Operating instructions
METTLER TOLEDO MultiRange
ID7sx-Sys weighing terminal
www.mt.com/support
Contents
ID7sx-Sys
Contents
Page
Operating instructions 22016401
1
Safety precautions .......................................................................
4
2
2.1
2.2
2.3
2.4
2.5
2.6
Introduction and commissioning ...................................................
Documentation .............................................................................
Applications .................................................................................
ID7sx-Sys weighing terminal..........................................................
Commissioning ............................................................................
Cleaning......................................................................................
Disposal ......................................................................................
5
5
5
6
8
8
8
3
3.1
3.2
3.3
3.4
3.5
Basic functions ............................................................................ 9
Setting to zero .............................................................................. 9
Taring ......................................................................................... 9
Weighing ..................................................................................... 10
Working with several weighing platforms......................................... 10
Displaying versions ...................................................................... 11
4
4.1
4.2
4.3
Additional functions .....................................................................
Display ID code and test weighing platform .....................................
Recall information.........................................................................
Recall data from Alibi Memory........................................................
12
12
13
14
5
5.1
5.2
5.3
5.4
5.5
Settings in the master mode ........................................................
Overview of the master mode .........................................................
Operating the master mode ............................................................
TERMINAL master mode block........................................................
SCALE master mode block .............................................................
INTERFACE master mode block ......................................................
17
17
18
20
22
25
6
What to do if....? ......................................................................... 27
7
7.1
7.2
Technical data and accessories .................................................... 29
Technical data.............................................................................. 29
Accessories.................................................................................. 31
8
Index .......................................................................................... 33
06/09
3
Safety precautions
1
ID7sx-Sys
Safety precautions
The ID7sx-Sys weighing terminal is approved for operation in zone 1 and 21
hazardous areas. It may only be used in areas in which the causes of static
electricity build-up, which lead to propagating brush discharges, have been
eliminated.
If the ID7sx-Sys weighing terminal is used in hazardous areas, special care must be
taken. The code of practice is oriented to the "Safe Distribution" concept drawn up by
METTLER TOLEDO.
Competence
▲ The weighing system may only be installed, maintained and repaired by
authorised METTLER TOLEDO service personnel.
▲ The mains connection may only be connected or disconnected by the owner’s
electrician.
Ex approval
▲ For the exact specification please refer to the statement of conformity.
▲ No modifications may be made to the terminal and no repair work may be performed on the modules. Any weighing platform or system modules that are used
must comply with the specifications. Non-compliant equipment jeopardises the
intrinsic safety of the system, cancels the Ex approval and renders any warranty or
product liability claims null and void.
▲ The safety of the weighing system is only guaranteed when the weighing system
is operated, installed and maintained in accordance with the respective instructions.
▲ Also comply with the following:
– the instructions for the system modules
– the regulations and standards in the respective country
– the statutory requirement for electrical equipment installed in hazardous areas
in the respective country
– all instructions related to safety issued by the owner
▲ The explosion-protected weighing system must be checked to ensure compliance
with the requirements for safety before being put into service for the first time,
following any service work and every 3 years, at least.
Operation
▲ Prevent the build-up of static electricity. Therefore:
– always wear suitable working clothes when operating or performing service
work on the system,
– do not rub or wipe off the keyboard surface with a dry cloth or glove.
▲ Do not use protective hoods.
▲ Prevent damage to the weighing terminal. Hairline cracks in the keyboard
membrane are also considered damage.
4
Operating instructions 22016401
06/09
Introduction and commissioning
ID7sx-Sys
2
Introduction and commissioning
2.1
Documentation
These operating instructions describe the operation of the ID7sx-Sys with the basic
functions of the Sys-690 application software and all the possible interfaces.
Information on your customer-specific ID7sx-Sys application can be found in the
corresponding operating instructions specially authored for you.
2.2
Applications
With the ID7sx-Sys weighing terminals the following applications are possible:
2.2.1
ID7sx-Sys with PSU power supply unit
RS232
CL
ID7sx
CL
I/O
PSU
K...x / DN...x/
PUA...x/PTA...x
Point Ex
system solution
PBA430x/
Analog scale
In this configuration the following possibilities are available:
• Multi-scale operation with up to 3 weighing platforms (K...x, DN...x, PUA...x,
PTA...x, PBA430x or system solution Point Ex)
• Up to 2 data interfaces and one I/O interface
– for printing,
– for data exchange with a computer,
– for control, e.g. of valves or flaps.
• Alibi memory module for storing the individual configuration data as well as
weighing data.
Operating instructions 22016401
06/09
5
Introduction and commissioning
2.2.2
ID7sx-Sys
ID7sx-Sys with PSUx power supply unit
ID7sx
PSUx
Only one weighing platform (K...x, DN...x, PUA...x, PTA...x, PBA430x or Point Ex
system solution) is permissible in this configuration.
