GW Instek GDM-541/533 Digital Multimeter User Manual


Add to my manuals
56 Pages

advertisement

GW Instek GDM-541/533 Digital Multimeter User Manual | Manualzz

Handheld Digital Multimeter

GDM-541/533

USER MANUAL

ISO-9001 CERTIFIED MANUFACTURER

This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company.

The information in this manual was correct at the time of printing.

However, Good Will continues to improve products and reserves the rights to change specification, equipment, and maintenance procedures at any time without notice.

Good Will Instrument Co., Ltd.

No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.

Table of Contents

Table of Contents

S AFETY INSTRUCTIONS .................................................. 4

G ETTING STARTED .......................................................... 7

GDM-541/533 Overview .......................................... 8

Function Dial and Function Buttons ...................... 12

O PERATION .................................................................. 14

Operating Instructions .......................................... 15

AC Voltage Measurement ...................................... 15

DC Voltage Measurement ...................................... 17

AC/DC Millivolt Voltage Measurement .................. 19

LoZ (low impedance) ACV Measurement (GDM-533)21

Resistance Measurement ....................................... 23

Continuity Test ...................................................... 25

Diode Test ............................................................. 27

Capacitance Measurement ..................................... 29

Transistor Magnification (hFE) Measurement

(GDM-541) ............................................................ 31

Frequency/Duty Ratio Measurement ...................... 32

Temperature Measurement (GDM-533) ................ 33

AC/DC Current Measurement ................................ 35

Non-Contact Voltage (NCV) Detection .................. 37

USB Data Transmission ......................................... 39

Others ................................................................... 40

S PECIFICATIONS ........................................................... 41

General Specifications ........................................... 41

Electrical specifications ......................................... 42

M AINTENANCE ............................................................. 55

General Meintenance ............................................ 55

Battery/Fuse Replacement ..................................... 56

3

GDM-541/533 User Manual

4

S

AFETY INSTRUCTIONS

This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition.

Electrical Symbols

These safety symbols may appear in this manual or on the instrument.

Warning or Caution

Caution, possibility of electric shock

Both direct and alternating current

Equipment protected throughout by DOUBLE

INSULATION or REINFORCED INSULATION

Earth (ground) Terminal

Complies with European Union standards

CAT III It is applicable to testing and measuring circuits connected to the distribution part of the building's low-voltage MAINS installation.

SAFETY INSTRUCTIONS

CAT IV It is applicable to testing and measuring circuits connected atthe source of the building’s low voltage MAINS installation.

Safety Instructions

The meter is designed and manufactured according to IEC61010-1 safety standard, and conforms to CAT III 1000V, CAT IV 600V, and pollution degree 2. If the meter is used in a manner not specified by the manufacturer, the protection provided by the meter may be impaired.

Before use, please check if there is any item which is damaged or behaving abnormally. If any abnormal item (such as bare test lead, damaged meter casing, broken LCD, etc.) is found, please do not use the meter.

Do not use the meter if the rear cover or the battery cover is not completely covered up, or it may pose a shock hazard!

Damaged test leads must be replaced with ones of the same model or same electrical specifications.

During measurement, do not touch any exposed wires, connectors, unused inputs or circuits being measured.

Use caution when working with voltages above AC 30Vrms,

42Vpeak or DC 60V. Keep fingers behind the finger guards of the test leads to prevent electric shock.

If the range of the measured value cannot be determined, the meter must be operated at the maximum range.

Do not apply more than the rated voltage or current marked on the meter between the terminals or between any terminal and earth ground.

5

GDM-541/533 User Manual

Place the function dial in the correct position before measurement.

Before measuring resistance, diode, continuity, or capacitance, switch off the power supply of the circuit, and fully discharge all capacitors.

Before measuring current, make sure the fuses are intact.

Do not use or store the meter in high temperature, high humidity, flammable, explosive, or strong magnetic field environments.

Do not change the internal circuit of the meter to avoid damage to the meter or user!

When “ ” is displayed, please replace the batteries in time to ensure measurement accuracy.

Turn off the meter in time after measurement. If the meter is not in use for a long time, please remove the batteries.

6

GETTING STARTED

G

ETTING STARTED

This chapter describes the GDM-541/533 in a nutshell, including its main features and front / rear panel introduction.

GDM-541 GDM-533

GDM-541/533 Overview .......................................... 8

Overview ............................................................................................. 8

Accessories .......................................................................................... 8

External Structure .............................................................................. 9

LCD Display ..................................................................................... 10

Function Dial and Function Buttons ...................... 12

Function Dial .................................................................................... 12

Function Buttons ............................................................................. 13

7

GDM-541/533 User Manual

GDM-541/533 Overview

Overview

The GDM-541/533 is a handheld true RMS digital multimeter with high reliability and security (GDM-533: 6000 counts; GDM-541:

22000 counts). With large screen, high resolution analog pointer display, full scale overload protection, and unique appearance design, it becomes a new practical electrical measuring meter. The meter can measure AC/DC voltage/current, resistance, diode, transistor hFE (GDM-541), continuity, capacitance, frequency, duty ratio, temperature (GDM-533), etc. Featuring data transmission, data hold, relative value measurement, peak measurement, internal temperature alarm, low battery indication, backlight, auto power off, and NCV functions, the meter is an ideal measuring tool for many application fields.

