User's Guide | Dome Sx User`s guide

Add to my manuals
52 Pages

advertisement

User's Guide | Dome Sx User`s guide | Manualzz

Dome

®

Sx Display

User’s Guide

0197

Dome S3c Display

Dome S10 Display

© 2014 NDS Surgical Imaging. All rights reserved.

Information in this document has been carefully checked for accuracy; however, no guarantee is given to the correctness of the contents. This document is subject to change without notice. Go to web site ndssi.com for the latest documentation. NDSsi provides this information as reference only. Reference to products from other vendors does not imply any recommendation or endorsement.

This document contains proprietary information protected by copyright. No part of this manual may be reproduced by any mechanical, electronic, or other means, in any form, without prior written permission of NDSsi.

NDS Surgical Imaging, NDSsi, the NDSsi logo, Dome, the Dome by NDSsi logo,

Dome S3, Dome S3c, Dome S6c, Dome S10, Dome CXtra, RightLight are either registered or unregistered trademarks of NDS Surgical Imaging. All other trademarks are the property of their respective owners.

020-0931-00 Rev. F

Contents

Display Package Contents . . . . . . . . . . . . . . . . . . . . . . . . . . 4

About the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Related Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Display Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Installing the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10

Technical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Important recycle instruction. . . . . . . . . . . . . . . . . . . 12

Symbol explanations. . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Unpacking and handling tips . . . . . . . . . . . . . . . . . . . 17

Preventing fire and injury . . . . . . . . . . . . . . . . . . . . . . 18

Cleaning the display . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Shipping/storing the display . . . . . . . . . . . . . . . . . . . 19

Disposal information . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Dome S3 Specification . . . . . . . . . . . . . . . . . . . . . . . . . 20

Dome S3c Specification . . . . . . . . . . . . . . . . . . . . . . . . 22

Dome S6c Specification . . . . . . . . . . . . . . . . . . . . . . . . 24

Dome S10 Specification . . . . . . . . . . . . . . . . . . . . . . . . 26

Windows Video Modes . . . . . . . . . . . . . . . . . . . . . . . . . 28

Power Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

LED Status Lights . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

Regulatory Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33

Warranty and Service Terms and Conditions. . . . . . . . .44

Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50

3

Display Package Contents

The Dome S10 display, shown here for illustrative purposes. Your display package may contain a Dome S3, Dome S3c, Dome S6c, or Dome S10 unit.

DC power adapter Power cord

DVI cable DisplayPort cable Mini USB cable Quick reference

NOTE: DVI-to-DVI and DisplayPort cables ship in the Dome S3 and Dome S3c display packages. DVI-to-DVI cables ship in the Dome S6c display package. DVI and mini USB cables ship in the Dome S10 display package.

4 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

About the Display

The Dome® Sx display systems, available in 3, 6, and 10 megapixels, consist of a TFT LCD panel and a graphics board and driver.The display’s thin film transistors, in a transmissive-type display, use an integrated cold cathode fluorescent tube (CCFT) backlight system.

The new model of the Dome S3c display and the Dome S6c display both use an LED backlight system.

The display is designed for medical imaging in diagnostic settings and comes fully tuned with gamma correction that complies with the

DICOM Part 14 Standard. The all-digital design enables the display to produce the sharp, crisp images critical to softcopy medical viewing.

Unwanted analog display image artifacts are eliminated.

For display classification and use; environmental guidelines; safety precautions; handling, cleaning and storage tip; and specifications,

see “Technical Information” on page 12.

System Requirements

• Windows 7 (x86 or x64) or Windows XP Service Pack 3 (x86 or x64)

• Intel or AMD multi-core processor

• PCI Express x16 lane slot

• 50 MB hard disk space

• 512 MB RAM

• CD-ROM drive

• Power supply, 350 watts or greater

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 5

Related Documents

ITEM

Dome CXtra

Quick Reference

ITEM NUMBER

020-0914-xx

Dome CXtra

User's Guide

MQSA Quality

Control Manual

020-0911-xx

020-0517-xx

Graphics Board &

Driver Installation

Quick Reference

020-0940-xx

DESCRIPTION

Quick reference booklet that covers installation of CXtra calibration software and describes DICOM, AAPM TG18, DIN, and MQSA quality control functions.

(Printed material in Dome CXtra CD

015-0081-xx envelope.)

In-depth user's guide that covers installation and configuration of CXtra calibration software and use of DICOM,

AAPM TG18, DIN, and MQSA quality control functionalities. (PDF file on

Dome CXtra CD 015-0081-XX.)

Instructional guide for performing mammography quality control tests on

Dome 5MP and 10MP displays. (PDF file on Dome CXtra CD 015-0081-XX.)

Quick reference booklet for installation of graphics board and driver. (Printed material in Dome Display Driver CD

012-0043-XX envelope.)

6 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Display Components

Review the following illustrations to identify controls and ports on the front and back panels of the Dome Sx displays.

Front sensor on Dome S3/Dome S3c MQSA button on Dome S10

The sensor on the front of the Dome S3, Dome S3c, and Dome S6c displays allows you to run an auto-conformance test without using the photometer.

On the S3 series the sensor is located on the side bezel as shown in the preceding illustration.

On the Dome S6c display, the sensor is underneath the front bezel at the bottom center.

The MQSA (Mammography Quality Standards Act) button (lower right, beneath display ID) on the Dome S10 display facilitates any daily, weekly, or annual tests that you have preset via calibration and monitoring software.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 7

Dome S3/Dome S3c connector plate

Dome S6c connector plate

Dome S10 connector plate

The Dome Sx displays include the following components on the connector plate.

• Power input.

– On Dome S3, drives power to the display via an 8-pin DIN connector. The power input is 12V ±5% (130 W).

– On Dome S3c, Dome S6c, and Dome S10, drives power to the display via a 6-pin DIN connector. The power input is 24V ±5%

(150 W).

• USB port.

– Dome S3/Dome S3c uplink B port. Usage to be defined at a future time.

– Dome S6c mini AB port. For service use only.

– Dome S10 uplink mini-B port. Usage assigned for MQSA button functionality.

8 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

• Video input. Drives the data to the display.

– Dome S3 and Dome S3c displays each offer two video connector options: a DVI connector or a DisplayPort connector.

– Dome S6c and Dome S10 displays each offer two DVI connectors.

(NOTE: For the Dome S6c display, a DisplayPort 1.1 video output port and a DisplayPort 1.2 video input port are future features.)

• Reset button. Restores the display configuration to default setting.

• LED lights. Provides information on the status of the display.

• Network connection port. Future feature of the Dome S6c display.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 9

Installing the Display

First, turn your computer off. Next, leave the power cord plugged into the grounded outlet. Then install the Dome Sx display using only the power adapter and video cable that shipped with the unit.

WARNING

No modification of this equipment is allowed.

In locations where 240V outlets are used, connect the

Dome Sx display only to a center-tapped, 240V, singlephase supply (for Canada and the United States only).

Connecting the Video Cable and Power Cord

IMPORTANT! Note the following:

• For the Dome S3 or Dome S3c display only: Install either the DVI or the DisplayPort cable, not both.

• If the power adapter that came with your display unit has an on/off switch, after installation, turn on the adapter before you turn on the computer.

To connect the cables

1 Plug the power cord of the adapter into the power input port on the connector plate, then plug the other end into the input socket of the adapter. Plug the power supply cord into a grounded AC outlet.

2 Plug one end of the video cable into the primary video port on the connector plate and secure the connection. Plug the other end of the cable into the video port on the installed graphics card and secure the connection. Repeat this step for a secondary video cable installation.

