Omega FMA1700 and FMA1800 Owner Manual
PDF
Download
Document
Advertisement
Advertisement
User’s Guide Shop online at www.omega.com e-mail: [email protected] FMA 1700 & FMA 1800 Mass Flowmeter OMEGAnet ® Online Service www.omega.com Internet e-mail info@ omega.com Servicing North America: USA: ISO 9001 Certified Canada: One Omega Drive, Box 4047 Stamford CT 06907-0047 Tel: (203) 359-1660 e-mail: [email protected] 976 Bergar Laval (Quebec) H7L 5A1 Tel: (514) 856-6928 e-mail: info@ omega.ca FAX: (203) 359-7700 FAX: (514) 856-6886 For immediate technical or application assistance: USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA® Customer Service: 1-800-622-2378 / 1-800-622-BEST® Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN® TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA Mexico: En Españ ol: (001) 203-359-7803 FAX: (001) 203-359-7807 e-mail: espanol@ omega.com info@ omega.com.mx Servicing Europe: Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643 Toll Free in Benelux: 0800 0993344 e-mail: sales@ omegaeng.nl Czech Republic: Rude´ arm-dy 1868, 733 01 Karvin- 8 Tel: +420 (0)59 6311899 Toll Free: 0800-1-66342 France: FAX: +420 (0)59 6311114 e-mail: info@ omegashop.cz 11, rue Jacques Cartier, 78280 Guyancourt, France Tel: +33 (0)1 61 37 29 00 FAX: +33 (0)1 30 57 54 27 Toll Free in France: 0800 466 342 e-mail: sales@ omega.fr Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany Tel: +49 (0)7056 9398-0 Toll Free in Germany: 0800 639 7678 e-mail: info@ omega.de FAX: +49 (0)7056 9398-29 United Kingdom: One Omega Drive, River Bend Technology Centre ISO 9002 Certified Northbank, Irlam, Manchester M44 5BD United Kingdom Tel: +44 (0)161 777 6611 Toll Free in United Kingdom: 0800-488-48 FAX: +44 (0)161 777 6622 e-mail: sales@ omega.co.uk It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification. The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. WARNING: These products are not designed for use in, and should not be used for, patient-connected applications. TABLE OF CONTENTS 1. UNPACKING THE FMA 1700/1800 MASS FLOW METER..........................1 1 1.1 Inspect Package for External Damage................................................. 1 1.2 Unpack the Mass Flow Meter............................................................... 1 1.3 Returning Merchandise for Repair....................................................... 2. INSTALLATION........................................................................................ 1 2.1 Primary Gas Connections................................................................. 1 2.2 Electrical Connections...................................................................... 2 2.3.1 Remote LCD Readouts..................................................................... 4 2.3.2 Panel Mounting Readouts................................................................ 4 3. PRINCIPLE OF OPERATION................................................................... 4 4. SPECIFICATIONS................................................................................... 5 4.1 CE Compliance................................................................................. 6 5. OPERATING INSTRUCTIONS.................................................................. 8 5.1 Preparation and Warm Up..................................................................8 5.2 Flow Signal Output Readings.......................................................... 8 5.3 Swamping Condition....................................................................... 9 6. MAINTENANCE.........................................................................................9 6.1 Introduction......................................................................................9 6.2 Flow Path Cleaning........................................................................... 9 6.2.1 Restrictor Flow Element (RFE)........................................................ 9 10 6.2.2 FMA 1700/1800 Series Max. Flow 15 L/min....................................... 7. CALIBRATION PROCEDURES................................................................. 11 7.1 Flow Calibration...............................................................................11 12 7.2 Calibration of FMA 1700/1800 Mass Flow Meters............................. 7.2.1 Connections and Initial Warm Up.....................................................12 7.2.2 ZERO Adjustment.............................................................................12 7.2.3 SPAN Adjustment............................................................................ 13 13 7.3 Linearity Adjustment.......................................................................... 7.3.1 Connections and Initial Warm Up....................................................13 7.3.2 ZERO Adjustment.............................................................................13 7.3.3 25% Flow Adjustment......................................................................14 7.3.4 50% Flow Adjustment......................................................................14 7.3.5 75% Flow Adjustment......................................................................14 7.3.6 100% Flow Adjustment....................................................................14 7.4 LCD Display Scaling.........................................................................14 7.4.1 Access LCD Display Circuit..............................................................14 7.4.2 Adjust Scaling..................................................................................15 7.4.3 Change Decimal Point......................................................................15 8. TROUBLESHOOTING............................................................................. 15 8.1 Common Conditions........................................................................15 8.2 Troubleshooting Guide.....................................................................16 8.3 Technical Assistance....................................................................... 20 9. CALIBRATION CONVERSIONS FROM REFERENCE GASES....................... 21 APPENDIX 1 COMPONENT DIAGRAM.........................................................22 APPENDIX 2 GAS FACTOR TABLE (“K” FACTORS)........................................ 23 APPENDIX 3 DIMENSIONAL DRAWINGS...................................................... 27 PARTS OF THE FLOW METER.............................................. 28 WARRANTY.............................................................................. 30 1. UNPACKING THE FMA 1700/1800 MASS FLOW METER 1.1 Inspect Package for External Damage Remove the Packing List and verify that you have received all equipment. If you have any questions about the shipment, please call the OMEGA7 Customer Service Department at 1-800-622-2378 or (203) 359-1660. Your FMA 1700/1800 Mass Flow Meter was carefully packed in a sturdy cardboard carton, with anti-static cushioning materials to withstand shipping shock. Upon receipt, inspect the package for possible external damage. In case of external damage to the package contact the shipping company immediately. 