Key facts
❙❙ Frequency range: 100 kHz to 1.6 GHz/3 GHz*1
❙❙ Spectral purity greater than -100 dBc/Hz (at 100 kHz)
❙❙ SWEEP from 20 ms to 1000 s
❙❙ Detectors: auto-, min-/max.-peak, sample, RMS, average, quasi-peak*2
❙❙ Miscellaneous marker/Δmarker and peak functions
❙❙ Tracking generator*3
Frequency range: 5 MHz to 1.6 GHz/3 GHz*1
Output level: -20 dBm to 0 dBm
❙❙ Directly export data to USB flash drive, RS-232/USB dual interface
for remote control
❙❙ Fanless design and fast boot time
*1 with R&S®HMS-3G (HV212) option
*2 with R&S®HMS-EMC (HV213) option
*3 with R&S®HMS-TG (HV211) option
R&S®HMS-EMC
R&S®HMS-3G
R&S®HMS-TG
Model overview
R&S®HMS-X with EMC Option
R&S®HMS-X basic unit
Amplitude measurement range
-114 dBm to +20 dBm
-104 dBm to +20 dBm
DANL
typ. -135 dBm
typ. -104 dBm
Resolution bandwidth
100 Hz to 1 MHz, 200 kHz (-3 dB),
200 Hz, 9 kHz, 120 kHz, 1 MHz (-6 dB)
10 kHz to 1 MHz, 200 kHz (-3 dB)
Video bandwidth
10 Hz to 1 MHz
1 kHz to 1 MHz
R&S®HMS-X
2 Rohde & Schwarz R&S®HMS-X
Your R&S®HMS-X Spectrum Analyzer
You can create your R&S®HMS spectrum analyzer by combining a basic unit with
any of three available options. In case of growing requirements, upgrade
vouchers allow you to upgrade your instruments with all options at any point in
time.
3 GHz
R&S®HMS-EMC
❙❙ This option activates all the functions that are
required for EMC precompliance measurements.
The preamplifier option has been integrated into the
new R&S®HMS-EMC option.
R&S®HMS-3G
❙❙ The frequency range is increased from
1.6 GHz to 3 GHz with this option.
R&S®HMS-TG
❙❙ This option activates the tracking generator
in the instrument.
We have used the first-class hardware from our HMS spectrum analyzer and
developed a new and flexible instrument concept. It can be individually
­configured, combined and upgraded for your applications.
HMS previous models
R&S®HMS-X
HMS1000E
HMS-X
HMS1000
HMS-X + EMC*
HMS1010
HMS-X + EMC* + TG
HMS3000
HMS-X + EMC* + 3G
HMS3010
HMS-X + EMC* + 3G + TG
* The preamplifier function is an integral part of the HMS-EMC option
Rohde & Schwarz R&S®HMS-X 3
EMC Precompliance
Not only do unexpected results in test labs during EMC compliance
measurements translate into extra costs, quite often they also cause a
substantial delay for your project. Rohde & Schwarz offers effective and costefficient tools for EMC precompliance measurements which allow you to
successfully prevent possible surprises before the actual onset of a problem.
Our HMExplorer software for your EMC measurements is included with every
HMS-X spectrum analyzer with activated EMC option.
EMC precompliance sets
Rohde & Schwarz offers product sets for your EMC precompliance
measurements, which include all necessary instruments to
analyse typical EMC problems. Depending on your requirements,
you can choose between a 1 GHz and a 3 GHz combination.
HMExplorer software for EMC
precompliance measurements
Spectrum analyzer R&S®HMS-X
1 GHz EMC-SET1
❙❙ Spectrum analyzer R&S®HMS-X incl. R&S®HMS-EMC option
❙❙ Probe set R&S®HZ530
❙❙ Line impedance stabilization network (LISN) R&S®HM6050-2
❙❙ HMExplorer software
3 GHz EMC-SET2
Differences to SET1:
❙❙ HMS-3G option additional
❙❙ 3 GHz probe set HZ540 instead of HZ530
4 Rohde & Schwarz R&S®HMS-X
Line impedance stabilization network
for line conducted measurements
LISN HM6050-2
1 GHz probe set
HZ530
3 GHz probe set
HZ540 (fig. similar)
Recommended Accessories
3 GHz VSWR bridge HZ547
Near-field probe set 3 GHz HZ540
This unit is used to measure the voltage standing wave ratio (VSWR) and
reflection coefficient of a device under test with an impedance of 50 Ω.
Typical test devices include attenuators, terminations, frequency switches,
amplifiers, cables and mixers.