3 Turn on the adapter power before you turn on the computer.

10 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Video Cable Installation on Dome S6c and Dome S10 Displays

For both the Dome S6c and the Dome S10 displays, connect the primary output from the graphics board (the bottom port) to the primary input of the Dome S6c and the Dome S10 displays. (See following illustration for the location of the primary input port on the connector plate of each display.) Improper connection may cause VGA boot messages not to appear on the display screen.

Primary video signal input on the Dome S6c connector plate

Primary video signal input on the Dome S10 connector plate

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 11

Technical Information

The design of the Dome® Sx digital display takes into account every known measure to ensure your personal safety. Improper use of the display can result in electric shock, fire, or damage to the display.

Read all instructions before setting up the display.

Classification

Shock Protection: Class I.

Degree of Protection Against Electric Shock: No applied part.

Degree of Protection Against Harmful Ingress of Water:

Ordinary equipment (IPX0).

Degree of Safety in the Presence of Flammable Anaesthetic

Mixture with Air or with Oxygen or Nitrous Oxide:

Not suitable for use in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide.

Mode of Operation: Continuous.

No applied part.

Important recycle instruction

HAZARDOUS SUBSTANCE. Lamp(s) inside this product contain mercury. This product may contain other electronic waste that can be hazardous if not disposed of properly. Recycle or dispose in accordance with local, state, or federal laws.

For more information, contact the Telecommunications

Industry Association at

WWW

.

ECYCLINGCENTRAL

.

COM

. For lampspecific disposal information, check

WWW

.

LAMPRECYCLE

.

ORG

.

DISPOSAL. Do not use household or municipal waste collection services for disposal of electrical and electronic equipment. EU countries require the use of separate recycling collection services.

ENVIRONMENT-FRIENDLY USE PERIOD. Hazardous substances are present. The number encircled by the recycling symbol indicates the safe-use period (in years).

China requires the use of recycling services at the end of product life.

12 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Symbol explanations

READ DOCUMENTATION. Follow operating instructions or consult instructions for use.

MANUFACTURER.

DATE OF MANUFACTURE.

SERIAL NUMBER.

AUTHORIZED REPRESENTATIVE. European

Community.

DANGEROUS VOLTAGE. Important precautions about electric shock. Read the accompanying text carefully, to prevent damage to display components and to guard your safety.

DIRECT CURRENT.

BAROMETRIC PRESSURE. Transport and storage

12,000 meters (39,400 feet), maximum in unpressurized container.

RELATIVE HUMIDITY. Transport and storage 5% to

90% (noncondensing) for Dome S3, Dome S3c, and

Dome S10 displays; 10% to 90% (noncondensing) for the Dome S6c display.

TEMPERATURE. Transport and storage -20° to 60° C.

WARNING.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 13

The Dome S series offers AMLCD units designed for viewing medical

X-ray images. A Dome Sx display should not be used near patients and should be kept outside of 1.5 m perimeter and 2.29 m vertical.

Intended use

Dome Sx line, models Dome S3, Dome S3c, and Dome S6c are intended for use in displaying and viewing radiography images for review, analysis, and diagnosis by trained medical practitioners.

Dome Sx line, model Dome S10 is intended for use in displaying and viewing radiography images for review and analysis by trained medical practitioners and for use in mammography display systems.

WARNING! In the Dome S series, only the Dome S10 display can be used for review and analysis in mammography. Do not use the

Dome S3, Dome S3c, or Dome S6c display in mammography diagnosis.

Federal law restricts this device to sale by or on the order of a medical practitioner.

14 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Safety precautions

External equipment intended for connection to signal input, signal output, or other connectors, must comply with the relevant IEC standard (EN 60601-1:2006 and IEC 60601-1:2005 series for medical electrical equipment). In addition, all such combinations (systems) must comply with the standard IEC 60601-1:2005, Safety requirements for medical electrical systems.

Equipment not complying to IEC 60601-1:2005 must be kept outside the patient environment, as defined in the standard as at least 1.5 meters from the patient or the patient support.

Any person who connects external equipment to signal input, signal output, or other connectors has formed a system and is therefore responsible for the system to comply with the requirements of

IEC 60601-1:2005. If in doubt, speak with a qualified technician.

The power cord is the only disconnect device that allows you to cut power from the AC main. Position the display in order to operate the disconnect device without difficulty.

Never touch the display and the patient at the same time.

WARNING! No modification of this equipment is allowed.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 15

Safety tips

• Never open the display case, even when the power is off.

Dangerous voltage inside may cause electric shock or death.

• To avoid damage to the display, use the grounded power supply and video cable supplied by NDSsi, or use certified replacements.

• Be sure the display is electrically grounded. You must connect the third grounding pin on the US power cord to a grounded outlet.

The European power cord does not have a third grounding pin, but it must be plugged into a grounded outlet.

• If you cannot insert the plug into the outlet you plan to use, have a licensed electrician replace the outlet with a properly grounded outlet. If the power cord connects directly into the computer, make sure the computer is grounded.

• Keep the display dry if it is part of a surgical system. The display lacks protection against liquids or spills.

• In locations where 240V outlets are used, connect the

Dome Sx display only to a center-tapped, 240V, singlephase supply (for Canada and the United States).

GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN

EQUIPMENT IS CONNECTED TO AN EQUIVALENT RECEPTABLE

MARKED “HOSPITAL ONLY” OR “HOSPITAL GRADE.”

WARNING! TO AVOID THE RISK OF ELECTRIC SHOCK, THIS EQUIPMENT

MUST ONLY BE CONNECTED TO A SUPPLY MAINS WITH PROTECTIVE

EARTH.

MAKE THE POWER SUPPLY CONNECTIONS FIRST, BEFORE YOU PLUG

IN ANY OTHER CABLES. THEN DISCONNECT THE POWER SUPPLY LAST,

AFTER YOU UNPLUG ALL OTHER CABLES.

16 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Unpacking and handling tips

The Dome Sx display is a precision instrument that requires proper care to maintain product operation and adherence to specification.

Unpack the display and components carefully, then set up and handle the unit properly to avoid damage to the LCD panel.

• Use both hands to grasp the display case when lifting it from the shipping carton, but avoid touching the screen.

• Do not apply pressure to the screen or touch the screen with bare fingers or objects. Pressure can affect image quality. Cosmetics and oils on the skin are both detrimental to the screen and difficult to remove.

• Allow the display to warm up to room temperature before turning it on. Avoid sudden temperature changes in the environment, as this may cause condensation, which damages the display.

• Secure the display properly onto a standard VESA 100-mm mounting unit if you elect not to use the desk stand.

• Do not set up the display near strong light or heat sources.

• Do not block the vents on the back of the display or install the display in a built-in enclosure. Blocked vents cause excessive heat to build up inside the display, increasing risk of fire.

• When installing components, turn off your computer, but leave it plugged into a grounded outlet.

• Do not remove the back cover or disassemble the display. There are no user-serviceable parts inside.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 17

Preventing fire and injury

• Replace the power supply or cables if damaged.

• Use only the power source indicated in this guide or listed on the display.

• Do not plug the power supply into an overloaded AC outlet or extension cord. Overloaded AC outlets and cords can result in electric shock or fire.

• Do not drop or push objects into the display case. Internal components contain high voltage.

• Unplug the power cord from the wall outlet during thunderstorms.

• Do not place magnetic devices, such as motors, near the display.

Cleaning the display

Observe the following guidelines to maintain the display and the LCD screen.

• Use a clean, lint-free, absorbent cotton cloth to clear off any residual glue from removal of the protective film or to remove surface dust. Apply light pressure to remove the dust.

• Dampen a clean cloth with a small amount of isopropyl alcohol to remove glue or dust if the screen is still not clean. Do not saturate the cloth; otherwise, alcohol may seep into the display case and collect in the enclosure. Use a clean, dry cloth to completely remove the alcohol residue.

