Quick Reference Guide For 01-Series

Note: Turn on and adjust the 01-Series gas monitor in a known fresh air area.

1. Press and hold POWER/MODE button for 3

seconds to turn instrument ON

2. Warm-up sequence a. All segments of LCD turn on.

b. LCD back light turns on.

c. Buzzer sounds and vibrator activates.

d. Alarm lights flash.

e. Instrument shows target gas and battery voltage.

f. Buzzer will sound after battery voltage is displayed.

g. Instrument is now in normal operating mode.

Note: If instrument gives a low battery or dead battery alarm, replace batteries or recharge (GP-01 only) before use.

3. Performing Fresh Air Adjustment a. Press and hold the AIR button. The LCD displays “hold”. b. Release the AIR button when LCD displays “Adj”.

c. The instrument will set reading to 0 ppm (H2S/CO), 0% LEL or 20.9% Vol. oxygen depending on the instrument.

4. Displaying PEAK, STEL and TWA (CO and H2S

versions) a. Press and release the POWER/MODE button to display the

PEAK gas reading.

b. Press and release the POWER/MODE button again to display STEL reading.

c. Press and release the POWER/MODE button again to display the TWA reading.

d. Press and release POWER/MODE button again to return to normal measuring mode.

Note: If no button is pressed for 20 seconds, the instrument will automatically return to measuring mode.

Quick Reference Guide For 01-Series

5. Displaying MIN and MAX readings (OX-01) a. Press and release POWER/MODE button to display the minimum oxygen reading.

b. Press and release the POWER/MODE button again to display the maximum oxygen reading.

c. Press and release POWER/MODE button again to return to normal mode.

6. Changing the Alarm Points a. Make sure that 01-Series is OFF.

b. Press and hold the AIR button, then press and hold the

POWER/MODE button. As soon as segments appear on the display (approximately 1 second), release the AIR button.

When the unit “beeps,” release the POWER/MODE button to put the 01-Series into alarm point adjustment mode. The

LCD should display the low alarm setting.

c. Use the the AIR button to increase the alarm point. Once selected, press the POWER/MODE button to select the next alarm level.

d. Press the POWER/MODE button repeatedly until the ROM number of your unit appears on the LCD. The 01-Series will then go into its startup sequence followed by the measuring mode.

Note: Consult manual for additional information.

RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587

Phone (510) 441-5656 • (800) 754-5165 • www.rkiinstruments.com

PN 72-0141RK Rev 0