Near field probe set for comparative measurements with built-in preamplifier
covering frequency ranges from 1 MHz to 3 GHz, designed for the 50 Ω
N-connectors of the HMS-X:
❙❙ E-field probe
❙❙ H-field probe
❙❙ High impedance probe
3 GHz VSWR bridge for R&S®HMS-X, option R&S®HMS-TG required, option HMS-3G recommended
HZ46
HZ99
R&S®HO732
R&S®HO740
HZ530
4 RU 19” rackmount kit
Carrying case for protection
and transport
Ethernet/USB dual interface
card
Interface IEEE-488 (GPIB),
galvanically isolated
Near-field probe set 1 GHz
Rohde & Schwarz R&S®HMS-X 5
Upgrade at any time
You can easily upgrade all three available options at any later point in time with
option upgrade vouchers available at your dealer.
The voucher number and the serial number of your R&S®HMS-X instrument
enable you to generate the respective licence key directly on our web page
http://voucher.rohde-schwarz.com.
Voucher Software Options
Voucher Software Options
Voucher Software Options
HV212
HV212
HV211
Activates EMC Option
Upgrade to 3 GHz
Unlocks TG
Accessories included:
Line cord, printed operating manual, CD, software
R&S®HMS-X options
Option code*1
Voucher code*2
EMC option incl. preamplifier
R&S® HMS-EMC
HV213
Bandwidth upgrade to 3 GHz
R&S® HMS-3G
HV212
Unlock built-in tracking generator
R&S® HMS-TG
HV211
*1 available only with purchase of R&S®HMS-X basic unit
*2 activate R&S®HMS-X options at any time after purchase of R&S®HMS-X basic unit
6 Rohde & Schwarz R&S®HMS-X
Recommended accessories:
HO732 Dual-interface ethernet/USB
HO740 Interface IEEE-488 (GPIB), galvanically isolated
HZ530 Near-field probe set 1 GHz for EMI diagnostics
HZ540 Near-field probe set 3 GHz for EMI diagnostics
HZ547 3 GHz VSWR bridge for HMS-X, incl. HMS-TG option
HZ46 4RU 19” rackmount kit
HZ72 GPIB-cable 2 m
HZ99 Carrying case for protection and transport
HZ520 Plug-in antenna with BNC connection
R&S®HMS-X Spectrum analyzer – Specifications
(from firmware version 2.250, all specifications at 23°C after 30 minutes warm-up)
Frequency
Frequency range
DANL (displayed average noise level)
(ref. level ≤-30 dBm, frequency range 10 MHz to 3 GHz)
Tracking generator output
Connector
N socket
Output impedance
50 Ω
basic unit
100 kHz to 1,6 GHz
10 kHz (RBW), 1 kHz (VBW)
with HMS-3G (HV212) option
100 kHz to 3 GHz
100 Hz (RBW), 10 Hz (VBW)
with HMS-EMC (HV213) option
-95 dBm (typ. -104 dBm)
-115 dBm (typ. -135 dBm)
Frequency range
Accuracy of the internal reference
±2 x 10-6
Temperature stability
±2 x 10-6 (0°C to +30 °C)
Preamplifier
with HMS-EMC (HV213) option
Aging (per year)
±1 x 10-6
Inherent spurious
Output level
Trigger input
Frequency counter (with HMS-EMC (HV213) option)
-124 dBm (typ.)
Resolution
1 Hz
ref.-level ≤-20 dBm,
f >30 MHz, RBW ≤100 kHz
Accuracy
±(frequency x tolerance of reference)
Input related spurious (mixer level ≤-40 dBm)
basic unit
0 Hz (Zero Span), 100 Hz to 1.6 GHz
Carrier offset:
1 MHz to 1.6 GHz
-70 dBc (typ.)
with HMS-3G (HV212) option
0 Hz (Zero Span), 100 Hz to 3 GHz
1.6 GHz to 3 GHz
with HMS-3G (HV212) option
-55 dBc (typ.)
Span range
Spectral purity, SSB phase noise (with HMS-EMC (HV213) option)
30 kHz from carrier
(500 MHz, +20°C to +30 °C)
<-85 dBc/Hz
100 kHz from carrier
(500 MHz, +20°C to +30 °C)
<-100 dBc/Hz
1 MHz from carrier
(500 MHz, +20°C to +30 °C)
<-120 dBc/Hz
2nd harmonic receive frequency
Mixer level: -40 dBm
fspan = 0 Hz (zero span)
2 ms to 100 s
fspan > 0 Hz
20 ms to 1000 s, min. 20 ms per 600 MHz
Level display
Reference level
-80 dBm to +20 dBm in 1 dB steps
basic unit
100 dB, 50 dB, 20 dB, 10 dB
with HMS-EMC (HV213) option
linear
level display error
(ref. level -50 dBm, 20°C to 30 °C)
<1.5 dB (typ. 0.5 dB)
5 MHz to 1.6 GHz
with HMS-3G (HV212) option
5 MHz to 3 GHz
basic unit
10 kHz to 1 MHz (in 1 to 3 steps), 200 kHz
logarithmical
with HMS-EMC (HV213) option
100 Hz to 1 MHz (in 1 to 3 steps), 200 kHz
linear,
with HMS-EMC (HV213) option
Tolerance
f ≤ 300 kHz
±5 % typ.
f > 1 MHz
±10 % typ.