1.2 Unpack the Mass Flow Meter Open the carton carefully from the top and inspect for any sign of concealed shipping damage. In addition to contacting the shipping carrier please forward a copy of any damage report to OMEGA7 directly. When unpacking the instrument please make sure that you have all the items indicated on the Packing List. Please report any shortages promptly. 1.3 Returning Merchandise for Repair Please contact an OMEGA7 customer service representative and request a Return Authorization Number (AR). It is mandatory that any equipment returned for servicing be purged and neutralized of any dangerous contents including but not limited to toxic, bacterially infectious, corrosive or radioactive substances. No work shall be performed on a returned product unless the customer submits a fully executed, signed SAFETY CERTIFICATE. Please request form from the Service Manager. 2. INSTALLATION 2.1 Primary Gas Connections Please note that the FMA 1700/1800 Mass Flow Meter will not operate with liquids. Only clean gases are allowed to be introduced into the instrument. If gases are contaminated they must be filtered to prevent the introduction of impediments into the sensor. 1 CAUTION: FMA 1700/1800 TRANSDUCERS SHOULD NOT BE USED FOR MONITORING OXYGEN GAS UNLESS SPECIFICALLY CLEANED AND PREPARED FOR SUCH APPLICATION. Attitude sensitivity of the Mass Flow Meter is ±15. This means that the gas flow path of the Flow Meter must be horizontal within those stated limits. Should there be need for a different orientation of the meter, re-calibration may be necessary. It is also preferable to install the FMA 1700/1800 transducer in a stable environment, free of frequent and sudden temperature changes, high moisture, and drafts. Prior to connecting gas lines inspect all parts of the piping system including ferrules and fittings for dust or other contaminant’s. Be sure to observe the direction of gas flow as indicated by the arrow on the front of the meter when connecting the gas system to be monitored. Insert tubing into the compression fittings (except FMA 1700/1800 Series Max Flow 1000 L/min) until the ends of the properly sized tubings home flush against the shoulders of the fittings. Compression fittings are to be tightened according to the manufacturer's instructions to one and one quarter turns. Avoid over tightening which will seriously damage the Restrictor Flow Elements (RFE's)! Using a Helium Leak Detector or other equivalent method perform a thorough leak test of the entire system. (All FMA 1700/1800 are checked prior to shipment for leakage within stated limits. See specifications in this manual.) 2.2 Electrical Connections FMA 1700/1800 transducers require a +12VDC (+24VDC optional) power supply with a minimum current rating of 200mA to operate. The operating power input is supplied via the DC power jack or the 9-pin “D” connector located at the side of the flow transducer enclosure. On FMA 1700/1800’s purchased without an LCD readout, a readout panel meter, digital multimeter, or other equivalent device is required to observe the flow signal. DO NOT CONNECT 24Vdc POWER SUPPLY UNLESS YOUR FMA 1700/1800 METER WAS ORDERED AND CONFIGURED FOR 24Vdc 2 FIGURE 2.a - 9-PIN “D” CONNECTOR PINOUTS FOR FMA 1700/1800 TRANSDUCER. PIN FUNCTION 1 Remote LCD display signal 2 0 to 5 VDC output indication 3 0 to 5 VDC common 4 Power supply, positive 5 Power supply, common 6 Remote LCD display reference 7 (unassigned) 8 4 to 20 mA output indication 9 4 to 20 mA common IMPORTANT NOTES: In general, “D” Connector numbering patterns are standardized. There are, however, some connectors with nonconforming patterns and the numbering sequence on your mating connector may or may not coincide with the numbering sequence shown in our pin configuration table above. It is imperative that you match the appropriate wires in accordance with the correct sequence regardless of the particular numbers displayed on your mating connector. Make sure power is OFF when connecting or disconnecting any cables in the system. When connecting power to the FMA 1700/1800 mass flow meter via the DC power jack, do not connect any power supply to the 9-pin 'D' Connector. The DC power jack has a center positive polarity. When battery use is required to power the FMA 1700/1800, use only the optional battery and accompanying charger available from OMEGA7. The power input is protected by a 750mA M (medium time-lag) resettable fuse. If a shorting condition or polarity reversal occurs, the fuse will cut power to the flow transducer circuit. Disconnect the power to the unit, remove the faulty condition, and reconnect the power. The fuse will reset once the faulty condition has been removed. Cable length may not exceed 9.5 feet (3 meters). Use of the FMA 1700/1800 flow transducer in a manner other than that specified in this manual or in writing from OMEGA7, may impair the protection provided by the equipment. 3 2.3.1 Remote LCD Readouts FMA 1700/1800 Mass Flow Meters are available with optional remote reading LCD displays supplied with a three foot long wire to accommodate most applications. This configuration includes the upper block element which serves as the LCD readout mounting. Special lengths of remote extension wiring (up to 9.5 feet [3 meters]) are available on request. 2.3.2 Panel Mounting Readouts Another option for the FMA 1700/1800 Mass Flow Meter is the Panel Mounting Remote Readout. In this configuration the LCD readout is supplied with a three foot long extension wire, and no aluminum housing around the LCD. The LCD readout for panel mounting includes a bezel with two plastic screws which conveniently fit into a rectangular cut-out for panel mounting (see Figure 2.b). FIGURE 2.b - CUTOUT DIMENSIONS FOR LCD PANEL MOUNTING. 3. PRINCIPLE OF OPERATION The stream of gas entering the Mass Flow transducer is split by shunting a small portion of the flow through a capillary stainless steel sensor tube. The remainder of the gas flows through the primary flow conduit. The geometry of the primary conduit and the sensor tube are designed to ensure laminar flow in each branch. According to principles of fluid dynamics the flow rates of a gas in the two laminar flow conduits are proportional to one another. Therefore, the flow rates measured in the sensor tube are directly proportional to the total flow through the transducer. In order to sense the flow in the sensor tube, heat flux is introduced at two sections of the sensor tube by means of precision wound heater-sensor coils. Heat is transferred through the thin wall of the sensor tube to the gas flowing inside. As gas flow takes place heat is carried by the gas stream from the upstream coil to the downstream coil windings. The resultant temperature dependent resistance differential is detected by the electronic control circuit. The measured gradient at the sensor windings is linearly proportional to the instantaneous rate of flow taking place. An output signal is generated that is a function of the amount of heat carried by the gases to indicate mass-molecular based flow rates. 