Agilent Technologies 6628A Operating Manual

Add to my manuals
123 Pages

advertisement

Agilent Technologies 6628A Operating Manual | Manualzz

OPERATING MANUAL

MULTIPLE OUTPUT LINEAR SYSTEM

DC POWER SUPPLY

AGILENT MODELS 6625A, 6626A,

6628A, and 6629A

Agilent Part No 06626-90001

Microfiche Part No. 06626-90002 Printed in Malaysia

September, 2004

CERTIFICATION

Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent

Technologies further certifies that its calibration measurements are traceable to the United States National Bureau of

Standards, to the extent allowed by the Bureau's calibration facility, and to the calibration facilities of other International

Standards Organization members.

WARRANTY

This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of one year from date of delivery. Agilent software and firmware products, which are designated by Agilent for use with a hardware product and when properly installed on that hardware product, are warranted not to fail to execute their programming instructions due to defects in material and workmanship for a period of 90 days from date of delivery. During the warranty period Agilent Technologies will, at its option, either repair or replace products which prove to be defective.

Agilent does not warrant that the operation of the software, firmware, or hardware shall be uninterrupted or error free.

For warranty service, with the exception of warranty options, this product must be returned to a service facility designated by Agilent. Customer shall prepay shipping charges by (and shall pay all duty and taxes) for products returned to Agilent for warranty service. Except for products returned to Customer from another country, Agilent shall pay for return of products to Customer.

Warranty services outside the country of initial purchase are included in Agilent's product price, only if Customer pays

Agilent international prices (defined as destination local currency price, or U.S. or Geneva Export price).

If Agilent is unable, within a reasonable time to repair or replace any product to condition as warranted, the Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent.

LIMITATION OF WARRANTY

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer,

Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation and maintenance. NO OTHER WARRANTY IS EXPRESSED

OR IMPLIED. Agilent SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND

FITNESS FOR A PARTICULAR PURPOSE.

EXCLUSIVE REMEDIES

THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER'S SOLE AND EXCLUSIVE REMEDIES. Agilent

SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL

DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.

ASSISTANCE

The above statements apply only to the standard product warranty. Warranty options, extended support contracts, product maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent

Technologies Sales and Service office for further information on Agilent's full line of Support Programs.

 Copyright 2000, 2004 Agilent Technologies

Edition 1___January, 1993

2

SAFETY SUMMARY

The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.

BEFORE APPLYING POWER.

Verify that the product is set to match the available line voltage and the correct fuse is installed.

GROUND THE INSTRUMENT.

This product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The instrument must be connected to the ac power supply mains through a threeconductor power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. For instruments designed to be hard-wired to the ac power lines (supply mains), connect the protective earth terminal to a protective conductor before any other connection is made. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury. If the instrument is to be energized via an external autotransformer for voltage reduction, be certain that the autotransformer common terminal is connected to the neutral (earthed pole) of the ac power lines

(supply mains).

FUSES.

Only fuses with the required rated current, voltage, and specified type (normal blow, time delay, etc.) should be used. Do not use repaired fuses or short circuited fuseholders. To do so could cause a shock or fire hazard.

DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE.

Do not operate the instrument in the presence of flammable gases or fumes.

KEEP AWAY FROM LIVE CIRCUITS.

Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified service personnel. Do not replace components with power cable connected. Under certain conditions, dangerous voltages may exist even with the power cable removed. To avoid injuries, always disconnect power, discharge circuits and remove external voltage sources before touching components.

DO NOT SERVICE OR ADJUST ALONE.

Do not attempt internal service or adjustment unless another person, capable of rendering first aid and resuscitation, is present.

DO NOT EXCEED INPUT RATINGS.

This instrument may be equipped with a line filter to reduce electromagnetic interference and must be connected to a properly grounded receptacle to minimize electric shock hazard. Operation at line voltages or frequencies in excess of those stated on the data plate may cause leakage currents in excess of 5.0 mA peak.

SAFETY SYMBOLS.

Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the instruction manual (refer to Table of Contents) .

Indicates hazardous voltages.

Indicate earth (ground) terminal.

The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.

The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and met.

DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT.

Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to an Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained.

Instruments which appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel

3

4

SAFETY SUMMARY (continued)

GENERAL

Any LEDs used in this product are Class 1 LEDs as per IEC 825-1.

This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 du Canada.

ENVIRONMENTAL CONDITIONS

This instrument is intended for indoor use in an installation category II, pollution degree 2 environment. It is designed to operate at a maximum relative humidity of 95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the ac mains voltage requirements and ambient operating temperature range.

Symbol

Direct current

Description

SAFETY SYMBOL DEFINITIONS

Symbol Description

Terminal for Line conductor on permanently installed equipment

Alternating current Caution, risk of electric shock

Both direct and alternating current Caution, hot surface

Three-phase alternating current

Earth (ground) terminal

Caution (refer to accompanying documents)

In position of a bi-stable push control

Protective earth (ground) terminal

Frame or chassis terminal

Out position of a bi-stable push control

On (supply)

Terminal for Neutral conductor on permanently installed equipment

Terminal is at earth potential

(Used for measurement and control circuits designed to be operated with one terminal at earth potential.)

Off (supply)

Standby (supply)

Units with this symbol are not completely disconnected from ac mains when this switch is off. To completely disconnect the unit from ac mains, either disconnect the power cord or have a qualified electrician install an external switch.

Herstellerbescheinigung

Diese Information steht im Zusammenhang mit den Anforderungen der

Maschinenl ä minformationsverordnung vom 18 Januar 1991.

* Schalldruckpegel Lp <70 dB(A) * Am Arbeitsplatz * Normaler Betrieb

* Nach EN 27779 (Typpr ü fung).

Manufacturer’s Declaration

This statement is provided to comply with the requirements of the German Sound Emission Directive,

from 18 January 1991.

* Sound Pressure Lp <70 dB(A) * At Operator Position * Normal Operation

* According to EN 27779 (Type Test).

Manufacturer’s Name and Address

DECLARATION OF CONFORMITY

According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014

Agilent Technologies, Inc.

550 Clark Drive, Suite 101

Budd Lake, New Jersey 07828

Agilent Technologies (Malaysia) Sdn. Bhd

Malaysia Manufacturing

Bayan Lepas Free Industrial Zone, PH III

USA 11900 Penang,

Malaysia

Declares under sole responsibility that the product as originally delivered

Product Names a) Multiple Output 40 W and 80 W system dc Power Supplies b) Multiple Output 25 W and 50 W precision system dc Power Supplies

Model Numbers a) 6621A, 6622A, 6623A; 6624A, 6627A

Product Options

b) 6625A; 6626A; 6628A; 6629A

(and other customized products based upon the above)

This declaration covers all options and customized products based on the above products.

Complies with the essential requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive

89/336/EEC (including 93/68/EEC) and carries the CE Marking accordingly.

EMC Information

As detailed in

ISM Group 1 Class A Emissions

Electromagnetic Compatibility (EMC), Certificate of Conformance Number

CC/TCF/00/076 based on Technical Construction File (TCF) HPNJ3, dated Oct. 29, 1997

Assessed by: Celestica Ltd, Appointed Competent Body

Westfields House, West Avenue

Kidsgrove, Stoke-on-Trent

Straffordshire, ST7 1TL

United Kingdom

Safety Information and Conforms to the following safety standards.

IEC 61010-1:2001 / EN 61010-1:2001

UL 1244

CSA C22.2 No. 1010.1:1992

This DoC applies to the above-listed products placed on the EU market after:

January 1, 2004

Date Bill Darcy/ Regulations Manager

For further information, please contact your local Agilent Technologies sales office, agent or distributor.

Authorized EU-representative: Agilent Technologies Deutschland GmbH, Herrenberger Stra

βe 130, D

71034 Böblingen, Germany

Revision: B.00.00 Issue Date: Created on 11/24/2003 3:33

PM

Document No. 662xA.11.24.doc

To obtain the latest Declaration of Conformity, go to http://regulations.corporate.agilent.com

and click on “Declarations of Conformity.”

5

WHAT THIS MANUAL CONTAINS

It contains information relating to the installation, operation, and programming of these supplies as outlined below.

Maintenance and troubleshooting instructions are given in a separate Service Manual (Agilent Part No. 06626-90003).

Chapter 1.--General Information

Chapter 1 contains a general description of the power supplies as well as instrument specifications and information concerning options and accessories.

Chapter 2.--Installation Procedures

Chapter 2 contains information to prepare the supply for use. Included in this chapter are power requirements, line voltage conversion, and GP-IB interface connections.

Chapter 3.--Getting Started

Chapter 3 contains a brief description of the supply’s front panel controls and indicators and describes how to turn on the supply and to check it’s operation. An introduction to remote operation over the GP-IB is also given to help a first time user get started quickly.

Chapter 4.--Output Connections and Operating Information

Chapter 4 contains information about making connections to the supply’s output terminals. General operating information is also provided.

Chapter 5.--Remote Operation

Chapter 5 contains all of the information required to operate the supply remotely via a GP-IB computer. All of the commands that can be used to program the supplies are described.

Chapter 6.--Local Operation

Chapter 6 contains instructions on using all of the front panel controls and indicators.

Appendix A--Calibration Procedure

Appendix A contains programming steps and procedures that are required to calibrate your power supply. It is recommended that the power supply be calibrated yearly.

Appendix B--Programming with Series 200/300 Computer

Appendix B contains Series 200/300 Computer programming examples (in Agilent extended BASIC language) for your

Power Supply’s most frequently used functions.

Appendix C--Command Summary

Appendix C contains an alphabetical listing of all commands that can be sent to a supply.

Appendix D--Error Messages

Appendix D contains a listing and brief explanation of all error codes and messages for all programming and hardware errors.

Appendix E - Manual Backdating

Appendix E contains backdating information for units with Serial numbers lower than those listed on the title page.

6

4

1

Table Of Contents

General Information

Introduction............................ .......................................................................................................................11

Safety Considerations....................................................................................................................................11

Instrument and Manual Identification.... .......................................................................................................11

Options................................. .........................................................................................................................11

Accessories....................................................................................................................................................12

Description ............................ .......................................................................................................................12

Basic Operation......................... ....................................................................................................................14

GP-IB Board..............................................................................................................................................14

Output Boards............................................................................................................................................15

Specifications......................... .......................................................................................................................15

Qualifying Conditions.................. .............................................................................................................15

Definitions............................ .....................................................................................................................15

2 Installation

Introduction............................ .......................................................................................................................25

Initial Inspection............................................................................................................................................25

Location and Cooling.................... ................................................................................................................25

Input Power Requirements................ ............................................................................................................26

Line Fuse............................... ........................................................................................................................27

Power Cord....................................................................................................................................................28

Line Voltage Conversion...............................................................................................................................29

GP-IB Interface Connection.............. ............................................................................................................29

Introduction ........................... .......................................................................................................................31

Front Panel Controls and Indicators..... .........................................................................................................31

Turning On Your Supply...............................................................................................................................36

Normal Self Test Indications.....................................................................................................................36

Self Test Errors..........................................................................................................................................37

Checking Out Your Supply Using Local Control..........................................................................................37

Voltage Test............................ ..................................................................................................................38

Overvoltage Test......................... ..............................................................................................................38

Current Test...............................................................................................................................................38

Changing Resolution Range ......................................................................................................................39

Introduction To Remote Operation....... ........................................................................................................39

Enter/Output Statements............................................................................................................................40

Reading the GP-IB Address.............. ........................................................................................................40

Changing the GP-IB Address............. .......................................................................................................40

Sending a Remote Command............... .....................................................................................................41

Getting Data from the Supply....................................................................................................................41

Often Used Commands..............................................................................................................................41

Returning the Supply to Local Mode..... ...................................................................................................43

Output Connections and Operating Information

Introduction ............................... ...................................................................................................................45

Output Ranges............................... ................................................................................................................45

Operating Quadrants..................................................................................................................................46

Operating Range........................................................................................................................................47

Protection Features......................... ...............................................................................................................48

Connecting The Load......................... ...........................................................................................................49

Wire Size Selection......................... ..........................................................................................................49

Multiple Loads.............................. ............................................................................................................50

7

8

5

6

Table Of Contents (continued)

Positive and Negative Voltages.................................................................................................................51

Remote Voltage Sensing..................... ..........................................................................................................51

Remote Sense Connections................... ....................................................................................................52

Output Noise Considerations.....................................................................................................................53

Programming Response Time with an Output Capacitor............................. .............................................53

Open Sense Leads......................................................................................................................................54

Overvoltage Trigger Connections............ .....................................................................................................54

External Trigger Circuit................... .........................................................................................................54

Power Supply Protection Considerations .......................... ...........................................................................55

Battery Charging........................... ............................................................................................................55

Capacitive Load Limitation.......................................................................................................................56

Parallel Operation..........................................................................................................................................56

CV Operation............................... .............................................................................................................57

CC Operation.............................................................................................................................................58

Remote Sensing........................... ..............................................................................................................58

Specifications for Parallel Operation.........................................................................................................58

Series Operation....................... .....................................................................................................................59

CV Operation............................... .............................................................................................................59

CC Operation.............................................................................................................................................60

Remote Sensing........................... ..............................................................................................................60

Specifications for Series Operation. ..........................................................................................................60

Bi-Polar Operation ....................................................................................................................................61

Remote Operation

Introduction ............................. .....................................................................................................................63

GP-IB Operation............................................................................................................................................63

Interface Functions......................... ...........................................................................................................63

GP-IB Address Selection...........................................................................................................................64

Power-On Service Request (PON) ............................................................................................................65

Programming Syntax......................... ............................................................................................................65

Numeric Data............................... .............................................................................................................67

Order of Execution......................... ...........................................................................................................70

Terminators................................ ...............................................................................................................70

Initial Conditions...........................................................................................................................................71

Power Supply Commands .............................................................................................................................71

Voltage Programming........................ .......................................................................................................71

Current Programming.............................. ..................................................................................................73

Range Programming.................................. ................................................................................................73

Output On/Off.................................... .......................................................................................................74

Overvoltage Protection........................... ...................................................................................................74

Overcurrent Protection........................... ...................................................................................................75

Multiple Output Storage and Recall............... ...........................................................................................75

The Clear Command..................................................................................................................................76

Status Reporting................................. .......................................................................................................76

Service Request Generation....................... ...............................................................................................79

Reprogramming Delay............................. .................................................................................................80

Display On/Off................................... .......................................................................................................81

Other Queries.............................................................................................................................................82

Local Operation

Introduction ..................................... .............................................................................................................85

Local Mode....................................................................................................................................................85

A

B

C

D

E

Table Of Contents (continued)

Local Control Of Output Functions...............................................................................................................85

General........................................... ...........................................................................................................85

Setting Voltage and Voltage Range...........................................................................................................86

Setting Current and Current Range............................................................................................................87

Enabling/Disabling an Output...................... .............................................................................................87

Setting Overvoltage Protection..................................................................................................................88

Resetting Overvoltage Protection..............................................................................................................88

Enabling/Disabling Overcurrent Protection........ ......................................................................................88

Resetting Overcurrent Protection................. .............................................................................................88

Local Control Of System Functions................ ..............................................................................................88

Setting the Supply’s GP-IB Address..........................................................................................................88

Displaying Error Messages........................................................................................................................89

Storing and Recalling Voltage and Current. Settings for all Outputs........................................................89

Alignment Procedures

Introduction...................................... ............................................................................. ................................91

Test Equipment and Setup Required................. ............................................................................. ...............91

Calibration Program.............................. .............................................................................. ..........................91

Alignment Program.............................. ................................................................................ .........................95

Programming With a Series 200/300 Computer

Introduction ....................... .........................................................................................................................103

I/O Path Names...................... .....................................................................................................................103

Voltage and Current Programming..... ........................................................................................................103

Voltage and Current Programming With Variables... .................................................................................104

Voltage and Current Readback....................................................................................................................104

Programming Power Supply Registers........................................................................................................105

Present Status...................... ....................................................................................................................105

Service Request and Serial Poll...............................................................................................................106

Error Detection .................... ...................................................................................................................107

Stored Operating States........... ................................................................................................................108

Command Summary

Introduction .................................................................................................................................................109

Error Messages

Introduction....................... ..........................................................................................................................117

Power-On Self Test Messages.....................................................................................................................117

Error Responses...........................................................................................................................................117

Test Responses...................... ......................................................................................................................117

Manual Backdating

Introduction ........................ ........................................................................................................................121

Make Changes.................…………………………………………………………………………………...121

Agilent Sales and Support Office

Contacts ...................................................................................................................... ................................122

9

1

General Information

Introduction

This chapter contains a general description of your power supply as well as its performance specifications. Information about options, accessories, and GP-IB cables is also provided. This manual describes the Agilent 6625A, 6626A, 6628A, and Agilent 6629A power supplies. Unless stated otherwise, the information in this manual applies to all of these models.

Information that is specific to one model only is identified as such in this manual.

Safety Considerations

This product is a Safety Class 1 instrument, which means that it is provided with a protective earth terminal. This terminal must be connected to a power source that has a 3-wire ground receptacle. Review the instrument and this manual for safety markings and instructions before operation. Refer to the Safety Summary page at the beginning of this manual for a summary of general safety information. Safety information for specific procedures is located at appropriate places in this manual.

Instrument and Manual Identification

Agilent Technologies power supplies are identified by a two-part serial number, i.e. 2601A-00101. The first part of the serial number (the prefix) is a number/letter combination that denotes either the date of manufacture or the date of a significant design change. It also indicates the country of origin. (Starting at 1960, 26 = 1986; 01 = the first week of the year; A = U.S.A.) The second part of the serial number is a different sequential number assigned to each instrument starting with 00101.

Options

Options 100,120, 220, and 240 simply determine which line voltage is selected at the factory. For information about changing the line voltage setting, see line voltage conversion paragraph on page 29.

Option 750 consists of a fault indicator (FLT) and remote inhibit (INH) circuit and relay control, which provide additional shutdown protection should either the GP-IB and/or controller fail. This Option is described in a separate document entitled, ’’Appendix E Option 750 Operating Instructions for the Multiple Output Linear System DC Power Supply, Agilent

Models 6621A, 6622A, 6623A, 6624A, and 6627A’’ (Agilent P/N 5957-6372).

#100 Input power, 100 Vac, 47-66 Hz

#120 Input power, 120 Vac, 47-66 Hz

#220 Input power, 220 Vac, 47-66 Hz

#240 Input power, 240 Vac, 47-66 Hz

#750 Fault (FLT) Remote Inhibit (INH) and Relay Control

#908 One rack mount kit (5062-3977)

#909 One rack mount kit with handles (5062-3983)

#0L2 One extra operating manual

#0B3 One service manual

General Information 11

Accessories

10833A GP-IB cable, 1 m (3.3 ft)

10833B GP-IB cable, 2 m (6.6 ft)

10833C GP-IB cable, 4 m (13.2 ft)

10833D GP-IB cable, 0.5 m (1.6 ft)

10834A GP-IB connector extender

Slide mount kit (1494-0059)

Description

The Agilent power supply features a combination of programming capabilities and linear power supply performance that make them ideal for systems applications. The models in this family offer a total of up to 200 watts of output power, with voltages up to 50 volts and currents up to 2 amps. The output combinations that correspond to each model are shown in

Table 1-1. Each isolated output has a high resolution and standard resolution range, as shown in Figure 1-1.

Model

Agilent 6625A

Agilent 6626A

Agilent 6628A

Agilent 6629A

Table 1-1 Output Combinations Available

Output 1 Output 2 Output 3

25 W

25 W

50 W

50 W

25 W

50 W

--

50 W

--

50 W 50 W 50 W

Output 4

--

50 W

--

50W

Figure 1-1. Output Operating Range

Output 2 of the Agilent 6625A and outputs 3 and 4 of the Agilent 6626A and all outputs of the Agilent 6628A and Agilent

6629A have an "L’’ shaped characteristic. The L shaped characteristic allows you to use the same output to power loads with different voltage and current requirements. The power supply automatically selects one of the operating ranges based on the last parameter (voltage or current) that is set. If a high resolution or a standard resolution application is required for either voltage or current, VRSET and IRSET should be used (see Appendix C-Command Summary). Additionally, each output contains an active downprogrammer, which means that voltage downprogramming can be accomplished as quickly as upprogramming, even without a load.

12 General Information

The power supplies allow up to a 10 volt sense lead drop. This feature makes them ideal for test system applications where remote sensing is used.

The output voltage and current for any output can be monitored with the front panel display. Output specific error messages are also displayed. Front panel annunciators show the operating status of the instrument. The front panel keypad lets you set and readback the voltage limit, current limit, and overvoltage trip level of any output. With the keypad, you can also enable or disable outputs, enable overcurrent protection, reset overvoltage and overcurrent protection, and return to local operating mode.

Your multiple output power supply can be both a listener and a talker on the GP-IB. (GP-IB is Agilent Technologies’ implementation of IEEE-488). The built-in interface is tailored to the supply, resulting in simpler programming. Voltage and current settings can be sent directly to the specified output in volts and amps.

Service can be requested from your power supply for up to 10 reasons. The supply responds to a serial poll by identifying the output on which the fault occurred. Self-contained measurement and readback capability eliminate the need for externally scanning the outputs using a separate DVM. Upon command, the supply will measure its output voltage or current and return the value on the GP-IB. The following functions are implemented via the GP-IB:

Voltage and current programming.

Programmable resolution range.

Voltage and current measurement and readback.

Present and accumulated status readback.

Programmable service request mask.

Programmable overvoltage and overcurrent protection.

Voltage and current range programming.

Storage and recall of programmed voltage and current values for all outputs.

Queries of programmed functions or settings.

Output enable or disable.

Programming syntax error detection.

Programmable delay time for service request and OCP mask.

Voltage, current, and overvoltage calibration.

GP-IB interface selftest.

Message display capability on the front panel.

Output connections are made to rear panel screw terminals. Either the positive or negative output terminal can be grounded, or the output can be floated up to

±

240 Vdc (including output voltage) from chassis ground. Output voltage can be locally or remotely sensed, and identical outputs can be operated in series or parallel combinations for increased output voltage or current capability. As shipped from the factory, the power supply is jumpered for local sensing.

Your power supply can be calibrated without having to remove the cover or even having to remove it from your system cabinet. This feature allows you to calibrate the supply at its normal operating temperature. The recommended calibration interval is one year. Refer to Appendix A of this manual for complete calibration details. A calibration security jumper is available inside the unit. Access is described in the service manual.

General Information 13

Basic Operation

Figure 1-2 is a block diagram that illustrates the major assemblies contained within the power supply. As shown in the figure, each supply includes a power transformer, two or more output boards, a GP-IB board, and front panel (display and control keys).

Figure 1-2. Block Diagram

The appropriate ac input voltage is applied to each output board where it is converted to a raw dc voltage which is subsequently linearly regulated to become the dc output voltage. The magnitude of the output and the mode of operation are determined by the load and the data received from the GP-IB computer or from the front panel.

Each power supply model contains one output board for each output that it provides.

GP-IB Board

The GP-IB board provides the interface between the user and the multiple outputs of the power supply. Each output board is actually an output channel that can be individually selected and controlled over the GP-IB or from the supply’s front panel. Circuits on the GP-IB board interpret commands from the GP-IB or from the front panel to control the selected output.

The GP-IB board also processes measurement and status data received from the output boards. This data may be read back over the GP-IB and/or displayed on the supply’s front panel.

The power supply has no potentiometers. Each output is individually calibrated over the GP-IB using calibration commands

(see Appendix A). Correction factors are calculated by the power supply during calibration and are stored in a non-volatile memory which is located on the supply’s GP-IB board. The supply contains no batteries.

14 General Information

Output Boards

The output boards are linear dc power supplies. Each isolated output operating boundary curve is shown in Figure 1-1.

The ac input to each output board is rectified and applied to a regulator circuit. Each output board employs series regulation techniques. A regulator element is connected in series with the load and operates in the linear region (between saturation and cutoff) of the transistor characteristic curve. Regulation is achieved by varying the conduction of the series element in response to a change in line voltage or circuit load.

The output board receives digital signals from the GP-IB board and converts them to analog signals which program the output voltage, current, and overvoltage values. The output may be programmed remotely over the GP-IB using commands

(see Chapter 5) or locally from the supply’s front panel using the control keys (see Chapter 6).

The output board can be commanded to send measurement and status data back over the GP-IB and/or front panel. The data is sent back via the supply’s GP-IB board. GP-IB readback capabilities include output voltage and current, present and accumulated status, and all programmed settings. The front panel LCD display can indicate the output voltage and current, the supply’s GP-IB address, error messages, and programmed values. Annunciators on the front panel indicate the operating status of the selected output (output board).

Specifications

Table l-2 lists the performance specifications for the Agilent 662xA power supplies. Performance specifications describe the instrument’s warranted performance. The service manual, Option 9l0, contains procedures for verifying the performance specifications.

Table 1-3 lists the supplemental characteristics for the Agilent 662xA supplies. Supplemental characteristics are type-tested or typical values, which are based on a product sample and, while representative, are not guaranteed.

Qualifying Conditions

All performance specifications apply over the full operating temperature range of the power supply (0 to 55°C) unless otherwise specified. All regulation, accuracy, etc. specifications are plus or minus the values listed. All measurements are made at the rear terminals of the supply with a resistive load and local sensing unless otherwise specified. Voltage measurements are made from the + S to the - S terminals. Overvoltage measurements are made from the + V to the - V terminals. + Current refers to the output acting as a current source while - Current refers to the output acting as a current sink.

Definitions

Load effect : Maximum steady state change in the regulated output parameter due to a change in load resistance on the output in question.

Source effect : Maximum steady state change in the regulated output parameter due to a change in the source voltage within rated values.

Cross regulation : Maximum steady state change in the regulated output parameter due to a change in load resistance on any other output(s).

Programming accuracy : (Calibration temp

±

5°C) Maximum difference between the programmed value and the actual output.

