TR5048 Programming Manual

TR5048 Programming Manual
TR5048 Vector Network Analyzer
Programming Manual COM/DCOM
First Edition
2013
TR5048
COM/DCOM Programming manual
TABLE OF CONTENTS
Introduction ................................................................................................................................................. 7
1
Registering COM Server.................................................................................................................... 7
2
COM Technology Overview.............................................................................................................. 7
3
Automation Server............................................................................................................................. 7
4
Automation Controllers ..................................................................................................................... 8
5
Local and Remote Server................................................................................................................... 9
6
Structure of COM Objects ............................................................................................................... 11
7
Accessing the Application Object.................................................................................................... 12
8
Object Methods................................................................................................................................ 14
9
Object Properties.............................................................................................................................. 14
10 Error Handling ................................................................................................................................. 15
11 COM Automation Data Types ......................................................................................................... 17
12 Measurement Data Arrays ............................................................................................................... 18
13 COM Server Commands.................................................................................................................. 19
NAME ....................................................................................................................................................19
Ready......................................................................................................................................................20
SCPI.ABORt ..........................................................................................................................................21
SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt). USER.FILename..........................22
SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe ..........................................24
SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt). USER.FILename..........................25
SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).STATe...........................................26
SCPI.CALCulate(Ch).FSIMulator.SENDed. ZCONversion.PORT(Pt).Z0.R .......................................27
SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.STATe ......................................................28
SCPI.CALCulate(Ch).PARameter.COUNt ............................................................................................29
SCPI.CALCulate(Ch).PARameter(Tr).DEFine......................................................................................30
SCPI.CALCulate(Ch).PARameter(Tr).SELect ......................................................................................32
SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion.....................................................................33
SCPI.CALCulate(Ch).SELected.CONVersion.STATe..........................................................................34
SCPI.CALCulate(Ch).SELected.CORRection.EDELay.TIME .............................................................35
SCPI.CALCulate(Ch).SELected.CORRection.OFFSet.PHASe.............................................................36
SCPI.CALCulate(Ch).SELected.DATA.FDATa ...................................................................................37
SCPI.CALCulate(Ch).SELected.DATA.FMEMory ..............................................................................38
SCPI.CALCulate(Ch).SELected.DATA.SDATa ...................................................................................39
SCPI.CALCulate(Ch).SELected.DATA.SMEMory ..............................................................................40
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.CENTer ...........................................................41
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SHAPe.............................................................42
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SPAN...............................................................43
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STARt .............................................................44
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STATe .............................................................45
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STOP ...............................................................46
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.TYPE...............................................................47
SCPI.CALCulate(Ch).SELected.FORMat .............................................................................................48
SCPI.CALCulate(Ch).SELected.FUNCtion.DATA...............................................................................49
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle .............................................................50
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STARt ...............................................................51
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STATe...............................................................52
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STOP.................................................................53
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute ..........................................................................54
SCPI.CALCulate(Ch).SELected.FUNCtion.PEXCursion......................................................................55
SCPI.CALCulate(Ch).SELected.FUNCtion.POINts..............................................................................56
SCPI.CALCulate(Ch).SELected.FUNCtion.PPOLarity.........................................................................57
SCPI.CALCulate(Ch).SELected.FUNCtion.TARGet ............................................................................58
SCPI.CALCulate(Ch).SELected.FUNCtion.TTRansition......................................................................59
SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE................................................................................60
SCPI.CALCulate(Ch).SELected.LIMit.DATA......................................................................................61
SCPI.CALCulate(Ch).SELected.LIMit.DISPlay.STATe.......................................................................62
SCPI.CALCulate(Ch).SELected.LIMit.FAIL ........................................................................................63
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.AMPLitude.................................................................64
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.STIMulus ...................................................................65
SCPI.CALCulate(Ch).SELected.LIMit.REPort.ALL.............................................................................66
2
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.REPort.DATA .........................................................................67
SCPI.CALCulate(Ch).SELected.LIMit.REPort.POINts ........................................................................68
SCPI.CALCulate(Ch).SELected.LIMit.STATe .....................................................................................69
SCPI.CALCulate(Ch).SELected.MARKer(Mk).ACTivate ....................................................................70
SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.DATA ..........................................................71
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.REFerence ...........................................................72
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.STATe .................................................................73
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.THReshold ..........................................................74
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.TYPE...................................................................75
SCPI.CALCulate(Ch).SELected.MARKer.COUPle ..............................................................................76
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STARt ...............................................77
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STATe...............................................78
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STOP.................................................79
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute...................................................80
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PEXCursion..............................................81
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PPOLarity .................................................82
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TARGet ....................................................83
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TRACking ................................................84
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TTRansition..............................................85
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE ........................................................86
SCPI.CALCulate(Ch).SELected.MARKer.REFerence.STATe .............................................................87
SCPI.CALCulate(Ch).SELected.MARKer(Mk).SET_...........................................................................88
SCPI.CALCulate(Ch).SELected.MARKer(Mk).STATe ........................................................................89
SCPI.CALCulate(Ch).SELected.MARKer(Mk).X.................................................................................90
SCPI.CALCulate(Ch).SELected.MARKer(Mk).Y.................................................................................91
SCPI.CALCulate(Ch).SELected.MARKer.COUNt ...............................................................................92
SCPI.CALCulate(Ch).SELected.MATH.FUNCtion ..............................................................................93
SCPI.CALCulate(Ch).SELected.MATH.MEMorize .............................................................................94
SCPI.CALCulate(Ch).SELected.MAXhold.STATe ..............................................................................95
SCPI.CALCulate(Ch).SELected.MSTatistics.DATA ............................................................................96
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.MARKer.STARt.............................................97
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.MARKer.STOP ..............................................98
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.STATe ............................................................99
SCPI.CALCulate(Ch).SELected.MSTatistics.STATe..........................................................................100
SCPI.CALCulate(Ch).SELected.RLIMit.DATA .................................................................................101
SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.LINE.....................................................................102
SCPI.CALCulate(Ch).SELected.RLIMit.FAIL ...................................................................................103
SCPI.CALCulate(Ch).SELected.RLIMit.REPort.DATA.....................................................................104
SCPI.CALCulate(Ch).SELected.RLIMit.STATe.................................................................................105
SCPI.CALCulate(Ch).SELected.SMOothing.APERture .....................................................................106
SCPI.CALCulate(Ch).SELected.SMOothing.STATe ..........................................................................107
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.CENTer ............................................................108
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh...............................................109
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.KBESsel ...........................................................110
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.LPFRequency ...................................................111
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.SPAN................................................................112
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt ..............................................................113
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STATe ..............................................................114
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STEP.RTIMe....................................................115
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STIMulus..........................................................116
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP ................................................................117
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.TYPE................................................................118
SCPI.CALCulate(Ch).TRACe(Tr).DATA.FDATa ..............................................................................119
SCPI.CALCulate(Ch).TRACe(Tr).DATA.FMEMory .........................................................................120
SCPI.CALCulate(Ch). TRACe(Tr).DATA.SDATa .............................................................................121
SCPI.CALCulate(Ch). TRACe(Tr).DATA.SMEMory ........................................................................122
SCPI.DISPlay.COLor.BACK...............................................................................................................123
SCPI.DISPlay.COLor.GRATicule .......................................................................................................124
SCPI.DISPlay.COLor.RESet................................................................................................................125
SCPI.DISPlay.COLor.TRACe(Tr).DATA ...........................................................................................126
SCPI.DISPlay.COLor.TRACe(Tr).MEMory .......................................................................................127
SCPI.DISPlay.FSIGn ...........................................................................................................................128
SCPI.DISPlay.IMAGe..........................................................................................................................129
3
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.SPLit.............................................................................................................................130
SCPI.DISPlay.UPDate_.IMMediate.....................................................................................................131
SCPI.DISPlay.WINDow(Ch).ACTivate ..............................................................................................131
SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.ALIGn.TYPE ...................................................132
SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe.................................................133
SCPI.DISPlay.WINDow(Ch).TITLe.DATA........................................................................................134
SCPI.DISPlay.WINDow(Ch).TITLe.STATe .......................................................................................135
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer. POSition.X ..................................136
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer. POSition.Y ..................................137
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.AUTO...............................................................138
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision.........................................................139
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RLEVel ............................................................140
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RPOSition ........................................................141
SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions .............................................................................142
SCPI.HCOPy.DATE.STAMp ..............................................................................................................143
SCPI.HCOPy.IMAGe ..........................................................................................................................144
SCPI.HCOPy.IMMediate .....................................................................................................................145
SCPI.HCOPy.PAINt ............................................................................................................................146
SCPI.IEEE4882.IDN............................................................................................................................147
SCPI.IEEE4882.RST............................................................................................................................148
SCPI.IEEE4882.TRG ...........................................................................................................................148
SCPI.IEEE4882.WAI ...........................................................................................................................148
SCPI.INITiate(Ch).CONTinuous .........................................................................................................149
SCPI.INITiate(Ch).IMMediate.............................................................................................................150
SCPI.MMEMory.COPY(Src, Dst) .......................................................................................................151
SCPI.MMEMory.DELete(File)............................................................................................................151
SCPI.MMEMory.LOAD.CKIT(Ck).....................................................................................................152
SCPI.MMEMory.LOAD.CALibration.................................................................................................153
SCPI.MMEMory.LOAD.LIMit............................................................................................................154
SCPI.MMEMory.LOAD.RLIMit .........................................................................................................155
SCPI.MMEMory.LOAD.SEGMent .....................................................................................................156
SCPI.MMEMory.LOAD.STATe .........................................................................................................157
SCPI.MMEMory.MDIRectory.............................................................................................................158
SCPI.MMEMory.STORe.CKIT(Ck)....................................................................................................159
SCPI.MMEMory.STORe.CALibration ................................................................................................160
SCPI.MMEMory.STORe.FDATa ........................................................................................................161
SCPI.MMEMory.STORe.IMAGe........................................................................................................162
SCPI.MMEMory.STORe.LIMit...........................................................................................................163
SCPI.MMEMory.STORe.RLIMit ........................................................................................................164
SCPI.MMEMory.STORe.SEGMent ....................................................................................................165
SCPI.MMEMory.STORe.SNP.DATA .................................................................................................166
SCPI.MMEMory.STORe.SNP.FORMat..............................................................................................167
SCPI.MMEMory.STORe.SNP.TYPE.S1P...........................................................................................168
SCPI.MMEMory.STORe.SNP.TYPE.S2P...........................................................................................169
SCPI.MMEMory.STORe.STATe.........................................................................................................170
SCPI.MMEMory.STORe.STYPe.........................................................................................................171
SCPI.OUTPut.STATe ..........................................................................................................................172
SCPI.SENSe(Ch).AVERage.CLEar.....................................................................................................173
SCPI.SENSe(Ch).AVERage.COUNt ...................................................................................................174
SCPI.SENSe(Ch).AVERage.STATe....................................................................................................175
SCPI.SENSe(Ch).BANDwidth.RESolution.........................................................................................176
SCPI.SENSe(Ch).CORRection.CLEar.................................................................................................177
SCPI.SENSe(Ch).CORRection.COEFficient.DATA(Str, Pt_r, Pt_s)..................................................178
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD.................................................................179
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN .................................................................180
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire. SHORt ...............................................................181
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU.................................................................182
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.LABel ......................................................................183
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.RESet.......................................................................184
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect .....................................................................184
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0 .........................................................186
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1 .........................................................187
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2 .........................................................188
4
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3 .........................................................189
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).DELay ..................................................190
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L0 .........................................................191
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L1 .........................................................192
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L2 .........................................................193
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L3 .........................................................194
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LABel ...................................................195
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS ....................................................196
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).TYPE ....................................................197
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0 .........................................................198
SCPI.SENSe(Ch).CORRection.COLLect.CLEar.................................................................................199
SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse ........................................................200
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN...............................................201
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.SHORt..............................................202
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.THRU ..............................................203
SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT1 ...............................................................204
SCPI.SENSe(Ch).CORRection.COLLect.METHod.TYPE .................................................................205
SCPI.SENSe(Ch).CORRection.COLLect.SAVE.................................................................................206
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).FREQuency(Ls).............................................207
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).INCLude(Ls).STATe.....................................208
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LDC ...............................................................209
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LOSS(Ls).......................................................210
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).TIME .............................................................211
SCPI.SENSe(Ch).CORRection.EXTension.STATe ............................................................................212
SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude.................................................................213
SCPI.SENSe(Ch).CORRection.STATe ...............................................................................................214
SCPI.SENSe(Ch).CORRection.TYPE(Tr)...........................................................................................215
SCPI.SENSe(Ch).FREQuency.CENTer...............................................................................................216
SCPI.SENSe(Ch).FREQuency.CW......................................................................................................217
SCPI.SENSe(Ch).FREQuency.DATA .................................................................................................218
SCPI.SENSe(Ch).FREQuency.SPAN..................................................................................................219
SCPI.SENSe(Ch).FREQuency.STARt.................................................................................................220
SCPI.SENSe(Ch).FREQuency.STOP ..................................................................................................221
SCPI.SENSe(Ch).ROSCillator.SOURce..............................................................................................222
SCPI.SENSe(Ch).SEGMent.DATA.....................................................................................................223
SCPI.SENSe(Ch).SWEep.POINt.TIME...............................................................................................224
SCPI.SENSe(Ch).SWEep.POINts........................................................................................................225
SCPI.SENSe(Ch).SWEep.TYPE..........................................................................................................226
SCPI.SERVice.CHANnel.ACTive.......................................................................................................227
SCPI.SERVice.CHANnel.COUNt .......................................................................................................227
SCPI.SERVice.CHANnel(Ch).TRACe.ACTive ..................................................................................228
SCPI.SERVice.CHANnel.TRACe.COUNt..........................................................................................228
SCPI.SERVice.PORT.COUNt .............................................................................................................229
SCPI.SERVice.SWEep.FREQency.MAXimum ..................................................................................229
SCPI.SERVice.SWEep.FREQency.MINimum ....................................................................................230
SCPI.SERVice.SWEep.POINts............................................................................................................230
SCPI.SOURce(Ch).POWer.CENTer....................................................................................................231
SCPI.SOURce(Ch).POWer.LEVel.IMMediate.AMPLitude................................................................232
SCPI.SOURce(Ch).POWer.LEVel.SLOPe.DATA ..............................................................................233
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.DATA ...............................................................234
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.STATe...............................................................235
SCPI.SOURce(Ch).POWer.SPAN .......................................................................................................236
SCPI.SOURce(Ch).POWer.STARt......................................................................................................237
SCPI.SOURce(Ch).POWer.STOP .......................................................................................................238
SCPI.SYSTem.CORRection.STATe....................................................................................................239
SCPI.SYSTem.DATE ..........................................................................................................................240
SCPI.SYSTem.PRESet.........................................................................................................................240
SCPI.SYSTem.TIME ...........................................................................................................................241
SCPI.SYSTem.LOCal ..........................................................................................................................242
SCPI.SYSTem.REMote .......................................................................................................................242
SCPI.SYSTem.RWLock ......................................................................................................................243
SCPI.SYSTem.HIDe ............................................................................................................................244
SCPI.SYSTem.SHOw ..........................................................................................................................244
5
TR5048
COM/DCOM Programming manual
SCPI.TRIGger.SEQuence.IMMediate..................................................................................................245
SCPI.TRIGger.SEQuence.SINGle .......................................................................................................246
SCPI.TRIGger.SEQuence.SOURce .....................................................................................................247
Appendix 1. Error Codes .................................................................................................................. 248
Appendix 2. Programming Examples............................................................................................ 249
6
TR5048
COM/DCOM Programming manual
Introduction
This Programming Manual contains information on remote control over TR5048
Network Analyzer (Analyzer) and its data communication by means of user programs
written with COM/DCOM technology.
COM technology is used when a user program runs together with an external
measurement instrument program on one PC. DCOM technology is used when a user
program runs on a PC connected with the measurement instrument by LAN.
Methods and techniques for writing of user programs are same for the both
technologies. The only difference between the technologies is that the DCOM
technology requires additional LAN setting performed by the LAN administrator.
Before reading this Manual, familiarize yourself with TR5048 Operating Manual.
1
Registering COM Server
To register COM server of the analyzer run the executable module from command
prompt with the /regserver keyword. To unregister COM server of the analyzer run the
executable module from command prompt with the /unregserver keyword.
Administrative rights is required to register/unregister COM server. Also user has
ability to register COM server during the software installation procedure.
Example of the COM server registration command:
PlanarTR5048.exe /regserver
2
COM Technology Overview
COM stands for Component Object Model. This programming technology was
developed by Microsoft for two purposes:
3

the model provides the specification for interaction of binary
modules created in different programming languages;

the model defines the interfacing between a client application and a
server application running either on the same PC or on two different
PCs. In the latter case, the technology has DCOM abbreviation –
Distributed COM.
