16716A User's Guide

16716A User's Guide
Help Volume
© 1992-2002 Agilent Technologies. All rights reserved.
Agilent Technologies 16716A
Logic Analyzer
Agilent Technologies 16716A Logic Analyzer
The Agilent Technologies 16716A 167 MHz State/2 GHz Timing Zoom
logic analyzer offers 512K deep memory and very fast sample rates - up
to 2 GHz for areas around the trigger - with up to 340 channels.
“Getting Started” on
page 11
“Task Guide” on
page 31
•
“Step 1. Connect the logic analyzer to the device under test” on page 13
•
“Step 2. Choose the sampling mode” on page 14
•
“Step 3. Format labels for the probed signals” on page 17
•
“Step 4. Define the trigger condition” on page 20
•
“Step 5. Run the measurement” on page 21
•
“Step 6. Display the captured data” on page 22
•
“Probing the Device Under Test” on page 33
•
“Choosing the Sampling Mode” on page 36
•
“Selecting the Timing Mode (Asynchronous Sampling)” on page 36
•
“Selecting the State Mode (Synchronous Sampling)” on page 43
•
“In Either Timing Mode or State Mode” on page 51
•
“Using 2 GHz Timing Zoom” on page 53
•
“Formatting Labels for Logic Analyzer Probes” on page 56
•
“Setting Up Triggers and Running Measurements” on page 63
2
•
“Using Trigger Functions” on page 64
•
“Using State Mode Trigger Features” on page 69
•
“Editing the Trigger Sequence” on page 71
•
“Editing Advanced Trigger Functions” on page 76
•
“Saving/Recalling Trigger Setups” on page 83
Agilent Technologies 16716A Logic Analyzer
•
“Reference” on
page 111
“Concepts” on
page 189
See Also
“Running Measurements” on page 84
•
“Displaying Captured Data” on page 87
•
“Using Symbols” on page 94
•
“Printing/Exporting Captured Data” on page 103
•
“Solving Logic Analysis Problems” on page 107
•
“Saving and Loading Logic Analyzer Configurations” on page 109
•
“The Sampling Tab” on page 113
•
“The Format Tab” on page 117
•
“The Trigger Tab” on page 144
•
“The Symbols Tab” on page 157
•
“Error Messages” on page 168
•
“Specifications and Characteristics” on page 184
•
“Understanding Logic Analyzer Triggering” on page 190
•
“Understanding State Mode Sampling Positions” on page 206
Main System Help (see the Agilent Technologies 16700A/B-Series Logic
Analysis System help volume)
Glossary (see page 209)
3
Agilent Technologies 16716A Logic Analyzer
4
Contents
Agilent Technologies 16716A Logic Analyzer
1 Getting Started
Step 1. Connect the logic analyzer to the device under test
Step 2. Choose the sampling mode
14
Step 3. Format labels for the probed signals
Step 4. Define the trigger condition
Step 5. Run the measurement
17
20
21
Step 6. Display the captured data
For More Information...
13
22
24
Example: Timing measurement on counter board
Example: State measurement on counter board
26
28
2 Task Guide
Probing the Device Under Test
Choosing the Sampling Mode
33
36
Selecting the Timing Mode (Asynchronous Sampling)
36
5
Contents
To select transitional timing or store qualified
37
More on Store Qualification in Transitional Timing 38
More on Storing Transitions 38
Transitional Timing Considerations 39
Selecting the State Mode (Synchronous Sampling) 43
In Either Timing Mode or State Mode 51
Using 2 GHz Timing Zoom 53
Formatting Labels for Logic Analyzer Probes
56
To assign pods to one or two analyzers 56
To set pod threshold voltages 57
To assign probe channels to labels 58
To change the label polarity 60
To reorder bits in a label 60
To turn labels off or on 61
Setting Up Triggers and Running Measurements
Using Trigger Functions 64
Using State Mode Trigger Features 69
Editing the Trigger Sequence 71
Editing Advanced Trigger Functions 76
Saving/Recalling Trigger Setups 83
Running Measurements 84
Displaying Captured Data
87
To open Waveform or Listing displays 87
To display Timing Zoom data 88
To use other display tools 89
If the captured data doesn't look correct 91
If there are filtered data holes in display memory 91
To display symbols for data values 92
To cancel the display processing of captured data 93
6
63
Contents
Using Symbols
94
To load object file symbols 95
To adjust symbol values for relocated code
To create user-defined symbols 97
To enter symbolic label values 98
To create an ASCII symbol file 99
To create a readers.ini file 100
Printing/Exporting Captured Data
Cross-Triggering
96
103
105
To cross-trigger between two analyzers 105
To cross-trigger with another instrument 106
Solving Logic Analysis Problems
107
To test the logic analyzer hardware
107
Saving and Loading Logic Analyzer Configurations
109
3 Reference
The Sampling Tab
113
Timing Mode 114
State Mode 115
The Format Tab
117
7
Contents
Importing Netlist and ASCII Files
119
Exporting ASCII Files 121
Importing ASCII Files 121
Termination Adapter 123
E5346A High Density Adapter 124
Mapping Connector Names 125
Import the Net List File 125
Verify Net to Label Mapping 126
Select/Create Interface Labels 127
Pod Assignment Dialog 128
Sampling Positions Dialog 129
The Trigger Tab
144
Trigger Functions Subtab 145
Settings Subtab 152
Overview Subtab 153
Default Storing Subtab 154
Status Subtab 155
Save/Recall Subtab 155
The Symbols Tab
157
Symbols Selector Dialog 159
Symbol File Formats 161
General-Purpose ASCII (GPA) Symbol File Format 162
8
Contents
Error Messages
168
Must assign Pod 1 on the master card to specify actions for flags 169
Branch expression is too complex 169
Cannot specify range on label with clock bits that span pod pairs 174
Counter value checked as an event, but no increment action specified 175
Goto action specifies an undefined level 175
Maximum of 32 Channels Per Label 175
Hardware Initialization Failed 176
Must assign another pod pair to specify actions for flags 176
No more Edge/Glitch resources available for this pod pair 176
No more Pattern resources available for this pod pair 177
No Trigger action found in the trace specification 177
Slow or Missing Clock 178
Timer value checked as an event, but no start action specified 178
Trigger function initialization failure 179
Trigger inhibited during timing prestore 180
Trigger Specification is too complex 180
Waiting for Trigger 182
Analyzer armed from another module contains no "Arm in from IMB"
event 183
Specifications and Characteristics
184
Agilent 16716A Logic Analyzer Specifications 184
Agilent 16716A Logic Analyzer Characteristics 184
What is a Specification? 187
What is a Characteristic? 188
9
Contents
4 Concepts
Understanding Logic Analyzer Triggering
The Conveyor Belt Analogy 190
Summary of Triggering Capabilities 192
Sequence Levels 192
Boolean Expressions 195
Branches 196
Edges 196
Ranges 196
Flags 197
Occurrence Counters and Global Counters
Timers 198
Storage Qualification 199
Strategies for Setting Up Triggers 201
Conclusions 205
190
197
Understanding State Mode Sampling Positions
Glossary
Index
10
206
1
Getting Started
After you have connected the logic analyzer probes to your device
under test (see “Step 1. Connect the logic analyzer to the device under
test” on page 13), any measurement will have the following basic steps:
11
Chapter 1: Getting Started
•
“Step 2. Choose the sampling mode” on page 14
•
“Step 3. Format labels for the probed signals” on page 17
•
“Step 4. Define the trigger condition” on page 20
•
“Step 5. Run the measurement” on page 21
•
“Step 6. Display the captured data” on page 22
If you have previously saved a logic analyzer setup to a configuration
file, or if configuration files are included with an analysis probe, you
can load the configuration file to set up the logic analyzer and define
the trigger condition.
Once you have made a logic analyzer measurement, the measurement
can be refined by repeating steps 4 - 6.
Next: “Step 1. Connect the logic analyzer to the device under test” on
page 13
12
Chapter 1: Getting Started
Step 1. Connect the logic analyzer to the device under test
Step 1. Connect the logic analyzer to the device
under test
Before you begin setting up the logic analyzer for a measurement, you
need to physically connect the logic analyzer to your device under test.
There are several ways to connect logic analyzer probes to the device
under test:
•
Using the general-purpose probes, the standard flying lead set, and
grabbers to connect to pins and leads in the device under test.
•
By designing connectors (headers) into the device under test so that logic
analyzer probe cables and termination adapters can plug in directly.
•
By designing connectors (headers) and terminations into the device under
test so that logic analyzer probe cables can plug in directly.
•
Using an analysis probe to connect to microprocessors and standard
buses.
When using an analysis probe, the Setup Assistant guides you through the
connection and setup process for your particular logic analyzer and
analysis probe.
When connecting logic analyzer probes to the device under test:
1. Attach the logic analyzer probes to the device under test in a way that
keeps logically-related channels together.
2. Be sure to ground each pod.
Next: “Step 2. Choose the sampling mode” on page 14
13
Chapter 1: Getting Started
Step 2. Choose the sampling mode
Step 2. Choose the sampling mode
There are two logic analyzer sampling modes to choose from: timing
mode and state mode.
In timing mode, the logic analyzer samples asynchronously, based on
an internal sampling clock signal.
In state mode, the logic analyzer samples synchronously, based on a
sampling clock signal (or signals) from the device under test. Typically,
the signal used for sampling in state mode is a state machine or
microprocessor clock signal.
To choose the sampling mode
1. In the Sampling tab, choose Timing Mode or State Mode.
14
Chapter 1: Getting Started
Step 2. Choose the sampling mode
If you chose Timing Mode
1. Select the timing analyzer full/half channel configuration.
Typically, you can choose a half-channel configuration with faster
sampling and greater memory depth, but with half of the channels.
2. Set the sample period.
To capture signal level changes reliably, the sample period should be less
than half (many engineers prefer one-fourth) of the period of the fastest
signal you want to measure.
If you chose State Mode
1. Select the state analyzer speed configuration (if there is a configuration
option).
2. In the Clock Setup, using the Master only mode, specify which clock signal
edges from the device under test will be used as the sampling clock.
15
Chapter 1: Getting Started
Step 2. Choose the sampling mode
You can also specify clock input signal levels (from the device under test)
that will enable (qualify) the sampling clock.
In either sampling mode
1. Specify the trigger position.
The trigger is the event in the device under test that you want to capture
data around.
Specify whether you want to look at data after the trigger (Start), before
and after the trigger (Center), before the trigger (End), or use a
percentage of the logic analyzer's memory for data after the trigger (User
Defined).
2. Set the acquisition memory depth.
If you need less data and want measurements to run faster, you can limit
the amount of trace memory that is filled with samples.
Next: “Step 3. Format labels for the probed signals” on page 17
16
Chapter 1: Getting Started
Step 3. Format labels for the probed signals
Step 3. Format labels for the probed signals
When a logic analyzer probes hundreds of signals in a device under
test, you need to be able to give those channels more meaningful
names than "pod 1, channel 1".
The Format tab is mainly for assigning bus and signal names (from the
device under test), to logic analyzer channels. These names are called
labels. Labels are used when setting up triggers and displaying
captured data.
The Format tab also lets you do things like assign pod pairs to one or
two logic analyzers and specify the logic analyzer threshold voltage.
The Format tab has activity indicators that show whether the signal a
channel is probing is above the threshold voltage (high), below the
threshold voltage (low), or transitioning.
17
Chapter 1: Getting Started
Step 3. Format labels for the probed signals
To assign pods to one or two logic analyzers
A logic analyzer's pod pairs can be assigned to one or two separate
logic analyzers or they can be left unassigned.
1. In the Format tab, select the Pod Assignment button.
2. In the Pod Assignment dialog, drag a pod pair to the appropriate logic
analyzer.
3. Select the Close button.
To specify threshold voltages
The threshold voltage is the voltage level that a signal must cross
before the logic analyzer recognizes a change in logic levels.
1. In the Format tab, select the button under the pod name.
2. In the Pod threshold dialog, select the desired Standard or User Defined
threshold voltage.
3. Select the Close button.
18
Chapter 1: Getting Started
Step 3. Format labels for the probed signals
To assign names to logic analyzer channels
1. Select a label button, and either:
•
Choose the Rename command, enter the label name, and select the OK
button.
•
Or, choose the Insert before or Insert after command, enter the label
name, and select the OK button.
2. In the label row, select the button of the pod that contains the channels
you want to assign.
3. Either choose one of the standard label assignments--dots (.) mean the
channel is unassigned, asterisks (*) mean the channel is assigned--or
choose Individual.
If you chose Individual:
a. In the "label - pod" dialog, select the channels you want to assign/
unassign.
b. Select the OK button.
Next: “Step 4. Define the trigger condition” on page 20
19
Chapter 1: Getting Started
Step 4. Define the trigger condition
Step 4. Define the trigger condition
The trigger is the event in the device under test that you want to
capture data around.
1. In the Trigger tab, and in the Trigger Functions subtab, choose the type of
trigger you want to specify, and select the Replace button.
2. In the Trigger Sequence portion of the Trigger tab, select the buttons to
define the label values and/or other conditions you want to trigger on.
Next: “Step 5. Run the measurement” on page 21
20
Chapter 1: Getting Started
Step 5. Run the measurement
Step 5. Run the measurement
Once the trigger condition has been defined, you can run the
measurement.
1. Select the Run Single button
.
When you run a measurement, the Stop button becomes available while
the logic analyzer looks for the trigger condition.
Logic analyzers with deep acquisition memory take a noticeable amount of
time to complete a run; however, messages like "Waiting in level 1" may
indicate you need to stop the measurement and refine the trigger
condition.
When the trigger condition is found, logic analyzer acquisition memory is
filled, the captured data is processed to the display tools, and the Run
Single button becomes available again.
Next: “Step 6. Display the captured data” on page 22
21
Chapter 1: Getting Started
Step 6. Display the captured data
Step 6. Display the captured data
Once you have run a measurement and filled the logic analyzer's
acquisition memory with captured data, you can display it with one of
the display tools.
To open Waveform or Listing displays
Waveform displays are typically used when data is captured with the
timing sampling mode, and Listing displays are used when data is
captured with the state sampling mode.
1. From the Window menu, select your logic analyzer and choose the
Waveform or Listing command.
To add display tools via the Workspace window
1. Select the Workspace button (or from the Window menu, select System
and Workspace).
2. In the Workspace window, scroll down to the Display portion of the tool
icon list.
22
Chapter 1: Getting Started
Step 6. Display the captured data
3. Drag the display tool icon and drop it on the analyzer icon.
4. To open the display tool, select its icon and choose the Display command.
Next: “For More Information...” on page 24
23
Chapter 1: Getting Started
For More Information...
For More Information...
On making
•
measurements on the
demo counter board: •
“Example: Timing measurement on counter board” on page 26
“Example: State measurement on counter board” on page 28
•
Making Basic Measurements for a self-paced tutorial
•
“Probing the Device Under Test” on page 33
•
Setup Assistant (see the Setup Assistant help volume) (when using
analysis probes).
•
Logic Analysis System and Measurement Modules Installation Guide
for probe pinout and circuit diagrams.
On choosing the
sampling mode:
•
“Choosing the Sampling Mode” on page 36
•
“The Sampling Tab” on page 113
On formatting labels
for probed signals:
•
“Formatting Labels for Logic Analyzer Probes” on page 56
•
“The Format Tab” on page 117
On defining the
trigger condition:
•
“Understanding Logic Analyzer Triggering” on page 190
•
“Setting Up Triggers and Running Measurements” on page 63
•
“The Trigger Tab” on page 144
On running
measurements:
•
“Running Measurements” on page 84
On displaying
captured data:
•
“Displaying Captured Data” on page 87
•
Using the Waveform Display Tool (see the Waveform Display Tool help
volume)
•
Using the Listing Display Tool (see the Listing Display Tool help volume)
•
Working with Markers (see the Markers help volume)
•
Using the Chart Display Tool (see the Chart Display Tool help volume)
•
Using the Distribution Display Tool (see the Distribution Display Tool
On connecting the
logic analyzer:
24
Chapter 1: Getting Started
For More Information...
help volume)
•
Using the Compare Analysis Tool (see the Compare Tool help volume)
25
Chapter 1: Getting Started
Example: Timing measurement on counter board
Example: Timing measurement on counter
board
This example uses the demo counter board that is supplied with the
Making Basic Measurements kit as the device under test. The kit is
supplied with every logic analysis system, or can be ordered from your
Agilent Technologies Sales Office.
To connect the logic analyzer to the device under test
1. Connect Pod 1 of the logic analyzer to J1 on the demo counter board.
The demo counter board has built-in terminations and header connectors.
To choose the sampling mode
1. In the Sampling tab, choose Timing Mode.
2. Enter a sample period of 3.0 ns.
To format labels for the probed signals
1. In the Format tab, select the button under the pod 1.
2. In the Pod threshold dialog, select TTL; then, select the Close button.
3. Select a label button.
4. Choose the Rename command, enter the label name "TCOUNT", and
select the OK button.
5. In the label row, select the button under pod 1.
6. Choose the "........********" standard label assignment to assign the lower 8
bits of pod 1 to the "TCOUNT" label.
To define the trigger condition
1. In the Trigger tab, and in the Trigger Functions subtab, choose the "Find
edge" trigger function, and select the Replace button.
2. In the Trigger Sequence portion of the Trigger tab, select the Edge button
and use the Specify Glitch/Edge dialog to specify the rising edge of bit 7
(and all other bits as "don't cares").
26
Chapter 1: Getting Started
Example: Timing measurement on counter board
To run the measurement
1. Select the Run Single button.
To display the captured data
1. From the Window menu, select your logic analyzer and choose the
Waveform command.
See Also
“For More Information...” on page 24
27
Chapter 1: Getting Started
Example: State measurement on counter board
Example: State measurement on counter board
This example uses the demo counter board that is supplied with the
Making Basic Measurements kit as the device under test. The kit is
supplied with every logic analysis system, or can be ordered from your
Agilent Technologies Sales Office.
To connect the logic analyzer to the device under test
1. Connect Pod 1 of the logic analyzer to J1 on the demo counter board.
The demo counter board has built-in terminations and header connectors.
To choose the sampling mode
1. In the Sampling tab, choose State Mode.
2. In the Clock Setup, using the Master only mode, specify the rising edge of
the J clock as the sampling clock.
To format labels for the probed signals
1. In the Format tab, select the button under the pod 1.
2. In the Pod threshold dialog, select TTL; then, select the Close button.
3. Select a label button.
4. Choose the Rename command, enter the label name "SCOUNT", and
select the OK button.
5. In the label row, select the button under pod 1.
6. Choose the "........********" standard label assignment to assign the lower 8
bits of pod 1 to the "SCOUNT" label.
To define the trigger condition
1. In the Trigger tab, and in the Trigger Functions subtab, choose the "Find
pattern n times" trigger function, and select the Replace button.
2. In the Trigger Sequence portion of the Trigger tab, enter "15" in the
occurrence count field, and enter enter "FX" in the label value field.
28
Chapter 1: Getting Started
Example: State measurement on counter board
To run the measurement
1. Select the Run Single button.
To display the captured data
1. From the Window menu, select your logic analyzer and choose the Listing
command.
See Also
“For More Information...” on page 24
29
Chapter 1: Getting Started
Example: State measurement on counter board
30
2
Task Guide
•
“Probing the Device Under Test” on page 33
•
“Choosing the Sampling Mode” on page 36
•
“Selecting the Timing Mode (Asynchronous Sampling)” on page 36
31
Chapter 2: Task Guide
•
“Selecting the State Mode (Synchronous Sampling)” on page 43
•
“In Either Timing Mode or State Mode” on page 51
•
“Using 2 GHz Timing Zoom” on page 53
•
“Formatting Labels for Logic Analyzer Probes” on page 56
•
“Setting Up Triggers and Running Measurements” on page 63
•
“Using Trigger Functions” on page 64
•
“Using State Mode Trigger Features” on page 69
•
“Editing the Trigger Sequence” on page 71
•
“Editing Advanced Trigger Functions” on page 76
•
“Saving/Recalling Trigger Setups” on page 83
•
“Running Measurements” on page 84
•
“Displaying Captured Data” on page 87
•
“Using Symbols” on page 94
•
“Printing/Exporting Captured Data” on page 103
•
“Cross-Triggering” on page 105
•
“Solving Logic Analysis Problems” on page 107
•
“Saving and Loading Logic Analyzer Configurations” on page 109
See Also
Measurement Examples (see the Measurement Examples help volume)
32
Chapter 2: Task Guide
Probing the Device Under Test
Probing the Device Under Test
The figures below shows a variety of simple probing connections. The
specific probe type, number of probes, and location on the device
under test circuit depends on your particular measurement.
For equivalent circuit diagrams and pinouts, see the description of the
probe type in the Logic Analysis System and Measurement Modules
Installation Guide. If you have misplaced the Logic Analysis System
and Measurement Modules Installation Guide, you can download
the latest version from the world-wide web at:
http://www.cos.agilent.com/manuals/logic_analyzers.html#la_16700b
Probe Lead-to-Board
Connection
The standard lead set plugs directly into any .1-inch grid with 0.026 to
0.033-inch diameter round pins or 0.025-inch square pins. All probe
tips work with the Agilent Technologies 5059-4356 surface mount
grabbers and the Agilent Technologies 5959-0288 through-hole
grabbers.
33
Chapter 2: Task Guide
Probing the Device Under Test
Adapter-to-Board
Connection
Both the 01650-63203 and the E5346A adapters include termination
for the logic analyzer. The 01650-63203 termination adapter plugs into
a 2 x 10 pin header with 0.1 inch spacing. The E5346A high-density
adapter connects to an AMP "Mictor 38" connector. If possible, use
support shrouds around the Mictor connector to relieve strain and
improve connections.
Direct Pod-to-Board
Connection
If you provide proper termination as part of the device under test
board, you can plug the pod directly into the ©3M 2520-series, or
similar alternative connector. Suggested termination is shown in the
Logic Analysis System and Measurement Modules Installation
Guide.
Also use this termination with the Agilent Technologies E5351A highdensity, non-terminated adapter.
Pod-to-Analysis
Probe Connection
Analysis probes (formerly called preprocessors) are microprocessorspecific interfaces that make it easier to probe buses. Generally,
analysis probes consist of a circuit board that attaches to the
microprocessor (possibly through an adapter) and a configuration file.
The configuration file sets up the logic analyzer's clocks and labels
34
Chapter 2: Task Guide
Probing the Device Under Test
correctly, and may include an inverse assembler. The circuit board
provides access to logical groups of pins through headers designed to
connect directly to the logic analyzer.
The easiest way to set up a measurement with an analysis probe is the
Setup Assistant (see the Setup Assistant help volume). The Setup
Assistant asks you questions about your measurement and then shows
you just the information you need to set up the probe correctly. It also
loads the proper configuration files.
See Also
http://www.agilent.com/find/Laaccessories/ for more information on
Agilent logic analyzer probing accessories.
35
Chapter 2: Task Guide
Choosing the Sampling Mode
Choosing the Sampling Mode
There are two logic analyzer sampling modes to choose from: timing
mode and state mode.
In timing mode, the logic analyzer samples asynchronously, based on
an internal sampling clock signal.
In state mode, the logic analyzer samples synchronously, based on a
sampling clock signal (or signals) from the device under test. Typically,
the signal used for sampling in state mode is a state machine or
microprocessor clock signal.
•
“Selecting the Timing Mode (Asynchronous Sampling)” on page 36
•
“Selecting the State Mode (Synchronous Sampling)” on page 43
•
“In Either Timing Mode or State Mode” on page 51
•
“Using 2 GHz Timing Zoom” on page 53
Selecting the Timing Mode (Asynchronous
Sampling)
In timing mode, the logic analyzer samples asynchronously, based on
an internal sampling clock signal.
•
“To select the timing mode” on page 40
•
“To select the full/half channel configuration” on page 41
•
“To select transitional timing or store qualified” on page 37
•
“To specify the sample period” on page 42
36
Chapter 2: Task Guide
To select transitional timing or store qualified
To select transitional timing or store qualified
1. In the Sampling tab with Timing Mode selected, select the Transitional
Timing with Store Qualification configuration.
Transitional Timing
In Conventional Timing Acquisition mode, the analyzer stores
measurement data at each sampling interval. In Transitional Timing
Acquisition mode, the timing analyzer samples data at regular intervals,
but only stores data when there is a threshold level transition. Each
time a level transition occurs on any of the bits, data on all channels is
stored. A time tag is stored with each stored data sample so the
measurement can be reconstructed and displayed later.
NOTE:
Transitional timing or store qualified timing, requires time tags to recreate the
data. Time tags are either stored in the memory resources of an unused pod
pair, or they are interleaved with the data in memory. If tags are interleaved
with the data, available memory depth is reduced by half.
Store Qualified Timing
Store qualified timing allows you to specify what data is stored during
the course of an acquisition. The level of data qualification can be
simple (Store Anything or Store Nothing), or more complex (Custom).
For information on setting up store qualification, refer to To specify
default storing (see page 70)
See Also
“More on Storing Transitions” on page 38
“Transitional Timing Considerations” on page 39
“Default Storing Subtab” on page 154
37
Chapter 2: Task Guide
To select transitional timing or store qualified
More on Store Qualification in Transitional
Timing
When Transitions is selected on the Default Storing subtab, the
default store qualification is setup to store data on all channels if an
edge/transition occurs on any one channel. Only active channels
(channels assigned to labels) are used. No further user action is
required.
If certain channels have a high occurrence of transitions that are
meaningless in the context of the measurement, they can be ignored
with the following procedure.
1. Select Select Labels.
2. From the Transitions Label Select dialog that appears, highlight the
desired label from the Available Labels list, then select the right-arrow to
move the selection to the Ignore Edges On list.
3. Repeat as needed for additional labels.
4. Select OK to save the selection and close the dialog. Selecting Cancel, will
undo any changes and close the dialog.
Unlike the Custom default storing mode, when Transitions is
selected, no other qualifier events like patterns, ranges, etc., are
available. Also, storing cannot be enabled or disabled in the sequence
branch action lists.
NOTE:
If you have a bit that is shared across multiple labels, all labels containing that
bit must be on the Ignore list before transitions on that bit will no longer cause
a sample to be stored.
More on Storing Transitions
Minimum Transitions Stored
Normally, transitions have not occurred at each sample period. This is
illustrated below with time-tags 2, 5, 7, and 14. When transitions
38
Chapter 2: Task Guide
To select transitional timing or store qualified
happen at this rate, two samples are stored (four at the fastest rate of
2.5 ns) for every transition. Therefore, with 2 K samples of memory, 1
K of transitions are stored. You must subtract one, which is necessary
for a starting point, for a minimum of 1023 stored transitions.
Maximum Transitions Stored
If transitions occur at a fast rate, such that there is a transition at each
sample point, only one sample is stored for each transition as shown by
time tags 17 through 21 below. If this continues for the entire trace, the
number of transitions stored is 2 K samples. Again, you must subtract
the starting point sample, which then yields a maximum of 2047 stored
transitions.
In most cases a transitional timing trace is stored by a mixture of the
minimum and maximum cases. Therefore, in this example the actual
number of transitions stored will be between 1023 and 2047.
Transitional Timing Considerations
Data Storage
When an edge is detected, two samples (four when sampling at 2.5 ns)
are stored across all channels assigned to the timing analyzer. The need
of two samples is to avoid loss of data if a second edge were to occur to
soon after the first edge for the edge detectors to reset.
39
Chapter 2: Task Guide
To select transitional timing or store qualified
Sequence level branching
In transitional timing, only 2 branches are available per sequence level.
Global counters
In transitional timing, only one global counter is available.
Storing Time Tags
Transitional timing requires time tags to recreate the data. Time tags
are either stored in the memory resources of an unused pod pair, or
they are interleaved with the data in memory. If tags are interleaved in
data, available memory depth is reduced by half.
Increasing Duration of Storage
Using the Transitional Label Select dialog to specify selected labels to
ignore can increase usable memory depth and acquisition time by
ignoring transitions on signals like clock or strobe that add little useful
information to the measurement when no other signals are
transitioning.
Invalid Data
The analyzer only looks for transitions on data lines on labels that are
turned on. Data lines on labels that are turned off store data, but only
when one of the lines that is turned on transitions. If the data line on a
label is turned on after a run, or the data line is assigned to a new label,
you would see data, but it is unlikely that every transition that occurred
was captured.
Trigger Position
In transitional timing, no data prestore (samples acquired before
trigger) is required. Therefore, much like state mode, the trigger
position (start/center/end) will indicate the percentage of memory
filled with samples after the trigger. The number of samples acquired/
displayed before trigger will vary between measurements.
To select the timing mode
1. Open the logic analyzer Setup window.
40
Chapter 2: Task Guide
To select transitional timing or store qualified
2. Select the Sampling tab.
3. Choose the Timing Mode option.
You can also select the timing sampling mode in the “Pod Assignment
Dialog” on page 128.
To select the full/half channel configuration
1. In the Sampling tab, with Timing Mode selected, select the timing analyzer
configuration. You can choose between:
•
512K Sample Full Channel 333 MHz
In this configuration, the total memory depth is 512K samples per
channel, with data being sampled and stored as often as every 3.0 ns.
You can set the sample rate to go slower with the Sample Period
control.
NOTE:
When the Sample Period is 3.0 ns, data is acquired at two times the trigger
sequencer rate. This means that data must be present for at least two samples
before the trigger sequencer can reliably detect it. The trigger sequencer
could miss data present for less than two sample periods.
The trigger sequencer treats the data as a group of two samples for each
sequencer clock. This means that the trigger point indication could be off by
one sample.
Although the trigger sequencer cannot detect all data, the analyzer will
correctly capture all data present for at least one sample period.
•
1M Sample Half Channel 667 MHz
In this configuration, only one pod of each pod pair is available.
Channels assigned to unavailable pods are ignored. You can specify
which pod to use by toggling the Pod field in Format.
The total memory depth is 1M samples per channel. Data is sampled
and stored every 1.5 ns; this rate cannot be changed.
41
Chapter 2: Task Guide
To select transitional timing or store qualified
NOTE:
When the Sample Period is 1.5 ns, data is acquired at four times the trigger
sequencer rate. This, along with other half-channel mode characteristics,
means that data must be present for at least five samples before the trigger
sequencer can reliably detect it. The trigger sequencer cannot detect data
present for less than two sample periods, and could miss data present for less
than five sample periods.
The trigger sequencer treats the data as a group of four samples for each
sequencer clock. This means that the trigger point indication could be off by
up to three samples.
Although the trigger sequencer cannot detect all data, the analyzer will
correctly capture all data present for at least one sample period.
2. If you chose the 667 MHz Half Channel 1M Sample configuration, select
the Format tab and choose which pod of the pod pair will be used to
sample data.
See Also
“To specify the sample period” on page 42
To specify the sample period
When the logic analyzer is in timing (asynchronous sampling) mode,
the Sample Period setting specifies how often the logic analyzer
samples the signals from the device under test.
1. In the Sampling tab, with Timing Mode selected, enter the desired time
between logic analyzer samples.
To capture signal level changes reliably, the sample period should be less
than half (many engineers prefer one-fourth) of the period of the fastest
signal you want to measure.
The sample rate is the inverse of the sample period.
