advertisement
OLY MPUS
POLARIZING MICROSCOPE
I
INSTRUCTION MANUAL
I
, (
- I I
OLYMPUS
This Instruction
manual
has been written
fartbs
use of Otympus Polarizing
Micro-
m p e Madel CHA-F. It is recommended to
read
the manual carafully in order to familierlze youmlf fully with the USB af the mimompe on the polarizing attachment
-we
IMPORTANT
tlta following points carefully:
1. Always handle the microscope with the care it deserves, and avoid abrupt
motions.
2. Avoid exposure of the microscope to direct sunlight, high temperature and humidity, dust and vibration.
If the microscope is used in ambient temperature higher than 40'~ impede its proper function.
3
3. Only use the tension adjustment ring for atbring the tension of the coarse adjustment.
Do not twist the two coarse adjustment knobs in the opposite directions simultane- ously, which might cause damage.
4. Ascertain that the line voltage selector switch on the base plate is set to
conform
with the local mains voltage.
1. Lenses must always i3e kept
clean.
Fine dust on
lens
surfaces should be blown or wiped
off
by maam of an air blower or a clean brush. Carefully wipe off oil or finger- prints deposited on the lens sutfaces with gauze moistened with a small amount of xylem, alcohol or
ether.
2. Do not use organic solutlans to wipe the surFaces of various camponants. Plastic parts, especially, should be cleaned with a neutral devteqent.
3. Never diassemble the microscope for repair.
4. The miwascope should be stored in its container immediately after uw. If this is not possible, it should be
mered
wivth a vinyl dust m r .
5. Disconnect the line cord from the AC power
source
k f o r e fuse replament.
CONTENTS
I
.
STANDARD EQUIPMENT
.....................................
It
.
NOMENCLATURE
..........................................
In
.
~ E M B L Y
N
. lDENTlFlCATlON AND FUNCTION OF VARIOUS COMPONENTS
. . . . . . . .
V
.
OPERATION
..............................................
1
.
Adiustment of Minimum Line Voltage
2
. lnterpupillary Distance and Diopter Adjustments
.
Polarirer Alignment
......................................
8
4
.
Centering the Stage
...................................... g
5
.
Centering the Objectives
6
.
Use of Apenure Irk Diaphragm
7
.
Focusing Adjustment
.....................................
10
8
.
Orthosoopic Observation
9
.
Conompic Observation
...................................
11
Vl
.
OPTICALDATA
............................................
VII
.
TROUBLESHOOTING
12
I. STANDARD EQUIPMENT
Model
Component
Microscope stand with circular rotatable stage
and
quadruple nmpiem
CHA-P-F
Intermediate polarizing attachment AH-PA
Quarter
wave
plate
(retardation 1 47.7.3W AH-TP147
SenltFw tint plate (retardation Wnyr) AH-TP530
Polarizing
monocular
tube (45") CH-PMO
Polarizing binocular tube (30°)
Swing-out polarizing
condenser
H a i m lamp
Halogen bulbs socket
PO
4x
BH-PSI
BH-PUC
CLSH-B
BVIOWHAL
Objectives (strain-f reel W10X
P040X
AH-WF1OX
Evepieces
AH-Miwo WF 1 QX
Spare fuses ( M A for 1110-1 113-12DV or 0.3A for 220-24\11
Vinyl dust c w e r
C H A - P a l
C H A - P a 1
1 1
1
2
1
1
1
1
0
1
1
1
1
1
1
2
1
1
1
0
1
2 t
1
1
I
0
I
I
1
2
1
I . NOMENCLATURE
Photo: Model CHA-P-651. Model CHA-P-051 is also available by modification of som components.
Revolving nosepiece
Objective axmatior, tube
Intermediate pojwiaing a t m h m t
Microscope stand
Condenser
1
Ill. ASSEMBLY
The picture balow illustrates the sequential procedure of assembly. The numbers indicate the assembly order of various components. Remove dust caps before mounting components.
Take care to keep all
glass
surfaces
ctean,
and avoid watching the surfaces.
Insert tha obbctb 10X into the fixed aperture of tha nos-
P .
