BUCHI B-414 Scrubber Operation Manual
The B-414 Scrubber is a laboratory device designed for the neutralization and adsorption of gases that arise during chemical reactions and syntheses. It is equipped with a pump that draws off the gases, a condenser or cold trap to cool and collect vapors and liquids, a neutralization stage and an adsorption stage to remove particles and aerosols. The Scrubber can be used with a variety of digestion units, including the Digestion Unit K-424 (6 places each), Digestion Unit K-435 (12 places), Digest Automat K-438 (20 places) and Digest System K-437 (20 places).
PDF
Download
Document
Advertisement
Advertisement
Operation Manual Scrubber B-414 96701 en Table of contents Table of contents 1 2 3 4 About this manual . . . . . . . . . . . . . . . . . . . . . . 1.1 Reference documents . . . . . . . . . . . . . . . . . 1.2 Abbreviations . . . . . . . . . . . . . . . . . . . . . Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.1 User qualification . . . . . . . . . . . . . . . . . . . 2.2 Proper use . . . . . . . . . . . . . . . . . . . . . . 2.3 Improper use . . . . . . . . . . . . . . . . . . . . . 2.4 Warning notices used in this manual . . . . . . . . . 2.5 Product safety. . . . . . . . . . . . . . . . . . . . . 2.5.1 Instrument-related hazards . . . . . . . . . . . . . . 2.5.2 Other hazards . . . . . . . . . . . . . . . . . . . . . 2.5.3 Safety measures . . . . . . . . . . . . . . . . . . . 2.5.4 Safety elements . . . . . . . . . . . . . . . . . . . . 2.6 General safety rules . . . . . . . . . . . . . . . . . . Technical data . . . . . . . . . . . . . . . . . . . . . . . . 3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . 3.1.1 Standard instrument. . . . . . . . . . . . . . . . . . 3.1.2 Scrubber with condenser . . . . . . . . . . . . . . . 3.1.3 Scrubber with cold trap . . . . . . . . . . . . . . . . 3.1.4 Scrubber complete with condenser and reaction stage 3.1.5 Standard accessories . . . . . . . . . . . . . . . . . 3.1.6 Optional accessories . . . . . . . . . . . . . . . . . 3.1.7 Upgrade. . . . . . . . . . . . . . . . . . . . . . . . 3.2 Technical data overview . . . . . . . . . . . . . . . . 3.3 Materials used. . . . . . . . . . . . . . . . . . . . . 3.4 Recommended consumables . . . . . . . . . . . . . Description of function . . . . . . . . . . . . . . . . . . . 4.1 Functional principle of the Scrubber . . . . . . . . . . 4.2 Scrubber capacity. . . . . . . . . . . . . . . . . . . 4.3 Condenser . . . . . . . . . . . . . . . . . . . . . . 4.4 Cold trap . . . . . . . . . . . . . . . . . . . . . . . 4.5 External receiving vessel (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 .5 .5 .6 .6 .6 .6 .7 .7 .7 .8 .8 .8 .9 10 10 10 10 11 11 12 12 13 13 13 14 15 15 16 17 17 18 Read this manual carefully before installing and running your system and note the safety precautions in chapter 2 in particular. Store the manual in the immediate vicinity of the instrument, so that it can be consulted at any time. No technical modifications may be made to the instrument without the prior written agreement of Buchi. Unauthorized modifications may affect the system safety or result in accidents. This manual is copyright. Information from it may not be reproduced, distributed, or used for competitive purposes, nor made available to third parties. The manufacture of any component with the aid of this manual without prior written agreement is also prohibited. If you need another language version of this manual, you can download it at www.buchi.com. 3 B-414 Operation Manual, Version D Table of contents 5 6 7 8 9 10 11 4 Putting into operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1 Installation site. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2 General commissioning procedure for all Scrubber models . . . . . . . . . . . 5.3 Commissioning the Scrubber with condenser / cold trap and reaction vessel . . 5.4 External receiving vessel (optional). . . . . . . . . . . . . . . . . . . . . . . . 5.5 Hose connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.5.1 Hose connections of the basic model . . . . . . . . . . . . . . . . . . . . . . 5.5.2 Hose connections for Scrubber with condenser / cold trap . . . . . . . . . . . 5.5.3 Hose connections for Scrubber with condenser / cold trap and reaction vessel . 5.6 Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.7 Preparing washing solutions. . . . . . . . . . . . . . . . . . . . . . . . . . . 5.7.1 Washing solutions for acidic vapors and gases . . . . . . . . . . . . . . . . . 5.7.2 Washing solutions for alkaline vapors and gases . . . . . . . . . . . . . . . . 5.7.3 Color indicator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.8 Filling up the prepared washing solution . . . . . . . . . . . . . . . . . . . . . Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1 Typical applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.2 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.1 Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.2 Glass parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.3 Tubes / tube connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4 Sealing system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4.1 Cleaning the seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.4.2 Replacing the seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.5 Washing insert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.6 Washing solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.7 Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.8 Power supply fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.9 Silencer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.10 Functional test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.11 Customer service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8.1 Malfunctions and their remedy . . . . . . . . . . . . . . . . . . . . . . . . . Shutdown, storage, transport and disposal . . . . . . . . . . . . . . . . . . . . . 