• Do not use chemically treated dust cloths, acetone, toluene, or harsh solvents on the display case or the screen. They can damage the polarizer and the display case.

• Do not expose the display to water or excessive moisture. Do not allow water or other stains to stand on the unit. Wipe liquids off immediately to prevent damage to the display case and the screen.

18 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

WARNING! DISCONNECT SUPPLY BEFORE SERVICING

AVERTISSEMENT

COUPER L’ALIMENTATION AVANT L’ENTRETIEN ET LE DEPANNAGE

Shipping/storing the display

Keep the display in its shipping container until installation. Return the display to its original container whenever you need to store the unit, move it to another location, or return it for repair. The packaging supplied by the manufacturer protects the display while it is in transit.

See environment specifications for more information.

Before returning the display to the container, do the following:

1 Swivel the display panel to landscape mode.

2 Push the panel down to the lowest position.

3 Use the stand lock to anchor the panel.

Disposal information

Follow your local governing ordinance and recycling policy for proper disposal or recycling of display components.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 19

Dome S3 Specification

In locations where 240V outlets are used, connect the Dome S3 display to a center-tapped,

240V, single-phase supply only (for Canada and the United States).

Category

LCD

Optical

Interface

Panel

Connectivity

Characteristic Item

Technology

Screen size, diagonal

Display resolution

Pixel pitch

Viewing angle

Response time (Tr+Tf)

Refresh rate

Brightness

Contrast ratio

DICOM LUT

Colors supported

Pixel arrangement

Display control -brightness/ contrast

Display identification

Display status

Video input

Pixel data sizes

Specification

TFT AMLCD Dual Domain IPS,

3 subpixels per pixel

541 mm (21.3 in.)

1536 x 2048 pixels (portrait)

2048 x 1536 pixels (landscape)

211.5 m, 120 DPI

176° horizontal/vertical

27 ms (typical)

60 Hz

1450 cd/m

2

(typical)

900:1 (typical)

16-bit architecture with 0.2 JND precision

Up to 1024 shades of gray, programmable gamma from palette of up to 3061 shades

LCR vertical stripe

DDC2B+

Physical

Power

Requirement

VGA compatibility

USB interface

Display size (without stand),

H x W x D

Display weight (without stand)

Display weight (with stand)

Mounting options

Adapter

EDID read using DDC2B+

Dual-stack tricolor LEDs on back panel

DisplayPort 1.1 and Dual Link DVI-D

8-bit, 10-bit per pixel grayscale,

24-bit color mapped to luminance

640 x 480 to 1280 x 1024

USB B upstream port

475 mm x 363 mm x 99 mm

(18.7 in. x 14.3 in. x 3.9 in.)

6.4 kg (14 lb)

9.3 kg (21 lb)

100 x 100 mm VESA mounting standard.

Desktop stand includes height, pivot, swivel, and tilt adjustment.

BridgePower BPM130S12F02 (with power switch)

(CAUTION: Use only the adapter supplied with unit.)

20 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Power Supply

WARNING: Use only the power adapter supplied with the

Dome S3 display unit; the BridgePower BPM130S12F02 (130W) model.

Category Characteristic Item Specification

Power input requirements

Voltage selection

Voltage

Current

Power output requirements

Frequency

Voltage

Current

Power consumption Wattage

Physical Size

Weight

Universal auto switching with medical approvals and PFC

100–240V AC

3.0 A (3.0–1.5 A)

50–60 Hz

12 V DC ± 5%

10.8 A (130 W)

70 W (typical)

228.6 mm x 76.2 mm x 50.8 mm

(9 in x 3 in x 2 in)

1.3 kg (2.75 lb)

Reliability

Characteristic item Specification

Display

Backlight

MTBF > 50,000 hours

MTBF 50,000 hours to 50% brightness with backlight on continuously

Environment

Characteristic item

EMI shielding

Temperature

Humidity

Altitude

Operating range

Storage range

Operating range

Storage range

Operating

Transport/storage

Specification

No emission of low-level radiation

10° C ~ 40° C

-20° C ~ 60° C

30% ~ 75% Relative Humidity

(noncondensing)

5% ~ 90% Relative Humidity

(noncondensing)

2,000 m (6,560 ft) maximum

12,000 m (39,400 ft) maximum in unpressurized container

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 21

Dome S3c Specification

In locations where 240V outlets are used, connect the Dome S3c display to a centertapped, 240V, single-phase supply only (for Canada and the United States).

Category

LCD

Optical

Interface

Characteristic Item

Technology

Screen size, diagonal

Display resolution

Pixel pitch

Viewing angle

Response time (Tr+Tf)

Refresh rate

Brightness

Contrast ratio

DICOM LUT

Colors supported

Specification

TFT AMLCD Dual Domain IPS,

3 subpixels per pixel

541 mm (21.3 in.)

1536 x 2048 pixels (portrait)

2048 x 1536 pixels (landscape)

211.5 m, 120 DPI

176° horizontal/vertical

40 ms LED backlight (typical)

24 ms CCFL backlight (typical)

60 Hz

800 cd/m

2

(typical)

1400:1 LED backlight (typical)

750:1 CCFL backlight (typical)

16-bit architecture with 0.3 JND precision

Up to 1.06 billion colors; up to 1024 shades of gray, programmable gamma from palette of up to

2041 shades

R,G,B vertical stripe

DDC2B+

Panel

Connectivity

Physical

Pixel arrangement

Display control -brightness/ contrast

Display identification

Display status

Video input

Pixel data sizes

VGA compatibility

USB interface

Display size (without stand),

H x W x D

Display weight (without stand)

Display weight (with stand)

Mounting options

Power

Requirement

Adapter

* 30-bit color is supported on DisplayPort only.

EDID read using DDC2B+

Dual-stack tricolor LEDs on back panel

DisplayPort 1.1 and Dual Link DVI-D

8-bit, 16-bit, 24-bit, and 30-bit per pixel color

*

640 x 480 to 1280 x 1024

USB B upstream port

475 mm x 363 mm x 99 mm

(18.7 in. x 14.3 in. x 3.9 in.)

6.4 kg (14 lb)

9.3 kg (21 lb)

100 x 100 mm VESA mounting standard.

Desktop stand includes height, pivot, swivel, and tilt adjustment.

BridgePower BPM150S24F05 (with power switch)

(CAUTION: Use only the adapter supplied with unit.)

22 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Power Supply

WARNING: Use only the power adapter supplied with the

Dome S3c display unit; the BridgePower BPM150S24F05 (150W) model.

Category Characteristic Item Specification

Power input requirements

Voltage selection

Voltage

Current

Power output requirements

Frequency

Voltage

Current

Power consumption Wattage

Physical Size

Weight

Universal auto switching with medical approvals and PFC

100–240V AC

2.5 A (2.5–1.5 A)

50–60 Hz

24 V DC ± 5%

6.25 A (150 W)

66 W LED backlight (typical)

81 W CCFL backlight (typical)

223.5 mm x 63.5 mm x 38.1 mm

(8.8 in x 2.5 in x 1.5 in)

0.8 kg (1.8 lb)

Reliability

Characteristic item Specification

Display

Backlight

MTBF > 50,000 hours

MTBF 50,000 hours to 50% brightness with backlight on continuously

Environment

Characteristic item

EMI shielding

Temperature

Humidity

Altitude

Operating range

Storage range

Operating range

Storage range

Operating

Transport/storage

Specification

No emission of low-level radiation

0° C ~ 40° C

-20° C ~ 60° C

30% ~ 75% Relative Humidity

(noncondensing)

5% ~ 90% Relative Humidity

(noncondensing)

2,000 m (6,560 ft) maximum

12,000 m (39,400 ft) maximum in unpressurized container

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 23

Dome S6c Specification

In locations where 240V outlets are used, connect the Dome S6c display to a centertapped, 240V, single-phase supply only (for Canada and the United States).

