LR Mtd. 153208 / 153209

LR Mtd. 153208 / 153209
Print on plain white paper - Part # 122919
Installation Instructions
Telescopic Arm for
Concept LR
PROCDURE:
1. Remove the existing set screw from the cuspidor
arm.
2. Disconnect the cuspidor umbilical from the "J"
box.
NOTE: Be sure to note the location of
the umbilical tubing connections
inside the "J" box. After the
installation, the umbilical tubing will
have to be reconnected in the "J" box.
3. Remove the cuspidor and umbilical from the
cuspidor arm by lifting the cuspidor up and
pulling the umbilical out of the arm
(See Figure 1).
4. Position the support ring on the cuspidor arm
and install the new adjusting set screw. Make
sure the set screw goes in the hole where the
old set screw was located (See Figure 2).
5. Level the support ring and tighten the two
locking set screws.
6. Assemble the remaining parts and squeeze
them together.
7. Tighten the three set screws in the cover ring.
8. Test the movement of the assembly and loosen
or tighten the new adjusting set screw if
necessary.
9. Replace the cuspidor and umbilical and
reconnect the umbilical in the "J" box.
CO VE R
R IN G
S U P P O RT
B E A R IN G
S U P P O RT
ARM
TE LE S C O P IC
ARM
C U S P ID O R
L OC K IN G
SET
SCREW S
U M B ILIC A L
E XIS TIN G
S ET SC R E W
S U P P O RT
R IN G
N E W A D JU S T IN G
SET SCR EW
C U S P ID O R A R M
C U S P ID O R A R M
A F111 8
A F 1117
Figure 1
Midmark Corporation P O Box 286 60 Vista Drive
Versailles Ohio 45380-0286 www.midmark.com
Phone: 937-526-3662
Fax: 937-526-5542
Style A
Figure 2
TPW1723 Rev. B