User’s Guide VetPro 1000 Mobile Dental

User’s Guide VetPro 1000 Mobile Dental
VetPro® 1000
Mobile Dental
Delivery System
For Models:
8000-001
8000-002
8000-003
8000-004
8000-005
8000-006
8000-006S
8000-007
8000-007S
8000-008
8000-008S
User’s Guide
003-1879-00 Rev T (11/7/15)
© 2011 Midmark Corp.
Style L
Product Information
(The information below is required when calling for service.)
Dealer :
Date of Purchase:
Model / Serial Number:
Midmark Authorized Service
Company:
Model / Serial Number Location
2
© 2011 Midmark Corp.
Table of Contents
Important Information
Safety Symbols.................................................................................................................... 4
Intended Use........................................................................................................................ 5
Electromagnetic Interference............................................................................................... 5
Disposal of Equipment......................................................................................................... 5
Authorized EU Representative............................................................................................. 5
Transportation / Storage / Operating Conditions.................................................................. 5
Transport
Transporting Dental System................................................................................................. 6
Operation
Electrical System.................................................................................................................. 7
Water System....................................................................................................................... 8
Air System............................................................................................................................ 9
Delivery System................................................................................................................... 10
Handpiece Care................................................................................................................... 11
Syringe Tip Replacement & Handpeice Installation............................................................. 12
Drive Air Adjustment & Locking Casters............................................................................... 13
Air Pressure Regulator & Water Flow Adjustment................................................................ 14
Coolant Air Adjustment......................................................................................................... 15
Cleaning / Maintenance
Cleaning............................................................................................................................... 16
Maintenance......................................................................................................................... 18
Calling for Service................................................................................................................ 22
Specifications
Drawing................................................................................................................................ 23
Weights / Dimensions........................................................................................................... 24
Electrical............................................................................................................................... 25
Model Identification / Compliance Chart............................................................................... 26
Warranty
Warranty Information............................................................................................................ 27
For installation information go to:
Technical Library at Midmark.com
3
© 2011 Midmark Corp.
Important Information
Safety Symbols
WARNING
Indicates a potentially hazardous situation which could result in serious injury.
Caution
Indicates a potentially hazardous situation which may result in minor or moderate injury. It may
also be used to alert against unsafe practices.
Equipment Alert
Note
Amplifies a procedure, practice, or condition.
Indicates a potentially hazardous
situation which could result in equipment
damage if not avoided.
Proper Shipping
Orientation
Handle With Care
Type B, Applied Part
Maximum stacking
height
(Do not stack)
Humidity Limit
Type BF, Applied Part
Fragile
Keep Dry
Refer to Instruction
Manual/Booklet
Protective Earth Ground
Fuse Rating
Specifications
LED Scaler Handpiece
Duty Cycle
Pressure Limit
100 F
38 C
Temperature Limit
TUVRheinland
C
US
TUV - Domestic
Models Only
23 F
-5 C
Do Not Tumble
4
© 2011 Midmark Corp.
Intended Use
The Midmark® VetPro® 1000 is a mobile dental delivery system intended to give qualified veterinarians
and/or technicians a self-contained, compact, portable unit for performing veterinary dental procedures.
Electromagnetic Interference
This product is designed and built to minimize electromagnetic interference with other devices.
However, if interference is noticed between another device and this cart:
• Remove interfering device from room
• Plug cart into an isolated circuit
• Increase separation between Midmark product and interfering device
• Contact Midmark if interference persists
Disposal of Equipment
At the end of product life, the unit(s), accessories, and other consumable goods may be
contaminated from normal use. Consult local codes and ordinances for proper disposal of
equipment and other consumable goods.
Authorized CE Representative
Countries in the EEC should direct all questions, incidents, and complaints to Midmark’s Authorized CE
representative listed below:
Midmark EMEA Ltd
Beech House
First Floor, East Wing
Ancells Business Park
Fleet
Hampshire GU51 2UN
United Kingdom
Tel: + 44 (0) 1252 360 940
Fax: + 44 (0) 1252 360 941
Transportation / Storage / Operating Conditions
Transportation / Storage Temperature Range:........................................23°F to 100°F (-5°C to 38°C)
Operating Temperature Range:...............................................................50°F to 104°F (+10°C to +40°C)
Transportation / Storage Relative Humidity:............................................10% to 90% (non-condensing)
Operating Relative Humidity:...................................................................30% to 75% (non-condensing)
Transportation / Storage Atmospheric Pressure:....................................50 kPA (7.2 psi) to 106 kPA (15.3 psi)
Operating Atmospheric Pressure:...................................................700 hPA (20 in Hg) to 1060 hPA (31 in Hg)
5
© 2011 Midmark Corp.
Transport
WARNING
Exercise care when transporting over thresholds and other low lying objects.
Failure to do so can lead to injury.
When transporting over large thresholds or steps...
A) Pull cart up and over onto level surface.
Threshold
6
© 2011 Midmark Corp.
Operation
Electrical System
On/Off Switch
On/Off Switch is located on the back of the cart. It controls the main power source and must be ON to
operate the delivery system.
Circuit Breakers
There are two circuit breakers located on the lower back panel of the cart. If the unit’s maximum load is
exceeded, the circuit breakers interrupt power to the unit. Maximum load is 115 VAC, 6 amps (domestic)
230 VAC, 3 amps (export).
( I ) On Switch
(O) Off Switch
Circuit Breakers
Tripped
(No Power)
Reset
(Power)
Mains
Disconnect
6
A
M
P
6
A
M
P
VA1227
Mains Disconnect
Power can be terminated by unplugging the mains disconnect or turning switch off.
7
© 2011 Midmark Corp.
Operation
Water System
Water Bottles
Caution
Failure to properly seal water bottles can lead to patient injury.
The unit has two 1 liter bottles. The water bottles supply water to the handpieces, air/water syringe and a
quick disconnect on the back of the cart. Air flows through the water regulator and the toggle switch to the
water bottles. The water system is regulated to 40 PSI. Use the water needle valve to adjust water flow
for each instrument. The toggle switch between the water bottles that controls which bottle will supply the
water. When the toggle is in the center neither water bottle is pressurized. Move the toggle toward the
left to pressurize the left bottle and depressurize the right bottle. Toggle to the right to pressurize the right
bottle and depressurize the left bottle.
Water Quick Disconnect
The water quick disconnect has a blue ring. You can attach accessories to the cart that will utilize the
water from the cart. You may need to add a regulator if attaching equipment that can not handle 40 PSI.
Water
Needle
Valve
Water Quick
Disconnect
Water
Regulator
Toggle
Water
Bottles
VA1229
To fill water bottles...
A) Move toggle to center.
B) Remove water bottle and fill. (Distilled Water Is Recommended)
C) Replace water bottle and move toggle toward filled water bottle.
Note: Water bottles should be checked in between each patient and filled if necessary.
8
© 2011 Midmark Corp.
Operation
Air System
Compressor
The air system is a 1/3 HP compressor with a 2 liter air tank and regulator. It supplies air for the
handpieces, air/water syringe, and a quick disconnect on the back of the cart. The compressor will cycle
between 70 to 100 PSI. The regulator is factory set at 65 PSI.
Drive Air
Adjustment Knob
Coolant Air
Adjustment Screw
Air Quick Disconnect
Foot Control Holder
Equipment Alert
Do not block air vent.
Air Vent
Compressor Compartment
Air Quick Disconnect
VA1231
The air quick disconnect has a red ring. You can attach accessories to the cart that will utilize the air from
the cart. You may need to add a regulator if attaching equipment that can not handle 100 PSI.
9
© 2011 Midmark Corp.
Operation
Delivery System
Handpieces
Caution
Handpieces and Syringe Tip should be changed
or disinfected between each patient.
The delivery system controls up to four handpieces. The handpiece pressure is controlled by handpiece
drive air adjustment valves. The gauge on front of the delivery system indicates the handpiece pressure.
