Agilent 16550A Help

Agilent 16550A Help
Help Volume
© 1992-2002 Agilent Technologies. All rights reserved.
Instrument: Agilent
Technologies 16550A Logic
Analyzer
Agilent Technologies 16550A 100 MHz State/500 MHz
Timing Logic Analyzer
The Agilent Technologies 16550A provides timing analysis at 500 MHz
in half-channel mode. In state mode, the analyzer can be clocked by an
external sample clock at a rate of 100 MHz. Memory depth is 4 KB per
channel in full-channel modes or 8 KB per channel in half channel
mode.
Getting Started
•
“Analyzer Probing Overview” on page 107
•
“Setting Up a Measurement” on page 10
•
“When Something Goes Wrong” on page 33
•
“Error Messages” on page 33
Measurement Examples
•
“Making a Basic Timing Measurement” on page 22
•
“Making a Basic State Measurement” on page 18
•
Advanced Measurement Examples (see the Measurement Examples help
volume)
•
“Interpreting the Data” on page 26
More Features
“Coordinating Measurements” on page 31
Using Inverse Assembly (see the Listing Display Tool help volume)
“Using Symbols” on page 110
Using Markers (see the Markers help volume)
“Loading and Saving Logic Analyzer Configurations” on page 32
2
Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
“Testing the Logic Analyzer Hardware” on page 39
Interface Reference
“The Sampling Tab” on page 40
“The Format Tab” on page 55
“The Trigger Tab” on page 75
“The Symbols Tab” on page 119
“Specifications and Characteristics” on page 103
Main System Help (see the Agilent Technologies 16700A/B-Series Logic
Analysis System help volume)
Glossary of Terms (see page 131)
3
Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
4
Contents
Agilent Technologies 16550A 100 MHz State/500 MHz Timing
Logic Analyzer
1 Agilent Technologies 16550A 100 MHz State/500 MHz Timing
Logic Analyzer
Setting Up a Measurement
10
Connect the Analyzer to the Target System
Define the Type of Measurement 11
Set Up the Bus Labels 13
Define Trigger Conditions 14
Run the Measurement 15
Examine the Data 16
Making a Basic State Measurement
Making a Basic Timing Measurement
Interpreting the Data
10
18
22
26
Analysis Using Waveform 26
Analysis Using Listing 28
Coordinating Measurements
31
Loading and Saving Logic Analyzer Configurations
When Something Goes Wrong
33
Interference with Target System
Error Messages 33
Nothing Happens 38
Suspicious Data 38
33
Testing the Logic Analyzer Hardware
32
39
5
Contents
The Sampling Tab
40
Acquisition Depth 40
Setting the Acquisition Mode 41
Performing Clock Setup (State only)
Naming the Analyzer 44
Turning the Analyzer Off 45
Sample Period (Timing Only) 45
Trigger Position Control 53
The Format Tab
41
55
Importing Netlist and ASCII Files
56
Exporting ASCII Files 58
Importing ASCII Files 58
Termination Adapter 60
E5346A High Density Adapter 61
Mapping Connector Names 62
Import the Net List File 62
Verify Net to Label Mapping 63
Select/Create Interface Labels 64
Activity Indicators 64
Assigning Pods to the Analyzers 65
Data On Clocks Display 66
To reorder bits in a label 69
Labels: Mapping Analyzer Channels to Your Target
Setting Up the Pod Clock 70
Pod Selection 71
Setting the Pod Threshold 72
State Clock Setup/Hold (State only) 73
6
70
Contents
The Trigger Tab
75
Understanding Logic Analyzer Triggering 76
Setting Up a Trigger 78
Inserting and Deleting Sequence Steps 79
Editing Sequence Steps 80
Setting Up Loops and Jumps in the Trigger Sequence 81
Saving and Recalling Trigger Sequences 82
Clearing Part or All of the Trigger 83
Overview of the Trigger Sequence 83
Trigger Functions 84
Working with the User-level Function 91
Defining Resource Terms 94
Tagging Data with Time or State Tags (State Only) 101
Arming Control 101
Specifications and Characteristics
103
What is a Specification 103
What is a Characteristic 103
What is a Calibration Procedure 104
What is a Function Test 104
Agilent Technologies 16550A Logic Analyzer Specifications 104
Agilent Technologies 16550A Logic Analyzer Characteristics 105
Analyzer Probing Overview
Using Symbols
107
110
To load object file symbols 111
To adjust symbol values for relocated code
To create user-defined symbols 113
To enter symbolic label values 114
To create an ASCII symbol file 115
To create a readers.ini file 116
112
7
Contents
The Symbols Tab
119
Symbols Selector Dialog 121
Symbol File Formats 123
General-Purpose ASCII (GPA) Symbol File Format 124
Glossary
Index
8
1
Agilent Technologies 16550A 100 MHz
State/500 MHz Timing Logic Analyzer
9
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Setting Up a Measurement
Setting Up a Measurement
After you have connected the logic analyzer probes to your target
system, (see page 10) there are five basic steps for any measurement.
1. “Define the Type of Measurement” on page 11
2. “Set Up the Bus Labels” on page 13
3. “Define Trigger Conditions” on page 14
4. “Run the Measurement” on page 15
5. “Examine the Data” on page 16
Refine measurement by repeating steps 3 - 5.
If you load a configuration file, it will set up the logic analyzer and
trigger. For your particular measurement, you may need to change
some settings.
See Also
“Making a Basic Timing Measurement” on page 22
“Making a Basic State Measurement” on page 18
Measurement Examples (see the Measurement Examples help volume)
Making Basic Measurements for a self-paced tutorial
Connect the Analyzer to the Target System
Before you begin setting up a measurement, you need to physically
connect the logic analyzer to your target system. Attach the pods in a
way that keeps logically related channels together and be sure to
10
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Setting Up a Measurement
ground each pod. Analysis probes, available for most common
microprocessors, can simplify the connection process.
The logic analyzer pods carry the signals to the logic analyzer from
your target system. Connect the pods either directly to the target
system or to an analysis probe. You can attach the pods either directly
to a 40-pin header, to a 20-pin header with an adapter, or use the
General Purpose Probes to attach to individual channels.
If you are using an analysis probe, Setup Assistant will guide you
through the process based on your logic analyzer and the analysis
probe.
Step 1: Describe the Measurement (see page 11)
See Also
“Analyzer Probing Overview” on page 107 for more detail on types of
probes
Setup Assistant (see the Setup Assistant help volume)
Logic Analysis System and Measurement Modules Installation Guide
for probe pinout and circuit diagrams.
Define the Type of Measurement
There are two types of measurements: state measurements and
timing measurements. Use the Sampling tab to select either type
and to specify the details particular to that type.
11
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Setting Up a Measurement
Choose State or Timing
In a state measurement, the analyzer uses an external clock to
determine when to sample. Each time the analyzer receives a state
clock pulse, it samples and stores the logic state of the target system.
In a timing measurement, the analyzer is analogous to an oscilloscope.
It samples at regular time intervals and displays the information in a
waveform similar to the oscilloscope.
Set Measurement Mode
Each measurement type has different measurement modes. In general,
there is a trade-off between number of signals and speed.
Because the measurement type and mode affect clocking and trigger
options, you must set the measurement type first.
Set up State Clock
For state measurements, you must specify a clock to match the
clocking arrangement used by your target system. It can be as simple
as a single rising edge, or a complex arrangement of up to four signals.
If the clock is incorrect, the trace data may indicate a problem where
there isn't one. Specify the state clock in Clock Setup.
The equivalent in timing mode of the state clock is the Sample Period.
The Sample Period sets the time between logic analyzer samples. For
reliable data, the sample period should be no more than half of your
clock period. Many engineers prefer setting it to one-fourth of the clock
period.
Set up the Trace
The remaining controls finish your description of how you want to
capture data. The trigger position determines where the events you
specify in the trigger sequence will be relative to the majority of the
data the logic analyzer captures.
Memory depth is affected by the measurement mode. Some logic
analyzers also let you limit how big the acquisition will be with an
Acquisiton Depth control.
Step 2: Set Up the Bus Labels (see page 13)
12
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Setting Up a Measurement
See Also
“The Sampling Tab” on page 40 for information on setting type and
assigning pods
“Setting the Acquisition Mode” on page 41 for links to this analyzer's
modes
“Performing Clock Setup (State only)” on page 41
Set Up the Bus Labels
The next step is to finish defining the physical connection between the
target system and the analzyer. Use the Format tab to tell the analyzer
what you want to measure on the target system. If you load a
configuration file, this step is taken care of for you.
Group and Label Signals
Because the logic analyzer can capture dozens or even hundreds of
signals, you need to organize the signals by grouping and labeling
channels. Labels are used to group these channels into logical signals;
for example, "addr bus". These groupings are then used in the trigger
tab and the data displays. A label can have up to 32 channels. Each
measurement can define 126 labels. Active channels are indicated like
so
.
Set Threshold Level
The logic analyzer needs to know what threshold level the target
system is using. You can set the analyzer to use a Standard or a User
13
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Setting Up a Measurement
Defined threshold voltage. The logic analyzer requires a minimum
voltage swing of 500 mV at the probe tip to recognize changes in logic
levels.
Step 3: Define Trigger Conditions (see page 14)
See Also
“Assigning Bits to a Label” on page 66
“The Format Tab” on page 55
Define Trigger Conditions
The third step is to define the trigger. The trigger settings tell the
analyzer when you want to capture data. Controls for this are located
under the Trigger tab. Configuration files saved from previous
measurements automatically define trigger settings.
Set Up a Trigger Sequence
The trigger sequence is like a small program that controls when the
logic analyzer stores data. There are trigger functions for the common
tasks, or you can set up your own. The logic analyzer starts at the first
trigger level, and stays there until the conditions described in that level
become true. When that happens, the logic analyzer goes to the next
level and follows the instructions there.
14
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Setting Up a Measurement
Define Terms
Trigger terms are like variables that you use in the trigger sequence.
Depending on what analyzer you are using, you can either assign the
values directly from within the trigger sequence or from the tabs
(pattern, range, edge).
Step 4: Run the Measurement (see page 15)
See Also
“Defining Resource Terms” on page 94
“Understanding Logic Analyzer Triggering” on page 76
“Setting Up a Trigger” on page 78
“The Trigger Tab” on page 75
Measurement Examples (see the Measurement Examples help volume)
Run the Measurement
You run the measurement by selecting the Run button. The Run button
is labeled either Run, Group Run, or Run All. The difference between
the three types is that Run starts only the instrument you are using,
Group Run starts all instruments attached to group run in the
Intermodule window, and Run All starts all instruments currently
placed in the workspace.
15
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Setting Up a Measurement
Select Single or Repetitive
Runs can be single or repetitive. Single runs gather data until the logic
analyzer memory is full, and then stop. Repetitive runs keep repeating
the same measurement and are useful for gathering statistics. To stop a
run, select the Stop button.
NOTE:
Repetitive runs on a logic analyzer don't do equivalent time sampling like
oscilloscopes do.
If Nothing Happens...
Analyzers with deep memory take a noticeable amount of time to
complete a run. Because data is not displayed until acquisition
completes, it may look like nothing is happening. Check the Run Status
window to see if the logic analyzer is still running. Messages such as
"Waiting in level 1" may indicate you need to refine your trigger. If the
status shows as "Stopped", the analyzer either finished the acquisition,
or was unable to run. The cause of the problem is listed in the bottom
half of the Run Status window, and the messages are explained in more
detail in “Error Messages” on page 33.
Step 5: Examine the Data (see page 16)
See Also
“When Something Goes Wrong” on page 33
Examine the Data
Data from your measurement can be viewed in various display windows
or offline. Some of the things you can do in the display windows are
16
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Setting Up a Measurement
•
Search for patterns
•
Display time-correlated data
•
Use markers to make measurements and gather statistics
Search for Patterns
You can search displays for certain values, and place markers on them.
There are two global markers which keep their place across all
measurement views, even across instruments.
Display Correlated Data
There are several tools for correlation. The Intermodule window allows
you to specify complex triggering configurations using several
instruments. It is also useful for starting acquisitions at the same time.
Global markers mark the same events in different displays, so you can
switch views without having to reorient yourself. The Compare tool lets
you compare two different acquisitions to look for changes.
Use Markers to Make Measurements
The markers can be positioned relative to the beginning, end, trigger,
or another marker, as well as set to a specific pattern, state, or time.
The Markers tab in the Display windows shows the time or state value
as you move the markers or take new acquisitions.
See Also
Working with Markers (see the Markers help volume)
Using the Chart Display Tool (see the Chart Display Tool help volume)
Using the Distribution Display Tool (see the Distribution Display Tool
help volume)
Using the Listing Display Tool (see the Listing Display Tool help volume)
Using the Digital Waveform Display Tool (see the Waveform Display Tool
help volume)
Using the Compare Analysis Tool (see the Compare Tool help volume)
“Interpreting the Data” on page 26
17
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Making a Basic State Measurement
Making a Basic State Measurement
This example uses the circuit board that is supplied with the Making
Basic Measurements kit as the target system. The kit is supplied with
every logic analysis system, or can be ordered from your Agilent
Technologies Sales Office.
There are six major steps to making a basic measurement.
Connect the Logic
Analyzer to your
Target System
1. Connect probes.
•
Connect Pod 1 of the logic analyzer to J1 on the target system.
The training board has terminations and headers already built in to the
system, so you can connect the logic analyzer pod directly to the board.
2. Define the type of measurement On the Agilent Technologies 16700A/B
logic analysis system, open a logic analyzer setup window.
a. In the main window, select the logic analyzer icon.
b. Choose Setup... from the menu.
c. Select the Sampling tab.
d. If the logic analyzer is not already set for State, change the type to
State.
3. Set up the clock to match the target system's clocking scheme.
a. In the bottom half of the Sampling tab window, choose the correct
edges to match your clock. For Pod 1 attached to the training board,
the correct clock is the falling edge on J.
1. Select the Off button under J and choose "Falling Edge" from the
menu.
4. Group and label bits.
a. Select the Format tab.
b. Optional - Insert a second label.
18
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Making a Basic State Measurement
1. Select the Label1 button.
2. Choose Insert after....
3. In the Enter Label Name box, select the OK button.
c. Optional - Rename Label1.
1. Select the Label1 button.
2. Choose Rename....
3. Enter a new name in the name field.
4. Select the OK button to close the Rename Label box.
d. Select the bit assignment button.
The bit assignment button is to the right of a label name, and under a
pod column.
e. Choose ........******** from the menu.
If none of the choices match your own system, choose Individual...
and select the individual bits to assign them (*) or ignore them (.).
5. Define trigger events for patterns on buses.
a. Select the Pattern subtab in the Trigger tab.
b. Optional - Rename Pattern1.
1. Select the Pattern1 field.
2. Enter a new name.
c. Select the appropriate label.
1. Select the label name button.
2. To define the event as a combination of labels, choose Insert... To
use a different label to define the event, choose Replace...
3. In the dialog box, select the label name you want to use and then
select the OK button.
d. Select the field with XX and enter the value you want to trigger on.
e. Optional - Repeat steps a - d for Pattern2.
6. Optional - Add additional trigger events to the trigger specification.
19
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Making a Basic State Measurement
The logic analyzer automatically triggers on Pattern1, the first trigger
event. You can set up more complex triggers by editing the sequence levels
and combining trigger events.
a. Select the sequence level number 1 button and choose Edit...
b. In the dialog box, select the Pattern1 button just after Trigger on and
choose Combo...
c. In the Combination box, select the Off option button next to Pattern2
and choose On.
d. To change the trigger to Pattern1 and Pattern2, select the Or option
button to the right of the events and choose And.
e. Select the OK button.
f. Select the Close button.
The analyzer is now set to trigger when it detects both the pattern
defined by Pattern1 and the pattern defined by Pattern2 on the target
system's buses. The trigger sequence windows shows
Trigger on "(Pattern1.Pattern2)" 1 time
See Also
“The Trigger Tab” on page 75
1. Select the Run button.
2. Examine the data.
a. Select the Window menu.
b. Select Analyzer<A> in the menu and choose Listing<1>.
Depending on what other instruments are active, there may be more
than one Analyzer<A>. Choose the one that refers to your analyzer.
c. To have the listing display appear automatically when you run the logic
analyzer, select Options -> Popup on Run -> On in the menu bar of the
listing display.
d. To insert additional labels, select the label name.
See Also
For Connection Information
Logic Analysis System and Measurement Modules Installation Guide
20
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Making a Basic State Measurement
For Details on the Training Board or More Tutorials
Making Basic Measurements
Examples of Typical Timing Measurements
The "Looking at State Events" group under Hardware Turn-On (see the
Measurement Examples help volume) measurements.
Firmware Development (see the Measurement Examples help volume)
measurements.
System Integration (see the Measurement Examples help volume)
measurements.
For Details on the Logic Analyzer Interface
“The Sampling Tab” on page 40
“The Format Tab” on page 55
“The Trigger Tab” on page 75
21
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Making a Basic Timing Measurement
Making a Basic Timing Measurement
This example uses the circuit board that is supplied with the Making
Basic Measurements kit as the target system. The kit is supplied with
every logic analysis system, or can be ordered from your Agilent
Technologies Sales Office.
There are six major steps to making a basic measurement.
Connect the Logic
1. Connect probes.
Analyzer to the Target
• Connect Pod 1 of the logic analyzer to J1 on the target system.
System
The training board has terminations and headers already built in to the
system, so you can connect the logic analyzer pod directly to the board.
2. Define the type of measurement. On the Agilent Technologies 16700A/B
logic analysis system, open a logic analyzer setup window.
a. In the main window, select the logic analyzer icon.
b. Choose Setup... from the menu.
c. Select the Sampling tab.
d. If the logic analyzer is not already set for Timing, change the type to
Timing.
3. Group and label bits.
a. Select the Format tab.
b. Optional - Insert a second label.
1. Select the Label1 button.
2. Choose Insert after....
3. In the Enter Label Name box, select the OK button.
c. Optional - Rename Label1
1. Select the Label1 button.
22
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Making a Basic Timing Measurement
2. Choose Rename....
3. Enter a new name in the name field.
4. Select the OK button to close the Rename Label box.
d. Select the bit assignment button.
The bit assignment button is to the right of a label name, and under a
pod column.
e. Choose ........******** from the menu.
If none of the choices match your own system, choose Individual...
and select the individual bits to assign them (*) or ignore them (.).
4. Define trigger events for a bus.
a. Select the Trigger tab.
b. Select the Pattern subtab.
c. Optional - Rename pattern Pattern1.
1. Select the Pattern1 field.
2. Enter a new name.
d. Select the appropriate label.
1. Select the label name button.
2. To define the event as a combination of labels, choose Insert... To
use a different label to define the event, choose Replace...
3. In the dialog box, select the label name you want to use and then
select the OK button.
e. Select the field with XX and enter the value you want to trigger on.
5. Define trigger events for an edge.
a. Select the Edge subtab.
b. Optional - Rename Edge1.
1. Select the Edge1 field.
2. Enter a new name.
c. Select the appropriate label.
23
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Making a Basic Timing Measurement
1. Select the label name button.
2. To define the event as a combination of labels, choose Insert... To
use a different label to define the event, choose Replace... Edges
within an event are always OR'd together, which means only one of
the edges on one of the labels needs to occur for the edge event to
become true.
3. In the dialog box, select the label name you want to use and then
select the OK button.
d. Select the edge assignment button (........) and enter the edge or
edges you want to trigger on. Remember, if more than one edge is
specified, then when the logic analyzer detects any of the edges the
event becomes true.
6. Add the edge event to the trigger specification.
a. Select the sequence level number 1 button and choose Edit...
b. In the dialog box, select the Pattern1 button and choose Combo...
c. In the Combination box, select the Off option button next to Edge1 and
choose On.
d. Select the Or option button where the path from Pattern1 and the
path from Edge1 come together, and choose And.
e. Select the OK button.
The analyzer is now set to trigger when it detects Edge1 and Pattern1
on the bus. The trigger sequence window shows
Trigger on Pattern1.Edge1 occurs 1 times
.
The logic analyzer automatically triggers on the first trigger event. You can
set up more complex triggers by editing the sequence levels and defining
additional trigger events.
See Also
“The Trigger Tab” on page 75
1. Select the Run button.
2. Examine the data.
a. Select the Window menu.
24
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Making a Basic Timing Measurement
b. Select Analyzer<A> in the menu and choose Waveform<1>.
Depending on what other instruments are active, there may be more
than one Analyzer<A>. Choose the one that refers to your analyzer.
c. To have the waveform display appear automatically when you run the
logic analyzer, select Options -> Popup on Run -> On in the menu bar
of the waveform display.
d. To insert additional labels, or expand overlaid signals, select the label
name.
See Also
For Connection Information
Logic Analysis System and Measurement Modules Installation Guide
For Details on the Training Board or More Tutorials
Making Basic Measurements
Examples of Typical Timing Measurements
Hardware Turn-On (see the Measurement Examples help volume)
measurements.