Accessories

Before using the GDM unit, check the package contents to make sure all the accessories are included.

Standard Accessories Description

User manual*1 pc

Test leads*1 pair

Adapter socket*1 pc

K-type thermocouple (GDM-533)*1 pc

USB cable*1 pc

1.5V AAA batteries*4

8

External Structure

GETTING STARTED

7

8

9

10

11

5

6

3

4

Item Index

1

2

Description

NCV detector

Indicator light

LCD display

Function buttons

Function dial

Input terminals

USB access port

Test lead slots

Nut for external holder

Battery compartment fixing screw

Tilt stand

9

LCD Display

GDM-541/533 User Manual

AC/DC k Ω , M Ω , Ω mV, V

μ A, mA, A nF, μ F, mF

Hz, %

GDM-533 GDM-541

Symbol Description

Measured voltage is >30V (AC or DC)

Data hold

Negative reading

AC/DC measurement

Low battery indication

Auto range

Diode test

Continuity test

Relative value measurement

Resistance units: ohm, kilohm, megaohm

Voltage units: millivolt, volt

Current units: microampere, milliampere, ampere

Capacitance units: nanofarad, microfarad, millifarad

Frequency, duty ratio

Low pass filter

10

NCV

P-MAX/P-

MIN

MAX/MIN

°C/°F

LoZ hFE

T RMS

GETTING STARTED

Data transmission

Transistor magnification (GDM-541)

Non-contact voltage detection

Peak measurement

Maximum/Minimum measurement

Celsius/Fahrenheit (GDM-533)

Low impedance measurement (GDM-533)

Transistor magnification measurement (GDM-541)

Auto power off

True RMS

11

GDM-541/533 User Manual

Function Dial and Function Buttons

Function Dial

Dial Position Description

Power off

AC voltage measurement/Low pass filter measurement/Frequency and duty ratio measurement (GDM-541)

DC voltage measurement/AC+DC measurement

(GDM-541)

AC voltage measurement/Frequency

AC/DC voltage measurement/Frequency and duty ratio measurement (GDM-533)

AC/DC millivolt voltage measurement/Frequency and duty ratio measurement

Diode test/Continuity test/Resistance measurement/Capacitance measurement

Transistor magnification measurement (GDM-541)

Frequency and duty ratio measurement

AC/DC microampere current measurement/Frequency and duty ratio measurement

AC/DC milliampere current measurement/Frequency and duty ratio measurement

AC/DC ampere current measurement/Frequencyand duty ratio measurement

Non-contact voltage detection

12

GETTING STARTED

Function Buttons

Button Description

Short press to switch between functions in each compound function position.

Short press to enter the manual range mode and change the range.

Long press to exit the manual range mode.

Short press to switch between frequency and duty ratio measurement.

Long press to turn on/off data communication (note: only available when USB communication module is inserted into the casing).

Short press to enter/exit the relative value measurement mode.

Short press to cycle through the measured maximum and minimum.

Long press to cycle through the peak maximum and peak minimum

Short press to hold the measurement on the display and “ ” will be displayed.

Short press again to cancel data hold.

Long press to turn on/off the backlight.

13

14

GDM-541/533 User Manual

O

PERATION

Operating Instructions .......................................... 15

AC Voltage Measurement ...................................... 15

DC Voltage Measurement ...................................... 17

AC+DC Voltage Measurement (GDM-541) .............................. 18

AC/DC Millivolt Voltage Measurement .................. 19

LoZ (low impedance) ACV Measurement (GDM-533)

.............................................................................. 21

Resistance Measurement ....................................... 23

Continuity Test ...................................................... 25

Diode Test ............................................................. 27

Capacitance Measurement ..................................... 29

Transistor Magnification (hFE) Measurement (GDM-

541) ....................................................................... 31

Frequency/Duty Ratio Measurement ...................... 32

Temperature Measurement (GDM-533) ................ 33

AC/DC Current Measurement ................................ 35

Non-Contact Voltage (NCV) Detection .................. 37

USB Data Transmission ......................................... 39

Others ................................................................... 40

OPERATION

Operating Instructions

Please check the internal batteries first. If “ ” is d isplayed, replace the batteries in time. Please also pay attention to the wa rning sign “ ” beside the input terminals, which indicates that the measured voltage or current must not exceed the values marked on the meter.

AC Voltage Measurement

Wire Connection

Diagram

Steps of

Measurement

1.

Insert the red test lead into the or terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

, , or

15

GDM-541/533 User Manual

Caution

3.

Short press the SELECT button to switch to

AC voltage measurement or LPF ACV measurement ( GDM-541 , manual maximum range by default) if required.