Automation Server
7
TR5048
COM/DCOM Programming manual
The PlanarTR5048.exe application contains a built-in COM server that enables other
programs to access its functionality. The PlanarTR5048.exe application COM server
was developed in conformity with the COM automation specification. COM automation
is a technology allowing control over the COM server by the programs written in both
traditional compiling programming languages and interpreting programming languages,
such as VBScript. This enables the server applications to make their functionality
accessible to many more clients.
4
Automation Controllers
Automation controllers are client programs, which use internal functionality of COM
servers. Automation controller programs are developed by users for writing their own
add-ons for the system.
8
TR5048
COM/DCOM Programming manual
User programs can be written in different languages:

programming languages with built-in COM support, such as Visual
Basic®, Delphi, Java;

universal programming languages, such as C, C++;

Microsoft Excel and Word office applications as they include built-in
programming language Visual Basic for Applications®;

program generators, such as National Instruments LabView®, or
HP–VEE.
Examples represented in this Manual are written in Visual Basic (VB). Appendix 3
contains examples written in VB, and C++ languages.
Examples\COM\VBA folder contains source codes for examples written in Visual Basic
for Applications® (Microsoft Excel files).
Examples\COM\CPP folder contains source codes for the C++ language examples.
5
Local and Remote Server
PlanarTR5048.exe application can function either as a local server or as a remote server
of COM automation.
Local server runs on the same PC with the automation controller and each of the
programs is executed as an individual application in a separate window. COM
technology is used in this case (Figure 1).
Remote server and the automation controller run on different PCs connected by LAN.
DCOM (Distributed COM) technology is used in this case (Figure 2). When using
DCOM it is necessary to configure the local network by means of DCOM Windows
tools.
COM technology is normally used to control PlanarTR5048.
The same automation controller is used for the both COM and DCOM technology.
Some changes to the user program may be required in operators, which establish
connection with the server. Moreover, DCOM technology requires additional settings of
the LAN performed by the LAN administrator.
9
TR5048
COM/DCOM Programming manual
User
Program
USB
COM
PlanarTR5048.exe
Figure 1. COM technology
PlanarTR5048.exe
DCOM
USB
User
Program
Ethernet
Figure 2. DCOM technology
10
TR5048
6
COM/DCOM Programming manual
Structure of COM Objects
COM server contains several objects, which provide different functionality of the
server. The COM objects of the PlanarTR5048.exe application are organized in a
hierarchical structure. Figure 3 shows the main COM objects, which comprise the first
three levels of the hierarchical structure of the PlanarTR5048.exe application COM
objects. COM objects provide various methods and properties, which allow access to
the server functions; besides, they allow access to the objects of the lower levels, which
are not shown in Figure 3.
Application
SCPI
CALCulate
DISPlay
IEEE4882
INITiate
MMEMory
SENSe
TRIGer
SOURce
SERVice
HCOPy
OUTPut
SYSTem
Figure 3. The structure of COM objects
The Object Application of the PlanarTR5048.exe application is in the top of the
hierarchy. Access to the lower level objects is implemented via higher level objects.
11
TR5048
COM/DCOM Programming manual
Note
The hierarchy of COM objects is organized in accordance with
the standard and syntax of the SCPI programming language.
Operators in SCPI have hierarchical chain structure, for
example:
CALCulate:PARameter:DEFine S11
The same COM command written in VB programming
language is as follows:
app.SCPI.CALCulate.PARameter.DEFine = "S11"
7
Accessing the Application Object
To establish connection with the COM server application, create an object reference in
the client program. In COM programming the object reference needs to be acquired
preliminarily, to be used later to access the object functionality. To define an object in
Visual Basic perform the following:
1) Declare a variable as an object.
2) Assign the object to this variable.
To declare a variable, use Dim operator or other declaration statement (Public, Private
or Static). The variables used for references should of the types Variant, Object, or a
type of a specific object. For example, the following three operators declare app
variable:
Dim app
Dim app as Object
Dim app as TR5048.Application
Use Set operator and CreateObject (ObjectName, HostName) function to assign a
specific object to a variable.
ObjectName
Automation object name is always equal to “TR5048.Applcation”
HostName
Network name of the PC hosting the PlanarTR5048 COM server. This
parameter is not specified in case of a local server.
For example, the following operators create Application object and assign it to app
variable:
12
TR5048
COM/DCOM Programming manual
Set app = CreateObject("TR5048.Applcation")
Set app = CreateObject("TR5048.Applcation", "Network_Name")
Note
The first form of the operator is used to create the reference to
the local COM server, the second one is used to create the
reference to the remote DCOM server.
To allow access to the objects of a lower level of the hierarchy, these objects are
specified after the reference to the higher level object and separated from it by a dot. For
example:
Dim SystObj
Set SystObj = app.SCPI.SYSTem
COM objects can have indices. For example, CALCulate, INITiate, SENSe, SOURce
objects represent various aspects of the 4 measurement channels of the Analyzer.
Therefore, it is necessary to write the channel index from 1 to 4 to acquire the data of
these objects. For example:
Set SensObj1 = app.SCPI.SENSe(1)
Set SensObj2 = app.SCPI.SENSe(2)
Visual Basic allows omitting of such indices; in this case the indices are considered as
equal to 1. For example, the following VB operators are equivalent:
Set SensObj = app.SCPI.SENSe(1)
Set SensObj = app.SCPI.SENSe
13
TR5048
8
COM/DCOM Programming manual
Object Methods
Objects have methods. Methods are actions that can be applied to objects. The object
methods are specified after the object name and separated from it by a dot.
The following example shows the PRESet method of SYSTem object. This method
performs setting of the Analyzer to the preset condition:
app.SCPI.SYSTem.PRESet
9
Object Properties
Along with methods, objects have properties. Properties are object characteristics that
can be set or read out. The object properties are specified after the object name and
separated from it by a dot.
To modify an object characteristic, write the value of the corresponding property. To
define an object characteristic, read out the value of its property. The following example
show the setting of the POINts property of SWEep object, i.e. the number of sweep
points:
app.SCPI.SENSe.SWEp.POINts = 201
Note
Some object properties cannot be written, and some object
properties cannot be read. In such cases, the properties are
indicated as “read only” or “write only”.
14
TR5048
10
COM/DCOM Programming manual
Error Handling
You can use different approaches to error handling in VB program:

check the value of Err.Number variable after execution of VB
operator, which contains the call to TR5048 object;

use On Error goto VB operator.
These approaches are represented in the examples below. The following operator causes
an error in VB program as "S13" value of the DEFine property is incorrect.
app.SCPI.PARameter.DEFine = "S13"
In the first example, the value of the Err.Number variable is checked after execution of
the VB operator, which contains the call to TR5048 object. On Error Resume Next
directive instructs VB not to interrupt the program execution when the error is detected
but to pass control to the next operator in natural order.
Dim app
Public Sub HandleError1()
Set app = CreateObject("TR5048.Application")
On Error Resume Next
app.SCPI.PARameter.DEFine = "S13"
If Err.Number <> 0 Then
Msg = "Error # " & Str(Err.Number) & " was generated by " &_
Err.Source & Chr(13) & Err.Description
MsgBox Msg,,"Error"
End If
...
End Sub
15
TR5048
COM/DCOM Programming manual
In the second example, On Error GoTo ErrHandler directive instructs VB to interrupt
the program execution when the error is detected and to pass control to ErrHandler
label.
Dim app
Public Sub HandleError2()
Set app = CreateObject("TR5048.Applcation")
On Error GoTo ErrHandler
app.SCPI.PARameter.DEFine = "S13"
...
Exit Sub
ErrHandler:
Msg = "Error # " & Str(Err.Number) & " was generated by " &_
Err.Source & Chr(13) & Err.Description
MsgBox Msg,,"Error"
End Sub
16
TR5048
11
COM/DCOM Programming manual
COM Automation Data Types
In COM automation, there are the following data types, which can be used for client-toserver communication:
Long
32-bit signed integer, value range from –2147483648 to 2147483647
Double
64-bit double-precision floating point,
value range from
–1.79769313486232E308 to –4.94065645841247E–324 for negative
values, and from 4.94065645841247E–324 to 1.79769313486232E308
for positive values
Boolean
16-bit integer, two values 0 – False, 1 – True
String
Variable-length string
Variant
Can be either a value of arbitrary type or an array of values of arbitrary
type. In this case, the term “arbitrary type” means any allowed type of
COM automation. A variable contains information about its type and
array size (if it is an array). It is used for communication of data arrays
between a client and a server.
17
TR5048
12
COM/DCOM Programming manual
Measurement Data Arrays
Measurement data can be either complex values or real values. This depends on the
format selected by the user. For example, the data is real in logarithmic magnitude
format and the data is complex in polar format.
The measurement data is transferred in a Variant type variable, which represents an
array of Double type. To transfer one complex measurement, two adjacent array cells
are used. To transfer one real measurement two adjacent array cells are used as well but
the second cell is always equal to 0. Thus, measurement data array size is a double
number of the measurement points.
Measurement 1
Real
Imag
Measurement 2
Real
Imag
Measurement N
…
Real
Imag
Figure 4. Array of complex measurements
Measurement 1
Measurement 2
Value
Value
0
0
Measurement N
…
Value
0
Figure 5. Array of real measurements
18
TR5048
13
COM/DCOM Programming manual
COM Server Commands
NAME
Object Type
Property (read only)
Data Type
String
Description
Instrument information string. String format: manufacturer, model, serial
number, number of firmware version and number of software version.
Range
up to 40 characters
Dim ID As String
Syntax
Equivalent Softkeys
ID = app.NAME
None
19
TR5048
COM/DCOM Programming manual
Ready
Object Type
Property (read only)
Data Type
Boolean
Description
Ready state of the instrument. Reads out the True value after successful
completion of the boot process (about 10 sec). The TR5048 must be
connected to PC by a USB cable.
Dim State as Boolean
Syntax
Equivalent Softkeys
State = app.Ready
None
20
TR5048
COM/DCOM Programming manual
SCPI.ABORt
Object Type
Method
Description
Aborts the sweep. Switches trigger mode from Single to Hold, or from
Continuous to waiting for a trigger. If the trigger source is set to Internal,
starts a new sweep.
Syntax
app.SCPI.ABORt
Equivalent Softkeys
None
21
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).
USER.FILename
Object Type
Property (read/write)
Data Type
String
Target
Port Pt of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Pt:
port number 1–2 (see Table 2 on page 23)
Description
De-embedding function file name (*.s2p). The file contains the circuit S–
parameters in Touchstone format.
Range
up to 256 characters
Preset Value
""
Dim File As String
File =
Syntax
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).USER.FILename
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).USER.FILename
=
"network.s2p"
Notes
If the full path to the file is not specified, the \FixtureSim subdirectory of the
main directory will be searched for the file.
Equivalent Softkeys
Analysis > Fixture Simulator > De–Embedding > S–parameters File
22
TR5048
COM/DCOM Programming manual
Table 1. Ch: Channel Number
Data Type
Long
Description
Channel number.
Range
from 1 to 4
Out of Range
An error occurs. Error code: 201.
Notes
If the channel number is not specified, it is taken as equal to 1.
Table 2. Pt: Port Number
Data Type
Long
Description
Port number.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 114.
Notes
If the port number is not specified, it is taken as equal to 1.
23
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe
Object Type
Property (read/write)
Data Type
Boolean
Port Pt of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Pt:
port number 1–2 (see Table 2 on page 23)
Description
The ON/OFF state of the e-embedding function.
Allowable Values
True: De-embedding function ON
False: De-embedding function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe = True
Equivalent Softkeys
Analysis > Fixture Simulator > De–Embedding > Port n
24
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).
USER.FILename
Object Type
Property (read/write)
Data Type
String
Target
Port Pt of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Pt:
port number 1–2 (see Table 2 on page 23)
Description
Embedding function file name (*.s2p). The file contains the circuit S–
parameters in Touchstone format.
Range
up to 256 characters
Preset Value
""
Dim File As String
File =
Syntax
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).USER.FILename
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).USER.FILename
=
"network.s2p"
Notes
If the full path to the file is not specified, the \FixtureSim subdirectory of the
main directory will be searched for the file.
Equivalent Softkeys
Analysis > Fixture Simulator > Embedding > User File
25
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.PMCircuit.PORT(Pt).STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Port Pt of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Pt:
port number 1–2 (see Table 2 on page 23)
Description
The ON/OFF state of the embedding function.
Allowable Values
True: Embedding function ON
False: Embedding function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.DEEMbed.PORT(Pt).STATe = True
Equivalent Softkeys
Analysis > Fixture Simulator > Embedding > Port n
26
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed. ZCONversion.PORT(Pt).Z0.R
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Pt:
port number 1–2 (see Table 2 on page 23)
Description
The value of the impedance for port impedance conversion function.
Range
from 1e–6 to 1e6
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
50
Unit
Ω (Ohm)
Dim Value As Double
Syntax
=
Value
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.PORT(Pt).Z0.R
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.PORT(Pt).Z0.R = 75
Equivalent Softkeys
Analysis > Fixture Simulator > Port Z Conversion > Port n Z0
27
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the port impedance conversion function.
Allowable Values
True: Port Z conversion function ON
False: Port Z conversion function OFF
Preset Value
False
Dim Status As Boolean
Syntax
=
Status
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.STATe
app.SCPI.CALCulate(Ch).FSIMulator.SENDed.ZCONversion.STATe = True
Equivalent Softkeys
Analysis > Fixture Simulator > Port Z Conversion > Port Z Conversion
28
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).PARameter.COUNt
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The number of traces in the channel.
Range
from 1 to 8
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Dim TraceNum As Long
Syntax
TraceNum = app.SCPI.CALCulate(Ch).PARameter.COUNt
app.SCPI.CALCulate(Ch).PARameter.COUNt = 2
Equivalent Softkeys
None
29
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).PARameter(Tr).DEFine
Object Type
Property (read/write)
Data Type
String
Target
Trace Tr of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Tr:
trace number 1–8 (see Table 3 on page 31)
Description
The measurement parameter of the trace.
"S11"
: S11 parameter
Allowable Values
"S21"
: S21 parameter
Out of Range
An error occurs. Error code: 208.
Depends on the trace number.
Preset Value
Tr 1, Tr 3, Tr 5, Tr 7:
Tr 2, Tr 4, Tr 6, Tr 8:
"S11"
"S21"
Dim Meas As String
Syntax
Meas = app.SCPI.CALCulate(Ch).PARameter(Tr).DEFine
app.SCPI.CALCulate(Ch).PARameter(Tr).DEFine = "S11"
Equivalent Softkeys
Response > Measurement > S11 | S21
30
TR5048
COM/DCOM Programming manual
Table 3. Tr: Trace Number
Data Type
Long
Description
Trace number
Range
from 1 to 8
Out of Range
An error occurs. Error code: 202.
Notes
If the trace number is not specified, it is taken as equal to 1.
31
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).PARameter(Tr).SELect
Object Type
Method
Target
Trace Tr of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Tr:
trace number 1–8 (see Table 3 on page 31)
Description
Sets the active channel and trace.
Syntax
app.SCPI.CALCulate(Ch).PARameter(Tr).SELect
Notes
If the channel number is greater than the number of the channels displayed,
an error occurs and the command is ignored. If the trace number is greater
than the number of the traces displayed in the channel, an error occurs and
the command is ignored.
Equivalent Softkeys
Channel > Active Channel
Trace > Active Trace
32
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch: channel number 1–4 (see Table 1 on page 23)
Description
The S–parameter conversion function type.
"IMPedance"
: Reflection or Transmission equivalent impedance
according to the trace measurement parameter (S11
or S21)
"ADMittance"
: Reflection or Transmission equivalent admittance
according to the trace measurement parameter (S11
or S21)
"INVersion"
: Inverse S–parameter
"CONJugation"
: S–parameter conjugate
Range
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
An error occurs. Error code: 217.
Preset Value
"IMP"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion
app.SCPI.CALCulate(Ch).SELected.CONVersion.FUNCtion = "INV"
Equivalent Softkeys
Analysis > Conversion > Function > Impedance Z | AdmittanceY | Inverse 1/S |
Conjugation
33
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.CONVersion.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the S-parameter conversion function.
Allowable Values
True: S–parameter conversion function ON
False: S–parameter conversion function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.CONVersion.STATe
app.SCPI.CALCulate(Ch).SELected.CONVersion.STATe = True
Equivalent Softkeys
Analysis > Conversion > Conversion
34
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.CORRection.EDELay.TIME
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The value of the electrical delay.
Range
from –10 to 10
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.CORRection.EDELay.TIME
app.SCPI.CALCulate(Ch).SELected.CORRection.EDELay.TIME = 1e–9
Equivalent Softkeys
Scale > Electrical Delay
35
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.CORRection.OFFSet.PHASe
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The value of the phase offset.
Range
from –360 to 360
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
° (degree)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.CORRection.OFFSet.PHASe
app.SCPI.CALCulate(Ch).SELected.CORRection.OFFSet.PHASe = 360
Equivalent Softkeys
Scale > Phase Offset
36
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.DATA.FDATa
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The formatted data array. The array elements contain measurements in the
current format, for example, in logarithmic magnitude format (Log Mag).
Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
real number in rectangular format, real part in polar
and Smith chart formats;
Data(2n–1)
0 in rectangular format, imaginary part in polar and
Smith chart formats.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.DATA.FDATa
None
37
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.DATA.FMEMory
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The formatted memory array. The array elements contain saved
measurements in the current format, for example, in logarithmic magnitude
format (Log Mag). Also, see section “Measurement Data Arrays” on page
19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
real number in rectangular format, real part in polar
and Smith chart formats;
Data(2n–1)
0 in rectangular format, imaginary part in polar and
Smith chart formats.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.DATA.FMEMory
None
38
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.DATA.SDATa
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The corrected data array. The corrected measurements are complex
numbers. Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the real part of corrected measurement;
Data(2n–1)
the imaginary part of corrected measurement.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.DATA.SDATa
None
39
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.DATA.SMEMory
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The corrected memory array. The corrected measurements are complex
numbers. Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the real part of corrected measurement memory;
Data(2n–1)
the imaginary part of corrected measurement memory.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.DATA.SMEMory
None
40
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.CENTer
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The gate center value of the gating function.
Range
Varies depending on the frequency span and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.CENTer
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.CENTer = 1e–8
Equivalent Softkeys
Analysis > Gating > Center
41
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SHAPe
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The gate shape of the gating function.
Range
"MAXimum"
: Maximum shape
"WIDE"
: Wide shape
"NORMal"
: Normal shape
"MINimum"
: Minimum shape
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
An error occurs. Error code: 218.
Preset Value
"NORM"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SHAPe
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SHAPe = "MAX"
Equivalent Softkeys
Analysis > Gating > Shape > Maximum | Wide | Normal | Minimum
42
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SPAN
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The gate span value of the gating function.
Range
Varies depending on the frequency span and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
2e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SPAN
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.SPAN = 1e–8
Equivalent Softkeys
Analysis > Gating > Span
43
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STARt
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The gate start value of the gating function.
Range
Varies depending on the frequency span and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
–1e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STARt
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STARt = 1e–7
Equivalent Softkeys
Analysis > Gating > Start
44
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the gating function.
Allowable Values
True: Gating function ON
False: Gating function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STATe
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STATe = Status
Equivalent Softkeys
Analysis > Gating > Gating
45
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STOP
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The gate stop value of the gating function.
Range
Varies depending on the frequency span and the number of points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STOP
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.STOP = 1e–7
Equivalent Softkeys
Analysis > Gating > Stop
46
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.TYPE
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The gate type of the gating function.
Range
"BPASs"
: Bandpass type
"NOTCh"
: Notch type
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
An error occurs. Error code: 219.
Preset Value
"BPAS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.TYPE
app.SCPI.CALCulate(Ch).SELected.FILTer.GATE.TIME.TYPE = "NOTC"
Equivalent Softkeys
Analysis > Gating > Type
47
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FORMat
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Data format.
Range
"MLOGarithmic"
"PHASe"
"GDELay"
"SLINear"
"SLOGarithmic"
"SCOMplex"
"SMITh"
"SADMittance"
"PLINear"
"PLOGarithmic"
"POLar"
"MLINear"
"SWR"
"REAL"
"IMAGinary"
"UPHase"
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
An error occurs. Error code: 209.
Preset Value
"MLOG"
: Logarithmic magnitude
: Phase
: Group delay time
: Smith chart format (Lin)
: Smith chart format (Log)
: Smith chart format (Real/Imag)
: Smith chart format (R + jX)
: Smith chart format (G + jB)
: Polar format (Lin)
: Polar format (Log)
: Polar format (Real/Imag)
: Linear magnitude
: Voltage standing wave ratio
: Real part
: Imaginary part
: Expanded phase
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FORMat
app.SCPI.CALCulate(Ch).SELected.FORMat = "PHAS"
Response > Format > Log Mag | SWR | Phase | Expand Phase | Group Delay |
Lin Mag | Real | Imag
Equivalent Softkeys
Response > Format > Smith > Log / Phase | Lin / Phase | Real / Imag | R+jX |
G+jB
Response > Format > Polar > Log / Phase | Ling / Phase | Real / Imag
48
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DATA
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The data array of analysis executed by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
The array size is 2N, where N is the number of points defined by the
SCPI.CALCulate(Ch).SELected.FUNCtion.POINts property.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the response value in the n–th measurement point;
Data(2n–1)
the stimulus value in the n–th measurement point.
Always set to 0 for the analysis of mean value,
standard deviation, and peak–to–peak value.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DATA
None
49
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF coupling state of the analysis range for the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Allowable Values
True: Coupling state ON
False: Coupling state OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle = Status
Equivalent Softkeys
None
50
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STARt
Object Type
Property (read/write)
Data Type
Double
Target
All traces of channel Ch (if the coupling is set to OFF by the
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle property), the active
trace of channel Ch (if otherwise),
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The start value of the analysis range set by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
0
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STARt
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STARt = 1e9
Equivalent Softkeys
None
51
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch (if the coupling is set to OFF by the
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle property), the active
trace of channel Ch (if otherwise),
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the arbitrary range when executing the analysis by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Allowable Values
True: Arbitrary range ON
False: Arbitrary range OFF (entire sweep range)
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STATe
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STATe = True
Equivalent Softkeys
None
52
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STOP
Object Type
Property (read/write)
Data Type
Double
Target
All traces of channel Ch (if the coupling is set to OFF by the
SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle property), the active
trace of channel Ch (if otherwise),
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The stop value of the analysis range set by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
0
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STOP
app.SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.STOP = 2e9
Equivalent Softkeys
None
53
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute
Object Type
Method
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Executes the analysis specified with the
Description
property.
The analysis result can then be read out with the
SCPI.CALCulate(Ch).SELected.FUNCtion.DATA property.
SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE
Syntax
app.SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute
Equivalent Softkeys
None
54
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.PEXCursion
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The lower limit for the peak excursion value when executing the peak
search by the SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
3
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.PEXCursion
app.SCPI.CALCulate(Ch).SELected.FUNCtion.PEXCursion = 1.5
Equivalent Softkeys
None
55
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.POINts
Object Type
Property (read only)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The number of points (data pairs) of the analysis result by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Always equal to 1, when the search is executed for the maximum, minimum,
mean, standard deviation, peak, and peak–to–peak values. The actual
number of points is read out, when the search is executed for all peaks or all
targets.
Dim Value As Long
Syntax
Equivalent Softkeys
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.POINts
None
56
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.PPOLarity
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The polarity selection when performing the peak search by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
"POSitive"
: Positive peaks
"NEGative"
: Negative peaks
"BOTH"
: Both positive peaks and negative peaks
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"POS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FUNCtion.PPOLarity
app.SCPI.CALCulate(Ch).SELected.FUNCtion.PPOLarity = "NEG"
Equivalent Softkeys
None
57
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.TARGet
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The target level when performing the search for the trace and the target level
crosspoints by the SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
0
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.FUNCtion.TARGet
app.SCPI.CALCulate(Ch).SELected.FUNCtion.TARGet = –10
Equivalent Softkeys
None
58
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.TTRansition
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The transition type selection when performing the search for the trace and
the target level crosspoints by the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
Range
"POSitive"
: Positive peaks
"NEGative"
: Negative peaks
"BOTH"
: Both positive peaks and negative peaks
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"POS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FUNCtion.TTRansition
app.SCPI.CALCulate(Ch).SELected.FUNCtion.TTRansition = "BOTH"
Equivalent Softkeys
None
59
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Range
The
selection
of
the
type
of
analysis
executed
by
the
SCPI.CALCulate(Ch).SELected.FUNCtion.EXECute method.
"PTPeak"
: Peak–to–peak (difference between the maximum value
and the minimum value)
"STDEV"
: Standard deviation
"MEAN"
: Mean value
"MAXimum"
: Maximum value
"MINimum"
: Minimum value
"PEAK"
: Search for the peak
"APEak"
: Search for all the peaks
"ATARget"
: Search for all targets
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"PTP"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE
app.SCPI.CALCulate(Ch).SELected.FUNCtion.TYPE = "STDEV"
Equivalent Softkeys
None
60
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.DATA
Object Type
Property (read/write)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The data array, which is the limit line in the limit test function. The array
size is 1 + 5N, where N is the number of measuring points.
For the n–th point, where n from 1 to N:
Data(0)
The number of limit line segments N is from 0 to 100.
Setting 0 clears the limit line;
Data(5n–4)
type of the n–th limit line segment;
0: OFF
1: Upper limit
Description
Notes
2: Lower limit
Data(5n–3)
the stimulus value in the start point of the n–th
segment;
Data(5n–2)
the stimulus value in the end point of the n–th segment;
Data(5n–1)
the response value in the start point of the n–th
segment;
Data(5n–0)
the response value in the end point of the n–th
segment.
If the array size is not 1 + 5N, where N is Data(0), an error occurs (error
code 214). If Data(5n – 4) is less than 0 or more than 2, an error occurs
(error code 214). When Data(5n–3), Data(5n–2), Data(5n–1) and Data(5n–
0) elements are out of allowable range, the value is set to the limit, which is
closer to the specified value.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.LIMit.DATA
app.SCPI.CALCulate(Ch).SELected.LIMit.DATA = Array(1,2,800,900,–10,–
10)
Equivalent Softkeys
Analysis > Limit Test > Edit Limit Line
61
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.DISPlay.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the limit line display of the limit test function.
Allowable Values
True: Limit line display ON
False: Limit line display OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.LIMit.DISPlay.STATe
app.SCPI.CALCulate(Ch).SELected.LIMit.DISPlay.STATe = True
Equivalent Softkeys
Analysis > Limit Test > Limit Line
62
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.FAIL
Object Type
Property (read only)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The limit test result.
Allowable Values
True: Fail
False: Pass
Dim Status As Boolean
Syntax
Equivalent Softkeys
Status = app.SCPI.CALCulate(Ch).SELected.LIMit.FAIL
None
63
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.AMPLitude
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The value of the limit line offset along Y–axis.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
0
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.AMPLitude
app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.AMPLitude = –10
Equivalent Softkeys
Analysis > Limit Test > Limit Line Offsets > Response Offset
64
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.STIMulus
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The value of the limit line offset along X–axis.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
0
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.STIMulus
app.SCPI.CALCulate(Ch).SELected.LIMit.OFFSet.STIMulus = 1e6
Equivalent Softkeys
Analysis > Limit Test > Limit Lines Offsets > Stimulus Offset
65
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.REPort.ALL
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The data array, which is the limit test results. The array size is 4N, where N
is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(4n–3)
the stimulus value in the n–th point
Data(4n–2) the limit test result in the n–th point
–1: No limit
0: Fail
1: Pass
Description
Data(4n–1)
the upper limit value in the n–th point (0 – if there is no
limit)
Data(4n–0)
the lower limit value in the n–th point (0 – if there is no
limit)
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.LIMit.REPort.ALL
None
66
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.REPort.DATA
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The data array, which is the stimulus values at all the measurement points
that failed the limit test. The array size is defined by the
SCPI.CALCulate(Ch).SELected.LIMit.REPort.POINts property.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.LIMit.REPort.DATA
None
67
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.REPort.POINts
Object Type
Property (read only)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The number of the measurement points that failed the limit test.
Description
The array of stimulus values of the points can be read out by the
SCPI.CALCulate(Ch).SELected.LIMit.REPort.DATA property.
Dim Cnt As Long
Syntax
Equivalent Softkeys
Cnt = app.SCPI.CALCulate(Ch).SELected.LIMit.REPort.POINts
None
68
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.LIMit.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the limit test function.
Allowable Values
True: Limit test function ON
False: Limit test function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.LIMit.STATe
app.SCPI.CALCulate(Ch).SELected.LIMit.STATe = True
Equivalent Softkeys
Analysis > Limit Test > Limit Test
69
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).ACTivate
Object Type
Method
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk: marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Sets the active marker. If a marker is OFF this function will turn it ON.
Description
Syntax
Turning ON a marker with the number from 1 to 15 will turn ON all the
markers of smaller numbers. Turning ON the reference marker with number
16 does not turn ON the markers with the numbers from 1 to 15, but
switches these markers to the relative measurement mode.
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).ACTivate
Markers > Active Marker > Marker n
Equivalent Softkeys
Markers > Reference Marker
Table 4. Mk: Marker Number
Data Type
Long
Description
Marker number. Numbers from 1 to 15 are for regular markers, number 16
is for the reference marker.
Range
from 1 to 16
Out of Range
An error occurs. Error code: 203.
Notes
If the marker number is not specified, it is taken as equal to 1.
70
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.DATA
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk: marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
The bandwidth search result. The bandwidth search can be performed
relatively to the marker Mk, or relatively to the absolute maximum value of
the trace (in this case the marker number is ignored), what is set by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.REFerence property.
Description
Notes
The array contains 4 elements:
Data(0)
Bandwidth;
Data(1)
Center frequency;
Data(2)
Q value;
Data(3)
Loss.
If the bandwidth search is impossible, all the read out values are 0. If the
search is performed relatively to a maker, which is OFF, an error occurs
(error code 204).
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.DATA
None
71
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.REFerence
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The selection of the reference point for the bandwidth search function:
reference marker or absolute maximum value of the trace.
"MARKer"
Range
: Bandwidth search relative to the reference marker
"MAXimum" : Bandwidth search relative to the absolute maximum of the
trace
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"MAX"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.REFerence
=
app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.REFerence
"marker"
Equivalent Softkeys
Markers > Marker Math > Bandwidth Search > Search Ref To
72
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the bandwidth search function.
Allowable Values
True: Bandwidth search function ON
False: Bandwidth search function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.STATe
app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.STATe = True
Equivalent Softkeys
Markers > Marker Math > Bandwidth Search > Bandwidth Search
73
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.THReshold
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The bandwidth definition value.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
–3
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
=
Value
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.THReshold
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.THReshold = –
6.0
Equivalent Softkeys
Markers > Marker Math > Bandwidth Search > Bandwidth Value
74
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.TYPE
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The type of the bandwidth search function.
Range
"BPASs"
: Bandpass
"NOTCh"
: Notch
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"BPAS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.TYPE
app.SCPI.CALCulate(Ch).SELected.MARKer.BWIDth.TYPE = "NOTC"
Equivalent Softkeys
Markers > Marker Math > Bandwidth Search > Type
75
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.COUPle
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the marker coupling function.
Allowable Values
True: Marker coupling ON
False: Marker coupling OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer.COUPle
app.SCPI.CALCulate(Ch).SELected.MARKer.COUPle = false
Equivalent Softkeys
Markers > Properties > Marker Couple
76
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STARt
Object Type
Property (read/write)
Data Type
Double
Target
All traces of channel Ch (if the marker search range coupling is set to OFF
by the SCPI.CALCulate(Ch).SELected.FUNCtion.DOMain.COUPle property), the
active trace of channel Ch (if otherwise),
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The start value of the marker search range.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
3e5
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STARt
app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STARt = 1e6
Equivalent Softkeys
Markers > Marker Search > Search Start
77
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
All traces of channel Ch (if the marker search range coupling is set to OFF
by
the
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.COUPle
property), the active trace of channel Ch (if otherwise),
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the arbitrary range when executing the marker search.
Allowable Values
True: Marker search range ON
False: Marker search range OFF (entire sweep range)
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STATe
app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STATe = True
Equivalent Softkeys
Markers > Marker Search > Search Range
78
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STOP
Object Type
Property (read/write)
Data Type
Double
Target
All traces of channel Ch (if the marker search range coupling is set to OFF
by
the
SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.COUPle
property), the active trace of channel Ch (if otherwise),
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The stop value of the marker search range.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Range
No limitation
Preset Value
1.5e9
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STOP
app.SCPI.CALCulate(Ch).SELected.MARKer.FUNCtion.DOMain.STOP = 1e6
Equivalent Softkeys
Markers > Marker Search > Search Stop
79
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute
Object Type
Method
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk: marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
Executes the marker search according to the specified criterion. The type of
the marker search is set by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE property.
Syntax
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute
Markers > Marker Search > Maximum | Minimum
Equivalent Softkeys
Markers > Marker Search > Peak > Search Peak | Search Max Peak | Search
Peak Left | Search Peak Right
Markers > Marker Search > Target > Search Target | Search Target Left |
Search Target Right
80
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PEXCursion
Object Type
Property (read/write)
Data Type
Double
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk:
marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
The peak excursion value, when the marker search for peak is performed by
the SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
1
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PEXCursion
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PEXCursion = 3.0
Equivalent Softkeys
Markers > Marker Search > Peak > Peak Excursion
81
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PPOLarity
Object Type
Property (read/write)
Data Type
String
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk:
marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
The peak polarity selection, when the marker search for peak is performed
by the SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
"POSitive"
: Positive polarity
"NEGative"
: Negative polarity
"BOTH"
: Both positive polarity and negative polarity
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"POS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PPOLarity
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.PPOLarity = "neg"
Equivalent Softkeys
Markers > Marker Search > Peak > Peak Polarity > Positive | Negative | Both
82
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TARGet
Object Type
Property (read/write)
Data Type
Double
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk:
marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
The target value, when the marker search for target is performed by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
Varies depending on the trace format.
