Fluke 3000 FC GM General Maintenance Kit Data Sheet

Fluke 3000 FC GM General Maintenance Kit Data Sheet
Fluke 3000 FC Series Wireless Multimeter
* For all specifications: accuracy is specified for one year after calibration, at operating
temperatures of 18 °C to 28 °C, with relative humidity at 0 % to 90 %. Accuracy specifications
take the form of ± ([ % of Reading ] + [ Number of least significant digits ]).
Detailed Specifications
AC Voltage
600.0 mV / 0.1 mV
6.000 V / 0.001 V
Range ¹ / Resolution
60.00 V / 0.01 V
600.0 V / 0.1 V
1000 V / 1 V
45 Hz to 500 Hz 1.0 % + 3
Accuracy 2 3 4
500 Hz to 1 kHz 2.0% + 3
1
All AC voltage ranges are specified from 1 % of range to 100 % of range.
2
Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor < 1.5 at 1000 V.
3
For non-sinusoidial waveforms, add ± (2 % of reading + 2 % full scale) typical, for crest factor
up to 3.
4
Do not exceed 107 V-Hz.
DC Voltage, Continuity, Resistance, Diode Test and Capacitance
Function
Range /
600.0 mV / 0.1 mV
Resolution
mV
Accuracy
0.09 % + 2
6.000 V / 0.001 V
60.00 V / 0.01 V
Range /
600.0 V / 0.1 V
Resolution
V
1000 V / 1 V
0.09 % + 2
Accuracy
0.15 % + 2
Range /
600 Ω / 1 Ω
Resolution
)))
Meter beeps at < 25 Ω, beeper detects opens or
Accuracy
shorts of 250 μs or longer.
600.0 Ω / 0.1 Ω
Range /
Ω
6.000 kΩ / 0.001 kΩ
Resolution
Accuracy
Diode test
Range /
Resolution
Accuracy
60.00 kΩ / 0.01 kΩ
60.00 kΩ / 0.1 kΩ
600.0 kΩ / 0.001 MΩ
50.00 MΩ / 0.01 MΩ
0.5 % + 2
0.5 % + 1
1.5 % + 3
2.000 V / 0.001 V
1%+2
1000 nF / 1 nF
10.00 μF / 0.01 μF
Range /
100.0 μF / 0.1 μF
Resolution
µF
9999 μF1 / 1 μF
1.2 % + 2
Accuracy
10 % typical
1
In the 9999 μF range for measurements to 1000 μF, the measurement accuracy is 1.2 % + 2.
AC and DC Current
Function
60.00 mA / 0.01 mA
Range 1 /
mA AC
400.0 mA 3/ 0.1 mA
Resolution
(45 Hz to 1 kHz)
Accuracy
1.5 % + 3
1
60.00 mA / 0.01 mA
Range /
2
400.0 mA 3/ 0.1 mA
Resolution
mA DC
Accuracy
0.5 % + 3
1
All AC current ranges are specified from 5 % of range to 100 % of range.
2
Input burden voltage (typical): 400 mA input 2 mV/mA.
3
400.0 mA accuracy specified up to 600 mA overload.
Frequency
99.99 Hz / 0.01 Hz
999.9 Hz / 0.1 Hz
Range / Resolution
9.999 kHz / 0.001 kHz
99.99 kHz / 0.01 kHz
1
Accuracy
0.1 % + 1
1
Frequency is specified up to 99.99 kHz in volts and up to 10 kHz in amps.
Input Characteristics
Input
Overload
Common mode rejection ratio Normal mode
Function
impedance
protection
(1 kΩ unbalance)
rejection
(nominal)
> 10 MΩ < 100
> 60 dB at 50
1100 V RMS
> 120 dB at DC, 50 Hz or 60 Hz
pF
Hz or 60 Hz
1100 V RMS
> 10 MΩ < 100 > 60 dB, DC to 60 Hz
> 60 dB at 50
1100 V RMS
pF
> 10 MΩ < 100
> 120 dB at DC, 50 Hz or 60 Hz
pF
Full scale voltage
Open circuit test voltage
To 6 MΩ
50 MΩ
Hz or 60 Hz
> 60 dB at 50
Hz or 60 Hz
Typical short
circuit current
Typical short
circuit current
< 350 mA
< 1.1 mA
1100 V RMS
< 2.7 V DC
< 0.7 V DC
< 0.9 V DC
1100 V RMS
< 2.7 V DC
2.000 V DC
mA Function
Overload protection
Fused, 44/100 A, 1000 V FAST Fuse
Overload
600 mA overload for 2 minutes maximum, 10 minutes rest minimum
MIN/MAX Recording Accuracy
The specified accuracy of the measurement function ± 12 counts for
DC functions
changes > 350 mS in duration.