Readback accuracy : (Calibration temp

±

5°C) Maximum error in reading back an output parameter. Expressed as a percentage of the reading, plus a constant.

General Information 15

Output response time : Beginning at the time the power supply has finished processing a VSET command (change output voltage), the maximum time for the output voltage to settle to within a settling band about the final value from any specified operating point. This value must be added to the command processing time to obtain total programming time (see

Figure 1-3). Time constant is the maximum time required for the voltage to reach 63% of its final value.

Figure 1-3. Output Response Characteristics

Temperature coefficient : Maximum change in the regulated output parameter per °C change in ambient temperature after a 30 minute warmup. Expressed as a percentage plus a constant per degrees C. (Plus a constant for readback temperature coefficient).

Long Term Drift : Maximum change of regulated output voltage or current during an 8-hour period following a 30 minute warmup, with all influence and control quantities maintained constant. Expressed as a percentage of the setting plus a constant.

Short Term Drift : Maximum change of regulated output voltage or current within 30 minutes after a line and/or load change. Expressed as a percentage of the setting plus a constant.

Output Noise (PARD): PARD replaces the former term ripple and noise. PARD is the periodic and random deviation of dc output voltage or current from its average value, over a specified bandwidth and with all influence and control quantities maintained constant.

Programming resolution : Average programming step size.

Current Sinking (- Current): Each output can sink as well as source current. The sinking capability is not programmable and depends upon the output voltage. The current sinking capability is described in greater detail in

Chapter 4.

Table 1-2. Specifications

PERFORMANCE SPECIFICATIONS : (0 to 55 °C unless otherwise specified)

DC Output Ranges : All outputs will accept voltage programming commands 1% higher than those listed and current programming commands 3% higher than those listed.

Output Power

Output Range

Output Volts

Output Amps

Lo Range

0-7 V

0-15 mA

Load Effect (Regulation):

Voltage 0.5 mV

25 Watt Output

Hi Range

0-50 V

0-500 mA

0.5 mV

Lo Range

0-16 V

0-200 mA

0.5 mV

50 Watt Output

Hi Range

0-50 V

0-2 A

0.5 mV

+ Current 0.005 mA 0.005 mA 0.01 mA 0.015 mA

16 General Information

Table 1-2. Specifications (continued)

Source Effect:

Voltage

+ Current

0.5 mV

0.005 mA

Programming Accuracy:

Voltage

+ Current

OVP

0.5 mV

0.005 mA

0.5 mV

0.01 mA

0.5 mV

0.01 mA

0.016% + 1.5 mV 0.016% + 10 mV

0.04% + 15

μA 0.04%

0.016% + 3 mV

185

0.016% + 10 mV

μA 0.04%

0.13% + 475 mV 0.13% + 475 mV 0.13% + 475 mV 0.13% + 475 mV

Measurement/Readback Accuracy*

Voltage

± Current

0.016% + 2 mV 0.016% + 10 mV 0.016% + 3.5 mV 0.016% + 10 mV

0.03% + 15

μA 0.03%

+

μA 0.04%

Load Transient Recovery Time :(all outputs)

75

μS maximum to recover to within 75 mV of nominal value following a load change from 0.1 A to 100% of the maximum rated current

Maximum Output Noise: (PARD) (20 Hz-20 MHz)

CV pk-to-pk 3 mV 3 mV

CV rms

+CC rms

0.5 mV

0.1 mA

AC Input Voltage and Frequency:

3 mV

0.5 mV

0.1 mA

3 mV

0.5 mV

0.1 mA

0.5 mV

0.1 mA

Nominal Line l00, 120, 220, or 240 Vac

Amplitude +6% to -13% of nominal line voltage

Frequency 47 to 66 Hz

Note at low line, the supply will operate with up to 3/4 ohm line resistance

*For a

± 5 ° C range about the calibration temperature

Table 1-3. Supplemental Characteristics

25 Watt Output 50 Watt Output Output Power

Output Range

Output Volts

Output Amps

Lo Range

0-7 V

0-15 mA

Temperature Coefficient-Programming:

Hi Range

0-50 V

0-500 mA

Lo Range

0-16 V

0-200 mA

Hi Range

0-50 V

0-2 A

+ Current

OVP per° C

(0.0035% + 1.5 per° C

(0.013% + 1 mV) per° C

μA) per° C

(0.0035% + 6 per° C

(0.013% + 2 mV) per° C

μA)

(0.003%+0.1 mV) per° C

(0.0035% + 30

μA) per° C

(0.013% + 1 mV) per° C

(0.003%+0.1 mV) per° C

(0.0035% + 50

μA) per° C

(0.013% + 2 mV) per° C

General Information 17

Voltage

+ Current

Table 1-3. Supplemental Characteristics (continued)

Output Power

Output Range

Output Volts

Output Amps

Lo Range

0-7 V

0-15 mA

25 Watt Output

Temperature Coefficient-Measurement:

Hi Range

0-50 V

0-500 mA

Lo Range

0-16 V

0-200 mA

50 Watt Output

Voltage

±

Current

(0.002% + 0.1 mV) per° C +0.5 mV

(0.0025% + 1.5

µ

A) per° C + 1

µ

A

(0.002% + 0.1 mV) per° C +4 mV

(0.0025% + 10

µ

A) per° C + 40

µ

A

(0.002% + 0.1 mV) per° C +1 mV

(0.0025% + 20

µ

A) per° C + 15

µ

A

Long Term Drift : (In an 8 hour period following a 30 minute warm up)

Voltage

+ Current

0.006% + 0.5 mV

0.01% + 5

µ

A

0.006% + 2 mV

0.01% + 20

µ

A

0.006% + 0.5 mV

0.01% + 40

µ

A

Short Term Drift: (Within 30 minutes after a line and/or load change):

Hi Range

0-50 V

0-2 A

(0.002% + 0.1 mV) per° C +4 mV

(0.0025% + 50

µ

A) per° C + 150

µ

A

0.006% + 2 mV

0.01% + 60

µ

A

0.002 + 0.5 mV

0.01% + 2

µ

A

0.002 + 1 mV

0.01% + 20

µ

A

0.002 + 0.5 mV

0.01% + 25

µ

A

0.002 + 1 mV

0.01% + 60

µ

A

DC Floating Voltage:

No output terminal may be more that 240 Vdc from any other terminal or from chassis ground. Also, no overvoltage terminal may be more than 240 Vdc from any other terminal or chassis ground.

Remote Sense Capability: (see pages 50 - 52):

Outputs can maintain specifications with up to 10 volt total in the remote sense leads.

AC Input Power : 550 Watts

Programming Resolution:

Voltage

+ Current

OVP

460

µ

V

1

µ

A

230 mV

3.2 mV

33

µ

A

230 mV

1 mV

13

µ

A

230 mV

Readback Resolution:

Voltage

+ Current

- Current

483

µ

V

1

µ

A

1

µ

A

3.3 mV

48

µ

A

37

µ

A

Fixed Overvoltage Protection: (Measured at output terminals +V and -V)

1.1 mV

14

µ

A

14

µ

A

Minimum

Nominal

Maximum

56 V

60 V

64 V

56 V

60 V

64 V

56 V

60 V

64 V

3.2 mV

131

µ

56 V

60 V

64 V

A

230 mV

3.3 mV

160

µ

A

151

µ

A

18 General Information

Table 1-3. Supplemental Characteristics (continued)

High Line Inrush Current:

Peak Value rms Value

Fuse Rating

100 V Opt

85 A

6.3A

8 A

120 V Opt

85 A

5.7 A

8 A

220 V Opt

50 A

3.0 A

4 A

GP-IB Interface Capabilities:

SH1. AH1, T6, L4, SR1, RL1, PP1, DC1, DT0, C0, E1

240 V Opt

50 A

3.0 A

4 A

Current Sink Capability:

25 Watt output: 0.50 A

50 Watt output: 1.0 A (2.0 A below 16 V)

Command Processing Time: (see Figure 1-3)

7 milliseconds typical (with front panel disabled). Using STD and RCL commands allows you to change all the voltage and current settings in about 10 mS (with front panel disabled).

Maximum Output Programming:

Response Time

Settling Band

6 mS

50 mV

Maximum Time

Constant 750

µ

S

6 mS

50 mV

6 mS

50 mV

6 mS

50 mV

750

µ

S 750

µ

S 750

µ

S

Series and Parallel operation:

Two outputs can be operated directly in parallel or can be connected for straight series operation. Refer to Chapter 4 for more information.

Reactive Load Capability:

All outputs have been designed with the ability to operate with significant reactive loads without instability (refer to

Figures 1-4 through 1-6)

Safety Agency Compliance:

This series of power supplies is designed to comply with the following standards:

IEC 348, UL 1244, and CSA 22.2 No. 231.

Dimensions (all models):

Height

Width

Depth

132.6mm (5.22 in)

425.5mm (16.75 in.)

497.8mm (19.6 in.)

General Information 19

Table 1-3. Supplemental Characteristics (continued)

Weight:

Net

Shipping

Agilent 6625A, 6628A

15.5 kg (34 lbs.)

20.8 kg (46 lbs.)

Load Cross Regulation:

Voltage

+ Current

0.25 mV

0.005 mA

0.25 mV

0.005 mA

Agilent 6626A, 6629A

17.7 kg (39 lbs.)

23 kg (51 lbs.)

0.25 mV

0.01 mA

0.25 mV

0.01 mA

20 General Information

Figure 1-4. CV Operation with Capacitive Load, Stability Graph for all Outputs

General Information 21

Figure 1-5. Output Noise (Typical) CC Mode

22 General Information

Figure 1-6. Output Noise (Typical) CV Mode

Figure 1-7. Output Impedance (Typical) Graph (See Supplemental Characteristics)

General Information 23

2

Installation

Introduction

This chapter contains instructions for checking and mounting your power supply, connecting your supply to ac power, converting it from one line voltage to another, and connecting the GP-IB cable.

The power supply generates operating magnetic fields which may affect the operation of other instruments. If your instrument is susceptible to magnetic fields, do not locate it in the immediate vicinity of the power supply. Typically, at three inches from the power supply, the electromagnetic field is less than 5 gauss.

Initial Inspection

Your instrument was thoroughly inspected and tested before it left the factory. As soon as you receive it, remove the power supply from its packing case and check to make sure it has not been damaged in shipment. Check that there are no broken connectors or keys, and that the cabinet and panel surfaces are free from dents and scratches. Check the rear panel terminal blocks and front panel display for any cracks. If damage is found, you should file a claim with the carrier immediately and notify the Agilent Technologies Sales and Service office nearest you.

Chapter 3 of this manual includes an electrical turn-on check-out procedure which, when carried out successfully, will give you a high level of confidence that the power supply is operating in accordance with its specifications. Detailed electrical checks complete with verification procedures are included in the Service Manual.

Keep the original packing materials for the carrier’s inspection if there was damage, or in case any equipment has to be returned to Agilent Technologies. Warranty information is printed on the inside cover of this manual. Remember to send a detailed description of the failure and symptoms when returning the power supply for service. Your Agilent Technologies

Sales and Service office will furnish the address of the nearest service office to which the instrument can be shipped.

Location and Cooling

Your power supply can operate without loss of performance within the temperature range of 0 to 55 ° C (measured at the fan intake). The fan, located at the rear of the unit, cools the supply by drawing air in through the openings on the rear panel and exhausting it through openings on the sides. Using Agilent Technologies rack mount kits will not impede the flow of air.

Because the power supply is fan cooled, it must be installed in a location that allows sufficient space at the rear and the sides for adequate circulation of air. Either side may be restricted to have as little as 1 inch (25 mm) space.

Figure 2-1 gives the dimensions of the power supply cabinet. These dimensions apply to both models. The cabinet has plastic feet that are shaped to ensure self-alignment when stacked with other Agilent Technologies System II cabinets. The feet may be removed for rack mounting.

The power supply can be mounted in a standard 19 inch rack panel or enclosure. Rack mounting accessories for this unit are listed under options (page 11) of Chapter 1. Complete installation instructions are included with each rack mounting kit.

Instrument support rails are required for non-stationary installations. These are normally supplied with the cabinet and are

not included with the rack mounting kits.

Installation 25

Figure 2-1. Outline Diagram

Input Power Requirements

You can operate this power supply from a nominal 100 V, 120 V, 220 V or 240 V single phase power source at 47 to 66

Hz. The input voltage range, maximum input current, high line inrush current (PK), and the fuse required for each of the nominal inputs are listed in Table 2-1. You can check the line voltage setting of your supply by examining the door on the line module. This is located on the rear panel of your supply as shown in Figure 2-2. The red mark that appears in one of the four windows on the line module indicates the line voltage setting for which your supply is set.

If necessary, you can convert the supply from one line voltage setting to another by following the instructions under LINE

VOLTAGE CONVERSION (see page 29).

Nominal

Voltage

100 V

120 V

220 V

240 V

Line Voltage Range

Nominal

-13%, +6%

Table 2-1. Input Power

Maximum Input Current

(rms.)

6.3 A

5.7 A

3.0 A

3.0 A

High Line Inrush

Current (PK)

85 A

85 A

50 A

50 A

Fuse

8 AM

8 AM

4 AM

4 AM

Line Fuse

The ac line fuse is located behind the door on the line module (see Figure 2-3). To access the fuse, remove the power cord and push against the tab on the line module in the direction of the ac input socket. The current rating of the fuse is based on the line voltage setting of your supply. Table 2-2 gives the Agilent part numbers for the fuses that should be used with specific line voltages.

Figure 2-2. Rear Panel Detail (Agilent 6626A Shown)

Line

Voltage

100/120 V

220/240 V

Fuse

Needed

8AM

4AM

Table 2-2 Line Fuses

Agilent Part Number (for ¼ X 1¼ in. fuses only)

2110-0342

2110-0055

Note: All fuses are rated for 250 V.

Figure 2-3. Line Module Detail

Installation 27

Power Cord

The power supply is shipped from the factory with a power cord that has a plug appropriate for your location. Figure 2-4 shows the standard configuration of plugs used by Agilent Technologies. Below each drawing is the Agilent part number for the replacement power cord equipped with a plug of that configuration. If a different power cord is required, contact the nearest Agilent Technologies Sales and Service office.

For your protection, the National Electrical Manufacturer’s Association (NEMA) recommends that the instrument panel and cabinet be grounded. This power supply is equipped with a three-conductor power cord; the third conductor being the ground. The power supply is grounded only when the power cord is plugged into an appropriate receptacle. Do not operate this power supply without adequate cabinet ground connection.

Figure 2-4. Power Cord Plug Configurations

SHOCK HAZARD. Connect the power cord to a grounded receptacle before you connect any external floating voltages to the supply.

The offset pin on the standard three-prong power cord connector is the ground connection. If a two contact receptacle is encountered, it must be replaced with a properly grounded three-contact receptacle in accordance with the National

Electrical Code, local codes and ordinances. The work should be done by a qualified electrician.

Line Voltage Conversion

You can change the supply to accept 100 V, 120 V, 220 V and 240 V ac input by adjusting the voltage selector card located inside of the line module (see Figure 2-3). After you have changed the line voltage, refer to Table 2-2 and check that the fuse inside the line module is the correct fuse for that line voltage. The procedure is as follows:

1. Turn off power and remove the power cord from the ac input socket on the back of the power supply.

2. To open the line module, move the plastic door on the module aside. If your line voltage change requires a change in the rating of the fuse, rotate FUSE PULL to the left and remove the fuse.

3. Grasp the voltage select pc board with a pair of needle-nose pliers and slide it out of its slot.

4. To select a voltage, orient the pc board so that the desired voltage appears on the top left side of the board. Push the board all the way back into its slot. The desired line voltage must be visible when the board is installed.

5. Install the correct fuse in the door of the line module if your line voltage change also requires a change in the rating of the fuse (see Table 2-2).

FIRE HAZARD. Make sure the replacement fuse is one of the same type (size) and rating (amps) that is consistent with the voltage level you are operating at. Do not use a substitute fuse; use a fuse with the same Agilent Part number listed in Table 2-2.

6. Close the door of the line module and insert the power cord in the ac input socket. Your power supply is now configured to operate at the voltage you selected.

GP-IB Interface Connector

The GP-IB connector on the rear panel connects your power supply to your computer and other GP-IB devices (see Figure

2-2). Accessories (page 12) in Chapter 1 lists the cables and cable accessories that are available from Agilent Technologies.

An GP-IB system can be connected together in any configuration (star, linear, or both) as long as the following rules are observed

1. The total number of devices, including the computer, is no more than 15.

2. The total length of all the cables used is no more than two meters times the number of devices connected together, up to a maximum of 20 meters.

Note

IEEE Std. 488-1978 states that you should exercise caution if your individual cable lengths exceed 4 m.

Do not stack more than three connector blocks together on any GP-IB connector. The resultant leverage can exert excessive force on the mounting panels. Make sure that all connectors are fully seated and that the lock screws are firmly finger tightened. Do not use a screwdriver. Use a screwdriver only for the removal of the screws.

Installation 29

3

Getting Started

Introduction

This chapter is intended for the first time user of the supply. It provides four main discussions:

Front Panel Controls and Indicators

Turning on Your Supply

Checking Out Your Supply Using Local Control

Introduction to Remote Operation

First, the supply’s front panel controls and indicators are briefly described. Some of the controls and indicators will be used in the Turn On And Checkout procedures that follow. Chapter 6 describes how to use all of the front panel controls.

Successful completion of the Turn On And Checkout procedures ensures with a high level of confidence that your supply is operating properly. Complete performance testing and troubleshooting procedures are given in the Service Manual (Agilent

Part No. 06626-90003).

The checkout procedures are performed locally from the front panel. In addition to checking the operation of your supply, these simple step-by-step checkout procedures will help the first time user become familiar with operating the supply from the front panel.

When you have completed the checkout procedures, you are then introduced to the fundamentals of operating the supply remotely from a computer. You will learn how to send a command to the supply from the computer and how to get data back to the computer from the power supply. A few of the most often used power supply commands will be described to help you get started and become familiar with the basics of programming your supply.

After completing this chapter, you can proceed to Chapter 4 to find out how to make load connections to your supply’s outputs and then to Chapter 5 (Remote Control) and/or Chapter 6 (Local Control) to learn all the details about operating your supply.

Front Panel Controls and Indicators

The power supply’s controls and indicators are shown in Figure 3-1 and are described in Table 3-1. Note that the front panel controls are identical for all models except for the number of OUTPUT annunciators (number 3 in Figure 3-1).

Getting Started 31

3 5 1 6

6626A SYSTEM DC POWER SUPPLY

VOLTS AMPS

45.153 14.235m

1 2 3 4 CV CC UNR OCP ERR RMT ADDR SRQ

-- OUTPUT - ENBLD

LCL

SYSTEM

ADDR

ERR

STO

RCL

OUTPUT

RANGE

V/I

VOLT

OVSET VOLT

OUTPUT

SELECT

VSET

OCP

RESET

CURR

ISET

CURR OUTPUT

ON/OFF

LINE

ON

OFF

7

7

4

1

0

5

2

.

8

ENTRY

9

6

3 ENTER

8

9 4 2

Figure 3-1. Agilent 6626A Front Panel

Number

1

2

3

4

Controls/Indicators

LCL key

GP-IB Status

Annunciators

(These three annunciators indicate the

GP-IB status of the power supply).

OUTPUT

Annunciators

Power Supply Status

Annunciators

(These five annunciators indicate the status of the power supply).

Table 3-1. Controls and Indicators

Description

Returns power supply to local mode (unless local lockout has been received via GP-IB). Also, turns the power supply’s display on if it was turned off via the

GP-IB.

RMT - Indicates that the power supply is operating under remote control (GP-IB)

ADDR - Indicates that the power supply is addressed to talk or to listen.

SRQ - Indicates that the power supply is requesting service.

Indicate which output channel has been selected for front panel control and/or display (Only one output annunciator can be on at a time.)

CV - Indicates that the selected output channel is in the constant voltage mode.

CC - Indicates that the selected output channel is in the positive constant current mode ( + CC) or the negative current limit ( - CC) mode.

Page

3-13, 5-2,

6-1

3-13, 5-2,

6-1,

3-10, 5-1

5-1, 5-2,

5-3, 5-17

3-7, 3-8,

6-1--6-4

3-8, 4-1,

6-2, 6-3

3-8, 4-1,

6-2, 6-3

32 Getting Started

Number

4 (cont)

5

6

Table 3-1. Controls and Indicators (continued)

Controls/Indicators

Alphanumeric LCD

Display (When power is turned on, all segments will be displayed for approximately 2

Description

UNR- Indicates that the selected output channel is unregulated; i.e., it is not regulated by CV or CC control loops.

OCP ENBLD - Indicates that the overcurrent protection function for the selected channel is enabled.

ERR - Indicates that a programming or hardware error has occurred and that the ERR bit in the serial poll register has not been cleared

Normally displays the measured output voltage and current for the selected channel. When programmed from the front panel, the function being programmed

(e.g. VSET), the output channel (e.g. 2), and the present value (e.g. 2.250) will be displayed. Error conditions seconds).

System Control Keys

(These four control keys affect the entire power supply and are independent of the output selected.) will be spelled out in alpha characters

ADDR - Displays the power supply’s GP-IB address.

You can change the address using the numeric entry keys. You cannot query or change the address remotely

(over the GP-IB).

ERR - Displays a programming or hardware error message and clears the ERR bit in the serial poll register

STO - Used in conjunction with the numeric entry keys to store the present output conditions for all outputs in the specified internal register (0-10). Each register stores the following parameters: VSET, VRSET, ISET,

IRSET, and OVSET. In addition, register 0 also stores

OCP, DLY, and MASK parameters. Register 0-3 are non-volatile.

The parameters stored in register 0 are the "power-up" parameters. These parameters are set upon applying AC power to the power supply. Note that the STO command cannot be used over the bus more than once per non-volatile register (0-3) without cycling the AC power (OFF and ON). However, these registers can be used for an unlimited number of times via the front panel keys without cycling the AC power .

RCL - Used in conjunction with the numeric entry keys to recall the settings from the specified internal register

(0 to 10). All outputs are set to the recalled values.

4-4

Page

3-8, 6-4,

5-13

4-4, 5-16

3-7--3-10,

4-4, 5-19,

5-20,

6-1--6-4

3-10, 5-2,

6-4

5-1, 5-20

5-1, 5-13

5-1, 5-13

Getting Started 33

Number

7

Table 3-1. Controls and Indicators (continued)

Controls/Indicators

Output Control Keys

(These twelve keys are

Description

OUTPUT SELECT - Selects one of the output output dependent).

channels for local control or display. This key allows the channels to be selected in forward (

Ð) or reverse (Ï) sequence.

VSET - Displays the selected output’s present voltage setting. The setting can be changed using the numeric entry keys,

Ñ VOLT, or Ò VOLT keys..

ISET - Displays the selected output’s present current setting. The setting can be changed using the numeric entry keys,

Ñ CURR, or Ò CURR keys.

OUTPUT ON/OFF - Toggles the selected output on and off. When off, "DISABLED’’ appears on the display.

OVSET - Displays the selected output’s overvoltage trip point. The setting can be changed using the numeric entry keys.

RANGE V/I - Displays the selected output’s full scale programming range for voltage and current. This key also enables the Up/Down keys to change the programming range.

Up/Down Keys - These keys increment or decrement the voltage or current settings of the output selected.

The display will be in the metering mode for the output selected. When one of the keys is pressed, it will change the output selected by one least significant bit (LSB). If the key is pressed and held, it will continually change the setting until it is released The

Ñ VOLT key is disabled when the supply is in the CC mode. The

Ò

CURRENT key is disabled when the supply is in the CV mode.

These keys will also allow you to change the programmed resolution range of voltage or current after the RANGE V/I key is depressed.

Ñ VOLT - Increases the selected output voltage by an

LSB and then a faster rate as the key is kept pressed, or after the RANGE V/I key has been pressed, sets the selected output to the high voltage range.

Ò VOLT - Decreases the selected output voltage by an

LSB and then at a faster rate as the key is kept pressed, or after the RANGE V/I key has been pressed, sets the selected output to the low voltage range.

Page

3-7, 3-8,

6-1--6-4

3-8, 3-9,

5-9, 5-11,

6-2

3-8, 5-11,

6-3

5-12, 6-3

3-8, 5-12,

6-4

3-9

3-9

3-9, 6-2,

6-3

3-9, 6-2

3-9, 6-2

34 Getting Started

Number

7 (cont)

Table 3-1. Controls and Indicators (continued)

Controls/Indicators Description

Ñ CURRENT - Increases the selected output current by an LSB and then at a faster rate as the key is kept pressed, or after the RANGE V/I key has been pressed, sets the selected output to the high current range.

Ò CURRENT - Decreases the selected output current by an LSB and then at a faster rate as the key is kept pressed, or after the RANGE V/I key has been pressed, sets the selected output to the low current range.

OCP - Toggles the selected output’s overcurrent protection circuit on and off.

RESET - This key performs the following:

1. Reset the selected output’s overvoltage crowbar (the cause of the overvoltage must be removed before reset is successful) .

2. Reset the selected output’s overcurrent condition and returns the output to its previous settings (the cause of the overcurrent must be removed before reset is

Page

3-9, 6-3

3-9, 6-3

3-8

3-8

8 Numeric Entry Keys

(These keys are used in conjunction with many of the System Control and Output Control keys to enter the desired values into the power the metering mode. Supply.

successful).

3. Return the display to the metering mode from any other mode (e.g. VSET). In the metering mode, the measured output voltage and current of the selected output are displayed .

0 to 9 - Set the value of the specified function and

and (e.g. VSET 1 6.550

.

Ï(backspace) - Erases the previous keystroke.

Depressing this key without setting a

value places the display in

ENTER - Enters the values on the display for the specified function, initiates the function, and returns the display to the metering mode. Pressing this key without setting a value will result in retention of the previous values and returning the display to the metering mode.