4 4. SPECIFICATIONS FLOW MEDIUM: Please note that FMA 1700/1800 Series Max Flow 15, 50, 100, 200, 500 and 1000 L/min Mass Flow Meters are designed to work with clean gases only. Never try to meter or control flow rates of liquids with any FMA 1700/1800. CALIBRATIONS: Performed at standard conditions [14.7 psia (1.01 bars) and 70 FF (21.1 FC)] unless otherwise requested or stated. ENVIRONMENTAL (per IEC 664): Installation Level II; Pollution Degree II. ACCURACY: ACCURACY %FS OPTIONAL ENHANCED ACCURACY %FS MODEL: FMA 1700 FMA 1700/1800 /1800 100, 200, 500, 15, 50 L/min 1000 L/min FLOW RANGE: 0-100% ACCURACY: ±1.5% MODEL: FLOW 20-100% 0-20% RANGE: ±1.5% ±3% FMA 1700/1800 FMA 1700 100, 200, 500, /1800 15, 50 L/min 1000 L/min 0-100% ACCURACY: ±1% 20-100% 0-20% ±1% REF DATA with ±1% REPEATABILITY: ±0.5% of full scale. TEMPERATURE COEFFICIENT: 0.15% of full scale/ FC. PRESSURE COEFFICIENT: 0.01% of full scale/psi (0.07 bar). RESPONSE TIME: 800 ms time constant; approximately 2 seconds to within ±2% of set flow rate for 25% to 100% of full scale flow rate. GAS PRESSURE: 1000 psig (69 bars) 10, 50 and 100 L/min; 500 psig (34.5 bars) 200, 500 and 1000 L/min; optimum pressure is 20 psig (1.4 bars). GAS AND AMBIENT TEMPERATURE: 32 FF to 122 FF (0 FC to 50 FC). 14 FF to 122 FF (-10 FC to 50 FC) - Dry gases only. RELATIVE GAS HUMIDITY: Up to 70%. LEAK INTEGRITY: 1 x 10- 7 sccs He max. to the outside environment. ATTITUDE SENSITIVITY: No greater than +15 degree rotation from horizontal to vertical; standard calibration is in horizontal position. OUTPUT SIGNALS: Linear 0-5 VDC (1000 Ω minimum load impedance) and 4-20 mA (0-500 Ω loop resistance); 20 mV peak to peak max noise. Contact Omega© for optional RS232 or IEEE488 interfaces. 5 TRANSDUCER INPUT POWER: +12 VDC, 200 mA maximum; +24 VDC optional. Power input is protected by a 750mA M (medium time-lag) resettable fuse, and a rectifier diode for polarity protection. WETTED MATERIALS: FMA 1700/1800 Series Max Flow 15, 50, 100, 200, 500 and 1000 L/min: Anodized aluminum, brass, and 316 stainless steel with FKM O-rings seals; BUNA, EPR or Perflouroelastomer O-rings are optional. FMA 1700/1800 Series Max Flow 15, 50, 100, 200, 500 and 1000 L/min: 316 stainless steel with FKM O-rings seals; BUNA, EPR or Perflouroelastomer O-rings are optional. OMEGA7 makes no expressed or implied guarantees of corrosion resistance of mass flow meters as pertains to different flow media reacting with components of meters. It is the customers' sole responsibility to select the model suitable for a particular gas based on the fluid contacting (wetted) materials offered in the different models. INLET AND OUTLET CONNECTIONS: FMA 1700/1800 Series Max. Flow 15 and 50 L/min - 1/4" compression fittings. Optional: 6mm compression, 1/4" VCR®, 3/8" or 1/8" compression fittings (Max. Flow 15). FMA 1700/1800 Series Max. Flow 100 and 200 L/min - 3/8" compression fittings. FMA 1700/1800 Series Max. Flow 500 L/min - 1/2" compression fittings. FMA 1700/1800 Series Max. Flow 1000 L/min - 3/4" FNPT fittings or 3/4" compression fittings. LCD DISPLAY: 3½ digit LCD (maximum viewable digits “1999”), 0.5 inch high characters. On FMA 1700/1800 aluminum or stainless steel models the LCD display is built into the upper block element and may be tilted over 90 degrees for optimal viewing comfort. Remote or panel mounting remote reading is optional. Standard readings are in direct engineering units for the given gas and flow rate (i.e. liters/minute [slpm], standard cubic centimeters/minute [sccm], standard cubic feet/hour [scfh], etc.). 0 to 100% LCD calibration scaling is available upon request at time of order. Contact OMEGA7 when non-standard display settings are desired. TRANSDUCER INTERFACE CABLE: Optional shielded cable is available mating to the FMA 1700/1800 transducer 9-pin “D” connector. [Cable length may not exceed 9.5 feet (3 meters)]. 4.1 CE Compliance Any model FMA 1700/1800 bearing a CE marking on it, is in compliance with the below stated test standards currently accepted. EMC Compliance with 89/336/EEC as amended; Emission Standard: EN 55011:1991, Group 1, Class B Immunity Standard: EN 55082-2:1992 6 FLOW RANGES TABLE I FMA 1700/1800 SERIES MAX FLOW 15 L/min LOW FLOW MASS FLOW METER* TABLE II FMA 1700/1800 SERIES MAX FLOW 50 L/min MEDIUM FLOW MASS FLOW METER* CODE mL/min [N2] CODE L/min [N2] 02 0 to 10 23 15 04 0 to 20 24 06 0 to 50 26 20 30 08 0 to 100 27 40 10 0 to 200 28 50 12 0 to 500 CODE L/min [N2] 14 0 to 1 16 0 to 2 18 0 to 5 CODE L/min [N2] 0 to 10 40 60 20 TABLE III FMA 1700/1800 SERIES MAX FLOW 100 L/min HIGH FLOW MASS FLOW METER* * Flow rates are stated for Nitrogen at STP conditions [i.e. 70 FF (21.1 FC) at 1 atm]. For other gases use the K factor as a multiplier from APPENDIX 2. 41 80 42 100 43 200 44 500 45 1000 TABLE IV PRESSURE DROPS MAXIMUM FLOW RATE SERIES FLOW RATE [std liters/min] 10 L/min MAXIMUM PRESSURE DROP [mm H2O] [psid] [mbar] up to 10 25 0.04 2.5 15 63 0.09 6.4 20 300 0.44 30 30 800 1.18 81 40 1480 2.18 150 50 2200 3.23 223 60 3100 4.56 314 100 5500 8.08 557 200 2720 4.0 280 500 L/min 500 3400 5.0 340 1000 L/min 1000 6120 9.0 620 50 L/min 100 L/min 200 L/min 7 5. OPERATING INSTRUCTIONS 5.1 Preparation and Warm Up It is assumed that the Mass Flow Meter has been correctly installed and thoroughly leak tested as described in section 2. Make sure the flow source is OFF. Apply power to the unit by plugging the power supply line into the DC power jack (or 9-pin 'D' connector) on the side of the meter. If you are using your own power supply, be sure it is between +12 and +15 VDC with at least 200 mA current capacity. Allow the Mass Flow Meter to warm-up for a minimum of 15 minutes. SUPPLYING DC POWER TO THE POWER JACK AND THE “D” CONNECTOR AT THE SAME TIME WILL DAMAGE THE METER. DC POWER JACK POLARITY IS CENTER POSITIVE. During initial powering of the FMA 1700/1800 transducer, the flow output signal will be indicating a higher than usual output. This is indication that the FMA 1700/1800 transducer has not yet attained it's minimum operating temperature. This condition will automatically cancel within a few minutes and the transducer should eventually zero. If after the 15 minutes warm-up period, the display still indicates a reading of less than ± 3.0 % of F.S., readjust the ZERO potentiometer [R34] through the access hole. Before zero adjustment it is good practice to temporarily disconnect the gas source, to ensure that no see page or leak occurs in to the meter. 