METEOROLOGICAL INSTRUMENTS
INSTRUCTIONS
ULTRASONIC ANEMOMETER
MODEL 86000
R.M. YOUNG COMPANY
2801 AERO PARK DRIVE, TRAVERSE CITY, MICHIGAN 49686, USA
TEL: (231) 946-3980 FAX: (231) 946-4772 WEB: www.youngusa.com
PN: 86000-90
REV: B010815
OPERATING INSTRUCTIONS
Model 86000 Ultrasonic Anemometer
Contents
SECTION
1.0
2.0
3.0
4.0
4.1
4.2
4.3
5.0
5.1
5.2
5.3
5.4
6.0
6.1
6.2
6.3
6.4
7.0
8.0
APPENDIX
A
B
DESCRIPTION
SPECIFICATIONS
INTRODUCTION
BEFORE INSTALLATION
INSTALLATION
Placement
Mounting and Alignment
Wiring Connections
OPERATION
Analog Outputs
With YOUNG Wind Tracker Display
Serial Output Formats and Protocols
Low Power Operation
SETTING SERIAL OUTPUTS AND OPERATING PARAMETERS
Setup using YOUNG 86SETUP program
Setup using terminal program
Command Overview
Command Details
EXAMPLE SETTINGS
WARRANTY
WIRING CONNECTIONS
SENSOR ORIENTATION AND DIMENSIONS
6
9
2
2
2
2
3
2
2
2
2
1
1
PAGE
1
5
5
3
4
3
3
86000-90(B)
MODEL 86000
ULTRASONIC ANEMOMETER
1.0 SPECIFICATIONS
WIND SPEED
Range:
Resolution:
Accuracy:
0-70 m/s (156 mph)
0.01 m/s
0 to 30 m/s ±2% or 0.1 m/s
30 to 70 m/s ±3%
WIND DIRECTION
Azimuth Range: 0-360 degrees
Resolution:
Accuracy:
0.1 degree
± 2 degrees
SERIAL OUTPUT (Selectable)
Interface Type:
Formats:
Baud Rates:
RS-232 or RS-485/422
ASCII Text (polled and continuous)
RMYT (YOUNG Wind Tracker)
NMEA
1200, 4800, 9600, and 38400
ANALOG OUTPUT (Selectable)
Polar Speed:
Polar Direction: 0-5000mV or 4-20 mA (0 to 360° or 0 to 540°)
Cartesian UV:
0-5000mV or 4-20 mA (0 to 100 m/s)
0-5000mV or 4-20 mA (-100 m/s to +100 m/s)
GENERAL
Output rate: 0.1-20 Hz, 1 Hz typical
Power Supply: 10 to 30 VDC,
20 mA typical, 85 mA max
Protection Class: IP65
Dimensions:
Weight:
34 cm high x 17 cm wide
0.7 kg (1.5 lb)
Shipping Weight: 1.6 kg (3.5 lb)
2.0 INTRODUCTION
The YOUNG 86000 Ultrasonic Anemometer is a 2-axis, nomoving-parts wind sensor. It is ideal for general meteorological applications requiring accurate and reliable measurement.
The sensor features wide operating range, compact size, easy installation and low power operation.
The 86000 measures wind speed and direction based on the transit time of ultrasonic pulses between three transducers.
Measurement results are available as calibrated analog output signals, or serial data using RS-232 or RS-485. Continuous serial output or polled operation may be used. Serial format options include direct connection to YOUNG Wind Tracker displays, marine NMEA systems, data loggers, or other compatible serial communication devices.
Operating parameters are easily set using the 86SETUP program provided. All 86000 parameter settings are stored internally in nonvolatile memory.
The sensor is constructed using ultraviolet-stabilized thermoplastic for superior environmental resistance. It is easily mounted on standard 1 inch (IPS) pipe. An orientation ring preserves mounting position when the sensor is removed.
3.0 BEFORE INSTALLATION
The 86000 arrives fully calibrated and ready to use. Unless otherwise specified, the sensor is configured as follows:
FACTORY DEFAULT CONFIGURATION
Analog Voltage Outputs:
- Channel VI1: Wind Speed
0-5000 mV = 0-100 m/sec
- Channel VI2: Wind Direction
0-5000mV = 0-360 Deg
If using the sensor with a datalogger or other device requiring
0-5000 mV outputs, no further action is required. Simply connect the device as shown in Appendix A, figure A1.