3-8,

6-1--6-5

6-1

3-8,

6-1--6-5

9 LINE switch Turns ac power on and off.

3-6

Table 3-1, in addition to providing a brief description of each control and indicator, lists the paragraphs in which the use of each control and indicator is described. Because most of the functions performed by the front panel controls can also be performed remotely by power supply commands, the corresponding paragraphs in Chapter 5 (Remote Operation) are listed in Table 3-1 where applicable.

Getting Started 35

Turning On Your Supply

The following paragraphs describe the power-on sequence which includes a self test of most of the power supply’s circuits.

Before you turn on your supply, make sure that:

The line module on the rear panel is set to match your input line voltage.

The proper fuse is installed and the line cord is plugged in.

If you have any questions concerning installation or power requirements, review Chapter 2.

To turn on your supply, press the front panel LINE switch. When the power is initially applied, the supply performs a series of self tests which last about 3 seconds. Included in these tests are checks of circuits on the GP-IB board and on each of the output boards.

Normal Self Test Indications

If the supply passes the self test, the display will first show all segments of the LCD display with annunciators on as illustrated in Figure 3-2.

Figure 3-2. Test Pattern of all Display Segments at Power-on

After all segments are displayed, the supply’s GP-IB address will appear for approximately 2 seconds as shown in Figure 3-

3. As shipped from the factory, the power supply’s address is set to 5. You must know this address before you can remotely program your supply.

Figure 3-3. Typical Address Display During Power-On Self Test

When self test is successfully completed, the power-up parameters stored in register 0 will be set (see page 89 for a complete description of the store and recall functions). As shipped from the factory, this non-volatile register contains zero volts and zero current (see Table 5-4 on page 72) for all outputs. Note that the CV annunciator will indicate that the output is in the constant voltage mode.

36 Getting Started

Figure 3-4. Typical Display at Power-On

Self-Test Errors

If the supply fails the power-on self-test, all power supply outputs will remain disabled (off) and the display will indicate the type of failure and the output channel on which it occurred. Figure 3-5 shows that self-test detected an error in output channel 3. Error messages that could appear on the display if self-test fails are listed below. Self-test error messages are explained in Appendix D and troubleshooting procedures are given in the Service Manual. You may also call your Agilent

Sales Office for help.

Power-On Self Test Error Messages

HDW ERR CH "N"

8291 FAILED

TIMER FAILED

CV DAC CH ’’N"

CC DAC CH "N"

OV DAC CH "N"

FUSE CH ’’N"

Note: "N" specifies the failed output channel number: 1,2,3, or 4 as applicable.

Figure 3-5. Sample Self-Test Failure Display

Checking Out Your Supply Using Local Control

The following procedures use the display and keys on the front panel to check each of your power supply’s outputs. No test equipment, other than a jumper wire (14 AWG), is required to perform these tests. The tests must be repeated for each output of your particular supply. The checkout consists of voltage, overvoltage, and current tests. It is assumed that power has already been turned on, the supply has passed the power-on self-test, loads are not connected to any of the supply’s outputs, and sense clips are connected between the sense terminals and the output terminals.

Note

The following procedures are identical for all models and for all outputs. Use the OUTPUT SELECT key to select an output to be tested. If an output fails any of the tests, refer to the troubleshooting section in the Service Manual.

Getting Started 37

Voltage Test

1. Set the voltage of the selected output to 10 V by pressing:

VSET 1 0 ENTER

2. Check that the display reads approximately 10 V and 0 A and the CV annunciator is on indicating that the supply is in the constant voltage mode of operation.

Overvoltage Test

1. Program the overvoltage protection (OVP) to 19 V by pressing:

OV

SET

2. Set the voltage to 16 V by pressing:

VSET

1

1

9

6

ENTER

ENTER

3. Check that the display reads approximately 16 V and 0 A.

4. Set the voltage to 20 V by pressing:

VSET 2

5. Check that the display reads "OVERVOLTAGE’’

6. Reset the supply by pressing:

0 ENTER

VSET 1 6

7. Check that the display reads approximately 16 V and 0 A.

ENTER RESET

Current Test

1. Turn off the supply.

2. Remove the barrier block cover from the output to be tested and connect a short circuit (jumper wire) between the + V and -V terminals of the output being tested.

3. Turn on the supply.

4. Use the to select the output being tested.

5. Set the voltage to a volts by pressing:

VSET 5

Set the current to 0.01 amps by pressing:

ISET .

ENTER

0 1 ENTER

38 Getting Started

6. Check that the display reads approximately 0 volts and the programmable current limit value. Also, check that the front panel CC annunciator is on indicating that the output is in the constant current mode of operation.

7. Set the current to 0.5 A by pressing:

ISET .

5 ENTER

8. Check that the display reads approximately 0 V and 0.5 A.

9. Enable the overcurrent protection circuit by pressing:

OCP

10. Check that the OCP ENBLD annunciator is on indicating that overcurrent protection is enabled and the display reads

’’OVERCURRENT’’. When in overcurrent, the output is disabled.

11. Disable the overcurrent protection circuit by pressing:

OCP

12. Reset the output by pressing:

RESET

13. Check that the display reads approximately 0 V and 0.5 A.

14. Turn off the supply and remove the jumper from the output terminals.

Repeat the voltage, overvoltage and current tests given on page 38 for the other output channel(s) using the key

Changing Resolution Range

1. Press the RANGE V/I key.

2. Observe the selected output’s full scale programming range for voltage and current in the front panel display.

3. Press either the

Ñ or the Ò key. (voltage or current)

4. Observe that the selected output’s full scale programming range for voltage and current changes to the high range (if the

Ñ key is pressed) or low range (if the Ò key is pressed). Note that no range change will occur if the Ñ or Ò key is pressed while the supply is already in the requested range.

Introduction To Remote Operation

The following paragraphs explain the fundamentals of operating the supply remotely from a computer. Only a few commonly used programming commands will be discussed. Refer to Chapter 5 for a detailed description of all the commands. The intent of this discussion is to help first time users to quickly become familiar with operating their supply from a computer.

The programming examples that follow assume that a computer is connected to the GP-IB connector on the rear of your supply (see Chapter 2), power is applied, and loads are not connected to any of the supply’s outputs. The examples used are primarily for Agilent Series 200/300 computers using Agilent BASIC language. Read the manuals for your particular computer to find out which statements you must use.

Getting Started 39

Enter/Output Statements

The programming statements you use to operate your supply from remote depend on your computer and its language. In particular, you need to know the statements your computer uses to output and enter information. For example, the Agilent

BASIC language statement that addresses the power supply to listen and sends the command to the power supply is:

OUTPUT

The Agilent BASIC language statement that addresses the power supply to talk and reads back data from the power supply is:

ENTER

The supply’s front panel ADDR annunciator is on when the supply is addressed to talk or to listen.

Reading the GP-IB Address

Before you can operate your power supply remotely, you need to know its GP-IB address. The address was displayed during the power on sequence described in page 36. To see the address, press:

ADDR

A typical address display is shown in Figure 3-6:

Figure 3-6. Typical Address Display

The displayed response is the power supply’s GP-IB address. When sending a remote command, you append this address to the computer’s GP-IB interface select code (normally 7). For example, if the select code is 7 and the power supply’s GP-IB address is 5, the combination is 705.

Changing the GP-IB Address

Note

All examples in this discussion assume a GP-IB address of 5. It is recommended that you retain this address to simplify programming.

Every device on the GP-IB must have an address. The supply’s address is factory set to decimal 5. Any address from 0 through 30 is a valid address. If you need to change the Agilent 662xA Supply’s address press:

ADDR

You can now enter a new address. For example, press:

1 4 ENTER

You have now changed the address from 5 to 14. If you want to change the address back to 5, repeat the above procedure but use 5 instead of 14 in the last step. Note that the address is stored in the power supply’s non-volatile memory and therefore will be retained through interruption of the ac line power.

40 Getting Started

Sending a Remote Command

To send the power supply a remote command, combine your computer’s output statement with the GP-IB interface select code, the GP-IB device address, and finally, the power supply command. For example, to set the output voltage of output channel 1 to 2 volts, send:

Getting Data From The Supply

The supply is capable of measuring the values of its output parameters in response to queries. In this example, the query asks the supply to measure the output voltage at output 1.

When you send a query from remote, the supply does not display the response as it did when you executed the command from the front panel. Instead, it holds the response in an output buffer. The output buffer is a register that holds information until it is read by the computer or is replaced with new information.

Note

On an Agilent Series 200/300 Computer, the A variable must be declared before you do the following steps. Refer to your computer’s operating manual for more information.

Use your computer’s enter statement to get the response from the output buffer. For example, execute:

ENTER 705; A

Followed by:

DISP A

The ENTER statement enters whatever is in the supply’s output buffer into the computer’s A variable. The DISP statement displays the A variable’s contents on the computer’s display.

Often Used Commands

The command set contains over sixty commands that allow you to program the power supply in a variety of applications.

Within this command set, however, is a small subset of commands that are all you need for most applications. These commands are: VRSET, IRSET, VSET, ISET, VOUT?, IOUT?, OUT, OVSET, and OCP.

Getting Started 41

Each of these commands is briefly discussed in the following paragraphs to help you get started in programming your supply. To know more about these commands, refer to Chapter 5.

The VRSET and IRSET commands select the range the power supply operates in. Two ranges are available for each output

- standard resolution, and high resolution (see Table 1-2).

If the value for voltage or current entered is not within the present operating range of the supply, the error annunciator will indicate an error. Sending ERR?, and addressing the supply to talk, will return the error message (in the case ’’number range" error would be displayed.

Voltage and Current Programming. You can send voltage and current values to the power supply directly in volts or amps. The following examples use voltage and current values that are within the range of any output that the power supply provides.

To set the voltage of output 1 to 5 volts, send:

OUTPUT 705; “VSET 1,5”

To set the current of output 2 to 450 milliamps, send:

OUTPUT 705; ''ISET 2,.450"

Output Voltage and Current Measurement. The power supply has built-in metering circuitry, which allows it to monitor the output voltage and current of the selected output. You can instruct the supply to return its present output voltage and current using VOUT? and IOUT? queries respectfully. The metering circuitry is auto-ranging, and will select the proper resolution.

To measure the output voltage at output 1, send:

OUTPUT 705; "VOUT? 1"

To get the measurement from the output buffer, send:

ENTER 705; A

DISP A

The computer should display a reading of approximately 5 volts .

To measure the output current at output 2, send:

OUTPUT 705: "IOUT? 2''

To get the measurement from the output buffer, send:

ENTER 705; A

DISP A

42 Getting Started

Output On/Off. You can turn a specified output on or off. Individual outputs can be controlled as shown below.

To turn off output 1, send:

OUTPUT 705; ’’OUT 1,0"

When an output is turned off, it is set to 0 volts and to the minimum current limit value

To turn on output 1, send:

OUTPUT 705; ’’OUT 1,1"

When an output is turned on it will return to the voltage and current settings determined by the present VSET and ISET values.

Overvoltage Setting. You can send an overvoltage setting value to the power supply directly in volts. If the output voltage exceeds this setting, the output crowbar is fired, and the output voltage is quickly downprogrammed and disabled (0 volts output).

To set the overvoltage value of output 2 to 3.5 volts, send:

OUTPUT 705; “OVSET 2,3.5”

Overcurrent Protection. The output will go to the off state (0 volts and 0 current) when the overcurrent protection (OCP) feature is enabled and the output is in the + CC mode. To enable the overcurrent protection mode for output 2 send:

OUTPUT 705; ''OCP 2,1"

To disable the overcurrent protection mode for output 2, send:

OUTPUT 705; ''OCP 2,0"

When overcurrent protection is disabled and the output is in + CC mode, the output current will be limited to and will stay at the ISET value.

Returning the Supply to Local Mode

In the remote mode (RMT annunciator on), the front panel keys have no effect on any of the supply's outputs and only the computer can control the supply. However, you can still use the front panel display to monitor the output voltage and current or check any of the present settings (VSET, ISET, OVSET, etc.) of the selected output channel.

If you want to use the front panel keys to change the output settings, you must return the supply to the local mode. You can return the supply to the local mode (provided that the local lockout command has not been received from the computer) by pressing the LCL key. A change between the local and remote modes (or vice versa) will not result in a change in the power supply outputs. Refer to Chapter 6 for additional details on using the LCL key and operating the supply in the local mode.

Getting Started 43

4

Output Connections and Operating Information

Introduction

This chapter explains how to make connections to the output terminals located on the rear of your power supply. Some general operating information is included in this chapter to help you understand how the power supply operates under various load conditions. This information applies whether you are operating the supply via the front panel or the GP-IB.

Output Ranges

Figure 4-1 identifies the output combinations that are available on the power supply. Each output can operate as a constant voltage (CV) or constant current (CC) source over a wide variety of output voltage and current combinations. In addition, each output has an active downprogrammer which operates at currents up to approximately 110% of the maximum positive current rating of the output. This means that each output can actively sink as well as source its maximum rated output current. At voltages below 2.5 V, a downprogramming resistor continues downprogramming until the voltage reaches approximately zero volts.

Figure 4-1. Output Combinations

Output Connections and Operating Information 45

Operating Quadrants

Figures 4-2A and 4-2B show the operating locus of your power supply in three quadrants. The area in quadrant 1 shows the operating locus defined by the voltage and current settings of each output. The area in quadrant 2 indicates the locus where each output can operate as a current sink. You cannot program current limit values in quadrant 2.

Figure 4-2. Typical Output Range Characteristics

46 Output Connections and Operating Information

Figure 4-3 shows the current sink characteristics lower voltages in greater detail. The area in quadrant 4 illustrates the reverse polarity diode characteristics of each output. Do not operate any output with reverse-voltage currents that are greater than the maximum rating of the output.

Figure 4-3. Typical Downprogramming Characteristic Below 2.0 V

Operating Ranges

Notice in Figure 4-2, there are four overlapping areas in quadrant one for both the 25 watt and 50 watt outputs. The Low

Range Voltage and Low Range Current ranges are higher precision than the High Range Voltage and High Range Current ranges. The voltage and current ranges can be set independently (see page 73). If the high resolution range is programmed, attempting to program values outside this range will result in a number range error.

The 50 watt output (Figure 4-2B) is different from the 25 watt output (Figure 4-2A), in that there is a power limiting boundary at 1 amp in the High Voltage/High Current range. As a result, the 50 watt output can be set to no more than 1.03

amps at voltage settings above 16.16 volts, and vice versa.

Setting parameters of voltage or current beyond this power limit boundary, but within the High Voltage/High Current range will cause the other parameter (voltage or current) to automatically be scaled down to the power limiting boundary (see pages 71 and 73).

The voltage and current programming resolution of the outputs is affected by the setting of the range. The voltage of the 25 watt output, when programmed to the Low Voltage Range, has a voltage programming resolution of 460

µ

V. When programmed above 7 volts, the High Voltage Range must be selected resulting in a programming resolution of 3.2 mV.

The current of the 25 watt output, when programmed to the High Current Range, has a current programming resolution of 1

µ

A. When programming above 15 mA, the High Current Range must be selected, resulting in a programming resolution of

33

µ

A.

The voltage of the 50 watt output, when programmed to the High Voltage Range, has a voltage programming resolution of

1 mV. When programming above 16.16 volts, the high range must be selected, resulting in a programming resolution of 3.2

mV.

The current of the 50 watt output, when programmed to the High Current Range, has a current programming resolution of

13

µ

A. When programmed above 200 mA, the High Current Range must be selected, resulting in a programming resolution of 131

µ

A.

The readback range of the outputs is not programmable, and is selected automatically by the power supply.

Output Connections and Operating Information 47

The readback resolution of the 25 watt outputs when metering voltages of 7 volts or below, will be 483

µ

V. For voltages above 7 volts, the readback resolution will be 3.3 mV.

The readback resolution of the 25 watt outputs when metering source currents of 15 mA or below, the readback resolution will be 1

µ

A. When metering source currents above 15 mA, the readback resolution will be 48

µ

A.

When metering sink currents of 15 mA or below, the readback resolution of the 25 watt outputs will be 1

µ

A. When metering sink currents above 15 mA, the readback resolution will be 37

µ

A.

When metering voltages of 16 volts, or below, the readback resolution of the 50 watt outputs will be 1.1

µ

V. For voltages above 16 volts. the readback resolution will be 3.3 mV.

When metering source currents of 200 mA or below, the readback resolution of the 50 watt outputs will be 14

µ

A. When metering source currents above 200 mA, the readback resolution will be 160

µ

A.

When metering sink currents of 200 mA or below, the readback resolution of the 50 watt outputs will be 14

µ

A. When metering sink currents above 200 mA, the readback resolution will be 151

µ

A.

Protection Features

Protective circuitry within the supply can limit or turn off an output in the event of an abnormal condition. The activated protection feature can be determined by observing the front panel display area. You can also read back the status of the supply over the GP-IB. The following protection features are implemented:

OVERVOLTAGE -- shorts the output by firing an SCR crowbar and sets zero volts and minimal current on an output if any of the following conditions are present:

1. The output voltage exceeds the programmed overvoltage trip point.

or

2. The sum of the voltage from the + V output terminal to the +S terminal plus the voltage from the -S terminal to the -V output terminal exceeds 10.0 V (applies to remote sensing only).

or

3. A trip signal is received on the output’s OV terminals.

or

4. The output’s fixed overvoltage circuit is activated.

The OV trip point can be programmed up to 55 V. When an overvoltage occurs, the word OVERVOLTAGE appears in the front panel display and the OV status bit is set for that output. Chapter 5 explains how to program the overvoltage trip level.

A fixed overvoltage threshold of approximately 120% of the maximum rated output voltage is built into each output.

Because the fixed overvoltage circuit is biased from the output terminals, it can be activated and provide protection even when the supply is not connected to the ac power line.

The OVRST command or front panel RESET key restores the programmed voltage and current values and clears the OV once the cause of the overvoltage has been eliminated.

OVERCURRENT -- when the overcurrent protection feature is enabled, and the output is sourcing current and enters the +

CC operating mode, the output will be disabled (set to zero volts and minimal current) and the word OVERCURRENT will appear on the front panel display. In addition, the OC status bit is set for that output. The OCRST command or front panel

RESET key restores the programmed voltage and current values and clears the OC once the cause of the overcurrent condition has been eliminated. Refer to Chapter 5 for programming details.

48 Output Connections and Operating Information

UNREGULATED OUTPUT -- the supply informs the user when output regulation is not guaranteed. This can occur when attempting to sink excessive currents below 4 volts on 25 W outputs and 2 volts on 50 W outputs or when operating outputs in parallel. The UNR annunciator on the front panel and the UNR bit in the status register indicate that the specified output is unregulated. Line voltage dropout or an incorrectly set ac power module can also cause the output to become unregulated. If line voltage dropout continues, the supply shuts down and will return to the power-up condition when normal line voltage is restored.

OVERTEMPERATURE -- shuts down the linear pass transistors and downprogrammer of the output that has reached an unsafe operating temperature. Operation of the other outputs is unaffected. An overtemperature can occur because of excessively high ambient temperature, a blocked fan, or insufficient space at the sides for adequate air circulation. When an overtemperature condition occurs, the word OVERTEMP appears in the front panel display and the OT status bit is set.

This circuit resets automatically and restores the output approximately 30 seconds after the temperature drops sufficiently for safe operation.

ERROR -- if the power supply receives an invalid command either through the front panel or the GP-IB, the ERR annunciator on the front panel comes on and the ERR bit in the serial poll register is set. The power supply does not execute the command and remains at previously set values. Pushing the ERR button in local mode displays the error message and clears the error. The error indicator may also indicate that an instrument failure has occurred. Refer to Appendix D for further details.

Connecting The Load

Each terminal block cover on the rear panel is secured by a locking tab which snaps into a slot at the left of the terminal block. To remove, insert a screwdriver into this rectangular slot and move the locking tab to the left. When the locking tab releases, gently pull the terminal block cover away from the terminal block. To reinstall the cover, align it over the terminal block and gently press it into position until the locking tab engages.

SHOCK HAZARD. Turn off ac power before making rear panel connections. All wires and straps must be properly connected with terminal block screws securely tightened. Replace terminal block covers before reapplying power.

Each rear terminal block has six M3.5 x 0.6 x 6 mm screws for attaching wires (see Figure 2-2). Load connections to the supply are made at the + V and -V terminals on each terminal block. Do not connect unterminated wires to the load terminals. Wires used for load connections must be properly terminated with termination connectors securely attached.

Remember to replace the impact resistant plastic covers (Agilent P/N 06624-20007) over the terminal blocks after making connections.

Consistent with good engineering practice, all sense and trigger leads connected to the rear terminal blocks should be twisted and shielded to maintain the instrument’s specified performance.

Wire Size Selection

FIRE HAZARD Select a wire size large enough to carry short-circuit current without overheating.

Two factors must be considered when selecting wire size for load connections: conductor temperature and voltage drop. To satisfy safety requirements, load wires must be heavy enough not to overheat while carrying the short-circuit output current of the unit. Table 4-1 lists the current-carrying capacity

(ampacity) for various sizes of stranded wire.

Note that the minimum wire size required to prevent overheating may not be large enough to prevent OV trip when remote sensing and to maintain good regulation. The load wires should be large enough to limit the total voltage drop in both leads to no more than 10 volts. See Figure 4-6 for total allowed load lead drop when programming to high voltage settings. With remote sensing, load regulation is degraded 1 mV per 1 V in the - V output terminal load lead. (See Remote Sense

Connections on page 52).

Output Connections and Operating Information 49

Note

To prevent tripping of the overvoltage circuit, pick a wire size sufficient to handle the FULL output current of the unit no matter what the intended load current or current limit setting.

Table 4-1 lists the resistivity for various wire sizes and the maximum lengths to limit the voltage drop to 1.0 volts for various currents.

Table 4-1. Stranded Copper Wire Ampacity and Maximum Wire Lengths to Limit Load Lead Voltage Drop

Wire Size

(AWG)

20

18

16

14

12

(Cross Section

Area in mm

0.5

0.75

1

1.5

2.5

2

)

Ampacity Per Wire (Amps)

2 Wire Bundled 4 Wire Bundled

7.8

14.5

18.2

29.3

37.6

7.8

9.4

12.7

15.0

23.5

6.9

12.8

16.1

25.9

33.2

6.9

8.3

11.2

13.3

20.8

Resistivity

(

(

/ft)

0.0102

0.0064

0.0040

0.0025

0.0016

/m)

0.0401

0.0267

0.0200

0.0137

0.0082

Max Length to Limit Voltage

Drop to 1 V Per Lead

5 A 10 A

(feet)

20 A

20

30

50

--

--

10

15

25

40

--

5

7.5

12.5

20

30

5

7.4

10

14.6

--

(meters)

2.4

3.8

5

7.2

12.2

1.2

Notes:

1. Ampacities for AWG wires are derived from MIL-W-5088B. Maximum ambient temp: 55°C. Maximum wire temp:

1.8

2.6

3.6

6

105°C.

2. Ampacities for metric wires are derived from IE Publication 335-1.

3. Ampacity of aluminum wire is approximately 84% of that listed for copper wire.

4. Because of wire inductance considerations, it is recommended that you keep your load leads twisted, tie wrapped, or bundled together and less than 50 feet (14.7 meters) in length per lead.

5. See Page 49 for information on wire gauge considerations with capacitive loads.

Note

The OVP circuit senses at the main output terminals and not on the sense leads. Thus, the voltage sensed by the OVP circuit could be as much as 10 V higher than the voltage being regulated at the load. Program the OVP trip voltage accordingly when using remote sensing. In addition, if the total voltage drop in both leads exceeds 10 V, a protective circuit will fire the OVP circuit regardless of the OVP setting.

Load lead resistance is an important factor relating to the CV stability of the supply with remote sensing of capacitive loads. If high capacitance loads are expected, you should not use wire gauges heavier than 12 to 14 AWG for long runs of load lead. See Figure 1-4 for more information about stability with output capacitors.

Multiple Loads

If you will be sensing the load locally (as shipped from the factory) and are connecting multiple loads to one output, connect each load to the output terminals using separate connecting wires (see Figure 4-4). This minimizes mutual coupling effects and takes full advantage of the power supply's low output impedance. Each pair of wires should be as short as possible and twisted or bundled to reduce lead inductance and noise pickup.

If load considerations require the use of distribution terminals that are located remotely from the supply, connect the power supply output terminals to the remote distribution terminals by a pair of twisted or bundled wires. Connect each load to the distribution terminals separately. Remote voltage sensing is recommended under these circumstances. Sense either at the remote distribution terminals or, if one load is more sensitive than the others, directly at the critical load.

50 Output Connections and Operating Information

Figure 4-4. Optimum Hookup for Multiple Loads, Local Sensing

Note

When a load is connected through relay or switch contacts, contact bounce may activate the overvoltage circuit and shut down the supply. Therefore, it is recommended that the output be downprogrammed to 0 or turned-off (disabled) before the relay (or switch) contact is opened or closed.

Positive and Negative Voltages

Either positive or negative voltages can be obtained from the supply by grounding (or "commoning") one of the output terminals. Always use two wires to connect the load to the supply regardless of where or how the system is grounded. This supply can be operated with any output terminal

±

240 Vdc (including output voltage) from ground.

Remote Voltage Sensing

The power supplies will allow up to a total load lead drop of 10 V.

Because of the unavoidable voltage drop developed in the load leads, the as-shipped terminal block strapping pattern shown in Figure 4-4 does not provide the best possible voltage regulation at the load. The remote sensing connections shown in

Figure 4-5 improve the voltage regulation at the load by monitoring the voltage there instead of at the supply’s outputs terminals. This allows the power supply to automatically compensate for the voltage drop in the load lead. Remote sensing is especially useful for CV operation with load impedances that vary or have significant lead resistance. It has no effect during CC operation. Because sensing is independent of other power supply functions, remote sensing can be used regardless of how the power supply is programmed. Note that with remote sensing, voltage readback monitors the load voltage at the sense points.