5.2 Flow Signal Output Readings Initiate a controlled gas flow after warm up. The flow signal output can be viewed either on the LCD display, remote panel meter, digital multimeter, or other display device. If an LCD display has been supplied with the FMA 1700/1800, the observed reading is in direct engineering units (0 to 100% indication is optional). Analog output flow signals of 0 to 5 VDC and 4 to 20 mA are attained at the appropriate pins of the 9-pin “D” connector (see Figure 2.a) on the side of the FMA 1700/1800 transducer. Meter signal output is linearly proportional to the mass molecular flow rate of the gas being metered. By default calibration is done against 0 to 5 VDC output signal. If 4-20 mA output signal is used for flow indication on the FMA 1700/1800, which was calibrated against 0 to 5 VDC, the total uncertainty of the reading may be in the range of +2.5% of full scale. Optional calibration for 4-20 mA output signal is available upon request at time of order. The full scale range and gas for which your meter has been calibrated are shown on the flow transducers front label. 8 5.3 Swamping Condition If a flow of more than 10% above the maximum flow rate of the Mass Flow Meter is taking place, a condition known as “swamping” may occur. Readings of a “swamped” meter cannot be assumed to be either accurate or linear. Flow must be restored to below 110% of maximum meter range. Once flow rates are lowered to within calibrated range, the swamping condition will end. Operation of the meter above 110% of maximum calibrated flow may increase recovery time. 6. MAINTENANCE 6.1 Introduction It is important that the Mass Flow Meter/Controller is used with clean, filtered gases only. Liquids may not be metered. Since the RTD sensor consists, in part, of a small capillary stainless steel tube, it is prone to occlusion due to impediments or gas crystallization. Other flow passages are also easily obstructed. Therefore, great care must be exercised to avoid the introduction of any potential flow impediment. To protect the instrument a 50 micron (FMA 1700/1800 Series Max Flow 15 L/min) or 60 micron (FMA 1700/1800 Series Max Flow 100 and 200 L/min) filter is built into the inlet of the flow transducer. The filter screen and the flow paths may require occasional cleaning as described below. There is no other recommended maintenance required. It is good practice, however, to keep the meter away from vibration, hot or corrosive environments and excessive RF or magnetic interference. If periodic calibrations are required they should be performed by qualified personnel and calibrating instruments, as described in section 7. It is recommended that units are returned to OMEGA7 for repair service and calibration. 6.2 CAUTION: TO PROTECT SERVICING PERSONNEL IT IS MANDATORY THAT ANY INSTRUMENT BEING SERVICED IS COMPLETELY PURGED AND NEUTRALIZED OF TOXIC, BACTERIOLOGICALLY INFECTED, CORROSIVE OR RADIOACTIVE CONTENTS. Flow Path Cleaning Before attempting any disassembly of the unit for cleaning, try inspecting the flow paths by looking into the inlet and outlet ends of the meter for any debris that may be clogging the flow through the meter. Remove debris as necessary. If the flow path is not unclogged, then proceed with steps below. Do not attempt to disassemble the sensor. If blockage of the sensor tube is not alleviated by flushing through with cleaning fluids, please return meter for servicing. 6.2.1 NOTE: DISASSEMBLY MAY COMPROMISE CURRENT CALIBRATION. Restrictor Flow Element (RFE) The Restrictor Flow Element (RFE) is a precision flow divider inside the transducer, which splits the inlet gas flow by a preset amount to the sensor and main flow paths. The particular RFE used in a given Mass Flow Meter depends on the gas and flow range of the instrument. 9 6.2.2 FMA 1700/1800 Series Max Flow 15 L/min models Unscrew the inlet compression fitting of meter. Note that the Restrictor Flow Element (RFE) is connected to the inlet fitting. Carefully disassemble the RFE from the inlet connection. The 50 micron filter screen will now become visible. Push the screen out through the inlet fitting. Clean or replace each of the removed parts as necessary. If alcohol is used for cleaning, allow time for drying. Inspect the flow path inside the transducer for any visible signs of contaminant. If necessary, flush the flow path through with alcohol. Thoroughly dry the flow paths by flowing clean dry gas through. Carefully re-install the RFE and inlet fitting, avoiding any twisting and deforming the RFE. Be sure that no dust has collected on the O-ring seal. NOTE: OVER TIGHTENING WILL DEFORM AND RENDER THE RFE DEFECTIVE. It is advisable that at least one calibration point be checked after re-installing the inlet fitting-See section (g). IT IS NOT RECOMMENDED TO ATTEMPT TO DISASSEMBLE, OR REPAIR MAXIMUM FLOW RATE SERIES MODELS 50 L/min, 100 L/min, 200 L/min, 500 L/min or 1000 L/min. DISASSEMBLY NECESSITATES RE-CALIBRATION. 10 Remove the block and connected screens. Clean each of the removed parts as necessary. If alcohol is used for cleaning, allow time for drying. Inspect the flow path inside the transducer for any visible signs of contaminants. If necessary, flush the flow path through with alcohol. Thoroughly dry the flow paths by flowing clean dry gas through. Re-install the inlet parts. Be sure that no dust has collected on the O-ring seal. It is advisable that at least one calibration point be checked after re-installing the inlet fitting - see section 7. 7. 7.1 CALIBRATION PROCEDURES NOTE: REMOVAL OF THE FACTORY INSTALLED CALIBRATION SEALS AND/OR ANY ADJUSTMENTS MADE TO THE METER, AS DESCRIBED IN THIS SECTION, WILL VOID ANY CALIBRATION WARRANTY APPLICABLE. Flow Calibration OMEGA7 Engineering Flow Calibration Laboratory offers professional calibration support for Mass Flow Meters, using precision calibrators under strictly controlled conditions. NIST traceable calibrations are available. Calibrations can also be performed at customers' site using available standards. Factory calibrations are performed using NIST traceable precision volumetric calibrators incorporating liquid sealed frictionless actuators. Generally, calibrations are performed using dry nitrogen gas. The calibration can then be corrected to the appropriate gas desired based on relative correction [K] factors shown in the gas factor table - see Appendix 2. A reference gas, other than nitrogen, may be used to better approximate the flow characteristics of certain gases. This practice is recommended when a reference gas is found with thermodynamic properties similar to the actual gas under consideration. The appropriate relative correction factor should be recalculated - see section 9. It is standard practice to calibrate Mass Flow Meters with dry nitrogen gas at 700 F (21.1 FC), 20 psig (1.4 bars) inlet pressure and 0 psig (0 bar) outlet pressure. It is best to calibrate the FMA 1700/1800 transducers to actual operating conditions. Specific gas calibrations of non-toxic and non-corrosive gases are available at specific conditions. Please contact your distributor or OMEGA7 for a price quotation. It is recommended that a flow calibrator of at least four times better collective accuracy than that of the Mass Flow Meter to be calibrated be used. Equipment required for calibration includes a flow calibration standard and a certified high sensitivity multimeter (which together have a collective accuracy of ±0.25% or better), an insulated (plastic) screwdriver, a flow regulator (example: metering needle valve) installed upstream from the Mass Flow Meter and a pressure regulated source of dry filtered nitrogen gas (or other suitable reference gas). The gas and ambient temperature, as well as inlet and outlet pressure conditions should be set up in accordance with actual operating conditions. 