Category

LCD

Optical

Characteristic Item

Technology

Screen size, diagonal

Display resolution (pixels)

Pixel pitch

Viewing angle

Response time (Tr+Tf)

Refresh rate

Brightness

Contrast ratio

DICOM LUT

Colors supported

Specification

TFT AMLCD Dual Domain IPS,

3 subpixels per pixel

762 mm (30.0 in.)

3280 x 2048 landscape

Dual 1640 x 2048 portrait

Dual 1536 x 2048 portrait (future implementation)

197 m, 129 DPI

170 ° horizontal/vertical

30 ms (typical)

50 Hz

800 cd/m

2

(typical)

1000:1 (typical)

16-bit architecture with 0.1 JND precision

Up to 1.07 billion colors; up to 1024 shades of gray, programmable gamma from palette of up to

2041 shades

Subpixel vertical stripe

DDC2B+ Interface

Panel

Connectivity

Physical

Pixel arrangement

Display control -brightness/ contrast

Display identification

Display status

Video input

Video output

Pixel data size

VGA compatibility

USB interface

Display size (without stand),

H x W x D

Display weight (without stand)

Display weight (with stand)

Mounting option

Power

Requirement

Adapter

* 30-bit color is supported on DisplayPort only.

24 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

EDID read using DDC2B+

Dual-stack tricolor LEDs on back panel

X2 Dual Link DVI-D

X1 DisplayPort 1.2 (future feature)

X1 DisplayPort 1.1 (future feature)

8-bit, 16-bit, 24-bit, and 30-bit per pixel color

*

640 x 480 to 1280 x 1024

Mini USB port (used for service)

699 mm x 473 mm x 80 mm

(27.5 in. x 18.6 in. x 3.1 in.)

10.0 kg (22 lb)

13.2 kg (29 lb)

100 mm x 100 mm VESA mounting standard.

Desktop stand capable of height, pivot, swivel, and tilt adjustment.

BridgePower Corp. BPM150S24F05 (150W) model

(CAUTION: Use only the adapter supplied with unit.)

Power Supply

WARNING: Use only the power adapter supplied with the

Dome S6c display unit; the BridgePower Corp. BPM150S24F05

(150W) model. Note that adapter has a power switch.

Category Characteristic Item Specification

Power input requirements

Voltage selection

Voltage

Current

Power output requirements

Frequency

Voltage

Current

Power consumption Wattage

Physical Size

Weight

Universal auto switching with medical approvals and PFC

100–240V AC

2.5 A (2.5–1.5 A)

50–60 Hz

24 V DC ± 5%

6.25 A (150 W)

85 W (typical)

223.5 mm x 63.5 mm x 38.1 mm

(8.8 in x 2.5 in x 1.5 in)

0.8 kg (1.8 lb)

Reliability

Characteristic item Specification

Display

Backlight

MTBF > 50,000 hours

MTBF 50,000 hours to 50% brightness with backlight on continuously

Environment

Characteristic item

EMI shielding

Temperature

Humidity

Altitude

Operating range

Storage range

Operating range

Storage range

Operating

Transport/storage

Specification

No emission of low-level radiation

5° C ~ 40° C

-20° C ~ 60° C

20% ~ 90% Relative Humidity

(noncondensing)

10% ~ 90% Relative Humidity

(noncondensing)

2,000 m (6,560 ft) maximum

12,000 m (39,400 ft) maximum in unpressurized container

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 25

Dome S10 Specification

In locations where 240V outlets are used, connect the Dome S10 display to a centertapped, 240V, single-phase supply only (for Canada and the United States).

Category

LCD

Optical

Interface

Characteristic Item

Technology

Screen size, diagonal

Display resolution

Pixel pitch

Viewing angle

Response time (Tr+Tf)

Refresh rate

Brightness

Contrast ratio

DICOM LUT

Colors supported

Pixel arrangement

Display control -brightness/ contrast

Display identification

Display status

MQSA

Specification

TFT AMLCD Dual Domain IPS, single 10-bit pixels

761 mm (30.0 in.)

4096 x 2560 pixels (landscape)

157 m, 161 DPI

170° horizontal/vertical

35 ms (typical)

50 Hz

1250 cd/m

2

(typical)

850:1 (typical)

16-bit architecture with 0.2 JND precision

Up to 1024 shades of gray, programmable gamma from palette of up to 4096 shades of gray

Single pixel, IPS structure

DDC2B+

Panel

Connectivity

Physical

Power

Requirement

Video input

Pixel data sizes

VGA compatibility

USB interface

Display size (without stand),

H x W x D

Display weight (without stand)

Display weight (with stand)

Mounting options

Adapter

EDID read using DDC2B+

Dual-stack tricolor LEDs on back panel

2 color bottom facing LEDs on front of unit with push-button activation

2 Dual Link DVI-D

8-bit and 10-bit per pixel grayscale,

24-bit color mapped to luminance

640 x 480 to 1280 x 1024

USB B upstream port

699 mm x 473 mm x 80 mm

(27.5 in. x 18.6 in. x 3.1 in.)

7.7 kg (17 lb)

10.9 kg (24 lb)

100 x 100 mm VESA mounting standard.

Desktop stand includes height, pivot, swivel, and tilt adjustment.

BridgePower BPM150S24F05 (with power switch)

(CAUTION: Use only the adapter supplied with unit.)

26 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Power Supply

WARNING: Use only the power adapter supplied with the

Dome S10 display unit; the BridgePower BPM150S24F05 (150W) model.

Category Characteristic Item Specification

Power input requirements

Voltage selection

Power output requirements

Voltage

Current

Frequency

Voltage

Current

Power consumption Wattage

Physical Size

Weight

Universal auto switching with medical approvals and PFC

100–240V AC

2.5A (2.5–1.5 A)

50–60 Hz

24 V DC ± 5%

6.25 A (150 W)

90 W (typical)

223.5 mm x 63.5 mm x 38.1 mm

(8.8 in x 2.5 in x 1.5 in)

0.8 kg (1.8 lb)

Reliability

Characteristic item Specification

Display

Backlight

MTBF >50,000 hours

MTBF 50,000 hours to 50% brightness with backlight on continuously

Environment

Characteristic item

EMI shielding

Temperature

Humidity

Altitude

Operating range

Storage range

Operating range

Storage range

Operating

Transport/storage

Specification

No emission of low-level radiation

10° C ~ 40° C

-20° C ~ 60° C

30% ~ 75% Relative Humidity (noncondensing)

5% ~ 90% Relative Humidity (noncondensing)

2,000 m (6,560 ft) maximum

12,000 m (39,400 ft) maximum in unpressurized container

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 27

Windows Video Modes

Resolutions are expressed in pixels, width by height (W x H).

Display Resolution Orientation

Dome S3

Dome S3c

Dome S6c

Dome S10

1536 x 2048

2048 x 1536

1536 x 2048

2048 x 1536

3280 x 2048

1640 x 2048 (dual)

1536 x 2048 (dual)

2x 2048 x 2560

(non-span mode)

4096 x 2560

(horizontal span mode)

*

32-bit palette is used for 256 and 1024 shades of gray.

† 32-bit palette is used for16.7 million and 1.06 billion colors.