@ Eyepiece
Quarter wave plete
Sensitive tint plate
Clamping screw with the lettEtrs "OLYMPUS' facing in front of the mi-
Clamping screw
Microscope stand
@ To AC outlet
Condenser
Q
Q
0 - 0
/
Halogen lamp Socket
Aligning red dots, on con- denser mount and condensas, in- the conderuar Into the
IV. IDENTI FlCATlON AND FUNCTION
OF
VAR
lOUS
COMPONENTS
Observation rube clamping screw
P
Main switch
R h m a t trimmer screw
After swltchlno r-te this wrew wlth a coin until
0mltIon. on. M tb bulb 18 dimly Ilt, n m r y , with the sliding cmtrol lrmrr at mlntmum
/
Line v o l w switch
Set the swltch to conform with the 1-1 mains volmp.
Bertrand tens turret ring
W llght path; "OUT" for remmal of the Barwand lens from the li&t path.
Automatic pre-fowsing lever
Swinwut knob for top lam
Codewer M.A. Is 0.9
(when top lms swln$s out N.A. k
Mechanical tube length adjust- ment rings
I
Bertrand lens focusing ring
Stage centering screw
Aperture iris diaphragm ring
Numerial aperture sale grrd-
Film mount
Amwe 45rnmdlm f llters.
Bulb mount
\
Lamp house clamping knob
The lamp h o w c o w Ean opened by kn&: or cl- pulllw down t
M o r e pushing, twr by pushing it up untll it snaps In plaes. escertaln knob is positiQned aa that the s h m in the picture rlght, markd wilh
Match the line voltage selector switch to Imal mains voltage
Switch on the light source.
Adjust the trimmer screw until the bulb is dimly lit (page 7).
Place a spacimn slide on the stage.
Remove the Bertrand lens and analyzer from the light path.
5).
Coarse
focus with the 1 O X objective.
Make interpupillary and diopter adjustments (paga 7).
Center the stage (page 9 ) .
C e n w objectiw other than 10X (page 9).
Swing in the desired objective.
Set
the
condenser, analyzer and Bertrand lens mrrectly according to your microscopic purpose (page 1
and
1 1).
Adjust light intensity.
Fine focus.
Adjust aperture
Iris
diaphragm {page 9).
Adjustment
of
Illumination System
Microampic r n e t h d
Obiective
O h o s m ~ i c o W a t i o n
ConmopIa observation
4X ,WX
20x to i ~ a x
Bmrand lens
OUT
IN
Condenmr top lens
O W
IN
For biological use, however,
remove
the analyzer, Bertrand lms and sensitive tint plates.
*
Cut off this paae a t dotted line and put it on the wall near the microscop for use as a re- minder of microscopic procadure.
OLYMPUS
V.
OPERATION
1. Adjustmm of Minimum Line Voltage
'
The
mhirnurn vol-
q u i r e d for the light source can be adjusted with
the
rheostat trim mer smw a? the microscope bas plate in accordanw with the line
voltage
and frequency.
The built-in rheostat incorporates a thyristor in its semiconductor circuit for the following advantages:
(a) E x t r e d y fine adjustment of light intensity mn be easily achied.
(b) Flickering of the bulb filament is eliminated and the light inmnsity is stabilized.
(cJ Increased life exp$ctanw of t h ~
For
adiustment
of the minimum line voltam,
ascartain
that the voltage selector switch is
E& to i n f o r m with the I m l mains voltaw, and the sliding control leuera is positioned closest to you (low voltage), and then activate the main switch Q. If the bulb Is dimly lit, the secondary voltage is correct. If it is not lit at all, rotate the rheostat trimmer screw @ gradually with a win, until the bulb is dimly lit; then push the stiding control lever forward in order to obtain optimum light intensity. (Fig. 1)
Fig. 1
2. Interpupiltary Distanm and Diopter A d j m n i s
1)
Insert
lhe
evwieae
@ with cross hairs into the right eyepiem tube positioning slot 21
-, of the binocular t u b , aligning the
Q and positioning an @ i ~ i ~
*
When the eyepiace positioning pin is i n e into the
Ioww dot on the tuba, tha m in the eyepieca
coincida
with the vibration direction of polarizer and analyrer at 0' settings. When slot, the mass lines
In- into the other am at 4 g to the direction of vibration. (This is the same with the monwlar tube.)
Them iwrt the eyepiece into the left tube.
Fig. 2
11
2) Looking through the right eyepiece (with cross hairs) with your right eye, rotate the diopter adjustment ring @ until the
cross
hairs are sharply focused.