Out of Range
No limitation
Preset Value
0
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
=
Value
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TARGet
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TARGet = –10
Equivalent Softkeys
Markers > Marker Search > Target > Target Value
83
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TRACking
Object Type
Property (read/write)
Data Type
Boolean
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk:
marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
The ON/OFF state of the marker search tracking function.
Allowable Values
True: Marker search tracking ON
False: Marker search tracking OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TRACking
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TRACking = True
Equivalent Softkeys
Markers > Marker Search > Tracking
84
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TTRansition
Object Type
Property (read/write)
Data Type
String
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk:
marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
The selection of the type of the target transition, when the marker search for
transition is performed by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
"POSitive"
: Positive target transition
"NEGative"
: Negative target transition
"BOTH"
: Both positive target transition and negative target transition
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"POS"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TTRansition
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TTRansition = "NEG"
Equivalent Softkeys
Marker > Marker Search > Target > Target Transition
85
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE
Object Type
Property (read/write)
Data Type
String
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk:
marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
The selection of the type of the marker search, which is performed by the
SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.EXECute method.
Range
"MAXimum"
"MINimum"
"PEAK"
"LPEak"
"RPEak"
"TARGet"
"LTARget"
"RTARget"
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"MAX"
:
:
:
:
:
:
:
:
Maximum value search
Minimum value search
Peak search
Peak search to the left from the marker
Peak search to the right from the marker
Target search
Target search to the left from the marker
Target search to the right from the marker
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).FUNCtion.TYPE = "MIN"
Markers > Marker Search > Maximum | Minimum
Equivalent Softkeys
Markers > Marker Search > Peak > Search Peak | Search Max Peak | Search
Peak Left | Search Peak Right
Markers > Marker Search > Target > Search Target | Search Target Left |
Search Target Right
86
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.REFerence.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the reference marker. When the reference marker is
turned ON, all the values of the other markers turn to relative values.
Allowable Values
True: Reference marker ON
False: Reference marker OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer.REFerence.STATe
app.SCPI.CALCulate(Ch).SELected.MARKer.REFerence.STATe = True
Equivalent Softkeys
Markers > Reference Marker
87
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).SET_
Object Type
Property (write only)
Data Type
String
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk: marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
The setting of the value of the specified item to the value of the position of
the marker.
"STARt" : Sweep start value set to the stimulus value of the marker
position.
"STOP"
Range
: Sweep stop value set to the stimulus value of the marker
position.
"CENTer" : Sweep center value set to the stimulus value of the marker
position.
"RLEVel" : Reference value set to the response value of the marker
position.
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Syntax
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).SET_ = "STOP"
Equivalent Softkeys
Markers > Marker Functions > Marker–>Start | Marker–>Stop |
Marker –>Center | Marker–>Ref Value
88
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk:
marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
The ON/OFF state of a marker. Turning ON a marker with the number from
1 to 15 will turn ON all the markers of smaller numbers. Turning OFF a
marker with the number from 1 to 15 will turn OFF all the markers of
greater numbers (except for the reference marker). Turning ON/OFF the
reference marker with number 16 does not turn ON/OFF the markers with
the numbers from 1 to 15, but switches these markers to the relative
measurement mode.
Allowable Values
True: Marker ON
False: Marker OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).STATe
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).STATe = True
Markers > Add Marker | Remove Marker
Equivalent Softkeys
Markers > Reference Marker
89
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).X
Object Type
Property (read/write)
Data Type
Double
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk: marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
Description
The stimulus value of the marker.
Range
From the stimulus current start value to the stimulus current stop value.
Out of Value
Sets the value of the limit, which is closer to the specified value.
Preset Value
Stimulus center value
Unit
Hz (Hertz) | s (second) | dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).X
app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).X = 1e9
Equivalent Softkeys
Markers > Edit Stimulus
90
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer(Mk).Y
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
Marker Mk of the active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Mk:
marker number 1–15, or reference marker number 16 (see
Table 4 on page 70)
The response value of the marker. If the reference marker is turned ON, the
values of the markers from 1 to 15 are read out as relative values to the
reference marker.
Description
The array includes 2 elements:
Data(0)
real number in rectangular format, real part in polar
and Smith chart formats;
Data(1)
0 in rectangular format, imaginary part in polar and
Smith chart formats.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).Y
None
91
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MARKer.COUNt
Object Type
Property (read/write)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The number of the turned ON markers.
Range
from 0 to 16
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Dim MarkerCnt As Long
Syntax
MarkerCnt = app.SCPI.CALCulate(Ch).SELected.MARKer.COUNt
app.SCPI.CALCulate(Ch).SELected.MARKer.COUNt = 5
Equivalent Softkeys
None
92
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MATH.FUNCtion
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The selection of the math operation between the measurement data and the
memory trace data. The math result replaces the data trace. If the data trace
is not saved, the command is ignored.
"DIVide"
: Division Data / Mem.
"MULTiply" : Multiplication Data x Mem.
Range
"ADD"
: Addition Data + Mem.
"SUBTract"
: Subtraction Data – Mem.
"NORMal"
: No math
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
An error occurs. Error code 210.
Preset Value
"NORM"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.MATH.FUNCtion
app.SCPI.CALCulate(Ch).SELected.MATH.FUNCtion= "DIV"
Equivalent Softkeys
Trace > Data Math > Data/Mem | Data*Mem | Data+Mem | Data–Mem | OFF
93
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MATH.MEMorize
Object Type
Method
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Saves the measurement data to the memory trace. Automatically turns on
the display of the memory trace.
Syntax
app.SCPI.CALCulate(Ch).SELected.MATH.MEMorize
Equivalent Softkeys
Trace > Memorize Data Trace
94
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MAXhold.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the Maximum Hold function.
Allowable Values
True: Maximum Hold ON
False: Maximum Hold OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MAXhold.STATe
app.SCPI.CALCulate(Ch).SELected.MAXhold.STATe = True
Equivalent Softkeys
Trace > Max Hold
95
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.DATA
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The math statistics data array. The statistics function is applied either over
the whole range (for all the trace), or within the range specified by the
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.STATe property (the range
limits are determined by two markers).
Description
The array includes 3 elements:
Data(0)
Mean value;
Data(1)
Standard deviation;
Data(2)
Peak–to–peak (difference between the maximum value
and the minimum value).
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.MSTatistics.DATA
None
96
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.MARKer.STARt
Object Type
Property (read/write)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The number of the marker, which specifies the start frequency of the math
statistics range.
Range
from 1 to 16
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Dim MkrNum As Long
Syntax
MkrNum
=
MSTatistics.DOMain.MARKer.STARt
app.SCPI.CALCulate(Ch).SELected.
app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.MARKer.STARt = 3
Equivalent Softkeys
Markers > Marker Math > Statistics > Statistics Start
97
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.MARKer.STOP
Object Type
Property (read/write)
Data Type
Long
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The number of the marker, which specifies the stop frequency of the math
statistics range.
Range
from 1 to 16
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
2
Dim MarkerNum As Long
MarkerNum =
Syntax
app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.MARKer.STOP
app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.MARKer.STOP =
4
Equivalent Softkeys
Markers > Marker Math > Statistics > Statistics Stop
98
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.DOMain.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the math statistics range.
Allowable Values
True: Statistics range ON
False: Statistics range OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.STATe
app.SCPI.CALCulate(Ch).SELected. MSTatistics.DOMain.STATe = True
Equivalent Softkeys
Markers > Marker Math > Statistics > Statistics Range
99
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.MSTatistics.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the math statistics display.
Allowable Values
True: Statistics display ON
False: Statistics display OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.MSTatistics.STATe
app.SCPI.CALCulate(Ch).SELected.MSTatistics.STATe = True
Equivalent Softkeys
Markers > Marker Math > Statistics > Statistics
100
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.DATA
Object Type
Property (read/write)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The data array, which is the limit line for the ripple limit function. The array
size is 1 + 4N, where N is the number of limit line segments.
For the n–th point, where n from 1 to N:
Data(4n–3)
the number of limit line segments N is the integer from
0 to 12. Setting 0 clears the limit line;
type of the n–th limit line segment;
0: Off
1: On
Data(4n–2)
the stimulus value in the beginning point of the n–th
segment;
Data(4n–2)
the stimulus value in the end point of the n–th segment;
Data(4n–0)
the ripple limit value of the n–th segment.
Data(0)
Description
Notes
If the array size is not 1 + 4N, where N is Data(0), an error occurs (error
code 214). If Data(4n – 3) is less than 0 or more than 1, an error occurs
(error code 214). When Data(4n–2), Data(4n–1), and Data(4n–0) elements
are out of allowable range, the value is set to the limit, which is closer to the
specified value.
Dim Data As Variant
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.RLIMit.DATA
app.SCPI.CALCulate(Ch).SELected.RLIMit.DATA = Array(1,1,800,900,10)
Equivalent Softkeys
Analysis > Ripple Limit > Edit Ripple Limit
101
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.LINE
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the ripple limit line display.
Allowable Values
True: Ripple limit line ON
False: Ripple limit line OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.LINE
app.SCPI.CALCulate(Ch).SELected.RLIMit.DISPlay.LINE = True
Equivalent Softkeys
Analysis > Ripple Limit > Ripple Limit
102
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.FAIL
Object Type
Property (read only)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Ripple limit test result.
Allowable Values
True: Fail
False: Pass
Dim Status As Boolean
Syntax
Equivalent Softkeys
Status = app.SCPI.CALCulate(Ch).SELected.RLIMit.FAIL
None
103
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.REPort.DATA
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The data array, which is the ripple limit test results. The array size is 1+3N,
where N is the number of ripple limit bands.
For the n–th point, where n from 1 to N:
Description
Data(0)
N total number of the bands;
Data(3n–2)
n number of the band;
Data(3n–1)
Ripple value in the n–th band;
Data(3n–0)
Ripple limit test result in the n–th band:
0: Pass
1: Fail
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).SELected.RLIMit.REPort.DATA
None
104
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.RLIMit.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the ripple limit test.
Allowable Values
True: Ripple limit test ON
False: Ripple limit test OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.RLIMit.STATe
app.SCPI.CALCulate(Ch).SELected.RLIMit.STATe = True
Equivalent Softkeys
Analysis > Ripple Limit > Ripple Test
105
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.SMOothing.APERture
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The smoothing aperture for the smoothing function.
Range
from 0.01 to 20
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Unit
%
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.SMOothing.APERture
app.SCPI.CALCulate(Ch).SELected.SMOothing.APERture = 1.5
Equivalent Softkeys
Response > Smo Aperture
106
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.SMOothing.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the trace smoothing function.
Allowable Values
True: Trace smoothing ON
False: Trace smoothing OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.SMOothing.STATe
app.SCPI.CALCulate(Ch).SELected.SMOothing.STATe = True
Equivalent Softkeys
Response > Smoothing
107
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.CENTer
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Еhe time domain center value, when the time domain transformation
function is turned ON.
Range
Varies depending on the specified frequency range and the number of
points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.CENTer
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.CENTer = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Center
108
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The impulse width (time domain transformation resolution), coupled with
the Kaiser–Bessel window shape β parameter. The impulse width setting
changes the β parameter, and setting of β parameter changes the impulse
width.
Range
Varies depending on the specified frequency range and the number of
points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
None
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Window > Impulse Width (when the transformation
type is set to Bandpass or Lowpass Impulse)
109
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.KBESsel
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The β parameter, which controls the Kaiser–Bessel window shape, when
performing time domain transformation.
Range
from 0 to 13
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
6
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.KBESsel
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.KBESsel = 13
Equivalent Softkeys
Analysis > Time Domain > Window > Kaiser Beta
110
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.LPFRequency
Object Type
Method
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Changes the frequency range to match with the lowpass type of the time
domain transformation function.
Syntax
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.LPFRequency
Equivalent Softkeys
Analysis > Time Domain > Set Frequency Low Pass
111
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.SPAN
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The time domain span value, when the time domain transformation function
is turned ON.
Range
Varies depending on the specified frequency range and the number of
points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
2e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.SPAN
app. SCPI.CALCulate(Ch).SELected.TRANsform.TIME.SPAN = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Span
112
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The start value used for the transformation function of the time domain
function.
Range
Varies depending on the specified frequency range and the number of
points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
–1e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Start
113
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the time domain transformation function.
Allowable Values
True: Time domain transformation ON
False: Time domain transformation OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STATe
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STATe = True
Equivalent Softkeys
Analysis > Time Domain > Time Domain
114
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STEP.RTIMe
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The rise time of the step signal (time domain transformation resolution),
coupled with the Kaiser–Bessel window shape β parameter. The impulse
width setting changes the β parameter, and setting of β parameter changes
the impulse width.
Range
Varies depending on the specified frequency range and the number of
points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
None
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh = 1e–8
Equivalent Softkeys
Analysis > Time Domain > Window > Impulse Width (when the transformation
type is set to Lowpass Step)
115
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STIMulus
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The selection of the stimulus type for the time domain transformation
function: impulse or step.
Range
"IMPulse"
Impulse
"STEP"
Step
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"IMP"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STIMulus
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STIMulus = "STEP"
Equivalent Softkeys
Analysis > Time Domain > Response Type > Bandpass | Lowpass Step |
Lowpass Impulse
116
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP
Object Type
Property (read/write)
Data Type
Double
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The time domain stop value, when the time domain transformation function
is turned ON.
Range
Varies depending on the specified frequency range and the number of
points.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1e–8
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STOP = 2e–8
Equivalent Softkeys
Analysis > Time Domain > Stop
117
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.TYPE
Object Type
Property (read/write)
Data Type
String
Target
The active trace of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The selection of the transformation type for the time domain transformation
function: bandpass response or direct current circuit.
Range
"BPASs"
Bandpass
"LPASs"
Lowpass
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"BPASs"
Dim Param As String
Syntax
Param = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.TYPE
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.TYPE = "STEP"
Equivalent Softkeys
Analysis > Time Domain > Response Type > Bandpass | Lowpass Step |
Lowpass Impulse
118
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).TRACe(Tr).DATA.FDATa
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The specified trace Tr of channel Ch,
Tr:
trace number 1–8 (see Table 3 on page 31)
Ch:
channel number 1–4 (see Table 1 on page 23)
The formatted data array. The array elements contain measurements in the
current format, for example, in logarithmic magnitude format (Log Mag).
Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
real number in rectangular format, real part in polar
and Smith chart formats;
Data(2n–1)
0 in rectangular format, imaginary part in polar and
Smith chart formats.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch).Trace(Tr).DATA.FDATa
None
119
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch).TRACe(Tr).DATA.FMEMory
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The specified trace Tr of channel Ch,
Tr:
trace number 1–8 (see Table 3 on page 31)
Ch:
channel number 1–4 (see Table 1 on page 23)
The formatted memory array. The array elements contain saved
measurements in the current format, for example, in logarithmic magnitude
format (Log Mag). Also, see section “Measurement Data Arrays” on page
19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
real number in rectangular format, real part in polar
and Smith chart formats;
Data(2n–1)
0 in rectangular format, imaginary part in polar and
Smith chart formats.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch). Trace(Tr).DATA.FMEMory
None
120
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch). TRACe(Tr).DATA.SDATa
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The specified trace Tr of channel Ch,
Tr:
trace number 1–8 (see Table 3 on page 31)
Ch:
channel number 1–4 (see Table 1 on page 23)
The corrected data array. The corrected measurements are complex
numbers. Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the real part of corrected measurement;
Data(2n–1)
the imaginary part of corrected measurement.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch). Trace(Tr).DATA.SDATa
None
121
TR5048
COM/DCOM Programming manual
SCPI.CALCulate(Ch). TRACe(Tr).DATA.SMEMory
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
The specified trace Tr of channel Ch,
Tr:
trace number 1–8 (see Table 1 on page 32)
Ch:
channel number 1–4 (see Table 1 on page 23)
The corrected memory array. The corrected measurements are complex
numbers. Also, see section “Measurement Data Arrays” on page 19.
The array size is 2N, where N is the number of measurement points.
Description
For the n–th point, where n from 1 to N:
Data(2n–2)
the real part of corrected measurement memory;
Data(2n–1)
the imaginary part of corrected measurement memory.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.CALCulate(Ch). Trace(Tr).DATA.SMEMory
None
122
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.COLor.BACK
Object Type
Property (read/write)
Data Type
Variant (Long array)
Target
Instrument
The background color for trace display.
The array contains 3 elements:
Description
Data(0)
Red value R;
Data(1)
Green value G;
Data(2)
Blue value B.
Range
For all the array elements from 0 to 255.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
16, 16, 16
Dim Data As Variant
Syntax
Data = app.SCPI.DISPlay.COLor.BACK
app.SCPI.DISPlay.COLor.BACK = Array(0, 0, 0)
Equivalent Softkeys
Display > Color > Background
123
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.COLor.GRATicule
Object Type
Property (read/write)
Data Type
Variant (Long array)
Target
Instrument
The grid and the graticule label color for trace display.
