completely new one.
1. Set up the TCG according to Table 26 below. In this example, 3mm SDHs
will be used for TCG recording. No TCG has been stored before.
Table 26: Calibration Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Setup
Reference Refl.
SDH
SDH Diameter
3mm
Reference Ampl.
8-%
Tolerance
5%
2. Enter the distance to the first reference reflector, then press Start.
Note: Gate A will automatically be set to cover the reference reflector (here the
SDH in 10 mm) after you press Start.
Table 27: Recording Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Record
Reference
Distance
10 mm
USM Vision+ User’s Manual
97
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
Figure 82: TCG Setup
3. Start changes to Record. A new window shows the measured amplitudes
for all shots versus the beam angle.The white curve shows the measured
echo amplitudes for the current probe position. Move the probe slowly to
record the maximum reference amplitude for every angle (the red curve).
The red curve will change to correspond as the SDH is hit by the different
beams.
98
USM Vision+ User’s Manual
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
Figure 83: Recording Amplitudes
When you have found all maximum amplitudes, press Record to let the system
calculate the necessary amplitude correction for every angle (beam).
Figure 84: Calculating Amplitudes
USM Vision+ User’s Manual
99
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
4. Re-check the amplitudes by recording the amplitudes again, and press
Record repeatedly, until all measured amplitudes (red curve) are finally
within the tolerance band. Then press Store.
Figure 85: Storing Amplitude Corrections
5. If one or more amplitudes have amplitudes outside the specified tolerance
band, the system will prompt an error message. If you press Yes, the system
will store the amplitude corrections even with values out of the tolerance. If
you select No, the system returns to the previous screen to let you record the
amplitudes of the last reference target again.
The system stores all dB-compensation values and returns to normal
operation. The reference number has incremented to 2.
100
USM Vision+ User’s Manual
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
Figure 86: Out of Tolerance Message
Figure 87: Storing Compensation Values
USM Vision+ User’s Manual
101
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
6. Enter the next reference distance — here, 25 mm. Press Start.
Note: Gate A will automatically be set to cover the reference reflector (here the
SDH in 25 mm) after you press Start.
Table 28: Recording Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Record
Reference
Distance
25 mm
Figure 88: TCG Setup for 25 mm
7. As with the first reference reflector, move the probe to pick up all maximum
echo amplitudes of the second reference reflector.
102
USM Vision+ User’s Manual
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
Figure 89: Recording 25 mm Amplitudes
8. Press Record repeatedly to adjust all maximum amplitudes into the
tolerance band.
9. Finally, press Store to apply the measured amplitude corrections for the
second reference reflector.
Figure 90: Storing Amplitudes
USM Vision+ User’s Manual
103
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
10. Enter the next reference distance of 40 mm. Press Start.
Note: Gate A will automatically be set to cover the reference reflector (here the
SDH in 40 mm) after you press Start.
Table 29: Recording Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Record
Reference
Distance
40 mm
Figure 91: TCG Setup for 40mm
Due to the geometry of the DAC block, a false indication runs into the recording
gate. Since the system will record the maximum amplitude of all echoes within the
gate, some amplitudes will be taken from the false indication, and finally this TCG
point will become wrong. Therefore, you must exclude all angles at which false
signals occur.
104
USM Vision+ User’s Manual
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
Figure 92: Amplitude Problem
11. Enter 60° to Beam Section. High angles from 61° to 70° will now be
excluded from the amplitude recording.
Table 30: Recording Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Record
Reference
Distance
40 mm
Beam Section
60°
USM Vision+ User’s Manual
105
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
Figure 93: Reducing the Angular Range of the Beam Section
12. As with the previous reference reflector, move the probe to pick up all
maximum echo amplitudes of the third reference reflector for the given
section. Press Record repeatedly to adjust all maximum amplitudes into the
tolerance band.
Figure 94: Recording Amplitudes
106
USM Vision+ User’s Manual
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
13. Press Store if all measured amplitudes (the red curve) are finally within the
tolerance band. The system is now ready to record the reference signals for
the remaining section (61° - 70°).
