Cuspidor
User Guide
Procenter Cuspidor
Control Buttons and Knobs
Intended Use......................................... 2
Electromagnetic Interference................ 2
Transportation / Storage /
Operating Conditions ............................ 2
Disposal of Equipment ......................... 2
Authorized EU Representative.............. 2
Calling for Service................................. 2
Ordering Parts....................................... 2
Symbols................................................. 3
Bowl Rinse Operation............................ 4
Cup Fill Operation................................. 4
Barriers.................................................. 5
Cleaning and Disinfecting...................... 5
Handpiece Accessories......................... 5
General Purpose Cleaning.................... 5
Visual Inspection................................... 5
Cleaning and Disinfecting Assistance... 5
Specifications........................................ 6
Warranty Information............................. 7
Cup Fill Button
Technical Library at Midmark.com
Language of origin: English
Cup Fill
Adjustment
Knob
KA938905i
Bowl Flush
Button
Bowl Flush
Adjustment
Knob
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
1
Part No. 003 -1110-00 REV S (9/9/15)
Style A
Compliance Information and Symbols
Authorized EU Representative
Intended Use
Countries in the EU should direct all questions, incidents, and complaints
to Midmark’s Authorized EU representative listed below.
To provide water for oral rinsing, and a drain to dispose of waste water and
oral debris during dental examinations.
Midmark EMEA Ltd
Beech House
First Floor, East Wing
Ancells Business Park
Fleet
Hampshire GU51 2UN
United Kingdom
Tel: + 44 (0) 1252 360 940
Fax: + 44 (0) 1252 360 941
Electromagnetic Interference
Midmark dental operatory components are designed and built to minimize electromagnetic interference with other devices. However, if interference is noticed
between another device and this operatory, remove the interfering device from
room and / or plug product into an isolated circuit.
Transportation / Storage / Operating Conditions
Calling For Service
Transportation / Storage Temperature:............23°F to 100°F (-5°C to 38°C)
Direct all service inquiries to your authorized Midmark dealer. When
calling for service, you must provide the following information:
Relative Humidity:........................................ 10% to 90% (non-condensing)
Atmospheric Pressure:.................. 7.2 PSI to 15.3 PSI (50 kPa to 106 kPa)
Operating Temperature Range: .......................59°F to 95°F (15°C to 35°C)
Disposal of Equipment
Ordering Parts
At the end of this product’s life, the unit, accessories and other consumable
goods may be contaminated from normal use. Consult local codes and ordinances for proper disposal of this equipment and other consumable goods.
The following information is required when ordering parts:
Model / serial number
Date of purchase
Symptom(s) of malfunction
Serial number & model number
Part number for desired part
(Refer to Exploded Views and Parts at midmark.com)
Non-warranty parts orders may be faxed to Midmark using the Fax Order
Form in the back of this manual.
For warranty parts orders, call Midmark’s Technical Service Department
with the required information.
Hours: 8:00 am to 5:00 p.m. EST (Monday thru Friday)
Phone: 1-800-Midmark
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
2
Style A
Symbols
These symbols may appear on your equipment and/or in the manuals.
Warnings and cautions are provided in the manuals where applicable.
Pressure
Limit
Consult Manuals
Type B,
Applied Part
Protective
Earth Ground
warning
Indicates a potentially hazardous situation which
could result in serious injury if not avoided.
100 F
38 C
23 F
-5 C
Caution
Temperature
Limit
Indicates a potentially hazardous situation which may result in minor or moderate injury if not avoided. It may
also be used to alert against unsafe practices
Humidity
Limit
Proper Shipping
Orientation
Maximum stacking
height (Refer to “n”
number on package.)
Fragile
Keep Dry
Equipment Alert
Indicates a potentially hazardous situation which
could result in equipment damage if not avoided.
NOTE
Amplifies a procedure, practice, or condition.
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
3
Style A
Operation and Adjustments
Bowl Rinse
Bowl Rinse
Flow Time
Adjustment
To rinse the cuspidor bowl...
Press Bowl Rinse Button.
Cup Fill
Flow Time
Adjustment
Cup Fill
To fill a cup with water...
Press Cup Fill Button.
Note: Water will run for approximately
6 seconds, then shut off automatically.
Note: Water will run for approximately
12 seconds,then shut off automatically.
To increase the flow time...
To increase the flow time...
Screw Cup Fill adjustment screw
counter-clockwise as necessary.
