ROVA-N5/10/20/50L Rotary Evaporator MRC.1.16

ROVA-N5/10/20/50L Rotary Evaporator  MRC.1.16
Operation Manual
Rotary Evaporator
ROVA-N5/10/20/50L
PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529
[email protected]
MRC.1.16
Content
Safety instructions ------------------------------------------------------------------------------------- 1
I. Purposes and features ----------------------------------------------------------------------------- 1
II. Technical parameters ------------------------------------------------------------------------------ 2
III. Instructions for main parts ------------------------------------------------------------------------ 4
3.1 ROVA-N5L complete machine drawing ------------------------------------------------ 4
3.2 ROVA-N10L complete machine drawing ------------------------------------------------- 4
3.3 ROVA-N20L complete machine drawing ------------------------------------------------- 5
3.4 ROVA-N50L complete machine drawing -------------------------------------------------- 5
IV. Instructions for panel ------------------------------------------------------------------------------ 6
V. Preparations for installation ---------------------------------------------------------------------- 6
VI. Instructions for installation ----------------------------------------------------------------------- 7
6.1 Upright pipe installation ----------------------------------------------------------------------- 7
6.2 Rotary shaft installation ----------------------------------------------------------------------- 7
6.3 Three-way flask installation ------------------------------------------------------------------ 8
6.4 Glass rotary flask ------------------------------------------------------------------------------- 8
6.5 Installation of auxiliary condenser and three-way flask ------------------------------- 9
6.6 Receiving flask installation ----------------------------------------------------------------- 10
6.7 Installation of main condenser and the rest parts ------------------------------------ 12
VII. Pipeline connection ----------------------------------------------------------------------------- 13
VIII. Operation ----------------------------------------------------------------------------------------- 14
8.1 Charging method ----------------------------------------------------------------------------- 14
8.2 Operating equipment ------------------------------------------------------------------------ 16
8.3 Emptying ---------------------------------------------------------------------------------------- 16
8.4 Halt ----------------------------------------------------------------------------------------------- 16
8.5 Treatment after halt -------------------------------------------------------------------------- 16
IX. Maintenance --------------------------------------------------------------------------------------- 16
XI. Failure cause and troubleshooting ----------------------------------------------------------- 18
Safety instructions/Purposes and features
Safety instructions

Please make sure the power connection is correct (see the technical parameters for
details) and well-grounded.

Cooling water pipeline and extraction line shall be maintained unobstructed without dead
bend in the running.

Method of rotational forward-push shall be adopted for the installation of pipeline for
glass parts to avoid damages caused to the glass parts.

If inflammable specimen or organic solvent is applied, please make sure no solution
splashing out and take all fire precaution measures

Apply a rag to wipe the glass parts clean after washing away the stubborn stains; do not
use hard objects to hit the glass.
I. Purposes and features
R-series large rotary evaporator is mainly used in the small-scale test, pilot-scale test and
production in biological, pharmaceutical, chemical, food and other fields. As with large capacity
and large-diameter rotary evaporating flask, when it is placed in a water bath, along with the
simultaneous performing of rotating and heating, it allows the solution spreading and
evaporating rapidly. What’s more, the product may be fitted with the multi-purpose water
circulating vacuum pump, diaphragm vacuum pump, low-temperature circulating (vacuum)
pump, circulating cooler, constant-temperature circulator, low-temperature cooling liquid
circulating pump and so on to form a system device.

Teflon(PTFE) and fluororubber dual spin seal, patented technology to ensure the high
vacuum.

Accurate and reliable temperature control system,.

On the premise of not affecting the vacuum and solution distillation, continuous receiving
allowable.

PTFE emptying valve, corrosion-resistant without any pollution.