NOTE:
In conventional timing mode the sample rate is fixed at 1.25 ns.
42
Chapter 2: Task Guide
To select transitional timing or store qualified
Selecting the State Mode (Synchronous
Sampling)
In state mode, the logic analyzer samples synchronously, based on a
sampling clock signal (or signals) from the device under test. Typically,
the signal used for sampling in state mode is a state machine or
microprocessor clock signal.
•
“To select the state mode” on page 43
•
“To change the sampling clock mode” on page 44
•
“To set up the sampling clock” on page 45
State Mode Sampling In order for a state mode logic analyzer to accurately capture data from
Position
a device under test, the logic analyzer's setup/hold time (window) must
fit within the device under test's data valid window.
Because the location of the data valid window relative to the bus clock
is different for different types of buses, the logic analyzer lets you
adjust the sampling position in order to accurately capture data on
high-speed buses (see “Understanding State Mode Sampling Positions”
on page 206).
When the device under test's data valid window is less than 2.5 ns
(roughly, for clock speeds >= 200 MHz), it's easiest to use eye finder
to locate the stable and transitioning regions of signals and to
automatically adjust sampling positions.
•
“To automatically adjust sampling positions” on page 46
When the device under test's data valid window is greater than 2.5 ns
(roughly, for clock speeds < 200 MHz), it's easiest to adjust the
sampling position manually, without using the logic analyzer to locate
the stable and transitioning regions of signals.
•
“To manually adjust sampling positions” on page 49
To select the state mode
1. Open the logic analyzer Setup window.
43
Chapter 2: Task Guide
To select transitional timing or store qualified
2. Select the Sampling tab.
3. Choose the State Mode option.
You can also select the state sampling mode in the “Pod Assignment
Dialog” on page 128.
To change the sampling clock mode
Normally, in the Master only sampling clock mode, there is one
sampling clock signal. When a clock edge occurs, data is captured and
saved into one sample of logic analyzer memory.
Two additional sampling clock modes let you capture data differently:
•
In the Master/Slave mode, you can save data captured on different clock
edges into the same sample of logic analyzer memory.
When the slave clock occurs, data captured on the pods that use the slave
clock is saved in a slave latch. Then, when the master clock occurs, data
captured on the pods that use the master clock, as well as the slave latch
data, are saved into logic analyzer memory.
•
44
In the Demultiplex mode, you can demultiplex data being probed by one
pod into the logic analyzer memory that is normally used for two pods.
Chapter 2: Task Guide
To select transitional timing or store qualified
When the slave clock occurs, data captured on the pod is saved into the
slave latch for the other pod in the pod pair. Then, when the master clock
occurs, data captured on the pod, as well as the slave latch data, are saved
in logic analyzer memory.
To set up the master sampling clock mode
1. In the Sampling tab, with State Mode selected, select the Master only
mode in the Clock Setup area.
To set up the master/slave sampling clock mode
1. In the Sampling tab, with State Mode selected, select the Master/Slave
mode in the Clock Setup area.
2. In the Format tab, select Slave Clock for each pod that should use the
slave clock, and select Master Clk for each pod that should use the master
clock.
To set up the demultiplex sampling clock mode
1. In the Sampling tab, with State Mode selected, select the Demultiplex
mode in the Clock Setup area.
2. In the Format tab, select Demultiplex for the pod pair that should use this
mode.
To set up the sampling clock
1. In the Sampling tab, with the State Mode selected, make sure the
Advanced Clocking box is unchecked.
2. For each clock input signal that will be used:
45
Chapter 2: Task Guide
To select transitional timing or store qualified
a. Select the pod's Master or Slave button (under the activity indicator).
b. If the signal edge will specify when to sample, choose Rising Edge,
Falling Edge, or Both Edges.
c. If the signal level will enable the sampling clock, choose Qualifier High or Qualifier - Low.
3. Make sure all unused clock inputs are turned Off.
To set up using advanced clocking
1. In the Sampling tab, with the State Mode selected, select the Advanced
Clocking check box.
2. Select the Master Clock button. In the Master clock dialog, select the
appropriate options for setting up the master clock.
3. If you have chosen the Master/Slave or Demultiplex clock mode, select the
Slave Clock button. In the Slave clock dialog, select the appropriate
options for setting up the slave clock.
See Also
“To change the sampling clock mode” on page 44
To automatically adjust sampling positions
When adjusting the state mode sampling position with eye finder, the
logic analyzer looks at signals from the device under test, figures out
the location of the data valid window in relation to the sampling clock,
and automatically sets the sampling position.
Because eye finder automatically runs on individual channels, it can
correct for the small delay effects caused by probe cables and circuit
board traces. This makes the logic analyzer's setup/hold window
smaller and lets you accurately capture data at higher clock speeds.
Eye finder requires:
•
At least 500 transitions on each signal during its run. (You can use the
advanced eye finder settings to cause longer or shorter runs.)
•
All devices which can drive each signal should contribute to the stimulus.
•
All device under test operating modes relevant to the eventual logic
analysis measurement should contribute to the stimulus as well.
46
Chapter 2: Task Guide
To select transitional timing or store qualified
NOTE:
Eye finder measurements and normal logic analyzer measurements cannot
run simultaneously.
To run eye finder
1. Probe the device under test by connecting the logic analyzer channels.
2. Select the state (synchronous sampling) mode (see “To select the state
mode” on page 43).
3. Format labels for those logic analyzer channels.
4. Make sure that the device under test and the logic analyzer have warmed
up to their normal operating temperatures.
5. In the Format tab, select the Setup/Hold button.
6. In the Sampling Positions dialog, select the Eye Finder option.
7. In the Eye Finder Setup tab, select the Use signals from Device Under
Test option.
The Use demo data (no probes required) option is for demonstration
purposes only.
8. Choose the labels that you wish to run eye finder on.
You may want to run eye finder on channel subsets, for example, when
certain bus signals transition in one operating mode (of the device under
test) and other bus signals transition in a different operating mode.
9. Select the Run Eye Finder button.
For more information on run messages, see “Eye Finder Run Messages” on
page 134.
When eye finder finds more than one stable region on a channel, it uses
the current sampling position as a hint about which stable region it should
suggest a position for.
If eye finder picks the wrong stable region, you can expand the label and
drag the blue Sampling Position line into the correct stable region. The
suggested sampling position for that region will be shown (see “How
Selected/Suggested Positions Behave” on page 133).
10. If you have moved the sampling position and wish to return to the
47
Chapter 2: Task Guide
To select transitional timing or store qualified
suggested positions, go to the Eye Finder Results tab, select a label button
or the Results menu, and choose the "set to suggested" command.
For more information on informational messages in the Eye Finder Results
tab, see “Eye Finder Info Messages” on page 137.
Eye finder finds optimal sampling positions for the actual specific
conditions -- amplitude, offset, slew rates, and ambient temperature.
Therefore, you will get the best results by running eye finder under
the same conditions that will be present when logic analysis
measurements are made.
To run eye finder repetitively
1. Select the Repetitive Run option in the Eye Finder Setup tab.
2. Select the Run Eye Finder (r) button.
In the Eye Finder Results tab, you can see how the stable and transitioning
areas vary over time.
3. Select the Stop Eye Finder button.
To view eye finder data as a bus composite
When you want a compressed, high-level view of the eye finder data:
1. In the Eye Finder Results tab, select the label button and choose the View
as Bus Composite command.
Average sampling positions as well as stable and transitioning areas are
displayed for the whole label. This is the default. Stable areas show
positions where every channel in the label is stable.
To view eye finder data as a stack of channels
When you want more resolution in your view of the eye finder data:
1. In the Eye Finder Results tab, select the label button and choose the View
48
Chapter 2: Task Guide
To select transitional timing or store qualified
as Stack of Channels command.
Individual sampling positions and stable and transitioning areas for all the
channels in a label are shown.
To save/load eye finder data
While the eye finder sampling positions are saved with the logic
analyzer configuration, eye finder measurement data is not; therefore,
eye finder data must be saved and loaded separately.
1. In the File Info tab, select the Save As... or Load... buttons.
You can also choose the Save Eye Finder or Load Eye Finder command
from the File menu.
2. In the file browser dialog, name the file to be saved or select the file to be
loaded.
For more information on save/load messages, see “Eye Finder Load/Save
Messages” on page 139.
See Also
“Understanding State Mode Sampling Positions” on page 206
“Eye Finder Advanced Settings Dialog” on page 131
“To manually adjust sampling positions” on page 49
To manually adjust sampling positions
To use the Eye Finder option
Although the Eye Finder option was intended for automatically
adjusting state mode sampling positions, you can also use it to
manually adjust sampling positions. You don't have to Run Eye Finder
to locate stable and transitioning regions on signals, just go directly to
the Eye Finder Results tab, and drag the sampling positions to the
49
Chapter 2: Task Guide
To select transitional timing or store qualified
proper locations.
1. Select the state (synchronous sampling) mode (see “To select the state
mode” on page 43).
2. In the Format tab, select the Setup/Hold button.
3. In the Sampling Positions dialog, select the Eye Finder option.
4. In the Eye Finder Results tab, drag the sampling positions to the proper
locations.
You can select bus labels to expand or collapse the channels in the label.
When using the Eye Finder option to manually adjust state mode
sampling positions, the sampling positions are saved with the logic
analyzer configuration (see “Saving and Loading Logic Analyzer
Configurations” on page 109).
To use the Manual Setup/Hold option
When adjusting the state mode sampling position with the Manual
Setup/Hold option, you adjust the logic analyzer's setup/hold window
relative to the sampling clock signal from the device under test. The
setup time is the front edge of the setup/hold window relative to the
sampling clock, and the hold time is the back edge of the setup/hold
window relative to the sampling clock.
1. Select the state (synchronous sampling) mode (see “To select the state
mode” on page 43).
2. In the Format tab, select the Setup/Hold button.
3. In the Sampling Positions dialog, select the Manual Setup/Hold option.
4. For each label, enter setup/hold values. The values are adjustable in 100 ps
increments, with a fixed window of 2.5 ns.
5. If you need to adjust bits individually:
a. Select a label containing the bit.
If a bit is used in more than one label, you will change its setup and
hold value in all labels.
b. Select the Individual bits option.
50
Chapter 2: Task Guide
To select transitional timing or store qualified
c. Enter the bit number you want to change.
d. Enter the setup/hold value.
6. Close the Sampling Positions dialog.
The Manual Setup/Hold sampling positions are saved and loaded along
with the logic analyzer configuration file.
Example
Suppose you're probing a bus in the device under test whose data valid
window is 3 ns. Suppose also that the bus clock edge occurs 1 ns into
the data valid window. To place the logic analyzer's setup/hold window
within the data valid window, you could specify a setup value of 800 ps
(and hold value of 1.7 ns).
(The actual sampling position is in the middle of the setup/hold
window.)
See Also
“Understanding State Mode Sampling Positions” on page 206
“To automatically adjust sampling positions” on page 46
In Either Timing Mode or State Mode
•
“To specify the trigger position” on page 52
•
“To set acquisition memory depth” on page 52
51
Chapter 2: Task Guide
To select transitional timing or store qualified
•
“To name an analyzer” on page 52
•
“To turn an analyzer off or on” on page 53
To specify the trigger position
1. In the Sampling tab (or in the Settings subtab of the Trigger tab), select
the trigger position.
Specify whether you want to look at data after the trigger (Start), before
and after the trigger (Center), before the trigger (End), or use a
percentage of the logic analyzer's memory for data after the trigger (User
Defined).
In Conventional Timing Mode, when a Run is started, the analyzer will
not look for a trigger until the specified percentage of pretrigger data
has been stored. After a trigger has been detected, the specified
percentage of posttrigger data is stored before the analyzer halts.
In State and Transitional Store Qualified modes, when a Run is started,
the analyzer immediately looks for the trigger condition. In other
words, the trigger position setting specifies the maximum amount of
data that should be stored before the trigger.
To set acquisition memory depth
If you need less data and want measurements to run faster, you can
limit the number of samples that are stored in logic analyzer acquisition
memory.
1. In the Sampling tab (or in the Settings subtab of the Trigger tab), select
the acquisition depth.
The number of samples that can be chosen for the Acquisition Depth are
approximations. The combination of count tags, pod assignments, and
configuration modes affect what choices are available.
To name an analyzer
You can give more descriptive names to a logic analyzer.
1. In the Sampling tab, select the Analyzer Name field.
2. Enter the new name.
52
Chapter 2: Task Guide
To select transitional timing or store qualified
The name now appears below the instrument tool icon in the workspace.
You can also name analyzers in the “Pod Assignment Dialog” on
page 128.
To turn an analyzer off or on
You may want to turn an analyzer off if you don't want it to be included
in further measurements.
To turn an analyzer off
1. In the Sampling tab, select the On box that is checked.
2. In the Analyzer Shutdown Options dialog, choose either:
•
Soft -- This will leave the logic analyzer window but turn off most
options.
•
Hard -- This will remove the logic analyzer and its display tools from
the Workspace.
You can also turn an analyzer off in the “Pod Assignment Dialog” on
page 128.
To turn an analyzer back on
1. If you used the Soft option when turning the logic analyzer off, you can
turn it on again by selecting the Off check box.
2. If you used the Hard option when turning the logic analyzer off, you can
turn it on again by selecting the Setup button in the System window or by
dragging the analyzer's instrument tool icon to the workspace in the
Workspace window.
Using 2 GHz Timing Zoom
Timing Zoom collects additional high-speed timing data around the
trigger of the logic analyzer. It uses a 16K-sample, 2 GHz timing
analyzer to sample data as closely as every 500 ps on all channels.
The Timing Zoom settings are accessed through the Timing Zoom
53
Chapter 2: Task Guide
To select transitional timing or store qualified
button under Sampling.
•
“To turn Timing Zoom on or off” on page 54
•
“To set the Timing Zoom trigger position” on page 54
•
“To specify which analyzer has Timing Zoom” on page 55
•
“To set the Timing Zoom sample period” on page 55
To turn Timing Zoom on or off
If you are not interested in the Timing Zoom data for a measurement,
you can turn off Timing Zoom and improve logic analyzer performance.
1. In the Sampling tab, select the Timing Zoom button.
2. In the Timing Zoom controls dialog, select the On/Off checkbox.
To set the Timing Zoom trigger position
1. In the Sampling tab, select the Timing Zoom button.
2. In the Timing Zoom controls dialog, select the trigger position.
Specify whether you want to look at Timing Zoom data after the trigger
(Start), before and after the trigger (Center), before the trigger (End), or
use a percentage of the logic analyzer's memory for data after the trigger
(User Defined).
NOTE:
When in 333 MHz State mode (16717,18,19), the 400 MHz State mode
(16750,51,52), or the 600 MHz State mode (16753, 54, 55, 56), the start of
Timing Zoom data may occur after the actual trigger point. The reason for this
data mis-alignment is due to how the trigger sequencer functions when in this
mode.
The analyzer sequencer works on Pairs of samples. It will not evaluate the
first sample of the pair until the second sample has entered the sequencer. If
for example the trigger point is determined to be on the first sample, the
analyzer displays the Timing Zoom data relative to the evaluation of the
second sample. What ever time difference is seen between the two samples
(of the pair) is reflected in the data display between the trigger point and the
start of the Timing Zoom data.
54
Chapter 2: Task Guide
To select transitional timing or store qualified
This time difference can be noticeable if your measurement is using bursted
clocks and the first sample (actual trigger point) is clocked on the last clock
signal of a burst, and the second sample (of the pair) is clocked with the first
clock of the next burst. The time difference between the clock bursts is
reflected as a mis-alignment between the trigger point and the start of the
Timing Zoom data.
The best thing to do to help mitigate this situation is to set the Timing Zoom to
"Trigger Position End" to capture as much data near the first sample as
possible.
To specify which analyzer has Timing Zoom
NOTE:
If you have both analyzers of the module turned on, you need to specify which
analyzer you want to use Timing Zoom with.
1. In the Sampling tab, select the Timing Zoom button.
2. In the Timing Zoom controls dialog, select the analyzer that Timing Zoom
will work with.
To set the Timing Zoom sample period
If you want to see more or less sampling resolution around the trigger,
you can adjust the Timing Zoom sample period.
1. In the Sampling tab, select the Timing Zoom button.
2. In the Timing Zoom controls dialog, select the sample period.
55
Chapter 2: Task Guide
Formatting Labels for Logic Analyzer Probes
Formatting Labels for Logic Analyzer Probes
The Format tab is mainly for assigning bus and signal names (from the
device under test) to logic analyzer channels. These names are called
labels. Labels are used when setting up triggers and displaying
captured data.
The Format tab also lets you do things like assign pod pairs to one or
two logic analyzers, specify the logic analyzer threshold voltage,
change the label polarity, reorder bits in a label, and turn labels off or
on.
The Format tab has activity indicators that show signal levels.
•
“To assign pods to one or two analyzers” on page 56
•
“To set pod threshold voltages” on page 57
•
“To assign probe channels to labels” on page 58
•
“To change the label polarity” on page 60
•
“To reorder bits in a label” on page 60
•
“To turn labels off or on” on page 61
To assign pods to one or two analyzers
A logic analyzer's pod pairs can be assigned to one or two separate
logic analyzers or they can be left unassigned.
1. In the Format tab, select the Pod Assignment button.
2. In the Pod Assignment dialog, drag a pod pair to the appropriate logic
analyzer.
3. Select the Close button.
When all pods are assigned and state or time counts are saved with the
captured data, logic analyzer acquisition memory is halved to 256K
samples.
56
Chapter 2: Task Guide
Formatting Labels for Logic Analyzer Probes
When using a multi-card logic analyzer:
•
When both analyzers are turned on, pods 1/2 and 3/4 of the master card
cannot be assigned to the same analyzer.
•
Each pod pair has two clock channels, but only the clock channels of
pods on the master card can be used in the analyzer's clocking setup.
(The master card's pods needn't be assigned in order to use their clock
channels).
To turn on an analyzer that is off
1. Select Off and choose State or Timing.
(Only one analyzer at a time can be set to Timing.)
A second analyzer window appears after a pause for setup.
You can also turn on an analyzer that is off by opening the Workspace
window and dragging the instrument tool icon onto the workspace.
See Also
“Pod Assignment Dialog” on page 128
To set pod threshold voltages
The threshold voltage is the voltage level that a signal must cross
before the logic analyzer recognizes a change in logic levels. In addition
to a user defined choice, you can also select a predefined level from the
standard choices.
Standard
1. In the Format tab, select the threshold button located just below the pod
name.
2. In the Pod threshold dialog, choose one of the standard threshold options:
•
TTL -- The threshold level is +1.50 volts.
•
LVTTL -- The threshold level is +1.40 volts.
•
SSTL2 -- The threshold level is +1.50 volts.
•
SSTL3 -- The threshold level is +1.25 volts.
57
Chapter 2: Task Guide
Formatting Labels for Logic Analyzer Probes
•
LVCMOS 1.5v -- The threshold level is +0.75 volts.
•
LVCMOS 1.8v -- The threshold level is +0.90 volts.
•
LVCMOS 2.5v -- The threshold level is +1.25 volts.
•
LVCMOS 3.3v -- The threshold level is +1.65 volts.
•
CMOS 5.0v -- The threshold level is +2.50 volts.
•
ECL -- The threshold level is -1.3 volts.
•
LVPECL -- The threshold level is 2.00 volts.
3. If you don't want the change to apply to all pods, deselect the checked box
next to Apply settings to all pods.
4. Select the Close button.
User Defined
When User Defined is selected, the threshold level is selectable from 6.0 volts to +6.0 volts.
NOTE:
The logic analyzer requires a minimum voltage swing of 500 mV at the probe
tip to recognize changes in logic levels.
NOTE:
The threshold voltage specified also applies to the pod's clock input.
To assign probe channels to labels
The logic analyzer lets you assign names (labels) to logic analyzer
channels so that it's easier to set up triggers and interpret the captured
data when displayed.
Typically, you give labels the names of the buses and signals in the
device under test that are are being probed.
1. In the Format tab, select a label button, and either:
•
58
Choose the Rename command, enter the label name, and select the OK
button.
Chapter 2: Task Guide
Formatting Labels for Logic Analyzer Probes
•
Or, choose the Insert before or Insert after command, enter the label
name, and select the OK button.
2. In the label row, select the button of the pod that contains the channels
you want to assign.
3. Either choose one of the standard label assignments or choose
Individual.
( * ) (asterisk) indicates an assigned bit.
( . ) (period) indicates an unassigned bit.
( R ) indicates an assigned bit in a reordered label.
If you chose Individual:
a. In the "label - pod" dialog, select the channels you want to assign/
unassign.
b. Select the OK button.
A maximum of 32 channels can be assigned to a label.
In the Format tab, least significant pod channels (bit 0) are on the right
and most significant pod channels (bit 15) are on the left. (The bit
numbers are shown just below the activity indicators.)
Labels can contain bits that are not consecutive; however, bits are
always numbered consecutively within a label.
A label can include data and clock channels from more than one pod,
but this places restrictions on the complexity of the trigger later.
To delete labels
1. Select the label name that you want to delete.
2. Choose Delete.
If only one label is defined, it cannot be deleted.
When you delete labels, their bit assignments are not saved. However,
you can make a label inactive and save its bit assignments by turning
the label off.
See Also
“To reorder bits in a label” on page 60
59
Chapter 2: Task Guide
Formatting Labels for Logic Analyzer Probes
“To turn labels off or on” on page 61
“To change the label polarity” on page 60
To change the label polarity
While negative logic is rare in circuits (the main exception at this time
is RAMBUS), you can change the label polarity if the device under test
uses negative logic.
1. In the Format tab, select the polarity button (next to the label button) to
toggle between positive (+) and negative (-) polarity.
Positive polarity means that a high voltage is a logic "1".
Negative polarity means that a high voltage is a logic "0".
Changing the label polarity will have the following effects:
•
"1" and "0" values flip in the trigger condition.
•
Waveforms and bus values (where shown) invert in the Waveform display
tool.
•
"1" and "0" values flip in the Listing display tool.
Changing the label polarity does not affect:
•
Edge definitions for clock setup and edge terms.
•
Symbol definitions for the logic analyzer.
•
Activity indicators.
To reorder bits in a label
In cases where buses in the device under test haven't been probed with
consecutive logic analyzer channels, you can reorder the bits in a label.
1. In the Format tab, select the label button whose bits you want to reorder.
2. Choose Reorder bits.
60
Chapter 2: Task Guide
Formatting Labels for Logic Analyzer Probes
3. In the Change Bit Order dialog:
•
To reorder the bits individually, enter the bit that the probe channel
should be mapped to.
•
To swap the high and low order bytes or words, select the button Big
Endian to Little Endian at the bottom of the dialog.
•
To return to sequentially ordered bits, select the button Default Order
at the bottom of the dialog.
4. Select the OK button.
The label now shows an "R" to indicate that the assigned bit has been
reordered.
NOTE:
Labels with reordered bits cannot be used as range terms or <, <=,>, >= in
triggers.
To turn labels off or on
When you temporarily want to remove a label and its data, you can turn
off the label. The label name and its bit assignments are preserved.
To turn a label off
1. In the Format tab, select the label button that you want to turn off.
2. Choose Label [ON] to toggle it off.
At least one label must remain on.
To turn a label on
1. In the Format tab, select the label button that you want to turn on.
2. Choose Label [OFF] to toggle it on.
To display a label that was off
1. Turn on the label.
2. At the bottom of the window, select the Apply button.
61
Chapter 2: Task Guide
Formatting Labels for Logic Analyzer Probes
The label's data appears in the display windows.
62
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
Setting Up Triggers and Running Measurements
The following information is a generic discussion about triggering in
logic analyzers. Depending on the logic analyzer type, and the state or
timing mode being used, some functionality may not be available.
In General...
•
“Using Trigger Functions” on page 64
•
“Using State Mode Trigger Features” on page 69
•
“Editing the Trigger Sequence” on page 71
•
“Editing Advanced Trigger Functions” on page 76
•
“Saving/Recalling Trigger Setups” on page 83
•
“Running Measurements” on page 84
Use trigger functions for basic measurements.
For more complicated measurements, where no trigger function exists,
start with a trigger function that's similar to the measurement you want
to make. Then, break down the trigger function and edit the advanced
trigger sequence levels.
Timing Analyzer
Triggers
Everything that looks like a button in the trigger definition gives you a
way to modify the trigger setup.
For example, to look for a edge instead of a pattern, select the button
that equates a label with a pattern and choose an edge comparison
instead.
State Analyzer
Triggers
For every state analysis sample, a logic analyzer needs to know two
things:
1. Should some action (like a trigger) be taken as a result of this sample?
2. What should be done with this sample? That is, should it be stored in logic
analyzer memory or should it be discarded? (This question doesn't need to
be asked when using a timing analyzer because all samples are stored.)
State analysis trigger definitions are made simpler with a default
storage qualifier. This makes it possible to ignore, at all trigger
63
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
sequence levels, the question about what to do with the captured data
samples.
Of course, sometimes it's useful to specify storage qualifiers at certain
levels in the trigger sequence. For this, you can insert storage actions
in the trigger sequence before trigger or goto actions. Storage actions
in the trigger sequence override the default storage qualifier for the
samples that cause the trigger or goto actions to occur. Storage actions
can also be used to turn on or off the default storing.
Using Trigger Functions
Many common measurement setups are provided with the logic
analyzer. These setups are called trigger functions, and you can use
them for quick measurement setup.
For more complicated measurements, where no trigger function exists,
start with a trigger function that's similar to the measurement you want
to make. Then, break down the trigger function and edit the advanced
trigger specification.
•
“To select a trigger function” on page 64
•
“To specify a label pattern event” on page 65
•
“To specify a label edge event” on page 65
•
“To break down a trigger function” on page 66
•
“To create a trigger function library” on page 67
To select a trigger function
1. In the Trigger tab's Trigger Functions subtab, select the appropriate
trigger function.
A picture describing the trigger function is shown.
2. Select the Replace button (or Insert before or Insert after button) to
move it to the Trigger Sequence below.
3. In the Trigger Sequence, select and/or enter the appropriate labels, values,
64
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
and options.
To specify a label pattern event
Label pattern events let you specify patterns or ranges on a bus.
1. Select the label name button and choose the label that you want to look for
a pattern on.
You can also insert other label events if you want to look for multiple
patterns on multiple labels. Once another label event is inserted, you can
choose And if both label events must occur in the same sample or Or if
only one of the label events must occur.
2. Select the operator button and choose the appropriate operator.
The In range and Not in range operators consider the values you enter to
be inside the range. Ranges cannot be set on labels whose bits have been
reordered.
3. Select the number base button, and choose the number base that you
want.
If the number base is changed in one window, the number base in other
windows may not change accordingly. For example, the number base
assigned to symbols is unique, as is the number base assigned in the
Listing window.
4. Enter the label value.
Xs mean you don't care about the value on the specified bits. Xs are not
allowed in ranges.
If you chose the Symbols or Line #s number base, select the Absolute
XXXX button, and use the Symbol Selector dialog to choose the symbol or
line number value.
See Also
“To specify a label edge event” on page 65
“To enter symbolic label values” on page 98
“Symbols Selector Dialog” on page 159
To specify a label edge event
Label edge events let you specify edges and glitches on a bus. Label
65
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
edge events are only available in certain timing mode trigger functions.
1. Select the label name button and choose the label that you want to look for
a pattern on.
You can also insert other label events if you want to look for multiple
patterns on multiple labels. Once another label event is inserted, you can
choose And if both label events must occur in the same sample or Or if
only one of the label events must occur.
2. Select the edge assignment button.
3. In the Specify Edge/Glitch dialog, select the edges or glitches that you're
looking for on particular logic analyzer channels.
When you select multiple edges or glitches, they are ORed together, and
any one of the edges or glitches in a sample will satisfy the label edge
event. If you want to AND edges or glitches on a label, insert multiple label
edge events and AND them together.
4. Select the OK button.
See Also
“To specify a label pattern event” on page 65
To break down a trigger function
When a trigger function doesn't quite let you set up the trigger you
want, you can break it down and edit the resulting advanced trigger
function.
1. In the Trigger tab, select the number button of the trigger sequence level
whose trigger function you want to break down.
2. Choose Break down function.
Breaking down the trigger function will be permanent (although you can
choose the Undo command from the Edit menu if no other editing has
taken place).
If you only want to look the advanced trigger function, without editing it,
you can expand the trigger function.
3. Select OK in the confirmation dialog.
66
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
To expand a trigger function
1. In the Trigger tab, select the number button of the trigger sequence level
whose trigger function you want to expand.
2. Choose Expand function.
To compress a trigger function
Expanded trigger functions can be compressed back into their original
form.
1. In the Trigger tab, select the number button of the trigger sequence level
whose trigger function you want to compress.
2. Choose Compress function.
See Also
“Editing Advanced Trigger Functions” on page 76 for information on
editing trigger functions that are broken down.
To create a trigger function library
You can create your own libraries of trigger functions that are separate
from logic analyzer configuration files (unlike saved/recalled trigger
setups).
1. In the Trigger tab's Trigger Functions subtab, select the Trigger function
libraries button.
2. In the Trigger function libraries dialog, select the Create button.
3. In the Create User Library dialog, enter the library name and description,
and select OK.
4. In the Edit Trigger Function Library dialog, choose the Add function
button.
5. In the Create User Function dialog, enter the function name and
description, and select the levels from the current trigger sequence that be
the trigger function; then, select OK.
Once you have created a trigger function library with trigger functions,
you can:
•
Load or unload the trigger function library.
67
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
NOTE:
•
Insert and break down trigger functions from the loaded library just like
normal trigger functions.
•
Copy trigger function libraries to other logic analysis systems and load
them into other logic analyzers that have trigger function library capability.
•
Edit the trigger function library, adding or deleting functions, or delete the
library.
If a trigger sequence or configuration file uses a trigger function library that
has been deleted, or a trigger function that has been deleted from a library,
the logic analyzer replaces the missing function with the default trigger
function.
To load/unload trigger function libraries
1. In the Trigger tab's Trigger Functions subtab, select the Trigger function
libraries button.
2. Select the library from the list.
Only libraries created in the same sampling mode are available.
3. Select the Load (or Unload) button.
All of the library's trigger functions are added to (or removed from) the list
of trigger functions.
To copy trigger function libraries between systems
1. Connect your logic analysis system to the network. (see the Agilent
Technologies 16700A/B-Series Logic Analysis System help volume)
2. Using a computer on your network, copy the appropriate files from the /
logic/trigger_functions/ directory to a central location, or
directly to other logic analyzers on the network.
See Also
“To break down a trigger function” on page 66
“Saving/Recalling Trigger Setups” on page 83
68
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
Using State Mode Trigger Features
When the logic analyzer sampling mode is state, you can specify
whether a state or time count is stored with samples and you can set up
the default storing options.
•
“To count states or time” on page 69
•
“To Specify Default Storing” on page 70
•
“To specify whether default storing is initially on or off” on page 71
To count states or time
When the logic analyzer sampling mode is state, you can specify
whether a state or time count is stored with samples.
1. In the Trigger tab's Settings subtab, select the Count option button and
choose either Off, Time, or States.
