Below you will find brief information for Pressure Transducer PX5000 Series. This high performance sputtered thin-film pressure transducer is designed for high-pressure environments. This transducer features a 4-active-arm bridge using sputtered-deposited thin-film elements for increased accuracy and reliability. The sensor is compensated for a temperature range of -65° to 250°F and can withstand a burst pressure up to 20,000 PSI.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

PX5000 Series Pressure Transducer Datasheet | Manualzz
 An OMEGA Technologies Company
PX5000 Series
High Performance Sputtered Thinfilm
Pressure Transducer
INSTRUCTION
M3003/1097
SPECIFICATIONS:
Electrical:
Excitation: 10 V dc
Output FSO: 30 mV typical, 2.6 mV minimum
input Resistance: 1000 ohms typical, 800
ohms minimum, 1500 ohms maximum
Output Resistance: 1000 ohms +200 ohms.
For 350 ohm bridge see custom configurations
Insulation Resistance: 500 megohms minimum
@ 45 Vdc over the compensated range
Sensing Element: 4 active-arm bridge using
sputtered-deposited thin-film elements
Performance:
Accuracy: Combined Linearity, Hysteresis and
Nonrepeatability: 10.10% FSO 0-15 thru 0-999
PSI +0.15% FSO 1,000 PSI and above
Calibration Stability: +0.1% for 18 months
Reliability: MTBF = 100,000 hours
Environmental:
Operating Temp Range: -54° to 150°C
(-65° to 300°F)
Compensated Temp Range:
-54° to +121°C (-65° to +250°F)
Thermal Effects: Span: +0.005% FSO/°F
Zero: +0.005% FSO/°F
Vibration Sensitivity: At 35g peak sinusoidal
vibration from 10 Hz to 2000 Hz (¥" D.A.),
the output shall not exceed 0.04% FSO/g for
15 PSI range to 0.003% FSO/g for 1000 PSI
and above
Natural Frequency: 50 kHz for 5,000 PSI range,
decreasing logarithmically to 5kHz for 15 PSI range
Shock: Qualification level of 100g, 11 msec
half sine wave without damage
Humidity: Qualified per MIL-STD-810
Electrostatic Discharge (ESD): Protected to
15kV
Pressure:
Pressure: 2.0 times rated pressure or 15,000
PSI whichever is less, will not cause changes in
performance beyond the specified tolerances
Burst Pressure: 3.0 times rated pressure or
20,000 PSI whichever is less, will not cause
rupture of the pressure containment cavity.
Mechanical:
Wetted Parts: 17-4 PH Stainless Steel or
15-5 PH Stainless Steel
Pressure Fitting: %-20 male per MS33656-4
Mounting Sensitivity: Less than 0.05% FSO
change in zero when mounted by the pressure
port to a manifold at a torque of 15 ft-lb on
a solid copper gasket
Mating Connector: (Not Included)
PTS06A-10-6S
Weight: 5 0z (145g) maximum
688 + .010 —
(17.47 + 25)
WRENCH FLATS
Pins Functions
|A|— POSITIVE INPUT
|B|—— POSITIVE OUTPUT
|C|—— NEGATIVE OUTPUT
D | —— NEGATIVE INPUT
E|—— SHORTED TO "D"
| F|—— SHORTED TO "C"
Dimensions shown in Inches (mm)
<— «— 196MAX _ >.
‚25 +.01 — > (49.8). > IDENTIFIED
(6.35 + .25) A INTHIS
| Т | DA | Ш
1.01 МАХ | |
(25.7) #—-—-—-— но
DIA.
| / qu
PRESSURE FITTING
PER MS33656-4
ELECTRICAL RECEPTACLE
BENDIX PTIH-10-6P (101)
(OR EQUIVALENT)
MATING CONNECTOR
BENDIX PTS06A-10-6S (SR)
(OR EQUIV.) NOT SUPPLIED
omega.com | Servicing Europe:
®
СЕОМЕСА Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands
| . . . Tel: (31) 20 6418405 FAX: (31) 20 6434643
ON net On-Line Service . Internet e-mail Toll Free in Benelux: 06 0993344
ttp://www.omega.com [email protected] e-mail: [email protected]
.. . Czech Republic: ul. Rude armady 1868, 733 01 Karvina-Hranice, Czech Republic
Servicing North America: Tel: 420 (69) 6311627 FAX: 420 (69) 6311114
USA: One Omega Drive, Box 4047 e-mail: [email protected]
ISO 9001 Certified Stamford, CT 06907-0047 France: 9, rue Denis Papin, 78190 Trappes
Tel: (203) 359-1660 FAX: (203) 359-7700 Tel: (33) 130-621-400 FAX: (33) 130-699-120
e-mail: [email protected] Toll Free in France: 0800-4-06342
Canada: 976 Bergar e-mail: [email protected]
Laval (Quebec) H7L 541 Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany
Tel: (514) 856-6928 FAX: (514) 856-6886 Tel: 49 (07056) 3017 FAX: 49 (07056) 8540
e-mail: [email protected] Toll Free in Germany: 0130 11 21 66
For immediate technical or application assistance: e-mail: [email protected]
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA™ United Kingdom: 25 Swannington Road, P.O. Box 7, Omega Drive,
Customer Service: 1-800-622-2378 / 1-800-622-BEST™™ ISO 9002 Certified Broughton Astley, Leicestershire, Irlam, Manchester,
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN™ LE9 6TU, England M44 5EX, England
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA Tel: 44 (1455) 285520 Tel: 44 (161) 777-6611
Mexico and FAX: 44 (1455) 283912 FAX: 44 (161) 777-6622
Latin America: Tel: (95) 800-TC-OMEGA™ FAX: (95) 203-359-7807 Toll Free in England: 0800-488-488
En Español: (203) 359-7803 e-mail [email protected] e-mail: [email protected]
К is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that
apply. OMEGA is constantly pursuing certification of its products to the European New Approach
Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct but OMEGA Engineering, Inc. accepts
no liability for any errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, patient connected applications.
a
A NC ит
; Are eee =
a - a
Fes
Da a
ADE NAAA
CTRA CN ER NANA A EE a
WARRAN I Y/DISCLAIMER A a TENA Ta ce
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months
from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year
product warranty to cover handling and shipping time. This ensures that OMEGAs customers receive maximum
coverage on each product.
If the unit should malfunction, it must be returned to the factory for evaluation. OMEGA Customer Service Department will
issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA, if the
unit is found to be defective it will be repaired or replaced at no charge. OMEGA's WARRANTY does not apply to defects
resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of
design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having
been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or
vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGAs control.
Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes
responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its
products in accordance with information provided by OMEGA, either verbal or written. OMEGA warrants only that
the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL —
IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are
exclusive and the total liability of OMEGA with respect to this order, whether based on contract, warranty,
negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon
which liability is based. In no event shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under
10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medica! applications or used on humans. Should
any Product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any
way, OMEGA assumes no responsibility as set forth in our basic WARRANTY/ DISCLAIMER language, and additionally,
purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising out of the use
of the Product(s) in such a manner.
nt BA
ER EE
RARE Tae ESS
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY
PRODUCTI(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGAS
CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then
be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.
фасона сея оо
CE eh
hh res
et
nn A ee
: A A O
FOR WARRANTY RETURNS, please have the following FOR NON-WARRANTY REPAIRS, consult OMEGA for
information available BEFORE contacting OMEGA: current repair charges. Have the following information
1. PO. number under which the product was PURCHASED, available BEFORE contacting OMEGA:
2. Model and serial number of the product under 1. PO. number to cover the COST of the repair,
warranty, and 2. Model and serial number of product, and
3. Repair instructions and/or specific problems relative to 3. Repair instructions and/or specific problems relative to
the product. the product.
OMEGA policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in
technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 1996 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced
to any electronic medium or machine-readable form, in whole or in part, without prior written consent of OMEGA ENGINEERING, INC.

advertisement

Key Features

  • High Performance
  • Sputtered Thin-film
  • 4-Active-Arm Bridge
  • Accuracy: ±0.1% FSO
  • Calibration Stability: ±0.1% for 18 months
  • Reliability: MTBF = 100,000 hours
  • Operating Temp Range: -65° to 300°F
  • Compensated Temp Range: -65° to +250°F
  • Burst Pressure: 3.0 times rated pressure or 20,000 PSI whichever is less
  • 17-4 PH or 15-5 PH Stainless Steel Wetted Parts

Frequently Answers and Questions

What is the compensated temperature range of the PX5000 Series pressure transducer?
The compensated temperature range of the PX5000 Series pressure transducer is -65° to +250°F.
What is the burst pressure of the PX5000 Series pressure transducer?
The burst pressure of the PX5000 Series pressure transducer is 3.0 times rated pressure or 20,000 PSI whichever is less.
What is the material used for the wetted parts of the PX5000 Series pressure transducer?
The wetted parts of the PX5000 Series pressure transducer are made of 17-4 PH or 15-5 PH Stainless Steel.
What is the pressure fitting used on the PX5000 Series pressure transducer?
The pressure fitting used on the PX5000 Series pressure transducer is 1/4-20 male per MS33656-4.