Resolution bandwidths (-6 dB)
CISPR: 200 Hz, 9 kHz, 120 kHz, 1 MHz
Video bandwidths
basic unit
1 kHz to 1 MHz (in 1 to 3 steps)
with HMS-EMC (HV213) option
10 Hz to 1 MHz (in 1 to 3 steps)
Amplitude
Average noise level displayed up to
+20 dBm
Amplitude measurement range
basic unit
-104 dBm to +20 dBm (typ.)
with HMS-EMC (HV213) option
-114 dBm to +20 dBm (typ.)
Max. voltage at HF input
80 V DC
Max. power at HF input
20 dBm (permanently),
30 dBm (temporarily for max. 3 min)
Intermodulation-free range
dBm, dBµV, dBmV
1 curve and 1 memory curve
Trace mathematics
A-B (curve-stored curve), B-A
Detectors
basic unit
auto-, min/max. peak, sample, RMS,
average
with HMS-EMC (HV213) option
same as basic unit, quasi-peak in addition
basic unit
free run, single trigger, external trigger
with HMS-EMC (HV213) option
same as basic unit, video trigger in
addition
External reference input / output
Connectors
BNC socket
Reference frequency
10 MHz
min. level (50 Ω)
10 dBm
Interfaces
for mass storage
2x USB-host (type A), FAT16/32
for remote control
HO720 dual interface:
RS-232 / USB-device (type B)
Optional interfaces
HO732 dual interface:
Ethernet (RJ45) / USB-device (type B)
HO740 interface: IEEE-488 (GPIB)
Video output
DVI-D (480p, 60Hz), HDMI compliant
Save and recall
on internal file system (up to 10 device
settings) or external USB memory (max.
4 GB)
Marker functions
8
peak, next peak, minimum,
center to marker, frequency,
reference level to marker level,
all marker on peak
Marker display
basic unit
mit HMS-EMC (HV213) option
Additional connectors
Supply output for field probes
6 V DC, max. 100 mA (2.5 mm DIN jack)
Audio output
3.5 mm DIN jack
Demodulation
AM and FM (via internal speaker)
General Characteristics
Display
Marker and delta marker
Number of markers
BNC socket (TTL)
Trigger types
percentaged from reference level
Measured curves
-20 dBm to 0 dBm (in 1 dB steps)
Connector
Display scaling
Resolution bandwidths (-3 dB)
Display range
-60 dBc (typ.)
Display range
Sweep time
with HMS-EMC (HV213) option
<-80 dBm
basic unit
normal (level, logarithmic), delta marker,
noise marker, normal (linear)
(frequency) counter
Connectors
HF Input
screen size / type
16.5 cm (6,5”) VGA color TFT
resolution
640 x 480 (LED)
Power supply
AC supply
105 V to 253 V, 50 Hz to 60 Hz, CAT II
power consumption
40 W (typ.)
Safety
safety class I (EN61010-1)
Temperature
operating temperature range
+5 °C to +40 °C
storage temperature range
-20 °C to +70 °C
Rel. humidity
5% to 80% (without condensation)
TOI products, 2 x -20 dBm (-10 dBm 66 dB (typ.)
ref. level)
(typ. +13 dBm third-order-intercept)
Connector
N socket
Input impedance
50 Ω
dimensions (W x H x D)
285 x 175 x 220 mm
signal distance ≤2 MHz
60 dB (typ., +10 dBm TOI)
VSWR (10 MHz to 3 GHz)
<1.5 (typ.)
weight
3.6 kg
signal distance >2 MHz
66 dB (typ., +13 dBm TOI)
Mechanical data
Rohde & Schwarz R&S®HMS-X 7
© 2015 Rohde & Schwarz GmbH & Co. KG
Mühldorfstr. 15, 81671 München, Germany
Phone: +49 89 41 29 - 0
Fax: +49 89 41 29 12 164
E-mail: [email protected]
Internet: www.rohde-schwarz.com
Subject to change – Data without tolerance limits is not binding.
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners. 3/16
PD 3607.0181.32 V 02.00 PDP1 en
3607018132
Was this manual useful for you? yes no
Thank you for your participation!

* Your assessment is very important for improving the work of artificial intelligence, which forms the content of this project

Download PDF

advertisement