Output Connections and Operating Information 51

Figure 4-5. Remote Voltage Sensing

The maximum voltage available at the power supply output terminals during remote sensing (see Figure 4-6) is 50.5 volts.

This allows the sum of the voltage across both load leads to equal 10 volts maximum. For lower output voltages refer to

Figure 4-3.

Remote Sense Connections

Remember to turn off the power supply before making or changing any connections on the rear panel terminal blocks.

Connect the unit for remote sensing by first disconnecting the straps between sense and load terminals. Then make your connections as shown in Figure 4-5. Connect the sense leads as close to the load as possible. See wire size selection

(pages 49 - 50) for information on selection of load lead wire gauge. Best results will be obtained by using the shortest load leads practical. It is recommended that you keep your load leads under 14.7 meters (50 feet) per lead because of inductance effects.

52 Output Connections and Operating Information

Figure 4-6. Total Allowable Load Lead Voltage Drop (total of both leads) with Remote Sensing

Output Noise Considerations

Any noise picked up on the sense leads will appear at the supply’s output and may adversely affect CV load regulation.

Twist the sense leads or use a ribbon cable to minimize the pickup of external noise. In extremely noisy environments it may be necessary to shield the sense leads. Ground the shield at the power supply end only; do not use the shield as one of the sensing conductors.

The noise specifications in Table 1-2 apply at the power supply output terminals when using local sensing. However, voltage transients may be produced at the load by noise induced in the leads or by load current transients acting on the inductance and resistance of the load lead. If it is desirable to keep voltage transient levels to a minimum, place an aluminum or a tantalum capacitor, with an approximate value of 10

µ

F per foot (30.5 cm) of load lead, right across the load

(see Figure 4-5). Refer to Figure 1-4 for capacitive load stability considerations.

Programming Response Time with an Output Capacitor

Because voltage programming into an external output capacitor may cause the supply to briefly enter CC operating mode, voltage programming response time may be longer than that specified in Table 1-2. Use the following formula to estimate the additional response time:

(

Added Output Capacitor)(Charge in Vout

)

Additional Response Time

=

Current Limit Setting

Output Connections and Operating Information 53

Open Sense Leads

The sense leads are part of the supply’s feedback path. Connect them in such a way so that they do not inadvertently become open circuited. The power supply includes protection resistors that reduce the effect of open sense leads during remote-sensing operation. If the sense leads open during operation, the supply returns to the local sensing mode, with the voltage at the output terminals approximately equal to the programmed value.

Overvoltage Trigger Connections

Each output of your power supply has two OV terminals on its rear panel terminal block. These terminals are labeled +OV and -OV. By connecting the OV terminals all in parallel as shown in Figure 4-7, an overvoltage shutdown on any one output will also trigger the overvoltage on the remaining outputs. Any number of OV terminals up to eight sets can be strapped together. Observe polarity when connecting the OV terminals in parallel.

Figure 4-7 Overvoltage Connections

The overvoltage trip point for each output can be set either from the front panel or by remote programming. You can also externally fire the overvoltage circuit of one or more outputs by applying a 5 volt pulse of at least 50

µ s to any pair of OV terminals (see Figure 4-8). As long as all OV terminals are wired together, the outputs will be crowbarred simultaneously .

External Trigger Circuit

Figure 4-8 illustrates a recommended external circuit that can be used to provide an OV trip signal to the OV terminals.

This circuit configuration provides good noise immunity and protects against the voltage pulse that is returned from the OV terminals every time that the overvoltage circuit fires. It can be operated from a wide range of bias voltages provided the input limiting resistors are chosen as tabulated in the figure. If it is not required to trip the OV with a TTL signal, then a bias supply, switch, current limiting resistor (R2), and protection diode are all that are required. Note that with the unit off

(ac power removed), the + OV and - OV terminals are inactive.

The internal equivalent OV circuit is shown in Figure 4-9. Note the internal DC blocking capacitor, bleed resistor, and noise bypass capacitors.

Do not exceed 50 volts maximum between the + OV and the - OV terminals. The OV terminals are rated at

±

240 Vdc (including external OV voltage) from chassis ground or any other output terminals.

54 Output Connections and Operating Information

Figure 4-8. External Trigger Circuit

Figure 4-9. Equivalent Internal OV Trigger Circuit

Power Supply Protection Considerations

Battery Charging

If you are using your supply in a battery charging application, it is recommended that a series protection diode be added to prevent damage to the supply during an overvoltage shutdown. Remember that each output has an overvoltage protection circuit that fires a crowbar to disable the output for any of the OVERVOLTAGE conditions described on page 48.

Figure 4-10 illustrates the recommended connections and protection circuit for a battery charging application. The diode will prevent damage to your supply that can result from excessive battery current flowing into the supply’s output in the event of an overvoltage shutdown.

Output Connections and Operating Information 55

Figure 4-10. Recommended Protection Circuit for Battery Charging

Capacitive Load Limitation

The programmable overvoltage protection circuit can be used to downprogram capacitive loads although it is primarily intended for use as a protection feature as described on page 48.

Repetitive (over 100 cycles) tripping of the overvoltage circuit with output capacitors greater than 2000

µ

F may result in eventual damage to the supply.

Parallel Operation

Connect in parallel only outputs that have equivalent voltage and current ratings.

Connecting outputs in parallel provides a greater current capability than can be obtained from a single output. Because each output contains an active downprogrammer that is capable of sinking current from only ONE identical output, you can parallel no more than two outputs. These outputs must have equivalent voltage and current capability. For example, on the

Agilent 6626A power supply, only outputs 1 and 2 or 3 and 4 may be connected together. On the Agilent 6625A supply, none of the outputs may be connected together. If you are mixing different model supplies (i.e. an Agilent 6625A and an

Agilent 6626A) be sure to connect only equivalent voltage and current outputs in parallel.

As an example, Figure 4-11 shows how to connect two outputs in parallel to a single load with local sensing. This configuration applies to both CV and CC operating modes. Connecting the load leads of output 2 directly to the + V and - V terminals of output 1 keeps the total length of the load leads to a minimum and reduces the number of wire connections that must be made at the load itself. Connecting the + S and -S terminals of output 2 directly to the sense terminals of output 1 compensates for the IR drop in the interconnecting load leads.

56 Output Connections and Operating Information

Figure 4-11. Parallel Connections with Local Sensing

CV Operation

For CV operation, one output must operate in CC mode and the other output must operate in CV mode. Although each output operates independently of the other, the output that is operating in CV mode will be ’’controlling" the voltage regulation of both outputs. Setting the output voltages as outlined in the following paragraph and configuring the outputs as shown in Figure 4-11 will allow output 1 to operate in CV mode and output 2 to operate in CC mode.

To assure that output 2 will be operating in CC mode, you must program output 2’s voltage to a higher value than the voltage of output 1. One way to accomplish this is to first program output 2 to the maximum allowable voltage setting for the desired operating range (see Table 4-2 or Figure 4-2). These values are 1% higher than the rated voltage for the operating range. Then, program output 1’s voltage to the desired operating voltage. The lower voltage setting of output 1 will determine the voltage that appears across the load. The current limit point of the paralleled outputs is the sum of both individual current limit points. The output current of the parallel combination is the algebraic sum of the individual current readbacks.

The + OV and - OV terminals of output 1 should be wired to the + OV and - OV terminals of output 2. When programming the overvoltage setpoint, set both outputs to the same overvoltage value. When resetting the overvoltage, first disable both outputs by using the OUTPUT ON/OFF key or OUT command. Next, reset both overvoltages. Finally, re-enable the outputs with the OUTPUT ON/OFF key or OUT command.

Output Type

25W

Table 4-2. Maximum Allowable Voltage Setting

Maximum Low Range Voltage Maximum High Range Voltage

7.07V

50.5V

50W 16.16V

50.5V

Note

Below 4 V on 25 W outputs and 2 V on 50 W outputs, the downprogrammer cannot sink the maximum rated current. (See Figures 4-2 and 4-3). To operate parallel outputs at voltages under 2.5 V, program both outputs to the same voltage setting. Depending on the load, one output may operate in the unregulated mode.

Output Connections and Operating Information 57

CC Operation

For CC operation, set the output voltages as outlined in CV operation (page 57), or alternatively, program the voltage settings of both outputs to the same voltage limit point. Then program the current of each output so that the sum of both currents equals the total desired operating current. The simplest way to accomplish this is to program each output to one half of the total desired operating current. Both outputs will operate in the CC mode.

Remote Sensing

If it is necessary to remote voltage sense at the load, parallel the sense leads of output 1 with the sense leads of output 2 and connect to the load as shown in Figure 4-12. The outputs can be programmed as previously described.

Figure 4-12. Parallel Connections with Remote Sensing

Specifications for Parallel Operation

Specifications for outputs operating in parallel can be obtained from the specifications for single outputs. Most specifications are expressed as a constant or as a percentage (or ppm) plus a constant. For parallel operation, the percentage portion remains unchanged while constant portions or any constants are changed as indicated below.

Current

Voltage

Load Transient Recovery

Time

All parallel specifications referring to current are twice the single output

specification except for programming resolution which is the same for both

single output and parallel output operation.

All parallel specifications referring to voltage are the same as for a single

output except for CV load effect, CV load cross regulation, CV source effect,

and CV short term drift. Below 4 V on 25 W outputs and 2 V on 50 W outputs, these are all twice the voltage programming accuracy (including the

percentage portion). CV load effect above 4 V on 25 W outputs and 2 V on

50 W outputs, could be twice the load effect specification for a single output.

CV output noise for output voltages less than 4 V on 25 W outputs, and 2 V on 50 W outputs, may be slightly higher than the output noise for a single

output.

350

µ s maximum to recover within 100 mV of nominal value following a load change within the range of 0 to full load.

58 Output Connections and Operating Information

Series Operation

SHOCK HAZARD. Floating voltages must not exceed 240 Vdc. No output terminal may be more than 240 Vdc from chassis ground.

Connect in series only outputs that have equivalent current ratings. Each output has a reverse voltage protection diode across its output terminals. The current conducted by this diode is not internally

limited by the output. Therefore, never connect an output in such a way that this diode will conduct

current in excess of the rated current of the output since damage could result.

Connecting outputs in series provides a greater voltage capability than can be obtained from a single output. Because the current is the same through each element in a series circuit, outputs connected in series must have equivalent current ratings. Otherwise, the higher rated output could potentially damage the lower rated output by forcing excessive current through it under certain load conditions.

Figure 4-13 shows an example of how to connect two outputs in series to a single load with local sensing. This configuration applies to both CV and CC operating modes. Connecting the + load lead of output 2 directly to the

- V terminal of output 1 completes the series connection between the two outputs. Connecting the + S terminal of output 2 directly to the - S terminal of output 1 and removing the sense jumper (between + S and + V) on output 2 compensates for the IR drop in the load lead from output 2 to output 1.

Figure 4-13. Series Connections with Local Sensing

CV Operation

For CV operation, first program the current setting of each output to the desired current limit point. Then program the voltage of each output so that the sum of both voltages equals the total desired operating voltage. The simplest way to accomplish this is to program each output to one half of the total desired operating voltage. Both outputs will operate in CV mode.

Output Connections and Operating Information 59

CC Operation

For CC operation, the current setting of each output must be programmed to the desired operating current. The sum of the voltage settings determines the voltage limit point. As an example, one way to program the voltage of the output is to set the voltage of each output to one half of the total voltage limit point. Then, at load voltages less than one half of the total voltage limit point, one output will operate in CC mode while the other output will be conducting through its internal reverse voltage protection diode. At load voltages greater than one half the total voltage limit point, the output that was in

CC mode will change to CV mode while the output that was conducting through its diode will regulate the current in CC mode and provide the balance of the voltage required by the load. Note that the total load voltage can be found by adding the results of reading back the individual series outputs only when neither reverse voltage protection diode is conducting.

When this diode is conducting, the corresponding output has reverse voltage across it so that its voltage readback may not be accurate.

When an output is conducting through its reverse voltage protection diode, the output will have a reverse voltage across its output terminals with the - V terminal more positive than the + V terminal. This voltage will be maximum at the rated current of the output. (See Figure 4-2 for reverse diode characteristic). Note that when an output conducts through this diode, it will indicate CC mode even though it is not regulating the current or voltage. Also, note that the voltage readback is not specified to indicate negative voltages although it will operate down to a limit of about - 0.176 V on the 25 W outputs and -0.392 V on the 50 W outputs. These values will still be indicated even if the actual voltage is more negative.

Remote Sensing

If it is necessary to remote voltage sense at the load, connect the sense leads of output 1 and output 2 as shown in Figure 4-

14. Note that the + sense lead of output 2 must remain connected to the -sense terminal of output 1. The outputs may be set as previously described.

Figure 4-14 Series Connections with Remote Sensing

Specifications for Series Operation

Specifications for outputs operating in series can be obtained from the specifications for single outputs. Most specifications are expressed as a constant or a percentage (or ppm) plus a constant. For series operation, the percentage portion remains unchanged while constant portions or any constants are changed as indicated below.

60 Output Connections and Operating Information

Voltage

Current

All series specifications referring to voltage are twice the single output specification except for programming resolution which is the same as for a single output.

All series specifications referring to current are the same as for a single output except for CC load effect, CC load cross regulation, CC source effect, and CC short term drift which are twice the current programming accuracy (including the percentage portion).

Load Transient

Recovery Time

Load transient recovery time is the same to within approximately twice the voltage setting band since the output impedances of the series combination add together.

Bi-Polar Operation

Figure 4-15 shows an example connecting the outputs of the power supply for bi-polar operation to test power FET’s

(PFET).

Great flexibility is achieved with this setup. The outputs of the power supply can be programmed as voltage or current sources providing positive or negative voltages or currents. By taking measurements via internal readback circuitry in the power supply, the outputs can be used as voltmeters or ammeters.

In order to properly measure the parameters of the PFET, it is desirable to be able to directly regulate any two of the bias parameters of the PFET, while being able to measure a third. The setup shown in Figure 4-15 allows control of Vds and Ids while measuring Vgs.

In this example, outputs 3 and 4 are connected to allow bipolar operation. Notice that the - S and - V leads, of outputs 3 and

4, are strapped together. The + S lead of V3 senses the voltage at the drain of the Device Under Test (DUT) and the + S lead of output 4 senses the voltage at the source of the DUT. Outputs 1 and 2 are strapped to sense the gate to source voltage.

Figure 4-15. Bipolar Operation

Output Connections and Operating Information 61

5

Remote Operation

Introduction

Chapter 3 introduced you to the basics of remote operation and provided a few simple examples using a Series 200/300 computer as the GP-IB controller. This chapter contains all the information required to control your power supply remotely and discusses in greater detail how each of the commands can be implemented. The material covered is intended for any controller capable of using the GP-IB interface functions mentioned in Interface Functions on page 63.

Four major sub-sections are discussed. These are:

1. GP-IB Operation

2. Programming Syntax

3. Initial Conditions

4. Power Supply Commands

The GP-IB section briefly describes the GP-IB interface functions to get you acquainted with remote programming using the GP-IB. Under Programming Syntax, the syntax of all commands, the numeric data formats and the programmable ranges for all models are given. Initial Conditions highlights the initial values of all the parameters at power-on. Power

Supply Commands will describe all the commands which can be used to program the supply’s functions including status reporting, error handling, protection features, and voltage and current programming. The power supply commands are summarized in Appendix C.

The examples are written in a generic form to make adaptation to your controller easy. You should be familiar with the use of your controller and its instruction set and how the power supply commands can be incorporated in your controller commands. If you are not familiar with the operation of the power supply, you are advised to read through Chapters 2 through 4 first.

GP-IB Operation

Interface Functions

Remote control is implemented by the GP-IB. It enables instructions to be sent from an external computer equipped with a

GP-IB interface. The power supply implements the following IEEE-488 Interface Functions:

SH1 - Source Handshake

AH1- Acceptor Handshake

SR1 - Service Request

RL1 - Remote/Local

T6 - Talker

L4 - Listener

RL1 - Parallel Poll

DC1 - Device Clear

The source handshake, acceptor handshake, talker and listener functions are implemented by the interface circuits of the power supply and the controller. The ADDR annunciator indicates when the power supply is addressed to listen or talk.

(The talker function includes the Serial Poll, see page 64).

Service Request . This is a message which can be initiated by the power supply to request service from the controller.

When the supply is requesting service, it asserts the service request (SRQ) line on the GP-IB to interrupt the controller providing the controller is configured to service interrupts. A service request can be generated whenever there is a fault on one of the outputs (up to 4 outputs), a programming error has occurred, or at power on providing certain commands are sent. Service request commands are discussed in detail in Service Request Generation on page 79.

The SRQ annunciator on the front panel display is turned on when the power supply is requesting service from the computer and remains on until the controller conducts a serial poll. A serial poll removes the service request and turns off

Remote Operation 63

the SRQ annunciator regardless of whether the condition that caused the service request continues to exist. The service request is also removed when you send the "CLR’’ command (see page 76).

Remote/Local . The power supply can receive programming information either from the GP-IB (remote) or from the front panel (local). When the power supply is in remote, the state of the supply cannot be changed by using the front panel keys, although the LCL key will remain enabled. Remote operation takes precedence over local operation, hence if the supply is accepting commands remotely and you attempt to change it to local operation, the supply will not allow any local settings and will remain in remote. You can prevent the front panel from sending programming information by sending the local lockout command. This command is sent only from the GP-IB. If you change from local to remote or vice-versa, there will be no change in the programmed settings.

Parallel Poll . Parallel Poll allows the controller to receive at the same time one bit of data from each of up to eight instruments connected to the bus. Agilent power supplies designate bit #6, the RQS bit of the serial poll register for this operation. By checking the status of this bit, the computer can quickly determine which instruments on the bus requested service. Once an instrument is identified, the computer can perform a serial poll to find out the exact cause of the request.

Parallel Poll does not reset this service request bit (RQS) in the power supply.

Note

IEEE-488 does not define what data an instrument should put on a bus in response to parallel poll. Many instruments such as Agilent Technologies power supplies indicate the state of their RQS bit, but the operator should not assume that all instruments on the bus respond to parallel poll with their RQS bit.

Unless remotely configured, the power supply will respond with a 1 on one of the GP-IB data lines if it is requesting service and its address is between 0 and 7. Addresses 0 through 7 define which data line (1 through 8) the supply will respond on. If the address is set to 8 or greater, the supply will not respond unless remotely configured.

The power supply may be remotely configured to respond with a 0 or 1, on any of the data lines, to indicate that it is requesting service. This is done in accordance with IEEE-488 1978.

Serial Poll . In a serial poll, the controller polls each instrument on the bus one at time. The power supply responds by placing the contents of the eight-bit serial poll register on the GP-IB data lines. The Serial Poll register is discussed on page 78 and defines the function of each of the bits. After the serial poll, the service request is cleared and the SRQ annunciator at the front panel is reset (off). However, the condition that generated the service request may still be present.

See page 79.

Device Clear . The Device clear command is typically used in systems to send all devices in the system to a known state with a single command. It may be implemented as an addressed or an unaddressed command. The power supply CLR command performs the same function as Device Clear (see page 76).

GP-IB Address Selection

You can find out the present address or change the address of the supply by using the front panel ADDR key as described in Chapter 3. Any address 0 through 30 is a valid address. If you program an address outside this range you will get a number range error.

Note

Care should be taken to not select the controller address.

64 Remote Operation

Power-On Service Request (PON)

The power supply can request service from the controller when the power is turned on. This request can be enabled or disabled by sending a PON command (see page 80). When the request is enabled, the supply can generate an SRQ at power-on or when there is a momentary loss in power. You can execute a serial poll to clear the service request. Table 5-7 details the conditions under which a PON command will generate an SRQ.

Note

The power supply has a non-volatile memory in which it stores certain system variables. Some of these variables are the calibration constants, the present supply address, and the present setting of the PON command.

Programming Syntax

The following paragraphs describe the syntax of the device command that is used to program your power supply. As shown in Figure 5-1, the device command is a specific part of the program statement that your computer will accept. The first part of the statement is computer as well as programming language specific. Figure 5-1 shows the structure of a typical programming statement for an Agilent Series 200/300 computer. If you are using a different computer or programming language, refer to your computer programming manual to determine the correct syntax for this portion of the program statement. This section of the manual is only concerned with the device command portion (the part inside the quotes for

Series 200/300 computers with BASIC) of the program statement.

Figure 5-1. Typical Program Statement for Series 200 Computers

Command

Query accumulated status register

Return power supply to turn on state

Turn calibration mode on or off

Query if calibration mode is on

Set the state of outputs at power on

Query outputs power on state

Set the reprogramming delay time

Query the setting of the delay time

Turn front panel display on or off

Display a string on front panel

Query if front panel display is on

Table 5-1. Power Supply Commands

Header

ASTS?

Output

Channel

1, 2, 3, 4

CLR

CMODE

---

---

CMODE?

DCPON

1, 2, 3, 4

---

DCPON?

DLY

DLY?

DSP

DSP

DSP?

1, 2, 3, 4

1, 2, 3, 4

---

---

---

---

0,1 (OFF,ON)

See Table 5-4

0,1 (OFF, ON) CC+

2,3 (OFF, ON) CC-

---

0-32

Data Range

(Fig. 5-2)

---

---

---

0,1(OFF,ON)

’’STRING’’

---

Syntax

Q2

C1

C2

Q1

C4

Q2

C2

C6

Q1

Remote Operation 65

Command

Query present hardware error

Query fault register

Query the model number of supply

Table 5-1. Power Supply Commands (continued)

Header

ERR?

Output

Channel

---

Data Range

(Fig. 5-2)

See Table 5-8

FAULT?

ID?

1, 2, 3, 4

---

---

---

Program the I DAC in counts

Query setting of I DAC in counts

Send data to calibrate I circuits

Sets output to high I cal. value

IDAC

IDAC?

IDATA

IHI

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

See Service Manual

---

See Table 5-4

See Table 5-4

Sets output to low I cal. value

Query measured I output

Used to Calibrate I readback circuits

Set output to + I readback high cal value

Set output to - I readback high cal value

Set output to - I readback low cal value

Set output to + I readback low cal value

Set full scale current range of output

ILO

IOUT?

IRDAT

IRHI

IRHN

IRLN

IRLO

IRSET

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

See Table 5-4

See Table 5-4

See Table 5-4

See Table 5-4

See Table 5-4

See Table 5-4

See Table 5-4

See Table 5-4

Query full scale current range

Set current of an output

Query current of an output

Increase or decrease output current by value

Select which output will be metered

Query which output is being metered

Send data to calibrate -I readback

Enable overcurrent protection

Query if OCP is enabled

Reset overcurrent protection

Turn output on or off

Query if output is on or off

Perform overvoltage calibration

Reset overvoltage circuit

Set overvoltage trip value

Query overvoltage trip value

Enable power on service request

Query if PON is enabled

Recall voltage and current settings

Set readback DAC to a number of counts

Query readback DAC count setting

Query revision date of ROM

Query revision date of secondary ROM

Set causes for generating a service request

Query causes which will generate an SRQ

METER?

METER

ISTEP

NIDAT

OCRST

ISET?

ISET?

ISET

OCP

OCP?

OUT

OUT?

OVCAL

OVRST

OVSET

OVSET?

PON

PON?

RCL

RDAC

RDAC?

ROM?

SROM?

SRQ

SRQ?

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

---

---

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

---

1, 2, 3, 4

---

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

---

---

0,1 (OFF,ON)

---

0-10

See Service Manual

---

---

---

01,2,3

---

See Table 5-4

See Table 5-4

See Table 5-4

See Table 5-4

1-4

---

See Table 5-4

0,1 (OFF,ON)

---

---

0,1 (OFF,ON)

---

See Table 5-4

---

See Table 5-4

See Table 5-4

Syntax

C3

C3

C4

Q2

Q2

C3

C4

Q2

C2

Q1

C5

C4

Q2

C4

Q2

C4

Q2

Q1

Q2

C2

Q1

C2

Q1

C2

C4

C3

C3

C3

C4

C3

Q2

C5

C3

Q1

Q2

Q1

C4

Q2

C5

C3

66 Remote Operation

Command

Store present output state

Table 5-1. Power Supply Commands (continued)

Header

STO

Output

Channel

Data Range

(Fig. 5-2)

0-10

Query preset status of output

Perform self test on GP-IB interface

STS?

TEST?

1, 2, 3, 4

.

Set bits in mask register

Query bits set in mask register

Program the voltage DAC in counts

Querys setting of voltage DAC in counts

UNMASK

UNMASK?

VDAC

VDAC?

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

0-255

See Service Manual

Send data to calibrate the voltage circuits

Set output to high V calibration value

Set output to low V calibration value

Query inputs of analog multiplexer

Query measured value of an output

Calibrate the voltage readback circuitry

Set output to V readback high cal value

Set output to V readback low cal value

VDATA

VHI

VLO

VMUX?

VOUT?

VRDAT

VRHI

VRLO

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1 2, 3, 4

1, 2, 3, 4

See Table 5-4

See Table 5-4

See Table 5-4 V

1-18

See Table 5-4

See Table 5-4

See Table 5-4

See Table 5-4

Syntax

Set full scale voltage programming range

Query full scale voltage programming range

Set output voltage

Query setting of output voltage

VRSET

VRSET?

VSET

VSET?

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

1, 2, 3, 4

See Table 5-4

See Table 5-4

See Table 5-4

See Table 5-4

C4

Q2

Q2

Q2

Increase or decrease output voltage by value VSTEP 1, 2, 3, 4 See Table 5-4 C4

Figure 5-2 shows the possible syntax forms for the device commands that are used to program the power supply. Syntax forms for the calibration commands that are discussed in Appendix A are also included. The oblong shape at the left of the syntax forms contains the command header which must be entered as shown in Tables 5-1 and 5-2. Commands are accepted in either uppercase or lowercase letters (ASCII characters). Circles contain characters that must be entered exactly as shown. Characters such as a space <SP> or a comma are used to separate elements in the command string. Characters such as a line feed < LF > or a semicolon are used to terminate the command string. Rectangles contain parameters that follow the command header lines and arrows indicate the correct paths through the syntax diagrams.