If using the sensor with a YOUNG Wind Tracker, jumpers must be moved as shown in figure A5. The sensor will then provide RS485 serial output in RMYT format.
Other options are described in APPENDIX A. These are accomplished using various jumper combinations and configuring settings using the 86SETUP program available at www.youngusa.
com.
Note: Always connect and bench test a complex system before installation in the field.
86000-90(B)
Page 1
4.0 INSTALLATION
4.1 PLACEMENT
Proper instrument placement is important. Eddies from buildings, trees, or other structures can influence measurements. For most applications, locate the sensor well above or upwind of obstructions. As a general rule, air flow around a structure is disturbed to 2 times the height of the structure upwind, 6 times the height downwind, and up to 2 times the height of the structure above ground.
4.2 MOUNTING AND ALIGNMENT
Mount the sensor to standard 1-inch (IPS) pipe that has an outside diameter of 1.34 inches (34 mm).
Most applications require aligning the sensor to geographic north
(0 degrees). In this orientation the sensor junction box faces
SOUTH (180 degrees). See the diagram in APPENDIX B.
4.2.1 Place orientation ring over pipe with guide pin up.
4.2.2 Place sensor mounting post over pipe.
4.2.3 Using the transducers as a sighting aid, align the sensor with a feature on the horizon that represents the proper orientation. After alignment, tighten the mounting post band clamp to secure the position. DO NOT OVER-
TIGHTEN.
4.2.4 Slide the orientation ring up so its guide pin is fully engaged in the sensor mounting post notch. Tighten the orientation ring band clamp to secure its position. DO
NOT OVER-TIGHTEN.
If the sensor needs to be removed later, leave the orientation ring on the pipe to preserve sensor alignment.
4.3 WIRING CONNECTIONS
With long cable lengths, resistance in the power supply wires reduces the available voltage at the sensor. Power at the sensor must be in the range of 10 to 30 VDC when the sensor is operating. See WIRING DIAGRAMS in APPENDIX A.
5.0 OPERATION
5.1 ANALOG OUTPUTS
As supplied, the sensor is configured for VOLTAGE OUTPUT,
Wind Speed and Wind Direction. The sensor may also be set up for 4-20 mA CURRENT output by changing internal settings.
Details are in section 6.0.
Analog VOLTAGE or 4-20mA CURRENT outputs may be connected to a datalogger or other device such as a YOUNG
26800 Meteorological Translator. See APPENDIX A for connection details.
Analog outputs may be used simultaneously with RS-232 serial connection. RS-485 serial output may not be used simultaneously with analog outputs since they share connection terminals in the junction box.
Analog outputs may be configured for either Polar (speed and direction) or Cartesian (UV) output format.
For voltage output with cable lengths greater than 3m (10 ft.), measure the signal differentially. Current output signals may be measured single-ended.
5.2 USE WITH YOUNG WIND TRACKER DISPLAY
The factory default serial output format is RMYT which is compatible with the YOUNG Model 06201 Wind Tracker display.
Set the Wind Tracker input to 'INP 09' and connect as shown in the Wiring diagram, Fig A5, Appendix A. Note that jumpers need to be moved so the RS-485 output is available at the connection terminals. Wind speed and direction measurements appear on the
Wind Tracker display. See the Wind Tracker manual for display options and other details.
5.3 SERIAL OUTPUT FORMATS
Available serial output formats include RMYT, ASCII, ASCII polled, and NMEA. The factory default format is RMYT for use with the
YOUNG Wind Tracker display. Other formats may be selected using the 86SETUP program described in Section 6.0.
5.3.1 RMYT
RMYT is a 6-byte binary data format sent at 9600 baud using
RS -485 OUTPUT ONLY mode. This is the factory default serial format for use with the YOUNG Model 06201 Wind Tracker.
5.3.2 ASCII
ASCII output format provides continuous wind measurement data in text format at any of the available baud rates.
ASCII output appears either in POLAR (default) or CARTESIAN
UV format. With POLAR format, the wind speed threshold, wind speed units, and resolution are user-selectable. With CARTESIAN the wind threshold is ignored and wind speed units are always meters per second (m/s).