(Fig. 3)
3) Looking through the both eyepiems with both
eyes,
adjust the lnterpupillary distance, sliding the knurled dovetail slid- @ of the right and left eyepiece tube,
until parfect binocular vision is obtained.
4) Memorize your interpupillary distanm wring by reading the scale @.
Fig. 3
5 ) Rotate the tube length adjustment ring @ on the right eyepiece tube to match your interpupillary distance wtting which you obtained from the scale.
5) at the image through the right eyepiece with your right eye and focus on the specimen with the coarse and fine adjustment knobs.
7) Look a t the image #rough the left eyepiece with your left eye and rotate the tube length
adjustment
r h g @ to
fucm
on the specimen without using
the
coarse and fine adjust-
ment
knobs.
3. Polariter Alignment
1) Push the analyzer @ into the light path, and maki sure that bod1 polarizer and analyzer are set at posi- tion
"0"
the "Crossed filter" position. Then
loosen
the clamping screw Q of the condenser.
(Fig. 4)
2) R e m * the specimen out of the light path
so
that a transparent area comes into the light path. Kwping the polarizer at
the
"0" position, rotate the polarizer rotation ring @ until the optimum extinction Is obtained, then clamp the ring. (Fig. 4)
Fig. 4
4. Cenbring aha Stage
1 1 Lmking through the eyepi-
and
ob-ive I O X , determine some pwtlarbr point, s you like,
In
the specimen image and coincide this point with the center of the
cross
hairs of the eyepiece.
Rotating
the stage, coincide the center of the rotation of a specimen pojnt with the can- ter of the crass haln by
means
of
the
two centering screws Q.
(Fig.
5)
*
Repear
this prcicdwe untif the m a t i o n is s e a i d .
Circular path of
5. antaring the O$@tim
This
centration is required
for
all PO objectives
except the
objective PO 1 OX.
1)
Imrt
a centering wrench into each centering swew of the
nosapiece.
(Fig. 8)
2 ) B y means of the two
wrenches,
coincide the center of the GTW hairs to the rotation center of the specimen.
I
C
Fig, 6
3) After all objectives
are centered,
remove the centering wrenches.
6. Clae of Aperture Iris Diaphragm
Adjust the opening of the aperture iris diaphragm m r d i n g to various conditions
such
as the numerical aperture of .th objective, i m a s contrast, depth of focus,
and
f l a t n a of field. Generally it is aften prefem b k
to
stop d w v n the aperture iris diaphragm to or 80% of the N. A. of the
objectiw.
70$6
After the tube, eyepisce is mrn& if necmary, look from
through
the &sewation the observation tube and check the opening at the objectiw pupil. of the aperture diaphragm
7. Focusing Mjusfment
1)
Tension
adjustment of coarse adjustment knobs
A Wnsion adjustment ring @ is pmvidd next to the right hand
coarse
adjustment knob. With this device the tension of the maw adjustment is freely adjust- able for either heavy or light movement depending on operator preferma. (Fig. 7)
However, do not l o o m the tension adjustment riw too much,
becausa
the stage drops, or the fins ad- justment
knobs
slip easily. f Be carefut not to mtate the right and I& m r s a adjustment knobs in the opposite directions simul- tanerwdy.
-
-r--l$
-
'
1. 5;
--
. t
Fig. 7
Crr
2) Pre-focusing
Iever
This lewr @ Is
locked
after mars focus has h n aecompiistaed. It prevents further upward travel of the stqe by means of the warse adjustment knobs, and automatically provides a limiting stop if the stage is lowered and then raised again. (Fig. 81
8. Orthaseodc Observation
Fig. 8
1) Swing out the top lens of the condenser.
In princfple, polarized light enters the light path paraitel to the optical axis, to enable obsewatjon of the optical characteristics of the m i m e n . Howww, this method will darken the field of view and lower the resolving power of the objective extremely. Them fore, swing out the top bns of the mndenser, using only the lawer aperture of the lower cundenser lens.
2)
Insert
the analyzer into the light path, and attain crossed filaer position with analymr
and
polarizer at o0 setting. At this posltion, the polarizer vibration is in the nortbouth direction, and the analyzer vibration in the east-west direction. To open the f i k tion, pull out the analyzer roQtion m.
3) the stage until extinction of the image is attained.
From this position, rotate the by 45' to obtain the d i w n a l posi'tion, at which position, the retardation angle is m u r e d .