Numeric Data

The power supply will accept numeric data in implicit point, explicit point, or scientific notation. A general syntax diagram for numeric data is included in Figure 5-2. Implicit point notation means that numbers do not contain a decimal point; integers for example. Numbers written in explicit notation contain a decimal point, such as 12.35. In scientific notation, the letter E stands for "10 raised to". For example, 1.2E3 is read as 1.2 times 10 raised to the 3rd power, which equals 1,200.

Plus and minus signs are considered numeric characters and are optional. If you program a number with an accuracy that is greater than the resolution of the supply, the number will automatically be rounded to the nearest multiple of the power supply’s resolution. Table 5-1 gives the ranges for numeric data that is sent to the supply.

The power supply will also return numeric data (ASCII characters) to your computer. The format of the numbers returned depends upon the type of data requested. Table 5-2 gives the format for data returned to the computer in response to any of the queries that are listed.

Q2

C5

C3

C3

C5

C3

C3

C4

C2

Q2

C1

C4

Q2

C4

Q2

Remote Operation 67

Figure 5-2 (Sheet 1 of 2). Syntax Forms For Power Supply Commands

68 Remote Operation

Figure 5-2 (Sheet 2 of 2). Syntax Forms For Power Supply Commands

Remote Operation 69

Query Header

(Note 7)

Table 5-2. Power Supply Queries

Channel

(Note 1)

Response

(Notes 5 and 6)

Initial Value

Voltage Setting

Current Setting

Full Scale Current Range

Full Scale Voltage Range

Voltage Output

Current Output

OVP Setting

OC Protection On/Off

Output On/Off

Unmask Setting

Delay Setting

Status

Accumulated Status

Fault

Error

Service Request Setting

Power-On SRQ On/Off

Display On/Off

Model Number

VSET?

ISET?

IRSET?

VRSET?

VOUT?

IOUT?

OVSET?

OCP?

OUT?

UNMASK?

DLY?

STS?

ASTS?

FAULT?

ERR?

SRQ?

PON?

DSP?

ID?

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

1,2,3,4

--

--

--

--

--

SZD.DDD(Note 3)

(Note 2)

(Note 2)

ZD.DDD (Note 8)

SZD.DDD (Note 3)

(Note 2)

SZZD.DD

ZZD

ZZD

ZZD

<sp>ZD.DDD

ZZD

ZD

ZZD

ZZD

ZZD

ZZD

ZZD

Agilent 662XA (Note

4)

0(Note 8)

10 mA (Note 8)

High (Note 8)

High (Note 8)

--

--

55 V (Note 8)

--

--

0 (Note 8)

.02 (Note 8)

--

--

--

--

0 (OFF)

0 (OFF) (Note 9)

1 (ON)

--

Selftest

Calibration Mode

DC Power On

TEST?

CMODE?

DCPON?

--

--

--

ZZD

ZZD

ZZD

--

0 (OFF)

1 (ON) (Note 9)

Q1

Q1

Q1

S = Sign Z = Digit with leading zeros put out as spaces D = Digit < sp > = space

NOTES:

1. Output channels 3 and 4 are not used in all modes. (See Table 5-4).

2. Current is SZD.DDDDD (0.5 A) & SZD.DDDDDD (15 mA) SZD.DDDD(2 A) & SZD.DDDDD

(0.2 A)

3. Voltage can be SZD.DDDD on 7 VR

4. "X" depends upon model.

5. A space is returned for a + sign.

6. All responses are followed by a <CR> and <LF> (EOI asserted with <LF>).

7. Spaces are allowed between the header and the question mark.

8. Unit powers up with values in state register 0. Table 5-3 gives initial factory settings.

9. Factory setting. Powers up to last stored value.

Order of Execution

When you send a set of instructions to the power supply, they are executed in the order in which they are received. The power supply completes the execution of the present command before executing another command. To send more than one command within the power supply command string, use a semicolon to separate the commands. This maximizes the rate at which the power supply accepts commands.

Terminators

Terminators mark the end of a command string. As shown in Figure 5-2, the semicolon, line feed < LF >, and carriage return line feed < CR > < LF > are the characters that indicate the end of a message to the power supply. When you are using the Agilent Series 200 computer with BASIC to send a command using the standard format (see Figure 5-1), the computer automatically sends < CR > < LF > on the data bus following the command.

Ql

Q1

Q1

Q1

Q1

Q2

Q2

Q2

Q2

Q2

Syntax

(Fig.

5-2)

Q2

Q2

Q2

Q2

Q2

Q2

Q2

Q2

Q2

70 Remote Operation

Initial Conditions

Immediately after power on from the factory, the power supply automatically undergoes a self-test and sets all parameters to the values contained in Table 5-3. The values in the first part of the table come from storage register 0 and were stored at the factory. They will remain until STO 0 is used to change settings of the power on state. When the power supply is again turned on, it will initialize its outputs to the values in storage register 0.

Table 5-3. Initial Conditions

Inital Value Parameter

Voltage

Current

Reprogramming Delay

Store/Recall registers:

0-10: Voltage

Current

Over Voltage (OV)

Over Current Protection (OCP)

UNMASK register

SRQ

Output Channels

0 V High Range

10 mA High Range

20 mS

0 V High Range

10 mA High Range

55 V (High Range)

Disabled

0 (cleared)

Front Panel Metering

Power Supply Address

Local Control

PON bit

PON SRQ

Cal Mode

0 (Off)

On

Output #1

Last stored value (Factory set to 5)

Enabled

On

Off (0)

Off

Power Supply Commands

This section discusses the commands which you will use to program the supply’s voltage and current, protection circuits, and enhanced features like storage and recall registers, and reprogramming delay. When programming, you should be aware that the current, voltage and overvoltage ranges for each output of your supply may differ. Table 5-1 shows these values for the power supply. If you send values out of these ranges, you will get a number range error. A summary of all commands appears in Appendix C.

The output voltage of some output channels exceeds the safe operating limit of 42.2 V. To avoid any electrical shock, program the voltage to zero volts or turn off ac input power before changing any rear panel connections. Make certain all straps are properly connected, terminal block screws are securely

tightened and terminal block covers are replaced before reapplying power.

Voltage Programming

To program voltage, send the output channel and the programmed value. In the example below, output l is programmed to 5

V.

VSET 1,5

The values you send must always be volts. For example if you want to program 450 millivolts, convert to volts and then send the command:

Remote Operation 71

VSET 1,.45

If the output channel is operating in constant voltage mode (CV annunciator on) then the actual voltage is the programmed voltage, but in CC mode of operation (CC annunciator on), the programmed voltage is the voltage limit for that output.

When programming a value of voltage, the current setpoint will be changed if the total power is greater than the power boundary (see Figure 4-2).

Example: ISET 4,1.5 sets current to 1.5 A

VSET 4,50 sets voltage to 50 V

Sending the VSET command caused the current ISET to be reduced to 1.03 A. This is an example of power limiting changing a supply setting. When this occurs, the "couple parameter" bit (see Table 5-5) in the status register is set.

Model

Output Configuration

Table 5-4. Programmable Output Range

Operating

Range *

Output Volt

(Avg. Res.)

Output Curr **

(Avg. Res.)

6625A / 6628A

OUTPUT

1

6626A / 6629A

OUTPUTS

1 & 2

Low

High

0-7 V

(0.460 mV)

0-50 V

(3.2 mV)

0-15 mA

(1

µ

A)

0-0.5 A

(33

µ

A)

OUTPUT OUTPUTS Low 0-16 V

(1.0 mV)

0.2 A

(13

µ

A)

2

(1,2-6628A)

3-4

1,2,3,4-6629A)

High 0-50 V

(3.2 mV)

0-2 A

(131

µ

A)

* The maximum programmable voltage values for each range are 1% higher than the rated voltage and the maximum programmable current values for each range are 3% higher than the rated current.

** Each channel wakes up to values stored in state register 0. The factory set values are shown in the first part of Table 5-3. All other registers are set to these settings. Volatile registers 4-10 always power-up to these conditions.

To readback the programmed voltage setting for output 1, send the query:

VSET? 1 and address the supply to talk. If you want to know the value of the actual output voltage of output 1, send the query:

VOUT? 1

The results are placed on the GP-IB and read into the controller when the supply is addressed to talk.

Note

The power supply will round the VSET and ISET settings to the nearest multiple of their resolution.

Table 5-4 lists the average resolution of these settings.

72 Remote Operation

Current Programming

To program the current, send the output channel and the programmed value in amps. In the example below, output 1 is programmed to 15 mA.

ISET 1,0.015

The value you send must always be in amps. For example if you want to program 95 milliamps, convert to amps and then send the command

ISET 1,.095

If the output channel is in constant current (CC) mode of operation, then the actual current is the programmed current but if the output is in the CV mode, the programmed current is the current limit of that output.

When programming a value of current, the voltage set point can be changed if the total power is greater then the power boundary (see Figure 4-2).

Example: VSET 4,50 Voltage set to 50 V

ISET 4,2 ISET to 2 A

Sending the ISET command caused the VSET to be reduced to 16.16 V. This is an example where power limiting changes a supply setting. When this occurs, the "coupled parameter" bit (see Table 5-5) in the status register is set.

To readback the programmed current for output 1, send the query and addressing the supply to talk.

ISET?

You can also instruct the supply to measure the actual output current at output channel 1 by sending the following query and address the supply to talk.

IOUT? 1

The results are placed on the GP-IB and read into the controller.

Range Programming

The range can be set by using the VRSET or IRSET commands. The power supply will automatically pick the range that the value sent will fit into.

Examples: VRSET 1,7 will select the 7 V range

VRSET 1,3.2 will select the 7 V range

VRSET 1,9.0 will select the 50 V range

VRSET 1,50.5 will select the 50 V range

IRSET 1,.015 will select the 15 mA range

IRSET 1,0 will select the 15 mA range

IRSET 1,.020 will select the 0.5 A range

IRSET 1,0.1 will select the 0.5 A range

Note

Ranges will accept values up to 1% greater then full scale for voltage and 3 % for current. Larger numbers will generate a number range error(s).

When going from the high range to the low range, the present setting must be within the boundary of the low range or the setting will be reduced to the maximum of the low range. If the setting is changed the "couple parameter" bit (see Table

5-5) in the status register will be set to indicate the change.

Remote Operation 73

Output On/Off

The OUT command disables/enables an output channel of the power supply. It will not disturb any other programmed function nor will it reset the protection circuits. You can control individual outputs with the OUT command as shown below. For example, to disable output channel 1 send the following:

OUT 1,0

To enable output channel 1 send the following command

OUT 1,1

You can find out the present state of output 1 by sending the query:

OUT? 1 and addressing the supply to talk. The response from the supply is either a "0" to indicate output 1 is off or a "1" to indicate that the output is on. When disabled, the output behaves as if it were programmed to zero volts on the high voltage range and 10 mA on the high current range.

Overvoltage (OV) Protection

The programmable OV is a protection feature which can be set by the operator to protect the load against excessive voltage.

When the actual voltage exceeds the programmed overvoltage setting for a given output channel, the OV is tripped. The

OV circuit will fire the SCR crowbar which shorts across the output and the output assumes a low impedance state.

For example, to program the OV of output channel 1 to 9.5 V send the following command:

OVSET 1,9.5

To find out the OV setting for output channel 1 send the following query and address the supply to talk:

OVSET? 1

To enable an output after it went into overvoltage, you must first remove the overvoltage condition and then send the OV reset command.

To reset output 1 send:

OVRST 1

If you send the reset command without first removing the OV condition, the supply will fire the OV again.

Note

If the programmable OV fails, the supply has a fixed OV circuit which will fire the SCR crowbar if the voltage exceeds 120% of the maximum rated output. The fixed OV circuit will also fire the SCR crowbar if the supply is off (line cord disconnected) and an external source is supplying voltage which exceeds

120% of the maximum rated output.

74 Remote Operation

Overcurrent Protection (OCP)

The OCP is a protection feature employed by the power supply to guard against excessive output currents. When the output enters the + CC mode and the OCP is enabled, the OCP circuit down programs the output voltage and disables the output.

To enable the OCP, for output channel 1, send the command

OCP 1,1

To disable the OCP, send the command

OCP 1,0

You can find out the OCP setting by sending the following query and addressing the power supply to talk.

OCP? 1

The response from the power supply is either a "0’’ to indicate that OCP is off or a ’’1" to indicate that it is on. To reset the output channel after an overcurrent trip, you can either disable the OCP and send the reset command, or you can reduce the output current below the programmed current and then send the reset command. To reset output 1, send the command:

OCRST 1

Note

The supply can report a fault condition when an output is in overvoltage or overcurrent. Although the

OVRST and OCRST commands re-enable the output, they do not clear the fault register. As a housekeeping measure, it is advisable to always clear the fault register by querying its value after an OV or OC reset.

Multiple Output Storage & Recall

The power supply has 11 internal registers (0-10) each of which can store the voltage and current settings of all the outputs.

By storing voltage, voltage range, current, current range and OV settings for all outputs and recalling them later, you can have significant savings in programming time. (See Supplemental Characteristics in Table 1-2). Upon shipment from the factory, registers (1-10) contain values of 0 volts and 0 current (see Table 5-4).

Registers 0-3 are non-volatile. They are limited to one STORE each. If more than one STORE is attempted, error 30

(STORE LIMIT) will result. The line must be turned off/on to re-enable these store locations. Register 0 is a special case that also contains Delay, Mask. When the power supply is turned on, it will access register 0 for power-up settings.

Registers 4-10 (volatile) will be reinitialized to 0 volts and 10 mA (max OV & V, I ranges = HIGH) when power is cycled on/off.

To store voltage and current settings, issue the store (STO) command and specify a register (0-10). For example, to store the present setting of current and voltage of all outputs in register 2, send the following command:

STO 2

To set the power supply to the voltage, voltage range, current, current range, and overvoltage settings stored in a register, issue the recall (RCL) command followed by the desired register. For example to recall settings stored in register 2, send the following command:

RCL 2

When a register is recalled, the outputs will be set sequentially (output 1, output 2, etc.). If you attempt to recall a register outside the 0 to 10 range, you will get a number range error (GP-IB code 5 in Table 5-8).

Remote Operation 75

The Clear Command

This command will return the power supply to its power-on state and all parameters are returned to their initial power-on values except for the following:

1. The store/recall registers are not cleared.

2. The power supply remains addressed to listen.

3. The PON bit in the serial poll register is cleared

To clear the power supply, send the following command:

CLR

Status Reporting

The power supply has the ability to report its internal status to the user whenever it is asked to do so. Depending on the type of status the user requested, the supply will interrogate the status, accumulated status, mask, or fault registers present in each output. The status register can report status independently (see page 76) or it can work together with the mask and fault registers to report a fault (see page 77). The accumulated status register records every status condition the output experienced since the time it was last read (see page 77). Figure 5-3 shows a conceptual model of the operation of these registers.

Figure 5-3. Functional Relationship of Status Registers

The supply has one serial poll register which services all outputs and provides the user with other power supply statusrelated information as discussed in page 78.

Status Register . Each output channel of the power supply maintains its present status in an eight bit register. This status register reports the status of the output channel whenever it is queried. A "1" in any of the bit positions indicates that the condition is true. As long as the condition continues to be true, the bit will remain set. Assignments for the bits are shown in

Table 5-5.

Decoding of the reading is based on the weighted number placed on each bit of the eight bit status registers. For example, bit position 5 in the register has a bit weight of 32 (see Table 5-5). Each bit is assigned to a particular condition and the corresponding bit weight is used to identify that condition. When set, bit 5 indicates that the associated output is in the unregulated state. If this is the only bit that is set, the number 32 will be returned when the output’s status register is queried.

76 Remote Operation

Bit Position

Bit Weight

Meaning

Table 5-5. Bit Assignment for the Status, Astatus, Fault, and Mask Registers

7

128

CP

6

64

OC

5

32

UNR

4

16

OT

3

8

OV

2

4

-CC

1

2

+CC

Where

CV =

+ CC =

- CC =

OV =

OT =

UNR =

Constant Voltage Mode

Positive Constant Current Mode

Negative Current Limit Mode

Overvoltage Protection circuit tripped

OC =

CP =

Over Temperature Protection circuit tripped

Unregulated Mode

Over Current Protection tripped

Coupled parameter (See Note)

Note: When the range is switched as discussed on page 73, or the power limit is exceeded (see pages 71 &

73), the CP bit is set. It is cleared when you send a voltage, current or range value that causes no other

0

1

CV changes.

To query an output channel for its status, you must specify the output channel. For example, to find out the status at output

2 send the following query and address the supply to talk:

STS? 2

Accumulated Status Register . Each output channel of the power supply also maintains a cumulative status in its accumulated status (astatus) register. This register records every status condition the power supply output entered since it was last queried. When queried, it returns a decimal number which is decoded as shown below. The astatus register is reset to the present value of the status register after it is queried. The bits are assigned as in Table 5-5. Here is an example to help you decode the decimal number (from 0 to 255) returned when the astatus register is queried. If the output channel was in overvoltage since the last reading of the astatus register and that channel is presently operating in constant voltage mode, the reading you will get when you query the register will be 9. To decode this we use Table 5-5.

9 = 8 + 1

OV + CV

For example, to query the astatus register of output 2, send the following query and address the supply to talk

ASTS? 2

The Mask and Fault Register . The fault register works in conjunction with the mask register. These are two eight bit registers which report any fault condition on a particular output channel. The mask register is used to set up the conditions that generate a fault which is latched into the fault register. The user can then read the fault register to determine the fault.

When a bit in the fault register is set, the power supply can generate a service request for that output providing the service request command on fault (SRQ 1 or SRQ 3) was previously sent. See page 79 for a discussion on service request.

To understand how these two registers work, we must include the status register in this discussion. Recall that the status register takes its input from the power supply and the user cannot change its contents. The mask register takes its inputs from the user, and the power supply cannot change its contents. The fault register takes its inputs from both the mask and the status registers. You can find out the setting of the mask register of output 2 by sending the following query and addressing the supply to talk:

UNMASK? 2

The response will be a numeric code between 0 and 255 which can be decoded by consulting Table 5-5. You can set the conditions to generate a fault by setting (unmasking) one or more bits in the mask register. The conditions will remain unmasked until you change them. To unmask conditions in output 2 for example, send the following command:

Remote Operation 77

UNMASK 2,XXX where XXX specifies the numeric code (0 to 255) for the unmasked conditions (see Table 5-5). If during operation, the output experiences any of the previously unmasked conditions, it will set the corresponding bit(s) in its fault register.

Remember that the bits in the fault register can be set when there is a change in either the status register or the mask register. Each output has its status, mask, and fault registers arranged as shown in Figure 5-3 and Table 5-5. The mask register, which is set by the user, is used to specify which bits in the status register are enabled (unmasked) to set bits in the fault register. A bit is set in the fault register when the corresponding bit in the status register changes from ’’0’’ to ’’1’’ and the corresponding bit in the mask register is a "1". Also, if a bit in the status register is already set and then the corresponding bit in the mask register is set (unmasked), the corresponding bit in the fault register will be set.

In addition, if both status and mask register bits remain set after the fault register was read (and cleared), the fault register will remain cleared as long as there are no changes in either the status or mask registers with the following exception.

Executing a VSET, ISET, RCL, OVRST, OCRST, or OUT on/off command, will cause the CV, + CC, - CC, or UNR bit

(as applicable) in the fault register to be set. Note that the fault register is cleared immediately after it is read.

As shown in Figure 5-3, if one or more bits in the fault register of a given output channel are set, then the FAU bit for that output in the serial poll register will also be set and a service request may be generated (see page 79). To read the fault register of output 2 and find out which bits are set, send the following query and address the supply to talk:

FAULT? 2

The power supply responds with a number which can be decoded from Table 5-5. For example, the number 9 (8 + 1) indicates that the OV and the CV bits in the fault register are set.

Note

If the condition(s) generating the fault(s) is (are) removed but the fault register is not read, the bit(s) in the fault register will remain set.

The Serial Poll Register . The serial poll register is an 8 bit register which the supply uses to keep track of its internal operating status and to determine the operating status of each of its outputs. Table 5-6 defines each bit.

Bit Position 7

Table 5-6. Bit assignment of the Serial Poll Register

6 5 4 3 2 1 0

Bit Weight

Meaning

128

PON

64

RQS

32

ERR

16

RDY

8

FAU 4

4

FAU 3

2

FAU 2

1

FAU 1

The first four bits (0 to 3) in the register tell whether or not a particular output has a fault. If there is a fault in one of the outputs, then the corresponding FAU bit will be set. Thus if output 1 has a fault, then FAU 1 will be set. In models with only three outputs, FAU 4 will always be zero and in two output models, FAU 3 and FAU 4 will always be zero.

The RDY bit is set when processing is complete and is cleared when the supply is processing commands.

The ERR bit is set when a programming or hardware error occurs and is cleared when the error query (ERR?) is received.

The error annunciator on the front panel informs the user when this bit is set or cleared.

The RQS bit is set when the power supply generates a service request and cleared after a serial poll is done (see page 79).

The PON bit is set at power on and cleared when a CLR command is sent.

78 Remote Operation

Service Request Generation

When operating your supply, you may want it to request service every time a fault or a programming error condition occurs. To do this you send a service request (SRQ) command. When the condition is true, the power supply responds by setting the RQS bit in the serial poll register, setting the SRQ annunciator on the front panel, and issuing an SRQ over the

GP-IB.

The power supply can generate a service request for any of the following reasons: (refer to Table 5-7)

An Output Fault . If there is a fault on one or more of the output channels and you previously sent the SRQ 1 or

SRQ 3 command (see Service Request Enable/Disable information below), then an SRQ will be generated.

An Error . If there is an error (see Tables 5-8) and you previously sent the SRQ 2 or SRQ 3 command, (see

Service Request Enable/Disable information below), then the supply will generate a service request.

Power-on . At power-on, the PON bit of the serial poll register is set but the supply will only generate an SRQ if you previously sent a PON 1 command.

Input Line Voltage Dropout . Same as power-on condition.

To find out the nature of the service request, you must do a serial poll. This will isolate the output that generated the request by checking which of the FAU bits are set in the case of a fault, or checking to see if the error bit is set in the case of an error. If the SRQ on faults was set, then send the fault query.

FAULT? 2 (using output 2 as an example) and address the supply to talk if you want to find out which of the conditions you unmasked in Figure 5-3 are true. For example if the supply was in overvoltage and that condition was unmasked then the response from the fault query will be

"8’’ (see Table 5-5).

Note

When you query the fault, the fault register is cleared. Performing a serial poll will reset the RQS bit but will not clear the fault register.

If the SRQ on error was set then you can send the error query ERR? and address the supply to talk. The response will identify the error by its code (see Table 5-8).

Service Request Enable/Disable . You can query the status of the service request enable/disable function by sending the query:

SRQ?

and addressing the power supply to talk. The response from the supply is one of the following:

0, 1, 2, or 3

0--indicates that the service request capability (except for power-on; see page 80) is disabled.

1--indicates that it is enabled for output fault conditions.

2--indicates that it is enabled for error conditions.

3--indicates that it is enabled for both fault and error conditions.

Remote Operation 79

The ability to generate service requests can be enabled or disabled using the SRQ command as described below.

To disable the service request capability, except for power-on, send:

SRQ 0

To enable the service request capability for all output faults, send:

SRQ 1

To enable the service request capability for errors, send:

SRQ 2

To enable the service request capability for both faults and errors, send:

SRQ 3

The Power-On Service Request . You can also cause the power supply to request service every time it is switched on or every time there is a temporary loss in power. To do this send the following command:

PON 1

If you want to disable this facility, send the command.

PON 0

If you want to find out if the power-on SRQ is enabled or disabled, send the following query:

PON?

and address the supply to talk. The supply will respond with a 1 or 0 as discussed above.

Note

The power-on (PON) SRQ mode is stored in the non-volatile memory of the supply so that although the supply may be switched off, it will remember the status of the last PON command at power-on and respond accordingly.

Table 5-7 summarizes all the conditions under which a service request will be generated.

Table 5-7. Conditions for Generating a Service Request

Condition

Any

Power-on

Error

Fault

Commands Sent

PON

0

1

-

-

SRQ

0

-

2 or 3

1 or 3

State of

RQS Bit

0

1

1

1

Reprogramming Delay

The power supply may switch modes or become unregulated momentarily after a new output value is programmed.

Because of their short duration, these cases may not ordinarily be considered a fault but the supply will recognize this deviation and generate a fault signal. To prevent this, the reprogramming delay feature is implemented.

80 Remote Operation

Reprogramming delay will delay the onset of certain fault conditions and prevent the power supply from registering a fault when these conditions are true. When the delay is in effect, the CV, + CC, - CC and UNR bits of the status register are masked and cannot communicate with the mask and fault registers and the OCP function. This will prevent the supply from registering a fault should any of these bits become set during the delay period. Reprogramming delay is initiated when any of the following functions are executed:

VSET, ISET, RCL, OVRST, OCRST, OUT on/off

At power-on reprogramming delay is set to 20 mS. You can specify new values between 0 and 32 S in steps of 4 mS. If you specify a value which is not a multiple of 4 mS, the supply will round off the set value to the nearest 4 mS multiple.

To program a new value of 80 mS in output 2 for example, send the following:

DLY 2,.08

If you send a value outside the 0 to 32 S range you will get a programming error. You may use the programming response times in the specifications table to give you an idea of a typical delay setting. However, the appropriate delay setting will also depend on load capacitance, load resistance, and current limit setting. See page 53 for output capacitor considerations.

To query the reprogramming delay setting of a particular output channel, send the following query:

DLY? 2 (using output 2 as an example) and address the supply to talk. The response will be a numeric value between 0 and 32.

