RAE Systems PGM-3000 EntryRAE Multi-Gas Monitor User Manual

RAE Systems PGM-3000 EntryRAE Multi-Gas Monitor User Manual

Below you will find brief information for Multi-Gas Monitor PGM-3000 EntryRAE. This device is a 4-gas plus PID (photoionization detector) gas monitor. It is easy to operate and calibrate, providing added protection without added complexity. The monitor can be used to detect a variety of gases and vapors, including carbon monoxide, hydrogen sulfide, combustibles, oxygen, and volatile organic compounds. The EntryRAE has a large, easy-to-read display and provides audible, visible, and vibration alarms to alert users of unsafe conditions.

advertisement

Assistant Bot

Need help? Our chatbot has already read the manual and is ready to assist you. Feel free to ask any questions about the device, but providing details will make the conversation more productive.

Multi-Gas Monitor PGM-3000 EntryRAE User Manual | Manualzz
EntryRAE
PGM-3000
Multi-Gas Monitor
User Manual
046-4001-000, Revision C, January 2006
Read Before Operating
This manual must be carefully read by all individuals who have
or will have the responsibility of using, maintaining, or servicing
this product. The product will perform as designed only if
it is used, maintained, and serviced in accordance with the
manufacturer’s instructions.
Caution!
To reduce the risk of electric shock, turn off power before removing the
monitor cover. Disconnect the battery before removing sensor modules for
service. Never operate this monitor while the cover is removed. Remove
monitor cover and sensor modules only in an area known to be nonhazardous.
Special Note
When the EntryRAE monitor is removed from the transport case and is
turned on for the first time, there may be residual vapors trapped inside the
monitor, and the initial toxic gas sensors values may indicate a few ppm.
After running the monitor for several minutes in clean air, the residual vapors
should clear and readings should return to near zero.
Attention! For European Applications
A. Recharge batteries in non-hazardous locations.
B . Do not connect external cable to serial interface jack in hazardous locations.
C. Use RAE Systems Charging cradle (P/N 046-3059-001) for connection to
communication port and charging jack only in a non hazardous area.
Note
Only the combustible gas detection portion of this instrument has been
assessed for performance in accordance with C22.2 No.152-M1984.
Seulement la partie combustible de détection de gaz de cet instrument a été
evaluée pour l’exécution selon C22.2 No. 152-M1984.
Protected by U.S. Patents 5,393,979 , 5,561,344 , 5,773,833 , 6,225,633,
6,313,638 , 6,333,632 , 6,320,388
WARNING
To prevent ignition of flammable or combustible atmospheres, disconnect
power before servicing.
WARNING
Do not mix old batteries with used batteries or mix batteries from different
manufacturers.
WARNING
Substitution of components may impair intrinsic safety.
Warnings
For safety reasons, this equipment must be operated and
serviced by qualified personnel only. Read and understand the
user manual completely before operating or servicing.
Battery Pack
Use only RAE Systems battery packs (P/N 046-3007-000 or 046-3051-000). This
instrument has not been tested in an explosive gas/air atmosphere having an
oxygen concentration greater than 21%. Substitution of components may impair
intrinsic safety. Recharge/replace batteries only in non-hazardous atmospheres.
Computer Interface
Do not transfer data by means of the computer interface cable in hazardous
atmospheres.
Static Hazard
Clean only with a damp cloth.
Calibration
While the EntryRAE has been factory calibrated prior to shipment, transport
and storage after leaving the factory can affect the sensors. Therefore, any
newly purchased RAE Systems Instrument should be fully calibrated by
exposing it to known concentration calibration gases before the instrument
is put into service for the first time. For safety, check the accuracy of the
monitor by exposing the sensors to calibration gas(es) before each day’s
use. (Refer to “Calibration” on page 10.)
Caution!
Before each day’s usage, sensitivity must be tested on a known
concentration of methane gas equivalent to 20% to 50% of full-scale
concentration. Accuracy must be within 0% to +20% of actual. Accuracy may
be corrected by calibration. Refer to “Calibration” on page 10.)
Long-Term Storage
Reliable performance of this gas detector is based upon regular usage.
For long term storage, the battery should be disconnected. Preparation for
use after long storage requires installation of the batteries and a warm-up
period of at least 10 minutes for the sensors to equilibrate. The user should
recognize that sensor life is based upon the purchase date.
Readings
Any rapid up-scale reading followed by a declining or erratic reading may indicate
a gas concentration beyond the upper scale limit, which may be hazardous.
CAUTION: HIGH OFF-SCALE READINGS MAY INDICATE AN EXPLOSIVE
CONCENTRATION!
Avertissements
Pour des raisons de sécurité, cet équipement doit être utilisé,
entretenu et réparé uniquement par un personnel qualifié.
Étudier le manuel d’instructions en entier avant d’utiliser,
d’entretenir ou de réparer l’équipement.
Ensemble de Batterie
Utiliser seulement de paquets batterie de RAE Systems (numéro de la pièce
046-3007-000 ou 046-3051-000). Cet instrument n’a pas été essayé dans une
atmosphère de gaz/air explosive ayant une concentration d’oxygène plus élevée
que 21%. La substitution de composants peut compromettre la sécurité intrinsèque.
Ne charger les batteries que dans l’emplacement désigné non dangereux.
Câble de Computer
Connecter pas le câble externe que dans environnements non dangereux.
Danger Risque D’origine Electrostatique
Nettoyer uniquement avec un chiffon humide.
La Calibration
Les indications de toute instruments de RAE Systems doit être testé en
exposant l’instrument à une concentration de gaz connue par une procédure
diétalonnage avant de mettre en service l’instrument pour la première fois.