Handpiece pressure should not exceed 60 PSI under normal operation. The scaler tool (optional) has an
adjustment knob for the intensity level.
Scaler
Gauge
Adjustment
Handpiece
Holders
Equipment Alert
Refer to handpiece manufacturer for suggested PSI
settings. Only check one handpiece
at a time.
VA1233
Air/Water Syringe
There are air and water buttons on top of the syringe. When both buttons are depressed you will get a
mist of both water and air. Air and water adjustments are located under the syringe holder. Insert flat head
screwdriver into adjustment screw(s). Counter-Clockwise to increase flow and Clockwise to decrease flow.
Turn screw until desired air/water flow is coming through the instrument.
Air
Water
Water
Air
10
© 2011 Midmark Corp.
Operation
Prophy
Angle
Prophy Angle Replacement
To Change Prophy Angle...
A) Rotate chuck to the right to unlock.
B) Install Prophy angle, turn chuck release ring left to tighten.
Burr Replacement
To Change Bur in High Speed Handpiece...
A)
B)
C)
Push button on back of handpiece and pull the bur out.
Push bur into handpiece. (The shank will only go in half-way)
Then push button on back of handpiece and push the bur in until
shank is all the way in.
Note: Friction grip burs (FG Style) are designed so that the entire straight portion of
the shaft should be inside the chuck of the handpiece. You should only see the tapered
portion of the bur if installed properly. New bur recommended for each procedure.
Finish Sliding Bur Into
Handpiece
Push Bur Into Handpiece
Push
Button
This Portion of
the Bur Should
be Inside the
Handpiece
Shaft Inside Handpiece
11
© 2011 Midmark Corp.
Syringe Tip Replacement
To Change Syring Tip...
A)
B)
C)
D)
Press down on large collar.
When a soft “click is felt pull tip straight out.
Hold collar down and insert new tip. Be sure to
press it all the way in, then release collar.
Test the installation by giving a firm pull on tip, to
assure it is inserted all the way and locked in place.
VA1807i
Handpiece Installation
Install handpieces.
Note
Not all units will be equipped with lamp
module.
Swivel Coupler
Connection
Line up components properly to
prevent damage resulting in hand
piece failure.
Lightly spray the exposed o-rings
with one step conditioner.
Tighten coupler firmly with wrench.
Handpiece
Connection
12
© 2011 Midmark Corp.
Drive Air Adjustment
& Locking Casters
To adjust drive air...
Insert flat head screwdriver into knob for Drive Air
Adjustment (for desired handpiece). Turn knob until
proper air pressure displays on the pressure gauge.
Drive Air
Adjustment Knob
OFF for Unlock
ON for Lock
13
© 2011 Midmark Corp.
Operation
Air Pressure Regulator Adjustment
Decrease
Increase
Adjustment
Knob
Unlock
Equipment alert
Do not adjust regulator over
65 PSI.
Lock
To Adjust Regulator...
A) Pull adjustment knob upward to unlock.
B) Rotate adjustment knob clockwise to
increase pressure, counterclockwise to
decrease pressure.
C) Push adjustment knob downward to lock.
Water Flow Adjustment
To adjust water flow...
Rotate Water Needle Valve (for desired handpiece) until desired flow is achieved.
Water Needle Valve
14
© 2011 Midmark Corp.
Operation
Coolant Air Adjustment
To adjust coolant air...
Insert flat head screwdriver into knob for Coolant Air Adjustment
(for desired handpiece). Counter-Clockwise to increase flow and
Clockwise to decrease flow. Turn screw until desired air pressure is
coming through the instrument.
Note: Drive air adjustment screws are located below tool holders for each instrument.
Coolant Air
Adjustment Screws
15
© 2011 Midmark Corp.
Cleaning
• Use cleaners that are appropriate for the situation, such as warm water and mild detergents,
or a 10% solution of bleach with 90% water.
• Clean the exterior surfaces with a clean, lint free cloth, dampened with disinfectant. Do not
drip any liquid into open vents, plugs or connectors. Dry the surfaces with a clean cloth or
paper towel.
• After cleaning, visually inspect the product for deterioration of covers. Do not use the delivery
system if excessive discoloration, cracking, or other signs of wear are noticeable (See Calling
for Service instructions).
Note:
Use only 5.25% Bleach Solution. Do not use bleach with
scents, color guards, or other additives. Mix fresh every day.
~90%
~10%
16
© 2011 Midmark Corp.
Cleaning
continued...
Note
Every veterinary practice setting is different and no one disinfectant is the best choice for every facility. There are several organizations educated in disinfection procedures. Refer to web sites of accredited sources that can assist veterinary personnel in choosing what is best for their practice. Listed below are a few of the sites.
Organization for Safety & Asepsis Procedures
http://www.osap.org
Dept. of Health & Human Resources Centers for Disease Control & Prevention (CDC)
http://www.cdc.gov
Read
carefully
DA148100i
When using disinfectants...
• Carefully read the product label and directions for use.
• Do not exceed the dilution rate.
17
Equipment Alert
Refer to handpiece manufacturer
for suggested cleaning procedures.
© 2011 Midmark Corp.
Maintenance
Note
Distilled water is not mandatory.
However, water must meet
requirements for drinking.
At the beginning of each work day...
• Fill water bottles with fresh distilled water.
• Flush handpieces.
To Flush Delivery System..
A)
B)
C)
D)
E)
F)
Remove all handpieces that use water.
Open water needle valve.
Hold end of tubing over a container and depress foot control.
Allow water to run 20 to 30 seconds.
Install newly disinfected handpieces and syringe tip.
Refill water bottle.
Water Needle Valve
Between each patient...
• Remove disposable tips,
instruments, etc..
• Flush delivery system.
• Disinfect handpieces and scaler
per manufacturers instructions.
• Disinfect syringe tip.
• Fill water bottles.
Equipment Alert
HANDPIECE STERILIZATION
Between each procedure complete
manufacture’s instructions for “Infection Control / Sterilization found in the documentation included
with the handpieces. Failure to do this will cause
premature failure of handpieces.
Equipment Alert
AUTOCLAVABLE SYRINGE TIP STERILIZATION
The autoclavable syringe tips supplied with the delivery system
must be sterilized prior to use with each patient, including initial use. Be sure to
thoroughly rinse and clean syringe tips prior to sterilization, any debris may reduce
the effectiveness of the sterilization. Recommended sterilization process is steam
autoclave. Recommended parameters are 125°C (250°F) and 106 kPa (15 PSI) for 40
minutes at temperature and pressure.
18
© 2011 Midmark Corp.
Maintenance
continued...
CH
EA
BL
At the end of each day...
• Bleach flush delivery system and
empty water system.
• Drain air regulator.
• Drain condensation from air tank.
• Clean system.
Equipment alert
Do not use colored/scented bleach, or concentrations greater than
recommended. Do not leave bleach solution in delivery system longer than
30 minutes, it could damage the delivery system.
To Bleach Flush Delivery System..
A) Remove all handpieces that use water.
B) Remove water bottles and fill with 9 parts water and 1 part bleach.
C) Open water needle valve.
D) Hold end of tubing over a container and depress foot control.
E) Allow water to run 20 to 30 seconds.
F) Leave bleach solution in delivery system for 10 - 20 minutes, then
remove / empty / rinse water bottles.
G) Fill water bottles with fresh water and repeat steps C thru E until bleach odor is gone.
H) Install newly disinfected handpieces and syringe tip.
I) Empty water bottles.
19
© 2011 Midmark Corp.
Maintenance continued...
To Drain Air Regulator...
Place a cloth under the regulator and depress
the button until there is no water coming out.
Button
Valve
Opened
VA1238
To Drain Condensation from Air Tank...
A) Turn power switch off.
B) Position end of red tube so it drains into a container.
C) Turn drain valve slowly to open.
D) When no water is coming out, slowly close valve.
20
© 2011 Midmark Corp.
Maintenance continued...
VA1241
Each Month...