ASCII POLAR FORMAT a www.w ddd ss*cc<CR> Low resolution a www.ww ddd.d ss*cc<CR> High resolution where a www.ww ddd.d ss
* cc
<CR>
= Sensor address
= Wind speed
= Wind direction
= Status code
= Asterisk (ASCII 42)
= Checksum
= Carriage return (ASCII 13)
ASCII CARTESIAN (UV) FORMAT a uu.uu vv.vv ss*cc<CR> where a
±uu.uu
±vv.vv ss
* cc
<CR>
= Sensor address
= U-axis wind speed (m/s)
= V-axis wind speed (m/s)
= Status code
= Asterisk (ASCII 42)
= Checksum
= Carriage return (ASCII 13)
CHECKSUM is a two-character hexadecimal value (in printable
ASCII format) generated by taking the exclusive-or of all characters up to the asterisk. STATUS CODE shows a non-zero value when the sensor cannot acquire sufficient samples or a measurement error has occurred.
5.3.3 ASCII POLLED
ASCII POLLED is like ASCII format described above except just one serial output string is sent for each polling command received.
The polling command is
Ma! where 'a' is the sensor address
(valid characters: 0-9, A-Z, a-z). The default address is '0' (ASCII
48).
86000-90(B)
Page 2
5.3.4 NMEA
NMEA format provides continuous wind measurements in standard
NMEA marine sentences at 4800 baud. Use RS-485 OUTPUT
ONLY serial output mode with YOUNG Model 06206 Marine Wind
Tracker or other NMEA-capable device.
NMEA FORMAT
$WIMWV,ddd,R,www.w,N,A*cc<CR><LF> where ddd
= Wind direction (degrees) www.w = Wind speed (knots)
* cc
= Asterisk (ASCII 42)
= Checksum
<CR><LF> = Carriage return, line feed (ASCII 13, 10)
CHECKSUM is the two-character printable hexadecimal value generated by taking the exclusive-or of all characters between '$' and '*'.
5.4 LOW POWER OPERATION
Average current consumption with default settings is about 20 mA.
This configuration uses minimal power and enables all features even though they may not be used. This is suitable for many low power applications.
To reduce current consumption further, additional strategies include disabling unused outputs, using polled serial operation, increasing the output interval, and limiting the sample count to the minimum optimal number. Faster baud rates also reduce power by limiting transmit duration.
6.0 SETTING OUTPUTS AND OPERATING
PARAMETERS
6.1 SENSOR CONFIGURATION WITH 86SETUP PROGRAM
(RECOMMENDED)
The YOUNG 86SETUP program is available from the factory web site at www.youngusa.com. It provides an easy method for checking and configuring sensor operation. Install the program on a Windows PC and follow instructions that appear on the program screen to retrieve current sensor settings or send new settings.
6.2 SENSOR CONFIGURATION USING A GENERAL-
PURPOSE COMMUNICATIONS PROGRAM
A general purpose text-based serial communications program like
HyperTerminal may be used to manually configure the sensor by sending simple text commands.
The YOUNG sensor and communication program must operate at the same baud rate and be properly connected. Sensor RS-232 mode must be enabled. See the RS-232 Wiring diagram, Fig A3, in Appendix A.
Factory default sensor baud rate is 9600, but may be set to 1200,
4800, or 38400. Configure the serial communications program for
NO handshaking and 1 start, 8 data, 1 stop bit.
The sensor must be in COMMAND MODE in order to set parameters. Enter COMMAND MODE by sending three ESC characters (ASCII 27) in quick succession while the sensor is running. When the sensor is in COMMAND MODE, it sends a '>' prompt character indicated that it is ready to accept commands.
If the prompt does not appear after sending three ESC characters, re-check wiring and communication program setup. If the sensor baud rate is unknown, try sending the ESC characters at each of the four available baud rates (1200, 4800, 9600, and 38400).
It is also possible that sensor parameters have been purposely configured to disable RS-232 mode. If this the case, the following method must be used.
In order to provide access under all conditions, the sensor always begins operation at power up with serial communications set to
38400 baud and RS-232 connections enabled. Immediately after power up, there is a short time window in which to send the ESC characters and enter COMMAND MODE.
To use this feature, set your serial communication program baud rate to 38400. Remove power then wait 5 seconds. Re-apply power to the sensor. The sensor will transmit four asterisks immediately after power up. After the asterisks appear, send three
ESC characters. The COMMAND MODE
'>' prompt should appear.
6.3 COMMAND OVERVIEW
After the '>' prompt appears, send '??' to display a list of available commands. Send
'RPTV' to report current settings.
(Note that some values in the report are for factory settings cannot be changed by the user.)
Commands are case sensitive and the exact format must be used.