4.

Connect the test leads with the measured load or power supply in parallel.

5.

Read the voltage value on the display (if the voltage is >1000V, the red indicator light will be on and the buzzer will sound an alarm).

6.

Short press the button to display the frequency/duty ratio of the measured voltage.

 Do not input a voltage over 1000V or it may damage the meter.

 Be cautious to avoid electric shock when measuring high voltages.

 After completing the measurement, disconnect the test leads from the circuit under test.

16

DC Voltage Measurement

Wire Connection

Diagram

OPERATION

Steps of

Measurement

1.

Insert the red test lead into the or terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

, , or

3.

Short press the SELECT button to switch to

DC voltage SELECT measurement if required.

4.

Connect the test leads with the measured load or power supply in parallel.

5.

Read the voltage value on the display (if the voltage is >1000V, the red indicator light will be on and the buzzer will sound an alarm).

17

GDM-541/533 User Manual

AC+DC Voltage Measurement (GDM-541)

Steps of

Measurement

1.

Insert the red test lead into the terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

3.

Short press the SELECT button to switch to

AC+DC voltage measurement.

4.

Connect the test leads with the measured load or power supply in parallel.

5.

Read the voltage values on the display. The

AC and DC voltages are displayed alternately.

18

OPERATION

AC/DC Millivolt Voltage Measurement

Wire Connection

Diagram

Steps of

Measurement

1.

Insert the red test lead into the or terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

3.

Short press the SELECT button to switch to

AC/DC millivolt voltage measurement if required.

4.

Connect the test leads with the measured load or power supply in parallel.

5.

Read the voltage value on the display.

6.

When measuring AC millivolt voltage, short press the button to display the frequency/duty ratio of the measured voltage.

19

20

Caution

GDM-541/533 User Manual

 Do not input a voltage over 1000V or it may damage the meter.

 Be cautious to avoid electric shock when measuring high voltages.

 After completing the measurement, disconnect the test leads from the circuit under test.

 Before each use, verify meter operation by measuring a known voltage.

 The input impedance of the AC mV range is about 10MΩ. This load effect may cause measurement errors in high-impedance circuits. In most cases, if the impedance of the circuit is below 10kΩ, the error ca n be ignored

(≤0.1%).

 The input impedance of the DC mV range is infinite (about 1GΩ), and it does not attenuate when measuring weak signals, so the measurement accuracy is high. When the test leads are open, there may be a value on the screen, but this is normal and will not affect the measuring result.

 Frequency measurement at 60mV range (AC voltage) is for reference only.

OPERATION

LoZ (low impedance) ACV Measurement

(GDM-533)

Wire Connection

Diagram

Steps of

Measurement

1.

Insert the red test lead into the or terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

3.

Connect the test leads with the measured load or power supply in parallel.

4.

Read the voltage value on the display.

5.

Short press the button to display the frequency/duty ratio of the measured voltage.

21

Caution

GDM-541/533 User Manual

 Do not input a voltage over 1000V or it may damage the meter.

 Be cautious to avoid electric shock when measuring high voltages.

 After completing the measurement, disconnect the test leads from the circuit under test.

 Before each use, verify meter operation by measuring a known voltage.

 After using the LoZ function, wait 3 minutes before next operation.

 LoZ ACV measurement eliminates ghost voltage for more accurate measurement.

22

Resistance Measurement

Wire Connection

Diagram

OPERATION

Steps of

Measurement

1.

Insert the red test lead into the or terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

or

3.

Touch the probes to the test points in the circuit.

4.

Read the resistance value on the display.

23

24

Caution

GDM-541/533 User Manual

 Use caution when working with voltages above AC 30Vrms, 42Vpeak or DC 60V. Such voltages pose a shock hazard.

 If the measured resistor is open or the resistance exceeds the maximum range, the

LCD will display “OL”.

 Before measuring resistance, switch off the power supply of the circuit, and fully discharge all capacitors.

 When measuring low resistance, the test leads will produce 0.1Ω~0.3Ω measurement error.

To obtain accurate measurement, short-circuit the test leads and use the relative value measurement (REL) mode.

 If the resistance is not less than 0.5Ω when the test leads are short-circuited, please check if the test leads are loose or abnormal.

 When measuring high resistance, it is normal to take a few seconds to stabilize the reading.

Continuity Test

Wire Connection

Diagram

OPERATION

Steps of

Measurement

1.

Insert the red test lead into the or terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

or

3.

Short press the SELECT button to switch to continuity test.

4.

Touch the probes to the test points in the circuit.

5.

Measured resistance <50Ω: The circuit is in good conduction status; the buzzer beeps continuously and the green indicator light is on.

25

26

Caution

GDM-541/533 User Manual

 Use caution when working with voltages above AC 30Vrms, 42Vpeak or DC 60V. Such voltages pose a shock hazard.

 Before testing continuity, switch off the power supply of the circuit, and fully discharge all capacitors.