Including two types: regular type and explosion-proof type, which can be adopted based
on the experimental environment.
1
Technical parameters
II. Technical parameters
Model
ROVA-N10L
ROVA-N5L
ROVA-N20L
Optimum ambient
temperature (℃)
Power (V/Hz)
5~25
Single-phase 220/50
Speed regulation
of main engine
Digital
variable-frequency
stepless speed
regulation
Rotate speed of
main engine (rpm)
10-140
Rotary motor
Digital direct current stepless speed regulation
20-130
Condenser pipe
Evaporative
capacity (L/h)
Digital temperature control, constant temperature -99℃
399.9Pa(under 3mmHg)
5
Flange mouthØ50
3
Vertical dual coil
pipe
Water ≥2
Alcohol ≥4
10
Flange mouthØ95
20
Flange mouthØ125
5
10
50
Flange mouth Ø125
20
Upright, main condenser + auxiliary condenser, efficient
tri-backflow condenser pipe
Water ≥3.2
Alcohol ≥6.5
Water bath pot
material
Water bath pot
size (mm)
20-110
250
Maximum vacuum
degree (Pa)
Receiving flask (L)
Three-phase 380/50
40
Temperature
control (℃)
Rotary flask (L)
ROVA-N50L
Water ≥5
Alcohol ≥11
Water ≥9
Alcohol ≥19
Stainless steel
Φ300×170
Lifting function
Φ350×220
Φ450×260
Φ550×320
Electric + manual
lifting
Electric lifting
Lifting stroke (mm)
0-150
0-160
0-190
0-180
Complete machine
power (kW)
2.2
4.8
6.3
8.3
765×400×1070
920×550×1700
1250×600×2100
1320×770×2340
Dimension (mm)
2
Instructions for main parts
III. Instructions for main parts
3.1 ROVA-N5L complete machine drawing
1.Rotary evaporator: main engine+
water bath kettle
2. Upright pipe
3. Rotary flask 5L
4. Rotary flask nut
5. Nose
6. Three-way flask retaining nut
7. Vacuum meter bracket
8. Vacuum meter
9. Condenser clamp holder
10. Silicon rubber binding
11. Condenser
12. Three-way flask
13. Condenser pallet
14. Condensate connector
15. Receiving flask
16. Receiving flask pallet
3.2 ROVA-N10L complete machine drawing
1. Rotary evaporator: main engine +
water bath kettle
2. Upright pipe
3. Rotary flask 10L
4. Rotary flask retaining nut
5. Nose
6. Three-way flask retaining nut
7. Vacuum meter bracket
8. Vacuum meter
9. Condenser clamp holder
10. Suction flange
11. Main condenser
12. Condenser clamp
13. Auxiliary condenser
14. Three-way flask
15. Condenser pallet
16. Recycling flask
17. Recycling flask pallet
4
Instructions for main parts
3.3 ROVA-N20L complete machine drawing
1. Rotary evaporator: main engine +
water bath kettle
2. Upright pipe
3. Rotary flask 20L
4. Rotary flask retaining nut
5. Nose
6. Three-way flask retaining nut
7. Glass elbow
8. Vacuum meter bracket
9. Vacuum meter
10. Condenser clamp holder
11. Suction flange
12. Main condenser
13. Condenser clamp
14. Auxiliary condenser
15. Condenser pallet
16. Recycling flask
17. Recycling flask pallet
3.4 ROVA-N50L complete machine drawing
1. Rotary evaporator: main engine+
water bath kettle
2. Rotary flask 50L
3. Rotary flask retaining nut
4. Nose
5. Three-way flask retaining nut
6. Vacuum meter bracket
7. Vacuum meter
8. Upright pipe
9. Condenser clamp holder
10. Suction flange
11.Condenser clamp
12. Main condenser
13. Main condenser pallet
14. Three-way flask
15. Auxiliary condenser
16. Auxiliary condenser pallet
17. Recycling flask
18. Recycling flask pallet
5
Instructions for panel
IV. Instructions for panel
Panel I:
1. Setting temperature-down value▽
2. Setting switch SET
3. Setting temperature-up value△
4. Power indicator lamp
5. Heating allowable indicator lamp
6. Power switch
7. Heating switch
8. Rotary switch
9. Water bath down key
10. Alarm indicator lamp
11. Water bath up key
12. Heating output indicator lamp
13. Rotate speed regulation knob
Panel II:
1. LCD display
2.“ ” key:Click on this key to start or shut down
the rotation function;
3.“▲”key:Click on this key, the bath intermittent
rise, long press this key, the bath continuous rise;
4.“ ” key:Click on this key to start or shut down
the heating function;
5.“▼”key:Click on this key, the bath intermittent
decline, long press this key, the bath continuous
decline;
6.Code switch:Click on this key to set the
temperature, the speed setting; Long press
to set the temperature, speed and other