2. If you chose States:
a. Select the Define button.
b. In the State count qualify dialog, select the Count if or Count if NOT
option.
c. Specify events that identify the states to be counted or not counted.
d. If you would like to specify the evaluation order of the event list, select
Group events. Then, in the Group Events dialog, either select the Add
parens button to group events or select the Remove parens button to
ungroup events. When you're done grouping events, select the OK
button.
When time or state counts are turned on and all pods are assigned,
logic analyzer acquisition memory is halved to 256K samples for the
16716A logic analyzer. Leaving one pod pair unassigned gives you full
memory depth.
See Also
“To assign pods to one or two analyzers” on page 56
69
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
To Specify Default Storing
You can set up default storing so that only the data samples you're
interested in are saved in logic analyzer acquisition memory.
NOTE:
Default storing in both state and timing modes require time tags to
reconstruct the data. These time tags will be stored either on a reserved tag
chip, or interleaved in the data.
1. In the Trigger tab's Default Storing subtab, select the store by default
option button and choose either Anything, Nothing, or Custom.
Anything means all samples are stored. Nothing means that no samples
are stored. Custom lets you specify which samples are stored.
2. If you chose Custom:
a. Select the Store if or Store if NOT option.
b. Specify events that identify the states to be stored or not stored.
c. If you would like to specify the evaluation order of the event list, select
Group events. Then, in the Group Events dialog, either select the Add
parens button to group events or select the Remove parens button to
ungroup events. When you're done grouping events, select the OK
button.
3. If you chose Transitions (Timing mode only).
Transitional timing is a subset of store qualified timing. In transitional
timing, the store qualification is set to look for edges (transitions) on all
active channels. Active channels are those that have been assigned to a
label in the Format tab. Samples are stored for all channels whenever a
transition occurs. No other store qualification events are allowed.
For more information on store qualification in transitional timing mode,
refer to More on Store Qualification in Transitional Timing. (see page 38)
In the trigger sequence, you can override default storing for the
samples that cause actions to occur, or you can turn default storing on
or off, by inserting store actions.
The Agilent Technologies 16716A logic analyzer does not use the
"Branches taken" feature of past logic analyzers. The best way to store
only the states that cause sequence level branches is by setting up
70
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
default storing to Nothing, inserting a Store sample action in each
sequence level, and inserting a Turn off default storing action in the
level that triggers.
To clear default storing changes
1. When the Trigger tab is displayed, select Clear Default Store from the
Clear menu.
See Also
“Storage Qualification” on page 199 in “Understanding Logic Analyzer
Triggering” on page 190
“To insert a store action (state mode)” on page 78
“To specify whether default storing is initially on or off” on page 71
To specify whether default storing is initially on or off
In the state sampling mode, you can specify whether the default
storing is initially on or off.
1. In the Trigger tab's Default Storing subtab, select the At start of
acquisition option button and choose either On or Off.
See Also
“Storage Qualification” on page 199 in “Understanding Logic Analyzer
Triggering” on page 190
“To insert a store action (state mode)” on page 78
“To Specify Default Storing” on page 70
Editing the Trigger Sequence
When you want to trigger on several events in the device under test
that follow one another, you need to use multiple levels in the trigger
sequence.
For example, multiple levels in the trigger sequence let you trigger on a
particular function calling sequence or capture only the execution
within a particular program loop.
•
“To insert/replace/delete sequence levels” on page 72
71
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
•
“To cut/copy-and-paste sequence levels” on page 73
•
“To specify a level's goto or trigger action” on page 73
•
“To send e-mail when the trigger occurs” on page 74
•
“To view a picture of the trigger sequence” on page 76
•
“To clear the trigger sequence” on page 76
See Also
“Sequence Levels” on page 192 in “Understanding Logic Analyzer
Triggering” on page 190
To insert/replace/delete sequence levels
To insert sequence levels
1. In the Trigger tab's Trigger Sequence area, select the level that you want to
insert before or after.
A yellow box appears around the level.
2. In the Trigger Functions subtab, select the trigger function you want to
insert.
A picture describing the trigger function is shown.
3. Select the Insert before or Insert after button, or select the level button
and choose Insert LEVEL before or Insert LEVEL after.
To replace sequence levels
1. In the Trigger tab's Trigger Sequence area, select the level that you want to
replace.
A yellow box appears around the level.
2. In the Trigger Functions subtab, select the trigger function you want to
insert.
A picture describing the trigger function is shown.
3. Select the Replace button, or select the level button and choose Replace
LEVEL.
72
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
To delete sequence levels
1. In the Trigger tab's Trigger Sequence area, select the level that you want to
delete.
A yellow box appears around the level.
2. Select the Delete button, or select the level button and choose Delete
LEVEL.
See Also
“To cut/copy-and-paste sequence levels” on page 73
To cut/copy-and-paste sequence levels
You can change the order of levels in the trigger sequence by cuttingand-pasting or you can copy levels by copying-and-pasting.
1. In the Trigger tab's Trigger Sequence area, select the level that you want to
cut or copy.
A yellow box appears around the level.
2. Select Cut level or Copy level from the Edit menu, or select the level
button and choose Cut LEVEL or Copy LEVEL.
3. Select the level that you want to paste before or after.
4. Select Paste level before or Paste level after from the Edit menu, or select
the level button and choose Paste LEVEL before or Paste LEVEL after.
See Also
“To insert/replace/delete sequence levels” on page 72
To specify a level's goto or trigger action
When using multiple levels in the trigger sequence, you specify the
event search order by setting the goto or trigger action in each
sequence level.
1. In the Trigger tab's Trigger Sequence area, select the level whose goto or
trigger action you want to specify.
A yellow box appears around the level.
2. Select the Trigger or Goto button and choose the appropriate Goto or
Trigger action.
73
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
3. If you chose the Goto or Trigger and goto action, select the level button
and choose the appropriate level.
Searching for events that trigger the analyzer always starts at the first
level. Searching stops after one of the Trigger and fill memory
actions.
One level can branch to one of several other levels depending on the
evaluation of the sample. You can set up multi-way branches using
advanced trigger functions or by selecting an If button and choosing
Insert BRANCH.
NOTE:
When you want to test a single sample for multiple conditions and take
different actions based on which is true, use branches within a trigger
sequence level. When you want to test different samples, use different
sequence levels.
See Also
“Understanding Logic Analyzer Triggering” on page 190
“To view a picture of the trigger sequence” on page 76
“Advanced Trigger Functions” on page 151
To send e-mail when the trigger occurs
You can set up the logic analyzer to send e-mail when the trigger
occurs. This is useful when triggering on an event that rarely occurs,
when you may not be around the logic analyzer to see that it triggered.
1. In the Trigger tab's Trigger Sequence area, select the level whose trigger
you want to send e-mail on.
A yellow box appears around the level.
2. Select the Trigger or Goto button and choose the Trigger, send e-mail,
and fill memory action.
3. Select the E-mail Setup button.
4. In the E-mail Setup dialog, enter the name of the SMTP (see page 75) mail
server (if you don't know this, contact your System Administrator), the
recipient's e-mail address (use spaces to separate multiple addresses), and
the text of the message.
74
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
If you want e-mail to be sent on each trigger of a repetitive run, select the
Send e-mail on repetitive run check box.
5. Select the OK button.
Note that the e-mail is sent when the trigger occurs and not after the
logic analyzer's acquisition memory is full.
You only need to specify one send e-mail action per trigger sequence.
As long as one trigger action sends e-mail, any trigger in the sequence
will result in e-mail being sent. (You cannot specify different send email setups in a trigger sequence.)
If the SMTP server has a problem with the default sender address
•
Message Format
You may need to specify a sender address that is recognizable by the
server. A possible address might be the one specified in the To: field.
The automatically generated text is shown as follows:
Example: system14 : Slot C : Analyzer C has triggered
Where system14 is the analysis system IP address or alias you have
assigned to it; Slot C is the frame slot the module is in; Analyzer C
identifies the specific analyzer module from others when configured in
a multi-module frame configuration.
Any text you add in the text entry area of the e-mail setup dialog will
appear after the automatically generated text.
What is SMTP? SMTP (Simple Mail Transfer Protocol) is a TCP/IP
protocol used in sending and receiving e-mail. A protocol is the special
set of rules for communicating the end points in a telecommunication
connection as they send signals back and forth.
Protocols exist at several levels in a telecommunication connection.
There are hardware telephone protocols. There are protocols between
the end points in communicating programs within the same computer
or at different locations. Both end points must recognize and observe
the protocol.
On the Internet, there are the following TCP/IP protocols:
•
TCP (Transmission Control Protocol), which uses a set of rules to
75
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
exchange messages with other Internet points at the information packet
level.
•
IP (Internet Protocol), which uses a set of rules to send and receive
messages at the Internet address level.
•
HTTP, FTP, SMTP and other protocols, each with defined sets of rules to
use with other Internet points relative to a defined set of capabilities.
To view a picture of the trigger sequence
1. In the Trigger tab, select the Overview subtab.
A picture of the trigger sequence is shown.
See Also
“Editing the Trigger Sequence” on page 71
To clear the trigger sequence
1. When the Trigger tab is displayed, select Trigger Sequence or All from the
Clear menu.
Selecting Trigger Sequence restores the default trigger sequence for the
selected sampling mode.
Selecting All restores the default trigger sequence, trigger settings, and
default storing if in the state sampling mode.
To restore default trigger settings
1. When the Trigger tab is displayed, select Settings from the Clear menu.
Settings (acquisition depth and trigger position) are returned to their
defaults. In the state sampling mode, time tags are turned back on. In the
timing sampling mode, the sample period returns to its fastest setting.
Editing Advanced Trigger Functions
After you break down a trigger function (if it didn't quite provide the
trigger capability you need), or after you select one of the advanced
trigger functions, you're ready to edit the advanced trigger function.
All trigger functions look for events and, if those events are found, take
76
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
actions.
Most often, the event is something that occurs on the probed signals
(label events), and the action is to trigger the logic analyzer. However,
events can also test timer, counter, and/or flag values that are set up in
the logic analyzer, and actions can include setting up timers, counters,
and flags as well as specifying special store actions.
•
“To specify a duration or occurrence count for events (timing mode)” on
page 77
•
“To insert a store action (state mode)” on page 78
•
“To insert timer actions/events” on page 78
•
“To insert counter actions/events” on page 79
•
“To insert flag actions/events” on page 80
•
“To insert a "reset occurrence counter" action” on page 82
•
“To group events” on page 82
•
“To use named events” on page 83
To specify a duration or occurrence count for events
(timing mode)
When working with advanced trigger functions or when you break
down other trigger functions, you can specify that an event be present
for a certain amount of time, or occur in a certain number of samples,
before the associated action is taken.
To specify a time duration for events
1. In the Trigger tab's Trigger Sequence area, if the present for > button is
not present, select the occurs button and choose present for >.
2. Enter a time duration value.
The event must be present for the specified period of time before the
action is taken.
To specify a < duration, break down the Find pattern present for <
duration trigger function. (It actually uses occurrence counts and four
sequence levels.)
77
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
To specify an occurrence count for events
1. In the Trigger tab's Trigger Sequence area, if the occurs button is not
present, select the present for > button and choose occurs.
2. Enter an occurrence count value.
3. If the occurrence count is greater than 1, select whether the event should
occur consecutively or eventually.
The event must occur the specified number of times before the action
is taken.
To insert a store action (state mode)
You can insert store actions to override the default storage qualifier for
the samples that cause actions to occur, and you can insert store
actions to turn default storing on or off.
1. In the Trigger tab's Trigger Sequence area, select one of the action buttons
(for example, Trigger or Goto), choose Insert ACTION, choose Store, and
choose either Store sample, Don't store sample, Turn on default
storing, or Turn off default storing.
You can use store actions to set up sequence level storage qualification.
See Also
“Storage Qualification” on page 199 in “Understanding Logic Analyzer
Triggering” on page 190
“To Specify Default Storing” on page 70
“To specify whether default storing is initially on or off” on page 71
To insert timer actions/events
Timers are like stopwatches. You can insert actions to start (from
zero), stop (and reset), pause, or resume a timer. You can insert timer
events in a different sequence level to test the value of a timer.
NOTE:
No timer is available for the first pod pair assigned to a logic analyzer. For
each additional pod pair assigned to the analyzer, an additional timer is
available.
78
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
To insert a timer action
1. In the Trigger tab's Trigger Sequence area, select one of the action buttons
(for example, Trigger or Goto), choose Insert ACTION, choose Timer, and
choose either Start from reset, Stop and reset, Pause, or Resume.
To insert a timer event
Timer events are like other events in that they evaluate to either true
or false.
1. In the Trigger tab's Trigger Sequence area, select one of the existing event
buttons (for example, a label name, Anything, Timer, Counter, or Flag) and
choose to insert or replace a Timer.
2. Select the timer number button and choose the number of the timer you
want to test.
3. Select the operator button and choose either >= or <.
4. Enter the time value.
The minimum value you can test a timer for depends on the timing/state
analyzer configuration.
See Also
“To assign pods to one or two analyzers” on page 56
To insert counter actions/events
Global counters are available in the trigger sequence. You can insert
actions to reset or increment a counter. You can insert counter events
in a different sequence level to test the value of a counter.
To insert a counter action
1. In the Trigger tab's Trigger Sequence area, select one of the action buttons
(for example, Trigger or Goto), choose Insert ACTION, choose Counter,
and choose either Reset or Increment.
To insert a counter event
Counter events are like other events in that they evaluate to either true
or false.
1. In the Trigger tab's Trigger Sequence area, select one of the existing event
79
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
buttons (for example, a label name, Anything, Timer, Counter, or Flag) and
choose to insert or replace a Counter.
2. Select the counter number button and choose the number of the counter
you want to test.
3. Select the operator button and choose either >= or <.
4. Enter the counter value.
To insert flag actions/events
Flags can be used to signal between modules in the logic analysis
system mainframe, an expansion frame, or in multiple frames
connected with the multiframe module.
There are 4 flags that are shared across all connected logic analysis
system frames. A flag may be driven or received by multiple modules.
Using flags, logic analyzer modules can communicate back and forth
with each other multiple times during a data acquisition, both before
and after their trigger events occur. (By comparison, the Intermodule
window lets one module arm another module one time when its trigger
occurs.)
By default, flags are cleared. You can insert actions to set, clear, pulse
set, or pulse clear a flag. You can insert flag events in different logic
analyzer modules to test whether a flag is set or clear.
NOTE:
In all but the slowest state speed, the logic analyzer can check flags by
inserting an event, but cannot change flag status with an action. Flag actions
are not available when not using the slowest state speed.
A flag that is set by a module remains set until that module clears it. If
multiple modules set the same flag, all of those modules must clear the
flag before it becomes clear.
Flags can also be used to drive the logic analysis system's Port Out
signal.
To insert a flag action
You can use the Set/clear/pulse flag trigger function to insert a flag
action. When editing advanced trigger functions, follow these steps to
80
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
insert a flag action:
1. In the Trigger tab's Trigger Sequence area, select one of the action buttons
(for example, Trigger or Goto), choose Insert ACTION, choose Flag, and
choose either Set, Clear, Pulse set, or Pulse clear.
Flags in Pulse mode sit in the opposite state when not being pulsed. If you
insert a Pulse set action for a flag in one analyzer, you cannot insert a
Pulse clear action for the same flag in a different analyzer.
NOTE:
Within an analyzer, the same flag cannot be used in both Pulse and Level (Set/
Clear) modes. If a flag action is inserted or modified with a different mode
than other actions for the same flag, all actions for that flag will change to
match the new mode.
2. If you chose Pulse set or Pulse clear, enter the width of the pulse.
In timing mode, pulse width is settable from 54 ns to 1.53 us in 6 ns steps.
In state mode, pulse width is settable from 50 ns to 1.275 ns in 5 ns steps.
NOTE:
Within an analyzer, a flag's pulse width must be the same in every action for
that flag. Whenever the pulse width is changed in a flag action, it changes in all
other actions for that flag.
3. Select the flag number button and choose the number of the flag you want
the action to occur on.
To insert a flag event
Flag events are like other events in that they evaluate to either true or
false.
You can use the Wait for flag trigger function to insert a flag event.
When editing advanced trigger functions, follow these steps to insert a
flag event:
1. In the Trigger tab's Trigger Sequence area, select one of the existing event
buttons (for example, a label name, Anything, Timer, Counter, or Flag) and
choose to insert or replace a Flag.
2. Select the flag number button and choose the number of the flag you want
to test.
3. Select whether you're testing if the flag is Set or Clear.
81
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
There is approximately 100 ns of delay before a flag action can be seen
by a flag event.
To drive the Port Out signal with a flag
1. In the main logic analysis system window, select the Port Out button.
2. In the Port Out dialog, select the Type, Polarity, and Output options.
When driving the Port Out signal with a flag, you can select the
Feedthrough type to pass the current state of the flag (set or clear)
directly to Port Out.
3. For the Armed by option, select the flag that will drive the Port Out signal.
4. Close the Port Out dialog.
5. Insert a flag action in one of the logic analyzer modules to drive the flag.
There is approximately 100 ns of delay between a flag action and the
signal on Port Out.
To insert a "reset occurrence counter" action
You can reset an occurrence counter if some event occurs by inserting
a "reset occurrence counter" action.
1. In the Trigger tab's Trigger Sequence area, select one of the action buttons
(for example, Trigger or Goto), choose Insert ACTION, and choose Reset
occurrence counter.
See Also
“To specify a duration or occurrence count for events (timing mode)” on
page 77
To group events
When you are working with advanced trigger functions (or when you
break down other trigger functions) and there are multiple events in an
event list, you can specify their evaluation order by grouping the
events.
1. In the Trigger tab's Trigger Sequence area, select the If, If not, Else if, or
Else if not button, and choose Group events.
2. In the Group Events dialog, either select the Add parens button to group
82
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
events or select the Remove parens button to ungroup events.
3. Select the OK button.
To use named events
When you are working with advanced trigger functions (or when you
break down other trigger functions), you can name an event list and
use it later when inserting or replacing events.
To give an event list a name
1. In the Trigger tab's Trigger Sequence area, select the If, If not, Else if, or
Else if not button, and choose Name event list.
2. In the Name Event List dialog, enter the name and select the OK button.
To insert a named event
1. In the Trigger tab's Trigger Sequence area, select a label name button and
choose to insert or replace a Named event.
2. In the Named Event selection dialog, select the named event, and select
the OK button.
To edit a named event
1. In the Trigger tab's Trigger Sequence area, select the named event button
and choose Edit locally or Edit globally.
Locally means to edit (and rename) this instance of the named event.
Globally means to edit all instances of the named event.
2. In the Edit dialog, edit the event list as you would edit it in the Trigger
tab's Trigger Sequence area.
Saving/Recalling Trigger Setups
You can save a trigger setup within a session by using trigger save/
recall.
•
“To save a trigger setup” on page 84
83
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
•
“To recall a trigger setup” on page 84
•
“To clear the trigger save/recall list” on page 84
See Also
“Save/Recall Subtab” on page 155
To save a trigger setup
1. Set up the trigger.
2. In the Trigger tab's Save/Recall subtab, select the Save button.
3. Select a memory location to store the trigger setup in.
4. In the Buffer Name dialog, enter a descriptive name for the trigger setup.
To recall a trigger setup
1. In the Trigger tab's Save/Recall subtab, select the Recall button.
2. Choose the trigger setup from one of the previous measurements or one of
the save/recall memories.
Recalling a trigger setup changes the trigger arming, memory depth,
and trigger position as well as the trigger sequence. Recalling a trigger
setup will not change the sampling mode configuration.
If one of the settings in the recalled trigger setup conflicts with the
sampling mode configuration, it will be set to the closest setting.
Also, if the trigger setup uses a trigger function library that does not
exist on this mainframe, it will not load correctly.
To clear the trigger save/recall list
1. When the Trigger tab is displayed, select Save/Recall Memories from the
Clear menu.
Running Measurements
After you set up a trigger, you're ready to run the logic analyzer
measurement.
84
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
•
“To start/stop measurements” on page 85
•
“If nothing happens when you start a measurement” on page 85
•
“To view the trigger status” on page 86
To start/stop measurements
To start measurements
1. Select the Run Single
, Run Repetitive
Group Run Repetitive, or Run All
, Group Run Single
,
button.
Run
starts only the instrument you are using. Single runs gather data
until the logic analyzer memory is full, and then stop.
Repetitive runs
keep repeating the same measurement and are
useful for gathering statistics.
Group Run
(or repetitive group run) starts all instruments attached
to group run in the Intermodule window.
Run All
starts all instruments currently placed in the workspace.
To stop a measurement
1. Select the Stop
or Stop All button.
If nothing happens when you start a measurement
•
Analyzers with deep memory take a noticeable amount of time to complete
a run. Because data is not displayed until acquisition completes, it may
look like nothing is happening. Check the Run Status window to see if the
logic analyzer is still running.
•
Messages such as "Waiting in level 1" may indicate you need to refine your
trigger.
•
If the status shows as "Stopped", the analyzer either finished the
acquisition, or was unable to run. The cause of the problem is listed in the
85
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
bottom half of the Run Status window.
•
Look for an error message in the message bar at the top of the window.
Common messages are "slow or missing clock" and "Waiting for trigger".
•
If Run briefly changed to Stop or Cancel, select the Window menu,
choose the logic analyzer's slot, then choose the Waveform or Listing
display.
See Also
“Slow or Missing Clock” on page 178
“Waiting for Trigger” on page 182
“Error Messages” on page 168
To view the trigger status
While a logic analyzer measurement is running, you can view the
trigger status to see the sequence level that is evaluating captured
data, occurrence and global counter values, and flag values.
1. In the Trigger tab, select the Status subtab.
See Also
Run status (in the system help volume).
86
Chapter 2: Task Guide
Displaying Captured Data
Displaying Captured Data
Once you have run a measurement and filled the logic analyzer's
acquisition memory with captured data, you can display the captured
data with one of the display tools.
You can use analysis tools to filter data and compare data sets.
You can also analyze captured data with toolsets like the Serial Analysis
Toolset and the System Performance Analysis Toolset.
•
“To open Waveform or Listing displays” on page 87
•
“To display Timing Zoom data” on page 88
•
“To use other display tools” on page 89
•
“If the captured data doesn't look correct” on page 91
•
“If there are filtered data holes in display memory” on page 91
•
“To display symbols for data values” on page 92
•
“To cancel the display processing of captured data” on page 93
To open Waveform or Listing displays
Waveform displays are typically used when data is captured with the
timing sampling mode, and Listing displays are used when data is
captured with the state sampling mode.
1. From the Window menu, select your logic analyzer and choose the
Waveform or Listing command.
87
Chapter 2: Task Guide
Displaying Captured Data
Waveform and Listing (and other) display tools provide global markers
that can be used to correlate data that is captured by different
instrument modules or displayed differently in other display tool
windows.
The Waveform and Listing display tools also give you the ability to
search for particular data values captured on labels.
Listing displays let you load inverse assemblers that will decode
captured data into assembly language mnemonics. From the Listing
display, you can also open Source Correlation Toolset (Source Viewer)
windows that can display the high-level language source code that is
associated with captured data.
See Also
Using the Digital Waveform Display Tool (see the Waveform Display Tool
help volume)
Using the Listing Display Tool (see the Listing Display Tool help volume)
Working with Markers (see the Markers help volume)
To display Timing Zoom data
1. From the Window menu, select your logic analyzer and choose the
TimingZoom command.
Tips for interpreting Timing Zoom Data
•
88
In a state measurement, Timing Zoom data may appear to be slightly off
from standard data. This is because the Timing Zoom data is asynchronous
to the clock signal, whereas the standard data is not sampled until after the
setup period.
Chapter 2: Task Guide
Displaying Captured Data
•
In Waveform displays, Timing Zoom and the regular data are in different
windows. To view them together, select Edit -> Insert from the menu bar.
•
In Waveform displays, if Timing Zoom and regular data do not correspond,
check that the display is set to Seconds/div. Because Timing Zoom has a
smaller sample period, when the display is set to Samples/div the data
become more divergent the farther from the trigger you scroll.
To use other display tools
You can add display tools to your logic analyzer via the Workspace
window.
1. Select the Workspace button (or from the Window menu, select System
and Workspace).
2. In the Workspace window, scroll down to the Display portion of the tool
icon list.
3. Drag the display tool icon and drop it on the analyzer icon.
4. To open the display tool, select its icon and choose the Display command.
You can use the Chart display tool to chart the data on a label over
89
Chapter 2: Task Guide
Displaying Captured Data
time. For example, if you use storage qualification (in the state
sampling mode) or the Pattern Filter analysis tool, you can chart
variable values.
You can use the Distribution display tool to show how often different
values (among the possible values) are captured on a label.
You can use the Compare analysis tool to show the differences between
two measurement data sets. For example, you can run a measurement
on one circuit board, then run the same measurement on a different
circuit board (or on the same circuit board in different environmental
conditions), and compare the results.
You can use the Pattern Filter analysis tool to remove samples from a
measurement data set before displaying or exporting the data. This lets
you look at selected samples without having to re-capture data.
You can use the Serial Analysis toolset to convert streams of serial data
into parallel words which are easier to view and analyze.
You can use the System Performance Analysis toolset to do things like:
isolate the root cause of performance bottlenecks, measure function
execution times, view the occurrence rate of an event, analyze bus
occupation and bandwidth, analyze bus stability, analyze jitter or time
dispersion, etc.
See Also
Using the Chart Display Tool (see the Chart Display Tool help volume)
Using the Distribution Display Tool (see the Distribution Display Tool
help volume)
Using the Compare Analysis Tool (see the Compare Tool help volume)
Using the Pattern Filter Analysis Tool (see the Pattern Filter Tool help
volume)
Using the Serial Analysis Tool (see the Serial Analysis Tool help volume)
Using the System Performance Analyzer (see the System Performance
Analyzer help volume)
Measurement Examples (see the Measurement Examples help volume)
90
Chapter 2: Task Guide
Displaying Captured Data
If the captured data doesn't look correct
Intermittent Data
Errors
Check for poor connections, incorrect signal levels on the hardware,
incorrect logic levels under the logic analyzer's Config tab, or marginal
timing for signals.
Unwanted Triggers
If you are using an inverse assembler or a pipeline, triggers can be
caused by instructions that were fetched but not executed. To fix, add
the prefetch queue or pipeline depth to the trigger address.
The depth of the prefetch queue depends on the processor that you are
analyzing, and can be quite deep.
Another solution which is sometimes preferred with very deep prefetch
queues is to add writes to dummy variables to your software. Put the
instruction just before the area you want to trigger on, then trigger on
the actual write to this variable. Although the instruction is prefetched,
the analyzer can be set to only trigger when the write is executed.
Capacitive Loading on Excessive capacitive loading can degrade signals, resulting in
the Device Under Test suspicious data or even system lockup. All analysis probes add
capacitive loading, as can custom probes you design for your device
under test. To reduce loading, remove as many pin protectors,
extenders, and adapters as possible.
Careful layout of your device under test can minimize loading problems
and result in better margins for your design. This is especially
important for systems running at frequencies greater than 50 MHz.
If there are filtered data holes in display
memory
When an analyzer measurement occurs, acquisition memory is filled
with data that is then transferred to the display memory of the analysis
or display tools you are using, as needed by those tools. In normal use,
this demand driven data approach saves time by not transferring
unnecessary data.
91
Chapter 2: Task Guide
Displaying Captured Data
Since acquisition memory is cleared at the beginning of a
measurement, stopping a run may create a discrepancy between
acquisition memory and the memory buffer of connected tools. Without
a complete trace of acquisition memory, the display memory will
appear to have 'holes' in it which appear as filtered data.
This situation will occur in these cases:
•
If you stop a repetitive measurement after analyzer data has been cleared
and before the measurement is complete.
•
If a trigger is not found by the analyzer and the run must be stopped to
regain control.
To make sure all of the data in a repetitive run is available for viewing:
1. In the workspace, attach a Filter tool to the output of the analyzer.
2. In the Filter, select "Pass Matching Data"
3. In the filter terms, assure the default pattern of all "Don't Cares" (Xs).
This configuration will always transfer all data from acquisition
memory. While this configuration will increase the time of each run, it
will guarantee that repetitive run data is available regardless of when it
is stopped.
To display symbols for data values
You can display data in symbolic form in some of the display tools, such
as the Listing display and the Waveform display.
To view symbolic values in a waveform display
1. Select the label name where you want to display symbolic values.
2. Select Properties....
3. In the Properties dialog:
92
•
Set ShowValue to On.
•
Set Base to Symbols or Line#.
Chapter 2: Task Guide
Displaying Captured Data
•
Select the OK button.
The symbolic names for the values now appear in the overlaid bus
waveform.
To view symbolic values in a listing display
1. Select the numeric base of the label where you want to display symbolic
values.
2. Set the numeric base to Symbols or Line#.
The symbolic names for the values now appear instead of numeric data.
See Also
“Using Symbols” on page 94
To cancel the display processing of captured
data
You can cancel the processing of captured data if it is taking too long.
1. Select the Cancel
button.
93
Chapter 2: Task Guide
Using Symbols
Using Symbols
You can use symbol names in place of data values when:
•
Setting up triggers
•
Displaying captured data
•
Searching for patterns in Listing displays
•
Setting up pattern filters
•
Setting up ranges in the System Performance Analyzer
Symbol names can be: variable names, procedure names, function
names, source file line numbers, etc.
You can load symbol name definitions into the logic analyzer from a
program's object file or from a general-purpose ASCII format symbol
file, or you can define symbol names in the logic analyzer.
•
“To load object file symbols” on page 95
•
“To adjust symbol values for relocated code” on page 96
•
“To create user-defined symbols” on page 97
•
“To enter symbolic label values” on page 98
•
“To create an ASCII symbol file” on page 99
•
“To create a readers.ini file” on page 100
See Also
To go to a pattern in the Listing (see the Listing Display Tool help
volume)
To modify the Source Viewer trace setup (see the Listing Display Tool
help volume)
To define System Performance Analyzer state interval ranges (see the
System Performance Analyzer help volume)
94
Chapter 2: Task Guide
Using Symbols
To load object file symbols
Object files are created by your compiler/linker or other software
development tools.
1. Generate an object file with symbolic information using your software
development tools.
2. If your language tools cannot generate object file formats that are
supported by the logic analyzer, create an ASCII symbol file (see page 99).
3. Select the Symbol tab and then the Object File tab.
4. Select the label name you want to load object file symbols for.
In most cases you will select the label representing the address bus of the
processor you are analyzing.
5. Specify the directory to contain the symbol database file (.ns) in the field
under, Create Symbol File (.ns) in This Directory. Select Browse... if
you wish to find an existing directory name.
6. In the Load This Object/Symbol File For Label field, enter the object file
name containing the symbols. Select Browse... to find the object file and
select Load in the Browser dialog.
If your logic analyzer is NFS mounted to a network, you can select object
files from other servers.
7. If your program relocates code, see “To adjust symbol values for relocated
code” on page 96.
The name of the current object file is saved when a configuration file is
saved. The object file will be reloaded when the configuration is loaded.
To reload object file symbols
1. Select the object file/symbol file to reload from the Object Files with
Symbols Loaded For Label field.
2. Select the Reload button.
The values of the object file symbols being used in the trigger sequence
or in SPA state-interval ranges will be updated automatically each time
the object file symbols are reloaded.
