IONTO-COMED IONTO-UNIVERSAL A1, A1 TREND, A2 TREND, A3 TREND Treatment Couch INSTRUCTION MANUAL
IONTO-UNIVERSAL A1, IONTO-UNIVERSAL A1 TREND, IONTO-UNIVERSAL A2 TREND and IONTO-UNIVERSAL A3 TREND are very comfortable treatment chairs for cosmetics, massage and foot care. The seat height is always continuously electrically adjustable. Depending on the equipment variants with up to three drives, the back and foot sections can be mechanically or electrically continuously adjusted. The seating angle is adjustable by means of a foot pedal. The treatment chairs are available with or without upholstery heating, by choice.
The low entry height of 51 cm as well as the adjustable arm rests offer simple, comfortable entry and exit into/out of the treatment chairs. The head and foot sections are extensible. In this way, the treatment chairs can be individually adjusted to your customer’s body size. The head section is rotated for massage. For the face down position, the arm rests are rotated forward, thereby providing a comfortable resting surface for the arms.
Advertisement
Advertisement
IONTO-COMED
INSTRUCTION MANUAL
IONTO-UNIVERSAL A1
Art. No. 8500, 8500.5
IONTO-UNIVERSAL A1 TREND
Art. No. 8502-1
IONTO-UNIVERSAL A2 TREND
Art. No. 8502-2
IONTO-UNIVERSAL A3 TREND
Art. No. 8502-3
Contents
Safety Instructions .......................................................................................................3
Fields of Use................................................................................................................4
Control Elements and Connections .............................................................................5
Installation ...................................................................................................................6
Operation of IONTO-UNIVERSAL A1, IONTO-UNIVERSAL A1 TREND ....................8
Operation of IONTO-UNIVERSAL A2 TREND ..........................................................10
Operation of IONTO-UNIVERSAL A3 TREND ..........................................................12
Cleaning and Maintenance ........................................................................................14
Fault Finding and Solutions .......................................................................................15
Repairs and Servicing ...............................................................................................16
Technical Data...........................................................................................................16
Transport/ Storage/ Operating Requirements............................................................16
Disposal.....................................................................................................................17
Delivery Contents ......................................................................................................17
Accessories ...............................................................................................................17
This instruction manual contains information intended to ensure safe operation of the treatment chair. Always keep it in close proximity where it is readily available for every user.
© IONTO-COMED 2011
IONTO-COMED GmbH
Durlacher Allee 109
D-76137 Karlsruhe
Telephone +49 (0) 721 / 97 70-0
Telefax +49 (0) 721 / 97 70-290
Email: [email protected] http://www.ionto.de
2 © by IONTO-COMED
Safety Instructions
Before installing and using the treatment chair, carefully read the instruction manual and ensure that all directions and instructions contained are fully understood. It includes important safety information for installation, operation and maintenance!
If the treatment chair is used for any purpose other than that for which it was designed and as described in this instruction manual, such use will invalidate the manufacturers warranty and guarantee!
The control unit may only be opened by the manufacturer or by an expert approved and authorised by the manufacturer.
When unpacked, check the contents of the delivery to ensure that all parts have been included and that nothing has been damaged during transit. In the event of any doubt, please contact our customer service department.
Never use the treatment chair with a damaged mains power supply cable. Do not jam or kink the cable and do not allow it to rest on any hot object (e.g. radiator, lamp, etc)!
If the mains cable is damaged in any way, immediately disconnect the treatment chair from the mains power supply. Please consult our customer service department before using again.
Before connecting to the mains power supply ensure that the voltage (V) shown on the type plate corresponds with the power supply available!
Under no circumstances may the mains plug be disconnected from the socket with wet hands. Never attempt to remove the plug from the socket by pulling the connecting cable!
Do not use in damp surroundings, nor splash or immerse in liquid!
In the event of any liquid or foreign body entering the inner of the control unit, immediately disconnect the treatment chair from the mains power supply! Please consult our customer service department before using again.
Always isolate from the mains power supply by disconnecting the mains plug before cleaning!
© by IONTO-COMED 3
Fields of Use
The treatment chairs IONTO-UNIVERSAL A1, IONTO-UNIVERSAL A1 TREND,
IONTO-UNIVERSAL A2 TREND and IONTO-UNIVERSAL A3 TREND are very comfortable treatment chairs for cosmetics, massage and foot care. The seat height is always continuously electrically adjustable. Depending on the equipment variants with up to three drives, the back and foot sections can be mechanically or electrically continuously adjusted. The seating angle is adjustable by means of a foot pedal. The treatment chairs are available with or without upholstery heating, by choice.