internal parameters.
V. Preparations for installation
1. Please refer to the packing list to check whether the components and parts are complete
after opening the packing box; if there is any missing part, please contact us.
2. Remove residues on the glass parts before the assembly and keep the glass contact
surface clean; apply vacuum grease to both sides of the seal ring before the installation.
3. Tools that might be used in the installation including allen wrench, screwdriver and so on
shall be well prepared.
6
Instructions for installation
VI. Instructions for installation
6.1 Upright pipe installation
Rotating in clockwise direction, the upper upright pipe is connected with the lower upright
pipe by screw threads (as shown below), on which, the cross clamp, vacuum meter bracket
and condenser clamp holder are successively installed.
1. Nose shell
2. Vacuum meter holder
3. Condenser clamp holder
4. Vacuum meter bracket
5. Upper upright pipe
6. Cross clamp
7. Lower upright pipe
8. Base seat
Upright pipe installation drawing
6.2 Rotary shaft installation
Set the O-ring on the glass rotary shaft which shall be placed into the nose with the O-ring
against the nose shell. (As shown below)
1. O-Ring
2. Glass rotary shaft
3. Nose shell
Rotary shaft installation drawing
7
Instructions for installation
6.3 Three-way flask installation
Set the retaining nut on the lower bore of glass three-way flask and set the retaining nut
check ring on the lower neck of glass three-way flask; then insert the flange seal washer into
the three-way flask bore and insert the three-way flask with the seal washer into the upper
internal opening of the glass rotary shaft, ensuring the center of three-way flask in the
horizontal direction on the left and properly tightening the retaining nut. (As shown below)
1. Three-way flask
2. Three-way flask retaining nut
3. Three-way flask retaining nut check ring
4. Flange seal washer
5. Nose shell
Three-way flask installation drawing
6.4 Glass rotary flask installation
Set the Teflon seal ring on the assembled glass rotary shaft as shown in the drawing and
then set the rotary flask nut on the bottom flask with the check ring on the flask neck. Rotary
flask positioning: apply the allen wrench to insert into the hole (inclined left) of nose seat,
turning the rotary shaft to make sure the shaft hole coincided with the seat hole, no inserting of
allen wrench when the rotary shaft cannot be twisted any more. Last, install the rotary flask
with retaining nut and check ring on the glass rotary shaft, tightening the rotary shaft nut. Turn
the rotary shaft, ensuring the rotation steady and consistent with the shaft center.
1. Coupling
2. Teflon seal ring
3. Check ring
4. Rotary flask nut
5. Rotary flask
Glass rotary flask installation drawing
8
Instructions for installation
6.5 Installation of auxiliary condenser and three-way flask
a. Connection of ROVA-N10L/ROVA-N50L condenser with three-way flask
First, secure the pallet on the upright pipe and then place the auxiliary condenser (ROVA-N10L
and ROVA-N50L provided with auxiliary condenser) on the mounted pallet, adjusting the height and
ensuring the side connecting mouth of auxiliary condenser and three-way flask connecting mouth at
the same level, fix the pallet and cross clamp screw .Then set the connecting flange on the bottom
connecting mouth of auxiliary condenser with the check ring on the connecting mouth neck of
auxiliary condenser. In the same way, set the other connecting flange and check ring on the threeway flask connecting mouth, place the flange seal washer in any of connecting bores of three-way
flask or auxiliary condenser and finally tighten the retaining screw and nut. (As shown below)
Note: ROVA-N5L is not provided with auxiliary condenser, for which, the below method shall be
directly adopted to connect the condenser and three-way flask.
1/3. 60# connecting flange
2. 60# flange seal washer
4. Three-way flask
5. 60# flange check ring
6. Auxiliary condenser
7. Cross clamp
8. Pallet