9.1 Storage and transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.1 Spare parts on the exploded instrument view . . . . . . . . . . . . . . . . . . 10.2 Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Declarations and requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1 FCC requirements (for USA and Canada) . . . . . . . . . . . . . . . . . . . . 11.2 Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 19 19 20 21 21 21 23 24 25 25 25 25 25 26 27 27 28 29 29 29 29 29 30 30 30 31 31 32 32 32 32 33 33 34 34 34 35 35 36 37 37 38 B-414 Operation Manual, Version D 1 1 About this manual About this manual This manual describes the Scrubber and provides all information required for its safe operation and to maintain it in good working order. It is adressed to laboratory personnel in particular. NOTE The symbols pertaining to safety (WARNINGS and ATTENTIONS) are explained in chapter 2. 1.1 Reference documents For information on the Digest System, the Digest Automat, and the Digest Unit, please refer to the corresponding manuals available in English, German, French, Italian and Spanish • • • • 1.2 Digest System K-437, Operation Manual numbers 96760 - 96764 Digest Automat K-438, Operation Manual numbers 96765 - 96769 Digestion Unit K-424, Operation Manual numbers 96710 - 96714 Digestion Unit K-435, Operation Manual numbers 96720 - 96724 Abbreviations CR: Chloroprene Rubber EPDM: Ethylene Propylene Dimonomer FEP: Fluorinated Ethylene Propylene FPM: Fluoroelastomer pa: pro analysis PA: Polyamides PMMA: Polymethylmethacrylate POM: Polyoxymethylene PP: Polypropylene PPS: Polyphenylenesulfide PTFE: Polytetrafluoroethylene (Teflon) PUR: Polyurethanes 5 B-414 Operation Manual, Version D 2 2 Safety Safety This chapter points out the safety concept of the instrument and contains general rules of behavior and warnings from hazards concerning the use of the product. The safety of users and personnel can only be ensured if these safety instructions and the safetyrelated warnings in the individual chapters are strictly observed and followed. Therefore, the manual must always be available to all persons performing the tasks described herein. 2.1 User qualification The instrument may only be used by laboratory personnel and other persons who on account of training or professional experience have an overview of the dangers which can develop when operating the instrument. Personnel without this training or persons who are currently being trained require careful instruction. The present Operation Manual serves as the basis for this. 2.2 Proper use The instrument has been designed and built for laboratories. It serves for the neutralization and adsorption of gases that arise during chemical reactions and syntheses. Only gases with known chemical composition may be drawn off. 2.3 Improper use Applications not mentioned above are improper. Also, applications, which do not comply with the technical data, are considered improper. In particular no gases with unknown chemical composition may be drawn off. The operator bears the sole risk for any damages caused by such improper use. The following uses are expressly forbidden: • Use of the instrument in rooms which require ex-protected instruments. • Use on samples, which can explode or inflame (example: explosives, etc.) due to shock, friction, heat or spark formation. • Use in overpressure situations. • Drawing off liquids. • Drawing off solvents. • Drawing off vapors of organic solvents. • Use for cleaning of room air. 6 B-414 Operation Manual, Version D 2 2.4 Safety Warning notices used in this manual WARNING Generally, the triangular warning symbol indicates the possibility of personal injury or even loss of life if the instructions are not followed. WARNING Hot surface. WARNING Electrical hazard. WARNING Biohazard. ATTENTION With the general “Read this” symbol, ATTENTIONs indicate the possibility of equipment damage, malfunctions or incorrect process results, if instructions are not followed. NOTE Useful tips for the easy operation of the instrument. 2.5 Product safety The Scrubber is designed and built in accordance with current state-of-the-art technology. Nevertheless, risks to users, property, and the environment can arise when the instrument is used carelessly or improperly. The manufacturer has determined residual dangers emanating from the instrument • if the instrument is operated by insufficiently trained personnel. • if the instrument is not operated according to its proper use. Appropriate warnings in this manual serve to make the user alert to these residual dangers. 2.5.1 Instrument-related hazards Pay attention to the following safety notices: WARNING Potential implosion risk if used with damaged glassware. • Beware of damaged or cracked glass parts. WARNING Potential explosion risk if solvent vapors accumulate within the instrument housing. • Always use the instrument in a well ventilated area. • Beware of damaged or cracked glass parts. • Beware of the fire hazard. 7 B-414 Operation Manual, Version D 2 2.5.2 Safety Other hazards WARNING Certain solvents in or in the vicinity of the Scrubber can form peroxides and/or are highly inflammable. • Always be aware of the explosion risk when working with hazardous substances or with substances of unknown composition. • Always provide a good ventilation within or in the vicinity of the system. WARNING Risk of implosion when operating the instrument with an external pump and excess pressure on the overpressure valve. • Always operated the instrument with the build-in pump and do not use an external pump. WARNING Risk of acidic vapors forming in the vicinity of the instrument when the washing solution is not exchanged regularly. • Always use a color indicator to judge the state of the washing solution and exchange the solution as soon as the color indicator undergoes a color change. 