The low entry height of 51 cm as well as the adjustable arm rests offer simple, comfortable entry and exit into/out of the treatment chairs. The head and foot sections are extensible. In this way, the treatment chairs can be individually adjusted to your customer’s body size. The head section is rotated for massage. For the face down position, the arm rests are rotated forward, thereby providing a comfortable resting surface for the arms.
The treatment chairs IONTO-UNIVERSAL A2 TREND and IONTO-UNIVERSAL
A3 TREND are equipped with a manual control with save function. Three individual treatment positions can be saved.
4 © by IONTO-COMED
Control Elements and Connections
1
2
3
4
9
10
11
Fig. 2: Manual control
12
13
7
8
5
6
Fig. 1: Treatment chair IONTO-UNIVERSAL A1 TREND
Legend for treatment chair Legend for manual control section rest rests section
(6) Foot switch (only with IONTO-
UNIVERSAL A1 / A1 TREND)
(7) Back rest adjustment (only with
IONTO-UNIVERSAL A1 / A1 TREND)
(8) Seat angle adjustment
(only with IONTO-UNIVERSAL A2
TREND and IONTO-UNIVERSAL A3
TREND) rest
(10) Leg rest adjustment (only with
IONTO-UNIVERSAL A3)
(11) Seating height adjustment
(12) Programme keys 1-3
(13) Save key S
© by IONTO-COMED 5
Installation
1. Install the treatment chair:
) Place the treatment chair on a level, solid underground.
) Arrest the treatment chair with the aid of the brakes on the rollers.
) Place the head section into the holding bracket on the back rest.
) Place the foot section into the holding bracket on the leg rest.
2. Connect the treatment chair to the mains voltage:
) Ensure that the mains power supply corresponds to the voltage (V) shown on the type plate.
) Plug the mains power plug into the mains socket.
3. Placement of the pushing rod for the leg rest:
) Move the treatment chair slightly upwards (see operation).
) Remove the transport lock from the pushing rod.
) Slightly raise the leg rest.
) Remove the splint from the bolt at the bracket for the pushing rod underneath the leg rest.
) Remove the bolt.
) Hold the pushing rod firmly.
) Loosen the black adjustment lever under the seat part so that the pushing rod can move freely.
) Guide the pushing rod into the bracket underneath the leg rest.
) Fasten the pushing rod with the bolt.
) Secure the bolt with the splint.
) Arrest the leg rest with the fixing lever.
The treatment chair is now ready for use.
6 © by IONTO-COMED
4. Readjusting the fixing lever for the leg rest:
The black fixing lever for adjusting the leg rest is underneath the seat part. If necessary, the fixing lever can be readjusted.
The fixing lever is fastened to a shaft. Pulling on the fixing lever changes the shaft position.
If the leg rest cannot be adjusted:
) Press on the centre of the fixing lever (see Fig. 3).
) Pull the fixing lever towards you.
The fixing lever lock opens and the fixing lever can move freely.
) Hold the rear part of the fixing lever (see Fig. 4).
) Turn the fixing lever anti-clockwise.
) Release the fixing lever.
) Rotate the fixing lever clockwise to arrest the leg rest.
If the leg rest cannot be arrested yet, repeat the process.
Leg rest cannot be released:
The fixing lever presses against the seat part.
) Press on the centre of the fixing lever (see Fig. 3).
) Pull the fixing lever towards yourself.
The lock of the fixing lever releases, and the fixing lever can move freely.
) Hold the rear part of the fixing lever (see Fig. 4).
) Rotate the fixing lever clockwise.
) Release the fixing lever.
) Rotate the fixing lever anti-clockwise to release the leg rest.
If the leg rest cannot be released yet, repeat the process. hold on
Fig. 3: Releasing the fixing lever Fig. 4: Readjusting the fixing lever
© by IONTO-COMED 7
Operation of IONTO-UNIVERSAL A1, IONTO-UNIVERSAL A1 TREND
Caution!
Risk of pinching!
The arms of the client must remain on the couch during adjustment of the reclining position.
Note
Do not allow that oil or any care products come in contact with the upholstery because this can damage it.
We recommend the use of an upholstery protection, e.g. a towelling cover.