Figure 95: Storing Amplitudes
Table 31: Recording Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Record
Reference
Distance
40 mm
Beam Section
70°
14. When you have found all maximum amplitudes, press Record to let the
system calculate the necessary amplitude correction for the angles 61° to
70°. Re-check the amplitudes by recording the amplitudes again, and press
Record repeatedly.
USM Vision+ User’s Manual
107
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
Figure 96: Recording Amplitudes
15. Press Store if all measured amplitudes (red curve) are finally within the
tolerance band.
Table 32: Recording Parameters
108
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Record
Reference
Distance
40 mm
USM Vision+ User’s Manual
Chapter 4. Calibration
4.6 Sector Scan TCG Recording (cont.)
Figure 97: Store Amplitudes
16. Press End to finalize the TCG calibration procedure.
The electronic scan now performs an equal sensitivity for all 31 angles
(beams), and all three reference reflectors from 10 mm to 40 mm. The End
button turns grey.
Figure 98: Ending TCG Calibration
USM Vision+ User’s Manual
109
Chapter 4. Calibration
4.6.1
Verifying TCG Calibration
Table 33: Recording Parameters
Menu
Sub-Menu
Parameter
Value
Calibration
TCG - Verify
Beam Angle
40° (60°)
1. Press Check TCG.
2. Scan all three reference reflectors: The green envelope curve shows proof
that all reference echoes reach 80% FSH.
3. Press Finish to return to normal operation.
Figure 99: Verifying TCG Calibration at 40°
110
USM Vision+ User’s Manual
Chapter 4. Calibration
4.6.1
Verifying TCG Calibration (cont.)
Figure 100: Verifying TCG Calibration at 60°
USM Vision+ User’s Manual
111
Chapter 4. Calibration
4.6.2
TCG Amplitude Evaluation Levels
Additional evaluation lines are displayed in the A-scan according to the entered
dB-differences to the original reference. The reading AdBC will directly display the
dB-difference of the echo in gate A to the reference (here, 80% FSH).
Figure 101: TCG Evaluation Levels
112
USM Vision+ User’s Manual
Chapter 4. Calibration
4.6.3
Echo Evaluation
This example shows inspection of a 30 mm thick V-Weld with an inclusion at half
thickness. The echo exceeds the reference level by 0.6 dB. Figure 103 shows the
same result in volume corrected view: the signals of the inclusion are hit directly
with 70° (leg 1), and after one reflection with 47° (leg 2), and with an x-shaped
indication in the sector scan.
Figure 102: Inspection of 30 mm V-Weld
USM Vision+ User’s Manual
113
Chapter 4. Calibration
4.6.3
Echo Evaluation (cont.)
Figure 103: Inspection in Volume Corrected View
114
USM Vision+ User’s Manual
Chapter 4. Calibration
4.7
Encoder Calibration
A quadrature wheel encoder provides a certain number of ticks (square wave
pulses) per revolution. In order to exactly measure scan distances, the system needs
to know the number of ticks per mm (inch). If this value is unknown, the encoder
calibration will establish this value.
To establish the number of ticks per mm (or inch), enter the wanted scan length into
the system, then move the encoder along this distance, and the system will calculate
the characteristic encoder value.
Menu
Scan
Table 34: Parameters for Encoder Calibration
Sub-Menu
Parameter
Value
Encoder Cal.
Scan Mode
Positional
Scan
Encoder Dir.
Clockwise
Scan
Encoder Counts
0.3 mm
Scan
Scan increment
1 mm
Scan
Cal. Distance
300 mm
1. Place the encoder at the Zero position of the calibration distance (here, 300
mm).
