Installation Instructions for Mindray Reagent Set

Required ONLY when changing to CDS reagents from other non-CDS products.

For proper performance DO NOT mix CDS reagents with non-CDS products

TM

To achieve optimal performance with CDS NextGeneration Reagents, it is important to

To prime Diluent and Rinse

1. Remove pickup tubes from diluent and rinse containers, press REAGENT

2.

PRIME CYCLE

When reagent empty alarm is activated, place pickup tubes in appropriate CDS

3.

Reagents (Diluent—CDS 501-201 or 501-202, Rinse—501-204 or 501-205), then press CLEAR ALARM to continue reagent prime

Press REAGENT PRIME three (3) more times

To prime Lyse

1.

2.

Remove lyse pickup line from container, press LYSE REAGENT PRIME

When reagent empty alarm is activated, place pickup tube in CDS Lytic Agent (CDS

501-203), then press CLEAR ALARM to continue reagent prime

3. Press Lyse prime three (3) more times

4. Run three (3) BACKGROUND CYCLES to complete reagent exchange

The instrument is now completely primed with

Proceed with normal instrument operation, complete reagent log.

If you experience any problems or have any questions, please call 800-453-3328.

TM

CDS NextGeneration is a trademark of Clinical Diagnostic Solutions, Inc., Plantation, Florida.

®

Mindray is a registered trademark of Shenzhen Mindray Bio-Medical Electronics Co. Ltd., China

advertisement

Key Features

  • Complete priming for optimal performance
  • CDS NextGeneration reagents compatibility
  • Easy-to-follow instructions
  • Clear alarm prompts during priming
  • Background cycles for complete reagent exchange

Frequently Answers and Questions

What should I do if I experience problems with the reagent set?
Contact Mindray at 800-453-3328 for assistance.
When do I need to prime the Mindray instrument with the CDS NextGeneration Reagent Set?
You need to prime the instrument only when switching from non-CDS reagents to CDS NextGeneration reagents.
Why is it important to completely prime the Mindray instrument before using CDS NextGeneration reagents?
Complete priming ensures all components of the instrument are properly filled with the new CDS NextGeneration reagents, which is vital for accurate and reliable analytical results.
What are the steps for priming the diluent and rinse?
Remove the pickup tubes from the diluent and rinse containers. Press REAGENT PRIME CYCLE. When the reagent empty alarm activates, place the pickup tubes in the appropriate CDS Reagents and press CLEAR ALARM to continue the reagent prime. Repeat this process three more times.
What are the steps for priming the lyse?
Remove the lyse pickup line from the container. Press LYSE REAGENT PRIME. When the reagent empty alarm activates, place the pickup tube in the CDS Lytic Agent and press CLEAR ALARM to continue the reagent prime. Repeat this process three more times and run three BACKGROUND CYCLES to complete the reagent exchange.