95
Chapter 2: Task Guide
Using Symbols
To delete object file symbol files
1. Select the Symbol tab, and then the Object File tab.
2. Select the file name you want to delete in the text box labeled, Object
Files with Symbols Loaded For Label.
3. Select Unload.
See Also
“Symbol File Formats” on page 161
To adjust symbol values for relocated code
Use this option to add offset values to the symbols in an object file. You
will need this if some of the sections or segments of your code are
relocated in memory at run-time. This can occur if your system
dynamically loads parts of your code so that the memory addresses
that the code is loaded into are not fixed.
To adjust symbol values for a single section of code
1. Select the Symbol tab and then the Object File tab.
2. In the Object Files with Symbols Loaded For Label list, select the file
whose symbols you wish to relocate.
3. Select the Relocate Sections... button.
4. In the Section Relocation dialog, select the field you wish to edit in the
section list.
5. Enter the new value for that field and press Enter on your keyboard.
6. Repeat steps 4 through 6 above for any other sections to be relocated.
7. Select Close.
To adjust all symbol values
1. Select the Symbol tab and then the Object File tab.
2. In the Object Files with Symbols Loaded For Label list, select the file
whose symbols you wish to relocate.
96
Chapter 2: Task Guide
Using Symbols
3. Select the Relocate Sections... button.
4. Enter the desired offset in the Offset all sections by field. The offset is
applied from the linked address or segment.
5. Select Apply Offset.
6. Select Close.
To create user-defined symbols
1. Under the Symbol tab, select the User Defined tab.
2. Select the label name you want to define symbols for.
3. At the bottom of the User Defined tab, enter a symbol name in the entry
field.
4. Select a numeric base.
5. Select Pattern or Range type for the symbol.
6. Enter values for the pattern or range the symbol will represent.
7. Select Add.
8. Repeat steps 3 through 7 for additional symbols.
9. You can edit your list of symbols by replacing or deleting them, if desired.
To replace user-defined symbols
1. Under the Symbol tab, select the User Defined tab.
2. Select the label you want to replace symbols for.
3. Select the symbol to replace.
4. At the bottom of the User Defined tab, modify the symbol name, numeric
base, Pattern/Range type, and value, as desired.
5. Select the Replace button.
6. Repeat steps 3 through 5 to replace other symbols, if desired.
97
Chapter 2: Task Guide
Using Symbols
To delete user-defined symbols
1. Under the Symbol tab, select the User Defined tab.
2. Select the label you want to delete symbols from.
3. Select the symbol to delete.
4. Select the Delete button.
5. Repeat steps 3 and 4 to delete other symbols, if desired.
To load user-defined symbols
If you have already saved a configuration file, and the configuration
included user-defined symbols, load the file with its symbols, as
follows:
1. In the menu bar of your analyzer window, select File and then Load
Configuration....
2. In the Load Configuration dialog, select the directory and filename to be
loaded.
3. Select the target of the load operation.
4. Select Load.
User-defined symbols that were resident in the logic analyzer when the
configuration was saved are now loaded and ready to use.
To enter symbolic label values
When entering label values in the trigger sequence:
1. Choose the Symbols or Line #s number base.
2. Select the Absolute XXXX button.
3. In the Symbol Selector dialog, select the symbol you want to use. All of
your symbols for the current label, regardless of type, will be available in
the dialog.
•
98
Use the Search Pattern (see page 160) field to filter the list of symbols
by name. You can use the Recall button to recall a desired Search
Chapter 2: Task Guide
Using Symbols
Pattern.
•
Use the Find Symbols of Type selections to filter the symbols by type.
4. Select the symbol you want to use from the list of Matching Symbols.
5. If you are using object file symbols, you may need to:
•
Set Offset By (see page 160) to compensate for microprocessor
prefetches.
•
Set Align to x Byte (see page 161) to trigger on odd-byte boundaries.
6. Select the Beginning, End, or Range of the symbol.
7. Select the OK button.
The name of your symbol now appears as the value of the label.
8. Select the Cancel button to exit the Symbol Selector dialog without
selecting a symbol.
NOTE:
When you use symbolic label values in trigger sequences, information about
the symbols is saved with the logic analyzer configuration (and in the trigger
Save/Recall buffer). If you re-compile your program and reload the logic
analyzer configuration (or reload symbols and recall a trigger), the symbol
values are automatically re-calculated.
Triggers set up by the Source Viewer now use symbolic values and are recalculated in the same way.
See Also
“To specify a label pattern event” on page 65
“Symbols Selector Dialog” on page 159
To create an ASCII symbol file
General-purpose ASCII symbol files are created with text editing/
processing tools.
See Also
“General-Purpose ASCII (GPA) Symbol File Format” on page 162
99
Chapter 2: Task Guide
Using Symbols
To create a readers.ini file
You can change how an ELF/Stabs, Ticoff or Coff/Stabs symbol file is
processed by creating a reader.ini file.
1. Create the reader.ini file on your workstation or PC.
2. Copy the file to /logic/symbols/readers.ini on the logic analysis system.
Reader options
C++Demangle
1= Turn on C++ Demangling (Default)
0= Turn off C++ Demangling
C++DemOptions
803=
203=
403=
800=
200=
400=
Standard Demangling
GNU Demangling
(Default Elf/Stabs)
Lucid Demangling
Standard Demangling without function parameters
GNU Demangling without function parameters
Lucid Demangling without function parameters
MaxSymbolWidth
80= Column width max of a function or variable symbol
Wider symbols names will be truncated.
(Default 80 columns)
OutSectionSymbolValid
0= Symbols whose addresses aren't within the
defined sections are invalid (Default)
1= Symbols whose addresses aren't within the
defined sections are valid
This option must be specified in the Nsr section of the Readers.ini file:
[Nsr]
OutSectionSymbolValid=1
ReadElfSection
2= Process all globals from ELF section (Default)
Get size information of local variables
1= Get size information of global and local variables
Symbols for functions will not be read, and
only supplemental information for those symbols in
the Dwarf or stabs section will be read.
0= Do not read the Elf Section
If a file only has an ELF section this will have no effect and the ELF
100
Chapter 2: Task Guide
Using Symbols
section will be read completely. This can occur if the file was created
without a "generate debugger information" flag (usually -g). Using the g will create a Dwarf or Stabs debug section in addition to the ELF
section.
StabsType
StabsType=0
StabsType=1
StabsType=2
Reader will determine stabs type (Default)
Older style stabs
(Older style stabs have individual symbol
tables for each file that was linked into
the target executable, the indexes of each
symbol table restart at 0 for each file.)
Newer style stabs
(New style stabs have a single symbol table
where all symbols are merged into a large
symbol array).
ReadOnlyTicoffPage
ReadOnlyTicoffPage tells the ticoff reader to read only the symbols
associated with the specified page (as an example
'ReadOnlyTicoffPage=0' reads only page 0 symbols). A value of -1 tells
the ticoff readers to read symbols associated with all pages.
ReadOnlyTicoffPage=-1 Read all symbols associated will all
ticoff pages (Default)
ReadOnlyTicoffPage=p Read only symbols associated with
page 'p' (where p is any integer
between 0 and n the last page of
the object file).
AppendTicoffPage
AppendTicoffPage tells the ticoff reader to append the page number to
the symbol value. This assumes that the symbol value is 16-bits wide
and that that page number is a low positive number which can be ORed
into the upper 16 bits of an address to create a new 32-bit symbol
address. For example, if the page is 10 decimal and the symbol address
is 0xF100 then the new symbol address will be 0xAF100.
AppendTicoffPage=1
AppendTicoffPage=0
Examples
Append the ticoff page to the symbol
address
Do not append the ticoff page to the
symbol address (Default)
Example for Elf/Stabs
[ReadersElf]
C
101
Chapter 2: Task Guide
Using Symbols
C
MaxSymbolWidth=60
StabsType=2
Example for Coff/Stabs (using Ticoff reader)
[ReadersTicoff]
C
C
MaxSymbolWidth=60
StabsType=2
Example for Ticoff
[ReadersTicoff]
C
C
MaxSymbolWidth=60
ReadOnlyTicoffPage=4
AppendTicoffPage=1
102
Chapter 2: Task Guide
Printing/Exporting Captured Data
Printing/Exporting Captured Data
To print captured data
You can print captured data from display tool windows.
1. In the display tool window, select Print this window from the File menu.
To export captured data
You can use the File Out tool to save measurement data to an ASCII
format file which can then be imported into a spreadsheet application,
a debugger, or some other post-processing tool.
1. Select the Workspace button (or from the Window menu, select System
and Workspace).
2. In the Workspace window, scroll down to the Utilities portion of the tool
icon list.
3. Drag the File Out tool icon and drop it on the analyzer icon.
4. To open the File Out tool, select its icon and choose the Display
command.
5. Select the file name, automatic file sequencing, and output file format
options.
6. Select the Save Data button.
To re-import captured data
You can use the File In tool to re-import measurement data into the
logic analysis system for further analysis.
1. Select the Workspace button (or from the Window menu, select System
and Workspace).
2. In the Workspace window, scroll down to the Utilities portion of the tool
icon list.
3. Drag the File In tool icon and drop it in the workspace.
4. To open the File In tool, select its icon and choose the Display command.
103
Chapter 2: Task Guide
Printing/Exporting Captured Data
5. Select the file name and automatic file sequencing options.
6. Select the Read File button.
7. Drag display, analysis, or toolset icons and drop them on the File In tool
icon to view the imported data.
See Also
Printing Windows - Configurations (see the Agilent Technologies 16700A/
B-Series Logic Analysis System help volume)
Using the File Out Tool (see the File Out Tool help volume)
Using the File In Tool (see the File In Tool help volume)
104
Chapter 2: Task Guide
Cross-Triggering
Cross-Triggering
An instrument must be armed before it can look for a trigger. By
default, instruments are set to be armed immediately when you Run
the measurement.
However, you can set an analyzer instrument to be armed either by the
second analyzer within the same instrument (if it's turned on) or by
another instrument (in a different slot or frame).
•
“To cross-trigger between two analyzers” on page 105
•
“To cross-trigger with another instrument” on page 106
To cross-trigger between two analyzers
1. Make sure both analyzers are turned on. To turn on the second analyzer:
a. In the Format tab, select the Pod Assignment button.
b. In the Pod Assignment dialog, change the analyzer type from Off to
either State or Timing.
The system pauses while setting up the second analyzer. When it is
done, a setup window for the second analyzer appears.
In the analyzer waiting for the arm signal
1. In the Trigger tab's Trigger Sequence area, select the sequence level that
that should wait for the other analyzer's trigger.
2. In the Trigger Functions subtab, select the Wait for second analyzer to
trigger trigger function and select either the Replace or Insert before
button.
(You can also use an advanced trigger function, edit it, and insert an
Analyzer<2> triggers event.)
In the analyzer driving the arm signal
1. Set up the logic analyzer trigger as you would normally.
105
Chapter 2: Task Guide
Cross-Triggering
2. Run the measurement.
To cross-trigger with another instrument
1. Select the Intermodule button (or from the Window menu, select System
and Intermodule).
2. In the Intermodule window, select the icon of the instrument to be armed,
and choose the instrument that will arm it.
When the logic analyzer waits for the arm signal
1. In the Trigger tab's Trigger Sequence area, select the sequence level that
should wait for the other analyzer's trigger.
2. In the Trigger Functions subtab, select the Wait for arm in trigger
function and select either the Replace or Insert before button.
(You can also use an advanced trigger function, edit it, and insert an Arm
in from IMB event.)
NOTE:
If the trigger sequence does not pass through the level containing the wait for
arm event, the logic analyzer will not wait for the arming signal.
When the logic analyzer drives the arm signal
1. In the Trigger tab's Settings subtab, select the analyzer (or analyzers) that
will be used to drive the arm signal.
2. Set up the logic analyzer trigger as you would normally.
In the Trigger tab's Trigger Sequence area, the trigger actions will show
arm out to indicate that the logic analyzer's trigger is driving the arm
signal.
3. Run the measurement.
See Also
Intermodule Window (see the Agilent Technologies 16700A/B-Series
Logic Analysis System help volume)
Group Run Arming Tree (see the Agilent Technologies 16700A/B-Series
Logic Analysis System help volume)
106
Chapter 2: Task Guide
Solving Logic Analysis Problems
Solving Logic Analysis Problems
See Also
•
“To test the logic analyzer hardware” on page 107
•
“If nothing happens when you start a measurement” on page 85
•
“If the captured data doesn't look correct” on page 91
•
“If there are filtered data holes in display memory” on page 91
To test the logic analyzer hardware
In order to verify that the logic analyzer hardware is operational, run
the Self Test utility. The Self Test function of the logic analysis system
performs functional tests on both the system and any installed
modules.
1. Disconnect all probes of the logic analyzer module.
2. If you have any work in progress, save it to a configuration file. (see the
Agilent Technologies 16700A/B-Series Logic Analysis System help
volume)
3. Disconnect all loads, adapters, or analysis probes from the probe cable
ends.
4. From the system window, select the System Admin icon.
5. Select the Admin tab, then Self Test....
The system closes all windows before starting up Self Test.
6. Select Master Frame.
If the module is in an expansion frame, select Expansion Frame.
7. Select the logic analyzer that you want to test.
8. In the Self Test dialog box, select Test All.
You can also run individual tests by selecting them. Tests that require you
to do something must be run this way.
107
Chapter 2: Task Guide
Solving Logic Analysis Problems
If any test fails, contact your local Agilent Technologies Sales Office or
Service Center for assistance.
See Also
Self Test (see the Agilent Technologies 16700A/B-Series Logic Analysis
System help volume)
Agilent Technologies 16716A 167 MHz State/2 GHz Timing Zoom
Service Guide
108
Chapter 2: Task Guide
Saving and Loading Logic Analyzer Configurations
Saving and Loading Logic Analyzer
Configurations
The Agilent Technologies 16716A logic analyzer settings and data can
be saved to a configuration file.
The configuration file will include references to any custom trigger
libraries you have created, but if the configuration is loaded into an
analyzer on a system that does not have the trigger libraries, they will
not work correctly.
You can also save any tools connected to the logic analyzer. Later, you
can restore your data and settings by loading the configuration file into
the logic analyzer.
The Agilent Technologies 16716A logic analyzer can load
configurations for Agilent Technologies 16715A, 16716A, and 16717A
logic analyzers with no restrictions.
The Agilent Technologies 16716A logic analyzer can also load
configuration files generated for the Agilent Technologies 16550A,
16554A, 16555A, 16555D, 16556A, 16556D, 16557D, 16710A, 16711A,
16712A, and 16600A-series logic analyzer modules, but much of the
triggering setup will not transfer. Labels, pod and clock assignments,
and the measurement mode will still work. Setup-and-hold values are
translated by preserving the hold value and adjusting setup
accordingly.
Note that while the eye finder (automatic sampling position
adjustment) selected sampling positions are saved and loaded with
logic analyzer configuration files, eye finder measurement data is not.
You can save eye finder data to separate files (see “To automatically
adjust sampling positions” on page 46).
109
Chapter 2: Task Guide
Saving and Loading Logic Analyzer Configurations
NOTE:
The Agilent Technologies 16700A/B logic analysis systems can translate
configuration files from Agilent Technologies 16500 and 16505A logic analysis
systems if the measurement module is the same.
If the modules are different, first load the configuration file into a module of
the same model number on the new logic analysis system. Re-save the
configuration, then load this configuration into the destination module on the
new system.
To save logic analyzer configurations
See Saving Configuration Files (see the Agilent Technologies 16700A/
B-Series Logic Analysis System help volume).
To load logic analyzer configurations
See Loading Configuration Files (see the Agilent Technologies
16700A/B-Series Logic Analysis System help volume).
110
3
Reference
•
“The Sampling Tab” on page 113
•
“The Format Tab” on page 117
•
“The Trigger Tab” on page 144
111
Chapter 3: Reference
•
“The Symbols Tab” on page 157
•
“Error Messages” on page 168
•
“Specifications and Characteristics” on page 184
112
Chapter 3: Reference
The Sampling Tab
The Sampling Tab
The Sampling tab lets you choose between the logic analyzer's
asynchronous sampling Timing Mode or its synchronous sampling State
Mode. This tab also lets you set controls for the selected mode as well
as Timing Zoom options.
See Also
•
“Timing Mode” on page 114
•
“State Mode” on page 115
“Choosing the Sampling Mode” on page 36
“Using 2 GHz Timing Zoom” on page 53
113
Chapter 3: Reference
The Sampling Tab
Timing Mode
When you select Timing Mode, the Timing Mode Controls area appears.
Full/Half
Channel
Configuration
Lets you configure the timing analyzer for faster sampling
and greater memory depth, but with half the channels.
Trigger Position Lets you specify where the sample that triggered the
analyzer should appear among all the other samples that
are stored in acquisition memory.
Acquisition
Depth
Sample Period
See Also
Lets you use a smaller portion of the acquisition memory
and speed up processing of the captured data.
Lets you specify how often the the logic analyzer samples
signals from the device under test.
“Selecting the Timing Mode (Asynchronous Sampling)” on page 36
“In Either Timing Mode or State Mode” on page 51
114
Chapter 3: Reference
The Sampling Tab
State Mode
When you select State Mode, the State Mode Controls area appears.
167 MHz / 512K
State
All pods are available.
Memory depth is 512K samples per channel. If time or
state count is turned on in Trigger Settings, the total
memory is split between data acquisition storage and time
or state count storage. To maintain the full memory depth
of 512K samples per channel, leave one pod pair
unassigned. (To unassign a pod pair, select the Pod
Assignment button in the Format tab, then drag a pod pair
to unassigned.)
State clock speed matches your device under test's clock,
up to 167 MHz.
Trigger Position Lets you specify where the sample that triggered the
analyzer should appear among all the other samples that
are stored in acquisition memory.
115
Chapter 3: Reference
The Sampling Tab
Acquisition
Depth
Clock Setup
Lets you use a smaller portion of the acquisition memory
and speed up processing of the captured data.
Lets you specify: the clock mode, the clock signal edges
from the device under test that will be used as the sampling
clock, and the clock input signal levels (from the device
under test) that will enable (qualify) the sampling clock.
Generally, the state mode sampling clock is taken from the
signals that clock valid data in the device under test.
The clock channel specifiers graphically show your clock
setup. Edges are ORed ("+") together, and qualifiers are
ANDed (".") to all edges. To qualify just one of the edges,
switch to Advanced Clocking.
All clock channels for the clock setup must be on the pods
of the master card of the module, but the pods do not
need to be part of the state measurement.
Advanced
Clocking
See Also
Lets you specify more complex clock setups than you can
with the normal Master (and perhaps Slave) buttons that
are available for each pod.
“Selecting the State Mode (Synchronous Sampling)” on page 43
“In Either Timing Mode or State Mode” on page 51
116
Chapter 3: Reference
The Format Tab
The Format Tab
The Format tab lets you assign bus and signal names (from the device
under test), to logic analyzer channels. These names are called labels.
Labels are used when setting up triggers and displaying captured data.
The Format tab also lets you assign pod pairs to one or two logic
analyzers, specify the logic analyzer threshold voltage, and adjust the
logic analyzer setup/hold (sampling positions).
The Data On Clocks display column shows all the clock inputs available
as data channels in the present configuration, including the clocks on
expander cards that cannot be used in the clock setup.
The Format tab has activity indicators that show whether the signal a
channel is probing is above the threshold voltage (high), below the
threshold voltage (low), or transitioning.
•
“Pod Assignment Dialog” on page 128
•
“Sampling Positions Dialog” on page 129
•
“Manual Setup/Hold Option” on page 142
•
“Eye Finder Option, Setup Tab” on page 130
•
“Eye Finder Option, Results Tab” on page 132
A label can have up to 32 channels.
Each measurement can define 126 labels.
Labels containing clock bits cannot be used in range terms where the
117
Chapter 3: Reference
The Format Tab
clock bits span more than one pod pair.
See Also
“Formatting Labels for Logic Analyzer Probes” on page 56
“To manually adjust sampling positions” on page 49
“To automatically adjust sampling positions” on page 46
“Importing Netlist and ASCII Files” on page 119
118
Chapter 3: Reference
Importing Netlist and ASCII Files
Importing Netlist and ASCII Files
Netlist Files
The Netlist Import feature provides a method for importing busses and
signals from ASCII netlists created by EDA tools. In order for the
feature to work, the device under test must either use the Agilent
E5346A high density adapter or the Agilent Technologies Termination
Adapter. The adapter must be included as a connector in the netlist.
You can create a Netlist file using an EDA tool. When importing a
Netlist file, you can specify which connectors are of interest;
connectors that are not specified will be ignored.
Example
NET'/Bus1(3)'J1-7
Bus1
Four bit bus
(3)
Bit 3
J1-7
Connector J1, pin 7
To Import a Net List
1. Configure the connector names referenced in the Netlist to logic analyzer
pods.
2. Specify the Net List file to import.
3. Verify that Nets were correctly mapped to Labels, Pods, and Channels.
4. Select or create labels to appear in the interface.
Map the Connector Names (see page 125).
Import the Net List File (see page 125)
Verify Net to Label Mapping (see page 126)
Select/Create Interface Labels (see page 127)
ASCII Files
You can create an ASCII file using any Windows, MS-DOS, or UNIX text
editor. The ASCII file should contain bus names, pod/clock numbers,
and channel definitions.
119
Chapter 3: Reference
Importing Netlist and ASCII Files
For Example
Label1;A2[15:5];A1[5,2]
Label1
Bus Name
A2 and A1
Pod Numbers
[15:5]
Channel 15 through Channel 5 ("***********.....")
[5,2]
Channel 5 and Channel 2 ("..........*..*..")
When setting up the ASCII file a comma (",") separates individual
channels, while a colon (":") creates a range of channels.
The following provides an explanation of how to setup and import
ASCII files into a logic analysis system.
Setting Up ASCII Files
NOTE:
If the analyzer is in state mode with the Clock Setup Mode set to
demultiplexer, slave pods will appear with an "S" in front of the pod
designation.
For example:
Label1;SA2[5] reads as Label1 maps to Slave Pod A2 Channel 5.
Individual channels
Label1;A2[5]
Label1 maps to Pod A2, Channel 5
Multiple channels
Label1;A1[5:2,0] Label1 maps to Pod A1, Channel 5 through Channel 2 and
Channel 0.
Individual channels on different pods
Label1;A2[1];A1
[0]
Label1 maps to Pod A2, Channel 1 and Pod A1, Channel 0.
Multiple channels on different pods
Label1;A3[15:5];
A2[5];A1[6]
Label1 maps to Pod A3, Channel 15 through Channel 5,
120
Chapter 3: Reference
Importing Netlist and ASCII Files
Pod A2 Channel 5, and Pod A1 Channel 6.
Clocks
Label1;CK[AK] Label1 maps to Slot A Clock K.
“Importing ASCII Files” on page 121
“Exporting ASCII Files” on page 121
See Also
“Termination Adapter” on page 123
“E5346A High Density Adapter” on page 124
Considerations when creating an Import file.
When updating a label ensure that the label is turned on. Labels that
are not active will not be updated.
Other considerations:
•
Maximum of 126 labels
•
Maximum of 20 characters per label. (The system truncates after 20
characters.)
•
Maximum of 32 channels per label.
Exporting ASCII Files
1. Select the Format tab.
2. Select File, then select Export Label...
Importing ASCII Files
Shown below is the default setup:
Label1;A1[15:0]
121
Chapter 3: Reference
Importing Netlist and ASCII Files
To Import an ASCII
file.
1. Create an ASCII file for importing into the logic analysis system. For
example:
Lab2;A2[15:10,6,3]
NewLabel2;A2[15]
Label1;A1[15:0]
2. Select the Format tab.
3. Select File, then select Import Labels...
4. Select the ASCII file you created.
5. Select OK.
122
Chapter 3: Reference
Importing Netlist and ASCII Files
Termination Adapter
The logic analyzer cable must have the proper RC network at its input
in order acquire data correctly. The Termination Adapter incorporates
the RC network into a convenient package. It also reduces the number
of pins required for the header on the target board from 40 pins to 20.
123
Chapter 3: Reference
Importing Netlist and ASCII Files
E5346A High Density Adapter
The E5346A high-density adapter provides a convenient and easy way
to connect an Agilent logic analyzer to the signals on your target
system for packages that are difficult to probe. An Amp "Mictor 38"
connector must be installed on your target system board.
124
Chapter 3: Reference
Importing Netlist and ASCII Files
Mapping Connector Names
1. Select the Format tab.
2. Select File, then select Import Netlist.
3. Select Next to go to the Mapping Connector Names dialog.
4. Enter a connector name from the Netlist.
5. Select the type of adapter.
6. Select the logic analyzer pods that are plugged into the adapters.
7. To map more connector names, select Add Connector and repeat the
steps above.
8. Select Next.
Import the Net List File
1. Select Select Netlist File.
125
Chapter 3: Reference
Importing Netlist and ASCII Files
2. Select the file from the File Selection dialog box.
3. Select OK
4. Select Next
Verify Net to Label Mapping
1. Verify that each net was properly imported into a label.
2. Select Next
The list provided, see example below, is the mapping from the logic
analyzer pods and channels to labels.
The label column is editable so you can make changes if necessary. If
you need to delete labels, this can be done more efficiently in the next
dialog box.
126
Chapter 3: Reference
Importing Netlist and ASCII Files
Select/Create Interface Labels
Select any additional labels to be copied into the Format tab. Typically
there is no need to add any more labels. However, this screen is useful
when you want to designate a signal bit in a bus as a separate label
name. It can also be used to delete unneeded labels.
Once you have completed editing the labels, select Done.
127
Chapter 3: Reference
Importing Netlist and ASCII Files
Pod Assignment Dialog
Name:
Lets you name the analyzers.
Type:
Lets you select the timing (asynchronous) sampling mode,
the state (synchronous) sampling mode, or turn the
analyzer off.
Pod Pairs
Can be dragged-and-dropped under one of the analyzers to
assign those channels to the analyzer or can be left
unassigned.
See Also
“To assign pods to one or two analyzers” on page 56
“To name an analyzer” on page 52
“To select the timing mode” on page 40
“To select the state mode” on page 43
“To turn an analyzer off or on” on page 53
128
Chapter 3: Reference
Importing Netlist and ASCII Files
Sampling Positions Dialog
The Sampling Positions dialog lets you position the logic analyzer's
setup/hold window (or sampling position) so that data on high-speed
buses is captured accurately, in other words, so that data is sampled
when it is valid.
When the device under test's data valid window is less than 2.5 ns
(roughly, for clock speeds >= 200 MHz), it's easiest to use eye finder
to locate the stable and transitioning regions of signals and to
automatically adjust sampling positions.
When the device under test's data valid window is greater than 2.5 ns
(roughly, for clock speeds < 200 MHz), it's easiest to adjust the
sampling position manually, without using the logic analyzer to locate
the stable and transitioning regions of signals.
•
“Eye Finder Option, Setup Tab” on page 130
•
•
See Also
“Eye Finder Advanced Settings Dialog” on page 131
“Eye Finder Option, Results Tab” on page 132
•
“Eye Finder Run Messages” on page 134
•
“Eye Finder Info Messages” on page 137
•
“Eye Finder Load/Save Messages” on page 139
•
“Eye Finder Option, File Info Tab” on page 140
•
“Manual Setup/Hold Option” on page 142
“Understanding State Mode Sampling Positions” on page 206
“Selecting the State Mode (Synchronous Sampling)” on page 43
129
Chapter 3: Reference
Importing Netlist and ASCII Files
Eye Finder Option, Setup Tab
File menu
Lets you save/load eye finder data.
EyeFinder menu Lets you run eye finder, choose the run mode, and access
the “Eye Finder Advanced Settings Dialog” on page 131.
Run Mode
Lets you look at eye finder with demo data or in normal
operating mode by sampling signals from the device under
test.
Repetitive Run Runs the eye finder repetitively, so you can see how stable
and transitioning signals vary over time.
Advanced
See “Eye Finder Advanced Settings Dialog” on page 131.
Label Selection Lets you choose the labels (channels) to run eye finder on.
You can expand or overlay the signals in a label, select all or
select none of the signals, select individual signals, or select
multiple signals.
130
Chapter 3: Reference
Importing Netlist and ASCII Files
If a channel appears in multiple labels, selecting that
channel will select it in each of those labels.
See Also
“Understanding State Mode Sampling Positions” on page 206
“To automatically adjust sampling positions” on page 46
Eye Finder Advanced Settings Dialog.
Short
Eye finder looks at 100,000 clock cycles on each channel
to determine the suggested sampling positions. This setting
requires frequent transitions on all channels.
Medium
Eye finder looks at 500,000 clock cycles on each channel
to determine the suggested sampling positions. Use this for
channels that transition at a normal rate.
Long
Eye finder looks at 2.5 million clock cycles on each
channel to determine the suggested sampling positions.
Use this setting if some channels have sporadic transitions.
Some things to consider when selecting among the eye finder
advanced settings are:
•
Upper address bits that don't transition as frequently as lower address bits.
•
Data buses that are driven by different circuitry at different times.
When different channels require different settings, you can run eye
finder on channel subsets to avoid using the Long setting on a large
number of channels.
131
Chapter 3: Reference
Importing Netlist and ASCII Files
Eye Finder Option, Results Tab
The Eye Finder Results display is a digital "eye" diagram in that it
represents many samples of data captured in relation to the sampling
clock. The transitioning edges measured before and after the sampling
clock result in a picture that is eye-shaped.
You should have already specified the logic analyzer threshold voltage,
but you may adjust it to maximize the width of the measured stable
regions.
Eye finder measures the location of the stable region boundaries and
places the logic analyzer's sampling position in the center of the stable
region.
File menu
Lets you save/load eye finder data.
EyeFinder menu Lets you run eye finder, choose the run mode, and access
the “Eye Finder Advanced Settings Dialog” on page 131.
132
Chapter 3: Reference
Importing Netlist and ASCII Files
Results menu
Let you expand/collapse the signals in a label, set the bus
view, set the sampling positions to the suggested sampling
positions, and remove all eye finder data.
Label buttons
Let you expand/collapse the signals in a label, set the bus
view, choose the suggested sampling position, and show
message or time stamp information.
Display Area
Shows:
•
Transitioning (dark) and stable (light) regions on the signals.
•
Suggested sampling positions (green triangles).
•
The current sampling positions (blue lines).
•
Informational message icons
. You can move the mouse pointer over the
icon to cause the message to pop up.
•
Time stamp icons
. You can move the mouse pointer over the icon to
see when the last eye finder measurement was run.
To give you more information about the signals, the display
covers +/-5 ns even though the sampling position may only
be set to +/-3.25 ns.
Sampling
Position
See Also
Lets you adjust the sampling position. You can also drag the
sampling position bar to a new location.
“Understanding State Mode Sampling Positions” on page 206
“How Selected/Suggested Positions Behave” on page 133
“Eye Finder Run Messages” on page 134
“Eye Finder Info Messages” on page 137
“Eye Finder Load/Save Messages” on page 139
“Eye Finder Option, Setup Tab” on page 130
“To manually adjust sampling positions” on page 49
How Selected/Suggested Positions Behave. The eye finder's
selected and suggested sampling positions behave as follows:
133
Chapter 3: Reference
Importing Netlist and ASCII Files
How the Selected Position Behaves
1. When eye finder is enabled, the selected position (blue line) is set based
on the manual setup/hold value.