2. Press Start Calibrate. The button changes to Stop Calibrate.
3. Move the encoder along the calibration distance.
Figure 104: Moving Encoder Along Calibration Distance
USM Vision+ User’s Manual
115
Chapter 4. Calibration
4.7 Encoder Calibration (cont.)
Figure 105: Encoder Calibration Screen
4. Press Stop Calibrate at the end of the calibration distance.The system now
calculates the encoder counts (mm/tick) and stores the value.
Figure 106: Encoder at End of Calibration
116
USM Vision+ User’s Manual
Chapter 4. Calibration
4.7 Encoder Calibration (cont.)
Figure 107: Encoder Calibration Screen, with
Value Calculated and Stored
USM Vision+ User’s Manual
117
Chapter 4. Calibration
[no content intended for this page]
118
USM Vision+ User’s Manual
Chapter 5. Specifications
Chapter 5. Specifications
5.1 General Specifications
Table 35: USM Vision+ Specifications
Information
Values
Unit
Comments
1. Configurations
Phased array configuration
16/128
number of simultaneously
controlled channels and
number of available channels
Conventional channel
1
Pulse/Echo or Dual mode
2. General features
Size, WxHxD
367 (310) x
mm3
250 x 100 (60)
Weight
4.6
kg
Display size
10.4
inch
Display resolution
1024x768
pixel
Power supply, input voltage
100 - 240
VAC
Power supply, output voltage
15
VDC
Maximum power consumption 45
W
Battery operational time
3
h
Number of batteries 2
Operating temperature range
0 - 45
°
Storage temperature range
-20 - 70
°
Pulse repetition frequencies
(PRF)
0,015 to 10
kHz
Protection grade
IP 54
Available measurement units
mm, inch
USM Vision+ User’s Manual
with one battery
TFT with LED backlight
hot swap possible
Lithium Ion
depending on settings
119
Chapter 5. Specifications
Table 35: USM Vision+ Specifications (cont.)
Information
Values
Unit
Comments
3. Input/output connectors
Phased Array Probe
Tyco
Conventional Probe
Lemo00 coax
and triax
Interface I/O
Lemo 2B
14pin
VGA
Lemo 0 9pin
Ethernet
RJ 45
1Gb/s
USB 2.0
3
Type A
Power connector
Lemo 0S 4pin
encoder (quadrature, 5V), SAP;
s. table
4. Display
Range of sound velocities
100 to 15000 m/s
Time base:
Delay
0 - 10000
mm
in steel long, IP delay
Width
6 -10000
mm
in steel long
Hz
depending on setting
Available views
A, B, C, D, E, S
Screen refresh rate
50
5. Beam forming
Maximum number of channels 16
active simultaneously
different configurations
Maximum number of delay
laws
256
individual cycles
Maximum time delay
20,000
ns
Step
5
ns
6. Phased Array Transmitter
Number of transmitters
available simultaneously
16
Shape of transmitter pulse
Negative
unipolar
120
depending on configuration
USM Vision+ User’s Manual
Chapter 5. Specifications
Table 35: USM Vision+ Specifications (cont.)
Information
Values
Unit
Comments
6. Phased Array Transmitter (cont.)
Transmitter voltage
3 -150
V
Fall time
<10
ns
Duration
20 - 1200
ns
Maximum time delay
0 to 20000
ns
Time delay resolution
5
ns
in 10V steps, supply voltage
200V
20ns steps
7. Conventional Transmitter (not accessible parameter)
Shape of transmitter pulse
Negative
unipolar
Transmitter voltage
3 -180
V
Fall time
<10
ns
Duration
20 - 1200
ns
20ns steps
800% of FSH available for
post processing
8. Phased Array Receiver
Number of receivers available
simultaneously
16
Input voltage at full screen
height (FSH)
0.5
Vpp
Maximum input voltage
4
Vpp
Linearity of vertical display
+/- 2
%
Frequency response
0,5 - 15
MHz
Digital Filters
8
Dead time after transmitter
pulse
<5
us
Dynamic range
0 to 90
dB
Maximum time delay
0 to 20000
ns
Time delay resolution
5
ns
Time Corrected Gain
90
dB
USM Vision+ User’s Manual
-3 dB without digital filter
digital gain, 0,1 dB step
16 points/90dB in 20ns steps,
90dB/80ns slope, 220ns delay
of start
121
Chapter 5. Specifications
Table 35: USM Vision+ Specifications (cont.)