Display On/Off

When the display is on, the commands sent across the GP-IB may experience a slower processing time because the processor must also spend time to monitor the outputs and update the display. You can shorten your command processing time by turning off the display. To turn off the display, send the command:

DSP 0

To re-enable the display send the command:

DSP 1

You can also find out the status of the display by sending the following query and addressing the supply to talk:

DSP?

The response will be either a "1’’ or a "0’’.

Message Display Capability. The display command can also be used to display messages on the front panel. Messages may consist of a maximum of twelve alphanumeric characters. Only upper case alpha characters, numbers, and spaces will be displayed.

For example, to display the message "OUTPUT 2 OK", send the following command:

DSP "OUTPUT 2 OK’’

Note

The BASIC programming statement for a series 200/300 computer would be as follows:

OUTPUT 705; "DSP""OUTPUT 2 OK"""

Remote Operation 81

Other Queries

In the examples discussed above, you saw how to use queries for each function discussed. The following paragraphs describe other queries which were not previously covered.

ERROR Query . The power supply can detect both programming and hardware errors. You can use either the front panel

(see page 89) or the GP-IB to find out the type of error. Upon detecting an error, the error annunciator on the front panel and the ERR bit in the serial poll register will be set. When in local mode the supply will display the error name in response to pressing the ERR key on the front panel. Over the GP-IB, only the error code will be returned. After a query, the error bit is cleared. A description of these codes is given in Tables 5-8. To find out what the error is, send the following query and address the supply to talk:

ERR?

The supply will respond with an error code number (see Table 5-8).

ID Query . If you want to know the model number of the power supply you are working with, you can send the ID? query over the GP-IB. To do this send the following over the GP-IB and address the supply to talk.

ID?

The supply will respond with its model number.

Test Query . You can get the power supply to perform a limited self-test at any time during its operation by sending the

TEST? query over the GP-IB. This test does not affect the analog control circuits of the supply and it can be performed while the outputs are connected to external circuits. For tests of the analog control circuits refer to Chapter 3. Responses to the test query are described in Table 5-9. This test cannot be done from the front panel. To instruct the power supply to carry out a self-test, send the following query and address the supply to talk:

TEST?

Calibration Mode Query . To be able to calibrate your power supply, the calibration mode (CMODE) must be turned on

(See Appendix A for a detailed description of the calibration procedure). To find out if the CMODE is on or off, send the following query over the GP-IB and address the supply to talk:

CMODE?

The supply will respond with a ’’1" which indicates that CMODE. is on or a "0" which indicates that CMODE is off.

DCPON.

The DCPON command sets the state of all outputs at power-on. You can specify if the outputs wake up enabled or disabled when the unit is turned on. To enable all outputs at turn-on send:

DCPON 1

To disable all outputs at turn-on send:

DCPON 0

Note that these commands set the output voltage to zero and the output current to a slightly positive value when the

OUTPUT OFF command is issued. Therefore, the constant voltage feedback loop is active and the outputs are in constant voltage mode when programmed “OFF”.

Starting with firmware revision A.00.04 and later, two additional commands let you set the output current to a slightly negative value when the OUPUT OFF command is issued, causing the outputs to be in constant current mode when programmed “OFF”. The benefit of being in constant current mode when the output is off is that if the load impedance and the voltage and current settings are such that the unit is forced into constant current mode at turn on, then the current feedback loop will be active during the transition to OUTPUT ON, and there will be no output current overshoot due to mode crossover.

82 Remote Operation

To enable all outputs in constant current mode at turn on send:

To disable all outputs in constant current mode at turn-on send:

Front Panel

Response

NO ERROR

INVALID CHAR

INVALID NUM

INVALID STR

SYNTAX ERROR

GP-IB

Code

0

1

2

3 or 28

4

DCPON 2

DCPON 3

Table 5-8. Error Messages

Explanation

This is the response to the ERR? query when there are no errors.

You sent the supply a character it did not recognize.

Format of your number is incorrect. Check number syntax.

Occurs when you send a command the supply does not understand.

Either too many parameters are sent without delimiters or the number representation is incorrect. Follow the Syntax Diagram in Figure 5-2. Check spaces and delimiters.

NUMBER RANGE

NO QUERY

5

6

DISP LENGTH

BUFFER FULL

EEPROM ERROR

HARDWARE ERR

7

8

9

10

An out of range number was sent. Send a new number within the legal range.

Computer addressed the supply to talk, but it did not first request data. Send query first then address supply to talk.

Quoted string exceeds the display length of 12 characters. Shorten string to a maximum of 12 characters.

This error may occur if too many numbers are sent. Error #4 or #5 may occur first.

EEPROM is not responding correctly to programming commands. An instrument failure has occurred and service is required.

An output error has occurred in an unknown output. Service is required.

Remote Operation 83

Front Panel

Response

HDW ERR CH 1

HDW ERR CH 2

HDW ERR CH 3

HDW ERR CH 4

NO MODEL NUM

CAL ERROR

UNCALIBRATED

CAL LOCKED

SKIP SELF TST

Code

0

20

21

27

29

30

GP-IB

Code

11

Table 5-8. Error Messages (continued)

Explanation

Errors 11 through 14 refer to a specific output where there is an output error.

Service is required.

12

13

14

15

Same as in Error #11.

Same as in Error #11.

Same as in Error #11.

The interface cannot find its model number. There may be a hardware failure or the instrument may require reprogramming. Service is required.

16

17

You tried to use either a calibration command with CMODE off or the calibration failed while in CMODE. Enable CMODE and check numbers sent during calibration. Also, there could be a hardware error.

There is an incorrect checksum in the EEPROM possibly as a result of incorrect calibration procedure. Recalibrate, and if the problem persists, your supply has a hardware failure.

18

22

Calibration was attempted with the Calibration Jumper on the GP-IB board in the lockout position. Reposition jumper if desired. See Service Manual.

The self test jumper on the GP-IB board is in the Skip Self Test position. No selftest was done. This is for diagnostics only. See Service Manual.

Table 5-9. TEST Responses

Explanation

This is the response to the TEST? query when there are no errors.

The timer has failed self-test. Refer to the troubleshooting section in the Service Manual.

The RAM has failed self-test. Refer to the troubleshooting section in the Service Manual.

The ROM has failed the checksum test. Refer to the troubleshooting section in the Service Manual.

There is an incorrect checksum in the EEPROM for non-volatile state storage. The non-volatile states 0-3 have not been loaded. Store new values, if problem persists, your supply has a hardware failure.

The store limit has been reached for CMODE, store states 0-3, DCPON, or SRQ. The power must be turned off/on to reenable.

84 Remote Operation

6

Local Operation

Introduction

Chapter 3 introduced you to the supply’s front panel controls and indicators to help you turn on the supply and perform the checkout procedures that were given in that chapter. The following paragraphs describe how to use all of the front panel controls and indicators. Most of the remote operations described in Chapter 5 can also be performed locally from the supply’s front panel.

Local Mode

In order to use the front panel keys to control the supply, the local mode must be in effect. The local mode is in effect immediately after power is applied. Table 5-3 lists the initial settings for all of the power supply’s functions when power is initially applied. When the local mode is in effect (RMT annunciator off), the Output Function, System Function, and

Numeric Entry keys on the front panel (see Figure 6-1) can be used to operate the power supply.

In the remote mode (front panel RMT annunciator on), the front panel keys will have no effect on any of the supply’s outputs and only the computer can control the supply. You can, however, still use the front panel display to view the output voltage and current readings or the present settings for the selected output channel while the supply is in the remote mode.

You can return the supply to the local mode from the remote mode by pressing the LCL key provided that the local lockout command has not been received from the GP-IB controller. Pressing the LCL key will also turn the supply’s display back on if it was turned off with a DSP command during remote operation (see page 81). A change between the local and remote modes will not result in a change in the power supply’s outputs.

Local Control Of Output Functions

The Output Function keys (see Figure 6-1) allow you to control the selected output. Figure 6-1 shows the annunciator arrow over OUTPUT 2 indicating that output channel 2 is selected. Pressing the OUTPUT SELECT key selects the output channels in forward (

>) or reverse (<) sequence. Note that Figure 6-1 illustrates the front panel for the Agilent 6626A supply which has four output annunciators. The front panel for Agilent Model 6625 is identical to Figure 6-1 except it has two output annunciators.

General

The power supply will accept programming values directly in volts and amps. The programmable voltage, current, and overvoltage ranges for the outputs of each model are given in Table 5-4. The power supply will round off the values received to the nearest multiple of the resolution for that particular output. If you send a value out of the valid range, it will not be executed and the ERR annunciator will come on. You can get a readout of the error on the display by pressing the

ERR key. For an out-of-range error, the error message ’’NUMBER RANGE" will be displayed.

When you press the VSET, ISET, or OVSET key, the output selected and the present setting for that function will be displayed. You can change the setting using the numeric entry keys. Pressing the number keys will cause the present numeric setting to become blank and be replaced with the new numbers on the display. You can use the

Ï key to erase previous keystrokes if you make a mistake.

Local Operation 85

OUTPUT

ANNUNCIATOR

CHANNEL 1

DISPLAY

6626A SYSTEM DC POWER SUPPLY

VOLTS AMPS

45.153 14.235m

1 2 3 4 CV CC UNR OCP ERR RMT ADDR SRQ

-- OUTPUT - ENBLD

LOCAL MODE

KEY

LCL

OUTPUT FUNCTION KEYS

SYSTEM

ADDR

ERR

STO

RCL

OUTPUT

RANGE

V/I

VOLT

OVSET VOLT

OUTPUT

SELECT

VSET

OCP

RESET

CURR

ISET

CURR

OUTPUT

ON/OFF

1

0

7

4

2

.

8

5

ENTRY

9

6

3 ENTER

LINE

ON

OFF

SYSTEM FUNCTION KEYS NUMERIC ENTRY KEYS

Figure 6-1. Front Panel (Model 6626A shown)

Pressing the ENTER key will enter the values displayed for the function indicated, initiate that function, and return the display to the metering mode in which the measured output voltage and current for the selected output are displayed.

Pressing the ENTER key without entering numbers will result in retention of the previous values and return to the metering mode. You can also return to the metering mode at anytime by pressing the RESET key.

The RANGE V/I key is used to display the voltage and current range settings. The key toggles; pressing it will display the range, pressing it again will put the display back to background metering. The range display shows the largest values of voltage and current that can be set for the ranges presently set.

The voltage step and current step (arrow keys) are used to change voltage & current settings and voltage & current ranges.

There are 3 modes:

1. To use the step keys to immediately change the output settings, the background metering mode must be in affect.

2. To use the step keys to change the setting, the VSET or ISET key must be pressed first. The step keys can then be used to adjust the setting. Press the ENTER key to change the actual setting.

3. To use the step keys to change the voltage or current ranges first press the RANGE V/I key. The step keys change the range as long as the RANGE display is in effect.

Note

The step keys change the voltage or current by 1 LSB each time they are pressed. If the key is held down more than 1 second, the automatic step mode goes into effect. The step rate will start to increase from the initial value after about 3 seconds.

Setting Voltage and Voltage Range

The selected output’s voltage is programmed locally using the VSET key. For example, program the voltage to 5.25 volts by pressing:

VSET 5 .

2 5 ENTER

The front panel display then indicates the actual output voltage and current for the selected output.

86 Local Operation

The voltage step keys can also be used to change the voltage setting.

VSET

ÑVOLT or

ÒVOLT then ENTER

The voltage can be set in the immediate execute mode. This mode is in effect when background metering is in effect. Pressing

ÑVOLT or

ÒVOLT will change the setting and the effect on the voltage and current can be observed on the display.

The voltage range can be changed by using the RANGE V/I key and the voltage step keys.

RANGE V/I

ÑVOLT or

ÒVOLT

The range will change as soon as one of the voltage step keys is pressed and the display will be updated. If the voltage setting is greater then the highest value of the low range and the range is changed from high to low, the VSET will be automatically reduced to the maximum voltage of the low range.

Setting Current and Current Range

The selected output’s current is programmed locally using the ISET key. For example, program the current to 1.5 amps by pressing:

ISET 1 .

5 ENTER

The power supply will accept any programmed current between zero and the minimum programmable current and automatically set the output to the minimum programmable current without causing a programming error. See Table 5-4.

The current step keys can also be used to change the current setting.

ISET

ÑCURR or

ÒCURR then ENTER

The current can be set in the immediate execute mode. This mode is in effect when background metering is in effect. Pressing

ÑCURR or

ÒCURR will change the setting and the effect on the voltage and current can be observed on the display.

The current range can be changed by using the RANGE V/I key and the current step keys.

RANGE V/I

ÑCURR or

ÒCURR

The range will change as soon as one of the current step keys is pressed and the display will be updated. If the current setting is greater then the highest value of the low range and the range is changed from high to low, the ISET will be automatically reduced to the maximum current of the low range.

Enabling/Disabling an Output

The selected output channel can be turned on and off from the front panel. The OUTPUT ON/OFF key toggles the selected output on and off. When an output is turned off, the message "DISABLED" will be displayed.

The OUTPUT ON/OFF key will not affect any other programmed functions nor will it reset an overvoltage or overcurrent condition. An output disabled by the OUTPUT ON/OFF key will behave as if it were programmed to zero volts and 10 mA.

Local Operation 87

Setting Overvoltage Protection

Programmable overvoltage protection (OVP) guards your load against overvoltage by crowbarring and downprogramming the power supply output if the programmed overvoltage setting is exceeded.

A fixed OV circuit with a trip level about 20 percent above the maximum programmable voltage acts as a backup to the programmable OVP. When overvoltage protection is activated, the output is shorted and the message ’’OVERVOLTAGE’’ will appear on the front panel display.

The selected output’s overvoltage setting is programmed locally using the OVSET key. For example, program the overvoltage to 10.5 volts by pressing:

OVSET 1 0 .

5 ENTER

Resetting Overvoltage Protection

The condition that caused the OVP must first be cleared and then the output can be returned to its previous state by pressing the RESET key.

Enabling/Disabling Overcurrent Protection

The overcurrent protection feature guards against excessive output currents. When the output goes into the + CC mode and

OCP is enabled, the OCP circuit is activated which downprograms the output voltage and disables the output. For this condition, the message ’’OVERCURRENT’’ appears on the front panel display.

The selected output’s overcurrent protection feature can be turned on and off from the front panel. The OCP key toggles the selected output’s overcurrent protection circuit on and off. When it is on (enabled), the OCP ENBLD annunciator will be on.

Resetting Overcurrent Protection

The condition that activated the OCP circuit must first be cleared and then the output can be returned to its previous state by pressing the RESET key.

Local Control Of System Functions

The System Function keys consist of the ADDR, ERR, STO, and RCL keys as shown in Figure 6-1. These keys are independent of the output selected and are used in setting the supply’s GP-IB address, displaying error messages, and storing/recalling voltage and current settings for all of the supply’s output channels.

Setting the Supply’s GP-IB Address

As described in page 40, before you can operate the supply remotely, you must know its GP-IB address. You can find this out locally from the front panel by pressing:

ADDR

The supply’s present address will appear in the display. Address 5 is the factory set address.

If you want to leave the address set at 5, you can return to the metering mode by pressing the METER key or you can press another function key.

If you want to change the address, you can enter a new value.

Any integer from 0 through 30 can be selected.

88 Local Operation

For example, you can change the address of your supply to 10 by pressing:

ADDR 1 0 ENTER

Displaying Error Messages

The power supply can detect both programming and hardware errors. Upon detecting an error, the ERR annunciator on the front panel comes on and the ERR bit in the serial poll register will be set (see page 78).

When an error is detected, you can display the error message by pressing the ERR key. The power supply will return the error message to the display and clear the error bit. For example, if you program a number that is not within the valid range, the ERR annunciator will come on. You can display the error message by pressing the ERR key. In this case, the error message "NUMBER RANGE" will be displayed. Errors generated either locally from the front panel or remotely from the

GP-IB computer can be displayed by pressing the ERR key only when the supply is in the local mode. Pressing the ERR key also clears the error bit so if you press ERR again, the message "NO ERROR" will appear. All error codes and associated display messages are listed in Table 5-8.

Storing and Recalling Voltage and Current Settings for All Outputs

As described on page 75, the power supply has 11 internal registers for storing voltage and current settings of all outputs.

At power on each location contains zero volts and the minimum current limit (see initial conditions on page 71).

The STO and RCL keys allow you to store and recall voltage and current settings for all your output channels from any of the 11 internal registers (numbered 0 through 10). For example, you can store the present settings of voltage and current for all the output channels in internal register 2 by pressing:

STO 2 ENTER

You can change the settings of any of your supply’s outputs any number of times as required and then program them to the settings stored in internal register 2 by pressing;

RCL 2 ENTER

The internal registers 0-3 will retain the settings when power is turned off. When power is turned off and then on again, registers 4-10 will be reset to zero voltage and the current to 10 mA. Register 0 is the power on state. The power will be programmed from values in register 0 at power on.

The advantages in using the internal registers are that command processing time is saved and repetitive programming of different settings is simplified. The STO key can be used in conjunction with the OUTPUT ON/OFF key to store settings while the outputs are disabled (OFF). These stored settings can be used later to program the outputs to the stored settings using the RCL and OUTPUT ON/OFF keys.

Local Operation 89

A

Alignment Procedures

Introduction

This appendix contains the software alignment procedures for the power supply. These supplies should be aligned twice a year, or whenever certain repairs are made (see Service Manual).

The equipment that you need and the test setups to perform the alignment are also shown. A listing of the alignment procedure appears at the end of Appendix A. Refer to Appendix C (Command Summary) for detailed explanations of the commands used within the alignment program.

Because there are no internal or external hardware adjustments, your power supply can be aligned without removing the covers or removing it from the cabinet if it is rack mounted. Alignment is performed by measuring actual output values and sending them to the supply over the GP-IB.

Security against accidental alignment is available. A jumper inside the unit may be moved from NORMAL to CAL

LOCKOUT to disable all alignment commands. Access to this jumper requires opening the unit (see Service Manual).

If a number returned to the power supply is out of range while running the alignment program, the front panel ERR annunciator will indicate an error. For example, an error will be created if the DMM is connected to measure a current when a voltage measurement is anticipated by the program.

Note

The memory used to store correction constants will accept and store data about 10,000 times, which is more than sufficient for normal calibrations over the life of the instrument. However, do not put the unit in a calibration loop that repeatedly turns the calibration mode on and off.

Test Equipment and Setup Required

The following test equipment is required for alignment:

1. A computer connected to the GP-IB connector on the back of the power supply.

2. An Agilent 3458A Digital Multimeter (DMM).

3. A precision 0.1 ohm

±

0.003%, 5 A shunt resistor (4 terminal). Recommended--Leeds & Northrop Part Number 4221-B

(in oil bath) or Guildline - Part Number 9330/.l/Agilent (in air).

In line 130, change the constant SHUNT_r to the actual measured value of the shunt being used.

Figures A-1 through A-5 show setups required for aligning the power supply. You will be instructed by the program to refer to a particular setup. Observe polarity when connecting the DMM.

Calibration Program

The following alignment program can be used as is, provided you have an HP Series 200/300 computer with the BASIC programming language and an Agilent 3458A DMM. Refer to Appendix C (Command Summary) for further information on the commands used within the program.

Alignment Procedures 91

The program contains embedded comments (identified by a leading asterisk {! } ) which explain various sections and procedures. To reduce keystroking, the program may be shortened to a minimum number of lines by eliminating these comments.

The alignment program is available on a 5¼ floppy (Agilent P/N 06626-10001) or 3½ inch microfloppy (Agilent P/N

06626-10002). Both disks also contain a verification program, which tests the power supply's output programming and readback accuracy to ensure that it is aligned correctly.

The program will automatically recognize what model power supply is being tested, the number of outputs the power supply has, and which type output it is testing. Using this information, the program will calibrate each of the power supply.

After the program has finished calibrating the power supply, it will store the calibration constants and turn off the calibration mode.

Type in the program and press RUN.

The program first does some "house keeping".

It will clear the screen, and assign a number of Common Variables (lines 100 - 130).

The program next assigns the address of the power supply to 705 (line 160) and the address of the DMM to 722 (line 170).

It will then assign the CRT as the output device (line 180) and the value of the current shunt. Enter the exact measured value at line 190.

The power supply is sent a CLR command (line 210), which will set it to the turn on state. It will then identify what model power supply is being tested (lines 220 and 230).

Next it will preset the DMM (line 250) and set the DMM sample (line 260).

It will now place the power supply in calibration mode (line 300).

The program is now ready to perform the calibrations on the power supply.

The program will first perform a subprogram (called at line 360 and performed in lines 3420 - 3630) which will set up the output parameters of the output under test.

The program will then calibrate its voltage programming and readback circuitry (the subprogram is called at lines 370 and

380, and performed in lines 1430 - 1920). During this subprogram, the controller will pause, and request that the user set up the power supply and DMM referring to a figure in this Appendix. Press CONTINUE when ready.

Next the overvoltage circuitry will be calibrated (the subprogram is called in line 390, and performed in lines 1960 - 2070).

The current alignment tests on the 2 Amp, 50 W outputs are made using a shunt, be sure that the DMM is connected to the

HI and LO voltage input terminals when the shunt is used.

The current circuitry will be calibrated next (the subprograrn is called in lines 400 and 410, and performed in lines 2110 -

2700). During this subprogram, the controller will pause and request that the user set up the power supply and DMM referring to a figure in this Appendix. Press CONTINUE when ready.

Finally the program will calibrate its current sink readback circuitry (performed in lines 2740 - 3380).

To perform the current sink readback alignment on the Agilent 6626A, outputs of equivalent current capability are used as pairs. For example outputs 1 and 2 are used to test each other, the outputs 3 and 4 are used to test each other.

In order to perform the current sink readback alignment on the Agilent 6625A, the program automatically limits the current of output 2 (the 50 W output) to 0.5 Amps (line 2780).

92 Alignment Procedures

After testing of a channel is completed, the program will check if any errors have occurred (the subprogram is called in line

850 and performed in lines 3670 - 3780). If an error has occurred, a message will be sent to the output device (CRT).

After all outputs are tested without errors, the CAL MODE will be turned off, and the calibration constants stored.

Figure A-1

Figure A-2

Figure A-3

Figure A-4

Figure A-1

Figure A-5

Figure A-2

Alignment Procedures 93

Figure A-3

Figure A-4

94 Alignment Procedures

Figure A-5

Alignment Program

10 ! This program called "ALIGN_6626" will align

20 ! the Agilent6625A, 26, 28 and 29A Power Supplies It

30 ! requires an Agilent3458A DMM and a four terminal

40 ! 0.1 ohm current shunt accurate to +/-50 ppm

50 !

60 ! May 06,l989 Rev A.03.01

70 !

80 Clear_screen

90 !

100 COM /Specs/ Vrng_lo,Vrng_hi,Irng_lo,Irng_hi

110 COM /Ps/ Chan,Other_chan,@Ps,Model$[7]

120 COM /Outputs/ No_of_outputs

130 COM /Vm/ @Vm

140 COM /Shunt/ Shunt_r

150 !

! Call program to clear display

160 ASSIGN @Ps TO 705

170 ASSIGN @Vm TO 722

180 PRINTER IS 1

190 Shunt_r= 1

200 !

210 OUTPUT @Ps;"CLR"

220 OUTPUT @Ps;"ID?"

230 ENTER @Ps;Model$

240 !

! Set Supply address to 705

! Set DMM address to 722

! Print to the CRT

! Set value of current shunt

! Preset Supply to turn on state

! Queries Supply model number

! Enter model number

250 OUTPUT @Vm;"PRESET NORM;TRIG HOLD"

260 OUTPUT @Vm; “NPLC 100”

270 !

! Preset DMM

! Set DMM sample

280 !

290 !

300 OUTPUT @Ps;"CMODE 1"

310 !

! Turn on Supply CAL mode

320 IF Model$="Agilent6625A" OR Model$='Agilent6628A" THEN No_of_outputs=2

330 IF Model$="Agilent6626A" OR Model$="Agilent6629A" THEN No_of_outputs=4

340 !

350 FOR Chan=1 TO No_of_outputs

360 Get_data

! Loop to test all outputs

! Call program with range data

370

380

390

Cal_voltage(Vrng_lo)

Cal_voltage(Vrng_hi)

Cal_overvoltage(Vrng_hi)

! Call program to cal low range voltage

! Call program to cal high range voltage

! Call program to cal overvoltage

400

410

Cal_current(Irng_lo)

Cal_current(Irng hi)

420 NEXT Chan

430 !

440 Chan=l

450 Other_chan=2

460 Get_data

470 Cal_sink(Irng_lo,1)

480

! Call program to cal low range current

! Call program to cal high range current

! Test next output

! Select channel to cal sink (-CC)

! Select current source channel

! Call program with range data

! Call program to cal low range sink

! 1 is + polarity for DMM

490 Chan=2

500 Other_chan=l

510 Get_data

520 Cal_sink(Irng_lo,-1)

! Select channel to cal sink (-CC)

! Select current source channel

! Call program with range data

! Call program to cal low range sink

Alignment Procedures 95

Alignment Program (continued)

530

540

550

560 Chan=1

570 Other_chan=2

580 Get_data

590 Cal_sink(Irng_hi,1)

600 !

610 Chan=2

620 Other_chan=1

630 Get_data

640 Cal_sink(Irng_hi,-1)

650 !

660 IF No_of_outputs=4 THEN

670 Chan=3

680

690

700

710

720

730

740

750

760

Other_chan=4

Get_data

Cal_sink(Irng_lo,1)

!

Chan=4

Other_chan=3

Cal_sink(Irng_lo,-1)

!

Chan=3

770

780

790

Other_chan=4

Cal_sink(Irng_hi,1)

!

800

810

820 Cal_sink(Irng_hi,-1)

830 END IF

840 :

Chan=4

Other_chan=3

850 Check_error

860 Cal_mode_off

870 !

880 END

890 !

900 !

910 !

