Quick start guide EasyClean 400
__________________________________________________________________________
Quick start guide for EasyClean 400 system installation
This checklist has to be used as a guideline to install and set up an
EasyClean 400 during commissioning. Please proceed step by step according to the 10 steps below and refer to the required instruction manuals for details.
Step
1.
Hardware and supplying installation
!
Installation EC 400 (Mounting of the system)
!
Installation M 700 (Mounting of the Transmitter)
!
Installation InTrac housing and sensor
!
Supplying installation (Power, air, water)
Step
2.
Installation of EC 400 and EC 700 module
!
Install Media adapter, Metering pump, and Media connection
!
Fill buffer and cleaning agent canister
!
Electrical installation (Metering pumps, Probe position, M 700, possibly DCS wiring and optional emergency button)
!
Wiring EC 700 module (Sensor cable and communication)
Step
3.
Tubing Housing
!
Tubing and installation of media connector on the housing.
Attention: Pneumatic position indicators: P ! feed
!
Tubing of InTrac housing (pneumatic connections for housing drive)
Manual
EC 400,
Page 11-14
M700,
Page 24-25
InTrac 7XXe,
Page 48
EC400, P 16
M700, P 22
Manual
EC 400,
Page 17-25
EC 400,
Page 26-30
EC 700,
Page 21-23
Manual
EC 400,
Page 31-35
InTrac 7XXe
Page 53
Step
4.
System start up
!
Power on
!
Execute automatic Hardware detection on EC 400
Parameter menu: Parameter setting / System control / Factory
setting EC 400! Plug & Play
!
Check logbook:
Parameter menu: Diagnostics / Logbook
Trouble shooting according to error list if required
Manual
EC 400,
Page 51
EC 700
Page 120-
123
__________________________________________________________________________
Mettler-Toledo GmbH, Process Analytics Page 1 of 2
Quick start guide EasyClean 400
__________________________________________________________________________
Step
Installation software settings
5.
Parameter menu: Parameter setting / EC400 / Installation
• Measurement procedure (continuous, short time)
• Monitoring
• Housing (if required, change displaced volume)
At the end:
!
Execute “Start up” (Tubes will be filled automatically)
Manual
EC 700,
Page 97-103
Manual
EC 400,
Page 53
Step
Manual Control and Function test
6.
Parameter menu: Maintenance / EC 400 / Manual control
!
Check housing, pumps and valves manually
Step
Adjust software programs
7.
Parameter menu: Parameter setting / EC400 / Program flow
!
Adjust programs to the customers need regarding to the application specification sheet.
Manual
EC 700,
Page 86-96
Step
8.
Define control mode
Parameter menu: Parameter setting / EC400 / Time control
!
Choose between „Fixed interval“ or „Week program”
!
Set the appropriate timing intervals
Attention: If the system is controlled via DCS, no time control is needed. Programs will be activated on event via DCS.
Step
Test run
9.
Parameter menu: Calibration / EC400/ Start program
!
All EasyClean 400 programs can be executed manually
Manual
EC 700,
Page 84-85
EC 400
Page 37-41
EC 700
Page 98-100
Manual
EC 700
Page 35
Step
System optimisation and customer approval
Manual
10.
!
Optimisation of the system (Timing, program etc.) if required
!
Staff training and customer approval
__________________________________________________________________________
Mettler-Toledo GmbH, Process Analytics Page 2 of 2