If oil comes in contact with the upholstery, remove immediately.
1. Head section adjustment:
) Pull the head section out until the required position is reached.
) Push the head section down to reset it.
2. Head section rotation:
For the massage position, the head section is rotated by 180º.
) Pull the head section out to the stop.
) Push on the stop on the head section bracket
(Fig. 5) and hold it down.
) Pull the head section out all the way.
) Place the head section into the holding bracket on the back rest after rotating it by 180º.
Fig. 5: Head section bracket
3. Seat height adjustment:
Note
Never put and use the foot switch onto the frame or protective cover of the treatment chair.
The entire treatment chair is electrically moved to the required height.
) Press the ▲ key on the foot switch to adjust the treatment chair upwards.
) Press the ▼ key on the foot switch to adjust the treatment chair downwards.
4. Back rest adjustment:
The back rest and leg rest are adjusted together.
) Press the left foot pedal (7).
) Hold the foot pedal down.
) Push the back rest backwards into the required position with your hand. The back rest and leg rest move together.
) To lock in the required position, release the foot pedal.
) Press the left foot pedal again to set the back rest back.
5. Seating angle adjustment:
The entire treatment chair is tilted backwards or forwards in its seating angle.
) Press the right foot pedal (8).
) Hold the foot pedal down.
) Push the treatment chair into the required position with your hand.
) To lock in the required position, release the foot pedal.
) Press the right foot pedal again to reset the seating angle.
8 © by IONTO-COMED
6. Leg rest adjustment:
The leg rest is adjusted independently from the back rest and seating angle.
) Turn the fixing lever beneath the seat section anti-clockwise to release the leg rest.
) Move the leg rest into the required position with your hand.
) Rotate the fixing lever clockwise to arrest the leg rest.
7. Foot section adjustment:
Note
The foot section must always be fully withdrawn in the entry and exit position (lowest position).
) Pull the foot section out until the required position is reached.
) Push the foot section back again.
Removing the foot section:
) Pull the foot section out to the stop.
) Press on the locking mechanism at the foot rest bracket and hold it down
(see rotating the head section).
) Pull the foot section out all the way.
8. Arm rests adjustment:
) Release the fixing screw under the arm rest.
) Swing the arm rest sideways until the required position is reached.
) Arrest the fixing screw again.
9. Switch the upholstery heating ON:
) Switch the button at the back of the back rest ON.
The green control lamp in the button will illuminate.
Note
For safety reasons, the upholstery heating will automatically switch OFF after 60 minutes of operation!
After approx. 5 seconds the upholstery heating can be switched ON again.
Switch the upholstery heating ON again:
) First switch the button on the back of the back rest OFF.
) Then switch the button on the back of the back rest ON again.
The green control lamp in the button will illuminate.
10. Switch the upholstery heating OFF:
) Switch the button on the back of the back rest OFF.
The green control lamp in the button won’t illuminate anymore.
© by IONTO-COMED 9
Operation of IONTO-UNIVERSAL A2 TREND
Caution!
Risk of pinching!
The arms of the client must remain on the couch during adjustment of the reclining position.
Note
Do not allow that oil or any care products come in contact with the upholstery because this can damage it.
We recommend the use of an upholstery protection, e.g. a towelling cover.
If oil comes in contact with the upholstery, remove immediately.
1. Head section adjustment:
) Pull the head section out until the required position is reached.
) Push the head section down to reset it.
2. Rotating the head section:
For the massage position, the head section is rotated by 180º.
) Pull the head section out to the stop.
) Push on the stop on the head section bracket
(Fig. 6) and hold it down.
) Pull the head section out all the way.
) Place the head section into the holding bracket on the back rest after rotating it by 180º.
Fig. 6: Head section bracket
3. Seat height adjustment:
The entire treatment chair is electrically moved to the required height.
) Press the left key on the manual control in the third row (11) to adjust the treatment chair upwards.
) Press the right key on the manual control in the third row (11) to adjust the treatment chair downwards.
4. Back rest adjustment:
Note
The leg rest must be released during adjustment (entry position).
The back rest and leg rest are adjusted together.
) Press the left key on the manual control in the first row (9) to adjust the treatment chair forwards.
) Press the right key on the manual control in the first row (9) to adjust the treatment chair backwards.
5. Seating angle adjustment:
The entire treatment chair is tilted backwards or forwards in its seating angle.
) Press the right foot pedal (8).
) Hold the foot pedal down.