Information
Values
Unit
Comments
8. Phased Array Receiver (cont.)
Linearity of time delays
<1
%
of full range
Gain linearity
+/-2
dB
of full range
Channel gain variation
3
dB
Maximum digitisation
frequency without processing
50
MHz
Digitisation frequency with
processing
200
MHz
with interpolation
Digitiser vertical resolution
20/24
bit
20/channel, 24 on formed
beam
Display Start Mode
IP, IF Start
display
Display start depending on
interface echo in gate I, gate A
and B also triggered with
interface echo
9. Conventional Receiver
Number of receivers
1
Input voltage at full screen
height (FSH)
0.5
Vpp
Maximum input voltage
4
Vpp
Linearity of vertical display
+/- 2
%
Linearity of the vertical display +/- 2
%
Frequency response
0,5 - 15
MHz
-3 dB without digital filter
Digital Filters
8
Dynamic range
0 to 90
dB
digital gain, 0,1 dB step
DAC
90
dB
16 points/90dB in 20ns steps,
90dB/80ns slope, 220ns delay
of start
Maximum digitisation
frequency without processing
100
MHz
122
800% of FSH available for
post processing
USM Vision+ User’s Manual
Chapter 5. Specifications
Table 35: USM Vision+ Specifications (cont.)
Information
Values
Unit
Comments
9. Conventional Receiver (cont.)
Digitisation frequency with
processing
200
MHz
Digitiser vertical resolution
20
bit
Display Start Mode
IP
with interpolation
10. Data acquisition
Maximum number of A-scans
stored per second
4000
A-scan 512 points with 16 bit
amplitude
Maximum number of samples
per A-scan
1024
16 bit amplitude
Number of gates
3
incl. IF (A, B, I)
Type of detection
2
coincidence or
anticoincidence
Measurement mode
3
flank, J-flank, peak
Synchronisation of gates
2
Initial pulse or with interface
echo in gate I
11. Gates
Characteristics of gates:
Threshold
0 - 95
%
screen height (+/- 95 % in RF
mode)
Start
0 to 4000
mm
in steel long
Width
0.1 to 4000
mm
in steel long
Resolution of TOF
measurements
5
ns
Resolution of Amp
measurements
1
bit
Start Mode
IP, IF Start
display
USM Vision+ User’s Manual
16 bit signed
123
Chapter 5. Specifications
Table 35: USM Vision+ Specifications (cont.)
Information
Values
Unit
Comments
12. Processing
Rectification
4
pos, neg, RF, full
Averaging
1,2,4,8,16
TOFD: max depth 500mm in
steel
Envelope, EchoMax
on/off
Scan Mode
pulse on
position
Video Filter
on/off
Phased Array mode
PC Module
1
COM Express compact,
1,6GHz
SSD
64
Input devices
4
13. PC
124
GB
SLC, SATA
2 Track balls, keypad, touch
screen
USM Vision+ User’s Manual
Chapter 5. Specifications
5.2 I/O Connector (LEMO ECG.2B.314.CLV)
Contact No.
Designation
Function
Signal Control
1
GND-EXT
Ground encoder
2
+5V_EXT
Supply encoder
Output
3
SAP
Transmitter trigger pulse
Output
4
INDX_Y
Encoder Y index
Input
5
Y_B
Encoder Y phase B
Input
6
PDF
Test data release
Input
7
X_A
Encoder X phase A
Input
8
X_B
Encoder X phase B
Input
9
INDX_X
Encoder X index
Input
10
Y_A
Encoder Y phase A
Input
11
Dig I/O-1
General purpose I/O
In/Output
12
Dig I/O-2
General purpose I/O
In/Output
13
Dig I/O-3
General purpose I/O
In/Output
14
Dig I/O-4
General purpose I/O
In/Output
USM Vision+ User’s Manual
125
Chapter 5. Specifications
[no content intended for this page]
126
USM Vision+ User’s Manual
Appendix A. Creating User Accounts
Appendix A. Creating User Accounts
To access and operate the USM Vision+, a typical user will need an assigned user
name and password. System administrators must create the list of user names and
passwords on a PC, and transfer them to the USM Vision+, either directly or from a
USB memory stick. For initial setup purposes, system administrators will receive an
active Administrator account and the USM Vision+ Guest Account.