4) the quarter wave plate or sensitive tint plate into the slot in the intermediate polarizing tube.
*
A Berek compensator is optionally available to measure the birefringence of a specimen.
Sensitha tint pbts
-
Quarter warn plata
7 ) Swing in the top
lens
of the
.wndenser,
and illuminate the mcirnen with no nmd to irnmerm betwen the condenser and specimen slide.
2) Bring the specimen inlo foeus, rotate the Bertrand lens turret ring into the IN position.
3)
Focus
on the interferenw figure formed a t the back fowl plane of the objectim from
20X to 1wx.
The
pinhob
cap provided may be used in place of the eyepiem to d i m l y view the interference figure m~ntioned In this case, the Bertrand lens is disengaged.
VI.
OPTICAL
DATA
ObJective Magnification
W. D. (mm)
F w l length (mml
Resolving pwver (id
K5X
(Field number 21)
WFlOX
(18)
Total magnification
Focfll demh IN)
Field of view Irnrn)
Total magnification
Focal depth ( p )
Field of viaw (mm)
P04X
0.10
18.77
28.45
3.4
20X
300.0
5.25
40X
172.6
4.5
P010X
0.25
6.78
16.08
1.3
50X
48.0
2.1
1KiX
27.60
1.8
W 2 0 X
0.110
1 -58
8.1 3
0.84 lOOX t 556
1.05
2 0 0 X
9.19
0.9
P040X
0.65
0.61
4.33
0.52
(Spring loaded)
200X
4 .W
0.53
400X
3.03
0.45
* W 1 0 0 X
1.30
0.1 1
1.81
0.26
(Spring loaded)
500X
1.05
0.21
1 ,000X
0.66
0.1 8
Immersion
objective. Resolving power is obtained when the objective is used at full aperture diaphragm.
T k eyepieces K5X and WFlOX incorporate a sliding eye
shield.
This shield can be pulled our to prevent glare and
loss
of contrast caused by ambient tight hitting the we.
W.D. (Working distance):
The
distance benveen
the s w i m e n crr
cover
glass and
the
nearsst point of the objective.
0 N.A. (Numerical aprture):
The numerical aperture represents a performance
number
which could be compared to the relative aperture (f-number) of a
camera lens.
M.A. values can t~ used for dimtly comparing the m l v i n g powers of all types of objectives. The larger N.A, the higher
the
resolving power.
0 Resolving power:
The ability of a lens to register small details. The resolving power of a tens is measured by its ability to separate two points.
0
Foal depth:
The distance between the upper and lower limits af sharpness in the image f o n d by an optical system.
0 Field number:
A number that represents the diarne'ter in mm of the image of the field diaphragm that is f o r d by the lens in front of it.
0 Field of view d i a d a r : The actual size of the field of view in mm.
Troubles
1. Optical System
Causes Remedies
(a) With the illuminator The switched on, condenser is lowerd exces- Raise the condenser to the upper the field sively. limit. of view cannot be seen.
Analyzer and polarizer are
"crossed filter" position in the Set them at the position
"0:90"
("0:O").
*'90:0".
(bl The field of view i s cut The nosepiece is
not
click stopped. Slightly rotate the nosepiece until off or illuminated irreg- it clicks into position. ularly.
The
condenser
is not correctly R e - i n ~ r t condenser all the way. mounted on the ring mount.
The sensitive tint plate is stopped Push the plate all the way until it midway. clicks.
In case of orthoscopic observation, Swing it out of the light path. the condenser top lens stays in the light path or stops midway.
(c 1 Dust or dirt is visible in Dust or dirt on the glass surface at the field of view. the light exit on the base.
Dust on condenser top lens.
Clean off the dust or dirt.
Dirty specimens.
Dust on eyepiece.
(d) Excessive imagecontrast. She condenser is lowered exces- Raise the condenser. sively.
The aperture iris diaphragm is stop- Open the diaphragm. ped down excekively.
(e)
0
Resolution problems:
0 Image i s not sharp.
0 Insufficient contrast.
The objective is not correctly posi- Slightly rotate the nosepiece until tioned in the lighf path.
Image
details
lack defini- Dirt on objective front lens. tion. it clicks into position.
Clean the objective.
The immersion objective is used Apply immersion oil. without immersion oil.
Bubbles in the immersion oil. Remove bubbtes.
The Olympus designated oil i s not Use the designated oil. used.
Troubles Causes
D i r t y specimen.