) Push the treatment chair into the required position with your hand.
) To lock in the required position, release the foot pedal.
) Press the right foot pedal again to reset the seating angle.
10 © by IONTO-COMED
6. Leg rest adjustment:
Note
The leg rest must be released during adjustment of the back rest (entry position).
The leg rest is adjusted independently from the back rest and seating angle.
) Turn the fixing lever beneath the seat section anti-clockwise to release the leg rest.
) Move the leg rest into the required position with your hand.
) Rotate the fixing lever clockwise to arrest the leg rest.
7. Foot section adjustment:
Note
The foot section must always be fully withdrawn in the entry and exit position (lowest position).
) Pull the foot section out until the required position is reached.
) Push the foot section back again.
Removing the foot section:
) Pull the foot section out to the stop.
) Press on the locking mechanism at the foot rest bracket and hold it down
(see rotating the head section).
) Pull the foot section out all the way.
8. Arm rests adjustment:
) Release the fixing screw under the arm rest.
) Swing the arm rest sideways until the required position is reached.
) Arrest the fixing screw again.
9. Saving the treatment position:
) Bring the treatment chair into the required position.
) Press the save key S (13).
An acoustic tone will sound.
) Press the required programme key 1, 2 or 3 (12).
Two acoustic tones will sound.
The treatment position is saved.
A new treatment position can be saved over the previous position.
10. Calling up the saved treatment position:
) Press the required programme key 1, 2 or 3 (12).
) Keep the programme key pressed until the treatment chair has reached the end position.
11. Switch the upholstery heating ON:
) Switch the button at the back of the back rest ON.
The green control lamp in the button will illuminate.
Note
For safety reasons, the upholstery heating will automatically switch OFF after 60 minutes of operation!
After approx. 5 seconds the upholstery heating can be switched ON again.
Switch the upholstery heating ON again:
) First switch the button on the back of the back rest OFF.
) Then switch the button on the back of the back rest ON again.
The green control lamp in the button will illuminate.
© by IONTO-COMED 11
12. Switch the upholstery heating OFF:
) Switch the button on the back of the back rest OFF.
The green control lamp in the button won’t illuminate anymore.
Operation of IONTO-UNIVERSAL A3 TREND
Caution!
Risk of pinching!
The arms of the client must remain on the couch during adjustment of the reclining position.
Note
Do not allow that oil or any care products come in contact with the upholstery because this can damage it.
We recommend the use of an upholstery protection, e.g. a towelling cover.
If oil comes in contact with the upholstery, remove immediately.
1. Head section adjustment:
) Pull the head section out until the required position is reached.
) Push the head section down to reset it.
2. Rotating the head section:
For the massage position, the head section is rotated by 180º.
) Pull the head section out to the stop.
) Push on the stop on the head section bracket
(Fig. 7) and hold it down.
) Pull the head section out all the way.
) Place the head section into the holding bracket on the back rest after rotating it by 180º.
Fig. 7: Head section bracket
3. Seat height adjustment:
The entire treatment chair is electrically moved to the required height.
) Press the left key on the manual control in the third row (11) to adjust the treatment chair upwards.
) Press the right key on the manual control in the third row (11) to adjust the treatment chair downwards.
4. Back rest adjustment:
The inclination angle of the back rest is adjusted.
) Press the left key on the manual control in the first row (9) to adjust the treatment chair forwards.
) Press the right key on the manual control in the first row (9) to adjust the treatment chair backwards.
5. Leg rest adjustment:
The leg rest is adjusted independently from the back rest and seating angle.
) Press the left key on the manual control in the second row (10) to adjust the treatment chair upwards.
) Press the right key on the manual control in the second row (10) to adjust the treatment chair downwards.
12 © by IONTO-COMED
6. Seating angle adjustment:
The entire treatment chair is tilted backwards or forwards in its seating angle.
) Press the right foot pedal (8).
) Hold the foot pedal down.
) Push the treatment chair into the required position with your hand.
) To lock in the required position, release the foot pedal.
) Press the right foot pedal again to reset the seating angle.
7. Foot section adjustment:
Note
The foot section must always be fully withdrawn in the entry and exit position (lowest position).
) Pull the foot section out until the required position is reached.
) Push the foot section back again.
Removing the foot section:
) Pull the foot section out to the stop.
) Press on the locking mechanism at the foot rest bracket and hold it down
(see rotating the head section).
) Pull the foot section out all the way.