‡ Future implementation

Portrait

Landscape

Portrait

Landscape

Landscape

Portrait

Portrait

Portrait

Landscape

Color Palette

32-bit

*

32-bit

32-bit

32-bit*

28 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Power Input

Power is supplied to DomeS3c, Dome S6c, and Dome S10 displays via the 6-pin DIN port on the connector plate, and via an 8-pin DIN port on the Dome S3 display. See pin assignments on the following illustrations.

For Dome S3c, Dome S6c, and dome S10, the power input is 24V ±5% (150 W); for

Dome S3, 12V ±5% (130 W).

IMPORTANT! Use only the power adapter that ships with the display unit.

Display

Dome S3

DIN Connector Power Input

8-pin 12V ±5% (130 W)

Dome S3c

Dome S6c

Dome S10

6-pin

6-pin

6-pin

24V ±5% (150 W)

24V ±5% (150 W)

24V ±5% (150 W)

Adapter

BPC BPM130S12F02 (with power switch)

BPC BPM150S24F05 (with power switch)

BPC BPM150S24F05 (with power switch)

BPC BPM150S24F05 (with power switch)

VCC 8

VCC 7

VCC 6

VCC 5

VCC 6

VCC 5

VCC 4

1 GND

4 GND

2 GND

3 GND

1 GND

2 GND

3 GND

8-pin connector on Dome S3 display

6-pin connector on Dome S3c, Dome S6c, and

Dome S10 displays

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 29

LED Status Lights

For Dome S3, Dome S3c, and Dome S10, the two LEDs on the back panel provide information on the status of the display.

• LED A describes the digital-link status between the display controller and the interface.

• LED B describes the display-panel status. it shows any faults currently in the panel.

Power-up sequence information from LED

LED A

Dark

Blink yellow

Dark

Action/Sequence LED B

Initial power-on

Self test

Flashing yellow

Solid yellow

One (1) second after power -on

Flashing green

Action/Sequence

Initial power-on

Power-on self test

Self test pass

LED A status information after initial power-on

LED Status

Solid green

Flashing yellow

Description

Functional link – normal operation

Link working, unrecognized sync information

LED B status information after initial power-on

LED Status

Solid green

Flashing green

Solid yellow

Alternating green/yellow

Description

Functional system – normal operation

Fault

DDC power-on, 12V/24V power not detected

POST failure

NOTE: A red LED or combination of red LEDs indicates a hardware or software fault.

Contact Dome Technical Support for assistance.

30 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

The LEDs on the back panel of the Dome S6c display provides information about the status of video and power supply connections to the display.

LED A (left) and LED B are located between the power port and the DisplayPort port on the connector plate

The LED A set describes the status of the video connection between the display controller and the display interface.

LED A

Top indicator

Bottom indicator

Both indicators

Color

Green

Green

Unlit

Video Cable Connection

Primary DVI connector is plugged into the computer.

Secondary DVI connector is plugged into the computer.

DVI cables are not connected.

The LED B set describes the source of the power supply.

LED B

Top indicator

Bottom indicator

Color

Unlit

Yellow

Green

Power Supply

Unassigned currently.

Power supply is on from DVI1, DVI2, or USB.

External power supply is on.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 31

Troubleshooting

Problem

No image appears on the screen

Possible Cause

Computer is OFF.

Power cord is not securely connected.

Video cable connected incorrectly.

Solution

Power on the machine.

Tighten power cord connection and turn on computer.

Make sure the first display of a dualheaded system or the only display of a single-headed system is connected to the primary port on the display controller. (NOTE: Both

DVI and DisplayPort functions cannot be used at the same time.)

Remove any existing display driver before you install a more recent version.

Conflict with driver version

Unable to set the maximum or native resolution of the display

A previous version of the display driver is still installed on the computer system.

The installation CD contains a file older than the one currently on your system.

The function Hide all modes that this monitor cannot display is selected on a subdialog of the

Windows Display

Properties dialog.

A single-link commodity graphics board is in use.

The Confirm File Replace dialog box appears if the installation CD contains a file older than the one currently on the system. Click No to the question, Overwrite the newer file? Or, click No to All.

When using commodity graphics boards on Windows XP systems, make sure the check box for the function is empty (unchecked).

When using commodity graphics boards with the Dome Sx displays, be sure to install a DualLink DVI graphics board.

32 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Regulatory Compliance

Canada, European Union, United States

This display has been tested and found to comply with IEC 60601-1:2005 and

EN 60601-1:2006, and IEC/EN 60601-1-2 standards, and is certified to meet medical standard CAN/CSA C22.2 No. 60601-1-A and UL 60601-1 ANSI/AAMI

ES60601-1:2005 (R2012).

The medical display, in addition to meeting medical requirements, has been tested and found to comply with the limits for Federal

Communications Commission

Dome S3, Dome S3c, Dome S6c, and

Dome S10

(FCC) Class A computing devices in a typically configured system since many medical offices are located in residential areas. It is the system integrator’s responsibility to test and ensure that the entire system complies with applicable electromagnetic compatibility (EMC) laws.

WARNING! Users of Dome Sx Displays

The Dome S3, Dome S3c, Dome S6c, and Dome S10 displays are Class A products. In a domestic environment, either of these products may cause radio interference, in which case the user may be required to take adequate measures.

Radio transmitting equipment, cellular phones, etc. shall not be used in close proximity of the devices, since this could influence the performance of the device.

Particular precaution must be considered during use of strong emission sources, such as High Frequency surgical equipment and similar, so that, for example, the HF-cables are not routed on or near the devices. If in doubt, contact a qualified technician or your local representative.

This equipment/system is intended for use by healthcare professionals only. This equipment/system may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as reorienting or relocating the Dome Sx display or shielding the location.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 33

We offer state-of-the-art displays suitable to the European market, CE-marked displays based on compliance with counsel directive 93/42/EEC—commonly referred to as the Medical Device Directive (MDD). The following summarizes our qualification of these displays as it relates to compliance with the MDD.

The European Medical Device Directive requires that the intended use of the device be defined. Dome Sx line, models Dome S3, Dome S3c, and Dome

S6c are intended for use in displaying and viewing radiography images for review, analysis, and diagnosis by trained medical practitioners. Dome Sx line, model Dome S10 is intended for use in displaying and viewing radiography images for review and analysis by trained medical practitioners, for use in displaying and viewing radiographs of the breast for review and analysis by trained medical practitioners, or for use in diagnostic and mammography display systems. These displays do not provide a measurement function in any way, and it is the device and systems manufacturer’s responsibility to verify its function in the integrated device or system.

For Dome S3, the display was classified as required by the MDD according to

Annex IX of the directive and the medical device (MEDDEV) guidance available at the time of classification. Because the display uses electrical energy and has no direct patient connections and—by itself—no medical utility, the display is classified according to Rule 12 as an MDD Class I device, component, or accessory. The MDD states that manufacturers of Class I medical devices or accessories shall satisfy the requirements in regard to design and manufacturing controls, that is, the applicable assessment route to be used for CE-marking under the MDD, and it shall carry the CE mark according to Annex XII of the directive, with no Notified Body annotation.

For Dome S6c, the display was classified as Class I medical device required by

MDD 93/42/EEC, according to Annex IX Classification Criteria rule 12.

Furthermore, the applicable safety standards for an MDD Class I display are

EN 60601-1:2006 and IEC 60601-1:2005. To help the medical device designer evaluate the suitability of these displays, NDSsi has also conducted EMC testing to IEC 60601-1-2 as it can be applied. The display with its power supply alone does not represent a functional medical device. Hence, NDSsi configured a minimal operating system to exercise the display. The resulting data are made available to interested parties.

For Dome S3c and Dome S10 each, the display was classified as required by the MDD 93/42/EEC, according to Annex IX Classification Criteria rule 10, dash three. The above intended use for review, analysis, and diagnosis is a Class IIa medical device. Thus, we believe the Dome S3c and the Dome S10 are

Class IIa medical devices, and subject to Notified Body’s approval.