Diode Test

Wire Connection

Diagram

OPERATION

Steps of

Measurement

1.

Insert the red test lead into the or terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

or

3.

Short press the SELECT button to switch to diode test if required.

4.

Connect the red probe with the diode anode, and black probe with the diode cathode.

5.

Read the forward bias value on the display.

27

GDM-541/533 User Manual

Caution

6.

Measured value <0.12V: The diode may be damaged; the red indicator light is on.

Measured value within 0.12V~2V: The diode is normal; the green indicator light is on (for reference only).

7.

If the diode is open or its polarity is reversed, the LCD will display “OL”. For silicon PN junction, the normal value is generally about

500~800 mV.

 Use caution when working with voltages above AC 30Vrms, 42Vpeak or DC 60V. Such voltages pose a shock hazard.

 Before testing the diode, switch off the power supply of the circuit, and fully discharge all capacitors.

28

Capacitance Measurement

Wire Connection

Diagram

OPERATION

Steps of

Measurement

1.

Insert the red test lead into the or terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

or

3.

Short press the SELECT button to switch to capacitance measurement.

4.

Touch the probes to the capacitor pins.

5.

Read the capacitance value on the display after it gets steady.

29

30

Caution

GDM-541/533 User Manual

 Use caution when working with voltages above AC 30Vrms, 42Vpeak or DC 60V. Such voltages pose a shock hazard.

 Before measuring, fully discharge all capacitors (especially high-voltage capacitors) to avoid damage to the meter and user.

 If the measured capacitor is short-circuited or the capacitance exceeds the maximum range, the LCD will display “OL”.

 When measuring high capacitance, it is normal to take a few seconds to stabilize the reading.

 For small capacitance measurement, the REL mode should be used to avoid the influence coming from distributed capacitance so as to obtain the correct reading.

OPERATION

Transistor Magnification (hFE) Measurement

(GDM-541)

Wire Connection

Diagram

Steps of

Measurement

1.

Turn the function dial to the hFE position.

2.

Insert the adapter socket into the input terminals.

3.

Insert the three pins of the transistor under test into the corresponding holes of the adapter socket.

4.

Read the magnification of the measured transistor.

31

GDM-541/533 User Manual

Frequency/Duty Ratio Measurement

Wire Connection

Diagram

Steps of

Measurement

Caution

1.

Insert the red test lead into the or terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

3.

Short press the button to switch to frequency/duty ratio measurement if required.

4.

Read the frequency/duty ratio value on the display.

 Use caution when working with voltages above AC 30Vrms, 42Vpeak or DC 60V. Such voltages pose a shock hazard.

32

OPERATION

Temperature Measurement (GDM-533)

Wire Connection

Diagram

Steps of

Measurement

1.

Turn the function dial to the °C°F position.

2.

Insert the K-type thermocouple into the adapter socket, and then insert the adapter socket into the input terminals.

3.

Bring the temperature sensing end of the thermocouple close to the object surface under test.

4.

Read the Celsius temperature value on the display after it gets steady.

5.

Short press the SELECT button to switch between °C and °F.

33

34

Caution

GDM-541/533 User Manual

 Only K-type thermocouple is applicable.

 The LCD displays “OL” when the meter is turned on.

 The measured temperature should be less than 230°C/446°F (°F = °C × 1.8 + 32)

AC/DC Current Measurement

Wire Connection

Diagram

OPERATION

Steps of

Measurement

1.

Insert the red test lead into the mA/μA or A terminal, and black test lead into the COM terminal.

2.

Turn the function dial to the position.

, , or

3.

Short press the SELECT button to switch to

AC/DC current measurement if required.

4.

Connect the test leads with the measured load or power supply in series.

5.

Read the current value on the display (if the current is >10A, the red indicator light will be on and the buzzer will sound an alarm).

35

GDM-541/533 User Manual

Caution

6.

When measuring AC current, short press the

button to display the frequency/duty ratio of the measured current.

 To prevent possible electric shock, fire, or personal injury, switch off the power supply of the circuit, and then connect the meter with the circuit in series before measuring the current.

 If the range of the measured current is unknown, select the maximum range and then accordingly reduce.

 There are fuses inside mA/μA and A input terminals. Do not connect the test leads with any circuit in parallel.

 When the measured current is >5A, each measurement time should be ≤10s and the rest inte rval should be ≥15 minutes.

 When the temperature in th e meter is ≥75°C after measurement of large current, the yellow indicator light will be on, the buzzer will beep, and the LCD will display “CUT”. When the temperature drops to <40°C, the yellow indicator light will be off, and the measurement can be made.

36

OPERATION

Non-Contact Voltage (NCV) Detection

Wire Connection

Diagram

Steps of

Measurement

1.

Turn the function dial to the NCV position.

2.

Bring the NCV detector (top left corner of the meter) close to the wire (AC) under test.

3.