• Clean screen underneath cart with a rag.
21
© 2011 Midmark Corp.
Maintenance continued...
To clean the air/oil separator...
A)
B)
C)
D)
E)
Turn master switch OFF.
Remove container by squeezing sides in while pulling out.
Remove fluid and saturated gauze.
Clean container and mounting cap.
Install clean gauze and reinstall the container.
VA1676i
Calling For Service
Direct all service inquiries to your authorized Midmark dealer. When
calling for service, you must provide the following information:
Model / serial number
Date of purchase
Symptom(s) of malfunction
Service and Parts Manual is available at www.midmark.com.
22
© 2011 Midmark Corp.
VetPro® 1000
Scaler Adjustment
Handpiece Pressure Gauge
Water Adjustment
Handpiece Air
Adjustment
Handpiece Holder
Water Regulator
Water Toggle Valve
Water Bottles
Quick Disconnect
Water Accessories
Quick Disconnect
Air Accessories
Air Pressure Regulator
Inside Back Door
Foot Control Storage Bracket
On/Off Power Switch
Circuit Breaker
Power Cord Receptacle
VA1247
Locking Casters (Rear Only)
23
© 2011 Midmark Corp.
Specifications
Cart
Width
23 inches (58.42 cm)
Length
18 1/2 inches (46.99 cm)
Height
36 1/2 inches (92.71 cm)
Weight
130 lbs (58.97 kg)
Shipping Weight
150 lbs (68.04 kg)
Top of Cart Weight Restriction
25 lbs (11.33 kg)
Compressor
Motor
Oil Free 1/3 HP
Working Pressure
70-100 PSI
Flow Rate
2.5 scfm @ 40 PSI
Sound Level (full load)
40 dB
Air Tank
1 Gallon
Water Module
Reservoirs
Two - 1 Liter - Toggled
Pressurized System
30-40 PSI
Aseptic
Non-Retracting HPC Coolant
Handpiece Flush System
Manual
Air and Water Connections
External Quick Connects
Other
Duty Cycle
Continuous Operation
Note: Exception, Units with LED Scaler,
10 min. ON, 5 min. OFF
Classifications
Class 1, Type B Applied Part, except as
noted for optional accessories,
Ordinary Equipment [IPXO]
Optional Accessories:
Acteon Satelec Scaler
Type BF Applied Part
High Speed Handpiece with and
without out Fiber Optics
Type B Applied Part
Low Speed Handpiece
Type B Applied Part
Air Water Syringe
Type B Applied Part
24
© 2011 Midmark Corp.
Specifications continued...
Electrical (See Regulatory Compliance Chart - Next Page)
Fuse (Scaler Transformer)
T0.8AL, 250V, 3AG
Circuit Breakers (2) 115VAC
6A 250V, Push to Reset
Circuit Breakers (2) 250VAC
3A 250V, Push to Reset
Power Cord (115VAC)
15A/125VAC x 10 ft. Hospital Grade
Power Cord (230VAC)
10A/250VAC x 2.5 meters
Handpiece
Connection Type
Midwest 4-Hole Type Handpiece
warning
Equipment not suitable for use in the presence of a flammable anesthetic mixture.
Do not modify this equipment without authorization of the manufacturer.
To avoid the risk of electric shock, this equipment must ONLY
be connected to supply mains with protective earth.
warning
warning
warning
The Acteon Newtron® LED scaler complies with current electromagnetic compatibility standards. However, it must not
be used by or on those with cardiac stimulators or other active implants
unless adequate electromagnetic interference shielding is present.
Check with a cardiologist to verify the EMI shielding capabilities of the
implant. See enclosed Newtron® LED manual for further advisories.
25
© 2011 Midmark Corp.