34 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Furthermore, the applicable safety standards for an MDD Class IIa display are

EN 60601-1:2006 and IEC 60601-1:2005. To help the medical device designer evaluate the suitability of these displays, NDSsi has also conducted EMC testing to IEC 60601-1-2 as it can be applied. The display with its power supply alone does not represent a functional medical device. Hence, NDSsi configured a minimal operating system to exercise the display. The resulting data are made available to interested parties.

This is informative data, not certification data. Certification data must be obtained by the device or system integrator according to Article 12 of the MDD titled “Particular procedure for systems and procedure packs.”

Paragraph 2 clearly outlines the device or system integrator’s responsibility in this matter.

In summary, NDS Surgical Imaging is CE-marking these displays under the

Medical Device Directive, which establishes compliance to the basic medical safety standards. However, EMC compliance can only be accomplished in the configured medical device or system and is the responsibility of the device or system manufacturer. NDSsi has the necessary documentation such as IEC 60601-1:2005, EN60601-1:2006, and IEC/EN60601-1-2 and thirdparty test reports and certifications, a risk management/hazard analysis, an essential requirements checklist, and NDS Surgical Imaging European

Community CE marking directive Declaration of Conformity.

NDS Surgical Imaging, located in San Jose, Calif., USA, is the manufacturer of these displays in the meaning of the directive. As required by the MDD in

Article 14, NDS Surgical Imaging, not residing in the European Economic

Area (EEA), has a European representative, NDS Surgical Imaging—located at

Nijverheidscentrum 28, 2761 JP Zevenhuizen (ZH), The Netherlands

(phone +31-180-63-43-56; fax +31-180-63-21-91).

In the opinion of NDS Surgical Imaging registration required to put this device into commerce is the responsibility of the medical device/system manufacturer, and NDS Surgical Imaging supports this requirement by providing a European Commission (EC) declaration of conformity. If NDS

Surgical Imaging supplies a display to an end user, rather than a device manufacturer, it is the end user’s responsibility to ensure continued compliance with the MDD of the system in which the display is integrated.

The supplier will make available on request, circuit diagrams, component part lists, etc.

For vigilance reporting as required under Article 10 of the MDD, NDS Surgical

Imaging will provide any information requested by competent authority to support any reported incident investigation by such an authority.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 35

EU Declaration of Conformity for Medical Application

A Declaration of Conformity has been filed for this product. For additional copies of the Declaration of Conformity document, contact NDS Surgical

Imaging.

Dome S3 and Dome S6c digital flat-panel displays meet the essential health and safety requirements of, is in conformity with, and the CE marking has been applied according to the relevant EU Directives listed below, using the relevant section of the following EU standards and other normative documents;

EU EMC Directive 2004/108/EC

EU Electromagnetic Compatibility Directive

EN 60601-1-2

Medical Electrical Equipment

Collateral standard electromagnetic compatibility requirements

EN 55011

IEC 1000-3-2

Limits and methods of measurements for radio interference characteristics of industrial, scientific, and medical equipment

Harmonic emissions

IEC 1000-3-3

IEC 1000-4-2

IEC 1000-4-3

Voltage fluctuations/flicker emissions

Electrostatic discharge requirements for industrial process measurement and control equipment

Radiated electromagnetic field requirements for industrial process measurement and control equipment

IEC 1000-4-4

IEC 1000-4-5

IEC 1000-4-11

IEC 1000-4-6

IEC 1000-4-8

Electrically fast transients for industrial process measurement and control equipment

Surge requirements

Voltage variations/dips/interrupts

Conducted immunity

Magnetic field immunity

Conformance to the Medical Device

Directive 93/42/EEC

EN 60601-1:2006

Medical Electrical Equipment

Part 1: General requirements for safety

36 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

EU Declaration of Conformity for Medical Application

A Declaration of Conformity has been filed for this product. For additional copies of the Declaration of Conformity document, contact NDS Surgical

Imaging.

The Dome S3c and Dome S10 digital flat-panel displays meet the essential health and safety requirements of, are in conformity with, and the CE marking has been applied according to the relevant EU Directives listed below, using the relevant section of the following EU standards and other normative documents;

EU EMC Directive 2004/108/EC

EU Electromagnetic Compatibility Directive

EN 60601-1-2

Medical Electrical Equipment

Collateral standard electromagnetic compatibility requirements

EN 55011 Limits and methods of measurements for radio interference characteristics of industrial, scientific, and medical equipment

IEC 1000-3-2

IEC 1000-3-3

IEC 1000-4-2

IEC 1000-4-3

Harmonic emissions

Voltage fluctuations/flicker emissions

Electrostatic discharge requirements for industrial process measurement and control equipment

Radiated electromagnetic field requirements for industrial process measurement and control equipment

IEC 1000-4-4

IEC 1000-4-5

IEC 1000-4-11

IEC 1000-4-6

IEC 1000-4-8

Electrically fast transients for industrial process measurement and control equipment

Surge requirements

Voltage variations/dips/interrupts

Conducted immunity

Magnetic field immunity

Conformance to the Medical Device

Directive 93/42/EEC

0197

EN 60601-1:2006

Medical Electrical Equipment

Part 1: General requirements for safety

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 37

U.S. FCC Compliance Statement

Class A Digital Device: Dome S3, Dome S3c, Dome S6c, or Dome S10 Display

This device complies with Part 15 of the FCC Rules.

NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment.

This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications.

Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment under FCC rules.

38 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Model Dome S series needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS;

Portable and mobile RF communications equipment can affect the Dome S series.

1. ELECTRO MAGNETIC EMISSIONS

Guidance and Manufacturer’s Declaration—Electromagnetic Emissions

The Dome S series is intended for use in the electromagnetic environment specified below.

The customer or the user of the display should assure that it is used in such an environment.

EMISSIONS COMPLIANCE ELECTROMAGNETIC ENVIRONMENT – GUIDANCE

RF emissions

CISPR 11

Group 1

RF emissions

CISPR 11

Harmonic emissions

IEC 61000-3-2

Voltage fluctuations/ flicker emissions

IEC 61000-3-3

Class A

Not Applicable

Not Applicable

The Dome S series uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

The Dome S series is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

WARNING! The use of accessories and cables other than those specified, with the exception of transducers and cables sold by the manufacturer of the

Dome S series as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the Dome S series.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 39

2. ELECTROMAGNETIC IMMUNITY

Guidance and Manufacturer’s Declaration 211; Electromagnetic Immunity

The model Dome S series is intended for use in the electromagnetic environment specified below. The customer or the user of the display should assure that it is used in such an environment.

IMMUNITY TEST

IEC 60601-1:2005

TEST LEVEL COMPLIANCE LEVEL

ELECTROMAGNETIC

ENVIRONMENT – GUIDANCE

Electrostatic discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

Voltage dips, short interruptions, and voltage variations on power supply input lines

IEC 61000-4-11

<5 % U

T

(>95 % dip in U

T for 0.5 cycle

)

40 % U

T

(60 % dip in U

T

) for 5 cycles

70 % U

T

(30 % dip in U

T

) for 25 cycles

<5 % U

T

(>95 % dip in U

T

) for 5s

Power frequency

(50/60 Hz) magnetic field IEC 61000-4-8

3 A/m

±6 kV contact

±8 kV air

Floors should be wood, concrete or ceramic tile.

If floors are covered with synthetic material, the relative humidity should be at least

30%.

Mains power quality should be that of a typical commercial or hospital environment.