A.1 Setting up User Accounts
1. From the Windows Start Menu on your PC, click on Settings>Control Panel.
When the Control Panel window (see Figure 108 below) opens, click on User
Accounts.
Figure 108: Windows Control Panel
USM Vision+ User’s Manual
127
Appendix A. Creating User Accounts
A.1 Setting up User Accounts (cont.)
2. When the User Accounts window (see Figure 109 below) opens, click Create a
New Account.
Figure 109: User Accounts Window
3. Select a name for the new account (Figure 110 below). This name will identify
the operator in all inspection tasks, and will be stored with all the user actions on
USM Vision+.
Figure 110: Naming the Account
128
USM Vision+ User’s Manual
Appendix A. Creating User Accounts
A.1 Setting up User Accounts (cont.)
4. Select an account type (see Figure 111 below) and click Create Account.
Figure 111: Account Type
5.
After creating the account you can create an initial password or change the
picture of the account. Detailed instructions are available from Microsoft at:
http://www.microsoft.com/windowsxp/using/setup/winxp/accounts.mspx
Figure 112: Changing an Account
USM Vision+ User’s Manual
129
Appendix A. Creating User Accounts
A.2 Setting up a User’s Access Rights
1. To set up a user’s access rights, return to the Windows Control Panel and click
on Administrative Tools (see Figure 113 below).
Figure 113: Control Panel (Administrative Tools Icon)
2. From the Administrative Tools window, click on Computer Management (see
Figure 114 below).
Figure 114: Administrative Tools (Computer Management Icon)
130
USM Vision+ User’s Manual
Appendix A. Creating User Accounts
A.2 Setting up a User’s Access Rights (cont.)
3. In the Computer Management option tree, expand the folder Local Users and
Groups (see Figure 115 below).
Figure 115: Local Users and Groups in Computer Management
USM Vision+ User’s Manual
131
Appendix A. Creating User Accounts
A.2 Setting up a User’s Access Rights (cont.)
4. Then, click on the Groups folder (see Figure 116 below).
Figure 116: The Groups List
5. From the Actions Menu, click New Group. You must create the following
groups to set access rights in USM Vision+:
• KisGuest: Guest logon that requires authentication before using the
instrument.The access rights are similar to KISScanoperator.
• KisScanOperator: This group is allowed to operate the instrument.
• KisPlanCreator: This group is allowed to create and validate inspection
plans as well as operate the instrument.
• KisProcAdmin: Besides operating the USM Vision+ and creating and
validating inspection plans, this group is also allowed to create and to revise
inspection procedures.
132
USM Vision+ User’s Manual
Appendix A. Creating User Accounts
A.2 Setting up a User’s Access Rights (cont.)
6. To assign a user to a specific group, return to Local Users and Groups (see
Figure 117 below).
Figure 117: The Users List
7. Select Users and then double-click on the respective user. The Properties
window (see Figure 118 below) for that user opens.
Figure 118: User Properties Window
USM Vision+ User’s Manual
133
Appendix A. Creating User Accounts
A.2 Setting up a User’s Access Rights (cont.)
8. To change group membership, click on the tab Member Of (see Figure 119
below). Click the Add button to add the member to a group.
Figure 119: Member Tab
9. When the Select Groups window opens (see Figure 120 below), click the
Advanced button.