Pour une sécurité maximale, la sensibilité du EntryRAE doit être vérifié en
exposant l’instrument. (Référez la Calibration à la page 10.)
Attention!
Avant chaque utilisation journalière verifier la sensibilité avec une concentration
connue de methane equivalante a 20% à 50% de la pleine échelle. La precision
doit être comprise entre 0% à +20% de la valeur vraie et peut être corrigée par
une procédure diétalonnage. (Référez la Calibration à la page 10.)
Stockage à Long Term
Le fonctionnement durable de ce détecteur de gas est conditionné par une
utilisation régulière de celui ci. Lors d’un stockage à long terme, la batterie
doit être déconnectée. Le rédémarrage aprés une longue période d’arrêt,
nécessite la réinstallation de la batterie, et une période de chauffage de 10 mn
afin que les capteurs se mettent à l’équilibre. L’utilisateur doit être conscient
que la durée de vie indiquée pour le capteur démarre à sa date d’achat.
Les Lectures
Toute lecture rapide et positive, suivie d’une baisse subite au erratique de la
valeur, peut indiquer une concentration de gaz hors gamme de détection qui
peut être dangereuse.
ATTENTION: DES LECTURES SUPÉRIEURES A L’ÉCHELLE PEUVENT
INDIQUER DES CONCENTRATIONS EXPLOSIVES.
Table of Contents
Read Before Operating. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . b
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . c
Avertissements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . d
General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Equipment List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Physical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Display and Legend. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Operating the EntryRAE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Turning the Monitor On. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Warm-up Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Turning the Monitor Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Monitor Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
The Pump Cycle. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Continuous Operation of the Pump. . . . . . . . . . . . . . . . . . . . . . . . . . 8
Detecting Gases and Vapors Real-time . . . . . . . . . . . . . . . . . . . . . . 8
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Calibration Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Calibration Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Optional Step: Zero Calibration for the Oxygen Sensor. . . . . . . . . 14
Calibration Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
EntryRAE Usage Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Alarm Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Testing Alarm Signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Alarm Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Preset Alarm Limits and Calibration . . . . . . . . . . . . . . . . . . . . . . . . 18
Back Light. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Sampling Pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Datalogging. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Charging the EntryRAE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
The Li-Ion Battery Pack. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Programming Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Security. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Programming Menus. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Entering Program Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Change Monitor Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Replacing the Li-Ion Battery Pack. . . . . . . . . . . . . . . . . . . . . . . . . . 27
Emergency Alkaline Battery Adapter. . . . . . . . . . . . . . . . . . . . . . . . 29
Sensor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
CO Sensor Charcoal Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Cleaning the PID. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Setting Pump Stall Threshold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Certifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Sensor Range and Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Service and Repair Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
SPECIAL NOTE: If the monitor needs to be serviced, contact either:
The RAE Systems distributor where the monitor was purchased; they will
return the monitor on your behalf,
or
The RAE Systems Technical Service department. Before returning the
monitor for service or repair, obtain a Returned Material Authorization (RMA)
number for proper tracking of your equipment. This number needs to be on
all documentation and posted on the outside of the box in which the monitor
is returned for service or upgrade. Packages without RMA Numbers will be
refused at the factory.
General Information
The EntryRAE is a 4-gas plus PID (photoionization detector) gas monitor. Reliable, easy to operate and simple to calibrate, the EntryRAE
delivers added protection without added complexity.
Simple, Modular, Durable PID
RAE Systems is the technology leader in PIDs. Our modular, plug &
play, self-cleaning (patented) PID is the most reliable and durable PID
available today.
Key Features
• Reliable self-cleaning PID
• Also includes CO, H2S, LEL and O2 sensors
• Built-in pump
• Ιnterchangeable Li-Ion and alkaline battery adapter
• Up to 16 hours of continuous operation
• Large, easy-to-read display
• Datalogging
• Visual alarm with bright red flashing LEDs
• Loud audible alarms (95dB at 30 cm)
• Rugged weather-resistant composite case
General Information Equipment List
Photo Item
Number
Part Name
Part Number
Monitor Only
*Shipping case
2
701-3040-000
4
Monitor
Sensors
Photoionization Detector
Oxygen (O2)
Combustible (%LEL, %Vol)
Carbon Monoxide (CO)
Hydrogen Sulfide (H2S)
Rechargeable Li-Ion Battery
Charging Cradle
- 120 V AC to 12 V DC wall charger, US plug or
- 230 V AC to 12 V DC wall charger, Euro plug
Alkaline Battery Adapter
5
Calibration Adapter/Tubing Assembly
046-3040-000
External Filters (5-pack)
046-3022-005
Charcoal Filters (for the CO Sensor)
008-3006-005
8
User Manual
046-4001-000
12
EntryRAE Resource CD
046-4013-000
3
7
023-0102-000
046-1161-000
014-0212-000
032-0200-000
032-0202-000
046-3007-000
046-3059-000
500-0036-000
500-0036-001
046-3051-000
Computer Interface Cable - RS232 to RS232 with USB
adapter
Optional Confined Space Kit II (CSK II)
046-0911-000
1
Hard Transport Case with precut foam
002-3009-000
9
Remote Sampling Probe with 15 feet (5 meters) of
self-coiling Teflon (tm) tubing
008-3015-002
5
Tool Kit
081-0005-000
10
Four-Gas Mix in a 34-Liter cylinder (50% LEL, 20.9%
O2, 10 ppm H2S, 50 ppm CO)
600-0050-004
11
6
100 ppm Isobutylene Gas in a 34 L cylinder
Regulator (male) with tubing
600-0002-000
007-3021-000
6
Regulator (female) with tubing
002-3011-000
Optional Guaranteed Cost of Ownership
4-year repair and replacement guarantee
SVC-PTC4-046
Optional Accessories (items sold separately)
not shown
AutoRAE Docking Station Starter Kit
048-5900-000
not shown
Additional AutoRAE Cradle
048-0154-000
not shown
PID Cleaning Kit
500-0014-010
* Different shipping cases are used for monitor-only and calibration kits.