Model Identification / Compliance Chart
Complies To:
Model
Description
Electrical Ratings:
UL
60601-1,
Ed. 1
CAN / CSA
22.2,
#601.1-M90
IEC
60601-1,
Ed. 2
CE
•
•
•
•
•
•
VAC
+/- 10%
Amps
Cycles
(Hz)
•
•
115
6
60
115
6
60
8000-001
VetPro 1000 w/ Scaler
8000-002
VetPro 1000 w/o Scaler
8000-003
VetPro 1000 w/ Scaler and
Fiber Optics
•
•
•
•
115
6
60
8000-004
VetPro 1000 w/o Scaler
and Fiber Optics
•
•
•
•
115
6
60
8000-005
VetPro 1000 w/ Scaler,
Export
•
230
3
50
8000-006
VetPro 1000 w/ LED Scaler
and Fiber Optics. Export
•
230
3
50
8000-006S
VetPro 1000 w/ LED Scaler
and Swivel Fiber Optics.
Export
•
230
3
50
8000-007
VetPro 1000 w/ LED Scaler
and 2 Position Fiber Optics
•
•
•
•
115
6
60
8000-007S
VetPro 1000 w/ LED Scaler
and 2 Position Swivel Fiber
Optics
•
•
•
•
115
6
60
8000-008
VetPro 1000 w/ Scaler and
Fiber Optics
•
•
•
•
115
6
60
8000-008S
VetPro 1000 w/ Scaler and
Swivel Fiber Optics
•
•
•
•
115
6
60
26
© 2011 Midmark Corp.
Limited Warranty
SCOPE OF WARRANTY
Midmark Corporation (“Midmark”) warrants to the original purchaser its Exam Room Casework, Veterinary products and components (except for components not warranted under “Exclusions”) manufactured
by Midmark to be free from defects in material and workmanship under normal interior use and service.
Midmark’s obligation under this warranty is limited to the repair or replacement, at Midmark’s option, of the
parts or the products the defects of which are reported to Midmark within the applicable warranty period
and which, upon examination by Midmark, prove to be defective.
APPLICABLE WARRANTY PERIOD
The applicable warranty period, measured from the date of delivery to the original user, shall be as
follows:
• The VetPro® 1000 unit is warranted against defect in material and workmanship for a period of two years from the time of delivery.
DENTAL PRODUCTS are warranted against defect in material and workmanship for the period indicated
below from the time of delivery:
a) 2 years for all products and components (except for the periods in b and c).
b) Accessories not manufactured by Midmark are excluded (i.e. Fiber optic systems, Scalers, Handpieces,
etc.).
c) Replacement parts and accessories carry a 90 day warranty
EXCLUSIONS
This warranty does not cover and Midmark shall not be liable for the following; (1) repairs and replacements required because of misuse, abuse, negligence, alteration, accident, freight damage, or tampering;
(2) changes in color caused by natural or artificial light; (3) products which are not installed, used, and
properly cleaned as required in the Midmark written “Installation” and or “Installation/Operation” Manual
applicable for this product; (4) products considered to be of a consumable nature such as light bulbs and
surge suppression product; (5) accessories or parts not manufactured by Midmark; (Any warranties on
these items are extended directly by the manufacturer of these items to the original purchaser.
Information on these manufacturers’ warranties will be enclosed with the applicable products. Also,
Midmark will furnish copies of any of the warranties extended by any such manufacturers upon request.)
(6) specially manufactured products; (7) damage caused by animals; (8) charges by anyone (including
Midmark’s authorized dealers) for adjustments, repairs, replacement parts, installation, or other work performed upon or in connection with such products which is not expressly authorized in writing in advance
by Midmark.
EXCLUSIVE REMEDY
Midmark’s only obligation under this warranty is the repair or replacement of defective parts. Midmark
shall not be liable for any direct, special, indirect, incidental, exemplary or consequential damages or delay
including, but not limited to, damages for loss of profits or loss of use.
NO AUTHORIZATION
No person or firm is authorized to create for Midmark any other obligation or liability in connection with the
products.
THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED. MIDMARK MAKES NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. THIS
WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS.
27
© 2011 Midmark Corp.
Language of origin: English
Midmark Corporation
60 Vista Drive
Versailles, OH 45380 USA
1-800-643-6275
1-937-526-3662
www.midmark.com
© 2011 Midmark Corp.