The specified accuracy of the measurement function ± 40 counts for
AC functions
changes > 900 mS in duration.
General Specifications
Maximum voltage
between any terminal and 1000 V DC or AC RMS
earth ground
Ω fuse protection from A 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part
inputs
only
4/sec
Update rate
Volts, amps,
6000 counts
ohms
Display (LCD)
Frequency
10,000 counts
Capacitance
1,000 counts
Battery type
Three AA Alkaline batteries, NEDA 15A IEC LR6
Battery life
250 hours minimum
RF communications
2.4 GHZ ISM Band
Up to 20m
Open air, unobstructed
Obstructed, sheetrock wall
Up to 6.5m
RF communication range
Obstructed, concrete wall, or steel electrical
Up to 3.5m
enclosure
-10 °C to 50 °C
Operating
Temperature
Storage
-40 °C to 60 °C
Temperature coefficient 0.1 X (specified accuracy) /°C (<18 °C or >28 °C)
0 % to 90 % (0 °C to 35 °C)
Relative humidity
0 % to 75 % (35 °C to 40 °C)
Altitude
Electromagnetic
compatibility
EMI, RFI, EMC, RF
Safety compliance
0 % to 45 % (40 °C to 50 °C)
2,000 m
Operating
Storage
12,000 m
EN 61326-1:2006, EN 61326-2-2:2006
ETSI EN 300 328 V1.7.1:2006, ETSI EN 300 489
V1.8.1:2008,
FCC Part 15 Subpart C Sections 15.207, 15.209, 15.249
FCCID : FCC: T68-FDMMBLE IC: 6627A-FDMMBLE
ANSI/ISA 61010-1 / (82.02.01): 3rd edition
CAN/CSA-C22.2 No 61010-1-12: 3rd edition
UL 61010-1: 3rd edition
IEC/EN 61010-1:2010
CSA, FCC, CE
Certifications
Ingress Protection (IP)
IP54
rating
Pollution Degree
2
Size (H x W x L)
4.75 cm x 9.3 cm x 20.7 cm (1.87 in x 3.68 in x 8.14 in)
Weight
340 g (12 oz)
Note: Not compatible with Fluke CNX test tools
Fluke v3000 FC Wireless AC Voltage Module
Accuracy is specified as ± ([% of reading] + [number of least significant digits]). All ranges are
autoranging. Accuracy is specified from 5 % to 100 % of the range obtained by autoranging,
from 18 °C to 28 °C.
Specifications
6.000 V / 0.001 V
60.00 V / 0.01 V
Range ¹ / Resolution
600.0 V / 0.1 V
1000 V / 1 V
1.0 % + 3
45 Hz to 500 Hz
Accuracy 2 3 4
500 Hz to 1 kHz
2.0% + 3
1
All AC voltage ranges are specified from 1 % of range to 100 % of range.
2
Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor < 1.5 at 1000 V.
3
For non-sinusoidial waveforms, add ± (2 % of reading + 2 % full scale) typical, for crest factor
up to 3.
4
Do not exceed 107 V-Hz.
General Specifications
LCD w/backlight
3½ digits, 6000 counts, updates 4/sec
Battery type
2 AA, NEDA 15 A, IEC LR6
Battery life
Memory
RF communications
400 Hours
Record up to 65000 readings
2.4 GHZ ISM Band
Up to 20 m
Open air, unobstructed
Obstructed, sheetrock wall
Up to 6.5 m
RF communication range
Obstructed, concrete wall or steel
Up to 3.5 m
electrical enclosure
Operating temperature
-10 °C to +50 °C
Storage temperature
-40 °C to +60 °C
Temperature coefficient
0.1 X (specified accuracy)/°C (< 18 °C or > 28 °C)
Operating humidity
90 % at 35 °C, 45 % at 40 °C, 45 % at 50 °C
2,000 m
Operating
Altitude
Storage
12,000 m
EMC
EN 61326-1:2006
ANSI/ISA 61010-1 / (82.02.01) 3rd edition
CAN/CSA-C22.2 No 61010-1-12 3rd edition
Safety compliance
UL 61010-1
3rd edition
IEC/EN 61010-1
2010
Safety rating
CAT IV 600 V, CAT III 1000 V
Certifications
CSA, CE, FCC: T68-FBLE IC: 6627A-FBLE
Ingress Protection (IP) rating
IP42
Pollution Degree
2
Size (HxWxD)
16.5 cm x 6.35 cm x 1.4 cm (6.5 in x 2.5 in x 1.4 in)
Weight
.22 kg (8 oz)
Note: Not compatible with Fluke CNX test tools
Fluke v3001 FC Wireless DC Voltage Module
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C,
with relative humidity at 0 % to 90 %. Accuracy specifications take the form of ±([ % of
Reading ] + [ Number of least significant digits ]).All ranges are autoranging. Accuracy is
specified from the range obtained by autoranging, from 18 °C to 28 °C.