Figure 120: Select Groups Window
134
USM Vision+ User’s Manual
Appendix A. Creating User Accounts
A.2 Setting up a User’s Access Rights (cont.)
10. Click Find Now and the Select Groups window opens (see Figure 121 below).
Then, select the correct group for the user and click OK.
Note: You can always use KisGuest with any account, but it probably makes most
sense to associate a user with KisScanOperator.
Figure 121: Select Groups with Group Listing
USM Vision+ User’s Manual
135
Appendix A. Creating User Accounts
A.2 Setting up a User’s Access Rights (cont.)
11. With the appropriate group highlighted in the Select Groups window (see
Figure 122 below), click OK. The user is now part of the selected group.
Figure 122: Select Groups Window with Highlighted Group
136
USM Vision+ User’s Manual
Appendix B. Calibrating the Touchscreen
Appendix B. Calibrating the Touchscreen
When you receive your USM Vision+, the touchscreen is calibrated. However, if
you need to recalibrate the touchscreen, complete the steps in this appendix.
B.1 Recalibrating the Touchscreen
1.
Start the USM Vision+ as an Administrator. From the USM Vision+ Control
Panel, launch the touch-base control panel applet by double-clicking the
Pointer Devices icon, as shown in Figure 123 below.
Figure 123: Control Panel with Pointer Devices Icon
USM Vision+ User’s Manual
137
Appendix B. Calibrating the Touchscreen
B.1 Recalibrating the Touchscreen (cont.)
The window shown in Figure 124 below opens.
Figure 124: The Pointer Device Properties Window
2.
To initiate the calibration, touch the Calibrate button to open the screen shown
in Figure 125 below.
Figure 125: Calibration Screen
138
USM Vision+ User’s Manual
Appendix B. Calibrating the Touchscreen
B.1 Recalibrating the Touchscreen (cont.)
3.
Go through all the points, touching each arrow tip or cross center as it appears.
Try to avoid parallax. When you have finished, the window shown in
Figure 126 below opens. Touch OK to confirm your calibration.
Figure 126: Calibration Confirmation Screen
USM Vision+ User’s Manual
139
Appendix B. Calibrating the Touchscreen
[no content intended for this page]
140
USM Vision+ User’s Manual
Appendix C. Environmental Compliance
Appendix C. Environmental Compliance
C.1 Waste Electrical and Electronic Equipment Directive
GE Measurement & Control is an active participant in Europe’s Waste Electrical
and Electronic Equipment (WEEE) take-back initiative, directive 2012/19/EU.
The equipment that you bought has required the extraction
and use of natural resources for its production. It may contain
hazardous substances that could impact health and the
environment.
In order to avoid the dissemination of those substances in our
environment and to diminish the pressure on the natural
resources, we encourage you to use the appropriate take-back
systems. Those systems will reuse or recycle most of the
materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please
contact your local or regional waste administration.
Visit www.ge.com/inspectiontechnologies for take-back instructions and more
information about this initiative.
C.2 Battery Disposal
This product contains a battery that cannot be disposed of as
unsorted municipal waste in the European Union. See the
product documentation for specific battery information. The
battery is marked with this symbol, which may include
lettering to indicate cadmium (Cd), lead (Pb), or mercury
(Hg). For proper recycling return the battery to your supplier
or to a designated collection point.
USM Vision+ User’s Manual
141
Appendix C. Environmental Compliance
C.2.1 What do the Markings Mean?
Batteries and accumulators must be marked (either on the battery or accumulator or
on its packaging, depending on size) with the separate collection symbol. In
addition, the marking must include the chemical symbols of specific levels of toxic
metals as follows:
• Cadmium (Cd) over 0.002%
• Lead (Pb) over 0.004%
• Mercury (Hg) over 0.0005%
C.2.2 The Risks and Your Role in Reducing Them
Your participation is an important part of the effort to minimize the impact of
batteries and accumulators on the environment and on human health. For proper
recycling you can return this product or the batteries or accumulators it contains to
your supplier or to a designated collection point.