2.5.3 Safety measures Always wear personal protective equipment such as protective eye goggles, protective clothing and gloves when working with the instrument. 2.5.4 Safety elements • Overpressure valve • The protective disc on top of the washing insert and the protective cover with the protective shield to be mounted at the neutralization vessel protects operators from broken glass and solvent splashes. 8 B-414 Operation Manual, Version D 2 2.6 Safety General safety rules Responsibility of the operator The head of laboratory is responsible for training his personnel. The operator shall inform the manufacturer without delay of any safety-related incidents which might occur during operation of the instrument. Legal regulations, such as local, state and federal laws applying to the instrument must be strictly followed. Duty of maintenance and care The operator is responsible for ensuring that the instrument is operated in proper condition only, and that maintenance, service, and repair jobs are performed with care and on schedule, and by authorized personnel only. Spare parts to be used Use only genuine consumables and genuine spare parts for maintenance to assure good system performance and reliability. Any modifications to the spare parts used are only allowed with the prior written permission of the manufacturer. Modifications Modifications to the instrument are only permitted after prior consultation with and with the written approval of the manufacturer. Modifications and upgrades shall only be carried out by an authorized Buchi technical engineer. The manufacturer will decline any claim resulting from unauthorized modifications. 9 B-414 Operation Manual, Version D 3 3 Technical data Technical data This chapter introduces the reader to the instrument specifications. It contains the scope of delivery, technical data, requirements and performance data. 3.1 Scope of delivery Check the scope of delivery according to the order number. NOTE For detailed information on the listed products, see www.buchi.com or contact your local dealer. 3.1.1 Standard instrument Table 3-1: Standard instrument Product Order number B-414 Scrubber, standard 3.1.2 230 V / 50 Hz 37876 120 V / 60 Hz 37877 100 V / 50/60 Hz 37878 1 adsorption vessel, complete 38128 1 hose, adsorption / pump 37885 Scrubber with condenser Table 3-2: Scrubber with condenser 10 Product Order number 230 V / 50 Hz 37882 120 V / 50 Hz 37883 100 V / 60 Hz 37884 1 condenser unit 37775 3 tube connectors, bent with screw cap 32885 1 silicone hose Ø 6/9 04133 1 spring ring 37872 1 holder ring for receiving vessel 36709 1 seal for receiving vessel 37873 1 receiving vessel 1.2 l 37776 1 hose, condenser / lye 37888 1 adsorption vessel, complete 38128 1 hose, adsorption / pump 37885 B-414 Operation Manual, Version D 3 3.1.3 Technical data Scrubber with cold trap Table 3-3: Scrubber with cold trap 3.1.4 Product Order number 230 V / 50 Hz 37879 120 V / 50 Hz 37880 100 V / 50/60 Hz 37881 1 cold trap complete 38083 3 tube connectors, bent with screw cap 32885 1 receiving vessel 1.2 l 37776 1 spring ring 37872 1 holder ring for receiving vessel 36709 1 seal for receiving vessel 37873 1 hose, condenser / lye 37888 1 adsorption vessel, complete 38128 1 hose, adsorption / pump 37885 Scrubber complete with condenser and reaction stage Table 3-4: Scrubber complete with condenser and reaction stage Product Order number 230 V / 50 Hz 37988 120 V / 50 Hz 37989 100 V / 50/60 Hz 37990 1 condenser unit 37775 3 hose connections bent with screw caps 32885 11 1 silicone hose Ø 6/9 04133 1 spring ring 37872 1 holder ring for receiving vessel 36709 1 seal for receiving vessel 37873 1 receiving vessel 1.2 l 37776 1 hose, condenser / lye 37888 1 adsorption vessel, complete 38128 1 hose, adsorption / reaction 37887 1 reaction stage 37956 B-414 Operation Manual, Version D 3 3.1.5 Technical data Standard accessories Table 3-5: Standard accessories Product Order number 1 neoprene hose 10021 1 power cable Type CH 10010 Type Schuko / Japan 10016 Type GB 17835 Type USA 33763 Type AUS 17836 Operation Manual: 3.1.6 English 96701 German 96700 French 96702 Italian 96703 Spanish 96704 Optional accessories Table 3-6: Optional accessories 12 Product Order number 1 condenser unit 37775 1 cold trap complete 38083 1 condensate drain 37785 1 reaction vessel 37797 1 lip gasket, acid-resistant 38064 1 hose, adsorption / pump 37885 1 hose, reaction / pump 37886 1 hose, adsorption / reaction 37887 1 hose, condenser / washing solution 37888 1 connection cable B-414 / Digestion Units, Control Unit 14738 Set large receiving vessel 4 l 48668 B-414 Operation Manual, Version D 3 3.1.7 Technical data Upgrade Table 3-7: Upgrade possibilities 3.2 Product Order number 1 condenser with receiving vessel 1.2 l, complete 37954 Cold trap with receiving vessel 1.2 l, complete 37955 1 reaction stage, complete 37956 Technical data overview Table 3-8: Technical data Scrubber Unit B-414 3.3 Dimensions (L x H x D) 260 x 450 x 480 mm Weight 13 kg Power system voltage / frequency 230 V ± 10 % / 50 Hz 120 V ± 10 % / 60 Hz 100 V ± 10 % / 50/60 Hz Ambient temperature For indoor use only, altitude up to 2000 m above seal level, maximum relative humidity 80 %, 5 - 40 °C Pump suction capacity 34 l/min Overvoltage category II Pollution degree 2 Power consumption 200 W Current consumption (230 V) 1.5 A 140 W 100 W 100 W Materials used Table 3-9: Materials used Component Material designation Housing PUR Minor hardware PP, PA, POM Lip gasket