Auxiliary condenser and three-way flask installation drawing
b. Connection of ROVA-N20L auxiliary condenser with three-way flask:
First set the connecting flange on the bottom connecting mouth of three-way flask and set
the check ring on the neck of three-way flask, insert half of the flange seal washer into the
three-way flask bore and then set the connecting flange nut and check ring on one end of glass
elbow, finally screw the nut and bolt together, adjust the condenser pallet to an appropriate
height with the glass elbow in right direction, and in the same way, connect the auxiliary
condenser and glass elbow with connecting flange and check ring which shall be tightened at
last.
9
Instructions for installation
1. Auxiliary condenser
2/7. 60# flange seal washer
3. Glass elbow
4/6/8/10. 60# connecting flange
5/9. 60# flange check ring
11. Three-way flask
Auxiliary condenser and three-way flask installation drawing
6.6 Receiving flask installation
a. ROVA-N5L receiving flask installation
Place the receiving flask on the pallet and lift it to a certain height; set the connecting
flange on the bottom connecting mouth of receiving flask with the check ring on the neck, insert
half of the flange seal washer into the receiving flask bore and then in the same way, set the
connecting flange bolt and check ring on the lower connecting mouth of condenser connector;
finally insert the lower interface of condenser connector into the other half of the flange seal
washer and screw the nut and bolt together. Afterwards, in the same way, successively set the
connecting flange and check ring on the lower connecting mouth of condenser and upper
connecting mouth of condenser connector, with half of the seal washer inserted into the
condenser connector bore. Then raise the height of receiving flask of the fixed condenser
connector, ensuring the other half of the flange seal washer inserted in the condenser bore to
secure the receiving flask pallet; tighten the connecting flange and make sure that the
connecting mouths of condenser, condenser connector and receiving flask are closely
meshing with each other without strain so as not to break the glass.
10
Instructions for installation
1. Condenser
2/4/6/8.40# connecting flange
3/7. 40# flange seal washer
5. Condenser Connector
9. Receiving flask
10. Receiving flask pallet
11. Emptying valve
12/13. 40# flange check ring
Receiving flask installation drawing
b. Methods for connection of ROVA-N10/20/50L auxiliary condenser with receiving flask
Place the receiving flask on the pallet and lift it to a certain height with the cross clamp
firmly fixed, then set the connecting flange on the connecting mouth neck of receiving flask and
put the automatic switching valve in the flange seal washer. (Note: Automatic switching valve
plane toward recycling flask, avoid to be inverted.) The connecting flange and check ring shall
also be successively set on the lower connecting mouth of auxiliary condenser. Then regulate
the height of receiving flask pallet, ensuring the other half of the switching valve right inserted
in the lower connecting mouth bore of auxiliary condenser. Secure the cross clamp and tighten
the bolt and nut. (As shown below)
11
Instructions for installation
1. Auxiliary condenser
2. Auxiliary condenser pallet
3/5. 50# connecting flange
4. 50# flange check ring
6. Receiving flask
7. Receiving flask pallet
8. Cross clamp
9. Upright pipe
10. 50# flange seal washer
11. Automatic switching valve
Receiving flask installation drawing
6.7 Installation of the rest parts
a. Connection of ROVA-N10/20/50L main condenser w ith auxiliary condenser as w ell as
connection of exhaust nozzle also includes connecting flange, check ring and flange seal
washer with the mounting method as described above. (As shown below)
b. Relief valve installation: it shall be directly installed in the receiving flask bore.
c. Vacuum meter installation: secure the vacuum meter on the vacuum meter holder and then
together into the vacuum meter bracket and tighten the retaining screw, adjusting the