If the voltage of the wire is ≥50Vrms

(frequency: 50Hz/60Hz), the red indicator light will be on and the buzzer will beep. If no voltage is detected, the LCD will display

“EF”. As the intensity of the detected voltage increases, more segments “ ” will be displayed, and the frequency for buzzer beeping and red indicator light flashing will be higher.

37

38

Caution

GDM-541/533 User Manual

 The detected voltage level varies with the distance between the NCV detector and the wire under test.

 The detected voltage level is for reference only, not for specific measurement. The frequency of the detected voltage should be

50Hz/60Hz.

 Hold the meter casing for non-contact voltage detection.

USB Data Transmission

Wire Connection

Diagram

OPERATION

Steps of

Measurement

Figure A Figure B

1.

Pull out the USB sealing cover at the back of the meter (Figure A).

2.

Insert the USB communication module into the USB access port of the meter and the LCD will display “ ” (Figure B).

3.

If USB data transmission is not needed during measurement, long press the button or pull out the USB module to disable data transmission, and “ ” will disappear.

4.

To recover this function, long press the button or insert the USB module.

5.

The USB communication software can be downloaded from the official website of

GW INSTEK (www.gwinstek.com).

39

GDM-541/533 User Manual

Others

Steps of

Measurement

1.

Auto power off: During measurement, if there is no operation for 15 minutes, the meter will automatically shut down to save power; before the automatic shutdown, the buzzer will beep for warning. Users can wake the meter up by pressing the SELECT button. To disable the auto-off function, press and hold the SELECT button in the off state, and then turn on the meter. To recover the function, restart the meter.

2.

Buzzer alarm during measurement: When the input voltage >1000V or current >10A, the buzzer will sound an alarm.

3.

Low battery indication: When the battery voltage is ≤4.6V ±0.2V, “ ” will be displayed.

40

SPECIFICATIONS

S

PECIFICATIONS

General Specifications

1

2

3

4

Max voltage between the input terminal and COM terminal: Please refer to the description of input protection voltage for each range. mA/ μ A input terminal protection: (CE) 600mA 1000V fast-acting fuse,

Φ 6x32mm

A input terminal protection: (CE) 11A 1000V fast-acting fuse, Φ 10x38mm

Max display: 22000 ( GDM-541 ), 6000 ( GDM-533 )

5 Analog bar: 46 segments ( GDM-541 ) (conversion rate: 30 times/s), 31 segments ( GDM-533 )

6

7

8

9

Refresh rate: 2~3 times/s

Range: Auto/Manual

Polarity display: Auto

Over-range Indication: OL

10 Low battery indication: “ ” is displayed.

11 Operating temperature: 0°C~40°C (32°F~104°F)

12 Storage temperature: -10°C~50°C (14°F ~122°F)

13 Relative humidity: ≤ 75% at 0˚C~30˚C; ≤ 50% at 30˚C~40˚C

14 Operating altitude: ≤ 2000m

15 Electromagnetic compatibility: Conforms to EN61326-1:2006 and

EN61326-2-2:2006 standards

16 Battery: 4×1.5V AAA

17 Dimensions: 186mm×89mm×49mm

18 Weight: 420g (GDM-541), 400g (GDM-533) (with battery)

19 Safety standard: IEC 61010-1: CAT III 1000V/CAT IV 600V

20 Certification: CE

21 Pollution degree: 2

22 Information of usage: Indoor and outdoor

41

GDM-541/533 User Manual

Electrical specifications

Accuracy: ± (a% of reading + b digits), 1 year warranty

Ambient temperature: 23°C ± 5°C (73.4°F ± 9°F) Relative humidity: ≤75%

Caution

 To ensure measurement accuracy, the operating temperature should be within

18°C~28°C and the fluctuation range should be within ±1°C. When the temperature is

<18°C or >28°C, add temperature coefficient error: 0.1 x (specified accuracy)/ °C.

Technical Specifications Index

DC Voltage ....................................................................................... 43

AC Voltage ........................................................................................ 44

AC+DC Voltage .............................................................................. 46

Resistance .......................................................................................... 47

Continuity and Diode ...................................................................... 48

Transistor Magnification (GDM-541) .......................................... 48

Capacitance ....................................................................................... 49

Temperature (GDM-533) ............................................................... 50

DC Current ....................................................................................... 50

AC Current ........................................................................................ 51

Frequency/Duty Ratio .................................................................... 53

Indicator Light .................................................................................. 54

42

SPECIFICATIONS

DC Voltage

Resolution

0.01mV

0.1mV

1mV

10mV

0.1V

GDM-541

Range

220.00mV

2.2000V

22.000V

220.00V

1000.0V

Accuracy

± (0.1%+5)

± (0.05%+5)

± (0.1%+5)

Resolution

0.01mV

0.1mV

0.001V

0.01V

0.1V

1V

GDM-533

Range

60.00mV

600.0mV

6.000V

60.00V

600.0V

1000V

Accuracy

± (0.8%+5)

± (0.8%+3)

± (0.5%+3)

± (0.5%+3)

± (1.0%+3)