acid- / lye-resistant FPM, EPDM Gasket PTFE, CR Protective disc PMMA Glass parts Borosilicate 3.3 Pump PPS, CR, FPM Hoses CR, FEP 13 B-414 Operation Manual, Version D 3 3.4 Technical data Recommended consumables Table 3-10: Active charcoal for adsorption vessel Product Order number Fluka / Merck Fluka, purum p.a. 4 - 8 mm 05110 Merck, granulated 102518 Table 3-11: Color indicator for washing solution Product Order number Merck Bromothymol blue, Merck 3026 14 B-414 Operation Manual, Version D 4 4 Description of function Description of function This chapter explains the basic principle of the instrument, shows how it is structured and gives a functional description of the assemblies. The pump draws off the gases and vapors produced in chemical reactions via a condenser or a cold trap providing a receiving vessel underneath. The condensation stage is used as a preliminary extractor for vapors, water steam (to prevent warming or an increase in volume of the washing solution) and for the liquids carried along with them, thus extending the service life of the neutralization stage. The acid or alkaline gases are washed and neutralized in the neutralization stage. The next stage, the adsorption stage, holds back most of the undesired particles by means of granules of activated charcoal or a universal adsorption granulate. It also enables the aerosols to recondensate. In the reaction stage that follows specific redox reactions are accomplished. The built-in overpressure valve makes sure that a certain pressure limit is not exceeded. The used air is directly lead into an exhaust or into the open air through a silencer. 4.1 Functional principle of the Scrubber 2 4 3 1 5 Receiving vessel Condenser (or cold trap) Reaction tubelet Adsorption vessel Neutralization vessel Fig. 4.1: Overview over the Scrubber 15 B-414 Operation Manual, Version D 4 Description of function Adsorption 4 1 PUMP 6 3 2 5 Condensation Neutralisation Condenser or cold trap Receiving vessel Neutralization vessel with washing solution Reaction stage Adsorption vessel with adsorbents Reaction stage Fume exhaust pump Fig. 4.2: Functional principle of the Scrubber Condensation stage • Condensation of vapors • Condensation of water steam (prevention of warming or an increase in volume of the washing solution) • Condensation of the liquids carried along Neutralization stage • Neutralization of acid or alkaline gases Adsorption stage • Specific adsorptions and retention of particles (using granules of activated charcoal or a universal adsorption granulate) • Recondensation of aerosols Reaction stage • Specific reactions (e.g., redox reactions) 4.2 Scrubber capacity The suction capacity of the Scrubber B-414 is designed for Digestion Units with a maximum of 20 sample tubes. Connection examples: • • • • 16 Connection of up to two Digestion Units K-424 (6 places each) Connection of one Digestion Unit K-435 (12 places) Connection of one Digest Automat K-438 (20 places) Connection of one Digest System K-437 (20 places) B-414 Operation Manual, Version D 4 4.3 Description of function Condenser 3 1 2 4 Cooling water out Connection to neutralization stage Emission in (connection to emission source) Cooling water in Fig. 4.3: Connections of the condenser 4.4 Cold trap 2 1 Emission in (connection to emission source) Connection to neutralization stage Fig. 4.4: Connections of the cold trap 17 B-414 Operation Manual, Version D 4 4.5 Description of function External receiving vessel (option) A large receiving vessel for up to 4 l is optionally available. It serves for the digestion of high volumes of water samples. Fig. 4.5: External receiving vessel 18 B-414 Operation Manual, Version D 5 5 Putting into operation Putting into operation This chapter describes how the instrument is installed and gives instructions on initial startup. NOTE Inspect the instrument for damages during unpacking. If necessary, prepare a status report immediately to inform the postal company, railway company or transportation company. Remove the transprotation lock at the bottom of the instrument. Keep the original packaging for future transportation. 5.1 Installation site Install the instrument on a clean, flat and stable base. ATTENTION To ensure the instrument safety do not place any object on top of it. ATTENTION Place the Scrubber B-414 on the left side of a digestion unit (e.g. Digestion Unit K-424, K-435, Digest System K-437 or Digest Automat K438). The Scrubber ventilation is located on the left side. In case the Scrubber is placed on the right side of another unit, cool air will hit the digestion sample and cool it down, so that no digestion can take place. 5.2 General commissioning procedure for all Scrubber models To commission the Scrubber B-414, proceed as follows: • Apply the protective cover and shield to the neutralization vessel. • Make sure that the silencer is filled with cotton wool and install it. • Check the glass for flaws and cracks. • Check the seals and gaskets (embrittlement, scratches). If they are not OK, take them out and replace them (see also chapter 7). • Carefully insert the glass parts using a clamping ring (without pressure from above). Otherwise there is a risk of damaging the glassware. NOTE When you carry out Kjeldahl digestions, the suction capacity might become too high. Therefore we recommend to fill activated charcoal into the adsorption vessel and to add glass wool at both ends. The glass wool prevents charcoal from being drawn into the vacuum pump. The activated charcoal should be grainy with a particle size of 2 to 6 mm. Fig. 5.1: Adsorption vessel with activated charcoal and glass wool 19 B-414 Operation Manual, Version D 5 5.3 Putting into operation Commissioning the Scrubber with condenser / cold trap and reaction vessel To prepare the complete Scrubber model for operation, proceed as follows: • Insert the white PTFE gasket into the holder ring for the receiving vessel, so that it fits properly into it on both sides. • Press the holder ring into the gap provided for condenser / cold trap and receiving vessel. • Apply the spring ring to the bottom of the condenser / cold trap (the cooling water / contaminated air inlet must be at the bottom) and introduce the condenser into the holder ring. • Screw the receiving vessel counterclockwise into the holder ring underneath the condenser. Spring ring Holder ring Gasket PTFE 3 1 2 Fig.5.2: Connection between condenser and receiving vessel • Insert the reaction vessel into the housing from below. • Insert the introduction tube of the reaction tubelet into the reaction vessel and screw it in tightly. 