11 FIGURE 7.a - CALIBRATION POTENTIOMETER AND JUMPER LOCATIONS 7.2 Calibration of FMA 1700/1800 Mass Flow Meters All adjustments in this section are made from the outside of the meter, there is no need to disassemble any part of the instrument. FMA 1700/1800 Mass Flow Meters may be field recalibrated/checked for the same range they were originally factory calibrated for. When linearity adjustment is needed, or flow range changes are being made proceed to step 7.3. Flow range changes may require a different Restrictor Flow Element (RFE). Consult OMEGA7 for more information. 7.2.1 Connections and Initial Warm Up At the 9-pin “D” connector of the FMA 1700/1800 transducer, connect the multimeter to output pins [2] and [3] for 0-5 VDC (or pins [8] and [9] for 4-20 mA)-(see Figure 2.a). Power up the Mass Flow Meter for at least 30 minutes prior to commencing the calibration procedure. 7.2.2 ZERO Adjustment Shut off the flow of gas into the Mass Flow Meter. To ensure that no seepage or leak occurs into the meter, it is good practice to temporarily disconnect the gas source. Using the multimeter and the insulated screwdriver, adjust the ZERO potentiometer [R34] through the access window for 0 VDC (or 4 mA respectively) at zero flow. 12 7.2.3 SPAN Adjustment Reconnect the gas source. Using the flow regulator, adjust the flow rate to 100% of full scale flow. Check the flow rate indicated against the flow calibrator. If the deviation is less than ±10% of full scale reading, correct the SPAN potentiometer [R33] setting by using the insulated screwdriver through the access window, to eliminate any deviation. If the deviation is larger than ±10% of full scale reading, a defective condition may be present. LIKELY REASONS FOR A MALFUNCTIONING SIGNAL MAY BE: ✓ ✓ ✓ ✓ Occluded or contaminated sensor tube. Leaking condition in the FMA 1700/1800 transducer or the gas line and fittings. For gases other than nitrogen, recheck appropriate “K” factor from the Gas Factor Table. Temperature and/or pressure correction errors. See also section 8. TROUBLESHOOTING. If after attempting to remedy the above conditions, a malfunction still persists, return the meter for factory service, see section 1.1. At this point the calibration is complete. However, it is advisable that several additional points between 0 and 100%, such as 25%, 50%, and 75% flow be checked. If discrepancies are found, proceed to step 7.3 for Linearity Adjustment. 7.3 Linearity Adjustment All adjustments in this section are made from the outside of the meter, there is no need to disassemble any part of the instrument. 7.3.1 Connections and Initial Warm Up At the 9-pin “D” connector of the FMA 1700/1800 transducer, connect the multimeter to output pins [2] and [3] for 0-5 VDC (or pins [8] and [9] for 4-20 mA)-(see Figure 2.a). If calibration to a new flow range or different gas is being performed, it may be necessary to remove any jumpers at J1, J2, and J3 before beginning linearizing procedure. Power up the Mass Flow Meter for at least 30 minutes prior to commencing the calibration procedure. 7.3.2 ZERO Adjustment Shut off the flow of gas into the Mass Flow Meter. To ensure that no seepage or leak occurs into the meter, it is good practice to temporarily disconnect the gas source.Using the multimeter and the insulated screwdriver, adjust the ZERO potentiometer [R34] through the access window for 0 VDC (or 4 mA respectively) at zero flow. 13 7.3.3 25% Flow Adjustment Reconnect the gas source. Using the flow regulator, adjust the flow rate to 25% of full scale flow. Check the flow rate indicated against the flow calibrator. Adjust the setting for potentiometer [R33] by using the insulated screwdriver through the access window, until the output of the flow meter reads 1.25VDC ±63mV (or 8mA ±0.25mA). 7.3.4 50% Flow Adjustment Using the flow regulator, increase the flow rate to 50% of full scale flow. Check the flow rate indicated against the flow calibrator. The output of the flow meter should read 2.50VDC ±63mV (or 12mA ±0.25mA). If the reading is outside of that range, place the jumper at [J1] as appropriate to increase or decrease the signal. Adjust the setting for potentiometer [R38] by using the insulated screwdriver through the access window, until reading is within specification. 7.3.5 75% Flow Adjustment Using the flow regulator, increase the flow rate to 75% of full scale flow. Check the flow rate indicated against the flow calibrator. The output of the flow meter should read 3.75VDC ±63mV (or 16mA ±0.25mA). If the reading is outside of that range, place the jumper at [J2] as appropriate to increase or decrease the signal. Adjust the setting for potentiometer [R39] by using the insulated screwdriver through the access window, until reading is within specification. 7.3.6 100% Flow Adjustment Using the flow regulator, increase the flow rate to 100% of full scale flow. Check the flow rate indicated against the flow calibrator. The output of the flow meter should read 5.00VDC ±63mV (or 20mA ±0.25mA). If the reading is outside of that range, place the jumper at [J3] as appropriate to increase or decrease the signal. Adjust the setting for potentiometer [R40] by using the insulated screwdriver through the access window, until reading is within specification. Repeat steps 7.3.3 to 7.3.6 at least once more. 7.4 LCD Display Scaling It may be desirable to re-scale the output reading on the LCD readout supplied with certain model FMA 1700/1800 transducers. Re-calibration for a new flow range or different engineering units are two examples of when this may be necessary. 7.4.1 Access LCD Display Circuit Carefully remove the LCD from the FMA 1700/1800 or panel mounted surface. Remove the aluminum housing on the side of the connection cable. Slide the LCD assembly out of the aluminum housing. 14 7.4.2 Adjust Scaling Using a digital multimeter connected to either the 0 to 5 VDC or 4 to 20 mA signal at the 9-pin “D” connector, set the flow rate on the FMA 1700/1800 to full scale flow (5 VDC or 20mA). Maintain full scale flow, and adjust the potentiometer [R3] on the LCD printed circuit board to desired full scale flow reading. 