 Input impedance: About 1GΩ for mV range, about 10MΩ for other ranges

 Accuracy guarantee: 1%~100% of range; short circuit allows least significant digit ≤5

 Max input voltage: 1000V (if the voltage is >1000V, the red indicator light will be on and the buzzer will sound an alarm; if the voltage is >1010V , the LCD will display “OL”)

 Overload protection: 1000V

43

GDM-541/533 User Manual

AC Voltage

Range

GDM-541

Resolution Frequency Response

220.00mV 0.01mV

40Hz~1kHz

1kHz~10kHz

40Hz~1kHz

2.2000V 0.1mV 1kHz~10kHz

22.000V 1mV

40Hz~100Hz (LPF)

40Hz~1kHz

1kHz~10kHz

40Hz~100Hz (LPF)

40Hz~1kHz

220.00V

1000.0V

10mV

0.1V

1kHz~10kHz

40Hz~100Hz (LPF)

40Hz~1kHz

1kHz~10kHz

40Hz~100Hz (LPF)

Accuracy

± (1.0%+10)

± (1.5%+30)

± (0.8%+10)

± (1.2%+50)

± (1.2%+50)

± (0.8%+10)

± (1.2%+50)

± (1.8%+50)

± (0.8%+10)

± (2.0%+50)

± (2.0%+50)

± (1.2%+10)

± (3.0%+50)

Range

60.00mV

600.0mV

GDM-533

Resolution Frequency Response Accuracy

0.01mV

0.1mV

40Hz~1kHz

40Hz~1kHz

± (1.2%+5)

± (1.2%+5)

6.000V

60.00V

0.001V

0.01V

600.0V

1000V

0.1V

1V

LoZ ACV 600.0V 0.1V

LoZ ACV 1000V 1V

40Hz~1kHz

40Hz~1kHz

40Hz~1kHz

40Hz~1kHz

40Hz~1kHz

40Hz~1kHz

± (1.0%+3)

± (1.0%+3)

± (1.0%+3)

± (1.2%+5)

± (2.0%+5)

± (2.0%+5)

44

SPECIFICATIONS

 Input impedance: About 10MΩ

 Display: True RMS

 Frequency response: 40Hz~1kHz ( GDM-533 ), 40Hz~10kHz

( GDM-541 )

 Accuracy guarantee ( GDM-533 ): 2%~100% of 60mV range, 1%~100% of other ranges; short circuit allows least significant digit ≤3

 Accuracy guarantee ( GDM-541 ): 1%~100% of range at

40Hz~1kHz, 10%~100% of range at 1kHz~10kHz; short circuit allows least significant digit ≤10

 The AC crest factor can be ≤3.0 at 3000 counts, an d can only be

≤1.5 at 6000 counts. The additional error should be added according to the crest factor of a non-sinusoidal wave as follows ( GDM-533 ): a) Add 4% when crest factor is 1~2 b) Add 5% when crest factor is 2~2.5 c) Add 7% when crest factor is 2.5~3

 The AC crest factor can be ≤2.0 at 10000 counts, and can only be

≤1 at 22000 counts. The additional error should be added according to the crest factor of a non-sinusoidal wave as follows ( GDM-541 ): a) Add 4% when crest factor is 1~2 b) Add 5% when crest factor is 2~2.5 c) Add 7% when crest factor is 2.5~3

 Frequency measurement range: 40Hz~1kHz ( GDM-533 ),

40Hz~10kHz ( GDM-541 ); input amplitude: ≥10% of voltage range Duty ratio is for reference only.

 Frequency accuracy: ± (0.1%+4), resolution: 0.1Hz (GDM-533)

 Frequency accuracy: ± (0.01%+5), resolution: 0.01Hz (GDM-541)

 Max input voltage: 1000V (if the voltage is >1000V, the red indicator light will be on and the buzzer will sound an alarm; if the voltage is >1010V, the LCD will display “OL”)

 Overload protection: 1000V

45

GDM-541/533 User Manual

AC+DC Voltage

Range

2.2000V

22.000V

220.00V

1000.0V

GDM-541

Resolution Frequency Response

0.1mV 40Hz~500Hz

1mV

10mV

0.1V

40Hz~500Hz

40Hz~500Hz

40Hz~500Hz

Accuracy

± (1.8%+70)

± (1.8%+70)

± (1.8%+70)

± (4.0%+70)

 AC voltage display: True RMS

 Input impedance: About 10MΩ

 Accuracy guarantee: 10%~100% of range

 For AC voltage, short circuit allows least significant digit ≤200

 Overload protection: 1000V

46

SPECIFICATIONS

Resistance

Resolution

0.01Ω

0.1Ω

10Ω

100Ω

1kΩ

10kΩ

GDM-541

Range

220.00Ω

2.2000kΩ

22.000kΩ

220.00kΩ

2.2000MΩ

22.000MΩ

220.00MΩ

Accuracy

± (0.5+10)

± (0.8+10)

± (1.5%+10)

± (3.0%+50)