The array contains 3 elements:
Description
Data(0)
Red value R;
Data(1)
Green value G;
Data(2)
Blue value B.
Range
For all array elements from 0 to 255.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Data(0)
63;
Data(1)
63;
Data(2)
63.
Dim Data As Variant
Syntax
Data = app.SCPI.DISPlay.COLor.GRATicule
app.SCPI.DISPlay.COLor. GRATicule = Array(128, 128, 128)
Equivalent Softkeys
Display > Color > Grid
124
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.COLor.RESet
Object Type
Method
Target
Instrument
Description
Restores the display settings to the default values.
Syntax
app.SCPI.DISPlay.COLor.RESet
Equivalent Softkeys
Display > Set Defaults
125
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.COLor.TRACe(Tr).DATA
Object Type
Property (read/write)
Data Type
Variant (Long array)
Target
Trace number Tr in all channels,
Tr:
trace number 1–8 (see Table 3 on page 31)
The data trace color.
The array contains 3 elements:
Description
Data(0)
Red value R;
Data(1)
Green value G;
Data(2)
Blue value B.
Range
For all array elements from 0 to 255.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the trace number.
Dim Data As Variant
Syntax
Data = app.SCPI.DISPlay.COLor.TRACe(Tr).DATA
app.SCPI.DISPlay.COLor.TRACe(Tr).DATA = Array(255, 255, 0)
Equivalent Softkeys
Display > Color > Data Trace
126
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.COLor.TRACe(Tr).MEMory
Object Type
Property (read/write)
Data Type
Variant (Long array)
Target
Trace number Tr in all channels
Tr:
trace number 1–8 (see Table 3 on page 31)
The memory trace color.
The array contains 3 elements:
Description
Data(0)
Red value R;
Data(1)
Green value G;
Data(2)
Blue value B.
Range
For all array elements from 0 to 255.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the trace number.
Dim Data As Variant
Syntax
Data = app.SCPI.DISPlay.COLor.TRACe(Tr).MEMory
app.SCPI.DISPlay.COLor.TRACe(Tr). MEMory = Array(255, 255, 0)
Equivalent Softkeys
Display > Color > Memory Trace
127
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.FSIGn
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the Fail sign display, when performing limit test or
ripple limit test.
Allowable Values
True: Fail sign display ON
False: Fail sign display OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.DISPlay.FSIGn
app.SCPI.DISPlay.FSIGn = True
Analysis > Limit Test > Fail Sign
Equivalent Softkeys
Analysis > Ripple Limit > Fail Sign
128
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.IMAGe
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
The inverted color display of the data traces.
Range
"NORMal"
: Normal display
"INVert"
: Inverted color display
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"NORM"
Dim Param As String
Syntax
Param = app.SCPI.DISPlay.IMAGe
app.SCPI.DISPlay.IMAGe = "INV"
Equivalent Softkeys
Display > Invert Color
129
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.SPLit
Object Type
Property (read/write)
Data Type
Long
Target
Instrument
Description
The number and layout of the channel windows on the screen. The channel
window layout is in Table 5 below.
Range
from 1 to 10
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1
Dim Value As Long
Value = app.SCPI.DISPlay.SPLit
Syntax
app.SCPI.DISPlay.SPLit = 2
Equivalent Softkeys
Channel > Allocate channels
Table 5. Channel Window Layout on the Screen
1:
2:
3:
4:
5:
6:
7:
8:
9:
10:
130
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.UPDate_.IMMediate
Object Type
Method
Target
Instrument
Description
Updates the display once, when the display update is set to OFF
(SCPI.DISPlay.ENABle property is set to False).
Syntax
app.SCPI.DISPlay.UPDate_.IMMediate
Equivalent Softkeys
None
SCPI.DISPlay.WINDow(Ch).ACTivate
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Sets the active channel.
Notes
The channel window must be displayed. At attempt to set to the active
channel the channel, which is not displayed, an error occurs.
Syntax
app.SCPI.DISPlay.WINDow(Ch).ACTivate
Equivalent Softkeys
Channel > Active Channel
Trace > Active Trace
131
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.ALIGn.TYPE
Object Type
Property (read/write)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The alignment mode of the marker display position of each trace, when the
only
active
trace
display
feature
is
turned
OFF
(SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe property is set
to False).
Range
"VERTical"
: Vertical alignment
"HORizontal"
: Horizontal alignment
"NONE"
: No alignment
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"NONE"
Dim Param As String
Syntax
Param = app.SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.ALIGn.TYPE
app.SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.ALIGn.TYPE = "VERT"
Equivalent Softkeys
Markers > Properties > Align > Vertical | Horizontal | OFF
132
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the marker display for the active trace only.
Allowable Values
True: Only active trace markers display ON
False: Only active trace markers display OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe
app.SCPI SCPI.DISPlay.WINDow(Ch).ANNotation.MARKer.SINGle.STATe = True
Equivalent Softkeys
Markers > Properties > Active Only
133
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TITLe.DATA
Object Type
Property (read/write)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The channel title label.
Range
up to 254 characters
Preset Value
""
Dim Text As String
Syntax
Text = app.SCPI.DISPlay.WINDow(Ch).TITLe.DATA
app.SCPI.DISPlay.WINDow(Ch).TITLe.DATA = "Example1"
Equivalent Softkeys
None
134
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TITLe.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the title label display.
Allowable Values
True: Title label display ON
False: Title label display OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.DISPlay.WINDow(Ch).TITLe.STATe
app.SCPI.DISPlay.WINDow(Ch).TITLe.STATe = True
Equivalent Softkeys
Display > Title Label
135
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.
POSition.X
Object Type
Property (read/write)
Data Type
Double
Target
Trace Tr of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Tr:
trace number 1–8 (see Table 3 on page 31)
Description
The display position of the marker value on the X–axis by a percentage of
the display width.
Range
from 0 to 100
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
%
Dim Value As Double
Syntax
=
Value
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.POSition.X
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.POSition.X = 50
Equivalent Softkeys
Markers > Properties > Data X Position
136
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.
POSition.Y
Object Type
Property (read/write)
Data Type
Double
Target
Trace Tr of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Tr:
trace number 1–8 (see Table 3 on page 31)
Description
The display position of the marker value on the Y–axis by a percentage of
the display height.
Range
from 0 to 100
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
%
Dim Value As Double
Syntax
=
Value
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.POSition.Y
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).ANNotation.MARKer.POSition.Y = 50
Equivalent Softkeys
Markers > Properties > Data Y Position
137
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.AUTO
Object Type
Method
Target
Trace Tr of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Tr:
trace number 1–8 (see Table 3 on page 31)
Description
Executes the auto scale function for the trace.
Syntax
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.AUTO
Equivalent Softkeys
Scale > Auto Scale
138
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision
Object Type
Property (read/write)
Data Type
Double
Target
Trace Tr of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Tr:
trace number 1–8 (see Table 3 on page 31)
Description
The trace scale. Sets the scale per division, when the data format is the
rectangular format. Sets the full scale value, when the data format is the
Smith chart format or the polar format.
Range
from 10E–18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the format.
Logarithmic Magnitude: 10 dB/Div
Phase: 40 °/Div
Expand Phase: 100 °/Div
Group Delay: 10e–9 s/Div
Smith Chart, Polar, SWR: 1 /Div
Linear Magnitude: 0.1 /Div
Real part, Imaginary part: 0.2 /Div
Unit
dB/Div (decibel per division), °/Div (degree per division), s/Div (second per
division)
Dim Value As Double
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision = 20
Equivalent Softkeys
Scale > Scale
139
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RLEVel
Object Type
Property (read/write)
Data Type
Double
Target
Trace Tr of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Tr:
trace number 1–8 (see Table 3 on page 31)
Description
The value of the reference line (response value on the reference line). For
the rectangular format only.
Range
from –1E–18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0 (except for SWR: 1)
Unit
dB (decibel) | ° (degree) | s (second)
Dim Value As Double
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RLEVel
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RLEVel = 10
Equivalent Softkeys
Scale > Ref Value
140
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RPOSition
Object Type
Property (read/write)
Data Type
Long
Target
Trace Tr of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Tr:
trace number 1–8 (see Table 3 on page 31)
Description
The position of the reference line. For the rectangular format only.
Range
From
0
to
the
number
of
the
scale divisions (set by
property, 10 by default).
the
SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
5 (except for SWR: 0)
Dim Value As Long
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RPOSition
app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.RPOSition = 10
Equivalent Softkeys
Scale > Ref Position
141
TR5048
COM/DCOM Programming manual
SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23).
Description
The number of the vertical scale divisions. For the rectangular format only.
Range
from 4 to 20
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
10
Resolution
2
Dim Value As Long
Syntax
Value = app.SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions
app.SCPI.DISPlay.WINDow(Ch).Y.SCALe.DIVisions = 12
Equivalent Softkeys
Scale > Divisions
142
TR5048
COM/DCOM Programming manual
SCPI.HCOPy.DATE.STAMp
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the current date and time printout in the upper right
corner.
Allowable Values
True: Date & time printout ON
False: Date & time printout OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.HCOPy.DATE.STAMp
app.SCPI.HCOPy.DATE.STAMp = False
Equivalent Softkeys
System > Print > Print Date & Time
143
TR5048
COM/DCOM Programming manual
SCPI.HCOPy.IMAGe
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
The inverted color image printout.
Range
"NORMal"
: Normal printout
"INVert"
: Inverted color printout
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"NORM"
Dim Param As String
Syntax
Param = app.SCPI.HCOPy.IMAGe
app.SCPI.HCOPy.IMAGe = "INV"
Equivalent Softkeys
System > Print > Invert Image
144
TR5048
COM/DCOM Programming manual
SCPI.HCOPy.IMMediate
Object Type
Method
Target
Instrument
Description
Prints out the image displayed on the screen without previewing.
Syntax
app.SCPI.HCOPy.IMMediate
Equivalent Softkeys
System > Print > Print Embedded
145
TR5048
COM/DCOM Programming manual
SCPI.HCOPy.PAINt
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
The color chart for the image printout.
Range
"COLor"
: Color printout
"GRAY"
: Grayscale printout
"BW"
: Black&white printout
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
The value is ignored.
Preset Value
"BW"
Dim Param As String
Syntax
Param = app.SCPI.HCOPy.PAINt
app.SCPI.HCOPy.PAINt = "COL"
Equivalent Softkeys
System > Print > Print Color
146
TR5048
COM/DCOM Programming manual
SCPI.IEEE4882.IDN
Object Type
Property (read only)
Data Type
String
Target
Instrument
Description
The instrument information string. The string format: ''{manufacturer},
{model}, {serial number}, {software version/firmware version}''.
Range
up to 40 characters
Dim ID As String
Syntax
Equivalent Softkeys
ID = app.NAME
None
147
TR5048
COM/DCOM Programming manual
SCPI.IEEE4882.RST
Object Type
Method
Target
Instrument
Description
Restores the default settings of the instrument. There is difference from
presetting the instrument with the SCPI.SYSTem.PRESet method – in this
case the trigger mode is set to Hold.
Syntax
app.SCPI.IEEE4882.RST
Equivalent Softkeys
None
SCPI.IEEE4882.TRG
Object Type
Method
Target
Instrument
Description
If the trigger source is set to LAN (SCPI.TRIGger.SEQuence.SOURce property
is set to ''BUS'' ), triggers a sweep.
If the trigger source is not set to the bus (SCPI.TRIGger.SEQuence.SOURce
property is not set to ''BUS'' ) or the instrument is not waiting for a trigger,
the method is ignored.
Syntax
app.SCPI.IEEE4882.TRG
Related Commands
SCPI.TRIGger.SEQuence.SOURce
SCPI.INITiate(Ch).CONTinuous
SCPI.INITiate(Ch).IMMediate
Equivalent Softkeys
None
SCPI.IEEE4882.WAI
Object Type
Method
Target
Instrument
Description
Waits for the execution of all commands sent before this command.
Syntax
app.SCPI.IEEE4882.WAI
Equivalent Softkeys
None
148
TR5048
COM/DCOM Programming manual
SCPI.INITiate(Ch).CONTinuous
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the continuous trigger initiation mode. If the
continuous trigger initiation mode is set to OFF, the channel turns to the
hold state.
Allowable Values
True: Continuous trigger initiation mode ON
False: Continuous trigger initiation mode OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.INITiate(Ch).CONTinuous
app.SCPI.INITiate(Ch).CONTinuous = False
Notes
The sweep start in continuous trigger initiation mode depends on the trigger
source. If the trigger is set to internal, the sweeps will go immediately one
after another. If the trigger is set otherwise, the sweep will start when the
trigger signal is received.
Equivalent Softkeys
Stimulus > Trigger > Continuous
Stimulus > Trigger > Hold
149
TR5048
COM/DCOM Programming manual
SCPI.INITiate(Ch).IMMediate
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Sets the channel to the single trigger mode. Before this method is called, the
channel must be in hold state, otherwise an error occurs (error code 213) and
the method is ignored.
On completion of the sweep, the channel goes back into the hold state.
The method returns control before the end of the sweep.
Syntax
app.SCPI.INITiate(Ch).IMMediate
Notes
The sweep start in the single trigger mode depends on the trigger source. If
the trigger is set to internal, the sweep will start immediately after the
method is called. If the trigger is set otherwise, the sweep will start when the
trigger signal is received.
Equivalent Softkeys
Stimulus > Trigger > Single
150
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.COPY(Src, Dst)
Object Type
Method
Target
Instrument
Description
Copies a file.
Syntax
app.SCPI.MMEMory.COPY(Src, Dst)
Src – Source file name. String data type.
Parameter
Equivalent Softkeys
Dst – Destination file name. String data type.
None
SCPI.MMEMory.DELete(File)
Object Type
Method
Target
Instrument
Description
Deletes a file.
Syntax
app.SCPI.MMEMory.DELete(File)
Parameter
File – File name. String data type.
Equivalent Softkeys
None
151
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.CKIT(Ck)
Object Type
Property (write only)
Data Type
String
Target
Calibration kit Ck,
Ck:
calibration kit number 1–11 (see Table 6 on page 151)
Description
Recalls the definition file for the calibration kit. The file must be saved by
the SCPI.MMEMory.STORe.CKIT(Ck) property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.CKIT(Ck) = File
Notes
If the full path to the file is not specified, the \CalKit subdirectory of the
main directory will be searched for the file. The calibration kit definition file
has *.ckd extension by default.
Equivalent Softkeys
None
Table 6. Ck: Calibration Kit Number
Data Type
Long
Description
Calibration kit number.
Range
from 1 to 11
Out of Range
An error occurs. Error code: 114.
Notes
If the calibration kit number is not specified, it is taken as equal to 1.
152
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.CALibration
Object Type
Property (write only)
Data Type
String
Target
User calibration of the active channel.
Description
Recalls the specified calibration file. The file must be saved by the
property.
SCPI.MMEMory.STORe.CALibration
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.CALibration = File
Notes
If the full path to the file is not specified, the \State subdirectory of the main
directory will be searched for the file. The calibration files have *.cal
extension by default.
Equivalent Softkeys
System > Recall > Calibration
153
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.LIMit
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel.
Description
Recalls the specified limit table file. The file must be saved by the
property.
SCPI.MMEMory.STORe.LIMit
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.LIMit = File
Notes
If the full path to the file is not specified, the \Limit subdirectory of the main
directory will be searched for the file. The limit table files have *.lim
extension by default.
Equivalent Softkeys
Analysis > Limit Test > Edit Limit Line > Restore Limit Table
154
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.RLIMit
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel.
Description
Recalls the ripple limit table file. The file must be saved by the
SCPI.MMEMory.STORe.RLIMit property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.RLIMit = File
Notes
If the full path to the file is not specified, the \Limit subdirectory of the main
directory will be searched for the file. The ripple limit files have *.rlm
extension by default.
Equivalent Softkeys
Analysis > Ripple Limit > Edit Ripple Limit > Restore Ripple Limit Table
155
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.SEGMent
Object Type
Property (write only)
Data Type
String
Target
Active channel
Description
Recalls the segment table file. The file must be saved by the
SCPI.MMEMory.STORe.SEGMent property.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.SEGMent = File
Notes
If the full path to the file is not specified, the \Segment subdirectory of the
main directory will be searched for the file. The segment files have *.seg
extension by default.
Equivalent Softkeys
Stimulus > Segment Table > Recall
156
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.LOAD.STATe
Object Type
Property (write only)
Data Type
String
Target
Instrument
Description
Recalls the specified instrument state file. The file must be saved by the
property.
SCPI.MMEMory.STORe.STATe
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.LOAD.STATe = File
Notes
If the full path to the file is not specified, the \State subdirectory of the main
directory will be searched for the file. The instrument state files have *.cfg
extension by default.
