Omron PMLLPAD Long Life Pads™ Instruction manual


Add to my manuals
9 Pages

advertisement

Omron PMLLPAD Long Life Pads™ Instruction manual | Manualzz
INSTRUCTION MANUAL
One Pair of Long Life Pads
PMLLPAD
(for use with Omron electroTHERAPY pain relief units)
Self-adhesive electrode pad with snap connection (98mm x 64mm)
www.OmronPainRelief.com
ENGLISH
Model:
TABLE OF CONTENTS
Save These Instructions..............................................................................................3
Safety Information........................................................................................................4
When To Replace The Pads.......................................................................................7
How To Clean The Pads..............................................................................................8
Storage.........................................................................................................................9
2
SAVE THESE INSTRUCTIONS
Please read this instruction manual thoroughly before using the pads.
3
SAFETY INFORMATION
Intended Use
These pads should be used only for applying electrical stimulation with the Omron
Electrotherapy Pain Relief Unit (PM3030) and applied on normal, healthy, dry, and clean skin
of adult patients only.
• 1 session - 15 min. with automatic shut off
• Max min. per session - 30 min. or 2 sessions
• Max times per day - 3 times/day
Do not use these pads for stimulation if you have a cardiac pacemaker, implanted
defibrillator, or other implanted metallic or electronic device. Such use could cause
electric shock, burns, electrical interference, or death.
Do not place the pads:
Over your neck because this could cause severe muscle spasms resulting in closure
of your airway, difficulty in breathing, or adverse effects on heart rhythm or blood
pressure.
Across your chest because the introduction of electrical current into the chest may
cause rhythm disturbances to your heart, which could be lethal.
Over open wounds or rashes, or over swollen, red, infected, or inflamed areas or skin
eruption, for example: phlebitis, thrombophlebitis, varicose veins.
Over, or in proximity to, cancerous lesion.
Over areas of skin that lack normal sensation.
On the opposite sides of your head since the effects of stimulation of the brain are
unknown.
4
SAFETY INFORMATION
or stimulation if you have a cardiac pacemaker, implanted defibrillator, or other
F
implanted metallic or electronic device. Such use could cause electric shock, burns,
electrical interference, or death.
he size, shape, and type of pads may affect the safety and effectiveness of electrical
T
stimulation.
The electrical performance characteristics of pads may affect the safety and
effectiveness of electrical stimulation.
Using pads that are too small or incorrectly applied could result in discomfort or skin
burns.
Precautions
• The long-term effects of pads for electrical stimulation are unknown.
• Since the effects of stimulation of the brain are unknown, the pads should not be placed
on opposite sides of your head.
• Use caution if pads are applied over areas of skin that lack normal sensation.
Adverse reactions
• You may experience skin irritation and burns beneath the pads applied to your skin if
used longer than indicated; 30 min./session, up to 3 times/day.
• You may experience headache and other painful sensations during or following the
application of electrical stimulation near your eyes and to your head and face.
5
HOW TO USE THE PADS
1. Remove and discard plastic film from pads.
2. Connect either electrode to any pad.
3. Place pads only on normal, intact, dry, clean, and healthy skin.
Note: These pads are intended for one-person use. Do not share pads with other persons because of the risks of adverse skin reactions and to prevent the spread of
disease.
6
WHEN TO REPLACE THE PADS
If the pad no longer sticks to your skin or if more than 25% of the pad’s surface is not in
contact with your skin.
7
HOW TO CLEAN THE PADS
1. Turn the power off on the unit and remove the cord from each pad.
2. Wash pads when adhesive becomes dirty and/or pads are not adhering to your skin.
3. Use your fingertips to gently wash under cold water.
• DO NOT use a sponge, cloth, detergents, chemicals or soap.
• Pads can be washed up to 10 times and used up to 150 uses.
• Air dry pads and DO NOT wipe with paper or cloth.
8
STORAGE
• Do not bend pads.
• In between uses, keep cord connected to the pads and place pads on the pad holder
with the adhesive side down onto each side of the holder.
• Wrap the cord around the pad holder.
• Store in a cool, dry place, 5-40°C (41-104°F.), 30-40% relative humidity.
• With proper storage and use, you may extend the life of your pads.
KEEP OUT OF REACH OF CHILDREN
If you have any questions or need assistance,
please call 1- 800-634-4350.
Manufactured for OMRON HEALTHCARE Co., Ltd.
Made in Japan
Distributed by: OMRON HEALTHCARE, INC.
1925 West Field Court
Lake Forest, IL 60045 U.S.A.
© 2012 OMRON HEALTHCARE, INC.

advertisement

Was this manual useful for you? Yes No
Thank you for your participation!

* Your assessment is very important for improving the workof artificial intelligence, which forms the content of this project

Related manuals

advertisement