Resolution

0.1Ω

10Ω

100Ω

1kΩ

10kΩ

GDM-533

Range

600.0Ω

6.000kΩ

60.00kΩ

600.0kΩ

6.000MΩ

60.00MΩ

Accuracy

± (1.2%+2)

± (1.0%+2)

± (1.2%+2)

± (2.0%+5)

 Measurement result = displayed value – resistance of shorted test leads

 Open circuit voltage: About 1V

 Accuracy guarantee: 1%~100% of range

 Overload protection: 1000V

47

GDM-541/533 User Manual

Continuity and Diode

Range Resolution

0.1Ω

0.001V

Remarks

Broken circuit: Resistance ≥70Ω, no beep

Wellconnected circuit: Resistance <50Ω, audio/visual alarm

Open circuit voltage: About 3V

For normal diodes, the buzzer will beep once.

For short circuit, the buzzer will beep for a long time.

 Overload protection: 1000V

 When the forward voltage drop is within 0.12V~2V, the buzzer will beep once. When the forward voltage drop is <0.12V, the buzzer will beep for a long time.

Transistor Magnification (GDM-541)

Range Resolution

1000β 1β

GDM-541

Remarks

Ib0: About 1.8μA; Vce: About 2.5V

 The displayed value of the transistor magnification is for reference only.

48

SPECIFICATIONS

Capacitance

Range

22.000nF

220.00nF

2.2000μF

22.000μF

220.00μF

2.2000mF

22.000mF

220.00mF

Range

60.00nF

600.0nF

6.000μF

60.00μF

600.0μF

6.000mF

60.00mF

GDM-541

Resolution

1pF

10pF

100pF

1nF

10nF

100nF

1μF

10μF

10pF

1nF

10nF

1μF

10μF

GDM-533

Resolution

100pF

100nF

Accuracy

± (3.0%+5)

± (4.0%+5)

± (10%+5)

± (20%+5)

Accuracy

± (3%+5)

± (10%+5)

 Overload protection: 1000V

 Measurement result = displayed value – capacitance of opencircuit test leads

 For capacitance ≤1μF ( GDM-533 ) and ≤22nF ( GDM-541 ), it is recommended to use the REL mode to deduct the open circuit reading.

 Accuracy guarantee: 1%~100% of range

 For range s of 2.2μF and below, when the accuracy is ≤3%, 10 digits should be added ( GDM-541 ).

 For ranges of 60mF ( GDM-533 ) and 220mF ( GDM-541 ), the measurement time is about 20s.

49

GDM-541/533 User Manual

Range

220.00μA

2200.0μA

22.000mA

220.00mA

20.000A

Range

600.0μA

6000μA

60.00mA

600.0mA

6.000A

20.00A

Temperature (GDM-533)

-40~1000°C

-40~1832°F

Range

-40~0°C

0~300°C

300~1000°C

-40~32°F

32~572°F

572~1832°F

Resolution

0.1°C~1°C

0.2°F~2°F

Accuracy

± (1.0%+3°C)

± (1.0%+2°C)

± (1.0%+3°C)

± (1.0%+6°F)

± (1.0%+4°F)

± (1.0%+6°F)

 The measured temperature should be less than 230°C/446°F.

DC Current

GDM-541

Resolution

0.01μA

0.1μA

1μA

10μA

1mA

GDM-533

Resolution

0.1μA

1μA

10μA

0.1mA

1mA

10mA

Accuracy

± (0.5%+10)

± (1.2%+50)

Accuracy

± (1.0%+2)

± (1.0%+3)

± (1.2%+5)

50

SPECIFICATIONS

 Overload protection: mA/μA range: F1 Fuse 600mA 1000V Φ6x32mm

A range: F2 Fuse 11A 1000V Φ10x38mm

 Open circuit allows least significant digit ≤5 ( GDM-533 ) and ≤10

( GDM-541 ).

 Accuracy guarantee: 1%~100% of range

AC Current

Range

220μA

2200μA

22mA

220mA

20A

GDM-541

Resolution Frequency Response

0.01μA

40Hz~1kHz

1kHz~10kHz

0.1μA

1μA

10μA

40Hz~1kHz

1kHz~10kHz

40Hz~1kHz

1kHz~10kHz

40Hz~1kHz

1mA

1kHz~10kHz

40Hz~1kHz

1kHz~10kHz

Range

600.0μA

6000μA

60.00mA

600.0mA

6.000A

20.00A

Resolution

0.1

μA

1μA

10μA

0.1mA

1mA

10mA

GDM-533

Frequency Response

40Hz~1kHz

40Hz~1kHz

40Hz~1kHz

40Hz~1kHz

40Hz~1kHz

40Hz~1kHz

Accuracy

± (0.8%+10)

± (3%+50)

± (0.8%+10)

± (3%+50)

± (1.2%+10)

± (3%+50)

± (1.2%+10)

± (3%+50)

± (1.2%+10)

± (3%+50)

Accuracy

± (1.2%+5)