Firmware Development (see the Measurement Examples help volume)
measurements.
System Integration (see the Measurement Examples help volume)
measurements.
For Details on the Logic Analyzer Interface
“The Sampling Tab” on page 40
“The Format Tab” on page 55
“The Trigger Tab” on page 75
25
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Interpreting the Data
Interpreting the Data
After you've acquired a trace with the logic analyzer, you can analyze it
in the display tools. The logic analysis system also provides filtering
and compare tools for more complex analysis.
The logic analyzer is automatically connected to the Waveform and
Listing displays when you set up a measurement. To move to that
display,
1. Select the Window menu.
2. Select the name of the analyzer whose data you want to view.
3. Choose Waveform or Listing.
Source Viewer brings up a Listing display but requires an inverse
assembler and an ADDR label.
•
“Analysis Using Waveform” on page 26
•
“Analysis Using Listing” on page 28
Analysis Using Waveform
Waveform is most useful for timing data. If you look at state data that
uses store qualification, you won't be able to easily see where samples
were not stored. Timing data, however, is periodic and stores all
samples and so works well with Waveform.
Example: Looking for a Missing Pattern
You can easily use the waveform tool to make timing measurements.
For example, if you were triggering when a pattern doesn't follow an
edge within a certain time (see the Measurement Examples help
volume), you would probably want to look at your data set to see if the
pattern ever did occur. This might be the case when you verifying that
the system is responding to an interrupt.
After triggering on an instance where the response did not appear
quickly enough, you might take these steps in the Waveform display:
26
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Interpreting the Data
1. Find the edge.
a. Select the Search tab.
b. Select the down arrow after the Label field, and select the label
containing the edge.
c. Select the Value field and enter 1.
d. Select the Next button to locate the edge transition.
2. Place a marker on the edge.
Select the Set G1 button. This sets global marker G1 at the location of the
edge you just found.
3. Search for the pattern. Searches start at your current location. Since you
just set the global marker G1, it indicates where the search starts from.
a. Select the down arrow after the Label field, and select the label
containing the pattern.
b. Select the Value field and enter the pattern you are searching for.
c. Select the down arrow after the When field and select Entering.
d. Select the Next button to find the next occurrence of that pattern after
G1.
If the logic analysis system cannot find the pattern, a "Value not found"
message pops up.
4. Place a marker on the pattern.
Select the Set G2 button. This will set global marker G2 at the location of
the pattern.
5. Find the time between the edge and the pattern.
a. Select the Markers tab.
b. In the G2 row, select the down arrow after from, and select G1.
The value after the from field changes to the time between G1 and G2. You
can toggle between time and samples by selecting the arrow after the Time
or Samples field.
See Also
Using the Digital Waveform Display Tool (see the Waveform Display Tool
27
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Interpreting the Data
help volume)
Using the Listing Display Tool (see the Listing Display Tool help volume)
Using the Chart Display Tool (see the Chart Display Tool help volume)
Using the Distribution Display Tool (see the Distribution Display Tool
help volume)
Using the Compare Analysis Tool (see the Compare Tool help volume)
Using the Pattern Filter Analysis Tool (see the Pattern Filter Tool help
volume)
Analysis Using Listing
Listing is more useful than Waveform when your target system is
running code because it shows the labels as states rather than
transitions. Listing is especially useful when you have defined
meaningful symbol names for your states. If you have an inverse
assembler, you might prefer Source Viewer which functions like
Listing.
Example: Examining a Subroutine
Listing is the preferred display tool for state measurements. for
example, if you were trying to see if a subroutine were exiting
abnormally, you might want to measure the number of states between
entering and exiting the subroutine. After acquiring data with the logic
analyzer, you could examine the data set in the Listing display like this:
1. Find the start of the subroutine.
Assume the subroutine starts at the address 0x58FC.
a. Select the Search tab.
b. Select the down arrow after the Label field, and select ADDR.
c. Select the Value field, and enter the starting address, 0x58FC.
d. Select the down arrow after the When field and select Present.
e. Select the Next or Prev button to move the display to the address.
28
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Interpreting the Data
2. Place a marker on the start of the subroutine.
Select the Set G1 button. This sets global marker G1 at the address you
just found.
3. Find the end of the subroutine.
Assume the end of the subroutine is at address 0x58FF. Searches always
start at the current location. Since you just set the global marker G1, it
indicates where the search starts from.
a. Select the Value field, and enter 58FF.
b. Select the Next button to find the next occurrence of 0x58FF after the
starting address.
4. Place a marker on the end of the subroutine.
Select the Set G2 button to set global marker G2 at this position. This lets
you refer to G2 when you want to know where the subroutine ends.
5. Find the number of states between the start and end of the subroutine.
Since you've placed markers at the start and end of the subroutine, all you
have to do is find the number of states between those markers.
a. Select the Markers tab.
b. In the G2 row, select the second down arrow and select Sample.
c. Select the down arrow after from, and select G1.
The value after the from field changes to the number of states between G1
and G2. You can toggle between time and states by selecting the arrow
after the Time or Samples field.
Now you know how long the execution stayed in the subroutine, and
can also examine the data set between G1 and G2 to look for unusual
data.
See Also
Using the Digital Waveform Display Tool (see the Waveform Display Tool
help volume)
Using the Listing Display Tool (see the Listing Display Tool help volume)
Using the Chart Display Tool (see the Chart Display Tool help volume)
Using the Distribution Display Tool (see the Distribution Display Tool
29
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Interpreting the Data
help volume)
Using the Compare Analysis Tool (see the Compare Tool help volume)
Using the Pattern Filter Analysis Tool (see the Pattern Filter Tool help
volume)
30
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Coordinating Measurements
Coordinating Measurements
When you want to coordinate the triggering of two measurements
happening on the same logic analyzer, you need to use arming control.
The two measurement machines are by default controlled by the same
run button. However, you can set machine 1 to arm machine 2, or the
other way around.
If the timing measurement and state measurement are being handled
by two different instruments, you can coordinate them by connecting
both instruments to the same display, and then setting the correlation
in the correlation dialog that appears.
You can tell which instrument is controlling a measurement by looking
at the instrument icon. The letter on the icon represents the slot of the
frame that the instrument is in. Icons with the same letter are
controlled by the same instrument.
See Also
Overview - Multiple Instrument Configuration (see the Agilent
Technologies 16700A/B-Series Logic Analysis System help volume)
Overview - Multiple Machine Configuration (see the Agilent Technologies
16700A/B-Series Logic Analysis System help volume)
31
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Loading and Saving Logic Analyzer Configurations
Loading and Saving Logic Analyzer
Configurations
The Agilent Technologies 16550A logic analyzer settings and data can
be saved to a configuration file. You can also save any tools connected
to the logic analyzer. Later, you can restore your data and settings by
loading the configuration file into the logic analyzer.
The Agilent Technologies 16550A logic analyzer can also load
configuration files generated for Agilent Technologies 16510B, 16511B,
and 16540A models of logic analyzer, but they must first be re-saved as
Agilent Technologies 16550A configuration files on the old logic
analysis system before being loaded into a Agilent Technologies
16550A on an Agilent Technologies 16700A/B-series logic analysis
system. Also, the data cannot be displayed across logic analyzer
models, and some setting may change when no comparable setting
exists in the Agilent Technologies 16550A logic analyzer.
See Also
Loading Configuration Files (see the Agilent Technologies 16700A/BSeries Logic Analysis System help volume)
Saving Configuration Files (see the Agilent Technologies 16700A/BSeries Logic Analysis System help volume)
32
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
When Something Goes Wrong
When Something Goes Wrong
•
“Nothing Happens” on page 38
•
“Error Messages” on page 33
•
“Suspicious Data” on page 38
•
“Interference with Target System” on page 33
Interference with Target System
Capacitive Loading on Excessive capacitive loading can degrade signals, resulting in
the Target System
suspicious data or even system lockup. All analysis probes add
capacitive loading, as can custom probes you design for your target. To
reduce loading, remove as many pin protectors, extenders, and
adapters as possible.
Careful layout of your target system can minimize loading problems
and result in better margins for your design. This is especially
important for systems running at frequencies greater than 50 MHz.
Error Messages
“Slow or Missing Clock” on page 34
“Waiting for Trigger” on page 37
“Timer is Off in Sequence Level Where It Is Used” on page 35
“Timer is Specified in Sequence, But Never Started” on page 35
“Measurement Initialization Error” on page 34
“Two Pod Pairs are Needed To Use Both Timers” on page 36
“Maximum of 32 Channels Per Label” on page 34
“Trigger inhibited during timing prestore” on page 36
33
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
When Something Goes Wrong
“Clock Lines Must Be Connected to the Card in Slot X” on page 37
Maximum of 32 Channels Per Label
The logic analyzer can only assign up to 32 channels for each label. If
you need more than 32 channels, assign them to two labels and use the
labels in conjunction.
See “Adding and Deleting Labels for Terms” on page 100 for how to use
more than one label with trigger terms.
Measurement Initialization Error
The logic analyzer module failed the internal calibration which it
performs when Run is selected. An internal calibration failure can
indicate either a software or a hardware problem.
Possible Causes
•
Hardware failure
•
Software failure
Run the Self-Test Utility (see page 39) on the logic analyzer and
contact your Agilent Technologies Sales Office for service or software
upgrades.
Slow or Missing Clock
The message "Slow or Missing Clock" only appears in state
measurements. However, if you have another instrument armed by the
state analyzer, a slow or missing clock on the state analyzer will prevent
the other instrument from triggering also.
Possible Causes
•
Target system is not running properly
Check that the system is running properly. The logic analyzer and other
probing fixtures such as pin extenders can place too much capacitive load
on a system.
•
34
Incorrect clock specification
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
When Something Goes Wrong
Make sure the target system clock matches the clock specified under
Sampling.
Also check that the probe's clock channels are attached to the target's
clock lines either directly or through an analysis probe.
If you are using an analysis probe, the probe's User's Guide should show
the correct connections and settings.
•
Bad probe connection
Check that the probe is securely attached to the clock line and is receiving
a signal. The logic analyzer shows activity indicators under the Sampling
and Format tabs.
•
Incorrect signal level
The clock's threshold level is set by the pod threshold. For the logic
analyzer's J clock, check the pod threshold of pod 1 of the master card.
See Also
“Performing Clock Setup (State only)” on page 41
“Setting the Pod Threshold” on page 72
Timer is Specified in Sequence, But Never Started
This error occurs because you have used a timer term in your trigger
sequence, but not turned the timer on using the timer controls. Timer
controls are available in all sequence levels except the first. You do not
need to turn on the timer in the same sequence level, but it does need
to be on when it is used in the trigger sequence.
This error message always occurs with the
Timer is off in sequence level where it is used
warning.
See “Using Timer Terms” on page 98 for more information on timer
control.
Timer is Off in Sequence Level Where It Is Used
Timer controls are available in all sequence levels except the first. You
do not need to turn on the timer in the same sequence level, but it does
need to be on when it is used in the trigger specification.
35
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
When Something Goes Wrong
Possible Causes
This warning occurs because you have used a timer term in your
trigger specification, but not turned the timer on using the timer
controls.
This warning message always occurs with the
Timer is specified in sequence, but never started
error message.
See “Using Timer Terms” on page 98 for more information on timer
control.
Trigger inhibited during timing prestore
The "trigger inhibited" informational message appears when you have a
logic analyzer making a timing measurement, and it is set to a slow
sample rate. The logic analyzer will fill the designated amount of pretrigger memory before checking for the trigger condition.
To calculate how long this should take, multiply the sample rate by the
percentage of pre-trigger memory and the acquisition depth. For
example, if
sample period = 1.0 ms (sample rate = 103 samples/sec.)
trigger position = center (percentage of pre-trigger memory = 50%)
acquisition depth = 64K (roughly 64 x 103 samples)
then the approximate time is 32 seconds.
Two Pod Pairs are Needed To Use Both Timers
Possible Causes
This message only occurs when all pod pairs are assigned, and the
machine which is using both timers has only one pod pair.
You can either assign another pod pair to the machine with both timers,
or change your trigger specification to use only one timer term.
See Also
“Using Timer Terms” on page 98
36
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
When Something Goes Wrong
Waiting for Trigger
This message indicates that the specified trigger pattern has not
occurred. This may be expected, as when you are waiting to trigger on
an unusual event.
Possible Causes
•
Misaligned boundaries for addresses
When the target is a microprocessor that fetches only from long-word
aligned addresses, if the trigger is set to look for an opcode fetch at an
address that is not properly aligned, the trigger will never be found.
•
Trigger set incorrectly
Some strategies you can use when verifying or debugging trigger
sequence levels are:
•
Look at the run status message line or open the Run Status window. It will
tell you what level of the sequence the logic analyzer is in.
•
Stop the measurement and look at the data that was captured. This is
particularly useful when you use store qualifiers to store "no states" (or
only the states you are interested in) and the branches taken are stored.
•
Save the trigger setup, then simplify it to see what part of the sequence
does get captured. When you learn what needs to be changed, you can
recall the original trigger setup and make changes to it.
See Also
“Branches Taken Stored / Not Stored (State only)” on page 94
“Saving and Recalling Trigger Sequences” on page 82
Clock Lines Must Be Connected to the Card in Slot X
This message occurs if the logic analysis system has detected a slow or
missing clock on a logic analyzer module that has an expander card.
Possible Causes
•
Using a clock channel from an expander card
Reconnect the logic analyzer probes to use a clock channel from the
master card of the module.
37
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
When Something Goes Wrong
Only the clock channels of the master card can be used in the state clock.
The master card of a module is identified in the system window.
•
Clock properly connected, but slow
See “Slow or Missing Clock” on page 34.
Nothing Happens
Look for an error message in the message bar at the top of the window.
Common messages are "slow or missing clock" and "Waiting for trigger".
If the Run button briefly grayed-out (and the Stop or Cancel buttons
briefly became active), select the Window menu, select the logic
analyzer's slot, then choose the Waveform or Listing display.
See Also
“Slow or Missing Clock” on page 34
“Waiting for Trigger” on page 37
Suspicious Data
Intermittent Data
Errors
Check for poor connections, incorrect signal levels on the hardware,
incorrect logic levels under the logic analyzer's Config tab, or marginal
timing for signals.
Unwanted Triggers
If you are using an inverse assembler or a pipeline, triggers can be
caused by instructions that were fetched but not executed. To fix, add
the prefetch queue or pipeline depth to the trigger address.
The depth of the prefetch queue depends on the processor that you are
analyzing, and can be quite deep.
Another solution which is sometimes preferred with very deep prefetch
queues is to add writes to dummy variables to your software. Put the
instruction just before the area you want to trigger on, then trigger on
the actual write to this variable. Although the instruction is prefetched,
the analyzer can be set to only trigger when the write is executed.
38
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Testing the Logic Analyzer Hardware
Testing the Logic Analyzer Hardware
In order to verify that the logic analyzer hardware is operational, run
the Self Test utility. The Self Test function of the logic analysis system
performs functional tests on both the system and any installed
modules.
1. Disconnect all probes of the logic analyzer module.
2. If you have any work in progress, save it to a configuration file. (see the
Agilent Technologies 16700A/B-Series Logic Analysis System help
volume)
3. Disconnect all loads, adapters, or preprocessors from the probe cable
ends.
4. From the system window, select the System Administration button.
5. Select the Admin tab, then select the Self Test... button.
The system closes all windows before starting up Self Test.
6. Select Master Frame.
If the module is in an expansion frame, select Expansion Frame.
7. Select the logic analyzer that you want to test.
8. In the Self Test dialog box, select Test All.
You can also run individual tests by selecting them. Tests that require you
to do something must be run this way.
If any test fails, contact your local Agilent Technologies Sales Office or
Service Center for assistance.
See Also
Self Test (see the Agilent Technologies 16700A/B-Series Logic Analysis
System help volume)
Agilent Technologies 16550A 100 MHz State/500 MHz Timing Service
Guide
39
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Sampling Tab
The Sampling Tab
The options under Sampling tell the analyzer the overall way in which
you want to make a measurement. The options include:
•
The acquisition mode.
•
The data sample rate.
•
How much data before and after the trigger.
•
How much data to acquire in all.
See Also
“Naming the Analyzer” on page 44
“Turning the Analyzer Off” on page 45
“Setting the Acquisition Mode” on page 41
“Sample Period (Timing Only)” on page 45
“Performing Clock Setup (State only)” on page 41
“Trigger Position Control” on page 53
“Acquisition Depth” on page 40
Acquisition Depth
Acquisition Depth, located under Sampling and also under Trigger
Settings, sets how much data will actually be acquired. While the
Agilent Technologies 16550A logic analyzer has a maximum memory
depth of 4K samples in State Mode and 8K in Timing Mode, sometimes
you may not want to wait for all that memory to fill up.
The numbers shown in the Acquisition Depth menu are
approximations. The combination of count tags, pod assignments, and
acquisition modes affect what choices are available. Also, the values
are memory depth per channel.
40
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Sampling Tab
Setting the Acquisition Mode
The measurement type affects all other areas of the logic analyzer
setup. It is set under the Sampling tab.
If you want the logic analyzer to sample data according to the target
system's clock, select State. Each time the clock signal becomes valid,
the analyzer will sample data from the system under test.
If you want the logic analyzer to sample the target system
independently of its clock, select Timing. Since in timing mode the
analyzer is clocked by a signal that is not related to the system under
test, timing measurements capture traces of electrical activity over
time. The Agilent Technologies 16550A logic analyzer can be split into
two analyzers, but only one of them may be a timing analyzer.
State Mode and Timing Mode offer different options for the
acquisition mode field in the top left corner under Controls. The
acquisition mode affects the channel width, memory depth, and sample
rate.
“State Acquisition Modes” on page 46
“Timing Acquisition Modes” on page 46
“Transitional Timing Acquisition Modes” on page 46
“Comparison of Transitional and Conventional Timing Modes” on page 51
“Glitch Acquisition Mode” on page 52
Performing Clock Setup (State only)
When you select State Mode, the Clock Setup area appears under the
State Mode Controls in Sampling. The Clock Setup contains three
controls and a display area. The clocks you specify here control when
the analyzer samples your data. Ideally, the logic analyzer's state clock
setup should be identical to the target system's clock. Differences
could result in invalid data.
41
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Sampling Tab
Mode field
The Mode field lets you select among Master only, Master/slave, and
Demultiplex. The default is Master only. When you select the others,
another control to set the slave clock appears at the bottom of the
Clock Setup area. It also enables the Pod Clock field under Format.
For more detail on the uses of Master/slave and Demultiplex clocking,
see “Clock Modes (State only)” on page 42.
Advanced Clocking
Advanced clocking allows you to specify clock qualifiers on individual
clock edges instead of the group of clock edges. When you select it, the
individual clock channels are replaced by Master Clock... or Slave
Clock... Selecting these brings up a dialog that lets you combine edges
and qualifiers in more complex Boolean expressions. When you switch
from Advanced Clocking to regular clocking, some of the qualifiers are
erased.
Clock Channel Specifiers
The clock channel specifiers graphically show your clock setup. Edges
are ORed ("+") together, and qualifiers are ANDed (".") to all edges. To
qualify just one of the edges, switch to Advanced Clocking.
All clock channels for the clock setup must be on the pods of the
master card of the module, but the pods do not need to be part of the
state measurement.
See Also
“Clock Modes (State only)” on page 42
Clock Modes (State only)
The Pod Clock field under Format appears when a clock mode other
than Master only is selected in Sampling. The Pod Clock field
indicates whether a pod's data lines are to be sampled into memory by
the master clock, slave clock, or both (demultiplex).
The Pod Clock field and the clocking arrangement are only available in
a state analyzer.
42
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Sampling Tab
Master
Data on all pods assigned to Master Clk is strobed into memory when
the status of the clock lines match the clocking arrangement specified
for Master in the clock setup.
Slave
Data on a pod designated Slave Clock is latched when the status of the
slave clock inputs meet the requirements of the slave clocking
arrangement. Then, followed by a valid master clock, the slave data is
strobed into analyzer memory along with the master data.
If multiple slave clocks occur between master clocks, only the data
latched by the last slave clock prior to a valid master clock is strobed
into analyzer memory.
Latching Slave Data
Demultiplex
The demultiplex mode is used to store two different sets of data that
occur at different times on the same channels. In demultiplex mode,
only one pod of the pod pair is used, and that pod is selectable (see
page 71). Channel assignments are displayed as Pod and Slave Pod.
Assign slave and master data to separate labels for easy recognition of
the two sets of data.
Both the master and slave clocks are used in the demultiplex mode.
When the analyzer sees a match between the slave clock input and the
slave clock arrangement, demux slave data is latched. Then, following a
valid master clock, the slave data is strobed into analyzer memory
43
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Sampling Tab
along with the master data. If multiple slave clocks occur between
master clocks, only the data latched by the last slave clock prior to the
master clock is strobed into analyzer memory.
Latching Demultiplex Data
Naming the Analyzer
The Analyzer Name field under Sampling lets you assign a specific
name to the analyzer. When you have stored several measurement
setups to disk and later reload them, having assigned a specific name to
an analyzer can help identify what the setup is for. The analyzer name
is also used in the workspace to label the analyzer icons and as part of
the title of the analyzer setup window.
There are two analyzers per logic analyzer module. To activate the
second one, go to the Workspace window and drag the second analyzer
on to the workspace.
The default names for the analyzers are Analyzer<N> and
Analyzer<N2>, where N is the slot of the analyzer module.
To Name an Analyzer
1. Select the Analyzer Name field.
2. Enter the new name.
The name now appears below the instrument tool icon in the
workspace.
44
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Sampling Tab
Turning the Analyzer Off
The On and Off checkboxes under the Sampling tab are a shortcut for
activating and de-activating the analyzer.
Analyzers come up in the On state. If you select the checkbox, the
label changes to Off and the Analyzer Shutdown Options dialog
appears.
Soft
"Soft" shutdowns have the same effect as though you selected Off in
the Type control of the Pod Assignment dialog under Format. The
analyzer window remains, but most options are unavailable. Soft
shutdowns are easily reversed by selecting the box next to Off.
Hard
"Hard" shutdowns have the same effect as deleting the analyzer icon
from the workspace. The analyzer window and the windows of its
display tools are closed, and the analyzer is removed from the
workspace. To turn the analyzer back on, select the analyzer icon in the
System window. You will need to restore any complex analysis or
display tools.
Sample Period (Timing Only)
Sample Period is used to set the time between data samples. The
inverse of sample period is sample rate. Every time a new sample is
taken, the analyzer will see updated measurement data. The choices
available for sample period depend on the acquisition mode.
If your analyzer is set to timing, the Sample Period control is located
under the Sampling tab's Timing Mode Controls and under the Trigger
tab's Settings sub-tab.
If your analyzer is set to state, the Sample Period control is not
available. Its functionality is handled by the Clock Setup.
45
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Sampling Tab
State Acquisition Modes
Full Channel/4K Memory/100 MHz
All pods are available. Memory depth is 4K samples per channel. If time
or state count is turned on, the total memory could be split between
data acquisition storage and time or state count storage. To maintain
the full 4K samples per channel depth, leave one pod pair unassigned.
State clock speed is 100 MHz.
Half Channel/8K Memory/100 MHz
Only one pod per pod pair is available, for 17 channels of width and a
total memory depth of 8192 bytes. Time and state count is not available
in this mode. State clock speed is 100 MHz.
Timing Acquisition Modes
In conventional timing acquisition mode the analyzer stores
measurement data at each sampling interval.
Conventional Full Channel 250 MHz
The total memory depth is 4K samples per channel, with data being
sampled and stored as often as every 4 ns.
Conventional Half Channel 500 MHz
Only one pod of each pod pair is available. The total memory depth is
8192 samples, with data being sampled and stored as often as every 2
ns.
See Also
“Sample Period (Timing Only)” on page 45
“Pod Selection” on page 71
Transitional Timing Acquisition Modes
In transitional timing modes, the timing analyzer samples data at
regular intervals just like it does in conventional timing mode.
However, in transitional timing mode, it only stores data when there is
a change on the lines assigned to a label within a pod pair.
46
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Sampling Tab
Data is collected on a pod pair basis, even when labels cross pod
boundaries. It is still displayed on a label basis, however. Because of
this, some pod pairs with less active channels may take longer to fill
memory and the end of the acquisition is delayed.
Each time a transition occurs on any of the assigned bits, all bits of the
pod pair are stored along with a time tag. If transitions are
exceptionally far apart, the timer which generates the time tags may
overflow.
NOTE:
The timer does not generate a warning when it overflows. If the time between
transitions in the display seems too short, it may be due to timer overflow.
“Transitional Full Channel 125-MHz Mode” on page 47
“Transitional Half Channel 250-MHz Mode” on page 48
“Other Transitional Timing Considerations” on page 50
Transitional Full Channel 125-MHz Mode. The total memory
depth is 4K samples per channel with a channel width of 34 channels
per pod pair. Data is sampled for new transitions every 8 ns. When the
timing analyzer runs in the 125-MHz mode, it operates very much like
the state analyzer with Count Time turned on. The only exceptions are
that the store qualification comes from transition detectors instead of
the sequencer and that the analyzer uses an internal clock.
In full-channel transitional timing mode, the analyzer uses half its
memory (2 K) to store time tags. Because each pod pair stores
transitions at its own rate, you do not have the option of using a free
pod to retain full memory as you have in the state mode.
When a transition is detected after a sample with no detected
transition, two samples are stored. One sample is a "before transition
sample" and the other is an "after transition sample." As long as there
are transitions in the subsequent sample, only one sample is stored.
When the next sample occurs without a transition, the two stored
sample sequences (one before, one after) repeats with the next
detected transition.
47
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Sampling Tab
Minimum Transitions Stored
Minimum and
Maximum Transitions
Normally, transitions have not occurred at each sample period. This is
Stored
illustrated below with time-tags 2, 5, 7, and 14. When transitions
happen at this rate, two cycles are stored for every transition.
Therefore, with 2 K samples of memory, 1 K of transitions are stored.
You must subtract one, which is necessary for a starting point, for a
minimum of 1023 stored transitions.
Maximum Transitions Stored
If transitions occur at a fast rate, such that there is a transition at each
sample point, only one sample is stored for each transition as shown by
time tags 17 through 21 below. If this continues for the entire trace, the
number of transitions stored is 2 K samples. Again, you must subtract
the starting point sample, which then yields a maximum of 2047 stored
transitions.
In most cases a transitional timing trace is stored by a mixture of the
minimum and maximum cases. Therefore, the actual number of
transitions stored will be between 1023 and 2047.
Transitional Half Channel 250-MHz Mode. The total memory
depth is 8 K samples with a channel width of 17 channels on one pod.
The pod used within the pod pair is selectable. Data is sampled for new
transitions every 4 ns. Transitional timing running at 250 MHz is the
same as the 125-MHz mode, except that two single-pod data samples
(17 bits x 2 = 34 bits) are stored instead of one full pod pair data
sample (34 bits). Storing two single-pod data samples occurs because
in half channel mode, data is multiplexed into the sequencer pipeline in
48
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Sampling Tab
two 17-bit samples. The first 17-bit sample is latched, the next 17-bit
sample is sent down the pipeline along with the latched 17-bit sample.
Note that the transition detector still looks at a full 34 bits. This means
it is looking at two samples at a time instead of one. In this mode,
between 682 and 4094 transitions are stored.
Minimum and
Minimum Transitions Stored
Maximum Transitions
The following example shows the kind of data stored from a data
Stored
stream with transitions that occur at a slow rate (more than 24 ns
apart).
The illustration depicts transitions stored in two different ways,
depending strictly on chance. Remember that the transition detector
only looks at the full 34 bits while the data is stored as two 17 bit
samples. So, the transition detector will not see time tag 3 (101/000) as
a transition. However, when it compares it to time tags 2 (101/101) or 4
(000/000), it sees a difference and detects them as transitions. For this
first set of time tags, the transition detector sees more transitions than
are really there. This causes the analyzer to store 6 samples per
transition (three 34-bit sample pairs), instead of just two, like in the
125-MHz mode. If all the transitions will be stored in this way
throughout the trace, the minimum number of stored transitions is 682
(4096/6).
With time tags such as 7 (000/000) and 8 (001/001), transitions can fall
between the pairs of samples. When this happens, only one transition is
49
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Sampling Tab
detected and only 4 samples (two sample pairs) are stored. If all
transitions will be stored in this way, 1023 (4096/4) transitions are
stored.
From run to run, the actual number of transitions stored for transitions
that occur at a slower rate will fall between these two numbers. This
storage occurs based on the probability of a transition falling between a
sample pair or falling within a sample pair.
Maximum Transitions Stored
The following example shows the case where the transitions are
occurring at a 4-ns rate:
In this case, transitions are being detected with each sample.
Therefore, they are all being stored. In addition, each sample pair
contains a transition. For example, time tag 1 (100/000) contains a
transition and is different from time tag 2 (111/011), which also
contains a transition. The difference between the two will trigger the
transition detector. If this same trend occurs throughout the trace, the
analyzer would store (4 K samples - 1) transitions, or 4095. As with the
125-MHz mode, the actual number of transitions stored will fall
somewhere between 682 and 4095, depending on the frequency of
transitions.
Other Transitional Timing Considerations. Pod Pairs are Independent
In single run mode, each pod pair runs independently. When one pod
50
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Sampling Tab
pair fills its trace buffer, it will not shut the others down. If there is a
pod pair with enabled data lines and with no transitions on its lines, you
get a message "Storing transitions after trigger for pods nn/nn." In
repetitive run mode, a full pod pair waits 2 seconds, then halts all other
pod pairs.
Increasing Duration of Storage
In the 125-MHz mode, a transition on any one of the 34 bits for each
sample (if they are all turned on) will cause storage. Reducing the
number of bits that are turned on for any one pod pair will more than
likely increase data storage time. Separating data lines which contain
fast occurring transitions from lines with slow occurring transitions
also helps. If separating the data lines, be sure to cross pod pair
boundaries. It does not help to move fast lines from pod 1 to pod 2;
they must be moved to pod 3, which is a different pod pair. In the 250MHz mode, a transition on any one of the 17 bits will cause storage.
Invalid Data
The analyzer only looks for transitions on data lines that are turned on.
Data lines that are turned off store data, but only when one of the lines
that is turned on transitions. If the data line is turned on after a run,
you would see data, but it is unlikely that every transition that occurred
was captured.
Comparison of Transitional and Conventional Timing
Modes
In Conventional Timing Acquisition mode, the analyzer stores
measurement data at each sampling interval. In Transitional Timing
Acquisition mode, the timing analyzer samples data at regular intervals,
but only stores data when there is a threshold level transition. Each
time a level transition occurs on any of the bits of a pod pair, data on all
channels is stored. A time tag is stored with each stored data sample so
the measurement can be reconstructed and displayed later.
When using transitional timing, you might consider how many
transitions can be stored. The number depends on the mode and
frequency of transition occurrence.
51
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Sampling Tab
See Also
“Timing Acquisition Modes” on page 46
“Transitional Timing Acquisition Modes” on page 46
Glitch Acquisition Mode
In Glitch Acquisition mode, a glitch is defined as a pulse with a
minimum width of 3.5 ns and a maximum width of 8 ns, or the sample
period, whichever is larger. As an example, if the sample period is 16
ns, then a glitch is defined as being between 3.5 ns and 16 ns. One
advantage of the glitch mode is that if you expand the sample rate, a
pulse that is less than the sample rate will still be displayed as a vertical
dashed line.
Glitch Half Channel 125 MHz
The total memory depth is split between data storage and glitch
storage. Data acquisition memory depth is 2048 per channel. Glitch
storage is 2048 per channel. Data is sampled for new transitions at
most every 8 ns.
52
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Sampling Tab
Trigger Position Control
The Trigger Position control, located under Sampling and also Trigger
Settings, determines how much data is stored before and after the
trigger occurs for all subsequent acquisitions. The trigger point is
placed at a specified position relative to the data in memory. The
analyzer triggers differently depending on if it is in Timing or State
mode.
Timing Mode
When a Run is started, the analyzer will not look for a trigger until at
least the proper percentage of pretrigger data has been stored. After a
trigger has been detected, the specified percentage of posttrigger data
is stored before the analyzer halts.
State Mode
When the Run is started, the analyzer immediately looks for the trigger
condition. The percentage of pretrigger data determines the
maximum amount of pretrigger data to save.
The trigger position choices are Start, Center, End, User Defined, or
Delay. Delay is only available in timing mode.
•
Start
The trigger position is set at the starting point of available memory. This
process results in maximum posttrigger data and minimum pretrigger
data. Note that there will be a small amount of pretrigger data stored.
•
Center
The trigger position is set at the center point of available memory. This
results in half pretrigger data and half posttrigger data.
•
End
The trigger position is set at the end point of available memory. This
results in maximum pretrigger data and minimum posttrigger data. Note
that there will be a small amount of posttrigger data stored.
•
User Defined
53
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Sampling Tab
When the trigger position is set to User Defined, a trigger position slider
appears. Use this slider to set the trigger position any where between 0%
and 100%. As the slider is adjusted, the % Post Store indicator shows the
amount of data that will be stored after the trigger point.
•
Delay
This option delays the start of data storage until some time after the
trigger. The range of the delay time is affected by the sample period, but it
could range between 16 ns to 8 ks.
54
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Format Tab
The Format Tab
Under Format, you specify the parts of the target system that you
want the logic analyzer to look at. You set up labels to match the buses
of the target system, and set the threshold level for the signals. For a
state measurement, you also adjust the setup and hold time.
For advanced measurements, Format is also where you assign pods
and specify whether to look at channels on the master, slave, or
demultiplexed clock.
Common
Measurement
Controls
“Labels: Mapping Analyzer Channels to Your Target” on page 70
“Setting the Pod Threshold” on page 72
“State Clock Setup/Hold (State only)” on page 73
Advanced
Measurement
Controls
“Assigning Pods to the Analyzers” on page 65
“Clock Modes (State only)” on page 42
“Setting Up the Pod Clock” on page 70
“Pod Selection” on page 71
See Also
“Data On Clocks Display” on page 66
“Activity Indicators” on page 64
“Assigning Bits to a Label” on page 66
“Inserting and Deleting Labels” on page 67
“Turning Labels On and Off” on page 68
“To reorder bits in a label” on page 69
“Importing Netlist and ASCII Files” on page 56
“Label Polarity” on page 68
55
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
Importing Netlist and ASCII Files
Netlist Files
The Netlist Import feature provides a method for importing busses and
signals from ASCII netlists created by EDA tools. In order for the
feature to work, the device under test must either use the Agilent
E5346A high density adapter or the Agilent Technologies Termination
Adapter. The adapter must be included as a connector in the netlist.
You can create a Netlist file using an EDA tool. When importing a
Netlist file, you can specify which connectors are of interest;
connectors that are not specified will be ignored.
Example
NET'/Bus1(3)'J1-7
Bus1
Four bit bus
(3)
Bit 3
J1-7
Connector J1, pin 7
To Import a Net List
1. Configure the connector names referenced in the Netlist to logic analyzer
pods.
2. Specify the Net List file to import.
3. Verify that Nets were correctly mapped to Labels, Pods, and Channels.
4. Select or create labels to appear in the interface.
Map the Connector Names (see page 62).
Import the Net List File (see page 62)
Verify Net to Label Mapping (see page 63)
Select/Create Interface Labels (see page 64)
ASCII Files
You can create an ASCII file using any Windows, MS-DOS, or UNIX text
editor. The ASCII file should contain bus names, pod/clock numbers,
and channel definitions.
56
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Importing Netlist and ASCII Files
For Example
Label1;A2[15:5];A1[5,2]
Label1
Bus Name
A2 and A1
Pod Numbers
[15:5]
Channel 15 through Channel 5 ("***********.....")
[5,2]
Channel 5 and Channel 2 ("..........*..*..")
When setting up the ASCII file a comma (",") separates individual
channels, while a colon (":") creates a range of channels.
The following provides an explanation of how to setup and import
ASCII files into a logic analysis system.
Setting Up ASCII Files
NOTE:
If the analyzer is in state mode with the Clock Setup Mode set to
demultiplexer, slave pods will appear with an "S" in front of the pod
designation.
For example:
Label1;SA2[5] reads as Label1 maps to Slave Pod A2 Channel 5.
Individual channels
Label1;A2[5]
Label1 maps to Pod A2, Channel 5
Multiple channels
Label1;A1[5:2,0] Label1 maps to Pod A1, Channel 5 through Channel 2 and
Channel 0.
Individual channels on different pods
Label1;A2[1];A1
[0]
Label1 maps to Pod A2, Channel 1 and Pod A1, Channel 0.
Multiple channels on different pods
Label1;A3[15:5];
A2[5];A1[6]
Label1 maps to Pod A3, Channel 15 through Channel 5,
57
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
Pod A2 Channel 5, and Pod A1 Channel 6.
Clocks
Label1;CK[AK] Label1 maps to Slot A Clock K.
“Importing ASCII Files” on page 58
“Exporting ASCII Files” on page 58
See Also
“Termination Adapter” on page 60
“E5346A High Density Adapter” on page 61
Considerations when creating an Import file.
When updating a label ensure that the label is turned on. Labels that
are not active will not be updated.
Other considerations:
•
Maximum of 126 labels
•
Maximum of 20 characters per label. (The system truncates after 20
characters.)
•
Maximum of 32 channels per label.
Exporting ASCII Files
1. Select the Format tab.
2. Select File, then select Export Label...
Importing ASCII Files
Shown below is the default setup:
Label1;A1[15:0]
58
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Importing Netlist and ASCII Files
To Import an ASCII
file.
1. Create an ASCII file for importing into the logic analysis system. For
example:
Lab2;A2[15:10,6,3]
NewLabel2;A2[15]
Label1;A1[15:0]
2. Select the Format tab.
3. Select File, then select Import Labels...
4. Select the ASCII file you created.
5. Select OK.
59
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
Termination Adapter
The logic analyzer cable must have the proper RC network at its input
in order acquire data correctly. The Termination Adapter incorporates
the RC network into a convenient package. It also reduces the number
of pins required for the header on the target board from 40 pins to 20.
60
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Importing Netlist and ASCII Files
E5346A High Density Adapter
The E5346A high-density adapter provides a convenient and easy way
to connect an Agilent logic analyzer to the signals on your target
system for packages that are difficult to probe. An Amp "Mictor 38"
connector must be installed on your target system board.
61
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
Mapping Connector Names
1. Select the Format tab.
2. Select File, then select Import Netlist.
3. Select Next to go to the Mapping Connector Names dialog.
4. Enter a connector name from the Netlist.
5. Select the type of adapter.
6. Select the logic analyzer pods that are plugged into the adapters.
7. To map more connector names, select Add Connector and repeat the
steps above.
8. Select Next.
Import the Net List File
1. Select Select Netlist File.
62
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Importing Netlist and ASCII Files
2. Select the file from the File Selection dialog box.
3. Select OK
4. Select Next
Verify Net to Label Mapping
1. Verify that each net was properly imported into a label.
2. Select Next
The list provided, see example below, is the mapping from the logic
analyzer pods and channels to labels.
The label column is editable so you can make changes if necessary. If
you need to delete labels, this can be done more efficiently in the next
dialog box.
63
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
Select/Create Interface Labels
Select any additional labels to be copied into the Format tab. Typically
there is no need to add any more labels. However, this screen is useful
when you want to designate a signal bit in a bus as a separate label
name. It can also be used to delete unneeded labels.
Once you have completed editing the labels, select Done.
Activity Indicators
Activity indicators are the arrows and dashes associated with the pods.
They appear in various places, primarily above the column of bit
assignment fields in Format and in the Clock Setup area.
When the logic analyzer is properly connected to an active target
system, you see arrows in the Activity Indicator displays for each
64
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Importing Netlist and ASCII Files
channel. An active channel looks like
.
A dash at the top of the activity indicator display indicates that the
signal connected to that channel is electrically high (above the
threshold voltage). A dash at the bottom indicates that the signal is
electrically low. An arrow indicates that the signal is transitioning.
Activity indicators are not affected by label polarity. (see page 68)
You can use these indicators to check whether there is proper probe
connection: bits that are stuck high or low may not be properly
connected. You can also verify that the signals in your target system are
active: bits that are stuck high or low are not crossing the threshold
voltage.
Activity indicators are not displayed during an acquisition.
See Also
“Setting the Pod Threshold” on page 72
Logic Analysis System and Measurement Modules Installation Guide
for details on probing
Assigning Pods to the Analyzers
The Pod Assignment... button under Format can be used to start the
second analyzer on the module and to assign pod pairs.
To Assign Pods to an Analyzer
1. In the Format tab, select the Pod Assignment... button.
The Pod Assignment window appears.
2. Drag a pod pair and drop it below the analyzer that you want to assign it to.
3. For pods that you do not want to use, drag and drop them in the
Unassigned Pods area. This preserves memory depth when count is turned
on.
NOTE:
When both analyzers are turned on, pods 1/2 and 5/6 of the master card can
not be assigned to the same analyzer.
65
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
Pods can only be assigned on a per-pair basis to either of the two
analyzers. Each pod pair has two clock channels, but only the clock
channels of pods on the master card can be used in the analyzer's
clocking setup. These pods do not need to be assigned to that
particular analyzer, however.
To turn on an analyzer that is off, select the Off option button and
choose State or Timing. (Only one analyzer at a time can be set to
Timing.) A second analyzer window appears after a pause for setup.
Data On Clocks Display
The Data On Clocks display column, to the left of the pods' bit
reference line, is a display of all clock inputs available as data channels
in the present configuration. This includes those clocks on expander
cards, which cannot be used in the clock setup.
To use a clock channel as a data channel, assign the clock bit to a
label. Labels containing clock bits cannot be used in range terms
where the clock bits span more than one pod pair.
Activity indicators above the clock identifier show signal activity.
See Also
“Assigning Bits to a Label” on page 66
“Activity Indicators” on page 64
Assigning Bits to a Label
The bits in a label correspond to the physical logic analyzer probe
channels. When you run the analyzer, data is gathered on all bits
(channels) that are assigned to labels. Unassigned bits are inactive.
( * ) (asterisk) indicates an assigned bit.
( . ) (period) indicates an unassigned bit.
( R ) indicates an assigned bit in a reordered label.
66
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Importing Netlist and ASCII Files
To Assign Bits
1. Select the bit assignment label to the right of the label name you want to
define.
Each bit assignment label corresponds to the data pod which is listed
above it.
2. Either choose one of the predefined groups, or Individual.
3. In Individual, select the bits to toggle them between an asterisk and a
period.
NOTE:
Labels can have a maximum of 32 channels assigned to them.
Bits assigned to a label are numbered from right to left. The least
significant assigned bit on the far right is numbered 0. The next
assigned bit to the left is numbered 1, and so on. Labels can contain
bits that are not consecutive; however, bits are always numbered
consecutively within a label. Above each column of bit assignment
fields is a bit reference line that shows you the bit numbers and activity
indicators.
See Also
“To reorder bits in a label” on page 69
“Activity Indicators” on page 64
Inserting and Deleting Labels
To Insert Additional Labels
1. Select the label name button that you want to insert another label next to.
2. Choose Insert before... or Insert after....
To Delete Labels
1. Select the button of the label name that you want to delete.
2. Choose Delete.
The bit assignments of deleted labels are not saved. You can make a
label inactive but not lose its assignment by turning it off (see page 68)
67
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
instead.
Turning Labels On and Off
You may want to turn off labels that you have created so that the label
is not displayed. When a label is turned off, its name and bit
assignments are preserved.
To Turn Off a Label
1. Select the button of the label name that you want to turn off.
2. Choose Label [ON] to toggle it off.
If the label is the only one, it cannot be turned off or deleted. If there is
more than one label but it is the only one on, turning it off turns the first
label back on.
To Turn On a Label
1. Select the button of the label name that you want to turn on.
2. Choose Label [OFF] to toggle it on.
To Display a Label that was Off
1. Turn on the label.
2. At the bottom of the window, select the Apply button.
The label's data appears in the display windows.
Label Polarity
The analyzer uses the label polarity to identify patterns when
triggering and displaying data.
To change the label polarity, select the polarity field, which toggles
between positive (+) and negative (-). Positive polarity means that a
high voltage is a logical "1". Negative polarity means that a high voltage
is a logical "0".
Changing the label polarity after you have set up other parts of the
measurement changes these things:
•
68
"1" and "0" values flip in trigger terms
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Importing Netlist and ASCII Files
•
waveforms and bus values (where shown) invert in the Waveform Display
tool
•
"1" and "0" values flip in the Listing Display tool
Changing the label polarity does not change these things:
•
Edge definitions for clock setup and edge terms
•
Symbol definitions for the logic analyzer
The default polarity for all labels is positive (+). The various display
tools, in particular the Waveform Display tool, all show logical values
and are affected by polarity changes.
NOTE:
Negative logic is rare in circuits. The main exception at this time is RAMBUS.
To reorder bits in a label
In cases where buses in the device under test haven't been probed with
consecutive logic analyzer channels, you can reorder the bits in a label.
1. In the Format tab, select the label button whose bits you want to reorder.
2. Choose Reorder bits.
3. In the Change Bit Order dialog:
•
To reorder the bits individually, enter the bit that the probe channel
should be mapped to.
•
To swap the high and low order bytes or words, select the button Big
Endian to Little Endian at the bottom of the dialog.
•
To return to sequentially ordered bits, select the button Default Order
at the bottom of the dialog.
4. Select the OK button.
The label now shows an "R" to indicate that the assigned bit has been
reordered.
69
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
NOTE:
Labels with reordered bits cannot be used as range terms or <, <=,>, >= in
triggers.
Labels: Mapping Analyzer Channels to Your
Target
Labels group and identify related channels, such as buses, in a way
that is relevant to your system under test.
A single label can include data and clock channels from more than one
pod pair, but these cannot be included in range terms in the trigger
specification.
You can define 126 labels per analyzer. Each label can contain up to 32
channels per label.
To Define a Label
1. (Optional) Select the label name button and choose Rename
2. Enter a new name and select the OK button.
3. Assign bits (see page 66) to the label.
4. If necessary, insert more labels (see page 67) in the list.
See Also
“Assigning Bits to a Label” on page 66
“To reorder bits in a label” on page 69
“Inserting and Deleting Labels” on page 67
“Turning Labels On and Off” on page 68
“Label Polarity” on page 68
Setting Up the Pod Clock
There is one Pod Clock field for each pod in the machine. It only
70
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Importing Netlist and ASCII Files
becomes visible when the Clock Setup under the Sampling tab is set
to Master/Slave or Demultiplex. The Pod Clock field is located just
below the Pod Threshold in Format.
The Pod Clock field is set to Master Clk by default. Use the Pod Clock
field to indicate if the pod's data is to be strobed into memory by the
master clock, slave clock, or both, in the demultiplex mode.
When the Pod Clock is set to Demultiplex, only one pod of a pod pair
is usable. That pod latches data on both the master and slave clocks, so
it appears twice in the label area. To select which pod of a pod pair you
want to demultiplex, select the Pod Field.
See Also
“Performing Clock Setup (State only)” on page 41
“Clock Modes (State only)” on page 42 for details on slave and demultiplex
clocks
“Pod Selection” on page 71
Pod Selection
The Pod field in Format identifies which pod of a pod pair the settings
of the bit assignment field, pod threshold field, and pod clock fields
effect. Most of the time it is simply informational. The exceptions are
noted below.
Half-Channel Mode
In the half-channel mode, the Pod field becomes a button that you can
use to select which pod of a pod pair all pod settings apply to. In the
full-channel modes, this field is simply an identifier and is not
selectable.
Demultiplex Clock Mode
In the demultiplex clock mode, use the pod field to select which pod of
a pod pair is to be used to sample data. In the master or slave clock
modes, this field is simply an identifier and is not selectable.
See Also
“Setting the Acquisition Mode” on page 41
71
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
“Performing Clock Setup (State only)” on page 41
Setting the Pod Threshold
The pod threshold is a voltage level which the data must cross before
the analyzer recognizes it as a change in logic levels. You can specify a
threshold level for each pod. The level specified for the master pod that
includes the clock is also used for the clock threshold.
Standard
1. In the Format tab, select the threshold button located just below the pod
name.
2. In the Pod threshold dialog, choose one of the standard threshold options:
•
TTL -- The threshold level is +1.50 volts.
•
LVTTL -- The threshold level is +1.40 volts.
•
SSTL2 -- The threshold level is +1.50 volts.
•
SSTL3 -- The threshold level is +1.25 volts.
•
LVCMOS 1.5v -- The threshold level is +0.75 volts.
•
LVCMOS 1.8v -- The threshold level is +0.90 volts.
•
LVCMOS 2.5v -- The threshold level is +1.25 volts.
•
LVCMOS 3.3v -- The threshold level is +1.65 volts.
•
CMOS 5.0v -- The threshold level is +2.50 volts.
•
ECL -- The threshold level is -1.3 volts.
•
LVPECL -- The threshold level is 2.00 volts.
3. If you don't want the change to apply to all pods, deselect the checked box
next to Apply settings to all pods.
4. Select the Close button.
72
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Importing Netlist and ASCII Files
User Defined
When User Defined is selected, the threshold level is selectable from 6.0 volts to +6.0 volts.
NOTE:
The logic analyzer requires a minimum voltage swing of 500 mV at the probe
tip to recognize changes in logic levels.
NOTE:
The threshold voltage specified also applies to the pod's clock input.
State Clock Setup/Hold (State only)
Setup/Hold in Format adjusts the relative position of the clock edge
with respect to the time period that data is valid. It is only available
when the analyzer is set up for a state measurement.
To Change Clock Setup/Hold
1. Select the Setup/Hold button.
2. For each pod pair, choose a Setup/Hold selection from the selection list.
3. Select the OK button.
With a single clock edge assigned, the choices range from 3.5 ns Setup/
0.0 ns Hold, to 0.0 ns Setup/3.5 ns Hold. With both edges of a single
clock assigned, the choices are from 4.0 ns Setup/0.0 ns Hold, to 0.0 ns
Setup/4.0 ns Hold. If multiple clocks are assigned, the choices range
from 4.5 ns Setup/0.0 ns Hold, to 0.0 ns Setup/4.5 ns Hold.
The relationship of the clock signal and valid data under the default
setup and hold is shown in the figure below for a generic logic analyzer.
73
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Importing Netlist and ASCII Files
Default Setup and Hold
If the relationship of the clock signal and valid data is such that the
data is valid for 1 ns before the clock occurs and 3 ns after the clock
occurs, you will want to use the 1.0 setup and 2.5 hold setting.
Clock Position in Valid Data
74
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
The Trigger Tab
The Trigger tab is used to set up a sequence that tells the analyzer
when to capture data. The key event is the trigger.
The Trigger tab has two main areas:
•
On top, tabs of functions and controls to build your trigger
•
Beneath the tabs, the current trigger sequence.
Some controls are also located in the logic analyzer window's menu bar.
See Also
•
“Understanding Logic Analyzer Triggering” on page 76
•
“Setting Up a Trigger” on page 78
•
“Inserting and Deleting Sequence Steps” on page 79
•
“Editing Sequence Steps” on page 80
•
“Setting Up Loops and Jumps in the Trigger Sequence” on page 81
•
“Saving and Recalling Trigger Sequences” on page 82
•
“Clearing Part or All of the Trigger” on page 83
“Overview of the Trigger Sequence” on page 83
“Trigger Functions” on page 84
“Working with the User-level Function” on page 91
“Defining Resource Terms” on page 94
“Trigger Position Control” on page 53
“Sample Period (Timing Only)” on page 45
“Tagging Data with Time or State Tags (State Only)” on page 101
“Arming Control” on page 101
75
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
Understanding Logic Analyzer Triggering
What is a Trigger (see page 76)
What does "Trigger Position" Mean (see page 76)
What can be Used to Specify a Trigger (see page 77)
When to use a Combination, a Branch, or a Level (see page 77)
What is a Trigger
In simplest terms, a trigger is an event that tells the logic analyzer to
finish filling its acquisition memory. The memory functions like a
conveyor belt: new samples are always coming in, and old samples
"falling off" (being overwritten). The logic analyzer has room for 4K
samples. When this is full, the only way to fit in new data is to discard
the old.
After you specify your trigger sequence and press run, the logic
analyzer searches incoming data for events in the trigger sequence.
Using the conveyor belt metaphor again, it is like someone tending the
conveyor belt who has been told to stop the belt when a certain sample
is seen.
The trigger is not like an oscilloscope trigger. Logic analyzers trigger
only once per run, even when more than one sample matches the
trigger event. Logic analyzer trigger events are like special switches to
stop the evaluation process and just fill memory.
What does "Trigger Position" Mean
Because the logic analyzer is continually looking at data from your
target system after you select Run, and because the trigger is a single
event, you can arrange to collect data relative to it. It is like the person
running the conveyor belt is told to stop the belt when the special
sample reaches a certain position.
The default trigger position is in the middle. This means there are
76
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
approximately as many samples before the trigger as after. You can also
arrange for the trigger to be at the beginning of the "conveyor belt"
(acquisition memory), the end, or any percentage along it.
What can be Used to Specify a Trigger
The trigger sequence can be as simple as one event to look for, or a
complicated set of branching levels that loop back and jump around.
Both types of triggers use a small set of standard resources.
Pattern Pattern terms are things such as ADDR=0880 or R/W=1.
Generally they represent values on buses. You can set patterns to
match imprecise events, too. See “Using Bit Pattern Terms” on page 95.
Range Range terms match a range of values on a label or bus. For more
detail, see “Using Range Terms” on page 96
Timer Timers are started in one sequence level and checked in another.
They act like stopwatches. See “Using Timer Terms” on page 98 for
more information.
Edge Edge terms are similar to edges in oscilloscopes. They are only
available for some types of measurements. Edge terms can check for
edges on more than one signal at a time, but not all edges have to occur
at the same time. To require that, combine edge terms with ANDs.
Combinations of Terms To check for more than one type of thing
happening in the same sample, combine terms within a sequence
level using AND and OR. There are restrictions on exactly which terms
can be ANDed and which ORed. The restrictions are covered in “Using
Combinations of Terms” on page 99.
When to use a Combination, a Branch, or a Level
To check for simultaneous occurrences, use combinations of terms. All
the events described by the terms must happen in the same sample.
To take different actions depending on which events happen in a
sample, use branches within a sequence level. A branch functions like
a set of "if statements" in programming. The sample is checked against
77
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
all branches, and the first branch that matches is taken. See “Setting
Up Loops and Jumps in the Trigger Sequence” on page 81 for more on
branching.
To look for a sequence of events (for example, first look for a memory
reference on ADDR, then a certain value on DATA, and when IRQ goes
low, trigger) use different sequence levels. When a sample matches the
event described in a sequence level, the analyzer goes to the next
sequence level and compares the rest of the incoming events. When
the logic analyzer reaches the trigger level and finds a sample that
matches the trigger event you specify, the logic analyzer triggers.
Setting Up a Trigger
When setting up a trigger sequence, you typically trigger first on a
simple pattern or edge. From that point, you execute an iterative
process of adding or fine-tuning sequence steps until the analyzer
consistently triggers at the desired point.
To Set Up a Trigger
1. Define resource terms. (see page 94)
2. Select the most appropriate trigger function and select the Replace button.
(see page 78)
3. Select the sequence level number button and choose Edit. (see page 80)
4. If necessary, insert and edit additional sequence steps. (see page 79)
See Also
“Trigger Functions” on page 84
“Working with the User-level Function” on page 91
“Overview of the Trigger Sequence” on page 83
Selecting a Function to Match Trigger Conditions
Review the list of trigger functions and select the one that matches the
event you are looking for. In most cases one of the predefined functions
provides a good starting point. If none of the predefined trigger
78
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
functions match choose User level at the end of the list.
For a picture corresponding to the trigger function, select the function
from the list. The area to the right shows a picture of the function's
effect. The function itself is not inserted into the trigger sequence
unless you select the Replace or Insert button.
See Also
“Trigger Functions” on page 84
Inserting and Deleting Sequence Steps
NOTE:
For state measurements, the last level of the trigger sequence is always a
Store level. It cannot be deleted. For timing measurements, the last level
must contain the TRIGGER action.
To Insert Sequence Steps
1. Select the sequence level that you want to insert other steps around.
2. Select a trigger function from the list in the Trigger Functions tab.
3. Choose the Insert Before or Insert After button.
To Delete Sequence Steps
1. Select the sequence step that you want to delete.
2. Select the Delete button.
Trigger Sequence
Editing Options
When you select a sequence level, you see a selection menu with
choices that allow you to modify the sequence. Choose the option you
want from the choices below.
•
Edit
Changes the contents of the sequence step. You can change the resource
terms or other assignment fields, such as durations and occurrences.
•
Copy
Copies the currently selected step. When you copy a level, the new level
contains the same function as the original.
79
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
•
Replace
Replaces the currently selected level with the currently highlighted trigger
function.
•
Delete
Deletes the level that is currently selected.
•
Insert Before / Insert After
Inserts an additional step before or after the selected step.
•
Trigger Level (State only)
Makes the current step the trigger level.
•
Default (State only)
Only appears in the menu for the last step in a state trigger sequence. The
last state always stores to fill memory. Default returns the state to Store
anystate.
Editing Sequence Steps
You can modify the contents of a step in the trigger sequence by
editing it.
To Edit a Sequence Step:
1. Select the sequence level number button and choose Edit....
2. Select a term name to choose a different term.
You can also choose negations of terms and combinations of terms.
3. Set the values of the other fields, such as durations and occurrence counts.
4. Select the Close button to close the sequence step editing dialog.
To set the value of a term used in a sequence step, see “Defining
Resource Terms” on page 94.
If you are editing a user-level function, refer to:
“Setting Pattern Durations and Occurrence Counts” on page 93 (Timing
80
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
Only)
“Using Occurrence Counters” on page 93 (State Only)
“Using Timer Terms” on page 98
“Setting Up Loops and Jumps in the Trigger Sequence” on page 81
Setting Up Loops and Jumps in the Trigger
Sequence
To set up loops and jumps in your trigger sequence, use Branches.
Branches are available in most of the Trigger Functions. You may
need to break down the functions (see page 90) in order to set up
branches.
NOTE:
If either the < or the > durations are selected, only the primary Find or
Trigger on selection is available. If the occurs duration is selected, the
secondary (Else on) branch becomes available. If the Find field is an edge,
only occurs is available.
To Set Up a Branch or Loop
1. Select the Find button and choose the term that you want to branch on. If
this term is found in the incoming data, the analyzer will follow this branch
and go to the next trigger sequence level.
2. If you want a secondary branch, select the Else on button and choose the
term that you want to branch on.
3. Select the goto level option button and select the sequence level that you
want to branch to.
If the first branch is taken, the analyzer goes to the next level. If the
term in the first branch is not found, the analyzer immediately
evaluates the Else on secondary branching term. The analyzer only
triggers when the TRIGGER primary branch is taken.
If the Else on term is found, the secondary branch is taken to the
designated sequence level, and the occurrence counter is reset even if
the branch loops back to the same level. If the Else on term is not
81
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
found, the analyzer stays at the same sequence level until one of the
two branches is found. If both branches are found true at the same
time, the first branch is taken.
See Also
“Breaking Down and Restoring Functions” on page 90
“Branches Taken Stored / Not Stored (State only)” on page 94
Saving and Recalling Trigger Sequences
You can save a trigger sequence independently of configuration files
within a session by using Save/Recall. Recalling the stored trigger
sequence can change the trigger arming, memory depth, and trigger
position as well as the trigger sequence and term definitions. The
trigger sequence specification will not change the acquisition mode
(full channel vs. half channel).
The Agilent Technologies 16550A logic analyzer can hold up to 10
trigger sequence specifications per mode per machine, for a total of 40
specifications. When you exit your Agilent Technologies 16700A/B
session, the trigger sequences are cleared. They can be saved across
sessions or be shared across logic analyzers as part of a configuration
file, however.
To Save a Trigger Sequence
1. In the Trigger tab, select the Save/Recall subtab.
2. Enter a descriptive name for the trigger sequence in the Title field.
3. Select the Save button.
4. Select a memory location to store the trigger sequence in.
To Recall a Trigger Sequence
1. In the Trigger tab, select the Save/Recall subtab.
2. Select the Recall button.
3. Choose the trigger sequence that you want.
82
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
If one of the settings in the recalled trigger sequence conflicts with the
acquisition mode, it will be set to the closest setting for that mode.
Clearing Part or All of the Trigger
To Clear Part or All of the Trigger:
1. Select the Trigger tab.
2. Select Clear from the menu bar.
3. Choose the option you want from the choices described below:
•
All
Clears sequence steps, resource terms, resource term names, and
trigger position back to their default values. Turns on Count Time.
•
Sequence Levels
Resets the trigger sequence to the default sequence for the analyzer
acquisition mode.
•
Resource Terms
Resets all resource term assignment fields, including labels and values,
back to their default values. The default is usually for all terms to be on
the first label, with value XXXX.
•
Resource Term Names
Resets all the resource term names to their default values, Pattern1
through Pattern9 and Patt10.
•
Save/Recall Memories
Deletes all saved trigger sequence specifications.
Overview of the Trigger Sequence
The trigger sequence is a sequence of steps that control the path that
the analyzer takes to find the trigger event. The path taken resembles a
83
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
flow chart, with each step in the sequence being an opportunity to
direct the analyzer's selection. You can edit the overall trigger
sequence by inserting or deleting sequence steps (see page 79).
Each step in the sequence is either a predefined trigger function (see
page 84) or a user-level step (see page 91). Both the trigger functions
and user-level steps contain variables that you define. The variables,
called resource terms (see page 94), represent edges and bit patterns
in the data.
When you run the analyzer, it searches for a match between the
resource term values and the measurement data. When a match is
found, the sequencing continues to the next step, loops back (see
page 81) to a previous step, or jumps ahead to another step.
Eventually, a path of "true" steps leads to the trigger event.
Each macro uses one or more of the analyzer's internal sequence levels
(see page 91). Each user-level step uses one internal sequence level.
See Also
“Understanding Logic Analyzer Triggering” on page 76 for more detail
“Setting Up a Trigger” on page 78 for actual steps
Trigger Functions
Trigger functions provide a simple way to set up the analyzer to trigger
on common events and conditions. A library of functions is available for
both state and timing measurements.
NOTE:
Each trigger function requires at least one internal sequence level (see
page 91), and in some cases, multiple levels. The number of levels used by
each function is described in the references below.
Timing Trigger
Functions
“Basic Timing Trigger Functions” on page 85
“Pattern/Edge Combinations” on page 86
“Time Violations” on page 87
“Timing User Level” on page 85
84
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
State Trigger
Functions
“Basic State Trigger Functions” on page 88
“Sequence-Dependent Trigger Functions” on page 88
“Time Violations” on page 89
“State User-level” on page 87
See Also
“Setting Up a Trigger” on page 78
“Defining Resource Terms” on page 94
“Breaking Down and Restoring Functions” on page 90
Timing User Level
The User level trigger function allows you to create a custom trigger
sequence using terms, a comparison function, and a secondary branch
if the comparison function is occurs. This trigger function uses one
internal sequence level.
See Also
“Working with the User-level Function” on page 91 for more information on
the user-defined mode.
Basic Timing Trigger Functions
The following basic trigger functions are found in Trigger Functions
when the analyzer is in timing mode. Each function uses one internal
sequence level.
•
Find edge.
This function becomes true when the edge you have designated is seen. It
uses one internal sequence level.
•
Find anystate n times.
This function becomes true with the nth state it sees. It uses one internal
sequence level. It is equivalent to having the analyzer wait in the sequence
level for (n x Sample Period) seconds.
85
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
•
Find nth occurrence of an edge.
This function becomes true when it finds the designated occurrence of an
edge you have designated. Note that the 500-MHz trigger sequencer may
not count edges that occur closer than 2 ns. This function uses one
internal sequence level.
•
Find pattern present/absent for > duration.
This function becomes true when it finds a pattern you have designated
that has been present or absent for greater than or equal to the set
duration. It uses one internal sequence level.
•
Find pattern present/absent for < duration.
This function becomes true when it finds a pattern you have designated
that has been present or absent for less than the set duration. It uses five
internal sequence levels.
Pattern/Edge Combinations
The following trigger functions are found in Trigger Function when
the analyzer is in timing mode. These predefined functions use a
pattern, edge, or a combination of both as the trigger element. The
functions use either one or two internal sequence levels.
•
Find edge AND pattern.
This function becomes true when a selected edge is seen within the time
window defined by a pattern you have designated. It uses one internal
sequence level.
•
Find pattern occurring too soon after edge.
This function becomes true when a pattern you have designated is seen
occurring within a set duration after a selected edge is seen. It uses two
internal sequence levels.
86
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
•
Find pattern occurring too late after edge.
This function becomes true when one edge you have selected occurs, and
for a designated period after that first edge is seen, a pattern is not seen. It
uses two internal sequence levels.
Time Violations
The following trigger functions are found in Trigger Functions when
the analyzer is in timing mode. These trigger functions are specifically
tailored to trigger on events occurring out of a predefined time range.
They use either one or two internal sequence levels.
•
Find width violation on a pattern/pulse.
This function becomes true when the width of a pattern violates minimum
and maximum width settings you have designated. It uses one internal
sequence level.
•
Find 2 edges too close together.
This function becomes true when a second selected edge is seen occurring
within a period you have designated after the occurrence of a first selected
edge. It uses two internal sequence levels.
•
Find 2 edges too far apart.
This function becomes true when a second selected edge occurs beyond a
period you have designated after the first selected edge. It uses two
internal sequence levels.
•
Wait t seconds
This function becomes true after a period you have designated has
expired. It uses one internal sequence level.
State User-level
The User-level trigger function allows you to create a custom trigger
sequence using terms, a comparison function, and a jump or loop. This
87
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
trigger function uses one internal sequence level.
See Also
“Working with the User-level Function” on page 91 for more information on
the user-defined mode.
Basic State Trigger Functions
The following basic trigger functions are found in Trigger Functions
when the analyzer is in state mode. Each macro uses one internal
sequence level.
•
Find Pattern n times.
This function becomes true when it sees a pattern you have designated
occurring a designated number of times. The pattern may occur
consecutively, but does not have to. It uses one internal sequence level.
•
Find anystate n times.
This function becomes true with the nth state it sees. It uses one internal
sequence level. It is equivalent to Wait n external clock states.
•
Find pattern2 occurring immediately after pattern1.
This function becomes true when the first pattern you have designated is
seen immediately followed by a second designated pattern. It uses two
internal sequence levels.
•
Find pattern n consecutive times.
This function becomes true when it sees a pattern you have designated
occurring a designated number of consecutive times. It uses one internal
sequence level.
Sequence-Dependent Trigger Functions
The following trigger functions are found in Trigger Functions when
the analyzer is in state mode. These functions each trigger on a
particular sequence of events.
88
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
•
Find too few states between pattern1 and pattern2.
This function becomes true when a designated pattern1 is seen, followed
by a designated pattern2, and with fewer than a selected number of states
occurring between the two patterns. It uses four internal sequence levels.
•
Find too many states between pattern1 and pattern2.
This function becomes true when a designated pattern1 is seen, followed
by more than a selected number of states, before a designated pattern2. It
uses two internal sequence levels.
•
Find n-bit serial pattern.
This function becomes true when a specified serial pattern of n bits is
found on the analyzed line. This function uses one internal sequence level
for each bit specified in the trigger sequence.
•
Find pattern2 n times after pattern1, before pattern3 occurs.
This function becomes true when it first finds a designated pattern1,
followed by a selected number of occurrences of a designated pattern2. In
addition, if a designated pattern3 is seen anytime while the sequence is not
yet true, the sequence starts over. This includes if pattern2's nth
occurrence is at the same time as pattern3, the sequence starts over. It
uses two internal sequence levels.
Time Violations
The following trigger functions are found in Trigger Functions when
the analyzer is in state mode. These predefined functions are
specifically tailored to trigger on events occurring out of a predefined
time range. These functions use either one or two internal sequence
levels.
•
Find pattern2 occurring too soon after pattern1.
89
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
This function becomes true when a designated pattern1 is seen, followed
by a designated pattern2, and with less than a selected period occurring
between the two patterns. It uses two internal sequence levels.
•
Find pattern2 occurring too late after pattern1.
This function becomes true when a designated pattern1 is seen, followed
by at least a selected period, before a designated pattern2 occurs. It uses
two internal sequence levels.
•
Wait n external clock states.
This function becomes true after a number of user clock states you have
designated have occurred. It uses one internal sequence level.
Breaking Down and Restoring Functions
When you break down a trigger function, you gain access to all the
resource assignment fields and branching options. You can change
these fields to change the structure of the trigger sequence. You might
need to do this to create a custom trigger sequence or to add jumps.
To Break Down Trigger Functions
1. Select Modify in the Trigger window menu bar.
2. Choose Break down functions. The trigger sequence area changes to
show the entire trigger sequence as a series of user-level steps.
The contents of broken down functions are displayed in the long form
used in a user-level sequence step. If the function uses two of the
analyzer's internal sequence levels, (see page 91) both levels are
separated out and displayed in the trigger sequence area.
To Restore Functions
1. Select Modify in the Trigger window menu bar.
2. Choose Restore functions.
Use Restore functions to restore all functions to their original
90
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
structure. Note that when the functions are restored, all changes are
lost and any branching that is part of the original structure is restored.
See Also
“Working with the User-level Function” on page 91 for information on
working with functions that are broken down.
How the Internal Sequence Levels Are Used
The analyzer has internal sequence levels that it uses to make up the
trigger sequence. There are a total of 16 sequence levels available.
The actual number of levels used in a trigger sequence can vary
depending on whether you elect to use predefined trigger functions or
use the user-defined steps (see page 91) to construct a more custom
trigger sequence.
When you use user-defined steps, all of the internal sequence levels are
available. Each user-defined sequence level corresponds to one
internal level. The only instance where multiple levels are used is when
the < duration is assigned.
When you use predefined trigger functions (see page 84), more than
one of the internal sequence levels may be required for a single trigger
function. Even though some trigger functions use multiple sequence
levels, trigger functions are easier to use, and they are the most
efficient way to construct a trigger specification.
Working with the User-level Function
NOTE:
Before you begin to set up user-level sequence steps, note that in most cases
one of the predefined trigger functions (see page 84) will work.
You might need to set up a user-level sequence step to accommodate a
condition not covered by the functions, or if you need to set up
additional loops and jumps in the sequence. Each user-level sequence
step has a "fill-in-the-blanks" type statement. You use resource terms
to fill in the statement with the appropriate values.
91
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
To Set Up a User-Defined Macro
1. In the Trigger Functions tab, select User-level at the end of the list.
2. Select the Replace or Insert button.
3. Select the new sequence level number button; choose Edit.
A Trigger Sequence Step window appears.
4. For state measurements, select the While storing button; choose the
resource term.
The values represented by this term will be stored in memory while the
logic analyzer is evaluating target system data against this step. The
default, anystate, stores everything. Nostate stores nothing.
5. Select the Trigger on button or the Find button; choose the resource
term.
6. In timing measurements, you have the choice of an occurrence counter or
duration. To use the secondary branch, you must set this to occurs.
7. Set the occurrence field or duration.
8. If you want to use a loop or jump:
a. Select the Else on button; select resource term.
b. Select the goto level option button; choose a sequence level.
9. If you want to use a timer:
a. In the appropriate sequence level, start the timer by selecting the Off
option button and choosing Start. Timers do not start automatically.
b. In the Timer tab, assign a time value you want to check against.
c. To check the value, assign the timer to a resource field either on its
own, or as part of a combination.
For more information on the functions available in a user-defined step,
refer to:
•
“Defining Resource Terms” on page 94
•
“Setting Pattern Durations and Occurrence Counts” on page 93 (Timing
Only)
•
“Using Occurrence Counters” on page 93 (State Only)
92
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
•
“Setting Up Loops and Jumps in the Trigger Sequence” on page 81
•
“Using Timer Terms” on page 98
Setting Pattern Durations and Occurrence Counts
(Timing Only)
When a bit pattern is found during a trigger sequence, you can
influence when the term actually becomes "true" by assigning a time
duration or an occurrence count.
The (>/</occurs) control may not be directly accessible if you are
editing a trigger function. To reveal it, break down (see page 90) the
function and then follow these instructions.
To Set a Pattern Duration
1. Select the (>/</occurs) option button and choose an option.
2. Set the duration or the number of occurrences.
When a greater-than or less-than duration is assigned, the secondary
branch (Else on) is not available.
When greater-than (>) is used, the analyzer continues sequence level
evaluation only after the resource term has been true for greater than
or equal to the amount of duration specified.
When less-than (<) is used, the analyzer continues sequence level
evaluation only after the resource term has been true for less than or
equal to the amount of duration specified. For each less-than
assignment, four internal sequence levels (see page 91) are used.
When occurs is selected, you can set an occurrence count. Use the
occurrence counter to delay the trigger sequence evaluation until a
resource term has occurred in a designated number of samples. These
samples may be interrupted by other values -- they do not need to be
consecutive.
Using Occurrence Counters
Use the occurrence counter to delay the trigger sequence evaluation
until a resource term has occurred in a designated number of samples.
93
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
The samples do not need to be consecutive. Whatever positive number
you assign to the counter, the pattern must be seen that number of
times before the term becomes true.
If the Else branch becomes true before all specified occurrences of the
primary (Trigger on, or Find) branch, the Else branch is taken.
Branches Taken Stored / Not Stored (State only). Branches
Taken sets the analyzer to store, or not to store, the resource terms of
the branch that the analyzer followed in the trigger sequence. The
Branches Taken control is located in Trigger, under the Settings tab.
Use Branches Taken if you want to maximize memory usage but have
a complex trigger. With Branches Taken set to stored, you can use
store qualifiers to not store the preliminary data ("While storing no
state") but still reconstruct the events leading up to your trigger.
When the analyzer is set to Branches Taken Stored, all branch events
(Find, Else On, or Trigger) are stored when they occur.
When the analyzer is set to Branches Taken Not Stored, the branch
events will only be stored if the states they represent are included in
the While Storing qualifier.
The While storing field specifies what is being stored in the analyzer's
memory before the trigger conditions are met. To set this, select the
While storing button and choose a pattern or range term for the store
qualification.
Defining Resource Terms
Resource terms are variables that you can use in defining a trigger
sequence. The terms available include patterns, edges, ranges, and
timers.
When the Agilent Technologies 16550A is configured as a state
analyzer, up to fourteen resource terms are available: 10 pattern
terms Pattern1 - Pattern9 and Patt10, two range terms, and two
timer terms. When configured as a timing analyzer, 16 resource
terms are available from the 14 terms already mentioned, plus two
94
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
edge terms.
When the module is set up as two analyzers, each resource term can
be used by either of the two analyzers, but not both at the same time.
To Define a Resource Term
1. In the Trigger tab, select the appropriate subtab to bring the group of
terms forward.
2. Optional - Highlight the term name and enter a new name.
For Timer terms, set the time value and skip the remaining steps.
3. Optional - Select the label button to the right of the term name and choose
Replace.
4. Select the label that you want to use.
5. Optional - add more labels (see page 100) to the term.
6. Optional - Set the numeric base.
7. Select the term value field, to the right of the label name.
8. For pattern and range terms, enter the term value. For edge terms, assign
edges or glitches to appropriate bits.
See Also
“Using Bit Pattern Terms” on page 95
“Using Edge Terms” on page 97
“Using Range Terms” on page 96
“Using Combinations of Terms” on page 99
“Using Timer Terms” on page 98
“Adding and Deleting Labels for Terms” on page 100
“Numeric Base” on page 100
Using Bit Pattern Terms
Bit pattern resource terms can be set to match a numeric value or bit
pattern on a group of data channels such as a bus. In order for a pattern
to be found by the analyzer, the input data must match all bits of the
pattern that are not defined as Don't Cares (X). Patterns can also be
95
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
used in a negated form.
To Define a Pattern Term
1. Select the label name button to the right of the term name and choose the
label that you want to use.
2. If necessary, add more labels (see page 100) to the term.
3. Select the term value field, to the right of the label name.
4. Enter the pattern for each label.
Depending on the base setting, such as hex or octal, some characters will
not be accepted. Don't cares are indicated by an X.
Right-click on any of the bit pattern value fields to quickly assign the
pattern term to a preset value. Clear (=X) sets the value to all X (don't
cares). Set (=1) sets the value to all 1s. Reset (=0) sets the value to all
0s.
Using Range Terms
Range terms bracket groups of data values between upper and lower
boundaries that you assign. The range term becomes true when the
data is numerically between or on the two specified boundaries.
The labels used by a range must be contained in a single pod pair, with
no clock channels or re-ordered bits allowed.
To Define a Range Term
1. Select the label name button to the right of the term name and choose the
label that you want to use.
Only one label can be used at a time for range terms.
2. Select the lower value field, to the right of the label name.
3. Enter the pattern for the low value of the range.
4. Select the upper value field, to the right of the label name.
96
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
5. Enter the pattern for the upper value of the range.
Depending on the base setting, such as Hex or Octal, some characters will
not be accepted.
Right-click on either of the value fields to quickly assign the value to a
preset value. Set (=1) sets the value to all 1s. Reset (=0) sets the value
to all 0s.
Using Edge Terms
Edge terms are only available in timing mode. You can set an edge
term to match transitions or glitches on one or more channels. When
you specify an edge or glitch on more than one channel, the analyzer
logically ORs the edges together. When the analyzer sees a transition
that matches any of the ones specified in the edge term, the term
becomes true. If you must have both edges occur in the same sample,
assign the edges to different terms and combine them (see page 99)
with an AND.
The following edge choices are available for each bit:
Rising edge (↑)
Falling edge (↓)
Either rising or falling
Glitch (*)
To Define an Edge Term
1. Select the label name button and choose the label that you want to use.
2. Select the edge value button, to the right of the label name.
97
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
3. Set the edge or glitch for each channel.
Don't cares are indicated by a (.).
NOTE:
After you close the edge term assignment dialog, you may see $ indicators in
the term value field. This symbol indicates that the value can't be displayed in
the selected base. Choose Binary base to see the actual assignments.
Using Timer Terms
Timers are like other resource terms in that they are either true or
false. They are unlike other terms, however, in that they are controlled
within the trigger sequence and act like a stopwatch. When a timer
reaches its assigned count (expires), it becomes true. When a timer
expires or stops, its count resets to zero.
Timers can be set to Start, Stop, Pause, or Continue as the analyzer
enters a trigger sequence level. The two timers are global, so each
sequence level except the first has the ability to control the same timer.
As more sequence levels are added, the timer status in the new levels
defaults to Off. If a timer is paused in one level, it must be continued in
another level before it will be restarted. If the trigger sequence returns
to the same sequence level again, the timer will be restarted.
Each Timer term can be used by either of the two analyzers, but not
both.
To Assign a Time Value to the Timer Terms
1. Select the Timer subtab in the Trigger tab.
2. Select the timer value field, next to the timer name.
3. Enter a timer value.
4. Use the up/down arrow buttons to scroll through timer values.
The minimum value a timer can have is 400 ns, which is also the default
value.
To Include a Timer in a Trigger Sequence
1. Assign a time value to the timer you want to use.
2. Select the sequence level you want to use it in, and choose Edit....
98
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
Timer control is not available in the first sequence level.
3. Select the Off option button at the bottom of the edit dialog box and
choose Start.
If the other analyzer is already using a timer, it won't appear in this
analyzer's edit dialog.
4. Select the event field where you want to use the timer.
5. Choose either Timer or Combo....
Use Combo for cases such as an edge occurring after so many
nanoseconds.
6. In the other trigger levels, you can Start, Pause, and Continue the timer.
When a timer reaches its assigned value, it automatically stops.
See Also
“Using Combinations of Terms” on page 99
Using Combinations of Terms
Resource terms can be used in combinations. Combinations use the
logical AND, NAND, OR, NOR, and XOR functions to combine
predefined resource terms.
A combined term is evaluated as a single events. Components that are
ANDed together must all occur in the same sample. Ones that are
ORed together only require one of the components to happen.
If you intend for the logic analyzer to decide between two actions, use
branches instead of combined terms. If you want the logic analyzer to
find the terms in sequence, use different levels or a trigger function in
the trigger sequence.
To Set Up a Combination
1. Edit the sequence step (see page 80) that you want a combination term to
appear in.
2. Select the resource term assignment field that you want to make into a
combination.
3. Choose Combo... from the menu of resource terms.
4. In the Combination dialog, select the On/Off/Not option button for each
term that you want to use and select On or Not.
99
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
Not selects the logical negation of the term.
5. Select the logical operator buttons and choose a logic operation.
Not all operations are available at all levels of the combination tree.
NOTE:
The combination of terms is now inserted into the term assignment field. If the
term is too long to fit, the display is truncated.
Adding and Deleting Labels for Terms
Labels are defined in Format, after which they are available throughout
the other analyzer areas and attached display tools.
When you use more than one label to define a trigger term, the term
conditions must be true on both labels for the term to be true.
To Add a Label to a Resource Term
1. Select the label name button.
2. Choose Insert.
3. Select the label that you want to add to the resource term.
To Delete a Label from a Resource Term
1. Select the label name button.
2. Choose Delete.
Numeric Base
All labels have a numeric base field next to them. The base choices are
Binary, Octal, Decimal, Hex, ASCII, Symbol and Twos Complement.
To Change the Numeric Base
1. Select the the base option button.
2. Choose the base that you want.
NOTE:
If the numeric base is changed in one window, the base in other windows may
not change accordingly. For example, the base assigned to symbols is unique,
as is the base assigned in the Listing window.
100
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Trigger Tab
Tagging Data with Time or State Tags (State
Only)
The Count field under Settings in Trigger accesses a selection menu
which is used to stamp the data at each memory location with either a
Time tag or a State Count tag. The tags reduce the memory depth by
half. To retain the full memory depth when using time or state tags
leave one pod pair unassigned.
State Count
When the State Count option is selected, numbered tags are placed on
all selected data. Pre-trigger data has negative numbers and posttrigger data has positive numbers. You select the data to be tagged
when you turn on State Count. A field appears to the right of States
that lets you define patterns.
State tag numbering can be set either relative to the previous tagged
sample or absolute from the trigger point. Selecting Absolute or
Relative is done by toggling the Absolute/Relative field in the Listing
Display window.
Time Count
Time Count places time tags on all data. Pre-trigger data has negative
time numbers and post-trigger data has positive time numbers. Time
tag numbering is set to be either relative to the previous memory
location or absolute from the trigger point. The time tag resolution is 4
ns.
Arming Control
An instrument must be armed before it can look for its trigger. When
you Run an instrument, it is armed immediately. When using logic
analyzer modules that provide two separate analyzers, you can set one
analyzer to arm the other within the same module by selecting Arming
Control... under Settings in Trigger.
101
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Trigger Tab
Selecting the Arming Control... button brings up the Machine
Arming Tree dialog. In this dialog you can set what starts each
analyzer, and which one sends a signal to Arm Out. Arm Out is used to
send signals to other instruments in the frame, or sent to Port Out.
Port Out can be used to control additional instruments external to the
logic analysis system.
To change the source of Arm In or the destination of Port Out, use the
Intermodule dialog (see the Agilent Technologies 16700A/B-Series
Logic Analysis System help volume).
Setting One Analyzer to Arm the Other
This procedure assumes you have both analyzers turned on. The
second analyzer can be turned on by dragging it onto the workspace in
the Workspace window.
1. Under the Trigger tab, select the Settings subtab.
2. Select the Arming Control... button.
3. Select the box of the machine that you want to have wait for its arm signal.
4. In the armed by menu, select the other machine.
The arming signal path forms a tree in the Machine Arming Tree dialog.
5. Set the sequence level number to the trigger sequence level that will wait
for the arm signal. It does not have to be the same as the trigger level.
6. Select the Close button.
NOTE:
If the trigger sequence does not pass through the level containing the wait for
arm term, the trigger will not wait for the arming signal.
See Also
Overview - Multiple Instrument Configuration (see the Agilent
Technologies 16700A/B-Series Logic Analysis System help volume)
Overview - Multiple Machine Configuration (see the Agilent Technologies
16700A/B-Series Logic Analysis System help volume)
102
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Specifications and Characteristics
Specifications and Characteristics
NOTE:
Definition of Terms
To understand the difference between specifications (see page 103) and
characteristics (see page 103), and what gets a calibration procedure (see
page 104) and what gets a function test (see page 104), refer to appropriate
links within this note.
“Agilent Technologies 16550A Logic Analyzer Specifications” on page 104
“Agilent Technologies 16550A Logic Analyzer Characteristics” on page 105
What is a Specification
A Specification is a numeric value, or range of values, that bounds the
performance of a product parameter. The product warranty covers the
performance of parameters described by specifications. Products
shipped from the factory meet all specifications. Additionally, the
products sent to Agilent Technologies Customer Service Centers for
calibration and returned to the customer meet all specifications.
Specifications are verified by Calibration Procedures.
What is a Characteristic
Characteristics describe product performance that is useful in the
application of the product, but that is not covered by the product
warranty. Characteristics describe performance that is typical of the
majority of a given product, but not subject to the same rigor
associated with specifications.
Characteristics are verified by Function Tests.
103
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Specifications and Characteristics
What is a Calibration Procedure
Calibration procedures verify that products or systems operate within
the specifications. Parameters covered by specifications have a
corresponding calibration procedure. Calibration procedures include
both performance tests and system verification procedure. Calibration
procedures are traceable and must specify adequate calibration
standards.
Calibration procedures verify products meet the specifications by
comparing measured parameters against a pass-fail limit. The pass-fail
limit is the specification less any required guardband.
The term "calibration" refers to the process of measuring parameters
and referencing the measurement to a calibration standard rather than
the process of adjusting products for optimal performance, which is
referred to as an "operational accuracy calibration".
What is a Function Test
Function tests are quick tests designed to verify basic operation of a
product. Function tests include operator's checks and operation
verification procedures. An operator's check is normally a fast test
used to verify basic operation of a product. An operation verification
procedure verifies some, but not all, specifications, and often at a lower
confidence level than a calibration procedure.
Agilent Technologies 16550A Logic Analyzer
Specifications
The specifications are the performance standards against which the
product is tested. These specifications apply only to the Agilent
Technologies 16550A 100 MHz State/500 MHz Timing logic analyzer:
Maximum State Speed:
Minimum State Clock Pulse Width:*
104
100 MHz
3.5 ns
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Specifications and Characteristics
Minimum Glitch Width:
3.5 ns
Threshold Accuracy:
+/- (100mV
Minimum Master-to-Master Clock Time:* 10.0 ns
Setup/Hold Time:*
Single Clock, Single Edge:
0.0/3.5 ns
adjustable
Single Clock, Multiple Edge:
0.0/4.0 ns
adjustable
Multiple Clock, Multiple Edge:
0.0/4.5 ns
adjustable
+ 3% of threshold setting)
through 3.5/0.0 ns,
in 500-ps increments
through 4.0/0.0 ns,
in 500-ps increments
through 4.5/0.0 ns,
in 500-ps increments
* Specified for an input signal VH=-0.9 V, VL=-1.7 V, threshold=-1.3 V,
slew rate=1 V/ns
Agilent Technologies 16550A Logic Analyzer
Characteristics
The characteristics are not specifications, but are included as
additional information.
General information
- Channel Counts:
1-card module
2-card module
- Memory Depth:
Half Channel
Full Channel
96 data, 6 clock
198 data, 6 clock
4 Ksamples
8 Ksamples
Probes
- Input Resistance:
- Parasitic tip capacitance:
- Minimum Voltage Swing:
- Maximum Voltage:
- Threshold Range:
100
1.5
500
+/+/-
Kohm, +/- 2%
pF
mV peak-to-peak
40 V peak, CAT I
6.0 V, adjustable in 50-mV increments
State Analysis
- State Clocks:
- State Clock Qualifiers:
- Time Tag Resolution:
- Maximum Time Count Between States:
- Maximum State Tag Count:
6
4
Each qualifier can be set to
recognize one of six clock
lines, either high or low.
8 ns
34 seconds
4.29e9
Timing Analysis
- Maximum Conventional Timing Rate:
Half Channel
500 MHz
Full Channel
250 MHz
- Maximum Transitional Timing Rate:
Half Channel
250 MHz
Full Channel
125 MHz
- Maximum Timing with Glitch Rate:
Half Channel
125 MHz
Full Channel
N/A
- Sample Period Accuracy:
0.01% of sample period
- Channel-to-Channel Skew:
2 ns, typical
- Time Interval Accuracy:
+/-( sample period + channel-tochannel skew + 0.01% of time
interval reading )
105
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Specifications and Characteristics
Triggering
- Sequencer Speed:
- State Sequence Levels:
- Timing Sequence Levels:
- Maximum Occurrence Count Value:
- Pattern Recognizers:
- Range Recognizers:
- Range Width:
- Timers:
- Timer Value Range:
- Glitch/Edge Recognizers:
- Minimum Detectable Glitch:
125 MHz, maximum
12
10
1,048,575
10
2
32 bits each
2
400 ns to 500 seconds
2 (timing only)
3.5 ns
Power Requirements
All necessary power is supplied through the backplane connector
of the logic analysis system mainframe.
Operating Environment Characteristics
- Indoor use only.
- Temperature
Instrument (except disk and media): 0 to 55 degrees C (+32
to 131 degrees F)
Probe lead sets and cables:
0 to 65 degrees C (+32 to
149 degrees F)
- Humidity
Instrument, probe lead sets, and cables: up to 95% relative humidity at
40 degrees C (+104 degrees F)
- Altitude
To 4600 m (15,000 ft)
- Vibration
Operating:
Random vibration 5 to 500 Hz, 10 minutes per axis,
approximately 0.3 g (rms)
Nonoperating: Random vibration 5 to 500 Hz, 10 minutes per axis,
approximately 2.41 g (rms); and swept sine resonant
search, 5 to 500 Hz, 0.75 g (0-peak), 5-minute
resonant dwell at 4 resonances per axis.
106
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Analyzer Probing Overview
Analyzer Probing Overview
The figures below shows a variety of simple probing connections. The
specific probe type, number of probes, and location on the target
circuit depends on your particular measurement.
For equivalent circuit diagrams and pinouts, see the description of the
probe type in the Logic Analysis System and Measurement Modules
Installation Guide. If you have misplaced the Logic Analysis System
and Measurement Modules Installation Guide, you can download
the latest version from the Web at <URL: http://www.agilent.com/find/
LogicAnalyzer-Manuals/ >
Probe Lead-to-Board Connection
The standard lead set plugs directly into any .1-inch grid with 0.026 to
0.033-inch diameter round pins or 0.025-inch square pins. All probe
tips work with the Agilent Technologies 5059-4356 surface mount
grabbers and the Agilent Technologies 5959-0288 through-hole
grabbers.
107
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Analyzer Probing Overview
Adapter-to-Board Connection
Both the 01650-63203 and the E5346A adapters include termination
for the logic analyzer. The 01650-63203 termination adapter plugs into
a 2 x 10 pin header with 0.1 inch spacing. The E5346A high-density
adapter connects to an AMP "Mictor 38" connector. If possible, use
support shrouds around the Mictor connector to relieve strain and
improve connections.
Direct Pod-to-Board Connection
If you provide proper termination as part of the target board, you can
plug the pod directly into the ©3M 2520-series, or similar alternative
connector. Suggested termination is shown in the Logic Analysis
System and Measurement Modules Installation Guide.
Also use this termination with the Agilent Technologies E5351A highdensity, non-terminated adapter.
Pod-to-Analysis Probe Connection
Analysis probes (formerly called preprocessors) are microprocessorspecific interfaces that make it easier to probe buses. Generally,
108
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Analyzer Probing Overview
analysis probes consist of a circuit board that attaches to the
microprocessor (possibly through an adapter) and a configuration file.
The configuration file sets up the logic analyzer's clocks and labels
correctly, and may include an inverse assembler. The circuit board
provides access to logical groups of pins through headers designed to
connect directly to the logic analyzer.
The easiest way to set up a measurement with an analysis probe is the
Setup Assistant. (see the Setup Assistant help volume) The Setup
Assistant asks you questions about your measurement and then shows
you just the information you need to set up the probe correctly. It also
loads the proper configuration files.
109
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Using Symbols
Using Symbols
You can use symbol names in place of data values when:
•
Setting up triggers
•
Displaying captured data
•
Searching for patterns in Listing displays
•
Setting up pattern filters
•
Setting up ranges in the System Performance Analyzer
Symbol names can be: variable names, procedure names, function
names, source file line numbers, etc.
You can load symbol name definitions into the logic analyzer from a
program's object file or from a general-purpose ASCII format symbol
file, or you can define symbol names in the logic analyzer.
•
“To load object file symbols” on page 111
•
“To adjust symbol values for relocated code” on page 112
•
“To create user-defined symbols” on page 113
•
“To enter symbolic label values” on page 114
•
“To create an ASCII symbol file” on page 115
•
“To create a readers.ini file” on page 116
See Also
To go to a pattern in the Listing (see the Listing Display Tool help
volume)
To modify the Source Viewer trace setup (see the Listing Display Tool
help volume)
To define System Performance Analyzer state interval ranges (see the
System Performance Analyzer help volume)
110
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Using Symbols
To load object file symbols
Object files are created by your compiler/linker or other software
development tools.
1. Generate an object file with symbolic information using your software
development tools.
2. If your language tools cannot generate object file formats that are
supported by the logic analyzer, create an ASCII symbol file (see
page 115).
3. Select the Symbol tab and then the Object File tab.
4. Select the label name you want to load object file symbols for.
In most cases you will select the label representing the address bus of the
processor you are analyzing.
5. Specify the directory to contain the symbol database file (.ns) in the field
under, Create Symbol File (.ns) in This Directory. Select Browse... if
you wish to find an existing directory name.
6. In the Load This Object/Symbol File For Label field, enter the object file
name containing the symbols. Select Browse... to find the object file and
select Load in the Browser dialog.
If your logic analyzer is NFS mounted to a network, you can select object
files from other servers.
7. If your program relocates code, see “To adjust symbol values for relocated
code” on page 112.
The name of the current object file is saved when a configuration file is
saved. The object file will be reloaded when the configuration is loaded.
To reload object file symbols
1. Select the object file/symbol file to reload from the Object Files with
Symbols Loaded For Label field.
2. Select the Reload button.
The values of the object file symbols being used in the trigger sequence
or in SPA state-interval ranges will be updated automatically each time
111
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Using Symbols
the object file symbols are reloaded.
To delete object file symbol files
1. Select the Symbol tab, and then the Object File tab.
2. Select the file name you want to delete in the text box labeled, Object
Files with Symbols Loaded For Label.
3. Select Unload.
See Also
“Symbol File Formats” on page 123
To adjust symbol values for relocated code
Use this option to add offset values to the symbols in an object file. You
will need this if some of the sections or segments of your code are
relocated in memory at run-time. This can occur if your system
dynamically loads parts of your code so that the memory addresses
that the code is loaded into are not fixed.
To adjust symbol values for a single section of code
1. Select the Symbol tab and then the Object File tab.
2. In the Object Files with Symbols Loaded For Label list, select the file
whose symbols you wish to relocate.
3. Select the Relocate Sections... button.
4. In the Section Relocation dialog, select the field you wish to edit in the
section list.
5. Enter the new value for that field and press Enter on your keyboard.
6. Repeat steps 4 through 6 above for any other sections to be relocated.
7. Select Close.
To adjust all symbol values
1. Select the Symbol tab and then the Object File tab.
2. In the Object Files with Symbols Loaded For Label list, select the file
112
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Using Symbols
whose symbols you wish to relocate.
3. Select the Relocate Sections... button.
4. Enter the desired offset in the Offset all sections by field. The offset is
applied from the linked address or segment.
5. Select Apply Offset.
6. Select Close.
To create user-defined symbols
1. Under the Symbol tab, select the User Defined tab.
2. Select the label name you want to define symbols for.
3. At the bottom of the User Defined tab, enter a symbol name in the entry
field.
4. Select a numeric base.
5. Select Pattern or Range type for the symbol.
6. Enter values for the pattern or range the symbol will represent.
7. Select Add.
8. Repeat steps 3 through 7 for additional symbols.
9. You can edit your list of symbols by replacing or deleting them, if desired.
To replace user-defined symbols
1. Under the Symbol tab, select the User Defined tab.
2. Select the label you want to replace symbols for.
3. Select the symbol to replace.
4. At the bottom of the User Defined tab, modify the symbol name, numeric
base, Pattern/Range type, and value, as desired.
5. Select the Replace button.
6. Repeat steps 3 through 5 to replace other symbols, if desired.
113
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Using Symbols
To delete user-defined symbols
1. Under the Symbol tab, select the User Defined tab.
2. Select the label you want to delete symbols from.
3. Select the symbol to delete.
4. Select the Delete button.
5. Repeat steps 3 and 4 to delete other symbols, if desired.
To load user-defined symbols
If you have already saved a configuration file, and the configuration
included user-defined symbols, load the file with its symbols, as
follows:
1. In the menu bar of your analyzer window, select File and then Load
Configuration....
2. In the Load Configuration dialog, select the directory and filename to be
loaded.
3. Select the target of the load operation.
4. Select Load.
User-defined symbols that were resident in the logic analyzer when the
configuration was saved are now loaded and ready to use.
To enter symbolic label values
When entering label values in the Pattern or Range subtabs of the
Trigger tab:
1. Choose the Symbols or Line #s number base.
2. Select the Absolute XXXX button.
3. In the Symbol Selector dialog, select the symbol you want to use. All of
your symbols for the current label, regardless of type, will be available in
the dialog.
•
114
Use the Search Pattern (see page 122) field to filter the list of symbols
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Using Symbols
by name. You can use the Recall button to recall a desired Search
Pattern.
•
Use the Find Symbols of Type selections to filter the symbols by type.
4. Select the symbol you want to use from the list of Matching Symbols.
5. If you are using object file symbols, you may need to:
•
Set Offset By (see page 122) to compensate for microprocessor
prefetches.
•
Set Align to x Byte (see page 123) to trigger on odd-byte boundaries.
6. Select the Beginning, End, or Range of the symbol.
7. Select the OK button.
The name of your symbol now appears as the value of the label.
8. Select the Cancel button to exit the Symbol Selector dialog without
selecting a symbol.
NOTE:
When you use symbolic label values in trigger sequences, information about
the symbols is saved with the logic analyzer configuration (and in the trigger
Save/Recall buffer). If you re-compile your program and reload the logic
analyzer configuration (or reload symbols and recall a trigger), the symbol
values are automatically re-calculated.
Triggers set up by the Source Viewer now use symbolic values and are recalculated in the same way.
See Also
“Symbols Selector Dialog” on page 121
To create an ASCII symbol file
General-purpose ASCII symbol files are created with text editing/
processing tools.
See Also
“General-Purpose ASCII (GPA) Symbol File Format” on page 124
115
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Using Symbols
To create a readers.ini file
You can change how an ELF/Stabs, Ticoff or Coff/Stabs symbol file is
processed by creating a reader.ini file.
1. Create the reader.ini file on your workstation or PC.
2. Copy the file to /logic/symbols/readers.ini on the logic analysis system.
Reader options
C++Demangle
1= Turn on C++ Demangling (Default)
0= Turn off C++ Demangling
C++DemOptions
803=
203=
403=
800=
200=
400=
Standard Demangling
GNU Demangling
(Default Elf/Stabs)
Lucid Demangling
Standard Demangling without function parameters
GNU Demangling without function parameters
Lucid Demangling without function parameters
MaxSymbolWidth
80= Column width max of a function or variable symbol
Wider symbols names will be truncated.
(Default 80 columns)
OutSectionSymbolValid
0= Symbols whose addresses aren't within the
defined sections are invalid (Default)
1= Symbols whose addresses aren't within the
defined sections are valid
This option must be specified in the Nsr section of the Readers.ini file:
[Nsr]
OutSectionSymbolValid=1
ReadElfSection
2= Process all globals from ELF section (Default)
Get size information of local variables
1= Get size information of global and local variables
Symbols for functions will not be read, and
only supplemental information for those symbols in
the Dwarf or stabs section will be read.
0= Do not read the Elf Section
If a file only has an ELF section this will have no effect and the ELF
116
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
Using Symbols
section will be read completely. This can occur if the file was created
without a "generate debugger information" flag (usually -g). Using the g will create a Dwarf or Stabs debug section in addition to the ELF
section.
StabsType
StabsType=0
StabsType=1
StabsType=2
Reader will determine stabs type (Default)
Older style stabs
(Older style stabs have individual symbol
tables for each file that was linked into
the target executable, the indexes of each
symbol table restart at 0 for each file.)
Newer style stabs
(New style stabs have a single symbol table
where all symbols are merged into a large
symbol array).
ReadOnlyTicoffPage
ReadOnlyTicoffPage tells the ticoff reader to read only the symbols
associated with the specified page (as an example
'ReadOnlyTicoffPage=0' reads only page 0 symbols). A value of -1 tells
the ticoff readers to read symbols associated with all pages.
ReadOnlyTicoffPage=-1 Read all symbols associated will all
ticoff pages (Default)
ReadOnlyTicoffPage=p Read only symbols associated with
page 'p' (where p is any integer
between 0 and n the last page of
the object file).
AppendTicoffPage
AppendTicoffPage tells the ticoff reader to append the page number to
the symbol value. This assumes that the symbol value is 16-bits wide
and that that page number is a low positive number which can be ORed
into the upper 16 bits of an address to create a new 32-bit symbol
address. For example, if the page is 10 decimal and the symbol address
is 0xF100 then the new symbol address will be 0xAF100.
AppendTicoffPage=1
AppendTicoffPage=0
Examples
Append the ticoff page to the symbol
address
Do not append the ticoff page to the
symbol address (Default)
Example for Elf/Stabs
[ReadersElf]
C
117
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
Using Symbols
C
MaxSymbolWidth=60
StabsType=2
Example for Coff/Stabs (using Ticoff reader)
[ReadersTicoff]
C
C
MaxSymbolWidth=60
StabsType=2
Example for Ticoff
[ReadersTicoff]
C
C
MaxSymbolWidth=60
ReadOnlyTicoffPage=4
AppendTicoffPage=1
118
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Symbols Tab
The Symbols Tab
The Symbols tab lets you load symbol files or define your own symbols.
Symbols are names for particular data values on a label.
Two kinds of symbols are available:
Multiple files
•
Object File Symbols. These are symbols from your source code and
symbols generated by your compiler.
•
User-Defined Symbols. These are symbols you create.
•
“Symbols Selector Dialog” on page 121
•
“Symbol File Formats” on page 123
•
“General-Purpose ASCII (GPA) Symbol File Format” on page 124
You can load the same symbol file into several different analyzers, and
you can load multiple symbol files into one analyzer. Symbols from all
the files you load will appear together in the object file symbol selector
that you use to set up resource terms.
119
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Symbols Tab
Object file versions
During the load process, a symbol database file with a .ns extension
will be created by the system. One .ns database file will be created for
each symbol file you load. Once the .ns file is created, the Symbol
Utility will use this file as its working symbol database. The next time
you need to load symbols into the system, you can load the .ns file
explicitly, by placing the .ns file name in the Load This Object/Symbol
File For Label field.
If you load an object file that has been loaded previously, the system
will compare the time stamps on the .ns file and the object file. If the
object file is newer, the .ns file will be created. If the object file has not
been updated since it was last loaded, the existing .ns file will be used.
See Also
“Using Symbols” on page 110
120
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Symbols Tab
Symbols Selector Dialog
Search Pattern: Lets you enter partial symbol names and the asterisk
wildcard character (*) to limit the symbols to choose from
(see “Search Pattern” on page 122). Use the Recall button
to select from previous search patterns.
Find Symbols of
Type
Lets you limit the types of symbols to choose from.
Matching
Symbols
Lists the symbols that match the search pattern. You
choose a symbol from this list.
121
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Symbols Tab
Offset By
Lets you add an offset value to the starting point of a
symbol. This can be useful when compensating for
microprocessor prefetches (see “Offset By Option” on
page 122).
Align to
Lets you mask the lower order bits of a symbol's value. This
can be useful for triggering on odd byte boundaries (see
“Align to x Byte Option” on page 123).
Beginning/End/
Range
When a symbol represents a range of addresses, you can
choose the beginning address of the range, the end address
of the range, or the whole range.
See Also
“To enter symbolic label values” on page 114
Search Pattern
Use this field to locate particular symbols in the symbol databases. To
use this field, enter the name of a file or symbol. The system searches
the symbol database for symbols that match this name. Symbols that
match appear in the list of Matching Symbols. You can also use
wildcard characters to find symbols.
Asterisk wildcard (*)
The asterisk wildcard represents "any characters." When you perform a
search on the symbol database using just the asterisk, you will see a list
of all symbols contained in the database. The asterisk can also be
added to a search word to find all symbols that begin or end with the
same letters. For example, to find all of the symbols that begin with the
letters "st", select the Search Pattern field and enter "st*".
Offset By Option
The Offset By option allows you to add an offset value to the starting
point of the symbol that you want to use. You might do this in order to
trigger on a point in a function that is beyond the preamble of the
function, or to trigger on a point that is past the prefetch depth of the
processor. Setting an offset helps to avoid false triggers in these
situations. The offset specified in the Offset By field is applied before
the address masking is done by the "Align to x Byte" option.
Example
An 80386 processor has a prefetch depth of 16 bytes. Assume functions
122
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Symbols Tab
func1 and func2 are adjacent to each other in physical memory, with
func2 following func1. In order to trigger on func2 without getting a
false trigger from a prefetch beyond the end of func1, you need to add
an offset value to your label value. The offset value must be equal to or
greater than the prefetch depth of the processor. In this case, you
would add an offset of 16 bytes to your label value. You would set the
value of the "Offset By" field to 10 hex. Now, when you specify func2 as
your label value, the logic analyzer will trigger on address func2+10.
Align to x Byte Option
Most processors do not fetch instructions from memory on byte
boundaries. In order to trigger a logic analyzer on a symbol at an oddnumbered address, the address must be masked off. The "Align to x
Byte" option allows you to mask off an address.
Example
Assume the symbol "main" occurs at address 100F. The processor being
probed is a 68040, which fetches instructions on long-word (4-byte)
boundaries. In order to trigger on address 100F, the Align to x Byte
option sets the two least-significant address bits to "don't cares". This
qualifies any address from 100C through 100F.
Symbol File Formats
The logic analysis system can read symbol files in the following
formats:
•
OMF96
•
OMFx86
•
IEEE-695
•
ELF/DWARF
•
ELF/stabs
•
TI COFF
For ELF/DWARF1, ELF/stabs, and ELF/stabs/Mdebug files, C++
symbols are demangled so that they can be displayed in the original
123
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Symbols Tab
C++ notation. To improve performance for these ELF symbol files, type
information is not associated with variables. Hence, some variables
(typically a few local static variables) may not have the proper size
associated with them. They may show a size of 1 byte and not the
correct size of 4 bytes or even more. All other information function
ranges, line numbers, global variables and filenames will be accurate.
These behaviors may be changed by creating a readers.ini (see
page 116) file.
See Also
“To load object file symbols” on page 111
“To create an ASCII symbol file” on page 115
“To create a readers.ini file” on page 116
General-Purpose ASCII (GPA) Symbol File
Format
General-purpose ASCII (GPA) format files are loaded into a logic
analyzer just like other object files.
If your compiler does not produce object files in a supported format, or
if you want to define symbols that are not included in the object file,
you can create an ASCII format symbol file.
Typically, ASCII format symbol files are created using text processing
tools that convert the symbol table information from a compiler or
linker map output file.
Different types of symbols are defined in different records in the GPA
file. Record headers are enclosed in square brackets, for example,
[VARIABLES]. For a summary of GPA file records and associated
symbol definition syntax, refer to the “GPA Record Format Summary”
on page 125 that follows.
Each entry in the symbol file must consist of a symbol name followed
by an address or address range.
While symbol names can be longer, the logic analyzer only uses the first
16 characters.
124
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Symbols Tab
The address or address range must be a hexadecimal number. It must
appear on the same line as the symbol name, and it must be separated
from the symbol name by one or more blank spaces or tabs. Address
ranges must be in the following format:
beginning address..ending address
The following example defines two symbols that correspond to address
ranges and one symbol that corresponds to a single address.
main
test
var1
00001000..00001009
00001010..0000101F
00001E22
#this is a variable
For more detailed descriptions of GPA file records and associated
symbol definition syntax, refer to the following topics:
•
“SECTIONS” on page 126
•
“FUNCTIONS” on page 127
•
“VARIABLES” on page 128
•
“SOURCE LINES” on page 128
•
“START ADDRESS” on page 129
•
“Comments” on page 129
GPA Record Format Summary
Format
[SECTIONS]
section_name
start..end
attribute
[FUNCTIONS]
func_name start..end
[VARIABLES]
var_name
start [size]
var_name
start..end
[SOURCE LINES]
File: file_name
line# address
125
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Symbols Tab
[START ADDRESS]
address
#comment text
Lines without a preceding header are assumed to be symbol definitions
in one of the [VARIABLES] formats.
Example
This is an example GPA file that contains several different kinds of
records.
[SECTIONS]
prog
00001000..0000101F
data
40002000..40009FFF
common
FFFF0000..FFFF1000
[FUNCTIONS]
main
00001000..00001009
test
00001010..0000101F
[VARIABLES]
total
40002000
value
40008000
4
4
[SOURCE LINES]
File: main.c
10
00001000
11
00001002
14
0000100A
22
0000101E
File: test.c
5
00001010
7
00001012
11
0000101A
SECTIONS
Use SECTIONS to define symbols for regions of memory, such as
sections, segments, or classes.
NOTE:
To enable section relocation, section definitions must appear before any other
definitions in the file.
126
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Symbols Tab
NOTE:
If you use section definitions in a GPA symbol file, any subsequent function or
variable definitions must be within the address ranges of one of the defined
sections. Functions and variables that are not within the range are ignored.
Format
[SECTIONS]
section_name
start..end
attribute
section_name A symbol representing the name of the section.
start
The first address of the section, in hexadecimal.
end
The last address of the section, in hexadecimal.
attribute
(optional) Attribute may be one of the following:
NORMAL (default) - The section is a normal, relocatable section, such as
code or data.
NONRELOC - The section contains variables or code that cannot be
relocated. In other words, this is an absolute segment.
Example
[SECTIONS]
prog
data
display_io
00001000..00001FFF
00002000..00003FFF
00008000..0000801F
NONRELOC
FUNCTIONS
Use FUNCTIONS to define symbols for program functions, procedures
or subroutines.
Format
[FUNCTIONS]
func_name start..end
func_name
A symbol representing the function name.
start
The first address of the function, in hexadecimal.
end
The last address of the function, in hexadecimal.
Example
[FUNCTIONS]
main
00001000..00001009
test
00001010..0000101F
127
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Symbols Tab
VARIABLES
You can specify symbols for variables using:
•
The address of the variable.
•
The address and the size of the variable.
•
The range of addresses occupied by the variable.
If you specify only the address of a variable, the size is assumed to be 1
byte.
Format
[VARIABLES]
var_name
start [size]
var_name
start..end
var_name
A symbol representing the variable name.
start
The first address of the variable, in hexadecimal.
end
The last address of the variable, in hexadecimal.
size
(optional) The size of the variable, in bytes, in decimal.
Example
[VARIABLES]
subtotal
total
data_array
status_char
40002000
4
40002004
4
40003000..4000302F
40002345
SOURCE LINES
Use SOURCE LINES to associate addresses with lines in your source
files.
Format
[SOURCE LINES]
File: file_name
line# address
file_name
The name of a file.
line#
The number of a line in the file, in decimal.
address
The address of the source line, in hexadecimal.
128
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic Analyzer
The Symbols Tab
Example
[SOURCE LINES]
File: main.c
10
00001000
11
00001002
14
0000100A
22
0000101E
See Also
Using the Source Viewer (see the Listing Display Tool help volume)
START ADDRESS
Format
[START ADDRESS]
address
address
The address of the program entry point, in hexadecimal.
Example
[START ADDRESS]
00001000
Comments
Use the # character to include comments in a file. Any text following
the # character is ignored. You can put comments on a line alone or on
the same line following s symbol entry.
Format
#comment text
Example
#This is a comment
129
Chapter 1: Agilent Technologies 16550A 100 MHz State/500 MHz Timing Logic
Analyzer
The Symbols Tab
130
Glossary
absolute Denotes the time period
or count of states between a captured
state and the trigger state. An
absolute count of -10 indicates the
state was captured ten states before
the trigger state was captured.
acquisition Denotes one complete
cycle of data gathering by a
measurement module. For example,
if you are using an analyzer with
128K memory depth, one complete
acquisition will capture and store
128K states in acquisition memory.
analysis probe A probe connected
to a microprocessor or standard bus
in the device under test. An analysis
probe provides an interface between
the signals of the microprocessor or
standard bus and the inputs of the
logic analyzer. Also called a
preprocessor.
analyzer 1 In a logic analyzer with
two machines, refers to the machine
that is on by default. The default
name is Analyzer<N>, where N is
the slot letter.
analyzer 2 In a logic analyzer with
two machines, refers to the machine
that is off by default. The default
name is Analyzer<N2>, where N is
the slot letter.
arming An instrument tool must be
armed before it can search for its
trigger condition. Typically,
instruments are armed immediately
when Run or Group Run is selected.
You can set up one instrument to arm
another using the Intermodule
Window. In these setups, the second
instrument cannot search for its
trigger condition until it receives the
arming signal from the first
instrument. In some analyzer
instruments, you can set up one
analyzer machine to arm the other
analyzer machine in the Trigger
Window.
asterisk (*) See edge terms,
glitch, and labels.
bits Bits represent the physical logic
analyzer channels. A bit is a channel
that has or can be assigned to a label.
A bit is also a position in a label.
card This refers to a single
instrument intended for use in the
Agilent Technologies 16700A/Bseries mainframes. One card fills one
slot in the mainframe. A module may
comprise a single card or multiple
cards cabled together.
channel The entire signal path from
the probe tip, through the cable and
module, up to the label grouping.
click When using a mouse as the
131
Glossary
pointing device, to click an item,
position the cursor over the item.
Then quickly press and release the
left mouse button.
clock channel A logic analyzer
channel that can be used to carry the
clock signal. When it is not needed
for clock signals, it can be used as a
data channel, except in the Agilent
Technologies 16517A.
context record A context record is
a small segment of analyzer memory
that stores an event of interest along
with the states that immediately
preceded it and the states that
immediately followed it.
context store If your analyzer can
perform context store
measurements, you will see a button
labeled Context Store under the
Trigger tab. Typical context store
measurements are used to capture
writes to a variable or calls to a
subroutine, along with the activity
preceding and following the events. A
context store measurement divides
analyzer memory into a series of
context records. If you have a 64K
analyzer memory and select a 16state context, the analyzer memory is
divided into 4K 16-state context
records. If you have a 64K analyzer
memory and select a 64-state
context, the analyzer memory will be
132
divided into 1K 64-state records.
count The count function records
periods of time or numbers of state
transactions between states stored in
memory. You can set up the analyzer
count function to count occurrences
of a selected event during the trace,
such as counting how many times a
variable is read between each of the
writes to the variable. The analyzer
can also be set up to count elapsed
time, such as counting the time spent
executing within a particular function
during a run of your target program.
cross triggering Using intermodule
capabilities to have measurement
modules trigger each other. For
example, you can have an external
instrument arm a logic analyzer,
which subsequently triggers an
oscilloscope when it finds the trigger
state.
data channel A channel that
carries data. Data channels cannot be
used to clock logic analyzers.
data field A data field in the pattern
generator is the data value associated
with a single label within a particular
data vector.
data set A data set is made up of all
labels and data stored in memory of
any single analyzer machine or
Glossary
instrument tool. Multiple data sets
can be displayed together when
sourced into a single display tool. The
Filter tool is used to pass on partial
data sets to analysis or display tools.
debug mode See monitor.
delay The delay function sets the
horizontal position of the waveform
on the screen for the oscilloscope and
timing analyzer. Delay time is
measured from the trigger point in
seconds or states.
demo mode An emulation control
session which is not connected to a
real target system. All windows can
be viewed, but the data displayed is
simulated. To start demo mode,
select Start User Session from the
Emulation Control Interface and
enter the demo name in the
Processor Probe LAN Name field.
Select the Help button in the Start
User Session window for details.
deskewing To cancel or nullify the
effects of differences between two
different internal delay paths for a
signal. Deskewing is normally done
by routing a single test signal to the
inputs of two different modules, then
adjusting the Intermodule Skew so
that both modules recognize the
signal at the same time.
device under test The system
under test, which contains the
circuitry you are probing. Also known
as a target system.
don't care For terms, a "don't care"
means that the state of the signal
(high or low) is not relevant to the
measurement. The analyzer ignores
the state of this signal when
determining whether a match occurs
on an input label. "Don't care" signals
are still sampled and their values can
be displayed with the rest of the data.
Don't cares are represented by the X
character in numeric values and the
dot (.) in timing edge specifications.
dot (.) See edge terms, glitch,
labels, and don't care.
double-click When using a mouse
as the pointing device, to double-click
an item, position the cursor over the
item, and then quickly press and
release the left mouse button twice.
drag and drop Using a Mouse:
Position the cursor over the item, and
then press and hold the left mouse
button. While holding the left mouse
button down, move the mouse to
drag the item to a new location. When
the item is positioned where you
want it, release the mouse button.
133
Glossary
Using the Touchscreen:
Position your finger over the item,
then press and hold finger to the
screen. While holding the finger
down, slide the finger along the
screen dragging the item to a new
location. When the item is positioned
where you want it, release your
finger.
edge mode In an oscilloscope, this
is the trigger mode that causes a
trigger based on a single channel
edge, either rising or falling.
edge terms Logic analyzer trigger
resources that allow detection of
transitions on a signal. An edge term
can be set to detect a rising edge,
falling edge, or either edge. Some
logic analyzers can also detect no
edge or a glitch on an input signal.
Edges are specified by selecting
arrows. The dot (.) ignores the bit.
The asterisk (*) specifies a glitch on
the bit.
emulation module A module
within the logic analysis system
mainframe that provides an
emulation connection to the debug
port of a microprocessor. An E5901A
emulation module is used with a
target interface module (TIM) or an
analysis probe. An E5901B emulation
module is used with an E5900A
emulation probe.
134
emulation probe The stand-alone
equivalent of an emulation module.
Most of the tasks which can be
performed using an emulation
module can also be performed using
an emulation probe connected to
your logic analysis system via a LAN.
emulator An emulation module or
an emulation probe.
Ethernet address See link-level
address.
events Events are the things you
are looking for in your target system.
In the logic analyzer interface, they
take a single line. Examples of events
are Label1 = XX and Timer 1 > 400
ns.
filter expression The filter
expression is the logical OR
combination of all of the filter terms.
States in your data that match the
filter expression can be filtered out or
passed through the Pattern Filter.
filter term A variable that you
define in order to specify which
states to filter out or pass through.
Filter terms are logically OR'ed
together to create the filter
expression.
Format The selections under the
logic analyzer Format tab tell the
Glossary
logic analyzer what data you want to
collect, such as which channels
represent buses (labels) and what
logic threshold your signals use.
frame The Agilent Technologies or
16700A/B-series logic analysis system
mainframe. See also logic analysis
system.
gateway address An IP address
entered in integer dot notation. The
default gateway address is 0.0.0.0,
which allows all connections on the
local network or subnet. If
connections are to be made across
networks or subnets, this address
must be set to the address of the
gateway machine.
glitch A glitch occurs when two or
more transitions cross the logic
threshold between consecutive
timing analyzer samples. You can
specify glitch detection by choosing
the asterisk (*) for edge terms under
the timing analyzer Trigger tab.
grouped event A grouped event is
a list of events that you have
grouped, and optionally named. It can
be reused in other trigger sequence
levels. Only available in Agilent
Technologies 16715A or higher logic
analyzers.
held value A value that is held until
the next sample. A held value can
exist in multiple data sets.
immediate mode In an
oscilloscope, the trigger mode that
does not require a specific trigger
condition such as an edge or a
pattern. Use immediate mode when
the oscilloscope is armed by another
instrument.
interconnect cable Short name for
module/probe interconnect cable.
intermodule bus The intermodule
bus (IMB) is a bus in the frame that
allows the measurement modules to
communicate with each other. Using
the IMB, you can set up one
instrument to arm another. Data
acquired by instruments using the
IMB is time-correlated.
intermodule Intermodule is a term
used when multiple instrument tools
are connected together for the
purpose of one instrument arming
another. In such a configuration, an
arming tree is developed and the
group run function is designated to
start all instrument tools. Multiple
instrument configurations are done in
the Intermodule window.
internet address Also called
Internet Protocol address or IP
address. A 32-bit network address. It
135
Glossary
is usually represented as decimal
numbers separated by periods; for
example, 192.35.12.6. Ask your LAN
administrator if you need an internet
address.
labels Labels are used to group and
identify logic analyzer channels. A
label consists of a name and an
associated bit or group of bits. Labels
are created in the Format tab.
line numbers A line number (Line
#s) is a special use of symbols. Line
numbers represent lines in your
source file, typically lines that have
no unique symbols defined to
represent them.
link-level address Also referred to
as the Ethernet address, this is the
unique address of the LAN interface.
This value is set at the factory and
cannot be changed. The link-level
address of a particular piece of
equipment is often printed on a label
above the LAN connector. An
example of a link-level address in
hexadecimal: 0800090012AB.
local session A local session is
when you run the logic analysis
system using the local display
connected to the product hardware.
logic analysis system The Agilent
Technologies 16700A/B-series
136
mainframes, and all tools designed to
work with it. Usually used to mean
the specific system and tools you are
working with right now.
machine Some logic analyzers allow
you to set up two measurements at
the same time. Each measurement is
handled by a different machine. This
is represented in the Workspace
window by two icons, differentiated
by a 1 and a 2 in the upper right-hand
corner of the icon. Logic analyzer
resources such as pods and trigger
terms cannot be shared by the
machines.
markers Markers are the green and
yellow lines in the display that are
labeled x, o, G1, and G2. Use them to
measure time intervals or sample
intervals. Markers are assigned to
patterns in order to find patterns or
track sequences of states in the data.
The x and o markers are local to the
immediate display, while G1 and G2
are global between time correlated
displays.
master card In a module, the
master card controls the data
acquisition or output. The logic
analysis system references the
module by the slot in which the
master card is plugged. For example,
a 5-card Agilent Technologies 16555D
would be referred to as Slot C:
Glossary
machine because the master card is
in slot C of the mainframe. The other
cards of the module are called
expansion cards.
menu bar The menu bar is located
at the top of all windows. Use it to
select File operations, tool or system
Options, and tool or system level
Help.
message bar The message bar
displays mouse button functions for
the window area or field directly
beneath the mouse cursor. Use the
mouse and message bar together to
prompt yourself to functions and
shortcuts.
module/probe interconnect cable
The module/probe interconnect cable
connects an E5901B emulation
module to an E5900B emulation
probe. It provides power and a serial
connection. A LAN connection is also
required to use the emulation probe.
module An instrument that uses a
single timebase in its operation.
Modules can have from one to five
cards functioning as a single
instrument. When a module has more
than one card, system window will
show the instrument icon in the slot
of the master card.
monitor When using the Emulation
Control Interface, running the
monitor means the processor is in
debug mode (that is, executing the
debug exception) instead of
executing the user program.
panning The action of moving the
waveform along the timebase by
varying the delay value in the Delay
field. This action allows you to
control the portion of acquisition
memory that will be displayed on the
screen.
pattern mode In an oscilloscope,
the trigger mode that allows you to
set the oscilloscope to trigger on a
specified combination of input signal
levels.
pattern terms Logic analyzer
resources that represent single states
to be found on labeled sets of bits; for
example, an address on the address
bus or a status on the status lines.
period (.) See edge terms, glitch,
labels, and don't care.
pod pair A group of two pods
containing 16 channels each, used to
physically connect data and clock
signals from the unit under test to the
analyzer. Pods are assigned by pairs
in the analyzer interface. The number
of pod pairs available is determined
137
Glossary
by the channel width of the
instrument.
pod See pod pair
point To point to an item, move the
mouse cursor over the item, or
position your finger over the item.
preprocessor See analysis probe.
primary branch The primary
branch is indicated in the Trigger
sequence step dialog box as either
the Then find or Trigger on
selection. The destination of the
primary branch is always the next
state in the sequence, except for the
Agilent Technologies 16517A. The
primary branch has an optional
occurrence count field that can be
used to count a number of
occurrences of the branch condition.
See also secondary branch.
probe A device to connect the
various instruments of the logic
analysis system to the target system.
There are many types of probes and
the one you should use depends on
the instrument and your data
requirements. As a verb, "to probe"
means to attach a probe to the target
system.
processor probe See emulation
probe.
138
range terms Logic analyzer
resources that represent ranges of
values to be found on labeled sets of
bits. For example, range terms could
identify a range of addresses to be
found on the address bus or a range
of data values to be found on the data
bus. In the trigger sequence, range
terms are considered to be true when
any value within the range occurs.
relative Denotes time period or
count of states between the current
state and the previous state.
remote display A remote display is
a display other than the one
connected to the product hardware.
Remote displays must be identified to
the network through an address
location.
remote session A remote session is
when you run the logic analyzer using
a display that is located away from
the product hardware.
right-click When using a mouse for
a pointing device, to right-click an
item, position the cursor over the
item, and then quickly press and
release the right mouse button.
sample A data sample is a portion of
a data set, sometimes just one point.
When an instrument samples the
target system, it is taking a single
Glossary
measurement as part of its data
acquisition cycle.
Sampling Use the selections under
the logic analyzer Sampling tab to tell
the logic analyzer how you want to
make measurements, such as State
vs. Timing.
secondary branch The secondary
branch is indicated in the Trigger
sequence step dialog box as the Else
on selection. The destination of the
secondary branch can be specified as
any other active sequence state. See
also primary branch.
session A session begins when you
start a local session or remote
session from the session manager,
and ends when you select Exit from
the main window. Exiting a session
returns all tools to their initial
configurations.
skew Skew is the difference in
channel delays between
measurement channels. Typically,
skew between modules is caused by
differences in designs of
measurement channels, and
differences in characteristics of the
electronic components within those
channels. You should adjust
measurement modules to eliminate
as much skew as possible so that it
does not affect the accuracy of your
measurements.
state measurement In a state
measurement, the logic analyzer is
clocked by a signal from the system
under test. Each time the clock signal
becomes valid, the analyzer samples
data from the system under test.
Since the analyzer is clocked by the
system, state measurements are
synchronous with the test system.
store qualification Store
qualification is only available in a
state measurement, not timing
measurements. Store qualification
allows you to specify the type of
information (all samples, no samples,
or selected states) to be stored in
memory. Use store qualification to
prevent memory from being filled
with unwanted activity such as noops or wait-loops. To set up store
qualification, use the While storing
field in a logic analyzer trigger
sequence dialog.
subnet mask A subnet mask blocks
out part of an IP address so that the
networking software can determine
whether the destination host is on a
local or remote network. It is usually
represented as decimal numbers
separated by periods; for example,
255.255.255.0. Ask your LAN
administrator if you need a the
subnet mask for your network.
139
Glossary
symbols Symbols represent
patterns and ranges of values found
on labeled sets of bits. Two kinds of
symbols are available:
•
Object file symbols - Symbols
from your source code, and
symbols generated by your
compiler. Object file symbols may
represent global variables,
functions, labels, and source line
numbers.
•
User-defined symbols - Symbols
you create.
Symbols can be used as pattern and
range terms for:
•
Searches in the listing display.
•
Triggering in logic analyzers and
in the source correlation trigger
setup.
•
Qualifying data in the filter tool
and system performance analysis
tool set.
system administrator The system
administrator is a person who
manages your system, taking care of
such tasks as adding peripheral
devices, adding new users, and doing
system backup. In general, the
system administrator is the person
you go to with questions about
implementing your software.
140
target system The system under
test, which contains the
microprocessor you are probing.
terms Terms are variables that can
be used in trigger sequences. A term
can be a single value on a label or set
of labels, any value within a range of
values on a label or set of labels, or a
glitch or edge transition on bits
within a label or set of labels.
TIM A TIM (Target Interface
Module) makes connections between
the cable from the emulation module
or emulation probe and the cable to
the debug port on the system under
test.
time-correlated Time correlated
measurements are measurements
involving more than one instrument
in which all instruments have a
common time or trigger reference.
timer terms Logic analyzer
resources that are used to measure
the time the trigger sequence
remains within one sequence step, or
a set of sequence steps. Timers can
be used to detect when a condition
lasts too long or not long enough.
They can be used to measure pulse
duration, or duration of a wait loop. A
single timer term can be used to
delay trigger until a period of time
after detection of a significant event.
Glossary
timing measurement In a timing
measurement, the logic analyzer
samples data at regular intervals
according to a clock signal internal to
the timing analyzer. Since the
analyzer is clocked by a signal that is
not related to the system under test,
timing measurements capture traces
of electrical activity over time. These
measurements are asynchronous
with the test system.
tool icon Tool icons that appear in
the workspace are representations of
the hardware and software tools
selected from the toolbox. If they are
placed directly over a current
measurement, the tools automatically
connect to that measurement. If they
are placed on an open area of the
main window, you must connect them
to a measurement using the mouse.
toolbox The Toolbox is located on
the left side of the main window. It is
used to display the available
hardware and software tools. As you
add new tools to your system, their
icons will appear in the Toolbox.
tools A tool is a stand-alone piece of
functionality. A tool can be an
instrument that acquires data, a
display for viewing data, or a postprocessing analysis helper. Tools are
represented as icons in the main
window of the interface.
trace See acquisition.
trigger sequence A trigger
sequence is a sequence of events that
you specify. The logic analyzer
compares this sequence with the
samples it is collecting to determine
when to trigger.
trigger specification A trigger
specification is a set of conditions
that must be true before the
instrument triggers.
trigger Trigger is an event that
occurs immediately after the
instrument recognizes a match
between the incoming data and the
trigger specification. Once trigger
occurs, the instrument completes its
acquisition, including any store
qualification that may be specified.
workspace The workspace is the
large area under the message bar and
to the right of the toolbox. The
workspace is where you place the
different instrument, display, and
analysis tools. Once in the workspace,
the tool icons graphically represent a
complete picture of the
measurements.
zooming In the oscilloscope or
timing analyzer, to expand and
contract the waveform along the time
base by varying the value in the s/Div
141
Glossary
field. This action allows you to select
specific portions of a particular
waveform in acquisition memory that
will be displayed on the screen. You
can view any portion of the waveform
record in acquisition memory.
142
Index
Symbols
, 93
&> duration trigger field, 93
*, bit assignment, 66
+, label polarity, 68
-, label polarity, 68
., bit unassignment, 66
Numerics
125 MHz acquisition mode, 47
16550A characteristics, 105
16550A logic analyzer, 2
16550A specifications, 104
250 MHz acquisition mode, 48
A
absolute tag, 101
acquisition depth control, 40
acquisition mode, 41
acquisition modes, about
transitional timing, 46
acquisition modes, comparison of
timing, 51
acquisition modes, overview, 11
acquisition modes, state, 46
acquisition modes, timing, 46
acquisition modes, timing, glitch,
52
acquisitions, interpreting, 26
acquisitions, processing, 26
activity indicator, in Format, 66
advanced clocking checkbox, 41
Align to x Byte option, 123
Align to x Byte option for symbols,
123
altitude characteristics, 105
analyzer 2, turning on, 65
analyzer name field, 44
analyzer name, where used, 44
analyzer probes, general-purpose,
107
analyzer probes, termination
adapter, 107
analyzer, arming, 101
analyzer, changing name, 44
analyzer, off button, 45
analyzer, on checkbox, 45
arming analyzer, 101
arming control, 31, 101
arrows, activity indicators, 64
ASCII, 56
ASCII format symbols, 126, 127,
128, 129
ASCII symbol file, 128
B
bad data in measurement, 38
base, numeric, 100
basic state measurements,
example, 18
basic timing measurements,
example, 22
bit activity indicator, 66
bit assignments, preserving, 68
bit numbering within a label, 66
bit order, changing, 69
bit pattern terms, 95
bit significance, 66
bits, assigning, 66
bits, reordering, 69
branches taken, 94
break down functions, 90
browsing, 122
browsing the symbol database, 122
buses, naming, 70
byte mode, numeric base, 100
C
cable activity indicators, 64
cable power, probe, characteristic,
105
capacitive loading, 33
center trigger position, 53
channel activity indicators, 64
channel counts, characteristic, 105
channel width, 46
channels, assigning to label, 70
channels, maximum per label, 66
channel-to-channel skew,
characteristic, 105
clear menu, 83
clear sequence levels, 83
clear sequence/resource/name, 83
clock bits as data channels, 66
clock channel specifiers, 41
clock channels, inputs available as
data, 66
clock edges, adjusting, 73
clock lines informational message,
37
clock mode field, 41
clock qualifiers, characteristic, 105
clock setup area, 41
clock threshold level note, 72
clock threshold note, 70
clock time, specification, 104
clocks, data/address same line, 42
clocks, master/slave/demultiplex,
42
clocks, overview, 11
code, assigning address offsets, 112
COFF symbol reader options, 117
combination terms, 99
combo pick, 99
comments, 129
config tab, use, 10
configuration files, loading, 32
configurations, compatibility
across models, 32
configurations, storing, 32
connection methods, 10
connections, 10
conventional timing acquisition
mode, 51
conventional timing acquisition
modes, 46
143
Index
count state, 101
count states timing function, 85
count time, 101
create labels for busses, 70
custom state function, 87
custom timing macros, 85
custom trigger macros, 82
D
dashes, activity indicators, 64
data channels, using clock bits, 66
data latching note, 42
data on clocks display, 66
data set, interpreting, 26
data, stamping, 101
defaulting the trigger, 83
definition, calibration procedure,
104
definition, characteristic, 103
definition, function test, 104
definition, operational accuracy
calibration, 104
definition, specification, 103
delay trigger position, 53
deleting terms, 83
demultiplex clock, 42
demultiplex clocks, 71
demultiplex clocks for pods, 70
depth of memory, characteristic,
105
duration, timing, 50
E
edge and pattern function, 86
edge terms, 97
edge trigger function, 87
edge trigger terms, 97
edit trigger sequence steps, 80
ELF symbol reader options, 116
ELF/DWARF file format, 123
ELF/stabs file format, 123
else branch, 81
144
end trigger position, 53
enviromental characteristics, 105
error messages, clock lines must be
connected to the card, 37
error messages, maximum of 32
channels per label, 34
error messages, measurement
initialization error, 34
error messages, slow or missing
clock, 34
error messages, timer is off, 35
error messages, timer never
started, 35
error messages, two pod pairs are
needed to use both timers, 36
error messages, waiting for trigger,
37
errors in data, 38
errors, error messages, 33
example, 122, 123
F
file versions, 111
files, 111
find early event state trigger
functions, 89
find edge function, 85
find event state functions, 88
find late event state trigger
functions, 89
find n times state functions, 88
find pattern state functions, 88
find sequence state triggers, 88
finding the symbol you want, 122
flowchart, 10
format tab, use, 13, 55
full channel 250 MHz, timing, 46
full channel state mode, 41
full channel timing mode, 41
full channel, timing mode, 47
full channel/4K Memory/100 MHz,
46
functions, 127
G
glitch, 97
glitch mode, 41
glossary of terms, 2
group bit assignment, 66
H
half channel 500 MHz, timing, 46
half channel state mode, 41
half channel timing mode, 41, 71
half channel timing, transitional, 49
half channel timing, transitions, 50
half channel transitional timing
mode, 48
half channel/8K Memory/100 MHz,
46
help, symbols, 119
help, trigger, 101
help, trigger sequence, 83
how to set up, 10
humidity characteristics, 105
I
id, naming analyzer, 44
IEEE-695 file format, 123
if branch, 81
import, 56
in ASCII format, 126, 127, 128, 129
in symbol browser, 122
input capacitance, probe,
characteristic, 105
input resistance, probe,
characteristic, 105
internal sequence, 83
internal sequence levels, 91
invalid data note, 71
L
label polarity, changing, 68
Index
label values, symbolic, 114
labels, 14
labels, activating, 68
labels, add after, 67
labels, add before, 67
labels, adding, 67
labels, assigning bits, 66
labels, defining, 70
labels, deleting, 67
labels, deleting from terms, 100
labels, inserting, 67
labels, inserting on terms, 100
labels, polarity, 68
labels, reordering bits, 69
labels, turning off, 68
labels, turning on, 68
latch, glitch, 94, 97
latch, glitch, characteristic, 105
least significant bit in label, 66
line numbers, 128
load, reducing, 33
loading, 111
loading files including symbols, 111
loading object file symbols, 111
loading user-defined symbols, 114
logic analyzer hangs, 38
logic analyzer probes, 10, 107
logic analyzer triggers, 76
logic analyzer, testing, 39
M
machines available, characteristic,
105
main system help page, 2
masking off addresses of symbols,
123
master clock, 42
master clocks for pods, 70
maximizing memory, 94
maximum pulse width trigger, 87
maximum transitions stored in halfchannel mode, 50
maximum transitions stored, full
channel, 48
measurement doesntrun', 38
measurement setup, 11
measurement, probing options, 107
measurements, overview of basic
state, 18
measurements, overview of timing,
22
memory and trigger, 53
memory depth, characteristic, 105
memory depth, setting, 40
message, trigger inhibited during
timing prestore, 36
minimum pulse width trigger, 87
minimum transitions in halfchannel mode, 49
minimum transitions stored, full
channel, 48
mixed signal, 31
mode and acquisition depth, 41
mode and channel width, 41
mode and sample rate, 41
mode, clocking, 42
most significant bit in label, 66
N
name, id for saving setup, 44
negative logic, note, 68
netlist, 56
note, activity indicators, 64
note, clock, characteristic, 105
note, clocking, 42
note, negative logic, 68
note, pattern recognizer, 105
note, state memory, 46
O
object file symbol files, 111
object file symbols, 122
occurrence counter, characteristic,
105
occurrence counter, state trigger,
93
occurrence counter, timing trigger,
93
occurs field, 93
occurs trigger field, 93
odd-numbered addresses, 123
odd-numbered addresses
represented by symbols, 123
offset addresses, assigning, 112
Offset By option of the symbol
browser, 122
OMF96 file format, 123
OMFx86 file format, 123
on/off switch, 45
operating environment
characteristics, 105
overview, measurement process,
10
P
parasitic tip capacitance, probe,
characteristic, 105
pattern after edge function, 86
pattern count, state trigger, 93
pattern duration function, 85
pattern duration, timing, 93
Pattern field, 122
pattern recognizers, characteristic,
105
pattern, state trigger functions, 88
pattern/edge trigger functions, 86
performance verification, 39
period, sample time, 45
pod assignment dialog, 65
pod clocking, demultiplex, 70
pod thresholds, setting, 72
pods, assigning, 65
pods, clocking, 42
pods, selecting, 71
pods, specifying state clock, 70
145
Index
power through pod cables,
characteristic, 105
predefined macros, 84, 91
predefined trigger functions,
modifying, 90
prefetch, triggering beyond, 122
probe leads, 107
probes, individual signal, 107
probing options, 10
probing tips, 10
probing, overview, 107
problems making measurements,
33
pulse width function, 87
Q
qualifiers, clocks, characteristic,
105
R
R, bit assignment, 66, 69
radix, numeric base, 100
range recognizers, characteristic,
105
range terms, 96
range trigger terms, 96
range width, characteristic, 105
ranges and reordered bits, 69
rate, sample period, 45
readers.ini file, 116
recalling trigger sequences, 82
relative tag, 101
relocating sections of code, 112
reset bit pattern, 95
restore functions, 90
results, 122
roadmap, 10
S
sample period control, 45
sample period, characteristic, 105
sample rate, setting, 41
146
sampling tab, use, 40
sampling tab, uses, 11
saving trigger sequences, 82
Search Pattern field, 122
searching the symbol database, 122
sections, 126
selecting macros, 78
self test, 39
sequence levels, characteristic, 105
sequencer, maximum levels,
characteristic, 105
set up, trigger sequence, 81
setting threshold, 72
settings, saving, 32
setup/hold field, 73
setup/hold time, specification, 104
shortcut, bit assignment, 66
skew, channel-to-channel,
characteristic, 105
slave clock, 42
slave clocks for pods, 70
slow clock message, 34
source line numbers, 128
specifications and characteristics,
103
speed, state/timing, characteristic,
105
Stabs symbol reader options, 117
stamp data, 101
start address, 129
start trigger position, 53
state channel width, 46
state clock pulse width,
specification, 104
state clocks, 41
state clocks, characteristic, 105
state clocks, master/slave/both, 70
state clocks, setup/hold, 73
state count, memory, 46
state measurements, basic
overview, 18
state memory, 46
state mode, 41
state modes, 46
state modes, characteristic, 105
state speed, characteristic, 105
state tags, 101
state trigger functions, 84, 88, 89
stored/not stored, 94
storing trigger setups, 82
symbol demangling, 116
symbol file formats, 123
symbol file versions, 111
symbol selector dialog, 121
symbolic label values, 114
symbols, 122
symbols, loading object file
symbols, 111
symbols, loading user-defined, 114
symbols, outside defined sections,
116
symbols, types and use, 119
symbols, user-defined, 113
symbols, using, 110
T
tab, symbols, 119
tag data, 101
temperature characteristics, 105
terms, 14
then branch, 81
threshold accuracy, specification,
104
threshold logic levels, ECL, 72
threshold logic levels, TTL, 72
threshold range, probe,
characteristic, 105
TI COFF file format, 123
time count, 46
time count, characteristic, 105
time interval accuracy,
characteristic, 105
time tag resolution, characteristic,
105
time tags, 101
Index
time violation functions, 87
timer error message, 35
timer warning message, 35
timer, characteristics, 105
timers, using, 98
timing acquisition mode, glitch, 52
timing analysis characteristics, 105
timing measurements, basic
overview, 22
timing mode, 41
timing modes, 46
timing modes, comparison, 51
timing modes, transitional, 46
timing speed, characteristic, 105
timing trigger functions, 84
timing, memory depth, 46, 51
trace buffer, 71
transitional timing acquisition
mode, 51
transitional timing acquisition
modes, about, 46
transitional timing half-channel
mode, 50
transitional timing mode, 41
transitional timing mode,
considerations, 50
transitional timing, full channel
mode, 47
transitional timing, half channel, 48
transitional timing, minimum
transitions stored, 49
trigger bit patterns, 94
trigger characteristics, 105
trigger edge terms, 94
trigger functions, 83
trigger functions, branching, 81
trigger functions, creating, 91
trigger functions, displaying parts,
90
trigger functions, modifying, 91
trigger functions, pattern/edge, 86
trigger functions, predefined, 84
trigger functions, state, 87, 88, 89
trigger functions, time violation, 87
trigger functions, timing, 85
trigger functions, user-level, 91
trigger functions, using, 78
trigger glitch, 94
trigger inhibited informational
message, 36
trigger macros, selecting, 78
trigger pattern functions, 85
trigger pattern terms, 94
trigger position control, 53
trigger range terms, 94
trigger resource terms, 94
trigger resource terms,
combination, 99
trigger sequence branches, 81
trigger sequence levels, goto, 81
trigger sequence steps, copying, 79
trigger sequence steps, editing, 80
trigger sequence steps, replacing,
79
trigger sequence, adding steps, 79
trigger sequence, clearing, 83
trigger sequence, customizing, 91
trigger sequence, default, 83
trigger sequence, deleting, 79
trigger sequence, editing, 79
trigger sequence, explanation, 83
trigger sequence, inserting, 79
trigger sequence, specifying, 78
trigger sequences, loading, 82
trigger sequences, saving, 82
trigger sequences, storing, 82
trigger set up, 81
trigger tab, reference, 75
trigger tab, use, 14
trigger terms, adding labels, 100
trigger terms, bit pattern, 95
trigger terms, combination, 99
trigger terms, deleting labels, 100
trigger terms, edge, 97
trigger terms, range, 96
trigger terms, timers, 98
trigger timer terms, 94
trigger variables, 94
trigger, continue timer, 98
trigger, pattern durations, 93
trigger, pause timer, 98
trigger, poststore, 53
trigger, resetting, 83
trigger, setting up, 78
trigger, start timer, 98
trigger, stop timer, 98
trigger, substeps, 14
triggering on a symbol, 122, 123
triggering on a symbol beyond
prefetch depth, 122
triggering, about, 76
troubleshooting the logic analyzer,
33
U
unassigned bits, 66
user macros, trigger, 82
user threshold logic level, 72
user-defined macros, 91
user-defined symbols, 113
user-defined symbols, loading, 114
user-defined trigger, 91
user-level function, 91
user-level trigger functions, 81
V
variables, 128
versions, 111
versions of symbol files, 111
vibration characteristics, 105
W
wait state functions, 89
wait time function, 87
warranty, what is covered, 103
what is triggering, 76
while storing, 94
wildcard characters, 122
147
Index
148
Publication Number: 5988-9023EN