Remedies
Clean.
Dust on condenser lens.
( f ) The field of view is par- tially out of focus.
The objective is n o t correctly po- sitioned in the light path.
Slightly rotate the nosepiece until it clicks into position.
The specimen i s n o t correctly posi- tioned on the stage.
Place the specimen o n the stage and secure it w i t h the specimen clips.
I
(g) The image goes out of focus eccentrically.
The objective is n o t correctly posi- tioned in the light path.
Stightly rotate the nosepiece until it clicks i n t o position.
(h) When objectives are changed, they are n o t parfocal.
The mechanical tube length is n o t correctly adjusted.
Adjust w i t h the tube length adjust- ment rings o n the observation tube.
1
( i ) Light intensity does n o t increase although the voltage is raised.
The condenser i s lowered exces- sively.
Raise the condenser.
Re-mount them correctly.
( j ) The condenser does n o t come to the correct posi- t i o n for optimum extinc- tion.
The observation tube and mndens- er are n o t correct1 y mounted.
I
I
(k) No conoscopic image can
I
The condenser t o p lens is n o t in
(
Swing it in. be seen. the light path.
1
1
1
( I ) The crossed fiiter p o s i t i o n is n o t attained.
I
2. Electric System
The analyzer i s out of the light path.
Push it in.
(a) The illuminator is t o o I bright (or too dark).
The rheostat trimmer screw is not matched to the mains voltage.
Adjust the trimmer screw t o match the mains voltage.
The mains voltage is too high (or t o o low).
Adjust the mains voltage with a variable voltage transformer.
I
I
The rheostat trimmer screw is n o t correct lv adjusted.
Adjust it correctly.
(b) O u t p u t voltage for the il- luminator cannot be reg- ulated.
The voltage selector switch is n o t matched to the mains voltage.
The mains voltage is too l o w or too high.
Adjusr the mains voltage selector
I switch t o the mains voltage.
Adjust the mains voltage w i t h a variable voltage transformer,
1
Troubles Causes Remedies
-.
Use a variable voltage transformer.
( c ) The light flickers and the intensity is unstable.
The mains voltage is unstable.
The filament o f the bulb is likely to burn out.
Replace the bulb.
I Loose electrical connection.
( d ) Fuse burns out t o o o f t e n .
I
The fuse is not a standard fuse.
Secure the connection.
I
Use a standard fuse.
(el Reduced bulb life
The voltage selector switch is not Match the switch t o the mains matched to the mains voltage. voltage.
-
The voltage selector switch is not matched t o the mains voltage.
Match the selector switch to t h e ' mains voltage.
The bulb is not a standard bulb.
I
Use a standard bulb.
1
Mains voltage i s too high.
1
Use variable voltage transformer. i3. Focusing
I
I
1
{a) C o a r ~ too tight.
1 1
Tension adjustment ring is tighten- Loosen the tension adjustment ring ed too much. properly.
,
I
I The user i s trying t o raise the stage over the upper focusing limit imposed by the engaged pre-focus- ing lever.
I
Unlock the pre-focusing lever.
I
(b) The stage drops and the specimen goes o u t of focus.
The tension adjustmer~t ring i s too loose.
Tighten the ring properly.
{ c ) The stage cannot be raised to the upper limit.
Pre-focusing lever is engaged in lower than focusing position.
Unlock the pre-focusing lever.
I
( ( d l The stage cannot be low- ered t o the lower limit of the working range.
{ e ) The objective front lens hits against the specimen.
The condenser mount is lowered too much.
The specimen is mounted o n the stage upside down.
Raise the condenser mount
Reverse the specimen.
4. Observation Tube
- -
(a1 Incomplete binocular vi- lnterpupillary distance is not cor- Correct the interpupillary distance. sion.
I rectly adjusted.
I LA
Troubles Causes Remedies
Diopter adjustment is incomplete. Complete the diopter adjustment.
Right and left eyepieces are not matched.
Use pair of matched eyepieces.
The user is unaccus?omed to binocular observation.
Prior t o looking a t the image of the specimen, try to look at a far away object.
5. Stage
(a 1 The image easily goes out of focus when you touch the stage.
The stage is not correctly clamped. Clamp the stage securely.
(b) The specimen stops mid- way on the east-west traverse.
The specimen is not correctly posi- tioned on the stage.
Adjust the specimen position.
advertisement
* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project