For example, the
SET01nn command requires two digits for the serial format code. If you send SET014 instead of SET0104 the sensor will reject the command and indicate an error. End all commands with a carriage return (ASCII 13). In HyperTerminal, do this by pressing the ENTER key.
After receiving the carriage return, the sensor will evaluated the command. Valid commands will be executed. Current settings can be evaluated at any time by sending RPTV to get a new report.
IMPORTANT NOTE:
The YOUNG 86SETUP program automatically saves all settings to flash memory when they are sent to the sensor.
Settings that are changed manually must be saved to flash with the SET77 command.
86000-90(B)
Page 3
6.4 COMMAND DETAILS
COMMAND DESCRIPTION
SET01nn Set OUTPUT MODE
01 Enable voltage output
03
Enable current output
04 Enable RS-232
08 Enable RS-485 output only
16 Enable RS-485 half duplex
24 Enable RS-485 full duplex
SET02n Set OUTPUT FORMAT
1 RMYT
2 ASCII
3 ASCII POLLED
4
NMEA
SET03nn Set BAUD RATE
12 1200
48 4800
96
9600
38 38400
SET04n Set ASCII WS UNITS
1 MPH
2
KNOTS
3 KMPH
4 M/S
SET05c ASCII character sensor address (0-9, A-Z, a-z)
SET06nnn
Wind speed threshold for polar output (cm/s)
SET07nnnnn Wind speed scale (nnnnn/10000)
SET08nnnnn Direction offset (±nnnnn degrees x 10)
SET09nn Damping factor
SET10nnnn Output interval (0-9999 milliseconds)
SET11nn
Direction VOUT (36=0-360, 54=0-540 degrees)
SET12nnn Sample count (3 to 200)
SET13n Wind output format (0/1=Polar/UV)
SET14n Analog error code (1/2/3=None/Lo/Hi)
SET15n ASCII serial resolution(0/1=Lo/Hi)
SET16n
Force analog out (0/1/2=Lo/Mid/Hi)
SET77 SAVE SETTINGS
XX
RPT
??
Go to OPERATE MODE
Report parameter settings
Command Help list
6.4.1 SET01nn SET OUTPUT MODE
This enables and disables 0-5000 mV output, 4-20 mA current output, RS-232, and RS-485. Only one of the two analog output types (voltage or current) may be enabled at one time. Both RS-
232 and RS-485 can be enabled at the same time but only one serial output may be used.
Add together values shown in 6.4 COMMANDS SET01 to configure multiple compatible modes. For example, to enable voltage output and RS-232, add the code for each one: 01 + 04 =
05, SET0105. To enable only RS-232, SET0104.
To conserve power, enable only those modes that are needed
6.4.2 SET02n OUTPUT FORMAT
This parameter determines the serial output format.
6.4.3 SET03nn BAUD RATE
Sets the baud rate for RS-232 and RS-485 serial communication.
Make sure this baud rate is the same as the connected device.
6.4.4 SET04n ASCII and NMEA WIND SPEED UNITS
Sets wind speed units for polar ASCII, ASCII POLLED, and NMEA serial outputs.
86000-90(B)
Page 4
6.4.5 SET05c POLL ADDRESS CHARACTER
Sets the sensor address for ASCII POLLED serial format. The default is '0' (ASCII 48). This is the address recognized when the
'Ma!' polling command is received ('a' is the address character).
Valid address characters include 0-9, A-Z, and a-z.
6.4.6 SET06nn WIND SPEED THRESHOLD
Sets the wind speed threshold for polar outputs (wind speed and direction) to minimize erratic wind direction indications at very low wind speeds. This allows the sensor output to mimic a mechanical wind vane that retains its orientation when there is no wind.
Wind below the threshold is reported as zero, while the wind direction angle is held at the last value when wind speed was above threshold. Set threshold in centimeters per second (m/s x
100). The default setting is 25 cm/s (0.25 m/s, 0.56 mph).
6.4.7 SET07nnnnn WIND SPEED MULTIPLIER
All wind speed measurements are multiplied by this parameter.
The default value is 10000 for a multiplier of 1.0000.
6.4.8 SET08nnnnn WIND DIRECTION OFFSET
Use this parameter to add or subtract a wind direction offset.
Value is degrees x 10 and may be positive or negative. Wind direction is always re-scaled to a 0-360 range after offset is applied. The default value is 00000.
6.4.9 SET09nn DAMPING FACTOR
Wind measurement outputs are damped using the following formula:
S damped
= [(d-1) * S damped
+ S sample
] / d where:
S damped
S sample d
= New or last damped wind speed
= New wind speed speed
= Damping factor
The default value is 00. This means no damping is applied. High damping values at long output intervals can slow the rate at which indicated wind values change.
6.4.10 SET10nnnn OUTPUT INTERVAL
Sets the time interval between measurements in one millisecond increments. Lower values increase power consumption when continuous measurements are taken.
6.4.11 SET11nn ANALOG DIRECTION SCALE
Sets wind direction analog output scale to 0-360 or 0-540 degrees.
Use the 0-540 scale whenever possible to avoid full-scale analog output swings between 0 and 360 in variable north wind. (Data loggers or display systems may otherwise obtain samples midpoint during these transitions causing erroneous readings.)
Logged data in 0-540 form may be re-scaled to 0-360 by subtracting 360 degrees from any value greater than or equal to 360. The default parameter setting is 0-360 for systems that cannot re-scale the 0-540 output.
6.4.12 SET12nnn SAMPLE COUNT
The rate at which the sensor internally takes complete sonic wind samples is greater than 200 times per second. This command sets the number of internal samples used to calculate the median measurement result.
More internal samples consume more power while providing greater immunity to conditions like turbulent high-speed wind.
Fewer samples consume less power while providing less immunity to disruptive conditions.
This command interacts with the OUTPUT INTERVAL setting.
More samples may require a longer output interval.
6.4.13 SET13n ASCII and ANALOG OUTPUT FORMAT
This setting determines whether ASCII and ASCII POLLED serial outputs and analog outputs provide wind data in either Polar
(speed and direction) or Cartesian (UV) form.
6.4.14 SET14n ANALOG OUTPUT STATUS
Serial ASCII and ASCII POLLED formats report a STATUS
CODE where non-zero values indicate insufficient samples or measurement error. SET14n determines how the STATUS CODE is indicated by the analog output.
6.4.15 SET15n ASCII RESOLUTION
Sets low or high wind speed and direction resolution.
6.4.16 SET16n FORCE ANALOG OUTPUT
This command forces both analog output channels to LO, MID, or
HI scale. This may be used to calibrate or check the operation of externally connected analog measurement devices.
6.4.17 SET77 SAVE SETTINGS
Use this command to save current parameter settings to nonvolatile memory. Any settings which have not been saved will be lost when power is removed. Saved settings are loaded at power up. This command may be used any time the sensor is in
COMMAND MODE.
6.4.16 XX, RPTV, and ??
XX
RPT
??
Returns the sensor to OPERATE MODE.
Reports the current parameter settings.
Shows a list of commands.
7.0 EXAMPLE SETTINGS
Suggested settings. Not all possible setting combinations are shown. Using YOUNG 86SETUP program for changing parameters is recommended.
7.1 FACTORY DEFAULT
RS-232 Enabled
RS-485 Output Only Enabled
Voltage Output:
Current Output:
Serial Output Format: RMYT
Output Interval
Sample Count:
Enabled
Disabled
250
16
7.2 MINIMUM POWER
RS-232
RS-485
Voltage Output:
Enabled
Disabled
Disabled
Disabled Current Output:
Serial Output Format: ASCII
Output Interval
Sample Count:
1000
16
7.3 4-20 mA OUTPUT
RS-232
RS-485
Voltage Output:
Current Output:
Output Interval
Sample Count:
Enabled
Disabled
Disabled
Enabled
1000
50
7.4 RS-485 POLLED ASCII OUTPUT
RS-232
RS-485
Voltage Output:
Current Output:
Enabled
Enabled
Disabled
Serial Output Format: ASCII POLLED
Sample Count:
Disabled
50
7.5 HIGH WINDS
Output Delay:
Sample Count:
1000
200
8.0 WARRANTY
This product is warranted to be free of defects in materials and construction for a period of 12 months from date of initial purchase.
Liability is limited to repair or replacement of the defective item.
A copy of the warranty policy may be obtained from R. M. Young
Company.
Page 5
86000-90(B)
APPENDIX A: WIRING CONNECTIONS
86000-90(B)
Page 6
APPENDIX A: WIRING CONNECTIONS
Page 7
86000-90(B)
APPENDIX A: WIRING CONNECTIONS
86000-90(B)
Page 8
APPENDIX B: SENSOR ORIENTATION AND DIMENSIONS
Page 9
86000-90(B)