Screw Bowl Rinse adjustment
counter-clockwise as necessary.
To decrease the flow time...
To decrease the flow time...
Screw Cup Fill adjustment
screw clockwise as necessary.
Screw Bowl Rinse adjustment
clockwise as necessary.
KA972001i
Cup Fill
Button
Cup Fill
Flow Knob
Bowl Rinse
Flow Knob
To increase the flow volume...
Rotate Bowl Rinse Flow Knob
counter-clockwise as necessary.
If cup is not full enough ...
KA938904i
Rotate Cup Fill Flow Knob clockwise
as necessary.
Bowl Rinse
Button
If cup overflows...
To decrease the flow...
Rotate Cup Fill Flow Knob counterclockwise as necessary.
Rotate Bowl Rinse Flow Knob
clockwise as necessary.
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
4
Style A
KA938903i
Cleaning, Disinfecting and Maintenance
General Purpose Cleaning
Barriers
For general purpose cleaning, use cleaners that are appropriate for the
situation, such as warm water and mild detergents, or a 10% solution of
bleach with water.
Midmark recommends the use of disposable barriers on all clinician controls
that may be in contact with clinician hands and fingers during dental procedures. The use of barriers significantly reduces the need for chemical cleaners, thus prolonging the life of the equipment.
Only use barrier material that is intended for use with dental equipment.
Midmark recommends the use of an FDA market-cleared barrier such as Pinnacle Cover-all™. Follow barrier manufacturer instructions for proper use of
these products.
Cleaning and Disinfecting
In addition to the use of barriers Midmark recommends the use of an EPA registered and FDA market-cleared cleaner/disinfectant such as Cavicide™ to be
used on all clinician controls or surfaces that may come in contact with dental
instruments during dental procedures.
Follow cleaner/disinfectant manufacturer
instructions for proper use of the product.
Care should be taken to avoid excessive
application and pooling of liquids.
Read all labels
carefully!
Visual Inspection
After cleaning, visually inspect the product for deterioration of covers and
touch pads. Do not use the delivery system if excessive discoloration,
cracking, or other signs of wear are noticeable (see Calling for Service
instructions).
Cleaning and Disinfecting Assistance
For assistance with cleaning and disinfecting instructions contact the
Midmark Technical Service Department at 1-800-Midmark, it is helpful to
provide the delivery system model number and serial number when asking
for assistance.
Additional information is available from the organizations listed below:
• Organization for Safety & Asepsis Procedures:
http://www.osap.org
Handpiece Accessories
Only use dental handpiece accessories with the delivery system that are FDA
market-cleared and refer to manufacturer’s instructions for proper cleaning
and disinfecting. Either an autoclavable syringe tip or a single use disposable
syringe tip may be used.
• American Dental Association:
http://www.ada.org
• Dept. of Health & Human Resources
Centers for Disease Control & Prevention (CDC):
http://www.cdc.gov
Equipment Alert
AUTOCLAVABLE SYRINGE TIP STERILIZATION
The autoclavable syringe tips supplied with the delivery system
must be sterilized prior to use with each patient, including initial use. Be sure to
thoroughly rinse and clean syringe tips prior to sterilization, any debris may reduce
the effectiveness of the sterilization. Recommended sterilization process is steam
autoclave. Recommended parameters are 125°C (250°F) and 106 kPa (15 PSI) for 40
minutes at temperature and pressure.
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
• European Dental Association:
http://www.eda-eu.org
Style A
Specifications
Electrical Rating for
Elevance Chair:
Power Supply Model No.:
153808-001, -002
Complies to the applicable
requirements of:
ES/IEC/EN 60601-1, CAN/CSA C22.2 No. 60601-1
(Safety Standards), EN/IEC 60601-1-2 (EMC
Standards)
Electrical Inputs for Knight
and Midmark Ultra Chairs:
24 VAC, 3.2 amp
13 VAC, 2 amp
Complies to the applicable
requirements of:
UL/IEC/EN 60601-1, CAN/CSA C22.2 No. 601.1,
IEC 60601-1-1
Air Supply
Requirements:
Maximum 80-100 PSI (5.52 - 6.9 bar)
Water Supply
Requirements:
Maximum 30 - 50 PSI (2.1 -3.4 bar)
Classifications:
Class I, Type B Applied Part, Ordinary Equipment
(IPX0), Continuous Operation
Certifications:
Midmark Corporation..........ISO-9001
warning
Do not modify this equipment without authorization of the manufacturer.
