RPS-65 65W Single Output Medical Type series Applications

65W Single Output Medical Type
RPS-65 series
EN 60601-1
IEC 60601-1
8
Features
Applications
3"×2" miniature size
Universal AC input / Full range
ClassⅡ(without FG) installations
Medical safety approved
(2 x MOPP between primary to secondary)
Suitable for BF application with appropriate system consideration
Low leakage current <100μA
No load power consumption<0.1W
High efficiency up to 91%
For 1U applications
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
-30~70℃wide range of operating temperature
Operating altitude up to 4000 meters(Note 6.)
LED indicator for power on
3 years warranty
Oral irrigator
Hemodialysis machine
Medical monitors
Sleep apnea devices
Description
RPS-65 is a 65W highly reliable green PCB type medical power supply with a high power density on the 3”
by 2” footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V. The
working efficiency is up to 91% and the extremely low no load power consumption is down below 0.1W.
RPS-65 is able to be used for ClassⅡ (no FG) system design. The extremely low leakage current is less
than 100 μ A. In addition, it conforms to international medical regulations (2*MOPP) and EMC EN55011,
perfectly fitting all kinds of BF rated “patient contact” medical system equipment.y
Model Encoding
RPS - 65 - 3.3
Output voltage
Output wattage
Series name
File Name: RPS - 65 -SPEC 2015-11-04
RPS-65 series
65W Single Output Medical Type
SPECIFICATION
ORDER NO.
OUTPUT
RPS-65-3.3
RPS-65-5
RPS-65-7.5
RPS-65-12
RPS-65-15
RPS-65-24
RPS-65-48
DC VOLTAGE
3.3V
5V
7.5V
12V
15V
24V
48V
RATED CURRENT
10A
10A
8A
5.42A
4.34A
2.71A
1.36A
CURRENT RANGE
0 ~ 11A
0 ~ 11A
0 ~ 8.8A
0 ~ 5.96A
0 ~ 4.77A
0 ~ 2.98A
0 ~ 1.49A
RATED POWER
33W
50W
60W
65W
65.1W
65W
65.3W
36.3W
PEAK LOAD(10sec.)
RIPPLE & NOISE (max.) Note.2 80mVp-p
55W
80mVp-p
66W
80mVp-p
71.5W
120mVp-p
71.6W
150mVp-p
71.5W
240mVp-p
71.5W
300mVp-p
VOLTAGE ADJ.RANGE
VOLTAGE TOLERANCE
4.7~5.5V
±2.0%
±0.5%
±2.0%
7.12~8.3V
±2.0%
±0.5%
±2.0%
11.4~13.2V
±2.0%
±0.5%
±2.0%
13.5~16.5V
±1.0%
±0.5%
±1.0%
22.8~27.6V
±1.0%
±0.5%
±1.0%
45.6~52.8V
±1.0%
±0.5%
±1.0%
89%
90%
91%
27.6~32.4V
55.2~64.8V
LOAD REGULATION
2.9~3.6V
±2.0%
±0.5%
±2.0%
SETUP, RISE TIME
500ms, 30ms / 230VAC
Note.3
LINE REGULATION
HOLD UP TIME (Typ.)
VOLTAGE RANGE
INPUT
FREQUENCY RANGE
47 ~ 63Hz
EFFICIENCY (Typ.)
80%
AC CURRENT (Typ.)
INRUSH CURRENT (Typ.)
1.5A / 115VAC
LEAKAGE CURRENT
OVERLOAD
PROTECTION
30ms / 230VAC
Note.4 80 ~ 264VAC
OVER VOLTAGE
WORKING TEMP.
WORKING HUMIDITY
500ms, 30ms / 115VAC at full load
12ms / 115VAC at full load
84%
85%
88%
1A / 230VAC
СOLD STAR 30A/115VAC 50A/230VAC
Note.5 Touch current< 100 A/264VAC
115 ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
3.8~4.46V
5.75~6.75V
8.62~11.3V
13.8~16.2V
17.25~20.25V
Protection type : Shut down o/p voltage, re-power on to recover
-30 ~ +70℃ (Refer to "Derating Curve")
20% ~ 90% RH non-condensing
ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85℃, 10 ~ 95% RH
TEMP. COEFFICIENT
±0.03% / ℃ (0 ~ 50℃)
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
ANSI/AAMI ES60601-1 , TUV EN60601-1 , I EС60601-1 approved
SAFETY STANDARDS
Primary-Secondary: 2xMOPP
SAFETY & ISOLATION LEVEL
WITHSTAND VOLTAGE
I/P-O/P: 4KVAC
EMC
(Note. 6) ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
OTHERS
EMC EMISSION
EMC IMMUNITY
Compliance to EN55011(CISPR11) class B, EN61000-3-2,3
MTBF
DIMENSION
959.1Khrs min. MIL-HDBK-217(25℃)
76.2*50.8*24mm or 3" * 2" *0.945" inch (L*W*H)
0.11Kg; 120pcs/14.2Kg/0.97CUFT
PACKING
NOTE
Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2 medical level, criteria A
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25℃ of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. Derating may be needed under low input voltages. Please check the derating curve for more details.
5. Touch current was measured from primary input to DC output.
6. The ambient temperature derating of 5℃/1000m is needed for operating altitude greater than 2000m (6500ft).
7. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that
it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
File Name: RPS - 65 -SPEC 2015-11-04
RPS-65 series
65W Single Output Medical Type
Block Diagram
fosc : 65KHz
EMI FILTER
&
RECTIFIER
I/P
RECTIFIERS
&
FILTER
RECTIFIERS
&
FILTER
POWER
SWITCHING
+V
-V
DETECTION
CIRCUIT
O.L.P.
CONTROL
O.V.P.
Derating Curve
Static Characteristics
100
100
90
80
80
70
LOAD (%)
LOAD (%)
60
40
20
-30
0
10
20
30
40
50
AMBIENT TEMPERATURE (℃)
60
70 (HORIZONTAL)
60
50
40
80
100
115
120
140
160
180
200
220
240
264
INPUT VOLTAGE (V) 60Hz
File Name: RPS - 65 -SPEC 2015-11-04
RPS-65 series
65W Single Output Medical Type
Mechanical Specification
Case No. Unit:mm
76.2
3.175
3.175
69.85
HS100
AC FUSE
T2A/250V
3
2
1
44.45
50.8
FS1
CN2
1
2
3
4
CN1
LED
SVR1
HS1
3 max.
24
4-ψ3.2
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
AC/N
No Pin
AC/L
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
DC Output Connector (CN2) : JST B2P-VH or equivalent
Pin No.
1
2
3
4
Assignment
+V
+V
-V
-V
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Installation Manual
Please refer to : http://www.meanwell.com/webnet/search/InstallationSearch.html
File Name: RPS - 65 -SPEC 2015-11-04