Some batteries or accumulators contain toxic metals that pose serious risks to
human health and to the environment. When required, the product marking includes
chemical symbols that indicate the presence toxic metals: Pb for lead, Hg for
mercury, and Cd for cadmium.
• Cadmium poisoning can result in cancer of the lungs and prostate gland.
Chronic effects include kidney damage, pulmonary emphysema, and bone
diseases such as osteomalcia and osteoporosis. Cadmium may also cause
anemia, discoloration of the teeth, and loss of smell (anosmia).
• Lead is poisonous in all forms. It accumulates in the body, so each exposure is
significant. Ingestion and inhalation of lead can cause severe damage to human
health. Risks include brain damage, convulsions, malnutrition, and sterility.
• Mercury creates hazardous vapors at room temperature. Exposure to high
concentrations of mercury vapor can cause a variety of severe symptoms. Risks
include chronic inflammation of mouth and gums, personality change,
nervousness, fever, and rashes.
Visit www.ge.com/inspectiontechnologies for take-back instructions and more
information about this initiative.
142
USM Vision+ User’s Manual
Appendix D. Glossary
Appendix D. Glossary
Several acronyms are used throughout this manual. Although these acronyms are
commonly used in the ultrasonic flaw detection industry, they are listed in Table 36
below for convenient reference.
Acronym
Table 36: Common Acronyms
Meaning
ACG
Angle Corrected Gain
DAC
Distance Amplitude Correction
DLC
Delay Law Calculator
ERS
Equivalent Reflector Size
FBH
Flat Bottom Hole
IP
Inspection Plan or Initial Pulse
MDI
Menu Directed Inspection
PA
Phased Array
PCF
Probe Center Separation
PRF
Pulse Repetition Frequency
SDH
Side Drilled Hole
SNR
Signal to Noise Ratio
TCG
Time Corrected Gain
TOFD
Time Of Flight Diffraction
UT
Ultrasonic Testing
V, X, J
Three different standard weld types
USM Vision+ User’s Manual
143
Appendix D. Glossary
[no content intended for this page]
144
USM Vision+ User’s Manual
Index
B
Base Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15, 16
Batteries, Inserting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Battery
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
C
Color of screen
Changing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Color Scheme . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Conventional Channel
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
D
Data entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Date and Time, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15, 16
Decimal option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Decimal point style, changing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
E
Environmental Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
I
Initialization screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Input devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
L
Language, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15, 16
M
March 2010. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .i
Measurement Units, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15, 16
Measurement Units, Switching . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
S
Specifications
Conventional Channel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
System Information screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
U
Unit Switching . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
USM Vision
Function of. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Inserting batteries. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
USM Vision+ User’s Manual
145
Index
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
System Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Typical Application for . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Unpacking. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
W
Waste Disposal
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Electronic Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
WEEE Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
146
USM Vision+ User’s Manual
Customer Support Centers
GE Sensing & Inspection Technologies GmbH
Service-Center
Robert-Bosch-Stra?e 3
50354 Hürth
Germany
or:
Postfach 1363
50330 Hürth
Germany
T +49 (0) 22 33 601 111
F +49 (0) 22 33 601 402
France
GE Measurement & Control France SAS
68 Chemin des Ormeaux
69578 Limonest Cedex
France
T +33 (0) 472 179 220
F +33 (0) 472 179 237
Great Britain
GE Inspection Technologies Ltd.
Building 4
Leicester Road
Rugby, CV21 1BD
UK
T +44 (0) 845 601 5771
USA
GE Inspection Technologies, LP
50 Industrial Park Road
Lewistown, PA 17044
USA
T +1 717 242 03 27
F +1 717 242 26 06
E-mail: [email protected]
www.gemeasurement.com
©2015 General Electric Company.
All rights reserved. Specifications subject to change without notice.
110N1532 Rev. 3
Printed in USA

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

Download PDF

advertisement