Equivalent Softkeys
System > Recall > State
System > Recall > Calibration
System > Recall > State and Calibration
157
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.MDIRectory
Object Type
Property (write only)
Data Type
String
Target
Instrument
Description
Creates a new directory (folder). Contains the full path to the folder being
created.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.MDIRectory = Path
Equivalent Softkeys
None
158
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.CKIT(Ck)
Object Type
Property (write only)
Data Type
String
Target
Calibration kit Ck,
Ck:
calibration kit number 1–11 (see Table 6 on page 151)
Description
Saves the definition file for the calibration kit parameters.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.CKIT(Ck) = File
Notes
If the full path to the file is not specified, the file will be saved to the \CalKit
subdirectory of the main directory. The calibration kit definition file has
*.dat extension by default.
Equivalent Softkeys
None
159
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.CALibration
Object Type
Property (write only)
Data Type
String
Target
User calibration of the active channel.
Description
Save the specified calibration file.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.CALibration = File
Notes
If the full path to the file is not specified, the \State subdirectory of the main
directory will be used for the file. The calibration files have *.cal extension
by default.
Equivalent Softkeys
System > Save > Calibration
160
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.FDATa
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel
Description
Saves the CSV formatted data into a file.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.FDATa = File
Notes
If the full path to the file is not specified, the file will be saved to the \CSV
subdirectory of the main directory. The files have *.csv extension by default.
Equivalent Softkeys
System > Save > Save Trace Data
161
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.IMAGe
Object Type
Property (write only)
Data Type
String
Target
Instrument
Description
Saves the display image in BMP or PNG format into a file.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.IMAGe = File
Notes
If the full path to the file is not specified, the file will be saved to the \Image
subdirectory of the main directory. If the file has *.png extension, the file
has PNG format, in all the other cases the file has BMP format.
Equivalent Softkeys
System > Print > Print Windows
162
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.LIMit
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel
Description
Saves the limit table into a file with the specified name.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.LIMit = File
Notes
If the full path to the file is not specified, the file will be saved to the \Limit
subdirectory of the main directory. The files have *.lim extension by default.
Equivalent Softkeys
Analysis > Limit Test > Edit Limit Line > Save Limit Table
163
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.RLIMit
Object Type
Property (write only)
Data Type
String
Target
Active trace of the active channel
Description
Saves the ripple limit table into a file with the specified name.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.RLIMit = File
Notes
If the full path to the file is not specified, the file will be saved to the \Limit
subdirectory of the main directory. The ripple limit files have *.rlm
extension by default.
Equivalent Softkeys
Analysis > Ripple Limit > Edit Ripple Limit > Save Ripple Limit Table
164
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SEGMent
Object Type
Property (write only)
Data Type
String
Target
Active channel
Description
Saves the segment table in a file with the specified name.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.SEGMent = File
Notes
If the full path to the file is not specified, the file will be saved to the
\Segment subdirectory of the main directory. The segment files have *.seg
extension by default.
Equivalent Softkeys
Stimulus > Segment Table > Save
165
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SNP.DATA
Object Type
Property (write only)
Data Type
String
Target
Active channel
Description
Saves the measured S–parameters of the active channel into a Touchstone
file with the specified name. The file type (1–port or 2–port) is set by the
SCPI.MMEMory.STORe.SNP.TYPE.S1P
and
SCPI.MMEMory.STORe.SNP.TYPE.S2P properties. 1–port type file saves one
reflection parameter: S11. 2–port type file saves the two parameters: S11,
S21.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.SNP.DATA = File
Notes
If the full path to the file is not specified, the file will be saved to the
\FixtureSim subdirectory of the main directory. The 1–port measurement
files have *.slp extension; the 2–port measurement files have *.s2p
extension.
Equivalent Softkeys
System >Save > Touchstone File > Save
166
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SNP.FORMat
Object Type
Property (read/write)
Data Type
String
Target
Active channel
Description
The data format for the S–parameters saving by the
SCPI.MMEMory.STORe.SNP.DATA property.
Range
" MA"
: Logarithmic Magnitude / Angle format
" DB"
: Linear Magnitude / Angle format
" RI"
: Real part /Imaginary part format
Out of Range
The value is ignored.
Preset Value
"RI"
Dim Param As String
Syntax
Param = app.SCPI.MMEMory.STORe.SNP.FORMat
app.SCPI.MMEMory.STORe.SNP.FORMat = "DB"
Equivalent Softkeys
System > Save > Touchstone File > Format
167
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SNP.TYPE.S1P
Object Type
Property (read/write)
Data Type
Long
Target
Active channel
Description
Selects 1-port measurements and sets port numbers (selects S11 parameter)
when
saving
the
measured
S-parameters
by
the
SCPI.MMEMory.STORe.SNP.DATA property.
Range
from 1 to 2
Out of Range
An error occurs. Error code 222.
Preset Value
1
Dim Value As Long
Syntax
Value = app.SCPI.MMEMory.STORe.SNP.TYPE.S1P
app.SCPI.MMEMory.STORe.SNP.TYPE.S1P = 2
System > Save > Touchstone File > Type > 1–Port (s1p)
Equivalent Softkeys
System >Save > Touchstone File > Select Port (s1p)
168
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.SNP.TYPE.S2P
Object Type
Property (read/write)
Data Type
Variant (Long array)
Target
Active channel
Selects 2-port measurements and sets port numbers when saving the
measured
S-parameters by the SCPI.MMEMory.STORe.SNP.DATA
property.
Description
The array contains 2 elements:
Data(0)
First port number (reads out 1, ignored when written);
Data(1)
Second port number
written).
(reads out 2, ignored when
Dim Data As Variant
Syntax
Data = app.SCPI.MMEMory.STORe.SNP.TYPE.S2P
app.SCPI.MMEMory.STORe.SNP.TYPE.S2P = Array(1, 2)
Equivalent Softkeys
System > Save > Touchstone File > Type > 2–Port (s2p)
169
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.STATe
Object Type
Property (write only)
Data Type
String
Target
Instrument
Description
Saves the instrument state into a file with the specified name.
Range
up to 254 characters
Syntax
app.SCPI.MMEMory.STORe.STATe = File
Notes
If the full path to the file is not specified, the file will be saved to the \State
subdirectory of the main directory. The state files have *.cfg extension by
default.
Equivalent Softkeys
System > Save > State
170
TR5048
COM/DCOM Programming manual
SCPI.MMEMory.STORe.STYPe
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
Selects the type of the instrument or channel state saving by the
SCPI.MMEMory.STORe.STATe
or
SCPI.MMEMory.STORe.CHANnel.STATe
property.
Range
"STATe"
: Measurement conditions
"CSTate"
: Measurement conditions and calibration tables
"DSTate"
: Measurement conditions and data traces
"CDSTate" : Measurement conditions, calibration tables and data traces
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
An error occurs. Error code 205.
Preset Value
"CST"
Dim Param As String
Syntax
Param = app.SCPI.MMEMory.STORe.STYPe
app.SCPI.MMEMory.STORe.STYPe = "STATe"
Equivalent Softkeys
None
171
TR5048
COM/DCOM Programming manual
SCPI.OUTPut.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
Sets the ON/OFF state of the stimulus signal output. Measurements cannot
be performed when the stimulus signal output is set to OFF.
Allowable Values
True:
False:
Preset Value
True
Stimulus signal output ON
Stimulus signal output OFF
Dim Status As Boolean
Syntax
Status = app.SCPI.OUTPut.STATe
app.SCPI.OUTPut.STATe = False
Equivalent Softkeys
Stimulus > Power > RF Out
172
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).AVERage.CLEar
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Resets the averaging data count to 0. Restarts the averaging process.
Syntax
app.SCPI.SENSe(Ch).AVERage.CLEar
Equivalent Softkeys
None
173
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).AVERage.COUNt
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The averaging factor, when the averaging function is set to ON by the
SCPI.SENSe(Ch).AVERage.STATe property.
Range
from 1 to 999
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
10
Dim Value As Long
Syntax
Value = app.SCPI.SENSe(Ch).AVERage.COUNt
app.SCPI.SENSe(Ch).AVERage.COUNt = 2
Equivalent Softkeys
Response > Avg Factor
174
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).AVERage.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the averaging function.
Allowable Values
True: Averaging ON
False: Averaging OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).AVERage.STATe
app.SCPI.SENSe(Ch).AVERage.STATe = False
Equivalent Softkeys
Response > Averaging
175
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).BANDwidth.RESolution
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The IF bandwidth value.
Range
from 10 to 30000
Resolution
In steps of 3. (10, 30, 100, 300, 1000, 3000, 10000, 30000)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
10000
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).BANDwidth.RESolution
app.SCPI.SENSe(Ch).BANDwidth.RESolution = 100
Equivalent Softkeys
Response > IF Bandwidth
176
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.CLEar
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Clears the calibration coefficient table.
Syntax
app.SCPI.SENSe(Ch).CORRection.CLEar
Equivalent Softkeys
None
177
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COEFficient.DATA(Str, Pt_r, Pt_s)
Object Type
Property (read)
Data Type
Variant (Double array)
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The calibration coefficient data array set by the type of the corrected error
Str, the number of the receiver port Pt_r and the number of the source port
Pt_s,
Str
: error type (see below)
Pt_r : the number of the receiver port 1–2 (see Table 2 on page 23)
Pt_s : the number of the source port 1–2 (see Table 2 on page 23)
The array size is 2N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Parameter
Data(2n–2)
real part of the calibration coefficients
Data(2n–1)
imaginary part of the calibration coefficients
String Str – corrected error type:
"ES": Source match
"ER": Reflection tracking
"ED": Directivity
"EL": Load match
"ET": Transmission tracking
"EX": Isolation
When ES, ER, or ED is used, the numbers of the ports Pt_r and Pt_s must
be the same. When EL, ET, or EX is used, the numbers of the ports Pt_r and
Pt_s must be different.
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).CORRection.COEFficient.DATA (Str, Pt_r, Pt_s)
Notes
The written calibration coefficients become effective only after the
SCPI.SENSe(Ch).CORRection.COEFficient.SAVE method is invoked.
Equivalent Softkeys
None
178
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Measures the calibration data of the load standard for the specified port.
Range
Port number is 1 or 2.
Out of Range
An error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD = 1
Notes
The property writing starts the measurement for the channel independently
of the trigger initiation and trigger source settings. The function of the
property writing waits for the completion of the measurement.
Calibration > Response (Open) > Load (Optional)
Calibration > Response (Short) > Load (Optional)
Equivalent Softkeys
Calibration > Full 1–Port Cal > Load
Calibration > One Path 2–Port Cal > Load
179
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Measures the calibration data of the open standard for the specified port.
Range
Port number is 1 or 2.
Out of Range
An error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN= 1
Notes
The property writing starts the measurement for the channel independently
of the trigger initiation and trigger source settings. The function of the
property writing waits for the completion of the measurement.
Calibration > Response (Open) > Open
Equivalent Softkeys
Calibration > Full 1–Port Cal > Open
Calibration > One Path 2–Port Cal > Open
180
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire. SHORt
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Measures the calibration data of the short standard for the specified port.
Range
Port number is 1 or 2.
Out of Range
An error occurs (error code: 222).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt = 1
Notes
The property writing starts the measurement for the channel independently
of the trigger initiation and trigger source settings. The function of the
property writing waits for the completion of the measurement.
Calibration > Response (Short) > Short
Equivalent Softkeys
Calibration > Full 1–Port Cal > Short
Calibration > One Path 2–Port Cal > Short
181
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU
Object Type
Property (write only)
Data Type
Variant (Long array)
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Measures the calibration data of the thru standard between the source port
and the receiver port.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. The array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If
the same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU= Array(1, 2)
Notes
The property writing starts the measurement for the channel independently
of the trigger initiation and trigger source settings. The function of the
property writing waits for the completion of the measurement.
Calibration > Response (Thru) > Thru
Equivalent Softkeys
Calibration > One Path 2–Port Cal > Thru
182
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.LABel
Object Type
Property (read/write)
Data Type
String
Target
Calibration kit, selected for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The calibration kit label.
Range
up to 254 characters
Varies depending on the number of the calibration kit.
1: Not Def 50Ohm
2: Not Def 75Ohm
3: 05CK10A-150 –F–
4: 05CK10A-150 –M–
5: N1.1 Type-N –F–
Preset Value
6: N1.1 Type-N –M–
7: Agil. 85032B –F–
8: Agil. 85032B –M–
9: Agil. 85036B –F–
10: Agil. 85036B –M–
11: Empty
Dim Lab As String
Syntax
Lab = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.LABel
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.LABel = "User1"
Equivalent Softkeys
Calibration > Edit Cal Kit > Label
183
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.RESet
Object Type
Method
Target
Calibration kit, selected for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Resets the calibration kit to the factory settings.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.RESet
Equivalent Softkeys
Calibration > Restore Cal Kit
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The selected calibration kit for the channel.
Range
from 1 to 11
Out of Range
An error occurs. Error code: 222.
Preset Value
1
Dim Value As Long
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect = 3
Equivalent Softkeys
Calibration > Cal Kit > Cal Kit n
184
TR5048
COM/DCOM Programming manual
Table 7. Std: Calibration Standard Number
Data Type
Long
Description
The number of the standard.
Range
Varies depending on the number of the standards in the calibration kit.
Out of Range
If the specified standard number is greater than the number of standards in
the kit, an error occurs (error code: 222).
Notes
If the standard number is not specified, it is taken as equal to 1.
185
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The C0 value of the open calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–15 F (Farad)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C0 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Capacitance > C0 [10–15 F]
186
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The C1 value of the open calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–27 F/Hz (Farad/Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Capacitance > C1 [10–27 F/Hz]
187
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std:
standard number (see Table 7 on page 188)
Description
The C2 value of the open calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–36 F/Hz2 (Farad/Hertz2)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Capacitance > C2 [10–36 F/Hz2]
188
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The C3 value of the open calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–45 F/Hz3 (Farad/Hertz3)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Capacitance > C3 [10–45 F/Hz3]
189
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).DELay
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The offset delay value of the calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).DELay
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).DELay = 93E–12
Equivalent Softkeys
Calibration > Edit Cal Kit > Offset Delay
190
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L0
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The L0 value of the short calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–12 H (Henry)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L0 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Inductance > L0 [10–12 H]
191
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L1
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The L1 value of the short calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–24 H/Hz (Henry/Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L1
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L1 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Inductance > L1 [10–24 H/Hz]
192
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L2
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The L2 value of the short calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–33 H/Hz2 (Henry/Hertz2)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L2
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L2 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Inductance > L2 [10–33 H/Hz2]
193
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L3
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The L3 value of the short calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
1E–42 H/Hz3 (Henry/Hertz3)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L3
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).L3 = 100
Equivalent Softkeys
Calibration > Edit Cal Kit > Inductance > L3 [10–42 H/Hz3]
194
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LABel
Object Type
Property (read)
Data Type
String
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The label of the calibration standard.
Range
up to 254 characters
Preset Value
Varies depending on the selected calibration kit and the standard.
Dim Lab As String
Syntax
Lab = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LABel
Equivalent Softkeys
None
195
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The offset loss value of the calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
Varies depending on the selected calibration kit and the standard.
Unit
Ω/s (Ohm/second)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).LOSS = 700E6
Equivalent Softkeys
Calibration > Edit Cal Kit > Offset Loss
196
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).TYPE
Object Type
Property (read)
Data Type
String
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The type of the calibration standard.
Range
"OPEN"
: Open
"SHORt"
: Short
"LOAD"
: Load
"THRU"
: Thru
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
An error occurs. Error code: 216.
Preset Value
Varies depending on the selected calibration kit and the standard.
Dim Param As String
Syntax
Param = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).TYPE
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).TYPE = "OPEN"
Equivalent Softkeys
Calibration > Edit Cal Kit > STD Type
197
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0
Object Type
Property (read/write)
Data Type
Double
Target
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Std: standard number (see Table 7 on page 188)
Description
The offset Z0 value of the calibration standard.
Range
from –1E18 to 1E18
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
50 or 75, depending on the selected calibration kit.
Unit
Ω (Ohm)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0 = 50
Equivalent Softkeys
Calibration > Edit Cal Kit > Offset Z0
198
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.CLEar
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Clears the measurement values of the calibration standards.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.CLEar
Equivalent Softkeys
Calibration > Calibrate > Response (Open) | Response (Short) | Response
(Thru) | One Path 2–Port Cal | Full 1–Port Cal > Cancel
199
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse
Object Type
Property (write only)
Data Type
Variant (Long array)
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Selects the ports and sets the one path 2–port calibration type for the
calculation of the calibration coefficients on completion of the calibration
executed by the SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Description
The array contains 2 elements:
Data(0)
Data(1)
the number of the receiver port;
the number of the source port.
Range
Port number is 1 or 2. The array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If
the same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse = Array(2, 1)
Equivalent Softkeys
None
200
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Selects the port and sets the response calibration (Open) type for the
calculation of the calibration coefficients on completion of the calibration
executed by the SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 222.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN =
1
Equivalent Softkeys
None
201
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.SHORt
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Selects the port and sets the response calibration (Short) type for the
calculation of the calibration coefficients on completion of the calibration
executed by the SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 222.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.SHORt =
1
Equivalent Softkeys
None
202
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.THRU
Object Type
Property (write only)
Data Type
Variant (Long array)
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Selects the ports and sets the response calibration (Thru) type for the
calculation of the calibration coefficients on completion of the calibration
executed by the SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Description
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Range
Port number is 1 or 2. Array elements can not contain the same port
numbers.
Out of Range
If an incorrect port number is specified, an error occurs (error code: 222). If
the same port numbers are specified, an error occurs (error code: 220).
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.THRU = Array(2,
1)
Equivalent Softkeys
None
203
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT1
Object Type
Property (write only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Selects the port and sets the full 1–port calibration type for the calculation
of the calibration coefficients on completion of the calibration executed by
the SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Range
from 1 to 2
Out of Range
An error occurs. Error code: 222.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT1 = 1
Equivalent Softkeys
None
204
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.METHod.TYPE
Object Type
Property (read only)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The calibration type selected for calculating of the calibration coefficients
on
completion
of
the
calibration
executed
by
the
SCPI.SENSe(Ch).CORRection.COLLect.SAVE method.
Range
"RESPO"
: Response (Open)
"RESPS"
: Response (Short)
"RESPT"
: Response (Thru)
"SOLT1"
: Full 1–port calibration
"1PATH"
: One path 2–port calibration
"NONE"
: Not defined
Dim Param As String
Syntax
Equivalent Softkeys
Param = app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.TYPE
None
205
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.COLLect.SAVE
Object Type
Method
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Calculates the calibration coefficients from the calibration standards
measurements depending on the selected calibration type.
Description
On completion of the method, all the calibration standards measurements
are cleared and the error correction automatically turns ON.
At the attempt to execute this method before all the needed standards are
measured, an error occurs and the method is ignored.
Syntax
app.SCPI.SENSe(Ch).CORRection.COLLect.SAVE
Calibration type selection:
Related Commands
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.SHORt
SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.THRU
SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse
SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT1
Calibration standards measurement:
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.LOAD
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.OPEN
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.SHORt
SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU
Equivalent Softkeys
Calibration > Response (Open) | Response (Short) | Response (Thru) | One
Path 2–Port Cal | Full 1–Port Cal > Apply
206
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).FREQuency(Ls)
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch: channel number 1–4 (see Table 1 on page 23)
Pt: port number 1–2 (see Table 2 on page 23)
Description
The value of the frequency at the point number Ls for calculation of the loss
for the port extension function,
Ls: point number 1–2 (see Table 8 on page 246)
Range
from 0.02 to 4800 MHz
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1e9
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).FREQuency(Ls)
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).FREQuency(Ls) = 100Е6
Equivalent Softkeys
Calibration > Port Extensions > Loss > Freq1 | Freq2
207
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).INCLude(Ls).STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Port Pt of channel Ch,
Ch: channel number 1–4 (see Table 1 on page 23)
Pt: port number 1–2 (see Table 2 on page 23)
Description
The ON/OFF state of the loss compensation at the point number Ls to calculate
the loss for the port extension function,
Ls: point number 1–2 (see Table 8 on page 246).
Allowable Values
True: Loss compensation ON
False: Loss compensation OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).INCLude(Ls).STATe
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).INCLude(Ls).STATe = True
Equivalent Softkeys
Calibration > Port Extensions > Loss > Loss1 | Loss2
208
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LDC
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch: channel number 1–4 (see Table 1 on page 23)
Pt: port number 1–2 (see Table 2 on page 23)
Description
The loss value at DC for the loss calculation of the port extension function.
Range
from -200 to 200
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
dB (decibel)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LDC
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LDC = 10
Equivalent Softkeys
Calibration > Port Extensions > Loss > Loss at DC
209
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LOSS(Ls)
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch: channel number 1–4 (see Table 1 on page 23)
Pt: port number 1–2 (see Table 2 on page 23)
Description
The loss value at the point number Ls to calculate the loss for the port
extension function,
Ls: point number 1–2 (see Table 7 on page 236)
Range
from -200 to 200
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
dB (decibel)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LOSS(Ls)
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).LOSS(Ls) = 10
Equivalent Softkeys
Calibration > Port Extensions > Loss > Loss 1 | Loss 2
210
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).TIME
Object Type
Property (read/write)
Data Type
Double
Target
Port Pt of channel Ch,
Ch: channel number 1–4 (see Table 1 on page 23)
Pt: port number 1–2 (see Table 2 on page 23)
Description
The electrical delay value for the port extension function.
Range
from -10 to 10
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).TIME
app.SCPI.SENSe(Ch).CORRection.EXTension.PORT(Pt).TIME = 1e-9
Equivalent Softkeys
Calibration > Port Extensions > Extension Port1 | Extension Port2
211
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.EXTension.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch: channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the port extension function.
Allowable Values
True: Port extension function ON
False: Port extension function OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).CORRection.EXTension.STATe
app.SCPI.SENSe(Ch).CORRection.EXTension.STATe = True
Equivalent Softkeys
Calibration > Port Extensions > Extension
212
TR5048
COM/DCOM Programming manual
SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude
Object Type
Property (read/write)
Data Type
Double
Target
Instrument
Description
The system impedance Z0.
Range
from 0.001 to 1000
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
50
Unit
Ω (Ohm)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude
app.SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude = 75
Equivalent Softkeys
Calibration > System Z0
213
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The ON/OFF state of the error correction.
Allowable Values
True: Error correction ON
False: Error correction OFF
Preset Value
False
Dim Status As Boolean
Syntax
Status = app.SCPI.SENSe(Ch).CORRection.STATe
app.SCPI.SENSe(Ch).CORRection.STATe = True
Equivalent Softkeys
Calibration > Correction
214
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).CORRection.TYPE(Tr)
Object Type
Property (read only)
Data Type
Variant (Variant array)
Target
Trace Tr of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Tr:
trace number 1–8 (see Table 3 on page 31)
The information about the applied calibration type and the port numbers for
the specified trace.
The array contains 3 elements:
Description
Data(0)
calibration type (see below);
Data(1)
the number of the receiver port to be calibrated;
Data(2)
the number of the source port to be calibrated.
Calibration type in the element Data(0):
Range
"RESPO"
: Response (Open)
"RESPS"
: Response (Short)
"RESPT"
: Response (Thru)
"SOLT1"
: Full 1–port calibration
"1PATH"
: One path 2–port calibration
"NONE"
: Not defined
Dim CalInfo As Variant
Syntax
Equivalent Softkeys
CalInfo = app.SCPI.SENSe(Ch).CORRection.TYPE(Tr)
None
215
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.CENTer
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The stimulus center value of the sweep range for linear or logarithmic sweep
types.
Range
from 3E5 to 1.3E9
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
650.15E6
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.CENTer
app.SCPI.SENSe(Ch).FREQuency.CENTer = 1Е9
Equivalent Softkeys
Stimulus > Center
216
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.CW
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The value of the fixed frequency for the power sweep.
Description
Before this method is called, the sweep type must be set to power sweep,
otherwise the method is ignored.
Range
from 3E5 to 1.3E9
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
650.15E6
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.CW
app.SCPI.SENSe(Ch).FREQuency.CW = 1Е9
Equivalent Softkeys
Stimulus > Power > CW Freq
217
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.DATA
Object Type
Property (read only)
Data Type
Variant (Double array)
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The array of the measurement points frequency for linear, logarithmic or
segment sweep type.
Description
The array size is N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
Data(n–1)
Syntax
the frequency value at the n–th measurement point.
Dim Data As Variant
Data = app. SCPI.SENSe(Ch).FREQuency.DATA
Equivalent Softkeys
None
218
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.SPAN
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The stimulus span value of the sweep range for linear or logarithmic sweep
types.
Range
from 3E5 to 1.3E9
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1.2997E9
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.SPAN
app.SCPI.SENSe(Ch).FREQuency.SPAN = 1Е9
Equivalent Softkeys
Stimulus > Span
219
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.STARt
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The stimulus start value of the sweep range for linear or logarithmic sweep
types.
Range
from 3E5 to 1.3E9
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
3E5
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.STARt
app.SCPI.SENSe(Ch).FREQuency.STARt = 1Е6
Equivalent Softkeys
Stimulus > Start
220
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).FREQuency.STOP
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The stimulus stop value of the sweep range for linear or logarithmic sweep
types.
Range
from 3E5 to 1.3E9
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
1.3E9
Unit
Hz (Hertz)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).FREQuency.STOP
app.SCPI.SENSe(Ch).FREQuency.STOP = 1Е8
Equivalent Softkeys
Stimulus > Stop
221
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).ROSCillator.SOURce
Object Type
Property (read/write)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Selects the internal or external source of the reference frequency of 10 MHz.
"INTernal" : Internal source of the reference frequency
Range
"EXTernal" : External source of the reference frequency
Notes
The short format of the parameter is indicated by upper case letters. There is no
distinction between upper and lower case letters when the property is written.
When the property is read out, the short format is indicated by upper case
letters.
Out of Range
An error occurs. Error code: 224.
Preset Value
"INT"
Dim Param As String
Syntax
Param = app.SCPI.SENSe(Ch).ROSCillator.SOURce
app.SCPI.SENSe(Ch).ROSCillator.SOURce = "EXT"
Equivalent Softkeys
None
222
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).SEGMent.DATA
Object Type
Property (read/write)
Data Type
Variant (Double array)
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
The array of the segment sweep table.
The array has the following format:
{ <Buf>, <Flag1>, <Flag2>, <Flag3>, <Flag4>, <Flag5>, <N>,
<Start(1)>, <Stop(1)>, <NOP(1)> [,<IFBW(1)>] [,<Pow(1)>] [,<Del(1)>] [,<Time(1)>],
<Start(2)>, <Stop(2)>, <NOP(2)> [,<IFBW(2)>] [,<Pow(2)>] [,<Del(2)>] [,<Time(2)>],
…
<Start(N)>, <Stop(N)>, <NOP(N)> [,<IFBW(N)>] [,<Pow(N)>] [,<Del(N)>] [,<Time(N)>] }
Description
<Buf>
<Flag1>
<Flag2>
<Flag3>
<Flag4>
<Flag5>
<N>
<Start n>
<Stop n>
<NOP n>
<IFBW n>
<Pow n>
<Del n>
: Always 5,
: Stimulus start setting (0 – start/stop, 1 – center/span),
: Setting of the <IFBW> field (0 – disabled, 1 – enabled),
: Setting of the <Pow> field (0 – disabled, 1 – enabled),
: Setting of the <Del> field (0 – disabled, 1 – enabled),
: Setting of the <Time> field (0 – disabled, 1 – enabled),
: Number of segments,
: Start value of the n–th segment,
: Stop value of the n–th segment,
: Number of points of the n–th segment,
: IF bandwidth of the n–th segment (if enabled),
: Power of the n–th segment (if enabled),
: Measurement delay of the n–th segment (if enabled),
<Time n>
: Reserved for future use (if enabled).
Dim Data As Variant
Syntax
Data = app.SCPI.SENSe(Ch).SEGMent.DATA
app.SCPI.SENSe(Ch).SEGMent.DATA = Data
Equivalent Softkeys
Stimulus > Segment Table
223
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).SWEep.POINt.TIME
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The value of the delay before measurement in each measurement point.
Range
from 0 to 0.3
Resolution
5E-6
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
s (second)
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).SWEep.POINt.TIME
app.SCPI.SENSe(Ch).SWEep.POINt.TIME = 5Е-6
Equivalent Softkeys
Stimulus > Meas Delay
224
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).SWEep.POINts
Object Type
Property (read/write)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The number of measurement points.
Range
from 2 to 10001
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
201
Dim Value As Long
Syntax
Value = app.SCPI.SENSe(Ch).SWEep.POINts
app.SCPI.SENSe(Ch).SWEep.POINts = 1001
Equivalent Softkeys
Stimulus > Points
225
TR5048
COM/DCOM Programming manual
SCPI.SENSe(Ch).SWEep.TYPE
Object Type
Property (read/write)
Data Type
String
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
Sets the sweep type.
"LINear"
: Linear frequency sweep
"LOGarithmic" : Logarithmic frequency sweep
Range
"SEGMent"
: Segment frequency sweep
"POWer "
: Power sweep
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
An error occurs. Error code: 206.
Preset Value
"LIN"
Dim Param As String
Syntax
Param = app.SCPI.SENSe(Ch).SWEep.TYPE
app.SCPI.SENSe(Ch).SWEep.TYPE = "LOG"
Equivalent Softkeys
Stimulus > Sweep Type
226
TR5048
COM/DCOM Programming manual
SCPI.SERVice.CHANnel.ACTive
Object Type
Property (read only)
Data Type
Long
Target
Instrument
Description
The number of the active channel.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.CHANnel.ACTive
Equivalent Softkeys
None
SCPI.SERVice.CHANnel.COUNt
Object Type
Property (read only)
Data Type
Long
Target
Instrument
Description
The maximum number of the channels.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.CHANnel.COUNt
Equivalent Softkeys
None
227
TR5048
COM/DCOM Programming manual
SCPI.SERVice.CHANnel(Ch).TRACe.ACTive
Object Type
Property (read only)
Data Type
Long
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The active trace number of the channel.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.CHANnel(Ch).TRACe.ACTive
Equivalent Softkeys
None
SCPI.SERVice.CHANnel.TRACe.COUNt
Object Type
Property (read only)
Data Type
Long
Target
Instrument
Description
The maximum number of the traces in the channel.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.CHANnel.TRACe.COUNt
Equivalent Softkeys
None
228
TR5048
COM/DCOM Programming manual
SCPI.SERVice.PORT.COUNt
Object Type
Property (read only)
Data Type
Long
Target
Instrument
Description
The number of the ports.
Dim Value As Long
Syntax
Value = app.SCPI.SERVice.PORT.COUNt
Equivalent Softkeys
None
SCPI.SERVice.SWEep.FREQency.MAXimum
Object Type
Property (read only)
Data Type
Double
Target
Instrument
Description
The upper limit of the measurement frequency.
Syntax
Dim Value As Double
Value = app.SCPI.SERVice.SWEep.FREQency.MAXimum
Equivalent Softkeys
None
229
TR5048
COM/DCOM Programming manual
SCPI.SERVice.SWEep.FREQency.MINimum
Object Type
Property (read only)
Data Type
Double
Target
Instrument
Description
The lower limit of the measurement frequency.
Syntax
Dim Value As Double
Value = app.SCPI.SERVice.SWEep.FREQency.MINimum
Equivalent Softkeys
None
SCPI.SERVice.SWEep.POINts
Object Type
Property (read only)
Data Type
Double
Target
Instrument
Description
The maximum number of the measurement points.
Syntax
Dim Value As Long
Value = app.SCPI.SERVice.SWEep.POINts
Equivalent Softkeys
None
230
TR5048
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.CENTer
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The center value of the power sweep range.
Range
from –55 to 3
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
–26
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.CENTer
app.SCPI.SOURce(Ch).POWer.CENTer = -5
Equivalent Softkeys
Stimulus > Center
231
TR5048
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.LEVel.IMMediate.AMPLitude
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The power level for the frequency sweep.
Range
from –55 to 3
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.LEVel.IMMediate.AMPLitude
app.SCPI.SOURce(Ch).POWer.LEVel.IMMediate.AMPLitude = -10
Equivalent Softkeys
Stimulus > Power > Output Power
232
TR5048
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.LEVel.SLOPe.DATA
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The power slope value for the frequency sweep.
Range
from 0 to 2
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
0
Unit
dB/GHz (decibel/gigahertz)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.LEVel.SLOPe.DATA
app.SCPI.SOURce(Ch).POWer.LEVel.SLOPe.DATA = 0.2
Equivalent Softkeys
Stimulus > Power > Power Slope
233
TR5048
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.DATA
Object Type
Property (read)
Data Type
Variant (Double array)
Target
Port Pt of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Pt:
port number 1–2 (see Table 2 on page 23)
The power correction table.
The array size is 1+2N, where N is the number of the table rows.
For the n–th point, where n from 1 to N:
Description
Data(0)
the number of the table rows N integer from 0 to
10001;
Data(2n–1)
the frequency of the n–th row of the table from 300
kHz to 1.3 GHz;
Data(2n)
power correction value of the n–th row of the table
from –10 to +10 dB.
Dim Data As Variant
Syntax
Equivalent Softkeys
Data = app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.DATA
None
234
TR5048
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Port Pt of channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Pt:
port number 1–2 (see Table 2 on page 23)
Description
The ON/OFF state of the power correction function.
Allowable Values
True: Power correction ON
False: Power correction OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.STATe
app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.STATe = True
Equivalent Softkeys
None
235
TR5048
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.SPAN
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The power span for the power sweep.
Range
from 0 to 58
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
58
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.SPAN
app.SCPI.SOURce(Ch).POWer.SPAN = 50
Equivalent Softkeys
Stimulus > Span
236
TR5048
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.STARt
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The power sweep start for the power sweep.
Range
from –55 to 3
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
–55
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.STARt
app.SCPI.SOURce(Ch).POWer.STARt = -5
Equivalent Softkeys
Stimulus > Start
237
TR5048
COM/DCOM Programming manual
SCPI.SOURce(Ch).POWer.STOP
Object Type
Property (read/write)
Data Type
Double
Target
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 23)
Description
The power sweep stop for the power sweep.
Range
from –55 to 3
Resolution
0.05
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
3
Unit
dBm (decibels above 1 milliwatt)
Dim Value As Double
Syntax
Value = app.SCPI.SOURce(Ch).POWer.STOP
app.SCPI.SOURce(Ch).POWer.STOP = 0
Equivalent Softkeys
Stimulus > Stop
238
TR5048
COM/DCOM Programming manual
SCPI.SYSTem.CORRection.STATe
Object Type
Property (read/write)
Data Type
Boolean
Target
Instrument
Description
The ON/OFF state of the system error correction.
Allowable Values
True: System error correction ON
False: System error correction OFF
Preset Value
True
Dim Status As Boolean
Syntax
Status = app.SCPI.SYSTem.CORRection.STATe
app.SCPI.SYSTem.CORRection.STATe = False
Equivalent Softkeys
None
239
TR5048
COM/DCOM Programming manual
SCPI.SYSTem.DATE
Object Type
Property (read/write)
Data Type
Variant (Long array)
Target
Instrument
The current date.
The array consists of three elements:
Description
Data(0)
year from 1900 to 2100;
Data(1)
month from 1 to 12;
Data(2)
day from 1 to 31.
Dim Data As Variant
Syntax
Data = app.SCPI.SYSTem.DATE
app. app.SCPI.SYSTem.DATE = Array(2009, 9, 9)
Equivalent Softkeys
None
SCPI.SYSTem.PRESet
Object Type
Method
Target
Instrument
Description
Resets the instrument to the factory settings. The difference from the
SCPI.IEEE4882.RST: method is that the trigger is set to the Continuous
trigger mode.
Syntax
app.SCPI.SYSTem.PRESet
Equivalent Softkeys
System > Preset > Apply
240
TR5048
COM/DCOM Programming manual
SCPI.SYSTem.TIME
Object Type
Property (read/write)
Data Type
Variant (Long array)
Target
Instrument
The current time.
The array consists of three elements:
Description
Data(0)
hours from 0 to 23;
Data(1)
minutes from 0 to 59;
Data(2)
seconds from 0 to 59.
Dim Data As Variant
Syntax
Data = app.SCPI.SYSTem.TIME
app. app.SCPI.SYSTem.TIME = Array(15, 20, 30)
Equivalent Softkeys
None
241
TR5048
COM/DCOM Programming manual
SCPI.SYSTem.LOCal
Object Type
Method
Target
Instrument
Description
Sets the instrument to the local operation mode, when all the keys on the
front panel, mouse and the touch screen are active.
Syntax
app.SCPI.SYSTem.LOCal
Related Commands
SCPI.SYSTem.RWLock
Equivalent Softkeys
None
SCPI.SYSTem.REMote
Object Type
Method
Target
Instrument
Description
Sets the instrument to the remote operation mode, when all the keys on the
front panel, mouse and the touch screen are not active, except for one key
labeled Return to Local. Pushing this button will reset the instrument to the
local operation mode.
Syntax
app. SCPI.SYSTem.REMote
Related Commands
SCPI.SYSTem.LOCal
Equivalent Softkeys
None
242
TR5048
COM/DCOM Programming manual
SCPI.SYSTem.RWLock
Object Type
Method
Target
Instrument
Description
Sets the instrument to the remote operation mode, when all the keys on the
front panel, mouse and the touch screen are not active. Only
SCPI.SYSTem.LOCal or SCPI.SYSTem.REMote command can release this
remote operation mode.
Syntax
app. SCPI.SYSTem.RWLock
SCPI.SYSTem.LOCal
Related Commands
SCPI.SYSTem.REMote
Equivalent Softkeys
None
243
TR5048
COM/DCOM Programming manual
SCPI.SYSTem.HIDe
Object Type
Method
Target
Instrument
Description
Make the main form of a application invisible.
Syntax
app.SCPI.SYSTem.HIDe
Related Commands
SCPI.SYSTem.SHOw
Equivalent Softkeys
None
SCPI.SYSTem.SHOw
Object Type
Method
Target
Instrument
Description
Return the main form of a application to visible state.
Syntax
app. SCPI.SYSTem.SHOw
Related Commands
SCPI.SYSTem.HIDe
Equivalent Softkeys
None
244
TR5048
COM/DCOM Programming manual
SCPI.TRIGger.SEQuence.IMMediate
Object Type
Method
Target
Instrument
Generates a trigger, independently of the trigger source setting (except for
the External). If the trigger source is set to External, an error occurs (error
code 221) and the command is ignored.
Description
If the instrument is not in the waiting for a trigger state (sweep is in progress
or all the channels are set to Hold), an error occurs (error code 211) and the
command is ignored.
The method returns control before the end of the sweep.
Syntax
app.SCPI.TRIGger.SEQuence.IMMediate
Related Commands
SCPI.TRIGger.SEQuence.SOURce
SCPI.INITiate(Ch).CONTinuous
SCPI.INITiate(Ch).IMMediate
Equivalent Softkeys
None
245
TR5048
COM/DCOM Programming manual
SCPI.TRIGger.SEQuence.SINGle
Object Type
Method
Target
Instrument
Generates a trigger, independently of the trigger source setting (except for
the External). If the trigger source is set to External, an error occurs (error
code 221) and the command is ignored.
Description
If the instrument is not in the waiting for a trigger state (sweep is in progress
or all the channels are set to Hold), an error occurs (error code 211) and the
command is ignored.
The method does not return control before the end of the sweep (waiting for
the completion of the sweep of all the channels).
Syntax
app.SCPI.TRIGger.SEQuence.SINGle
Related Commands
SCPI.TRIGger.SEQuence.SOURce
SCPI.INITiate(Ch).CONTinuous
SCPI.INITiate(Ch).IMMediate
Equivalent Softkeys
None
246
TR5048
COM/DCOM Programming manual
SCPI.TRIGger.SEQuence.SOURce
Object Type
Property (read/write)
Data Type
String
Target
Instrument
Description
Selects the sweep trigger source.
"INTernal" : Internal
Range
"EXTernal" : External
"BUS"
: Bus
Notes
The short format of the parameter is indicated by upper case letters. There is
no distinction between upper and lower case letters when the property is
written. When the property is read out, the short format is indicated by upper
case letters.
Out of Range
An error occurs. Error code: 205.
Preset Value
"INT"
Dim Param As String
Syntax
Param = app.SCPI.TRIGger.SEQuence.SOURce
app.SCPI.TRIGger.SEQuence.SOURce = "BUS"
Related Commands
SCPI.TRIGger.SEQuence.IMMediate
SCPI.TRIGger.SEQuence.SINGle
SCPI.IEEE4882.TRG
Equivalent Softkeys
Stimulus > Trigger > Trigger Source > Internal | External
247
Appendix 1. Error Codes
114
"Header suffix out of range"
200
"Execution error"
211
"Trigger ignored"
213
"Init ignored"
220
"Parameter Error"
222
"Data out of range"
224
"Illegal parameter value"
201
"Invalid channel index"
202
"Invalid trace index"
203
"Invalid marker index"
204
"Marker is not active"
205
"Invalid save type specifier"
206
"Invalid sweep type specifier"
207
"Invalid trigger source specifier"
208
"Invalid measurement parameter specifier"
209
"Invalid format specifier"
210
"Invalid data math specifier"
214
"Invalid limit data"
215
"Invalid segment data"
216
"Invalid standard type specifier"
217
"Invalid conversion specifier"
218
"Invalid gating shape specifier"
219
"Invalid gating type specifier"
300
"Device-specific error"
302
"Status reporting system error"
248
Appendix 2. Programming Examples
Example 1. Instrument Information String Readout
The following program reads out and displays on the screen the instrument information
string – the Name property of the COM object. The string contains the following fields:
Manufacturer, Model, Serial Number, Software Version/Firmware Version
For example:
Copper Mountain Technologies, TR5048, 00000001, 0.7.9/ 2.0
Dim app As Object
Sub Example1()
Set app = CreateObject("TR5048.Application")
ID = app.Name
MsgBox ("Information string read out: " + ID)
End Sub
249
Example 2. Checking the Instrument Ready State
Normally, the user control program starts when the PlanarTR5048.exe application is
running, the instrument booting is completed, and the instrument is ready for use. In
some cases, it is recommended to check if the instrument is ready for use. The
instrument may be not ready for use if it is not connected to PC via USB cable.
Moreover, if the PlanarTR5048.exe application has not been started in advance, the
CreateObject function will automatically start the application and then within about 10
seconds the instrument booting will be in progress. The instrument will not be ready for
use until the booting is completed. The Ready property is used to check if the
instrument is ready for use.
The following program checks the Ready property right after a COM object has been
created. If the PlanarTR5048.exe application has been started in advance and the
booting is completed, “Analyzer is ready” will be displayed. If the Ready property
value is False, 10 second delay is activated for the case the PlanarTR5048.exe
application has been started by the COM object creation. In 10 seconds the program
rechecks the Ready property. If the value is True, “Analyzer is ready” will be displayed,
if otherwise, “Analyzer is not ready” will be displayed, what means the instrument is
not connected to LAN or it is not connected to PC via USB cable.
Dim app As Object
Sub Example2()
Set app = CreateObject("TR5048.Application")
If app.Ready = False Then
Application.Wait (Now + TimeValue("0:00:10"))
If app.Ready = False Then
MsgBox ("Analyzer is not ready")
Exit Sub
End If
End If
MsgBox ("Analyzer is ready")
End Sub
250
Example 3. Setting the Measurement Parameters
The following program shows the setting of some measurement parameters. First, the
instrument is reset to the factory settings. Then the following parameters are set:
 Two channel windows are opened and allocated one above the other.
 The number of traces is set to 2 in the first channel window.
 For the first channel the stimulus parameters are set as follows: the frequency
range from 100 MHz to 1.2 GHz, the number of measurement points 401.
 For the second channel the stimulus parameters are set as follows: the frequency
range from 800 MHz to 900 MHz, the number of points 51, IF bandwidth 100 Hz,
stimulus power –10 dBm.
 In the first channel window: S11 measurement is set for the trace 1, S21
measurement is set for the trace 2. The Smith chart format is set for the both traces.
 In the second channel window: S21 measurement and logarithmic magnitude
format are set for the single trace. Then the auto scale function is called for this trace.
Dim app As Object
Public Sub Example3()
Set app = CreateObject("TR5048.Application")
app.SCPI.SYSTem.PRESet
app.SCPI.DISPlay.Split = 2
app.SCPI.Calculate(1).Parameter.Count = 2
app.SCPI.SENSe(1).Frequency.Start = 100000000
app.SCPI.SENSe(1).Frequency.STOP = 1200000000
app.SCPI.SENSe(1).SWEep.Points = 401
app.SCPI.SENSe(2).Frequency.Start = 800000000
app.SCPI.SENSe(2).Frequency.STOP = 900000000
app.SCPI.SENSe(2).SWEep.Points = 51
app.SCPI.SENSe(2).BANDwidth.RESolution = 100
app.SCPI.Source(2).Power.LEVel.IMMediate.AMPlitude = -10
app.SCPI.Calculate(1).Parameter(1).DEFine = "S11"
app.SCPI.Calculate(1).Parameter(2).DEFine = "S21"
app.SCPI.Calculate(1).Parameter(1).Select
app.SCPI.Calculate(1).Selected.Format = "SMIT"
app.SCPI.Calculate(1).Parameter(2).Select
app.SCPI.Calculate(1).Selected.Format = "SMIT"
251
app.SCPI.Calculate(2).Parameter(1).DEFine = "S21"
app.SCPI.Calculate(2).Parameter(1).Select
app.SCPI.Calculate(2).Selected.Format = "MLOG"
app.SCPI.DISPlay.Window(2).TRACe(1).Y.SCALe.AUTO
End Sub
252
Example 4. Measurement Data Acquisition
The following program shows data array acquisition with further writing into a file. The
program also shows the method of a sweep triggering and waiting for the sweep
completion.
Three variables F, M, P are declared in the second string of the code. They are used for
arrays of frequency values (Hz), magnitude values (dB), and phase values (degree)
respectively.
After the instrument has been reset to the factory settings, two operators are used for the
sweep triggering and waiting for the sweep completion:
app.SCPI.TRIGger.SEQuence.Source = "BUS"
app.SCPI.TRIGger.SEQuence.Single
The first operator sets the LAN bus command or the COM/DCOM interface command
as a trigger source. It aborts the sweep and switches the instrument to waiting for a
trigger. The second operator is used for a new sweep triggering and waiting for the
sweep completion.
Note
Unlike
the
SCPI.TRIGger.SEQuence.IMMediate
and
SCPI.IEEE4882.TRG commands, which are completed
immediately
after
a
trigger
generation,
the
SCPI.TRIGger.SEQuence.Single command is not completed
until
the
end
of
the
sweep.
Using
the
SCPI.TRIGger.SEQuence.Single command is the simplest way
to set the waiting for the sweep completion.
On completion of the sweep, three arrays are read out: frequency values, magnitude
values and phase values. Before the magnitude and phase arrays are read out, the
corresponding trace format is set.
The array size of frequency F is equal to the number of measurement points, and the
array size of magnitude M and phase P is equal to the double number of measurement
points (see section 11 “Measurement Data Arrays”). In rectangular formats (for
magnitude and phase) the measurement data are real numbers located in even cells of
the array. Odd cells of the array contain 0.
On completion of the program, the frequency, magnitude and phase values for each
measurement point are written string by string into the file named TESTFILE.
253
Dim app As Object
Dim F, M, P
Public Sub Example4()
Set app = CreateObject("TR5048.Application")
app.SCPI.SYSTem.PRESet
app.SCPI.TRIGger.SEQuence.Source = "BUS"
app.SCPI.TRIGger.SEQuence.Single
F = app.SCPI.SENSe.Frequency.Data
app.SCPI.Calculate.Selected.Format = "MLOG"
M = app.SCPI.Calculate.Selected.Data.FDATa
app.SCPI.Calculate.Selected.Format = "PHASe"
P = app.SCPI.Calculate.Selected.Data.FDATa
Open "TESTFILE" For Output As #1
For i = LBound(F) To UBound(F)
Print #1, F(i), M(i * 2), P(i * 2)
Next i
Close #1
End Sub
254
Example 5. Program Written in C++
The following C++ program represents an example of the measurement parameter
setting, as well as acquisition and display of the measurement data array. The program
also shows a method of the sweep triggering and waiting for the sweep completion.
//--------------------------------------------------------------------------// Simple example of using COM object of PlanarTR5048.exe application.
//
// This example is console application. GUI is not used in this example to
// simplify the program. Error proccessing is very restricted too.
//
#include "stdafx.h"
//--------------------------------------------------------------------------// Generate description of COM object of PlanarTR5048.exe application.
#import "PlanarTR5048.exe" no_namespace
//--------------------------------------------------------------------------int _tmain(int argc, _TCHAR* argv[])
{
ITR5048Ptr pNWA;
// Pointer to COM object of PlanarTR5048.exe
CComVariant Data;
// Variable for measurement data
// Init COM subsystem
HRESULT hr = CoInitialize(NULL);
if(hr != S_OK) return -1;
// Create COM object
hr = pNWA.CreateInstance(__uuidof(TR5048));
if(hr != S_OK) return -1;
// Preset network analyzer
pNWA->SCPI->SYSTem->PRESet();
// Set frequency start to 1 GHz
pNWA->SCPI->SENSe[1]->FREQuency->STARt = 1e8;
// Set frequency stop to 1.2 GHz
pNWA->SCPI->SENSe[1]->FREQuency->STOP = 1.2e8;
// Set number of measurement points to 51
pNWA->SCPI->SENSe[1]->SWEep->POINts = 51;
// Set measured parameter to S21
pNWA->SCPI->CALCulate[1]->PARameter[1]->DEFine = "S21";
// Set trigger source to GPIB/LAN bus or COM interface
pNWA->SCPI->TRIGger->SEQuence->SOURce = "bus";
// Trigger measurement and wait
pNWA->SCPI->TRIGger->SEQuence->SINGle();
// Get measurement data (array of complex numbers)
Data = pNWA->SCPI->CALCulate[1]->SELected->DATA->FDATa;
//
//
//
//
//
//
Display measurement data.
Data is array of NOP * 2 (number of measurement points).
Where n is an integer between 0 and NOP - 1.
Data(n*2) : Primary value at the n-th measurement point.
Data(n*2+1) : Secondary value at the n-th measurement point. Always 0
when the data format is not the Smith chart or the polar.
CComSafeArray<double> mSafeArray;
if (mSafeArray.Attach(Data.parray) == S_OK)
{
for (unsigned int n = 0; n < mSafeArray.GetCount() / 2; ++n)
{
printf("%+.9E\t%+.9E\n",
mSafeArray.GetAt(n*2),
mSafeArray.GetAt(n*2+1));
}
mSafeArray.Detach();
}
255
printf("Press ENTER to exit.\n");
getc(stdin);
// Release COM object
pNWA.Release();
CoUninitialize();
return 0;
}
256