Electrical fast transient/burst

IEC 61000-4-4

±2 kV for power supply lines

Surge IEC 61000-4-5 ±1 kV line(s) and neutral

±2 kV for power supply lines

±1 kV line(s) and neutral

<5 % U

T

(>95 % dip in U

T for 0.5 cycle

)

40 % U

T

(60 % dip in U

T

) for 5 cycles

70 % U

T

(30 % dip in U

T

) for 25 cycles)

<5 % U

T

(>95 % dip in U

T

) for 5s

Not applicable

Mains power quality should be that of a typical commercial or hospital environment.

Mains power quality should be that of a typical commercial or hospital environment. If a dips or an interruption of mains power occurs, the current of the model Dome S series may be dropped off from normal level, it may be necessary to use uninterruptible power supply or a battery.

Not applicable

NOTE : U

T

is the AC mains voltage prior to application of the test level.

40 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

3. ELECTROMAGNETIC IMMUNITY

Guidance and Manufacturer’s declaration–Electromagnetic Immunity

The model Dome S series is intended for use in the electromagnetic environment specified below.

The customer or the user of the Dome S series should assure that it is used in such an environment.

IMMUNITY TEST

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

IEC 60601-1:2005

TEST LEVEL

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

COMPLIANCE

LEVEL

3 Vrms

3 V/m

ELECTROMAGNETIC ENVIRONMENT – GUIDANCE

Portable and mobile RF communications equipment should be used no closer to any part of the Dome S series, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance d = 1 2 P d = 1 2 P 80 MHz to 800 MHz d = 2 3 P 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters

(m).

Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,

(a) should be less than the compliance level in each frequency range; (b) interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the model Dome S series is used exceeds the applicable RF compliance level above, the display should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the display.

(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

WARNING! The Dome S series should not be used adjacent to or stacked with other equipment.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 41

4. RECOMMENDED SEPARATION DISTANCES

Recommended Separation Distances

Between Portable and Mobile RF Communications Equipment and the Dome S Series

The model Dome S series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the display can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the displays as recommended below, according to the maximum output power of the communications equipment.

RATED MAXIMUM

OUTPUT POWER OF

TRANSMITTER

(W)

0.01

10

100

0.1

1

SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER

150 kHz to 80 MHz

0.12

0.38

1.2

3.8

12

80 MHz to 800 MHz

0.12

0.38

1.2

3.8

12

(m)

800 MHz to 2.5 GHz

0.23

0.73

2.3

7.3

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.

42 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

Each Dome Sx display has one or more of the following certifications.

China

Class A

China Compulsory Certification regulating safety and

EMC.

GB4943

GB9254

GB17625

Dome S3, Dome S3c, Dome S6c, and Dome S10

Germany

TÜV Rheinland GM Mark for medical products.

Japan

Voluntary Control Council for Interference by information technology equipment sold in Japan.

Class A Dome S3, Dome S3c, Dome S6c, and Dome S10

Class B

Korea

Korea Certification.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 43

NDSsi Dome

®

Display Products

Warranty and Service Terms and Conditions

STANDARD LIMITED WARRANTY

COVERAGE: NDS Surgical Imaging, LLC (hereinafter "NDSsi") warrants this product to be in compliance with the specifications provided by NDSsi and to be free from defects in material and workmanship as defined in such specifications. Subject to the conditions set forth below, NDSsi agrees to repair or replace any defective part of the enclosed unit for the length of period indicated on the chart below.

NDSSI PRODUCT

Dome® E Series

Radiology Displays

Dome® S Series

Radiology Displays

STANDARD WARRANTY COVERAGE

• Standard 5-year “repair and return” warranty

• 10-year backlight warranty on grayscale displays

*

• 5-year backlight warranty on color displays

† ‡

• Standard 5-year “repair and return” warranty

• 10-year backlight warranty on grayscale displays

*

• 5-year backlight warranty on color displays

Dome® GX4MP Display • Standard 5-year “repair and return” warranty

• 5-year backlight warranty

**

Standard 3-year “repair and return” warranty Dome® GX2MP Display

VitalScreen® Display

* If, within 10 years of initial purchase, the maximum output of the Dome Ex/Dome Sx grayscale display is determined by NDSsi to be less than 300 cd/m², NDSsi will repair or replace the display at its sole discretion. The display must have been run at or below the CXtra default brightness setting as configured from the factory.

† If, within 5 years of initial purchase, the maximum output of the Dome Ex/Dome Sx color display

(excluding the Dome E4c) is determined by NDSsi to be less than 250 cd/m², NDSsi will repair or replace the display at its sole discretion. The display must have been run in native color mode at or below the

CXtra default brightness setting as configured from the factory.

‡ If, within 5 years of initial purchase, the maximum output of the Dome E4c color display is determined by NDSsi to be less than175 cd/m², NDSsi will repair or replace the display at its sole discretion.

The display must have been run in native color mode at or below the CXtra default brightness setting as configured from the factory.

** If, within 5 years of initial purchase, the maximum output of the Dome GX4MP color display is determined by NDSsi to be less than 175 cd/m², NDSsi will repair or replace the display at its sole discretion. The display must have been run in native color mode at or below the CXtra default brightness setting as configured from the factory.

44 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

EXCLUSIONS – WHAT IS NOT COVERED

1 Any product with a defaced, modified, or removed serial number.

2 Damage, deterioration or a malfunction resulting from accident, misuse, neglect, fire, water, lightning, or other acts of nature, unauthorized product modification, or failure to follow instructions supplied with the product.

3 Cosmetic damage including, but not limited to scratches, cracks, dents, markings, glue and adhesive residue.

4 Any damage of the product due to shipment.

5 Any damage caused by factors external to the product, such as electric power fluctuation or failure.

6 Normal wear and tear, including backlights dimming over time, or image retention resulting from displaying fixed images for long periods of time.

7 Removal, installation, and set-up service charges.

8 Failures not reported within the warranty term.

9 Any NDSsi products purchased through a distributor, reseller, or medical device manufacturer other than NDSsi (each, an “Intermediary”), where such Intermediary provides direct warranty service to its end-user customers in connection with such product.

*

DISCLAIMER

This limited product warranty sets forth your sole and exclusive remedy and NDSsi’s sole and exclusive liability under the Standard Limited Warranty described herein.

THERE ARE NO WARRANTIES, EXPRESS OR IMPLIED, WHICH EXTEND BEYOND THE

DESCRIPTION CONTAINED HEREIN INCLUDING, BUT NOT LIMITED TO, THE IMPLIED

WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, OR

NONINFRINGEMENT.

* Note that NDSsi sells its products through distributors, resellers, and other medical device manufacturers which prefer to provide their end-user customers with direct warranty support. Contact

NDSsi Customer Care to determine if the product that you purchased is covered by this Standard Limited

Warranty or whether you need to contact the Intermediary through which you purchased the product for warranty service.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 45

NDSSI’S LIABILITY IS LIMITED TO THE COST OF REPAIR OR REPLACEMENT OF THE

PRODUCT. NDSSI SHALL NOT BE LIABLE FOR THE FOLLOWING:

1 DAMAGE TO OTHER PROPERTY CAUSED BY ANY DEFECTS IN THE PRODUCT,

DAMAGES BASED UPON INCONVENIENCE, LOSS OF USE OF THE PRODUCT, LOSS OF

TIME, LOSS OF PROFITS, LOSS OF BUSINESS OPPORTUNITY, LOSS OF GOODWILL,

INTERFERENCE WITH BUSINESS RELATIONSHIPS, OR OTHER COMMERCIAL LOSS,

EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.

2 ANY OTHER DAMAGES, WHETHER INCIDENTAL, INDIRECT, CONSEQUENTIAL OR

OTHERWISE.