2.3
2.3.1
ID7sx-Sys weighing terminal
Display
1
2
MODE
3
4
5
6
7
6
1
Weight display BIG WEIGHT® with sign and decimal point
2
Stability monitor: lights up until the weighing platform has levelled out, then the
weight unit appears here
3
Range display for multi-range weighing platforms
4
Number of the weighing platform: shows the weighting platform just selected
5
Symbol * for identifying weight values in the second unit or in a higher
resolution
6
NET symbol for marking net weight values
7
Assignment of the function keys
Operating instructions 22016401
06/09
Introduction and commissioning
ID7sx-Sys
2.3.2
Keypad
1
CODE A ... CODE D keys – enter identification data
2
FUNCTION CHANGE key – display additional functions
when entering weight values: switch over unit
3
INFO key – recall memory contents and system information
4
Function keys F1 … F6 – the current assignment is shown in the display above
the key
5
SCALE key – select scale
6
ZERO-SET key – set scale to zero, test scale
7
TARA key – tare scale
8
TARE SPECIFICATION key – enter known tare values numerically
9
CLEAR key – clear entries and values
10 ENTER key – accept and transfer data
11 Numeric keypad with decimal point and signs
Operating instructions 22016401
06/09
7
Introduction and commissioning
2.4
2.4.1
2.5
ID7sx-Sys
Commissioning
Control drawings and guides for installers
The explosion-protected weighing system with the ID7sx-Sys weighing terminal may
only be installed according to the respective guide for installers and the
accompanying control drawing, depending on the power supply unit.
Component
Guide for installers
Control drawing
ID7sx
ME-22008316
PSU
ME-22006472
PSU-ID7sx
ME-22006478
PSUx/230V
ME-22006386
PSUx/230V-ID7sx
ME-22006397
PSUx/120V
ME-22006395
PSUx/120V-ID7sx
ME-22006399
Cleaning
DANGER OF SHOCK
➜ Do not open ID7sx-Sys weighing terminal to clean.
CAUTION
➜ Make sure that unused connection sockets are covered with the corresponding
protective caps. Ensure the correct position of the seals when doing so.
➜ Do not use high-pressure cleaners.
Cleaning
➜ Wipe off ID7sx-Sys weighing terminal with a commercially available glass or
plastic cleaner.
➜ When used in category 2, zone 21 hazardous areas, the weighing terminal must
be cleaned regularly. See also standard EN 50281-1-1 / 50281-1-2.
2.6
Disposal
In conformance with the European Directive 2002/96 EC on Waste Electrical and
Electronic Equipment (WEEE) this device may not be disposed of with domestic
waste. This also applies to countries outside the EU, per their specific requirements.
➜ Please dispose of this product in accordance with local regulations at the
collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor
from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the
content of this regulation must also be related.
Thank you for your contribution to environmental protection.
8
Operating instructions 22016401
06/09
Basic functions
ID7sx-Sys
3
Basic functions
3.1
Setting to zero
Setting to zero corrects the influence of minor dirt on the load plate.
In the case of excessive dirt which cannot be compensated by setting to zero, the
display shows OUT OF RANGE.
Manual zero set
1. Relieve weighing platform.
2. Press ZERO-SET key.
The display shows 0.000 kg.
Automatic zero set
3.2
3.2.1
On certified weighing platforms the zero point of the weighing platform is automatically corrected when the weighing platform is relieved.
The automatic zero set can be switched off in the master mode on non-certified
weighing platforms.
Taring
Manual taring
1. Place empty container on scale.
2. Press TARE key.
The tare weight is saved and the weight display set to zero.
The display shows the NET symbol.
Notes
• When the weighing platform is relieved, the saved tare weight is displayed with a
negative sign.
• The weighing platform only saves one tare value.
3.2.2
Enter numerically
Specify tare weight
This function is only available when ordered expressly.
1. Press TARE SPECIFICATION key.
2. Enter tare weight (container weight) and confirm with ENTER.
When weighing platform is relieved, the entered tare weight is displayed with a
negative sign.
Note
With the FUNCTION CHANGE key you can select the weight unit for entering the tare
weight.
Correct entry
Operating instructions 22016401
➜ Clear the entry character by character with CLEAR key and repeat correctly.
06/09
9
Basic functions
ID7sx-Sys
3.2.3
Recall currently saved tare weight
The saved tare weight can be recalled at any time.
➜ Enter INFO, TARE SPECIFICATION key sequence.
The saved tare weight is displayed.
3.2.4
Clear tare weight
➜ Relieve weighing platform and tare.
– or –
➜ Recall tare weight and press CLEAR key.
3.3
Weighing without taring
Weighing with taring
Weighing
➜ Lay weighing sample on weighing platform.
Gross weight (total weight) is displayed.
1. Place the empty container on the weighing platform and tare.
2. Pour in weighing sample.
The display shows the net weight and the NET symbol.
Weighing with tare
specification
1. Place filled container on weighing platform.
The display shows the gross weight (total weight).
2. Specify tare weight or recall tare memory.
The display shows the net weight (container content) and the NET symbol.
3.4
Working with several weighing platforms
Up to 3 weighing platforms can be connected to the ID7sx-Sys.
Depending on the setting in master mode, only the currently active scale appears in
the display (serial Multi-scale mode) or all scales are operated at the same time
(parallel multi-scale mode). A constantly updated sum scale is also available in
parallel multi-scale mode.
