Assembly instructions induSENSOR VIP series

Assembly instructions induSENSOR VIP series
>> Damage to or destruction of the sensor
Avoid shocking and knocking the sensor
>> Damage to or destruction of the sensor
>> Avoid bending the sensor rod.
>> Damage to or destruction of the sensor
Sensor
housing
Sensor rod
Notes on CE Identification
The following applies for the induSENSOR, VIP series:
EC regulation 2004/108/EC
The induSENSOR, VIP series, satisfies the requirements of the standards EN
61326-1: 2006-10 and DIN 61326-2-3: 2007-05.
Proper Environment
-- Protection class for sensor:IP 67 (only with plug connected)
-- Operating temperature: -40 up to +85 °C (RL ≤ 500 Ohm) -40 to +185 °F
-- Storage temperature: -40 up to +100 °C (-40 up to +212 °F)
-- Humidity: 5 - 95 % (non condensing)
-- Ambient pressure: Atmospheric pressure
-- EMC acc. to: EN 61326-1: 2006-10 and DIN 61326-2-3: 2007-05
Mounting of the Housing GA
A
Wrench
Plug
7-contact size 32
48
(1.89)
13.8
(.54)
C
12
40
(1.57) (.47)
Start of measuring
range, I OUT = 4 mA
dia. .47 x
.04)
ø12x1
ø30.5
End of measuring
range, I OUT = 20 mA
Dimensional drawing housing ZA, dimensions in mm (inches), not to scale
Bending radius sensor cable:
> 30 mm (once)
> 90 mm (permanent)
D
B
End of measuring
range, I OUT = 20 mA
Dimensional drawing housing GA, dimensions in mm (inches), not to scale
The sensor is screwed to the mounting plate using the M18 thread, see
figure on the right side.
12
(.47)
Start of measuring
range, I OUT = 4 mA
B
Wrench size 32
Mounting plate
Measuring range A
50 (1.97)
105 (4.13)
B
24 (.94)
C
11.5 (.45)
D
19.5 (.77)
100 (3.94)
150 (5.91)
27 (1.06)
30 (1.18)
22 (.87)
33 (1.30)
26 (1.02)
29 (1.14)
175 (6.89)
242 (9.53)
2x M4,
length of thread
(min.) 12 mm (.47 “)
Sensor mounting with mounting kit MBS 12/8
(dia. 1.20)
53 (2.09)
40
(1.57)
Measuring range
B
C
ø8
(dia..31)
An electrically conductive mounting material (e.g. measuring ring
mounting facility) influences the measuring result. The specified
technical data are no longer observed!
2x M4,
length of thread
(min.) 16 mm (.63 “)
Measuring range
Plug
B
7-contact
ø12
Mounting Kit MBS 12/8
A
(dia. .47)
i
Mounting of the Housing ZA
The sensor is fastened at both ends of the sensor rod (area B and D, see
the lower figure) by means of a clamping / mounting facility.
ø8
(dia. .31)
Power supply must be connected in accordance with the safety regulations
for electrical equipment.
>> Danger of injury
(dia. .47
x 0.4)
>> Damage to or destruction of the sensor
ø12x1
Warnings
The power supply may not exceed the specified limits.
>> Danger of injury
M18x1.5
Assembly Instructions
induSENSOR, VIP series
Installation and Mounting
Centering and Mounting the Measuring Ring
Install the measuring ring in an electrically non-conductive mounting facility
( e.g. plastics). The dimensions of the measuring ring are shown in the
following figures. MICRO-EPSILON recommends to use the mounting kit
MBS 12/8, which is available as an accessory, see Chap. 8.1 of the instruction manual and Figure “Sensor mounting with mounting kit MBS 12/8”.
Please observe the position of the measuring ring at the zero point (= 4 mA
output), see also the following figures.
A slightly eccentrical mounting of the measuring ring has no negative influence on the sensor signal. The measuring ring is attached on the target to
be measured by means of circumferential clamping or gluing.
Precautionary Measure
The measuring ring must not contact the sensor rod during operation.
>> Damage to or destruction of the sensor through abrasion.
Avoid bending or shortening the measuring ring.
>> Loss of specified technical data
The minimum bending radius is 30 mm (1.2 “) once, repeated: 90 mm
(3.5 “) for the sensor cable C 703x.
>> Damage to or destruction of the sensor cable
i
The specified technical data are valid only, if the measuring ring is
used supplied from MICRO-EPSILON!
i
Output signal (mA)
Measuring Principle
20
1/2
0
1
Measuring range
Measuring ring
Sensor rod
Ue
Ia
Power Supply and Display / Output Device
Power supply and signal output are effected through the 7-contact connector at the sensor’s electronic housing. The pin assignment is shown in
drawing and table stated below.
54
3
2
6
7
1
View: Solder
pin side,
female cable
connector
Pin Assignment
Color C703
1
Supply + (18 ... 30 VDC)
white
2
0 V Ground
I OUT 4 ... 20 mA
(U OUT 1 ... 5 V) 1
brown
4
Signal ground
yellow
5
SCL (sensor calibration)
grey
6
SDA (sensor calibration)
pink
7
not connected (n.c.)
blue
3
green
Pin and color assignment of the connector and sensor cable C703-5 respectively C703-5/U
1) With sensor cable C703-5/U
exceed 5 mVss, if the sensors are supplied through switched-mode
power supply units,
1
A small load resistor loads the sensor electronics more thermical. With a
maximum operating temperature of 85 °C (+185 °F) the minimum load
resistor RL permitted is calculated as:
82.5 Ohm * U B
- 1625 Ohm (If the result is negative: R L = 0 )
R L min =
V
With a preset load resistor the maximum operating temperature permitted
is calculated as:
T max = 150 °C -
3.3 °C * U
V
B
+
0.04 °C * R L
;
Ohm
R L = Load resistor
U B = Operating voltage
T max = Maximum operating temperature
Note: T max ≤ 85 °C
18 ... 30 VDC
VIP
Pin 2 is connected with pin 4 on the sensor electronics. The screen of the
C703 sensor cable is connected with the connector housing.
Connect the screen of the C703 sensor cable with the protective earth
conductor on power side.
The sensors are connected according to the pin assignment, see opposite
table and Figures “Signal monitoring“. Notice the different criterions:
The maximum load resistor RL is limited by the operating voltage UB.
(U B - 10 V)
R L max =
20 mA
12
4
Make sure, that the output noise of the power supply units does not
+
UB
Read the detailed instruction manual before using the sensor. The manual
is available online on:
www.micro-epsilon.com/download/manuals/man--induSENSOR-Serie-VIP-de-en.pdf
_
2
3
RL
+
I OUT
A
_
4
Signal monitoring with amperemeter
i
R L can be inserted as an option for adaptation of the power loss to
high ambient temperatures.
If the signal is monitored with a voltmeter, the load resistor RL is dimensioned in accordance with the desired output voltage U OUT.
Formula: U OUT = R L * I Signal
1
18 ... 30 VDC
VIP
+
UB
_
2
3
+
RL
U OUT
V
_
4
Signal monitoring with load resistor and voltmeter
R L = Load resistor
U B = Operating voltage
T max = Maximum operating temperature
MICRO-EPSILON MESSTECHNIK GmbH
& Co. KG
Königbacher Str. 15 · 94496 Ortenburg
www.micro-epsilon.com
X9771052-A021014HDR
*X9771052-A02*