holder to an appropriate height to enable operators to obtain vacuum meter readings by
horizontal check.
1. Retaining clamp
2. Main condenser
3/5. 80# connecting flange
4. 80# flange seal washer
6. Auxiliary condenser
7.80# flange check ring
Auxiliary condenser and main condenser installation drawing
12
Pipeline connection
VII. Pipeline connection
ROVA-N5L pipeline connection
ROVA-N10L pipeline connection
13
Operations
VIII. Operations
8.1 Charging method
The rotary evaporator is pumped into the state of negative pressure, with a hose to the
feeding valve is connected with the sample, open the feeding valve, sample will be pumped
directly into rotary bottle, feeding close after charging valve, adjusting rotary bottle height,
which is placed in the water bath pot.
8.2 Operating equipment
8.2.1 Panel I:
1. Add pure water into the water bath (submerged rotating bottle of 1/2, not overflow),
connecting the power supply with the power indicator lamp on.
2. Click on the “Power” key, the window will display the present measured temperature in
water bath.
3. Click on the “SET” key to enter the temperature setting state; then click the up key “▲” or
down key “▼” to increase or decrease the thermal value until to get the desired testing
temperature. Afterwards save settings with a click of “SET” key and exit the setup program.
(One number will be modified by one click of up key or down key, with long press resulting
in continuous numerical jumping).
4. Click on the “
” key with the heating allowable indicator lamp on. When the set
temperature is higher than or equal to the actual measured temperature, the heating
operation proceeds with the heating output indicator lamp on; when the measured
temperature is 5℃ higher than the set temperature, the alarm indicator lamp will be lit with
the buzzer beeping (Click the up key or the down key to adjust the sound).
5. Click on the “
” key and the characters on the rotate-speed display window stop blinking,
obtain the desired rotate speed by adjusting the rotate-speed knob.
8.2.2 Panel II:
1. Add pure water into the water bath (submerged rotating bottle of 1/2, not overflow),
2. Turn on the power switch, the temperature display shows "CC2P", the speed display
shows "S1.1", all identifiers light; after 4 seconds entering the normal display.
14
Operations
3. Temperature、speed setting
① Click on the "code switch" button, the temperature display flashing display
temperature settings, and then click on the "code switch" button, the speed display
shows flashing speed setting, rotary the "code switch" can modify the settings
(dextral increase , L decreases). After modify the desired setting and then click on
"code switch" button, the setting is automatically saved, the controller exits the
current state.
" key to start heating control, ○
S identifier lights, heating output,
② Click on the "
HEAT identifier lights, and then click on the "
" key, stop heating control,
S identifiers off.
○
③Click on the "
the "
" key to start the speed control, RUN identifier lights, and then click on
" key to stop the speed control, STOP identifier lights,
4.Click on the "▲" key, bath intermittent rises, click "▲" key, bath consecutive rise.
5.Click on the " ▼ " key, bath intermittent decline, long press " ▼ " key, bath consecutive
decline.
6. System self-tuning
Click on "
" key to start heating control. In the non-state,press "code switch" button
for 3 seconds,temperature window displays the password prompt "Lc",speed window
displays the password value. Rotate "code switch" button to modify the password to
"58", then click on "code switch" to enter into the self-tuning selection state, the
temperature window displays the prompt "AT", the speed window displays value.
Rotate "code switch", modify the parameters, if modify the value from "0" to "1", click on
"code switch" button, then enter the self-tuning, AT NOW identifier lights, self-tuning
ends, AT NOW identifier extinguished; If modify the value from "1" to "0", click on "code
switch" button, then stop self-tuning.