3.4.1
Switch over weighing platform
The weighing platform currently selected is shown on the terminal.
➜ Press SCALE key.
The next weighing platform is selected.
– or –
➜ Enter number of weighing platform and press SCALE key.
The desired weighing platform is selected.
Note
The sum scale can only be operated non-verifiably. It is therefore identified by the
symbol Σ.
10
Operating instructions 22016401
06/09
Basic functions
ID7sx-Sys
3.5
Displaying versions
When connected to a power supply, the versions of the ID7sx-Sys and the installed
components are displayed.
Version code for the ID7sx-Sys with customer-specific application program
METTLER-TOLEDO ID7sx-Sys
ID7sx 0.13
CC-AA-XXXXXXXX-WV-DD.MM.YYYY
CC
AA
XXXXXXXXXX
VVV
DD.MM.YYYY
Operating instructions 22016401
06/09
Country code, e.g. DE = Germany
Software type, e.g. 00 = Base
Order No.
Software version
Date of manufacture, day.month.year
11
Additional functions
4
ID7sx-Sys
Additional functions
The assignment of the 6 function keys of the ID7sx-Sys weighing terminal differs
depending on the weighing task. The current assignment is shown above the function
keys.
Independent of the customer specific application software, the ID7sx-Sys has the
following additional functions:
4.1
INFO
MODE
Display
information,
see 4.2
Activate
master mode,
see Chapter 5
Display ID code and test weighing platform
Each time the weighing platform configuration is changed the ID code counter is
increased by 1. On certified weighing platforms the displayed ID code must match
the ID code on the ID code sticker, otherwise the calibration is no longer valid.
Display ID code
➜ Press ZERO SET key and hold until IDENTCODE = ... appears in the display.
Test weighing platform
➜ Press ZERO SET key again.
The connected weighing platform is checked. The display shows CHECK SCALE
and then SCALE IS OK after completing the test.
Note
If weighing platform is defective, display shows SCALE ERROR.
12
Operating instructions 22016401
06/09
Additional functions
ID7sx-Sys
4.2
Recall information
On the weighing terminal memory contents and system information can be recalled.
1. Press INFO key.
Then the following function key assignment appears:
DATE
TARE
SCALE
ALIBI
END
VERS
Display date
and time
Display tare
weight
Call up the
data of the
current scale:
Version,
max. load,
ID code, etc.
Recall content
of alibi
memory, see
section 4.3.
This selection
only appears
when Alibi
Memory-ID7sx
is installed.
Exit
information
mask
Display
version
numbers of
installed
software
modules
W&M
PERFO
ERROR
Display
checksum of
the software
relevant to
calibration.
The correct
checksum is
documented
in the calibration approval.
System
utilisation
Fault /
Event memory
display
2. Select desired information.
The information is displayed for about 5 seconds, then the weighing terminal
changes to the weighing mode again.
Note
When several values are displayed, the ID7sx-Sys automatically changes to the next
value after approx. 5 seconds.
Operating instructions 22016401
06/09
13
Additional functions
ID7sx-Sys
4.3
Recall data from Alibi Memory
With the Alibi-Memory-ID7sx memory module you can fulfill your recording
obligations in certified operation without having to archive paper.
Alibi-Memory-ID7sx automatically assigns every weighing operation a consecutive
data record number that also appears on the printout, saves the net and tare value
and also the date and the time.
The Alibi-Memory-ID7sx operates according to the principle of a ring memory: When
the capacity limit of 12 Mbytes is reached, the oldest record block having the size of
64 Kbytes is deleted and overwritten by the data of the subsequent weighings.
By entering suitable search criteria you can quickly access the data of a very specific
weighing.
4.3.1
Initiate
➜ Press INFO, ALIBI key sequence.
The function keys change to the following assignment:
4.3.2
FIND
>>...
<
>
->Num
END
Enter search
criteria
Search for
next
matching
data record
starting with
oldest
Show the
data record of
the weighing
carried out
directly
beforehand
Show the
data record of
the weighing
carried out
directly
afterwards
Search for
data record
with known
data record
number
Exit Info Alibi
and return to
normal mode
Fast search with entry of data record number
1. Press ->Num key.
2. Enter number of data record to be searched for and confirm with ENTER.
Alibi-Memory-ID7sx now searches for the desired data record.
Notes
• The search may take up to 10 seconds.
• If no data record with the entered number is found, the message NO MATCHING
DATA RECORD appears.
14
Operating instructions 22016401
06/09
Additional functions
ID7sx-Sys
4.3.3
Search with other search criteria
➜ Press FIND key.
The function keys are given the following assignment:
DATE
TIME
NET
TARE
START
END
Enter
date as
search
criterion
Enter
time as
search
criterion
Enter net
value as
search
criterion
Enter
tare value as
search
criterion
Start search
with entered
search
criteria
Terminate
search
All offered search criteria can be combined with each other.
The entered search criteria are shown in the display in clear text.
This enables you to search for a find a specific weighing.
Enter date
➜ Press DATE key and enter complete date in DD.MM.YY form.
Enter time
➜ Press TIME key and enter desired time in one of following formats.