2. Whenever the selected position is moved, the manual setup/hold value is
also updated. They always track each other.
3. When the manual setup/hold is enabled again, the position changes made
while eye finder was enabled can be kept or discarded.
NOTE:
If eye finder changes are discarded, this includes any setup/hold settings
loaded from a configuration file while eye finder was enabled.
4. The selected position is "snapped" to the suggested position (green
triangle) each time the channel is measured.
How the Suggested Position Behaves
1. There is only a suggested position (green triangle) on channels that have
been measured.
2. The suggested position is always in the center of the stable region closest
to the selected position (blue line).
3. If the selected position is moved to a different stable region, the suggested
position "hops" to the center of that region.
4. If a stable region is open-ended, the suggested position is placed 1.25 ns
from the closed end (the visible boundary). If more than 1 clock edge is
active, the suggested position is placed 1.5 ns from the closed end.
Eye Finder Run Messages. These messages can appear in the status
area after you run a eye finder measurement.
"XX% complete"
The indicated percentage of the channels selected for the current eye
finder measurement are completed.
"Cannot run the Eye Finder at this time."
A logic analyzer measurement is currently running. Stop the logic
analyzer or wait for it to complete before running eye finder.
134
Chapter 3: Reference
Importing Netlist and ASCII Files
An eye finder measurement is currently running. Stop the eye finder
or wait for it to complete before running the eye finder.
The eye finder is already running on the other machine defined for
this analyzer. Eye finder cannot run on both machines at the same
time.
"Cannot run the Logic Analyzer at this time."
An eye finder measurement is currently running. Stop the eye finder
measurement or wait for it to complete before running the logic
analyzer measurement.
Eye finder uses the same hardware as an ordinary logic analyzer
measurement uses. Therefore, they cannot be performed at the same
time.
"Characterizer cannot be loaded"
This is an internal error. It means that a software unit responsible for
measuring one of the channels did not fit the hardware setup and the
other characterizers which were already loaded. Try running eye
finder on just one label (or one channel) at a time to attempt to clear it
up. Contact support if this persists.
"Complete: DATE"
The measurement completed successfully on the date and time given.
This time may also be accessed by expanding a label in the results
display and selecting the clock icon or selecting "Show Time Stamp" on
the popup menu raised by selecting the channel name in the results
display.
"Eye Finder only operates when the analyzer clocking is master clock (slave and/or
demux clocking are not supported)."
See the "Clock Setup" section of the "Sampling" tab in the analyzer
setup window.
"Eye Finder only operates when the analyzer is setup for state analysis."
See the "Pod Assignment" dialog accessed from the "Format" tab in the
analyzer setup window.
135
Chapter 3: Reference
Importing Netlist and ASCII Files
"From Eye Finder: After hardware calibration, the sampling positions for the
following channels may have shifted out of the selected stable region by the
amount shown: CHANNEL: AMOUNT ps ... (NNN more)"
Each time a measurement is started, the hardware is re-calibrated. The
new calibration values are checked against those used when the eye
finder measurements were taken. This message indicates that the
sampling positions for the given CHANNELs may have drifted out of
the stable region. The measurement is taken anyway, but you may want
to treat the results with caution and run eye finder again (or manually
adjust the sample position away from the indicated unstable region).
"Hardware calibration failed"
Something isn't as expected about the hardware and/or the cables and
connections between boards. To get detailed messages, start an
ordinary run or run PV.
"Measurement Canceled"
The measurement was stopped. No change was made to the results
displayed. A run is stopped by user request or when the Sampling
Positions dialog is closed or iconified.
"No labels defined with channels for the analyzer."
Define one or more labels with channels in the Format tab of the
analyzer's Setup window.
No labels are defined for the analyzer. Eye finder cannot be run until
one or more labels are defined with one or more channels assigned.
"No labels or channels selected for running Eye Finder."
Select one or more labels in the "Eye Finder Setup" tab.
All labels defined for the analyzer are listed in the Eye Finder Setup
page. None are currently selected (selected labels are highlighted).
Select one or more labels for measurement by eye finder.
"Repetitive runs stopped"
The measurement was stopped. Data from the last measurement which
completed fully are retained. Repetitive runs are stopped by user
136
Chapter 3: Reference
Importing Netlist and ASCII Files
request or when the Sampling Positions dialog is closed or iconified.
"Timeout: < N K clocks in 5 sec"
Eye finder requires stimulus at a minimum rate to perform its
measurements. Too few state clocks were seen in the time allotted.
Check clock inputs, clock definition, threshold voltage settings, and the
operation of the device under test.
Eye Finder Info Messages. These messages appear in the Eye
Finder Results tab after an eye finder measurement is run.
"Example measurement for demo. Results and settings will not be used for
analysis."
This channel was measured when "Use demo data (no probes
required)" was selected in the Settings tab. The data shown are typical
of eye finder operation, but the sample position setting shown is NOT
used. (The manual setting is still in use.)
"No activity present. Confirm connection, stimulus, and threshold."
This channel appears to be completely quiet.
•
Check the probe connection between the analyzer and the device under
test.
•
Check the threshold voltage setting in the Format tab.
•
Check that the device under test is turned on and is running the
appropriate diagnostic or other stimulus program.
If all these things are set up correctly, activity will be shown in the
Format tab.
"No stable regions. Is this the correct clock for this channel?"
Two common possibilities exist:
1. The signal on this channel is asynchronous to the clock defined for the
logic analyzer. If this is the case, there is no stable relationship between
the times when the signal switches and when the clock arrives.
If you expect the signal to be sampled synchronously you must redefine
the clock for this signal.
137
Chapter 3: Reference
Importing Netlist and ASCII Files
2. The stable region(s) are too small for eye finder to detect.
In this case you must resort to adjusting the sample position manually and
checking its validity by running an ordinary analyzer measurement to see if
the data values you expect are sampled. You can adjust the sample position
manually by selecting the arrow buttons or by dragging the blue sampling
position indicator in the display.
"Only a few transitions detected. Change stimulus or increase measurement
duration (Advanced Settings)."
The signal on this channel was observed to toggle fewer than 500
times. The characterization may be accepted as it stands or you may
wish to change the stimulus program or diagnostic in the device under
test to increase the toggle rate.
Another option is to select "Long" in the Eye Finder Advanced Settings
dialog (accessed from the "Advanced..." button on the Eye Finder
Setup tab or the "Advanced Settings..." menu pick under the EyeFinder
pull down menu). Using the "Long" setting won't necessarily make the
message go away, but it will ensure that eye finder has the opportunity
to observe a more significant number of transitions on the channel.
"Run Eye Finder to characterize this channel."
Select this channel (or a label that contains it) and run eye finder to
characterize the channel. Only channels selected for a measurement
are updated when the measurement finishes. Information about other
channels is not changed.
"See individual channels for message(s)"
There is a message for one or more the the channels assigned to this
bus label. Expand the label (using either the popup menu on the label
in the display or the "Results" pulldown menu), then scroll down to the
channel with a message icon and display its message (either by
selecting the yellow message icon or by using the popup menu).
"The stable region extends beyond the limits of the display."
This channel is active, but the signal does not switch within 5 nsec
before or after the clock. For example, this could occur if the
propagation delay in the device under test from clock to data is greater
138
Chapter 3: Reference
Importing Netlist and ASCII Files
than 5 nsec and the clock period is greater than 10 nsec (slower than
100 MHz).
Eye Finder Load/Save Messages. These messages can appear when
saving or loading eye finder data.
"... (at line XX in the file)"
Indicates where the error occurred in the file being read. Since eye
finder data files are ASCII text, you can use a text editor to examine
the file at the indicated line to determine how to repair the problem.
"Bad assignment for XXX"
The numerical value for the item XXX could not be read.
"Cannot load Eye Finder data for a different module type. Data in file is from a
OTHERMODEL. This module is a THISMODEL."
"Cannot load Eye Finder data for a module installed in different slots. Data in file is
from a module installed in slots B A C. This module is installed in slots E D F".
"Cannot load Eye Finder data from a different instrument. Data in file is from
OTHERINSTRUMENT. This instrument is THISINSTRUMENT."
"Channel name in the data file ("OTHERNAME") does not match the expected
channel name ("THISNAME")."
"Channel number in the data file (NN) does not match the expected channel number
(MM).",
"Did not find: XXX (Search began at line LLL)"
The eye finder data file has a nested block structure. The next block
(or item) to be read was XXX, but it was not found before the end of
the file. The line after the last item successfully read was LLL.
"Error in writing"
Disk is probably full.
139
Chapter 3: Reference
Importing Netlist and ASCII Files
"Failed to open file for reading/writing: NAME"
The selected file could not be opened. Check access and file
permissions.
"File NAME already exists. Overwrite?"
The selected file exists. Answering "Yes" will cause the existing
contents of the file to be replaced with current eye finder information.
"Invalid true/false flag"
The boolean value for the item could not be read. Boolean values start
with either a 't' or 'f' ('T' or 'F'are also accepted).
"Master slot in the data file ('j') does not match the expected master slot ('E')."
"The number of channels in the data file (NNN) does not match the number of
channels in the analyzer (XXX)."
Eye finder characterizes the unique combination of the instrument,
module, cabling, probes, and Device Under Test. Results are not
transferable from one situation to another.
"Unsupported revision level (AA.BB)"
Something else is probably wrong because there aren't any
unsupported versions.
Eye Finder Option, File Info Tab
When the Eye Finder option is selected, the File Info subtab lets you
save and load eye finder data.
140
Chapter 3: Reference
Importing Netlist and ASCII Files
File:
Name of the eye finder data file.
Created:
Date and time the eye finder data file was created.
Saved:
Date and time the eye finder data file was last saved.
You are notified if the eye finder data has changed since the
last time it was saved.
See Also
Load...
Loads eye finder data that was previously saved to a file.
Reload
Reloads eye finder data from the named file, deleting
unsaved changes.
Save
Saves current eye finder data to the named file.
Save As...
Saves current eye finder data to a new file.
“To automatically adjust sampling positions” on page 46
141
Chapter 3: Reference
Importing Netlist and ASCII Files
Manual Setup/Hold Option
When you select Manual Setup/Hold, the following options appear.
Label Selection
List
Lets you select the label whose setup/hold window will be
positioned.
All bits
Specifies that the setting is for all bits on the label.
Individual bits
Specifies that the setting is for a single bit on the label.
Bit:
When Individual bits is selected, this field identifies the
single bit.
Setup:
Specifies the front edge of the setup/hold window relative
to the sampling clock. Setup times are positive when the
position is before the sampling clock.
Hold:
Specifies the back edge of the setup/hold window relative
to the sampling clock. Hold times are positive when the
142
Chapter 3: Reference
Importing Netlist and ASCII Files
position is after the sampling clock.
See Also
“To manually adjust sampling positions” on page 49
143
Chapter 3: Reference
The Trigger Tab
The Trigger Tab
The Trigger tab is used to tell the analyzer when to capture data. The
key event is the trigger.
In the Agilent Technologies 16716A logic analyzer, you can insert
multiple trigger actions. When you insert multiple trigger actions, the
trigger marker in the display windows is placed on the first sample
whose evaluation caused a branch through an associated trigger action.
The Trigger tab has two main areas: On top, tabs of functions and
controls to build your trigger; and beneath the tabs, the current trigger
sequence.
Some controls are also located in the logic analyzer window's menu bar.
•
“Trigger Functions Subtab” on page 145
•
“Settings Subtab” on page 152
•
“Overview Subtab” on page 153
•
“Default Storing Subtab” on page 154
•
“Status Subtab” on page 155
144
Chapter 3: Reference
The Trigger Tab
•
See Also
“Save/Recall Subtab” on page 155
“Understanding Logic Analyzer Triggering” on page 190
“Setting Up Triggers and Running Measurements” on page 63
“Editing the Trigger Sequence” on page 71
Trigger Functions Subtab
Trigger functions provide a simple way to set up the analyzer to trigger
on common events and conditions. A library of functions is available for
both state and timing measurements.
NOTE:
See Also
Each trigger function requires at least one internal sequence level (see
page 71), and in some cases, multiple levels. The number of levels used by
each function can be seen by breaking down or expanding the trigger
function.
•
“General Timing Trigger Functions” on page 146
•
“General State Trigger Functions” on page 148
•
“Advanced Trigger Functions” on page 151
•
“Using Trigger Functions” on page 64
•
“Editing Advanced Trigger Functions” on page 76
145
Chapter 3: Reference
The Trigger Tab
General Timing Trigger Functions
The following general trigger functions are found in the Trigger
Functions tab when the logic analyzer is in the timing sampling mode.
You can edit most of the trigger functions to specify particular pattern
and edge events.
You can break down a trigger function to see how many advanced
sequence levels are used.
•
Find pattern
Becomes true when a pattern occurs one time.
•
Find edge
Becomes true when the specified edge occurs in one sample.
•
Find edge AND pattern
Becomes true when the specified pattern and the specified edge occurs in
one sample.
•
Find width violation on a pattern/pulse
Becomes true when the width of a pattern violates minimum and
maximum width specifications.
•
Find Nth occurrence of an edge
Becomes true when the specified edge occurs in the specified number of
samples.
•
Find pattern present/absent for > duration
Becomes true when the specified pattern is present or absent for greater
than the amount of time specified.
•
Find pattern present/absent for < duration
Becomes true when the specified pattern is present or absent for less than
the amount of time specified.
•
Run until user stop
Sets up to never trigger. You must select the stop button to view the
captured data.
146
Chapter 3: Reference
The Trigger Tab
•
Find 2 edges too close together
Becomes true when the second specified edge occurs within a specified
time after the first specified edge.
•
Find 2 edges too far apart
Becomes true when the second specified edge does not occur within a
specified time after the first specified edge.
•
Find pattern occurring too soon after edge
Becomes true when a specified pattern occurs within a specified time after
the first specified edge.
•
Find pattern occurring too late after edge
Becomes true when a specified pattern does not occur within a specified
time after the first specified edge.
•
Find glitch
Becomes true when the specified glitch occurs in one sample.
•
Wait t seconds
Becomes true when the specified amount of time has expired.
•
Wait for arm in
When the logic analyzer is armed by another instrument (as specified in
the Intermodule window), this trigger function becomes true when the
arm signal is received.
•
Wait for second analyzer to trigger
When the logic analyzer's pods are assigned to two analyzers, this trigger
function becomes true when the other analyzer triggers.
•
Wait for flag
Becomes true when the specified flag has the specified value. This trigger
function tests for a flag event.
•
Set/clear/pulse flag
Becomes true on any sample and sets, clears, pulse sets, or pulse clears
the specified flag. This trigger function inserts a flag action.
147
Chapter 3: Reference
The Trigger Tab
•
OR Trigger
When the logic analyzer is armed by another instrument (as specified in
the Intermodule window), this trigger function becomes true when a
pattern occurs a specified number of times OR when the arm signal is
received.
See Also
“To specify a label pattern event” on page 65
“To specify a label edge event” on page 65
“To break down a trigger function” on page 66
“To cross-trigger with another instrument” on page 106
“To cross-trigger between two analyzers” on page 105
General State Trigger Functions
The following general trigger functions are found in the Trigger
Functions tab when the logic analyzer is in the state sampling mode.
You can edit most of the trigger functions to specify particular pattern
events.
You can break down a trigger function to see how many advanced
sequence levels are used.
•
Find pattern n times
Becomes true when the specified pattern occurs in the specified number
of samples (eventually).
•
Store range until pattern occurs
Becomes true when the specified pattern occurs in the specified number
of samples (eventually) and only stores samples in the specified range
until then.
•
Store pattern2 until pattern1 occurs
Becomes true when the first specified pattern occurs in the specified
number of samples (eventually) and only stores samples with the second
specified pattern until then.
•
148
While storing pattern2, find pattern1
Chapter 3: Reference
The Trigger Tab
This trigger function has been replaced by the "Store range until pattern
occurs" and "Store pattern2 until pattern1 occurs" trigger functions.
•
Store nothing until pattern occurs
Becomes true when the specified pattern occurs one time and doesn't
store any samples until then.
•
Run until user stop
Sets up to never trigger. You must select the stop button to view the
captured data.
•
Find pattern2 occurring immediately after pattern1
Becomes true when the second specified pattern occurs in the sample
immediately after a sample in which the first specified pattern occurs.
•
Find pattern1 eventually followed by pattern2
Becomes true when the second specified pattern occurs in a sample
(eventually) after a sample in which the first specified pattern occurs.
•
Find pattern2 occurring too soon after pattern1
Becomes true when the second specified pattern occurs within a specified
time after the first specified pattern.
•
Find pattern2 occurring too late after pattern1
Becomes true when the second specified pattern does not occur within a
specified time after the first specified pattern.
•
Find too few states between pattern1 and pattern2
Becomes true when the second specified pattern occurs within a specified
number of samples (states) after the first specified pattern.
•
Find too many states between pattern1 and pattern2
Becomes true when the second specified pattern does not occur within a
specified number of samples (states) after the first specified pattern.
•
Find n-bit serial pattern
Becomes true when a specified serial pattern of N bits is found on the
analyzed line.
•
Find pattern n consecutive times
149
Chapter 3: Reference
The Trigger Tab
Becomes true when the specified pattern occurs in the specified number
of samples consecutively.
•
Find pattern2 n times after pattern1, before pattern3 occurs
Becomes true when the second specified pattern occurs in a specified
number of samples after the the first specified pattern but without the
third specified pattern occurring anywhere in between.
•
Store n samples
Becomes true when the specified number of samples are stored.
•
Wait n external clock states
Becomes true when the specified number of external clocks have
occurred.
•
Wait for arm in
When the logic analyzer is armed by another instrument (as specified in
the Intermodule window), this trigger function becomes true when the
arm signal is received.
•
Wait for second analyzer to trigger
When the logic analyzer's pods are assigned to two analyzers, this trigger
function becomes true when the other analyzer triggers.
•
Wait for flag
Becomes true when the specified flag has the specified value. This trigger
function tests for a flag event.
•
Set/clear/pulse flag
Becomes true on any sample and sets, clears, pulse sets, or pulse clears
the specified flag. This trigger function inserts a flag action.
•
OR Trigger
When the logic analyzer is armed by another instrument (as specified in
the Intermodule window), this trigger function becomes true when a
pattern occurs a specified number of times OR when the arm signal is
received.
See Also
“To specify a label pattern event” on page 65
150
Chapter 3: Reference
The Trigger Tab
“To break down a trigger function” on page 66
“To cross-trigger with another instrument” on page 106
“To cross-trigger between two analyzers” on page 105
Advanced Trigger Functions
The advanced trigger functions let you create a custom trigger
sequence level using events, comparison functions, and up to 4
branches.
The advanced trigger functions are available in both the timing and
state sampling modes and are the same, except that in the timing
mode, you can specify edges on labels and event durations.
The types of events include labels, timers, flags, and counters.
•
Advanced - If/then
This trigger function has only one branch. If the events in the "If" event list
are true, it executes the actions after "then".
•
Advanced - 2-way branch
This trigger function has two branches, of the form "If - then; else if - then".
For each sample, the events in the first "If" branch are checked. If all
events are true, the "then" portion is executed. If they are not true, the
events in the "else if" branch are checked. If the "else if" events are true, its
"then" portion is executed.
If neither branch is true, the logic analyzer remains in this sequence level
and repeats the comparison with the next sample.
•
Advanced - 3-way branch
This function has three branches, of the form
If (events1)
then (actions1)
Else if (events2)
then (actions2)
Else if (events3)
then (action3)
The logic analyzer evaluates each sample against the clauses in the order
151
Chapter 3: Reference
The Trigger Tab
they are specified. The logic analyzer executes the set of actions in the
"then" clause associated with the first listed "if" or "else if" clause that
becomes true.
•
Advanced - 4-way branch
Like the 3-way branch, but with 3 "Else if" clauses.
•
Advanced - pattern1 AND pattern2
Searches for two different patterns occurring in the same sample. If you
set it to look for more than 1 occurrence, you can specify whether
occurrences are consecutive or not. You can also add other events,
including labels, to be searched for.
•
Advanced - pattern1 OR pattern2
Finds either pattern1 or pattern2 or both in a sample. If you set it to look
for more than 1 occurrence, you can specify whether the occurrences are
consecutive or not. You can also add other events, including labels, to be
searched for.
See Also
“Editing Advanced Trigger Functions” on page 76 for more information on
using the advanced trigger functions.
Settings Subtab
Sample Period
Acquisition
Depth
152
(Timing mode only). Lets you specify how often the the
logic analyzer samples signals from the device under test.
Lets you use a smaller portion of the acquisition memory
and speed up processing of the captured data.
Chapter 3: Reference
The Trigger Tab
Trigger Position Lets you specify where the sample that triggered the
analyzer should appear among all the other samples that
are stored in acquisition memory.
Count
(State mode only). Lets you save time or state count
information with the captured data samples.
Arm out from:
When two logic analyzers are turned on, this option lets
you choose which of them (or both) should drive the arm
signal.
Intermodule
Control
See Also
Lets you configure multiple instrument measurements and
adjust the order of trigger arming, as well as compensate
for timing skew between modules.
“To specify the sample period” on page 42
“To set acquisition memory depth” on page 52
“To specify the trigger position” on page 52
“To count states or time” on page 69
“To cross-trigger with another instrument” on page 106
Overview Subtab
This tab gives a picture of the trigger sequence.
See Also
“To view a picture of the trigger sequence” on page 76
153
Chapter 3: Reference
The Trigger Tab
Default Storing Subtab
Store by default Lets you specify that Anything, Nothing, Custom, or
selected Transitions events be stored by default.
At start of
acquisition
Event
specification
list
Group events
See Also
Lets you choose whether default storing is initially On or
Off.
When you choose Custom events, this area lets you specify
the custom events.
When you choose Custom events, you can group events in
the list to specify the evaluation order.
“To Specify Default Storing” on page 70
“To specify whether default storing is initially on or off” on page 71
“To group events” on page 82
“To specify a label pattern event” on page 65
154
Chapter 3: Reference
The Trigger Tab
Status Subtab
The Status subtab shows you the sequence level that is evaluating
captured data, occurrence and global counter values, and flag values.
See Also
“To view the trigger status” on page 86
Save/Recall Subtab
The Save/Recall subtab lets you save trigger setups within a session.
The Agilent Technologies 16716A logic analyzer provides memory
locations to store up to 15 trigger sequences for both state and timing
sampling modes. Five of the 15 memory positions are reserved for the 5
most recent runs.
When you exit your Agilent Technologies 16700 session, the trigger
save/recall list is cleared. However, the trigger save/recall list can be
saved as part of a configuration file.
155
Chapter 3: Reference
The Trigger Tab
You can also save trigger sequences outside of configuration files by
creating trigger function libraries.
See Also
“Saving/Recalling Trigger Setups” on page 83
“Saving and Loading Logic Analyzer Configurations” on page 109
“To create a trigger function library” on page 67
156
Chapter 3: Reference
The Symbols Tab
The Symbols Tab
The Symbols tab lets you load symbol files or define your own symbols.
Symbols are names for particular data values on a label.
Two kinds of symbols are available:
Multiple files
•
Object File Symbols. These are symbols from your source code and
symbols generated by your compiler.
•
User-Defined Symbols. These are symbols you create.
•
“Symbols Selector Dialog” on page 159
•
“Symbol File Formats” on page 161
•
“General-Purpose ASCII (GPA) Symbol File Format” on page 162
You can load the same symbol file into several different analyzers, and
you can load multiple symbol files into one analyzer. Symbols from all
the files you load will appear together in the object file symbol selector
that you use to set up resource terms.
157
Chapter 3: Reference
The Symbols Tab
Object file versions
During the load process, a symbol database file with a .ns extension
will be created by the system. One .ns database file will be created for
each symbol file you load. Once the .ns file is created, the Symbol
Utility will use this file as its working symbol database. The next time
you need to load symbols into the system, you can load the .ns file
explicitly, by placing the .ns file name in the Load This Object/Symbol
File For Label field.
If you load an object file that has been loaded previously, the system
will compare the time stamps on the .ns file and the object file. If the
object file is newer, the .ns file will be created. If the object file has not
been updated since it was last loaded, the existing .ns file will be used.
See Also
“Using Symbols” on page 94
158
Chapter 3: Reference
The Symbols Tab
Symbols Selector Dialog
Search Pattern: Lets you enter partial symbol names and the asterisk
wildcard character (*) to limit the symbols to choose from
(see “Search Pattern” on page 160). Use the Recall button
to select from previous search patterns.
Find Symbols of
Type
Lets you limit the types of symbols to choose from.
Matching
Symbols
Lists the symbols that match the search pattern. You
choose a symbol from this list.
159
Chapter 3: Reference
The Symbols Tab
Offset By
Lets you add an offset value to the starting point of a
symbol. This can be useful when compensating for
microprocessor prefetches (see “Offset By Option” on
page 160).
Align to
Lets you mask the lower order bits of a symbol's value. This
can be useful for triggering on odd byte boundaries (see
“Align to x Byte Option” on page 161).
Beginning/End/
Range
When a symbol represents a range of addresses, you can
choose the beginning address of the range, the end address
of the range, or the whole range.
See Also
“To enter symbolic label values” on page 98
Search Pattern
Use this field to locate particular symbols in the symbol databases. To
use this field, enter the name of a file or symbol. The system searches
the symbol database for symbols that match this name. Symbols that
match appear in the list of Matching Symbols. You can also use
wildcard characters to find symbols.
Asterisk wildcard (*)
The asterisk wildcard represents "any characters." When you perform a
search on the symbol database using just the asterisk, you will see a list
of all symbols contained in the database. The asterisk can also be
added to a search word to find all symbols that begin or end with the
same letters. For example, to find all of the symbols that begin with the
letters "st", select the Search Pattern field and enter "st*".
Offset By Option
The Offset By option allows you to add an offset value to the starting
point of the symbol that you want to use. You might do this in order to
trigger on a point in a function that is beyond the preamble of the
function, or to trigger on a point that is past the prefetch depth of the
processor. Setting an offset helps to avoid false triggers in these
situations. The offset specified in the Offset By field is applied before
the address masking is done by the "Align to x Byte" option.
Example
An 80386 processor has a prefetch depth of 16 bytes. Assume functions
160
Chapter 3: Reference
The Symbols Tab
func1 and func2 are adjacent to each other in physical memory, with
func2 following func1. In order to trigger on func2 without getting a
false trigger from a prefetch beyond the end of func1, you need to add
an offset value to your label value. The offset value must be equal to or
greater than the prefetch depth of the processor. In this case, you
would add an offset of 16 bytes to your label value. You would set the
value of the "Offset By" field to 10 hex. Now, when you specify func2 as
your label value, the logic analyzer will trigger on address func2+10.
Align to x Byte Option
Most processors do not fetch instructions from memory on byte
boundaries. In order to trigger a logic analyzer on a symbol at an oddnumbered address, the address must be masked off. The "Align to x
Byte" option allows you to mask off an address.
Example
Assume the symbol "main" occurs at address 100F. The processor being
probed is a 68040, which fetches instructions on long-word (4-byte)
boundaries. In order to trigger on address 100F, the Align to x Byte
option sets the two least-significant address bits to "don't cares". This
qualifies any address from 100C through 100F.
Symbol File Formats
The logic analysis system can read symbol files in the following
formats:
•
OMF96
•
OMFx86
•
IEEE-695
•
ELF/DWARF
•
ELF/stabs
•
TI COFF
For ELF/DWARF1, ELF/stabs, and ELF/stabs/Mdebug files, C++
symbols are demangled so that they can be displayed in the original
161
Chapter 3: Reference
The Symbols Tab
C++ notation. To improve performance for these ELF symbol files, type
information is not associated with variables. Hence, some variables
(typically a few local static variables) may not have the proper size
associated with them. They may show a size of 1 byte and not the
correct size of 4 bytes or even more. All other information function
ranges, line numbers, global variables and filenames will be accurate.
These behaviors may be changed by creating a readers.ini (see
page 100) file.
See Also
“To load object file symbols” on page 95
“To create an ASCII symbol file” on page 99
“To create a readers.ini file” on page 100
General-Purpose ASCII (GPA) Symbol File
Format
General-purpose ASCII (GPA) format files are loaded into a logic
analyzer just like other object files.
If your compiler does not produce object files in a supported format, or
if you want to define symbols that are not included in the object file,
you can create an ASCII format symbol file.
Typically, ASCII format symbol files are created using text processing
tools that convert the symbol table information from a compiler or
linker map output file.
Different types of symbols are defined in different records in the GPA
file. Record headers are enclosed in square brackets, for example,
[VARIABLES]. For a summary of GPA file records and associated
symbol definition syntax, refer to the “GPA Record Format Summary”
on page 163 that follows.
Each entry in the symbol file must consist of a symbol name followed
by an address or address range.
While symbol names can be longer, the logic analyzer only uses the first
16 characters.
162
Chapter 3: Reference
The Symbols Tab
The address or address range must be a hexadecimal number. It must
appear on the same line as the symbol name, and it must be separated
from the symbol name by one or more blank spaces or tabs. Address
ranges must be in the following format:
beginning address..ending address
The following example defines two symbols that correspond to address
ranges and one symbol that corresponds to a single address.
main
test
var1
00001000..00001009
00001010..0000101F
00001E22
#this is a variable
For more detailed descriptions of GPA file records and associated
symbol definition syntax, refer to the following topics:
•
“SECTIONS” on page 164
•
“FUNCTIONS” on page 165
•
“VARIABLES” on page 166
•
“SOURCE LINES” on page 166
•
“START ADDRESS” on page 167
•
“Comments” on page 167
GPA Record Format Summary
Format
[SECTIONS]
section_name
start..end
attribute
[FUNCTIONS]
func_name start..end
[VARIABLES]
var_name
start [size]
var_name
start..end
[SOURCE LINES]
File: file_name
line# address
163
Chapter 3: Reference
The Symbols Tab
[START ADDRESS]
address
#comment text
Lines without a preceding header are assumed to be symbol definitions
in one of the [VARIABLES] formats.
Example
This is an example GPA file that contains several different kinds of
records.
[SECTIONS]
prog
00001000..0000101F
data
40002000..40009FFF
common
FFFF0000..FFFF1000
[FUNCTIONS]
main
00001000..00001009
test
00001010..0000101F
[VARIABLES]
total
40002000
value
40008000
4
4
[SOURCE LINES]
File: main.c
10
00001000
11
00001002
14
0000100A
22
0000101E
File: test.c
5
00001010
7
00001012
11
0000101A
SECTIONS
Use SECTIONS to define symbols for regions of memory, such as
sections, segments, or classes.
NOTE:
To enable section relocation, section definitions must appear before any other
definitions in the file.
164
Chapter 3: Reference
The Symbols Tab
NOTE:
If you use section definitions in a GPA symbol file, any subsequent function or
variable definitions must be within the address ranges of one of the defined
sections. Functions and variables that are not within the range are ignored.
Format
[SECTIONS]
section_name
start..end
attribute
section_name A symbol representing the name of the section.
start
The first address of the section, in hexadecimal.
end
The last address of the section, in hexadecimal.
attribute
(optional) Attribute may be one of the following:
NORMAL (default) - The section is a normal, relocatable section, such as
code or data.