Note: when the temperature control effect is not ideal, the user can self-tuning. In
self-tuning process the temperature will have a large overshoot, in the system
15
Operations/Maintenance
self-tuning before users should give full consideration to this factor. In the system
of the system process if there is an over temperature alarm, ALM1 indicator
lights, heating alarm relay automatically disconnect.
8.3 Emptying
1. For the ROVA-N5L, close the valve of the condenser connector and open the exhaust valve on
the receiving flask to depressurize the internal receiving flask, and then open the emptying
valve to discharge the materials.
2. For ROVA-N10/20/50L, open the exhaust valve on the receiving flask to depressurize
the internal receiving flask, and then
open the emptying valve to discharge the materials.
8.4 Halt
1. Stop the specimen bottle from rotating, adjust the lift height and remove the specimen
bottle out of water bath.
2. Open the feed pipe tap, eliminating the inside vacuum.
3. Turn the decompressor off.
4. If no specimen is added, please turn off the cooling water circulating device along with the
water bath. (*As the water bath or specimen bottle would not immediately cool down after
the concentration, please be careful with the high temperature).
5. Rotary flask disassembly: apply the cross screwdriver to insert in the hole of nose seat
(inclined left) to secure the rotary shaft, loosen the clamp nut with bottle-removal device
and detach the rotary flask.
8.5 Treatment after halt
For long periods without using it, please shut off the power switch and disconnect the
power.
IX. Maintenance
1. Please shut off the power switch before the maintenance, disconnecting the power cord
and power.
2. Please use wrung-out wet soft cloth to wipe clean. Stubborn stains shall be cleaned by
neutral detergents which shall be then wiped off with a rag.
3. The maintenance of internal electrical and heating parts must be conducted by
professionals or trained electricians.
4. Do not directly splash water over the product or use abrasive powder, diluent, oil,
kerosene, acidic material and similar substance during the site clearing, or else the shock
or other accidents would occur.
16
Cause and troubleshooting
XI. Cause and troubleshooting
Failure
Cause
Troubleshooting
Connect the power supply, the
power indicator lamp doesn't
light or LCD panel does not
light.
Check the power line and find out whether it’s
through or misconnected.
Connect the power line.
Failure of power switch.
Replace power switch.
The power indicator lamp lights
or rotation allows identifier lights,
but does not rotate.
Failure of motor.
The heating output, but not
warming up.
The solid state relays fault.
Replace the solid state relays
Heater fault.
Replace heater.
The temperature display shows
up□□□
The bath of temperature measurement sensor
short circuit or open circuit.
Check sensor and wiring.
The temperature display shows
“Er-1”,ALM1 identifier lights
The bath of temperature measurement sensor
short circuit or open circuit.
The temperature display shows
“Er-2”,ALM1 identifier lights
The bath of protection sensor short circuit or
open circuit.
The temperature display shows
“Er-3”,ALM1 identifier lights
The bath of protection sensor exceeds the set
value.
The speed display shows
“Er-1”,ALM2 identifier lights.
Power module fault.
The speed display shows
“Er-2”,ALM2 identifier lights.
Motor stall.
The speed display shows
“Er-3”,ALM2 identifier lights.
Holzer logic error.
The speed display shows
“Er-4”,ALM2 identifier lights.
Bus voltage under voltage.
The speed display shows
“Er-5”,ALM2 identifier lights.
Bus voltage over voltage.
The speed display shows
“Er-6”,ALM2 identifier lights.
Serial communication fault
Stop using immediately and
contact us.
Abrasion of seal ring.
Replace the seal ring.
Abnormal noise
Failure of circuit board.
Failure of circuit board.
Abrasion of internal gear.
Lack of oil in drive part.
Failure of motor.
Leakage of pressure and poor
degree of vacuum
Abnormal lifting
Stop using immediately and
contact us.
Check sensor and wiring.
Restart the power switch.
Remove the rotary bottle,
no-load operation. The fault has
not been lifted, discontinue use
immediately and contact the
company.
Check the power supply
voltage value.
Stop using immediately and
contact us.
Abrasion of glass rotary shaft.
Replace the glass rotary shaft.
Abrasion of seal ring.
Replace the seal ring.
Improper installation of seal ring (Opposite
direction)
Remount the seal ring.
Aging of the seal ring of pressure-relief air tap.
Replace the air tap seal washer
Aging of vacuum hose.
Replace the vacuum hose.
Failure of circuit board or motor.
Stop operating immediately
and contact us.
Abrasion or rusting of sliding bearing.
18
Was this manual useful for you? yes no
Thank you for your participation!