920 SUB Clear_screen

930 OUTPUT 2;CHR$(255)&CHR$(75);

940 SUBEND

950 !

960 !

970 !

980 DEF FNDcv

990

1000

COM /Vm/ @Vm

OUTPUT @Vm;"DCV"

1010

1020

1030

TRIGGER @Vm

ENTER @Vm;Volts

RETURN Volts

1040 FNEND

96 Alignment Procedures

! -1 is reverse polarity, this saves

! reversing leads as sink channel

! becomes the source for other channel

! Select channel to cal sink (-CC)

! Select current source channel

! Call program with range data

! Call program to cal high range sink

! Select channel to cal sink (-CC)

! Select current source channel

! Call program with range data

! Call program to cal high range sink

! Cal outputs 3 & 4

! Select channel to cal sink (-CC)

! Select current source channel

! Call program with range data

! Call program to cal low range sink

! Select channel to cal sink (-CC)

! Select current source channel

! Call program to cal low range sink

! Select channel to cal sink (-CC)

! Select current source channel

! Call program to cal high range sink

! Select channel to cal sink (-CC)

! Select current source channel

! Call program to cal high range sink

! Call error checking program

! Call program to turn cal mode off

! Subprogram to clear display

! Clear screen

! Function to r at voltage

! Set DMM to read DC voltage

! Take a voltage reading

! Enter the voltage

! Return voltage reading

! End of voltage function

Alignment Program (continued)

1050 !

1060 !

1070 !

1080 DEF FNDci(I_range) ! Function to read current

1100 COM /Shunt/ Shunt_r

1110 IF I_range=2 THEN

1120 Amps=FNDcv/Shunt_r

1130 ELSE!

1140

1150

OUTPUT @Vm;"DCI"

TRIGGER @Vm

1160 ENTER @Vm;Amps

! 2A must read across shunt

! Convert shunt voltage to Amps

! <2A read direct from DMM

! Set DMM to read DC current

! Take a current reading

! Enter the current

1490

1500

1510

1520

1530

1540

1550

1560

1180 RETURN Amps

1190 FNEND

1200 !

1210 !

1220 !

1230 SUB Output_on

1240 COM /Ps/ Chan,Other_chan, @Ps, Model$[7]

! Return the current reading

! Subprogram to turn ON outputs

1250

1260

1270

1280

COM /Outputs/ No_of_outputs

OUTPUT @Ps; OUT l,l;OUT 2,1 ! Turn on outputs 1 & 2

IF No_of_outputs=4 THEN OUTPUT @Ps;"OUT 3,1;OUT 4,1"

! Turn on outputs 3 & 4

1290 SUBEND

1300 !

1310 !

1320 !

1330 SUB Output_off

1340

1350

1360

COM /Ps / Chan, Other_chan, @Ps, Mode1$ [ 7 ]

COM /Outputs/ No_of_outputs

OUTPUT @Ps;"OUT 1,0,OUT 2,0"

! Subprogram to turn OFF outputs

! Turn off outputs 1 & 2

1370 IF No_of_outputs=4 THEN OUTPUT @Ps; “OUT 3,0;OUT 4,0”

1380 ! Turn off outputs 3 & 4

1390 SUBEND

1400 !

1410 !

1420 !

1430 Cal_voltage : SUB Cal_voltage(V_range)

1440 COM /Ps/ Chan,Other_chan,@Ps,Model$ [7]

1450 !

1460

1470

1480

IF V_range=50 THEN GOTO Cal_v

Output_off

Clear_screen

! Subprogram to cal voltage

! Skip setup instructions

! Turn off all outputs

PRINT "Set up output";Chan;"for voltage calibration

PRINT

PRINT " Use Fig A-1 of the Operating Manual"

PRINT

PRINT “ 3458A input is connected to VOLTAGE terminals”

PRINT

PRINT " The current shunt is NOT used"

PRINT

Alignment Procedures 97

Alignment Program (continued)

1570

1580

1590

PRINT "Press CONTINUE when ready

PAUSE

Output_on

1600 !

1610 Cal_v: !

1620

1630

Clear_screen

PRINT "Calibrating the";V_range;"volt range programming”

1640

1650

1660

1670

!

OUTPUT @Ps;"VRSET" ;Chan,V_range

!

OUTPUT @Ps; VLO";Chan

1680 WAIT 1

1690 Vlo=FNDcv

1700 !

1710 OUTPUT @Ps; “VHI” ;Chan

! Set voltage range

! Set to low output voltage

! Wait for supply to stabilize

! Read low voltage

1720 WAIT 1

1730 Vhi=FNDcv

1740 !

1750 OUTPUT @Ps; “VDATA” ;Chan, Vlo, Vhi

1760 !

! Set to high output voltage

! Wait for supply to stabilize

! Read high voltage

! Send output voltage data

1770 Clear_screen

1780 PRINT "Calibrating the";V_range,'volt range readback"

1790 !

1800 OUTPUT @Ps; VRLO";Chan ! Set low readback voltage

1810

1820

1830

REPEAT

UNTIL BIT(SPOLL(@Ps),4)

Vrlo=FNDcv

1840 !

1850 OUTPUT @Ps; “VRHI”;Chan

1860 REPEAT

1870 UNTIL BIT(SPOLL(@Ps),4)

1880 Vrhi=FNDcv

! Wait for supply to finish

! Finished when Bit 4 goes true

! Read low output voltage

! Set high readback voltage

! Wait for supply to finish

! Finished when Bit 4 goes true

! Read high output voltage

1890 !

1900 OUTPUT @Ps; “VRDAT”; Chan, Vrlo, Vrhi

1910 !

1920 SUBEND

1930 !

1940 !

1950 !

1960 Cal_overvoltage:SUB Cal_overvoltage(V_range)

1970 COM /Ps/ Chan,Other_chan,@Ps,Model$[7]

1980 !

1990 Clear_screen

2000 PRINT "Calibrating the Overvoltage circuit

2010 !

2020 OUTPUT @Ps;"VRSET";Chan,V_range

2030 OUTPUT @Ps;"OVCAL";Chan

2040 REPEAT

2050 UNTIL BIT(SPOLL(@Ps),4)

2060 !

2070 SUBEND

2080 !

! Send voltage readback data

! Sub program to cal OV

! Set voltage range

! Calibrate Overvoltage

! Wait for supply to finish

! Finished when Bit 4 goes true

98 Alignment Procedures

Alignment Program (continued)

2090 !

2100 !

2110 Cal_current: SUB Cal_current(I_range)

2120 COM /Ps/ Chan,Other_chan,@Ps,Model$[7]

2130 !

2140 IF I_range= .5 THEN GOTO Cal_i

2150 !

! Subprogram to cal current

! Skip setup instructions

2160 Output_off

2170 !

2180 Clear_screen

2190 PRINT "Set up output";Chan;"for";I_range;”A current calibration”

2200 PRINT

2210 IF I_range=2 THEN

2220

2230 PRINT " Use Fig A-3 of the Operating Manual"

! If the next cal range is 2

! amps then print this message

2240

2250

2260

2270

2280 ELSE

PRINT

PRINT “3458A input is connected to VOLTAGE terminals”

PRINT

PRINT “The current shunt IS used”

! If not 2 amps then print this

2290

2300

2310

2320

PRINT “Use Fig A-2 of the Operating Manual”

PRINT

PRINT “3458A input is connected to CURRENT terminals”

PRINT

2330 PRINT “The current shunt is NOT used”

2340 END IF

2350 PRINT

2360 PRINT “Press CONTINUE when ready”

2370 PAUSE

2380

2390

Output_on

!

2400 Cal_i : !

2410 Clear_screen

2420 PRINT "Calibrating the";I_range;"A current programming

2430 !

2440 OUTPUT @Ps;"IRSET";Chan,I_range ! Set current range

2450 OUTPUT @Ps;''ILO'';Chan

2460 WAIT 1

2470 Ilo=FNDci(I_range)

! Set output current low

! Wait for supply to stabilize

! Read low output current

2480 !

2490 OUTPUT @Ps;"IHI";Chan

2500 WAIT 1

2510 Ihi=FNDci(I_range)

2520 !

! Set output current high

! Wait for supply to stabilize

! Read high output current

2530 OUTPUT @Ps;”IDATA”;chan,Ilo,Ihi

2540 !

! Set output current data

2550 Clear_screen

2560 PRINT 'Calibrating the";I_range;"A current readback"

2570 !

2580 OUTPUT @Ps;"IRLO";Chan

2590 REPEAT

2600 UNTIL BIT(SPOLL(@Ps),4)

! Set low readback current

! Wait for supply to finish

! Finished when Bit 4 goes true

Alignment Procedures 99

Alignment Program (continued)

2610

2620

Irlo=FNDci(I_range)

!

2630 OUTPUT @Ps;"IRHI";Chan

2640 REPEAT

2650 UNTIL BIT(SPOLL(@Ps),4)

2660 Irhi=FNDci(I_range)

2670 !

2680 OUTPUT @Ps;"IRDAT";Chan,Irlo,Irhi

2690 !

2700 SUBEND

2710 !

2720 !

2730 !

2740 Cal_sink SUB Cal_sink(I_range,Polarity)

2750 COM /Ps/ Chan,Other_chan,@Ps,Model$[7]

! Read low output current

! Set high readback current

! Wait for supply to finish

! Finished when Bit 4 goes true

! Read high output current

! Send current readback data

! Subprogram to cal -CC

2760

2770

!

IF I_range= .5 THEN GOTO Sink_rb ! Skip setup instructions

2780 IF Model$="Agilent6625A" AND I_range=2 THEN I_range= .5

2790

2800

! If a 6625A, 2A output will be

! set to 5A to match output 1

! Skip setup instructions 2810 IF Chan=2 OR Chan=4 THEN GOTO Sink_rb

2820 !

2830 Output_off

2840 Clear_screen

2850 PRINT “Set up output';Chan;”for ;I_range; “A current SINK cal”

2860 PRINT

2870 PRINT “ Connect outputs”;Chan; “and” ;Other_chan; “together as in”

2880 !

2890 IF I range=2 THEN

2900 PRINT " Fig A-5 of the Operating Manual'

2910 PRINT

! If it is 2A output use setup 5

2920 PRINT " 3458A input is connected to VOLTAGE terminals"

2930 PRINT

2940 PRINT " The current shunt IS used for this part"

2950

2960

PRINT " of the alignment"

ELSE

2970

2980

2990

PRINT “Fig A-4 of the Operating Manual”

PRINT

! If not 6626A 2A use setup 4

PRINT “3458A input is connected to CURRENT terminals”

3000

3010

3020

3030

PRINT

PRINT " The current shunt IS NOT used"

END IF

PRINT

3040 PRINT “The sink readback of both outputs”

3050 PRINT “will be aligned with this setup”

3060 !

3070 PRINT

3080 PRINT "Press CONTINUE when ready”

3090

3100

PAUSE

!

3110 Sink_rb

3120 Clear_screen

100 Alignment Procedures

Alignment Program (continued)

3130 PRINT "Calibrating output";Chan;"";I_range;"A current sink readback"

3140 !

3150 OUTPUT @Ps;"IRSET";Chan,I_range;";IRSET";Other chan,I_range

3160

3170

! Set I range for source and

! sink range

3180 OUTPUT @Ps; “ISET”;Other_chan,0;”;VSET”;Other_chan,7

3190 ! Set source V & I output

3200

3210

3220

3230

Output_on

WAIT 1

!

OUTPUT @Ps;"IRLN";Chan

3240

3250

REPEAT

UNTIL BIT(SPOLL(@Ps),4)

3260 Ilo=(FNDci(I_range))*Polarity

3270 !

! Wait for source to stabilize

! Set sink readback low

! Wait for supply to finish

! Finished when Bit 4 goes true

! Read low sink current

! Set source output current 3280 OUTPUT @Ps;"ISET";Other_chan,I_range

3290 WAIT 1

3300

3310

3320

!

OUTPUT @Ps;"IRHN";Chan

REPEAT

3330 UNTIL BIT(SPOLL(@Ps),4)

3340 Ihi=(FNDci(I_range))*Polarity

3350 !

3360 OUTPUT @Ps; NIDAT”;chan,Ilo,Ihi

3370 !

3380 SUBEND

3390 !

3400 !

3410 !

3420 Get_data SUB Get_data

3430 COM /Specs/ Vrng_lo,Vrng_hi,Irng_lo,Irng_hi

3440 COM /Ps/ Chan,Other_chan,@Ps,Model$[7]

! Set sink readback high

! Wait for supply to finish

! Finished when Bit 4 goes true

! Read high sink current

! Send current sink data

! Subprogram with range data

3450

3460

3470

3480

3490

3500

3510

3520

3530

3540

3550

3560

IF Model$=“Agilent6628A" OR Model$=“Agilent6629A" THEN RESTORE Type_2_data

!

READ Vrng_lo,Vrng_hi,Irng_lo,Irng_hi

!

!

3570

3580

!

Output variables range data

!

Vrng Irng

3590 !

3600 Type 1 data:DATA

Lo Hi

7, 50,

Lo Hi

.015, .500

3610 Type_2_data:DATA

3620 !

3630 SUBEND

16, 50,

! Read the Range data

.200, 2

! Type 1 data (25W output)

! Type 2 data (50W output)

3640 !

!

IF Model$="Agilent6625A" AND Chan=1 THEN RESTORE Type_1_data

IF Model$="Agilent6625A" AND Chan=2 THEN RESTORE Type_2_data

IF Model$=”Agilent6626A" AND Chan=1 THEN RESTORE Type_1_data

IF Model$="Agilent6626A" AND Chan=2 THEN RESTORE Type 1 data

IF Model$=“Agilent6626A" AND Chan=3 THEN RESTORE Type_2_data

IF Model$=“Agilent6626A" AND Chan=4 THEN RESTORE Type_2_data

Alignment Procedures 101

Alignment Program (continued)

3650 !

3660 !

3670 Check_error:SUB Check_error

3680

3690

3700

3710

COM /Ps/ Chan,Other_chan,@Ps,Model$[7]

Clear_screen

PRINT “Checking for errors”

OUTPUT @Ps;”ERR?”

3720

3730

3740

3750

! Subprogram to check for errors

! Query supply for errors

ENTER @Ps;Err

IF Err>0 THEN

! Enter the error number

! If an error then print msg

PRINT “Error”;Err;”occurred. Supply may not be calibrated”

PRINT "The alignment program is stopped"

3760

3770

STOP

END IF

3780 SUBEND

3790 !

3800 !

3810 !

3820 Cal_mode_off:SUB Cal_mode_off

3830 COM /Ps/ Chan,Other_chan,@Ps,Model$[7]

3840 Clear_screen

! Subprogram to turn off cal mod

3850 PRINT "wait while calibration constants are stored"

3860 OUTPUT @Ps;”CMODE 0”

3870

3880

OUTPUT @Ps;”CLR”

Clear_screen

3890 PRINT "Calibration complete"

3900 SUBEND

3910 !

! Turn cal mode off, store data

! Clear supply

102 Alignment Procedures

B

Programming With a Series 200/300 Computer

Introduction

The purpose of this appendix is to serve as an introduction to programming your power supply with an HP Series 200/300 computer using the BASIC language. Examples are included that employ some of the most frequently used functions.

These examples have been written so that they will run on all power supplies. The values used in the examples (5 V and 0.5

A for instance), are within the operating locus of all outputs on all models. The examples program only channels one and two because both models contain at least two channels (outputs).

You must be familiar with the BASIC language to understand the examples. If you do not recognize a programming statement, look up the keyword in the BASIC Language Reference document that was supplied with your computer, and look up the device command in Chapter 5 of this manual. Questions about program structure and selection are discussed in the BASIC Programming Techniques manual.

I/O Path Names

Throughout this appendix, I/O path names are used in place of interface and device select codes. In a large program, I/O path names simplify changing the address of an instrument if necessary. Reading and writing the program is easier as well.

The I/O path name can be carried in a common block and changed by a single assign statement.

In the programming examples in this appendix, the I/O path name @Ps is used for the power supply. The ASSIGN statement that defines the I/O path must precede any statements that use the I/O path name. Therefore, instead of using the statement OUTPUT 705;’’VSET1,5’’ in the following programs, the equivalent OUTPUT @Ps;"VSET1,5" statement is used. The examples assume that the power supply is at address 5 and the GP-IB interface in the computer is at select code

7(factory default).

Voltage and Current Programming

The power supply normally functions in one of two modes, either constant voltage with current limit or constant current with voltage limit. The operating mode is determined by a combination of voltage and current settings and load resistance.

For example, with a 50

load connected to output 1, the following program will put output 1 in constant voltage mode at 5 volts out with a 0.5 amp current limit. In this case the output current would be 0.1 A.

10 ASSIGN @Ps TO 705

20 OUTPUT @Ps;’’VSET1,5;ISET1,0.5’’

30 END

Line 10: Assigns the I/O pathname to the power supply.

Line 20: Sets output voltage and current. Note the use of the semicolon to separate multiple device commands.

If a 4

load were used instead of a 50

load, output 1 would have been operating in constant current mode at 0.5 amp out with a voltage limit of 5 volts. In this case the output voltage would be 2 V.

Programming with a Series 200/300 Computer 103

VOLTAGE AND CURRENT PROGRAMMING WITH VARIABLES

You can use variables in a program to represent data values in the device commands. This is useful in applications that require changing the voltage and current values to different predetermined settings. The following program uses a variable in a FOR NEXT loop to ramp up output voltage in 0.1 volt steps from 0 to 5 volts.

10 ASSIGN @Ps TO 705

20 OUTPUT @Ps;"CLR;ISET1,0.5"

30 FOR Voltage=0 TO 5 STEP 0.1

40 OUTPUT @Ps;’’VSET1,";Voltage

50 WAIT 0.2

60 NEXT Voltage

70 END

Line 10:

Line 20:

Line 30,60:

Line 40:

Assigns the l/O pathname to the power supply.

Initializes the power supply to its power on state and sets the current limit.

Increments the voltage in 0.1 V steps to 5 volts.

Sets the voltage of output 1 to the value of the variable ’’Voltage’’. The comma inside the quotes is required because it separates numbers in the device command (the output channel number from the voltage value in this case). A space < SP > may also be used instead of the comma. The semicolon outside the quotes is used because it suppresses the <CR> < LF> that the computer would normally send to the power supply if a comma were used as a separator after a string item. Using a comma in this case would produce a syntax error in the power supply.

Waits 0.2 seconds between steps.

Line 50:

Another way to use variables to represent data values in device commands is when using input statements to program the power supply. The following program uses input statements to program the voltage and current settings of output 1 and output 2.

10 ASSIGN @Ps TO 705

20 INPUT "ENTER A VOLTAGE FOR OUTPUT #1",V1

30 INPUT ’’ENTER A CURRENT LIMIT FOR OUTPUT #1",I1

40 INPUT "ENTER A VOLTAGE FOR OUTPUT #2",V2

50 INPUT "ENTER A CURRENT LIMIT FOR OUTPUT #2",I2

60 OUTPUT @Ps;"VSET1,";V1;";ISET1,’’;I1;’’;VSET2,";V2;";ISET2,’’;I2

70 END

Line 10:

Line 20,30:

Line 40,50:

Line 60:

Assigns the I/O pathname to the power supply.

Enters voltage and current values for output 1.

Enters voltage and current values for output 2.

Sets the voltage and current of outputs 1 and 2 to the values entered into the variables. The previous example explained the use of the comma inside the quotes and the semicolon that precedes the variable.

The semicolon that follows the variable suppresses the comma that the computer would normally send to the power supply if a comma were used as a separator after a numeric item. The leading semicolons inside the quotes separate multiple device commands (the VSET commands from the ISET commands in this case).

Voltage and Current Readback

Reading back data from the power supply requires two statements. First, an output statement is used to query the power supply. A list of queries appears in Table 5-2 on page 70. The power supply responds to the query by entering the requested data into a buffer. Next, an enter statement is used to read the data from the buffer on the power supply into a variable in the computer. The following program queries the power supply for the voltage and current settings of output 1 and prints the results on the screen.

104 Programming with a Series 200/300 Computer

10 ASSIGN @Ps TO 705

20 OUTPUT @Ps;’’VSET?1’’

30 ENTER @Ps;Vsl

40 OUTPUT @Ps;"ISET?1"

50 ENTER @Ps;Isl

60 PRINT ’’VOLTAGE SETTING OF OUTPUT #1 = ";Vsl

70 PRINT "CURRENT LIMIT SETTING OF OUTPUT #1 = ";Isl

80 END

Line 10:

Line 20,30:

Line 40,50:

Line 60,70:

Assigns the I/O pathname to the power supply.

Queries the supply for output 1’s voltage setting. You cannot string multiple queries together in a single device command because the power supply can only return the most recently queried data.

Queries the supply for output 1’s current limit.

Prints the results of the queries on the screen.

Programming Power Supply Registers

Present Status

The power supply makes available several forms of status information. Chapter 5 discusses the different registers and their functions. The present status register contains continuously updated information. The following example shows how to query bit position 0 of output 1’s status register to see if output 1 is in CV mode. In this example, the program references the function as a variable in a conditional execution statement. Note that instead of printing a message, line 30 could be used to branch to another part of the program in the event that the supply is in CV mode.

10 ASSIGN @Ps TO 705

20 COM /Ps/ @Ps

30 IF FNCv _ mode THEN

40 PRINT "OUTPUT 1 IS IN CV MODE"

50 END IF

60 END

70 !

80 !

90 DEF FNCv _ mode

100 COM /Ps/ @Ps

110 OUTPUT @Ps;’’STS?1"

120 ENTER @Ps;Sts

130 RETURN BIT(Sts,0)

140 FNEND

Line 10:

Line 20:

Assigns the I/O path name to the power supply.

Declares a common block for the I/O path name. The COM statement must be used for the @Ps variable to preserve its value in the function subprogram.

Line 30,40,50: If FNCv _ mode is true, print the message.

Line 90:

Line 100:

Defines the Cv _ mode function.

Brings in the common block for the I/O pathname.

Line 110,120: Reads the present status of output 1 into the variable Sts.

Line 130: Returns the value of bit position 0 of Sts.

Programming with a Series 200/300 Computer 105

Service Request and Serial Poll

The fault and mask registers, when used in conjunction with the service request and serial poll functions, allow you to select which conditions can cause computer interrupts. The fault and mask registers can also be used independently of the serial poll or service request if so desired. The following example shows how to enable an interrupt to the computer in the case of an overvoltage condition. After the interrupt has occurred, this example includes an interrupt routine that conducts a serial poll to determine on which output the overvoltage occurred. Note that this example assumes that terminal block external OV trip lines are not wired together.

10 ASSIGN @Ps TO 705

20 COM /Ps/ @Ps

30 OUTPUT @Ps;’’CLR;UNMASK1,8;UNMASK2,8;SRQ1’’

40 ON INTR 7,1 CALL Err _ trap

50 ENABLE INTR 7;2

60 OUTPUT @Ps;"OVSET1,4;OVSET2,4"

70 OUTPUT @Ps;’’VSET1,5;VSET2,5"

80 Lbl: GOTO Lbl

90 END

100 !

110 !

120 SUB Err _ trap

130 OFF INTR

140 COM /Ps/ @Ps

150 IF BIT(SPOLL(@Ps),0) THEN

160 OUTPUT @Ps;’’OUT1,0;OVRST1"

170 PRINT "OVERVOLTAGE ON OUTPUT #1"

180 END IF

190 IF BIT(SPOLL(@Ps),1) THEN

200 OUTPUT @Ps;"OUT2,0;OVRST2’’

210 PRINT ’’OVERVOLTAGE ON OUTPUT #2"

220 END IF

230 OUTPUT @Ps;’’FAULT?1;FAULT?2’’

240 SUBEND

LINE 10:

LINE 20:

LINE 30:

Assigns the I/O pathname to the power supply.

Declares a common block for the I/O pathname. The COM statement must be used for the @Ps variable to preserve its value for use in the service routine.

Returns the power supply to its power on state, unmasks output 1’s and output 2’s OV status bits to generate faults, and enables the service request capability on all outputs.

Defines interrupt at interface 7, with GP-IB priority 1.

Enables interrupt at interface 7 for service request type interrupts only.

LINE 40:

LINE 50:

LINE 60:

LINE 70:

LINE 80:

LINE 120:

LINE 130:

LINE 140:

Sets the overvoltage of outputs 1 and 2 to 4 volts.

Sets the voltage of outputs 1 and 2 above the OV setting so that both outputs will overvoltage when the program is run.

Waits for the computer to receive the interrupt. This simulates conditions that would normally exist when a program is running.

Defines the error handling routine.

Disables interrupt capability while processing.

Brings in the common block for the I/O pathname.

LINE 150-180: Conducts a serial poll. If bit 0 in serial poll register indicates a fault for output 1, output 1 is disabled and the overvoltage circuit is reset.

LINE 190-220: Checks bit 1 in serial poll register for a fault on output 2. If true, output 2 is disabled and the overvoltage circuit is reset.

LINE 230: Reads fault registers to clear FAU bits in serial poll register.

106 Programming with a Series 200/300 Computer

Error Detection

The power supply can recognize programming errors and can inform you when a programming error occurs. When an error is detected, no attempt is made to execute the command. Instead, a bit in the serial poll register is set. If SRQ2 or SRQ3 is set, an interrupt will be generated. The following program checks for programming errors and can be entered and run as is.

While it is running, commands can be sent to the power supply from the keyboard. If the ERROR annunciator on the power supply’s front panel indicates that an error has been detected, depress the labeled softkey to display the error on your computer screen.

10 ASSIGN @Ps TO 705

20 COM /Ps/ @Ps

30 ON KEY 0 LABEL "ERROR?’’ CALL Err_trap

40 Lbl: GOTO Lbl

50 END

60 !

70 !