Format HH
all weighings between HH.00.00 and HH.59.59 are found
Format HH.MM
all weighings between HH.MM.00 and HH.MM.59 are found
Format HH.MM.SS only the weighing at the time HH.MM.SS is found
Enter net/tare value
1. Press NET or TARE key.
2. Enter weight value and confirm with ENTER.
The function key assignment changes back again for selection of the search
criteria.
Note
With the FUNCTION CHANGE key you can select the weight unit for entering the tare
weight.
Start search
➜ Press START key.
Alibi-Memory-ID7sx searches for the oldest data record which meets the entered
search criteria.
Notes
• The search may take up to 10 seconds.
• If no data record with the entered values is found, the message NO MATCHING
DATE RECORD appears.
• If no search criterion has been entered, the oldest data record is displayed.
Operating instructions 22016401
06/09
15
Additional functions
ID7sx-Sys
4.3.4
Displaying data records
Found data records are shown in the display.
DATE:
NUM:
NET:
TARE:
Scroll forward/back
02.04.06
000987
25.000 KG
100.346 KG
TIME
09.25.25
The keys >>..., < and > enable you to scroll within the found data records.
Note
When, during scrolling with the key >>... all entries of the Alibi-Memory-ID7sx have
been searched through, the message END OF FILE appears.
16
Operating instructions 22016401
06/09
Settings in the master mode
ID7sx-Sys
5
Settings in the master mode
5.1
Overview of the master mode
In the master mode you adapt the ID7sx-Sys weighing terminal to meet your needs.
Depending on the configuration, the master mode is divided into 4 or 5 master mode
blocks, which are in turn divided into further blocks.
TERMINAL
PAC
For system settings, such as entering the date and time or loading permanent texts,
see section 5.3.
To set application-specific parameters.
SCALE
To select one of the connected weighing platforms. For each selected weighing
platform the parameters are then set which concern the weight value, e.g. stability
detector, unit, etc., see section 5.4.
INTERFACES
To select an interface. The communication parameters are then set for each interface,
see section 5.5.
SERVICE
For configuring the weighing platform(s).
On IDNet weighing platforms only for METTLER TOLEDO service technicians.
On weighing platforms with an analog signal output, see service manual A/D
converter Point ME-22004256.
Operating instructions 22016401
06/09
17
Settings in the master mode
5.2
5.2.1
ID7sx-Sys
Operating the master mode
Enter the master mode
1. Press MODE key.
If the current function keys assignment does not contain MODE, press the
FUNCTION CHANGE key as often as necessary until the MODE key appears.
2. Enter personal code if configured.
The display shows the first master mode block TERMINAL.
5.2.2
Assignment of function keys in the master mode
Assignment on the top level
On the top level of the master mode the function keys are assigned as follows:
←
→
↑
END
OK
Change to
previous block
within a level
Change to next
block within a
level
Exit level and
return to
higher-level
block
Exit the master
mode and
return to
normal mode
Recall lowerlevel block or
confirm
selection
➜ Select the function by pressing the function key.
When the function keys are otherwise allocated
➜ Press the FUNCTION CHANGE key repeatedly until the function key assignment
shown above appears.
Example
5.2.3
Example
18
➜ Press the END key to exit the master mode and return to the normal mode.
Orientation in the master mode
For improved orientation the display shows the last steps in the path of the current
master mode block.
The upper 3 lines of the display show the following path for selecting the language:
Operating instructions 22016401
06/09
Settings in the master mode
ID7sx-Sys
5.2.4
Entries in the master mode
The following basic rules apply to entries made in the master mode:
• Confirm (alpha)numeric entries with ENTER.
• Alphanumeric entries with the ID7sx-Sys: see below.
• To accept the displayed value: Press ENTER key.
Enter
alphanumeric
identification
1. Press one of the keys CODE A … CODE F.
The functions keys are given the following assignment:
ABCDE
FGHIJ
KLMNO
PQRST
UVWXY
Z/-()
Selection of
letters A to E
Selection of
letters F to J
Selection of
letters K to O
Selection of
letters P to T
Selection of
letters U to Y
Selection of
letter Z or a
special
character
2. Select desired group of letters, e.g. press KLMNO key.
3. Select desired letter.
The display changes again to the above selection.
4. Repeat entry in steps 2 and 3 for additional characters.
Note
Letters and numbers can be combined as desired.
5.2.5
Emergency entrance into the master mode
If a personal code has been assigned for entering the master mode and you have
forgotten your code, you can still enter the master mode:
➜ Enter the character sequence C, L, E, A, R as your personal code.
Operating instructions 22016401
06/09
19
Settings in the master mode
5.3
5.3.1
Legend
5.3.2
RESET TERMINAL
ID7sx-Sys
TERMINAL master mode block
Overview of the TERMINAL master mode block
In the TERMINAL master mode block you enter the following system settings:
Blocks highlighted in grey are described in detail in the following.
Settings in the TERMINAL master mode block
Reset all terminal functions to the factory setting
• Resets the password MINIMUM to the factory setting: 2234
• LANGUAGE
LANGUAGE
German
Select dialog language
Possible setting: English, Deutsch, Français, Nederlands, Italiano, Español, Polski,
Russian, Slovakian, Portuguese, Hungarian, Slovenian, Czech, Croatian
LAYOUT MFII KEYBOARD
Select keyboard layout of connected external keyboard
This setting is not available at ID7sx-Sys.
It is available at IND690 only.
20
Operating instructions 22016401
06/09
Settings in the master mode
ID7sx-Sys
DATE / TIME
Enter date and time
• Enter DATE in European notation: Day.Month.Year.
• Enter TIME in European notation: (24) Hours.Minutes.Seconds.
Comments
• Enter single-place numbers with a preceding zero.
• The clock continues to run after the terminal is switched off.
APPL. BLOCK
Editing application blocks
Assign application blocks or combinations of application blocks to keys or change
key assignments for application blocs.
This setting is not available at ID7sx-Sys.
PERSONAL CODE
Codes for entering various master mode menus
Enter codes with a maximum of 8 alphanumeric characters.
MINIMUM
Access to master mode
GENERAL
Access to all personal codes
PAC
Access to the PAC / RESET PAC and PAC / ERROR-INFO menus
PAC-DATAB.
Access to the PAC / DATABASE menu
PAC-APPL.
Access to the PAC / APPLICATION menu
INTERFACE
Access to all parameters in the INTERFACE menus; the test functions are freely
accessible
OPTION 1
OPTION 2
OPTION 3
Program dependent
Comment
If only zeros are entered, the password is not used
MODE SCALES
Select between serial and parallel operating mode for the connected scales
SCALES SERIAL
Serial operation of the connected scales: Only the weight value of the current scale
is displayed.
SCALES PARALLEL
Parallel operation of the connected scales: All weight values of the connected scales
are displayed simultaneously.
SUM SCALE
A sum scale can be defined in parallel scale operation.
1. SUM SCALE: Select ACTIVATED.
2. With , change to SCALE 1 and select YES with < or > if this scale is to be the
sum scale.
3. Repeat the procedure for SCALE 2 - SCALE 3.
Factory setting: SUM SCALE DEACTIVATED
Operating instructions 22016401
06/09
21
Settings in the master mode
5.4
ID7sx-Sys
SCALE master mode block
In the first block the weighing platform is selected: SCALE 1 … SCALE 3.
The other setting possibilities are the same for all connected weighing platforms.
5.4.1
Overview of the SCALE master mode block
...
Legend
• Blocks highlighted in grey are described in detail in the following.
• Factory settings are printed in bold print.
• Blocks which only appear under certain conditions have a dotted outline.
22
Operating instructions 22016401
06/09
Settings in the master mode
ID7sx-Sys
5.4.2
RESET SCALE
Settings in the SCALE master mode block
Reset weighing platform to factory setting
WEIGHING-PROC ADAPT
VIBRATION ADAPTER
STABILITY DETECTOR
AUTOZERO
RESTART
universal weighing
average conditions
ASD = 2
on
off
WEIGHING-PROC
ADAPT
Adapt weighing platform to weighing sample
UNIVERSAL WEIGHING
For solid bodies, coarse filling or checkweighing (factory setting).
STATIC WEIGHING
For solid bodies and weighing under extreme conditions,
e. g. strong vibrations or weighing animals.
FINE FILLING
For liquid or powdered weighing samples.
VIBRATION ADAPTER
Adapt weighing platform to the vibration influences of the environment
AVERAGE CONDITIONS
Factory setting.
EXTREME CONDITIONS
The weighing platform operates more slowly, however is less sensitive,
e. g. suitable with building vibrations and vibrations at the weighing location.
IDEAL CONDITIONS
The weighing platform operates very quickly, however is very sensitive,
e. g. suitable with very calm and stabile weighing location.
STABILITY DETECTOR
Adapt automatic stability detector
Possible settings:
ASD = 0
Stability detector switched off
(only possible with non-certified weighing platforms)
ASD = 1
fast display
good reproducibility
ASD = 2
▲
▼ (factory setting)
ASD = 3
▲
▼
ASD = 4
slow display
very good reproducibility
AUTOZERO
Switch automatic zero-point correction on or off
The automatic zero-point correction corrects the weight of minor dirt with the
weighing platform unloaded.
Factory setting: AUTOZERO ON
Operating instructions 22016401
06/09
23
Settings in the master mode
RESTART
ID7sx-Sys
Switch restart function on or off
When RESTART ON is set, the zero point and tare value remain stored after the
power supply is interrupted. When the weighing platform is switched on again, the
terminal shows the current weight.
Factory setting: RESTART OFF
DISPLAY UPDATE
Set display speed of the weight display
Select number of updates per second (UPS).
Possible values: 6, 10, 15, 20 UPS
Comments
• This block only appears when the DISPLAY UPDATE function is supported by the
connected weighing platform.
• The possible settings are dependent on the connected weighing platform.
24
Operating instructions 22016401
06/09
Settings in the master mode
ID7sx-Sys
5.5
Select the interface
connection
Possible
interface types
INTERFACE master mode block
➜ Select the interface connection in the first block:
COM1 through COM7.
• SERIAL
Only for COM1 (standard) and COM5 (optional). A
corresponding slotcard interface has to be installed for each
channel in the PSU/..Ex in order to ensure communication in the
secure area (RS232 or CL20mA). The wiring is described in the
Guide for installers. For additional settings, see Page 25.
• 4 I/O and
RELAY BOX 8
Only for COM6. The module designation has been retained for
reasons of compatibility. In reality only the optional 8 I/O
module on which the functionalities "4 I/O" and "Relay box 8"
are mapped exists. For addtitional settings see Page 26.
• 8 I/O
Only for COM6. Alternative dialog for the 8 I/O interface.
• PROFIBUS-DP
Only for COM1 and COM5. An ID7sx Profibus module has to be
connected at the PSU/..Ex via RS232. This is a top-hat rail
module with an external 24-V supply voltage.
• ALIBI MEMORY
Only for COM7. An Alibi-Memory-ID7sx must be installed in the
ID7sx. No further settings are required in the master mode.
SERIAL
Parameters of the standard COM interface driver for serial interfaces. Some
parameters are not editable (dependent on the program)
SWITCH ON/OFF
Switches the function of this driver on or off.
MODE
Changes interface parameters.
BAUDRATE
Possible settings: 150, 300, 600, 1200, 2400, 4800, 9600 Baud
PARITY
Possible settings: Parity even, Parity odd, No parity
DATABITS
Possible settings: 7 databits, 8 databits
STOPBITS
Possible settings: 1 stopbit, 2 stopbits
PROTOCOL
Possible settings: Non, XonXoff, CL-Handshake
END-CHARACTERS
Possible settings: End-character 1, end-character 2
TEST
Operating instructions 22016401
Test function for communication testing.
All characters received over the interface are shown in the display. Simple character
strings can be sent, if desired.
06/09
25
Settings in the master mode
ID7sx-Sys
4 I/O
Parameters of the I/O module. Some parameters are not editable (dependent on
the program). At ID7sx this module is mapped to the 8 I/O module.
SWITCH ON/OFF
Switches the function of this driver on or off.
TEST
Test function for testing the 4 I/O module
The display shows the signal over inputs 1–4.
Outputs 1–4 can be switched on or off using keys 1–4.
Important
Please ensure that no system functions are activated inadvertently!
RELAY BOX 8
At ID7sx this module is mapped to the 8 I/O module.
SWITCH ON/OFF
Switches the function of this driver on or off.
I/O TEST
Tests function and state of inputs/outputs of 1 or 2 connected relay boxes 8-690
When an input or output is set (high), its number is shown in the display.
When an input or output is not set (low), – is shown in the display.
Setting outputs
Switch outputs using keys 1 to 8 of the numeric keypad; .
Setting inputs
Inputs can be set by connecting a voltage supply (+24 V, current limited to 50 mA),
for example.
Comment
The relay box 8-690 corresponds with the BIU Binary Interface Unit. For additional
information, see the operating instructions and installation information for the
505918A Binary Interface Unit.
8 I/O
Parameters of the I/O module. Some parameters are not editable (dependent on
the program)
SWITCH ON/OFF
Switches the function of this driver on or off.
TEST
Test function for testing the 8 I/O module
The display shows the signal over inputs 1–8.
Outputs 1–8 can be switched on or off using keys 1–8.
Important
Please ensure that no system functions are activated inadvertently!
PROFIBUS-DP
Configuring Profibus-DP-ID7sx
SWITCH ON/OFF
Switches the function of this driver on or off.
NODE ADDRESS
Selects the desired node address from a range of 001 to 126.
OPERATING MODE
Possible settings:
4 words
8 words
TEST
Test function for testing the Profibus I/O values
26
Operating instructions 22016401
06/09
What to do if....?
ID7sx-Sys
6
What to do if....?
Error / Message
Possible causes
Rectification
Display dark
• No supply voltage
➜ Check power supply unit
• Incorrect wiring
➜ Carry out wiring according to control
drawing
• Restless installation location
➜ Adjust vibration adapter
• Draft
➜ Block draft
• Interference between load plate and/or
materials being weighed and
environment
➜ Rectify interference
• Power supply interference
➜ Check power supply
• Incorrect zeroing of weighing platform
➜ Unload weighing platform, repeat
zeroing and weighing
• Incorrect tare value
➜ Delete tare or enter correct tarevalue
• Interference between load plate and/or
materials being weighed and
environment
➜ Rectify interference
• Weighing platform tilted
➜ Level out weighing platform
• Incorrect weighing platform selected
➜ Select correct weighing platform
IDENTCODE =
• Test cycle started
➜ Complete test by pressing the
ZERO-SET key
-----------
• Load plate not in place
➜ Put load plate in place
• Pre-load not in place
➜ Load pre-load
• Weighing range dropped below
➜ Set to zero
• Weighing range exceeded
➜ Unload weighing platform
• Weighing platform locked
➜ Release lock
• Error in weighing platform
➜ Test weighing platforms
Weight display unstable
Incorrect weight display
SCALE ERROR
➜ If message appears again: Inform
METTLER TOLEDO Customer Service
SCALE OVERLOAD
• Weighing range exceeded
➜ Unload weighing platform
➜ Release lock
TOTAL IS NEGATIVE
• Total weight is negative
➜ Unload weighing platform and set to
zero
ERROR TARE SET
• Taring not possible, as total weight is
negative
➜ Unload weighing platform and set to
zero, repeat taring
Operating instructions 22016401
06/09
27
What to do if....?
ID7sx-Sys
Error / Message
Possible causes
Rectification
ERROR ZERO SETTING
• Zero-set range exceeded
➜ Unload weighing platform and set to
zero
SCALE IN MOTION
• Weight value does not remain
constant
➜ Wait until weighing platform shows
weight value which remains
constant
• Restless environment
➜ Ensure low-vibration environment
➜ Check setting of vibration adapter
➜ If message appears again: Inform
METTLER TOLEDO Customer Service
SCALE NOT ZERO
• Zero point has shifted
➜ Set to zero
PRINTER OFFLINE
• Printer (GA46) switched off
➜ Switch on printer
PRINTER NO PAPER
• No paper in printer (GA46)
➜ Insert paper
PRINTER ERROR
• General printer error
➜ Inform METTLER TOLEDO Customer
Service
RAM CLEARED
• New software in ID7sx-Sys, all data
reset to factory setting
NO ALIBI MEMORY !!
• No alibi memory found
➜ Inform METTLER TOLEDO Customer
Service
ALIBI ERROR
• Error in alibi memory
➜ Inform METTLER TOLEDO Customer
Service
NO SCALE FOUND
• Weighing platform cable not
connected properly
➜ Connect the weighing platform cable
according to the control drawing
• Load-cell cable not connected
properly
➜ Inform METTLER TOLEDO Customer
Service
• Not all scales connected
➜ Connect all weighing platforms
• Error in weighing cell
➜ Repeat test
ERROR SCALE NUMBER
–
➜ If message appears again: Inform
METTLER TOLEDO Customer Service
PERMISSION DENIED
DATABASE INIT ERROR
28
• Two or more connected weighing
platforms have identical scale
number
➜ Inform METTLER TOLEDO Customer
Service
• Incorrect personal code
➜ Enter correct personal code
• Incorrect information in FTP SECURITY
or FTP FILE
➜ Enter permissible values for
username, password, drive,
directory and file name
• Database could not be initialised
➜ Inform METTLER TOLEDO Customer
Service
Operating instructions 22016401
06/09
Technical data and accessories
ID7sx-Sys
7
Technical data and accessories
7.1
Technical data
Terminal
Display
• Backlit LC display, with graphics capabilities,
64 240 pixels, display field 39 x 132 mm
• Cover of scratch-resistant, hardened, antireflection glass
Keypad
• Tactile-touch membrane keypad with acoustic acknowledgement
• Scratch-resistant marking, 3-color
• 4 keys A to D for identification data, 6 function keys with function change and info
key, 4 scale function keys, numerical keypad
• Alphanumeric input possible with function keys
Housing
• All nickel chromium steel DIN X5 CrNi 1810
• Weight: net 3.2 kg, without cables
Type of protection
• II 2G EEx ib IIC T4
–10 °C ... +40 °C
• II 2D IP65 T55 °C
Protection type
(IEC 529)
• Dust and water-tight as per IP65/68
Pollution degree
2
Installation category
II
Power supply
connectioni
• via PSU/ID..-Ex power supply unit (in the safe area)
Ambient temperature
• operation:
–10 °C – +40 °C
0 – +40 °C
• storage
–25 °C – +60 °C
• via PSUx/.. (in the hazardous area)
for weighing platforms of certification class III
for weighing platforms of certification class II
Relative humididy
20 – 80 %, non condensing
Weighing platform
connection
• 1 IDNet connection standard for METTLER TOLEDO weighing platforms of the
DN...x, PUA..x, PTA...x, PBA430x series, Point Ex system solution and analog
weighing platforms with Point Ex
• In the configuration with the PSU power supply unit 2 additional IDNet
connections (IDNet-ID7sx) are possible
Serial interface
connection
Operating instructions 22016401
• 1 CL20mA connection standard
• In the configuration with the PSU power supply unit max. 2 additional serial
interface connections (CL20mA-ID7sx or RS232-ID7sx) possible, see guide for
installers and control drawing ME-22006478
06/09
29
Technical data and accessories
ID7sx-Sys
Terminal
Digital inputs/outputs
8 I/O-ID7sx module as an option
• 8 digital inputs / 8 digital outputs, galvanically separated, passive
• Signal level input
logically 0 = not supplied with current < 0.3 mA
logically 1 = supplied with current
> 4 mA
• Intrinsically safe connection values see PSU guide for installers and control
drawing ME-22006478
Storage of certificationrelevant weighing data
Alibi-Memory-ID7sx module as an option
• Paperless archiving of certification-relevant weighing data
• Capacity: approx. 700000 data records with net weight, tare weight, date and time
• Convenient search function
• Ring data buffer which overwrites the oldest data record when the memory
capacity is exceeded
Profibus-DP-ID7sx
Housing
Aluminium, DIN locking foot
Dimensions
42 mm x 106 mm x 115 mm (W x H x D)
Weight
2,400 g
Protection type
IP20
Supply
24 V DC +/– 20 %
Current consumption
100 mA (at 24 V DC)
CPU core
80C52, 32 K RAM, 128 K flash
Displays
2 status LEDs
UART interface
RS232
UART buffer
max. 15 KB
Temperature range
0 – 50 °C
Max. humidity
90 %, non-condensing
Profibus
DP connection, 9-pin Sub-D
Electrical isolation
DP interface and 24 V DC
Baud rate
up to 12 Mbit/s
Node address
adjustable to between 0 and 126 in the master mode
Factory setting: 126
Data width
2/4 IN and 2/4 OUT words, consistent for 2 words at operating mode 16-bit integer
or for 4 words at operating mode 32-bit floating point
Status indication
2 LEDs
30
Operating instructions 22016401
06/09
Technical data and accessories
ID7sx-Sys
7.2
Accessories
Weighing platform connections
Order No.
IDNet-ID7sx
22 008 443
• Connection for a weighing platform
• Max. of 2 additional connections possible
Point Ex-ID7sx
• Connection for an analog weighing platform
22 008 443
• Max. of 2 additional connections possible
Active CL/IDNet-ID7sx
• Connection for a weighing platform
22 008 443
• Connection of a second ID7sx for use as a second
display
Serial data interfaces
Order No.
CL20mA-ID7sx
CL 20 mA interface, for installation in ID7sx
22 008 444
RS232-ID7sx
RS232 interface, for installation in ID7sx
22 008 445
Digital inputs/outputs
8 I/O-ID7sx
Order No.
8 digital inputs, 8 digital outputs
Network connection
Profibus-DP-ID7sx
Order No.
Field bus module: Connection via PSU/ID..Ex
Alibi Memory module
Alibi-Memory-ID7sx
22 008 446
22 008 649
Order No.
• Paperless archiving of certification-relevant weighing
data
22 008 447
• Fail-safe storage of configuration data
Barcode reader
Barcode kit ID7sx
Operating instructions 22016401
Order No.
barcode scanner (II2G EEx ib IIC T4; II2D T 70°C)
complete with RS232-ID7sx and Viper-Ex (SW) slotcard
06/09
22 008 640
31
Technical data and accessories
ID7sx-Sys
Other accessories
Order No.
ID7sx installation set
for cabinet installation
22 008 439
Wall bracket
Completely rust-proof
00 504 130
Floor stand
Completely rust-proof
00 504 132
Stand socket
Completely rust-proof
00 503 701
Bracket stand
Completely rust-proof
00 504 128
32
Operating instructions 22016401
06/09
Index
ID7sx-Sys
8
Index
A
Accessories...................31
Additional functions .......12
Alibi memory...........14, 25
Alphanumeric
identification ..............19
Applications ....................5
Autozero ...................9, 23
B
Basic functions................9
Big Weight Display ..........6
C
Cleaning.........................8
Commissioning ...............8
D
Date.............................21
Digital inputs/
outputs................30, 31
Display.....................6, 29
Display update ..............24
Documentation................5
E
Entries..........................19
Error messages .............27
F
Function keys......6, 12, 29
H
Housing .......................29
I
I/O test .........................26
ID code ........................12
Interface
4 I/O......................26
8 I/O................26, 31
CL20mA-ID7sx .......31
Interface
connections .........25, 29
Introduction.....................5
K
Keypad.....................7, 29
M
Master mode ................ 17
Master mode block
INTERFACE ............ 25
SCALE ................... 22
TERMINAL.............. 20
Multi-scale mode .......... 10
P
Personal code .............. 21
Profibus-DP 25, 26, 30, 31
R
Recall information ......... 13
Relay box 8 ................. 26
Reset scale .................. 23
Reset terminal .............. 20
Restart......................... 24
S
Safety precautions........... 4
Scale modes ................ 21
Serial (RS232, CL,
RS422/485) ............. 25
Set zero ......................... 9
Stability detector ........... 23
Switch over weighing
platform.................... 10
T
Tare .............................. 9
Technical data .............. 29
Terminal ...................... 29
Test weighing platform .. 12
Time............................ 21
V
Vibration adapter .......... 23
W
Weighing ..................... 10
Weighing platform
connections ........ 29, 31
Weighing process
adapter..................... 23
What to do if....?........... 27
L
Language .....................20
Operating instructions 22016401
06/09
33
Congratulations on choosing the quality and precision of METTLER TOLEDO. Proper
use according to these instructions and regular calibration and maintenance by our
factory-trained service team ensure dependable and accurate operation, protecting
your investment. Contact us about a ServiceXXL agreement tailored to your needs and
budget.
We invite you to register your product at www.mt.com/productregistration so we can
contact you about enhancements, updates and important notifications concerning
your product.
*22016401*
22016401
Subject to technical changes
© Mettler-Toledo (Albstadt) GmbH 06/09
Mettler-Toledo (Albstadt) GmbH
D-72458 Albstadt
Tel. ++49-7431-14 0, Fax ++49-7431-14 232
Internet: http://www.mt.com
Printed in Germany 22016401