Caution
Federal law restricts this device to sale by or on
the order of a licensed dental practitioner.
warning
Equipment is not suitable for use in the presence of a
flammable anesthetic mixture with oxygen, air, or nitrous oxide.
Clarification: Equipment is suitable for use in the presence of oxygen,
air, or nitrous oxide.
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
6
Style A
Warranty Information
SCOPE OF WARRANTY
Midmark Corporation (“Midmark”) warrants to the original purchaser its Dental
products and components (except for componentsnot warranted under “Exclusions”) manufactured by Midmark to be free from defects in material and
workmanship under normaluse and service. Midmark’s obligation under this
warranty is limited to the repair or replacement, at Midmark’s option, of the
parts or the products the defects of which are reported to Midmark within the
applicable warranty period and which, upon examination by Midmark prove to
be defective.
APPLICABLE WARRANTY PERIOD
The applicable warranty period, measured from the date of delivery to the
original user, shall be as follows:
1. DENTAL PRODUCTS are warranted against defect in material and workmanship for the period indicated below from the time of delivery:
(a) Two (2) years for all products and components (except for the
periods in b thru e).
(b) “KINK-VALVE” module carries a ten (10) year warranty.
(c) The original light bulb on a new light carries a one (1) year warranty.
(d) Accessories not manufactured by Midmark are excluded.
(e) Replacement parts and accessories carry a ninety (90) day warranty.
2. Oral Surgery PRODUCTS are warranted against defect in material and
workmanship for a period of one (1) year from the time of delivery.
3. STERILIZERS PRODUCTS are warranted against defect in material and
workmanship for a period of one (1) year from the time of delivery.
4. ULTRASONIC CLEANER PRODUCTS when used in accordance with
manufacturer’s instructions and under normal use, are guaranteed for
two (2) years after date of shipment. Within the period guaranteed, the
manufacturer will repair or replace free of charge, at its sole discretion,
all parts that are defective because of material or workmanship including costs for removing or installing parts. Repairs are made on a FOB
repair depot basis. Manufacturer’s liability, whether based on warranty,
negligence or other cause, arising out of and/or incidental to sale, use or
operation of the transducer elements, or any part thereof, shall not in any
case exceed the cost of repair or replacement of the defective equipment, and such repair or replacement shall be the exclusive remedy of
the purchaser, and in no case shall manufacturer be responsible for any
and/or all consequential or incidental damages including without limitation, and/or all consequential damages arising out of commercial losses.
© 2010 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
5. INTEGRA™ DENTAL CASEWORK AND ARTIZAN™ DENTAL FURNITURE PRODUCTS are warranted against defect in material and workmanship for the period indicated below from the time of delivery:
(a) Five (5) years for all products and components (except for the shorter
periods in (b and c), including plastic laminate top surfaces, door and
drawer fronts, casters, and electrical components such as task lights
and cords. (b) Three (3) years for task light ballast. (c) One year (1) for
plastic laminate, solid surface countertops and accessories.
EXCLUSIONS
This warranty does not cover and Midmark shall not be liable for the following;
(1) repairs and replacements because of misuse, abuse, negligence, alteration, accident, freight damage, or tampering; (2) matching of color, grain, or
texture except to commercially acceptable standards; (3) changes in color
caused by natural or artificial light; (4) products which are not installed, used,
and properly cleaned as required in the Midmark “Installation” and or “Installation/Operation” Manual for this applicable product; (5) products considered
to be of a consumable nature; (6) accessories or parts not manufactured by
Midmark; (7) specially manufactured products; (8) charges by anyone for
adjustments, repairs, replacement parts installation, or other work performed
upon or in connection with such products which is not expressly authorized in
writing in advance by Midmark.
EXCLUSIVE REMEDY
Midmark’s only obligation under this warranty is the repair or replacement of
defective parts. Midmark shall not be liable for any direct, special indirect, incidental, exemplary or consequential damages or delay including, but not limited
to, damages for loss of profits or loss of use.
NO AUTHORIZATION
No person or firm is authorized to create for Midmark any other obligation or
liability in connection with the products.
THIS WARRANTY IS MIDMARK’S ONLY WARRANTY AND IS IN LIEU OF
ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED. MIDMARK MAKES
NO IMPLIED WARRANTIES OF ANY KIND INCLUDING ANY WARRANTIES
OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.
THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT OF
DEFECTIVE PARTS.
7