20 B-414 Operation Manual, Version D 5 5.4 Putting into operation External receiving vessel (optional) To commission the optional external receiving vessel, proceed as follows: • Connect the one end of the adapter (FEP hose) to the bottom of the condenser. • Connect other side of the adapter to the external receiving vessel and fasten the connection with the black screw caps. Fig.5.3: Connecting the external receiving vessel ATTENTION Do not use plastic tanks for the collection of condensate but only original glass vessels from Buchi. • The diameter of the connection between the outlet and the receiving vessel must not be less than 10 mm. • The materials used must be resistant to the condensate and must withstand a vacuum of 100 mbar. 5.5 Hose connections NOTE Use clips and cable ties where necessary to secure the cables in place. 5.5.1 Hose connections of the basic model The basic model consists of the neutralization and the adsorption vessel. To establish the hose connections, proceed as follows: • Connect the adsorption vessel to the distribution head and secure with tube connectors GL-14 at both ends. • Connect the emission source via a Woulff bottle to the neutralization vessel. 21 B-414 Operation Manual, Version D 5 3 Putting into operation 1 2 4 5 Neutralization stage Connection of the neutralization vessel Adsorption vessel Distribution head Silencer Fig. 5.4: Hose connections of the basic model 22 B-414 Operation Manual, Version D 5 5.5.2 Putting into operation Hose connections for Scrubber with condenser / cold trap To establish the hose connections, proceed as follows: • • • • Put tube connectors GL-14 on all PTFE hoses. Connect the emission source to the condenser / cold trap inlet (position in Fig. 5.2 ). Connect the condenser / cold trap outlet (position in Fig. 5.2) to the neutralization vessel. Connect the adsorption vessel to the distribution head. 1 2 10 3 6 7 9 8 4 5 Neutralization stage Adsorption vessel Reaction stage Distribution head Silencer Inlet from emission source Outlet to neutralization vessel Cooling water inlet (bottom) Cooling water outlet (top) Condenser Fig. 5.5: Hose connections for Scrubber with condenser / cold trap 23 B-414 Operation Manual, Version D 5 5.5.3 Putting into operation Hose connections for Scrubber with condenser / cold trap and reaction vessel To establish the hose connections, proceed as follows: • • • • • Put tube connectors GL-14 on all PTFE hoses. Connect the emission source to the condenser / cold trap inlet (position in Fig. 5.3 ). Connect the condenser / cold trap outlet (position in Fig. 5.3) to the neutralization vessel. Connect the adsorption vessel to the top of the reaction tubelet. Connect the reaction tubelet (connection at the front) to the distribution head 2 1 10 3 6 7 9 8 4 5 Neutralization stage Adsorption vessel Reaction stage Distribution head Silencer Inlet from emission source Outlet to neutralization vessel Cooling water inlet (bottom) Cooling water outlet (top) Condenser Fig. 5.6: Hose connections for Scrubber with condenser / cold trap and reaction vessel 24 B-414 Operation Manual, Version D 5 5.6 Putting into operation Electrical connections ATTENTION Make sure that the voltage on the socket corresponds to the voltage given on the type plate of the instrument. Always connect the instrument to an earthed socket. External connections and extension cables must be provided with an earthed conductor lead (3-pole couplings, cable or plug equipment) as the mains lead has a molded plug, thus avoiding risks due to inadvertent defective wiring. Make sure that no electric sparks form in the instrument or its surroundings as they might damage the instrument. Connect the Scubber B-414 to the mains with the power cord contained in the scope of delivery. To start the Scrubber automatically at the same time the digestion process starts, connect it to the Control Unit B-436, to a Digest System K-437 or to a Digest Automat K-438 with the optional connection cable. 5.7 Preparing washing solutions ATTENTION Always wear personal protective equipment such as protective eye goggles, protective clothing and gloves when working with washing solutions. 5.7.1 Washing solutions for acidic vapors and gases • Sodium hydroxide 8-10 %, max. 20 % • Sodium carbonate – dissolve 600 g Na2CO3 in 3 l distilled warm water, or – dissolve 1.7 kg Na2CO3 · 10H2O in 3 l distilled warm water 5.7.2 Washing solutions for alkaline vapors and gases • Hydrochloric acid, max. 15% • Sulphuric acid, max. 20% Depending on the area of application, it is recommended to use different washing solutions or concentrations. The B-414 Scrubber is supplied ex works with a lye-resistant lip gasket. For operating with acid washing solutions, an acid-resistant lip gasket should be used. 5.7.3 Color indicator A color indicator is added to the washing solution to assess its washing power visually as well. We recommend to use 100 mg of color indicator for 3 l of washing solution. Bromothymol blue serves as a standard indicator. The transition area here is between pH 6.0 to 7.6. Thus the fresh solution has a blue color while the neutralized or acidic solution has an orange yellowish color. Depending on the application purpose, other suitable color indicators should be used. 25 B-414 Operation Manual, Version D 5 5.8 Putting into operation Filling up the prepared washing solution To fill up the washing solution, proceed as follows: Remove the adsorption vessel by gently pulling it upwards. Loosen the tube connector GL-14 at the neutralization vessel and pull off the hose. Remove the protective cover with the protective shield. Remove the protective disc with the washing insert from the neutralization vessel. Carefully pour the washing solution into the washing vessel and fill up to the wavy maximum filling mark. • Reassemble the instrument in the reverse order. • • • • • 1 2 3 5 4 Adsorption vessel Nut GL-14 Protective cover with protective shield Washing insert ON/OFF switch Fig. 5.7: Filling up the washing solution NOTE For information on how and when to exchange the washing solution, see chapter 7.6 26 B-414 Operation Manual, Version D 6 6 Operation Operation This chapter gives examples of typical instrument applications and instructions on how to operate the instrument properly and safely. WARNING Risk of injury. • Never operate the instrument when glassware is damaged. ATTENTION Switch on the B-414 before a connected instrument produces gases. 6.1 Typical applications Table 6-1: Typical applications Gases and vapors Chemical formula Neutralization (stage 2) Kjeldahl digestions SO2, H2SO4 saturated soda / NaOH 8 - 10 % (max. 20 %) Sulfur dioxide SO2 saturated soda / NaOH 8 - 10 % (max. 20 %) Sulfuryl chloride SO2Cl2 NaOH 8 - 10 % (max. 20 %) Hydrochloric or hydrobromic acids HCI, HBr NaOH 8 - 10 % (max. 20 %) Thionyl chloride SOCl2 NaOH 8 - 10 % (max. 20 %) Cyanuric chloride C3N3Cl3 NaOH 8 - 10 % (max. 20 %) Carbonic acid chloride R-COCI NaOH 8 - 10 % (max. 20 %) Phosgene COCI2 NaOH 8 - 10 % (max. 20 %) Specific reactins (stage 4) Ammonia NH3 HCI max. 15 % / H2SO4 max. 20 % Nitrous oxides (nitrous gases) Nitric acid / aqua regia NOX NaOH 20 % FESO4 saturated Chlorine, bromine, iodine Cl2, Br2, J2 NaOH 8 - 10 % (max. 20 %) FESO4 saturated Thiophenols Ar-SH NaOH 8 - 10 % (max. 20 %) KMnO4 saturated Thioalcohols, mercaptan R-SH NaOH / javelle water KMnO4 saturated Hydrogen sulphide H 2S NaOH 8 - 10 % (max. 20 %) KMnO4 saturated Carbon disulphide CS2 NaOH 8 - 10 % (max. 20 %) KMnO4 saturated Prussic acid HCN NaOH 8 - 10 % (max. 20 %) KMnO4 saturated 27 B-414 Operation Manual, Version D 6 6.2 Operation Operation 2 • Switch on the power switch . • Make sure that the cooling water flow rate does not exceed 60 l/h. • The suction performance can be regulated during operation by drilling a 0.1 - 0.3 mm wide hole in the screw cap GL-14 (e.g. when samples are getting dry). • Carry out a functional test as described in chapter 7.10. 1 Fig. 6.1.: Operation 28 B-414 Operation Manual, Version D 7 7 Maintenance and repairs Maintenance This chapter gives instructions on all maintenance work to be performed in order to keep the instrument in good working condition. WARNING Electrical hazard: • Prior to all maintenance work on the instrument switch off the power supply and remove all sources of flammable vapor. ATTENTION Always wear personal protective equipment such as protective eye goggles, protective clothing and gloves when maintaining the instrument. 7.1 Housing Check the housing for defects (switches, plugs) and clean it regularly with a damp cloth. ATTENTION Never use solvents as cleaning agents as these might damage the instrument. 7.2 Glass parts The glass parts can be taken out and cleaned with commercially available cleaning agents or in an ultrasonic bath. After the glass parts have been cleaned and fully dried, check each part visually for cracks, scratches and for any parts or sections that might have splintered off. Take out and replace any damaged glass parts. 7.3 Tubes / tube connectors Visually examine the tube connections regularly. When tubes become cracked and brittle, replace them with new tubes. Grease all joints on the condenser side regularly to achieve optimum tightness of the system. Flush out the tubes regularly with water or ethanol. 7.4 Sealing system ATTENTION When removing and reinstalling the seals and gaskets, make sure not to damage them. Always move them perpendicularly to the axis of the glass parts and ensure no damage occurs to the sealing lip. Never apply grease to the seals and never touch them with sharp object, otherwise they will get damaged. 29 B-414 Operation Manual, Version D 7 7.4.1 Maintenance and repairs Cleaning the seals To prolong the lifetime of the seals, rinse them regularly with water, especially if working with crystalline products. Afterwards, dry them with a solft cloth. To remove the seals, see Fig. 7.1. 7.4.2 Replacing the seals Seals are subject to wear and tear, thus you should check them regularly and replace them, if necessary, e.g. if they do not pass the functional test described in chapter 7.8 anymore. We recommend to replace the seals as follows: Table 8-1: General malfunction and their remedy 7.5 Position in Fig. 9.1 Order no. Description Replacing interval -- 40029 Set of seals EPDM for tube connectors once a year 8 37928 O-ring to adapter (37777) once a year 13 03575 Gasket SVL 42 x 30 PTFE once a year 17 03576 Gasket SVL 22 x 16 PTFE once a year 19 38225 Hose seal once a year 23 37873 Seal for receiving vessel once a year 34 37925 Gasket ring once a year 38 37871 Silencer unit once a year Washing insert The washing insert must be cleaned from time to time and the lip gasket must be replaced when brittle. 7 6 5 4 2 3 2 1 Nut M20 Swirl disc Spacer tube Swirl disc holder Lip gasket Protective disc Tube connector adapter M20 GL-14 Fig. 7.1. Exploded view of the washing insert 30 B-414 Operation Manual, Version D 7 Maintenance and repairs To (re)assemble the washing insert, proceed as follows: • • • • • • • • • • 7.6 Insert the lip gasket into the protective disc . Place the protective disc on the swirl disc holder . Screw on the adapter and tighten it. Insert the first swirl disc . Insert the spacer tube . Insert the second swirl disc . Screw the washing insert into the adapter . Introduce the washing insert into the neutralisation vessel. Put on the protective disc. Put on the protective cover with shield and center the neutralization vessel. Washing solution Used washing solutions may impair the functionality of the B-414 Scrubber especially in the range of the neutralization point as a result of strong foaming. Changing the washing solution in appropriate time (when the color indicator ongoes a color change) keeps the adsorption stage and/or reaction stage from getting soiled and prevents the possibility of the pump getting damaged. For a discription of how to prepare and fill up the washing solution, see chapter 5.7. NOTE Make sure to dispose of the used washing solution according to your laboratory guidelines. 7.7 Pump To flush the pump, proceed as follows: • Take off the silencer and flush out the pump with distilled water through the secondary air valve . • Switch on the instrument and collect the waste water in a suitable vessel . 2 1 4 3 Fig. 7.2: Flushing the pump 31 B-414 Operation Manual, Version D 7 7.8 Maintenance and repairs Power supply fuses 2 1 To replace the power supply fuses, proceed as follows: • Pull out the power supply cable. • Remove the fuse holder . • Replace the defective fuse with a new fuse of the same type: – 100 V: T 2.0 A L 250 V – 120 V: T 1 A L 250 V – 230 V: T 1 A L 250 V • Put the fuse holder back. Fig. 7.3: Replacing the power supply fuses 7.9 Silencer Replace the silencer once a year. 7.10 Functional test To check whether the overpressure valve functions properly, screw off the left tube connector at the distribution head and put a finger on the opening. After the overpressure valve has switched, a change of sound is audible. The system is tight, when the washing solution in the neutralization vessel bubbles when the instrument is switched on. 7.11 Customer service Only authorised service personnel are allowed to perform repair work on the instrument. These persons have a comprehensive technical training and knowledge of possible dangers which might arise from the instrument. Addresses of official Buchi customer service offices are given on the Buchi website under: www.buchi.com. If malfunctions occur on your instrument or you have technical questions or application problems, contact one of these offices. The customer service offers the following: • Spare part delivery • Repairs • Technical advice 32 B-414 Operation Manual, Version D 8 8 Troubleshooting Troubleshooting This chapter helps to resume operation after a minor problem has occurred with the instrument. It lists possible occurrences, their probable cause and suggests how to remedy the problem. The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective measures are listed in the column “Corrective measure”. The elimination of more complicated malfunctions or errors is usually performed by a Buchi technical engineer who has access to the official service manuals. In this case, please refer to your local Buchi customer service agent. 8.1 Malfunctions and their remedy Table 8-1: General malfunction and their remedy Malfunction Possible cause Corrective measure B-414 does not work No mains connection Fuses defective Check whether the instrument is connected to the mains Replace the defective fuses Contact the Buchi customer service Mains switch defective Pump does not work Wiring defective Pump defective Contact the Buchi customer service Suction system leaking Fumes are not exhausted during extraction (Scrubber suction capacity Hose bent Adsorption vessel clogged too weak) Washing solution dirty Adsorbents clogged Pump dirty Silencer unit clogged Overpressure valve dirty Tighten the tube connections Check the hoses Check the adsorption vessel Exchange the washing solution Renew the adsorbents Clean the pump Replace the silencer unit Contact the Buchi customer service Fumes are exhausted too fast during Suction system clogged extraction (Scrubber suction capacity too strong) Adsorbents clogged Silencer unit defective Overpressure valve defective Check all stages and the connected digestion unit as well Exchange the adsorbents Replace the silencer unit Contact the Buchi customer service Washing solution turns yellow Exchange the washing solution 33 Washing solution dirty B-414 Operation Manual, Version D 9 9 Shutdown, storage, transport and disposal Shutdown, storage, transport and disposal This chapter instructs how to shut down the instrument, how to pack it for storage or transport, and specifies the storage and shipping conditions. 9.1 Storage and transport WARNING Biohazard: • Remove all dangerous substances from the instrument and clean it thoroughly. Store and transport the instrument in its original packaging. 9.2 Disposal To dispose of the instrument in an environmentally friendly manner, a list of materials is given in chapter 3. This helps to ensure that the components are separated and recycled correctly. Make especially sure to dispose of the gas springs appropriately. Please follow valid regional and local laws concerning disposal. 34 B-414 Operation Manual, Version D 10 10 Spare parts Spare parts This chapter lists spare parts, accessories, and options including their ordering information. Order the spare parts from Buchi. Always state the product designation and the part number when ordering spare parts. Use only genuine Buchi consumables and genuine spare parts for maintenance and repair to assure good system performance and reliability. Any modifications to the spare parts used are only allowed with the prior written permission of the manufacturer. 10.1 Spare parts on the exploded instrument view 21 29 26 28 22 20 18 23 19 24 30 17 27 9 12 35 36 25 37 13 14 15 38 11 10 9 16 6 31 39 7 8 33 5 4 3 34 32 2 1 Fig. 10.1: Spare parts on the exploded instrument view 35 B-414 Operation Manual, Version D 10 Spare parts Table 10-1: Spare parts on the exploded instrument view Position Product Order Position Product Order number number 10.2 1 Protective shield 37820 20 Hose adsorption / pump 37885 2 Neutralization vessel 37786 21 Condenser complete, including: 38022 3 Lip gasket lye-resistant EPDM 37874 22 - Spring ring 37872 4 Lip gasket acid-resistant FPM 38064 23 - Gasket 37873 5 Swirl disc 37821 24 - Holder ring 36709 6 Set of 4 tube connectors with cap bent FPM 40295 25 Receiving vessel, 1.2 l 7 Adapter for GL-14 37777 26 Hose condenser / washing solution 37888 8 O-ring to adapter (37777) 37928 27 Cold trap 37781 9 Screw cap GL-14 (set of 10) 41956 28 O-ring FPM for cold trap 38067 - Set of seals EPDM for tube connectors (set of 10) 40029 29 Trap unit for cold trap 37782 10 Hose seal 38225 30 Connector GL-14 straight 37642 11 Screw cap GL-14 40624 31 Introduction tube assembly 37778 12 Adsorption vessel 37774 32 Reaction vessel, 1.0 l 37797 13 Gasket SVL 42 x 30 PTFE 03575 33 Screw cap SVL 42 03551 14 Adapter complete for adsorption vessel 37971 34 Gasket ring 37925 15 O-ring 47.2 for adsorption vessel 37857 35 Hose adsorption / reaction 37887 16 O-ring 39.7 for adsorption vessel 38068 36 Hose reaction / pump 37886 17 Gasket SVL 22 x 16 PTFE 03576 37 Condensate drain 37785 18 Adapter SVL 22 / GL14, complete 37972 38 Silencer unit 37871 19 Seal for FEP tube 38225 39 Distribution head, complete 37952 37776 Miscellaneous Table 10-2: Miscellaneous Product Order number Set of sealings FPM (10 pieces) 40040 Axial flow fan B-414 38026 36 B-414 Operation Manual, Version D 11 11 Declarations and requirements 11.1 FCC requirements (for USA and Canada) Declarations and requirements English: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to both Part 15 of the FCC Rules and the radio interference regulations of the Canadian Department of Communications. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. Français: Cet appareil a été testé et s'est avéré conforme aux limites prévues pour les appareils numériques de classe A et à la partie 15 des réglementations FCC ainsi qu’à la réglementation des interférences radio du Canadian Department of Communications. Ces limites sont destinées à fournir une protection adéquate contre les interférences néfastes lorsque l’appareil est utilisé dans un environnement commercial. Cet appareil génère, utilise et peut irradier une énergie à fréquence radioélectrique, il est en outre susceptible d’engendrer des interférences avec les communications radio, s’il n’est pas installé et utilisé conformément aux instructions du mode d’emploi. L’utilisation de cet appareil dans les zones résidentielles peut causer des interférences néfastes, auquel cas l’exploitant sera amené à prendre les dispositions utiles pour palier aux interférences à ses propres frais. 37 B-414 Operation Manual, Version D 11 11.2 Declarations and requirements Declaration of conformity We BÜCHI Labortechnik AG do hereby declare on our responsibility that the product: Scrubber B-414 which is the object of this certification, is in accordance with the following norms: EN 61010-1:2001 (~ IEC 61010-1) Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements EN 55011:1998 + A1:1999 (IEC/CISPR 11:1997, modified + A1:1999) Industrial, scientific and medical (ISM) radio-frequency equipment - Radio disturbance characteristics - Limits and methods of measurement EN 60335-1:2002: Household and similar electrical appliances - Safety Part 1: General requirements EN 61000-3-2:2000 Limits for harmonic current emissions EN 61000-3-3:2000 Limitations of voltage fluctuations and flicker In accordance with the regulations of the EU guidelines 73/23/EEC (electrical operating equipment/low-voltage guidelines) 89/336/EEC (electromagnetic compatibility) Flawil, 12.01.06 BÜCHI Labortechnik AG Meierseggstrasse 40 Postfach CH-9230 Flawil 1 Switzerland Tel.: +41 (0)71 394 63 63 Fax: +41 (0)71 394 65 65 [email protected] www.buchi.com 38 Guido Worch Quality Manager B-414 Operation Manual, Version D BÜCHI Labortechnik AG CH-9230 Flawil 1 / Switzerland T +41 71 394 63 63 F +41 71 394 65 65 www.buchi.com 096701 Quality in your hands ">
Advertisement
Key features
- Neutralizes and adsorbs gases
- Condenser or cold trap for vapor and liquid collection
- Integrated fume exhaust pump
- Overpressure valve for safety
- Reusable and replaceable components
- Multiple connection options for digestion units
- Uses activated charcoal for adsorption
- Compatible with various washing solutions
- Includes a silencer for noise reduction
Frequently asked questions
The Scrubber is designed for the neutralization and adsorption of gases that arise during chemical reactions and syntheses.
The Scrubber can handle gases with known chemical composition. It is not recommended for gases with unknown composition.
The Scrubber has a suction capacity designed for Digestion Units with a maximum of 20 sample tubes.
The Scrubber is available in various models, including the basic model, models with condenser or cold trap, and models with condenser, cold trap, and reaction vessel.
The manual recommends sodium hydroxide (8-10%, max 20%) and sodium carbonate for acidic vapors and gases.