80 SUB Err _ trap

90 OFF KEY

100 COM /Ps/ @Ps

110 OUTPUT @Ps;"ERR?’’

120 ENTER @Ps:Err

130 OUTPUT 2 USING "#,K";CHR$(255)&CHR$(75)

140 IF Err THEN

150 PRINT "POWER SUPPLY PROGRAMMING ERROR:’’

160 END IF

170 SELECT Err

180 CASE 0

190 PRINT ’’NO ERROR OCCURRED"

200 SUBEXIT

210 CASE 1

220 PRINT "INVALID CHARACTER’’

230 CASE 2

240 PRINT "INVALID NUMBER"

250 CASE 3

260 PRINT "INVALID STRING"

270 CASE 4

280 PRINT ’’SYNTAX ERROR"

290 CASE 5

300 PRINT ’’NUMBER OUT OF RANGE"

310 CASE 6

320 PRINT "DATA REQUESTED WITHOUT QUERY’’

330 CASE 7

340 PRINT ’’STRING EXCEEDS DISPLAY LENGTH"

350 CASE 8

360 PRINT "NUMBER TOO LARGE FOR INPUT BUFFER"

370 CASE 28

380 PRINT "INVALID CHARACTERS IN STRING"

390 CASE ELSE

400 PRINT "UNRECOGNIZED ERROR NUMBER’’; Err

410 END SELECT

420 PRINT "RE-ENTER STATEMENT AND TRY AGAIN"

430 SUBEND

Programming with a Series 200/300 Computer 107

LINE 10:

LINE 20:

LINE 30:

LINE 40:

LINE 80:

Assigns the I/O path name to the power supply.

Declares a common block for the I/O path name.

Defines interrupt on softkey depression and branch to error routine.

Idles on softkey definition.

Defines subprogram Err _ trap

LINE 90:

LINE 100:

Disables interrupt capability while processing.

Brings in the common block for the I/O pathname.

LINE 110,120: Enters error code from power supply.

LINE 130: Clears computer screen.

LINE 140,150: If an error occurred, print message.

LINE 170-420: Prints message based on error code. Prompt user to try again. If no error occurred, print message saying no error occurred.

Stored Operating States

Your power supply has registers that can store up to 11 operating states (note registers 0-3 are non-volatile, see Chapter 3).

These states can be recalled in arbitrary order. Storing a state involves taking a "snapshot" of the voltage and current settings that are in effect when the command is received. The following example uses stored operating states to set up an output. Using this method of setting up outputs saves processing time and facilitates repeating the same commands.

10 ASSIGN @Ps TO 705

20 OUTPUT @Ps;’’OUT1,0;OUT2,0"

30!

40 OUTPUT @Ps;"VSET1,1;ISET1,0.01;VSET2,5;ISET2,0.01;STO1"

50 OUTPUT @Ps;"VSETl,2;ISETl,0.02;VSET2,4;ISET2,0.02;STO2"

60 OUTPUT @Ps;"VSETl,3;ISETl,0.03;VSET2,3;ISET2,0.03;STO3"

70 OUTPUT @Ps;"VSETl,4;ISETl,0.04;VSET2,2;ISET2,0.04;STO4"

80 OUTPUT @Ps;’’VSET1,5;ISET1,0.05;VSET2,1;ISET2,0.05;STO5"

90 !

100 OUTPUT @Ps;"CLR"

110 FOR State = 1 to 5

120 OUTPUT @Ps;"RCL";State

130 WAIT 2

140 NEXT State

150 END

LINE 10:

LINE 20:

Assigns the I/O pathname to the power supply.

Disables Output 1 and Output 2.

LINE 40-80: Stores 5 operating states for Output 1 and Output 2 in storage registers 1 through 5. Outputs not explicitly programmed will store the settings that are in effect when the store command is received.

LINE 100: Clears the supply. All outputs are enabled and set the initial power on state.

LINE 110-140: Loops through the sequence of five states with a two second wait between states.

108 Programming with a Series 200/300 Computer

C

Command Summary

Introduction

For convenience, a “quick reference” listing tabulating general command headers, and calibration command headers is shown at the beginning of Table C-1 (page 110). (Note that calibration commands are described in detail in Appendix A.)

Table C-2 (pages 111-115) provides an alphabetical listing and a brief description of each command that may be sent to the power supplies. Chapter 5 gives a complete description (including syntax) of most of the commands listed.

All of the commands can be executed remotely over the GP-IB. Many of the commands may also be executed locally from the supply's front panel (marked by single asterisks in Table C-1).

Several commands will generate a response which will be indicated on the front panel of the supply. These commands are indicated by double asterisks in Table C-1.

Command headers are accepted in upper or lower case letters although only upper case letters are used in this summary.

The brackets < > indicate a number to be sent.

Note that <ch> must specify an output channel number from 1 through 4 as applicable. Commas or spaces are required between numbers. Spaces are optional between the command header and the first number. Use semicolons between multiple commands sent in one statement.

Command Summary 109

*

*

**

*

**

**

**

*

ASTS? < ch >

CLR

CMODE < 1 (on) or 0 (off) >

CMODE?

DCPON < 1 (on) or 0 (off) >

DLY < ch >, < delay time >

DLY? < ch >

DSP < 1 (on) or 0 (off)>

DSP <"string">

DSP?

ERR?

FAULT? < ch >

ID?

IOUT? <ch>

IRSET <ch>, <x>

IRSET? < ch >

ISET < ch >, < current setting >

ISET? <ch>

ISTEP < ch >, < step value >

METER < ch >

METER?

OCP < ch >, < 1, (on) or 0 (off) >

OCP? < ch >

OCRST <ch>

Table C-1 Quick Reference Listing of Commands

General Commands

*

**

*

**

*

**

**

**

**

**

OUT < ch >, < 1 (on) or 0 (off) >

OUT? < ch >

OVRST < ch >

OVSET < ch > < x >

OVSET? < ch >

PON < 1 (on) or 0 (off) >

PON?

RCL < register >

ROM?

SROM? < ch >

SRQ <x>

SRQ?

STO < register >

STS? < ch >

TEST?

UNMASK < ch >, < sum of bit weights >

UNMASK? < ch >

VOUT? < ch >

VRSET < ch >, <x >

VRSET? < ch >

VSET < ch >, < voltage setting >

VSET? < ch >

VSTEP < ch >, < step vaiue >

IDATA < ch >, < llo >, < ihi >

IHI < ch >

ILO < ch >

IRDAT < ch > , < Ilo >, < Ihi >

IRHI < ch >

IRLO < ch >

IRLN < ch >

IRHN < ch >

IDAC < ch >, < x >

IDAC? < ch >

RDAC < ch >,< x >

RDAC? < ch >

Calibration Commands

NIDAT < ch >, < Ilo >, < Ihi >

OVCAL <ch>

VDATA < ch >, < VLo >, < Vhi >

VHI < ch >

VLO < ch >

VRDAT < ch >, < Vlo >, < Vhi >

VRHI < ch >

VRLO < ch >

Servlce Commands

VDAC < ch >,< x >

VDAC? < ch >

VMUX? < ch >

*· = Command can be executed from the front panel.

** = Data returned from these cormnands is also displayed on the front panel after command is issued. Note that the display must be enabled (i.e. DSP = 1).

110 Command Summary

ASTS? < ch >

CLR

CMODE < on/off >

CMODE?

DCPON < on/off >

DLY < ch >,< x >

DLY? < ch >

DSP < on/off>

DSP?

DSP “xxxxxxxxxxxx”

ERR?

FAULT? < ch >

IDAC < ch >, < x >

IDAC? < ch >

Table C-2 Command Summary

Queries the accumulated status (ASTS) of the specified output < ch > . The response

(integer 0-255) represents the sum of the binary weights of the ASTS register bits (see page 77). The ASTS register is automatically set to the present status after being queried.

Returns the entire power supply (all outputs) to the power on state, except that the supply is not unaddressed and its store/recall registers are not changed (see page 76).

Turns the calibration mode on or off. On/off is a 1 to turn the calibration mode on; a 0 turns it off (see Appendix A).

Queries if the calibration mode is on or off. Response is either a 1 (on) or 0 (off), (see page 82).

Sets the state of the outputs at power-on. For on/off = 0, all outputs will be off when the power supply is turned on. For on/off = 1 all outputs will be on when the power supply is turned on.

Sets the reprogramming delay time (x) for the specified output < ch > . This delay is used to mask the CV, + CC, - CC, and UNR status bits from the fault register and the

OCP function for the specified delay period (0 to 32 seconds). The delay time is initiated after a VSET, ISET, OUT, RCL, over voltage reset, or over current reset commands is sent (see page 81).

Queries the present reprogramming delay of the specified output <ch> (see page 81).

The response can be a real number from 0.000 to 32 seconds (e.g. 0.450 seconds).

Turns the power supply’s front panel display either on or off (see page 81). On/off equals

1 to turn the display on or a 0 to turn it off.

Queries the present status of the display (see page 81). Response is either a 1 (on), or a 0

(off).

Puts the quoted string on the power supply's front panel display (see page 81). Only numerals, upper case letters, and spaces are allowed (12 characters max) in the quoted string.

Queries the present programming or hardware error (see page 82). An error code number is returned over the GP-IB to identify the error. In the local mode, pressing the error key will cause the appropriate error message (not the error code) to be displayed on the front panel. The error register is cleared after being read.

Queries the fault register of the specified output < ch > (see page 77) . A bit is set in the fault register when the corresponding bit is set in both the status and the mask registers.

The response is an integer 0 to 255. The fault register is cleared after being read.

Programs the current DAC to a specified number of counts < x > for the selected channel < ch > . The number returned is in DAC counts.

Queries the LSB setting of the current DAC for the selected channel < ch > . The number returned is in DAC counts.

Command Summary 111

ID?

Table C-2 Command Summary (continued)

Queries the identification (model number) of the supply. (see page 82).

IDATA < ch >, < Ilo > , < Ihi > Sends data to calibrate the current setting circuits of the specified output < ch >. Ilo and

Ihi are measured values which the supply uses to calculate correction constants (see

Appendix A).

IHI < ch >

ILO < ch >

IOUT? < ch >

Causes the current of the specified output < ch > to go to the high calibration point (see

Appendix A).

Causes the current of the specified output < ch > to go to the low calibration point (see

Appendix A).

Queries the measured output current of the specified output < ch > (see page 73). The response is a real number. The front panel display can be used to monitor the measured output current (and voltage) of the selected output < ch >.

IRDAT < ch >, < Ilo >, < Ihi > Used to calibrate the + current readback for the specified output < ch >. This command must be sent after IRHI and IRLO commands.

IRLO < ch > Finds the minimum + current readback value for the specified output < ch > ( see

Appendix A.)

IRHI < ch >

IRLN <ch>

IRHN <ch>

Funds the maximum + current readback value for the specified output < ch > (see

Appendix A.)

Finds the minimum - current readback value for the specified output < ch > (see

Appendix A.)

Finds the maximum - current readback value for the specified output < ch > ( see

Appendix A).

IRSET < ch >, < x >

IRSET? < ch >

ISET < ch >,< x >

ISET? < ch >

ISTEP < ch >,< x >

*

METER < ch >

METER?

Sets the full scale current range programming for an output < ch > . The power supply will select the lowest operating range that will contain the value (x). If the present operating current of the supply is greater than the full scale range selected, the present output current will be reduced to the full scale value of the new range selected. If the current entered is larger than the highest possible range, an error message will result.

Queries the full scale current range setting of an output < ch > .

Sets the current of the specified output < ch > (see page 73).

Queries the present current setting of the specified output < ch > (see page 73). The response is a real number.

Increases or decreases an output < ch > current in steps. If the step value < x > is positive, output current will increase by the step value, if the step value is negative, the output current will decrease by the step value. Trying to step beyond the active operating range will cause a number range error.

Selects which output < ch > of the supply the front panel will display.

Returns which output the supply is displaying on the front panel.

112 Command Summary

Table C-2 Command Summary (continued)

NIDAT < ch > . < Ilo > . < Ihi > Sends data to calibrate - current readback for the selected output < ch > . This command must be sent after IRLN and IRHN commands.

OCP < ch >, < x >

OCP? < ch >

OCRST < ch >

Enables the over current protection circuit protection circuit for the specified output

< ch >. This circuit, when enabled, causes the output to go to the off state when the output is on the + CC mode. On/off is a 1 to turn on (enable) or a 0 to turn off (disable) the circuit. (see page 75).

Queries the over current protection circuit on/off status for the specified output < ch >

(see page 75). Response is either a 1 (on) or a 0 (off). The OCP ENBLD annunciator on the front panel displays the on/off status of the OCP circuit for the selected output.

Returns the specified output < ch > to the previous settings after it had been turned off by the over current protection circuit (see page 75).

OUT < ch >, < on/off >

OUT? < ch >

Turns the specified output < ch > on or off. On/off equals 1 to turn the output on and equals 0 to turn the output off (see page 77).

Queries whether the specified output < ch > is turned on or off (see page 77). The response is either 1 (on) or 0 (off). The front panel will display the message

’’DISABLED’’ when the selected output channel is turned off.

OVCAL < ch >

OVRST < ch >

OVSET <ch>,<x>

OVSET? < ch >

PON < on/off >

PON?

RCL < x >

RDAC < ch >, < counts >

RDAC? < ch >

Causes the specified output < ch > to go through the overvoltage calibration routine (see

Appendix A).

Attempts to reset the overvoltage crowbar circuit in the specified output < ch > (see page

77).

Sets the overvoltage trip point < x > for the specified output < ch > ( see page 77).

Queries the present overvoltage setting of the specified output < ch > (see page 77). The response is a real number.

Enables the power on service request (SRQ). On/off = 1 causes the power supply to generate a service request when power is applied (see page 80). On/off = 0 disables the

PON SRQ. The on/off settings is retained in the supply’s memory through interruption of ac line power.

Queries the present state of the power on SRQ function (see page 80). The response is either 1 (on) or 0 (off).

Recalls the voltage and current settings for all output channels from the specified internal register < x >. These settings were previously stored using the STO command.

RCL programs all output channels in sequential order (1, 2, 3, 4) to these stored settings

(see page 80).

Sets the readback DAC of an output < ch > to a specified number of counts.

Queries the setting of the readback DAC of an output < ch > . The number returned is in

DAC counts.

Command Summary 113

ROM?

SROM?

SRQ <X>

SRQ?

STO < x >

STS? < ch >

TEST?

UNMASK < ch >, < x >

UNMASK? < ch >

VDAC < ch >, < x >

VDAC? < ch >

VDATA< ch >, < Vlo >,

< Vhi >

VHI <ch>

VLO < ch >

Table C-2 Command Summary (continued)

Queries the revision date of the power supply’s firmware. See service manual.

Queries the revision date of the secondary ROM. See service manual.

Sets the causes for generating SRQ. Setting < x > can be a 0, 1, 2, or 3 as described on page 79.

Queries the present setting of the reasons for issuing an SRQ (see page 79). Response is

0, 1, 2, or that corresponds with the settings for SRQ.

Stores the present operating state of all outputs in one of eleven (0-10) registers < x > .

Register 0 will store values of: VSET, VRSET, ISET, IRSET, OVSET, OCP, DLY, and

MASK.

Registers 1-10 will store values of: VSET, VRSET, ISET, IRSET, and OVSET.

Registers 1 - 3 are contained in the EEPROM (U230) and are non-volatile. Registers 6-

10 are located in RAM (U207) and are volatile. Whenever the supply is powered up, it initializes all outputs using the values stored in register 0. STO <0,1,2,3> can only be sent once. An error will result if an attempt is made to store more than once into each location. The supply must be cycled ON/OFF to re-enable.

Queries the present status of the specified output < ch > . The response (integer 0-255) represents the sum of the binary weights of the status register bits (see page 76).

Causes the power supply to perform a self test of its GP-IB interface. The response is 0 if the test passes or an integer failure code (see page 82).

Sets the bits in the mask register of the specified output channel to the setting (integer from 0 to 255). The mask register operates in conjunction with the status and fault registers (see page 78).

Queries the present setting of the mask register of the specified output channel (see page 77). The response is an integer from 0 to 255.

Programs the voltage DAC to a specified number of counts < x > for the selected channel < ch >.

Queries the LSB setting of the voltage DAC for the selected channel < ch >. The number returned is in DAC counts.

Sends data to calibrate the voltage setting circuits of the specified output < ch > . Vlo and Vhi are measured values which the supply uses to calculate correction constants (see

Appendix A).

Causes the voltage of the specified output < ch > to go to the high calibration point (see

Appendix A).

Causes the voltage of the specified output < ch > to go to the low calibration point (see

Appendix A).

114 Command Summary

VMUX? < ch > , < x >

VOUT? < ch >

VRDAT < ch >, < Vlo >,

< Vhi >

VRLO < ch >

VRHI < ch >

VRSET < ch > ,x

VRSET < ch >

VSET < ch >, < x >

VSET? < ch >

VSTEP< ch > , < x >

Table C-2 Command Summary (continued)

Queries the measurement of the input < x > to the analog multiplexer on the specified output < ch > (see Appendix A).

Queries the measured output voltage of the specified output < ch > (see page 70). The response is a real number. The front panel display can be used to monitor the measured output voltage (and current) of the selected output channel.

Used to calibrate the voltage readback for the specified output < ch > . This command must be sent after VRHI and VRLO commands.

Finds the minimum voltage readback value for the specified output < ch > .

Finds the maximum voltage readback value for the specified output < ch >.

Sets the full scale voltage range for prograrnming an output < ch > . The power supply will select the lowest operating range that will contain the value (x). If the present operating voltage of the supply is greater than the full scale range selected, the present output voltage will be reduced to the full scale value of the new range. If the voltage entered is larger than the highest possible range, an error message will result.

Queries the full scale voltage range setting of an output < ch >.

Sets the voltage of the specified output < ch > (see pages 71-72).

Queries the present voltage setting of the specified output < ch > (see page 72). The response is a real number.

Increases or decreases an output’s < ch > voltage in steps. If the step value < x > is positive, the output voltage will increase by step value; if the step value is negative, the output voltage will decrease by the step value. Trying to step beyond the active operating range will cause a number range error.

Command Summary 115

D

Error Codes and Messages

Introduction

This appendix describes the GP-IB error codes that can be readback to the controller and the error messages that can be displayed on the power supply’s front panel. A brief explanation of each code and message is also given. The error codes and/or messages fall into three categories: Power-on Self Test Messages, responses to the ERR? query, and responses to the

TEST? query.

Power-On Self Test Messages

The applicable message listed in Table D-1 is displayed on the supply’s front panel if the corresponding function has failed the power-on self test. No error code numbers can be returned for these errors. If any of these error messages appear, refer to the Troubleshooting section in the Service Manual.

Error Responses

Table D-2 describes the error codes and messages that can be generated for various programming, calibration, and hardware errors. The error codes can be read back over the GP-IB in response to the ERR? query. The applicable error message is displayed on the supply’s front panel when the ERR key is pressed. Sending the ERR? or pressing the ERR key clears the error.

Test Responses

Table D-3 describes the codes that can be read back over the GP-IB in response to the TEST? query. The TEST? query initiates a self test of the supply.

Message

TIMER FAILED

8291 FAILED

CV DAC CH < ch >

CC DAC CH < ch >

OV DAC CH < ch >

FUSE CH < ch >

HDW ERR CH < ch >

Table D-1. Power-On Self Test Error Messages

Explanation

The timer on the GP-IB board failed.

GP-IB control chip on the GP-IB board failed.

The voltage DAC on the specified output board failed.

The current DAC on the specified output board failed.

The overvoltage DAC on the specified output board failed.

The return fuse on the specified output board is opened.

The specified output board is down (no output).

Error Code

(ERR? query)

0

1

2

Message

(ERR key)

NO ERROR

Table D-2. ERROR Responses

Explanation

Indicates there are no errors.

INVALID CHAR You sent the supply a character it did not recognize.

INVALID NUM The format of your number is incorrect. Check syntax (see Chapter 5).

Error Codes and Messages 117

Error Code

(ERR? query)

3

4

5

6

7

Message

(ERR key)

Table D-2. ERROR Responses (continued)

Explanation

INVALID STR You sent a command the supply does not understand. Resend recognizable command.

SYNTAX ERROR You sent a command with improper syntax. Check syntax of your command (see Chapter 5).

NUMBER RANGE An out of range number was sent. Send a new number within the legal range.

NO QUERY The computer addressed the supply to talk, but it did not first request data. Send query first and then address the supply to talk.

DISP LENGTH Quoted string in the DSP command exceeds the display length of 12 characters.

8

9

10

11

12

13

14

15

16

17

BUFFER FULL May occur if too many numbers are sent. Error code 4 or error code 5 are more likely to occur for this condition.

EEPROM ERROR The EEPROM on the GP-IB board is not responding correctly to programming commands. An instrument failure has occurred and service is required. Refer to the Troubleshooting Section in the Service

Manual.

HARDWARE ERR An output error has occurred on an unknown output. Service is required.

Refer to the Troubleshooting Section in the Service Manual.

HDW ERR CH 1 Error codes 11 through 14 refer to a specific output where an output error has occurred. Service is required. Refer to the Output Board

Troubleshooting Section in the Service Manual.

HDW ERR CH 2 Same as in error 11.

HDW ERR CH 3

HDW ERR CH 4

Same as in error 11.

Same as in error 11.

NO MODEL NUM The supply’s model number cannot be found. The GP-IB Interface board may be defective or the supply may require reprogramming. Refer to the

Troubleshooting Section in the Service Manual.

CAL ERROR An error has occurred during calibration. This may be the result of out of range numbers sent. If recalibration (see Appendix A) does not fix this there may be a hardware failure (Refer to the Troubleshooting Section in the Service Manual).

UNCALIBRATED Unexplained EEPROM error; possibly as the result of incorrect calibration procedure. Recalibrate as described in Appendix A. If the problem persists, a hardware failure exists (Refer to the Troubleshooting

Section in the Service Manual)

118 Error Codes and Messages

Error Code

(ERR? query)

18

Message

(ERR key)

CAL LOCKED

Table D-2. ERROR Responses (continued)

Explanation

Calibration was attempted with the calibration jumper on the GP-IB board in the lockout position (See Section IV in the Service Manual).

Reposition the jumper and re-calibrate if this is desired.

22 SKIP SLF TST The self test jumper on the GP-IB board is in the Skip Self Test Position

(See Section IV in the Service Manual). Reposition the jumper and carry out self-test if this is desired.

Response Code

(TEST? query)

0

20

Table D3. TEST? Response

Explanation

No errors were detected.

The timer on the GP-IB board failed self test. Refer to the Troubleshooting Section in the

Service Manual.

21

27

The RAM on the GP-IB Board failed self test. Refer to the Troubleshooting Section in the

Service Manual. Note that if this failure occurred at power-on, it cannot be displayed on the front panel.

The RAM on the GP-IB Board failed self test. Refer to the Troubleshooting Section in the

Service Manual. Note that if this failure occurred at power-on, it cannot be displayed on the front panel.

29

30

See Table 5-9 on page 84.

See Table 5-9 on page 84.

Error Codes and Messages 119

E

Manual Backdating

Introduction

The backdating information in this chapter applies to units that have the following serial numbers:

Agilent Model 6625A. serials 2831A-00101 to 00663

Agilent Model 6626A, serials 2831A-00101 to 00823

Agilent Model 6628A, serials 2922A-00101 to 00232

Agilent Model 6629A, serials 2920A-00101 to 00292

Make Changes

On page 29, replace the information in Line Voltage Conversion paragraph under steps number 2, 3, and 4 as follows:

2. To open the line module push against the tab on the line module in the direction of the ac input socket.

3. Remove the voltage selector card from its receptacle. This card is about one inch square and has a red notch in each of its four edges.

4. To select a voltage, insert the card into its receptacle so that after the door is closed, the red notch shows through the window next to the voltage level you require.

It is possible to insert the voltage selector card so that a red notch is not visible through any of the four windows. Do NOT operate the power supply while the selector card is in this position.

Also on page 28, replace figure 2-3 with the following figure;

Figure 2-3. Line Module Detail

Manual Backdating 121

Agilent Sales and Support Office

For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir

You can also contact one of the following centers and ask for a test and measurement sales representative.

United States:

Agilent Technologies

Test and Measurement Call Center

P.O. Box 4026

Englewood, CO 80155-4026

(tel) 1 800 452 4844

Latin America:

Agilent Technologies

Latin American Region Headquarters

5200 Blue Lagoon Drive, Suite #950

Miami, Florida 33126

U.S.A.

(tel) (305) 267 4245

(fax) (305) 267 4286

Canada:

Agilent Technologies Canada Inc.

5150 Spectrum Way

Mississauga, Ontario

L4W 5G1

(tel) 1 877 894 4414

Europe:

Agilent Technologies

Test & Measurement European Marketing Organisation

P.O. Box 999

1180 AZ Amstelveen

The Netherlands

(tel) (31 20) 547 9999

Australia/New Zealand:

Agilent Technologies Australia Pty Ltd

347 Burwood Highway

Forest Hill, Victoria 3131

(tel) 1-800 629 485 (Australia)

(fax) (61 3) 9272 0749

(tel) 0 800 738 378 (New Zealand)

(fax) (64 4) 802 6881

Asia Pacific:

Agilent Technologies

24/F, Cityplaza One, 1111 King’s Road,

Taikoo Shing, Hong Kong tel: (852)-3197-7777 fax: (852)-2506-9284

Japan:

Agilent Technologies Japan Ltd.

Measurement Assistance Center

9-1, Takakura-Cho, Hachioji-Shi,

Tokyo 192-8510, Japan

(tel) (81) 426 56 7832

(fax) (81) 426 56 7840

Technical data is subject to change.

122 Agilent Sales and Support Office

Manual Updates

The following updates have been made to this manual since the print revision indicated on the title page.

2/01/00

All references to HP have been changed to Agilent.

All references to HP-IB have been changed to GPIB.

9/20/04

The Declaration of Conformity has been updated.

5/11/09

A URL has been added to the declaration page to obtain the latest declaration of conformity.

The RF field annotation note has been removed from Table 1-2.