7.4.3 Change Decimal Point To change the decimal place on the LCD display readout, simply move the jumper to the appropriate location on the 8-pin header block. The numbers are printed to the side of the connections. Do not attempt to place more than one jumper for decimal setting. JUMPER POSITION MAXIMUM SCALABLE DISPLAY READING “3” 1999 “1” 199.9 “2” 19.99 “3” 1.999 8. TROUBLESHOOTING 8.1 Common Conditions Your Mass Flow Meter was thoroughly checked at numerous quality control points during and after manufacturing and assembly operations. It was calibrated according to your desired flow and pressure conditions for a given gas or a mixture of gases. It was carefully packed to prevent damage during shipment. Should you feel that the instrument is not functioning properly please check for the following common conditions first: Are all cables connected correctly? Are there any leaks in the installation? Is the power supply correctly selected according to requirements? When several meters are used a power supply with appropriate current rating should be selected. Were the connector pinouts matched properly? When interchanging with other manufacturers' equipment, cables and connectors must be carefully wired for correct pin configurations. Is the pressure differential across the instrument sufficient? 15 8.2 Troubleshooting Guide INDICATION lack of reading or output unstable or no zero reading LIKELY REASON REMEDY power supply off check connection of power supply fuse blown disconnect FMA 1700/1800 transducer from power supply; remove the shorting condition or check polarities; fuse resets automatically filter screen obstructed at inlet flush clean or disassemble to remove impediments or replace occluded sensor tube flush clean or or return to factory for replacement pc board defect return to factory for replacement power supply problem check power supply for appropriate output gas leak locate and correct pc board defective return to factory for replacement full scale output at “no flow” defective sensor condition or with valve closed gas leak return to factory for replacement calibration off gas metered is not the same as what meter was calibrated for use matched calibration composition of gas changed see K factor tables in APPENDIX 2 gas leak locate and correct pc board defective return to factory for replacement RFE dirty flush clean or disassemble to remove impediments occluded sensor tube flush clean or return to factory for replacement filter screen obstructed at inlet flush clean or disassemble to remove impediments or replace transducer is not mounted properly check for any tilt or change in the mounting of the transducer; generally, units are calibrated for horizontal installation (relative to the sensor tube) 16 locate and correct INDICATION LIKELY REASON REMEDY No zero reading after 15 Embedded temperature has Readjust ZERO minute warm up time and no been changed. potentiometer R34 through flow condition. the access hole (see page 12 for details). No zero reading after 15 Power supply voltage is minute warm up time and no less than 11.0 Vdc. flow condition. Display reading does not response on zero adjustment. Measure voltage on pins 4 and 5 of the 9 pin D-connector. If voltage is less than 11.0Vdc replace power supply with new one (regulated 12.0 Vdc, 250 mA minimum is recommended). No zero reading after 15 minute warm up time and no flow condition. Display reading does not response on zero adjustment. But analog output 0-5 Vdc can be adjusted from 10mV up to 0.5 Vdc with zero potentiometer R34. Wire is disconnected inside Carefully remove the LCD of the LCD Display. from FMA 1700/1800. Remove the aluminum housing on the side of the connection cable. Side the LCD assembly out of the aluminum housing. Check connection for all four wires. If any wire is disconnected restore connection and readjust zero potentiometer R34 to get zero reading on the display (ensure gas source is disconnected and no seepage or leak occurs in to the meter). LCD Display remains blank when unit is powered up. But flow can be observed on analog output 0-5 Vdc (pis 2 and 3 of the D-connector). LCD Display connector is Check LCD connector not attached to the FMA (remove and reinstall LCD 1700/1800 or connection is connector) loose. Wire is disconnected inside See instructions for pos. 3 of the LCD Display. LCD Display remains blank Power supply is bad or when unit is powered up. No polarity is reversed. response on the flow from analog output 0-5 Vdc (voltage is less than 15 mV). PC board is defective. 17 Measure voltage on pins 4 and 5 of the 9 pin D-connector. If voltage is less than 11.0Vdc replace power supply with new one (regulated 12.0 Vdc, 250 mA minimum is recommended). If polarity is reversed (reading is negative) make correct connection. Return FMA 1700/1800 to factory for repair. INDICATION LIKELY REASON REMEDY LCD Display reading does LCD Display is adjusted for not correspond the correct wrong flow range or flow range according analog engineering units. output 0-5 Vdc signal. Readjust LCD Display scaling for required full scale flow (see 7.4 on page 14). LCD Display reading and analog output 0-5Vdc signal are fluctuate in wide range during the flow measurement. Check external connections to pins 2 and 3 of the D-connector. Make sure the load resistance is more than 1000 Ohm. Output 0-5 Vdc signal (pins 2 and 3 of the D-connector) is shorted on the GND or overloaded. LCD Display reading does Output 0-5Vdc schematic is correspond the correct flow burned out or damaged. range, but 0-5 Vdc output signal does not change (always the same reading or around zero). Return FMA 1700/1800 to factory for repair. LCD Display reading and 0-5 Vdc output voltage do correspond the correct flow range, but 4-20 mA output signal does not change (always the same or reading around 4.0 mA). External loop resistance is open or more than 500 Ohm. Check external connections to pins 8 and 9 of the D-connector. Make sure the loop resistance is less than 500 Ohm. Output 4-20 mA schematic is burned out or damaged. Return FMA 1700/1800 to factory for repair. Calibration is off (no more than 3.0 % F.S.). FMA 1700/1800 has initial zero shift. Shut off the flow of gas into the FMA 1700/1800 (ensure gas source is disconnected and no seepage or leak occurs in to the meter). Wait for 15 minute. with no flow condition and readjust zero potentiometer R34 to get zero reading on the display. 18 INDICATION LIKELY REASON REMEDY LCD Display reading is above maximum flow range and output voltage 0-5 Vdc signal is more than 5.5 Vdc when gas flows through the FMA 1700/1800. Sensor under swamping conditions (flow is more than 10% above maximum flow rate for particular FMA 1700/1800). Lower the flow through FMA 1700/1800 within calibrated range or shut down the flow completely. The swamping condition will end automatically. PC board is defective. Return FMA 1700/1800 to factory for repair. Gas flows through the FMA The gas flow is too low for 1700/1800, but LCD Display particular model of reading and output voltage FMA 1700/1800. 