General Information
1
9
8
12
3
5, 6
5
10
7
2
4
11
Figure 1. The EntryRAE Optional Confined Space Kit.
General Information 4
Physical Description
1. Display
2. Operation / Program keys
3. Charge status
2
4. Visual alarm
5. Gas plate
6. Buzzer
7. Gas inlet with external filter
8. Charging contacts
1
9. Charging cradle
6
10. Power jack
11. RS-232 port
5
9
11
10
3
8
monitor in cradle
7
bottom of monitor
Figure 2. EntryRAE and Charging Cradle.
General Information
Display and Legend
1. PID lamp alarm
2. Alarm conditions (page 18)
3. Pump on (page 19)
4. Pump off
5. Time (page 25)
6. Date (Day, Month, Year)
7. Apply calibration gas (page 12)
8. Battery charge status (page 19)
9. Password protected (page 11, 26)
1
2
3
4
5
6
7
8
9
Figure 3. Display Overview.
General Information Operating the EntryRAE
The external filter must always be used with the monitor.
Turning the Monitor On
To turn the monitor on:
1.0 Press the @ (mode) key for 1 second.
2.0 The warm-up sequence begins:
2.1 The EntryRAE firmware version displays.
2.2 All of the display segments appear.
2.3 The buzzer sounds twice (if the unit has not
been set to run silently; see page 24), the LEDs
blink twice, and the vibration alarm turns on and
off.
2.4 The preset alarm limits appear in order:
HIGH, LOW, TWA, STEL.
3.0 The Time and Date settings are checked:
3.1 The time is shown in 24-hour format, and the
date is shown in “ddmmyy” format.
3.2 If the monitor has never been programmed, you
will see a “date error” message. Press @ to
acknowledge the error. (Refer to “Adjust Date?”
on page 20 to reset the time and date.)
4.0 PID Warm-up:
4.1 If the monitor has not been used recently, PID
warm-up of up to three minutes (180 seconds)
occurs. During this time, the PID lamp icon
appears, as well as a countdown from 180 to 1.
Operating the EntryRAE
5.0 The date of the last calibration is checked:
5.1 Calibration Needed
This message appears if the number of days
since the last calibration is greater than the “Cal
Due Day” setting. (Refer to “Calib Due Day?” in the Calibration section on page 24.)
NOTE: If the Time and Date are invalid or
were recently updated, then calibration is
automatically due. Press @ to acknowledge
the message.
6.0 A Fresh Air Calibration is requested
6.1 Perform a fresh air calibration before each use.
6.2 The monitor must be in clean ambient air during
fresh air calibration.
6.3 Press Y and the calibration begins. The display
counts down from 5 to 0 (6 seconds), at which
time the process is complete.
6.4 “Cal Updating.” The calibration data is updated,
the monitor warm-up is complete, and the
EntryRAE is in Monitor Mode.
Turning the Monitor Off
The monitor cannot be turned off until after the warm-up sequence is
completed.
To turn the monitor off:
1.0 Press the @ button for 5 seconds.
1.1 The display counts down from 3 to 1
(3 seconds), and flashes the LED for each
count.
1.2
Release @ when “Unit Off” appears on the display.
Operating the EntryRAE Monitor Mode
In Monitor Mode, the EntryRAE samples the environment and displays
real-time readings for each enabled sensor.
The Pump Cycle
Under normal operating conditions, when the monitor is not in an alarm
state, the pump cycles on and off about every 8 seconds. This on/off cycle
improves the reliability of the PID and saves battery life.
The pump runs continuously when:
1.0 The concentration of any measured gas or vapor triggers an alarm
condition.
2.0 When the concentration of VOCs is nearing an alarm condition.
Continuous Operation of the Pump
Pressing the Y key and the N key simultaneously causes the pump to run
continuously for 5 minutes. This value can be changed using the ProRAE
Studio software.
Detecting Gases and Vapors Real-time
CAUTION: HIGH OFF-SCALE READINGS MAY INDICATE AN
EXPLOSIVE CONCENTRATION.
ATTENTION: LES LECTURES HORS ÉCHELLE ÉLEVÉES PEUVENT
INDIQUER UNE CONCENTRATION EXPLOSIVE.
Operating the EntryRAE
In Monitor Mode, the EntryRAE samples the environment and displays
the current measured readings for each enabled sensor.
It also calculates and stores the High, Low, STEL and TWA for the
sensors. Press @ to scroll through these readings:
TWA - The Time Weighted Average for CO, H2S, and
VOC gases only; the accumulated reading of the
gas concentration since the monitor was turned on,
divided by 8 hours. Updated every minute.
STEL - The Short Term Exposure Limit for CO, H2S,
and VOC gases only; the average reading of the
gas concentration for the last 15 minutes, which is
updated every minute. Dashes (“- - -”) appear for the
first 15 minutes.
HIGH - The highest reading for each gas
concentration since the monitor was turned on;
updated every second. Press Y to reset the high
values.
LOW – The lowest reading for each gas
concentration since the monitor was turned on;
updated every second. Press Y to reset the
minimum values.
Operating the EntryRAE Calibration
While EntryRAEs are calibrated prior to leaving the factory, temperature
extremes and/or shocks during shipment can cause sensor drift. Therefore,
the accuracy of any newly purchased RAE Systems monitor should be
tested by exposing the sensor(s) to known concentration calibration gas
before the monitor is used or put into service. For maximum safety, the
accuracy of the monitor should be checked by exposing the sensor(s) to
known concentration calibration gas before each day’s use.
Les indications de toute instruments de RAE Systems doit être testé
en exposant l’instrument à une concentration de gaz connue par une
procédure dietalonnage avant metre en service l’instrument pour la
première fois. Pour une sécurité maximale, la sensibilité du EntreRAE doit
être vérifie en exposant l’instrument à une concentration de gaz connue
par une procédure diétalonnage avant chaque utilisation journaliere.
Information regarding sensor expiration is listed in Technical Note TN114, available at www.raesystems.com.
Calibration Equipment
The sensors are calibrated using both fresh air and calibration (span)
gas. The EntryRAE span gas mixtures are fixed. To calibrate an
EntryRAE, the following items are needed:
A. A cylinder of four-gas mix containing:
 50% LEL methane
 20.9% O2
 10 ppm H2S
 50 ppm CO
B. A cylinder of 100 ppm isobutylene.
C. A calibration adapter to connect the monitor to the outlet of the gas
cylinder.
Calibration Procedure
The calibration procedure is broken down into three steps. All steps
should be performed whenever calibration is needed.
Step One: Enter Programming Mode
1.0 The monitor must be in Programming Mode
for calibration.
1.1 Press and hold @ and N for 3 seconds to
enter Programming.
1.2 Password Enabled Monitors.
10 Operating the EntryRAE
1.2.1 If a password is enabled, you are
prompted to enter one.
1.2.2 Use Y to increase the digit and N to
decrease the digit.
1.2.3 Use @ to move to the next digit.
1.2.4 Press @ and Y to enter the password.
1.2.5 If the password is incorrect, the monitor
returns to Monitor mode.
Step Two: Fresh Air Calibration
The monitor must be in clean ambient air during fresh air calibration.
2.1 Press Y and the calibration begins. The display
counts down from 5 to 1 (5 seconds), at which
time the process is complete.
2.2 Calibration Updating
The monitor is in the process of calibrating the
sensors. As it completes, the display shows
the sensors registering zero (or 20.9% for O2),
before it moves to the next step, “Span Cal?”
Using External Zeroing Organic Filters
Use an external zero organic filter (P/N 008-3024-000, 3-pack) if the
ambient air may be contaminated with hydrocarbons. Attach the filter
to the EntryRAE during fresh air calibration. The filter can be used up
to 20 times before disposing. This filter removes most heavier organic
and inorganic compounds, but may not completely remove lighter
compounds such as methane, propane and CO. Note that the filter
should be attached to the EntryRAE before the calibration adapter.
Step Three: Span Gas Calibration for Multiple Sensors
3.0 For this calibration step, the LEL, CO and H2S sensors are calibrated
using the 4-gas mix (P/N 600-0050-004).
Operating the EntryRAE 11
Figure 4. Gas Cylinded with Regulator.
Figure 5. Calibration Adapter.
3.1 Connect the calibration adapter to the EntryRAE and put a regulator
(500 cc per minute flow rate) on the 4-gas mix cylinder.
Figure 6. EntryRAE Connected to Calibration Gas.
3.2 After “Span Cal?” appears on the display with the
CO, H2S, and LEL sensors indicated, press Y.
3.3 Turn the gas all the way on.
3.3 Connect the open end of the calibration
adapter to the regulator. The monitor starts to
countdown from 70 seconds.
3.4 When the countdown timer reaches zero, turn
off the gas, and disconnect the calibration
adapter from the regulator.
3.5 Compare the readings displayed to the span
gas values indicated on the gas cylinder. The
readings should closely agree with the span gas
values.
12 Operating the EntryRAE
3.6 If calibration fails, the monitor briefly displays “Err”
above each failed sensor.
NOTE: Stop calibration in the event that the
gas runs out or is disconnected. To interrupt
calibration, press @ . When calibration stops, the
sensors revert to their previous calibration values.
3.7 After briefly displaying the concentration, the
monitor moves on to calibrating the PID.
Step Four: Span Gas Calibration for the PID Sensor
4.1 The final step is to calibrate the PID sensor.
4.2 Connect the EntryRAE calibration adapter to the monitor (see Figure
6) and put a regulator (500 cc per minute flow rate) on the cylinder of
isobutylene.
4.3 Connect the open end of the calibration adapter to the regulator.
4.4 When “Span Cal?” appears on the display
with the VOC sensor indicated, press
Y and turn the gas all the way on. The
monitor counts down from 30 seconds.
4.5 Turn off the gas, dicsconnect from the
adapter, and remove the calibration
adapter from the monitor.
If the calibration fails, the monitor briefly
displays “Err” above the failed sensor.
NOTE: Stop calibration if the gas runs out
or is disconnected. To interrupt calibration,
press @. When calibration stops, the
sensors revert to their previous calibration
values.
Operating the EntryRAE 13
Optional Step: Zero Calibration for the
Oxygen Sensor
Zero calibration for the oxygen sensor is not required
under normal use. A span calibration using ambient
air (20.9% O2) is usually sufficient to establish its
performance. However, zero calibration may be required
for special applications. Zero gas (100% nitrogen) is
not included in the standard calibration kit and may be
ordered separately (P/N 600-0062-000).
Zeroing the O2 Sensor
 Connect the calibration adapter to the EntryRAE and
put a regulator (500 cc per minute flow rate) on the
cylinder of nitrogen.
 Connect the open end of the calibration adapter to
the regulator.
The display reads “Zero Cal?” and only the O2 segment
appears.
 Press Y and turn on the gas.
If the calibration fails, the monitor briefly displays “Err”
above the failed sensor.
NOTE: Stop calibration in the event that gas runs out
or is disconnected. To interrupt calibration, press @ .
When calibration stops, the sensors will revert back to
their previous calibration values.
Note that each sensor has cross-sensitivities to several gases. These
gases, sensor sensitivity, and other sensor specifications are listed in
Technical Note TN-114, Sensor Specifications And Cross-Sensitivities,
available at www.raesystems.com. Information regarding LEL sensor
poisons is located in Technical Note TN-144, Handling LEL Sensor
Poisons, also available at www.raesystems.com.
14 Operating the EntryRAE
After the optional zero calibration for the O2 sensor, “Calib due day?” is
displayed. Press Y to set a new date, or N to skip.
When setting a new day, Y increases the digit, N decreases the digit,
and @ moves to the next digit. After the third digit is set, the question
mark (“?”) blinks. Press Y to accept the changes. Press N to discard the
changes. Press @ to change the due date.
Calibration Period
The EntryRAE should have a fresh air calibration before each use and
should be fully calibrated:
 no less than every 30 days
 if it does not pass a fresh air reading
 if it does not pass a field calibration.
Operating the EntryRAE 15
EntryRAE Usage Overview
After the EntryRAE is warmed up, calibrated, and charged, it is ready for
use to enhance your personal safety.
CAUTION: HIGH OFF-SCALE READINGS MAY INDICATE AN
EXPLOSIVE CONCENTRATION.
Storage
Always keep the EntryRAE on its charging
cradle in a dry indoor area when it is not in use.
See page c for long-term storage warnings and
procedures.
Confined Space Pre-Entry Test
It is important to always test the atmosphere inside a confined space before
entering. Be aware of any hazardous chemicals you may be bringing into
the space. Remember that many cleaners, paints, adhesives, degreasers
and other modern industrial products (even treated wood) contain volatile
organic compounds that may be hazardous to your health and safety.
Test the atmosphere in the confined space by sampling air at three levels
(top, middle and bottom) in the confined space. Give the instrument time
to sample the gas at each level - the correct sampling time is 60 seconds
plus 1 second for every foot of tubing.
Alarms
If there are any gas ALARMS at any level in the confined space, that
space is not safe. DO NOT ENTER! Identify the alarm condition and then
start your preventive actions according to your company’s confined space
entry procedures.
No Alarms
If there are no alarms, the confined space may be safe to enter. Disconnect
the tubing before carrying the EntryRAE into the confined space. If
monitoring is to be done by a confined space attendant, they should
continue to monitor while you enter the area. If the monitor alarms and
“BAT” is displayed, the battery needs to be charged: There are 15 minutes
16 Operating the EntryRAE
or less of run time remaining. Leave the Confined Space immediately!
(Refer to the Applications and Technical Notes Guide, P/N 000-4001-000,
for more information regarding confined space entry and other applications.)
After Usage
 Press the @ button for 5 seconds.
 The display counts down from 3 to 1
(3 seconds).
 Release @ when “Unit Off” appears on the
display.
 Return the EntryRAE to the dry, indoor storage
area and put it in its charger.
Alarm Signals
When an alarm condition occurs, the monitor provides audible, visible
and vibration alarms to alert users of unsafe conditions.
Auto Reset and Latched Alarm
The EntryRAE comes from the factory with “Auto Reset Alarm” turned
on. This means that the alarms cease when the alarm condition is no
longer present. Alternatively, this feature can be changed and set up as a
“Latched Alarm.” When the alarm system is “Latched,” Y must be pressed
to acknowledge the alarm condition and to reset the alarms.
The Auto Reset function does not automatically reset if a sensor fails
calibration and goes into alarm. The sensor must be successfully
calibrated in order to clear this alarm.
Testing Alarm Signals
Press Y while in monitoring mode. If the alarm is functional, the buzzer
beeps once (if the monitor is not set to run silently; see page 24), the
alarm LED flashes once, and the vibration alarm turns on and off.
Operating the EntryRAE 17
Alarm Conditions
Alarm
Condition
Beep/sec
Flash/sec
Vibration
Alarm
Screen Message
Reset Alarm
Pump
Failure
3
Yes
alarm symbol on, pump
symbol blinks
unblock inlet, press Y
to restart pump
Lamp
Failure
3
Yes
alarm symbol on, lamp
symbol blinks
turn off/on or wait
Over Range
3
Yes
alarm symbol on, OVER
symbol on, blink “999”
move away from gas
Max
3
Yes
alarm symbol on, OVER possible sensor failure;
symbol on, blink reading
cal sensor to verify
High
3
Yes
alarm and HIGH symbol
on, sensor name blinks
move away from gas
Low
2
Yes
alarm and LOW symbol
on, sensor name blinks
move away from gas
TWA
1
Yes
alarm and TWA symbol
on, sensor name blinks
move away from gas &
turn unit off/on
STEL
1
Yes
alarm and STEL symbol move away from gas &
on, sensor name blinks
wait 15 min.
Negative
1
Yes
alarm symbol on, blink
“nEg”
perform zero
calibration
No
battery symbol blinks
charge battery
Battery Low 1 beep/min
Preset Alarm Limits and Calibration
Unit
TWA
STEL
LOW
HIGH
CO
Cal Gas/
Balance
50/Air
ppm
35
100
35
200
H 2S
10/N2
ppm
10
15
10
20
Gas
O2
20.9/N2
%Vol
-
-
19.5
23.5
LEL
50/Air
%LEL
-
-
10
20
VOC
100/Air
ppm
-
-
-
100
Back Light
The back light makes it easy to read the display in poor lighting conditions.
It automatically turns on in low lighting conditions.
NOTE: Heavy back light usage can shorten battery life 20% to 30%.
18 Operating the EntryRAE
Sampling Pump
Under normal operating conditions, when the monitor is not in an alarm
state, the pump cycles on and off about every eight seconds. This on/off
cycle improves the reliability of the PID and saves battery life. Press the Y
key and the N key simultaneously and the pump will run continuously for
five minutes. The pump also runs continuously when:
1.0 The concentration of any measured gas or vapor triggers an alarm
condition.
2.0 The concentration of VOCs approaches alarm conditions.
The monitor can detect any obstructions in the external filter that causes
a pump stall. The alarm activates and the pump symbol blinks in the
upper right corner of the display. Clear the obstruction.
Press Y to start the pump again.
Datalogging
Datalogging occurs automatically for all enabled sensors at one-minute
intervals, allowing five days' continuous data storage. When the end of
the storage memory is reached, the monitor automatically wraps around
to the beginning of the log and overwrites the oldest stored data.
NOTE: Datalogging pauses when the monitor goes into Programming
Mode.
Charging the EntryRAE
The battery charge is indicated on the EntryRAE screen by the battery
charge status icon on the monitor’s LCD. When the icon’s segments are
filled in, the battery is charged. As the battery is discharged, the segments
disappear, indicating that the unit should be charged before use. Refer to
“Maintenance” for instructions on changing/removing the battery.
The Li-Ion battery pack can be charged alone or while it is installed in
the monitor. To charge an installed battery, simply plug the transformer
into the cradle and into an AC electrical outlet. Place the monitor in the
cradle. �����������������������������������������������
The monitor (or battery) is fully charged when:
 The LED on the charging cradle is green
 The “Fully Charged” message appears on the display
 The segments of the battery icon are blinking.
Operating the EntryRAE 19
The monitor (or battery) is still in charging mode when:
 The LED on the charging cradle is red
 The “Monitor Charging” message appears on the display
 The segments of the battery icon are scrolling.
Figure 6. Installing the battery pack into the charging cradle.
To charge the battery pack
alone, place the battery
pack on the charging cradle,
matching the contacts on the
battery pack to the contacts
on the charging cradle.
Figure 7. Contacts on the battery pack.
20 Operating the EntryRAE
The Li-Ion Battery Pack
The factory-supplied Li-ion battery pack is designed to last for 16 hours
of normal operation between charges (without alarm or back light
conditions). The rechargeable batteries have a 1-year warranty. Age,
ambient temperature, and heavy usage may impact battery life.
Battery packs slowly drain even if the monitor is turned off. Therefore,
it is strongly recommended that the EntryRAE be stored on its charging
cradle. The battery can trickle-charge when not in use.
If the battery packs have not been charged for 10 days, the battery
voltage will be low. Fully charge battery packs before going into the field,
and recharge them upon returning from the field.
To reduce the risk of ignition of hazardous atmospheres, recharge
battery only in areas known to be non-hazardous. Remove and replace
battery only in areas known to be non-hazardous.
Ne charger les batteries que dans emplacement désignes non
dangereux.
Programming Mode
In addition to calibration, authorized users may change the monitor
settings to their requirements using the Programming Mode.
Note: Monitoring gas concentrations pauses during Programming Mode
and during calibration. Datalogging also pauses during Programming
Mode, but resumes when programming is finished.
Security
EntryRAE provides the option of securely enabling a password to
access Programming Mode. The default password is “0000.” To enable
password access, change the “0000” to a desired password. (Refer to
“Change Pass?” on page 26.)
Programming Mode 21
Programming Menus
22 Programming Mode
Entering Program Mode
1. Press @ and N together until the first program
menu, “Fresh Air Calibration?” appears.
2. Press N to cycle through the submenu options.
3. Press Y to enter a submenu.
4. To modify a value, use Y to increase and N to
decrease.
5. Use @ to move from character to character and/or
from sensor to sensor.
6. When the “?” appears at the bottom of the display,
press Y to save changes (or N to cancel changes)
and exit the submenu.
7. To discard changes or exit the Programming Mode,
press @ to return to the Reading Display.
Change Monitor Setup
The following submenus represent the programmable setup of the
monitor.
Fresh Air Calibration? Refer to “Fresh Air Calibration”
on page 11.
Span Calibration CO,
�����
H2S, LEL�
? Refer
���������������
to “Span
Gas Calibration for Multiple Sensors” on page 11 to
calibrate the CO, H2S and LEL sensors.
Programming Mode 23
Span Calibration VOC? Refer to “Span Gas Calibration
for the PID Sensor” on page 13.
Zero Calibration? Refer to “Zero Calibration for the
Oxygen Sensor” on page 14.
Calib Due Day? Use this feature to set a reminder to
calibrate regularly. Enter the number of days between
calibrations per your company’s standards. After 30 days
of no calibration, the monitor sounds. Once the monitor
is calibrated, the counter resets.
Set Type? Choose either automatic alarm reset or
latched (constant) alarm. To acknowledge and reset a
latched alarm, press Y.
Set Run Silent? Press Y to let the monitor run in silent
mode. In this mode, the audible alarm does not sound
during alarm conditions.
Press N to let the audible alarm sound during alarm
conditions.
Adjust Alarm Limits?
HIGH - The highest reading for each gas concentration
since the monitor was turned on; updated every second.
Press Y to reset the high values.
24 Programming Mode
LOW – The lowest reading for each gas concentration
since the monitor was turned on, updated every second.
Press Y to reset the minimum values.
TWA - The Time Weighted Average for CO, H2S, and
VOC gases only; the accumulated reading of the gas
concentration, divided by 8 hours, since the monitor was
turned on. Updated every minute.
STEL - The Short Term Exposure Limit for CO, H2S,
and VOC gases only; the average reading of the gas
concentration for the last 15 minutes, which is updated
every minute. Dashes “- - -” appear in the display for the
first 15 minutes.
Enable/Disable Sensor? Enable or disable sensor(s);
a disabled sensor does not measure or display the
gas concentration. Use this if a sensor has failed or is
providing erroneous readings. Use @ to move from
sensor to sensor. “Yes” means the sensor is enabled;
likewise, “No” means the sensor is disabled. Press Y to
enable or press N to disable.
Adjust Date? Set the date and time using the standard
24-hour format.
Programming Mode 25
Set LEL Unit? The EntryRAE provides the option of
displaying the concentration of combustibles gases in
%LEL or %Vol.
Change Pass? (Password) Set a four-digit password,
which is used to access Programming Mode. (Refer to
“Security” on page 21.)
Secure Beep? If “yes,” a security beep sounds every
minute when the monitor is on.
26 Programming Mode
Maintenance
Replacing the Li-Ion Battery Pack
To reduce the risk of ignition of hazardous atmospheres, recharge the
battery only in areas known to be non-hazardous. Remove and replace
the battery only in areas known to be non-hazardous.
Ne charger les batteries que dans emplacements désignes non
dangereux.
Do not use the battery outside the operating temperature range of -4º to
122º F / -20º to 50º C (UL/cUL) or -4º to -116º F / -20º to 47º C (ATEX).
Ne faites aucune utilisation l’extérieur de batterie la gamme de
température de fonctionnement entre -4º à 122º F / -20º à 50º C (UL/cUL)
or -4º à -116º F / -20º à 47º C (ATEX).
Do not store the battery outside the temperature range of -4º to 122º F
(-20º to 50º C).
Ne faites aucune magasin la batterie en dehors de la température entre
-4º à 122º F (-20º à 50º C).
case screws for battery compartment
Figure 8. Battery Compartment Screw Locations
The battery compartment is located on the back of the monitor.
Remove the battery compartment cover by loosening the three case
screws with a screwdriver or a coin.
Maintenance 27
Remove the battery pack from the monitor.
Figure 9. Removing the Battery Pack Cover.
Figure 10. Removing the Li-Ion Battery Pack.
Replace the battery compartment cover and tighten the screws.
The spent battery pack may be charged on the charging cradle by itself.
(Refer to “Charging the Li-Ion Battery Pack” on page 19.)
28 Maintenance
Emergency Alkaline Battery Adapter
Figure 11. Installing the Alkaline Battery Pack.
Use the adapter when there isn't time to recharge
the Li-Ion battery pack. The adapter (P/N 0463051-000) accepts four AA alkaline batteries (use
only Duracell MN1500 or Energizer 91) to provide
approximately 12 to 14 hours of operation.
Remove the Li-Ion battery pack from the monitor.
Install four AA alkaline batteries into the battery
adapter, making sure the battery polarity is correct.
Replace the battery cover and tighten the screws.
The monitor automatically detects the alkaline batteries.
Maintenance 29
Sensor Replacement
Under normal operating conditions, most sensors lose their original
sensitivity after the expected operating life and need to be replaced.
Warranties: The oxygen (O2), combustible gas (LEL), hydrogen sulfide
(H2S) and carbon monoxide (CO) sensors all have a 2-year warranty.
The PID sensor has 1 year warranty.
Replace a sensor when it fails to calibrate.
The sensors are located inside the front of the monitor.
See Technical Note TN-114, Sensor Specifications And CrossSensitivities, for additional information, available at www.raesystems.
com.
PID
H 2S
O2
CO
LEL
Figure 12. Sensor Locations.
30 Maintenance
Figure 13. Loosening rear
screws.
Figure 14. Pushing screws to
remove sensor cover.
To replace a sensor:
1. Turn the monitor off.
2. Remove the front sensor cover by
loosening the two screws on the back
of the monitor below the battery cover.
See Figure 13.
3. Push the screws from the back of the
monitor to pop the cover off the front
(Figure 14).
4. Using the sensor puller, remove the
sensor by carefully pulling straight out
(Figure 15).
The colored dots marked next to each
sensor socket should match the color
of the sensors.
Figure 15. Using
sensor puller to
handle sensor.
5. To install new sensors, re-align the
dots on the sensors with the dots on
the monitor.
6. Press the sensors all the way into the
socket.
7. Replace the monitor cover and tighten
the screws in the back.
Maintenance 31
8. Turn the monitor on and the newly installed sensors should be
properly identified by the EntryRAE in the start-up screen. Let the
monitor run for 15 minutes before calibration.
9. Calibrate all sensors prior to use.
CO Sensor Charcoal Filter
The carbon monoxide (CO) sensor may be sensitive to hydrocarbons, so
use a charcoal filter (P/N 008-3006-005) to reduce or eliminate organic
vapor contamination or cross-sensitivity. The charcoal filter is installed in
the gas plate above the CO sensor. Under normal operation conditions,
the charcoal filter lasts from four to six weeks before it needs to be
replaced. If the monitor has been exposed to high concentrations of
VOC gases for a prolonged period of time, the carbon filter needs to be
replaced more frequently.
Place the carbon filter above the CO sensor on the gas plate.
NOTE: The CO charcoal filter lowers the reading if it is accidentally used
on other sensors.
CO Sensor
Charcoal filter location
Figure 16. CO and charcoal filter locations.
32 Maintenance
Cleaning the PID
If the PID does not calibrate, it may need to be cleaned. Be sure to use
the specially shaped swab included in the PID Cleaning Kit (p/n 5000014-010), shown in the Equipment List, to clean the PID.
1. Open the sensor compartment.
2. Remove sensor with sensor puller (follow procedure on page 31).
3. Using a swab, drip a little bit methanol on the lamp window through
the openings on the top of the PID.
4. Use the swab to carefully clean the lamp window through the two
openings.
5. Leave the PID to dry by itself for five minutes.
Figures 19. Cleaning the PID.
Pump Replacement
The pump in the EntryRAE is not field replaceable. Call RAE Systems
Technical Support or an authorized service center when the pump needs
to be replaced.
Maintenance 33
Setting Pump Stall Threshold
Note: This section should only be used by authorized technical users.
Monitoring continues while EntryRAE is in Diagnostic Mode.
Diagnostic mode allows users to set the pump stall setpoint. This should
only be done if your EntryRAE is going into a pump alarm when the
pump is not blocked, or if the EntryRAE does not go into pump alarm
when you block the pump inlet by holding your finger over it.
Entering Diagnostic Mode
1. Press @ and Y simultaneously until the first
diagnostic screen blanks. Then release the keys
and the values reappear.
2. Press @ 7 times to cycle through the diagnostic
screens until you reach the pump diagnostic
screen. The first, CO, H2S, O2, LEL, VOC, and run
time screens are used for factory operation and
testing only. The pump diagnostic screen displays
a high and low pump rate. The first digit of the last
number blinks.
3. To set the pump stall alarm, write down the “Hi”
value as found, and then block the pump inlet
while the pump is running (about 5 seconds), until
the “Hi” reading stabilizes.
4. Write down the new “Hi” value. Add the two “Hi”
values together, and then divide by two to get the
average value.
5. Enter the average into the bottom number on the
screen using the Y and N keys to increase and
decrease the digits, and @ to move between
digits.
6. Once the average is entered, press @ until you
see the blinking “?”. Press Y to save the pump
stall threshold value.
7. Exit the diagnostic mode by pressing @ and Y
together until the screen changes. Gas readings
may take a second to appear.
34 Maintenance
Specifications
Size 5.9" L x 3.3" W x 1.9" H (15 x 8.5 x 1.9 cm) without clip
Weight 20 oz with battery (567 g) and clip
Sensors (5) Protected catalytic bead for combustible gases
Electrochemical sensors for oxygen and toxic gases
Photoionization detector for VOC broadband detection using 10.6 eV lamp
Battery Drop-in rechargeable Li-ion battery pack
Standard alkaline battery adapter
Charging cradle doubles as external battery changer
Operating Hours 16 hours continuous with Li-ion (typical), 12 hours with alkaline
Battery Operation -4º to 122º F (-20º to 50º C) UL/cUL
Temperature -4º to 116º F (-20º to 47º C) ATEX
Battery Storage -4º to 122º F (-20º to 50º C)
Temperature
Display Large 1.4" x 1.8" (3.5 x 4.5 cm) segmented display with LED back light
Keypad Three-button operation
Direct Readout Instantaneous for 5 values:
Oxygen as percentage by volume
Combustible gas as %LEL or %Vol
VOCs, CO and H2S as parts per million
TWA and STEL values for VOCs, CO and H2S
High and low values for all gases
Battery life displayed as icon in 1/4 increments
Alarms Audible: 95dB at 30 cm
Visible: Bright LED bar visible from top, front and sides
Sensory: Built-in vibration alarm
High: 3 beeps and flashes per second
Low: 2 beeps and flashes per second
STEL and TWA: 1 beep and flash per second
Low battery displays empty battery symbol, beep per second
EMI/RFI Highly resistant to EMI/RFI. Compliant with EMC directive 89/336/EEC
IP Rating IP-55: protected against dust, protected against low pressure jets of
water from all directions
Datalogging Wrapping 120 hours (five days) of data at one minute intervals for all
five sensors
Communication PC-to-EntryRAE via RS-232 through charging cradle, with USB adapter
Calibration Two-point field calibration for zero and span gas
Sampling Pump Internal pump, flow rate typical 200 cc per minute
Low Flow Alarm Auto shut-off at low-flow condition
Minimum Pump 100 cc per minute
Flow
Operation -4o to 122o F (-20o to 50o C), UL/cUL
Temperature -4o to 116o F (-20o to 47o C), ATEX
Storage -4o to 122o F (-20o to 50o C)
Temperature
Humidity 0% to 95% relative humidity (non-condensing)
Pressure 1 Atmosphere ±10%
Attachments Stainless steel alligator clip (installed), wrist strap
Warranty Lifetime on non-consuming components (per RAE Standard Warranty),
2 years for O2, LEL, CO, and H2S sensors, 1 year for PID, 1 year for
pump, 1 year for battery
Specifications 35
Service and Repair Record
Date
Service / Repair Type
Specifications 37
Thank you for reading this data sheet.
For pricing or for further information, please contact us at our UK Office, using the details
below.
UK Office
Keison Products,
P.O. Box 2124, Chelmsford, Essex, CM1 3UP, England.
Tel: +44 (0)1245 600560
Fax: +44 (0)1245 808399
Email: [email protected]
Please note - Product designs and specifications are subject to change without notice. The user is responsible for determining the
suitability of this product.

advertisement

Key Features

  • Reliable self-cleaning PID
  • Includes CO, H2S, LEL and O2 sensors
  • Built-in pump
  • Large, easy-to-read display
  • Datalogging
  • Visual alarm with bright red flashing LEDs
  • Loud audible alarms
  • Rugged weather-resistant composite case

Frequently Answers and Questions

What types of gases can this monitor detect?
The EntryRAE can detect carbon monoxide, hydrogen sulfide, combustibles, oxygen, and volatile organic compounds.
How often does the monitor need to be calibrated?
The EntryRAE should be calibrated every 30 days, and before each use.
How long does the battery last on a single charge?
The factory-supplied Li-ion battery pack is designed to last for 16 hours of normal operation between charges.
What are the alarm conditions?
The EntryRAE has several alarm conditions, including high, low, TWA, STEL, pump failure, lamp failure, over range, max, negative, and battery low.
How do I reset the alarms?
The alarms can be reset by pressing the Y key.

Related manuals

Download PDF

advertisement