3 ANY CLAIM AGAINST THE CUSTOMER BY A THIRD PARTY.

End-users are cautioned that system configuration, software, the application, customer data and operator control of the system, among other factors, affect the product performance. While NDSsi products are considered to be compatible with many systems, specific functional implementation by the customers of the product may vary.

Therefore, suitability of a product for a specific purpose or application must be determined by consumer and is not warranted by NDSsi.

This warranty gives you specific legal rights. You may have other rights, which may vary from locality to locality. Some localities do not allow limitations on implied warranties and/or do not allow the exclusion of incidental or consequential damages, so the above limitations might not apply to you.

WARRANTY SERVICE - REPAIR & RETURN – (U.S., LATIN AMERICA, AND

CANADA ONLY)

• Our standard warranty service is “Repair and Return”. Repair and Return requires the defective unit to be returned to our service location for repair.

• Our service location will repair your unit within ten (10) business days. The service time-period does not include in-transit shipping time to or from our service location.

• Cost of shipment of the defective unit to NDSsi is the responsibility of the customer.

Cost of shipment of the repaired unit to Customer is the responsibility of NDSsi

(duties and taxes to Canada and Latin America are not included).

• NDSsi reserves the right in its sole discretion to provide Customer with a comparable refurbished replacement unit in lieu of repair of customer’s defective unit.

• If NDSsi is unable within a reasonable time to repair or replace the defective unit, it shall refund the purchase price for the product paid by the customer (exclusive of taxes, installation and shipping related fees).

• NDSsi provides a complimentary 90-day Out of Box Failure (OBF) replacement warranty from the date of shipment from NDSsi.

46 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

CONTACTING NDSSI CUSTOMER CARE

PRODUCT LINE

TOLL FREE

LOCAL

E-MAIL

HOURS

Dome®

(866) 961-9340

(503) 620-3787 [email protected]

6 A .

M . – 5 P .

M . Pacific Time, Monday – Friday

1 Contact NDSsi Medical Support during the hours listed above. Have your NDSsi model number and serial number available. You may be required to provide proof of purchase as a condition for receiving warranty service.

2 Our representative may perform troubleshooting and diagnosis of the problem by telephone or e-mail. If our representative is unable to fix the problem by telephone or e-mail, we will issue a Return Material Authorization (RMA) for the nonfunctioning unit and provide return instructions.

3 Upon receipt of the returned product, a service technician will evaluate the unit and confirm the failure description provided by the customer.

(a) If the found failure type is covered under warranty, the repair, parts, labor and shipment back to the customer will be at no charge.

(b) If the found failure type is not covered under warranty, NDSsi’s then current evaluation fee will apply in addition to the cost of parts, labor and the return shipment.

(c) If the unit is within the warranty period but the customer does not provide a problem description and the service technician cannot determine the failure

(no problem found – “NPF”) NDSsi’s then current evaluation fee will apply in addition to return shipment.

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 47

PAYMENT FOR NONWARRANTY EVALUATIONS AND REPAIRS AND OTHER EVALUATION

CHARGES: Nonwarranty repairs requested by customer will be performed at NDSsi’s then current rates for any nonwarranty repairs and will include applicable evaluation fees and the cost of return shipping (if customer requests that NDSsi ship the repaired product). NDSsi further reserves the right in its sole discretion to request pre-payment for any nonwarranty repairs, NPF evaluations, and return shipping charges prior to return of the product to customer. Any invoiced amount not paid when due shall be subject to a service charge equal to the lesser of one and one-half percent (1.5%) per month or the maximum rate permitted by law. If NDSsi undertakes collection or enforcement efforts, customer shall be liable for all costs thereof, including attorney’s fees. NDSsi reserves the right to suspend warranty coverage to any customer who is in default of payment until such payment has been made.

48 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

ITEM

Dome S3 display

Dome S3c display

Dome S6c display

Dome S10 display

Power adapter, 6-pin

Power adapter, 8-pin

ITEM NUMBER

933-0993

933-0992

933-1014

933-0989

902-0176

902-0174

902-0173

DESCRIPTION

3-megapixel grayscale display.

3-megapixel color display.

6-megapixel color display.

10-megapixel grayscale display.

Power block (24V, 150W) with 6-pin DIN connector for use with the Dome S6c display.

Power block (24V, 150W) with 6-pin DIN connector for use with the Dome S3c,

Dome S6c, or Dome S10 display.

Power block (12V, 130W) with 8-pin DIN connector for use with the Dome S3 display.

Medical-grade power cord for use in the

United States.

Medical-grade power cord for use in the

European Union.

Medical-grade power cord for use in China.

Power cord, US

Power cord, EU

Power cord, CN

DVI cable

903-1160

903-0500

903-1159

903-0499

903-1158

903-0497

903-1161

903-0538

903-1138

Video cable for use with any Dome Sx display.

DisplayPort cable

Mini USB cable

Quick reference

903-1139

020-0915

Video cable for use with the Dome S3 or

Dome S3c display.

Connectivity cable for use with the Dome S10 display.

Installation booklet for Dome display systems supported by NVIDIA graphics controllers.

The use of ACCESSORIES , transducers, and cables other than those specified, with the exception of transducers and cables sold by the MANUFACTURER of the ME EQUIPMENT or ME SYSTEM as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the

ME EQUIPMENT or ME SYSTEM .

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 49

Index

A about display

5

auto-conformance

7

C certifications

43

cleaning display 18

components, packaged

4 ,

49

connecting power cord

10

video cable

10

connector plate 8

, 29

customer support

47

D

DIN connector 29

display about

5

classification

12

cleaning

18

components

7

connector plate 8

DIN power connector 29

documents related

6

install

10

intended use

5 ,

14

LED status lights

30

orientation

28

package contents

4

packaged components

4

packaging symbols 13

preventing fire/injury

18

proper disposal

19

proper handling

17

recycling instructions 12

regulatory compliance

33

resolution 28

safety precautions 15

shipping/storing

19

system requirements

5

technical information

12

troubleshooting

32

warranty and service terms

44

disposal information

19

E environmental conditions

21 ,

23 ,

27

EU declaration of conformity

36 ,

37

F

FCC compliance 38

I information, technical

12

installing display

10

intended use

5 ,

14

L

LED status lights

30

M

MQSA function 7

O orientation

28

50 | Dome  Sx Display User’s Guide, 020-0931-00 Rev. F

P package contents, display

4

power adapter

21 ,

23 ,

27 ,

29

DIN connector 29

input

29

power cord, connecting 10

power supply

21 ,

23 ,

27

precautions

15

preventive measures

18

product handling

17

R recycling instructions

12

regulatory compliance 33

Canada, EU, USA

33

China 43

electromagnetic emissions

39

electromagnetic immunity

40

, 41

Germany 43

Japan

43

Korea 43

separation distances

42

related documents

6

reliability

21 ,

23 ,

27

resolutions 28

S

safety precautions 15

shipping/storing display 19

specification

Dome S10 display

26

Dome S3 display 20

Dome S3c display

22

Dome S6c display

24

symbols, explained

13

system requirements

5

T technical information

12

tips, safety 16

troubleshooting

32

U use, intended

5 ,

14

V video modes

28

video cable

connecting 10

connecting to Dome S10 11

W

warranty and service terms 44

Dome  Sx Display User’s Guide, 020-0931-00 Rev. F | 51

NDS Surgical Imaging, LLC

5750 Hellyer Avenue

San Jose CA 95138-1000 USA www.ndssi.com

NDS Surgical Imaging BV

Nijverheidscentrum 28

2761 JP Zevenhuizen (ZH)

The Netherlands

+31 180 63 43 56 Tel

+31 180 63 21 91 Fax

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals