Craftsman 82003 Owner's Manual
Add to my manuals
25 Pages
Craftsman 82003 is a versatile digital multimeter that can measure various electrical parameters such as DC and AC voltage, current, resistance, capacitance, frequency, and temperature. It features a 4,000-count backlit LCD display, autoranging and manual range selection, relative mode, data hold, and auto power-off functions. With its IP67 waterproof enclosure, the device offers protection against dust and water, making it suitable for use in harsh conditions.
advertisement
Owner's Manual
True RMS
Multimeter
Model No.
82003
CAUTION: Read, understand and follow Safety Rules and Operating
Instructions in this manual before using this product.
Safety
Operation
Maintenance
Español
© Sears, Roebuck and Co., Hoffman Estates, IL 60179 U.S.A. www.craftsman.com 042508
TABLE OF CONTENTS
Warranty
Safety Instructions
Safety Symbols
Control and Jacks
Symbols and Annunciators
Specifications
Battery Installation
Operating Instructions
DC Voltage Measurements
AC Voltage Measurements
DC Current Measurements
AC Current Measurements
Resistance Measurements
Continuity Check
Diode Test
Capacitance Measurements
Frequency (Duty Cycle) Measurements
Temperature Measurements
Autoranging/Manual range selection
Relative Mode
Display Backlight
Hold
Auto Power Off
Low Battery Indication
Battery Replacement
Fuse Replacement
Service and Parts
2
16
17
17
18
18
19
20
20
21
21
21
21
Page 3
4
5
6
6
7
11
12
12
13
14
15
23
24
25
ONE YEAR FULL WARRANTY
CRAFTSMAN PROFESSIONAL ONE YEAR FULL
WARRANTY. If this product fails due to a defect in materials or workmanship within one year from the date of purchase,
RETURN IT TO ANY SEARS STORE OR OTHER
CRAFTSMAN OUTLET IN THE UNITED STATES for free replacement. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
Sears, Roebuck and Co., Hoffman Estates, IL 60179
For Customer Assistance Call 9am - 5pm (ET)
Monday through Friday 1-888-326-1006
WARNING: USE EXTREME CAUTION IN THE USE OF THIS
DEVICE. Improper use of this device can result in injury or death. Follow all safeguards suggested in this manual in addition to the normal safety precautions used in working with electrical circuits. DO NOT service this device if you are not qualified to do so.
3
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules listed below must be carefully followed for safe operation.
1. NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Limits
V DC or V AC mA AC/DC
A AC/DC
1000VDC/AC rms
400mA AC/DC
10A AC/DC (20A for 30 seconds max every 15 minutes)
Resistance, Diode Test,
Continuity, Capacitance,
Frequency, Temperature
1000VDC/AC rms
2. USE EXTREME CAUTION when working with high voltages
3. DO NOT measure voltage if the voltage on the "COM" input jack exceeds 1000V above earth ground
4. DO NOT measure current of circuits whose voltage is greater than 600V above earth ground
5. NEVER connect the meter leads across a voltage source while the function switch is in the resistance or diode mode.
Doing so can damage the meter
6. ALWAYS turn off the power and disconnect the test leads before opening the cover to replace the fuse or battery
7. NEVER operate the meter unless the back cover is in place and fastened securely
4
SAFETY SYMBOLS
This symbol adjacent to another symbol, terminal or operating device indicates that the operator must refer to an explanation in the
Operating Instructions to avoid personal injury or damage to the meter.
WARNING
CAUTION
MAX
1000V
This WARNING symbol indicates a potentially hazardous situation, which if not avoided, could result in death or serious injury.
This CAUTION symbol indicates a potentially hazardous situation, which if not avoided, may result damage to the product.
This symbol advises the user that the terminal(s) so marked must not be connected to a circuit point at which the voltage with respect to earth ground exceeds (in this case)
1000 VAC or VDC.
This symbol adjacent to one or more terminals identifies them as being associated with ranges that may, in normal use, be subjected to particularly hazardous voltages. For maximum safety, the meter and its test leads should not be handled when these terminals are energized.
This symbol indicates that a device is protected throughout by double insulation or reinforced insulation.
5
CONTROLS AND JACKS
1. 4,000 count LCD display
3. Hz and % button
6. mA, µA and 10A input jacks
7. COM input jack
10. REL button
11. HOLD button
Note: Tilt stand and battery compartment are on rear of unit.
SYMBOLS AND ANNUNCIATORS
•))) Continuity
Diode test
Battery status n nano
(10 -6
) (capacitance)
) (amps) m milli
A Amps
) (ohms)
(capacitance)
6 ) (ohms)
Ohms
Hz Hertz V Volts
% Percent (duty ratio) REL Relative
DC
ºF
Direct current
Degrees Fahrenheit
HOLD Display hold
ºC Degrees Centigrade
6
SPECIFICATIONS
Function Range Resolution
DC Voltage
Accuracy
400mV 0.1mV (0.5% reading + 2 digits)
4V 0.001V
40V 0.01V (1.2% reading + 2 digits)
AC Voltage
400V 0.1V
1000V 1V
40V 0.01V
400V 0.1V
(1.5% reading + 2 digits)
400mV 0.1mV (2.0% reading + 10 digits)
4V 0.001V
(2.0% reading + 5 digits)
DC Current
AC Current
1000V 1V
400 A 0.1A
4000 A 1A
40mA 0.01mA
(2.5% reading + 5 digits)
(1.0% reading + 3 digits)
400 A 0.1A
4000 A 1A
(1.5% reading + 3 digits)
400mA 0.1mA
10A 0.01A (2.5% reading + 5 digits)
(2.5% reading + 10 digits)
40mA 0.01mA
400mA 0.1mA
(2.5% reading + 5 digits)
Resistance
10A 0.01A (3.0% reading + 7 digits)
400 0.1 (1.2% reading + 4 digits)
(1.0% reading + 2 digits) 4k 0.001k
40k 0.01k
400k 0.1k (1.2% reading + 2 digits)
4M 0.001M
40M 0.01M (2.0% reading + 3 digits)
7
Function Range Resolution Accuracy
Capacitance 40nF 0.01nF (5.0% reading + 7 digits)
Frequency
400nF 0.1nF
4 F 0.001F
40 F 0.01F
100 F 0.1F
5.999Hz 0.001Hz
59.99Hz 0.01Hz
(3.0% reading + 5 digits)
(5% reading + 5 digits)
(1.5% reading + 1 digits)
599.9Hz 0.1Hz
5.999kHz 0.001kHz
59.99kHz 0.01kHz
599.9kHz 0.1kHz
(1.2% reading + 3 digits)
Duty Cycle
Temp
(type-K)
5.999MHz 0.001MHz
9.999MHz 0.001MHz
(1.5% reading + 4 digits)
Sensitivity: 0.5V rms <500kHz; 3V rms >500kHz
0.1 to 99.9% 0.1% (1.2% reading + 2 digits)
Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz
-4 to 1382 F
-20 to 750 C
0.1
<400
1 ≥ 400
(3.0% reading + 5C /9F )
(probe accuracy not included)
Supplied Probe range: -58.0 to 482 o F (-50.0 to 250 o C)
Note: An optional probe that is rated to the full range of the meter is available from the www.craftsman.com
website.
Note: Accuracy is stated at 65 o than 75% RH.
F to 83 o F (18 o C to 28 o C) and less
Note: Accuracy specifications consist of two elements:
(% reading) – This is the accuracy of the measurement circuit.
(+ digits) – This is the accuracy of the analog to digital converter.
8
Enclosure
Shock (Drop Test)
Diode Test
Continuity Check
Double molded, waterproof (IP67)
6.5 feet (2 meters)
Test current of 0.3mA typical, open circuit voltage 1.5V DC typical
Audible signal will sound if the resistance is less than 100 (approx.), test current <0.3mA
Temperature Sensor Requires type K thermocouple
Input Impedance 7.8MΩ
AC Response
ACV Bandwidth
Crest Factor
True rms
50Hz to 400Hz
≤3 at full scale up to 500V, decreasing linearly to ≤1.5 at 1000V
4,000 count backlit liquid crystal Display
Overrange indication “OL” is displayed
Auto Power Off 30 minutes (approximately)
Polarity Automatic (no indication for positive);
Minus (-) sign for negative
Measurement Rate 2 times per second, nominal
Low Battery Indication “ ” is displayed if battery voltage drops below operating voltage
Battery One 9 volt (NEDA 1604) battery
Fuses mA, µA ranges; 0.5A/1000V ceramic fast blow
A range; 10A/1000V ceramic fast blow
Operating Temperature 41 F to 104F (5C to 40C)
Storage Temperature -4 F to 140F (-20C to 60C)
Operating Humidity Max 80% up to 87 F (31C) decreasing linearly to 50% at 104 F (40C)
Storage Humidity <80%
Operating Altitude
Weight
Size
7000ft. (2000meters) maximum.
0.908lb (412g) (includes holster).
7.2” x 3.2” x 2.2” (182 x 82 x 55mm)
9
Safety This meter is intended for origin of installation use and protected, against the users, by double insulation per EN61010-1 and IEC61010-1 2 nd
Edition (2001) to Category IV 600V and Category III
1000V; Pollution Degree 2. The meter also meets
UL 61010-1, 2 nd Edition (2004), CAN/CSA C22.2
No. 61010-1 2 nd Edition (2004), and UL 61010B-2-
031, 1 st Edition (2003)
Approvals UL CE
UL Listed The UL mark does not indicate that this product has been evaluated for the accuracy of its readings.
10
BATTERY INSTALLATION
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover.
1. Turn power off and disconnect the test leads from the meter.
2. Open the rear battery cover by removing two screws (B) using a Phillips head screwdriver.
3. Insert the battery into battery holder, observing the correct polarity.
4. Put the battery cover back in place. Secure with the screws.
WARNING: To avoid electric shock, do not operate the meter until the battery cover is in place and fastened securely.
NOTE: If your meter does not work properly, check the fuses and batteries to make sure that they are still good and that they are properly inserted.
A A
B
B A
F1 F2
11
OPERATING INSTRUCTIONS
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very dangerous and should be measured with great care.
1. ALWAYS turn the function switch to the OFF position when the meter is not in use.
2. If “OL” appears in the display during a measurement, the value exceeds the range you have selected. Change to a higher range.
NOTE: On some low AC and DC voltage ranges, with the test leads not connected to a device, the display may show a random, changing reading. This is normal and is caused by the high-input sensitivity. The reading will stabilize and give a proper measurement when connected to a circuit.
DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure DC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter.
1. Set the function switch to the V position.
2. Press the MODE button to select DC (if necessary).
3. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
4. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
5. Read the voltage in the display.
12
AC VOLTAGE (Frequency, Duty Cycle) MEASUREMENTS
AC VOLTAGE: The probe tips may not be long enough to contact the live parts inside some 240V outlets for appliances because the contacts are recessed deep in the outlets. As a result, the reading may show 0 volts when the outlet actually has voltage on it.
WARNING: To avoid electric shock, make sure the probe tips are touching the metal contacts inside the outlet before assuming no voltage is present
CAUTION: Do not measure AC voltages if a motor on the circuit is being switched ON or OFF. Large voltage surges may occur that can damage the meter.
1. Set the function switch to the V position.
2. Press the MODE button to select AC (if necessary)
3. Insert the black test lead banana plug into the negative COM jack.
Insert red test lead banana plug into the positive V jack.
4. Touch the black test probe tip to the neutral side of the circuit.
Touch the red test probe tip to the “hot” side of the circuit.
5. Read the voltage in the display.
6. Press the HZ/% button to indicate “Hz”.
7. Read the frequency in the display.
8. Press the Hz/% button again to indicate “%”.
9. Read the % of duty cycle in the display.
13
DC CURRENT MEASUREMENTS
CAUTION: Do not make current measurements on the 20A scale for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative COM jack.
2. For current measurements up to 4000µA DC, set the function switch to the µA position and insert the red test lead banana plug into the µA/mA jack.
3. For current measurements up to 400mA DC, set the function switch to the mA position and insert the red test lead banana plug into the µA/mA jack.
4. For current measurements up to 10A DC, set the function switch to the 10A position and insert the red test lead banana plug into the 10A jack.
5. Press the MODE button to indicate “DC” on the display.
6. Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
7. Touch the black test probe tip to the negative side of the circuit.
Touch the red test probe tip to the positive side of the circuit.
8. Apply power to the circuit.
9. Read the current in the display.
14
AC CURRENT (Frequency, Duty Cycle) MEASUREMENTS
CAUTION: Do not make current measurements on the 20A scale for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert the black test lead banana plug into the negative COM jack.
2. For current measurements up to 4000µA
AC, set the function switch to the µA position and insert the red test lead banana plug into the µA/mA jack.
3. For current measurements up to 400mA
AC, set the function switch to the mA position and insert the red test lead banana plug into the µA/mA jack.
4. For current measurements up to 10A AC, set the function switch to the 10A position and insert the red test lead banana plug into the 10A jack.
5. Press the MODE button to indicate “AC” on the display.
6. Remove power from the circuit under test, then open up the circuit at the point where you wish to measure current.
7. Touch the black test probe tip to the neutral side of the circuit.
Touch the red test probe tip to the “hot” side of the circuit.
8. Apply power to the circuit.
9. Read the current in the display.
10. Press the Hz/% button to indicate “Hz”.
11. Read the frequency in the display.
12. Press the Hz/% button again to indicate “%”.
13. Read the % duty cycle in the display.
14. Press the Hz/% button to return to current measurement.
15
RESISTANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the line cords.
1. Set the function switch to the Ω CAP position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive jack.
3. Press the MODE button to indicate “ "on the display.
4. Touch the test probe tips across the circuit or part under test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the resistance reading.
5. Read the resistance in the display.
16
CONTINUITY CHECK
WARNING: To avoid electric shock, never measure continuity on circuits or wires that have voltage on them.
1. Set the function switch to the Ω CAP position.
2. Insert the black lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive jack.
3. Press the MODE button to indicate” "and
“Ω” on the display
4. Touch the test probe tips to the circuit or wire you wish to check.
5. If the resistance is less than approximately
100 , the audible signal will sound. If the circuit is open, the display will indicate “OL”.
DIODE TEST
1. Set the function switch to the Ω CAP position.
2. Insert the black test lead banana plug into the negative COM jack and the red test lead banana plug into the positive V jack.
3. Press the MODE button to indicate and V
on the display.
4. Touch the test probes to the diode under test.
Forward voltage will typically indicate 0.400 to 0.700V. Reverse voltage will indicate “OL”.
Shorted devices will indicate near 0V and an open device will indicate “OL” in both polarities.
17
CAPACITANCE MEASUREMENTS
WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any capacitance measurements. Remove the batteries and unplug the line cords.
1. Set the rotary function switch to the Ω CAP position.
2. Insert the black test lead banana plug into the negative COM jack.
Insert the red test lead banana plug into the positive V jack.
3. Press the MODE button to indicate “nF” on the display.
4. Touch the test leads to the capacitor to be tested.
5. The test may take up to 3 minutes or more for large capacitors to charge. Wait until the readings settle before ending the test.
6. Read the capacitance value in the display
FREQUENCY (Duty Cycle) MEASUREMENTS (Electronic)
1. Set the rotary function switch to the “Hz %” position.
2. Press the Hz/% button to indicate “Hz” in the display.
3. Insert the black lead banana plug into the negative COM jack and the red test lead banana plug into the positive Hz jack.
4. Touch the test probe tips to the circuit under test.
5. Read the frequency on the display.
6. Press the Hz/% button again to indicate “%” on the display.
7. Read the % of duty cycle on the display.
18
TEMPERATURE MEASUREMENTS
1. Set the function switch to the ºF or ºC position.
2. Insert the Temperature Probe into the input jacks, making sure to observe the correct polarity.
3. Touch the Temperature Probe head to the part whose temperature you wish to measure. Keep the probe touching the part under test until the reading stabilizes (about 30 seconds).
4. Read the temperature in the display.
Note: The temperature probe is fitted with a type K mini connector.
A mini connector to banana connector adaptor is supplied for connection to the input banana jacks.
Note: The supplied thermocouple is rated for 482°F (250°C). For higher temperatures you can purchase a high temperature thermocouple from www.craftsman.com
.
FREQUENCY SENSITIVITY (ELECTRICAL)
The frequency sensitivity is range dependent when the function is selected from while in the voltage or current measuring function. Below are typical sensitivities for the “electrical” measurement modes.
Range
( DC/AC )
4V ≥1.5V rms 5Hz~10kHz
40V, 400V
400mA
10A
≥10V rms
≥20V rms
1000V/1000V ≥420V rms
≥45mA rms
≥4A rms
5Hz~20kHz
5Hz~200kHz
50Hz~1kHz
5Hz~5kHz
5Hz~1kHz
19
AUTORANGING/MANUAL RANGE SELECTION
When the meter is first turned on, it automatically goes into
Autoranging. This automatically selects the best range for the measurements being made and is generally the best mode for most measurements. For measurement situations requiring that a range be manually selected, perform the following:
1. Press the RANGE key. The “Auto” display indicator will turn off.
2. Press the RANGE key to step through the available ranges.
Observe the decimal point and units displayed until the preferred range is located.
3. To exit the Manual Ranging mode and return to Autoranging, press and hold the RANGE key for 2 seconds.
Note: Manual ranging does not apply for the Capacitance and
Frequency functions.
RELATIVE MODE
The relative measurement feature allows you to make measurements relative to a stored zero reference value. A reference voltage, current, etc. can be stored and measurements made in comparison to that value. The displayed value is the difference between the reference value and the measured value.
1. Perform the measurement as described in the operating instructions.
2. Press the REL button to store (zero) the reading in the display and the "REL" indicator will appear on the display.
3. The display will now indicate the difference between the stored value and the measured value.
4. Press the REL button to exit the relative mode.
Note: The Relative function does not operate in the Frequency function.
DISPLAY BACKLIGHT
Press the backlight button to turn on. The light will slowly fade to off after approxamately 10 to 15 seconds.
HOLD
The hold function freezes the reading in the display. Press the
HOLD key momentarily to activate or to exit the HOLD function.
20
AUTO POWER OFF
The auto off feature will turn the meter off after 30 minutes. To disable the auto power off feature, hold down the MODE button and turn the meter on.
LOW BATTERY INDICATION
The icon will appear in the display when the battery voltage becomes low. Replace the battery when this appears.
21
MAINTENANCE
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the back cover or the battery cover.
WARNING: To avoid electric shock, do not operate your meter until the battery cover is in place and fastened securely.
This multimeter is designed to provide years of dependable service, if the following care instructions are performed:
1. KEEP THE METER DRY. If it gets wet, wipe it off.
2. USE AND STORE THE METER IN NORMAL
TEMPERATURES. Temperature extremes can shorten the life of the electronic parts and distort or melt plastic parts.
3. HANDLE THE METER GENTLY AND CAREFULLY. Dropping it can damage the electronic parts or the case.
4. KEEP THE METER CLEAN. Wipe the case occasionally with a damp cloth. DO NOT use chemicals, cleaning solvents, or detergents.
5. USE ONLY FRESH BATTERIES OF THE RECOMMENDED
SIZE AND TYPE. Remove old or weak batteries so they do not leak and damage the unit.
6. IF THE METER IS TO BE STORED FOR A LONG PERIOD
OF TIME, the batteries should be removed to prevent damage to the unit.
UL LISTED
The UL mark does not indicate that this product has been evaluated for the accuracy of its readings.
22
BATTERY INSTALLATION
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover.
1. Turn power off and disconnect the test leads from the meter.
2. Open the rear battery cover by removing two screws (B) using a Phillips head screwdriver.
3. Insert the battery into battery holder, observing the correct polarity.
4. Put the battery cover back in place. Secure with the screws.
WARNING: To avoid electric shock, do not operate the meter until the battery cover is in place and fastened securely.
NOTE: If your meter does not work properly, check the fuses and batteries to make sure that they are still good and that they are properly inserted.
REPLACING THE FUSES
WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the fuse cover.
1. Disconnect the test leads from the meter.
2. Remove the battery cover
(two “B” screws) and the battery.
A
B
A
F1 F2
B A
3. Remove the six “A” screws securing the rear cover.
4. Gently remove the old fuse and install the new fuse into the holder.
23
5. Always use a fuse of the proper size and value (0.5A/1000V fast blow for the 400mA range [SIBA 70-172-40], 10A/1000V fast blow for the 20A range [SIBA 50-199-06]).
6. Replace and secure the rear cover, battery and battery cover.
WARNING: To avoid electric shock, do not operate your meter until the fuse cover is in place and fastened securely.
24
TROUBLESHOOTING
There may be times when your meter does not operate properly.
Here are some common problems that you may have and some easy solutions to them.
Meter Does Not Operate:
1. Always read all the instructions in this manual before use.
2. Check to be sure the battery is properly installed.
3. Check to be sure the battery is good.
4. If the battery is good and the meter still doesn’t operate, check to be sure that both ends of the fuse are properly installed.
If You Do Not Understand How the Meter Works:
Call our Customer Service Line 1-888-326-1006.
SERVICE AND PARTS
Item Number
82378
82003-D
82003-C
82003-CS
Description
Set of black and red Test Leads
Replacement battery cover cover
Rear cover screws
For replacement parts shipped directly to your home
Call 9 am – 5 pm Eastern Time, M - F
1-888-326-1006
25
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project
Key Features
- True RMS measurements for accurate readings of non-sinusoidal waveforms
- 4,000 count LCD display provides clear and precise readings
- Autoranging function for automatic selection of the appropriate measurement range
- Relative mode for zeroing out lead resistance
- Backlight function for easy reading in low-light conditions
- Low battery indication alerts the user when the battery needs to be replaced
- Fuse replacement without opening the case
- Tilt stand and magnetic hanger for hands-free operation
Related manuals
Frequently Answers and Questions
What is the maximum AC voltage that the Craftsman 82003 can measure?
Can the Craftsman 82003 measure temperature?
What is the accuracy of the resistance measurement function?
Does the Craftsman 82003 have a backlight?
What type of battery does the Craftsman 82003 use?
What is the warranty period for the Craftsman 82003?
advertisement