0-5 Vdc signal do not responds on the flow. FMA 1700/1800 Series Max Flow 15 L/min models: RFE is not connected properly to the inlet fitting. Sensor or PC board is defective. Check maximum flow range on transducer's front panel and make required flow adjustment. Unscrew the inlet compression fitting of the meter and reinstall RFE (see 6.2.2 on page 10). NOTE: Calibration accuracy can be affected. Return FMA 1700/1800 to factory for repair. Gas does not flow through Filter screen obstructed at inlet. the FMA 1700/1800 with inlet pressure applied to the inlet fitting. LCD Display reading and output voltage 0-5 Vdc signal show zero flow. Flush clean or disassemble to remove impediments or replace the filter screen (see 6.2 on page 9). NOTE: Calibration accuracy can be affected. Gas flows through the FMA Direction of the gas flow is 1700/1800, but LCD Display reversed. reading is negative and output voltage 0-5 Vdc signal do not responds on the flow (reading near 10mV). FMA 1700/1800 is installed in the installation with back pressure conditions and gas leak exist in the system. Check the direction of gas flow as indicated by the arrow on the front of the meter and make required reconnection in the installation. Gas flows through the FMA 5.00 Vdc reference voltage 1700/1800, but LCD Display (green wire) is disconnected inside of the LCD Display. reading is negative and does not change according to gas flow. Output voltage 0-5 Vdc signal corresponds correct gas flow. Carefully remove the LCD from FMA 1700/1800. Remove the aluminum housing on the side of the connection cable. Side the LCD assembly out of the aluminum housing. Check connection for green wire (5.00 Vdc reference voltage). If any wire is disconnected restore connection. 19 Locate and correct gas leak in the system. If FMA 1700/1800 has internal leak return it to factory for repair. INDICATION LIKELY REASON REMEDY Sensor or PC board is FMA 1700/1800 is defective. disconnected from the source of the gas (no flow conditions) but LCD Display reading is fluctuating in wide range. Output voltage 0-5 Vdc signal also fluctuating. The power supply voltage is 12 Vdc and stable. Return FMA 1700/1800 to factory for repair. Reading on the LCD Display Decimal point jumper is installed in wrong position ten (hundreds) times less on the LCD Display Circuit. or more than actual gas flow. Output voltage 0-5 Vdc signal corresponds correct gas flow. Carefully remove the LCD from FMA 1700/1800. Remove the aluminum housing on the side of the connection cable. Side the LCD assembly out of the aluminum housing. Reinstall jumper to the appropriate location on the 8-pin header block (see 7.4.3 on page 15). For best results it is recommended that instruments are returned to the factory for servicing. See section 1.3 for return procedures. 8.3 Technical Assistance OMEGA7 Engineering will provide technical assistance over the phone to qualified repair personnel. Please call our Flow Department at 800-872-9436 Ext. 2298. 20 9. CALIBRATION CONVERSIONS FROM REFERENCE GASES The calibration conversion incorporates the K factor. The K factor is derived from gas density and coefficient of specific heat. For diatomic gases: 1 d X Cp where d = gas density (gram/liter) Cp = coefficient of specific heat (cal/gram) K gas = Note in the above relationship that d and Cp are usually chosen at the same conditions (standard, normal or other). If the flow range of a Mass Flow Meter remains unchanged, a relative K factor is used to relate the calibration of the actual gas to the reference gas. K = Qa Qr where Qa Qr Ka Kr = = = = = Ka Kr mass flow rate of an actual gas (sccm) mass flow rate of a reference gas (sccm) K factor of an actual gas K factor of a reference gas For example, if we want to know the flow rate of oxygen and wish to calibrate with nitrogen at 1000 SCCM, the flow rate of oxygen is: QO2 = Qa = Qr X K = 1000 X 0.9926 = 992.6 sccm where K = relative K factor to reference gas (oxygen to nitrogen) 21 APPENDIX 1 COMPONENTS DIAGRAM PHONE JACK 9-PIN “D” CONNECTOR DC POWER JACK FMA 1700/1800 METERING PC BOARD 22 APPENDIX 2 GAS FACTOR TABLE (“K” FACTORS) CAUTION: K-Factors at best are only an approximation. K factors should not be used in applications that require accuracy better than +/- 5 to 10%. ACTUAL GAS K FACTOR Relative to N2 Cp [Cal/g] Density [g/I] .5829 1.0000 .4346 .7310 1.4573 1.205 .6735 .4089 .5082 .8083 .38 .26 .3855 .3697 .3224 .2631 .2994 .324 .291 .7382 .658 .4036 .240 .352 .492 .1244 .1244 .1167 .1279 .1778 .0539 .0647 .1369 .1161 .1113 .3514 .4007 .3648 .336 .374 .2016 .2016 1.162 1.293 1.787 .760 1.782 1.782 3.478 5.227 3.025 7.130 11.18 7.803 6.108 6.644 2.413 2.593 2.503 2.503 2.503 1.964 1.964 .6026 1.00 .31 .42 .5428 .6606 .86 .4016 .4589 .3912 .2418 .3834 .61 .6130 .4584 .1428 .2488 .1655 .1654 .1710 .1651 .114 .1650 .1544 .1309 .164 .153 .2613 .1739 .3177 3.397 1.250 6.860 3.926 2.945 2.680 3.163 4.125 3.858 5.326 6.892 4.660 2.322 2.742 1.877 Acetylene C2H2 Air Allene (Propadiene) C3H4 Ammonia NH3 *Argon Ar *Argon AR-1 (>10 L/min) Arsine AsH3 Boron Trichloride BCl3 Boron Trifluoride BF3 Bromine Br2 Boron Tribromide Br3 Bromine PentaTrifluoride BrF5 Bromine Trifluoride BrF3 Bromotrifluoromethane (Freon-13 B1) CBrF3 1,3-Butadiene C4H6 Butane C4H10 1-Butene C4H8 2-Butene C4H8 CIS 2-Butene C4H8 TRANS *Carbon Dioxide CO2 *Carbon Dioxide CO2-1 (>10 L/min) Carbon Disulfide CS2 Carbon Monoxide C0 Carbon Tetrachloride CCl4 Carbon Tetrafluoride (Freon-14)CF4 Carbonyl Fluoride COF2 Carbonyl Sulfide COS Chlorine Cl2 Chlorine Trifluoride ClF3 Chlorodifluoromethane (Freon-22)CHClF2 Chloroform CHCl3 Chloropentafluoroethane(Freon-115)C2ClF5 Chlorotrifluromethane (Freon-13) CClF3 CyanogenC2N2 CyanogenChloride CICN Cyclopropane C3H5 * Flow rates indicated ( ) is the maximum flow range of the Mass Flow meter being used. 23 ACTUAL GAS K FACTOR Relative to N2 Cp [Cal/g] Density [g/I] Deuterium D2 Diborane B2H6 Dibromodifluoromethane CBr2F2 Dichlorodifluoromethane (Freon-12) CCl2F2 Dichlofluoromethane (Freon-21) CHCl2F Dichloromethylsilane (CH3)2SiCl2 Dichlorosilane SiH2Cl2 Dichlorotetrafluoroethane (Freon-114) C2Cl2F4 1,1-Difluoroethylene (Freon-1132A) C2H2F2 Dimethylamine (CH3)2NH Dimethyl Ether (CH3)2O 2,2-Dimethylpropane C3H12 Ethane C2H6 Ethanol C2H6O Ethyl Acetylene C4H6 Ethyl Chloride C2H5Cl Ethylene C2H4 Ethylene Oxide C2H4O Fluorine F2 Fluoroform (Freon-23) CHF3 Freon-11 CCl3F Freon-12 CCl2F2 Freon-13 CClF3 Freon-13B1 CBrF3 Freon-14 CF4 Freon-21 CHCl2F Freon-22 CHClF2 Freon-113 CCl2FCClF2 Freon-114 C2Cl2F4 Freon-115 C2ClF5 Freon-C318 C4F8 Germane GeH4 Germanium Tetrachloride GeCl4 *Helium He *Helium He-1 (>50 L/min) *Helium He-2 (>10-50 L/min) Hexafluoroethane C2F6 (Freon-116) Hexane C6H14 *Hydrogen H2-1 *Hydrogen H2-2 (>10-100 L) *Hydrogen H2-3 (>100 L) 1.00 .4357 .1947 .3538 .4252 .2522 .4044 .2235 .4271 .3714 .3896 .2170 .50 .3918 .3225 .3891 .60 .5191 .9784 .4967 .3287 .3538 .3834 .3697 .4210 .4252 .4589 .2031 .2240 .2418 .1760 .5696 .2668 1.454 2.43 2.05 .2421 .1792 1.0106 1.35 1.9 1.722 .508 .15 .1432 .140 .1882 .150 .1604 .224 .366 .3414 .3914 .420 .3395 .3513 .244 .365 .268 .1873 .176 .1357 .1432 .153 .1113 .1654 .140 .1544 .161 .160 .164 .185 .1404 .1071 1.241 1.241 1.241 .1834 .3968 3.419 3.419 3.419 1.799 1.235 9.362 5.395 4.592 5.758 4.506 7.626 2.857 2.011 2.055 3.219 1.342 2.055 2.413 2.879 1.251 1.965 1.695 3.127 6.129 5.395 4.660 6.644 3.926 4.592 3.858 8.360 7.626 6.892 8.397 3.418 9.565 .1786 .1786 .1786 6.157 3.845 .0899 .0899 .0899 * Flow rates indicated ( ) is the maximum flow range of the Mass Flow meter being used. 24 ACTUAL GAS K FACTOR Relative to N2 Cp [Cal/g] Density [g/I] 1.000 1.000 .764 .9998 .9987 .7893 .80 .2492 .27 .2951 1.453 .7175 .75 .0861 .1912 .3171 .3479 .0545 .1025 .2397 .1108 .3872 .3701 .0593 .5328 .5328 3.610 1.627 1.206 .893 5.707 3.613 1.520 9.90 3.593 2.503 3.739 .7175 .7175 .5843 .4313 .5835 .6299 .68 .5180 .2499 .2126 .3512 .51 1.46 .990 1.000 .737 .4802 .6134 .7128 .176 .9926 .6337 .446 .2554 .2134 .3950 .174 .4438 .759 .3274 .3547 .1106 .1926 .3221 .2459 .164 .1373 .387 .4343 .246 .2328 .2485 .1933 .1797 .1632 .2088 .185 .2193 .1917 .195 .38 .398 .1514 .197 .1394 .2374 1.429 1.787 4.236 2.253 1.518 2.146 6.669 9.366 2.011 1.386 .900 1.339 1.25 2.052 3.168 2.920 1.964 8.397 1.427 2.406 2.144 2.816 3.219 4.571 8.388 4.418 1.517 Hydrogen Bromide HBr Hydrogen Chloride HCl Hydrogen Cyanide HCN Hydrogen Fluoride HF Hydrogen Iodide HI Hydrogen Selenide H2Se Hydrogen Sulfide H2S Iodine Pentafluoride IF5 Isobutane CH(CH3)3 Isobutylene C4H6 Krypton Kr *Methane CH4 *Methane CH4-1 (>10 L/min) Methanol CH3 Methyl Acetylene C3H4 Methyl Bromide CH2Br Methyl Chloride CH3Cl Methyl Fluoride CH3F Methyl Mercaptan CH3SH Methyl Trichlorosilane (CH3)SiCl3 Molybdenum Hexafluoride MoF6 Monoethylamine C2H5NH2 Monomethylamine CH3NH2 Neon NE Nitric Oxide NO Nitrogen N2 Nitrogen Dioxide NO2 Nitrogen Trifluoride NF3 Nitrosyl Chloride NOCl Nitrous Oxide N2O Octafluorocyclobutane (Freon-C318) C4F8 Oxygen O2 Oxygen Difluoride OF2 Ozone Pentaborane B5H9 Pentane C5H12 Perchloryl Fluoride ClO3F Perfluoropropane C3F8 Phosgene COCl2 Phosphine PH3 * Flow rates indicated ( ) is the maximum flow range of the Mass Flow meter being used. 25 ACTUAL GAS K FACTOR Relative to N2 Cp [Cal/g] Density [g/I] .36 .3021 .30 .35 .40 .5982 .284 .3482 .69 .2635 .3883 .5096 .3237 .3287 .3278 .1324 .1610 .1250 .399 .366 .3189 .1270 .1691 .1488 .1592 .1543 .127 .182 .1357 .1380 6.843 5.620 6.127 1.967 1.877 1.433 7.580 4.643 2.858 6.516 4.562 4.224 4.64 6.129 6.043 .2031 .161 8.36 Phosphorous Oxychloride POCl3 Phosphorous Pentafluoride PH5 Phosphorous Trichloride PCl3 Propane C3H8 Propylene C3H6 Silane SiH4 Silicon Tetrachloride SiCl4 Silicon Tetrafluoride SiF4 Sulfur Dioxide SO2 Sulfur Hexafluoride SF6 Sulfuryl Fluoride SO2F2 Tetrafluoroethane (Forane 134A) CF3CH2F Tetrafluorohydrazine N2F4 Trichlorofluoromethane (Freon-11) CCl3F Trichlorosilane SiHCl3 1,1,2-Trichloro-1,2,2 Trifluoroethane (Freon-113) CCl2FCClF2 Triisobutyl Aluminum (C4H9)AL Titanium Tetrachloride TiCl4 Trichloro Ethylene C2HCl3 Trimethylamine (CH3)3N Tungsten Hexafluoride WF6 Uranium Hexafluoride UF6 Vinyl Bromide CH2CHBr Vinyl Chloride CH2CHCl Xenon Xe .0608 .2691 .32 .2792 .2541 .1961 .4616 .48 1.44 26 .508 .120 .163 .3710 .0810 .0888 .1241 .12054 .0378 8.848 8.465 5.95 2.639 13.28 15.70 4.772 2.788 5.858 APPENDIX 3 DIMENSIONAL DRAWINGS 2.38 0.95 FLOW MAXIMUM FLOW SERIES DIMENSION (INCH) CONNECTION Compression Fitting (Except Model FMA 1700/1800 Series Max Flow 1000 L/min) NO LCD LCD VERSION A B C/*C D/*D E/*E F G H 10 L/min ¹/₄" Tube O Diameter 5.60 1.00 1.00 3.00 5.02 0.69 2.69 4.50 50 L/min ¹/₄" Tube O Diameter 5.98 1.37 1.25 4.13 6.15 0.69 2.69 4.88 100 L/min ³/₈" Tube O Diameter 5.98 1.37 1.25 4.13 6.27 0.69 2.69 4.88 200 L/min ³/₈" Tube O Diameter 6.60 2.00 1.75 6.69 8.83 0.99 4.69 5.50 500 L/min ¹/₂" Tube O Diameter 7.60 3.00 3.00 7.25 9.67 2.250 6.750 6.50 1000 L/min ³/₄" NPT Female 8.60 4.00 4.00 7.30 - 3.000 6.800 7.50 NOTES: OMEGA7 reserves the right to change designs and dimensions at its sole discretion at any time without notice. For certified dimensions please contact OMEGA7. 27 PARTS OF THE FLOW METER LEFT AND RIGHT VIEWS FRONT VIEW 28 NOTES: 29 NOTES: 30 WARRANTY/DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product. If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs. OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages. CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner. RETURN REQUESTS/INQUIRIES Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence. The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit. FOR WARRANTY RETURNS, please have the following information available BEFORE contacting OMEGA: 1. Purchase Order number under which the product was PURCHASED, 2. Model and serial number of the product under warranty, and 3. Repair instructions and/or specific problems relative to the product. FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the following information available BEFORE contacting OMEGA: 1. Purchase Order number to cover the COST of the repair, 2. Model and serial number of the product, and 3. Repair instructions and/or specific problems relative to the product. OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. OMEGA is a registered trademark of OMEGA ENGINEERING, INC. © Copyright 2001 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC. 31 Where Do I Find Everything I Need for Process Measurement and Control? OMEGA …Of Course! Shop online at www.omega.com TEMPERATURE Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies Wire: Thermocouple, RTD & Thermistor Calibrators & Ice Point References Recorders, Controllers & Process Monitors Infrared Pyrometers PRESSURE, STRAIN AND FORCE Transducers & Strain Gages Load Cells & Pressure Gages Displacement Transducers Instrumentation & Accessories FLOW/LEVEL Rotameters, Gas Mass Flow meter & Flow Computers Air Velocity Indicators Turbine/Paddlewheel Systems Totalizers & Batch Controllers pH/CONDUCTIVITY pH Electrodes, Testers & Accessories Benchtop/Laboratory Meters Controllers, Calibrators, Simulators & Pumps Industrial pH & Conductivity Equipment DATA ACQUISITION Data Acquisition & Engineering Software Communications-Based Acquisition Systems Plug-in Cards for Apple, IBM & Compatibles Datalogging Systems Recorders, Printers & Plotters HEATERS Heating Cable Cartridge & Strip Heaters Immersion & Band Heaters Flexible Heaters Laboratory Heaters ENVIRONMENTAL MONITORING AND CONTROL Metering & Control Instrumentation Refractometers Pumps & Tubing Air, Soil & Water Monitors Industrial Water & Wastewater Treatment pH, Conductivity & Dissolved Oxygen Instruments M-1680/0710 ">

Public link updated
The public link to your chat has been updated.
Advertisement