8. Arm rests adjustment:
) Release the fixing screw under the arm rest.
) Swing the arm rest sideways until the required position is reached.
) Arrest the fixing screw again.
9. Saving the treatment position:
) Bring the treatment chair into the required position.
) Press the save key S (13).
An acoustic tone will sound.
) Press the required programme key 1, 2 or 3 (12).
Two acoustic tones will sound.
The treatment position is saved.
A new treatment position can be saved over the previous position.
10. Calling up the saved treatment position:
) Press the required programme key 1, 2 or 3 (12).
) Keep the programme key pressed until the treatment chair has reached the end position.
11. Switch the upholstery heating ON:
) Switch the button at the back of the back rest ON.
The green control lamp in the button will illuminate.
Note
For safety reasons, the upholstery heating will automatically switch OFF after 60 minutes of operation!
After approx. 5 seconds the upholstery heating can be switched ON again.
Switch the upholstery heating ON again:
) First switch the button on the back of the back rest OFF.
) Then switch the button on the back of the back rest ON again.
The green control lamp in the button will illuminate.
© by IONTO-COMED 13
12. Switch the upholstery heating OFF:
) Switch the button on the back of the back rest OFF.
The green control lamp in the button won’t illuminate anymore.
Cleaning and Maintenance
Caution!
Danger of electric shock!
Disconnect from the mains power supply before cleaning.
) Clean the upholstery of the treatment chair with warm water and a cloth. We recommend the use of a microfiber cloth (art. no. 8299.61) or a cleaning mitt
(art. no. 9168.9).
) Remove larger areas of soiling with a mild cleaning agent.
) Remove stubborn stains with Incidin Foam (art. no. 4660). Observe the instructions for use of this special cleaner.
Stains should be removed immediately to avoid them penetrating the material.
) Clean the frame of the treatment chair and the manual control or the foot switch with a lightly moistened cloth. Ensure that no water enters the control unit, the manual control or foot switch.
Note
For disinfection purposes we recommend the use of Bacillol plus (art. no. 9220.300). Please note that an intensive use of Bacillol plus can cause slight deformations of the surfaces it is applied to. All types of synthetic leather can embrittle after repeated application over a long period of time.
Do not rub the disinfectant in the upholstery.
Do not use any solvents, chlorides, polish, synthetic cleaning agents, wax polish, aerosol sprays or aggressive cleaning agents which could cause damage to the upholstery and the treatment chair.
Do not use any substances which contain oil or grease to clean the upholstery.
The upholstery is not resistant to chemical cleaning and dry-cleaning methods.
Incidin Foam is not practical for daily use. The cleaner should be used only in areas of extreme soiling.
14 © by IONTO-COMED
Fault Finding and Solutions
Fault Cause Solution
Leg rest cannot be locked.
Readjusted fixing lever for the leg rest.
Readjust the fixing lever for the leg rest (see page 7).
Leg rest cannot be released.
The fixing lever for the leg rest presses against the seat part.
Readjust the fixing lever for the leg rest (see page 7).
Fixing lever for the leg rest is too tightly turned into place.
Turn the fixing lever anticlockwise to release the leg rest.
Upholstery heating does not function.
Following automatic cut-out, the upholstery heating has not switched OFF.
Switch the button on the back of the back rest OFF, wait 5 seconds and switch it
ON again.
Manual control does not function; electrical adjustment of the treatment chair does not work.
Mains power cable not correctly connected.
Socket without power.
Check connection of the mains power cable. Mains power plug must be firmly inserted.
Check the function of the socket with another electrical device
(e.g. a table lamp). control defective.
Contact our customer service department.
In the event of any other problems occurring, please contact our Customer Service
Department (telephone number on the rear page).
© by IONTO-COMED 15
Repairs and Servicing
Your treatment chair does not require any special servicing when used correctly and in compliance with the manufacturer’s directions and instructions. However, in the event of any problem arising, please contact our customer service department.
Only persons authorised by the manufacturer may perform repairs. Any alterations or repairs performed by non-authorised persons will invalidate any guarantee, warranty or responsibility of the manufacturer. In addition, the safety of the treatment chair could be adversely affected.
Technical Data
Article number: 8500 (IONTO-UNIVERSAL A1)
8502-1 (IONTO-UNIVERSAL A1 TREND)
8502-2 (IONTO-UNIVERSAL A2 TREND)
8502-3 (IONTO-UNIVERSAL A3 TREND)
Operating voltage: 230 V~ / 50-60 Hz
110 V~ / 50-60 Hz
Power: UNIVERSAL
110 VA
190 VA (with seat heating)
UNIVERSAL
170 VA
Protection class:
Maximum loading:
250 VA (with seat heating)
II
150 kg
Length: 1660 - 1900 mm
Seat height: 510 - 850 mm
Weight: approx.
Transport/ Storage/ Operating Requirements
Transport only in the manufacturer’s original packaging.
Transport and storage temperature: -20 °C to +50 °C
Operating temperature:
Relative air humidity:
+10 °C to +40 max. 90
°C
%, non-condensing
To avoid the formation of condensation, abrupt changes of temperature should be avoided. Wait until the treatment chair has acclimatised to the ambient temperature before putting into operation.
16 © by IONTO-COMED
Disposal
Dispose of the packaging, treatment chair, upholstery and accessories in an environmentally friendly manner and in compliance with local regulations.
All electrical and electronic products should be disposed of separately from the municipal waste.
This crossed-out wheeled bin symbol on the product means the product is covered by the European Directive 2002/96/EC. The product should not be mixed with general household waste. There exists a separate collection system for these products.
The correct disposal and separate collection of your old appliance will help prevent potential negative consequences for the environment and human health. It is a precondition for reuse and recycling of used electrical and electronic equipment.
Delivery Contents
IONTO-UNIVERSAL A1
1x Treatment Chair IONTO-UNIVERSAL A1
1x Foot Switch
IONTO-UNIVERSAL A1 TREND
1x Treatment Chair IONTO-UNIVERSAL A1 TREND
1x Foot Switch
IONTO-UNIVERSAL A2 TREND
1x Treatment Chair IONTO-UNIVERSAL A2 TREND
1x Manual Control
IONTO-UNIVERSAL A3 TREND
1x Treatment Chair IONTO-UNIVERSAL A3 TREND
1x Manual Control
Accessories
Art. No. 4660
Art. No. 8299.61
Art. No. 9168.9
Art. No. 8301.69
Art. No. 8301.71
Art. No. 8302.60
Art. No. 9220.300
Incidin Foam, 750 ml
Microfiber Cloth
Single-use Cleaning Gloves, 30 pcs.
Towelling Cover, 5-piece, white
Towelling Cover for Foot Section, white
Towelling Cover for Arm Rests, 2-piece, white
Bacillol Plus, 500 ml
For details of further accessories, please refer to the current IONTO-COMED
COSMETIC TECHNOLOGY catalogue.
© by IONTO-COMED 17
EG-KONFORMITÄTSERKLÄRUNG /
DECLARATION OF CONFORMITY
gemäß den Bestimmungen der EG-Richtlinie über / in accordance with the EU-Directives for
Elektromagnetische Verträglichkeit 2004/108/EG /
Electromagnetic Compatibility Directive 2004/108/EC
Elektrische Betriebsmittel zur Verwendung innerhalb bestimmter Spannungsgrenzen
2006/95/EG / Low Voltage Directive 2006/95/EC
Hersteller / Manufacturer:
Adresse / Address:
IONTO-COMED GmbH
Durlacher Allee 109
D-76137 Karlsruhe
Hiermit bestätigen wir, dass das Produkt /
We certifies that the product
Geräteart /
Purpose of equipment:
Universalbehandlungsliege /
Universal Treatment Couch
Gerätetyp / Type of equipment: IONTO-UNIVERSAL A1
IONTO-UNIVERSAL A1 TREND
Artikelnr. / Article no: 8500, 8500.5, 8502-1, 8502-2, 8502-3 mit oben genannten EG-Richtlinien übereinstimmt / is in accordance with the EU Directives above mentioned.
Karlsruhe, 2011-01-03
IONTO-COMED GmbH
Manager
Martin La Fontaine
IONTO-COMED GmbH
Durlacher Allee 109 y D-76137 Karlsruhe
Telephone: +49 (0)7 21 / 97 70-0 y Telefax: +49 (0)7 21 / 97 70-290 http://www.ionto.de y Email: [email protected]
Advertisement
Key features
- Electrically adjustable seat height
- Adjustable back and foot sections
- Adjustable seating angle
- Upholstery heating option
- Low entry height
- Adjustable arm rests
- Extensible head and foot sections
- Rotating head section for massage
- Face down position with forward rotating arm rests