Mindray NextGeneration Reagent Set Installation Instructions
The Mindray NextGeneration Reagent Set is a critical component for achieving optimal performance with your Mindray instrument. These reagents are designed to provide accurate and reliable results in your laboratory analyses.
advertisement
◉
Touch to zoom
advertisement
Key Features
- Complete priming for optimal performance
- CDS NextGeneration reagents compatibility
- Easy-to-follow instructions
- Clear alarm prompts during priming
- Background cycles for complete reagent exchange
Frequently Answers and Questions
What should I do if I experience problems with the reagent set?
Contact Mindray at 800-453-3328 for assistance.
When do I need to prime the Mindray instrument with the CDS NextGeneration Reagent Set?
You need to prime the instrument only when switching from non-CDS reagents to CDS NextGeneration reagents.
Why is it important to completely prime the Mindray instrument before using CDS NextGeneration reagents?
Complete priming ensures all components of the instrument are properly filled with the new CDS NextGeneration reagents, which is vital for accurate and reliable analytical results.
What are the steps for priming the diluent and rinse?
Remove the pickup tubes from the diluent and rinse containers. Press REAGENT PRIME CYCLE. When the reagent empty alarm activates, place the pickup tubes in the appropriate CDS Reagents and press CLEAR ALARM to continue the reagent prime. Repeat this process three more times.
What are the steps for priming the lyse?
Remove the lyse pickup line from the container. Press LYSE REAGENT PRIME. When the reagent empty alarm activates, place the pickup tube in the CDS Lytic Agent and press CLEAR ALARM to continue the reagent prime. Repeat this process three more times and run three BACKGROUND CYCLES to complete the reagent exchange.