6003A Safety Information

6003A Safety Information
6003A
Three Phase Power Calibrator
Safety Information
Introduction
Safety Specifications
This document supplies safety information for the
6003A Three Phase Power Calibrator (the Product).
Input Power
Contact Fluke Calibration
To contact Fluke Calibration, call one of the following
telephone numbers:
•
Technical Support USA: 1-877-355-3225
•
Calibration/Repair USA: 1-877-355-3225
•
Canada: 1-800-36-FLUKE (1-800-363-5853)
•
Europe: +31-40-2675-200
•
Japan: +81-3-6714-3114
•
Singapore: +65-6799-5566
•
China: +86-400-810-3435
•
Brazil: +55-11-3759-7600
•
Anywhere in the world: +1-425-446-6110
To see product information and download the latest
manual supplements, visit Fluke Calibration’s website
at www.flukecal.com.
Voltage ...................................... 115 V/230 V ±10 %, 47 Hz to
63 Hz, 1875 VA max.
Temperature
Operating .............................. 5 °C to 40 °C
Calibration ............................ 21 °C to 25 °C
Storage ................................. -10 °C to 55 °C
Transit .................................. -15 °C to 60 °C
Relative Humidity (non-condensing)
Operating ............................. <80 % 5 °C to 31 °C ramping
linearly down to 50 % at 35 °C
Storage ................................ <90 % -10 °C to 55 °C
Altitude
Operating .............................. 2000 m maximum
Storage ................................. 0 m to 12000 m
Safety ............................................. IEC 61010-1, IEC 61010-2030, Overvoltage category II,
Pollution Degree 2
To register your product, visit
http://flukecal.com/register-product.
PN 4406535
May 2014
©2014 Fluke Corporation. All rights reserved.
Product specifications are subject to change without notice. All product names are trademarks of their respective companies.
1
6003A
Safety Information
Safety Information
A Warning identifies conditions and procedures that
are dangerous to the user.
•
Do not apply more than the rated voltage,
between the terminals or between each
terminal and earth ground.
•
Do not touch voltages > 30 V ac rms, 42 V ac
peak, or 60 V dc.
•
Use only cables with correct voltage ratings.
•
Never connect the Product current output
terminals to a voltage source other than the
Product voltage output terminals.
Warning
To prevent possible electrical shock, fire, or
personal injury:
•
Read all safety information before you use
the Product.
•
Carefully read all instructions.
•
•
Two people are necessary to move or lift the
Product. Always use correct lifting
techniques. The Product has a mass of
62 kg.
Use extreme caution around the output
terminals. Lethal voltages may be present.
•
Make sure the Product is in standby mode
and external circuits are not energized
before you connect or disconnect cables
between the Product and the equipment
under test.
•
Do not enable voltage outputs unless the
cables between the Product and equipment
under test are connected or disconnected at
both ends of the cable.
•
Do not connect any connector or terminal,
other than the mains power inlet, to line
power.
•
Whenever it is likely that safety protection
has been impaired, the Product must be
made inoperative and be secured against
any unintended operation. Inform qualified
maintenance or repair personnel. Safety
protection is likely to be impaired if, for
example, the Product shows visible damage
or fails to operate normally.
•
Use the Product only as specified, or the
protection supplied by the Product can be
compromised.
•
Do not use the Product if it operates
incorrectly.
•
Do not operate the Product with covers
removed or the case open. Hazardous
voltage exposure is possible.
Do not connect to live output terminals. The
Product can supply voltages that can cause
death. Standby mode is not sufficient to
prevent electrical shock.
•
Remove the input signals before you clean
the Product.
•
Use only specified replacement parts.
Before changing the fuse, turn the Product
off and remove the mains power cord. Stop
for two minutes to let the power assemblies
discharge before you open the fuse door.
•
Use only specified replacement fuses.
•
Have an approved technician repair the
Product.
•
Use this Product indoors only.
•
Do not use the Product around explosive
gas, vapor, or in damp or wet environments.
•
Do not use an extension cord or adapter
plug.
•
Do not put the Product where access to the
mains power cord is blocked.
•
Use only the mains power cord and
connector approved for the voltage and plug
configuration in your country and rated for
the Product.
•
Make sure the ground conductor in the
mains power cord is connected to a
protective earth ground. Disruption of the
protective earth could put voltage on the
chassis that could cause death.
•
Replace the mains power cord if the
insulation is damaged or if the insulation
shows signs of wear.
•
Examine the case before you use the
Product. Look for cracks or missing plastic.
Carefully look at the insulation around the
terminals.
•
•
2
Three Phase Power Calibrator
Safety Information
The symbols used in the manuals and on the Product
are shown below.
Symbols
Symbol
Description

Hazardous voltage. Risk of electric shock.

Risk of Danger. Important information. See
Manual.

Earth Terminal

Conforms to relevant North American
Safety Standards.

Conforms to European Union directives

Conforms to relevant Australian EMC
standards

Conforms to relevant South Korean EMC
Standards.

This product complies with the WEEE
Directive (2002/96/EC) marking
requirements. The affixed label indicates
that you must not discard this
electrical/electronic product in domestic
household waste. Product Category: With
reference to the equipment types in the
WEEE Directive Annex I, this product is
classed as category 9 "Monitoring and
Control Instrumentation" product. Do not
dispose of this product as unsorted
municipal waste. Go to Fluke’s website for
recycling information.
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and
workmanship for one year from the date of purchase. This
warranty does not cover fuses, disposable batteries, or
damage from accident, neglect, misuse, alteration,
contamination, or abnormal conditions of operation or
handling. Resellers are not authorized to extend any other
warranty on Fluke’s behalf. To obtain service during the
warranty period, contact your nearest Fluke authorized
service center to obtain return authorization information,
then send the product to that Service Center with a
description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS
NOT LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR
LOSSES, ARISING FROM ANY CAUSE OR THEORY.
Since some states or countries do not allow the exclusion or
limitation of an implied warranty or of incidental or
consequential damages, this limitation of liability may not
apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
11/99
3
6003A
Safety Information