Voltage Specifications
Function
mV DC
Range ¹ / Resolution
Accuracy
V DC
Range ¹ / Resolution
600.0 mV / 0.1 mV
0.09 % + 3
6.000 V / 0.001 V
60.00 V / 0.01 V
Accuracy
Range ¹ / Resolution
Accuracy
General Specifications
LCD w/backlight
Battery type
Log rate/interval
Battery life
Memory
RF communications
600.0 V / 0.1 V
0.09 % + 3
1000 V / 1 V
0.15 % + 2
3 ½ digits, 6000 counts, 4/sec update rate
2 AA Alkaline batteries, NEDA 15A, IEC LR6
Adjustable by PC from 1 sec to 1 hr, default 1 min
400 hours minimum
Record up to 65,000 readings
2.4 GHz ISM Band
Up to 20 m
Open air, unobstructed
Obstructed, sheetrock wall
Up to 6.5 m
RF communication range
Obstructed, concrete wall or steel
Up to 3.5 m
electrical enclosure
Operating temperature
-10 °C to +50 °C
Storage temperature
-40 °C to +60 °C
Temperature coefficient
0.1 X (specified accuracy)/°C (< 18 °C or > 28 °C)
0 % to 90 % (0 °C to 35 °C), 0 % to 75 % (35 °C to 40 °C),
Relative humidity
0 % to 45 % (40 °C to 50 °C)
EMC
IEC 61236-1, Portable
IEC 61010-1, 600 V CAT IV / 1000 V CAT III, 3rd edition,
Safety compliance
Pollution Degree 2
Certifications
CSA, CE, FCC: T68-FBLE IC: 6627A-FBLE
Ingress Protection (IP) rating IP42
Size (HxWxD)
160 mm x 66 mm x 38 mm (6.3 in x 2.6 in x 1.5 in)
Weight
0.255 kg (9 oz)
Note: Not compatible with Fluke CNX test tools
Fluke a3001 FC Wireless iFlex® AC Current Module
Specifications
Range
Resolution
Accuracy
Crest factor (50 Hz/60 Hz)
0.5 A to 2500 A AC
0.1 A
3 % ± 5 digits (5 Hz to 500 Hz)
3.0 at 1100 A
2.5 at 1400 A
1.42 at 2500 A
add 2 % for C.F. > 2
LCD w/backlight
Log rate/interval
Battery type
Battery life
Memory
RF communications
3½ digits
1 sec minimum/adjustable by PC or front panel
2 AA, NEDA 15 A, IEC LR6
400 hours
Record up to 65,000 readings
2.4 GHZ ISM Band
Up to 20 m
Open air, unobstructed
Obstructed, sheetrock wall
Up to 6.5 m
RF communication range
Obstructed, concrete wall or
Up to 3.5 m
steel electrical enclosure
Operating temperature
-10 °C to +50 °C
Storage temperature
-40 °C to +60 °C
Temperature coefficient
Add 0.1 X (specified accuracy) / °C (<18 °C or >28 °C)
90 % at 35 °C
Operating humidity
75 % at 40 °C
45 % at 50 °C
2,000 m
Operating
Altitude
Storage
12,000 m
EMC
EN 61326-1:2006
Safety compliance
IEC 61010-1, 600 V CAT IV/1000 V CAT III, 3rd edition
Safety rating
CAT IV 600 V, CAT III 1000 V
Certifications
CSA, CE, FCC: T68-FBLE IC: 6627A-FBLE
Ingress Protection (IP) rating IP42
Pollution Degree
2
Jaw opening
25.4 cm (10 in) coil
Size (HxWxD)
16.5 cm x 6.35 cm x 1.4 cm (6.5 in x 2.5 in x 1.4 in)
Weight
.22 kg (8 oz)
Note: Not compatible with Fluke CNX test tools