NONRELOC - The section contains variables or code that cannot be
relocated. In other words, this is an absolute segment.
Example
[SECTIONS]
prog
data
display_io
00001000..00001FFF
00002000..00003FFF
00008000..0000801F
NONRELOC
FUNCTIONS
Use FUNCTIONS to define symbols for program functions, procedures
or subroutines.
Format
[FUNCTIONS]
func_name start..end
func_name
A symbol representing the function name.
start
The first address of the function, in hexadecimal.
end
The last address of the function, in hexadecimal.
Example
[FUNCTIONS]
main
00001000..00001009
test
00001010..0000101F
165
Chapter 3: Reference
The Symbols Tab
VARIABLES
You can specify symbols for variables using:
•
The address of the variable.
•
The address and the size of the variable.
•
The range of addresses occupied by the variable.
If you specify only the address of a variable, the size is assumed to be 1
byte.
Format
[VARIABLES]
var_name
start [size]
var_name
start..end
var_name
A symbol representing the variable name.
start
The first address of the variable, in hexadecimal.
end
The last address of the variable, in hexadecimal.
size
(optional) The size of the variable, in bytes, in decimal.
Example
[VARIABLES]
subtotal
total
data_array
status_char
40002000
4
40002004
4
40003000..4000302F
40002345
SOURCE LINES
Use SOURCE LINES to associate addresses with lines in your source
files.
Format
[SOURCE LINES]
File: file_name
line# address
file_name
The name of a file.
line#
The number of a line in the file, in decimal.
address
The address of the source line, in hexadecimal.
166
Chapter 3: Reference
The Symbols Tab
Example
[SOURCE LINES]
File: main.c
10
00001000
11
00001002
14
0000100A
22
0000101E
See Also
Using the Source Viewer (see the Listing Display Tool help volume)
START ADDRESS
Format
[START ADDRESS]
address
address
The address of the program entry point, in hexadecimal.
Example
[START ADDRESS]
00001000
Comments
Use the # character to include comments in a file. Any text following
the # character is ignored. You can put comments on a line alone or on
the same line following s symbol entry.
Format
#comment text
Example
#This is a comment
167
Chapter 3: Reference
Error Messages
Error Messages
•
“Analyzer armed from another module contains no "Arm in from IMB"
event” on page 183
•
“Branch expression is too complex” on page 169
•
“Cannot specify range on label with clock bits that span pod pairs” on
page 174
•
“Counter value checked as an event, but no increment action specified” on
page 175
•
“Goto action specifies an undefined level” on page 175
•
“Hardware Initialization Failed” on page 176
•
“Maximum of 32 Channels Per Label” on page 175
•
“Must assign another pod pair to specify actions for flags” on page 176
•
“Must assign Pod 1 on the master card to specify actions for flags” on
page 169
•
“No more Edge/Glitch resources available for this pod pair” on page 176
•
“No more Pattern resources available for this pod pair” on page 177
•
“No Trigger action found in the trace specification” on page 177
•
“Slow or Missing Clock” on page 178
•
“Timer value checked as an event, but no start action specified” on
page 178
•
“Trigger function initialization failure” on page 179
•
“Trigger inhibited during timing prestore” on page 180
•
“Trigger Specification is too complex” on page 180
•
“Waiting for Trigger” on page 182
168
Chapter 3: Reference
Error Messages
Must assign Pod 1 on the master card to specify
actions for flags
When using a 16760A analyzer in 200Mb/s state mode, Pod 1 on the
master card must be assigned in order to add actions for the flags in a
branch action list. To assign the pod, go to the format tab and select the
"Pod Assignment" button. Drag Pod 1 of the master card from the
unassigned pods list to the analyzer assigned list.
Branch expression is too complex
The "Branch expression is too complex" message means that the event
list expression for the indicated branch contains more event terms to
logically combine than the hardware is capable of combining on a single
branch.
Other branches in the sequence may also be too complex. The trigger
sequence compiler stops compiling at the first convenient place after it
encounters a fatal error.
Because the trigger sequence compiler tries to optimize the event list
expression to best fit the capabilities of the hardware, a precise
description of the event list limits cannot be easily enumerated, but
listed below are some general guidelines for all acquisition modes and
some specific suggestions for particular modes.
General Guidelines
•
Labels that span multiple pod pairs (split labels) greatly increases the
compiled hardware expression complexity as compared with labels that
are entirely contained within a single pod pair.
Whenever possible try to arrange the probing such that labels to not span
pod pairs. This is the single most effective way to reduce the complexity
required to implement the event list expression.
169
Chapter 3: Reference
Error Messages
NOTE:
For labels that do span pod pairs, the complexity can be reduced to the same
as that of the non-split label case if all bits in the label on all but one pod pair
can be set to Xs in the event list expression for the measurement.
For example, if label ADDR has its 16 most significant bits on pod A3 and
16 least significant bits on pod A2 (spanning pod pairs A4/A3 and A2/A1),
the complexity of the compiled expression will be reduced if all 16 most
significant bits or all 16 least significant bits are set to Xs in the pattern
event.
•
Inequality compares (<,<=,>,>=) of split labels increases the expression
complexity compared to equality (=,!=) compares of split labels. There is
no difference in complexity for non-split labels.
•
Ranges are implemented as two inequality compares which doubles the
required complexity for non-split labels but compounds the complexity to
an even greater extent for ranges on split patterns.
•
Equivalent event list expressions compile to a MUCH greater hardware
complexity in 333/400/800/1250 Mb/s state modes than in 167/200 Mb/s
state mode. This is due to the way the hardware implements these faster
state modes. The hardware parallelizes the data to allow the internal
sequencer to still run at <= 167/200 Mb/s. However, this requires the
trigger compiler to allocate additional sequence levels, branches, and
pattern resources and combine them in complex expressions to deparallelize the trigger expression. Using split labels in these faster modes
further multiplies the complexity of these compiler generated expressions.
•
The trigger compiler first expands all expression lists to sum-of-products
form (for example, A(B+C) is expanded to AB+AC). The trigger compiler
then does rudimentary boolean reduction on the expanded expression.
However, the compiler does make some trade-offs between complete
reduction and compile speed. Manually expanding and reducing a complex
expression may help the trigger compiler to better fit the expression into
the hardware resources.
Specific Guidelines - 167/200 Mb/s State and all Timing Modes
•
Cannot OR more than 16 non-split pattern events if the pattern events are
all on the same pod pair.
•
Cannot OR more than 4 non-split pattern events if each pattern event is on
a different pod pair. You can, however, OR 4 patterns together on each of 4
different pod pairs to make a total of 16 patterns ORed across 4 pod pairs.
170
Chapter 3: Reference
Error Messages
•
Cannot AND more than 16 non-split pattern events if the pattern events
are all on the same pod pair.
•
Can AND up to 160 non-split pattern events if the pattern events are
evenly distributed across all 10 pod pairs on a 5 card set (16 pattern events
per pod pair).
Specific Guidelines - 333/400 Mb/s State Modes
•
Cannot AND or OR more than 8 non-split pattern events if the pattern
events are all on the same pod pair.
•
Cannot OR more than 4 non-split pattern events if each pattern event is on
a different pod pair. You can, however, OR 2 patterns together on each of 4
different pod pairs to make a total of 8 patterns ORed across 4 pod pairs.
•
Cannot AND or OR more than 4 non-split ranges if the pattern events are
all on the same pod pair.
•
Cannot AND or OR more than 2 split equality (=,!=) pattern events.
•
Cannot specify more than 1 split inequality (<,<=,>,>=) pattern events.
•
Cannot specify any range on a split label.
•
In 333/400 Mb/s State Mode, the trigger sequence compiler must combine
elements of the trigger events of the previous sequence level and the next
sequence with the current sequence level, thereby increasing the total
complexity of the current level. A sequence level that may compile fine
when its the only level in the sequence, may be too complex to compile
another level is inserted before or after it.
One possible workaround to this problem is to insert a simple "If anything"
sequence level in between two complex levels. The disadvantage to this
approach, of course, is that the trigger sequence will miss one state in
between the two complex sequence levels.
If the following sequence does not compile:
1 If (complex event list)
occurs 1 time
then Goto Next
2 If (complex event list)
occurs 1 time
then Trigger and fill memory
This one may:
171
Chapter 3: Reference
Error Messages
1 If (complex event list)
occurs 1 time
then goto next
2 If anything
occurs 1 time
then Goto Next
3 If (complex event list)
occurs 1 time
then Trigger and fill memory
•
In 333/400 Mb/s State Modes, the trigger sequence compiler must always
add some additional complexity to the compiled expression for the first
sequence level that is not needed in subsequent sequence levels.
Additional complexity is also required in the first sequence level for the
following 2 conditions:
a. When using the "Find pattern1, or reset on pattern2" trigger function in
333/400 Mb/s State Modes, the event list of the first sequence level
must be combined with reset branch of each subsequent sequence
level by the trigger compiler in order to evaluate the parallelized
samples.
b. When using double edge clocking mode (J Clk rising and falling edges)
in 333 Mb/s State Mode, an additional pattern resource is allocated and
combined with the event list in the first sequence level by the trigger
compiler to prevent triggering on an initial garbage state.
Inserting an "If anything" state as the first state can simplify the complexity
of the compiled event list in the first sequence level and subsequent "If/
Else" sequence levels. The disadvantage is that the sequence will then miss
the first state after the reset condition is met in an "If/Else" sequence level.
If the following sequence does not compile:
1 If (complex event list)
occurs 1 time
then Goto Next
3 If (complex event list)
then Trigger and fill memory
Else if (complex event list)
then Goto 1
This one may:
1 If anything
occurs 1 time
then Goto Next
2 If (complex event list)
occurs 1 time
then Goto Next
3 If (complex event list)
then Trigger and fill memory
Else if (complex event list)
then Goto 1
172
Chapter 3: Reference
Error Messages
Specific Guidelines - 800 Mb/s State Mode
•
Labels that span pods (split labels) require more combiner resources than
labels with bits that all belong to a single pod. Whenever possible, define
labels that do not span pods. In some cases, the compiler will be able to
combine 2 non-split labels that are ANDed together even though it fails to
compile a pattern on a single label that spans pods.
•
Cannot specify more than 4 patterns or 2 ranges per pod.
Non-split patterns may use operations: =, !=, <, <=,>, >=, In range, Not
in range.
Split patterns may only use operations: =, !=.
Patterns on labels with re-ordered bits may only use operations: =, !=
(same as Timing and 200 and 400 Mb/s modes).
•
Restrictions on the way event resources can be used and combined vary
with the type of trigger function selected.
The "Find Pattern" and "Find n Patterns in Immediate Sequence"
trigger functions can use all of the available pattern events in any
sequence level. However, some combinations (ANDing and ORing) of
pattern events, while not exceeding the maximum number of pattern
resources, can use too many combiner resources. Any combination of 3
of the flag events and "Wait for arm in from IMB" event can also be
combined with the pattern events in each sequence level.
When using the "Find n Patterns in Eventual Sequence" trigger
functions, only non-split labels may be used and only one pattern event
per sequence level may be used. The interface allows insertion of other
events in each sequence level to allow ANDing or ORing of the pattern
event with other event types (flags or arm in) or replacement of the
pattern event with another event type. However, if more than one
pattern event is specified the trigger compiler will be unable to compile
it.
•
Pattern events used in the Default Storing Control count against the
number of available resources.
Specific Guidelines - 1250 Mb/s State Mode
•
Labels that span pods (split labels) require more combiner resources than
labels with bits that all belong to a single pod. Whenever possible, define
labels that do not span pods. In some cases, the compiler will be able to
173
Chapter 3: Reference
Error Messages
combine 2 non-split labels that are ANDed together even though it fails to
compile a pattern on a single label that spans pods.
•
Cannot specify more than 3 patterns or 1 range per pod.
Non-split patterns may use operations: =, !=, <, <=,>, >=, In range, Not
in range.
Split patterns may only use operations: =, !=.
Patterns on labels with re-ordered bits may only use operations: =, !=
(same as Timing and 200 and 400 Mb/s modes).
•
Restrictions on the ways event resources can be used and combined vary
with the type of trigger function selected.
The "Find Pattern" and "Find 2 Patterns in Immediate Sequence"
trigger functions can use all of the available pattern events in any
sequence level. However, some combinations (ANDing and ORing) of
pattern events, while not exceeding the maximum number of pattern
resources, can use too many combiner resources. Any combination of 3
of the flag events and "Wait for arm in from IMB" event can also be
combined with the pattern events in each sequence level.
When using the "Find 2 Patterns in Eventual Sequence" trigger
function, only non-split labels may be used and only one pattern event
per sequence level may be used. The interface allows insertion of other
events in each sequence level to allow ANDing or ORing of the pattern
event with other event types (flags or arm in) or replacement of the
pattern event with another event type. However, if more than one
pattern event is specified the trigger compiler will be unable to compile
it.
•
Pattern events used in the Default Storing Control count against the
number of available resources.
Cannot specify range on label with clock bits
that span pod pairs
A label that contains clock bits being used as data bits, can only be
included in a range term if the clock bits are confined to a single pod
pair.
174
Chapter 3: Reference
Error Messages
Counter value checked as an event, but no
increment action specified
This warning occurs because you have used a counter in your trigger
sequence, but do not have Counter Increment as an action. You do not
need to increment the counter in the same sequence level. The counter
event will still function, but will not change value. The default value for
both counters is 0.
Goto action specifies an undefined level
The "undefined level" messages mean that the trigger sequence
contains goto statements that point to non-existent levels. This is
detected when the trigger sequence is evaluated. The logic analyzer
will not run if there are undefined levels, even if there is no possibility
of the goto sequence being called.
Possible Causes
•
The last sequence level calls "goto next"
To check this, select the Overview subtab under the Trigger tab.
To fix, find the "goto next" statement and change it to point to an
existing level.
Maximum of 32 Channels Per Label
The logic analyzer can only assign up to 32 channels for each label. If
you need more than 32 channels, assign them to two labels and use the
labels in conjunction.
175
Chapter 3: Reference
Error Messages
Hardware Initialization Failed
Please go to System Administration Tools and run the Self-Test Utility
(see page 107) on the logic analyzer. If you have failures, contact your
Agilent Technologies Sales Office for service or software upgrades.
Must assign another pod pair to specify actions
for flags
In state sampling mode, when there is only one pod pair assigned to an
analyzer, flags are not available. You must assign another pod pair to
the analyzer in order for flags to be available.
In the timing sampling mode, flags are are always available.
See Also
“To assign pods to one or two analyzers” on page 56
No more Edge/Glitch resources available for
this pod pair
This error occurs when you have used more than 2 edges or glitches
per pod pair in the trigger specification.
Possible Solutions
•
Phrase some of the edges as patterns.
For example, if you are looking for a rising edge on a read/write line, you
can check for R/W = 0 in one level followed by R/W = 1 in the next level.
•
Move some of the edges to another pod pair.
Even if a label spans pod pairs, only the edge resources of the pod pair the
specific channel is on are used.
176
Chapter 3: Reference
Error Messages
No more Pattern resources available for this
pod pair
This error occurs when you have used up all the pattern resources
available. Each pod pair has about 28 pattern resources. Some pattern
events use more than 1 resource.
Possible Solutions
•
Keep labels within a pod pair
If a label (bus) spans pod pairs (for example, pods 2 and 3) then when you
use the label in a trigger sequence it will use up at least one pattern
resource on both pod pairs. If you hook up your probes in such a way that
the signals are on a single pod pair, you can free up a pattern resource on
the other pod pair.
•
Move some labels to another pod pair
Each pod pair has its own set of pattern resources. Putting your two mostused labels on different pod pairs can improve your resource usage.
You may find more information in the Branch expression is too
complex (see page 169) error message help for the 800 and 1250 Mb/s
modes.
No Trigger action found in the trace
specification
This warning occurs when the trigger sequence you specified does not
have at least one trigger and fill memory or trigger and goto action.
The analyzer will still acquire data, but you will need to manually stop
it. For example, when you use the Run until user stop function, to
avoid popping up the status window, set the popup level to errors only.
177
Chapter 3: Reference
Error Messages
Slow or Missing Clock
The message "Slow or Missing Clock" only appears in state
measurements. However, if you have another instrument armed by the
state analyzer, a slow or missing clock on the state analyzer will prevent
the other instrument from triggering also.
Possible Causes
•
Target system is not running properly
Check that the system is running properly. The logic analyzer and other
probing fixtures such as pin extenders can place too much capacitive load
on a system.
•
Incorrect clock specification
Make sure the device under test clock matches the clock specified under
Sampling.
Also check that the probe's clock channels are attached to the device
under test's clock lines either directly or through an analysis probe.
If you are using an analysis probe, the probe's User's Guide should show
the correct connections and settings.
•
Bad probe connection
Check that the probe is securely attached to the clock line and is receiving
a signal. The logic analyzer shows activity indicators under the Sampling
and Format tabs.
•
Incorrect signal level
The clock's threshold level is set by the pod threshold. For the logic
analyzer's J clock, check the pod threshold of pod 1 of the master card.
Timer value checked as an event, but no start
action specified
This warning occurs because you have used a timer in your trigger
sequence, but have not started it with either Start from reset or
178
Chapter 3: Reference
Error Messages
Resume in any action. You do not need to start the timer in the same
sequence level. The timer will still function if not started, but will not
change value.
Trigger function initialization failure
The "trigger function initialization failure" messages mean that you
tried to insert a trigger function which required a change to the trigger
function default state.
Possible Causes
•
Tried to insert "Wait for arm in" trigger function
A "Wait for arm in" trigger level causes the logic analyzer to wait for a
signal from another module or Port In. These signals are passed through
the Intermodule Bus. To prevent the logic analyzer from hanging, it must
be added to the Group Run Arming Tree. To do this, open the Intermodule
window by selecting the Intermodule icon in the System window. In the
Intermodule window, select the analyzer icon and select any option except
for Independent.
•
Tried to insert "Wait for other machine to trigger" function
A "Wait for the other machine to trigger" trigger level causes the logic
analyzer to wait for a signal from the other logic analyzer machine in the
module. If this machine is not on, the current logic analyzer will hang. To
turn the other machine on, select Pod Assignment under Format. Set the
type of the other machine to State or Timing.
•
In Format, no labels have bits assigned to them.
When you insert a trigger function, the logic analyzer sets up a field for you
to enter values. The field length is based on the number of bits assigned to
the first active label, or the label you specify. If there are no bits assigned
to the label, the logic analyzer cannot complete the value field.
See Also
Using the Intermodule Window (see the Agilent Technologies 16700A/BSeries Logic Analysis System help volume)
“To assign probe channels to labels” on page 58
179
Chapter 3: Reference
Error Messages
Trigger inhibited during timing prestore
The "trigger inhibited" informational message appears when you have a
logic analyzer making a conventional timing measurement, and it is
set to a slow sample rate. The logic analyzer will fill the designated
amount of pre-trigger memory before checking for the trigger
condition.
To calculate how long this should take, multiply the sample rate by the
percentage of pre-trigger memory and the acquisition depth. For
example, if
sample period = 1.0 ms (sample rate = 103 samples/sec.)
trigger position = center (percentage of pre-trigger memory = 50%)
acquisition depth = 64K (roughly 64 x 103 samples)
then the approximate time is 32 seconds.
Trigger Specification is too complex
The "Trigger Specification is too complex" message means that the
trigger sequence contains more unique event list expressions than can
be allocated to the available combiner resources in the analyzer
hardware.
The analyzer has a maximum limit of 16 event list combiner resources.
Each unique event list expression requires the use of at least one of
these combiner resources. A complex event list may require more than
one combiner resource.
The message does not mean that any single event list expression was
too complex to combine (see Branch expression is too complex), but
that the overall number of unique branch expressions specified has
exceeded the limit of 16.
In order to compile and run, the total number of unique event list
180
Chapter 3: Reference
Error Messages
expressions must be reduced to 16 and the complexity of some of the
expressions may have to also be reduced.
Branch expressions that are identical (and simple enough to be
combined by a single combiner resource) share the same combiner
resource. Reusing identical event list equations where possible will
optimize the use of combiner resources (see page 181).
You may find more information in the Branch expression is too
complex (see page 169) error message help for the 800 and 1250 Mb/s
modes.
Combiner resource allocation guidelines:
•
Labels that span multiple pod pairs (split labels) increases the number of
required combiner resources as compared with labels that are entirely
contained within a single pod pair.
Whenever possible try to arrange the probing such that labels do not span
pod pairs. This is the single most effective way to reduce the number of
required combiner resources.
NOTE: For labels that do span pod pairs the complexity can be reduced to
the same as that of the non-split label case if all bits in the label on all but
one pod pair can be set to Xs in the event list expression for the
measurement.
For example, if label ADDR has it's 16 most significant bits on pod A3 and
16 least significant bits on pod A2 (spanning pod pairs A4/A3 and A2/A1),
the complexity of the compiled expression will be reduced if all 16 most
significant bits or all 16 least significant bits are set to Xs in the pattern
event.
•
Event lists with up to 4 unique pattern events can be combined in any
combination of ANDs and ORs by a single combiner resource if all of the
pattern labels are non-split and contained on the same pod pair.
Combining more than 4 labels on the same pod pair will require another
combiner resource.
•
Non-split label pattern events from different pod pairs that are ORed
together require an additional combiner resource for each additional pod
pair included in the event list. (ANDing on non-split patterns from
different pod pairs does not increase the required number of combiner
181
Chapter 3: Reference
Error Messages
resources).
•
An inequality compare (<,<=,>,>=) with a split label pattern event
requires 2 combiner resources.
•
A range on a split label pattern event requires 4 combiner resources.
•
The event list in the custom store qualification dialog also allocates
combiner resources from the same pool of 16 resources. If the store
qualification event list equation is the same as one of the branch event list
equations in the trigger sequence, the combiner resource will be shared. A
unique store qualification event list requires the allocation of 1 (or more)
of the combiner resources.
•
333/400/800/1250 MHz state modes requires many more combiner
resources to implement the same trigger sequence as compared to 167/200
MHz state modes and all timing modes. Refer the discussion of complexity
in the Branch expression is too complex help topic.
Waiting for Trigger
This message indicates that the specified trigger pattern has not
occurred. This may be expected, as when you are waiting to trigger on
an unusual event.
Possible Causes
•
Misaligned boundaries for addresses
When the device under test is a microprocessor that fetches only from
long-word aligned addresses, if the trigger is set to look for an opcode
fetch at an address that is not properly aligned, the trigger will never be
found.
•
Trigger set incorrectly
Some strategies you can use when verifying or debugging trigger
sequence levels are:
•
Look at the run status message line or open the Run Status window. It will
tell you what level of the sequence the logic analyzer is in.
•
Stop the measurement and look at the data that was captured. This is
182
Chapter 3: Reference
Error Messages
particularly useful when you use store qualifiers to store "no states" (or
only the states you are interested in) and the branches taken are stored.
•
See Also
Save the trigger setup, then simplify it to see what part of the sequence
does get captured. When you learn what needs to be changed, you can
recall the original trigger setup and make changes to it.
“To Specify Default Storing” on page 70
“To save a trigger setup” on page 84
Analyzer armed from another module contains
no "Arm in from IMB" event
This warning is displayed when a 16715A and newer analyzer machine
is in the group run arming tree, armed from another module (not
directly from the group run), and no sequencer event list in the
analyzer contains an "Arm in from IMB" event or no level in a "Wait for
arm in" trigger function.
In this case the analyzer will not wait for the module above it to trigger
before it triggers.
183
Chapter 3: Reference
Specifications and Characteristics
Specifications and Characteristics
NOTE:
For a complete comparison of all logic analyzer specifications and
characteristics refer to the "Agilent Technologies 16700 Series Logic Analysis
System Product Overview".
•
“Agilent 16716A Logic Analyzer Specifications” on page 184
•
“Agilent 16716A Logic Analyzer Characteristics” on page 184
•
“What is a Specification?” on page 187
•
“What is a Characteristic?” on page 188
Agilent 16716A Logic Analyzer Specifications
The specifications are the performance standards against which the
product is tested. These specifications apply only to the Agilent
Technologies 16716A 167 MHz State/2 GHz Timing Zoom logic
analyzer:
Maximum State Clock Speed:
167 MHz
Threshold Accuracy:
+/-(65 mV + 1.5% of threshold setting)
Minimum Master-to-Master Clock Time: 5.988 ns
Setup/Hold Time:
*Single Clock, Single Edge:
4.5/-2.0 ns through -2.0/4.5 ns,
adjustable in 100-ps increments
*Multiple Edges:
5.0/-2.0 ns through -1.5/4.5 ns,
adjustable in 100-ps increments
* Specified for an input signal VH=-0.9 V, VL=-1.7 V, threshold=-1.3 V,
slew rate=1 V/ns
Agilent 16716A Logic Analyzer Characteristics
The characteristics are not specifications, but are included as
additional information.
General information
- Channel Counts:
1-card module
184
64 data, 4 clock
Chapter 3: Reference
Specifications and Characteristics
2-card
3-card
4-card
5-card
module
module
module
module
132
200
268
336
data,
data,
data,
data,
4
4
4
4
Probes (at end of flying lead set)
- Input Resistance:
100
- Parasitic Tip Capacitance: 1.5
- Minimum Voltage Swing:
500
- Minimum Input Overdrive:
250
- Maximum Voltage:
+/- Threshold Range:
+/- Power through pod cables:
1/3
clock
clock
clock
clock
Kohm, +/- 2%
pF
mV peak-to-peak
mV
40 V peak CAT I
6.0 V, adjustable in 10-mV increments
amp per pod
State Analysis
- Maximum State Clock Speed:
- Maximum Memory Depth
- *Minimum Setup/Hold Time:
-
Minimum State Clock Width:
Minimum Master-to-Master Clock:
Minimum Master-to-Slave Clock:
Minimum Slave-to-Slave Clock:
State Clocks:
State Clock Qualifiers:
**Time Tag Resolution:
Maximum Time Count Between States:
**Maximum State Tag Count:
Store qualification
167 MHz
512 K per channel
4.5/-2.0 ns through -2.0/4.5 ns,
adjustable in 100-ps increments
1.2 ns
5.988 ns
2 ns
5.988 ns
4
4
4 ns
17 seconds
2e32
Default and per sequence level
*
Specified for single-edge, single-clock acquisition.
Multi-edge setup/hold window is 3.0 ns.
** When all pods are being used, time or state tags halve the memory
depth.
Timing Analysis
Timing Zoom
- Sample Rates:
- Memory Depth:
- Sample Period Accuracy:
- Channel-to-channel Skew:
- Time Interval Accuracy:
Conventional Timing
Maximum Sample Rate:
Half Channel
Full Channel
Memory Depth:
Half Channel
Full Channel
Sample Period:
Half Channel
Full Channel
Sample Period Accuracy:
Channel-to-Channel Skew:
Time Interval Accuracy:
2 GHz/1 GHz/500 MHz/250 MHz
16 K samples per channel, both analyzers
+/- 50 ps
less than 1.0 ns
+/- (sample period + channel-tochannel skew + 0.01% of time
interval reading)
667 MHz
333 MHz
1 M samples per channel
512 K samples per channel
1.5 ns
3 ns to 1.0 ms
+/- (100 ps + 0.01% of sample period)
less than 1.5 ns, typical
+/-( sample period + channel-tochannel skew + 0.01% of time
interval reading)
Transitional Timing
Maximum timing analysis sample rate:
Number of channels:
34 x
333 MHz
For sample rates <333 MHz:
(number of modules)
For sample rates =>333 MHz: 34 x
(number of modules) - 17
Maximum channels on a single time base and trigger:
170 (5 modules)
Sample period:
3.0 ns to 1 ms
Sample period accuracy:
+/- (100ps + 0.01% of sample
185
Chapter 3: Reference
Specifications and Characteristics
period)
Channel-to-channel skew:
Time interval accuracy:
to-chan
time interval)]
Minimum data pulse width:
Maximum trigger sequencer speed:
Trigger resources:
< 1.5 ns
+/- [sample period + (chanskew) + (0.01% of
4.0 ns
167 MHz
16 patterns(=,/=,<,>,<=,>=),
15 ranges(in range,not in
range),
Trigger resource conditions:
combinations
Maximum global counter:
Maximum occurrence counter:
Timer value range:
Timer resolution:
Timer accuracy:
Greater than duration:
increments
Less than duration:
increments
Timer rest latency:
Data in to BNC port out delay:
Flag set/reset to evaluation:
Triggering
- Maximum Trigger Sequencer Speed:
- State Sequence Levels:
- Timing Sequence Levels:
- Sequence Level Branching:
- Maximum Occurrence Count Value:
- Pattern Recognizers:
- Range Recognizers:
- Range Width:
- Occurrence Counters:
- Global Counters:
- Flags:
- Flag set/reset to evaluation:
- Timers:
- Timer Value Range:
- Timer Resolution:
- Timer Accuracy:
- Timer Reset Latency:
- Glitch/Edge Recognizers:
- Minimum Detectable Glitch:
- Greater Than Duration:
- Less Than Duration:
- Data In to Trigger Out:
2 edge/glitch,
(2 timers per module) -1
2 global counters,
1 occurrence counter per
sequence level,
4 flags
Arbitrary boolean
16,777,215
16,777,215
100 ns to 5497 sec
5 ns
+/- (10 ns + 0.01%)
5 ns to 83 ms in 5 ns
10 ns to 83 ms in 5 ns
75 ns
150 ns
110 ns
167 MHz
16
16
Arbitrary 4-way "If/then/else"
16,777,215
16
15
32 bits
1 per sequence level
2
8, can be used between modules
110 ns, typical
(2 x number of cards) - 1
100 ns to 5497 seconds
5 ns
+/- 10 ns + 0.01%
70 ns
2 per pod pair (timing only)
1.5 ns
6 ns to 100 ms in 6-ns increments
12 ns to 100 ms in 6-ns increments
150 ns, typical
Power Requirements
All necessary power is supplied by the backplane connector of
the logic analysis system mainframe.
Operating Environment Characteristics
- Indoor use only.
- Temperature
Instrument (except disk and media): 0 to 50 degrees C (+32
to 122 degrees F)
Probe lead sets and cables:
0 to 65 degrees C (+32 to
149 degrees F)
- Humidity
Instrument, probe lead sets, and cables: up to 80% relative humidity at
40 degrees C (+104 degrees F)
- Altitude
Operating, to 4600 m (15,000 ft)
186
Chapter 3: Reference
Specifications and Characteristics
Non-operating, to 15,300 m (50,000 ft)
- Vibration
Operating:
Random vibration 5-500 Hz, 10 minutes per axis,
approximately 0.2 g rms
Nonoperating: Random vibration 5 to 500 Hz, 10 minutes per axis,
approximately 2.41 g rms; and swept sine resonant
search, 5 to 500 Hz, 0.50 g (0-peak), 5-minute
resonant dwell at 4 resonances per axis.
Reliability is enhanced when operating within the following ranges:
- Temperature
+20 to 35 degrees C (+68 to 95 degrees F)
- Humidity
20% to 80% non-condensing
Storage
Store or ship the logic analyzer in environments with the following
limits:
- Temperature
-40 to +75 degrees C
- Humidity
up to 90% at 65 degrees C
- Altitude
up to 15,300 meters (50,000 feet)
Protect the module from temperature extremes which cause condensation
on the instrument.
What is a Specification?
A Specification is a numeric value, or range of values, that bounds the
performance of a product parameter. The product warranty covers the
performance of parameters described by specifications. Products
shipped from the factory meet all specifications. Additionally, the
products sent to Agilent Technologies Customer Service Centers for
calibration and returned to the customer meet all specifications.
Specifications are verified by Calibration Procedures.
What is a Calibration
Procedure?
Calibration procedures verify that products or systems operate within
the specifications. Parameters covered by specifications have a
corresponding calibration procedure. Calibration procedures include
both performance tests and system verification procedure. Calibration
procedures are traceable and must specify adequate calibration
standards.
Calibration procedures verify products meet the specifications by
comparing measured parameters against a pass-fail limit. The pass-fail
limit is the specification less any required guardband.
The term "calibration" refers to the process of measuring parameters
and referencing the measurement to a calibration standard rather than
the process of adjusting products for optimal performance, which is
referred to as an "operational accuracy calibration".
187
Chapter 3: Reference
Specifications and Characteristics
What is a Characteristic?
Characteristics describe product performance that is useful in the
application of the product, but that is not covered by the product
warranty. Characteristics describe performance that is typical of the
majority of a given product, but not subject to the same rigor
associated with specifications.
Characteristics are verified by Function Tests.
What is a Function
Test?
Function tests are quick tests designed to verify basic operation of a
product. Function tests include operator's checks and operation
verification procedures. An operator's check is normally a fast test
used to verify basic operation of a product. An operation verification
procedure verifies some, but not all, specifications, and often at a lower
confidence level than a calibration procedure.
188
4
Concepts
•
“Understanding Logic Analyzer Triggering” on page 190
•
“Understanding State Mode Sampling Positions” on page 206
189
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
Understanding Logic Analyzer Triggering
Setting up logic analyzer triggers can be difficult and time-consuming.
You could assume that if you know how to program, you should be able
to set up a logic analyzer trigger with no difficulty. However, this is not
true because there are many concepts that are unique to logic analysis.
The purpose of this section is to describe these key concepts and how
to use them effectively.
•
“The Conveyor Belt Analogy” on page 190
•
“Summary of Triggering Capabilities” on page 192
•
“Sequence Levels” on page 192
•
“Boolean Expressions” on page 195
•
“Branches” on page 196
•
“Edges” on page 196
•
“Ranges” on page 196
•
“Flags” on page 197
•
“Occurrence Counters and Global Counters” on page 197
•
“Timers” on page 198
•
“Storage Qualification” on page 199
•
“Strategies for Setting Up Triggers” on page 201
•
“Conclusions” on page 205
See Also
“Setting Up Triggers and Running Measurements” on page 63
The Conveyor Belt Analogy
The memory of a logic analyzer can be compared to a very long
conveyor belt, and the samples acquired from the Device Under Test
(DUT) as boxes on the conveyor belt. At one end, new boxes are
190
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
placed on the conveyor belt, and at the other end the boxes fall off. In
other words, because logic analyzer memory is limited in depth
(number of samples), whenever a new sample is acquired the oldest
sample currently in memory is thrown away if the memory is full. This
is shown in the following figure.
The conveyor belt analogy
A logic analyzer trigger is similar to someone standing at the beginning
of the conveyor belt placing more boxes on it. They are told to “look for
a special box and to stop the conveyor belt when that box reaches a
particular position on the belt”. Using this analogy, the special box is
the trigger. Once a logic analyzer detects a sample that matches the
trigger condition, this is the indication that it should stop acquiring
more samples when the trigger is located appropriately in memory.
The location of the trigger in memory is known as the trigger position.
Normally, the trigger position is set to the middle so that the maximum
number of samples that occurred before and after the trigger are in
memory. However, you can set the trigger position to any point in
memory.
The concepts in this analogy are summed up in the following table.
Mapping of concepts in the Conveyor Belt Analogy
to a Logic Analyzer
Conveyor Belt Analogy
=====================
Boxes on the belt
--------------------Number of boxes that
will fit on the belt
---------------------
Logic analyzer
==============================
Samples acquired from the
device under test
-----------------------------Memory depth
------------------------------
191
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
Special box
---------------------
Trigger point
------------------------------
Next: “Summary of Triggering Capabilities” on page 192
Summary of Triggering Capabilities
Because logic analyzer triggering provides a great deal of functionality,
the following table provides a brief summary of the capabilities covered
in this article. Each of these capabilities will be described.
Summary of Logic Analyzer Triggering Capabilities
Capability
=====================
Edges
--------------------Boolean expressions
--------------------Ranges
--------------------Storage qualification
--------------------Counters
--------------------Timers
---------------------
Examples
===============================================
If there is rising edge on SIG1 then Trigger
If there is falling edge on SIG1 then Trigger
----------------------------------------------If ADDR = 1000 and DATA = 2000
----------------------------------------------If ADDR in range 1000 to 2000
----------------------------------------------1. If..
Else If ADDR in range 1000 to 2000 then
Store Sample
Go to 1
Else If ADDR not in range 1000 to 2000 then
Don't Store Sample
Go to 1
----------------------------------------------1. If DATA = 1000 Then
Increment Counter 1
Go to 2
2. If Counter 1 > 2 Then
Trigger
----------------------------------------------1. If DATA = 1000 Then
Start Timer 1
Go to 2
2. If Timer 1 > 500 ns Then
Trigger
-----------------------------------------------
Next: “Sequence Levels” on page 192
Sequence Levels
While logic analyzer triggers are often simple, they can require
complex programming. For example, you may want to trigger on the
rising edge of one signal that is followed by the rising edge of another
signal. This means that the logic analyzer must first find the first rising
192
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
edge before it begins looking for the next rising edge. Because there is
a sequence of steps to find the trigger, this is known as a trigger
sequence. Each step of the sequence is called a sequence level.
Each sequence level consists of two parts; the conditions and the
actions. The conditions are Boolean expressions such as “If ADDR =
1000” or “If there is a rising edge on SIG1”. The actions are what the
logic analyzer should do if the condition is met. Examples of actions
include triggering the logic analyzer, going to another sequence level,
or starting a timer. This is similar to an If/Then statement in
programming.
Each sequence level in the trigger sequence is assigned a number. The
first sequence level to be executed is always Sequence Level 1, but
because of the Go To actions, the rest of the sequence levels can be
executed in any order.
When a sequence level is executed and none of the Boolean
expressions are true, the logic analyzer acquires the next sample and
executes the same sequence level again. As a simple example, consider
the following trigger sequence:
1. If DATA = 7000 then Trigger
If the following samples were acquired, the logic analyzer would trigger
on sample #6.
Sample #
1
2
3
4
5
6
7
ADDR
1000
1010
1020
1030
1040
1050
1060
DATA
2000
3000
4000
5000
6000
7000
2000
<- This is where the logic analyzer triggers
In essence, Sequence Level 1 is equivalent to “Keep acquiring more
samples until DATA=7000, then trigger”.
If a Boolean expression in a sequence level is met, another sample is
always acquired before the next sequence level is executed. In other
words, if a sample meets the condition in Sequence Level 1, another
sample will be acquired before executing Sequence Level 2. This
means that it is not possible for a single sample to be used to meet the
conditions of more than one sequence level. Each sequence level can
be thought of as representing events that occur at different points in
193
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
time. Two sequence levels can never be used to specify two events that
happen simultaneously.
For example, consider the following trigger sequence:
1. If ADDR = 1000 then Go to 2
2. If DATA = 2000 then Trigger
If the following samples were acquired, the logic analyzer would trigger
on sample #7.
Sample #
1
2
3
4
5
6
7
ADDR
1000
1010
1020
1030
1040
1050
1060
DATA
2000
3000
4000
5000
6000
7000
2000
<- This sample meets the condition in Sequence level #1
<- This is where the logic analyzer triggers
Note that the logic analyzer will not trigger on Sample #1 because a
new sample is acquired between the time that the condition in
Sequence level 1 is met and when the condition in Sequence Level #2 is
tested. A good way to think of this trigger sequence is “Find ADDR =
1000 followed by DATA = 2000 and then trigger”. Multiple sequence
levels in a trigger sequence imply a “followed by”.
Once a logic analyzer triggers, it does not trigger again. In other words,
even if more than one sample meets the trigger condition, the logic
analyzer still only triggers once. For example, using “ADDR=1000” as
our trigger, if the logic analyzer acquires the following samples, it will
trigger on Sample #2 and only on Sample #2.
Sample #
1
2
3
4
5
ADDR
0000
1000
2000
1000
1040
<- The logic analyzer triggers here
<- The logic analyzer does NOT trigger again here
A frequently asked question is “What happens if the conditions in a
sequence level are not met?” For example, if there is a condition that
says “If ADDR = 1000 Then Trigger”, what happens if the current
sample has ADDR = 2000? The logic analyzer simply acquires the next
sample and tries to execute this sequence level again. In essence, if the
trigger condition is “ADDR = 1000”, this is equivalent to “Keep
acquiring more samples until you find one that has ADDR=1000”.
Therefore, if you set up a trigger condition that is never met, the logic
194
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
analyzer will never trigger.
When the conditions are met in a sequence level, it is clear which
sequence level will be executed next when a “Go To” action is used, but
it is not necessarily clear if there is no “Go To”. On some logic
analyzers, if there is no “Go To”, this means that the next sequence
level should be executed. On other logic analyzers, it means the same
sequence level should be executed again. Because of this confusion, it
is good practice to always use a “Go To” action rather than relying on
the default. The new Agilent Technologies 16715/16/17/18/19A state
and timing modules deal with this problem by automatically including a
“Go To” or “Trigger” action in every sequence level. For example:
If ADDR = 1000 and DATA = 2000 then
Go to 1
<- This is automatically added on the Agilent 16715/16/17/18/19A
Next: “Boolean Expressions” on page 195
Boolean Expressions
While multiple sequence levels imply a “followed by”, within a
sequence level Boolean expressions can be used. An example is:
If ADDR = 1000 and DATA = 2000
This expression means that for this expression to be met, ADDR must
equal 1000 in the same sample that DATA equals 2000. In other words,
ADDR equals 1000 at the same time that DATA equals 2000. Therefore,
if you want to trigger on two events that occur at the same time, a
Boolean expression should be used.
It's a common mistake to try to use two sequence levels when a
Boolean expression should be used or to use a Boolean expression
when two sequence levels should be used.
NOTE:
Boolean expressions are used for events that happen at the same time, and
multiple sequence levels are used when one event follows another.
Next: “Branches” on page 196
195
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
Branches
Branches are similar to the Switch statement in the C programming
language and the Select Case statement in Basic. They provide a
method for testing multiple conditions. Each branch has its own
actions. An example of multiple branches is shown below:
1. If ADDR < 1000 then Go To 2
Else If ADDR > 2000 then Go To 3
Else If DATA = 2000 then Trigger
2. If DATA <= 7000 then Trigger
3. If there is a Rising Edge on SIG1,
<- This is a branch of Level 1
<- This is a 2nd branch of Level 1
<- This is a 3rd branch of Level 1
then Trigger
In sequence level 1, there are three branches, so there are three
possible actions that can be taken.
When the condition of one branch is met, none of the branches below it
are tested. In other words, there is no way for more than one branch to
be executed based upon a single sample, even if the sample causes the
conditions for more than one branch to be met. In other words, each
branch is an “Else If”.
Next: “Edges” on page 196
Edges
Edges represent a transition from low to high or high to low on a single
signal. Typically, edges are specified as “rising edge”, “falling edge”, or
“either edge”, where “rising edge” indicates a transition from a low to a
high. On most logic analyzers, up to two edges can be included in the
trigger sequence although some allow only one.
Next: “Ranges” on page 196
Ranges
Ranges are a convenient method for specifying a range of values, such
as “ADDR in range 1000 to 2000”. Most logic analyzers also support a
196
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
“not in range” function as well. Ranges are a convenient shortcut so
that you don't have to specify “ADDR >= 1000 and ADDR <= 2000”.
Next: “Flags” on page 197
Flags
Flags are Boolean variables that are used to send signals from one
module to another. They can be set when a condition occurs in one
module and tested later by another module. In the example below, flag
1 is used to keep track of what happens in the trigger sequence of
Module 1 so that this information can be used in Module 2.
Trigger Sequence for Module 1
1. If ADDR < 5000 then
Set Flag 1
Trigger and fill memory
Trigger Sequence for Module 2
1. If DATA = 5000 and Flag 1 is set then Trigger
Else if DATA = 1000 and not Flag 1 then Trigger
Next: “Occurrence Counters and Global Counters” on page 197
Occurrence Counters and Global Counters
Occurrence Counters are used in situations where you want to find the
Nth occurrence of an event. For example, if you want to trigger on the
5th time that ADDR = 1000, you could set up the trigger as:
If ADDR = 1000 occurs 5 times then Trigger
Global Counters are like integer variables. They are more flexible than
Occurrence Counters because they can be used to count complex
events such as an edge followed by another edge. Global Counters can
be incremented, tested, and reset. By default, Global Counters begin
with zero and don't need to be reset unless they have already been
used in the trigger sequence. In general, Occurrence Counters should
197
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
be used in place of Global Counters, if possible, because they are easier
to use and because there is a limited number of Global Counters.
Next: “Timers” on page 198
Timers
Timers are used to check the amount of time that has elapsed between
events. For example, if you want to trigger on one edge followed by
another edge that occurs within 500ns, use a timer. The most critical
point to remember in using timers is that they need to be started
before they are tested. In other words, timers do not start
automatically.
The key to setting up a timer is to identify where it should be started
and where it should be tested. Consider the example in the following
figure. The timer should be started when the rising edge on SIG1 is
detected and it should be tested when the rising edge occurs on SIG2.
An edge followed by an edge with a time limit
An example trigger sequence to set up this measurement is:
1. If there is a Rising Edge on SIG1, then
Start Timer1
Go to 2
2. If there is a Rising Edge on SIG2 AND Timer1 < 500ns then
Trigger
While the above trigger sequence seems correct, it actually has a
critical flaw. What happens if there is a rising edge on SIG1 but SIG2
doesn't occur within 500ns? The logic analyzer will never trigger,
198
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
because timer1 will keep running and condition “Timer1 <500 ns” will
never be met. There might be another rising edge on SIG1 that is
followed within 500ns by the rising edge on SIG2 that occurs later on,
so this situation is unacceptable.
To fix this problem, whenever the timer exceeds 500ns without
triggering, the sequence should loop back to Level 1 to look for another
rising edge on SIG1. The following shows an example of the correct
sequence:
1. If there is a Rising Edge on SIG1, then
Start Timer1
Go to 2
2. If there is a Rising Edge on SIG2 AND Timer1 < 500ns then
Trigger
Else If Timer1 >= 500ns then
Reset Timer1
Go to 1
Occasionally, you may run out of timers. A counter can be used in place
of a timer if the logic analyzer is sampling at regular intervals (that is, if
it's in the timing sampling mode). A timer can be simulated by counting
the number of samples that are acquired. For example, if the logic
analyzer acquires a new sample every 10ns and seven samples are
acquired, this represents 70ns.
Next: “Storage Qualification” on page 199
Storage Qualification
Storage qualification is used to determine if an acquired sample should
be stored (that is, placed in memory) or thrown away. This keeps the
logic analyzer memory from being filled with samples that are not
needed.
Default Storage
The simplest method to set up storage qualification is by setting up the
Default Storage. This is specified separate from the trigger sequence,
such as in a separate tab or another dialog. Default Storage means
“unless a sequence level specifies otherwise, this is what should be
stored”. As an example, you may want to only store samples if ADDR is
in the range 1000 to 2000, so you should set the Default Storage to:
199
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
ADDR In Range 1000 to 2000
By default, the Default Storage is set to store all samples acquired. You
can also set the Default Storage to store nothing, which means that no
samples will be stored unless a sequence level overrides the default
storage.
Sequence Level
Storage
Sequence level storage qualification means that within a particular
sequence level only certain samples will be stored. This means that
until a “Go To” or “Trigger” action is used to leave this sequence level,
the storage qualification applies. This is useful when you want different
storage qualification for each sequence level. For example, you may
want to store nothing until ADDR = 1000 and then store only samples
with ADDR in the range 1000 to 2000 for the rest of the measurement.
Setting up sequence level storage requires the use of an additional
branch. For example, if you want to store only samples with ADDR in
the range 5000 to 6FFF while looking for DATA = 005E, the following
sequence level could be used in some situations:
1. If DATA
Else If
Store
Go to
= 005E then Trigger
ADDR in range 5000 to 6FFF then
Sample
1
Note the use of the store sample action. This means “store the most
recently acquired sample in memory now”. It does not mean, “From
now on, start storing”. It should be noted that since the store sample
action is never executed unless ADDR is in the range 5000 to 6FFF,
this branch essentially means “While in this sequence level, store only
samples with ADDR between 5000 and 6FFF”.
The above example seems to imply that only samples with ADDR
between 5000 and 6FFF will be stored. However, this depends upon
how the default storage has been set up. Using the previous example, if
the default storage is set to “Store Everything”, and a sample is outside
of the range 5000 to 6FFF, then the Else If branch is not executed and
the Default Storage is applied. In essence, the sequence level has said
what to do when a sample has a value in a particular range, but it
doesn't say what to do for samples outside the range. Therefore, if you
want to specify the sequence level storage unambiguously, use the
following:
200
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
1. If DATA
Else If
Store
Go to
Else If
Don't
Go to
= 005E then Trigger
ADDR in range 5000 to 6FFF then
Sample
1
ADDR not in range 5000 to 6FFF then
Store Sample
1
Alternatively, if the default storage is set to “Store Everything”, use the
following:
1. If DATA
Else If
Don't
Go to
= 005E then Trigger
ADDR not in range 5000 to 6FFF then
Store Sample
1
In summary, Sequence Level Storage always overrides the Default
Storage, but only for the conditions specifically mentioned in the
Sequence Level Storage. You must be very careful that you account for
the interaction between Default Storage and Sequence Level Storage.
Next: “Strategies for Setting Up Triggers” on page 201
Strategies for Setting Up Triggers
•
“Trigger Functions” on page 201
•
“Setting Up Complex Triggers” on page 204
•
“Document Your Trigger Sequences” on page 204
Trigger Functions
While setting up logic analyzer triggers can be difficult, trigger
functions can greatly simplify the process. Trigger functions are
commonly-needed building blocks that can be combined to set up a
trigger. Because the functions cover most common triggers, you can set
up your trigger simply by selecting the appropriate function and filling
in the data. The Agilent Technologies 16715A logic analyzer trigger
user interface is shown in the following figure. Note that trigger
functions are prominently located at the top of the screen.
201
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
The Agilent 16715A trigger user interface
Note that a picture (which corresponds to the selected function) is
provided to the right of the trigger function list.
For example, if you want to trigger when a bus pattern is immediately
followed by another bus pattern, you can use the “Find Pattern2
occurring immediately after Pattern1” trigger function, shown in the
following figure.
Pattern2 occurring immediately after Pattern1
Once you have selected this function, you simply fill in the names of
the buses and the patterns. Contrast the previous figure with the
following figure, which is the same trigger created using If/Then
statements. The trigger function is easier to use because the additional
details of the If/Then statements have been hidden. However, if you
want to see the details, you can break down the function.
202
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
The same trigger as If/Then statements
Trigger functions can be modified. For example, if you start with the
function “Find Edge”, you can add another event, and it becomes the
same as “Find Edge and Pattern”. Therefore, a function that is not
exactly correct can often be converted into the desired trigger. It is also
possible to break down a function into the underlying If/Then
statements and modify them.
The functions “Store range until pattern occurs” and “Store nothing
until pattern occurs” make storage qualification much easier. These
functions completely override the Default Storage. The “Store range
until pattern occurs” function is shown in the following figure.
Store range until pattern occurs
Trigger functions are like building blocks because they can be used
together in a trigger sequence. For example, if you want to set up a
trigger as “Find edge followed by pattern”, you can use a “Find Edge”
function for Level 1 and a “Find Pattern” function for Sequence Level 2
(see the following figure). So, functions are useful both as an entire
trigger sequence and as one step in a trigger sequence.
203
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
“Find Edge” and “Find Pattern” together
Next: “Setting Up Complex Triggers” on page 204
Setting Up Complex Triggers
Frequently, the most difficult part of setting up a complex trigger is
breaking down the problem. In other words, how do you map a
complex trigger into sequence levels, branches, and Boolean
expressions? Here are step by step instructions:
1. Break down the problem into events that don't happen simultaneously.
These correspond to the sequence levels.
2. Scan the list of trigger functions to try to find some that match the events
identified in Step #1.
3. Within all remaining events, break them down into Boolean expressions
and their corresponding actions. Each Boolean expression/Action pair
corresponds to a separate branch within a sequence level. Remember that
“Store” branches may exist that are used only to handle storage
qualification for that sequence level.
Next: “Document Your Trigger Sequences” on page 204
Document Your Trigger Sequences
If a trigger sequence is important at one time, it is likely to be
important again. This is why documenting trigger sequences is so
valuable. Complex trigger sequences generally are too difficult to
understand without some accompanying explanation. The following
figure shows an example of the inline documentation on an Agilent
Technologies logic analyzer. Inline means that the documentation is
included in the trigger definition itself. This allows you to document
204
Chapter 4: Concepts
Understanding Logic Analyzer Triggering
different parts of the trigger to describe how they work.
Inline documentation on an Agilent logic analyzer
Next: “Conclusions” on page 205
Conclusions
Setting up logic analyzer triggers is very different than writing
software. The job can be greatly simplified if other work can be
leveraged by using pre-defined trigger functions and well-documented
triggers that were written earlier. Only write your own trigger setup if
there's nothing else available. Finally, when faced with a difficult
trigger to set up, break down the problem into smaller chunks and deal
with each one individually.
205
Chapter 4: Concepts
Understanding State Mode Sampling Positions
Understanding State Mode Sampling Positions
Synchronous sampling (state mode) logic analyzers are like edgetriggered flip-flops in that they require input logic signals to be stable
for a period of time before the clock event (setup time) and after the
clock event (hold time) in order to properly interpret the logic level.
The combined setup and hold time is known as the setup/hold window.
A device under test (because of its own setup/hold requirements)
specifies that data be valid on a bus for a certain length of time. This is
known as the data valid window. The data valid window on most buses
is generally less than half of the bus clock period.
To accurately capture data on a bus:
•
The logic analyzer's setup/hold time must fit within the data valid window.
•
Because the location of the data valid window relative to the bus clock is
different for different types of buses, the position of the logic analyzer's
setup/hold window must be adjustable (relative to the sampling clock, and
with fine resolution) within the data valid window. For example:
206
Chapter 4: Concepts
Understanding State Mode Sampling Positions
To position the setup/hold window (sampling position) within the data
valid window, a logic analyzer has an adjustable delay on each sampling
clock input (to position the setup/hold window for all the channels in a
pod).
Sample Position
Adjustments on
Individual Channels
Some logic analyzers let you adjust the position of the setup/hold
window (sampling position) on each channel. When you can make
sampling position adjustments on individual channels, you can make
the logic analyzer's setup/hold window smaller because you can correct
for the delay effects caused by the probe cables and the logic analyzer's
internal circuit board traces, and you are left with the setup/hold
requirements of the logic analyzer's internal sampling circuitry.
However, the process of manually positioning the setup/hold window
for each channel is time consuming. For each signal in the device under
test and each logic analyzer channel, you must measure the data valid
window in relation to the bus clock (with an oscilloscope), repeatedly
position the setup/hold window and run measurements to see if the
logic analyzer captures data correctly, and finally position the setup/
hold window in between the positions where data was captured
incorrectly.
In Agilent Technologies logic analyzers which have the eye finder
feature, you can automatically position the setup/hold window on each
207
Chapter 4: Concepts
Understanding State Mode Sampling Positions
channel in a small fraction of the time (and without the extra test
equipment) that it takes to make the adjustments manually. Eye finder
is an easy way to get the smallest possible logic analyzer setup/hold
window.
See Also
“To automatically adjust sampling positions” on page 46
“To manually adjust sampling positions” on page 49
“Selecting the State Mode (Synchronous Sampling)” on page 43
“Sampling Positions Dialog” on page 129
208
Glossary
absolute Denotes the time period
or count of states between a captured
state and the trigger state. An
absolute count of -10 indicates the
state was captured ten states before
the trigger state was captured.
acquisition Denotes one complete
cycle of data gathering by a
measurement module. For example,
if you are using an analyzer with
128K memory depth, one complete
acquisition will capture and store
128K states in acquisition memory.
analysis probe A probe connected
to a microprocessor or standard bus
in the device under test. An analysis
probe provides an interface between
the signals of the microprocessor or
standard bus and the inputs of the
logic analyzer. Also called a
preprocessor.
analyzer 1 In a logic analyzer with
two machines, refers to the machine
that is on by default. The default
name is Analyzer<N>, where N is
the slot letter.
analyzer 2 In a logic analyzer with
two machines, refers to the machine
that is off by default. The default
name is Analyzer<N2>, where N is
the slot letter.
arming An instrument tool must be
armed before it can search for its
trigger condition. Typically,
instruments are armed immediately
when Run or Group Run is selected.
You can set up one instrument to arm
another using the Intermodule
Window. In these setups, the second
instrument cannot search for its
trigger condition until it receives the
arming signal from the first
instrument. In some analyzer
instruments, you can set up one
analyzer machine to arm the other
analyzer machine in the Trigger
Window.
asterisk (*) See edge terms,
glitch, and labels.
bits Bits represent the physical logic
analyzer channels. A bit is a channel
that has or can be assigned to a label.
A bit is also a position in a label.
card This refers to a single
instrument intended for use in the
Agilent Technologies 16700A/Bseries mainframes. One card fills one
slot in the mainframe. A module may
comprise a single card or multiple
cards cabled together.
channel The entire signal path from
the probe tip, through the cable and
module, up to the label grouping.
click When using a mouse as the
209
Glossary
pointing device, to click an item,
position the cursor over the item.
Then quickly press and release the
left mouse button.
clock channel A logic analyzer
channel that can be used to carry the
clock signal. When it is not needed
for clock signals, it can be used as a
data channel, except in the Agilent
Technologies 16517A.
context record A context record is
a small segment of analyzer memory
that stores an event of interest along
with the states that immediately
preceded it and the states that
immediately followed it.
context store If your analyzer can
perform context store
measurements, you will see a button
labeled Context Store under the
Trigger tab. Typical context store
measurements are used to capture
writes to a variable or calls to a
subroutine, along with the activity
preceding and following the events. A
context store measurement divides
analyzer memory into a series of
context records. If you have a 64K
analyzer memory and select a 16state context, the analyzer memory is
divided into 4K 16-state context
records. If you have a 64K analyzer
memory and select a 64-state
context, the analyzer memory will be
210
divided into 1K 64-state records.
count The count function records
periods of time or numbers of state
transactions between states stored in
memory. You can set up the analyzer
count function to count occurrences
of a selected event during the trace,
such as counting how many times a
variable is read between each of the
writes to the variable. The analyzer
can also be set up to count elapsed
time, such as counting the time spent
executing within a particular function
during a run of your target program.
cross triggering Using intermodule
capabilities to have measurement
modules trigger each other. For
example, you can have an external
instrument arm a logic analyzer,
which subsequently triggers an
oscilloscope when it finds the trigger
state.
data channel A channel that
carries data. Data channels cannot be
used to clock logic analyzers.
data field A data field in the pattern
generator is the data value associated
with a single label within a particular
data vector.
data set A data set is made up of all
labels and data stored in memory of
any single analyzer machine or
Glossary
instrument tool. Multiple data sets
can be displayed together when
sourced into a single display tool. The
Filter tool is used to pass on partial
data sets to analysis or display tools.
debug mode See monitor.
delay The delay function sets the
horizontal position of the waveform
on the screen for the oscilloscope and
timing analyzer. Delay time is
measured from the trigger point in
seconds or states.
demo mode An emulation control
session which is not connected to a
real target system. All windows can
be viewed, but the data displayed is
simulated. To start demo mode,
select Start User Session from the
Emulation Control Interface and
enter the demo name in the
Processor Probe LAN Name field.
Select the Help button in the Start
User Session window for details.
deskewing To cancel or nullify the
effects of differences between two
different internal delay paths for a
signal. Deskewing is normally done
by routing a single test signal to the
inputs of two different modules, then
adjusting the Intermodule Skew so
that both modules recognize the
signal at the same time.
device under test The system
under test, which contains the
circuitry you are probing. Also known
as a target system.
don't care For terms, a "don't care"
means that the state of the signal
(high or low) is not relevant to the
measurement. The analyzer ignores
the state of this signal when
determining whether a match occurs
on an input label. "Don't care" signals
are still sampled and their values can
be displayed with the rest of the data.
Don't cares are represented by the X
character in numeric values and the
dot (.) in timing edge specifications.
dot (.) See edge terms, glitch,
labels, and don't care.
double-click When using a mouse
as the pointing device, to double-click
an item, position the cursor over the
item, and then quickly press and
release the left mouse button twice.
drag and drop Using a Mouse:
Position the cursor over the item, and
then press and hold the left mouse
button. While holding the left mouse
button down, move the mouse to
drag the item to a new location. When
the item is positioned where you
want it, release the mouse button.
211
Glossary
Using the Touchscreen:
Position your finger over the item,
then press and hold finger to the
screen. While holding the finger
down, slide the finger along the
screen dragging the item to a new
location. When the item is positioned
where you want it, release your
finger.
edge mode In an oscilloscope, this
is the trigger mode that causes a
trigger based on a single channel
edge, either rising or falling.
edge terms Logic analyzer trigger
resources that allow detection of
transitions on a signal. An edge term
can be set to detect a rising edge,
falling edge, or either edge. Some
logic analyzers can also detect no
edge or a glitch on an input signal.
Edges are specified by selecting
arrows. The dot (.) ignores the bit.
The asterisk (*) specifies a glitch on
the bit.
emulation module A module
within the logic analysis system
mainframe that provides an
emulation connection to the debug
port of a microprocessor. An E5901A
emulation module is used with a
target interface module (TIM) or an
analysis probe. An E5901B emulation
module is used with an E5900A
emulation probe.
212
emulation probe The stand-alone
equivalent of an emulation module.
Most of the tasks which can be
performed using an emulation
module can also be performed using
an emulation probe connected to
your logic analysis system via a LAN.
emulator An emulation module or
an emulation probe.
Ethernet address See link-level
address.
events Events are the things you
are looking for in your target system.
In the logic analyzer interface, they
take a single line. Examples of events
are Label1 = XX and Timer 1 > 400
ns.
filter expression The filter
expression is the logical OR
combination of all of the filter terms.
States in your data that match the
filter expression can be filtered out or
passed through the Pattern Filter.
filter term A variable that you
define in order to specify which
states to filter out or pass through.
Filter terms are logically OR'ed
together to create the filter
expression.
Format The selections under the
logic analyzer Format tab tell the
Glossary
logic analyzer what data you want to
collect, such as which channels
represent buses (labels) and what
logic threshold your signals use.
frame The Agilent Technologies or
16700A/B-series logic analysis system
mainframe. See also logic analysis
system.
gateway address An IP address
entered in integer dot notation. The
default gateway address is 0.0.0.0,
which allows all connections on the
local network or subnet. If
connections are to be made across
networks or subnets, this address
must be set to the address of the
gateway machine.
glitch A glitch occurs when two or
more transitions cross the logic
threshold between consecutive
timing analyzer samples. You can
specify glitch detection by choosing
the asterisk (*) for edge terms under
the timing analyzer Trigger tab.
grouped event A grouped event is
a list of events that you have
grouped, and optionally named. It can
be reused in other trigger sequence
levels. Only available in Agilent
Technologies 16715A or higher logic
analyzers.
held value A value that is held until
the next sample. A held value can
exist in multiple data sets.
immediate mode In an
oscilloscope, the trigger mode that
does not require a specific trigger
condition such as an edge or a
pattern. Use immediate mode when
the oscilloscope is armed by another
instrument.
interconnect cable Short name for
module/probe interconnect cable.
intermodule bus The intermodule
bus (IMB) is a bus in the frame that
allows the measurement modules to
communicate with each other. Using
the IMB, you can set up one
instrument to arm another. Data
acquired by instruments using the
IMB is time-correlated.
intermodule Intermodule is a term
used when multiple instrument tools
are connected together for the
purpose of one instrument arming
another. In such a configuration, an
arming tree is developed and the
group run function is designated to
start all instrument tools. Multiple
instrument configurations are done in
the Intermodule window.
internet address Also called
Internet Protocol address or IP
address. A 32-bit network address. It
213
Glossary
is usually represented as decimal
numbers separated by periods; for
example, 192.35.12.6. Ask your LAN
administrator if you need an internet
address.
labels Labels are used to group and
identify logic analyzer channels. A
label consists of a name and an
associated bit or group of bits. Labels
are created in the Format tab.
line numbers A line number (Line
#s) is a special use of symbols. Line
numbers represent lines in your
source file, typically lines that have
no unique symbols defined to
represent them.
link-level address Also referred to
as the Ethernet address, this is the
unique address of the LAN interface.
This value is set at the factory and
cannot be changed. The link-level
address of a particular piece of
equipment is often printed on a label
above the LAN connector. An
example of a link-level address in
hexadecimal: 0800090012AB.
local session A local session is
when you run the logic analysis
system using the local display
connected to the product hardware.
logic analysis system The Agilent
Technologies 16700A/B-series
214
mainframes, and all tools designed to
work with it. Usually used to mean
the specific system and tools you are
working with right now.
machine Some logic analyzers allow
you to set up two measurements at
the same time. Each measurement is
handled by a different machine. This
is represented in the Workspace
window by two icons, differentiated
by a 1 and a 2 in the upper right-hand
corner of the icon. Logic analyzer
resources such as pods and trigger
terms cannot be shared by the
machines.
markers Markers are the green and
yellow lines in the display that are
labeled x, o, G1, and G2. Use them to
measure time intervals or sample
intervals. Markers are assigned to
patterns in order to find patterns or
track sequences of states in the data.
The x and o markers are local to the
immediate display, while G1 and G2
are global between time correlated
displays.
master card In a module, the
master card controls the data
acquisition or output. The logic
analysis system references the
module by the slot in which the
master card is plugged. For example,
a 5-card Agilent Technologies 16555D
would be referred to as Slot C:
Glossary
machine because the master card is
in slot C of the mainframe. The other
cards of the module are called
expansion cards.
menu bar The menu bar is located
at the top of all windows. Use it to
select File operations, tool or system
Options, and tool or system level
Help.
message bar The message bar
displays mouse button functions for
the window area or field directly
beneath the mouse cursor. Use the
mouse and message bar together to
prompt yourself to functions and
shortcuts.
module/probe interconnect cable
The module/probe interconnect cable
connects an E5901B emulation
module to an E5900B emulation
probe. It provides power and a serial
connection. A LAN connection is also
required to use the emulation probe.
module An instrument that uses a
single timebase in its operation.
Modules can have from one to five
cards functioning as a single
instrument. When a module has more
than one card, system window will
show the instrument icon in the slot
of the master card.
monitor When using the Emulation
Control Interface, running the
monitor means the processor is in
debug mode (that is, executing the
debug exception) instead of
executing the user program.
panning The action of moving the
waveform along the timebase by
varying the delay value in the Delay
field. This action allows you to
control the portion of acquisition
memory that will be displayed on the
screen.
pattern mode In an oscilloscope,
the trigger mode that allows you to
set the oscilloscope to trigger on a
specified combination of input signal
levels.
pattern terms Logic analyzer
resources that represent single states
to be found on labeled sets of bits; for
example, an address on the address
bus or a status on the status lines.
period (.) See edge terms, glitch,
labels, and don't care.
pod pair A group of two pods
containing 16 channels each, used to
physically connect data and clock
signals from the unit under test to the
analyzer. Pods are assigned by pairs
in the analyzer interface. The number
of pod pairs available is determined
215
Glossary
by the channel width of the
instrument.
pod See pod pair
point To point to an item, move the
mouse cursor over the item, or
position your finger over the item.
preprocessor See analysis probe.
primary branch The primary
branch is indicated in the Trigger
sequence step dialog box as either
the Then find or Trigger on
selection. The destination of the
primary branch is always the next
state in the sequence, except for the
Agilent Technologies 16517A. The
primary branch has an optional
occurrence count field that can be
used to count a number of
occurrences of the branch condition.
See also secondary branch.
probe A device to connect the
various instruments of the logic
analysis system to the target system.
There are many types of probes and
the one you should use depends on
the instrument and your data
requirements. As a verb, "to probe"
means to attach a probe to the target
system.
processor probe See emulation
probe.
216
range terms Logic analyzer
resources that represent ranges of
values to be found on labeled sets of
bits. For example, range terms could
identify a range of addresses to be
found on the address bus or a range
of data values to be found on the data
bus. In the trigger sequence, range
terms are considered to be true when
any value within the range occurs.
relative Denotes time period or
count of states between the current
state and the previous state.
remote display A remote display is
a display other than the one
connected to the product hardware.
Remote displays must be identified to
the network through an address
location.
remote session A remote session is
when you run the logic analyzer using
a display that is located away from
the product hardware.
right-click When using a mouse for
a pointing device, to right-click an
item, position the cursor over the
item, and then quickly press and
release the right mouse button.
sample A data sample is a portion of
a data set, sometimes just one point.
When an instrument samples the
target system, it is taking a single
Glossary
measurement as part of its data
acquisition cycle.
Sampling Use the selections under
the logic analyzer Sampling tab to tell
the logic analyzer how you want to
make measurements, such as State
vs. Timing.
secondary branch The secondary
branch is indicated in the Trigger
sequence step dialog box as the Else
on selection. The destination of the
secondary branch can be specified as
any other active sequence state. See
also primary branch.
session A session begins when you
start a local session or remote
session from the session manager,
and ends when you select Exit from
the main window. Exiting a session
returns all tools to their initial
configurations.
skew Skew is the difference in
channel delays between
measurement channels. Typically,
skew between modules is caused by
differences in designs of
measurement channels, and
differences in characteristics of the
electronic components within those
channels. You should adjust
measurement modules to eliminate
as much skew as possible so that it
does not affect the accuracy of your
measurements.
state measurement In a state
measurement, the logic analyzer is
clocked by a signal from the system
under test. Each time the clock signal
becomes valid, the analyzer samples
data from the system under test.
Since the analyzer is clocked by the
system, state measurements are
synchronous with the test system.
store qualification Store
qualification is only available in a
state measurement, not timing
measurements. Store qualification
allows you to specify the type of
information (all samples, no samples,
or selected states) to be stored in
memory. Use store qualification to
prevent memory from being filled
with unwanted activity such as noops or wait-loops. To set up store
qualification, use the While storing
field in a logic analyzer trigger
sequence dialog.
subnet mask A subnet mask blocks
out part of an IP address so that the
networking software can determine
whether the destination host is on a
local or remote network. It is usually
represented as decimal numbers
separated by periods; for example,
255.255.255.0. Ask your LAN
administrator if you need a the
subnet mask for your network.
217
Glossary
symbols Symbols represent
patterns and ranges of values found
on labeled sets of bits. Two kinds of
symbols are available:
•
Object file symbols - Symbols
from your source code, and
symbols generated by your
compiler. Object file symbols may
represent global variables,
functions, labels, and source line
numbers.
•
User-defined symbols - Symbols
you create.
Symbols can be used as pattern and
range terms for:
•
Searches in the listing display.
•
Triggering in logic analyzers and
in the source correlation trigger
setup.
•
Qualifying data in the filter tool
and system performance analysis
tool set.
system administrator The system
administrator is a person who
manages your system, taking care of
such tasks as adding peripheral
devices, adding new users, and doing
system backup. In general, the
system administrator is the person
you go to with questions about
implementing your software.
218
target system The system under
test, which contains the
microprocessor you are probing.
terms Terms are variables that can
be used in trigger sequences. A term
can be a single value on a label or set
of labels, any value within a range of
values on a label or set of labels, or a
glitch or edge transition on bits
within a label or set of labels.
TIM A TIM (Target Interface
Module) makes connections between
the cable from the emulation module
or emulation probe and the cable to
the debug port on the system under
test.
time-correlated Time correlated
measurements are measurements
involving more than one instrument
in which all instruments have a
common time or trigger reference.
timer terms Logic analyzer
resources that are used to measure
the time the trigger sequence
remains within one sequence step, or
a set of sequence steps. Timers can
be used to detect when a condition
lasts too long or not long enough.
They can be used to measure pulse
duration, or duration of a wait loop. A
single timer term can be used to
delay trigger until a period of time
after detection of a significant event.
Glossary
timing measurement In a timing
measurement, the logic analyzer
samples data at regular intervals
according to a clock signal internal to
the timing analyzer. Since the
analyzer is clocked by a signal that is
not related to the system under test,
timing measurements capture traces
of electrical activity over time. These
measurements are asynchronous
with the test system.
tool icon Tool icons that appear in
the workspace are representations of
the hardware and software tools
selected from the toolbox. If they are
placed directly over a current
measurement, the tools automatically
connect to that measurement. If they
are placed on an open area of the
main window, you must connect them
to a measurement using the mouse.
toolbox The Toolbox is located on
the left side of the main window. It is
used to display the available
hardware and software tools. As you
add new tools to your system, their
icons will appear in the Toolbox.
tools A tool is a stand-alone piece of
functionality. A tool can be an
instrument that acquires data, a
display for viewing data, or a postprocessing analysis helper. Tools are
represented as icons in the main
window of the interface.
trace See acquisition.
trigger sequence A trigger
sequence is a sequence of events that
you specify. The logic analyzer
compares this sequence with the
samples it is collecting to determine
when to trigger.
trigger specification A trigger
specification is a set of conditions
that must be true before the
instrument triggers.
trigger Trigger is an event that
occurs immediately after the
instrument recognizes a match
between the incoming data and the
trigger specification. Once trigger
occurs, the instrument completes its
acquisition, including any store
qualification that may be specified.
workspace The workspace is the
large area under the message bar and
to the right of the toolbox. The
workspace is where you place the
different instrument, display, and
analysis tools. Once in the workspace,
the tool icons graphically represent a
complete picture of the
measurements.
zooming In the oscilloscope or
timing analyzer, to expand and
contract the waveform along the time
base by varying the value in the s/Div
219
Glossary
field. This action allows you to select
specific portions of a particular
waveform in acquisition memory that
will be displayed on the screen. You
can view any portion of the waveform
record in acquisition memory.
220
Index
Symbols
&, 77
*, bit assignment, 58
+, label polarity, 60
-, label polarity, 60
., bit unassignment, 58
Numerics
1.5 ns sample rate, 41
16716A 167 MHz State/2 GHz
Timing Zoom Logic Analyzer, 2
16716A characteristics, 184
16716A specifications, 184
2 GHz Timing Zoom, 53
3.0 ns sample rate, 41
A
acquisition depth control, 52
acquisition memory depth, 14,
114, 152
acquisition mode, 36
acquisition modes, timing, 41
actions, 76
actions, counter, 79
actions, flag, 80, 147, 150
actions, reset occurrence counter,
82
actions, store, 70, 78
actions, timer, 78
activity indicators, 17, 56, 58, 117
advanced clocking, 115
advanced settings, eye finder, 131
advanced trigger function, after
break down, 66
advanced trigger functions, 151
advanced trigger functions,
editing, 76
AGP-2 threshold level, 57
Align to x Byte option, 161
Align to x Byte option for symbols,
161
altitude characteristics, 184
analogy, conveyor belt, 190
analysis probe, 13
analyzer (logic), understanding
triggering, 190
analyzer 2, turning on, 56
analyzer name, 52, 128
analyzer probes, general-purpose,
33
analyzer probes, termination
adapter, 33
analyzer shutdown options dialog,
53
analyzer, arming, 105
analyzer, turning back on, 53
analyzer, turning off, 53
And, combining label events, 65
arm in from IMB event, 106
arm out signal, 152
arming an analyzer, 105
arming one analyzer with the
otherstrigger', 105
ASCII, 119
ASCII format symbols, 164, 165,
166, 167
ASCII symbol file, 166
assembly language mnemonics, 87
assign pod pairs, 17
asynchronous sampling, 14
automatic sampling position
adjustment, 46, 130
B
bad data in measurement, 91
basic state measurement, example,
28
basic steps, 11
basic timing measurement,
example, 26
bit assignments, preserving, 61
bit numbering within a label, 58
bit numbers of logic analyzer
channels, 58
bit order, changing, 60
bit significance, 58
bits, reordering, 60
boolean expressions, 195
branches, 196
branches taken, storing, 70
break down trigger functions, 201
breaking down a trigger function,
66
breaking down trigger functions,
64
browsing, 160
browsing the symbol database, 160
C
cable power, probe, characteristic,
184
canceling data processing, 93
capacitive loading, 91
captured data, canceling
processing of, 93
captured data, displaying, 87
center trigger position, 52
channel counts, characteristic, 184
channels, assigning labels to, 17
channels, assigning to labels, 58
channels, maximum per label, 58
channel-to-channel skew,
characteristic, 184
Chart display tool, 89
clear flag, 80
clear menu, 76
clearing the trigger save/recall list,
84
clearing the trigger sequence, 76
clock bits as data channels, 117
clock bits warning message, 174
clock channel specifiers, 115
clock channels, inputs available as
data, 117
clock qualifier, 14
clock qualifiers, characteristic, 184
221
Index
clock setup, 14
clock setup area, 115
clock speeds and sampling
positions, 43
clock threshold level note, 57
clock threshold note, 44
clock time, specification, 184
code, assigning address offsets, 96
COFF symbol reader options, 101
comments, 167
Compare analysis tool, 89
complex triggers, 204
compressing a trigger function, 66
configuration files, 11
configuration files, loading, 109
configurations, compatibility
across models, 109
configurations, storing, 109
connectors, header, 13
consecutive occurrence counts, 77
conventional timing acquisition
modes, 41
conveyor belt analogy, 190
copying trigger function libraries,
67
copying-and-pasting trigger
sequence levels, 73
counter (occurrence), reset action,
82
counter 1 value checked as an
event, but no increment action
specified, 175
counter 2 value checked as an
event, but no increment action
specified, 175
counter actions, 79
counter events, 79
counter warning message, 175
counters, global, 79
counters, occurrence and global,
197
counting states, 69
counting states or time, 152
222
counting time, 69
cross-triggering, 105
cutting-and-pasting trigger
sequence levels, 73
D
data channels, using clock bits, 117
data displayed in symbolic form, 92
data on clocks display, 117
data valid window, 43, 206
data, displaying, 87
default storage, 199
default storing, 70, 78, 154
default storing, initially on/off, 71
default trigger sequence, 76
definition, calibration procedure,
187
definition, characteristic, 188
definition, function test, 188
definition, operational accuracy
calibration, 187
definition, specification, 187
deleting labels, 58
deleting trigger sequence levels, 72
demand driven data, 91
demultiplex clocks for pods, 44
depth of memory, characteristic,
184
digital "eye" diagram, 132, 140
display tools, other, 89
displaying captured data, 87
displaying symbols to represent
data, 92
displaying Timing Zoom data, 88
displays, listing, 87
displays, waveform, 87
Distribution display tool, 89
documenting trigger sequences,
204
duration, 77
E
edges, 65, 196
editing a named event, 83
editing advanced trigger functions,
76
ELF symbol reader options, 100
ELF/DWARF file format, 161
ELF/stabs file format, 161
else branch, 73
e-mail notify, 74
end trigger position, 52
enviromental characteristics, 184
error messages, branch expression
is too complex, 169
error messages, goto action
specifies an undefined level,
175
error messages, hardware
initialization failed, 176
error messages, maximum of 32
channels per label, 175
error messages, must assign
another pod pair to specify
actions for flags, 176
error messages, no more edge/
glitch resources, 176
error messages, no more pattern
resources available, 177
error messages, slow or missing
clock, 178
error messages, trigger function
initialization failure, 179
error messages, trigger
specification is too complex,
180
error messages, waiting for trigger,
182
errors in data, 91
evaluation order, 82
event evaluation order, 82
event list, naming, 83
events, 20, 76
events, counter, 79
Index
events, flag, 80, 147, 150
events, grouping, 82
events, label edge, 65
events, label pattern, 65
events, timer, 78
eventual occurrence counts, 77
example, 160, 161
expanding a trigger function, 66
expansion frames, using flags, 80
exporting captured data, 103
expressions, boolean, 195
eye finder, 46, 130, 206
eye finder advanced settings, 131
eye finder data, 109
eye finder selected/suggested
sampling positions, 133
F
Feedthrough type, Port Out, 80
File In tool, 103
File Out tool, 103
file versions, 95
files, 95
Find 2 edges too close together
trigger function, 147
Find 2 edges too far apart trigger
function, 147
Find edge AND pattern trigger
function, 146
Find edge trigger function, 146
Find glitch trigger function, 147
Find n-bit serial pattern, 149
Find Nth occurrence of an edge
trigger function, 146
Find pattern n consecutive times,
150
Find pattern n times, 148
Find pattern occurring too late
after edge trigger function,
147
Find pattern occurring too soon
after edge trigger function,
147
Find pattern present/absent for &,
146
Find pattern present/absent for >
duration trigger function, 146
Find pattern trigger function, 146
Find pattern1 eventually followed
by pattern2, 149
Find pattern2 n times after
pattern1, before pattern3
occurs, 150
Find pattern2 occurring
immediately after pattern1,
149
Find pattern2 occurring too late
after pattern1, 149
Find pattern2 occurring too soon
after pattern1, 149
Find too few states between
pattern1 and pattern2, 149
Find too many states between
pattern1 and pattern2, 149
Find width violation on a pattern/
pulse trigger function, 146
finding the symbol you want, 160
flag actions, 80, 147, 150
flag events, 80, 147, 150
flags, 80, 197
flags, pod pair requirements in
state mode, 176
flying lead set, 13
for more information, 24
format tab, 17
full channel, timing, 41
full/half channel configuration, 114
functions, 165
functions (trigger), advanced, 151
functions (trigger), general state,
148
functions (trigger), general timing,
146
functions, trigger, 201
G
general state trigger functions, 148
general timing trigger functions,
146
general-purpose probes, 13
getting started, 11
glitches, 65
global counters, 79, 197
global markers, 87
glossary of terms, 2
group run repetitive, 85
group run single, 85
group run with OR Trigger, 148,
150
grouping events, 82
H
half channel, timing, 41
half/full channel configuration, 114
half-channel configuration, 14
hard shutdown option, 53
header connectors, 13
help, symbols, 157
high-level language source viewer,
87
hold time, 206
HSTL threshold level, 57
humidity characteristics, 184
I
IEEE-695 file format, 161
if branch, 73
import, 119
importing previously exported
data, 103
in ASCII format, 164, 165, 166, 167
in symbol browser, 160
increment counter, 79
information, for more, 24
223
Index
input capacitance, probe,
characteristic, 184
input resistance, probe,
characteristic, 184
inserting a named event, 83
inserting labels, 58
inserting trigger sequence levels,
72
instruments, triggering other, 105
Intermodule window, 80, 106
inverse assemblers, 87
L
label edge events, 65
label pattern events, 65
label polarity, 60
label values, 65
label values, symbolic, 98
labels, 20, 56
labels, assigning channels to, 58
labels, assigning to logic analyzer
channels, 17
labels, rename/insert/delete, 58
labels, reordering bits, 60
labels, turning off or on, 61
least significant bit in label, 58
levels, sequence, 192
line numbers, 166
line numbers number base, 65
listing displays, 87
load, reducing, 91
loading, 95
loading files including symbols, 95
loading object file symbols, 95
loading trigger function libraries,
67
loading user-defined symbols, 98
logic analyzer channels, assigning
labels to, 17
logic analyzer hangs, 85
logic analyzer probes, 33
224
logic analyzer triggering,
understanding, 190
logic analyzer, testing, 107
LVCMOS threshold level, 57
LVTTL threshold level, 57
M
machines available, characteristic,
184
mail on trigger, 74
main system help page, 2
manual sampling position
adjustment, 49, 142
markers, global, 87
masking off addresses of symbols,
161
master clocks for pods, 44
maximum transitions stored, full
channel, 38
measurement doesntrun', 85
measurement, probing options, 33
measurement, state example, 28
measurement, timing example, 26
measurements, starting, 85
measurements, stopping, 85
memory and trigger, 52
memory depth, 14, 114, 152
memory depth, characteristic, 184
memory depth, setting, 52
message, trigger inhibited during
timing prestore, 180
microprocessors, probing, 13
minimum transitions stored, full
channel, 38
mode and acquisition depth, 36
mode and channel width, 36
mode and sample rate, 36
more information, 24
most significant bit in label, 58
Multiframe, using flags, 80
N
name, analyzer, 52, 128
named events, 83
naming an event list, 83
negative logic, 60
netlist, 119
no more pattern resources
message, 177
note, clock, characteristic, 184
number base, 65
O
object file symbol files, 95
object file symbols, 160
occurrence count, 77
occurrence counter, characteristic,
184
occurrence counter, reset action,
82
occurrence counters, 197
odd-numbered addresses, 161
odd-numbered addresses
represented by symbols, 161
offset addresses, assigning, 96
Offset By option of the symbol
browser, 160
OMF96 file format, 161
OMFx86 file format, 161
operating environment
characteristics, 184
operators in label events, 65
OR Trigger trigger function, 148,
150
Or, combining label events, 65
order of event evaluation, 82
other instruments, triggering, 105
override default storing, 70
P
pasting trigger sequence levels, 73
Pattern field, 160
patterns, 65
Index
pause timer, 78
performance verification, 107
period, sample, 42
pod assignment dialog, 56
pod clocking, demultiplex, 44
pod pairs, assigning, 17
pod thresholds, setting, 57
pods, assigning, 56
pods, specifying state clock, 44
polarity, label, 60
Port Out, using flags to drive, 80
positive logic, 60
power through pod cables,
characteristic, 184
predefined trigger functions, 145
prefetch, triggering beyond, 160
present for >, 77
preserving bit assignments, 61
previous trigger setup, recalling, 84
printing captured data, 103
probe leads, 33
probes, general-purpose, 13
probes, individual signal, 33
probing, overview, 33
problems making measurements,
107
processing of captured data,
canceling, 93
pulse clear flag, 80
pulse set flag, 80
Q
qualifier, clock, 14
qualifiers, clocks, characteristic,
184
R
R, bit assignment, 58, 60
ranges, 65, 196
ranges and reordered bits, 60
rate, sample, 42
readers.ini file, 100
recalling trigger setups, 83
refining measurements, 11
re-importing captured data, 103
relocating sections of code, 96
renaming labels, 58
reordered bits, 65
repetitive data display, 91
repetitive group run, 85
repetitive run, 85
replacing trigger sequence levels,
72
replacing with a named event, 83
reset counter, 79
reset occurrence counter action,
82
results, 160
resume timer, 78
run all, 85
run repetitive, 85
run single, 85
Run until user stop trigger
function, 146, 149
S
sample period, 14, 42, 114, 152
sample period, characteristic, 184
sample rate, 1.5 ns, 41
sample rate, 3.0 ns, 41
sample rate, setting, 36
samples, storing, 70
sampling mode, 36, 128
sampling mode, state, 70
sampling modes, 14
sampling position, automatic
adjustment, 46, 130
sampling position, manual
adjustment, 49, 142
sampling positions, 43
sampling positions, eye finder
behavior, 133
save/recall list (trigger), clearing,
84
Save/Recall subtab, 155
saving trigger setups, 83
Search Pattern field, 160
searching captured data, 87
searching the symbol database, 160
sections, 164
Select Case statement, 196
selected sampling positions, eye
finder, 133
selecting a trigger function, 64
self test, 107
sequence (trigger), clearing, 76
sequence level storage, 199
sequence levels, 192
sequence levels, characteristic, 184
sequence levels, cutting/copyingand-pasting, 73
sequence levels, inserting/
replacing/deleting, 72
sequencer rate, fastest, 41
sequencer, maximum levels,
characteristic, 184
Serial Analysis tool, 89
set flag, 80
set up, trigger sequence, 73
Set/clear/pulse flag trigger
function, 80, 147, 150
setting threshold voltages, 57
settings (trigger), clearing, 76
settings, eye finder advanced, 131
settings, saving, 109
Setup Assistant, 13
setup time, 206
setup/hold, 49, 142
setup/hold (logic analyzer), 43
setup/hold time, specification, 184
setup/hold window, 46, 49, 130,
142, 206
Signal Seeker, 43
single group run, 85
single run, 85
skew, channel-to-channel,
characteristic, 184
225
Index
slave clocks for pods, 44
slow clock message, 178
SMTP, 75
soft shutdown option, 53
source line numbers, 166
source viewer, 87
specifications and characteristics,
184
speed, state/timing, characteristic,
184
SSTL2 threshold level, 57
SSTL3 threshold level, 57
stable regions, 133
Stabs symbol reader options, 101
standard buses, probing, 13
standard lead set, 13
start address, 167
start timer, 78
start trigger position, 52
starting measurements, 85
state analyzer configuration, 14
state clock speed, specification,
184
state clocks, 115
state clocks, characteristic, 184
state clocks, master/slave/both, 44
state count, 152
state measurement, example, 28
state mode, 14
state mode clocks, setup/hold, 49,
142
state mode controls, 14
state mode trigger features, 69
state modes, characteristic, 184
state sampling mode, 43, 70, 128
state trigger functions, 145
state trigger functions, general, 148
states, counting, 69
stop timer, 78
stopping measurements, 85
storage qualification, 199
store action, 78
store actions, 70
226
Store n samples, 150
Store nothing until pattern occurs,
149
Store pattern2 until pattern1
occurs, 148
store qualification, 70
store qualified, 37
Store range until pattern occurs,
148
storing trigger setups, 83
storing, default, 70, 78, 154
strategies for setting up triggers,
201
suggested sampling positions, eye
finder, 133
summary of triggering capabilities,
192
Switch statement, 196
symbol demangling, 100
symbol file formats, 161
symbol file versions, 95
Symbol Selector dialog, 65
symbol selector dialog, 159
symbolic label values, 98
symbols, 160
symbols number base, 65
symbols, displaying to represent
data, 92
symbols, loading object file
symbols, 95
symbols, loading user-defined, 98
symbols, outside defined sections,
100
symbols, types and use, 157
symbols, user-defined, 97
symbols, using, 94
synchronous sampling, 14
System Performance Analysis
toolset, 89
T
tab, symbols, 157
temperature characteristics, 184
termination adapters, 13
terminations, 13
then branch, 73
threshold accuracy, specification,
184
threshold range, probe,
characteristic, 184
threshold voltages, 17, 132, 140
threshold voltages, setting, 57
TI COFF file format, 161
time count, 152
time count, characteristic, 184
time duration, 77
time interval accuracy,
characteristic, 184
time tag resolution, characteristic,
184
time, counting, 69
timer 1 value checked as an event,
but no start action specified,
178
timer 2 value checked as an event,
but no start action specified,
178
timer actions, 78
timer events, 78
timer warning message, 178
timers, 78, 198
timing analysis characteristics, 184
timing analyzer configuration, 14
timing measurement, example, 26
timing mode, 14
timing mode controls, 14
timing modes, 37, 41
timing sampling mode, 40, 128
timing speed, characteristic, 184
timing trigger functions, 145
timing trigger functions, general,
146
Timing Zoom data, 88
timing, memory depth, 41
Index
tip capacitance, probe,
characteristic, 184
transitional timing, 37
transitional timing considerations,
39
trigger and send e-mail, 74
trigger characteristics, 184
trigger function libraries, 67
trigger function, breaking down, 66
trigger function, compressing, 66
trigger function, expanding, 66
trigger functions, 64, 201
trigger functions, advanced, 151
trigger functions, branching, 73
trigger functions, general state, 148
trigger functions, general timing,
146
trigger functions, predefined, 145
trigger functions, selecting, 64
trigger inhibited informational
message, 180
trigger position, 14, 114, 152, 190
trigger position control, 52
trigger save/recall list, clearing, 84
trigger sequence, 192
trigger sequence branches, 73
trigger sequence label values, 65
trigger sequence levels, goto, 73
trigger sequence, cutting/copyingand-pasting levels, 73
trigger sequence, default, 76
trigger sequence, editing, 71
trigger sequence, inserting/
replacing/deleting levels, 72
trigger sequencer rate, fastest, 41
trigger sequences, documenting,
204
trigger set up, 73
trigger setup, recalling, 84
trigger setup, saving, 84
trigger setups, saving/recalling, 83
trigger tab, reference, 144
trigger tab, use, 20
trigger, poststore, 52
trigger, substeps, 20
triggering capabilities, summary,
192
triggering on a symbol, 160, 161
triggering on a symbol beyond
prefetch depth, 160
triggering other instruments, 105
triggering, understanding logic
analyzer, 190
triggers, complex, 204
triggers, strategies for setting up,
201
troubleshooting the logic analyzer,
107
TTL threshold level, 57
U
unassigned bits, 58
understanding logic analyzer
triggering, 190
Undo command, 66
ungrouping events, 82
unloading trigger function
libraries, 67
user threshold logic level, 57
user-defined symbols, 97
user-defined symbols, loading, 98
user-level trigger functions, 73
Wait for flag trigger function, 80,
147, 150
wait for other machine to trigger,
179
wait for second analyzer to trigger,
105
Wait for second analyzer to trigger
trigger function, 147, 150
Wait n external clock states, 150
Wait t seconds trigger function, 147
warning messages, clock bits, 174
warning messages, counter not
incremented, 175
warning messages, flags, 169
warning messages, no arm in from
IMB event, 183
warning messages, no trigger
action found in the trace
specification, 177
warning messages, timer never
started, 178
warranty, what is covered, 187
waveform displays, 87
While storing pattern2, find
pattern1, 149
wildcard characters, 160
V
variables, 166
versions, 95
versions of symbol files, 95
vibration characteristics, 184
voltages, threshold, 17
W
wait for arm in, 106, 179
Wait for arm in trigger function,
147, 150
227
Index
228
Publication Number: 5988-9032EN