RKI Instruments 30-0951RK-IR1S8 Owner Manual

Add to my manuals
3 Pages

advertisement

RKI Instruments 30-0951RK-IR1S8 Owner Manual | Manualzz

Part Number: 71-0374

Revision: P1

Released: 6/11/15

30-0951RK-IR1S8

Flow Through Adapter www.rkiinstruments.com

Description

The attached detector head operator’s manual can be used in conjunction with this supplement for the 30-0951RK-IR1S8 flow through adapter. The flow through adapter includes a sensor adapter and a sensor chamber with two compression fittings. A sensor adapter O-ring seals to the detector when installed. A detector chamber O-ring seals the sensor chamber to the sensor adapter. Calibration instructions are the same as those outlined in the attached detector head manual with the exception that the calibration gas must be connected either to the inlet fitting tubing or pump inlet as opposed to a calibration cup.

Installation instructions are slightly different and are outlined below.

Installation

Sensor Adapter

O-Ring

Sensor Chamber

O-Ring

Sensor Adapter

Sensor Chamber

Thumbscrew,

3x

8 mm OD Tube

Sample Exhaust

Fitting

Screw Onto Sensor 8 mm OD Tube Sample Inlet Fitting

2.56

3.20 Max

3.27

Figure 1: Outline and Mounting Dimensions/Component Location

1.

The flow through adapter is normally shipped installed to the detector. If it is not installed, screw the flow through adapter onto the detector tightening firmly by hand.

CAUTION: Make sure the flow through adapter is installed tightly and after installation is complete, verify that the flow through adapter is still tightened firmly onto the detector to avoid possible leaks.

2.

If necessary, re-orient the sensor chamber so the inlet fitting faces away from the mounting surface by unscrewing the three thumbscrews, rotating the sensor chamber, and tightening the three thumbscrews.

3.

Install and startup the detector head as described in the detector head operator’s manual. Make sure there is sufficient room between the detector and the mounting surface for the flow through adapter.

4.

The inlet and exhaust fittings accept 8 mm OD rigid metal tubing such as copper, aluminum, or stainless steel. The inlet fitting is located on the side of the sensor chamber and the exhaust fitting is located on the bottom. Insert the desired length of tubing into each fitting and tighten the fittings so that the ferrules crimp onto the

1

tubing.

5.

Install appropriate device(s) (pump, aspirator, etc.) to draw sample to the detector.

NOTE: Be sure to use the inlet and exhaust ports as intended to ensure correct sensor response and to allow any accumulated liquid to drain out the bottom of the flow through adapter.

6.

Calibrate the detector head as described in the attached manual and in the next section.

NOTE: Calibrate the detector head with the flow through adapter installed to ensure an accurate reading.

Calibration

A demand flow regulator should be used for calibration.

1.

Follow the calibration instructions outlined in the attached detector head manual.

2.

When the instructions call for applying gas, connect the cylinder.

• If you are drawing sample through the system, connect the cylinder to the inlet fitting of the flow through adapter.

• If you are pushing sample through the system, connect the cylinder to the pump inlet.

3.

Perform a calibration as outlined in the attached detector head manual.

4.

When the instructions call for removing the gas, disconnect the cylinder.

Parts List

See the table below for spare parts not listed in the attached detector manual.

Table 1: Parts List

Part Number

07-7218RK

07-7225RK

13-1070RK

17-2665

17-2666

17-2667

17-2668

81-1054RK

81-1055RK

Description

O-ring, 0.734 x .139, buna, for IR adapter

O-ring, 1.243 ID x .139, buna, for sensor chamber

Captive panel screw, 10-32 x 1.75

Front ferrule (for inlet and exhaust fittings)

Nut for ferrules

Back ferrule (for inlet and exhaust fittings)

Front and back ferrule set (for inlet and exhaust fittings)

Regulator, demand flow, for 58- and 103-liter calibration cylinders

Regulator, demand flow, for 17- and 34-liter calibration cylinders

2