± (1.5%+5)

± (2.0%+5)

51

GDM-541/533 User Manual

 Display: True RMS

 Frequency response: 40Hz~1kHz ( GDM-533 ), 40Hz~10kHz

( GDM-541 )

 Accuracy guarantee ( GDM-533 ): 5%~100% of 600.0μA range,

1%~100% of other ranges; open circuit allows least significant digit ≤5

 Accuracy guarantee (GDM-541): 1%~100% of range at

40Hz~1kHz, 10%~100% of range at 1kHz~10kHz (the minimum measured current at μA ranges is 30μA); open circuit allows least significant digit ≤10

 The AC crest factor can be ≤3.0 at 3000 counts, and can only be

≤1.5 at 6000 counts. The additional error should be added according to the crest factor of a non-sinusoidal wave as follows

( GDM-533 ): a) Add 4% when crest factor is 1~2 b) Add 5% when crest factor is 2~2.5 c) Add 7% when crest factor is 2.5~3

 The AC crest factor can be ≤2.0 at 10000 counts, and can only be

≤1 at 22000 counts. The additional error should be added according to the crest factor of a non-sinusoidal wave as follows

( GDM-541 ): a) Add 4% when crest factor is 1~2 b) Add 5% when crest factor is 2~2.5 c) Add 7% when crest factor is 2.5~3

 Frequency measurement range: 40Hz~1kHz ( GDM-533 ),

40Hz~10kHz ( GDM-541 ); input amplitude: ≥50% of current range. Duty ratio is for reference only.

 Frequency accuracy: ± (0.1%+4), resolution: 0.1Hz (GDM-533)

 Frequency accuracy: ± (0.01%+5), resolution: 0.01Hz (GDM-541)

 Overload protection: Same as that for DC current

52

SPECIFICATIONS

Frequency/Duty Ratio

Accuracy

± (0.01%+5)

± (2%+5)

Accuracy

± (0.1%+4)

± (2%+5)

Range

10Hz~220MHz

0.1%~99.9%

GDM-541

Resolution

0.01Hz~0.01MHz

0.1%

Range

GDM-533

Resolution

10.00Hz~10.00MHz 0.01Hz~0.01MHz

0.1%~99.9% 0.1%

 Frequency input amplitude:

≤100kHz: 200mVrms ≤ input amplitude ≤ 20Vrms

>100kHz~1MHz: 600mVrms ≤ input amplitude ≤ 20Vrms

>1MHz ( GDM-533 ): 1Vrms ≤ input amplitude ≤ 20Vrms

>1MHz~40MHz ( GDM-541 ): 1Vrms ≤ input amplitude ≤ 20Vrms

>40MHz ( GDM-541 ): Not specified

 Duty ratio measurement is only applicable to square waves.

1Vpp ≤ input amplitude ≤ 20Vpp

Frequency ≤10kHz, duty ratio: 10.0% ~ 90.0%

 Overload protection: 1000V

53

GDM-541/533 User Manual

Indicator Light

Function

NCV

Continuity

Diode

AC/DC voltage

Status

Off

On, red

Off

On, red

On, green

Off

On, red

On, green

Off

On, red

Off

On, red

Current

Internal temperature during

AC/DC current measurement

Off

On, yellow

Description

<36V

50V~1000V (the red indicator light flashes from slow to fast)

OL

No continuity (≥70Ω)

Continuity (<50Ω)

>2V

Breakdown (<0.12V)

Conduction (0.12V~2V)

≤1000V

>1000V

≤10A

>10A

The temperature in the meter drops to <40°C after measurement of large current.

The temperature in the meter is

≥75°C after measurement of large current.

54

MAINTENANCE

M

AINTENANCE

 Before opening the rear cover or battery cover of the meter, switch off the power supply and remove the test leads.

Caution

General Maintenance

 Clean the meter casing with a damp cloth and mild detergent.

Do not use abrasives or solvents!

 If there is any malfunction, stop using the meter and send it for maintenance.

 The maintenance and service must be implemented by qualified professionals or designated departments.

55

GDM-541/533 User Manual

Battery/Fuse Replacement

 Battery: 4×1.5V AAA batteries

 Fuse: F1 Fuse 600mA 1000V Φ6x32mm (mA/μA input terminal)

F2 Fuse 11A 1000V Φ10x38mm (A input terminal)

 When “ ” is displayed, please replace the batteries in time to ensure measurement accuracy.

 Resistance measurement can be used to check the built-in

600mA and 11A fuses. Operation: Insert the red test lead into the or terminal. Insert the red probe into the mA/μA input terminal to measure the resistance. If the LCD displays

“OL”, the 600mA fuse is blown. Insert the red probe into the A input terminal to measure the resistance. If the LCD displays

“OL”, the 11A fuse is blown.

Steps of

Replacement

1.

Turn the function dial to the "OFF" position, and remove the test leads.

2.

Unscrew and remove the battery cover to replace the batteries and fuses.

56

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals