Campbell Scientific PVS4100/4120/4150 Portable Automatic Samplers Owner Manual
Below you will find brief information for Portable Sampler PVS 4100, Portable Sampler PVS 4120, Portable Sampler PVS 4150. These samplers are designed to automatically gather water samples from a variety of sources, including containers, open channels, sewers, and pipes. They can be programmed to take samples at specific intervals, triggered by external events, or even based on the flow rate of the water. Samples can be collected in either composite or discrete systems, depending on your specific needs. All machines are designed to keep the temperature of the deposited liquid at 4°C for up to 24 hours using crushed ice or ice packs until the samples are collected and brought back to the laboratory for analysis.
Advertisement
Advertisement
PVS4100/4120/4150
Portable Samplers
Revision: 4/12
C o p y r i g h t © 2 0 1 1 - 2 0 1 2
C a m p b e l l S c i e n t i f i c , I n c .
Warranty
The PVS4100/4120/4150 Portable Samplers are warranted for thirty-six (36) months subject to this limited warranty:
“PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless otherwise specified in the corresponding
Campbell pricelist or product manual. Products not manufactured, but that are re-sold by Campbell, are warranted only to the limits extended by the original manufacturer. Batteries, fine-wire thermocouples, desiccant, and other consumables have no warranty. Campbell's obligation under this warranty is limited to repairing or replacing (at Campbell's option) defective products, which shall be the sole and exclusive remedy under this warranty. The customer shall assume all costs of removing, reinstalling, and shipping defective products to Campbell. Campbell will return such products by surface carrier prepaid within the continental United States of America. To all other locations, Campbell will return such products best way CIP (Port of Entry)
INCOTERM® 2010, prepaid. This warranty shall not apply to any products which have been subjected to modification, misuse, neglect, improper service, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied. The warranty for installation services performed by Campbell such as programming to customer specifications, electrical connections to products manufactured by Campbell, and product specific training, is part of Campbell’s product warranty. CAMPBELL
EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. Campbell is not liable for any special, indirect, incidental, and/or consequential damages.”
Assistance
Products may not be returned without prior authorization. The following contact information is for US and international customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit www.campbellsci.com
to determine which Campbell Scientific company serves your country.
To obtain a Returned Materials Authorization (RMA), contact CAMPBELL
SCIENTIFIC, INC., phone (435) 227-9000. After an applications engineer determines the nature of the problem, an RMA number will be issued. Please write this number clearly on the outside of the shipping container. Campbell
Scientific's shipping address is:
CAMPBELL SCIENTIFIC, INC.
RMA#_____
815 West 1800 North
Logan, Utah 84321-1784
For all returns, the customer must fill out a "Statement of Product Cleanliness and Decontamination" form and comply with the requirements specified in it.
The form is available from our web site at www.campbellsci.com/repair . A completed form must be either emailed to [email protected]
or faxed to
(435) 227-9106. Campbell Scientific is unable to process any returns until we receive this form. If the form is not received within three days of product receipt or is incomplete, the product will be returned to the customer at the customer's expense. Campbell Scientific reserves the right to refuse service on products that were exposed to contaminants that may cause health or safety concerns for our employees.
PVS4100/4120/4150
Table of Contents
PDF viewers: These page numbers refer to the printed version of this document. Use the
PDF reader bookmarks tab for links to specific sections.
1.
Product Overview......................................................1-1
PVS4100 and PVS4120 Sampler Features............................... 1-2
PVS4150 Sampler Features...................................................... 1-4
Sampler Vacuum System Features........................................... 1-6
Signal Panel.............................................................................. 1-8
PVS4100 Portable Sampler Specifications............................... 1-9
PVS4120 Lightweight Portable Sampler Specifications ........ 1-10
PVS4150 Ultra-Portable Sampler Specifications ................... 1-11
Controller Specifications........................................................ 1-12
Vacuum System Specifications .............................................. 1-13
Sample Container Options...................................................... 1-14
Composite and Discrete Overview......................................... 1-14
Sample Transport Velocity..................................................... 1-15
1.3.8.2 Horizontal/Vertical Combinations.............................. 1-16
1.3.9 Special Systems...................................................................... 1-17
1.3.9.2 MISA Systems – Teflon and Glass............................. 1-17
Portable Sampler Model Selection Guide ........................................ 1-18
2.
Installation .................................................................2-1
Cabinet Positioning ............................................................................ 2-1
Installation Checklist.......................................................................... 2-4
3.
Operation ...................................................................3-1
Operating Sequence ........................................................................... 3-1
Sampling Sequence .................................................................. 3-1
Line Voltage Failure................................................................. 3-2
Operating Instructions........................................................................ 3-2
Sample Volume Adjustment .................................................... 3-2
Liquid Sensing Rod.................................................................. 3-3
Battery: Operating and Backup (optional)................................ 3-4
Battery: Microprocessor ........................................................... 3-5
i
PVS4100/4120/4150 Table of Contents
4.
Maintenance.............................................................. 4-1
General Maintenance.......................................................................... 4-1
Testing System Vacuum..................................................................... 4-1
5.
Programming ............................................................ 5-1
General Programming ........................................................................ 5-1
Guidelines................................................................................. 5-1
5.1.1.2 Real Time Clock ........................................................... 5-1
5.1.1.3 Total Bottles.................................................................. 5-1
Touchpad Keys......................................................................... 5-2
General Terms .......................................................................... 5-4
Quick Start Guide to Programming.................................................... 5-6
Automatic Sampling Program .................................................. 5-6
Taking a Manual Sample .......................................................... 5-7
Viewing Program Parameters ................................................... 5-7
Setting Program Parameters Individually ................................. 5-8
Programming START DELAY.......................................................... 5-8
START DELAY Overview ...................................................... 5-8
START DELAY using Time/Day ............................................ 5-9
START DELAY using Pulse Input ........................................ 5-11
START DELAY using 4-20mA Input.................................... 5-12
START DELAY using External Contact................................ 5-14
START DELAY using Level Control .................................... 5-15
Programming SAMPLE INITIATION............................................. 5-16
SAMPLE INITIATION Overview ......................................... 5-16
SAMPLE INITIATION using Interval Time ......................... 5-17
SAMPLE INITIATION using Pulse Input ............................. 5-19
SAMPLE INITIATION using 4-20mA Input......................... 5-20
SAMPLE INITIATION using External Contact .................... 5-22
Programming PROGRAM TYPE .................................................... 5-23
PROGRAM TYPE Overview................................................. 5-23
PROGRAM TYPE - Composite............................................. 5-24
PROGRAM TYPE - Daily Cycle ........................................... 5-26
PROGRAM TYPE - Daily Cycle for Dual Station ................ 5-27
PROGRAM TYPE - Consecutive........................................... 5-29
PROGRAM TYPE - Multi-Composite................................... 5-31
PROGRAM TYPE - Timed Step............................................ 5-32
Programming OTHER OPTIONS.................................................... 5-34
OTHER OPTIONS Overview ................................................ 5-34
OTHER OPTIONS - Clock .................................................... 5-36
OTHER OPTIONS - Purge Time ........................................... 5-37
OTHER OPTIONS - Pinch Valve .......................................... 5-39
OTHER OPTIONS - Fault Shutdown .................................... 5-40
Viewing Information ........................................................................ 5-42
Viewing Programmed Information......................................... 5-42
Viewing Generated Information ............................................. 5-44
ii
PVS4100/4120/4150 Table of Contents
Appendices
A. Principles of Operation........................................... A-1
B. Parts List.................................................................. B-1
C.
Programming 4-20mA for Flow Proportional
Sampling ............................................................... C-1
List of Figures
1-1. Highlights of the PVS4100 and PVS4120 Samplers ........................... 1-2
1-2. Highlights of the PVS4150 Sampler.................................................... 1-4
1-3. Diagram of the PVS Vacuum System ................................................. 1-6
2-2. External Signal Cable for PVS4100 and PVS4120 ............................. 2-3
3-1. Battery Performance Curve ................................................................. 3-4
List of Tables
1-1. PVS4100 and PVS4120 Sampler Features ......................................... 1-3
1-2. PVS4150 Sampler Features ................................................................. 1-5
1-3. Vacuum System Features .................................................................... 1-7
1-4. PVS4100 Sampler Specifications........................................................ 1-9
1-5. PVS4120 Sampler Specifications...................................................... 1-10
1-6. PVS4150 Sampler Specifications...................................................... 1-11
1-8. Vacuum System Specifications ......................................................... 1-13
1-9. Sample Container Options – PVS4100 and PVS4120....................... 1-14
1-10. Sample Container Options – PVS4150............................................. 1-14
1-13. MISA System Changes.................................................................... 1-17
5-1. Touchpad Button Descriptions ............................................................ 5-2
B-1. PVS Replacement Parts ......................................................................B-1 iii
PVS4100/4120/4150 Table of Contents iv
Section 1. Product Overview
1.1 Introduction
The PVS4100, PVS4120 and PVS4150 Portable Samplers are automatic liquid samplers for water and wastewater applications. PVS Samplers are capable of gathering fluid automatically from a variety of sources, including containers, open channels, sewers, pipes, and any open source of water.
Samplers are designed for reliable unattended sample collection. Portable units are capable of keeping the temperature of the deposited liquid at 4ºC (39.2ºF) for up to 24 hours using crushed ice or ice packs until the samples are gathered and brought back to the laboratory for analysis.
There are a variety of methods for depositing samples. Composite sampling is used where samples are deposited, over time, into one container. Discrete systems are used when multiple bottles are needed. These are also called
“sequential” systems, and involve a stepper with distributor arm which dispenses the liquid into a bottle, then moves to the next bottle.
Operating temperature for portable samplers is 10ºC to 50ºC (50ºF to 122ºF), adaptable down to 0ºC (32ºF) upon request.
Samples can be triggered by a variety of means. The internal clock on the controller can be set to sample based on time/day (e.g. sample every hour).
There are also a variety of external inputs that can be connected to control sampling using the optional external signal cable. Pulse count is useful for sampling after a certain number of pulses have been reached (e.g. using a rain gauge to trigger sampling). The 4-20 mA option is useful for flow-based sampling (e.g. using a flow meter to trigger sampling after a certain volume of water has passed by). External contact is used to control the sampler from another data logger or PLC, and is useful when full external control is desired.
Level control is the option to choose when the application has starts and stops
(e.g. using a float switch to trigger sampling when water is present, then stop sampling when the water drops below the set level).
When sampling is initiated, liquid travels through the intake tube into the metering chamber. The amount of water taken is set mechanically using the liquid sensing rod and the volume control tube, which means sample accuracy is precise every time, usually within +/- 2% or +/- 2ml.
Once the pre-set amount has been reached, all excess liquid is purged from the system, and the sample is dropped into a container. Sample containers range from 500 ml (500 cc or 2 cup) wedges in discrete systems, to 9 liters (2.3
Gallon) containers for composite systems.
Intake tube is offered in either 3/8” (9.5 mm) ID or 5/8” (15.9 mm) ID, and can be either Nylon-reinforced PVC or Teflon-lined PVC. Transport velocity varies depending on height and distance being sampled. For most situations the sampler pulls at over 1.5 m/s (5 ft/sec). For an in-depth speed chart, refer to
Section 1.3.7 Sample Transport Velocity
1-1
1
2
3
4
5
6
Section 1. Product Overview
1.2 Features
1.2.1 PVS4100 and PVS4120 Sampler Features
31.875”
(810 mm)
10
11
12
7
8
9
Diameter
16.375”
(416 mm)
15 16
13
14
17 18 19 20
FIGURE 1-1. Highlights of the PVS4100 and PVS4120 Samplers
1-2
Section 1. Product Overview
TABLE 1-1. PVS4100 and PVS4120 Sampler Features
Number Item
1 Top Handle
2 Enclosure
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
Description
For easy transport.
Intake Hose Slot
Clasps
Signal Panel
Folding Handles
Intake Hose Connection construction. Provides protection from wind and rain. Not submersible.
The hose must lie in the slot for lid to close properly. A secondary notch is provided for lining up the sampler properly.
There are three clasps for each section, for a total of six. The clasps have ringlets for attaching a suspension harness.
Red Light: Reverse Polarity, Green Light: Power
Left Plug: External Signals (to be used with optional signal cable)
Right Plug: Power – 120VAC or 12VDC
Left Fuse: 120VAC, Right Fuse: 12VDC
Toggle Switch: Power On/Off
For detailed chart, see page 1-8.
For easy transport.
The volume control tube is where the intake hose is connected to the sampler. This stainless steel tube is raised or lowered
manually using fitting to set the sample volume (see FIGURE 1-3
Wiring
Nuts to adjust volume
Connects electricity to the rods for sampling. When replacing metering chamber, these rods can be easily disconnected.
To adjust sample volume, twist top nut while holding bottom nut in place. Hand tightening works for many applications, but a wrench is advised.
Metering Chamber
Battery Standard on all samplers. The PVS4100 battery is 15 lbs, 17AH.
The PVS4120 battery (shown) is 4 lbs, 7 AH.
Multi-Function Input Controller This is where sampler is controlled and programmed.
Sinker. Optional Strainer. Keeps the end of the intake tube in the source liquid. Optional strainer can raise collection point above sinker.
Intake Hose
Stepper Assembly
This chamber is where the sample liquid is drawn into before dropping into the final container. The rods inside are raised and lowered to the sample volume desired.
Distributor Arm
Sample Container(s)
Retaining Plate
T-Bar
Standard samplers come with 25 Feet of 3/8” ID PVC tube.
Moves the distributor arm for multiple bottle configurations. This is not present on composite (single bottle) samplers.
Dispenses liquid into bottles. Moves to next bottle after sampling. On composite (single bottle) samplers, this is replaced with discharge tube.
The container(s) that the sample is deposited in can be made from a variety of materials, shapes, and sizes. In discrete samplers, there is a distributor arm that deposits samples into multiple containers.
Holds sample bottles down tightly (discrete samplers only).
Bottle-Guide Notch
This bar is needed to hold the retaining plate down. It is imperative that the bottles do not lift up even a little bit, as they can interfere with the mobility of the distributor arm (discrete samplers only).
This notch is the location of bottle number one. It lines up the middle section of the sampler so that the distributor arm can be placed at the same location.
1-3
6
7
8
9
3
4
5
1
2
Section 1. Product Overview
1.2.2 PVS4150 Sampler Features
14
15
10
11
12
13
16
17
18
14.40”
(366 mm)
24.60”
(625 mm)
19
19.70”
(500 mm)
FIGURE 1-2. Highlights of the PVS4150 Sampler
1-4
Section 1. Product Overview
TABLE 1-2. PVS4150 Sampler Features
Number Item Description
Green Light: Integral Battery Charging (AC power connected)
Red Light: External Battery Reverse Polarity
Toggle Switch: Power On/Off
Plug: AC power (coupled with optional signal cable when supplied).
2 Volume Control Tube This stainless steel tube is raised or lowered manually using the
fitting to set the sample volume (see FIGURE 1-3 on page 1-6).
Multi-Function Input Controller This is where sampler is controlled and programmed. 3
4 Metering This chamber is where the sample liquid is drawn into before dropping into the final container. The rods inside are raised and lowered to the sample volume desired.
This valve shuts during sampling, and then releases once desired liquid has entered the chamber.
8 Latches
Attached to the side of the door with Velcro. During transport of full sample container, this plug screws into the hole where the discharge tube enters the bottle so that the bottle can be carried upright by its handle.
The 2.3 Gal (9L) container that the sample is deposited into is
HDPE, or can be upgraded to PP. Container has a side notch with plug for discharge tube to enter.
4 Press-and-Pull Latches provide airtight seal of sampler for transport. Sampler can operate with door open or closed.
9
10
12
Cavity Space for Cooling Pack This space fits one large Zero-Pak cooling pack, to keep sample cooled for 24 hours.
Quick Connectors Optional connectors for quick connecting and disconnecting.
Connection Intake hose is connected to volume control tube in the absence of quick connectors.
LED Indicator Lights Green light: Power On/Sampling.
Red light: Fault.
13 Enclosure Hardigg HPX® high performance resin. Enclosure is fully watertight and submersible, depending on options (check with factory).
14
15
19
Intake Hose
Sinker. Optional Strainer.
Wheels
Standard samplers come with 25 Feet of 3/8” ID PVC tube.
Keeps the end of the intake tube in the source liquid. Optional strainer can raise collection point above sinker.
16 Telescoping Retractable handle slides up and down for easy transport by pressing in the small lever on the right side.
Four durable soft-grip handles, one on each side, can be folded up or down.
Optional gauge on the side of the enclosure (not shown). This is helpful for knowing what stage the sampler is at when sampling with the door closed.
Two wheels for rolling on smooth surfaces.
1-5
Section 1. Product Overview
1.2.3 Sampler Vacuum System Features
1-6
FIGURE 1-3. Diagram of the PVS Vacuum System
Section 1. Product Overview
Number Item
1
2
5
6
3
4
7
8
9
10
11
12
13
14
15
16
Solenoid Valves
Pump
Touchpad Controller
Sample Distributor
Distributor Arm
Discrete Sample Containers
Pressure Gauge
Liquid Sensing Rod
Barrier Valve
Metering Chamber
Pinch Valve
Cap with “Container Full” Shutoff
Composite Sample Container
Intake Hose
Sinker. Optional Strainer.
TABLE 1-3. Vacuum System Features
Volume Control Tube
Description
Control the air flow from pump to sampler, either purging or sucking.
Located behind a sheet of metal, the pump does not come into contact with any liquid whatsoever. It does all the drawing and purging through using a vacuum and compressor.
Controls sampler program and offers status feedback on LCD.
Rotates distributor arm between multiple discrete containers.
Dispenses liquid from metering chamber into discrete container.
Multiple containers, always in a quantity divisible into 24
(PVS4100 and PVS4120 only).
Visually describes sampling process in terms of vacuum/pressure. Useful for troubleshooting a plugged/kinked line, or signals leaks. Optional (PVS4150 only).
This rod must remain above the volume control tube. When the sample liquid comes into contact with the two rods it signals the controller to stop sampling and begin purging.
Prevents metering chamber overflow in case the liquid sensing rod fails (e.g. completely coated with oils/grease).
Mechanically set the volume required for sample by using a wrench on the fitting at the base of this stainless steel tube.
Sample is drawn into chamber up to level set by volume control tube, then line is purged, followed by dropping sample into containers. Metering Chambers come in glass or acrylic, from
250cc to 500cc.
This valve shuts during sampling, then opens during sampling to drop sample into container, then closes to purge hose.
Optional cap contains Overflow Protection Probes which signal the sampler to halt when container is full. Can be installed in maximum two containers, or into a discrete bottle tray.
A single container to hold sample liquid.
Standard samplers come with 25 Feet of 3/8” ID PVC tube.
Keeps the end of the intake tube in the source liquid. Optional strainer can raise collection point above sinker.
1-7
Section 1. Product Overview
1.2.4 Signal Panel
PVS4100 and PVS4120
Red Light: Reverse Polarity, Green Light: Power
Left Plug: External Signals (to be used with optional signal cable)
Right Plug: Power – 120VAC or 12VDC
Left Fuse: 120VAC, Right Fuse: 12VDC
Toggle Switch: Power On/Off
PVS4150
Green Light: Integral Battery Charging (AC power connected)
Red Light: External Battery Reverse Polarity
Toggle Switch: Power On/Off
Plug: AC power (coupled with optional signal cable when supplied).
1-8
Section 1. Product Overview
1.3 Specifications
1.3.1 PVS4100 Portable Sampler Specifications
TABLE 1-4. PVS4100 Sampler Specifications
Dimensions Height: 809 mm (31.875 in)
Diameter: 428 mm (16.85 in)
Extended Base:
Height: 962 mm (37.875 in)
Diameter: 428 mm (16.85 in)
11.8 kg (26 lbs) Weight
(without battery)
Enclosure
Power
Requirements
Cooling System
Operating
Temperature
Storage
Temperature
Molded medium density linear polyethylene, 3 piece construction, all SS fittings.
Protection Rating: IP 55, Dust protection, water jets.
Sampler: DC Output: 13.6V, 10A. AC Input: 88-
264VAC, 50/60Hz, 2.5A (max 3A)
Integral Battery: 12VDC, 17AH, 15 lbs.
External Receptacle: 12 VDC.
Optional AC only model available (no battery).
Insulated container wall. Cavity space for ice.
Standard: 10ºC to 50ºC (50ºF to 122ºF)
* Can be modified to operate down to 0ºC (32ºF) upon request.
-30ºC to +60ºC (-22ºF to +140ºF)
1-9
Section 1. Product Overview
1.3.2 PVS4120 Lightweight Portable Sampler Specifications
TABLE 1-5. PVS4120 Sampler Specifications
Dimensions Height: 31.875 in (809 mm)
Diameter: 16.85 in (428 mm)
Extended Base:
Height: 37.875 in (962 mm)
Diameter: 16.85 in (428 mm)
10.4 kg [23 lbs] Weight
(without battery)
Enclosure
Power
Requirements
Cooling System
Operating
Temperature
Storage
Temperature
Molded medium density linear polyethylene, 3 piece construction, all SS fittings.
Protection Rating: IP 55, Dust protection, water jets.
Integral Battery: 12 VDC, 7 AH, 4 lbs.
External Receptacle, 12 VDC.
External Charger for 115VAC (optional up to
240VAC).
Optional 120VAC/12VDC power supply with external battery charger.
Insulated container wall. Cavity space for ice.
Standard: 10ºC to 50ºC (50ºF to 122ºF)
* Can be modified to operate down to 0ºC (32ºF) upon request.
-30ºC to +60ºC (-22ºF to +140ºF)
1-10
Section 1. Product Overview
1.3.3 PVS4150 Ultra-Portable Sampler Specifications
TABLE 1-6. PVS4150 Sampler Specifications
Dimensions H: 24.6” x W: 19.7” x D: 14.4”
[H: 625mm x W: 500mm x D: 366mm]
35.5 lbs [16.1 kg] Weight
(without battery)
Enclosure
Power
Requirements
Cooling System
Operating
Temperature
Storage
Temperature
HPX high performance resin. Press & Pull latches, and soft-grip handles.
Protection Rating: IP 67, Dust-tight, water-tight
(depending on options chosen).
Integral Battery: 12 VDC, 7 AH, 4 lbs.
External Charger for 115VAC (optional up to
240VAC).
Cavity space for two Zero-Pack (#12396).
Standard: 10ºC to 50ºC (50ºF to 122ºF)
* Can be modified to operate down to 0ºC (32ºF) upon request.
-30ºC to +60ºC (-22ºF to +140ºF)
1-11
Section 1. Product Overview
1.3.4 Controller Specifications
TABLE 1-7. Controller Specifications
START DELAY Disabled
Time/Day
Pulse Count
4-20mA
No start delay.
Adjustable, up to 1 week in advance.
Adjustable, up to 9,999,999.
Adjustable, up to 9,999,999 (4-20mA = 0-100 Pulses/min).
External Contact
Level Control
SAMPLE INITIATION Disabled
Interval Time
PROGRAM TYPE
CLOCK
PINCH VALVE
PURGE CYCLE
SUCTION CYCLE
Vacuum
Momentary, 25 millisecond dry contact closure.
Adjustable up to 99 second contact duration.
No sample initiation.
Pulse Count
4-20mA
External Contact
Adjustable up to 999 hours, 99 minutes
Adjustable, up to 9,999,999.
Adjustable, up to 9,999,999 (4-20mA = 0-100 Pulses/min).
Momentary, 25 millisecond dry contact closure.
Composite Terminate after up to 9,999,999 samples.
Multi-Composite Adjustable, up to 99 cycles per bottle.
Consecutive
Daily Cycle
Timed Step
Real Time Clock
Sample release
Draw and purge time
Variable
Adjustable, up to 9 bottles per cycle.
Adjustable, up to 9 bottles per day.
Adjustable, up to 99 hours, 99 minutes per step.
Real time operating system.
Adjustable, normally open / normally closed.
Adjustable, 1 to 99 seconds.
Adjusts automatically to double the value of the purge time setting or until liquid contacts level electrode in metering chamber.
System pressure range is -14 psi to +20 psi, which can be shown on the Optional Pressure Gauge.
ALARM OUTPUTS
STATUS OUTPUTS
DIRECT FUNCTION
KEYS
AVAILABLE
DISPLAYS
Independent
Independent
Manual Sample
Manual Purge
Manual Bottle Advance
Restart
Real-Time Clock
Process Timing
Container Full (Latched. Any key resets. NPN*)
Sample Fault (Latched. Any key resets. NPN*)
Cycle Abandoned (Pulsed. NPN*)
*NPN (sinking) – see Technical Appendix for details.
Sample Taken (DC relay driver, sinking)
Samples manually when pressed twice. Does not interrupt program.
Purges system during second press as long as button is pressed.
Moves distributor arm to next bottle.
Re-initiates program when pressed twice.
Elapsed, remaining.
AUTOMATIC
DISPLAYS
Multi-Level
Descriptions
Container Full
Fault
Power Interrupt –
Program Resumed
Sample program complete.
Program not completed.
1-12
Section 1. Product Overview
1.3.5 Vacuum System Specifications
TABLE 1-8. Vacuum System Specifications
Feature Description
Switches Run/Off (SPST Toggle).
Sample Volume Adjustable, 50cc to 500cc [PVS4100 and
PVS4120]
Adjustable, 50cc to 250cc [PVS4150]
Sample Transport
Velocity
Metering Chamber
PVS4100: Minimum of 3 ft/sec at 20 ft of lift (3/8”
ID intake line).
PVS4100: Minimum of 3 ft/sec at 16 ft of lift (5/8”
ID intake line).
PVS4100: Maximum Vertical 27.5 ft.
PVS4120 and PVS4150: Minimum of 3 ft/sec at 13 ft of lift.
PVS4120 and PVS4150: Maximum Vertical: 18 ft
For complete charts, see page 1-15.
500cc capacity, clear acrylic, calibrated in 100cc’s.
500cc capacity, glass (Pyrex), calibrated in 100cc’s.
[OPTIONAL]
250cc capacity, clear acrylic, calibrated in 50cc’s.
[PVS4150]
Metering Chamber
Cover
Nylon
Reinforced Teflon Top. [OPTIONAL]
Volume Control Tube 316 Stainless Steel
316 Stainless Steel Metering Chamber
Level Electrode
Intake Hose
Discharge Hose
Nylon-Reinforced PVC, 3/8” ID (standard 25 ft with sinker)
Nylon-Reinforced PVC, 5/8” ID [OPTIONAL]
Teflon-lined Tygon, 1/2" ID [OPTIONAL]
Teflon-lined Tygon, 3/4" ID [OPTIONAL]
Latex, 3/8” ID
Latex, 5/8” ID [OPTIONAL]
Silicone, 3/8” ID [OPTIONAL]
Silicone, 5/8” ID [OPTIONAL]
Lead Sinker Sinker/Strainer
Stainless Steel Sinker/Strainer [OPTIONAL]
Stainless Steel Sinker [OPTIONAL]
1-13
Section 1. Product Overview
1.3.6 Sample Container Options
TABLE 1-9. Sample Container Options – PVS4100 and PVS4120
Feature Description
Composite (single) containers
9 liter (2.3 US Gal) high density polyethylene (HDPE)
9 liter (2.3 US Gal) polypropylene (PP)
10 liter (2.5 US Gal) Glass [with extended base]
Discrete (multiple) containers
500cc (0.5 L) Wedges (PP) [24 bottles]
500cc (0.5 L) Wedges (PP) [24 bottles] [with extended
base]
1 liter Glass [8 bottles] [removable bottle tray c/w
handles]
1 liter high density polyethylene (HDPE) [8 bottles]
TABLE 1-10. Sample Container Options – PVS4150
Feature Description
Composite (single) containers
9 liter (2.3 US Gal) high density polyethylene (HDPE)
9 liter (2.3 US Gal) polypropylene (PP)
1.3.7 Composite and Discrete Overview
Discrete Sampling (PVS4100 and PVS4120 only) is sampling wherein samples are taken into more than one container. Inside of the cooling chamber is a stepper assembly which revolves 360° and delivers samples into separate containers, ranging from 8 to 24 bottles. Discrete sampling is beneficial in situations where change over time needs to be measured, such as measuring different water characteristics over 24 hours. Labs and monitoring personnel tend to rely on discrete portable sampling.
1-14
Section 1. Product Overview
Composite Sampling is for drawing water samples into one large container.
This is the simplest way of taking samples and typical for most situations where a sampler is set up to measure effluent in one location. It is also significantly less expensive than discrete sampling.
1.3.8 Sample Transport Velocity
TABLE 1-11. Vertical Velocity
3/8” ID Large pump
(BVS 4300, CVS 4200, PVS4100)
3/8" ID Small pump
(PVS4120, PVS4150)
5/8" ID Large pump
0’ 5’ 10’ 15’ 18’ 20’ 22’ 25’ 27’ 28’ Height
7.1 7.1 6.0 5 4.4 4.1 3.6 3 2.6 0 Ft/sec
5
5.1 4.7 3.6 2.7 1.6 0
4.6 3.9 3.1 2.7 1.8 0
1-15
Section 1. Product Overview
3/8” ID Large pump
(BVS 4300, CVS 4200, PVS4100)
3/8" ID Small pump
(PVS4120, PVS4150)
5/8" ID Large pump
TABLE 1-12. Horizontal Velocity
25’ 50’ 75’ 100’ 175’ 200’ 250’ Distance
7.1 6.2 5.6 5.0 4.0 3.7 2.6 Ft/sec
5.1 4.7 4.2 4.1 3.4 3.1 2.3
5.0 4.7 4.3 4.2 3.7 3.3 2.4
1-16
1.3.8.1 Using Velocity to Calculate Purge Time
Purge time of the sampler needs to be programmed based on the length of hose and the velocity at which the liquid will travel through the hose. The formula is
l / v = p (length / velocity = min. purge time). Adding a few second to the purge time is recommended to ensure the line is fully cleared of any obstructions.
Example: 100 ft of hose, at 5 ft/sec, requires a minimum 20 second purge time.
100 / 5 = 20 seconds. The number input for purge time should be a minimum of 20, but preferably 24.
Standard purge time for 25 ft of intake tube is 10 seconds. Although a standard
25 ft hose will sample in less than 4 seconds, 10 seconds is the minimum recommended for proper clearing of the line.
1.3.8.2 Horizontal/Vertical Combinations
The velocity charts above measure only horizontal or only vertical. Most applications will have combinations of both. With 200 feet of intake tubing,
PVS Samplers are capable of drawing a sample above 2 ft/sec at 20 feet of vertical. At 23 feet of vertical with 200 feet of intake tubing, sampling may or may not be successful, depending on altitude and other factors. For more
Section 1. Product Overview detailed information for your specific application, please contact a Campbell
Scientific applications engineer.
1.3.9 Special Systems
1.3.9.1 5/8 Systems
In applications with large particles or materials in the source liquid, a 5/8” ID system will help prevent clogging. The added diameter adds 66% more volume to the entire system. As of 2010, both composite and discrete samplers are available in 5/8”. Smaller pump systems (PVS4120) have minimal capacity running the larger volume of the 5/8” ID so the larger pump systems
(PVS4100) are recommended.
For a sampler to increase to a 5/8” ID, the following parts and components are changed to allow for more volume: intake tube, volume control tube, all fittings, metering chamber, metering chamber lid, discharge tube, sample container cover, sinker or strainer, and the distributor assembly (for discrete models).
1.3.9.2 MISA Systems – Teflon and Glass
In applications wherein the water sample must be prevented from coming into contact with any plastics, a MISA system is recommended. For example, when testing for acid/base/neutral extractable organics and pesticides, the MISA system will keep the final sample clean from any contaminants.
MISA stands for “Municipal/Industrial Strategy for Abatement,” and includes changing all “wetted” components of the sampling system, i.e., everything that comes in contact with the final sample. The chart below outlines the key changes made to the sampler for a MISA system.
TABLE 1-13. MISA System Changes
Intake Tube PVC Teflon-Lined PVC
Metering Chamber Acrylic
Metering Chamber Cover Delrin
Discharge Tube
Sample Container(s)
O-Rings
Latex
HDPE (or PP)
Buna-N (or Viton)
Pyrex
Teflon with Steel Bracing Ring
Silicone
Glass
Silicone
1-17
Section 1. Product Overview
1.4 Portable Sampler Model Selection Guide
PVS Samplers come in a variety of models designed for variations in weight, pump strength, battery, larger intake hose (5/8” ID), ease-of-transport, fitting in manholes, discrete or composite sampling, protection rating for dust and water, signal options, and budgets.
1-18
Section 2. Installation
2.1 Cabinet Positioning
Place the sampler on a level surface as close as practical to the sample source.
Refer to the speed charts for maximum expected lift and draw. Manhole installation may require 3-point suspension harness.
Sampler must be located above sample source, or liquid will flood the machine.
For situations where this is not possible, please contact a Campbell Scientific application engineer for solutions on pressurized sources.
FIGURE 2-1. Sampler Installation
2-1
Section 2. Installation
2.2 Intake Hose
Ensure the intake hose is submerged at all times throughout different flow velocities.
CAUTION Twenty-five feet of intake hose is provided with the sampler. Shortening the hose is not recommended since this length of hose provides sufficient back pressure to the metering chamber, allowing the pump to efficiently expel all solids into the sampler container. Coil any excess intake hose in a manner to provide natural drainage away from the sampler. To install intake hose, heat the end in hot water and slip over volume control tube. Secure with hose clamp provided.
Sample line should be routed so that it has a near continuous slope from the sampler to the source liquid. This will help keep the sample line cleared and fully drained. All excess line should be coiled neatly and without any vertical loops.
2.3 Sinker / Strainer
Sinker or sinker/strainer is intended to keep sample line fully submerged in the source liquid. A sinker/strainer should be used in situations with material that clogs up a normal sinker, or where the standard sinker could stir up bottom sediment.
2-2
Stainless Steel Sinker/Strainer: SAM-23-28-11
2.4 Storage
If the sampler is not to be used for an extended period of time, store the unit in an upright position in a warm, dry location. If the unit has an integral battery,
recharge the unit prior to storage.
Acceptable storage temperature: -30ºC to +60ºC (-22ºF to +140ºF)
Section 2. Installation
2.5 Signal Wiring
External input capabilities of the sampler are implemented by the use of an optional external signal cable plugged into an external receptacle on the side of the sampler. External inputs of different types can be used by choosing the appropriate pair of wires in the cable.
On the PVS4150 the signal cable is combined with the AC Power cable, thus limiting the number of connections possible.
FIGURE 2-2. External Signal Cable for PVS4100 and PVS4120
IMPORTANT Wiring to remote/external functions should AVOID ALL AC
POWER LINES if possible and/or be run in shielded cable terminating the shield at the AC ground terminal at the remote site.
2-3
Section 2. Installation
2.6 Installation Checklist
Check the following items prior to use of sampler:
1) Sampler is mounted securely and level.
2) Intake Hose: - Free of kinks.
- Properly installed into liquid.
- Properly connected to volume control tube on
3) Discharge hose: - Free of kinks.
- Natural downward slope to sample container.
- Properly connected to (or in) sample container.
4) Power requirements: - Check battery electrical condition.
- Recharge if necessary.
2-4
Section 3. Operation
3.1 Operating Sequence
3.1.1 Sampling Sequence
SAMPLING PROCESS:
1. High pressure air purge of intake hose.
2. Liquid is drawn into the metering chamber, up to the liquid sensing rod.
3. All excess liquid is purged from the system down to the level set by the volume control tube.
4. The sample is then released into either one composite container or one of several discrete containers.
The sampling sequence begins with a high pressure air purge of the intake assembly to remove residual liquid and obstructions. Upon completion of the pre-purge cycle, the system converts to a vacuum state, drawing the sample through the intake hose into the metering chamber. The system then
pressurizes, ejecting excess fluid back through the intake line until the predetermined sample volume is achieved. The sample is then deposited under pressure into the sample container while the post purge again clears the intake line of any residual liquid.
3-1
Section 3. Operation
Should the sampler, for any reason, not be able to draw a sufficient volume of fluid to obtain a sample, the unit automatically initiates a second attempt.
Should a sample still not be delivered, the sequence will be abandoned and the unit will await the next initiation. Upon two consecutive failures, the sampler will suspend the sampling program until manually RESTARTed.
If programmed with the FAULT SHUTDOWN “disabled”, the sampler will not make a second attempt to draw the sample, but will simply abandon it and await the next sample initiation. Neither will the unit suspend the sampling program after consecutive failures. This function is provided for use in the event that the sample source may be lacking sufficient fluid from which to draw, for a period of time , yet allows the sampler to continue operating without a “FAULT SHUTDOWN” occurring. The second attempt is not made to prevent unnecessary wear on the sampler.
3.1.2 Line Voltage Failure
Should the sampler have a factory installed internal battery or have an external battery connected, the sampler will continue operating. The duration of operation will depend on the capability and charge level of either battery. The frequency and the length of each sample cycle will also have an impact on how long the batteries will last.
3.2 Operating Instructions
3.2.1 Sample Volume Adjustment
3-2
Setting the desired sample volume is accomplished by adjusting the height of the volume control tube within the metering chamber. The tube is mounted through the top of the chamber with a gland nut fixing the position. To adjust the sample volume, loosen the nut until the volume control tube may be moved freely. Raise or lower the bottom end of the tube to the desired volume using the lines provided on the side of the chamber as a guide (lines are spaced at 100 cc intervals with the exception of one at 50 cc). Tighten the gland nut to hold the volume control tube at the desired position.
Section 3. Operation
IMPORTANT
NOTE
The volume control tube should always be located below the liquid sensing rod.
Hold the bottom nut while loosening / tightening the top nut, or it may become loosened from the metering chamber cover and create an imperceptible leak in the vacuum system.
3.2.2 Liquid Sensing Rod
This probe, also called the “level control rod”, is used to stop the sample intake. Always ensure that its lower end is located above the volume control tube. Approximately 1” difference is sufficient. If the fluid intake is turbulent within the metering chamber, more than 1” may be required to ensure splashing of fluid does not trigger probe.
In applications with substantial oil or grease, the rods can become coated and lose their conductivity. This is prevented by cleaning the rods regularly. In extreme cases, extra SS wire can be wrapped around the liquid sensing rod to increase its surface area.
IMPERATIVE: THE LIQUID SENSING ROD AND VOLUME CONTROL
TUBES MUST BE KEPT CLEAN TO ENSURE CONDUCTIVITY
NECESSARY TO DETECT THE PRESENCE OF THE FLUID.
Most PVS Samplers incorporate a Barrier Valve in the metering chamber cover, where the tubing from the pump enters. It consists of a cage containing a ball that will float if the sample should rise to the top of the chamber without detection. Should rod conductivity fail, the fluid brings the float into contact with an O-ring surrounding the pressure / vacuum port, sealing the entry to the tubing and the pump (where the fluid may cause serious damage). This O-ring
Barrier Valve should be inspected regularly and replaced as necessary.
Due to the restriction of Wetted Materials (i.e. stainless steel, glass and fluorocarbons etc.), some models of the sampler do not contain this barrier valve. In these units, a secondary liquid-sensing circuit may be added as a precaution. This circuit is connected to the pump tubing fitting on the Metering
Chamber cover.
3-3
Section 3. Operation
3.3 Battery
3.3.1 Battery: Operating and Backup (optional)
CHARGING AND REVERSE POLARITY PROTECTION
The sampler will charge ONLY the factory installed internal battery. This charging takes place continually as long as there is incoming line power.
Should the need arise to only charge the internal battery, as would be required to store the sampler for an extended period of time, simply place the “ RUN /
OFF “ toggle switch in the OFF position, and leave the sampler power breaker on. Twenty-four (24) hours should be sufficient to fully charge the battery.
The sampler is equipped with REVERSE POLARITY PROTECTION for checking the connection of an external battery. When attaching an external battery, be sure to check the reverse polarity indicator. If it is ON, reverse the connections at the battery.
3-4
FIGURE 3-1. Battery Performance Curve
Section 3. Operation
3.3.2 Battery: Microprocessor
SAMPLER CONTROLLER BACKUP BATTERY
The controller contains a 1/2AA, 3.6V lithium backup battery to maintain user settings during loss of system power. If power is removed for any reason, the controller will start a planned shutdown procedure which will save all user settings while its operating voltage is reduced from 5V to approximately 3.3V.
The rate at which this voltage drops is slowed by the presence of a supercapacitor. By the time the voltage has reached 3.3V, the controller has safely stored all user settings and entered a “sleep” mode. This is an extremely low-power mode which is maintained by a trickle of current from the lithium battery, and can be maintained for many years under normal circumstances.
The battery is located on the top left hand side of the controller. It is accessible by the removal of the clear cover, and should be changed under powered
conditions. Since the controller is a low-power device, this uncovering can be safely done, taking care that no conductive implement contacts sensitive circuit components.
If the controller starts to exhibit certain operating anomalies such as loss of user settings after sustained power outages or an inability to “wake up” after a normal shutdown, it may be due to a low or totally discharged backup battery.
To predict the probability of these events, regular examination of the battery
condition is encouraged. The battery status is easily determined while the controller is active. The process will not affect a running program. Battery status can be checked by use of the following touchpad sequence:
VIEW, OTHER OPTIONS, select MAINTENANCE, ENTER, then select B/U BATTERY TEST, ENTER
The display will then show “PASSED”, “LOW” or “FAULT”. The latter two require battery replacement.
TO REPLACE BATTERY
1. Make sure the controller is powered.
2. Remove touchpad (clear) cover from the controller, remembering to handle internal ribbon cable and connector with care.
3. Locate the battery holder on the normal left side of the circuit board. The battery is a 1/2” cylinder about 1” in length. The positive (+) end of the battery has a raised button. Note the button’s relative position in the holder. It should be pointing away from the display side of the board.
4. Place new battery in holder, noting position of button with respect to the polarity indicators in the holder.
5. Check battery status, as above. If necessary, locate small white button at top of the control board (under display) and push to restore factory defaults.
6. Replace cover securely and re-enter user settings.
3-5
Section 3. Operation
3.4 Test Procedure
1. Set volume control tube to 200 cc.
2. Set level probe 1” above bottom of volume control tube.
3. Turn on power. Place the “RUN/OFF” switch in the “RUN” position.
After an initial delay of 15 to 20 seconds, the display will show a two line message, the top line displaying SAMPLER HALTED and an alternating message on the second line displaying why the sampling procedure was interrupted as well as the event time and date.
4. Enter the following sampling program: a) Set purge time to 10 seconds. See OTHER OPTIONS b) Set interval time to 2 minutes. See SAMPLE INITIATION c) Set program type to composite. See PROGRAM TYPE d) Set to terminate after 2 samples. e) Press RESTART, RESTART (to confirm)
5. View the following displays: a) Samples taken - should read 0 b) Remaining time - should be counting down from 2 minutes.
6. Sampling should begin when remaining time indicator reaches 0.
7. Upon completion of sample, view the following displays: a) Samples taken - should read 1. b) Remaining time - should be counting down from 5 minutes.
8. Press MANUAL PURGE. Press again to confirm.
9. Press MANUAL ADVANCE. Press again to confirm.
10. Press MANUAL SAMPLE. Press again to confirm.
11. If equipped with sample container full option, short circuit level probes in container (no dangerous voltage present - 16 Vdc). The message
“SAMPLER HALTED External Stop” should appear on the display, the bottom line flashing.
12. Press RESTART, RESTART, the message “RESTART <Completed>” should appear on the display.
3.5 Troubleshooting
SAMPLER INOPERATIVE: Check supply voltage.
POWER ON BUT PUMP WILL NOT START: Check wiring from sampler controller to pump. Ensure controller is properly connected into harness. a) Pump defective. b) Sampler controller defective.
3-6
Section 3. Operation
SAMPLER WILL NOT TAKE TIMED SAMPLE: a) Sampler controller defective.
SAMPLER WILL NOT INITIATE FROM AN EXTERNAL
CONTACT: Check wiring from terminal strip to sampler controller plug. (Terminals 12 & 13) a) Sampler controller is defective. b) Sampler controller not programmed for External Contact input.
PUMP IS OPERATING, NO AIR PURGE OF INTAKE LINE: Check for blockage of intake hose by removing hose from the metering chamber volume control tube. Initiate manual sample and check for pressure/vacuum throughout sample cycle. a) If pressure/vacuum is present throughout sample cycle, intake hose is plugged. b) Pinch valve may not be closing the discharge hose with sufficient force to ensure an adequate seal. Increase tension by tightening the lock nuts on the pinch valve tension springs and/or replace discharge hose. c) Check for disconnected air lines from pump to metering chamber. d) Check for loose gland nuts. e) Pump flapper valves defective.
PURGE CYCLE OPERATIVE, NO SUCTION: Pinch valve may not be closing the discharge hose with sufficient force to ensure an adequate seal.
Increase tension by tightening the lock nuts on the pinch valve tension springs and/or replace discharge hose. a) Check air lines, metering chamber O-rings and fittings for leakage. b) Solenoid valve clogged or not working. c) Intake tube, not below water level.
SAMPLER HAS HAD AN “EXTERNAL STOP”: Contact not supplied via terminal block. a) Sample container Full Level Probe has been triggered.
LEVEL SENSING PROBE INOPERATIVE: Check wire contact connections on volume control tube and level sensing probe. Check wiring to the sampler controller plug. a) Sampler controller defective b) Coating on probe and/or Volume Control Tube.
3-7
Section 3. Operation
3-8
Section 4. Maintenance
The following maintenance procedure should be performed at regular
intervals:
4.1 General Maintenance
1. Disconnect power.
2. Open metering chamber by removing wing nuts and chamber cover.
3. Clean volume control tube and level sensing probe with mild detergent.
Alternatively, exchange tube and probe with clean set. Do not use any cleaner which may be harmful to the metering chamber cover. Do not use solvents such as acetone, benzene, carbon tetrachloride or lacquer thinners.
Grease and oil may be removed with kerosene or aliphatic naphtha (nonaromatic).
4. Check and clean O-rings in metering chamber cover.
Replace if damaged, worn or brittle.
5. Clean metering chamber using mild detergent.
Do not use any cleaner which may be harmful to the clear acrylic, (e.g. petrochemical solvents, as noted above.) Do not use abrasives or
“scouring” compounds.
6. Check discharge tubing for wear and replace as necessary.
7. Check pinch valve to ensure free movement.
8. If possible, run sampler through several sampling sequences in clean water.
4.2 Testing System Vacuum
Using the (optional) built-in pressure / vacuum gauge, take a reading to ensure system has no leaks. Optimal pressure is above 28 psi. Optimal vacuum should be 12 psi or better.
If the system is not performing at its peak, try the following:
1. Check intake hose for leaks/kinks.
2. Check discharge tube, ensure it has no leaks and is in good shape.
3. Check all fittings to ensure they are tight.
4. Make sure when tightening and loosening the gland nuts on the top of the metering chamber that the bottom nut is held secure and does not move on the cover. Ensure the top nut is securely tightened, and use a wrench if necessary.
4-1
Section 4. Maintenance
5. If still system is still not performing at its peak, inspect pump and all pump tubing.
4-2
Section 5. Programming
5.1 General Programming
5.1.1 Guidelines
Controller settings may be changed at any time. Changes are termed NEW
ENTRIES. No NEW ENTRIES will be acted upon unless the controller is
RESTARTed. Once RESTARTed, all NEW ENTRIES become ACTIVE
SETTINGS.
Every time the controller is RESTARTed, all accumulators (i.e. SAMPLES
TAKEN, TIME REMAINING, REMAINING PULSES, etc.) are cleared and the ACTIVE SETTINGS are reloaded unless NEW ENTRIES have been made.
Remember - Start Delay is reloaded too !!
5.1.1.1 Flashing Text
Flashing text is the system wide prompt that indicates an input is required from the user. Flashing words or duel flashing digits prompt for arrow keys to be pressed to scroll through available options. A single flashing digit prompts for a numeric key to be pressed. When the desired option or number is shown on the display, press the ENTER key.
5.1.1.2 Real Time Clock
The controller has two basic timing modes. The simplest of these requires no maintenance; it simply provides a “heartbeat” for various timed functions. The other timing mode is the REAL TIME CLOCK that is used in several functions and must be correctly set. This is likely the first item requiring
programming. Although time may have been set at the factory, time zone shifts may require adjustment of the Real Time Clock.
5.1.1.3 Total Bottles
Since the number of bottles is usually determined by customer requirement at the time of purchase, this variable will normally be set at the factory to match the actual container hardware. Choices are restricted to a single container (as in composite) or 2, 3, 4, 6, 8, 12 or 24. These all form instructions to the stepper motor in how it will behave when the internal command is given to step to the next container (as each step increment is 15°).
5-1
Section 5. Programming
5.1.2 Touchpad Keys
TABLE 5-1. Touchpad Button Descriptions
Button Description
The VIEW key is used to review alterable parameters currently in use. It has no effect on the program being executed at the time. Once pressed, the user is prompted for a FUNCTION to be viewed. The parameters visible under the function can be stepped through using the ENTER key.
The SET key is used to change program settings or the entire sampling program. Changes made have no effect on the program being executed at the time until the RESTART key is pressed twice.
To leave a programming sequence before entering it in memory either press SET or VIEW and the sequence is aborted.
The ENTER key is used to complete either a VIEW or SET sequence, where sub-menu items are available. Under the control of the VIEW key, parameters are scrolled onto the display, changing with each use of the ENTER key until a complete display of the parameter is completed. Under the control of the SET key, parameters can be displayed, with the added ability to change their values, using the ENTER key to accept the new value until the entire parameter is displayed. (Note: New values are not operational at this time.)
5-2
Section 5. Programming
TABLE 5-1. Touchpad Button Descriptions
Button Description
The RESTART key is used to load any new parameters into the operating program. Pressing it twice will initialize the program and terminate any existing sample program. Any parameters altered under the SET command are updated to the active program. If no parameters have been changed, the program is reset to its first instruction and the same sampler program is started again.
This key requires a confirming second activation to complete its function. WARNING: Any program in progress is ended and all data is lost.
SAMPLES TAKEN [VIEW]. The total number of samples taken can be shown on the display.
REMAINING PULSES [VIEW]. In modes using internal or external pulse counting, the current status of the pulse count can be displayed.
REMAINING TIME [VIEW]. Various views are available dependent on the method used to gather samples. Program variables will determine whether the displayed time is REMAINING
TIME, ELAPSED TIME or START DELAY.
START DELAY [VIEW/SET]. The start of a sample program can be made to occur at a fixed time or event. Options: DISABLE, TIME/DAY, PULSE INPUT, 4-20mA INPUT, EXTERNAL
CONTACT, LEVEL CONTROL.
SAMPLE INITIATION [VIEW/SET]. A sample program may be initiated and controlled by various internal and external parameters. These parameters determine how the program will begin its actions and how the results will be recorded. Options: DISABLE, INTERVAL TIME, PULSE
INPUT, 4-20mA INPUT, EXTERNAL CONTACT.
PROGRAM TYPE [VIEW/SET]. A sample program can be made to collect samples in a fixed style so that the results are useable in different ways. The type of program used may be hardware dependent. This will determine the sampler’s ability to collect and store the desired samples.
Options: COMPOSITE, DAILY CYCLE, CONSECUTIVE, MULTI-COMPOSITE, TIMED
STEP.
ACTIVE SETTINGS [VIEW]. Current sample program parameters can be reviewed by scrolling through them using the ENTER key as a toggle.
NEW ENTRIES [SET]. Program all major program settings at once (including START DELAY,
SAMPLE INITIATION, PROGRAM TYPE, and PURGE TIME).
[VIEW]. Review parameters that have been changed since the sample program was started (only if the changes have been properly ENTERED). Scroll through them using the ENTER key as a toggle.
OTHER OPTIONS [VIEW/SET]. Various options relating to equipment and information retrieval are available under this key. Changes in equipment setup can be entered here, and certain status information is also available here. Options: CLOCK, PURGE TIME, PINCH VALVE, FAULT
SHUTDOWN, SAMPLER STATUS, CYCLES ABANDONED, BOTTLE POSITION,
MAINTENANCE.
5-3
Section 5. Programming
TABLE 5-1. Touchpad Button Descriptions
Button Description
MANUAL PURGE. Purges the intake line independent of program control, as long as a programmed cycle has not started. Sampler starts its pump, creating pressure in the sample intake tube to purge it of any excess material that may be present. Button must be pressed twice to purge line. Sustained pressure on the key during the second press will cause purging to continue until the key is released.
MANUAL ADVANCE. Distributor arm advances one position (e.g. to next bottle), dependent on the equipment available (discrete samplers only). This action is NOT updated to any current sampler program. Button must be pressed twice to initiate manual advance.
MANUAL SAMPLE. Initiate a single Sample Cycle. Sampler must not be engaged in a sampling event at the time. This action and any resulting sample collected are NOT updated to any current sampler program. The Bottle Position is NOT advanced. Program will continue uninterrupted.
Button must be pressed twice to initiate manual sample. Whether successful or not, the display will read “MANUAL SAMPLE Completed”.
5.1.3 General Terms
Many of the functions available on the Touchpad have a variety of options to enhance their capabilities. These options are programmable from the Touchpad and require only that the sampler have the correct equipment configuration to utilize them.
DISABLE
The display showing disabled will reflect the status of any function not being used.
TIME/DAY
The basis for several timed functions is the Real Time Operating System. Time
(of) Day will be a means of setting the timing period for the START DELAY function. The format is on a weekly basis, requiring hour, minute, AM/PM and day inputs (HH:MM AM SUN). This means the Start Delay can be set to any particular minute in a week.
INTERVAL TIME
Sampler operation can be controlled by fixed time intervals which do not require Time/Day setting. SAMPLE INITIATION has an option whereby an interval time can be set between sample cycles. The controller will cause samples to be taken on a timed interval basis, continuing until the sample program is completed by a full jar or operator intercession.
PULSE INPUT
This option will allow the controller to determine the sampler operation based on external criteria. Pulses fed to an internal accumulator in the controller will be compared to the setting entered by the operator and will cause a sample cycle to start. The accumulator will reset immediately and counting of pulses
5-4
Section 5. Programming will begin again. There is no loss of count the sample cycle. Pulse requirements of the system are detailed in the specifications.
4-20mA INPUT
Where external devices do not themselves generate pulses in any relation to their process but generate a current signal of 4-20 mA, this input option will generate internal pulses proportional to the incoming 4-20 mA signal. These can then be treated the same as the Pulse Input option and accumulated in the controller to determine when a sample cycle will occur.
EXTERNAL CONTACT
The sampler controller can react to an external dry contact, otherwise known as a zero-voltage contact, to activate a sample cycle on demand. This will generally be when external conditions have caused a relay to close, requiring a sample be taken at that time.
LEVEL CONTROL
The START DELAY function is a special case of the external contact option. where the contact signal is required to be present for a pre-programmed time.
This enables verification of the signal where fluctuations may occur in the level which would trigger samples at unwanted times. This is the only case in which
the START DELAY is not a single timed event. The operation of the sampler after the level signal is verified will be controlled by whatever function is set in the SAMPLE INITIATION. It will continue until the level drops or the function is terminated by the controller. If the level drops before the function is finished, any sample cycle already in progress will be completed and then the system will shut down until the next verified level control signal.
COMPOSITE
A program option which determines that all the samples that are gathered will be placed in a single container. The sample program terminates after a specific number of samples.
MULTI-COMPOSITE
This option is used for discrete sampling applications, to deposit multiple samples to one container before advancing the distributor mechanism to the next container. The interval between each sample is controlled by the
SAMPLE INITIATION options. The multi-composite setting is programmable up to 99 samples per container, for up to 24 containers depending on the hardware configuration.
CONSECUTIVE
This option is used for discrete sampling applications, to successively deposit one sample to each of a programmed number of containers on any given sample initiation. The consecutive setting is programmable up to 99 containers per sample initiation, although this may be severely limited by hardware configuration.
5-5
Section 5. Programming
DAILY CYCLE
Allows the sampler to deposit equal sample volumes into a predetermined number of containers per programmed day. Each day may have any number of samples taken, dependent on the SAMPLE INITIATION mode chosen.
Deposits are made to as many as 9 containers per day, to a cumulative total of
24 containers. (e.g. Choosing a 24 bottle format, the sampler may be programmed to deposit to Three (3) bottles on any six days of the week, together with up to Six (6) bottles on the seventh). Timing is dependent on the crystal-controlled Real Time Clock in the controller. The first program day will be the current day the programming is done, unless the START DELAY option is chosen to determine when sampling will begin.
TIMED STEP (Override)
This option will cause the sample distributor to step to a new container
regardless of the status of the SAMPLE INITIATION setting. For example, the actual sampling may be under the control of a flowmeter and taking samples based on the flow rate as determined by pulses or 4-20mA input to the controller. When the user-programmed Timed interval has elapsed, the controller will Step to a new container. The Timed Step can be set for any interval up to 99 hours 59 minutes. Progress of the step timer can be viewed by selecting View, Program Type and pressing ENTER twice. Thus you may view the REMAINING TIME or the step timing.
5.2 Quick Start Guide to Programming
5.2.1 Automatic Sampling Program
To begin a new, quick program:
Press “SET”
Press “NEW ENTRIES” . Press “ENTER”
START DELAY (how you will be delaying the sample program until certain external conditions are met). Select, using arrows, which parameter you would like, and adjust settings (see 5.3
Programming Start Delay). Options: DISABLE;
TIME/DAY; PULSE INPUT; 4-20mA INPUT;
EXTERNAL CONTACT; LEVEL CONTROL.
Press “ENTER” twice
5-6
Section 5. Programming
SAMPLE INITIATION (parameters for frequency of samples). Select, using arrows, which parameter you would like, and adjust settings (see 5.4
Programming Sample Initiation). Options:
DISABLE; INTERVAL TIME; PULSE INPUT; 4-
20mA INPUT; EXTERNAL CONTACT.
Press “ENTER” twice.
PROGRAM TYPE (which type of sampling program). Select, using arrows, which parameter you would like, and adjust settings (see 5.5
Programming Program Type). Options:
COMPOSITE; MULTI-COMPOSITE;
CONSECUTIVE; DAILY CYCLE; TIMED STEP
(override).
Press “ENTER” twice.
PURGE TIME (set how long sampler will purge between samples, minimum of 10 seconds). Using # keys, enter the purge time needed for application
(e.g. 100 ft draw at 5 ft/sec = 20 sec). Press
“ENTER”.
Press “RESTART” twice.
Sampling is ready to go.
5.2.2 Taking a Manual Sample
To take a sample manually, simply press the
“Manual Sample” button twice. Manual samples will not interrupt the current automatic sampling program.
5.2.3 Viewing Program Parameters
To view the program or remaining time, press the “VIEW” button, followed by the button representing what you want to see, e.g. “REMAINING TIME”.
5-7
Section 5. Programming
5.2.4 Setting Program Parameters Individually
To modify any of the settings individually press the “SET” button followed by the appropriate button based on what parameter is being changed.
5.3 Programming START DELAY
5.3.1 START DELAY Overview
START DELAY is the function which will delay the beginning of a sample program until certain external conditions are met. Upon meeting those conditions, the sampler will initiate a sample cycle and then operate based on the SAMPLE INITIATION parameters. Under START DELAY, flashing text prompts the user to scroll through available options by pressing arrow keys. These options only require that the correct equipment is present to utilize them. The last option selected in previous programming will be the first to appear on the display.
The display showing disabled will reflect the fact that the function is not being used.
The basis for several timed functions is the
Real Time Operating System. Time (of) Day will be a means of setting the timing period for the START DELAY function. The format is on a weekly basis, requiring hour, minute, AM/PM and day inputs (HH:MM
AM SUN). This means the start of a sampler program can be delayed up to seven days.
This option will allow the controller to determine the sampler’s start of operation based on external pulses. Pulses fed to an internal accumulator in the controller will be compared to the setting entered by the operator. Pulse requirements of the system are detailed in the specifications.
Where external devices do not themselves generate pulses in any relation to their process but generate a current signal of
4-20mA, this input option will generate internal pulses proportional to the incoming
4-20mA signal. These can then be treated the same as the Pulse Input option and accumulated in the controller to determine when a sample program should start.
5-8
Section 5. Programming
The sampler controller can react to an external dry contact, otherwise known as a zero-voltage contact, to activate a sample program on demand. This will generally be when external conditions have caused a relay to close, requiring a sample program be started at that time.
This option is a special case of the external
contact option. The key difference is that the contact closure must be present for a pre-programmed time, thus enabling verification of the signal. This will accommodate fluctuations as seen in a level switch, thereby avoiding triggering of samples at unwanted times. This is the only
case in which the START DELAY is not a
single timed event. Should the contact open for the same pre-programmed time, the sampler will (after completing any sample cycle already in progress), halt the sampling initiation and await the next verified signal.
At this time, the sampling program will resume.
5.3.2 START DELAY using Time/Day
The following sequence of entries are made on the Touchpad to create a future starting time for the operation of the Sampler. The
ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the START DELAY key.
3. Press an ARROW key. Continue until
Time/Day is shown on the display.
4. Press the ENTER key.
5-9
Section 5. Programming
When setting the time, a single flashing digit indicates an input from a numeric key is required. Press a number key to enter a value. The next digits flash in succession.
Enter each as required. The format is
HH:MM.
When the four digits are entered, press
ENTER. Any wrong entries will require re- entry. There are two methods of correcting a mistake. The digit flashing “wraps around” and begins again, at which time the correct entry may be pressed. Alternately, the arrow keys can be used to reposition the flashing prompt over the error, which can then be replaced with the correct value.
The flashing prompt advances to the
AM/PM indicator. Press the ARROW key until the right indicator is shown. Press
ENTER.
The flashing prompt advances to the day indicator. Press the ARROW key repeatedly until the correct day appears. Press ENTER.
The display will echo the last entry with
<ENTERED>.
The controller has now been given a new value for START DELAY. The new values reside in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice. The controller will then wait until the designated time before starting its sampling program.
Summary of Sequence:
SET, START DELAY, ARROW(S), ENTER, #, #, #, #, ENTER,
ARROW(S), ENTER, ARROW(S), ENTER, RESTART, RESTART.
5-10
Section 5. Programming
5.3.3 START DELAY using Pulse Input
The following sequence of entries are made on the Touchpad to create a future starting time for the operation of the Sampler. The
ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the START DELAY key.
3. Press an ARROW key. Continue until
Pulse Input is shown on the display.
4. Press the ENTER key.
The display will show a new screen containing the option title PULSE INPUT on the top line and a 7 digit number with the leftmost digit flashing as an input prompt.
To set the number of pulses required to be input before a sample program is started, use the ARROW keys to move the flashing prompt until it is over the digit requiring change.
Press a number key (0-9) to replace any existing number and advance the flashing prompt to the next digit to change. Replace digits as required, then press ENTER.
The display will echo the last entry with
<ENTERED>.
Any wrong entries will require re-entry. If
ENTER has not been pressed, reposition the prompt over the incorrect digit and replace it. After ENTER has been pressed, the entire entry must be redone from the beginning
(press SET).
5-11
Section 5. Programming
The controller has now been given a new value for START DELAY. The new values reside in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice. The controller will then wait until the required pulses have been received before starting its sampling program.
Summary of Sequence:
SET, START DELAY, ARROW(S), ENTER, #######, ENTER,
RESTART, RESTART.
5.3.4 START DELAY using 4-20mA Input
The following sequence of entries are made on the Touchpad to create a future starting time for the operation of the Sampler. The
ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the START DELAY key.
3. Press an ARROW key. Continue until 4-
20mA Input is shown on the display.
4. Press the ENTER key.
5-12
Section 5. Programming
The display will show a new screen containing the option title 4-20mA INPUT on the top line and a 7 digit number with the leftmost digit flashing as a prompt for input.
The 4-20 mA input will be converted by the controller to pulses, proportional to the span of the input, at the rate set in the specifications. To set the number of pulses required to be input before a sample program is started, use the ARROW keys to move the flashing prompt until it is over the digit requiring change.
Press a number key (0-9) to replace any existing number and advance the flashing prompt to the next digit to change. Replace digits as required, then press ENTER.
The display will echo the last entry with
<ENTERED>.
Any wrong entries will require re-entry. If
ENTER has not been pressed, reposition the flashing prompt over the incorrect digit and replace it. After ENTER has been pressed, the entire entry must be redone from the beginning (press SET).
The controller has now been given a new value for START DELAY. The new values reside in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice. The controller will then wait until the required pulses have been received before starting its sampling program.
Summary of Sequence:
SET, START DELAY, ARROW(S), ENTER, #######, ENTER,
RESTART, RESTART.
5-13
Section 5. Programming
5.3.5 START DELAY using External Contact
The following sequence of entries are made on the Touchpad to create a future starting time for the operation of the Sampler. The
ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the START DELAY key.
3. Press an ARROW key. Continue until
External Contact is shown on the display.
4. Press the ENTER key.
The display will echo the last entry with
<ENTERED>.
The controller has now been given a new value for START DELAY. The new value resides in the NEW ENTRIES area of the controller memory. To make this change active, press the RESTART key, then again press it to confirm your choice. The controller will then wait until it receives a contact closure (at the External Start inputs on the terminal block) before starting its sampling program.
Summary of Sequence:
SET, START DELAY, ARROW(S), ENTER, RESTART, RESTART.
5-14
Section 5. Programming
5.3.6 START DELAY using Level Control
The following sequence of entries are made on the Touchpad to create a future starting time for the operation of the Sampler. The
ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the START DELAY key.
3. Press an ARROW key. Continue until
Level Control is shown on the display.
4. Press the ENTER key.
The display will change to read “Minimum
Contact Time: 01 seconds”. The actual time shown may be any two digit number. The first digit will be flashing as a prompt for input.
Press a number key (0-9). The number will replace the current number and advance the flashing prompt to the next digit. Press a second number key (0-9). Repeat this procedure if number is wrong, until desired time is displayed. Press the ENTER key.
The display will echo the last entry with
START DELAY, <ENTERED>.
5-15
Section 5. Programming
The controller has now been given a new value for START DELAY. The new values reside in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice. The controller will then wait until it receives a contact closure (at the External Start inputs on the terminal block). The contact must remain closed for the length of time programmed in the steps above.
Summary of Sequence:
SET, START DELAY, ARROW(S), ENTER, ##, ENTER, RESTART,
RESTART.
5.4 Programming SAMPLE INITIATION
5.4.1 SAMPLE INITIATION Overview
SAMPLE INITIATION is the function that will determine the frequency that samples are drawn. There is available on the Touchpad a variety of options to enhance the capabilities of this function. When the Sample Initiation has been chosen to be set, a list of options is presented as flashing text below the main heading of the function selected. The list is advanced using the ARROW keys
(any direction) until the desired option is displayed. These options are programmable from the Touchpad and require only that the sampler have the correct equipment configuration to utilize them. The last option selected in previous programming will be the first to appear on the display.
The display showing disabled will reflect the fact that the function is not being used.
Sampler operation can be started at uniform intervals. This option allows an interval time to be set between sample cycles.
5-16
Section 5. Programming
This option will allow the controller to determine the SAMPLE INITIATION based on external pulses. Pulses fed to an internal accumulator in the controller will be compared to the setting entered by the operator and will cause a sample cycle to start. The accumulator will reset immediately and counting of pulses will begin again. There is no loss of count during the sample cycle. Pulse requirements of the system are detailed in the specifications.
Where external devices do not themselves generate pulses in any relation to their process but generate a current signal of
4-20mA, this input option will generate internal pulses proportional to the incoming
4-20mA signal. These can then be treated the same as the Pulse Input option and accumulated in the controller to determine when a sample cycle should occur.
The sampler controller can react to an external dry contact, otherwise known as a zero-voltage contact, to activate a sample cycle on demand. This will generally be when external conditions have caused a relay to close, whose contact will cause a sample to be taken.
5.4.2 SAMPLE INITIATION using Interval Time
The following sequence of entries are made on the Touchpad to form a basic operating parameter for operation of the sampler. This will determine the time from the start of a sample cycle to the start of the next sample cycle. No time is lost during the actual sample cycle. The ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the SAMPLE INITIATION key.
5-17
Section 5. Programming
3. Press an ARROW key. Continue until
Interval Time is shown on the display.
4. Press the ENTER key.
When setting the time, a flashing digit prompts for input from a numeric key. To set the time, press a numeric key to enter a value and advance to each digit in succession. The format is HHH:MM. The minimum time can be set to 1 minute, however, practical considerations, such as equipment duty cycle, maintenance and service life suggest times of 3 minutes or longer.
When the five digits are entered, press
ENTER. Any wrong entries will require re-entry. There are two methods of correcting a mistake. The digit flashing
“wraps around” and begins again, at which time the correct entry may be pressed.
Alternately, the arrow keys can be used to reposition the flashing prompt over the error, which can then be replaced with the correct value.
The display will echo the last entry with
<ENTERED>.
The controller has now been given a new value for SAMPLE INITIATION. The new values reside in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then wait the designated time before taking a sample.
Summary of Sequence:
SET, SAMPLE INITIATION, ARROW(S), ENTER, ###:##, ENTER,
RESTART, RESTART.
5-18
Section 5. Programming
5.4.3 SAMPLE INITIATION using Pulse Input
The following sequence of entries are made on the Touchpad to form a basic operating parameter for operation of the sampler. This will determine the number of pulses from the start of a sample cycle to the start of the next sample cycle. The ACTIVE
SETTINGS are not being altered.
1. Press the SET key.
2. Press the SAMPLE INITIATION key.
3. Press an ARROW key. Continue until
Pulse Input is shown on the display.
4. Press the ENTER key.
The display will show a new screen containing the option title PULSE INPUT on the top line and a 7 digit number with the leftmost digit flashing to prompt for a numeric input. To set the number of pulses required to be input before a sample is taken, by use of the ARROW keys, move the flashing prompt until it is over the digit requiring change.
Press a number key (0-9) to replace any existing number and advance the flashing prompt to the next digit to change. Replace digits as required, then press ENTER.
The display will echo the last entry with
<ENTERED>.
Any wrong entries will require re-entry. If
ENTER has not been pressed, reposition the prompt over the incorrect digit and replace it. After ENTER has been pressed, the entire entry must be redone from the beginning
(press SET).
5-19
Section 5. Programming
The controller has now been given a new value for SAMPLE INITIATION. The new values reside in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then wait until the required pulses have been received before taking a sample.
Summary of Sequence:
SET, SAMPLE INITIATION, ARROW(S), ENTER, #######, ENTER,
RESTART, RESTART.
5.4.4 SAMPLE INITIATION using 4-20mA Input
The following sequence of entries are made on the Touchpad to form a basic operating parameter for operation of the sampler. This will determine the number of pulses from the start of a sample cycle to the start of the next sample cycle. The ACTIVE
SETTINGS are not being altered.
1. Press the SET key.
2. Press the SAMPLE INITIATION key.
3. Press an ARROW key. Continue until 4-
20mA Input is shown on the display.
4. Press the ENTER key.
5-20
Section 5. Programming
The display will show a new screen containing the option title 4-20mA INPUT on the top line and a 7 digit number with the leftmost digit flashing as a prompt for input.
The 4-20 mA input will be converted by the controller to pulses, proportional to the span of the input, at the rate set in the specifications. To set the number of pulses required to be input before a sample is taken, use the ARROW keys to move the flashing prompt until it is over the digit requiring change.
Press a number key (0-9) to replace any existing number and advance the flashing prompt to the next digit to change. Replace digits as required, then press ENTER.
The display will echo the last entry with
<ENTERED>.
Any wrong entries will require re entry. If
ENTER has not been pressed, reposition the prompt over the incorrect digit and replace it. After ENTER has been pressed, the entire entry must be redone from the beginning (press SET).
The controller has now been given a new value for SAMPLE INITIATION. The new values reside in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then wait until the required pulses have been received before taking a sample.
Summary of Sequence:
SET, SAMPLE INITIATION, ARROW(S), ENTER, #######, ENTER,
RESTART, RESTART.
5-21
Section 5. Programming
5.4.5 SAMPLE INITIATION using External Contact
The following sequence of entries are made on the Touchpad to form a basic operating parameter for operation of the sampler. This will determine the time between samples being taken. The ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the SAMPLE INITIATION key.
3. Press an ARROW key. Continue until
External Contact is shown on the display.
4. Press the ENTER key.
The display will echo the last entry with
<ENTERED>.
The controller has now been given a new value for SAMPLE INITIATION. The new value resides in the NEW ENTRIES area of the controller memory. To make this change active, press the RESTART key, then again press it to confirm your choice. The controller will then wait until a contact closure has been received before taking a sample.
Summary of Sequence:
SET, SAMPLE INITIATION, ARROW(S), ENTER, RESTART,
RESTART.
5-22
5.5 Programming PROGRAM TYPE
Section 5. Programming
5.5.1 PROGRAM TYPE Overview
PROGRAM TYPE is the function which determines how the sampler will perform its program. A variety of options are available. These options are programmable from the Touchpad and require only that the sampler have the correct equipment configuration to utilize them. The basic function of the
Program Type is to determine the movement of the distributor.
The program option which determines that all the samples that are gathered will be placed in a single container. The sampler program terminates after the specified number of samples have been taken.
This option is used for discrete sampling applications, to deposit one or a number of samples to one container before advancing to the next container. The interval between samples is controlled by the SAMPLE
INITIATION function. The multi-composite setting is programmable up to 99 samples per container, for up to 24 containers depending on the hardware configuration.
This option is used for discrete sampling applications, to successively deposit one sample to each of a programmed number of containers on any given sample initiation.
The consecutive setting is programmable up to 99 containers per sample initiation.
Allows the sampler to deposit equal sample volumes into a predetermined number of containers per programmed day. Each day may have any number of samples taken, dependent on the SAMPLE INITIATION mode chosen. Deposits are made to as many as 9 containers per day, to a cumulative total of 24 containers. Timing is dependent on the crystal-controlled Real Time Clock in the controller. The first program day will be the current day the programming is done.
5-23
Section 5. Programming
This option will cause the sampler to step to a new container regardless of the status of the SAMPLE INITIATION setting. For example, the actual sampling may be under the control of a flowmeter and taking samples based on the flow rate as determined by pulses or 4-20mA input to the controller. When the user-programmed timed interval has elapsed, the controller will step to a new container. The Timed
Step can be set for any interval up to 99 hours 59 minutes. Progress of the step timer can be viewed by selecting View, Program
Type and pressing ENTER twice. Thus you may view the step timing and the amount of time until the next step occurs.
5.5.2 PROGRAM TYPE - Composite
The following sequence of entries are made on the Touchpad to describe how the
Sampler controller is to store the samples it takes, in the hardware specified in its configuration. The ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the PROGRAM TYPE key.
3. Press an ARROW key. Continue until
Composite is shown on the display (for storage in single container).
4. Press the ENTER key.
5-24
Section 5. Programming
The display will respond with the message
“Terminate After 0000009 Samples”. The numerical value will be whatever value was last placed in the controller’s memory, usually after previous programming. To keep the previous value press ENTER, or, to set a new value, use the ARROW keys to advance the flashing prompt to the desired location and replace the digits under the prompt by using the digits (0-9) on the
Touchpad. Each new entry will automatically advance the prompt to the next location. In this way, the entire 7 digit number can be changed. The ARROW keys can be used to skip already correct digits, in either direction.
When the 9 digit number is correctly entered, press ENTER.
The display will echo the last entry with
PROGRAM TYPE <ENTERED>.
The controller has now been given a new value for PROGRAM TYPE. The new value resides in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then be set to perform as a Composite Sampler in conjunction with the parameters programmed under the
START DELAY and SAMPLE
INITIATION variables.
Summary of Sequence:
SET, PROGRAM TYPE, ARROW(S), ENTER, #######, ENTER,
RESTART, RESTART.
5-25
Section 5. Programming
5.5.3 PROGRAM TYPE - Daily Cycle
The following sequence of entries are made on the Touchpad to describe how the
Sampler controller is to store the samples it takes, in the hardware specified in its configuration. The ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the PROGRAM TYPE key.
3. Press an ARROW key. Continue until
Daily Cycle is shown on the display (for storage in a single container or a multiple container array).
4. Press the ENTER key.
The display will respond with the message
“DAILY CYCLE Total Bottles `nn’“. The two digits will be flashing as a prompt that they can be changed using the arrow keys.
Press ENTER. The second line of the display will change to show a day of the week. The first day that shows will be the day the programming is being done.
Following the day of the week will be a flashing digit, which is prompting for a numeric input. Using the number keys (0-9) enter the number of bottles to be utilized on the displayed day, to a maximum of 9 or the total number of containers not yet allocated from the array.
Press ENTER. This will register the value for the displayed day and change that display to show the next day. Again, enter a number for that particular day, remembering that only containers not previously allocated
5-26
Section 5. Programming can be chosen. If no change in the displayed value is required, the value has been changed or the value is zero, pressing
ENTER will advance the day of the week.
Therefore...
Press ENTER. Repeat the above procedures until all the required containers, on their respective days, have been allocated, or the
7 days of the week are all selected. The total number available is never allowed to be more than the amount registered as the two
digit representation of the hardware configuration in the “Total bottles 00” entry.
The display will echo the last entry with
PROGRAM TYPE <ENTERED>.
The controller has now been given a new value for PROGRAM TYPE. The new value resides in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then be set to perform as a Daily Cycle Sampler in conjunction with the parameters programmed under the
START DELAY and SAMPLE
INITIATION variables.
Summary of Sequence:
SET, PROGRAM TYPE, ARROW(S), ENTER, ARROW(S), ENTER, #,
ENTER, #, ENTER, #, ENTER, #, ENTER, #, ENTER, #, ENTER, #,
ENTER, RESTART, RESTART.
5.5.4 PROGRAM TYPE - Daily Cycle for Dual Station
Dual Station – Flip Flop Application Only (Single Controller, Two Metering
Chambers)
The following sequence of entries are made on the Touchpad to describe how the
Sampler controller is to store the samples it takes, when the hardware specified is configured to deliver Samples from two separate sources. The ACTIVE SETTINGS are not being altered.
5-27
Section 5. Programming
1. Press the SET key.
2. Press the PROGRAM TYPE key.
3. Press an ARROW key. Continue until
Daily Cycle is shown on the display.
4. Press the ENTER key.
The display will show the response “DAILY
CYCLE Total Bottles nn“. The two digits,
nn, will flash, indicating they can be changed . In this application, the number of bottles must be set to 24.
Press ENTER. The second line of the display will change to show a day of the week. The first day that shows will be the day the programming is being done.
Following the day of the week will be a flashing single digit, which is the prompt for a numeric input. Using the number keys
(0-9) enter the number of bottles to be utilized on the first day. In this application this will be set to 2.
Press ENTER. This will register the value for the displayed day and change that display to show the next day. Again, enter a number for that particular day, remembering that in this application this will be set to 2. If no change in the displayed value is required, the value has been changed or the value is zero, pressing ENTER will advance the day of the week. Therefore...
Press ENTER. Repeat the above procedures until all the required containers, on their respective days, have been allocated, or the
7 days of the week are all selected. If all 7 days have been selected, then continuous sampling will occur for up to 12 days.
5-28
Section 5. Programming
The display will echo the last entry with
PROGRAM TYPE <ENTERED>.
The controller has now been given a new value for PROGRAM TYPE. The new value resides in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then be set to perform as a Dual Station Sampler in conjunction with the parameters programmed under the
START DELAY and SAMPLE
INITIATION variables.
Summary of Sequence:
SET, PROGRAM TYPE, ARROW(S), ENTER, ARROW(S), ENTER, #,
ENTER, #, ENTER, #, ENTER, #, ENTER, #, ENTER, #, ENTER, #,
ENTER, RESTART, RESTART.
5.5.5 PROGRAM TYPE - Consecutive
The following sequence of entries are made on the Touchpad to describe how the
Sampler controller is to store the samples it takes, in the hardware specified in its configuration. The ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the PROGRAM TYPE key.
3. Press an ARROW key. Continue until
Consecutive is shown on the display (for storage in single multiple container array).
4. Press the ENTER key.
5-29
Section 5. Programming
The display will show the response
“CONSECUTIVE Total Bottles nn“. The two digits, nn, will be flashing, indicating they can be changed by scrolling with the arrow keys.
Press ENTER. The display now reads “nn bottles per Sample Cycle”. A previously set value will be displayed. One of the digits is flashing. Using the number keys (0-9), enter the first digit of the number of bottles that will be used at each sampling time determined by the programming setting, i.e., enter the number of samples to be taken at each predetermined time. After the first digit is entered, the second digit will flash prompting for the remaining digit of the entry. The sampler will repeat this quantity each time the sampling is initiated, until the
“Total-Bottles” setting is reached.
Press ENTER. The display now shows
“PROGRAM TYPE <ENTERED>”.
The controller has now been given a new value for PROGRAM TYPE. The new value resides in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then be set to perform as a Consecutive Sampler in conjunction with the parameters programmed under the
START DELAY and SAMPLE
INITIATION variables.
Summary of Sequence:
SET, PROGRAM TYPE, ARROW(S), ENTER, ENTER, ##, ENTER,
RESTART, RESTART.
5-30
Section 5. Programming
5.5.6 PROGRAM TYPE - Multi-Composite
The following sequence of entries are made on the Touchpad to describe how the
Sampler controller is to store the samples it takes, in the hardware specified in its configuration. The ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the PROGRAM TYPE key.
3. Press an ARROW key. Continue until
Multi-Composite is shown on the display
(for storage in multiple container array).
4. Press the ENTER key.
The display will respond with the message
“MULTI-COMPOSITE Total Bottles nn”.
The two digits, nn, will be flashing, indicating they can be changed by scrolling with the arrow keys.
Press ENTER. The display now reads
“nn Cycles per bottle”. A previously set value is displayed. One of the digits is flashing, prompting for a numeric entry.
Using the number keys (0-9), select the number of times the same bottle will be used at the times determined by the programming, i.e., enter the number of samples to be taken before the stepper advances to the next bottle. After the first digit is entered, the second digit will flash, prompting for the second digit of the entry. Each sample placed in a bottle will be a complete cycle.
The program will repeat this action each time the sampling is initiated, until the
“Total Bottles” setting is reached.
5-31
Section 5. Programming
Press ENTER. The display now reads
PROGRAM TYPE <ENTERED>.
The controller has now been given a new value for PROGRAM TYPE. The new value resides in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then be set to perform as a Multi-Composite Sampler in conjunction with the parameters programmed under the
START DELAY and SAMPLE
INITIATION variables.
Summary of Sequence:
SET, PROGRAM TYPE, ARROW(S), ENTER, ENTER, ##, ENTER,
RESTART, RESTART.
5.5.7 PROGRAM TYPE - Timed Step
The following sequence of entries are made on the Touchpad to describe how the
Sampler controller is to store the samples it takes, in the hardware specified in its configuration. The ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the PROGRAM TYPE key.
3. Press an ARROW key. Continue until
Timed Step is shown on the display (for storage in multiple container array).
5-32
Section 5. Programming
4. Press the ENTER key.
Press ENTER. The display will show the response “TIMED STEP Total Bottles nn”.
The two digits, nn, will be flashing, indicating they can be changed by scrolling with the arrow keys.
Press ENTER. The bottom line of the display now reads “Step Intvl 00:00”. A previously set value is displayed. One of the digits is flashing, prompting for first digit of a numeric input. The format is HH:MM.
Using the number keys (0-9) enter the time interval at which the stepper MUST advance, regardless of SAMPLE
INITIATION settings. The program will repeat this action at the set interval except during an ongoing sample cycle, when it will advance the stepper after the sample cycle is complete.
Press ENTER. The display now reads
PROGRAM TYPE <ENTERED>.
The controller has now been given a new value for PROGRAM TYPE. The new value resides in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then be set to perform as a Timed Step Sampler in conjunction with the parameters programmed under the
START DELAY and SAMPLE
INITIATION variables.
Summary of Sequence:
SET, PROGRAM TYPE, ARROW(S), ENTER, ENTER, ####, ENTER,
RESTART, RESTART.
5-33
Section 5. Programming
5.6 Programming OTHER OPTIONS
5-34
5.6.1 OTHER OPTIONS Overview
SET or VIEW
This feature allows the user to SET or
VIEW the internal Real-Time clock of the microprocessor.
This feature allows the user to SET or
VIEW the duration for which the sampler will purge the intake line prior to drawing in a sample to the chamber. The maximum allowable setting is 99 seconds.
This feature allows the user to change how the pinch valve will operate during sampling cycles. Depending on which generation of sampler the controller is to be used on determines the setting. If this is a new unit, the pinch valve action will have been factory set. Should the controller be used as a retrofit into an older model, the setting may have to be changed. If the sampler has a pinch valve that squeezes shut the discharge tube even during an inactive state, this model is termed as normally closed. Should the tubing be shut only when the pinch valve is energized, it is termed normally open.
This is the ONLY setting for the
PVSsamplers.
This feature will enable or disable the ability of the controller to cease operations when it encounters repeated difficulties in the drawing of samples. The controller normally will attempt to obtain a valid sample by repetition of its programming with extended purge times and vacuum cycles, also extending the time allowed for the acquisition of the sample. When a sample is not obtained, this fact is noted in the controller memory and the program resumes.
If after two (2) successive attempts have failed, the controller will Shut Down, halting sampling until operator intervention clears any reason for fault and RESTARTs the program. This is not always a required course of action. If FAULT SHUTDOWN is disabled, the program will record all failures to obtain samples and without further
VIEW ONLY
Section 5. Programming attempts being made to obtain the failed sample, will wait until the next sample initiation.
The controller will remember conditions encountered during normal operation.
Reasons for premature ending of a set program will also be saved in memory. By
VIEWing this feature, this information can be obtained at the time the sampler is checked.
Values retained by the controller to indicate number of missed samples.
When equipped with the appropriate hardware and with the controller running the proper program (i.e. Multi-Composite), the current position of the distributor arm can be determined by VIEWing this option. The position information is relative to the original position of the arm at the beginning of the program start. Note: There is no physical “Bottle 1”, any bottle can be determined to be #1 at the beginning of a
sample program.
The following selections are all available under the maintenance heading and are all for VIEWing only. To check any of these values or perform any tests, press VIEW, then OTHER OPTIONS. Arrow left or right as required until the flashing text
MAINTENANCE appears and press ENTER.
Once more, arrow left or right until the desired flashing text appears, and press
ENTER.
Displays the microprocessor’s serial number.
Tests the controller’s on-board lithium battery.
Tests the touchpad keys.
Checks the main IC’s read / write integrity.
Technicians Only! Digital Feedback from two on-board A/D channels.
Channel 1: 4-20mA Input
Channel 2: Displays Float Voltage
5-35
Section 5. Programming
5.6.2 OTHER OPTIONS - Clock
The following sequence of entries are made on the Touchpad to alter the Real Time
Clock, running internally in the controller, which is the basis for all timed functions.
The ACTIVE SETTINGS are not being altered and there are no NEW ENTRIES generated.
1. Press the SET key.
2. Press the OTHER OPTIONS key.
3. Press an ARROW key. Continue until
Clock is shown on the display (for updating the internal Real Time Clock).
4. Press the ENTER key.
The display shows a time / date response in the form of “01:23 AM SUN 01-Jan-92”.
The flashing digits are changed, if necessary, by use of the number keys (0-9) in the same manner as a standard watch, in the HH:MM format. Maximum values are 01 - 12 for the hours pair and 00 to 59 for the minutes.
However, each digit is set separately.
Press ENTER. The display will shift its flashing prompt to the AM/PM pair. Since both characters are flashing, the selection is made by use of the ARROW keys. The selection cycles through AM and PM repeatedly. Choose one.
Press ENTER. The display will shift its flashing prompt to the three characters forming the day of the week. Since all three characters are flashing, the selection is made by use of the
ARROW keys. The selection cycles through the 7 days. Choose one.
5-36
Section 5. Programming
Press ENTER. The bottom line of the display shows a date in the format
DD-MM-YY. The DD pair is a pair of digits with the normal range of 00 - 31. Set by number keys (0-9) individually.
Press ENTER. The MM characters are set using the ARROW keys for the choice of one regular calendar month.
Press ENTER. The YY pair of digits are set using the number keys again, corresponding to the last two digits in the year. Set each digit individually. Press ENTER. The date as entered is now checked by the internal clock.
An invalid date will return the prompt to the beginning of the date setting, after an error message is displayed. It can then be corrected and re-entered.
The display now reads CLOCK <ENTERED>.
The controller has now been given a new value for the REAL TIME CLOCK. The new value is in use immediately, once set.
Summary of Sequence:
SET, OTHER OPTIONS, ARROW(S), ENTER, ##, ##, ENTER, ,
ARROWS, ENTER, ARROWS, ENTER, ##, ENTER, ARROWS,
ENTER, ##, ENTER.
5.6.3 OTHER OPTIONS - Purge Time
The following sequence of entries are made on the Touchpad to describe how the
Sampler controller is to operate some of the hardware specified in its configuration. The
ACTIVE SETTINGS are not being altered.
1. Press the SET key.
5-37
Section 5. Programming
2. Press the OTHER OPTIONS key.
3. Press an ARROW key. Continue until
Purge Time is shown on the display (for operating time when the sampler is in the purge mode of the Sample cycle, pressurizing the inlet hose to clear it of obstructions and fluid).
4. Press the ENTER key.
The display shows the response “PURGE
TIME SS seconds”. The first digit of SS is flashing. A previously set value may be displayed. Using the number keys (0-9) enter the time in seconds that will represent the basic purge time. Multiples of this time may be used by the controller to facilitate the removal of obstructions in the inlet hose.
Press ENTER. The display now reads
PURGE TIME <ENTERED>.
The controller has now been given a new value for PURGE TIME. The new value resides in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice. The controller will then be set to perform in conjunction with the parameters programmed under the START DELAY,
SAMPLE INITIATION and PROGRAM
TYPE settings.
Summary of Sequence:
SET, OTHER OPTIONS, ARROW(S), ENTER, ##, ENTER, RESTART,
RESTART.
5-38
Section 5. Programming
5.6.4 OTHER OPTIONS - Pinch Valve
The following sequence of entries are made on the Touchpad to describe how the
Sampler controller is to operate some of the hardware specified in its configuration. The
ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the OTHER OPTIONS key.
3. Press an ARROW key. Continue until
Pinch Valve is shown on the display.
4. Press the ENTER key.
The Pinch Valve option is for control of the solenoid activated pinch bar determining vacuum / pressure modes in the sampler operation. This is a factor usually set at the factory because it relates directly to the equipment specification. It is alterable only to facilitate the use of the controller in plants that have a variety of equipment configurations.
The display will show the response “PINCH
VALVE Normally Open / Closed”. All the characters on the bottom line are flashing, therefore the choice is made with the
ARROW keys. The choice is a toggle between ‘Open’ and ‘Closed’. It represents the state of the pinch valve when the equipment is idle. Note: The sampler will not perform properly if this setting is in
error.
Press ENTER. The display now reads
PINCH VALVE <ENTERED>,
5-39
Section 5. Programming
The controller has now been given a new value for PINCH VALVE. The new value resides in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice. The controller will then be set to perform in conjunction with the parameters programmed under the START DELAY,
SAMPLE INITIATION and PROGRAM
TYPE settings.
Summary of Sequence:
SET, OTHER OPTIONS, ARROW(S), ENTER, ARROW(S), ENTER,
RESTART, RESTART.
5.6.5 OTHER OPTIONS - Fault Shutdown
The following sequence of entries are made on the Touchpad to describe how the
Sampler controller is to operate some of the hardware specified in its configuration. The
ACTIVE SETTINGS are not being altered.
1. Press the SET key.
2. Press the OTHER OPTIONS key.
3. Press an ARROW key. Continue until
Fault Shutdown is shown on the display..
4. Press the ENTER key.
Fault Shutdown is used to control whether the sampler will cease taking samples after a predetermined number of unsuccessful attempts.
5-40
Section 5. Programming
The display will show the response “FAULT
SHUTDOWN Enabled / Disabled”. All the characters on the bottom line are flashing, therefore the choice is made with the
ARROW keys. The choice is a toggle between ‘Enabled’ or ‘Disabled’. If disabled, the controller will make a lengthy attempt to obtain a sample, then return control to the SAMPLE INITIATION to try again. The controller will accumulate a count of unsuccessful (abandoned) attempts.
Press ENTER. The display now reads
FAULT SHUTDOWN <ENTERED>.
The controller has now been given a new value for FAULT SHUTDOWN. The new value resides in the NEW ENTRIES area of the controller memory. To make these changes active, press the RESTART key, then again press it to confirm your choice.
The controller will then be set to perform in conjunction with the parameters programmed under the START DELAY,
SAMPLE INITIATION and PROGRAM
TYPE settings.
Summary of Sequence:
SET, OTHER OPTIONS, ARROW(S), ENTER, ARROW(S), ENTER,
RESTART, RESTART.
5-41
Section 5. Programming
5.7 Viewing Information
5.7.1 Viewing Programmed Information
To see current settings, press the VIEW button, followed by the appropriate button as described on its label.
The display will show current parameter settings, beginning with the requested major category.
START DELAY
Press the VIEW key.
Press the START DELAY key.
Sequence: VIEW, START DELAY, ENTER,
(ENTER).
SAMPLE INITIATION
Press the VIEW key.
Press the SAMPLE INITIATION key.
Sequence: VIEW, SAMPLE INITIATION,
ENTER, (ENTER).
PROGRAM TYPE
Press the ENTER key. If more information is available for a given parameter, it will be displayed. Continue pressing ENTER until no new information is presented. The display will “wrap-around” to its first message.
Where timing or counting are used, active values will be shown which can be used to monitor the progress of the parameter.
Press the VIEW key.
Press the PROGRAM TYPE key.
Sequence: VIEW, PROGRAM TYPE, ENTER,
(ENTER).
5-42
Section 5. Programming
OTHER OPTIONS
Press the VIEW key.
Press the OTHER OPTIONS key.
Use ARROW keys to navigate to desired option on the flashing display.
Press ENTER to view.
Available options are:
Clock - Time, Date (including Day)
Purge Time - Time in seconds
Pinch Valve - Normally Open or Closed
Fault Shutdown - Enabled or Disabled
Sampler Status - Error and system messages that have been lost from the display by keyboard entry.
Cycles Abandoned - Counter
Bottle Position - Relative position of distributor,
Maintenance
Serial Number - unit identification No.,
Analog channels - A/D output display,
- battery,
- locations.
Keypad Check - test of touchpad
(Under Maintenance, ARROW to selection, then display with ENTER.)
ACTIVE SETTINGS
Press the VIEW key.
Press the ACTIVE SETTINGS key.
The display will show “ACTIVE SETTINGS
‘ENTER’ to list”.
Press the ENTER key. The display will show the
START DELAY programming.
Continuously pressing the ENTER key will display all of the active program selections and return to the original display.
Sequence: VIEW, ACTIVE SETTINGS,
ENTER(S)
5-43
Section 5. Programming
NEW ENTRIES
Press the VIEW key.
Press the NEW ENTRIES key.
If no “NEW ENTRIES” have been made, the display will show “No New Entries View Active
Set”. If new parameters have been set, but the unit hasn’t been RESTARTed, the display will show
“NEW ENTRIES ‘ENTER’ to list”. Press the
ENTER key. The display will show the START
DELAY programming. Continuously pressing the
ENTER key will display all of the program selections, SUBSTITUTING new parameters where they’ve been changed, and return to the original display.
Sequence: VIEW, NEW ENTRIES, ENTER(S)
5.7.2 Viewing Generated Information
The following sequence of entries are made on the Touchpad to examine the sample information collected or generated by the controller and stored in its memory.
SAMPLES TAKEN
Press the VIEW key.
Press the SAMPLES TAKEN key.
The display will show a count of all samples taken during the current program. To make these changes active, press the RESTART key, then again press it to confirm your choice. The controller will then wait until the designated time before starting its sampling program.
REMAINING PULSES
Press the VIEW key.
Press the REMAINING PULSES key.
The display will show a countdown of incoming pulses, decreasing from the programmed value.
Only available when either START DELAY or
SAMPLE INITIATION are using their Pulse Input options or pulses generated by the 4-20mA input option. The information is updated continuously and can be left on the display as a progress indicator.
Sequence: VIEW, REMAINING PULSES.
5-44
Section 5. Programming
REMAINING TIME
Press the VIEW key.
Press the REMAINING TIME key.
The display will show various time counters dependent on the programming of the START
DELAY and SAMPLE INITIATION parameters.
Priority goes to START DELAY, which will show an incrementing time for event related delays or decrementing time for time related delays. The display will then yield to SAMPLE INITIATION for an elapsed time display for event related inputs and Remaining Time display for time related inputs.
Sequence: VIEW, REMAINING TIME.
5-45
Section 5. Programming
5-46
Appendix A. Principles of Operation
Switching Methods (Sinking / NPN)
LOAD
Sensor
Sinking (NPN) Switch
The Sinking method connects or switches one side of the load to the negative
(-) side of the power supply. The positive (+) side is connected directly to the other side of the load as shown. “NPN” refers to the type of transistor used to act as a switch in this type of solid-state sensor.
Switching Methods (Sourcing / PNP)
Sensor
LOAD
Sourcing (PNP) Switch
The Sourcing method connects or switches one side of the load to the positive
(+) side of the power supply. The negative (-) side is connected directly to the other side of the load as shown. “PNP” refers to the type of transistor used to act as a switch in this type of solid-state sensor.
A-1
Appendix A. Principles of Operation
A-2
Appendix B. Parts List
This is a partial list of most frequently requested PVS Sampler replacement parts.
TABLE B-1. PVS Replacement Parts
Part No.
Old Part No.
(Prior to 8-1-11) Description
SAMPLE CONTAINERS
26897
28258
27956
27-03-05
27-03-05P
2.3 Gallon (9L) Nalgene with side plug (Polyethylene)
2.3 Gallon (9L) Nalgene with side plug (Polypropylene)
27-03-05-4150 2.3 Gallon (9L) Nalgene with side plug for PVS4150 (HDPE)
27-03-07 2.5 Gallon (10 L) Glass with Teflon Cap
26900 22-10-32 Discrete Bottle Tray (24-Bottle x 500cc)
SINKER / STRAINER
26915 23-28-01-3/8 Sinker (Lead): 3/8 System
27820
27821
26914
27821
27938
23-28-01-5/8 Sinker (Lead): 5/8 System
23-28-10 Sinker Strainer (Stainless Steel): 5/8 System
23-28-11
23-28-12
23-28-13
Sinker Strainer (Stainless Steel): 3/8 System
Sinker (Stainless Steel): 3/8 System
Sinker (Stainless Steel): 5/8 System
INTAKE TUBE
26904
27819
26-01-16
26-01-18
PVC Standard: 3/8" ID (per foot)
PVC Standard: 5/8" ID (per foot)
26-01-14
26-01-09
Teflon: 1/2" ID (Minimum 25 Ft)
Teflon: 3/4"ID (Minimum 25 Ft)
INTAKE TUBE WITH SINKER/STRAINER
27949 26-02-01 PVC: 3/8"ID: 25 Ft with Lead Sinker
26925-L50-E1 26-02-01-050 PVC: 3/8"ID: 50 Ft with Lead Sinker
26925-L100-E1 26-02-01-100 PVC: 3/8"ID: 100 Ft with Lead Sinker
26925-L150-E1 26-02-01-150 PVC: 3/8"ID: 150 Ft with Lead Sinker
26926-L25-E1 26-02-02 PVC: 5/8"ID: 25 Ft with Lead Sinker
B-1
Appendix B. Parts List
28224
26909
26908
26910
26907
26911
26905
27942
26918
26919
26926-L50-E1 26-02-02-050 PVC: 5/8"ID: 50 Ft with Lead Sinker
26926-L100-E1 26-02-02-100 PVC: 5/8"ID: 100 Ft with Lead Sinker
26926-L150-E1 26-02-02-150 PVC: 5/8"ID: 150 Ft with Lead Sinker
26-02-03 Teflon: 1/2" ID: 25 Ft with SS Sinker
26-02-03-050 Teflon: 1/2" ID: 50 Ft with SS Sinker
26-02-11 Teflon: 3/4" ID: 25 Ft with SS Sinker/Strainer
Teflon: 1/2" ID: 25 Ft with SS Sinker/Strainer 26-02-21
DISCHARGE TUBE
26898
27957
28251
26899
26-03-01
26-03-06
26-03-11
26-03-12
METERING CHAMBER
26906
27941
50-01-01-MR
50-01-08-MR
24-01-01
24-01-02
Discharge Tubing (Latex): 3/8" ID: 3 Ft
Discharge Tubing (Latex): 5/8" ID: 3 Ft
Discharge Tubing (Silicone): 5/8" ID: 3 Ft
Discharge Tubing (Silicone): 3/8" ID: 3 Ft
Metering Chamber Assembly (complete): 3/8 System, 500cc
Metering Chamber Assembly (complete): 5/8 System, 500cc
Metering Chamber (Acrylic): 3/8 System, 500cc
Metering Chamber (Acrylic): 5/8 System, 500cc
24-01-08
24-01-10
50-21-01
50-21-04
50-21-06
50-21-07
23-03-01
23-03-04
23-37-01
28-05-01
28-05-02
28-05-03
28-05-04S
28-05-04
Metering Chamber (Pyrex): 3/8 System, 500cc
Metering Chamber (Pyrex): 5/8 System, 500cc
Metering Chamber Cover: 3/8 Delrin
Metering Chamber Cover: 5/8 Delrin
Metering Chamber Cover: 3/8 Teflon
Metering Chamber Cover: 5/8 Teflon
Volume Control Tube: 3/8 System, 500cc
Volume Control Tube: 5/8 System, 500cc
Liquid Sensing Rod for Metering Chamber: 500cc
O-Ring: Barrier Valve (Buna-N)
O-Ring: Metering Chamber (Buna-N)
O-Ring: Barrier Valve (Viton)
O-Ring: Metering Chamber (Silicone)
O-Ring: Metering Chamber (Viton)
B-2
Appendix B. Parts List
VACUUM PUMP
28006
28009
32-01-01
32-02-05
26895 32-08-10
32-10-01
DISTRIBUTOR / STEPPER
Vacuum Pump - 12VDC
Pump Assembly (including solenoids and fixtures) - 12VDC
Brush & Lead Wire Kit (for 32-01-01)
Pump Kit with Flap & Diaphragms
OTHER COMPONENTS
28005
28012
30-DC-MFCB Multi-Function Input Controller (12VDC)
50-02-13 Pinch Valve Assembly (12VDC All Systems)
27997
27998
28-11-41
28-11-42
28-11-43
28-11-44
Quick Connector Stem (SS): 1/2"
Quick Connector Body (SS): 1/2"
Quick Connector Stem (SS): 3/4"
Quick Connector Body (SS): 3/4"
26926
6053
MANUAL
55-15-12
55-15-40
Hard Copy of Manual
Battery: Large 12VDC, 17AH, 15 lbs
Battery: Small 12VDC, 7AH, 4 lbs
Switching Power Supply c/w Filtering Capacitor:
INPUT: 88 - 264 VAC, 50/60 Hz. 2.5 Amps
OUTPUT: 13.6 VDC, 10 Amps.
B-3
Appendix B. Parts List
B-4
Appendix C. Programming 4-20mA for
Flow Proportional Sampling
In order to use the 4-20mA interface with a PVS Sampler, calculations must be made based on flow. The 4-20mA input is a signal that corresponds to the flow meter’s output. 20mA is equal to the maximum flow, and 4mA is equal to the minimum flow. The controller requires a number which reflects the maximum flow going through the sampler.
The PVS Controller generates 100 pulses per minute internally at the maximum flow. This number decreases with the amount of flow proportional to the 4-20mA scale. The Controller requires the number of pulses at maximum flow. In order to calculate this, use the following formula:
1. Calculate Q. Q = Average flow rate divided by the maximum flow rate.
Q =
Average Flow
Maximum Flow
Rate
Rate
2. Calculate t. t =
Volume between samples
Average volume per minute
t is the number of minutes per sample you would like for an average flow rate.
Either choose how long between samples you’d like for average flow, or calculate based on volume above.
3. Multiply Q x t x 100 (100 pulses at max flow)
This is the number you will input into the Controller at the 4-20mA dialogue.
Example
You want to collect samples every 30 minutes. On average 175gal/min flows by. Maximum is 300gal/min.
1. Calculate Q.
Q =
Average Flow Rate
Maximum Flow Rate
=
175 gal/min
300 gal/min
= .
58333
2. Calculate t.
t = 30 min/sample or
5250 Gallons btwn samples
Average 175 gal/min
3. Multiply Q x t x 100 pulses = 1750 pulses/sample
Enter 1750 into the Controller at the 4-20mA dialogue.
C-1
Appendix C. Programming 4-20mA for Flow Proportional Sampling
C-2
Campbell Scientific Companies
Campbell Scientific, Inc. (CSI)
815 West 1800 North
Logan, Utah 84321
UNITED STATES www.campbellsci.com • [email protected]
Campbell Scientific Africa Pty. Ltd. (CSAf)
PO Box 2450
Somerset West 7129
SOUTH AFRICA www.csafrica.co.za • [email protected]
Campbell Scientific Australia Pty. Ltd. (CSA)
PO Box 8108
Garbutt Post Shop QLD 4814
AUSTRALIA www.campbellsci.com.au • [email protected]
Campbell Scientific do Brazil Ltda. (CSB)
Rua Luisa Crapsi Orsi, 15 Butantã
CEP: 005543-000 São Paulo SP BRAZIL www.campbellsci.com.br • [email protected]
Campbell Scientific Canada Corp. (CSC)
11564 - 149th Street NW
Edmonton, Alberta T5M 1W7
CANADA www.campbellsci.ca • [email protected]
Campbell Scientific Centro Caribe S.A. (CSCC)
300 N Cementerio, Edificio Breller
Santo Domingo, Heredia 40305
COSTA RICA www.campbellsci.cc • [email protected]
Campbell Scientific Ltd. (CSL)
Campbell Park
80 Hathern Road
Shepshed, Loughborough LE12 9GX
UNITED KINGDOM www.campbellsci.co.uk • [email protected]
Campbell Scientific Ltd. (France)
3 Avenue de la Division Leclerc
92160 ANTONY
FRANCE www.campbellsci.fr • [email protected]
Campbell Scientific Spain, S. L.
Avda. Pompeu Fabra 7-9, local 1
08024 Barcelona
SPAIN www.campbellsci.es • [email protected]
Please visit www.campbellsci.com to obtain contact information for your local US or International representative.
Advertisement
Key Features
- Automatic sampling
- Programmable sampling intervals
- External event triggering
- Flow-based sampling
- Composite and discrete sampling
- Temperature control
- Durable and portable design